Erbe Erbotom ICC300 User manual

5 (1)

ERBE

ERBOTOM ICC 300

Instruction manual 08.00

ERBOTOM ICC 300-H V 4.X

10128-070, 10128-077, 10128-306, 10128-075

ERBOTOM ICC 300-E V 4.X

10128-071, 10128-078, 10128-076, 10128-307, 10128-401

Instruction manual 08.00

ISO 9001

EN 46001

All rights to this instruction manual, particularly the right to reproduction, distribution and translation, are reserved. No part of this instruction manual may be reproduced in any form (including photocopying, microfilm or other means), or processed, reproduced or distributed by means of electronic systems without prior written permission from ERBE ELEKTROMEDIZIN GmbH.

The information contained in this instruction manual may be revised or extended without prior notice and represents no obligation on the part of ERBE ELEKTROMEDIZIN GmbH.

©ERBE ELEKTROMEDIZIN GmbH, Tübingen 2000

Printed by: ERBE ELEKTROMEDIZIN, Tübingen

Instruction manual no. 80104-351

Chapter

Title

Page

1

INTRODUCTION ................................................................

1-1

1.1

Intended purpose of the ICC 300 ................................................

1-1

1.2

Explanation of the safety instructions .........................................

1-1

2

INITIAL OPERATION ........................................................

2-1

3

RISKS AND SAFETY

 

 

OF HIGH-FREQUENCY SURGERY ..................................

3-1

3.1

Unintentional thermal tissue damage ..........................................

3-1

3.1.1

- due to HF leakage currents .......................................................

3-1

3.1.2

- due to unintentional activation of an HF generator ...................

3-2

3.1.3

- due to inappropriate application ................................................

3-3

3.1.4- due to inappropriate or nonapplication of

 

the neutral electrode ....................................................................

3-3

3.1.5

- due to unsuitable and/or faulty accessories ...............................

3-4

3.1.6

- due to inattentiveness ................................................................

3-5

3.1.7

- due to an output error ................................................................

3-5

3.1.8- due to the ignition of flammable liquids, gases

 

and/or vapors ...............................................................................

3-5

3.1.9

Unintentional burns due to hot electrodes ...................................

3-6

3.2

Electric shock ..............................................................................

3-6

3.3

Stimulation of nerves and muscles ..............................................

3-6

3.4

Cardiac pacemaker ......................................................................

3-7

3.5

Danger of explosion ....................................................................

3-7

3.6

Interference with other electronic equipment ..............................

3-7

4

DESCRIPTION OF THE HIGH-FREQUENCY

 

 

SURGICAL UNIT ...............................................................

4-1

4.1

General description .....................................................................

4-1

4.2

Description of the controls ..........................................................

4-2

1

Power switch ..........................................................................................

4-2

2

AUTO CUT function field .....................................................................

4-3

3

AUTO COAG function field ..................................................................

4-4

4

AUTO BIPOLAR function field ............................................................

4-6

5

Connecting socket for neutral electrodes ...............................................

4-7

6

Connecting socket for the AUTO CUT and AUTO COAG function

 

 

fields, fingerswitch activation ................................................................

4-7

7

Connection socket for the AUTO CUT and AUTO COAG function

 

 

fields, fingerswitch and footswitch activation .......................................

4-7

8

Connecting socket for the AUTO BIPOLAR function field ..................

4-8

9

Safety field .............................................................................................

4-8

10

Terminal for potential equalization ........................................................

4-8

11

Connecting socket for a dual-pedal footswitch ......................................

4-8

12

Connecting socket for a single-pedal footswitch ...................................

4-8

13

Volume for acoustic signals ...................................................................

4-9

14

Loudspeaker for acoustic signals ...........................................................

4-9

15

Power connection ...................................................................................

4-9

16

Power fuses ............................................................................................

4-9

4.3

Description of the safety features ..............................................

4-10

5

TECHNICAL DATA, SIGNALS, DIAGRAMS .....................

5-1

5.1

Technical data ..............................................................................

5-1

5.2

Visual and acoustic signals ..........................................................

5-5

5.3

Diagrams .....................................................................................

5-6

6

INSTALLATION .................................................................

6-1

7

CLEANING AND DISINFECTION OF THE UNIT ..............

7-1

8

PERFORMANCE CHECKS ...............................................

8-1

8.1

Automatic performance test after switching on the unit .............

8-1

8.2

Automatic performance check during activation ........................

8-1

8.3

Automatic error documentation ...................................................

8-2

 

Error list .......................................................................................

8-4

9

SAFETY CHECKS .............................................................

9-1

10

MAINTENANCE, CARE, DISPOSAL ..............................

10-1

11

GUARANTEE ..................................................................

11-1

 

ADDRESSES ...................................................................

12-1

1 INTRODUCTION

1.1Intended purpose of the ICC 300

The ICC 300 is a high-frequency surgical unit for cutting and coagulation. Due to its performance features, it has universal application. The ICC 300 is available in the versions -H with HIGH CUT and -E without HIGH CUT.

1.2Explanation of the safety instructions

The WARNING safety instruction indicates a danger which can result in personal injury.

The CAUTION safety instruction indicates a danger which can result in property damage.

The IMPORTANT safety instruction indicates a danger which can cause functional failure of the unit.

1-1

1-2

2 INITIAL OPERATION

Read carefully before initial operation of the unit.

In the development and production of this high-frequency surgical unit, the relevant, generally recognized rules of technology, as well as the valid occupational safety and accident prevention regulations have been taken into consideration. This ensures that patients, employees and third parties are protected from dangers to life and health during intended application of the highfrequency surgical unit, to the extent permitted by the type of application intended.

Initial operation

Before delivery, every high-frequency surgical unit is tested by the manufacturer in regard to its function and safety. To ensure that the unit also functions safely after shipping and installation at the operator’s site, the following points should be observed:

The operator should only operate the high-frequency surgical unit if the manufacturer or supplier

1.has subjected the unit to a performance test on site

2.has instructed the parties responsible for operation of the unit in handling of the unit by means of the instruction manual.

2-1

2-2

3 RISKS AND SAFETY OF HIGH-FREQUENCY SURGERY

3.1Unintentional thermal tissue damage

High-frequency surgery is associated in principle with various risks for the patient, the personnel and surroundings. In order to avoid these risks in practice, the surgeon and his/her assistants must recognize these risks and observe the appropriate rules for prevention of damage. In the following, these risks and rules for prevention of damage are explained.

3.1.1 Unintentional thermal tissue damage due to HF leakage currents

During high-frequency surgery, the patient unavoidably conducts high-frequency electrical current to ground potential. If the patient makes contact with electrically conductive objects during high-frequency surgery, a high-frequency electrical current can result at the contact point between the patient and this object, which can in turn cause thermal necroses. Not just objects made of metal are electrically conductive objects, but also wet cloths.

WARNING

The patient must be insulated against electrically conductive objects during high-frequency surgery. The black elastic table covers on operating tables demonstrate a certain electrical conductivity for diverting electrical charges. Therefore they are never suitable for ensuring the required insulation of the patient against metal parts of the operating table. For this reason, an electrically insulating intermediate layer, for example dry cover cloths, must be laid between the patient and this black operating table cover during the application of high-frequency surgery.

Electrically insulated

Grounded

surface

operating

 

table

Fig.: Insulated positioning of the patient on the operating table

If it is possible for this intermediate layer to become wet during the operation, for example due to perspiration, irrigation liquid, urine etc., wetting of these intermediate layers must be prevented by a watertight sheet of plastic. Urine should be carried away via catheter.

Extremities lying against the trunk or skin-to-skin contact points should be insulated from one another by laying dry cover cloths between them.

Do not apply ECG electrodes closer than 15 cm next to the operating field.

Needle electrodes or injection cannulae should not be used as ECG electrodes during high-frequency surgery.

3-1

3.1.2 Unintentional activation of an HF generator

Unintentional activation of an HF generator can lead to burns on the patient if the active electrode hereby touches the patient directly or indirectly through electrically conductive objects or wet cloths.

Unintentional activation of an HF generator can, for example, be caused by:

Unintentionally pressing a footswitch pedal

Unintentionally pressing a fingerswitch

Defective fingerswitches, footswitches or cables

Penetration of electrically conductive liquids (blood, amniotic fluid, urine, physiological saline solution, irrigation fluids etc.) into fingerswitches or footswitches.

Errors within the high-frequency surgical unit

WARNING

To prevent burns on the patient due to unintentional activation of a high-frequency generator, the following application rules should be heeded:

Never lay active electrodes onto or beside a patient in such a way that they can touch the patient directly or indirectly through electrically conductive objects or wet cloths.

The lines to the active electrodes should be positioned in such a way that they touch neither the patient nor other lines.

Always set the acoustic signal, which indicates the active status of the high-frequency generator, so that it can be easily heard.

For operations in which the cutting or coagulation electrode unavoidably remains in contact with the patient even in a nonactive condition, e.g. for endoscopic operations, particular care is required. If such an electrode is unintentionally activated due to an error, this activated electrode should then not be removed from the body without special supervision. When removing the activated electrode from the patient’s body, burns can result on all areas within the body which come into contact with the activated electrode. For this reason, in case such errors occur, the power switch for the high-frequency surgical unit should be switched off immediately before an attempt is made to remove the activated electrode from the body.

3-2

3.1.3 Unintentional thermal tissue damage due to inappropriate application

Generally speaking, the bipolar coagulation technique should be applied in preference to the monopolar coagulation technique. This particularly applies to coagulations on straight organs, on which the high-frequency current flows over longer areas through diameters which are approximately equal or become even smaller.

Fig.: Thermal damage of lateral tissue

The tissue is always first heated at places on the tissue where the diameter is smallest. If the HF current flows through the same diameter (a) over longer distances, the tissue coagulates over this entire distance. If the diameter of the tissue next to the application point of the coagulation electrode is smaller than at the point of application, coagulation will also occur next to the application point (b).

WARNING

Always make certain that the HF current does not flow through thin tissue structures or vessels with a small diameter.

3.1.4Unintentional thermal tissue damage due to inappropriate or nonapplication of the neutral electrode

With inappropriate or even nonapplication of the neutral electrode, there is a large risk of unintentional thermal tissue damage both at the application point of the neutral electrode as well as to other areas on the patient’s body.

The neutral electrode must be applied with its entire surface as closely and reliably as possible to the operating field on the patient’s body.

3-3

WARNING

The effective contact surface, i.e. the electrical conductive value between the neutral electrode and the patient must correspond to the HF capacity used, meaning the intensity of the HF current. Here the effective contact surface means the surface of the neutral electrode which has electrically conductive contact to the skin of the patient during high-frequency surgery.

a)

b)

Fig.: The neutral electrode must be applied at an appropriate location on the patient’s skin using the entire contact surface available (a). If the neutral electrode has only partial contact to the patient’s skin (b), there is a risk that burning will occur at this location

3.1.5Unintentional thermal tissue damage due to unsuitable and/or faulty accessories

It must be ensured that only accessories in perfect condition are used for high-frequency surgery. Only accessories that are compatible or tested by the unit manufacturer must be used. This applies both to the active electrodes including cable and plugs, as well as to the neutral electrodes including cables and plugs.

When using an instrument with electric insulation, it is necessary to be certain that these insulations are not overloaded and destroyed by overly high electric voltages. The electric output voltages for the high-frequency surgical unit are indicated for the various cutting and coagulation modes relative to the possible settings in this instruction manual. The electric strength of the instrument insulation can be found in the technical data for the instruments or, in case of doubt, can be requested from the manufacturer of the respective instrument.

WARNING

All insulation on electrodes, electrode holders, cables, plugs etc. must be in perfect condition.

3-4

3.1.6 Unintentional thermal tissue damage due to inattentiveness

Like a scalpel, high-frequency surgery is always a potential source of danger if handled without care.

WARNING

The cutting or coagulation electrodes should always be handled with care and laid aside in the intervals between use so that neither the patient nor other persons can come in contact with the electrodes.

Laying unused electrode handles or coagulation forceps on the patient, next to the patient or within folds on the cover cloths is dangerous. Cases of burns on patients are known which were caused by laying the coagulation forceps within folds on the cover cloths which penetrated through the cloths into the patient’s skin and resulted in burns without being noticed.

3.1.7 Unintentional thermal tissue damage due to output error

The risk of unintentional thermal tissue damage is proportionate to the intensity and time limit set on the unit for cutting or coagulation.

WARNING

The intensity for cutting or coagulation should only be set and only activated for as long as necessary for the intended purpose.

An insufficient effect at a standard setting can, for example, be caused by poor attachment of the neutral electrode, poor contact in the connectors, defective cables or electrically insulating tissue remnants on the active electrode. This must be checked before setting at a higher power.

3.1.8Unintentional thermal tissue damage due to the ignition of flammable liquids, gases and/or vapors

During high-frequency surgery, electric sparks or arcs that can ignite flammable liquids, gases or vapors occur at the active electrode.

3-5

WARNING

Make certain during high-frequency surgical operations that anesthetics, skin cleaning agents and disinfectants are nonflammable. If their use is unavoidable, they must have completely evaporated and the vapor must be removed from the area of spark formation before switching on the high-frequency surgical unit.

Before application of high-frequency surgery in the gastro-intestinal tract, it must be ensured that no flammable (endogenous) gases are present here. There is danger of explosion if flammable gases are present. For this reason, these gases must be extracted and/or eliminated by flushing out the affected lumen with CO2before using high-frequency surgery.

During transurethral resection (TUR), H2O molecules may dissociate into H2 and O2 in the arc between the resection loop and the irrigation liquid. These gases may collect on the roof of the urinary bladder as a highly explosive gas mixture. If resection is performed in this gas mixture, dangerous explosions may occur.

3.1.9 Unintentional burns due to hot electrodes

Cutting and/or coagulation electrodes become hot during cutting and/or coagulation procedures indirectly through the heated tissue and through the electric arc.

WARNING

Tissue can be unintentionally burnt immediately after cutting and/or coagulation procedures if electrodes that are still hot touch the tissue. Attention must be especially paid to this during endoscopic operations, such as during pelviscopic fallopian tube coagulation or during endoscopic polypectomy.

3.2Electric shock

An electric shock may occur if the high-frequency surgical unit delivers a too heavy low-frequency current or if a too heavy low-frequency current flows through the patient into the high-frequency surgical unit from another voltage source.

3.3Stimulation of nerves and muscles

A known risk of high-frequency surgery is the unintentional electric stimulation of the patient’s nerves and muscles. This stimulation can result from low-frequency electrical currents that are caused either by low-frequency current sources or due to electrical arcs between an active electrode and the patient’s tissue.

Electric alternating current with a frequency above 300 kHz is unable to stimulate nerves and muscles.

3-6

During cutting procedures, forced coagulation and spray coagulation, the unavoidable electric arcs between an active electrode and the tissue nevertheless have the effect that a portion of the high-frequency alternating current is rectified, from which more or less strongly modulated, low-frequency current components result which stimulate electrically stimulable structures such as nerves and muscles.

This can result in more or less strong spasms or muscle contractions.

WARNING

When using high-frequency surgery on electrically stimulable structures, contractions of the affected muscles must be taken into account. This can occur, for example, during endoscopic operations in the urinary bladder in the vicinity of the obturator nerve and during operations in the area of the facial nerve.

3.4Cardiac pacemaker

For patients with implanted cardiac pacemakers or pacemaker electrodes, irreparable damage to the pacemaker and disturbance of the pacemaker function, which can lead to ventricular fibrilation, must be reckoned with.

3.5Danger of explosion

High-frequency surgical units always generate sparks during operation on the active electrode. For this reason, it is necessary to make certain during interventions that anesthetics, degreasers and disinfectants are neither flammable nor explosive. They should at least have evaporated completely before switching on the high-frequency surgical unit and be removed from the area of spark formation.

3.6Interference with other electronic equipment

High-frequency surgical units normally generate high-frequency electrical voltages and currents which can interfere with other electronic equipment.

When installing or arranging sensitive electronic equipment in the operating room, this problem should be taken into consideration. In principle, sensitive electronic equipment should be set up as far as possible from the high-frequency surgical unit and particularly from the cables providing HF current. In addition, the cables providing HF current, which act like broadcast antennas, should not be unnecessarily long and should never be positioned parallel or too close to cables from sensitive electronic equipment.

The unit has been fitted with a special generator in consideration of the disturbance of sensitive electronic equipment, which generates a relatively low interference level as compared to conventional high-frequency surgical units.

3-7

3-8

Erbe Erbotom ICC300 User manual

Standard

International

4 DESCRIPTION OF THE HIGH-FREQUENCY SURGICAL UNIT

4.1General description

Cutting with automatic control of the HF voltage (Auto Cut)

The ERBOTOM ICC is equipped with automatic open and closed loop control systems which control and regulate the parameters relevant to the cutting quality so that each respectively selected cutting quality is guaranteed to be reproducible and constant.

Adjustable power limitation in the cutting mode

Since the ICC units are equipped with automatic control of the HF voltage in the cutting mode, a power setting in regard to cutting quality is not required. The adjustable power limitation is primarily intended to guarantee the safety of the patient from unintentional thermal tissue damage, and to protect fine cutting instruments, such as fine needle electrodes, from destruction due to overly high HF currents if these come in contact in activated condition with other metallic instruments. The latter, for example, is a risk during laparoscopic operations. This adjustable power limitation must not be confused with the power setting for conventional high-frequency surgical units, where the cutting quality is directly dependent upon the power setting.

Cutting with automatic control of the electric arc (HIGH CUT)

The HIGH CUT key only functions on the ICC 300-H. This key has no function on the ICC 300- E. HIGH CUT is especially suitable for cutting in tissue with poor electric conduction. Since the HF voltage in the HIGH CUT mode can attain high values on the active electrode, HIGH CUT should however only be used with electrically well insulated instruments when cutting in tissue with poor electric conduction. For bipolar or quasibipolar cutting, HIGH CUT is not recommended inasmuch as the insulation paths for bipolar and quasibipolar instruments are generally very small and must not be exposed to higher voltages. For bipolar and quasibipolar instruments, automatic voltage control is recommended.

PPS (Power Peak System)

The initial incision phase can represent a special problem during an incision, particularly if the cutting electrode is firmly pressed against the tissue to be cut before activating the HF generator, so that the cutting electrode has a relatively large-surface, and therefore low-resistance, contact. This is the case, for example, for TUR and for endoscopic polypectomy. In such cases, the HF generator must provide higher-than-normal power so that the initial incision can proceed without delay, for otherwise a very large coagulation necrosis may result at the cutting location. The ICC is equipped with an automatic power control which recognizes low-resistance loads and controls the HF generator in such a way that it briefly provides enough power so that the HF voltage, i.e. the intensity of the electric arc, required for the set cutting quality is ensured even for a lowresistance load. Thanks to this device, the average power can be limited to relatively small quantities, which corresponds to an improvement in protection from unintentional thermal tissue damage.

4-1

Soft coagulation

Soft coagulation can be activated by key or pedal.

Forced coagulation

Forced coagulation is advantageous if an efficient hemostasis is to be achieved with relatively small-surface electrodes, such as TUR resection loops.

Spray coagulation

The Spray coagulation mode on the ICC is also suitable for Argon Gas coagulation.

Adjustable power limitation in the various coagulation modes

For the ERBOTOM ICC units, the surgically relevant coagulation qualities, i.e. the coagulation effects Soft Coag., Forced Coag. Spray Coag. and Bipolar Coag., are delimited by definition from one another and selectable by the press of a key. Nevertheless, the intensity of the different effects can be varied by power limitation.

Bipolar coagulation

In this coagulation mode, the HF voltage is, similar to Soft coagulation, automatically and constantly controlled, and its peak value always remains lower than 200 Vp, so that the current density and thus the coagulation effect is, for the most part, independent of the effective contact surface between the coagulation electrode and the tissue, provided the effective contact surface is not too large relative to the currently set power limitation.

The adjustable power limitation serves the purpose of protecting fine bipolar coagulation instruments, such as pointed bipolar coagulation forceps, from being thermally destroyed in case of a short between the two forcep tips.

The footswitch or Auto Start is used for activation.

In the Auto Start mode, the HF generator is automatically activated if both poles of the bipolar coagulation instruments used contact electrically conductive tissue simultaneously. Auto Start can occur either immediately at the moment of contact with the tissue or more or less temporally delayed. With immediate activation, it is possible to work very quickly, especially if several coagulations must be performed one after another. Delayed activation offers the operator the advantage that he/she can prepare and securely grasp the tissue to be coagulated with the bipolar coagulation forceps before the HF generator is automatically activated. Automatic activation of the HF generator only occurs if both poles of the bipolar coagulation instruments contact the tissue uninterruptedly for at least as long as the respectively selected delay lasts. If the contact is interrupted before the respectively selected period of delay is over, the respective period of delay restarts as of the next contact.

4-2

4.2Description of the controls

This symbol, in accordance with EN 60 601-1, is intended to indicate to the user that this unit must only be used on the patient if the user is acquainted with the operation and features of this unit.

The figures set in cursive relate to the ICC illustration for this chapter, or to the function fields in the text.

1Power switch

Using this power switch, the unit is switched on and off.

Each time after being switched on, the unit automatically proceeds with various performance checks. If an error in the unit or in the accessories is recognized here, a warning signal sounds and the determined error is indicated by a corresponding error number. (See Chapter 8.1, Automatic performance checks after switching on the unit). If no error is determined, the unit is ready to operate.

If the unit was switched off for less than approx. 15 seconds, all settings for the program used before switching off appear after the automatic performance check on the front panel, and the unit can be immediately reactivated. This is advantageous if, for example, the power supply briefly fails.

If the unit was switched off for longer than approx. 15 seconds, the basic setting of the program used before switching off appears after the automatic performance check on the front panel, whereby all relevant visual displays continue to blink and the unit cannot be activated until any key on the front panel is briefly pressed as confirmation that this program should be used. Then the relevant displays are continously illuminated and the unit can be activated using the available settings. These settings can be changed or adapted to the current requirements at any time. However, other programs can be selected as well.

2-4 Function fields

The AUTO CUT, AUTO COAG and AUTO BIPOLAR function fields can be adjusted separately from one another, although not activated simultaneously for reasons of safety.

WARNING

Function fields that are not used may be switched off completely to prevent unintentional activation. To do this, the power limitation must be set down so far in the corresponding function field until a beep is heard and “ —“ appears on the digital display. The corresponding function field cannot be activated in this condition..

4-3