This document supplements the operation manual of Centrifuge 5810/5810 R and does not replace it. Therefore, please also read the operating manual before starting up the rotors for the first time. You can find the current version on the Internet at www.eppendorf.com/worldwide
.

1 Product description

1.1 Features

The rotor A-4-81, A-4-81-MTP/Flex is a swing-bucket rotor for the Centrifuge 5810/ 5810 R. The rotor has a maximum capacity of 4 × 500 mL.

2Safety

2.1 Warnings for intended use

CAUTION! Risk of injury from asymmetric loading of rotors.
Load rotors symmetrically with identical tubes and/or buckets and plates.Only load adapters with suitable tubes or plates.Always use the same type of tubes and plates (weight, material/density and
volume). Always observe the max. g-force of the tubes and plates as indicated by the manufacturer.
Check that loading is symmetrical by balancing the adapters and tubes and/or
plates used with scales.
NOTICE! If handled incorrectly, the rotor can fall over.
The buckets of the rotor may not be used as a handle.
Before moving the rotor, remove the buckets.Always grip the rotor on the rotor cross using both hands.

3Operation

Observe the instructions in the operating manual of the centrifuge on inserting the rotor and loading the buckets symmetrically as well as for information on the max. load per bucket and the max. loading height/tube length for the buckets.

3.1 Inserting the rotor

Use the supplied rotor key for the rotor A-4-81/A-4-81-MTP/Flex.

3.2 Loading the swing-bucket rotor

Use the rotor, bucket and adapter combinations approved by Eppendorf.Check the grooves in the buckets for cleanliness. Lightly lubricate the grooves with
pivot grease.
Check that all buckets have been fully attached and can swing out freely.
Instructions for use
Rotor A-4-81, Rotor A-4-81-MTP/Flex English (EN)

3.2.1 Loading the rectangular bucket

Insert the adapters into the rectangular bucket with the black clamp facing the
"Eppendorf" logo on the bucket.
For tubes with a length of more than 100 mm perform a manual swing-out test with
empty tubes. The tubes must not come into contact with the rotor cross.

3.2.2 Loading the plate buckets

Observe the maximum load and the maximum height when using stacked plates.

4 Maintenance

4.1 Cleaning

• All buckets, buckets, caps and adapters can be autoclaved (121°C, 20 min).
• The rotor cross is not autoclavable.
Clean the rotor and accessories with a mild cleaning agent.

5 Technical data

Rectangular bucket 500 mL
Bucket for concical tubes 7 × 50 mL
MTP/Flex bucket
Max. g-force (rcf) 3 220 × g 3 220 × g 2 900 × g
Max. speed (rpm) 4 000 rpm 4 000 rpm 4 000 rpm
Max. load per bucket (adapter, vessel and
780 g 7 × 75 g 380 g
content)
Max. centrifugation radius
Max. loading height/ vessel length
Your local distributor: www.eppendorf.com/contact
Eppendorf AG · 22331 Hamburg · Germany
eppendorf@eppendorf.com · www.eppendorf.com
18 cm 18 cm 16.3 cm
Please note the data in the operating manual of the centrifuge.