Read this operating manual completely before using the device for the first time. Also observe the
instructions for use of the accessories.
This operating manual is part of the product. Thus, it must always be easily accessible.
Enclose this operating manual when transferring the device to third parties.
You will find the current version of the operating manual for all available languages on our webpage
under www.eppendorf.com
.
1.2Danger symbols and danger levels
The safety instructions of this operating manual indicate the following danger symbols and danger levels:
7
1.2.1Danger symbols
Electric shockExplosion
Toxic substancesHazard point
Material damage
1.2.2Danger levels
DANGERWill lead to severe injuries or death.
WARN INGMay lead to severe injuries or death.
CAUTIONMay lead to light to moderate injuries.
NOTICEMay lead to material damage.
1.3Symbols used
DepictionMeaning
1.
2.
Actions without a specified order
• List
Actions in the specified order
Press this key to perform the described action.
or sample
Page 8
Operating instructions
Eppendorf BioSpectrometer
8
English (EN)
DepictionMeaning
Press this softkey to perform the described action.
or [Copy]
Additional information
1.4Abbreviations used
A
Absorbance
DNA
Deoxyribonucleic acid
®
kinetic
dsDNA
Double-stranded DNA
Dye methods
Methods for dye labels group for measuring dyed biomolecules
FOI
Frequency of Incorporation: measure for the amount of dye molecules in relation to the amount of
nucleotides in dyed biomolecules
M
mol/L (molar)
OD600
Optical density at a wavelength of 600 nm
RNA
Ribonucleic acid
ssDNA
Single-stranded DNA
UV
Ultraviolet radiation
Vis
Visible light
CV
Coefficient of variation (standard deviation/average value) in percent
Page 9
2Product description
2.1Main illustration
Abb. 2-1: Front and rear view
Product description
Eppendorf BioSpectrometer
English (EN)
®
kinetic
9
1
1
2
3
abc
def
method
4
5
ghi
6
jkl
mno
function
7
pqrs
8
9
tuv
0
wxyz
µ %
exit
delete
standard
enter
10
Fig. 2-1:Front and rear view
1Display
2Cuvette shaft
3
23
absorbance
height
8.5 mm
blank
sample
absorbance
8
9
7
5
6
4
6Fuse holder
7 Mains/power connection
3Cuvette shaft cover
4 USB port for USB stick and printer
5 Mains/power switch
8USB port for PC
9 Connection for RS-232 printer
10 Operating controls
The name plate is located at the rear left on the bottom of the device.
2.2Delivery package
QuantityDescription
1BioSpectrometer kinetic
1Power cord
44 UVettes
Original Eppendorf plastic cuvette, individually packaged, PCR clean, protein-free
1Operating manual, in multiple languages
Page 10
10
Product description
Eppendorf BioSpectrometer
English (EN)
®
kinetic
2.3Features
The BioSpectrometer kinetic is a UV/Vis spectrophotometer for measuring liquids in cuvettes in a
wavelength range of 200 nm to 830 nm. It is intended for use in development and research in the fields of
molecular biology, biotechnology, biochemistry and cell biology. Glass and plastic cuvettes in a volume
range of 1 μL to 3000 μL can be used.
2.3.1Methods
Numerous methods for concentration determination of nucleic acids, proteins, and dye-marked nucleic
acids and proteins, and the OD 600 method for determining bacterial density via turbidity measurement,
are already preprogrammed. Furthermore, method templates for various measurement and evaluation
procedures (single and multiple wavelength measurements, acquisition of spectra, kinetic methods,
evaluations with factor, standard and standard curve) are preprogrammed. It is possible to create individual
methods on the basis of the preprogrammed methods and templates. The templates in the Absorbance
method group can be used to quickly measure absorbances or spectra without an additional evaluation.
2.3.2Operation
The preprogrammed methods and templates are combined into clearly arranged groups from which the
desired method can be quickly selected. After calling up the method, you are guided through the
measuring procedure in clear steps. A help box in the display provides hints upon request. The 3 round
measuring keys (standard, blank, sample) allow users to quickly start a measurement.
2.3.3Result output
The BioSpectrometer kinetic outputs the results using the device display and a printer available from
Eppendorf. With a USB connection, you can transfer result data from the device to a USB stick, a printer or
directly to a PC.
2.3.4Device self test
The device automatically checks the function of the spectrometer unit and thermal module immediately
after it has been switched on. Access the Device calibration function for a more comprehensive test (see
Device self-test on p. 72).
Page 11
Safety
®
Eppendorf BioSpectrometer
kinetic
English (EN)
3Safety
3.1Intended use
The BioSpectrometer kinetic is to be used in molecular biology, biochemistry and cell biology research
laboratories. The BioSpectrometer kinetic is exclusively intended for use indoors. All country-specific
safety requirements for operating electrical equipment in the laboratory must be observed.
The BioSpectrometer kinetic is used for photometric concentration determination of analytes in liquids and
recording of absorbance wavelength spectra in cuvettes.
Only use Eppendorf accessories or accessories recommended by Eppendorf.
3.2User profile
The device and accessories may only be operated by trained and skilled personnel.
11
Before using the device, read the operating manual carefully and familiarize yourself with the device's
mode of operation.
3.3Warnings for intended use
3.3.1Personal injury
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
DANGER! Risk of explosion.
Do not operate the device in areas where work is completed with explosive substances.
Do not use this device to process any explosive or highly reactive substances.
Do not use this device for processing any substances which could generate an explosive
atmosphere.
WARNING! Electric shock due to damage to device or mains cable.
Only switch on the device if the device and mains cable are undamaged.
Only use devices that have been properly installed or repaired.
In case of danger, disconnect the device from the mains supply by pulling the power plug
from the device or the mains socket or, by using the isolating device intended for this
purpose (e.g., emergency stop switch in the laboratory).
Page 12
12
Safety
Eppendorf BioSpectrometer
English (EN)
WARNING! Damage due to UV radiation.
Microliter cuvettes, e.g., Hellma® TrayCell (or microliter cuvettes with a similar design) divert
the radiation from the light source within the cuvette so the radiation can escape upward
when the lid is not closed.
Before starting a measurement, ensure that the lid on the microliter cuvette is not open.
WARNING! Damage to health from toxic, radioactive or aggressive chemicals as well as
infectious liquids and pathogenic germs.
Observe the national regulations for handling these substances, the biological security
level of your laboratory, the material safety data sheets and the manufacturer's application
notes.
Wear personal protective equipment.
For comprehensive regulations about handling germs or biological material of risk group II
or higher, please refer to the "Laboratory Biosafety Manual" (source: World Health
Organization, Laboratory Biosafety Manual, in its respectively current valid version).
®
kinetic
WARNING! Damage to health due to contaminated device and accessories.
Decontaminate the device and the accessories before storage and shipping.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may
impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or
accept any liability for damage resulting from the use of incorrect or non-recommended
accessories and spare parts, or from the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
Page 13
3.3.2Damage to device
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device has been contaminated by aggressive chemicals, immediately clean it by
means of a mild cleaning agent.
NOTICE! Damage to the device from fumigating with aggressive chemicals.
Do not use fumigation to disinfect the device.
NOTICE! Corrosion from aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive polishes.
Do not incubate the accessories in aggressive cleaning agents or disinfectants for a longer
period of time.
Eppendorf BioSpectrometer
English (EN)
Safety
®
kinetic
13
NOTICE! Damage and measurement errors due to condensation water.
With high humidity, condensation water can form on a cuvette with a temperature
significantly lower than ambient temperature. The condensate may cause damage to the
optics as well as incorrect measuring results.
Do not insert any cuvettes into the cuvette shaft with a temperature that is significantly
lower than the ambient temperature.
The temperature of the cuvette should not remain significantly below the ambient
temperature for a longer period of time.
If required, observe the actual dew point.
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a
warmer environment.
After installing the device, wait at least for 3 h. Only then connect the device to the mains.
NOTICE! Function impairment due to mechanical damage.
After mechanical damage to the device, ensure that the measuring and evaluation
functions of the device are operating correctly by completing an inspection.
NOTICE! Damage from overheating.
Do not install the device near to any heat sources (e.g., heating, drying cabinet).
Do not expose the device to direct sunlight.
Ensure unobstructed air circulation. Keep free a clearance of at least 5 cm around all
ventilation grilles.
Page 14
14
Gerät nach dem Öffnen
justieren!
Adjust device after
opening!
Safety
Eppendorf BioSpectrometer
English (EN)
NOTICE! Material damage from incorrect use.
Only use the product for its intended purpose as described in the operating manual.
Ensure adequate material resistance when using chemical substances.
In case of doubt, contact the manufacturer of this product.
NOTICE! Damage as a result of incorrect packing.
Eppendorf AG is not liable for damage caused by improper packing.
The device may only be stored and transported in its original packaging.
NOTICE! Damage due to improper cleaning of the cuvette shaft.
Only clean the cuvette shaft using a moist cotton swab (see Cleaning on p. 65).
Do not allow any liquid to enter the cuvette shaft.
Do not reach with your fingers into the cuvette shaft.
®
kinetic
3.4Information on product liability
In the following cases, the designated protection of the device may be compromised. Liability for any
resulting property damage or personal injury is then transferred to the operator:
• The device is not used in accordance with the operating manual.
• The device is used outside of its intended use.
• The device is used with accessories or consumables which are not recommended by Eppendorf.
• The device is maintained or repaired by people not authorized by Eppendorf.
• The user makes unauthorized changes to the device.
3.5Safety instructions located on the device
DepictionMeaningLocation
Hazard point
Follow the operating manual.
The device needs to be readjusted
after it has been opened.
Do not open the device.
Rear side of the device
Bottom of the device
Page 15
Installation
®
Eppendorf BioSpectrometer
kinetic
English (EN)
4Installation
4.1Preparing installation
Keep the transport carton and the packing material for subsequent safe transport or storage.
Check the completeness of the delivery using the information in the delivery package (see Delivery
package on p. 9).
Check all parts for any transport damage.
4.2Selecting the location
Select the location for the BioSpectrometer kinetic according to the following criteria:
• 2 grounded sockets for the BioSpectrometer kinetic and for the printer.
• Solid laboratory bench with horizontal work surface
Space requirement of the device: 50 cm (with printer: 75 cm) width, 50 cm depth.
• Temperature: 15°C to 35°C.
• Avoid temperature fluctuations (e.g, caused by open windows).
• Avoid direct sunlight.
• Humidity: 25% to 70% relative humidity.
15
Ensure that no objects (e.g., loose sheets, notebooks) that could impede the flow of air are
positioned under the device.
4.3Connecting the device to the mains
1. Place the BioSpectrometer kinetic on a suitable work surface.
2. Verify that the mains/power supply voltage and mains/power frequency match the information on the
name plate.
3. Connect the device to the mains/power line and switch it on with the power switch.
4. Remove the protective film from the display.
4.4Connecting the printer
4.4.1Thermal printer DPU-S445
Prerequisites
Software version 3.4.4.0 or higher is installed on the device.
Connect the thermal printer DPU-S445 to the USB port for printers.
1. Connect the printer cable with the USB port for printers 4 (see Main illustration on p. 9).
2. Connect the printer cable with the printer.
3. Connect the printer to the mains/power line using the supplied mains/power adaptor and mains/power
cord (printer accessory) and switch it on.
For information on the printer, refer to the operating manual of the printer.
Page 16
16
Installation
®
Eppendorf BioSpectrometer
kinetic
English (EN)
4.4.2Thermal printer DPU-414
Connect the thermal printer DPU-414 to the serial printer connection.
1. Connect the printer cable to the serial printer connection 9 and tighten the locking screws. (see Main illustration on p. 9).
2. Connect the printer cable to the printer and tighten the locking screws as well.
3. Connect the printer to the mains/power line using the supplied mains/power adaptor and mains/power
cord (printer accessory) and switch it on.
Information about modifying printer settings can be found in the operating manual for the printer.
The DIP switches are preset for the BioSpectrometer kinetic according to the following table.
Tab. 4-1: Setting the DIP SW for the thermal printer
DIP SW-1Meaning
1 (OFF)Input = Serial
2 (ON)Printing Speed = High
3 (ON)Auto Loading = ON
4 (OFF)Auto LF = OFF
5 (ON)Setting Command = Enable
6 (OFF)Printing
7 (ON)Density
8 (ON)= 100%
DIP SW-2Meaning
1 (ON)Printing Columns = 40
2 (ON)User Font Back-up = ON
3 (ON)Character Select = Normal
4 (ON)Zero = Normal
5 (ON)International
6 (ON)Character
7 (ON)Set
8 (OFF)= U.S.
DIP SW-3Meaning
1 (ON)Data Length = 8 bits
2 (ON)Parity Setting = NO
3 (ON)Parity Condition = Odd
4 (OFF)Busy Control = XON/XOFF
Page 17
Installation
Eppendorf BioSpectrometer
English (EN)
DIP SW-3Meaning
5 (OFF)Baud
6 (ON)Rate
7 (ON)Select
8 (ON)= 9600 bps
4.5Connecting PC or USB stick for data export
You can connect a FAT 32-formatted USB stick to the USB port 4 (see Main illustration on p. 9).
Alternatively, you can connect the device for the data export directly to a PC by using a USB cable:
Prerequisites
• PC with Windows, version XP, SP2 or higher version.
• USB cable with a type A and type B plug each.
®
kinetic
17
Connect the device to the PC by using the USB cable on the USB port 8 (see Main illustration on p. 9).
• You do not need any special PC software for the data transmission: the transferred data
packets are recognized by the PC like a USB stick as a removable medium. For viewing the
data, you only need to open the registered data packet.
• The transmission of data to the USB stick or to the PC is started after completing the series
of measurement in the print & export (see Print & export on p. 55) method step.
Page 18
18
Installation
Eppendorf BioSpectrometer
English (EN)
®
kinetic
Page 19
5Operation
5.1Overview of operating controls
Abb. 5-1: Control panel of the BioSpectrometer kinetic
Eppendorf BioSpectrometer
English (EN)
Operation
®
kinetic
19
Fig. 5-1:Control panel of the BioSpectrometer kinetic
Key:Function
Keypad: Enter digits and text.
Keys 1 to 9 as well as 0: When entering text, next to numbers you also can
enter letters and special characters by pressing the key several times.
Alternatively, you can switch to a displayed keyboard with the [Keyboard] key.
Outside of entry fields: Call up method selection.
Outside of entry fields: Call up function selection.
Softkey: Select functions.
The key assignment changes along with the software dialog. The current
function is displayed directly above the key on the display.
Page 20
20
Operation
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Key:Function
Move the cursor to the left, right, up, down.
• Navigation between input fields.
• and keys inside an entry field: Navigate within the character string.
• and keys in a result display: Navigate between the sample results of
the series of measurement.
• and keys within a graph: Navigate on the x-axis of the graph, e.g. for
displaying the wavelength-dependent absorbance values in a scan.
and keys in an absorbance wavelength spectrum: Change image
section (SpectraZoom procedure) (see Tab. on p. 51).
Exit the current selection for the next higher level.
Delete entry. Within a sequence of signs, the sign on the left of the cursor is
deleted
•Call up selected method or function.
• Open the selection list.
• Confirm entry or selection.
Start standard measurement.
Start blank measurement.
Start sample measurement.
Page 21
Operation
®
Eppendorf BioSpectrometer
kinetic
English (EN)
5.1.1Entering text
You can enter texts when assigning method names and result units. Restriction: Only digits, letters and the
underscore "_" are allowed for method names.
Entry via keyboard:
Use the and cursor keys to navigate within the
entry field and to change single positions in the
name.
Softkeys:
• [Keyboard]: Display keyboard.
• [abc]: Change between upper and lower case
letters when making entries with the keypad.
• [Save]: Save entered text.
• [Cancel]: Cancel text input.
21
Entry via the displayed keyboard:
Use the cursor keys to select the displayed signs and
respectively confirm your selection with the enter
key. As for a PC key pad, you can use the "Shift"
resp. the "Caps Lock" key for changing the
capitalization for the next entry or for all following
entries.
Softkeys:
• [Numbers]: Switch to entry using the keyboard.
• [Save]: Save entered text.
• [Cancel]: Cancel text input.
5.2Inserting the cuvette
Standard rectangular glass or plastic cuvettes can be inserted in the cuvette shaft:
• External dimensions: 12.5 mm × 12.5 mm
• Height of light path: 8.5 mm higher than cuvette base
• Total height: min. 36 mm
The cuvettes must be optically transparent for the respective measuring wavelength. For measurements in
the UV range, Eppendorf offers the plastic cuvette UVette which is transparent for wavelengths of 220 nm
and higher and therefore also is suitable for measuring nucleic acids.
Page 22
22
Operation
Eppendorf BioSpectrometer
English (EN)
®
kinetic
Cuvettes
Basic area 12.5 mm × 12.5 mm
Min. overall height 36 mm
Min. filling level 10 mm
Light path 8.5 mm
Max. height of base 7 mm
0 mm
Min. volume Photometry
Eppendorf
µCuvette G1.0
See manufacturer
information
®
Hellma
See manufacturer
* or similar microliter cuvette
TrayCell *
information
®
50 µL70 µL400 µL1000 µL
Prerequisites
• The cuvette is free from contamination by dust or fingerprints and free from scratches.
• The cuvette shaft is free from particles, dust and liquid.
• The measuring volume in the cuvette is sufficient. Ensure that the minimum measuring volume has
been reached.
• The measuring solution is free from particles and bubbles.
The direction of the light path is marked with an arrow on the housing.
MacroSemi-microUltra-microUVette
1. Position the cuvette so that the optical window of the cuvette is pointing towards the direction of the
light path.
2. When inserting the cuvette, press it completely to the bottom against the slight resistance.
5.3Summary of the measuring procedure
5.3.1Preparing the measurement
1. Switch on the device and, if required, the printer.
The device performs a self test (taking approx. 1 minute) and displays the method selection.
2. Make ready the cuvettes for the measurements (see Inserting the cuvette on p. 21).
3. Prepare the measuring solutions for measuring the blank values, if required, also the standards and the
samples.
4. Open the cover of the cuvette shaft. The cover can remain open during the measurements.
You should not use any measuring solution for standards and samples with a lower
absorbance than 0.02 to 0.03 A (e.g. dsDNA concentration between 1.0 and 1.5 μg/mL). The
detection limit of the device may be significantly lower, nevertheless, the impact of
disturbances from the measuring solutions (particles, bubbles, turbidity) on the reliability of
the result is very high for these low absorbance values.
Page 23
5.3.2Measuring procedure
5.3.2.1 Selecting a method
Operation
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Use the cursor keys to select the desired
method and call up the method with the enter
key.
For an overview and a detailed description of the
methods, refer to the next chapter (see Methods on p. 29).
Wizard: The wizard at the top of the display will
take you through the method procedure
step-by-step.
Help box: You will receive help texts in the lower
right of the display during each step of the
procedure.
Softkeys: The [< Back] and [Next >] softkeys allow
you to move between method steps in the wizard.
23
5.3.2.2 Checking parameters
Check the parameter setting. The [Page dn] and
[Page up] softkeys allow you to call up the
parameter list pages. You can modify and save
parameters using [Edit].
Page 24
24
Operation
Eppendorf BioSpectrometer
English (EN)
®
kinetic
5.3.2.3 Measuring the blank and standards
For evaluations without standards (e.g. DNA measurements), this method step is omitted.
1. Start by measuring a blank (blank key).
2. Then measure all standards one by one
(standard key).
The display always marks the standard that is to be
measured next. Use the [Graph] resp. [Table]
softkey to change the result view.
5.3.2.4 Measuring samples
Press [Next] to accept the evaluation calculated
from the standard results.
The sample key is used for measuring your
samples consecutively.
Blank results will remain saved for the duration of
one series of measurement. However, a new blank
measurement always is possible. (The adjacent
figure shows a measuring procedure with
evaluation via the standard curve and, in addition
to the sample result, the graph of the standard
evaluation.)
Page 25
5.3.2.5 Finalizing the method
5.3.2.6 Optional: process results
Operation
®
Eppendorf BioSpectrometer
1. Press [Finish], to complete the measuring series
and return to the method selection.
2. After all measurements have been completed,
switch off the device and close the cuvette shaft
cover to protect the cuvette shaft from
contamination.
kinetic
English (EN)
25
5.3.2.7 Printing and exporting
For some methods, you can postprocess the results
in the process results method step. For example,
you can use the SpectraZoom zoom function in the
spectra.
Use the and cursor keys for selecting
systematically any results of the series of
measurement for postprocessing.
1. Compose data packets for all samples or for
selected samples.
2. Print the data, save them to a USB stick or
transfer them to a PC via a USB cable.
Page 26
26
Operation
Eppendorf BioSpectrometer
English (EN)
®
kinetic
5.3.3Important measurement instructions
Check for each measurement:
• For plastic cuvettes: How many consecutive measurements can be reliably carried out in
the cuvette?
• Measure the cuvette blank value before the sample or standard measurements in order to
compensate the cuvette blank value in addition to the reagent blank.
• Blank results remain saved for one measuring series, but a new blank result measurement
can be performed at any time, even between sample measurements.
• The displayed absorbance values always correspond to the directly measured values. The
dilution or cuvette factor as well as background absorbances only will be incorporated for
the following result calculation (see Absorbance values on p. 89).
• The measuring result is typically displayed 2 to 3 seconds after a measurement has been
started. If a small amount of light reaches the receiver, the measuring time automatically
can be extended to 9 seconds in order to increase the precision of the measurement. For
kinetic measurements, the automatic extension of the measuring time is not applied in
order to prevent any conflicts with the preprogrammed interval time for the measuring
point recording.
• Observe that the measured absorbance values do not exceed the upper limit of the
photometric measuring range. In this case, reject the measuring result. The upper limit of
the photometric measuring range does not only depend on the wavelength (see
Photometric properties on p. 86) but also on the cuvette blank. Ultra-micro cuvettes with a
small diaphragm, such as TrayCell (Hellma), may have a cuvette blank of approx. A = 1.
The available photometric measuring range is reduced by this amount. You can estimate
the cuvette blank by measuring the cuvette filled with demineralized water as a sample in
comparison with the empty cuvette shaft as a blank. The cuvette blank of the Eppendorf
μCuvette G1.0 is negligible (approximately A = 0).
• After the measurement, remove the measuring solution completely before filling in the
next measuring solution in order to minimize carry-over. If a carry-over from one sample to
the next sample can be expected due to a high concentration difference, rinse the cuvette
between the measurements.
• If the temperature between the lamp and the ambience differs, photometric drift may
occur. Therefore a device from a colder ambience first has to be adjusted to the ambient
temperature.
Avoid quick changes of temperature. Carry out a new blank measurement for a long series
of measurements or measurements over a long period of time.
5.3.4Notes on working with cuvette temperature control
Temperature control is regulated using a measurement on the cuvette holder. The temperature in the
measuring solution may deviate from the temperature on the cuvette holder.
The extent of the deviation is dependent on the measuring volume, cuvette material, cuvette shape and
ambient temperature. The temperature control speed is also dependent on these factors. The temperature
control of plastic cuvettes is slower compared to glass cuvettes. The surface of the cuvette in direct contact
with the wall of the cuvette holder should be as large as possible to ensure quick temperature control.
Therefore, the temperature control of plastic semi-micro cuvettes as well as, for example, the UVette is only
performed slowly.
Page 27
Operation
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Measured values typical for Eppendorf for temperature control in closed cuvettes, with closed cuvette shaft
covers, are shown in the following tables. The temperature was measured in the measuring solution; the
ambient temperature was 24.5°C.
Tab. 5-1: The temperature to be set for temperature control of a measuring solution
Cuvette type/measuring volume Target temperatureTemperature to be set in the
method parameters
Quartz glass macrocuvette,
1 500 μL
25°C25°C
30°C30.4°C
37°C37.7°C
Plastic macrocuvette,
1 000 μL
25°C25°C
30°C30.4°C
37°C37.7°C
Quartz glass semimicrocuvette,
500 μL
25°C25°C
30°C30.4°C
37°C37.7°C
Quartz glass ultra-micro cuvette,
60 μL
25°C25°C
30°C30.3 °C
37°C37.6°C
27
Tab. 5-2: Duration for temperature control of a measuring solution
Cuvette type/measuring volume Initial temperature and target
temperature
Quartz glass macrocuvette,
1 500 μL
25°C - 37°CApprox. 7 min
37°C - 25°CApprox. 11 min
25°C - 30°CApprox. 7 min
Plastic macrocuvette,
1 000 μL
Quartz glass semimicrocuvette,
500 μL
Quartz glass ultra-micro cuvette,
60 μL
25°C - 37°CApprox. 13 min
37°C - 25°CApprox. 19 min
25°C - 37°CApprox. 7 min
37°C - 25°CApprox. 12 min
25°C - 37°CApprox. 7 min
37°C - 25°CApprox. 9 min
25°C - 30°CApprox. 5 min
Temperature control duration
Page 28
28
Operation
Eppendorf BioSpectrometer
English (EN)
• For efficient temperature control, the volume of the measuring solution in the cuvette
should not project beyond the edge of the cuvette holder.
• To speed up the measuring procedure during series measurements, you can pre-cool
cuvettes with reagents in a thermostat outside the BioSpectrometer before inserting the
cuvette in the cuvette holder and adding the sample.
• When switching from a method with temperature control to one without temperature
control, please note that the temperature of the cuvette holder will slowly drift back to
room temperature. The results of the method without temperature control could be
affected by this.
®
kinetic
Page 29
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
6Methods
6.1Selecting a method
Methods and method templates are delivered preprogrammed. The methods are organized in main groups
and subgroups.
29
Write-protected methodsThe most important methods in molecular biology.
Parameters can be modified, but the modified parameters
must be saved under a new method name.
Non-write-protected methodsYou can change parameters any number of times and start
the measurement right after saving.
New methods ("templates")Each method group contains a template which is
preprogrammed with complete parameter sets to facilitate
the programming of new methods. The parameters can be
changed and saved under new names any number of times.
To call a method, first use the cursor keys to select the main group, subgroup and the method. Confirm
each with enter.
Tab. 6-1: Photometric methods
AbsorbanceMethods for fast, simple absorbance measurements without additional
evaluations
RoutineFrequently used molecular biology methods. The methods are
preprogrammed. However, the parameters can be modified if saved under
a new name.
BasicMethods for the evaluation of absorbance measurements with factor,
standard or standard curve/line. Method templates for measuring and
evaluating kinetics.
AdvancedMethods for the evaluation of two wavelength measuring methods and for
kinetics with more demanding evaluation options.
FavoritesIn Favorites, you can set up your own folders using <New folder>, and
copy your frequently used methods to this folder in order to quickly
access them when needed.
Page 30
30
Methods
Eppendorf BioSpectrometer
English (EN)
You can create new methods in all folders using <New Method>.
In Favorites, you can create your own folders (e.g., to allocate folders to specific people), and rename and
delete the folders.
Tab. 6-2: Softkeys in method selection
[Cut] and [Paste]Cut and paste methods.
[Copy] and [Paste]Copy and paste methods.
[Delete]Delete methods.
[Rename]Rename methods.
Copied or cut methods can be added to a different folder under Favorites, or added to the original folder
under a new name. Use the cursor keys to navigate to the Methods column of the desired folder and press
[paste] for adding the method.
®
kinetic
6.2Photometry method description
The preprogrammed methods and method templates are described in this section.
6.2.1Absorbance method group
Single λ
• Absorbance measurement on a wavelength.
• No subsequent evaluation.
Single λ - continuous
• Repeated absorbance measurement on a wavelength.
• Parameters can be entered for temperature control between 20 and 42°C (presetting: 37°C).
• Entry of parameters for total time and interval time for the measuring points. A premature stop cannot
be made during the measurement.
• Evaluation as kinetic via linear regression. A subsequent change cannot be made to the time frame for
the evaluation.
• The measured values are displayed in an absorbance time graph.
• To subsequently evaluate measured values as linear regression kinetics, press the [Next >] softkey and
go to the process results method step.
Multi λ
• Absorbance measurement at two to six wavelengths.
• No subsequent evaluation.
Scan
• Absorbance wavelength spectra measurement via a defined wavelength range.
• Display of wavelength and absorbance in the spectrum by navigation with a wavelength cursor.
• The spectra section can be modified using 3 different zoom variants.
• Peak detection possible.
Page 31
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
6.2.2Routine method group
The methods for the Routine group are preprogrammed as fixed methods. Therefore, a new method name
is required after the method parameters in the fixed preprogrammed methods have been modified.
Nucleic acids
• Determination of the concentration of nucleic acids through measurement at 260 nm and evaluation via
factor.
• Various nucleic acid methods, such as dsDNA or RNA, are preprogrammed. The parameters vary
according to the factor.
• Preprogrammed method for microliter cuvettes: Measuring DNA in sample volumes within the
microliter range with 1 mm light path (with microliter cuvettes as Eppendorf μCuvette G1.0 or Hellma
TrayCell).
• Additional information on the purity of the measured nucleic acid: Ratios A260/A280, ratios A260/A230,
absorbance wavelength spectrum of nucleic acid, absorbance of the background wavelength (preset:
320 nm; the absorbance of the pure nucleic acid should be close to zero here).
• Partial turbidity correction can be performed via the Background parameter.
• Concentrations can be converted to molar concentrations and (after the sample volume has been
entered) to nucleic acid quantities (process results method step).
31
®
Proteins direct UV
• Determination of the concentration of proteins via measurement at 280 nm and factor or standard
evaluation.
• Preprogrammed methods for direct absorbance output as a result (Protein A 280) and for evaluation via
albumin-specific absorbance coefficients (Albumin A 280).
• Preprogrammed method for microliter cuvettes: Measuring protein in sample volumes in the microliter
range with 1 mm light path (with microliter cuvettes as Eppendorf μCuvette G1.0 or Hellma
®
TrayCell).
• Additional information on the purity of the measured protein: Absorbance of the background
wavelength (preset: 320 nm; the absorbance of the pure protein should be close to zero here).
• Partial turbidity correction can be performed via the Background parameter.
• When programming the methods, the corresponding factor is imported through the simple selection of
the protein from a predefined list. The factors are separately defined in the functions of the Gen. method param. group. Various proteins are preprogrammed in Gen. method param.; additional
proteins can be added.
Proteins (with reagent)
• Concentration determination of proteins via measurement according to color reactions and evaluation
using standards or factors (typical: evaluation with standard curve).
•The Bradford, Bradford micro, Lowry, Lowry micro, BCA and BCA micro methods are already
preprogrammed. According to the reagent manufacturer, the "Curve fit" (standard curve type) must be
changed as necessary.
Dye labels
• For dye-labeled biomolecules: Concentration determination of the biomolecule (nucleic acid or protein)
via measurement at 260 or 280 nm and measurement of the dye in one measuring procedure.
• Evaluation with factor. In addition to the biomolecule, up to two dyes can be measured at the same time
as two different wavelengths.
• Additional: evaluation of the frequency of incorporation (FOI) of the dye. Selection between two
different FOI calculation procedures.
• Already preprogrammed methods: ssDNA, labeled with Cy 3 or Cy 5.
Page 32
32
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
• Correction of the influence of the dye spectrum on the accuracy of the biomolecule measurement is
possible.
• Partial turbidity correction can be performed via the Background parameter.
• Additional information on the purity of the measured materials: Ratios A260/A280 and ratios A260/A230
(ratio values only for nucleic acids), absorbance wavelength spectrum.
• When the methods are programmed, various associated parameters, such as wavelengths and
evaluation factors, are imported by simply selecting the biomolecule and the dye from predetermined
lists. These parameters are separately defined in the functions of the Gen. method param. group.
Various nucleic acids, proteins and dyes are preprogrammed in Gen. method param.. You may add
further nucleic acids, proteins and dyes.
• Only for labeled nucleic acids: Concentrations can be converted to molar concentrations and (after the
sample volume has been entered) to nucleic acid and dye quantities (process results method step).
Bacterial density
• Turbidity measurement to determine the bacteria density.
• Measurement on a wavelength and factor or standard evaluation.
• Methods for factor and standard evaluation are preprogrammed.
Calibration curve
• Measurement on a wavelength and subsequent evaluation with a series of 2 to 12 standards.
• You can select between different evaluation procedures ("Curve fit") as linear regression, non-linear
regression.
• Graphical and tabular display of the standard results.
• The last saved standard evaluation can be used.
• A method for standard curve evaluation is preprogrammed.
Simple kinetics
• Measurement of a kinetics at one wavelength and subsequent evaluation by factor.
• 3 different measurement methods are available:
– "Linear regression": Evaluation in a series of measuring points recorded in intervals.
– "Two point": Calculation of ΔA/min using 2 measuring points at defined times.
– "Endpoint": Recording of a measuring point at a defined time.
• Parameters can be entered for temperature control between 20 and 42°C (presetting: 37°C).
• For the "linear regression" measurement method, the absorbance/time curve is depicted as a graph.
• A method for the evaluation with linear regression is preprogrammed and can be changed for other
evaluation procedures ("endpoint", "two.point") via simple selection.
Page 33
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
6.2.4Advanced method group
Dual wavelength
• Measurement at two wavelengths and evaluation of the measured absorbance values via two basic
formulas (subtraction, division)
• Variants of the basic formulas can be defined.
• The result can be evaluated with a factor, with a standard or with a standard series.
• Methods are preprogrammed for calculation, subtraction and division, and subsequent factor
evaluation.
Advanced kinetics
The following options are available via the description of the Simple kinetics method (Basic method group):
• Measurement of a reagent blanks kinetics. The blank result is subtracted from all sample results prior to
the evaluation.
• An alternative to evaluation via factor is evaluation via a standard.
33
6.3Method parameters
This chapter illustrates the parameters for programming the methods. The order of the parameters in the
device display may slightly differ from the order in the table in order to display the parameters more clearly.
The table displays all parameters available for the various methods. Only a small portion of these
parameters are required for the corresponding method and will be shown in the display.
ParameterEntryExplanation
CuvetteSelection:
10 | 5 | 2 | 1 | 0.5 | 0.2 |
0.1 mm
No. of
wavelengths
WavelengthValue input:
Value input:
Range: 2 to 6.
Measurement
wavelength in nm.
Range: 200 to 830 nm.
Optical path length of the cuvette. The device always
automatically converts absorbance values to the 10 mm path
length of a standard cuvette (see Absorbance values on p. 89).
Therefore, there is no need to change factors such as "50" for
the calculation of dsDNA concentrations when modifying the
Cuvette parameter.
Only for the Multi λ method group.
Number of wavelengths at which the measurement is to be
performed.
Measurement wavelength: The concentration is calculated
based on the absorbance measured with this wavelength.
For the Multi λ and Dual wavelength method groups, enter
more than one wavelength. For some method groups (e.g.,
Nucleic acids and Proteins direct UV), the wavelengths are
preprogrammed.
For the Dye labels method group the measuring wavelengths
are not entered into the method procedure one by one. They
are automatically imported simply by selecting the biomolecule
and dye from the General Method Parameters function.
Page 34
34
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
ParameterEntryExplanation
UnitSelection:
mg/mL | μg/mL |
ng/mL | pg/mL | μg/μL
Unit for the concentration result.
In the preprogrammed methods of the Routine group, the
selection is restricted to units that are useful for these methods.
| mg/dL | μmol/mL |
nmol/mL | pmol/mL |
pmol/μL | U | U/mL |
U/L | % | Abs | A/min
In addition, further
units are freely
programmable in the
General Method
Parameters/Units
function. Max. 7 digits.
Formula typeSelection:
division | subtraction
Only for the Dual wavelength method group.
Formula type for calculating the absorbances at the two
measurement wavelengths prior to evaluation with factor or
standard.
Formula: aValue input:
Value for a in the
evaluation formula.
Only for the Dual wavelength method group.
Value for a in the formulas:
[(a*A1) / (b*A2)] * c + d and [(a*A1) - (b*A2)] * c + d.
Limit: max. of 5 digits
including decimal
point.
Formula: bValue input:
Value for b in the
evaluation formula.
Only for the Dual wavelength method group.
Value for b in the formulas:
[(a*A1) / (b*A2)] * c + d and [(a*A1) - (b*A2)] * c + d.
Limit: max. of 5 digits
including decimal
point.
Formula: cValue input:
Value for c in the
evaluation formula.
Only for the Dual wavelength method group.
Value for c in the formulas:
[(a*A1) / (b*A2)] * c + d and [(a*A1) - (b*A2)] * c + d.
Limit: max. of 5 digits
including decimal
point.
Formula: dValue input:
Value for d in the
evaluation formula.
Only for the Dual wavelength method group.
Value for d in the formulas:
[(a*A1) / (b*A2)] * c + d and [(a*A1) - (b*A2)] * c + d.
Limit: max. of 5 digits
including decimal
point.
Calc
ulationSelection:
factor | standard
Evaluation procedure for the calculation of the sample
concentration from the measured absorbance.
Page 35
ParameterEntryExplanation
FactorValue input:
Factor.
Limit: max. of 6 digits
including decimal
point.
Factor for converting absorbance values into the concentration.
Negative factors can also be entered for the following method
groups: Simple kinetics,Advanced kinetics,Dual
wavelength,Factor.
For the Dye labels method group the factors are not entered
into the method procedure individually. They are automatically
imported simply by selecting the biomolecule and dye from the
General Method Parameters function.
ProteinSelection:
List of protein types
which are stored in the
General Method
Parameters/Proteins
Only for the Dye labels and Proteins direct UV method
groups.
When selecting the protein, the corresponding Factor
parameter programmed in the General Method Parameters/
Proteins function also will be imported from that function.
function.
StandardsValue input:
Number of standards.
Range: 1 to 12.
Number of different standard concentrations for the evaluation
with standards.
For some methods the range for the number of standards is
restricted to a smaller range than 1 to 12.
ReplicatesValue input:
Number of replicates
Number of repeated measurements for the various standard
concentrations.
per standard.
Range: 1 to 3.
Std. conc.Value input:
Concentration values
Based on the number of standards, this parameter is available
for all standards (e.g.,: std. conc. 1, std. conc. 2, ...).
of the standards.
Limit: max. of 6 digits
including decimal
point.
Decimal placesValue input:
Number of decimal
Number of decimal points for the calculated concentration
result.
points for the result.
Range: 0 to 3.
Dye 1Selection:
List of dyes that are
stored in the General
Method Parameters/
Dyes function.
Only for the Dye labels method group.
When the dye is selected, the parameters programmed in the
General Method Parameters/Dyes function, which correspond
to the dye, are also imported: factor, wavelength and, if
necessary, correction factors for the measurement at 260 or
280 nm (see the description of the following parameter).
Eppendorf BioSpectrometer
English (EN)
Methods
®
kinetic
35
Page 36
36
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
ParameterEntryExplanation
Correct A260 1Selection:
on | off
Only for the Dye labels method group.
Correction of the influence of the dye spectrum on the
absorbance with the measuring wavelength of the biomolecule
(260 nm or 280 nm). Some of the dye spectra have a low
absorbance at 260 and 280 nm. These absorbances distort the
calculations for the nucleic acids or the proteins of these
methods. To minimize the distortion correction factors are
used, as long as these are known for the corresponding dyes.
When the parameter is switched on, the correction factor is
imported from the General Method Parameters/Dyes function.
Correct A 280 1 Selection:
on | off
Only for the Dye labels method group.
For an explanation see the description of the Correct A260 1
parameter above.
Dye 2 activeSelection:
on | off
Only for the Dye labels method group.
Option to measure a second dye at the same time. Application:
Labeling a biomolecule with two dyes.
Dye 2Selection:
List of dyes that are
Only for the Dye labels method group when measuring 2 dyes.
Selection of the second dye (see Dye 1 parameter).
stored in the General
Method Parameters/
Dyes function.
Correct A260 2Selection:
on | off
Correct A280 2Selection:
on | off
Show scanSelection:
on | off
Start λValue input:
Only for the Dye labels method group when measuring 2 dyes.
Analog to parameter Correct A260 1.
Only for the Dye labels method group when measuring 2 dyes.
Analog to parameter Correct A280 1.
Display of a scan (absorbance wavelength graph) in addition to
the result of the sample measurement.
Starting wavelength for recording the scan.
Wavelength in nm.
Range: 200 to 830 nm.
Stop λValue input:
Stopping wavelength for recording the scan.
Wavelength in nm.
Range: 200 to 830 nm.
Value must be higher
than the value for Start λ.
A260/A280Selection:
on | off
Only for nucleic acids.
Display of the A260/A280 ratio in addition to the result of the
sample measurement.
A260/A230Selection:
on | off
Only for nucleic acids.
Display of the A260/A230 ratio in addition to the result of the
sample measurement.
Page 37
ParameterEntryExplanation
FOISelection:
none | dye/kb | pmole/
μg
Only for the Dye labels method group.
Display of the FOI in addition to the result of the sample
measurement.
The FOI (frequency of incorporation) is a measure for the
number of dye molecules per nucleic acid molecule that are
integrated into the nucleic acid. Units are "dye/kb" (dye
molecules per 1000 bases) or "pmole/μg" (pmol dye per μg
nucleic acid). "None": no FOI calculation.
BackgroundSelection:
on | off
Prior to the calculation of the results of a sample the
absorbance of a background wavelength, during which the
analyte to be measured should exhibit the absorbance value
zero, is subtracted from the absorbance of the measuring
wavelength. Frequent application: Partial correction of
turbidity for measurement of nucleic acids (background
wavelength in this case: 320 nm or 340 nm).
WavelengthWavelength in nm.
Range: 200 to 830 nm.
Wavelength at which the background is to be measured. The
analyte to be measured should have the absorbance value zero
in pure form here.
Background for
dyes
Selection:
on | off
Only for the Dye labels method group.
Application of the background correction to the measurement
of a dye (see Background parameter).
WavelengthWavelength in nm.
Range: 200 to 830 nm.
Only for the Dye labels method group.
Wavelength at which the background for the dye is to be
measured. The pure, not contaminated dye to be measured
should have the absorbance value zero for this wavelength.
Temperature on Selection:
on | off
TemperatureValue input:
Temperature in °C.
Range: 20 to 42°C.
For kinetic methods only.
Use of cuvette temperature control.
For kinetic methods only.
Temperature input for the cuvette temperature control when
the Tem perature parameter is set to "on".
Notice! Damages and measurement errors due to
condensate.
At high air humidity, condensate may form on a cuvette having
a significantly lower temperature than the ambient
temperature. The condensate may cause damage to the optics
and lead to incorrect measuring results.
Eppendorf BioSpectrometer
English (EN)
Methods
®
kinetic
37
The temperature of the cuvette should not remain
significantly below the ambient temperature for a longer
period of time. If required, observe the actual dew point.
Page 38
38
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
ParameterEntryExplanation
Measuring
procedure
Selection:
lin.regr. | endpoint |
two point
For kinetic methods only.
"Linear regression": Measurement over several periods in set
time intervals within a defined period of time. Evaluation via
linear regression of the absorbance time graph within the
measuring time. Absorbance result: ΔA/min.
"Endpoint": Measurement of a measuring point after a defined
period of time. Absorbance result: A.
"Two point": Measurement of 2 measuring points at defined
times. Evaluation: Via linear interpolation between these
measuring points in the absorbance time graph. Absorbance
result: ΔA/min.
Reagent blankSelection:
on | off
For kinetic methods of the Advanced kinetics method group
and evaluation with factor only.
Measurement of a reagent blank (RB). The RB is measured
according to the same measurement method as a sample. The
absorbance result in A or ΔA/min is subtracted from the
absorbance result of the sample before the sample
concentration is calculated. Application: Correction of sample
results for kinetics with reagent drift. The reagent blank
contains the reagent and demineralized water as sample.
DelayValue input:
Time from the start to
the first measuring
For kinetic methods only.
Time from the start of the measuring procedure to recording
the same measuring point.
point.
Range: 00:00 to 10:00
min:sec.
Measuring time Value input:
Time between first and
last measuring point.
Only for kinetic methods and the "lin.regr." and "two point"
measurement methods.
Time between recording the first and the last measuring point.
Range: 00:05 to 59:59
min:sec.
Total timeValue input:
Time between first and
For the Single λ - cont kinetic method only.
Time between recording the first and the last measuring point.
last measuring point.
Range: 00:05 to 59:59
min:sec.
IntervalValue input:
Time between two
Only for kinetic methods and "lin.regr." measurement methods.
Time intervals between the measuring points.
measuring points.
Range: 00:05 to 10:00
min:sec.
AutoprintSelection:
on | off
Printing a measuring result immediately following
measurement with the thermal printer.
Only the main result data will be printed. To output detailed
data, the required data packets can be compiled and printed in
the print & export method step at the end of a measuring
series.
Page 39
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
6.4Method procedure
Wizard: the wizard at the top of the display will take
you through the method procedure. The currently
active method step is highlighted.
A method procedure is composed of a maximum of 5 steps. The currently active step is highlighted visually.
After the last step, print & export, of a measuring series, the start of a new measuring series is offered as a
next step. It once again starts with the sample measurement.
39
Method stepExplanation
Check parametersCheck method parameters. Carry out changes if required.
Measure standardsOnly for methods with standard evaluation:
Measure and evaluate standards. Alternatively, the last saved standard
evaluation can be used.
Measure samplesMeasuring samples
Process resultsOnly for some methods: postprocess results, e.g. zoom scan graphs.
Print & exportAssemble data packets for printing or exporting the data.
Use the [Next >] and [< Back] softkeys to navigate between method steps. With [Abort] and [Finish] you can
cancel or finish the measuring procedure. The name of this softkey changes from [Abort] to [Finish] after
the first sample measurement.
Page 40
40
Methods
Eppendorf BioSpectrometer
®
English (EN)
6.4.1Check parameters
kinetic
Softkeys
• [Page dn] and [Page up]: Change between 1 to
3 parameter list pages.
• [Edit]: Switch to the parameter edit mode.
Editing mode for parameters:
Modified parameters are marked with a red star until
the modification has been saved.
Softkeys
• [Save] and [Save as]: Save changes. When using
[Save as] you have to rename the method. This is
always the case when modifying the methods
preprogrammed by Eppendorf in the Routine
group.
• [Cancel]: Exit the edit mode without saving
changes.
Saving the method under a new name:
You can save the method in the same folder from
which you called up the method or in any folder in
the Favorites method group.
The name (maximum 20 letters) can be entered
using a displayed keyboard ([Keyboard] softkey) or
directly using the keyboard (see Entering text on p. 21).
After saving you will return to the check parameters
display.
Page 41
Eppendorf BioSpectrometer
English (EN)
6.4.2Measure standards
The first standard to be measured is marked on the
display. After the blank value (blank key) measure all
standards (standard key) one by one.
When measuring more than one replicate per
standard, the average value for each standard is
calculated and displayed automatically.
By using the and cursor keys, you also can
select certain standards for measurement. Individual
standards can be remeasured as well.
Softkeys
• [Last cal]: Call up the last saved standard evaluation for this method in order to use it for sample
measurements.
• [Curve fit]: Select the procedure for the standard evaluation. If the result has not been saved, the
method can also be entered later. Instructions for selecting the evaluation procedure can be found in
the Evaluation procedure chapter (see Evaluation with standard curve/line on p. 91).
• [Graph]: Switch to the graphic display of the standard results.
As soon as the minimum number of results for the
evaluation with the selected method (curve fit) is
available, the evaluation result will be shown on the
right side of the display. You can now save the
evaluation and switch to sample measurements via
the [Next >] key.
Methods
®
kinetic
41
Graphical view of the standard evaluation.
With the and cursor keys, you navigate between
the standards to view the results. With more than one
replicate per standard, you can switch between the
replicate results using and . You can also select
individual standards from the graphical display and
measure or remeasure them.
Softkeys
• [Table]: Switch to the table display of the standard results.
• [Next >]: Save the standard evaluation and switch to the sample measurement.
Page 42
42
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
6.4.3Measure samples
The sample key is used for measuring your samples consecutively. Blank results remain saved for one
measuring series, but a new blank result measurement can be performed at any time. With the and
keys you can navigate between the sample results that have been achieved in the measuring series up to
this point.
Results display:
• The concentration result (6 digits with floating
point) is clearly emphasized.
• With graphic: Result to the right of the display.
• Without graphic: Result in the middle of the
display.
• In addition to the result, the basic absorbance
value is shown at a smaller scale.
Additional data
• Upper right; first row:
Sample number: Counted sequentially and reset to "1" for each new series of measurements.
Sample dilution (if provided)
• Upper right; second row:
Sample identification (ID) (if provided)
•Top left:
File name with which the data in the print and export method step can be exported as Excel file (see
p. 55).
Softkeys
• [Dilution]: Enter sample dilution.
• [Edit ID]: Enter sample ID
• [Data]: Display additional result data (not available with all methods).
• [Finish]: End series of measurements and return to method selection.
The displayed absorbance values always correspond to the directly measured values. The
dilution or cuvette factor as well as background absorbances will be incorporated for the
following result calculation (see Absorbance values on p. 89).
Page 43
Enter dilution
The [Dilution] softkey is activated after the blank
(blank key) has been measured.
1. Press the [Dilution] softkey.
2. Enter the volumes for the sample (up to 3 digits)
and for the dilution buffer (up to 4 digits).
The device will multiply the following sample results
by the calculated dilution factor.
Softkeys
• [Clear dil.]: Delete values for sample dilution.
• [OK]: Confirm sample dilution and return to sample measurement.
• [Cancel]: Cancel entry and return to sample measurement.
Eppendorf BioSpectrometer
English (EN)
Methods
®
kinetic
43
The dilution is used for all following sample results until it is changed by a new entry.
Enter sample ID
The ID will be applied to the following sample result. When an ID is being entered the last entered ID will
be displayed as a default template to allow the quick entry of IDs with a consecutive structure. A single ID
can only be assigned once for the same measuring series.
1. Press the [Edit ID] softkey.
2. Enter the sample ID (up to 12 digits).
Alternatives for character input:
• Keypad: If the key is pressed several times in a
row, the possible entries for this key will be
shown consecutively.
• Display keyboard with softkey [Keyboard]: Select
character with the cursor keys and confirm with
enter.
Softkeys
• [Keyboard]: Display keyboard.
• [abc]: Change between upper and lower case letters when making entries with the keypad.
• [OK]: Confirm ID entry and return to sample measurement.
• [Cancel]: Cancel entry and return to sample measurement.
Page 44
44
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Result image with dilution and ID
Result image with dilution and sample ID
Page 45
Eppendorf BioSpectrometer
English (EN)
6.4.4Measure samples: Results displays
This section contains a display of typical results displays for all method groups and an overview of
additional results data, which can be accessed using the [Data] softkey.
Method group Results displayExplanation
Absorbance main group
Single λResults display:
• Absorbance at the measuring
wavelength
• Only for dilution or with cuvettes other
than 10 mm: additional display of the
absorbance value before the
conversion.
Methods
®
kinetic
45
Single λ continuous
Prior to measurement:
• Check parameters method step:
Display temperature control. When the
temperature set in the parameters has
been reached and the display changes
from Temperi ng to Ready, you can go
to the measurement.
During the measurement:
• You can prematurely cancel the
measurement using the [Stop] softkey.
Results display:
• Graph with absorbance time display
and plotted regression line.
• Value for "A/min" calculated from the
linear regression.
Additional data ([Data] softkey):
• Absorbance time value pairs for the
first and last measuring points.
• Quality parameters for linear
regression
• Navigate between the measuring
points in the graph with and .
• If necessary, change the time frame
for evaluation with linear regression in
the process results method step.
Page 46
46
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Method group Results displayExplanation
Multi λResults display:
• Absorbance values at the wavelengths
Additional data ([Data] softkey):
• Only for dilution or with cuvettes other
than 10 mm: absorbance value before
the conversion.
ScanResults display:
• Scan (graph with absorbance
wavelength display)
• Navigate between the measuring
points in the graph with and .
Routine main group
Nucleic acidsResults display:
• Concentration result with absorbance
at the measuring wavelength
• If activated in the parameters: Ratios
A260/A280
• If activated in the parameters: Scan.
Navigate between the measuring
points on the graph which are used for
the result calculation with and .
Additional data ([Data] softkey):
If the corresponding parameters have
been activated:
• Absorbance value for 280 nm.
• Ratios A260/A230 and absorbance
value for 230 nm.
• Absorbance value for the background
wavelength.
Page 47
Eppendorf BioSpectrometer
Method group Results displayExplanation
Proteins
direct UV
Results display:
• Concentration result with absorbance
at the measuring wavelength
• If activated in the parameters: Scan.
Navigate between the measuring
points on the graph which are used for
the result calculation with and .
Additional data ([Data] softkey):
If the corresponding parameters have
been activated:
• Absorbance value for 260 nm.
• Absorbance value for the background
wavelength.
Proteins (with
reagent)
Results display:
• Concentration result with absorbance
at the measuring wavelength.
• For evaluation with standard series:
graph of the standard evaluation with
plotted sample result.
Methods
®
kinetic
English (EN)
47
Dye labelsResults display:
• Concentration results with absorbance
at the measuring wavelength of the
biomolecule.
• If activated in the parameters: Scan.
Navigate between the measuring
points in the graph with and .
Additional data ([Data] softkey):.
If the corresponding parameters have
been activated:
• Ratios A260/A280 and A260/A230.
• Absorbance values for 280 nm and
230 nm and for the measuring
wavelength of the dye.
•FOI value.
• Absorbance values for the background
wavelengths.
Ratios and FOI are not displayed for the
measurement of the dye-labeled proteins.
Page 48
48
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Method group Results displayExplanation
Bacterial
density
Results display:
• Calculated result with absorbance at
the measuring wavelength.
• If activated in the parameters: Scan.
Navigate between the measuring
points in the graph with and .
Basic main group
Factor,
standard
Analog to Protein direct UV (see above)Results display:
• Concentration result with absorbance
at the measuring wavelength.
Calibration
curve
Analog to Proteins (with reagent) (see above) Results display:
• Concentration result with absorbance
at the measuring wavelength.
• Graph of the standard evaluation with
plotted sample result.
Simple
kinetics
Prior to measurement:
• Check parameters method step:
Display temperature control. When the
temperature set in the parameters has
been reached and the display changes
from Temperi ng to Ready, you can go
to the measurement.
Results display:
• Concentration result with absorbance
or A/min with the measuring
wavelength.
• Only for the "linear regression"
measuring procedure: graph with
absorbance time display and plotted
regression line.
Navigate between the measuring
points in the graph with and .
Additional data ([Data] softkey):
Dependent on the measuring procedure:
• Absorbance time value pairs for the
first and last measuring points
("endpoint" method: only one
measuring point).
• Quality parameters for linear
regression (omitted for kinetics with a
slope below 0.050 E/min).
Page 49
Eppendorf BioSpectrometer
Method group Results displayExplanation
Advanced main group
Dual
wavelength
Results display:
• Concentration result: calculated from
A
with factor or standard
calc.
evaluation.
• A
: calculated using the formula,
calc.
defined in the parameters, created
from the absorbances measured on
both wavelengths.
• Absorbance values that were
measured at the two measuring
wavelengths.
Additional data ([Data] softkey):
If the corresponding parameters have
been activated:
• Absorbance value for the background
wavelength.
Advanced
kinetics
Analog to Simple kinetics (see above).In addition to the data similar to the
Simple kinetics method group:
Additional data ([Data] softkey):
If the corresponding parameter was
activated:
• Absorbance or A/min for the reagent
blank.
Methods
®
kinetic
English (EN)
49
Page 50
50
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
6.4.5Process results
The sample measurement is followed by two optional steps in the method sequence: process results and
print & export.
In the process results step, you can postprocess the results for some methods. Example: Changing the
spectra section of a scan.
As for the result display, you can navigate between the sample results of the measurement series with the
and cursor keys and select results for postprocessing.
Tab. 6-3: Options: Overview
OptionExplanationAvailable in methods
ZoomChange the axis limit of the absorbance
wavelength graph to limit the view to
enlarged sections of the graph.
More calculations Convert concentration results into molar
concentrations and (after entering the
volume) into total amounts.
Peak detectionIdentifying peaks in absorbance wavelength
spectra.
Linear regressionModifying time windows for the evaluation of
a kinetic via linear regression.
Generally, all methods for which the
Scan parameter is offered and has
been activated.
• Multi λ
• Scan
• Nucleic acids
• Proteins direct UV
• Dye labels
• Bacterial density
• Dual wavelength
• Nucleic acids
• Dye labels (with nucleic acids as
biomolecule)
• Scan
In principle, all kinetics methods for
which the "Linear regression"
measuring procedure was used.
• Single λ - continuous
• Simple kinetics
• Advanced kinetics
Options for postprocessing are presented on the two
softkeys at left. In this example: [Zoom] and [More
Calculations].
Page 51
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
After changes have been made, you can exit the
current mode using the two softkeys at right:
• [Save]: Save changes and return to the process results method step.
• [Cancel]: Cancel and return to the process results
method step.
After the changes have been saved you can apply
them to all samples of the measuring series with
[Yes].
51
6.4.6Process results: Options
Zoom
Press the [Zoom] softkey and select one of the following versions.
Variant [spectra]:
• Cursor keys and : Move the wavelength
cursor. It determines the zoom center above the
x-axis.
• Cursor keys and : Gradually zoom in and out
of the displayed section of the x-axis using the
SpectraZoom procedure.
The shown section of the y-axis is automatically
adjusted with each step so that maximum and
minimum of the data to be displayed make
optimum use of the section.
Page 52
52
Methods
Eppendorf BioSpectrometer
English (EN)
®
kinetic
Variant [spectra-0]:
Same as the [spectra] variant, with one exception:
The lower limit of the displayed section of the y-axis
always equals "0 A".
Variant [free]:
User-defined values for interval limits can be entered
for both axes.
Navigation between the entry fields by means of the
cursor keys ( , , ).
For all 3 versions, the [reset zoom] softkey will bring you back to the original display of the spectrum.
Page 53
More calculations
Press the [More calc.] softkey.
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Nucleic acids method group:
• After the molar mass has been entered (in base/
base pairs or in kDa): Convert the concentration
result to the molar concentration.
• After the sample volume has been entered:
Calculate the total amount in the sample.
53
Dye labels method group:
Nucleic acid:
• After the molar mass has been entered (in base/
base pairs or in kDa): Convert the concentration
result to the molar concentration.
• After the sample volume has been entered:
Calculate the total amount in the sample.
Dye:
• After entering the volume of the sample:
Calculate the total amount in the sample.
•For dsDNA the calculation of the molar concentration is based on the assumption of a
double-stranded nucleic acid. For the ssDNA, RNA and Oligo methods, a single-stranded
nucleic acid is assumed.
• For methods which have been reprogrammed via <New Method> in the Routine main
group, Nucleic acids method group, always double-stranded nucleic acids are assumed
for calculating the molar concentration.
Peak detection
Press the [Peaks] softkey. For the peak detection you can alternate between two criteria:
• λ grid: Evaluation grid on the wavelength scale for the peak detection (e.g., 10 nm).
10 nm example: The spectra section from -5 nm to +5 nm is evaluated in relation to the peak to be
detected.
• Min. Δ Abs: Minimum difference between the peak to be detected and the lowest absorbance in the
evaluation grid. No absorbance value in the grid may be higher than the value of the peak at any given
time (e.g.: 0.5).
Page 54
54
Methods
Eppendorf BioSpectrometer
English (EN)
Examples:
®
kinetic
λ grid: 100 nm, min. Δ Abs: 0.050:
The peak is not detected because the λ grid is too
large: The absorbance values on the left edge of the
grid are higher than the absorbance of the peak.
λ grid: 20 nm, min. Δ Abs: 0.200:
The peak is not detected because the predetermined
value for min. Δ abs is too high. The difference of the
absorbance of the peak and the lowest absorbance in
the grid is less than 0.2 A.
λ grid: 20 nm, min. Δ Abs: 0.050:
The peak is detected.
Page 55
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Linear regression
For kinetic methods evaluated using the "Linear regression" measurement method, you can subsequently
define the start and end times for the regression evaluation in the process results method step.
Selecting measuring points
• [First pt.] softkey: Define the first measuring
point for the regression evaluation. Select the
measuring point using and . Confirm with
enter.
• [Last pt.] softkey: Define the last measuring point
for the regression evaluation. Select the
measuring point using and . Confirm with
enter.
• At the same time that the times are modified, the
result is recalculated. The reagent blank is also
recalculated in the newly defined time window for
this purpose.
55
6.4.7Print & export
In the last optional method step, you can assemble data packets for all samples of a series of
measurements, or selected samples of a series of measurements, for printing to the printer, export to a USB
stick or export to a PC using a USB cable.
Select data packets
• Use the cursor keys for navigating and confirm
with enter.
Softkeys
• [Print]: Start printing.
• [Export]: Start export.
• [Sample]: Select individual sample results.
Select samples
• Press the [Samples] softkey to call up the sample
selection.
• Use the cursor keys for navigating and confirm
with enter.
Softkeys
• [Select all]: Select all samples
• [De-Sel. all]: Cancel selection.
Page 56
56
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Data export
The data will be transferred as Excel files (.xls) and can be read with Excel versions Excel 97 and later. For
each of the selected data packets, a worksheet is created in Excel. The file name consists of the method
name, the time and the date of the measuring series.
Select export version
If no USB stick is connected, the first variant cannot
be selected.
Export to USB stick
1. Connect a FAT 32-formatted USB stick to the USB port 4 (see Main illustration on p. 9).
2. Start with [Export] "export to an external storage medium".
Export to PC
Requirement for the PC operating system: Windows XP, SP2 or a higher version.
1. Connect the device to the PC by using the USB cable on the USB port 8 (see Main illustration on p. 9).
2. Prior to beginning a new export make sure that any data that has been exported previously has been
saved to the PC hard drive. Otherwise, the new export will overwrite the data.
3. Start with [Export] "export to PC".
4. The exported data packet will be displayed on your PC as a removable medium named "eppendorf".
Open the Excel file on this drive and save it to the hard drive.
Select data packets
ResultsPrimary result data; cannot be selected because they are always transferred.
DataAdditional results data that are displayed during the measurement using the
[Data] softkey.
GraphAbsorbance-wavelength-spectrum.
(Only for kinetic methods with the "linear regression" measuring procedure:
absorbance time graph.)
Graph dataThe basic numeric data for the graph.
"export only": Only available for export, i.e., not for printing.
ParametersMethod parameters
Standards/resultsResults data of the standard evaluation.
Standards/graph(Only for standard evaluations with several standards:) absorbance concentration
graph.
Based on the method and parameter setting, only the available data packets are presented.
Page 57
Methods
®
Eppendorf BioSpectrometer
kinetic
English (EN)
6.4.8Finish the series of measurements
After the print & export method step has been finished, you can start a new series of measurements using
the selected method or select a new method.
Finish the series of measurements and start a new series of measurements
• [Next >] softkey: Call up new series method step
• [New] softkey: Call up measure samples method
step and start a new series of measurements.
57
Finish the series of measurements and select a new method
• [Finish] softkey: Close the series of measurements and call up the method selection.
Page 58
58
Methods
Eppendorf BioSpectrometer
English (EN)
®
kinetic
Page 59
Functions
®
Eppendorf BioSpectrometer
kinetic
English (EN)
7Functions
7.1Functions of the User main group
With the function key or the [Function] softkey, you reach a menu containing functions like device settings
or calling up saved results.
The functions are structured in 3 columns analog to the method selection. The functions in the User main
group are accessible to you. As for the method selection, you navigate with the cursor keys for selecting
first the desired subgroup and then the desired function in the right column. Press enter to open the
function.
59
[Info] softkey: call up the firmware version and serial number of the BioSpectrometer kinetic.
Tab. 7-1: Overview of the functions
SubgroupExplanation
Results memoryDisplays saved results.
The results can be accessed structured according to methods and measuring
series and can be printed and exported directly from the memory.
Device calibration• Option for checking the spectrophotometer; An Eppendorf filter set is
InfoOpen Source Licenses and information on registered trademarks.
Parameters which are used for different methods in common are stored
centrally in the Functions area.
Here it is possible to edit (change or create anew) these parameters. In the
Check parameters method step, the comprehensive parameters can be easily
selected using drop-down menus.
• Proteins, nucleic acids, dyes include parameters that are directly applied
for methods of the Dye labels and Proteins direct UV groups.
• Units: Units for concentration results which can be used for many methods.
Absorbance wavelength spectra of important substances, e.g., DNA.
The spectra serve as information and can be used for comparison to a
spectrum of a sample result.
required for this.
• Option for checking the thermal module.
Page 60
60
Functions
Eppendorf BioSpectrometer
English (EN)
7.1.1Results memory
®
kinetic
In the right column, select the method for which
you would like to call up saved results.
Confirm with enter.
Select the desired series of measurement with
the cursor keys.
Confirm with enter.
As in the method procedure, you can also
successively switch between the display of the
parameters, standards, sample results and, finally,
the data packets for print and export.
The assignment of the softkeys matches the
assignment in the method procedure.
If you would like to print or export results,
select the data packets.
The procedure for print and export and the
meaning of the function keys corresponds to the
print & export method step.
Page 61
7.1.2General method parameters
Functions
®
Eppendorf BioSpectrometer
kinetic
English (EN)
In the right column, select the parameter group
for which you would like to edit parameters.
Confirm with enter.
In this example, parameter groups are summarized
for various dyes (dye components for the dye
methods) and stored under a name. With this
name, the required parameter group can be
imported into the method program during the
editing of a dye method.
Display:
• Left: Name of the dye. Select via and .
• Right: corresponding parameters
61
Softkeys
• [Edit]: Edit selected parameter group.
• [New]: Create new parameter group.
• [Delete]: Delete selected parameter group.
• [OK]: Return to the function selection.
To edit a parameter group, use and to
select the parameter which you would like to
edit.
Confirm with enter.
Softkeys
• [OK]: Save entry and return to the parameter group selection.
• [Cancel]: Return to the parameter group selection without making any changes.
Page 62
62
Functions
®
Eppendorf BioSpectrometer
kinetic
English (EN)
When programming a method of the Dye labels or Proteins direct UV method groups, you can access the
entries in General Method Parameter:
Select the name of the dye to import the
corresponding parameter group into the method
program. By using the "edit" selection of the
"Nucleic acid" parameter, you also can get directly
to the General Method Parameter function and
view and edit the parameters.
Tab. 7-2: Parameter in General Method Parameter
ParameterExplanation
ProteinsThese parameters are loaded into the method parameters when a protein is
selected during the programming of a method in the Dye labels and Proteins
direct UV groups.
• Protein name
•Factor
•A
0.1%
In order to define a factor for calculating the concentration on the basis of the
absorbance, enter the following data in addition to the name and wavelength:
Factor or A
or absorbance coefficient and molar mass.
0.1%
•Ext.coeff.
•Molecular mass
Nucleic acidsThese parameters are loaded into the method parameters when a nucleic acid
is selected during the programming of a method in the Dye labels group.
•NA name
•Factor
• Double-stranded
The factor is used to calculate the concentration on the basis of the
absorbance.
The double-stranded parameter affects the calculation of the molar nucleic
acid concentration. (see Conversion to molar concentrations and nucleic acid
quantities on p. 94)
DyesThese parameters are loaded into the method parameters when a dye is
selected during the programming of a method in the Dye labels group.
•Dye name
• Wavelength
•Ext.coeff.
•Factor
• Corr. A260
• Corr. A280
In order to define a factor for calculating the concentration on the basis of the
absorbance, enter the following data in addition to the name:
Factor or absorbance coefficient.
The correction factors for the absorbance values at 260 or 280 nm are used
when the correction function in the method parameters is active. For more
details, refer to the chapter on evaluation (see Correction A
A
on p. 93).
280
and correction
260
UnitsYou can select a unit from all available units when programming method
parameters.
• UnitEntering a unit that has not yet been programmed for the concentration result.
Page 63
• Specifications for proteins which are not preset at the factory can be determined in the
expasy database: http://www.expasy.org/tools/protparam.html.
• A table with A
values for many proteins can also be found in: C.N.Pace et al., Protein
1%
Science (1995), 4: 2411–2423 (Table 5). The A
return the required A
0.1%
7.1.3Absorbance spectra library
values.
Functions
®
Eppendorf BioSpectrometer
kinetic
English (EN)
values must be multiplied by 0.1 to
1%
In the right column, you select the spectrum which
you would like to call up and confirm with enter.
63
7.1.4Device settings
Softkeys
• [Export] and [Print]: Export to a USB stick or
print to a PC using a USB cable (see Print & export on p. 55).
• Frequency of the automatic self test of the
device after switching on.
Softkeys
• [Save]: Save changes and return to the function
selection.
• [Cancel]: Return to the parameter group
selection without making any changes.
Page 64
64
Functions
Eppendorf BioSpectrometer
English (EN)
®
kinetic
7.1.5Device calibration
Information on checking the device is provided separately (see Checking the device on p. 66).
7.1.6Info
The Copyright menu item contains license
information on the Open Source software.
Page 65
8Maintenance
8.1Cleaning
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
NOTICE! Corrosion from aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive polishes.
Do not incubate the accessories in aggressive cleaning agents or disinfectants for a longer
period of time.
Maintenance
Eppendorf BioSpectrometer
English (EN)
®
kinetic
65
1. Wipe down the surfaces with a cloth moistened with a mild cleaning agent.
Cleaning the cuvette shaft
2. The cuvette shaft may only be cleaned with a lint-free cotton swab that has been dampened with ethanol
or isopropanol. Prevent liquid from entering the cuvette shaft. If the shaft needed to be dampened with
water to remove contamination, follow this up by cleaning the shaft with a cotton swab dampened with
ethanol or isopropanol to accelerate the drying process.
8.1.1Cleaning the cuvette shaft cover
If you would like not only to clean the directly accessible surface of the cuvette shaft cover, you can remove
the cover.
Do not soak the cuvette shaft cover in cleaning agent.
Clean the cuvette shaft cover as described.
1. Lift the cuvette shaft cover with one hand.
2. With the other hand, hold the cover at the height
of the locking pin and pull the cover to the right
until the locking pin has been removed.
• Pull the cover to the right at a 90 degree angle.
Page 66
66
Maintenance
Eppendorf BioSpectrometer
English (EN)
3. Clean the cover with a cloth or lint-free cotton swab dampened with a mild cleaning agent.
4. Slide the locking pin back into the housing as far as it will go.
The locking pin has completely disappeared in the housing.
When the photometer is not being used, close the cuvette shaft using the blue cuvette shaft
cover to protect it from dust and other contamination.
®
kinetic
8.2Disinfection/Decontamination
DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.
Do not spray clean/spray disinfect the housing.
Only plug the device back in if it is completely dry, both inside and outside.
1. Clean the device with a mild cleaning agent before the disinfection (see Cleaning on p. 65).
2. Choose a disinfection method which corresponds to the legal regulations and guidelines in place for
your range of application.
3. For example use alcohol (ethanol, isopropanol) or other alcoholic disinfectants.
4. Wipe the surfaces with a cloth which you have moisturized with a disinfectant.
5. If the cuvette shaft cover needs to be removed for the disinfection, proceed as follows for the
disassembly and assembly (see Cleaning the cuvette shaft cover on p. 65).
6. You can use spray disinfection to disinfect the disassembled cuvette shaft cover.
8.3Checking the device
Requirements
• Observe the ambient conditions (see Ambient conditions on p. 85).
• Carry out an inspection at approx. 20 °C. Avoid temperature variations (e.g, due to opened windows).
• The filter may only be removed from the filter box for a short period of time, and the filter surfaces must
be protected from contamination or damage.
• Protect the filter from dust, heat, liquids and aggressive vapors.
• The filter is inserted so that the label, with the name of the filter, points toward the detector.
When the spectrometer unit is checked: label points forward.
• The cuvette shaft is free from contamination.
Page 67
Maintenance
®
Eppendorf BioSpectrometer
kinetic
English (EN)
8.3.1Checking the spectrometer unit
Eppendorf offers a filter kit (BioSpectrometer reference filter kit) for checking the photometric accuracy
and wavelength systematic error. The kit contains one blank filter (A0) and three filters (A1, A2 and A3) for
checking the photometric accuracy, and 3 filters for checking the wavelength systematic error in the range
of 260 nm to 800 nm. The filter absorbances are measured against blank filter A0. In addition to
information on accuracy, you also obtain information on precision: average value and variation coefficient
(CV value) are calculated from the sets of 15 measurements per wavelength.
For the measurement, place the blank filters (for blank measurement) and sample filters into the cuvette
shaft, using the same procedure as with cuvettes. The absorbance values measured for the test filters will
be compared to the permitted range of values. For the individual filters, the limit values for the permitted
area are printed in a table in the lid of the filter box.
Please connect the Eppendorf thermal printer if you would like to document the values.
67
Page 68
68
Maintenance
Eppendorf BioSpectrometer
English (EN)
Abb. 8-1: Inside of the filter box lid (sample)
®
kinetic
Fig. 8-1:Inside of the filter box lid (sample)
Page 69
8.3.1.1 Checking photometric accuracy
Maintenance
®
Eppendorf BioSpectrometer
kinetic
English (EN)
1. Select the Spectrometer unit function in the
Device calibration group and confirm with enter.
2. Select whether you want to check the wavelength
systematic error or photometric accuracy, and
confirm with enter. Press [Next >] to switch to
the measurement.
69
3. Follow the instructions on the device display and
start by measuring the A0 blank filter, and then
the first test filter A1.
The device measures the test filter 15 times at
9 wavelengths, and then shows the average value
and the VC values for the series of measurements
for all 9 wavelengths.
Page 70
70
Maintenance
Eppendorf BioSpectrometer
English (EN)
®
kinetic
4. Results display after measuring a test filter for
testing photometric accuracy.
Measure the other two test filters A2 and A3.
5. Results display after measuring all 3 test filters
for testing the photometric accuracy.
With the and keys, you can view the results
for the different test filters again. Press [Finish] to
complete the test.
6. Compare the average values and CV values to the
supplied table.
If the measured values do not agree with the
permitted range of values, contact Eppendorf
Service.
Have the filter recertified by Eppendorf Service after 2 years.
8.3.1.2 Checking wavelength systematic error
To check the wavelength systematic error, proceed as follows: Measure the 3 test filters at the
corresponding wavelength.
Page 71
8.3.2Checking the thermal module
• Carry out an inspection at approx. 20 °C.
• Make sure that the cuvette shaft is empty for the check.
Maintenance
Eppendorf BioSpectrometer
®
English (EN)
1. Select the Temperature unit function in the Device calibration group and confirm with enter.
kinetic
71
2. Make sure the cuvette holder is empty, and start
the test procedure with [Next >].
The following temperature control test at
5 temperatures takes approx. 30 minutes. The
estimated remaining time is displayed during the
test.
3. The results are displayed after the test has been
completed.
A positive test result is indicated by "OK" behind
the temperature value. If the result "failed" is
displayed for at least one temperature please
contact Eppendorf Service.
Page 72
72
1
2
3
Maintenance
Eppendorf BioSpectrometer
English (EN)
®
kinetic
8.3.3Device self-test
The frequency of the automatic self-test (approx. 1 minute) can be set using the Device settings function
(see Device settings on p. 63). The factory setting for Self-test interval is "Weekly".
The self-test checks the following:
• Verification of the detector
– Determination of the random error across the entire available spectrum
• Verification of the light source
– Verification of the maximum available energy of the light source and quality of the light transmission
through the device
– Determination of the signal random error at the reference sensor
– Determination of the signal level at the reference sensor
– Separate determination of light intensity in UV range
• Determination of the systematic and random error of the wavelength
– Position of an intensity peak in the UV range of the spectrum
– Precision of the position of an intensity peak in the UV range of the spectrum
• Random error of the temperature sensor
• Verification of the temperature control rate
Select the Perform self-test function in the Device calibration group and confirm with enter.
At the end of the self-test, the display shows the message PASSED.
If the display shows the message FAILED, the self-test has failed. If this error cannot be corrected (see
Error messages on p. 76), contact Eppendorf Service.
8.4Replacing fuses
DANGER! Electric shock.
Switch off the device and disconnect the power plug before starting maintenance or
cleaning work.
The fuse holder is located between the mains connection socket and the mains power switch.
1. Disconnect the power plug.
2. Press the upper and lower end of the plastic springs 1 together and pull the fuse holder 2 fully out.
3. Replace faulty fuses and reinsert the fuse holder. Make sure that the guiding rail 3 is positioned
correctly.
Page 73
Maintenance
®
Eppendorf BioSpectrometer
kinetic
English (EN)
8.5Decontamination before shipment
If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer
for disposal please note the following:
WARNING! Risk to health from contaminated device
1. Follow the instructions in the decontamination certificate. You find it as a PDF file on our
website (www.eppendorf.com/decontamination
2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.
).
73
Page 74
74
Maintenance
Eppendorf BioSpectrometer
English (EN)
®
kinetic
Page 75
9Troubleshooting
9.1General errors
ErrorPossible causeRemedy
Measuring results are
imprecise.
• Reagent is past its shelf
life.
• Reagent has not been
prepared properly.
• Pipetting is not correct.
• Incubation procedure
before measurement is
incorrect.
• The cuvette is
contaminated.
• The cuvette is not filled
completely with
measuring solution, and it
contains bubbles.
• Turbidity of the
measuring solution.
• Spectrophotometer is
drifting.
• Cuvette shaft is dirty.
Ensure that the reagent is still within its
shelf life and properly prepared.
Use clean demineralized water of adequate
quality for preparation if required.
Ensure that the pipette is calibrated and
that pipetting is being performed correctly.
If the method procedure requires
incubation before the measurement,
ensure that the temperature and time for
incubation are correctly observed.
Clean and rinse the cuvette. When
replacing a cuvette, pay attention that the
optical window of the cuvette remains
clean and that you do not touch it with your
fingers.
If the cuvette window has become soiled
from fingerprints, wipe it clean using a
lint-free lab cloth soaked in ethanol or
isopropanol.
Ensure that the required minimum volume
of the cuvette for a measurement is
reached and that no bubbles are in the
measuring solution.
Centrifuge the turbid measuring solutions
containing particles and use the clear
supernatant.
Contact Eppendorf Service.
Observe the ambient conditions.
Prevent temperature changes.
Clean the cuvette shaft.
Troubleshooting
Eppendorf BioSpectrometer
English (EN)
®
kinetic
75
Page 76
76
Trou bleshooting
Eppendorf BioSpectrometer
English (EN)
ErrorPossible causeRemedy
The measuring results are
not correct.
®
kinetic
• The method has not been
programmed correctly.
• The standard solution has
not been prepared
correctly.
• The absorbance of the
reagent is drifting.
• The cuvette is not
positioned correctly.
Ensure that the method parameters are
entered correctly.
Ensure that the correct standard is used
and that the measuring solution for the
standard is prepared correctly.
For instable reagent absorbance and end
point methods: When measuring a long
series of samples, measure the reagent
blank value not only at the beginning but
also during the sample series. If the
reagent blank value drifts strongly, the
reagent is not appropriate for error-free
measurements and has to be replaced by a
new reagent.
Position the cuvette in the cuvette shaft so
that the optical window points towards the
direction of the light path.
Photometry light path: from back to front
9.2Error messages
You can exit device displays with error messages using the [OK] softkey.
System errors require an evaluation by the Technical Service. These errors are shown in English (System error …). Please contact Technical Service in these cases. Other error messages, for which you can carry
out troubleshooting measures, are illustrated in the table below.
ProblemCauseSolution
Self test failed.• Cuvette shaft cover was open during
self test.
• The cuvette shaft was not empty
during the self test.
• Device is faulty.
File export failed.During data export:
• USB stick improperly formatted or
faulty.
• USB stick removed from the device
too early (during the export).
Failed to initialize
printer.
• Printer not connected or switched off.
• Printer not configured correctly.
Repeat the self test with empty
cuvette shaft and the cuvette shaft
cover closed.
Contact Eppendorf Service.
Reformat or replace the USB stick.
Reconnect the USB stick and repeat
the export.
Connect the printer and switch it on.
Reconfigure the printer.
For a correct configuration of the printer
settings refer to the installation
description (see Connecting the printer on p. 15).
Page 77
Eppendorf BioSpectrometer
ProblemCauseSolution
Blank
measurement: An
intensity on a pixel
that influences the
main, auxiliary or
scan wavelength is
too low.
The entered name
is not valid.
A method (or
folder, dye, protein,
nucleic acid, or
unit) with this name
already exists.
The following
parameter values
are not defined in
General Method
Parameter:
The value of the
parameter marked
with * is not defined
in the Gen. Param.
Please correct the
parameter.
Invalid zoom
interval.
The entered
standard
concentrations are
not monotonically
increasing resp.
monotonically
falling. Correct the
standard
concentrations.
• The absorbance of the blank solution
used for the blank measurement is too
high.
• Incorrect or turbid blank solution.
• For scans: Wavelength range is too
large, because the sample is very
strongly absorbed in part of the
wavelength range.
• Error when entering the name.
Different causes are possible. For the
precise cause please see the
information in the help box.
• The name under which the method
was saved has already been used for a
different method in the same folder.
• The message also appears after
editing names already given to a
folder or to a nucleic acid (dye,
protein, concentration unit) (under
General Method Parameter).
• When opening a method with
parameters which access General Method Parameter, the system
determined that at least one
parameter (dye, nucleic acid, protein,
unit) does not exist there anymore, so
probably has been deleted.
This error message appears when editing
method parameters.
• Parameter in General Method Parameter is not defined.
During the Zoom process with free entry
of limits ([Free]) softkey):
• The zoom area is below the lower
limit.
• See the error text.
Check the blank solution and
remeasure the blank if required.
For scans: Match the wavelength
range to the sample spectrum.
See information in the help box.
Assign a different name.
Select a different parameter from the
existing list. If necessary, program a
new list entry in General Method Parameter in order to be able to use it
when programming a method.
Select a different parameter from the
existing list. If necessary, program a
new list entry in General Method Parameter in order to be able to use it
when programming a method.
Enter the values so that the interval
does not fall below the range limits of
0.02 A and 10 nm.
Enter the standard concentrations so
that the first standard receives the
lowest concentration and the other
standard concentrations form an
increasing sequence.
Troubleshooting
®
kinetic
English (EN)
77
Page 78
78
Trou bleshooting
Eppendorf BioSpectrometer
English (EN)
ProblemCauseSolution
At least two of the
entered standard
concentrations are
identical. Correct
the standard
concentrations.
The measured
values are not
strictly
monotonous!
The ID cannot be
set.
The dilution cannot
be set.
Calculation not
possible because of
division by zero.
Absorbance result
or Formula "b"
parameter is zero.
There is only one
measurement left to
be performed in
this series of
measurement.
The maximum
number of
measurements
within one series of
measurements has
been reached.
®
kinetic
• See the error text.
• Error when measuring a standard
series: The measured absorbance
values of the standard series are not
continuously increasing or
decreasing.
• Error when entering the sample ID.
Different causes are possible. For the
precise cause please see the
information in the help box.
• Error when entering the dilution.
Different causes are possible. For the
precise cause please see the
information in the help box.
• An absorbance result was divided by
a "zero" value during the evaluation
of a Division type method (Dual wavelength method group). This is
not mathematically permissible.
• The number of measurements in one
measuring series is limited to 99.
Enter the standard concentrations so
that the first standard receives the
lowest concentration and the other
standard concentrations form an
increasing sequence.
Repeat the standard measurements or
delete the single, incorrectly
measured standard result.
See information in the help box.
See information in the help box.
Check the reagents and samples used
and repeat the measurement.
Do not enter "zero" as a value for the
Formula b parameter.
Start a new series of measurement
after maximally 99 measurements.
Page 79
Eppendorf BioSpectrometer
ProblemCauseSolution
Invalid zoom
interval!
Device
configuration was
changed from …
to … .
The temperature
control system is
faulty. Please
program the
method without
temperature control
or cancel the
method.
The ambient
temperature is too
high.
Linear regression
could not be
applied to all
measurements.
Error in the process results method step
in the Zoom mode.
Permissible zoom range for the
wavelength scale:
• Wavelength interval at least 10 nm
• Entries for wavelengths only within
the range programmed in the
parameters for the method.
Permissible zoom range for the
absorbance scale:
• Absorbance interval at least 0.02 A
• Upper and lower limit for absorbance
interval +3 A or –3 A
• A BioSpectrometer kinetics is not
detected as the kinetic variant but as a
BioSpectrometer basic. The kinetic
methods are therefore not displayed.
• The temperature control of the device
is faulty.
For kinetic methods with temperature
control:
• The ambient temperature measured
by the device is above the specified
range.
• For kinetic methods, the time frame
for evaluation with linear regression
was changed in the process results
method step, and the change should
be extended to all measuring results.
The required number of measuring
points, however, was not available for
at least one sample result.
Please observe the stated limits in the
zoom procedure.
Switch the device off and back on
again. If the error occurs again:
Contact Technical Service.
Contact Eppendorf Service. Until the
temperature control has been
repaired only use methods for which
no temperature control has been
programmed.
Make sure that the ambient
temperature is within the range
specified for the operation of the
device.
Only change the time frame for
evaluation with linear regression in
the process results method step for
samples with sufficient measuring
points.
Troubleshooting
®
kinetic
English (EN)
79
Page 80
80
Trou bleshooting
Eppendorf BioSpectrometer
English (EN)
®
kinetic
9.3Result flags
Warnings and error messages for results are displayed in the bottom right of the help box. The header bar
of the Help box is highlighted yellow for warnings and red for error messages.
Warnings: Decide whether the result is useful for you while taking the displayed warning into
consideration.
Error messages: No result is displayed; the reason is shown in the error message.
ProblemCauseSolution
The standard curve
is not monotone.
Please select
another Curve Fit.
Some absorbance
values for
secondary
wavelengths are too
high or are not
displayed.
The result is
outside the range of
the standard
concentrations.
The coefficient of
determination is
<0.8.
• No usable result was returned during the
evaluation of a standard curve using the
"spline interpolation", "quadratic
regression" or "cubic regression" Curve Fit procedures.
• For at least one secondary wavelength,
the absorbance exceeded the measuring
range.
• Secondary wavelengths are not needed
for calculating the concentration result.
They are used for different purposes. For
example, dsDNA method: absorbance at
280 nm for the calculation of ratios 260/
280.
• Turbidity of the measuring solution
• Measurements at the limits of the
photometric measuring range.
• For methods with evaluation via standard
curves (nonlinear evaluation method):
The sample result is up to 5 % outside of
the standard concentration range.
• For methods with evaluation of standard
series via the regression procedure: The
coefficient of determination for the
regression evaluation indicates a
significant deviation of the measuring
points from the regression line.
• Turbidity of the measuring solution.
• Measurements at the limits of the
photometric measuring range.
Select a different Curve Fit
procedure.
If the absorbance values of the
secondary wavelengths are
relevant: Dilute the sample or
remove the turbidity via
centrifugation and repeat the
measurement.
Accept the measurement result, or
remeasure the sample under
conditions under which the result
is within the range of the standard
concentrations (dilute sample or
modify standard concentrations
and remeasure).
Accept the result of the standard
evaluation or remeasure the
standards.
Make sure the measuring
solutions are clear.
Page 81
Eppendorf BioSpectrometer
ProblemCauseSolution
The coefficient of
determination for
the regression
evaluation of the
standard series is
< 0.8.
Scan: Some of the
measured
absorbances are
too high and are
not displayed.
The measurement
is not complete.
The kinetics are
nonlinear: The
coefficient of
determination is
< 0.95.
The kinetics are
nonlinear for at
least one standard:
The coefficient of
determination is
<0.95.
• For methods with evaluation of standard
series via the regression procedure: If the
regression evaluation for the standard
series was nonlinear, but the standard
evaluation was accepted by the user, a
warning appears after samples have been
measured.
• For at least one scan wavelength, the
absorbance exceeded the measuring
range.
• Turbidity of the measuring solution.
• Measurements at the limits of the
photometric measuring range.
• Kinetic procedure: You have prematurely
canceled the measurement using the
[Stop] softkey. If at least 2 measuring
points are available, a result is calculated
and displayed.
• Kinetic procedure with "linear
regression" measuring procedure: The
coefficient of determination for the
regression evaluation indicates a
significant deviation of the measuring
points from the regression line.
• Turbidity of the measuring solution.
• Measurements at the limits of the
photometric measuring range.
• Activity concentration of the enzyme too
high.
• Kinetic procedure with "linear
regression" measuring procedure and
evaluation procedure via standards: The
coefficient of determination for the
regression evaluation of at least one
standard measurement indicates a
significant deviation of the measuring
points from the regression line.
• Turbidity of the measuring solution.
• Measurements at the limits of the
photometric measuring range.
• Activity concentration of the enzyme too
high.
Use the sample results with the
reservation mentioned or repeat
the measurement of the standard
series and samples.
If the non-displayed areas of the
scan are relevant: Dilute the
sample or remove the dilution via
centrifugation and repeat the
measurement.
Accept the measuring result with
the reduced measuring time or
measure again with a longer
measuring time.
Accept measuring result or
remeasure the sample. Before
repeating the measurement,
evaluate the reason for the
non-linearity and act accordingly
(e.g., clear the measuring solution
by centrifugation or dilute the
sample).
Accept measuring result or
remeasure the standard. Before
repeating the measurement,
evaluate the reason for the
non-linearity and act accordingly
(e.g., clear the measuring solution
via centrifugation or use the
standard with low concentration).
Troubleshooting
®
kinetic
English (EN)
81
Page 82
82
Trou bleshooting
Eppendorf BioSpectrometer
English (EN)
ProblemCauseSolution
During the kinetics
measurement, the
temperature was
outside of the
allowable range.
Absorbance at the
measuring
wavelength is too
high.
The calculated
result is negative.
At least one of the
results is negative.
The result has more
than 6 pre-decimal
places.
The result is more
than 5 % outside of
the standard
concentration
range.
• Calculation not
possible
because of
division by zero.
Absorbance
result is zero.
•Calculation
error. Division
by zero.
Calculation not
possible because of
division by zero.
Absorbance result
or parameter
formula b is zero.
®
kinetic
• Ambient temperature outside the
specified range.
• The temperature control is faulty.
• Turbidity of the measuring solution.
• Optical surfaces of the cuvette are soiled.
• Cuvette has been inserted into the
cuvette shaft facing the wrong direction.
• Too high absorbance of measuring
solution.
• Measuring solution not prepared
correctly.
• The incorrect factor has been entered
(wrong algebraic sign).
• For methods with several results (e.g.,
Dye labels).
• Measuring solution not prepared
correctly.
• The incorrect factor has been entered
(wrong algebraic sign).
•Very high sample concentration.
• Concentration unit does not match the
expected range of the sample
concentrations.
• For methods with evaluation via standard
curves (nonlinear evaluation method):
The sample result is more than 5 %
outside of the standard concentration
range.
• The evaluation required dividing by an
absorbance result with the value of
"zero". This is not mathematically
permissible.
Examples: Calculation of a factor at
one-point calibration; calculation of a
260/280 ratio with nucleic acid
measurements.
• An absorbance result was divided by a
"zero" value during the evaluation of a
Division type method (Dual wavelength
method group). This is not
mathematically permissible.
Measure the sample at ambient
temperature within the specified
range (15°C to 35°C). If the
warning nevertheless appears,
contact Eppendorf Service.
Measure again considering the
possible causes.
Measure again considering the
possible causes.
Measure again considering the
possible causes.
Dilute sample and measure again.
Change the concentration unit
(Parameter Unit) and measure
again.
Remeasure the sample under
conditions under which the result
is within the range of the standard
concentrations (dilute sample,
modify standard concentrations
and remeasure).
Check the reagents and samples
used and repeat the measurement.
Check the reagents and samples
used and repeat the measurement.
Do not enter "zero" as a value for
the Formula b parameter.
Page 83
Transport, storage and disposal
Eppendorf BioSpectrometer
English (EN)
10Transport, storage and disposal
10.1Transport
Use the original packaging for transport.
Air temperatureRelative humidityAtmospheric pressure
General transport-25 °C – 60 °C10 % – 95 %30 kPa – 106 kPa
Air freight-40 °C – 55 °C10 % – 95 %30 kPa – 106 kPa
10.2Storage
Air temperatureRelative humidityAtmospheric pressure
In transport packaging-25 °C – 55 °C25 % – 75 %70 kPa – 106 kPa
Without transport packaging-5 °C – 45 °C25 % – 75 %70 kPa – 106 kPa
®
kinetic
83
10.3Disposal
In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European Community:
Within the European Community, the disposal of electrical devices is regulated by national regulations
based on EU Directive 2012/19/EU pertaining to waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005, in the business-to-business
sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. To
document this, they have been marked with the following identification:
Because disposal regulations may differ from one country to another within the EU, please contact your
supplier if necessary.
In Germany, this is mandatory from March 23, 2006. From this date, the manufacturer has to offer a
suitable method of return for all devices supplied after August 13, 2005. For all devices supplied before
August 13, 2005, the last user is responsible for the correct disposal.
Page 84
84
Transport, storage and disposal
®
Eppendorf BioSpectrometer
kinetic
English (EN)
Page 85
Technical data
Eppendorf BioSpectrometer
English (EN)
11Technical data
11.1Power supply
Power supply100 V to 240 V ±10 %, 50 Hz to 60 Hz
Overvoltage categoryII
Degree of pollution2
Power consumptionMaximum power consumption according to name plate: 50 W
Approx. 30 W during operation
Approx. 5 W with the display dimmed and temperature control
switched off
Permitted mains interruptionApprox. 10 ms at 90 V
Approx. 20 ms at 230 V
Protection classI
FusesT 2.5 A/250 V, 5 mm × 20 mm (2 pcs.)
®
kinetic
85
11.2Ambient conditions
OperationAmbient temperature: 15°C to 35°C
Rel. humidity: 25% to 70%
Air pressure: 86 kPa to 106 kPa
Air pressureUse up to an altitude of 2000 m above MSL
Do not expose to direct sunlight.
11.3Weight/dimensions
Weight5.3 kg
DimensionsWidth: 295 mm
Depth: 400 mm
Height:150 mm
Space requiredWidth: 500 mm (with thermal printer: 750 mm)
Depth: 500 mm
Page 86
86
Technical data
Eppendorf BioSpectrometer
English (EN)
®
kinetic
11.4Photometric properties
Measuring principleSingle beam absorption spectrophotometer with reference beam
Maximum temperature
uncertainty (in relation to the
temperature in the sample)
Systematic temperature error±0.2 °C at 25 °C to 37 °C
Random temperature error±0.15 °C at 25 °C to 37 °C
(see Tab. on p. 27)
Page 87
Technical data
®
Eppendorf BioSpectrometer
kinetic
English (EN)
11.6Further technical parameters
Cuvette materialFor measurements in the UV:
Quartz glass or UV transparent plastic (Eppendorf UVette, 220 nm to
1600 nm)
For measurements in the visible range:
Glass or plastic material
Cuvette shaft12.5 mm × 12.5 mm, tempered
Overall cuvette heightMin. 36 mm
Height of the light beam in the
cuvette
Key pad22 foil keys
Result outputAbsorbance, concentration, scan (absorbance wavelength spectrum)
DisplayVGA TFT display 5.7”
Operator guidance languageEnglish, French, Spanish, Italian, German
InterfacesUSB master: for USB stick and thermal printer DPU-S445
8.5 mm
6 foil keys as softkeys
Additional, method-dependent data (ratio, FOI, background
absorbances)
USB slave for connecting to a PC
Serial interface RS 232: for thermal printer DPU-414
Connected devices must meet the safety requirements specified in
IEC 60950-1.
87
Page 88
88
Technical data
Eppendorf BioSpectrometer
English (EN)
®
kinetic
11.7Application parameters
MethodsPreprogrammed and freely programmable methods for all measuring
and evaluation procedures:
• Absorbance measurements with one or more wavelengths, scans
• Nucleic acids and proteins, OD600, dye methods (parallel
measurement of biomolecule and dye marking)
• Methods with evaluation via factor, standard and standard series
• Dual wavelength procedure with subtraction and division
evaluation
• Kinetic method: End point, two-point, linear regression
Method-dependent evaluationAbsorbance, concentration via factor and standard.
Concentration via standard series:
• Linear regression
• Nonlinear regression (2nd and 3rd degree polynoms)
• Spline evaluation
• Linear interpolation (point-to-point evaluation)
Absorbance calculations via subtraction and division
Additional data for nucleic acids: Ratios 260/280 and 260/230; molar
concentration, total yield
Additional data for dye methods: FOI (frequency of incorporation,
labeling density)
Scans: zoom, peak evaluation
Kinetics: Subsequent modification of the time window for the
regression evaluation
Method memory>100 method programs
Measured value memory and
calibration memory
Memory for >1 000 results with all data of the results evaluation and
standard evaluation, sample number, sample name, date and used
parameter set of the method program.
(The number of saved results depends on the number of saved
methods.)
Page 89
Evaluation procedure
Eppendorf BioSpectrometer
English (EN)
12Evaluation procedure
This chapter describes the evaluation procedures available in the method programs as well as the
calculation of a dilution using the device software.
When comparing the measuring results to the results of other photometers/
spectrophotometers, note that the values may be dependent on the bandwidth of the devices.
In the following cases the differences may be significant:
• The absorbance spectrum shows a narrow peak in the measurement wavelength.
• The measurement is carried out not at the maximum but at the edge of a peak.
Therefore, check the accuracy of the methods by measuring standards.
12.1Absorbance values
®
kinetic
89
Absorbance values are displayed as A
(XXX represents the wavelength). These displays always match
XXX
the directly measured values, i.e., without corrections, which are incorporated in the final evaluation, e.g.,
corrections for optical path lengths of the cuvette, or background corrections.
12.1.1Blank
All absorbance values are always related to the last measured blank (blank). Therefore, a blank
measurement is compulsory at the start of every series of measurements and can be completed at any time
during a series of measurements. Ideally, the blank measurement should be able to compensate for any
influences on the absorbance value of the measuring solution. The blank should therefore be measured
with the same buffer that was used for the sample measurement and the same cuvette that was used to
measure the sample value – unless the cuvettes used for the blank and sample measurements are optically
aligned and thus have the same absorbance value at the measuring wavelength.
12.1.2Background correction
Main application: Partial correction of distortions of the absorbance for nucleic acid measurements due to
turbidity in the measuring solution. For example, the absorbance at 320 nm, which should be approx. 0 A
with pure nucleic acids, is subtracted from the absorbance at 260 nm, (the measuring wavelength for
nucleic acids).
AAA
,
BkgrXXXcorrBkgrXXX
A
XXX, corrBkgr
= measured absorbance at a wavelength of XXX nm.
A
XXX
= measured absorbance at the background wavelength.
A
Bkgr
= calculated corrected absorbance at a wavelength of XXX nm.
Page 90
90
FACu
Evaluation procedure
Eppendorf BioSpectrometer
English (EN)
®
kinetic
12.1.3Cuvette correction
All absorbance values which are used for result calculation are standardized to the cuvette layer thickness
of 10 mm. If a cuvette with a different path length is used, this path length must be defined in the cuvette
parameter. In this case, the measured absorbances are corrected to match measuring results with a cuvette
layer thickness of 10 mm before converting them to sample results.
This correction is applied to:
• Methods with evaluation by factor.
•Methods of the Absorbance group, for which only absorbance values are output.
The correction is not applied to:
• Methods with evaluation by standards, as we presume that standards and samples are measured in
cuvettes of the same layer thickness.
• Calculations with division: Division method (Dual wavelength method group) and calculation of ratios
such as A
260/A280
(for nucleic acid measurements).
AA
u
XXXcorrCuvXXX10,
Cuv
A
XXX, corrCuv
= measured absorbance at a wavelength of XXX nm.
A
XXX
Cuv = path length of the cuvette.
= calculated corrected absorbance at a wavelength of XXX nm.
12.2Evaluation with factor or standard
C = calculated concentration.
A = absorbance.
F = factor.
The factor is programmed in the parameter list and can be modified. It always relates to an optical path
length of the cuvette of 10 mm. If you change the Cuvette parameter the device will take the modification
into account when calculating the results. Therefore you do not need to change the factor for the
evaluation.
If, on the other hand, you modify the concentration unit, you have to ensure that the factor is adjusted for
the selected unit.
Page 91
Evaluation procedure
S
S
A
C
F
®
Eppendorf BioSpectrometer
kinetic
English (EN)
The factor is either entered directly as a parameter during the "Factor" evaluation procedure or calculated
during the "Standard" evaluation procedure (evaluation with a standard concentration):
F = calculated factor.
C
= concentration of the standard (enter as parameter).
S
A
= measured absorbance of the standard.
S
If multiple measurement (2 or 3 replicates) has been programmed for the standard, the average value is
calculated from the measured absorbance values and inserted as A
.
S
91
12.3Evaluation with standard curve/line
If evaluations are made with more than one standard, the following evaluation procedures for the standard
curve/line can be selected with the [Curve fit] in the measure standards/new method step:
Evaluation procedureDescriptionMinimum required number of
standard points
Linear interpolationLinear point-to-point connection in the
absorbance concentration graph of the
standard evaluation.
Linear regressionPolynome regression for first degree
polynomial.
Quadratical regressionPolynome regression for second degree
polynomial.
Cubical regressionPolynome regression for third degree
polynomial.
Spline interpolationInterpolation via natural cubic splines.3 standards minimum.
For the regression procedure, one can select that the regression line (regression curve) goes through the
zero point.
2 standards minimum.
3 standards minimum.
4 standards minimum.
5 standards minimum.
Page 92
92
Evaluation procedure
Eppendorf BioSpectrometer
English (EN)
• Use the "linear regression" procedure for calibration lines.
• With curvilinear gradients, test which evaluation procedure (quadratic regression, cubic
regression, spline interpolation) produces the function that is most suitable to the standard
evaluation. Spline interpolation connects the measuring points by cubic polynomials,
whereas the regression methods position a quadratic or cubic function between the
measuring points in such a way, that the smallest possible deviation from the function
results for the measuring points.
• Aside from the calculated regression equation, the regression method also displays the
coefficient of determination as a measure for the scattering of the measuring points around
the calculated function. At a value of < 0.8 for the coefficient of determination the result is
issued with a warning.
• If the first standard hat a concentration of "0", select the setting in which the regression
line (regression curve) goes through the zero point.
• If none of the procedures recommended for curvilinear gradients produce satisfactory
results, select the "linear interpolation" procedure.
®
kinetic
12.4Dilution
In the measure samples method step. entered dilutions are considered in the result calculation:
C
V
V
= result converted using the dilution factor
Dil, corr
= volume of the sample in the measuring solution
S
= volume of the diluent in the measuring solution
Dil
Page 93
Evaluation procedure
®
Eppendorf BioSpectrometer
kinetic
English (EN)
12.5Special evaluation procedures for nucleic acids and protein UV
This section covers the evaluation of nucleic acids or proteins in the Nucleic acids and Proteins direct UV
method groups, as well as the corresponding biomolecular components in the Dye labels method group.
93
12.5.1Correction A
and correction A
260
280
Application: correction of the influence of dye absorbance on the nucleic acid or protein absorbance at 260
and 280 nm for the methods of the Dye labels group.
The application of the evaluation procedure can be activated in the parameters Correct A260 or Correct A280.
ACFAAu
,
A
XXX, corr
A
XXX
= calculated corrected absorbance for a wavelength of 260 nm or 280 nm
= measured absorbance for a wavelength of 260 nm or 280 nm
YYYXXXcorrXXX
CF = correction factor for a wavelength of 260 nm or 280 nm (the correction factors for 260 nm and 280 nm
are both dye-specific and are programmed in General Method Parameter: Dyes in the Functions area).
A
= measured absorbance at the dye wavelength.
YYY
The absorbance values displayed in the results are the directly measured, not the corrected
absorbance values.
12.5.2Ratios A260/A280 and A260/A230
Application: Information on the purity of the measured nucleic acid. Application of the evaluation
procedure can be activated in the A260/280 or A260/A230 parameters.
"Ratio" refers to the quotients of the measured absorbances at the listed wavelengths.
Literature values for ratio values with pure nucleic acids:
A260/A280
• DNA: 1.8 to 1.9
• RNA: 1.9 to 2.0
(Current Protocols in Molecular Biology, 1994)
A260/A230
For the ratios A260/A230, different information can be found in the literature for pure nucleic acids:
• DNA: 2.3 to 2.5
(The Nucleic Acids, 1955)
•DNA: 1.9
(Current Protocols in Molecular Biology, 1994)
The values are highly dependent on the pH value. Therefore, nucleic acids should not be measured in
water, but in a buffer with a pH of 7 to 7.2 (e.g., TE buffer).
Page 94
94
Evaluation procedure
Eppendorf BioSpectrometer
English (EN)
®
kinetic
12.5.3Conversion to molar concentrations and nucleic acid quantities
The conversion only can be applied to nucleic acids and dye methods with nucleic acids as biomolecule
components. It is realized in the process results/More calculations method step.
12.5.3.1 Calculation of amount
Application: calculating the amount (mass) of nucleic acid in the total sample volume.
M = calculated total amount (mass) of nucleic acid in the sample tube. Unit: μg.
C = nucleic acid concentration calculated from the measurement. Unit: μg/mL or ng/μL.
= total volume of the sample in the sample tube. Enter this value in More calculations. Unit: μL.
V
S, total
12.5.3.2 Calculation of the molar concentration
Application: calculation of the molar concentration of the nucleic acid from the mass concentration and
relative molar mass. The molar mass is either entered directly or calculated by the device from the entered
number of bases or base pairs per nucleic acid molecule.
3
C
C
Mol
= calculated molar concentration of the nucleic acid. Unit: pmol/mL.
C
Mol
C = nucleic acid concentration calculated from the measurement. Unit: μg/mL or ng/μL.
MM = relative molar mass. Unit: kDa
If the number of bases or base pairs per nucleic acid molecule are entered in More calculations instead of
the relative molar mass, the MM is calculated from the number of the bases or base pairs:
For dsDNA:
For ssDNA, RNA, Oligo:
10u
MM
MM = calculated relative molar mass; unit: kDa
bp = entered number of base pairs per molecule
b = entered number of bases per molecule
Page 95
Evaluation procedure
%1.0
1
A
F
P
P
P
MM
A
H
%1.0
®
Eppendorf BioSpectrometer
kinetic
English (EN)
•For dsDNA the calculation of the molar concentration is based on the assumption of a
double-stranded nucleic acid. For the ssDNA, RNA and Oligo methods, a single-stranded
nucleic acid is assumed.
• For methods which have been reprogrammed via <New Method> in the Routine main
group, Nucleic acids method group, always double-stranded nucleic acids are assumed
for calculating the molar concentration.
12.5.4Calculating the factor for protein in "General Method Parameter"
This section only covers the calculation of the protein components in the Dye labels and Proteins direct UV method groups. For these method groups, the protein component is selected in the parameters (see
Method parameters on p. 33). The protein component is assigned a factor that will be entered in the
General Method Parameter/Proteins function for each protein. Alternatively, A
coefficient plus the molar mass of the protein can be entered instead of the factor. In this case, the factor is
calculated as follows:
or the absorbance
0.1%
95
F = factor for the protein; unit: g/L.
= absorbance of the protein at a concentration of 0.1 % (1 g/L).
A
0.1%
When entering the molar absorbance coefficient and the relative molar mass of the protein A
calculated on this basis:
ε
= molar absorbance coefficient of the protein; unit: cm-1M-1.
P
= relative molar mass of the protein; unit: Da (entry in General Method Parameter in kDa).
MM
P
0.1%
can be
Page 96
96
NAXXX
nt
Dye
YYY
FA
MMA
FOI
u
u
u
6
10
H
NAXXXDye
YYY
FA
A
FOI
u
u
9
10
H
Evaluation procedure
Eppendorf BioSpectrometer
English (EN)
®
kinetic
12.6Special evaluation procedures for the dye methods
12.6.1Calculating the factor for the dye from the absorbance coefficient
For the dye methods the concentration of the dye is calculated using a factor from the measured
absorbance (see Evaluation with factor or standard on p. 90). The factor is entered for each dye in the
General Method Parameter/Dyes function. Alternatively, you can enter the absorbance coefficient. In this
case, the factor is calculated as follows:
6
Dye
10
H
Dye
F
F = factor for the dye; unit: pmol/μL.
ε = absorbance coefficient for the dye, unit: cm
-1M-1
12.6.2Calculation of the FOI
As a value for the ratio of dye molecules to the number of nucleotides in the nucleic acid the frequency of
incorporation (FOI) is calculated and displayed for the dye methods. The calculation can be selected for two
different result units:
MOLECULE dye/kb unit
pmole/μg DNA (or RNA) unit
A
= absorbance of the dye.
YYY
= absorbance of the nucleic acid.
A
XXX
= average molar mass of the nucleotides: 330 g/mol.
MM
nt
= factor for calculating the nucleic acid
F
NA
ε
= absorbance coefficient for the dye, unit: cm-1M
dye
-1
Page 97
Evaluation procedure
totalP
VCM
,
u
dc
Ab
Aa
A
calc
u
u
u
2
1
Eppendorf BioSpectrometer
®
kinetic
English (EN)
12.6.3Conversion to amounts of dye
The quantity (dimensions) of dye in the entire sample volume is calculated in the process results/More calculations method step.
M = calculated total amount (mass) of dye in the sample tube. Unit: pmol.
C = dye concentration calculated from the measurement. Unit: pmol/μL.
V
= total volume of the sample in the sample tube; entered by the user under More calculations.
S, total
Unit: μL.
12.7Dual wavelength
97
For methods of the Dual Wavelength group absorbances that were measured at two wavelengths can be
calculated with each other before the calculated absorbance is evaluated further with the factor or
standard.
To determine the calculated absorbance a division or subtraction evaluation can be defined in the
parameters:
>@
calc
, A2 = measured absorbance.
A
1
a, b, c, d = factors that are entered in the parameters. Negative figures may also be entered.
21
dcAbAaA
uuu
Page 98
98
RBcorrRB
AAA
,
Evaluation procedure
Eppendorf BioSpectrometer
English (EN)
®
kinetic
12.8Kinetics
An absorbance value A or an absorbance difference standardized to a minute ΔA/min is determined using
the measurement procedures available for selection. This determined absorbance value is entered in the
concentration calculation using a factor or (only for Advanced kinetics) with an end-point standard.
12.8.1Measurement procedures
Endpoint
At the end of the waiting period (incubation period,Delay parameter), an absorbance value is measured
and used for the concentration calculation.
Two point
After the delay the measuring point is taken and after the measuring time has expired a second one is
taken. ΔA/min is calculated from the absorbance and the time difference.
AAA
'
mintt
A
, t1 = absorbance and time for the first measuring point.
1
, t2 = absorbance and time for the second measuring point.
A
2
Linear regression
After the delay has expired measuring points are recorded at fixed time intervals from the start to the end of
the measuring time. A linear regression is carried out via the measuring points in the absorbance/time
diagram. Result: absorbance value in ΔA/min.
12
12
12.8.2Reagent blank value
For the methods of the Advanced kinetics group, the measurement of a reagent blank can be
preprogrammed in the parameters. Primary application: compensation of a reagent drift in the "Linear
regression" measurement procedure. Instead of a sample, the reagent blank contains the reagent and
demineralized water, and is measured according to the same measurement procedure as the sample. The
measurement is made at the start of the sample series; the absorbance result is then subtracted from the
sample absorbance result before the evaluation:
A
A = measured absorbance result (in A or ΔA/min) of the sample.
A
= corrected absorbance result (in A or ΔA/min) of the sample.
RB, corr
= absorbance result (in A or ΔA/min) of the reagent blank.
RB
Page 99
Evaluation procedure
®
Eppendorf BioSpectrometer
kinetic
English (EN)
If the time frame for the kinetic evaluation with linear regression is reduced in the process results method step, the reduced measuring window is automatically used for calculating the
reagent blank which was incorporated in the calculation of this sample result. That means that
the reagent blank is recalculated (only) for the calculation of this sample result. For the
calculation of the other sample results the reagent blank result that was calculated on the
basis of the originally used time window is used.
When the kinetic method is evaluated with standards the measurement of a reagent blank is not available.
Here, the reagent blank can alternatively be defined as the first standard ("zero" concentration).
99
Page 100
100
Evaluation procedure
Eppendorf BioSpectrometer
English (EN)
®
kinetic
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.