Eppendorf 5430 R User Manual [en, es]

Page 1
Register your instrument!
www.eppendorf.com/myeppendorf
Centrifuge 5430/5430 R
Operating manual Manual de instrucciones
Page 2
©
2013 Eppendorf AG, Hamburg. No part of this publication may be reproduced without the prior
permission of the copyright owner.
Trademarks
Eppendorf
®
, the Eppendorf logo. CombiSlide®, and QuickLock® are registered trademarks of
Eppendorf AG, Hamburg, Germany.
Centriplus
Microtainer
Trademarks are not marked in all cases with ™ or
®
is a registered trademark of Millipore Corporation, Billerica, USA.
®
is a registered trademark of Becton Dickinson, Franklin Lakes, NJ, USA.
®
in this manual.
Operating manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Manual de instrucciones . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Declarations and Certificates . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Menu structure of Centrifuge 5430/5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 127
Rotor F-35-6-30: Adapters for round-bottom tubes and blood taking systems . . . . . . . . . . . . . . . . . . . . . 128
5427 901.124-04/072013
Page 3
Table of contents
Centrifuge 5430/5430 R
English (EN)
Operating manu alCentrifuge 5430/5430 Rseeon p.Fig.Tab.p.

English (EN)Operating manual

Table of contents
1 Operating instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2 Danger symbols and danger levels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.1 Danger symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.2.2 Danger levels. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.3 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
1.4 Abbreviations used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
2 Product description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.1 Main illustration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
2.2 Delivery package. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.1 Centrifuge 5430. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.2.2 Centrifuge 5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2.3 Features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 11
2.4 Rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
2.4.1 rcf display and calculation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
3
3 Safety. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.1 Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.2 User profile . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3 Application limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3.1 Declaration concerning the ATEX directive (94/9/EC) . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.3.2 Maximum service life for accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
3.4 Information on product liability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5 Warnings for intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.1 Personal injury or damage to the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
3.5.2 Incorrect handling of the centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
3.5.3 Incorrect handling of the rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
3.5.4 Extreme strain on the centrifuging tubes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
3.5.5 Aerosol-tight centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
3.6 Safety notes on the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.1 Selecting the location . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.2 Preparing installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
4.3 Installing the instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
5 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.1 Overview of operating controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28
5.2 Menu navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
5.3 Configure centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.1 Set menu language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.3.2 Setting the date and time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
5.4 Preparing for centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.1 Switching on the centrifuge . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.2 Inserting the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.3 Automatic rotor detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.4 Manual rotor detection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
5.4.5 Loading the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
5.4.6 Closing the centrifuge lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Page 4
Table of contents
Centrifuge 5430/5430 R
4
English (EN)
5.5 Cooling (only 5430 R) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.6 Centrifuging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.7 Standby mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.8 User instructions on rotors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.5.1 Temperature adjustment. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.2 Temperature display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.3 Temperature monitoring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.4 FastTemp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
5.5.5 FastTemp pro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
5.5.6 Continuous cooling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38
5.6.1 Centrifugation with time setting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 39
5.6.2 Centrifuging in continuous operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.6.3 Short-spin centrifugation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
5.6.4 Removing the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
5.8.1 Rotor F-35-6-30: rotor removal tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
5.8.2 Rotor A-2-MTP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.8.3 Rotor FA-45-24-11-HS: using the special rotor key . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
5.8.4 QuickLock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
6 Operating controls and function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.1 Device menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
6.2 Settings in the device menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2.1 Programs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2.2 Use program keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
6.2.3 Other menu items . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
6.2.4 Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 49
7 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.1 Prepare cleaning/disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.2 Cleaning/disinfection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
7.2.1 Cleaning and disinfecting the device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.2.2 Cleaning and disinfecting the rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
7.3 Additional service instructions for Centrifuge 5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 52
7.4 Glass breakage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.5 Fuses . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
7.6 Decontamination before shipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
8 Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.1 General errors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 54
8.2 Error messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 55
8.3 Emergency release . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
9 Transport, storage and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.1 Transport. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.2 Storage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
9.3 Disposal. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
10 Technical data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10.1 Power supply. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10.2 Ambient conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
10.3 Weight/dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
10.4 Application parameters. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 60
Page 5
Table of contents
Centrifuge 5430/5430 R
English (EN)
11 Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11.1 Centrifuge 5430 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11.2 Centrifuge 5430 R . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 62
11.3 Rotors, rotor lids and seals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
11.3.1 Rotors with QuickLock rotor lid . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
11.3.2 Rotors with rotor lid thread. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
11.3.3 Rotors with rotor lid for attaching. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.4 Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.4.1 Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
11.4.2 Other accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
11.4.3 Fuses for Centrifuge 5430. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68
5
Page 6
Table of contents
Centrifuge 5430/5430 R
6
English (EN)
Page 7
Operating instructions
Centrifuge 5430/5430 R
English (EN)

1 Operating instructions

1.1 Using this manual

Read this operating manual completely before using the device for the first time. Please also note the
operating instructions for the accessories, if applicable.
This operating manual is part of the product. Thus, it must always be easily accessible.Enclose this operating manual when transferring the device to third parties.If this manual is lost, please request another one. For the current version, please refer to our webpage
www.eppendorf.com
The Centrifuge 5430/5430 R is available in two versions: key pad or rotary knobs. This operating manual generally describes how to operate the keypad version but it also applies to the rotary knob version.
(international) or www.eppendorfna.com (North America).

1.2 Danger symbols and danger levels

The safety instructions in this manual appear with the following danger symbols and danger levels:
7
1.2.1 Danger symbols
Biohazard Explosion
Electric shock Crushing
Hazard point Material damage
1.2.2 Danger levels
DANGER Will lead to serious injuries or death. WARNING May lead to serious injuries or death. CAUTION May lead to light to moderate injuries. NOTICE May lead to material damage.

1.3 Symbols used

Example Meaning
You are requested to perform an action.
1.
2.
• List.
Perform these actions in the sequence described.
Press this key to perform the described action.
Text Terms from the display of the device.
References useful information.
Page 8
Operating instructions
Centrifuge 5430/5430 R
8
English (EN)

1.4 Abbreviations used

MTP
Micro test plate
PCR
Polymerase chain reaction
PTFE
Polytetrafluorethylene
RZB/rcf
Relative centrifugal force – g-force in m/s
rpm
Revolutions per minute – in rpm
2
UV
Ultraviolet radiation
Page 9

2 Product description

1
2
3
4
5
678
1
2
3
4
5
1
9
6
8
7

2.1 Main illustration

Abb. 2-1: Front and rear view of Centrifuge 5430
Product description
Centrifuge 5430/5430 R
English (EN)
9
Fig. 2-1: Front and rear view of Centrifuge 5430
Abb. 2-2: Front and rear view of Centrifuge 5430 R
Fig. 2-2: Front and rear view of Centrifuge 5430 R
Page 10
10
Product description
Centrifuge 5430/5430 R English (EN)
1 Emergency lid release
On both sides of the device (see Emergency release on p. 57).
2 Centrifuge lid
3 Control panel
Keys and dials (dependent on the device version) for operating the centrifuge(see p. 28).
4Window
Visual control for rotor stop or option for speed check via stroboscope.
5Display
Depiction of the current centrifuging parameters and device settings (see p. 28).
6 Mains connection
Connection socket for the mains cable supplied. Only 5430: The fuse holder is located beneath (see Fuses on p. 53).
7 Mains switch
Switch for switching the device on and off. Switch position 0: The device is switched off. Switch position I: The device is switched on.
8ID plate
9USB port
Interface for error analysis and software updates by the Technical Service.

2.2 Delivery package

2.2.1 Centrifuge 5430
Quantity Order no.
(International)
1– See chapter Ordering Information for corresponding device
1 5301 850.249 022654403 4 A (230 V), 2 pieces oder 5427 850.341 022654381 8.0 AT UL (120 V/100 V), 2 pieces
1 5416 301.001 022634305 Standard 1– Mains power cable
1 5427 901.124 5427901124 Languages: EN, DE, FR, ES, IT, PT 1 5427 901.132 5427901132 Languages: DA, FI, EL, NL, SV (230 V devices only)
2.2.2 Centrifuge 5430 R
Quantity Order no.
(International)
1– See chapter Ordering Information for corresponding device
1 5416 301.001 022634305 Standard 1– Mains power cable
1 5427 901.124 5427901124 Languages: EN, DE, FR, ES, IT, PT 1 5427 901.132 5427901132 Languages: DA, FI, EL, NL, SV (230 V devices only)
Order no. (North America)
Order no. (North America)
Description
Centrifuge 5430
version, equipment and order number
Fuse
Rotor key
Operating manual Centrifuge 5430/5430 R
Description
Centrifuge 5430 R
version, equipment and order number
Rotor key
Operating manual Centrifuge 5430/5430 R
Page 11
Product description
Centrifuge 5430/5430 R
English (EN)

2.3 Features

The multifaceted Centrifuge 5430 has a capacity of 48 × 2.0 mL and reaches max. 30,130 × g/17,500 min-1. The versatility is reflected in the available rotor options. You can select from 12 different rotors to centrifuge the following tubes for your various applications:
• Tubes (0.2 to 5.0 mL)
•PCR strips
• Microtainers
•Spin columns
•Cryo tubes
• Conical tubes (15/50 mL)
•Microplates
•PCR plates
• Deepwell plates (max. height 29 mm)
• Slides (with CombiSlide adapter)
Five program keys for rapid loading and saving of parameters, as well as another 45 program places, a large display and menu-controlled operation all make it easier to use the centrifuge. The Centrifuge 5430 has been designed based on latest ergonomic studies. This facilitates an intuitive and easy operation.
11
The Centrifuge 5430 is available with two different operator panels: one easy to clean keypad or blue rotary knobs to quickly set the centrifugation parameter.
The Centrifuge 5430 R has an additional temperature control function for centrifugation between -11°C and +40°C. The Fast Temp function is used to start a temperature control run without samples in order to quickly bring the rotor chamber, and rotor, bucket and adapter, to the set temperature. This temperature control cycle can also be started automatically at specified times using the Fast Temp pro function.
Page 12
12
Product description
Centrifuge 5430/5430 R English (EN)

2.4 Rotors

The Centrifuge 5430/5430 R can be operated with the following rotors. Before using tubes, observe the manufacturer's recommended specifications on resistance to centrifugation (max. g-force).
Rotor FA-45-48-11
With aerosol-tight QuickLock rotor lid
Rotor F-45-48-11
With polypropylene rotor lid
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
48 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
Outer ring: 18,213 × g Inner ring: 16,048 × g/ 12,700 rpm
≤ 20 s/20 s ≤ 61 s/65 s (soft)
48 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
Outer ring: 18,213 × g Inner ring: 16,048 × g/ 12,700 rpm
≤ 20 s/20 s ≤ 61 s/65 s (soft)
Max. load per rotor
(1)
bore
(2)
3.75 g
3.75 g
Notes
• Aerosol-tight QuickLock rotor lid (aluminum).
(3)
Rotor FA-45-30-11
With aerosol-tight QuickLock rotor lid
Rotor F-45-30-11
With polypropylene rotor lid
30 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
30 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
20,817 × g/ 14,000 rpm
≤ 15 s/15 s ≤ 61 s/65 s (soft)
20,817 × g/ 14,000 rpm
≤ 15 s/15 s ≤ 61 s/65 s (soft)
3.75 g
3.75 g
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
•PTFE-coated (particularly resistant to chemicals), marked: coated
•Spin columns available, better with rotor FA-45-24-11-kit.
Page 13
Product description
Centrifuge 5430/5430 R
English (EN)
13
Rotor FA-45-24-11-Kit
With aerosol-tight QuickLock rotor lid
Rotor FA-45-16-17
With aerosol-tight QuickLock rotor lid
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
24 spin columns or 1.5/2.0 mL tubes. With adapters:
• 0.2 mL PCR tubes
19.090 × g/ 13,200 rpm
≤ 15 s/16 s ≤ 78 s/90 s (soft)
• 0.4 mL tubes
• 0.5 mL tubes
• 0.6 mL Microtainer
16 tubes 5.0 mL. 21,191 × g/
14,200 rpm ≤ 20 s/20 s
≤ 61 s/66 s (soft)
Max. load per rotor
(1)
bore
(2)
3.75 g
9.5 g
Notes
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
• Uniquely high edge, for all commercial spin columns.
Also observe the note on centrifugation with open tube lids (see
Fixed-angle rotors on p. 33)
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
Rotor S-24-11-AT
With aerosol-tight QuickLock rotor lid
24 tubes 1.5/2.0 mL. 16,049 × g/
12,700 rpm
This rotor exclusively is to be used with 1.5/2.0 mL reaction
≤ 13 s/16 s
≤ 61 s/66 s (soft) vessels. Spin Columns and the adapters including the corresponding 0.2 mL, 0.4 mL,
0.5 mL and 0.6 mL vessels must not be used in this rotor.
3.75 g
• Aerosol-tight
(3)
QuickLock rotor lid (aluminum).
•The rotor must always be operated with a rotor lid.
Page 14
14
Product description
Centrifuge 5430/5430 R English (EN)
Rotor FA-45-24-11-HS
Rotor F-45-64-5-PCR
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
24 tubes 1.5/2.0 mL. With adapters:
• 0.2 mL PCR tubes
• 0.4 mL tubes
30,130 × g/ 17,500 rpm
≤ 21 s/16 s ≤ 61 s/65 s (soft)
• 0.5 mL tubes
• 0.6 mL Microtainer
64 PCR tubes (0.2 mL) or
13,543 × g/
11,800 rpm eight 5-tube or 8-tube PCR strips, each with the enclosed
≤ 12 s/15 s
≤ 62 s/65 s (soft) adapters.
Max. load per rotor
(1)
bore
(2)
3.75 g
3.4 g (without adapter)
Notes
• Aerosol-tight
(3)
rotor
lid (aluminum).
• Max. g-force/speed (30,130 × g/17,500 rpm) only with tubes approved by the man­ufacturer for this speed.
• PTFE-coated (particu­larly resistant to chemicals), marked:
coated
• Spin columns avail­able, better with rotor FA-45-24-11-kit.
•The rotor can only be tightened and loos­ened using the special rotor key for rotor FA-45-24-11-HS (see p. 42).
Rotor F-45-18-17-Cryo
18 Cryo tubes or 18 sealable centrifugation tubes, max. Ø: 16.9 mm. With supplied adapters: max. Ø: 13.4 mm, max. tube length: 50 mm.
8,324 × g/ 8,900 rpm
≤ 8 s/11 s ≤ 77 s/85 s (soft)
8.7 g • Setting the g-force/ speed in increments of 10 × g or 10 rpm.
Page 15
Product description
Fixed-angle
rotor
Aerosol-tight
version
Angle of
bores
Number of
bores
of
bores (mm)
AT
24 11
Aerosol-tight
version
Number of
bores
of
bores (mm)
Swing-bucket
rotor
Centrifuge 5430/5430 R
English (EN)
15
Rotor F-35-6-30
Rotor A-2-MTP
Max. capacity Max. g-force
(rcf)/speed (rpm) without adapter
Acceleration/ deceleration time (soft): with soft ramp
6 conical tubes 50 mL, with or without skirted bottom or 6 conical tubes 15 mL, with
7,745 × g/ 7,830 rpm
≤ 23 s/23 s ≤ 62 s/67 s (soft)
enclosed adapters or 6 Centriplus centrifuge filter units with adapters.
Two buckets to hold:
•Microplates
2,204 × g/ 4,680 rpm
• Cell culture plates
•PCR plates
• Deepwell plates (max.
≤ 18 s/21 s ≤ 63 s/67 s (soft)
height 29 mm)
• Slide (with CombiSlide adapter)
Max.
Notes load per rotor
(1)
bore
(2)
110 g • The rotor can only be
removed and inserted using the supplied removal tool.
•Centrifugation of round-bottom tubes and blood taking systems can be completed using additional adapters (see p. 128).
170 g (per bucket)
• PCR plates can only be centrifuged with appropriate adapters.
• Max. loading height: 29 mm.
Only 5430 R: More efficient cooling via centrifugation without wind shield upper shell (see
Swing-bucket rotor on p. 34).
(1) Maximum load per rotor bore for adapter + tube + contents.
(2) According to DIN 58 970 (device versions: 230 V, 120 V and 100 V, 50 to 60 Hz).
(3) Aerosol tightness tested and certified by the Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down (UK) (see certificates at the end of this operating manual).
For the rotors and rotor lids labeled coated, color fluctuations may occur as a result of the production process. These fluctuations have no effect on service life or resistance to chemicals.
Page 16
16
Product description
Centrifuge 5430/5430 R English (EN)
2.4.1 rcf display and calculation
Use the rpm/rcf key to switch the speed of centrifugation display from rpm to g-force (rcf). Ensure that the g-force displayed during switching is standardized to suit the rotor in question without an adapter. When adapters are used, the following maximum g-forces (rcf) can be achieved at maximum speed:
Rotor Adapter Max.centrifugation
radius r
Rotor FA-45-48-11/ Rotor F-45-48-11
Rotor FA-45-30-11/ Rotor F-45-30-11
Rotor FA-45-24-11-Kit
Rotor FA-45-16-17 For 5.0 mL micro test tubes 9.4 21,191 Rotor S-24-11-AT Without adapter 8.9 16,049 Rotor
FA-45-24-11-HS
Rotor F-45-64-5-PCR
Rotor F-45-18-17-Cryo
Without adapter Outer ring: 10.1
Inner ring: 8.9
For 0.2 mL PCR tubes Outer ring: 8
Inner ring: 6.8
For 0.4 mL tubes Outer ring: 10.1
Inner ring: 8.9
For 0.5 mL tubes Outer ring: 9
Inner ring: 7.8
For 0.6 mL Microtainer Outer ring: 10.1
Inner ring: 8.9 Without adapter 9.5 20,871 For 0.2 mL PCR tubes 7.4 16.215 For 0.4 mL tubes 9.5 20,871 For 0.5 mL tubes 8.4 18,407 For 0.6 mL Microtainer 9.5 20,817 Without adapter 9.8 19,090 For 0.2 mL PCR tubes 7.7 15,000 For 0.4 mL tubes 9.8 19,090 For 0.5 mL tubes 8.7 16,950 For 0.6 mL Microtainer 9.8 19,090
Without adapter 8.8 30,130 For 0.2 mL PCR tubes 6.7 22,940 For 0.4 mL tubes 8.8 30,130 For 0.5 mL tubes 7.7 26,364 For 0.6 mL Microtainer 8.8 30,130 For PCR strips, inside 7.7 11,987 For PCR strips, outside 8.7 13,543 Without adapter 9.4 8,320 For Cryo tubes 9.0 7,970
max
[cm]
Max. g-force (rcf)
Outer ring: 18,210 Inner ring: 16,048
Outer ring: 14,425 Inner ring: 12,261
Outer ring: 18,210 Inner ring: 16,048
Outer ring: 16,229 Inner ring: 14,065
Outer ring: 18,210 Inner ring: 16,048
Page 17
Product description
Centrifuge 5430/5430 R
English (EN)
17
Rotor Adapter Max.centrifugation
radius r
max
[cm]
Max. g-force (rcf)
Rotor F-35-6-30* For 15 mL conical tubes 11.0 7,540
For 50 mL conical tubes 10.5 7,197 for Centriplus centrifuge filter units 11.1 7,567
A-2-MTP rotor Without adapter 9.0 2,204
for 384 well PCR plates 7.7 1,885 for 96 well PCR plates 7.3 1,788 CombiSlide adapter 7.7 1,885
*) Centrifugation of round-bottom tubes and blood taking systems can be done using additional adapters (see p. 128).
To determine the g-force (rcf) for a special adapter, you can perform a calculation according to DIN 58 970 with the following formula:
-5
rcf = 1.118 · 10
· n2 · r
max
n: speed in min-1 (rpm)
: max. centrifuging radius in cm
r
max
Example:
In Rotor FA-45-30-11, the 0.5 mL adapter has a maximum radius of 8.4 cm. At 7,000 rpm, a maximum g-force of 4,600 × g is achieved.
Page 18
18
Safety
Centrifuge 5430/5430 R English (EN)

3Safety

3.1 Intended use

The Centrifuge 5430/5430 R is intended exclusively for indoor use and for separating aqueous solutions and suspensions of various densities in approved test tubes.

3.2 User profile

This device may only be operated by trained specialist staff. They must have carefully read the operating manual and be familiar with the function of the device.

3.3 Application limits

3.3.1 Declaration concerning the ATEX directive (94/9/EC)
DANGER! Risk of explosion.
Do not operate the device in areas where work is completed with explosive substances.Do not use this device to process any explosive or highly reactive substances.Do not use this device to process any substances which may generate an explosive
atmosphere.
Due to its design and the environmental conditions inside the device, the Centrifuge 5430/5430 R is not suitable for use in a potentially explosive atmosphere.
The device only must be used in a safe environment, such as the open environment of a ventilated laboratory or fume hood. The use of substances which could create a potentially explosive atmosphere is not permitted. The final decision on the risks associated with the use of these types of substances is the responsibility of the user.
3.3.2 Maximum service life for accessories
WARNING! Risk of injury from chemically or mechanically damaged accessories.
Even minor scratches and cracks can lead to serious internal material damage.
Protect all accessory parts from mechanical damage.Inspect the accessories for damage before each use. Replace any damaged accessories.Do not use rotors, rotor lids or buckets with signs of corrosion or mechanical damage (e.g.,
deformations).
Do not use any accessories whose maximum service life has been exceeded.When inserting the buckets and rotors, ensure that they do not become scratched.
CAUTION! Risk of injury due to chemically damaged rotor lids or caps.
Transparent rotor lids or caps made from PC, PP or PEI may lose their strength under the impact of organic solvents (e.g., phenol, chloroform).
If rotor lids or caps have come into contact with organic solvents, they should be cleaned
immediately.
Regularly check the rotor lids and caps for damages and cracks.Immediately replace rotor lids or caps that have cracks or milky stains.
Page 19
Safety
Centrifuge 5430/5430 R
English (EN)
The rotors listed below, and the corresponding buckets and rotor lids, have a maximum service life of the number of years or cycles listed in the table. Since 2012, Eppendorf has been stating the maximum service life of the rotors and accessories both in years and in the maximum number of cycles. The decisive factor for the service life is which case occurs first, generally this is the number of years in operation.
Rotor Maximum service from commissioning onward
FA-45-48-11 (With QuickLock)
F-45-48-11 (With rotor lid thread)
FA-45-30-11 (With QuickLock)
FA-45-24-11-Kit (With QuickLock)
FA-45-16-17 (With QuickLock)
S-24-11-AT (With QuickLock)
A-2-MTP With the corresponding bucket and upper shell of the wind shield
QuickLock rotor lid - 3 years Rotor lid and caps made of polycarbonate (PC),
polypropylene (PP) or polyetherimide (PEI) Plastic adapters - 1 year
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 10 years
100,000 mechanical cycles 7 years
-3 years
19
For all other rotors and rotor lids of this centrifuge there is no service life limit as long as the following requirements are met:
•Proper use,
• Recommended maintenance
• Undamaged condition
The date of manufacture is stamped on the rotors in the format 03/10 (= March 2010) or on the inside of the plastic rotor lids in the form of a clock .
To ensure aerosol tightness, the following applies:
• Replace aerosol-tight rotor lids and caps after 50 autoclaving cycles.
• Replace the seal of QuickLock rotor lids after 50 autoclaving cycles.
Page 20
20
Safety
Centrifuge 5430/5430 R English (EN)

3.4 Information on product liability

In the following cases, the designated protection of the device may be compromised. Liability for any resulting property damage or personal injury is then transferred to the operator:
• The device is not used in accordance with the operating manual.
• The device is used outside of its intended use.
• The device is used with accessories or consumables which are not recommended by Eppendorf.
• The device is maintained or repaired by people not authorized by Eppendorf.
• The user makes unauthorized changes to the device.

3.5 Warnings for intended use

Read the operating manual and observe the following general safety instructions before using the Centrifuge 5430/5430 R.
3.5.1 Personal injury or damage to the equipment
WARNING! Electric shock due to damage to device or mains cable.
Only switch on the device if the device and mains cable are undamaged.Only use devices that have been properly installed or repaired.In case of danger, disconnect the device from the mains supply by pulling the power plug
from the device or the mains socket or, by using the isolating device intended for this purpose (e.g. emergency stop switch in the laboratory).
WARNING! Lethal voltages inside the device.
Ensure that the housing is always closed and undamaged so that no parts inside the device
can be contacted by accident.
Do not remove the housing of the device.Do not allow any liquids to penetrate the inside of the housing.Do not allow the device to be opened by anyone except service personnel who have been
specifically authorized by Eppendorf.
WARNING! Risk from incorrect supply voltage
Only connect the device to voltage sources which correspond to the electrical
requirements on the name plate.
Only use sockets with a protective earth (PE) conductor and suitable power cable.
WARNING! Damage to health due to infectious liquids and pathogenic germs.
When handling infectious liquids and pathogenic germs, observe the national regulations,
the biological security level of your laboratory, the material safety data sheets, and the manufacturer's application notes.
Use aerosol tight sealing systems for the centrifugation of these substances.When working with pathogenic germs belonging to a higher risk group, more than one
aerosol-tight bioseal must be used.
Wear personal protective equipment.For full instructions regarding the handling of germs or biological material of risk group II
or higher, please refer to the "Laboratory Biosafety Manual" (Source: World Health Organization, current edition of the Laboratory Biosafety Manual).
Page 21
Safety
Centrifuge 5430/5430 R
English (EN)
WARNING! Risk of injury when opening or closing the centrifuge lid.
There is a risk of crushing your fingers when opening or closing the centrifuge lid.
When opening or closing the centrifuge lid, do not reach between the lid and device or into
the latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
CAUTION! Poor safety due to incorrect accessories and spare parts.
The use of accessories and spare parts other than those recommended by Eppendorf may impair the safety, functioning and precision of the device. Eppendorf cannot be held liable or accept any liability for damage resulting from the use of incorrect or non-recommended accessories and spare parts, or from the improper use of such equipment.
Only use accessories and original spare parts recommended by Eppendorf.
NOTICE! Damage to device due to spilled liquids.
1. Switch off the device.
2. Disconnect the device from the power supply.
3. Carefully clean the device and the accessories in accordance with the cleaning and disinfection instructions in the operating manual.
4. If a different cleaning and disinfecting method is to be used, contact Eppendorf AG to ensure that the intended method will not damage the device.
21
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a warmer environment.
After installing the device, wait for at least 4 h. Only then connect the device to the mains
power supply.
3.5.2 Incorrect handling of the centrifuge
NOTICE! Damage from knocking against or moving the device during operation.
If the rotor bangs against the rotor chamber wall, it will cause considerable damage to the device and rotor.
Do not move or knock against the device during operation.
Page 22
22
Safety
Centrifuge 5430/5430 R English (EN)
3.5.3 Incorrect handling of the rotors
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with rotor and rotor lid firmly tightened.If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
properly secured. Immediately press the start/stop key to stop centrifuging.
CAUTION! Risk of injury due to asymmetric loading of a rotor.
Load rotors symmetrically with identical tubes or plates and buckets.Only load adapters with suitable tubes or plates.Always use tubes or plates of the same type (weight, material/density and volume).Check that loading is symmetrical by balancing the adapters and tubes or plates used with
scales. The device automatically detects imbalances during operation and stops the run immediately with an error message and a signal tone. Check the loading, balance the tubes and re-start the centrifugation.
CAUTION! Risk of injury from overloaded rotor.
The Centrifuge 5430/5430 R is designed for the centrifugation of material with a max. density of 1.2 g/mL at maximum speed and filling volume and/or load.
Do not exceed the maximum load of the rotor.
NOTICE! Damage to rotors from aggressive chemicals.
Rotors are high-quality components which withstand extreme stresses. This stability can be impaired by aggressive chemicals.
Avoid using aggressive chemicals, including strong and weak alkalis, strong acids,
solutions with mercury, copper and other heavy metal ions, halogenated hydrocarbons,
concentrated saline solutions and phenol.
If the rotor is contaminated by aggressive chemicals, clean it immediately using a neutral
cleaning agent. This applies to the rotor bores, in particular.
Due to the manufacturing process, color variations may occur on rotors marked "coated".
These color variations do not effect service life or resistance to chemicals.
NOTICE! If handled incorrectly, the rotor can fall over.
The rotor can fall over if the buckets are used as a handle.
Remove the buckets before inserting and/or removing a swing-bucket rotor.Always use both hands to carry the rotor cross.
CAUTION! Risk of injury due to chemically damaged rotor lids or caps.
Transparent rotor lids or caps made from PC, PP or PEI may lose their strength under the impact of organic solvents (e.g., phenol, chloroform).
If rotor lids or caps have come into contact with organic solvents, they should be cleaned
immediately.
Regularly check the rotor lids and caps for damages and cracks.Immediately replace rotor lids or caps that have cracks or milky stains.
Page 23
3.5.4 Extreme strain on the centrifuging tubes
CAUTION! Risk of injury from overloaded tubes.
Note the loading limits specified by the tube manufacturer.Only use tubes which are approved by the manufacturer for the required rcf.
NOTICE! Risk from damaged tubes.
Damaged tubes must not be used, as this could cause further damage to the device and the accessories and loss of the samples.
Before use, visually check all of the tubes for damage.
NOTICE! Risk from open tube lids.
Open tube lids can brake off during centrifugation and damage the rotor and the centrifuge.
Carefully seal all tube lids before centrifuging.
Exception: Note the information on the centrifugation of spin columns in the rotor FA-45-24-11-Kit (see p. 34).
Safety
Centrifuge 5430/5430 R
English (EN)
23
NOTICE! Hazard to plastic tubes from organic solvents.
The density of plastic tubes is reduced when organic solvents (e.g., phenol, chloroform) are used, i.e. the tubes could become damaged.
Observe the manufacturer's specifications for chemical resistance of the tubes.
NOTICE! Sample tubes heat up.
In uncooled centrifuges, the temperature in the rotor chamber, rotor and sample can increase to above 40 °C, based on the run time, g-force (rcf)/speed and ambient temperature.
Note that this can reduce the centrifugation resistance of the sample tubes.Please note the temperature resistance of the samples.
Page 24
24
Safety
Centrifuge 5430/5430 R English (EN)
3.5.5 Aerosol-tight centrifugation
WARNING! Risk to health due to limited aerosol tightness with incorrect rotor/rotor lid combination.
Aerosol-tight centrifugation is guaranteed only if the rotors and rotor lids intended for this purpose are used. The designation of aerosol-tight fixed-angle rotors always starts with FA.The aerosol-tight rotors and rotor lids of this centrifuge are additionally marked with a red ring on the rotor and a red rotor lid screw. Aerosol-tight swing-bucket rotors are marked with AT (aerosol-tight).
For aerosol-tight centrifugation, always simultaneously use rotors and rotor lids which are
marked as aerosol-tight in the centrifuge intended for the corresponding purpose. The
details specifying in which centrifuge you may use the aerosol-tight rotors and rotor lids
can be found on the rotor and, beginning from production date of October 2003, on the
upper side of the rotor lid.
Only use aerosol-tight rotor lids in combination with rotors which are marked on the rotor
lid.
Only use aerosol-tight buckets with the corresponding caps.
WARNING! Damage to health as a result of limited aerosol tightness and incorrect usage.
Autoclaving, mechanical stresses and contamination by chemicals or other aggressive solvents can impair the aerosol-tightness of the rotors and rotor lids.
Check the integrity of the seals of the aerosol-tight rotor lids or caps before each use.Only use aerosol-tight rotor lids or caps if the seals are undamaged and clean.Lightly grease the threads of the rotor lid screw with pivot grease after every proper
autoclaving (121 °C, 20 min.) (int. order no. Int. 5810 350.050, North America
022634330).
Replace aerosol-tight rotor lids and caps after 50 autoclaving cycles.For QuickLock rotor lids, the seal must be replaced after 50 autoclaving cycles.Never store aerosol-tight rotors or buckets closed.

3.6 Safety notes on the device

Symbol Meaning Location
Hazard point. 5430: Rear of the device
CAUTION
Always tighten the rotor with the enclosed rotor key.
CAUTION
Close all tubes and use a rotor lid.
5430 R: Right side of the device
Upper side of device, under the centrifuge lid.
Upper side of device, under the centrifuge lid.
Page 25

4 Installation

4.1 Selecting the location

NOTICE! If an error occurs, the objects in the immediate proximity of the device will be damaged.
In accordance with recommendations in EN 61010-2-020, leave a safety clearance of
30 cm around the device during operation.
Please remove all materials and objects from this area.
NOTICE! Damage from overheating.
Do not install the device near heat sources (e.g. heating, drying cabinet).Do not expose the device to direct sunlight.Ensure unobstructed air circulation. Maintain a clearance of at least 30 cm around all
ventilation grilles.
Installation
Centrifuge 5430/5430 R
English (EN)
25
Select the location for the device according to the following criteria:
• Suitable power connection as per the name plate (230 V/120 V/100 V).
• Stable, horizontal and resonance-free lab bench.
• A well ventilated environment which is protected from direct sunlight to prevent the device from heating up more.

4.2 Preparing installation

Prerequisites The weight of the centrifuge is 29 kg (5430) or 56 kg (5430 R). A second person is needed to unpack and
position the device.
Retain the packaging material and the transport protection device for subsequent transport or storage. See also the instructions relating to transport (see p. 58).
Perform the following steps in the sequence described.
Centrifuge 5430 Centrifuge 5430 R
1. Open the packaging board.
2. Remove the covering cardboard.
3. Remove the accessories.
4. Grip from the strap retainers and have two persons lift the centrifuge out of the box.
5. Pull off the strap retainers, do not cut.
6. Remove the front and rear transport securing devices from the centrifuge.
7. Remove the plastic sleeve.
8. Carefully lift the centrifuge on one side and pull off the transport securing device of the motor on the underside of the centrifuge.
1. Open the packaging board.
2. Remove the accessories.
3. Lift and remove the front and rear transport securing devices.
4. Grip from the fabric straps and have two people lift the centrifuge out of the box.
5. Pull off the textile straps, do not cut.
6. Remove the plastic sleeve.
Page 26
26
Installation
Centrifuge 5430/5430 R English (EN)

4.3 Installing the instrument

Prerequisites The device is on a suitable lab bench.
NOTICE! Damage to electronic components due to condensation.
Condensate can form in the device after it has been moved from a cool environment to a warmer environment.
After installing the device, wait for at least 4 h. Only then connect the device to the mains
power supply.
NOTICE! Centrifuge 5430 R: compressor damage after improper transport.
Only switch on the centrifuge 4 h after installation.
Perform the following steps in the sequence described.
1. Let the device warm up to ambient temperature for at least 4 hours to avoid damaging electronic assemblies as a result of condensation and avoid damaging the compressor (only 5430 R).
2. Check that the mains voltage and frequency match the requirements on the device type plate.
3. Connect the centrifuge to the mains and switch it on using the mains/power switch on the rear of the device (Centrifuge 5430) or at the right side of the device (Centrifuge 5430 R).
• The standby key lights green.
•The display is active.
Only 5430: Lid opens automatically
4. Only 5430: Remove the transport securing device of the motor shaft.
5. If the scope of delivery includes a rotor, dismantle and remove it using the supplied rotor key.
6. Only 5430: Remove the transport securing device of the air guide ring.
Page 27
Centrifuge 5430/5430 R
7. Use the details included in the scope of delivery to check that the delivery is complete.
8. Check all parts for any transport damage. Contact your dealer if any damage is found.
Retain the packaging material and the transport protection device for subsequent transport or storage. See also the instructions relating to transport (see p. 58).
Installation
27
English (EN)
Page 28
28
1
2
3
4
5
6
7891011
13 12
Operation
Centrifuge 5430/5430 R English (EN)

5 Operation

5.1 Overview of operating controls

The Centrifuge 5430/5430 R is available in two versions: keypad or rotary knobs. This operating manual generally describes how to operate the keypad version. However, it also applies to the rotary knob version.
Before using the Centrifuge 5430/5430 R for the first time, familiarize yourself with the operating controls and the display.
Abb. 5-1: Control panel and display of Centrifuge 5430/5430 R (keypad version).
Fig. 5-1: Control panel and display of Centrifuge 5430/5430 R (keypad version).
1 Select program
Press briefly: load the stored centrifuging
parameters. Press and hold (> 2 sec): save the current centrifuging parameters (see p. 47).
2 Short Spin centrifugation
(see p. 40)
3Release lid
4 Start and stop centrifugation
5 Activate/deactivate standby mode
Key lights green: centrifuge is ready for operation. Key lights red: standby mode active (see p. 41).
6Display
8 Adjust the temperature (only 5430 R)
9 Adjust the centrifuging duration
Dependent on device version, designed as key or rotary knob.
10 Start the temperature control cycle Fast Temp
(only 5430 R)
(see p. 36)
11 Switch the displayed centrifuging speed (rpm/
rcf)
12 Call and select the menu parameters
(see p. 47)
13 Navigating the menu
(see p. 29)
7 Set speed of centrifugation
Dependent on device version, designed as key or rotary knob.
Page 29
Abb. 5-2: Display of the Ce ntrifuge 5430/5430 R
1
2 3 4 5 7
6
8
10
9
111213
Fig. 5-2: Display of the Centrifuge 5430/5430 R
Operation
Centrifuge 5430/5430 R
English (EN)
29
1 Program number (if enabled)
2 Status of the function At set rpm
: Start of operation when reaching 95% of the
preset g-force (rcf) or speed.
: Start of run time immediately.
3 Status of the key lock
: Centrifuging parameters cannot be modified
unintentionally.
: No key lock.
4 Status of the loudspeaker
: Switched on.
: Switched off.
5 Status of the centrifuge
: Centrifuge lid unlocked.
: Centrifuge lid locked.
(flashing): Centrifuging in progress.
6 Temperature control cycle programming (only
5430 R)
: Fast Temp pro is enabled, i.e., start time and temperature of the temperature control run have been programmed (see p. 37).
8Soft ramp
: Rotor accelerates and brakes slowly.
No symbol: Rotor accelerates and brakes rapidly.
9Standard display
10 Extended display (if enabled)
11 g-force/speed
12 Temperature (only 5430 R)
13 Centrifuging duration
7Time
Please also read the precise description of the individual functions (see p. 47).

5.2 Menu navigation

The menu consists of two levels. To change settings, generally proceed as follows.
1. Open menu.
Page 30
30
Operation
Centrifuge 5430/5430 R English (EN)
2. Select the desired menu item.
3. Confirm selection.
4. Select the setting of the parameters in question.
5.
6.
Confirm changed setting. A tick appears in front of the confirmed setting.
Keep pressing the key until you reach the desired menu level or exit the menu. Some menus can only be exited by selecting and confirming the menu item Back /
Zurück / Retour / Atrás.

5.3 Configure centrifuge

5.3.1 Set menu language
Proceed as follows to set menu language.
1. Open the menu.
2. Select Settings.
3. Confirm your selection.
4. Select Language.
5. Confirm your selection.
6. Select English, Deutsch, Francais or Espanol.
7.
8. Press key several times to exit the menu.
Confirm your selection. A tick appears in front of the selected language. The setting takes effect immediately.
5.3.2 Setting the date and time
Proceed as follows to set date and time.
1. Open the menu.
Page 31
2. Select Settings.
3. Confirm your selection.
4. Select Date/time.
5. Confirm your selection.
6. Set date.
7. Confirm setting.
Operation
Centrifuge 5430/5430 R
English (EN)
31
8. Set time format (12 h/24 h).
9. Confirm setting.
10. Set time.
11. Confirm setting.
12. Press key several times to exit the menu.
There is no automatic switch between summer time and winter time.
Page 32
32
Operation
Centrifuge 5430/5430 R English (EN)

5.4 Preparing for centrifugation

5.4.1 Switching on the centrifuge
1. Switch on the centrifuge using the mains power switch or the standby key. Only 5430: After switching on at the mains power switch, the centrifuge lid opens automatically.
2. Open the closed centrifuge lid by pressing the open key. The parameter settings of the last run are displayed.
5.4.2 Inserting the rotor
Swing-bucket rotors: remove the buckets before inserting and/or removing the rotor. Use
both hands to pick up the rotor cross.
Rotor F-35-6-30: only use the rotor removal tool supplied to insert or remove the rotor.
1. Fit the rotor vertically on the motor shaft.
2. Insert the supplied rotor key into the rotor nut. Rotor FA-45-24-11-HS: Use the special rotor key.
3. Turn rotor key clockwise until the rotor nut is firmly tightened.
5.4.3 Automatic rotor detection
The centrifuge has automatic rotor detection. It detects a newly inserted rotor during centrifugation and displays its name for approx. 2 seconds. The set g-force (rcf)/speed (rpm) is automatically limited to the maximum permissible value of the rotor, if necessary.
If you start centrifuging immediately after a rotor change, the centrifuge has not carried out an automatic rotor detection yet. The speed set for the previous rotor may exceed the maximum permitted speed for the new rotor. In this case the centrifuge stops after the automatic rotor detection and displays the error message Note C. The new maximum permitted speed appears in the display. You can then restart the centrifuging with this setting or adjust the speed as necessary.
Always check the set g-force (rcf)/speed (rpm) after a rotor change and adjust it if necessary.
5.4.4 Manual rotor detection
CAUTION! Risk of injury when turning the rotor manually.
When turning a swing-bucket rotor, pay special attention to ensure that your fingers do not
get jammed or get caught on the swinging buckets.
In order to trigger the rotor detection manually, turn the rotor counterclockwise by hand.
• The name of the rotor appears in the display.
• The g-force (rcf)/speed (rpm) automatically is limited to the maximum value of the rotor.
Page 33
5.4.5 Loading the rotor
CAUTION! Risk of injury due to asymmetric loading of a rotor.
Load rotors symmetrically with identical tubes or plates and buckets.Only load adapters with suitable tubes or plates.Always use tubes or plates of the same type (weight, material/density and volume).Check that loading is symmetrical by balancing the adapters and tubes or plates used with
scales. The device automatically detects imbalances during operation and stops the run immediately with an error message and a signal tone. Check the loading, balance the tubes and re-start the centrifugation.
CAUTION! Risk from damaged or overloaded tubes.
When loading the rotor, observe the safety precautions on dangers as a result of
overloaded or damaged tubes (see Warnings for intended use on p. 20).
Operation
Centrifuge 5430/5430 R
English (EN)
33
The device automatically detects imbalances during operation and stops the run immediately with an error message and a signal tone.
Check the load, balance the tubes and restart the run.
5.4.5.1 Fixed-angle rotors
Rotor lid
• Fixed-angle rotors may only be operated with the appropriate rotor lid in each case. This is clearly shown by the identical rotor name labeling on the rotor and on the rotor lid.
• To carry out an aerosol-tight centrifugation, an aerosol-tight rotor (label: red ring) and the corresponding aerosol-tight rotor lid (label: aerosol-tight and red lid screw) must be used.
To load the rotor, proceed as follows:
1. Check the maximum load (adapter, tube and contents) per rotor bore. The information about this can be found on the rotor and in this operating manual (see Rotors on p. 12).
2. Load rotors and adapters only with the tubes intended for them.
3. Insert tubes opposite each other in pairs into the rotor bores. To ensure symmetric loading, tubes that are arranged opposite each other must be of the same type and contain the same filling quantity.
To minimize weight differences between filled sample tubes, we recommend taring with a scale. This will reduce wear on the drive and reduce running noise.
4. Attach and tighten rotor lid.
Page 34
34
Operation
Centrifuge 5430/5430 R English (EN)
Spin columns
For centrifuging spin columns in the rotor FA-45-24-11-Kit, you can leave the tube lids open. However, this can only be done using the tubes provided by kit manufacturers for this purpose. For reliable centrifugation, you must lean the open tube lids against the edge of the rotor. Ensure that this does not involve the lids projecting beyond the edge of the rotor, and then put on the associated rotor lid.
5.4.5.2 Swing-bucket rotor
Prerequisites
• A combination of rotor, carrier and adapter, approved by Eppendorf.
• Matching and tested tubes and plates.
NOTICE! Filling the plates too high can cause overflowing.
During the run the meniscuses in the tubes along the edges of the plates are at an angle. This is due to the centrifugal forces and cannot be avoided.
Fill the wells of the plates to a maximum of 2/3 of the max. capacity.
To load the rotor, proceed as follows:
1. Check the carrier grooves for cleanliness and grease lightly with pivot grease (order no. Int.: 5810 350.050 / North America: 022634330).
Dirty grooves and pivots prevent carriers from swinging out evenly.
2. Hang the buckets into the rotor. All rotor positions must be loaded with carriers.
3. Check that all carriers are hanging properly and can swing freely.
4. Carry out a manual loading and swing test the first time a tube or plate type is used.
5. Check the maximum load per carrier (adapter, tube or plate and contents) and the loading height. The information about this can be found on the rotor and in this operating manual (see Rotors on p. 12).
Page 35
Operation
Centrifuge 5430/5430 R
English (EN)
6. Load the buckets symmetrically.
Abb. 5-3: Incomplete, but symmetric loading of the buckets. The pegs of each bucket must be uniformly loaded .
Fig. 5-3: Incomplete, but symmetric loading of the buckets. The pegs of each bucket must be uniformly
loaded.
Abb. 5-4: Symmetrical loa ding of the plates.
35
Fig. 5-4: Symmetrical loading of the plates.
The plate arrangement shown on the right-hand side is incorrect, as the buckets will not swing properly.
The plates have some play in the buckets.
7. Check the loading of the bucket.
Carry out a brief centrifugation test at low speed (e.g., 1000 rpm) when you use a tube or
plate type for the first time.
Only 5430 R: When using the A-2-MTPs, centrifuge without the upper shell of the wind shield in order to guarantee precise and quick temperature control of samples. Note that the centrifugation noise will increase slightly in this case. This does not apply to the S-24-11-AT. The S-24-11-AT must always be operated with a rotor lid.
5.4.6 Closing the centrifuge lid
WARNING! Risk of injury when opening or closing the centrifuge lid.
There is a risk of crushing your fingers when opening or closing the centrifuge lid.
When opening or closing the centrifuge lid, do not reach between the lid and device or into
the latching mechanism of the lid.
Always open the centrifuge lid completely to prevent it from falling.
1. Check the correct attachment of the rotor and rotor lid.
2. Push down the centrifuge lid until the lid latch engages and the lid is automatically closed.
The centrifuge will close automatically.
The open key lights up blue. The symbol appears in the display.
Page 36
36
Operation
Centrifuge 5430/5430 R English (EN)

5.5 Cooling (only 5430 R)

5.5.1 Temperature adjustment
Set the temperature using the temp arrow keys between -11°C and +40°C.
You can also modify the temperature during centrifugation.
At higher ambient temperature a brief fan noise is possible until the desired temperature has been reached. This indicates a heavy cooling performance.
5.5.2 Temperature display
If the rotor is stopped: Target temperature During centrifugation: Actual temperature
The set temperature appears in the extended display.
5.5.3 Temperature monitoring
After the target temperature has been reached, the centrifuge reacts to temperature deviations during centrifugation as follows:
Deviation from set value Action
± 3 °C Temperatures on the display are flashing. ± 5 °C Periodic warning tone and display Error 18. Centrifugation is stopped
automatically.
5.5.4 FastTemp
This function can be used to start a temperature control run directly without samples with a rotor and temperature-specific speed in order to quickly adjust the rotor chamber, including the rotor, buckets and adapters, to the previously set nominal temperature.
The FastTemp pro function for programming the temperature control run with defined start times is described in the next section.
Prerequisites
• The centrifuge is switched on.
• The rotor and rotor lid are properly attached.
• The centrifuge lid is closed.
• Temperature and g-force (rcf)/speed (rpm) for the centrifugation are set (see Centrifuging on p. 39).
1. Press the fast temp key.
The display shows Fast Temp, the remaining duration of the temperature control run as well as the current temperature and g-force (rcf)/speed (rpm).
The temperature control run automatically ends when the set temperature is reached. A periodic signal tone sounds.
2. Press the start/stop key to terminate the temperature control run early.
After the set temperature has been reached and the temperature control run is complete, the centrifuge keeps the rotor chamber with the centrifuge lid closed at the set target temperature if the temperature is below the ambient temperature. However, independent of the target temperature, 4 °C must be met via this continuous cooling in order to prevent the rotor chamber from freezing.
Page 37
Operation
Centrifuge 5430/5430 R
English (EN)
The centrifuge stops the cycle automatically if the rotor or the buckets have reached the set temperature. Therefore, there may be a delay of approx. 30 min between the display of the set temperature and the automatic end of the temperature control run.
Only 5430 R: When using the A-2-MTPs, centrifuge without the upper shell of the wind shield in order to guarantee precise and quick temperature control of samples. Note that the centrifugation noise will increase slightly in this case. This does not apply to the S-24-11-AT. The S-24-11-AT must always be operated with a rotor lid.
5.5.5 FastTemp pro
You can have the previously described temperature control run FastTemp (see p. 36) start automatically at a specified time. Two options are available:
Once The temperature control cycle is started once at the set time. Repeatedly The temperature control cycle is started at the set time on the next specified weekday.
This is repeated for an unlimited period of time with each weekday specified.
37
5.5.5.1 Programming the start time
1. Select Fast Temp pro in the device menu.
2. Select Once or Repeatedly. This selection only appears as long as the FastTemp pro function has not already been activated.
Otherwise it is only possible to edit or delete the programmed start time.
3. Only with Repeatedly: Activate/deactivate weekdays with menu/enter, select Continue and confirm with menu/enter.
4. Enter date and time for the one-time or repeated start of the temperature control run as well as the set temperature and confirm with menu/enter.
An overview of the current settings is displayed.
5. Edit the settings again or save.
6. Exit the menu.
FastTemp pro is now activated. In the display the symbol appears as long as an automatic start of a temperature control run is still outstanding. In the standby mode Fast Temp pro is displayed.
• The temperature control run FastTemp (see p. 36) starts automatically at the set time.
• After a one-off programmed temperature control run, the following symbol is extinguished . With several programmed temperature control runs, the FastTemp pro remains active until you deactivate it. To do this, select Fast Temp pro in the device menu and delete the settings.
5.5.5.2 Preparing the centrifuge
Ensure that the centrifuge is switched on or in the standby mode during the start time set and the rotor
and rotor lid are properly attached and the centrifuge lid is closed.
5.5.5.3 Automatic start of the temperature control cycle
1. If the centrifuge is in standby mode, it switches to the operating mode 1 min before the set start time.
2. At the start time the temperature control run FastTemp (see FastTemp on p. 36) begins. Fast Temp pro appears in the display.
Automatically starting the temperature control cycle is not possible during centrifugation.
Page 38
38
Operation
Centrifuge 5430/5430 R English (EN)
5.5.6 Continuous cooling
If the rotor stops, the rotor chamber will be maintained at the target temperature if the following requirements have been met:
• The centrifuge is switched on.
• The centrifuge lid is closed.
• The target temperature is lower than the ambient temperature.
• The centrifuge is not in standby mode.
During continuous cooling the following applies:
• The set and actual temperature are displayed alternately.
• Irrespective of the set temperature, the temperature does not go below 4 °C to prevent the rotor chamber from freezing and from increased condensation in the device.
• The temperature adjustment is slower because the rotor does not rotate during this process.
To end continuous cooling, open the centrifuge lid or press the standby key.
If the centrifuge is not used for more than 8 hours, the continuous cooling is switched off automatically. The device then switches to standby mode. This protects against ice formation in the rotor chamber and increased condensation in the device. With FastTemp you can quickly reach the desired temperature again (see p. 36).
You can change the continuous cooling to endless operation at your own risk. To do so, in the device menu under Continuous cooling enable the (see p. 48) item.
Page 39
Operation
Centrifuge 5430/5430 R
English (EN)

5.6 Centrifuging

CAUTION! Risk from incorrectly-loaded rotors and damaged/overloaded tubes!
Before commencing centrifugation, follow the safety instructions relating to risks from
asymmetrically loaded and/or overloaded rotors and from overloaded, damaged and/or open tubes (see Warnings for intended use on p. 20).
WARNING! Risk of injury from improperly attached rotors and rotor lids.
Only centrifuge with rotor and rotor lid firmly tightened.If unusual noises occur when the centrifuge starts, the rotor or the rotor lid may not be
properly secured. Immediately press the start/stop key to stop centrifuging.
Each of the centrifuging variants described here must be preceded by the preparation described above (see Preparing for centrifugation on p. 32).
39
Only 5430 R: Please also note the instructions on cooling (see Cooling (only 5430 R) on p. 36).
5.6.1 Centrifugation with time setting
Perform the following steps in the sequence described.
1. Use the time arrow keys to set run time.
2. Only 5430 R: Use the temp arrow keys to adjust the temperature.
3. Use the speed arrow keys to the g-force (rcf)/speed.
4. Press start/stop to start centrifuging.
During centrifugation:
blinks in the display as long as the rotor is running.
• The remaining run time will be displayed in minutes. The last minute is counted down in seconds.
Only 5430 R: The current temperature will be displayed.
• The current g-force (rcf) or rotor speed is displayed.
• The shortcut keys, the , open and short keys, and all menu items which directly affect
centrifugation, are blocked during centrifugation.
End of centrifugation
• The centrifuge automatically stops after the set time has elapsed. The elapsed centrifugation will be
shown in a blinking display during the braking process. A signal tone sounds when the rotor is at a standstill.
Only 5430: The centrifuge lid opens automatically. The display shows the symbol .
Only 5430 R: The centrifuge lid remains closed to maintain the sample temperature. It can be
opened lid by pressing the blinking open key.
5. Remove the material for centrifuging.
• During the run you can modify the total run time, the temperature (only 5430 R), the speed and the rpm/rcf indication. The new parameters are adopted immediately. Please note that the shortest new total run time that can be set is the elapsed time plus 2 minuts.
• You can also terminate the centrifugation before the set run time has elapsed by pressing the start/stop key.
Page 40
40
Operation
Centrifuge 5430/5430 R English (EN)
5.6.2 Centrifuging in continuous operation
Perform the following steps in the sequence described.
1. Use the time arrow keys to set the continuous run. The continuous operation function can be set above 99:59 h or below 30 seconds. The timer shows to indicate continuous operation.
2. Only 5430 R: Use the temp arrow keys to adjust the temperature.
3. Use the speed arrow keys to the g-force (rcf)/speed.
4. Press start/stop to start centrifuging.
blinks in the display as long as the rotor is running. Time is counted upwards, first in 30-second increments and then in minute increments from ten minutes.
5. Press start/stop to end centrifuging after the desired time period.
• The centrifugation time will be shown in a blinking display during the braking process.
• A signal tone sounds when the rotor is at a standstill.
Only 5430: The centrifuge lid opens automatically. The display shows the symbol .
Only 5430 R: The centrifuge lid remains closed to maintain the sample temperature. It can be opened lid by pressing the blinking open key.
6. Remove the material for centrifuging.
5.6.3 Short-spin centrifugation
You can carry out a short-spin cycle with the currently set or with the maximum g-force (rcf)/speed of the rotor used. This is set in the device menu (see Other menu items on p. 48) before executing the following steps in the sequence specified:
1. For short-spin cycle with the current g-force (rcf)/speed, set this directly with the arrow keys speed.
2. Only 5430 R: Use the temp arrow keys to adjust the temperature.
3. Start the short run: Hold the short key pressed down.
blinks in the display as long as the rotor is running.
• The time is counted upwards in seconds.
• During short run centrifuging all other keys are blocked. However, short run centrifuging is interrupted if another key is pressed simultaneously.
4. End short spin run: Release the short key.
• The centrifugation time will be shown in a blinking display during the braking process.
• A signal tone sounds when the rotor is at a standstill.
Only 5430: The centrifuge lid opens automatically. The display shows the symbol .
Only 5430 R: The centrifuge lid remains closed to maintain the sample temperature. It can be opened lid by pressing the blinking open key.
5. Remove the material for centrifuging.
During the braking process, you can restart the centrifugation up to two times by pressing the short key again.
The soft ramp does not work during short-spin centrifugation.
Page 41
Operation
Centrifuge 5430/5430 R
English (EN)
5.6.4 Removing the rotor
Swing-bucket rotors: remove the buckets before inserting and/or removing the rotor. Use
both hands to pick up the rotor cross.
Rotor F-35-6-30: only use the rotor removal tool supplied to insert or remove the rotor.
1. Turn the rotor nut counterclockwise using the supplied rotor key. Rotor FA-45-24-11-HS: use special rotor key.
2. Remove the rotor vertically in an upward motion.
3. Only 5430 R: Switch off the centrifuge after use and empty the condensation water tray (remove it from the left side of the device). Leave centrifuge lid fully opened and protect it against closing.

5.7 Standby mode

The centrifuge automatically changes from the ready state to the standby mode if the following prerequisites are met:
• The centrifuge was not used for the time set in the device menu (1 to 60 min)(see Settings on p. 49)
Only 5430 R: The centrifuge lid is open.
41
In the Standby mode, the following applies :
• The standby key lights red.
Only 5430 R: The rotor chamber is not cooled (see Continuous cooling on p. 38).
In the Ready state, the following applies:
• The centrifugation parameters are displayed.
• The standby key lights green.
Only 5430 R: The rotor chamber is cooled when the centrifuge lid is closed (see Continuous cooling on p. 38).
You can switch between standby mode and ready state at any time when centrifugation is not performed by pressing the standby key.
Page 42
42
1. 2.
1. 2.
Operation
Centrifuge 5430/5430 R English (EN)

5.8 User instructions on rotors

5.8.1 Rotor F-35-6-30: rotor removal tool
5.8.1.1 Inserting the rotor
Prerequisites The rotor nut is loose.
Use the rotor removal tool to insert the rotor in the centrifuge and remove it again.
1. Position the rotor removal tool with the narrow side on the rotor thread.
2. Tighten the rotor removal tool clockwise with approx. 3 revolutions.
3. Grip the rotor removal tool and insert the rotor.
4. Loosen and remove the rotor removal tool by turning it counterclockwise.
5. Inserting the rotor: Tighten the rotor using the enclosed rotor key (see Inserting the rotor on p. 32).
5.8.1.2 Loosening the rotor lid
Use the rotor removal tool to loosen a firmly tightened rotor lid screw.
1. Position the rotor removal tool with the wide side on the rotor lid screw.
2. Loosen the rotor lid screw by turning the rotor removal tool counterclockwise.
Page 43
Operation
Centrifuge 5430/5430 R
English (EN)
5.8.2 Rotor A-2-MTP
5.8.2.1 Inserting the rotor
NOTICE! If handled incorrectly, the rotor can fall over.
The rotor can fall over if the buckets are used as a handle.
Remove the buckets before inserting and/or removing a swing-bucket rotor.Always use both hands to carry the rotor cross.
5.8.2.2 Position and remove the upper shell of the wind shield
The upper shell of the wind shield is used for noise reduction.
1. Before it is used for the first time, remove the attachment that is mounted on the knob of the upper shell of the wind shield .
2. Position the upper shell of the wind shield and rotate a max. 1/4 revolution until it is lowered onto the rotor hub.
The lock indicator must protrude so far over the knob that the red marking is clearly visible:
43
The loose positioning of the upper shell is intentional and aids in self-centering.
3. Pull on the knob of the upper shell of the wind shield in order to lift it.
Only 5430 R: When using the A-2-MTPs, centrifuge without the upper shell of the wind shield in order to guarantee precise and quck temperature control of samples. Note that the centrifugation noise will increase slightly in this case. This does not apply to the S-24-11-AT. The S-24-11-AT must always be operated with a rotor lid.
5.8.3 Rotor FA-45-24-11-HS: using the special rotor key
5.8.3.1 Tightening the rotor
1. Insert the rotor key for rotor FA-45-24-11-HS in the rotor nut.
2. Turn the rotor key clockwise until it spins through ("click"). The rotor is correctly tightened.
5.8.3.2 Loosening the rotor
Turn the rotor nut and rotor key for Rotor FA-45-24-11-HS counterclockwise.
The rotor key for Rotor FA-45-24-11-HS can only be used to tighten and loosen this rotor. For the other rotors described in this operating manual, use the rotor key that is delivered with the Centrifuge 5430/5430 R.
Page 44
44
Operation
Centrifuge 5430/5430 R English (EN)
5.8.4 QuickLock
Aerosol-tight rotors FA-45-48-11, FA-45-30-11, FA-45-24-11-Kit, FA-45-16-17 and S-24-11-AT have a quick lock for the rotor lid (QuickLock).
Rotors FA-45-30-11 and FA-45-24-11-Kit have been redesigned and are now only available with a quick lock (QuickLock) instead of a rotor lid thread.
Closing the rotor lid
1. Check the correct positioning of the external sealing ring in the groove.
2. Place the rotor lid on the rotor in a vertical motion.
3. To lock the rotor, turn the red rotor lid screw clockwise as far as it will go, and after an audible "click" is heard.
The rotor is correctly locked after the audible "click" is heard!
If the locking system is stiff, lightly lubricate the pins in the rotor lid screw with pivot grease.
Page 45
Operating controls and function
Centrifuge 5430/5430 R
English (EN)

6 Operating controls and function

6.1 Device menu

Most menu levels contain the additional menu item Back / Zurück / Retour / Atrás. Detailed information on the menu can be found in the following chapters.
Tab. 6-1: Menu structure of the Centrifuge 5430/5430 R in four different menu languages.
English Deutsch Français Español Display Programs
• Load program
•Save program
• Delete program
Soft ramp
•On
•Off
Key lock
•On
•Off
Programme
•Programm laden
•Programm speichern
•Programm löschen
Softrampe
•An
•Aus
Tastensperre
•An
•Aus
Programmes
•Charger prog.
• Enregistrer prog.
• Supprimer prog.
Rampe douce
•Marche
•Arrêt
Verrouilla. de touches
•Marche
•Arrêt
Programas
•Cargar programa
•Guardar programa
• Borrar programa
Rampa suave
• Encendido
•Apagado
Bloqueo del teclado
• Encendido
•Apagado
45
At set rpm
•On
•Off
Short spin
• Maximum speed
•Current speed
Continuous cooling (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•One-time-use
• Repeated use
Settings
•Display –S
andard display
t
– Extended display
•Alarm –On –Off
At set rpm
•An
•Aus
Short Spin
• Maximaler Wert
• Aktueller Wert
Dauerkühlung (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Einmal
•Mehrmals
Einstellungen
•Anzeige –Standardanzeige – Erweiterte
Anzeige
•Lautsprecher –An –Aus
At set rpm
•Marche
•Arrêt
Short Spin
• Vitesse max
• Vitesse actuelle
Refrigération continue (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Une fois
•Plusieurs fois
Réglages
•Affichage –Affichage
standard
–Affichage large
•Signal sonore –Marche –Arrêt
At set rpm
• Encendido
•Apagado
Short Spin
• Velocidad máximo
• Velocidad actual
Refrigeración continua (5430 R)
•8 h
•∞
Fast Temp pro (5430 R)
•Una vez
• Varias veces
Ajustes
•Indicador –Indicador
estándar
–Indicador
extendido
•Altavoz – Encendido –Apagado
Page 46
46
Operating controls and function
Centrifuge 5430/5430 R English (EN)
English Deutsch Français Español Display
•Volume –Cancel –Save –Default
• Date/Time • Datum/Uhrzeit • Date/Heure • Fecha/Hora
• Contrast –Cancel –Save –Default
• Language –English –Deutsch –Français –Español
•Standby –On –Off –Set time
•Lid release
(5430) –Automatic –Manual
• Lautstärke –Abbrechen –Speichern –Lieferzustand
•Kontrast –Abbrechen –Speichern –Lieferzustand
•Sprache –English –Deutsch –Français –Español
•Standby –An –Aus – Zeit einstellen
•Deckelöffnung
(5430) –Automatisch –Manuell
•Volume – Annuler – Enregistrer –Réglage usine
• Contraste – Annuler – Enregistrer –Réglage usine
•Langue –English –Deutsch –Français –Español
• Mise en veille –Marche –Arrêt –Réglage du temps
• Ouverture
couvercle (5430) –Automatique –Manuelle
•Volumen –Cancelar –Guardar – Est. de Fábrica
• Contraste –Cancelar –Guardar – Est. de Fábrica
•Idioma –English –Deutsch –Français –Español
• Standby – Encendido –Apagado –Establecer tiempo
• Apertura de la tapa
(5430) –Automática –Manual
5430: only for Centrifuge 5430 5430 R: only for Centrifuge 5430 R
Page 47
Operating controls and function
Centrifuge 5430/5430 R
English (EN)

6.2 Settings in the device menu

6.2.1 Programs
The Centrifuge 5430/5430 R has more than 50 programmable memory locations.
Load program Load the selected program. This appears in the display with number and name
and can be started immediately using the <>start/stop key. When selecting a program with a too high g-force (rcf)/speed for the rotor used, it flashes and a safety message appears.
Save program Save the set centrifugation parameters (centrifugation duration, temperature
(only 5430 R), g-force (rcf)/speed, soft ramp and At set rpm) under the selected number. In addition, you can assign a program name with a maximum of 20 characters. The centrifuging duration, the temperature (only 5430 R) and the g-force/rotational speed can still be changed in this menu using the corresponding arrow buttons time, temp (only 5430 R) and speed.
Delete program Delete the selected program.
Program numbers 1 to 5 cannot be deleted.
47
These functions are only available with the centrifuge at a standstill.
If the program memory is empty the menu item Delete program is exited automatically after the last program has been deleted. You will furthermore be unable to call up this menu item if the program memory is empty.
Program numbers which are already occupied can be overwritten.
6.2.2 Use program keys
You can also save and load Programs 1 to 5 directly by pressing the program keys:
6.2.2.1 Load program
Press the desired program key briefly.
The pressed program key illuminates in blue, the parameters are displayed.
By pressing again, you can exit the selected program again. The blue light of the key will then go out. The parameters of the most recent centrifugation are then displayed again.
6.2.2.2 Save program
1. Set the centrifugation parameters (centrifugation duration, temperature (only 5430 R), g-force (rcf)/ speed, soft ramp and At set rpm).
2. Press the desired program key for at least 2 seconds.
A signal tone sounds and the program key you pressed lights up.
The centrifuging parameters are saved under the appropriate program number (1-5).
Page 48
48
Operating controls and function
Centrifuge 5430/5430 R English (EN)
6.2.3 Other menu items
Menu item/meaning Setting Function Display Soft ramp
Reduce speed of acceleration and braking ramp. Not used for Short Spin centrifuging.
Key lock
Set the current centrifugation parameters permanently to prevent the time, temperature (only 5430 R), g-force (rcf) or speed, soft ramp and At set rpm from being unintentionally modified.
At set rpm
Set start of centrifuging run time.
Short Spin
Before the start of a short run (see Short-spin centrifugation on p. 40) it is possible to switch between the maximum and currently set g-force (rcf) or speed. Soft ramp is not used for Short Spin centrifuging.
Continuous cooling (only 5430 R) Time limitation of continuous cooling (see p. 38) . Continuous cooling is only activated when the rotor is stopped and the centrifuge lid is closed.
Fast Temp pro (only 5430 R) Time and temperature programming for the automatic temperature control run. The selection Once/Repeatedly only appears if no Fast Temp pro has been set (see FastTemp pro on p. 37).
on Rotor accelerates and brakes slowly.
off Rotor accelerates and brakes rapidly.
on Set the centrifugation parameters
permanently.
off Release the permanent settings.
on The set time is counted down only
once 95% of the specified g-force (rcf) or speed has been reached.
off The set time is counted down
immediately.
Maximum value
Current value
8 h Preset value. Endless operation of continuous
Once Set the date and time for the start of a
Repeatedly Set the week days and the common
Short-spin run at maximum g-force (rcf) or speed of the rotor used.
Short run at set g-force (rcf)/speed.
cooling. Caution! Icing possible! Set at own risk!
temperature control run.
start time for several repeated temperature control runs. Fast Temp pro is active from the date set.
Page 49
Operating controls and function
Centrifuge 5430/5430 R
English (EN)
6.2.4 Settings
Menu item/meaning Setting Function Display Display
Select standard display or extended display of centrifuging parameters.
Loudspeaker
Switch loudspeaker on and off. In the event of error messages, a signal tone sounds even if the loudspeaker is switched off.
Volume
Adjust the speaker volume using the menu arrow keys and in 5 stages.
The signal tone for error messages is always issued at least at medium volume.
Date/time
Set date and time. The system does not switch automatically between summer and winter time.
Contrast
Adjust the display contrast using the menu arrow keys and .
Language Set menu language (English, Deutsch,
Standby
Switch standby mode on and off. If the centrifuge is not used during the set time period and no continuous cooling takes place (only 5430 R), it switches to the standby mode (see p. 41).
Lid release (only 5430) Automatic Lid opens automatically at the end of
Standard display
Extended display
On Switch on loudspeaker.
Off Switch off loudspeaker.
Cancel Exit menu item without saving. Save Save volume just set. Default Restore default volume.
Cancel Exit menu item without saving. Save Save the contrast just set. Default Restore default contrast.
On Switch on standby mode. Off Switch off standby mode. Set time Using the arrow keys, set the time
Manual Lid remains closed at the end of
If the centrifuge is at rest the target values are displayed, and during centrifugation the actual values of run time, temperature (only 5430 R) and the g-force (rcf)/speed.
In addition to the standard display, specified values are always shown at the bottom of the display.
In the date display, set year (YYYY), month (MM) and day (DD). In the time display, set hours (hh) and minutes (mm). Before setting the clock time the time format is selected (12 h / 24 h).
Français or Español) . (see Set menu language on p. 30).
after which the centrifuge should automatically change to the standby mode (1 to 60 min).
centrifuging when the rotor stops.
centrifuging when the rotor stops and can be opened using the now flashing key open.
49
Page 50
50
Maintenance
Centrifuge 5430/5430 R English (EN)

7 Maintenance

7.1 Prepare cleaning/disinfection

Clean all accessible surfaces of the device and the accessories at least weekly and when contaminated.Clean the rotor regularly. This way the rotor is protected and the durability is prolonged.Furthermore, observe the notes on decontamination (see Decontamination before shipment on p. 53)
when the device is sent to the authorized Technical Service for repairs.
The procedure described in the following chapter applies to the cleaning as well as to the disinfection or decontamination. The table below describes the steps required on top of this:
Cleaning Disinfecting/decontamination
1. Use a mild cleaning fluid to clean the accessible surfaces of the device and the accessories.
2. Carry out the cleaning as described in the following chapter.
1. Choose the disinfection method which corresponds to the legal regulations and guidelines in place for your range of application. For example, use alcohol (ethanol, isopropanol) or alcohol-based disinfectants.
2. Carry out the disinfection or decontamination as described in the following chapter.
3. Then clean the device and the accessories.
If you have any further questions regarding the cleaning and disinfection or decontamination or regarding the cleaning fluid to be used, contact the Eppendorf AG Application Support. The contact details are provided on the back of this manual.

7.2 Cleaning/disinfection

DANGER! Electric shock as a result of penetration of liquid.
Switch off the device and disconnect the power plug before starting cleaning or
disinfection work.
Do not allow any liquids to penetrate the inside of the housing.Do not spray clean/spray disinfect the housing.Only plug the device back in if it is completely dry, both inside and outside.
NOTICE! Damage from the use of aggressive chemicals.
Do not use any aggressive chemicals on the device or its accessories, such as strong and
weak bases, strong acids, acetone, formaldehyde, halogenated hydrocarbons or phenol.
If the device has been contaminated by aggressive chemicals, clean it immediately using a
mild cleaning agent.
NOTICE! Corrosion due to aggressive cleaning agents and disinfectants.
Do not use corrosive cleaning agents, aggressive solvents or abrasive polishes.Do not incubate the accessories in aggressive cleaning agents or disinfectants for longer
periods.
Page 51
Maintenance
Centrifuge 5430/5430 R
English (EN)
NOTICE! Damage from UV and other high-energy radiation.
Do not use UV, beta, gamma, or any other high-energy radiation for disinfecting.Avoid storage in areas with strong UV radiation
Autoclaving
All rotors, rotor lids and adapters, except the A-2-MTP rotor, can be autoclaved (121 °C, 20 min). Replace the lids of the aerosol-tight rotors after a maximum of 50 autoclaving cycles. Only QuickLock rotor lid: Only the seal needs to be replaced after a maximum of 50 autoclaving cycle.
Swing-bucket rotor
• Before cleaning the rotor, remove old pivot grease from grooves and pivots.
• Make sure that the grooves and pivots are clean. Dirty grooves and pivots prevent carriers from swinging out evenly.
• After cleaning, lubricate the pivots of the rotor and the grooves of the buckets with pivot grease (order no. Int.: 5810 350.050/North America: 022634330) so that the carriers can move freely in a swinging manner.
51
7.2.1 Cleaning and disinfecting the device
1. Open the lid. Switch off the device with the mains/power switch. Disconnect the power plug from the power supply.
2. Loosen the rotor nut by turning the rotor key counterclockwise.
3. Remove the rotor.
4. Clean and disinfect all accessible surfaces of the device, including the power cable, using a damp cloth and the recommended cleaning agents.
5. Thoroughly clean the rubber seals of the rotor chamber with water.
6. Rub the dry rubber seal with glycerine or talcum powder to prevent it from becoming brittle. Other components of the device, such as the lid latch, lid springs, motor shaft and rotor cone, must not be lubricated.
7. Clean the motor shaft with a soft, dry and lint-free cloth. Do not lubricate the motor shaft.
8. Check the motor shaft for damage.
9. Inspect the device for corrosion and damage.
10. Leave the centrifuge lid open when the device is not being used.
11. Only connect the device to the power supply if it is fully dry inside and out.
7.2.2 Cleaning and disinfecting the rotor
1. Inspect the rotor and accessories for damage and corrosion. Do not use any damaged rotors or accessories.
2. Clean and disinfect the rotors and accessories with the recommended cleaning agents.
3. Use a bottle brush to clean and disinfect the rotor bores.
Page 52
52
Maintenance
Centrifuge 5430/5430 R English (EN)
4. Clean and disinfect the rotor lids. ONLY QuickLock: It is imperative that the sealing ring be removed during this process so that the groove below it can be thoroughly cleaned.
5. Rinse the rotors and accessories thoroughly with distilled water. Rinse the rotor bores of fixed-angle rotors particularly thoroughly.
Do not immerse the rotor in liquid as liquid can get trapped inside the cavities.
6. Place rotors and accessories on a cloth to dry. Place fixed-angle rotors with the rotor bores facing downwards to allow the bores to also dry.
7. Correctly reinsert the rotor lid sealing ring in the clean and dry groove.
8. Clean the rotor cone with a soft, dry and lint-free cloth. Do not lubricate the rotor cone.
9. Inspect the rotor cone for damage.
10. Place the dry rotor onto the motor shaft.
11. Tighten the rotor nut firmly by turning it clockwise with the rotor key.
12. Load the fixed-angle rotor with the cleaned adapters or the swing-bucket rotor with the cleaned buckets and adapters, if necessary.
13. Leave the rotor lid open when the rotor is not being used.

7.3 Additional service instructions for Centrifuge 5430 R

Empty and clean the condensation water tray regularly and especially after liquid spillage in the rotor
chamber. Pull out the tray for condensation water from the left side of the centrifuge.
Clean the condensation water drain on a regular basis, e.g., using a bottle brush.Regularly free the rotor chamber ice formations via thawing, by leaving the centrifuge lid open or
carrying out a short temperature control run at approx. 30 °C.
Leave the centrifuge lid open when not in use for a long period.Wipe up condensate in the rotor chamber using a soft, absorbent cloth.Remove dust deposits from the ventilation slits of the centrifuge using a brush or swab at the latest
every six months. First switch off the device and remove the power plug.
Regularly check the gas spring of the centrifuge lid for proper functioning.
A defective gas spring is an insufficient support for the centrifuge lid and could cause injury if the centrifuge lid falls down. We recommend that the gas spring be replaced by a service technician every 2 years.
Page 53
Maintenance
Centrifuge 5430/5430 R
English (EN)

7.4 Glass breakage

When using glass tubes there is a risk of glass breakage in the rotor chamber. The resulting glass splinters are swirled around in the rotor chamber during centrifugation and have a sandblasting effect on the rotor and accessories. The smallest glass particles become lodged in the rubber parts (e.g., the motor guide, the rotor chamber gasket, and the rubber mats of adapters).
NOTICE! Glass breakage in the rotor chamber
Glass tubes in the rotor chamber may break if the g-force is too high. Broken glass can damage the rotor, accessories and samples.
Please note the manufacturer's information on the recommended centrifugation
parameters (load and speed).
Effects of glass breakage in the rotor chamber:
• Fine black metal abrasion in the rotor chamber (in metal rotor chambers)
• The surfaces of the rotor chamber and accessories are scratched.
• The chemical resistance of the rotor chamber is reduced.
• Contamination of samples
• Wear on rubber parts
53
How to proceed in case of glass breakage
1. Remove all splinters and glass powder from the rotor chamber and accessories.
2. Thoroughly clean the rotor and rotor chamber. Thoroughly clean the bores of the fixed-angle rotors, in particular.
3. If required, replace rubber mats and adapters to prevent any further damage.
4. Regularly check the rotor bores for deposits and damage.

7.5 Fuses

The fuse holder is located under the mains power socket.
1. Pull out the power plug.
2. Remove the fuse holder towards the rear. The two fuses can now be reached. The fuses can be replaced.
Instead of removable fuses, the Centrifuge 5430 R features a thermal overcurrent protective switch. The mains power switch jumps to the '0' switch setting if the overcurrent protective switch is triggered.
1. Use the mains power switch to switch the device back on after more than 20 s.
2. If the mains power switch returns to the '0' switch setting, contact Technical Support.

7.6 Decontamination before shipment

If you are shipping the device to the authorized Technical Service for repairs or to your authorized dealer for disposal please note the following:
WARNING! Risk to health from contaminated device
1. Follow the instructions in the decontamination certificate. It is available as a PDF file on our webpage (www.eppendorf.com/decontamination
2. Decontaminate all the parts you would like to dispatch.
3. Include the fully completed decontamination certificate in the package.
).
Page 54
54
Troubleshooting
Centrifuge 5430/5430 R English (EN)

8 Troubleshooting

If you cannot remedy an error with the recommended measures, please contact your local Eppendorf partner. The contact address can be found online at: www.eppendorf.com
.

8.1 General errors

Symptom/message Cause Remedy
No display. No power connection. No display. Power failure.
Lid of the device cannot be opened.
Lid of the device cannot be opened.
Device cannot be started.
Device shakes when it starts up.
Centrifuge brakes during a short run centrifugation, although the short key is pressed.
Temperature display flashes. (only for 5430 R)
Rotor is still running.
Power failure. 1. Check the mains fuse of the device
Lid of the device is not closed.
Rotor loaded unsymmetrically. 1. Stop the device and load
The short key was released briefly more than twice (protective function for the drive).
Temperature deviation from the set value: ±3 °C.
Check mains/power connection.Check the mains fuse of the device
(see Fuses on p. 53).
Check the mains fuse of the
laboratory.
Wait for rotor to stop.
(see Fuses on p. 53).
2. Check the mains fuse of the laboratory.
3. Activate the emergency lid release (see p. 57).
Close device lid.
symmetrically.
2. Restart device.
Press the short key continuously
during a short run centrifugation.
Check the settings.Check unhindered air circulation
through the air slots.
Thaw ice or switch off device and
allow it to cool down.
Page 55
Troubleshooting
Centrifuge 5430/5430 R
English (EN)

8.2 Error messages

If one of the following error messages appears, proceed as follows:
1. Remove fault (see Remedies).
2. Press open key to clear the error message.
3. If necessary, repeat centrifugation.
Some errors can have various causes. The actual cause is described in the message in the device display.
Symptom/message Cause Remedy
Note A Lid latch
Note B Imbalance
Note C Rotor detection
Centrifuge lid could not be locked.
Rotor loaded unsymmetrically.
Set g-force/speed too high, e.g. after a rotor change (see Automatic rotor detection on p. 32).
Try again to close centrifuge lid.
Load the rotor symmetrically and
balance it.
1. Check the g-force/speed.
2. Repeat the run.
55
Symptom/message Cause Remedy
Error 1 Rotor detection
Error 2 Electronics fault
Error 3 Speed control
Error 3 Speed control
Error 5 Lid latch
Error 6 Drive fault
Error 6 Drive fault
Error 7 Speed control
Error 8 Speed control
Error 9 to Error 14 Electronics fault.
Rotor not detected.
Check rotor.If this error message appears again,
test with a different rotor.
Electronics fault.
Error in speed measuring system.
Error in speed measuring system.
Prohibited opening of lid or lid switch is defective during a run.
Drive fault.
Switch centrifuge off and back on
again after > 20 s.
Insert and tighten rotor.
Wait for displayed time.
1. Wait for rotor to stop.
2. Open and close again the lid of the device.
3. Repeat the run.
Repeat the run.If this error message appears again,
switch centrifuge off and back on again after > 20 s.
Drive overheated.
Major deviation in the speed control. 1. Wait for rotor to stop.
•Drive fault.
• Rotor loose.
•Incorrect rotor.
Allow the drive to cool down for at
least 15 min.
2. Tighten rotor.
1. Wait for rotor to stop.
2. Tighten rotor.
3. Repeat the run.
Switch centrifuge off and back on
again after > 20 s.
Page 56
56
Troubleshooting
Centrifuge 5430/5430 R English (EN)
Symptom/message Cause Remedy
Error 16 to Error 17 Electronics fault
Error 18 Room Temp. of rotor chamber
(only 5430 R)
Error 18 Room Temp. of rotor chamber
(only 5430 R)
Error 22 Electronics fault
(only 5430 R)
Error 25 Mains/power failure
Error 26 Electronics fault
(only 5430 R)
Error 27 Electronics fault
(only 5430 R)
Error 28 Electronics fault
Error 30 Lid latch
Error 30 Lid latch
Error 30 Lid latch
Electronics fault. Switch centrifuge off and back on
again after > 20 s.
Temperature deviation from set value in the rotor chamber: ΔT > 16 °C.
Temperature deviation from set value in the rotor chamber: ΔT > 50 °C.
Electronics fault.
Mains/power failure during a run.
Electronics fault.
Electronics fault.
Electronics fault.
Centrifuge lid could not be locked.
Centrifuge lid could not be released.
Centrifuge lid has not been opened wide enough.
Allow the device to cool down and
repeat cycle.
Allow the device to cool down and
repeat cycle.
Switch centrifuge off and back on
again after > 20 s.
Check the power supply.
Switch centrifuge off and back on
again after > 20 s.
Switch centrifuge off and back on
again after > 20 s.
Press the open key.
Try again to close centrifuge lid.
Switch the device off and back on
again.
If the error occurs again:
1. Switch off the device.
2. Activate the emergency lid release (see Emergency release on p. 57).
Open the centrifuge lid wider by
hand.
Page 57
Troubleshooting
Centrifuge 5430/5430 R
English (EN)

8.3 Emergency release

If the centrifuge lid cannot be opened, you can activate the emergency release manually.
WARNING! Risk of injury from rotating rotor.
If the emergency release of the lid is operated, the rotor may continue rotating for several minutes.
Wait until the rotor stop before operating the emergency release.Check the monitoring glass in the centrifuge lid.
Use the rotor key delivered with the Centrifuge 5430 for the emergency release. The rotor key for the rotor FA-45-24-11-HS is not suitable for this purpose.
1. Pull the mains/power plug.
2. Carry out the following steps for the emergency release on both the left side and right side of the centrifuge (see Fig. 1 and Fig. 2).
3. Only 5430: Remove the plastic cover for the emergency release.
4. Insert the centrifuge rotor key in the rear hexagonal opening until a noticeable resistance is felt.
5. Slightly press and turn the rotor key counterclockwise five to ten revolutions, as depicted on the openings of the emergency release.
This will release the centrifuge lid.
6. Open the centrifuge lid.
7. Remove the rotor key and put the plastic covers back on (Centrifuge 5430).
57
Page 58
58
Transport, storage and disposal
Centrifuge 5430/5430 R English (EN)

9 Transport, storage and disposal

9.1 Transport

CAUTION! Risk of injury due to lifting and carrying heavy loads
The device is heavy. Lifting and carrying the device can lead to back injuries.
Transport and lift the device with an adequate number of helpers only.Use a transport aid to transport the device.
Remove the rotor from the centrifuge before transport.Use the original packaging for transport.
Air temperature Relative humidity Atmospheric pressure
General transport -25 °C – 60 °C 10 % – 75 % 30 kPa – 106 kPa Air freight -20 °C – 55 °C 10 % – 75 % 30 kPa – 106 kPa

9.2 Storage

Air temperature Relative humidity Atmospheric pressure
In transport packaging -25 °C – 55 °C 10 % – 75 % 70 kPa – 106 kPa Without transport
packaging
-5 °C – 45 °C 10 % – 75 % 70 kPa – 106 kPa

9.3 Disposal

In case the product is to be disposed of, the relevant legal regulations are to be observed.
Information on the disposal of electrical and electronic devices in the European Community:
Within the European Community, the disposal of electrical devices is regulated by national regulations based on EU Directive 2002/96/EC pertaining to waste electrical and electronic equipment (WEEE).
According to these regulations, any devices supplied after August 13, 2005, in the business-to-business sphere, to which this product is assigned, may no longer be disposed of in municipal or domestic waste. To document this, they have been marked with the following identification:
Because disposal regulations may differ from one country to another within the EU, please contact your supplier if necessary.
Page 59

10 Technical data

10.1 Power supply

Centrifuge 5430
Mains connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz 100 V, 50 to 60 Hz
Current consumption 3 A (230 V)
6 A (120 V)
7 A (100 V) Power consumption Maximum 475 W EMC: interference emission (radio
interference) EMC: noise immunity EN 61326 Overvoltage category II Protection class 1 Fuses 4.0 AT (230 V)
EN 61326 - category B
8.0 AT (120 V/100 V)
Technical data
Centrifuge 5430/5430 R
English (EN)
59
Centrifuge 5430 R
Mains connection 230 V, 50 to 60 Hz
120 V, 50 to 60 Hz
100 V, 50 to 60 Hz Current consumption 6 A (230 V)
12 A (120 V/ 100 V) Power consumption Maximum 1050 W EMC: Interference emission (radio
interference) EMC: Noise immunity EN 61326 Overvoltage category II Protection class 1 Fuses Thermal overcurrent protective switch 7 A (230 V)
EN 61326 – Class B (230 V/120 V)
EN 61326 – Class A (100 V)
Thermal overcurrent protective switch 15 A (120 V/100 V)

10.2 Ambient conditions

Ambience For indoor use only. Ambient temperature Centrifuge 5430: 4 to 40°C
Centrifuge 5430 R: 10 to 40°C Maximum relative humidity 75 %, non-condensing humidity Atmospheric pressure Use up to an altitude of 2000 m above MSL. Degree of pollution 2
Page 60
60
Technical data
Centrifuge 5430/5430 R English (EN)

10.3 Weight/dimensions

Centrifuge 5430
Dimensions Width: 335 mm (11.2 in.)
Depth: 415 mm (16.3 in.)
Height: 250 mm (9.84 in.) Weight without rotor 29 kg (63.9 lbs.) Noise level < 60 dB(A) *
Centrifuge 5430 R
Dimensions Width: 380 mm (15.0 in.)
Depth: 640 mm (25.2 in.)
Height: 296 mm (11.7 in.) Weight without rotor 56 kg (123.5 lbs.) Noise level < 60 dB(A) *
*) The noise level was measured frontally in a sound measuring room with accuracy class 1, 1 m from the device and at lab bench height.

10.4 Application parameters

Run time 30 s to 99:59 h, infinity ( )
Adjustable to 10 min in 0.5 min increments, then increments of
1 min Temperature of 5430 R -11°C to 40°C Relative centrifugal force (or rcf) 1 to 30.130 × g
Adjustable in increments of 10 × g up to 3,000 × g, then in
increments of 100 × g Speed 100 to 17,500 rpm
Adjustable in increments of 10 rpm up to 5,000 rpm, then in
increments of 100 rpm Maximum load 48 tubes with 2.0 mL each or 6 conical tubes with 50 mL each Maximum kinetic energy 10,000 nm Compulsory test log book No Permitted density of the material for
centrifuging at max. g-force/rpm and max. load
1.2 g/mL
Page 61
Technical data
Centrifuge 5430/5430 R
English (EN)
61
Rotor Lowest achievable temperature
-11°C set 23°C ambient temperature
60 min run time FA-45-48-11 < 0 °C 12 700 rpm F-45-48-11 < 0 °C 12 700 rpm FA-45-30-11 < 0 °C 14 000 rpm F-45-30-11 < 0 °C 14 000 rpm FA-45-24-11-HS < 5 °C ? rpm FA-45-24-11-Kit < 0 °C 13 200 rpm F-45-64-5-PCR < 0 °C 11 800 rpm F-45-18-17-Cryo < 0 °C 8 900 rpm FA-45-16-17 < 0 °C 14 200 rpm F-35-6-30 < 0 °C 7 8300 rpm A-2-MTP < 0 °C 4 680 rpm S-24-11-AT < 0 °C 12 700 rpm
Acceleration times and braking times of Centrifuge 5430/5430 R
Rotor Acceleration time/Deceleration time Without soft ramp
Acceleration time/Deceleration time With soft ramp
230 V 120 V 100 V
FA-45-48-11 F-45-48-11
FA-45-30-11 F-45-30-11
FA-45-24-11-Kit ≤ 15 / 16 s
FA-45-16-17 ≤ 20 / 20 s
S-24-11-AT ≤ 13 / 16 s
FA-45-24-11-HS ≤ 21 / 16 s
F-45-64-5-PCR ≤ 12 / 15 s
F-45-18-17-Cryo ≤ 8 / 11 s
F-35-6-30 ≤ 23 / 23 s
A-2-MTP ≤ 18 / 21 s
≤ 20 / 20 s
≤ 61 / 65 s (SOFT)
≤ 15 / 15 s
≤ 61 / 65 s (SOFT)
≤ 78 / 90 s (SOFT)
≤ 61 / 66 s (SOFT)
≤ 61 / 66 s (SOFT)
≤ 60 / 65 s (SOFT)
≤ 62 / 65 s (SOFT)
≤ 77 / 85 s (SOFT)
≤ 62 / 67 s (SOFT)
≤ 63 / 67 s (SOFT)
≤ 20 / 20 s ≤ 61 / 65 s (SOFT)
≤ 15 / 15 s ≤ 61 / 65 s (SOFT)
≤ 15 / 16 s ≤ 78 / 90 s (SOFT)
≤ 20 / 20 s ≤ 61 / 66 s (SOFT)
≤ 13 / 16 s ≤ 61 / 66 s (SOFT)
≤ 21 / 16 s ≤ 60 / 65 s (SOFT)
≤ 12 / 15 s ≤ 62 / 65 s (SOFT)
≤ 8 / 11 s ≤ 77 / 85 s (SOFT)
≤ 23 / 23 s ≤ 62 / 67 s (SOFT)
≤ 18 / 21 s ≤ 63 / 67 s (SOFT)
Speed to safely maintain 4°C sample temperature
4°C set 23 °C ambient temperature
≤ 27 / 20 s ≤ 61 / 65 s (SOFT)
≤ 20 / 15 s ≤ 61 / 65 s (SOFT)
≤ 20 / 16 s ≤ 78 / 90 s (SOFT)
≤ 30 / 20 s ≤ 61 / 66 s (SOFT)
≤ 16 / 16 s ≤ 61 / 66 s (SOFT)
≤ 30 / 16 s ≤ 60 / 65 s (SOFT)
≤ 15 / 15 s ≤ 62 / 65 s (SOFT)
≤ 8 / 11 s ≤ 77 / 85 s (SOFT)
≤ 27 / 27 s ≤ 62 / 67 s (SOFT)
≤ 18 / 21 s ≤ 63 / 67 s (SOFT)
Page 62
62
Ordering Information
Centrifuge 5430/5430 R English (EN)

11 Ordering Information

11.1 Centrifuge 5430

Order no. (International)
5427 000.011 022620533 230 V / 50 – 60 Hz – 022620584 120 V / 50 – 60 Hz, with US-plug
5427 000.216 022620525 230 V / 50 – 60 Hz – 022620509 120 V / 50 – 60 Hz, with US-plug
022620541 120 V / 50 – 60 Hz, with US-plug
022620568 120 V / 50 – 60 Hz, with US-plug
5427 000.615 022620540 230 V / 50 – 60 Hz – 022620596 120 V / 50 – 60 Hz, with US-plug
5427 000.410 - 230 V / 50 – 60 Hz – 022620511 120 V / 50 – 60 Hz, with US-plug
022620557 120 V / 50 – 60 Hz, with US-plug
022620572 120 V / 50 – 60 Hz, with US-plug
Order no. (North America)
Description
Centrifuge 5430, Keypad
without rotor
Centrifuge 5430, Keypad
with rotor FA-45-30-11 incl. rotor lid
Centrifuge 5430, Keypad
with rotor FA-45-24-11-Kit incl. rotor lid
Centrifuge 5430, Keypad
with rotor A-2-MTP incl. wind shield upper shell
Centrifuge 5430, Knobs
without rotor
Centrifuge 5430, Knobs
with rotor FA-45-30-11 incl. rotor lid
Centrifuge 5430, Knobs
with rotor FA-45-24-11-Kit incl. rotor lid
Centrifuge 5430, Knobs
with rotor A-2-MTP incl. wind shield upper shell

11.2 Centrifuge 5430 R

Order no. (International)
5428 000.210 022620678 230 V / 50 – 60 Hz – 022620667 120 V / 50 – 60 Hz, with US-plug
5428 000.015 022620612 230 V / 50 – 60 Hz – 022620601 120 V / 50 – 60 Hz, with US-plug
Order no. (North America)
Description
Centrifuge 5430 R, Keypad
without rotor
Centrifuge 5430 R, Keypad
with rotor FA-45-30-11 incl. rotor lid
Page 63
Ordering Information
Centrifuge 5430/5430 R
English (EN)
63
Order no. (International)
022620645 120 V / 50 – 60 Hz, with US-plug
5428 000.619 022620690 230 V / 50 – 60 Hz – 022620689 120 V / 50 – 60 Hz, with US-plug
5428 000.414 022620634 230 V / 50 – 60 Hz – 022620623 120 V / 50 – 60 Hz, with US-plug
022620656 120 V / 50 – 60 Hz, with US-plug
Order no. (North America)
Description
Centrifuge 5430 R, Keypad
with rotor A-2-MTP incl. wind shield upper shell
Centrifuge 5430 R, Knobs
without rotor
Centrifuge 5430 R, Knobs
with rotor FA-45-30-11 incl. rotor lid
Centrifuge 5430 R, Knobs
with rotor A-2-MTP incl. wind shield upper shell

11.3 Rotors, rotor lids and seals

11.3.1 Rotors with QuickLock rotor lid
Rotor FA-45-48-11
Order no. (International)
5427 754.008 5427754008 aerosol-tight, angle 45°, 48 places, max. tube diameter 11 mm,
5427 762.000 5427762000 aerosol-tight, aluminum
5409 718.002 5409718002 set of 5 pieces
Order no. (North America)
Description
Fixed-angle rotor FA-45-48-11
incl. rotor lid (aluminum)
Rotor lid for FA-45-48-11
Seal for rotor lid
FA-45-48-11 (5427 R/5430/5430 R)
Rotor FA-45-30-11
Order no. (International)
5427 753.001 5427753001 aerosol-tight, angle 45°, 30 places, max. tube diameter 11 mm,
5427 761.004 5427761004 aerosol-tight, aluminum
5820 762.004 5820762004 Set of 5 pieces
Order no. (North America)
Description
Fixed-angle FA-45-30-11
incl. rotor lid (aluminum)
Rotor lid for FA-45-30-11
Seal for rotor lid
FA-45-30-11 (5427 R/5430/5430 R)
Page 64
64
Ordering Information
Centrifuge 5430/5430 R English (EN)
Rotor FA-45-24-11 Kit
Order no. (International)
5427 752.005 5427752005 aerosol-tight, angle 45°, 24 places, max. tube diameter 11 mm,
5427 760.008 5427760008 aerosol-tight, aluminum
5820 767.006 5820767006 Set of 5 pieces
Rotor FA-45-16-17
Order no. (International)
5427 750.002 5427750002 aerosol-tight, angle 45°, 16 places, max. tube diameter 17 mm,
5427 751.009 5427751009 aerosol-tight, aluminum
5409 717.006 5409717006 Set of 5 pieces
Order no. (North America)
Order no. (North America)
Description
Fixed-angle rotor FA-45-24-11-Kit
incl. rotor lid (aluminum)
Rotor lid for FA-45-24-11-Kit
Seal for rotor lid
FA-45-24-11-Kit (5427 R/5430/5430 R), FA-45-48-11, FA-45-20-17 (5804/5804 R/5810/5810 R)
Description
Fixed-angle rotor FA-45-16-17
incl. rotor lid (aluminum)
Rotor lid for FA-45-16-17
Seal for rotor lid
FA-45-24-11 (5427 R), FA-45-16-17 (5430/5430 R)
Rotor S-24-11-AT
Order no. (International)
5427 757.007 5427757007 aerosol-tight, steel, angle 90°, 24 places, max. tube diameter
5427 758.003 5427758003 aerosol-tight, aluminum
5409 719.009 5409719009 set of 5 pieces
5409 721.003 5409721003 set of 2 pieces
Order no. (North America)
Description
Swing-bucket rotor S-24-11-AT
11 mm, incl. rotor lid (aluminum)
Rotor lid for S-24-11-AT
Seal for rotor lid
S-24-11-AT (5427 R/5430/5430 R)
Tube holder for S-24-11-AT
for 4 × 1,5 mL/2,0 mL Eppendorf tubes
Page 65
11.3.2 Rotors with rotor lid thread
Rotor F-45-48-11
Ordering Information
Centrifuge 5430/5430 R
English (EN)
65
Order no. (International)
5409 712.004 5409712004 aluminum, angle 45°, 48 places, max. tube diameter 11 mm,
5409 713.000 5409713000 Polypropylene
Rotor FA-45-30-11
Order no. (International)
5427 719.008 022654063 aerosol-tight, PTFE-coated, aluminum
Rotor F-45-30-11
Order no. (International)
5427 712.003 022654004 PTFE-coated, angle 45°, 30 places, max. tube diameter
5427 718.001 022654021 Polypropylene
Order no. (North America)
Order no. (North America)
Order no. (North America)
Description
Rotor F-45-48-11
incl. rotor lid (polypropylene)
Rotor lid for F-45-48-11
Description
Rotor lid
for F-45-30-11
Description
Rotor F-45-30-11
11 mm, incl. rotor lid (polypropylene)
Rotor lid
for F-45-30-11
Rotor FA-45-24-11-HS
Order no. (International)
5427 710.000 022654080 aerosol-tight, PTFE-coated, angle 45°, 24 places, max. tube
5427 711.007 022654101 aerosol-tight, PTFE-coated, aluminum
Order no. (North America)
Description
Rotor FA-45-24-11-HS
diameter 11 mm, incl. rotor lid (aluminum), incl. rotor key
Rotor lid
for FA-45-24-11-HS
Page 66
66
Ordering Information
Centrifuge 5430/5430 R English (EN)
Rotor FA-45-24-11-Kit
Order no. (International)
5427 704.000 022654144 aerosol-tight, aluminum
Rotor F-45-64-5-PCR
Order no. (International)
5427 714.006 022654209 angle 45°, 64 places, max. tube diameter 5 mm, incl. rotor lid
5427 720.006 022654225 aluminum
Rotor F-45-18-17-Cryo
Order no. (International)
5427 705.007 022654161 angle 45°, 18 places, max. tube diameter 17 mm, incl. rotor lid
5427 707.000 022654187 Polypropylene
Order no. (North America)
Order no. (North America)
Order no. (North America)
Description
Rotor lid
for FA-45-24-11-Kit
Description
Rotor F-45-64-5-PCR
(aluminum) and adapters
Rotor lid
for F-45-64-5-PCR
Description
Rotor F-45-18-17-Cryo
(polyproplene) and adapters
Rotor lid
for F-45-18-17-Cryo
Rotor F-35-6-30
Order no. (International)
5427 716.009 022654306 angle 35°, 6 places, max. tube diameter 30 mm, incl. rotor lid
5427 739.009 5427739009 angle 35°, 6 places, max. tube diameter 30 mm, incl. rotor lid
5427 715.002 022654322 aluminum
Order no. (North America)
Description
Rotor F-35-6-30
(aluminum) and adapters for 15/50 mL conical tubes
Rotor lid
for F-35-6-30
Page 67
11.3.3 Rotors with rotor lid for attaching
A-2-MTP rotor
Ordering Information
Centrifuge 5430/5430 R
English (EN)
67
Order no. (International)
5427 700.005 022634403 with 2 buckets and windshield upper shell
5427 722.009 022634420 Set of 2
5427 725.008 022654446 aluminum
The Aerosol tightness is tested and certified by the Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down (UK).
Order no. (North America)
Description
Rotor A-2-MTP
Spare MTP buckets for A-2-MTP
Wind shield upper shell for A-2-MTP

11.4 Accessories

11.4.1 Adapter
Order no. (International)
5425 715.005 022636260 for 1 PCR tube (0.2 mL, max. Ø 6 mm), set of 6 5425 717.008 022636243 for 1 sample tube (0.4 mL, max. Ø 6 mm), set of 6 5425 716.001 022636227 for 1 sample tube (0.5 mL, max. Ø 6 mm) or 1 Microtainer
5427 717.005 022654241 for PCR strips, set of 4 pieces
5702 752.002 022639498 for cryo tubes (max. Ø 13 mm) and sealable centrifuge tubes
5427 708.006 5427708006 for 1.5 mL HPLC vials, 18 pieces
5427 740.007 5427740007 13 × 65-89, set of 2 5427 741.003 5427741003 13 × 90-110, set of 2 5427 746.005 5427746005 for Eppendorf Tubes 5.0 mL, set of 2 5427 726.004 022654365 for 15 mL conical tubes, set of 2 5427 732.004 022654512 for 7 - 15 mL round-bottom tubes and blood collection tubes,
5427 735.003 022654538 for 9 - 15 mL round-bottom tubes and blood collection tubes,
5427 742.000 5427742000 13 × 65-89, set of 2
Order no. (North America)
Description
Adapter
used in FA-45-48-11, F-45-48-11, FA-45-30-11, F-45-30-11, FA-45-24-11-HS and FA-45-24-11-Kit
(0.6 mL, max. Ø 8 mm), set of 6
Adapter
used in F-45-64-5-PCR
Adapter
used in F-45-18-17-Cryo
(max. Ø 12.2 mm), max. length 50 mm, set of 6
Adapter
used in F-35-6-30, small tube bore
set of 2
set of 2
Adapter
used in F-35-6-30, large tube bore
Page 68
68
Ordering Information
Centrifuge 5430/5430 R English (EN)
Order no. (International)
5427 743.006 5427743006 13 × 90-110, set of 2 5427 747.001 5427747001 for Eppendorf Tubes 5.0 mL, set of 2 5427 727.000 022654349 for 50 mL conical tubes, set of 2 5427 723.005 022654331 for Centriplus centrifugal filter units, set of 6 5427 734.007 022654524 for 7 - 15 mL round-bottom tubes and blood collection tubes,
5427 738.002 022654545 for 9 - 15 mL round-bottom tubes and blood collection tubes,
5427 736.000 022654556 for 20 - 30 mL round-bottom tubes, set of 2 5427 737.006 022654567 for 50 mL round-bottom tubes, set of 2
5825 711.009 022638947 for 96-well PCR plates, set of 2 5825 713.001 022638955 for 384-well PCR plates, set of 2 5825 706.005 022638963 CombiSlide Adapter, set of 2
Order no. (North America)
Description
set of 2
set of 2
Adapter
used in A-2-MTP
11.4.2 Other accessories
Order no. (International)
5416 301.001 022634305 Standard 5427 730.001 5427730001 for rotor FA-45-24-11-HS
5427 728.007 5427728007
5810 350.050 022634330 Tube 20 mL
5428 850.418 022680452
Order no. (North America)
Description
Rotor key
Rotor removal tool for Rotor F-35-6-30
Pivot grease
Tray for condensation water
11.4.3 Fuses for Centrifuge 5430
Order no. (International)
5301 850.249 022654403 4 A (230 V), 2 pieces 5427 850.341 022654381 8.0 AT UL (120 V/100 V), 2 pieces
Order no. (North America)
Description
Fuse
Page 69
Índice
Centrifuge 5430/5430 R
Español (ES)
Manual de instrucciones Centrifuge 5430/5430 Rveren pág.Fig.Tab.en pág.

Español (ES)Manual de instrucciones

Índice
1 Instrucciones de empleo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
1.1 Utilización de estas instrucciones. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
2 Descripción del producto. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
2.1 Ilustración general . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
2.2 Rotores . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 73
3 Instrucciones generales de seguridad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.1 Uso de acuerdo con lo previsto. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.2 Requerimiento para el usuario . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.3 Límites de aplicación . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.3.1 Aclaración de la directriz ATEX (94/9/EG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.3.2 Máxima vida útil de los accesorios . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 78
3.4 Indicaciones relativas a la responsabilidad del producto . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
3.5 Peligros durante el uso previsto . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
3.5.1 Daños personales o al equipo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
3.5.2 Manipulación errónea de la centrífuga . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
3.5.3 Manipulación errónea de los rotores . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
3.5.4 Utilización extrema de los tubos de centrifugado . . . . . . . . . . . . . . . . . . . . . . . . . . . . 83
3.5.5 Centrifugado hermético a los aerosoles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
3.6 Indicaciones de seguridad en el equipo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 84
69
4 Instalación. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
4.1 Seleccionar ubicación . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
4.2 Instalación del aparato . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
5 Manejo. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
5.1 Visión general de los elementos de control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 87
5.2 Configuración de la centrífuga . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.2.1 Ajuste del idioma del menú . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.3 Preparación del centrifugado . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.3.1 Conexión de la centrífuga . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.3.2 Inserción del rotor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89
5.3.3 Detección automática del rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.3.4 Detección manual del rotor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.3.5 Carga del rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90
5.3.6 Cierre de la tapa de la centrífuga . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 92
5.4 Refrigeración (sólo 5430 R) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.4.1 Ajuste de temperatura. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.4.2 Indicación de temperatura . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.4.3 Supervisión de temperatura . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.4.4 FastTemp. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 93
5.4.5 FastTemp pro . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 94
5.4.6 Refrigeración continua . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 95
5.5 Centrifugado . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
5.5.1 Centrifugado con ajuste de tiempo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 96
5.5.2 Otras posibilidades de centrifugado: ciclo permanente y short spin . . . . . . . . . . . . . . 96
5.5.3 Extracción del rotor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
5.6 Modo standby . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Page 70
70
Índice
Centrifuge 5430/5430 R Español (ES)
5.7 Advertencias de uso de los rotores . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
5.7.1 Rotor F-35-6-30: herramienta de extracción . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 98
5.7.2 Rotor A-2-MTP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
5.7.3 Rotor FA-45-24-11-HS: uso de la llave especial del rotor. . . . . . . . . . . . . . . . . . . . . . . 99
5.7.4 QuickLock . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 100
6 Mantenimiento . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
6.1 Preparación de la limpieza / desinfección . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
6.2 Realización de la limpieza / desinfección . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 101
6.2.1 Limpieza y desinfección del equipo . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 102
6.2.2 Limpieza y desinfección del rotor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 103
6.3 Indicaciones de cuidado adicionales para la Centrifuge 5430 R . . . . . . . . . . . . . . . . . . . . . . . 103
6.4 Rotura de vidrio. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
6.5 Fusibles . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 104
6.6 Descontaminación antes del envío . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 105
7 Solución de problemas. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
7.1 Errores generales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 106
7.2 Mensajes de error . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 107
7.3 Desbloqueo de emergencia. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 109
8 Transporte, almacenaje y eliminación. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
8.1 Transporte. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
8.2 Almacenamiento . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
8.3 Eliminación . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 110
Page 71
Instrucciones de empleo
1
2
3
4
5
678
Centrifuge 5430/5430 R
Español (ES)

1 Instrucciones de empleo

1.1 Utilización de estas instrucciones

Lea detenidamente este manual de instrucciones antes de poner en funcionamiento el equipo por
primera vez. Tenga en cuenta, en caso necesario, los manuales de instrucciones de los accesorios.
Encontrará una descripción detallada del equipo en la versión en inglés y en alemán del manual de
instrucciones.
Igualmente, puede encontrar las características técnicas e información de pedidos en la versión inglesa y alemana.
Este manual de instrucciones es parte del producto. Consérvelo en un lugar accesible.Incluya siempre este manual de instrucciones cuando entregue el equipo a terceros.En caso de pérdida del manual de instrucciones, solicite otro. Encontrará una versión actual de este
manual en nuestra página de internet www.eppendorf.com (Norteamérica).
(Internacional) o www.eppendorfna.com
71
La Centrifuge 5430/5430 R está disponible en dos variantes: con teclado o selectores giratorios. En este manual de instrucciones se describe por regla general el manejo de la variante con teclado de membrana. Este manual, sin embargo, también es válido para la variante con mandos giratorios.

2 Descripción del producto

2.1 Ilustración general

Abb. 2-1: Vista anterior y posterior de la Centrifuge 5430
Imag. 2-1: Vista anterior y posterior de la Centrifuge 5430
Page 72
72
1
2
3
4
5
1
9
6
8
7
Descripción del producto
Centrifuge 5430/5430 R Español (ES)
Abb. 2-2: Vista anterior y posterior de la Centrifuge 5430 R
Imag. 2-2: Vista anterior y posterior de la Centrifuge 5430 R
1 Desbloqueo de emergencia
En ambos lados del aparato (ver Desbloqueo de emergencia en pág. 109).
2 Tapa de la centrífuga
3Panel de control
Teclas y selectores giratorios (según variante) para el manejo de la centrífuga(ver en pág. 87).
4 Mirilla
Control visual de la parada del rotor y/o posibilidad de controlar las revoluciones mediante estroboscopio.
5Indicación
Representación de los parámetros de centrifugación y ajustes de aparato actuales (ver en pág. 87).
6 Alimentación eléctrica
Toma hembra para la conexión del cable de red suministrado. Sólo 5430: Debajo de la toma hembra se encuentra el portafusibles (ver Fusibles en pág. 104).
7 Interruptor de alimentación
Interruptor para conectar y desconectar el aparato. Posición 0: el aparato está desconectado. Posición I: el aparato está conectado.
8 Placa de características
9Interfaz USB
Interfaz para análisis de fallos y actualizaciones de software a través del servicio técnico.
Page 73
Descripción del producto
Centrifuge 5430/5430 R
Español (ES)

2.2 Rotores

Puede utilizar la Centrifuge 5430/5430 R con los siguientes rotores. Antes de utilizar tubos de muestras, observe las especificaciones respecto a la resistencia al centrifugado (fuerza de la gravedad máx.) recomendadas por el fabricante.
73
Rotor FA-45-48-11
con tapa del rotor hermética a los aerosoles QuickLock
Rotor F-45-48-11
con tapa del rotor de polipropileno
Rotor FA-45-30-11
con tapa del rotor hermética a los aerosoles QuickLock
Capacidad máx. Fuerza de la gra-
vedad (FCR) / n.° de revoluciones (rpm) máx. sin adaptador
Tiempo de arranque/parada (SOFT): con rampa suave
48 tubos de reacción de 1,5/2,0 mL. Con adaptadores:
• Tubos PCR de 0,2 mL
• Tubos de reacción de 0,4 mL
• Tubos de reacción
Anillo exterior:
18.213 × g Anillo interior:
16.048 × g /
12.700 rpm ≤ 20 s/20 s
≤ 61 s/65 s (soft)
de 0,5 mL
• Tubos Microtainer de 0,6 mL
48 tubos de reacción de 1,5/2,0 mL. Con adaptadores:
• Tubos PCR de 0,2 mL
• Tubos de reacción de 0,4 mL
• Tubos de reacción
Anillo exterior:
18.213 × g Anillo interior:
16.048 × g /
12.700 rpm ≤ 20 s/20 s
≤ 61 s/65 s (soft)
de 0,5 mL
• Tubos Microtainer de 0,6 mL
30 tubos de reacción de 1,5/2,0 mL. Con adaptadores:
• Tubos PCR de 0,2 mL
20.817 × g /
14.000 rpm ≤ 15 s/15 s
≤ 61 s/65 s (soft)
• Tubos de reacción de 0,4 mL
• Tubos de reacción de 0,5 mL
• Tubos Microtainer de 0,6 mL
Carga
Indicaciones máx. por cada orificio del
(1)
rotor
(2)
3,75 g • Tapa de rotor
hermética a los aerosoles
(3)
QuickLock (aluminio).
3,75 g
3,75 g • Tapa de rotor her-
mética a los aero-
(3)
soles
QuickLock
(aluminio).
•Recubrimiento de PTFE (especial­mente resistente a las sustancias quí­micas), identifica­ción: coated.
•Uso de columnas de centrifugado posible, mejor con rotor FA-45-24-11-Kit.
Page 74
74
Descripción del producto
Centrifuge 5430/5430 R Español (ES)
Rotor F-45-30-11
con tapa del rotor de polipropileno
Rotor FA-45-24-11-Kit
con tapa del rotor hermética a los aerosoles QuickLock
Rotor FA-45-16-17
con tapa del rotor hermética a los aerosoles QuickLock
Capacidad máx. Fuerza de la gra-
vedad (FCR) / n.° de revoluciones (rpm) máx. sin adaptador
Tiempo de arranque/parada (SOFT): con rampa suave
30 tubos de reacción de 1,5/2,0 mL. Con adaptadores:
• Tubos PCR de 0,2 mL
20.817 × g /
14.000 rpm ≤ 15 s/15 s
≤ 61 s/65 s (soft)
• Tubos de reacción de 0,4 mL
• Tubos de reacción de 0,5 mL
• Tubos Microtainer de 0,6 mL
24 columnas de centrifugado o tubos de reacción de 1,5/2,0 mL. Con adaptadores:
19.090 × g /
13.200 rpm ≤ 15 s/16 s
≤ 78 s/90 s (soft)
• Tubos PCR de 0,2 mL
• Tubos de reacción de 0,4 mL
• Tubos de reacción de 0,5 mL
• Tubos Microtainer de 0,6 mL
16 tubos de reacción de 5,0 mL.
21.191 × g /
14.200 rpm ≤ 20 s/20 s
≤ 61 s/66 s (soft)
Carga
Indicaciones máx. por cada orificio del
(1)
rotor
(2)
3,75 g
3,75 g • Tapa de rotor
hermética a los aerosoles
(3)
QuickLock (aluminio).
•Borde especialmente alto para columnas de centrifugado habituales.
Tenga en cuenta la indicación sobre el centrifugado con tapas de tubo abiertas (ver
Rotores de ángulo fijo en pág. 90)
9,5 g • Tapa de rotor
hermética a los aerosoles
(3)
QuickLock (aluminio).
Page 75
Descripción del producto
Centrifuge 5430/5430 R
Español (ES)
75
Rotor S-24-11-AT
con tapa del rotor hermética a los aerosoles QuickLock
Rotor FA-45-24-11-HS
Capacidad máx. Fuerza de la gra-
vedad (FCR) / n.° de revoluciones (rpm) máx. sin adaptador
Tiempo de arranque/parada (SOFT): con rampa suave
24 tubos de reacción de 1,5/ 2,0 mL.
Este rotor está previsto exclusivamente para tubos
16.049 × g /
12.700 rpm ≤ 13 s/16 s
≤ 61 s/66 s (soft) de ensayo de 1,5/2,0 mL. Las columnas (spin columns) y los adaptadores, incluyendo los respectivos tubos de 0,2 mL, 0,4 mL, 0,5 mL y 0,6 mL, no se deben utilizar en este rotor.
24 tubos de reacción de 1,5/ 2,0 mL. Con adaptadores:
• Tubos PCR de 0,2 mL
30.130 × g /
17.500 rpm
≤ 21 s/16 s
≤ 61 s/65 s (soft)
• Tubos de reacción de 0,4 mL
• Tubos de reacción de 0,5 mL
• Tubos Microtainer de 0,6 mL
Carga
Indicaciones máx. por cada orificio del
(1)
rotor
(2)
3,75 g • Tapa de rotor
hermética a los aerosoles
(3)
QuickLock (aluminio).
• El rotor debe funcionar siempre con la tapa cerrada.
3,75 g • Tapa de rotor her-
mética a los aero-
(3)
soles
(aluminio).
• Fuerza de la grave­dad (FCR) / n.° de revoluciones (rpm) máx. (30.130 × g /
17.500 rpm) sólo con recipientes autorizados por el fabricante.
•Recubrimiento de PTFE (especial­mente resistente a las sustancias quí­micas), identifica­ción: coated.
•Uso de columnas de centrifugado posible, mejor con rotor FA-45-24-11-Kit.
• Apriete y afloje el rotor sólo con una llave de rotor espe­cial para el rotor FA-45-24-11-HS (ver en pág. 98).
Page 76
76
Descripción del producto
Centrifuge 5430/5430 R Español (ES)
Rotor F-45-64-5-PCR
Rotor F-45-18-17-Cryo
Rotor F-35-6-30
Capacidad máx. Fuerza de la gra-
vedad (FCR) / n.° de revoluciones (rpm) máx. sin adaptador
Tiempo de arranque/parada (SOFT): con rampa suave
64 tubos PCR (0,2 mL) o 8 tiras PCR de 5 u 8 tubos, con los respectivos
13.543 × g /
11.800 rpm ≤ 12 s/15 s
≤ 62 s/65 s (soft)
adaptadores adjuntos.
18 criotubos o 18 tubos de centrifugado con cierre,
8.324 × g /
8.900 rpm ≤ 8 s/11 s
≤ 77 s/85 s (soft)
máx. Ø: 16,9 mm. Con adaptadores adjuntos: máx. Ø: 13,4 mm, máx. longitud del tubo: 50 mm.
6 tubos cónicos de 50 mL con o sin borde de pie o 6 tubos cónicos de 15 mL
7.745 × g /
7.830 rpm ≤ 23 s/23 s
≤ 62 s/67 s (soft)
con los respectivos adaptadores adjuntos o 6 unidades de filtro con adaptadores para centrífuga Centriplus.
Carga
Indicaciones máx. por cada orificio del
(1)
rotor
(2)
3,4 g (sin adaptador)
8,7 g • Ajuste de la fuerza
de la gravedad / n.° de revoluciones en incrementos de 10 × g o 10 rpm.
110 g • Retire o introduzca
el rotor sólo con la herramienta que se suministra.
•Centrifugado de tubos de base circular con sistemas de extracción de sangre posible con adaptadores adicionales (ver pág. 128).
Page 77
Descripción del producto
Fixed-angle
rotor
Aerosol-tight
version
Angle of
bores
Number of
bores
of
bores (mm)
AT
24 11
Aerosol-tight
version
Number of
bores
of
bores (mm)
Swing-bucket
rotor
Centrifuge 5430/5430 R
Español (ES)
77
Rotor A-2-MTP
Capacidad máx. Fuerza de la gra-
vedad (FCR) / n.° de revoluciones (rpm) máx. sin adaptador
Tiempo de arranque/parada (SOFT): con rampa suave
Dos cestillos para colocar:
• Microplacas
• Placas de cultivo celular
•Placas PCR
2.204 × g /
4.680 rpm ≤ 18 s/21 s
≤ 63 s/67 s (soft)
• Placas Deepwell (altura máx. 29 mm)
• Portaobjetos (con adaptador CombiSlide)
Carga máx. por cada orificio del
(1)
rotor
(2)
170 g (por cestillo)
Indicaciones
•Centrifugado de placas PCR sólo con los adaptadores correspondientes.
• Máx. altura de carga permisible: 29 mm.
Sólo 5430 R: enfriamiento efectivo mediante centrifugado sin tapa de la cámara de aire (ver Rotores
basculantes en pág. 91).
(1) Carga máxima por cada orificio del rotor para adaptador + tubo + contenido.
(2) Según la norma DIN 58 970 (variante del equipo: 230 V, 120 V y 100 V, de 50 a 60 Hz).
(3) Hermético a los aerosoles comprobado y certificado por el Centre of Emergency Preparedness and Response, Health Protection Agency, Porton Down (UK), (véanse los certificados al final de este manual de instrucciones).
En los rotores y las tapas de rotor identificados con "revestido" ("coated") pueden surgir cambios de color debido al proceso de producción. Estos cambios no tienen ninguna repercusión sobre la durabilidad o sobre la resistencia a los productos químicos.
Page 78
78
Instrucciones generales de seguridad
Centrifuge 5430/5430 R Español (ES)

3 Instrucciones generales de seguridad

3.1 Uso de acuerdo con lo previsto

La Centrifuge 5430/5430 R está destinada exclusivamente para su uso en interiores y sirve para la separación de soluciones acuosas y suspensiones de diferente densidad en tubos de muestras autorizados.

3.2 Requerimiento para el usuario

Este aparato debe ser manejado exclusivamente por personal técnico debidamente formado. El personal debe haber leído atentamente el manual de instrucciones y estar familiarizado con el funcionamiento del aparato.

3.3 Límites de aplicación

3.3.1 Aclaración de la directriz ATEX (94/9/EG)
¡PELIGRO! Peligro de explosión.
No utilice el equipo en salas en donde se trabaje con sustancias explosivas.No procese con este equipo sustancias explosivas o que reaccionen bruscamente.No procese con este equipo sustancias que puedan crear una atmósfera explosiva.
La Centrifuge 5430/5430 R no está indicada para su utilización en una atmósfera potencialmente explosiva debido a su construcción y a las condiciones en el interior del equipo.
El equipo sólo puede utilizarse en un ambiente seguro, es decir en el ambiente abierto de un laboratorio ventilado o una campana extractora. No está permitido el uso de sustancias que puedan originar una atmósfera potencialmente explosiva. La decisión definitiva respecto a los riesgos relacionados con el uso de tales sustancias es responsabilidad del usuario.
3.3.2 Máxima vida útil de los accesorios
¡ADVERTENCIA! Peligro de lesión a causa de accesorios dañados química o mecánicamente.
Tanto arañazos como grietas pequeñas pueden provocar graves daños en los materiales internos.
Proteja todas las piezas de los accesorios contra los daños mecánicos.Controle la presencia de daños en los accesorios antes de cada uso. Cambie los accesorios
dañados.
No utilice rotores, tapas de rotor o cestillos con huellas de corrosión o daños mecánicos
(p.ej. torceduras).
No utilice ningún accesorio, cuya máxima vida útil haya caducado.Preste atención a no causar arañazos al insertar los cestillos y los rotores.
Page 79
Instrucciones generales de seguridad
Centrifuge 5430/5430 R
Español (ES)
¡ATENCIÓN! Peligro de lesiones a causa de tapas o cubiertas de rotor químicamente deterioradas.
Las tapas de rotor o cubiertas transparentes de PC, PP o PEI pueden perder resistencia por el efecto de disolventes orgánicos (p.ej. fenol, cloroformo).
Si las tapas de rotor o cubiertas entran en contacto con disolventes orgánicos, límpielas
inmediatamente.
Controle con regularidad la presencia de daños o fisuras en las tapas de rotor o las
cubiertas.
Sustituya inmediatamente las tapas de rotor o cubiertas con fisuras o decoloraciones
lechosas.
Los rotores siguientes, incluidos los frascos, cestillos y tapa del rotor correspondientes, tienen una vida útil máxima de los años indicados en la tabla a partir de la primera puesta en marcha o el número de ciclos indicados en la tabla (según lo que tenga lugar primero).
79
Rotor Máxima vida útil a partir de la puesta en marcha
FA-45-48-11 (con QuickLock) 100.000 ciclos mecánicos 10 años F-45-48-11 (con tapa del rotor con rosca) 100.000 ciclos mecánicos 10 años FA-45-30-11 (con QuickLock) 100.000 ciclos mecánicos 10 años FA-45-24-11-Kit (con QuickLock) 100.000 ciclos mecánicos 10 años FA-45-16-17 (con QuickLock) 100.000 ciclos mecánicos 10 años S-24-11-AT (con QuickLock) 100.000 ciclos mecánicos 10 años A-2-MTP
Incluidos los cestillos y la cubierta superior de la tapa del rotor correspondientes
Tapa del rotor QuickLock - 3 años Tapas de rotor y cubiertas de policarbonato (PC),
polipropileno (PP) o polieterimida (PEI) Adaptador de plástico - 1 año
Los otros rotores y tapas de rotor de esta centrífuga no tienen una vida útil limitada, siempre y cuando se cumplan las condiciones siguientes:
•Uso correcto
• Mantenimiento recomendado
• Estado libre de daños.
100.000 ciclos mecánicos 7 años
-3 años
La fecha de fabricación está grabada en los rotores como 03/10 (= marzo 2010) o en la parte interior de la tapa del rotor como un reloj .
Para garantizar la hermeticidad a los aerosoles se aplica lo siguiente:
• Cambie las tapas de rotor y cubiertas herméticas a los aerosoles después de 50 ciclos de autoclave.
• Cambie el anillo de obturación de la tapa de rotor QuickLock después de 50 ciclos de autoclave.
Page 80
80
Instrucciones generales de seguridad
Centrifuge 5430/5430 R Español (ES)

3.4 Indicaciones relativas a la responsabilidad del producto

Los siguientes casos pueden reducir la protección prevista del aparato. La entidad explotadora asume entonces la responsabilidad por las lesiones personales y daños materiales producidos:
• El equipo no es utilizado de acuerdo con lo especificado en el manual de instrucciones.
• El equipo no es utilizado de acuerdo con lo prescrito.
• El equipo se utiliza con accesorios o consumibles no recomendados por Eppendorf.
• El equipo es mantenido o reparado por personas que no han sido autorizadas por Eppendorf AG.
• El usuario realiza modificaciones no autorizadas en el equipo.

3.5 Peligros durante el uso previsto

Antes de utilizar la Centrifuge 5430/5430 R, lea primero el manual de instrucciones y tenga en cuenta las siguientes indicaciones generales de seguridad.
3.5.1 Daños personales o al equipo
¡ADVERTENCIA! Descarga eléctrica por daños en el equipo o cable de alimentación.
Solo encienda el equipo si éste y el cable de alimentación no presentan ningún daño.Utilice únicamente equipos que hayan sido instalados o reparados correctamente.En caso de peligro, separe el equipo de la red eléctrica extrayendo el enchufe de
alimentación del equipo o con ayuda del dispositivo de separación previsto para ello (p.ej., el interruptor de emergencia del laboratorio).
¡ADVERTENCIA! Descargas de tensión mortales en el interior del equipo.
Cerciórese de que la carcasa siempre esté cerrada y no presente daños, de modo que no
sea posible tocar piezas en el interior del equipo por descuido.
No desmonte el revestimiento del equipo.No deje que ningún líquido entre al interior de la carcasa.Solo deje que personal de mantenimiento autorizado por Eppendorf abra el equipo.
¡ADVERTENCIA! Peligro a causa de un suministro de energía eléctrica equivocado.
Solo conecte el equipo a fuentes de alimentación que cumplan los requisitos eléctricos
especificados en la placa de características.
Solo utilice enchufes con toma de tierra y un cable de red adecuado.
¡ADVERTENCIA! Daños para la salud a causa de líquidos infecciosos y gérmenes patógenos.
Siempre tenga en cuenta las disposiciones nacionales, el nivel de seguridad biológica de
su laboratorio, así como las fichas de datos de seguridad y las instrucciones de uso del fabricante cuando maneje líquidos infecciosos y gérmenes patógenos.
Utilice sistemas de cierre hermético a los aerosoles al centrifugar estas sustancias.Cuando trabaje con gérmenes patógenos que pertenezcan a un grupo de riesgo superior,
debe disponer de más de una impermeabilización biológica hermética a los aerosoles.
Póngase su equipo de protección personal.Encontrará amplias prescripciones respecto al manejo de gérmenes o material biológico
del grupo de riesgo II o superior en el "Manual de Bioseguridad en el Laboratorio" (fuente: World Health Organisation, Laboratory Biosafety Manual, en la versión actualmente vigente).
Page 81
Instrucciones generales de seguridad
Centrifuge 5430/5430 R
Español (ES)
¡ADVERTENCIA! Peligro de lesiones al abrir o cerrar la tapa de la centrífuga.
Puede aplastarse los dedos al abrir o cerrar la tapa de la centrífuga.
Al abrir y cerrar la tapa de la centrífuga, no ponga la mano entre la tapa y el equipo o en el
mecanismo de cierre de la tapa.
Abra siempre la tapa de la centrífuga por completo para que ésta no pueda cerrarse de
golpe.
¡ATENCIÓN! Riesgos de seguridad debido a accesorios y piezas de recambio equivocados.
Los accesorios y piezas de recambio no recomendados por Eppendorf merman la seguridad, el funcionamiento y la precisión del equipo. Por daños producidos por accesorios y piezas de recambio no recomendados por Eppendorf o por un uso incorrecto, Eppendorf queda eximido de cualquier responsabilidad o garantía.
Utilice exclusivamente accesorios y piezas de recambio recomendados por Eppendorf.
¡AVISO! Daños en el equipo a causa de líquidos derramados.
1. Apague el equipo.
2. Desconecte el equipo del suministro de corriente.
3. Lleve a cabo una limpieza cuidadosa del equipo y sus accesorios según las indicaciones de limpieza y desinfección del manual de instrucciones.
4. Si debe utilizarse otro método de limpieza o desinfección, consulte a Eppendorf AG para asegurarse de que el método previsto no dañe el equipo.
81
¡AVISO! Daños en los componentes electrónicos debido a la formación de condensación.
Después de transportar el equipo de un entorno frío a un entorno más caliente se puede formar condensado en el equipo.
Después de emplazar el equipo, debe esperar por lo menos 4 h. Una vez transcurrido este
tiempo, puede conectar el equipo a la red eléctrica.
3.5.2 Manipulación errónea de la centrífuga
¡AVISO! Daños por golpes o movimientos del equipo en funcionamiento.
Un rotor que golpea contra la pared de la cámara produce daños considerables en el equipo y en el rotor.
No mueva o golpee el equipo mientras éste está en funcionamiento.
3.5.3 Manipulación errónea de los rotores
¡ADVERTENCIA! Peligro de lesión por rotores y tapa del rotor incorrectamente fijados.
Centrifugue solo cuando el rotor y la tapa del rotor estén bien fijados.Si al arrancar la centrífuga se producen ruidos no habituales, puede que el rotor o la tapa
del rotor no estén fijados correctamente. Finalice el centrifugado de inmediato pulsando la tecla start/stop.
Page 82
82
Instrucciones generales de seguridad
Centrifuge 5430/5430 R Español (ES)
¡ATENCIÓN! Peligro de lesión por carga asimétrica de un rotor.
Cargue los rotores simétricamente con tubos o placas y cestillos.Cargue los adaptadores solo con los tubos o placas adecuados.Utilice siempre tubos o placas del mismo tipo (peso, material/densidad y volumen).Compruebe la carga simétrica tarando los tubos o placas y el adaptador utilizados con una
báscula. Durante el funcionamiento, el equipo detecta desequilibrios automáticamente y finaliza el ciclo de inmediato emitiendo un mensaje de error y una señal acústica. Compruebe la carga, tare los tubos y arranque de nuevo el centrifugado.
¡ATENCIÓN! Peligro de lesión por sobrecarga del rotor.
La Centrifuge 5430/5430 R está diseñada para el centrifugado de sustancias con una densidad máxima de 1,2 g/mL a máxima velocidad y a máximo volumen de llenado o carga.
No exceda la carga máxima del rotor.
¡AVISO! Deterioro de los rotores a causa de productos químicos agresivos.
Los rotores son componentes de alta calidad que resisten cargas extremas. Esta estabilidad puede verse perjudicada por productos químicos agresivos.
Evite el uso de productos químicos agresivos; a esta clase de productos pertenecen, entre
otros, álcalis fuertes y débiles, ácidos fuertes, soluciones con iones de mercurio, cobre u
otros metales pesados, hidrocarburos halogenados, soluciones salinas concentradas y
fenol.
En caso de contaminación por sustancias químicas agresivas, limpie inmediatamente el
rotor con un producto de limpieza neutro. Esto es especialmente aplicable a los orificios
del rotor.
En los rotores identificados con "revestido" ("coated") pueden surgir cambios de color
debido al proceso de producción. Éstos no tienen ninguna repercusión sobre la
durabilidad o la resistencia a los productos químicos.
¡AVISO! El rotor se puede caer en caso de una manipulación errónea.
El rotor puede caerse si se utilizan los cestillos como asas.
Antes de colocar o extraer un rotor basculante, extraiga los cestillos.Cargue siempre la cruz del rotor con ambas manos.
¡ATENCIÓN! Peligro de lesiones a causa de tapas o cubiertas de rotor químicamente deterioradas.
Las tapas de rotor o cubiertas transparentes de PC, PP o PEI pueden perder resistencia por el efecto de disolventes orgánicos (p.ej. fenol, cloroformo).
Si las tapas de rotor o cubiertas entran en contacto con disolventes orgánicos, límpielas
inmediatamente.
Controle con regularidad la presencia de daños o fisuras en las tapas de rotor o las
cubiertas.
Sustituya inmediatamente las tapas de rotor o cubiertas con fisuras o decoloraciones
lechosas.
Page 83
3.5.4 Utilización extrema de los tubos de centrifugado
¡ATENCIÓN! Peligro de lesión por tubos sobrecargados.
Observe los valores límite especificados por el fabricante de los tubos sobre la capacidad
de carga de los mismos.
Utilice exclusivamente tubos autorizados por el fabricante para la fuerza de la gravedad
(FCR) deseada.
¡AVISO! Peligro por tubos dañados.
No deben utilizarse tubos dañados. La consecuencia pueden ser daños adicionales en el equipo y en los accesorios así como la pérdida de muestras.
Inspeccione visualmente todos los tubos respecto a posibles daños antes de su uso.
¡AVISO! Peligro por tapas de tubo abiertas.
Las tapas de tubo abiertas pueden romperse en el centrifugado y dañar tanto el rotor como la centrífuga.
Cierre cuidadosamente todas las tapas de tubo antes del centrifugado.
Excepción: consulte la advertencia para el centrifugado de columnas de centrifugado en el rotor FA-45-24-11-Kit.
Instrucciones generales de seguridad
Centrifuge 5430/5430 R
Español (ES)
83
¡AVISO! Daños de los tubos de plástico por disolventes orgánicos.
En caso de utilización de disolventes orgánicos (p.ej. fenol, cloroformo) se reduce la resistencia de los tubos de plástico, de forma que se pueden dañar los tubos.
Observe las indicaciones del fabricante sobre la resistencia química de los tubos.
¡AVISO! Los tubos de muestras se calientan.
En las centrífugas no refrigeradas, dependiendo de la duración del ciclo, la fuerza de la gravedad (FCR)/número de revoluciones y la temperatura ambiente, la temperatura en la cámara del rotor, el rotor y la muestra puede aumentar por encima de 40 °C.
Tenga en cuenta que en este caso la resistencia al centrifugado de los tubos de muestras
se reduce.
Observe la resistencia a la temperatura de las muestras.
Page 84
84
Instrucciones generales de seguridad
Centrifuge 5430/5430 R Español (ES)
3.5.5 Centrifugado hermético a los aerosoles
¡ADVERTENCIA! Riesgos para la salud debido a una hermeticidad a los aerosoles limitada por una combinación incorrecta de rotor/tapa del rotor.
El centrifugado hermético a los aerosoles solo está garantizado en caso de utilizar los rotores y tapas de rotor previstos para ello. Las denominaciones de los rotores de ángulo fijo herméticos a los aerosoles siempre empiezan con FA. Los rotores y las tapas de rotor herméticos a los aerosoles de esta centrífuga están marcados adicionalmente con un anillo rojo en el rotor y un tornillo de fijación rojo para la tapa del rotor. Los rotores basculantes herméticos a los aerosoles tienen la marca AT (aerosol tight).
Para el centrifugado hermético a los aerosoles en la centrífuga prevista, utilice siempre al
mismo tiempo rotores y tapas de rotor con marca de hermético a los aerosoles. La
indicación sobre qué tipo de rotores y tapas de rotor herméticos a los aerosoles pueden
utilizarse en una determinada centrífuga la encontrará en el rotor y a partir de la fecha de
producción de octubre de 2003 en la parte superior de la tapa del rotor.
Utilice tapas de rotor herméticas a los aerosoles exclusivamente en combinación con
rotores que estén indicados en la tapa del rotor.
¡ADVERTENCIA! Riesgos para la salud debido a una hermeticidad a los aerosoles limitada por una aplicación incorrecta.
Las autoclaves, las cargas mecánicas y la contaminación por sustancias químicas u otras soluciones agresivas pueden perjudicar la hermeticidad a los aerosoles de los rotores y las tapas de rotor.
Controle antes de cada uso la integridad de los anillos de obturación de las tapas o
cubiertas herméticas a los aerosoles.
Utilice tapas de rotor o cubiertas herméticas a los aerosoles solo con anillos de obturación
limpios y en buen estado.
Unte ligeramente la rosca del tornillo de la tapa de rotor después de cada autoclave
correcta (121 °C, 20 min) con grasa para gorrones (n.° de pedido. int. 5810 350.050,
Norteamérica 022634330).
Cambie las tapas de rotor y cubiertas herméticas a los aerosoles después de 50 ciclos de
autoclave.
Con las tapas de rotor QuickLock, el anillo de obturación debe cambiarse después de 50
ciclos de autoclave.
Nunca almacene cerrados los rotores herméticos a los aerosoles o los frascos.

3.6 Indicaciones de seguridad en el equipo

Representación Significado Lugar
Lugar peligroso. 5430: lado posterior del equipo
5430 R: lado derecho del equipo
ATE NCIÓN
Apriete siempre el rotor con la llave de rotor adjunta.
ATE NCIÓN
Cierre todos los tubos y utilice una tapa de rotor.
Lado superior del equipo, debajo de la tapa de la centrífuga.
Lado superior del equipo, debajo de la tapa de la centrífuga.
Page 85

4 Instalación

4.1 Seleccionar ubicación

¡AVISO! En caso de error se pueden dañar objetos que se encuentran junto al equipo.
Respete las recomendaciones de la norma EN 61010-2-020, dejando durante el
funcionamiento un espacio de seguridad de 30 cm alrededor del equipo.
Retire todos los materiales y objetos que se encuentren en esta área.
¡AVISO! Daños por sobrecalentamiento.
No coloque el equipo cerca de fuentes de calor (p. ej., calefacción, armario de desecación).No exponga el equipo a la radiación solar directa.Asegúrese de que la circulación de aire no se obstaculice. Mantenga una distancia mínima
de 30 cm de todas las rendijas de ventilación.
Instalación
Centrifuge 5430/5430 R
Español (ES)
85
Determine la ubicación del equipo según los siguientes criterios:
• Alimentación eléctrica apropiada conforme a la placa de características (230 V/120 V/100 V).
• Mesa de laboratorio estable, horizontal y libre de resonancias.
• Entorno bien ventilado y protegido de la radiación solar directa para evitar un calentamiento adicional.

4.2 Instalación del aparato

Requisitos El equipo está colocado sobre una mesa de laboratorio apropiada.
¡AVISO! Daños en los componentes electrónicos debido a la formación de condensación.
Después de transportar el equipo de un entorno frío a un entorno más caliente se puede formar condensado en el equipo.
Después de emplazar el equipo, debe esperar por lo menos 4 h. Una vez transcurrido este
tiempo, puede conectar el equipo a la red eléctrica.
¡AVISO! Centrifuge 5430 R: daño del compresor tras un transporte inadecuado.
No conecte la centrífuga hasta 4 horas después de haberla colocado en su respectivo sitio.
Ejecute los siguientes pasos en el orden indicado:
1. Deje que el equipo se caliente a temperatura ambiente durante mínimo 3 horas (5430) o 4 horas (5430 R) para evitar que los componentes electrónicos resulten dañados por la formación de condensación y para evitar daños en el compresor (solo 5430 R).
2. Verifique que la tensión y frecuencia de red coinciden con los requisitos especificados en la placa de características del equipo.
3. Conecte la centrífuga a la red y enciéndala con el interruptor de alimentación que se encuentra en el lado posterior (Centrifuge 5430) o en el lado derecho del equipo (Centrifuge 5430 R).
• La tecla standby se ilumina de color verde.
•El indicador está activo.
Sólo 5430: la tapa se abre automáticamente.
Page 86
86
Instalación
Centrifuge 5430/5430 R Español (ES)
4. Sólo 5430: retire el dispositivo de seguridad para el transporte del eje del motor.
5. En el caso de que se incluya un rotor en el alcance de suministro, suéltelo con ayuda de la llave de rotor adjunta y retírelo.
6. Sólo 5430: retire los dispositivos de seguridad para el transporte del anillo de guía de aire.
7. Compruebe todas las piezas respecto a posibles daños de transporte. Póngase en contacto con su proveedor en caso de que exista algún daño.
Guarde el material de embalaje y los bloqueadores de transporte para un transporte posterior o almacenaje. Observe también las indicaciones para el transporte (ver en pág. 110).
Page 87

5 Manejo

1
2
3
4
5
6
7891011
13 12

5.1 Visión general de los elementos de control

Abb. 5-1: Panel de mando e indicador de la Centrifuge 5430/5430 R (variante con teclado).
Manejo
Centrifuge 5430/5430 R
Español (ES)
87
Imag. 5-1: Panel de mando e indicador de la Centrifuge 5430/5430 R (variante con teclado).
1 Seleccionar el programa
Pulsación breve: cargar los parámetros del
centrifugado memorizados. Pulsación larga (> 2 s): guardar los parámetros de centrifugado actuales .
2 Centrifugado short spin
(ver en pág. 96)
3Desbloquear la tapa
4 Iniciar y detener el centrifugado
5Activar/desactivar el modo standby
La tecla se ilumina de color verde: la centrífuga está lista para el funcionamiento.
8 Ajustar la temperatura (sólo 5430 R)
9 Ajustar la duración de centrifugado
Según la variante del equipo, se trata de una tecla o de un selector giratorio.
10 Iniciar el proceso de atemperado FastTemp
(sólo 5430 R)
11 Conmutación de la velocidad de centrifugado
indicada (rpm/FCR)
12 Llamar y seleccionar parámetros de menú
13 Navegar en el menú
La tecla se ilumina de color rojo: modo standby activo .
6Indicación
7 Ajustar la velocidad de centrifugado
Según la variante del equipo, se trata de una tecla o de un selector giratorio.
Page 88
88
1
2 3 4 5 7
6
8
10
9
111213
Manejo
Centrifuge 5430/5430 R Español (ES)
Abb. 5-2: Indicación de Centrifug e 5430/5430 R
Imag. 5-2: Indicación de Centrifuge 5430/5430 R
1 Número de programa (si está activado)
2 Estado de la función At set rpm
: inicio del ciclo a partir del 95% de la fuerza de la gravedad (FCR) o del número de revoluciones predefinidos.
: inicio del ciclo inmediato.
3 Estado del bloqueo de teclas
: los parámetros de centrifugado no se pueden
cambiar inintencionadamente.
: el teclado no está bloqueado.
4 Estado del altavoz
: conectado.
: desconectado.
5 Estado de la centrífuga
: tapa de centrífuga desbloqueada.
: tapa de centrífuga bloqueada.
(parpadea): centrifugado en marcha.
6 Programación del ciclo de atemperación (sólo
5430 R)
: FastTemp pro está activada, es decir, el tiempo de inicio y la temperatura del ciclo de atemperación están programados .
8Rampa suave
: arranque y frenado lentos del rotor. Ningún símbolo: Arranque y frenado rápidos del rotor.
9 Indicación estándar
10 Indicación ampliada (si está activa)
11 Fuerza de la gravedad / número de
revoluciones
12 Temperatura (sólo 5430 R)
13 Duración del centrifugado
7Hora
Page 89

5.2 Configuración de la centrífuga

5.2.1 Ajuste del idioma del menú
Acceda a la selección del idioma del menú del modo siguiente:
1. Abra el menú
2. Seleccione Ajustes (Settings).
3. Confirme la selección.
4. SeleccioneIdioma (Language).
5. Confirme la selección.
Manejo
Centrifuge 5430/5430 R
Español (ES)
89
6. Seleccione English, Deutsch, Francais o Español.
7.
8. Pulse la tecla varias veces para salir del menú.
Confirme la selección. Delante del idioma seleccionado aparece una señal. El ajuste tiene efecto inmediato.

5.3 Preparación del centrifugado

5.3.1 Conexión de la centrífuga
1. Conmute la centrífuga con el interruptor de red o la tecla standby . Sólo 5430: después de la conexión con el interruptor de alimentación, la tapa de la centrífuga se abre
automáticamente.
2. Abra la tapa de la centrífuga pulsando la tecla open. Los ajustes de los parámetros del último ciclo se visualizan.
5.3.2 Inserción del rotor
Rotores basculantes: retire los cestillos antes de insertar o extraer el rotor. Sujete la cruz
del rotor con las dos manos.
Rotor F-35-6-30: para insertar o extraer el rotor, debe utilizar exclusivamente la
herramienta auxiliar de extracción adjunta.
1. Coloque el rotor verticalmente en el eje del motor.
2. Inserte la llave de rotor suministrada en la tuerca del rotor. Rotor FA-45-24-11-HS: Utilice una llave de rotor especial.
3. Gire la llave de rotor en el sentido de las agujas del reloj hasta que la tuerca del rotor esté totalmente apretada.
Page 90
90
Manejo
Centrifuge 5430/5430 R Español (ES)
5.3.3 Detección automática del rotor
Compruebe, después de un cambio de rotor, el valor de la fuerza de la gravedad (FCR)/número de
revoluciones (rpm) determinado y ajústelo si fuera necesario.
5.3.4 Detección manual del rotor
¡ATENCIÓN! Peligro de lesión en caso de giro manual del rotor.
Preste especial atención al girar un rotor basculante de no pillarse los dedos o quedarse
enganchado en los cestillos.
Para iniciar la detección manual del rotor, gire el rotor con la mano en el sentido contrario a las agujas
del reloj.
• En el indicador se muestra el nombre del rotor.
• La fuerza de la gravedad (FCR)/número de revoluciones (rpm) se limita automáticamente al valor máximo del rotor.
5.3.5 Carga del rotor
¡ATENCIÓN! Peligro de lesión por carga asimétrica de un rotor.
Cargue los rotores simétricamente con tubos o placas y cestillos.Cargue los adaptadores solo con los tubos o placas adecuados.Utilice siempre tubos o placas del mismo tipo (peso, material/densidad y volumen).Compruebe la carga simétrica tarando los tubos o placas y el adaptador utilizados con una
báscula. Durante el funcionamiento, el equipo detecta desequilibrios automáticamente y finaliza el ciclo de inmediato emitiendo un mensaje de error y una señal acústica. Compruebe la carga, tare los tubos y arranque de nuevo el centrifugado.
Durante el funcionamiento, el equipo detecta desequilibrios automáticamente y finaliza el ciclo de inmediato emitiendo un mensaje de error y una señal acústica.
Compruebe la carga, tare los tubos e inicie el ciclo de nuevo.
5.3.5.1 Rotores de ángulo fijo
Tapa del rotor
• Los rotores de ángulo fijo sólo se deben utilizar con la tapa de rotor adecuada. Esto se indica mediante la inscripción del nombre del respectivo rotor tanto en el rotor como en la tapa correspondiente.
• Para realizar un centrifugado hermético a los aerosoles, debe instalarse un rotor hermético a los aerosoles (identificación: anillo rojo) y la tapa de rotor hermética a los aerosoles correspondiente (identificación: aerosol-tight y tornillo rojo de la tapa).
Para cargar el rotor, proceda del siguiente modo:
1. Compruebe la carga máxima (adaptador, tubo y contenido) permisible para cada orificio del rotor. La indicación de ésta se encuentra en el rotor y en el manual de instrucciones(ver Rotores en pág. 73).
2. Cargue los rotores y los adaptadores sólo con los tubos previstos para ello.
Page 91
Manejo
Centrifuge 5430/5430 R
Español (ES)
3. Introduzca los tubos por parejas en posición opuesta en los orificios del rotor. Para una carga simétrica, los tubos en posición opuesta tienen que ser del mismo tipo y contener la misma cantidad de sustancia.
Para que la diferencia de peso entre los tubos de muestras llenos sea la mínima posible, es recomendable pesarlos con una báscula. A través de ello se protege el accionamiento y se reducen los ruidos de funcionamiento.
4. Coloque y fije la tapa del rotor.
Columnas de centrifugado
Durante el centrifugado de columnas de centrifugado en el rotor FA-45-24-11-Kit se pueden dejar abiertas las tapas de estos tubos. No obstante, esto sólo está permitido con tubos previstos para ello por sus fabricantes. Para un centrifugado seguro, tiene que apoyar las tapas abiertas en el borde del rotor. Preste atención a que las tapas no sobresalgan verticalmente del borde del rotor y coloque entonces la tapa de rotor correspondiente.
5.3.5.2 Rotores basculantes
Requisitos
• Una combinación de rotor, cestillos y adaptadores autorizada por Eppendorf.
• Tubos y placas adecuados y comprobados.
91
¡AVISO! Un llenado en exceso de las placas conlleva reboses.
Durante el ciclo, los meniscos dentro de los tubos en los bordes de las placas se encuentran en posición oblicua. Esto está condicionado por la fuerza centrífuga y es algo inevitable.
Llene los pocillos de las placas como máximo hasta 2/3 del máximo volumen de llenado.
Para cargar el rotor, proceda del siguiente modo:
1. Compruebe si las ranuras del cestillo están limpias y engráselas moderadamente con grasa para gorrones (n.° de pedido int.: 5810 350.050/Norteamérica: 022634330).
Las ranuras y gorrones sucios impiden la oscilación uniforme de los cestillos.
2. Enganche los cestillos en el rotor. Todas las posiciones del rotor deben estar ocupadas con cestillos.
3. Compruebe que todos los cestillos estén bien colgados y que puedan oscilar libremente.
4. Al utilizar por primera vez un tipo de placas o tubos, realice una prueba manual de carga y oscilación.
5. Compruebe la carga máxima por cestillo (adaptador, tubo o placa y contenido) y la altura de carga. La indicación de ésta se encuentra en el rotor y en el manual de instrucciones(ver Rotores en pág. 73).
Page 92
92
Manejo
Centrifuge 5430/5430 R Español (ES)
6. Cargue los cestillos simétricamente.
Abb. 5-3: Carga simétrica pero incompleta de los frascos. Los gorrones de cada frasco tienen que estar expuestos a una carga homogénea.
Imag. 5-3: Carga simétrica pero incompleta de los frascos. Los gorrones de cada frasco tienen que estar
expuestos a una carga homogénea.
El equipamiento de placas mostrado en la parte derecha es erróneo, ya que los cestillos no pueden oscilar correctamente así.
Las placas tienen un ligero juego en los cestillos.
7. Compruebe la carga del cestillo.
Realice una prueba de centrifugado breve con un número de revoluciones bajo (p.ej.
1000 rpm) si utiliza un tipo de placas o tubos por primera vez.
Sólo 5430 R: al utilizar el A-2-MTP, realice el centrifugado sin la cubierta superior de la tapa del rotor, para garantizar un atemperado preciso de las muestras. Tenga en cuenta que en este caso el ruido producido por el proceso de centrifugado es un poco más fuerte. Esto no es aplicable al S-24-11-AT. El S-24-11-AT debe funcionar siempre con la tapa del rotor cerrada.
5.3.6 Cierre de la tapa de la centrífuga
1. Compruebe la fijación correcta del rotor y de su tapa.
2. Presione la tapa de la centrífuga hacia abajo hasta que el bloqueo de la tapa enganche y la tapa se cierre automáticamente.
La centrífuga se cierra automáticamente.
La tecla open parpadea en color azul. En el indicador se muestra el símbolo .
Page 93
Manejo
Centrifuge 5430/5430 R
Español (ES)

5.4 Refrigeración (sólo 5430 R)

5.4.1 Ajuste de temperatura
Ajuste la temperatura con las teclas de flecha temp de -11 a +40 °C.
También puede modificar la temperatura durante el centrifugado.
Cuando la temperatura ambiente sea elevada, puede que se emitan ruidos breves del ventilador hasta alcanzar la temperatura deseada. Estos sirven para indicar una capacidad refrigerante elevada.
5.4.2 Indicación de temperatura
En caso de parada del rotor: Temperatura ajustada Durante el centrifugado Temperatura real
En la indicación ampliada se muestra la temperatura ajustada.
5.4.3 Supervisión de temperatura
Después de alcanzar la temperatura ajustada, la centrífuga reacciona de la siguiente manera a desviaciones de temperatura que ocurren durante el centrifugado:
93
Desviación del valor teórico Acción
± 3 °C Las temperaturas parpadean en el indicador. ± 5 °C Tono de advertencia periódico e indicación de Error 18.
El centrifugado finaliza automáticamente.
5.4.4 FastTemp
Con esta función se inicia directamente un ciclo de atemperado sin muestras con un número de revoluciones específico del rotor y la temperatura, para llevar rápidamente la cámara del rotor con rotor, cestillos y adaptadores, a la temperatura ajustada anteriormente.
La función FastTemp pro para la programación del ciclo de atemperado con tiempos de inicio definidos, se describe en la sección siguiente.
Requisitos
• La centrífuga está encendida.
• El rotor y la tapa del rotor están montados correctamente.
• La tapa de la centrífuga está cerrada.
• La temperatura y la fuerza de la gravedad (FCR)/número de revoluciones (rpm) están definidos para el centrifugado posterior (ver Centrifugado en pág. 96).
1. Pulse la tecla FastTemp. En el indicador aparecen FastTemp, el tiempo restante del ciclo de atemperado, así como la temperatura
actual y la fuerza de la gravedad (FCR)/número de revoluciones (rpm).
El ciclo de atemperado finaliza automáticamente al alcanzar la temperatura ajustada. Un tono de aviso se escucha periódicamente.
2. Pulse la tecla start/stop para finalizar el ciclo de atemperado antes de tiempo.
Después de alcanzar la temperatura teórica y después de finalizar el ciclo de atemperado, la centrífuga mantiene la temperatura de la cámara del rotor con la tapa de la centrífuga cerrada a la temperatura teórica ajustada si ésta se encuentra por debajo de la temperatura ambiente. Independientemente de la temperatura ajustada, durante esta refrigeración continua no se baja de 4 °C para evitar que la cámara del rotor se congele.
Page 94
94
Manejo
Centrifuge 5430/5430 R Español (ES)
La centrífuga finaliza el ciclo automáticamente, en cuanto el rotor y los cestillos estén completamente atemperados. Por esta razón, es posible que se produzca un retardo de aprox. 30 min entre la indicación de la temperatura ajustada y el fin automático del ciclo de atemperado.
Sólo 5430 R: al utilizar el A-2-MTP, realice el centrifugado sin la cubierta superior de la tapa del rotor para garantizar un atemperado preciso de las muestras. Tenga en cuenta que en este caso el ruido producido por el proceso de centrifugado es un poco más fuerte. Esto no es aplicable al S-24-11-AT. El S-24-11-AT debe funcionar siempre con la tapa del rotor cerrada.
5.4.5 FastTemp pro
Puede ajustar el ciclo de atemperado FastTemp (ver en pág. 93) descrito previamente para que empiece de forma automática a una hora fija. En este caso, tiene dos opciones:
Una vez El ciclo de atemperado comienza a la hora fijada una sola vez. Varias veces El ciclo de atemperado comienza a la hora fijada al siguiente día indicado. Esto se repite
ilimitadamente con cada día indicado de la semana.
5.4.5.1 Programación de la hora de inicio
1. Seleccione en el menú del equipo FastTemp pro.
2. Seleccione Una vez o Varias veces.
Esta selección se muestra sólo si la función FastTemp pro no ha sido activada. En caso contrario, puede modificar o borrar el tiempo de inicio programado.
3. Sólo con varias veces: Active/desactive días de la semana con menu/enter, seleccione seguir y confirme
con menu/enter.
4. Introduzca la fecha y la hora para el inicio de uno o varios ciclos de atemperado así como la
temperatura ajustada y confírmelo con menu/enter. Los ajustes actuales se visualizan en una visión de conjunto.
5. Modifique los ajustes nuevamente o guárdelos.
6. Salga del menú.
FastTemp pro está activado. En el indicador aparece el símbolo siempre y cuando quede pendiente el inicio automático de un ciclo de atemperado. En el modo standby se visualiza
FastTemp pro.
• El ciclo de atemperado FastTemp (ver en pág. 93)comienza automáticamente a la hora fijada.
• Después de un ciclo de atemperado programado una vez, el símbolo desaparece. Con varios ciclos de atemperado programados, la función FastTemp pro permanece activa hasta que usted la desactive. Seleccione para ello FastTemp pro en el menú del equipo y borre los ajustes.
Page 95
Manejo
Centrifuge 5430/5430 R
Español (ES)
5.4.5.2 Preparación de la centrífuga
Cerciórese de que a la hora de inicio ajustada la centrífuga esté encendida o se encuentre en el modo
standby, el rotor y la tapa del rotor estén montados correctamente y la tapa de la centrífuga esté cerrada.
5.4.5.3 Inicio automático del ciclo de atemperado
1. Si la centrífuga se encuentra en el modo standby, ésta conmuta 1 minuto antes de la hora de inicio ajustada al modo de servicio.
2. El ciclo de atemperado FastTemp (ver FastTemp en pág. 93)comienza a la hora de inicio. En el indicador se muestra FastTemp pro.
Un arranque automático del ciclo de atemperado no es posible durante el centrifugado.
5.4.6 Refrigeración continua
Cuando el rotor está detenido, la cámara del rotor se mantiene a la temperatura ajustada, siempre y cuando se den las siguientes condiciones:
• La centrífuga está encendida.
• La tapa de la centrífuga está cerrada.
• La temperatura ajustada es inferior a la temperatura ambiente.
• La centrífuga no se encuentra en el modo standby.
95
Durante la refrigeración continua se aplica lo siguiente:
• La temperatura ajustada y real se muestran alternativamente.
• Independientemente de la temperatura ajustada, no se baja de 4 °C para evitar así que la cámara del rotor se congele y se produzca demasiada condensación en el interior del equipo.
• Como el rotor no gira, la adaptación de la temperatura sucede con más lentitud.
Para finalizar la refrigeración continua, abra la tapa de la centrífuga o presione la tecla standby.
Si la centrífuga no es utilizada durante más de 8 horas, la refrigeración continua se desconecta automáticamente. El equipo conmuta entonces al modo standby. Esto protege contra la formación de hielo en la cámara del rotor y contra la formación aumentada de condensación en el interior del equipo. Con FastTemp puede volver a alcanzar la temperatura deseada rápidamente (ver en pág. 93).
Puede conmutar la refrigeración continua a funcionamiento ilimitado por cuenta propia. Para ello, active en el menú del equipo bajo Refrigeración continua la entrada .
Page 96
96
Manejo
Centrifuge 5430/5430 R Español (ES)

5.5 Centrifugado

La condición para cada variante de centrífuga aquí descrita es la preparación (ver Preparación del centrifugado en pág. 89) mencionada anteriormente.
Sólo 5430 R: tenga en cuenta también las advertencias para el enfriamiento (ver Refrigeración (sólo 5430 R) en pág. 93).
5.5.1 Centrifugado con ajuste de tiempo
Ejecute los siguientes pasos en el orden indicado:
1. Ajuste la duración del ciclo con las teclas de flecha time.
2. Sólo 5430 R: ajuste con las teclas de flecha temp la temperatura
3. Con las teclas de flecha speed ajuste la fuerza de la gravedad (FCR)/número de revoluciones.
4. Pulse start/stop para iniciar el centrifugado.
Durante el centrifugado
• En el indicador parpadea mientras el rotor esté en funcionamiento.
• El tiempo restante del ciclo se indica en minutos. El último minuto se cuenta hacia atrás en
segundos.
Sólo 5430 R: se indica la temperatura actual.
• Se indica la fuerza de la gravedad (FCR) y/o el número de revoluciones actuales del rotor.
• Las teclas de acceso rápido, las teclas , open y short, así como todos los elementos del menú
relacionados directamente con el centrifugado están bloqueados durante el centrifugado.
Finalización del centrifugado
• Una vez transcurrido el tiempo ajustado, la centrífuga se detiene automáticamente. Durante el
proceso de frenado se indica el tiempo de duración del centrifugado de forma parpadeante. En cuanto el rotor se haya detenido por completo, se escucha una señal acústica.
Sólo 5430: la tapa de la centrífuga se abre automáticamente. En el indicador se muestra el símbolo
.
Sólo 5430 R: la tapa de la centrífuga permanece cerrada para conservar la temperatura de las
muestras. Puede abrirla pulsando la tecla parpadeante open.
5. Extraiga el material de centrifugado.
• Durante el ciclo puede modificar el tiempo de funcionamiento total, la temperatura (sólo 5430 R), el número de revoluciones y la indicación de rpm/FCR. Los parámetros nuevos se aceptan inmediatamente. Tenga en cuenta que el tiempo de ejecución total nuevo más breve que se puede ajustar es el tiempo ya transcurrido más 2 minutos.
• También puede acabar el centrifugado antes de que se termine el tiempo establecido pulsando la tecla start/stop.
5.5.2 Otras posibilidades de centrifugado: ciclo permanente y short spin
Para un ciclo permanente, ajuste la duración del ciclo con las teclas de flecha time por encima de 99:59
h o por debajo de 30 s.
En la indicación de tiempo aparece el símbolo ∞.
Puede iniciar un short spin (ciclo breve) con fuerza de la gravedad (FCR)/número de revoluciones
máximo o actual pulsando la tecla short, siempre que la tapa de la centrífuga esté cerrada. Para detenerlo, suelte la tecla.
Page 97
Manejo
Centrifuge 5430/5430 R
Español (ES)
5.5.3 Extracción del rotor
Rotores basculantes: retire los cestillos antes de insertar o extraer el rotor. Sujete la cruz
del rotor con las dos manos.
Rotor F-35-6-30: para insertar o extraer el rotor, debe utilizar exclusivamente la
herramienta auxiliar de extracción adjunta.
1. Gire la tuerca del rotor con la llave del rotor suministrada en el sentido contrario a las agujas del reloj. Rotor FA-45-24-11-HS: utilice una llave de rotor especial.
2. Extraiga el rotor sacándolo verticalmente hacia arriba.
3. Sólo 5430 R: apague la centrífuga después del uso y vacíe el recipiente colector de agua condensada (en el lado izquierdo del equipo). Deje la tapa de la centrífuga completamente abierta y asegúrese de que ésta no pueda cerrarse sola de golpe.

5.6 Modo standby

La centrífuga conmuta automáticamente del estado operativo al modo standby si se dan las siguientes condiciones:
• La centrífuga no ha sido utilizada por el tiempo ajustado en el menú del equipo (de 1 a 60 min).
Sólo 5430 R: la tapa de la centrífuga está abierta.
97
En el modo standby se aplica lo siguiente:
• La tecla standby se ilumina de color rojo.
Sólo 5430 R: la cámara del rotor no se refrigera (ver Refrigeración continua en pág. 95).
En el estado operativo se aplica lo siguiente:
• Se visualizan los parámetros de centrifugado.
• La tecla standby se ilumina de color verde.
Sólo 5430 R: la cámara del rotor se refrigera cuando la tapa de la centrífuga está cerrada (ver Refrigeración continua en pág. 95).
Puede conmutar en cualquier momento entre el modo standby y el estado operativo del equipo pulsando la tecla standby, excepto durante un centrifugado.
Page 98
98
1. 2.
1. 2.
Manejo
Centrifuge 5430/5430 R Español (ES)

5.7 Advertencias de uso de los rotores

5.7.1 Rotor F-35-6-30: herramienta de extracción
5.7.1.1 Cambio del rotor
Requisitos La tuerca del rotor está suelta.
Utilice la herramienta de extracción para colocar el rotor en la centrífuga y volverlo a sacar.
1. Inserte la herramienta de extracción con la parte estrecha en la rosca del rotor.
2. Apriete la herramienta de extracción con aprox. 3 giros en el sentido de las agujas del reloj.
3. Sujete la herramienta de extracción y cambie el rotor.
4. Suelte la herramienta de extracción girando en el sentido contrario a las agujas del reloj y retírela.
5. Colocación del rotor: apriete el rotor con la llave de rotor incluida (ver Inserción del rotor en pág. 89).
5.7.1.2 Desmontaje de la tapa del rotor
Utilice la herramienta de extracción para aflojar un tornillo de la tapa del rotor apretado fuertemente.
1. Inserte la herramienta de extracción con el lado ancho en el tornillo de la tapa del rotor.
2. Suelte el tornillo de la tapa del rotor girando la herramienta de extracción en el sentido contrario a las agujas del reloj.
Page 99
Manejo
Centrifuge 5430/5430 R
Español (ES)
5.7.2 Rotor A-2-MTP
5.7.2.1 Cambio del rotor
¡AVISO! El rotor se puede caer en caso de una manipulación errónea.
El rotor puede caerse si se utilizan los cestillos como asas.
Antes de colocar o extraer un rotor basculante, extraiga los cestillos.Cargue siempre la cruz del rotor con ambas manos.
5.7.2.2 Colocación y retirada de la cubierta superior de la tapa del rotor
La cubierta superior de la tapa del rotor sirve para reducir el ruido.
1. Antes del primer uso, retire la etiqueta del pomo de la cubierta superior de la tapa del rotor.
2. Coloque la cubierta superior de la tapa del rotor y gírela máx. 1/4 de giro hasta que se asiente en el cubo del rotor.
El indicador de cierre debe sobresalir del pomo de modo que la marca roja se pueda ver bien:
99
El asiento aflojado de la cubierta superior es intencionado y sirve para el auto-centrado.
3. Tire del pomo de la cubierta superior de la tapa del rotor para quitarla.
Sólo 5430 R: al utilizar el A-2-MTP, realice el centrifugado sin la cubierta superior de la tapa del rotor para garantizar un atemperado preciso y rápido de las muestras. Tenga en cuenta que en este caso el ruido producido por el proceso de centrifugado es un poco más fuerte. Esto no es aplicable al S-24-11-AT. El S-24-11-AT debe funcionar siempre con la tapa del rotor cerrada.
5.7.3 Rotor FA-45-24-11-HS: uso de la llave especial del rotor
5.7.3.1 Fijación del rotor
1. Inserte la llave del rotor para el FA-45-24-11-HS en la tuerca del rotor.
2. Gire la llave del rotor en el sentido de las agujas del reloj hasta el tope ('clic'). El rotor se ha fijado correctamente.
5.7.3.2 Desmontaje del rotor
Gire la tuerca del rotor con la llave del rotor para el FA-45-24-11-HS en el sentido contrario a las agujas
del reloj.
Con la llave del rotor para el FA-45-24-11-HS sólo puede apretar y aflojar este rotor. Para los otros rotores descritos en este manual de instrucciones, utilice la llave del rotor que acompañe a la Centrifuge 5430/5430 R.
Page 100
100
Manejo
Centrifuge 5430/5430 R Español (ES)
5.7.4 QuickLock
Los rotores herméticos a los aerosoles FA-45-48-11, FA-45-30-11, FA-45-24-11-Kit, FA-45-16-17 y S-24-11-AT tienen un cierre rápido de la tapa del rotor (QuickLock).
Los rotores FA-45-30-11 y FA-45-24-11-Kit han sido retocados y, de momento, sólo están disponibles con un cierre rápido (QuickLock) en lugar de una tapa de rotor con rosca.
Cierre de la tapa del rotor
1. Compruebe el asiento correcto del anillo de obturación exterior en la ranura.
2. Coloque la tapa del rotor perpendicularmente sobre el rotor.
3. Para cerrar el rotor, gire el tornillo rojo de la tapa del rotor en el sentido de las agujas del reloj hasta el tope, cuando se escuche un clic.
Sólo después de haber escuchado el clic, el rotor está cerrado correctamente
Si el sistema de cierre no se acciona bien, unte ligeramente las clavijas del tornillo de la tapa del rotor con grasa para gorrones.
Loading...