Endress+Hauser FMX 570 Operating Instructions Manual

silometer
FMX 570 Level Measurement
Operating Instructions
BA 119F/00/en/06.03 016338-1000 Software Version 1.x
Hauser+Endress
The Power of Know How
Level Measurement at a Glance
*Can be performed in reverse order
Function Matrix Action
1 Reset transmitter V9H5 Enter 671: »+« and »-« keys, changes digit
Press »E« to register entry
- Omit if commissioned as in Section 4.1
2 »Empty« calibration* V0H1 ●Fill vessel 040% full (probe covered)
Enter level in %, m, ft, etc. Press »E« to register entry
3 »Full« calibration* V0H2 ●Fill vessel 60100% full
Enter level in %, m, ft, etc. Press »E« to register entry
4 0/4 mA signal V0H3
V0H5
V0H6
Enter 0 for 020 mA signal, 1 for 420 mA signal
Press »E« to register entry
Enter level for 0/4 mA signal (if not 0)
Press »E« to register entry
Enter level for 20 mA signal (if not 100)
Press »E« to register entry
Silometer FMX 570 Measurement at a Glance
Silometer FMX 570
Selects vertical matrix position
Selects horizontal matrix position
Select position V0H0
Selects next digit
Move decimal point
Increases value of digit
Decreases value of digit
Registers entry
0 100
VH 00
VH
+
E
FMX 570
m
Measured value
Bar chart
Matrix selection
Parameter entry
Alarm relay lit: fault
Matrix position
Test sockets for
0/420 mA output
+
+
See also »Controls«, Chapter 3
Measurement at a Glance Silometer FMX 570
1
Table of Contents
Measurement at a Glance
Notes on Safety . . . . . . . . . . . . 3
1 Introduction . . . . . . . . . . . . . 5
1.1 Application . . . . . . . . . . . . . . . 6
1.2 Measuring principle . . . . . . . . . . . 7
2 Installation . . . . . . . . . . . . . . 8
2.1 Probes and sensors . . . . . . . . . . . 8
2.2 Silometer installation . . . . . . . . . . . 9
2.3 Transmitter wiring . . . . . . . . . . . . 11
2.4 Sensor connection . . . . . . . . . . . . 13
2.5 Technical data: Silometer FMX 570 transmitter . 14
3 Controls . . . . . . . . . . . . . . . 15
3.1 Commutec operating matrix . . . . . . . . 15
3.2 Configuration and display . . . . . . . . . 16
4 Calibration and Operation . . . . . . . 17
4.1 Commissioning . . . . . . . . . . . . . 17
4.2 Calibration for level measurement . . . . . . 18
4.3 Calibration for linear volume or weight
measurement . . . . . . . . . . . . . . 19
4.4 »Dry calibration« for open vessels . . . . . 20
4.5 Level offset value . . . . . . . . . . . . 22
4.6 Measured value display . . . . . . . . . 23
4.7 Locking the parameter matrix . . . . . . . 23
5 Linearization . . . . . . . . . . . . . 24
5.1 Linearization for a horizontal cylindrical tank . 25
5.2 Linearization for a tank with conical outlet . . 26
5.3 Other modes . . . . . . . . . . . . . . 29
6 Analogue Outputs . . . . . . . . . . 30
6.1 Analogue output settings . . . . . . . . . 31
7 Trouble-Shooting . . . . . . . . . . . 33
7.1 Trouble-shooting tables . . . . . . . . . 33
7.2 Simulated operating mode . . . . . . . . 35
7.3 Exchanging transmitters, probes and electronic
inserts . . . . . . . . . . . . . . . . 36
7.4 Repairs . . . . . . . . . . . . . . . . 37
8 Quick programming guide . . . . . . . 38
8.1 Level measurement . . . . . . . . . . . 38
8.2 Continuous volume measurement (linearization) 39
Index . . . . . . . . . . . . . . . . 40
Operating matrix
Silometer FMX 570 Table of Contents
2
Notes on Safety
The Silometer FMX 570 is a level measurement transmitter which can be used with a variety of probes and sensors. It must be installed by qualified personnel according to the instructions in this manual.
CertificatesThe Silometer FMX 570 transmitter is available with certificate. The Table below indicates
the combinations available and conditions for installation. Full details can be taken from the certificates. Please note that where quoted technical data differs from that listed in Section 2.5, that in the certificate applies.
Certificate Instruments Notes
TÜV 00 ATEX 1640 Silometer FMC 671 Z/676 Z II (1) GD,
[EEx ia] IIC/IIB,
install outside Ex-area
PTB 98 ATEX 2215 X DC 12 TE, DC .. TE .,
DC .. E ., DC .. Capacitance probes 11500 Z(M), 11961 (Z), 21561 (Z) with electronic insert EC 16/17/27/37/47 Z, FEC 12, HTC 16/17/27 Z, HTC 10 E, HMC 37/47 Z
II 1/2 G, II 2 G, EEx ia IIC/IIB T6
PTB 98 ATEX 2215 X DC 12 TE, DC .. TE .,
DC .. E ., DC .. Capacitance probes 11500 Z(M), 11961 Z, 21561 Z with electronic insert EC 17/37/47 Z, FEC 12
II 1 G, EEx ia IIC/IIB T6
PTB 98 ATEX 2094 DB 50, DB 50 L,
DB 51, DB 52, DB 53
II 1/2 G, II 2 G, EEx ia IIC T4…T6
DIBt No. Z-65.11-29 Silometer FMX 570,
DB 50…52 with electronic insert FEB 17 / FEB (17) P
Continuous level measurement for overspill protection in stationary vessels (for storage of non-combustible, water-polluting liquids)
Silometer FMX 570 Notes on Safety
3
Safety conventions
In order to highlight safety-relevant or alternate operation procedures in the manual the following conventions have been used, each indicated by a corresponding icon in the margin.
Note!
A note highlights actions or procedures which, if not performed correctly, may
indirectly affect operation or may lead to an instrument response which is not planned.
Caution!
Caution indicates actions or procedures which, if not performed correctly, may lead
to personal injury or incorrect functioning of the instrument.
Warning!
A warning indicates actions or procedures which, if not performed correctly, will
lead to personal injury, a safety hazard or destruction of the instrument.
Notes on Safety Silometer FMX 570
4
1 Introduction
Quick Operating GuidesThe front cover contains short instructions for continuous level measurement with the
default parameters.
In this manualUsers unfamiliar with the Silometer FMX 570 must read the operating instructions, which
are structured as follows:
Chapter 1: Introduction;
contains general information including application, measurement principle and functional description.
Chapter 2: Installation;
contains hardware configuration, installation instructions. connection diagrams and technical data for the plug-in card.
Chapter 3: Controls;
describes the front panel keys and operating matrix.
Chapter 4: Calibration and Operation;
tells you how to commission the Silometer for level measurement.
Chapter 5: Linearization;
tells you how to calibrate the Silometer to measure volume in a horizontal cylindrical tank or a tank with a conical outlet.
Chapter 6: Analogue Outputs; describes in detail the setting of the 0/420 mA signal line.
Chapter 7: Trouble-Shooting;
contains a description of the self-checking system with error messages, the simulation feature as well as instructions for configuration on replacement of the transmitter, probe or electronic insert.
Chapter 8: Short Operating Guide
contains a flowcharts for level and volume measurements
Chapter 9: Index;
lists key words to help you find information quickly.
Chapter 10: Operating Matrix
contains the operating matrix, the default parameters and a table to enter your operating parameters,
Further documentationInstallation of the probes, electronic inserts and accessories are described in the
documentation accompanying these articles - see text for references. When installing probes in explosion hazardous areas the instructions included in the accompanying probe certification must also be observed.
Silometer FMX 570 Chapter 1: Introduction
5
1.1 Application
The Silometer FMX 570 is designed for level measurement with a capacitance or hydrostatic pressure probe in safe or explosion hazardous areas. It possesses an intrinsically-safe sensor circuit conforming to EEx ia IIC and IIB. A list of certificated combinations is to be found in »Notes on Safety« preceding this chapter. A working system for level measurement comprises:
Silometer FMX 570 transmitter,
Capacitance, Multicap or Deltapilot S probe
Electronic insert
Silometer function The capacitance or pressure measured by the sensor is converted into a frequency
signal by the electronic insert located in its head. The Silometer FMX 570 supplies the power and receives a level-proportional frequency signal over a two-core cable. The signal is then processed to provide a level or volume measurement.
Fail-safe operation If a fault condition is detected, e.g. a break in sensor - transmitter cable, the analogue
signal switches to -10 % or +110 % level or holds the last measured value. In addition, the alarm relay de-energises.
2
1
1
FMX 570
mA1
0/4...20 mA, 0/2...10V output
Silometer FMX 570 transmitter
Liquids
Bulk solids
Channel 1
Ex non-Ex
or
or
BA119Y01
Fig. 1.1:
Standard application showing
Silometer FMX 570 controlling
level measurement
Capacitance probe
Deltapilot
frequency
analog
digital
current voltage
sens. [Hz/pF]offset [Hz]
offset [pF] sens. [pF/cm]
range upper limit
V2
range lower limit
V1
frequency
electronic­insert
electronic­insert
capacit./ pressure
volume
V2 V1
level
volume
H2 H1
level
cap./press.
level
zero point shift
BA119E08
Fig. 1.2
Signal processing in the
Silometer FMX 570
Chapter 1: Introduction Silometer FMX 570
6
1.2 Measuring principle
The Silometer FMX 570 measures level on the basis of the capacitance and hydrostatic measurement principles. In both cases the measured value is processed by the electronic insert and passed on as a frequency signal.
The probe and vessel form the two plates of a capacitor, the total capacitance of which can then be calculated from the formula:
C
tot
=C
1
+ 2πε
r
x L pF (1)
--------------------------------
ln (D/d)
whereby C
tot
= total capacitance
C
1
= capacitance or feed through
ε
0
= dielectric constant of air
ε
r
= rel. dielectric constant of product D= diameter of vessel d= diameter of probe L= length of probe immersed in
product in meters
If the product conducts, the capacitance is determined by the thickness and properties of the insulating material surrounding the probe. Equation (1) applies, whereby the variable D is now the diameter of the probe with insulation. In this case the capacitance varies by approx. 300 pF/m.
Measurement is independent of dielectric constant and not affected by changes in this variable.
In an open vessel, the level is derived from the hydrostatic pressure exerted by a column of liquid on a probe placed at its foot. The pressure exerted is:
p
1
= ρ x g x h (2)
whereby p
1
= hydrostatic pressure
ρ = density of the liquid
g = acceleration due to gravity h = height of the liquid column.
Assuming a constant density, the level of the liquid can be calculated from the pressure measured by the Deltapilot.
BA119Y05
d
L
D
C
1
Fig. 1.3 Capacitance measurement principle
BA119E06
d
Conducting product
L
D
Fig. 1.4 Measurement in conducting media
BA119Y07
p1 = ρ x g x h
h
atmospheric pressure
p
1
Fig. 1.5 Hydrostatic measurement principle
Capacitance measurement
Hydrostatic measurement
Measurement in conducting media
Silometer FMX 570 Chapter 1: Introduction
7
2 Installation
This Chapter describes:
The probes for use with the Silometer FMX 570
Silometer installation in a rack or Monorack housing
Transmitter wiring
Sensor connection.
Technical data.
Warning!
The Silometer FMX 570 transmitter must be installed outside explosion hazardous
areas.
2.1 Probes and sensors
Table 2.1 lists the probes most frequently used with the Silometer FMX 570 transmitter. In addition to those listed, all probes which can be used with an EC 37 Z or EC 47 Z electronic insert can be connected to the transmitter. Installation hints can be taken from the appropriate Technical Information Sheet.
Sensor constants Deltapilot S sensors and EC 37 Z/47 Z inserts for capacitance probes are supplied with
the sensor constants zero frequency »fo« and sensitivity »f« or »S«. For Deltapilot S sensors the constants are printed on a label stuck inside the sensor head, for inserts they are printed on the name plate, see Fig. 7.1, Section 7.3.
Note these constants and enter them into fields V3H5 and V3H6 during commissioning, Section 4.1. This dispenses with the need for a recalibration of the transmitter on replacement of the sensor or insert.
Principle Probe TI sheet Insert
Capacitance, Multicap
11 500 Z TI 161F Multicap DC 11 TI 169F Multicap DC 16 TI 096F Multicap DC 21 TI 208F Multicap DC 26 TI 209F Multicap TA TI 239F Multicap TE TI 240F Multicap E TI 242F Multicap A TI 243F
EC 37 Z EC 47 Z FEC 12
Hydrostatic pressure Deltapilot S TI 031F
DB 50...53 TI 257P
FEB 17 (P)
Table 2.1:
Selection of probes suitable for
use with the Silometer FMX 570
Cell type
Electronic insert FEB 17/FEB 17 P Range f
0
fRange f
0
f
0.1 bar BA 0…100 mbar 200 10 DA -100…100 mbar 200 5
0.4 bar BB 0…400 mbar 200 2.5 DB -400…400 mbar 200 1.25
1.2 bar BC 0…1200 mbar 200 0.833 DC -900…1200 mbar 200 0.476
4.0 bar BD 0…4000 mbar 200 0.25 DD -900…4000 mbar 200 0.204
Table 2.2:
Measuring ranges and sensor
constants of the
Deltapilot S DB 5x
Silometer FMX 570 Chapter 2: Installation
8
2.2 Silometer installation
There are three possibilities for installing Silometer transmitters:
Standard 19" rack with space for 12 7HP cards,
Field housing with space for up to 6 7HP cards,
Monorack housings for single transmitters.
Rack installationA Racksyst system can be ordered fully wired, in which case the sensors and the external
power supply only need to be wired. Planning hints can be found in Publication SD 041/00/e, »Racksyst Assembly Racks« .
For non-Racksyst installations and for installations including non-Racksyst cards, fill the rack as follows (see also Fig. 2.1):
Rack arrangement
Step Procedure
1 Allocate the power supply (NT 470) at the rightmost position.
- If two NT 470s are used, install a 2 HP dummy panel between them.
2 Install non-intrinsically safe transmitters next to the power supply.
- Install a 2 HP dummy panel between all foreign transmitters and between Racksyst cards and foreign transmitters
3 Install intrinsically safe transmitters to the left of the rack.
- Install foreign cards first.
- Install dummy panels between all foreign transmitters and between Racksyst cards and foreign transmitters in accordance with the instructions on the Ex-Certificate.
- No spacer is required between Racksyst cards.
Racksyst field housingInstructions for installing Commutec transmitters in the Racksyst field housing with half
19" rack are to be found in Publication PI 003.
Check that the field housing is not installed in direct sunlight.
- If appropriate fit a protective sun cover.
The maximum permissible ambient temperature for the field housing varies between +50+60 °C according to the power consumption of the cards (020 W)
BA119E09
Non-E+H (Ex)i devices
E+H (Ex)i devices
E+H devices Non-E+H
devices
NT power supply
Spacing between non-E+H (Ex)i devices as per certificate
2 HP dummy panel between E+H and non-E+H devices
Fig. 2.1: Recommended arrangement for Racksyst rack assemblies
Chapter 2: Installation Silometer FMX 570
9
Monorack housing The Silometer FMX 570 transmitter and Monorack housing are supplied separately. The
system must be assembled as shown in Fig. 2.2 before use.
The Monorack is prepared for wall-mounting, degree of protection IP 40.
The site must be chosen such that the operating temperature of
-20°C+60°C for one Monorack and -20°C+50°C for Monorack banks is
not exceeded.
Full details of the Monorack installation procedure can be taken from the manual supplied with it.
Monorack protective housing
If the Silometer FMX 570 transmitter and Monorack housing are to be mounted at an exposed site, then it is recommended that they be installed in the protective housing, degree of protection IP 55, which is available as an accessory.
The protective housing accomodates two Silometer FMX 570 transmitters.
The permissible ambient temperature is -20°C…+50°C for one Monorack and
-20°C+40°C for two.
Dimensions and instructions for installation are to be found in the Technical Information sheet TI 099/00/e.
BA119E10
Plug-in card
Monorack housing
Monorack base (terminal block)
Retaining screws
Retaining flap
Fig. 2.2:
Assembly and disassembly of
the Monorack housing
Power pack
Fig. 2.3:
Monorack protective housing
BA119E60
Silometer FMX 570 Chapter 2: Installation
10
2.3 Transmitter wiring
Warning!
Make electrical connections with the power supply switched off!
When wiring up probes and sensors in explosion hazardous areas, observe the
instructions on the certificate and other appropriate regulations.
Rack wiringFig. 2.4 is a pin assignment diagram for the Silometer FMX 570.
Terminals z 30, b 14 and d 14 are connected internally
Inputs d2, d4 are electrically isolated from the circuit and each other.
The circuit zero of the unit ( ) is connected to the negative terminal of the
supply voltage.
Note!
Two indexing pins, at positions 2 and 9 in the rack connector ensure that Silometer
FMX 570 transmitters only can be inserted at these points. The pins must be inserted if the rack is not custom built by Endress+Hauser.
Analogue outputsThe negative terminal of the current output, of the voltage output and of the supply voltage
are connected to the circuit zero of the Silometer FMX 570 module.
Any number of measurement and control units can be connected in parallel to
the voltage output, provided that all potentials are related to negative terminal of the 24 V supply (R
L
10 kOhm).
- There is no limit to the number of floating devices, apart from that imposed
by considerations of maximum or minimum load.
Only one non-floating device can be connected to each of the current
outputs.
d b z
+
2 4
6 8
2 1
+
-
EC 47 Z
d2 d4
3 2
+
-
FEB 17 (P)
d2 d4
2 1
+
-
EC 37 Z
d2 d4
L L+
20 V...30 V
z30 d32
+
+
d12 d14
0/4...20 mA
+
+
b12 b14
0/2...10 V
r
z26 d28 z28
u a
10 12 + + 14 16
32 L+
18
20 22
24 26 r 28 u a 30 -L
Analogue outputs
Alarm relay
Power supply
Level, volume
BA119Y11
Alarm relay
FMX 570
Non-Ex- area
Input
DC supply
Deltapilot S
Ex-area
Current Voltage
Capacitance/ Multicap probe
Capacitance/ Multicap probe
Input
Fig. 2.4: Pin assignment diagram for Silometer FMX 570
Chapter 2: Installation Silometer FMX 570
11
Monorack wiring Fig. 2.5 shows the layout in the base of the Monorack II housing, the pin assignments
correspond to those in Fig. 2.4. When connecting together several Monoracks, follow the instructions supplied with the housing.
Set the jumper to position "Racksyst I"
Insert the coding pins supplied at positions 2 and 9 in the female connector
at the base of the housing.
The pin assignments printed in black are valid for the Silometer FMX 570.
Note! If you are installing the Silometer FMX in a Monorack I housing, please note that there is no jumper switch. In addition, for the 24 VDC version, the dummy card in the power control slot must be replaced by the 24 V card supplied.
PC board slot
BA119E14
Power
Outputs see Fig. 2.4
Input, d2, d4
Insert coding pins for Z-version at Positions 2 and 9
Hole for mounting screw
Hole for mounting screw
Jumper positioned for Racksyst I card
Fig. 2.5:
Layout of Monorack terminal
blocks
Silometer FMX 570 Chapter 2: Installation
12
2.4 Sensor connection
The Silometer FMX 570 can be operated with a variety of sensor types, each requiring a different electronic insert, e.g.:
EC 37 Z or EC 47 Z for capacitance and Multicap probes
EB 17 Z or EB 27 Z for Deltapilots
Sensor cableUse commercial 2-core installation cable, max. line resistance 25 /core, for the sensor/
transmitter cable. If electromagnetic interference is to be expected, we recommend
that the PFM negative line be grounded at the sensor (check Ex-regulations)
in case of heavy interference, that shielded cable be used, grounded at both
ends.
EC 37 Z and EC 47 ZThe electronic inserts EC 37 Z and EC 47 Z have two measuring ranges which can be
selected by inserting a bridge between terminals 4 and 5 of the insert, see Fig 2.6. Full instructions on the selection of the insert are to be found in Publication E 07.80.06/1c.
Note the zero frequency f
o
____ and sensitivity S_____on the insert.
FEB 17 (P)The FEB 17 (P) electronic insert can be used with Deltapilot S sensors to measure level
and volume in open vessels.
Note the zero frequency f
o
_____ and sensitivity f _____ of the probe
(see Table 2.2 on page 8)
BA119E15
Insert bridge for Range II (standard) Remove for Range I
Ground connection
FMX 570 input
Fig. 2.6: Connection diagram for electronic inserts EC 37 Z/EC 47 Z
FEB 17/FEB 17 P
1234
BA119Y18
FMX 570 input
Fig. 2.7: Connection diagram for electronic insert FEB 17 (P)
Chapter 2: Installation Silometer FMX 570
13
Loading...
+ 33 hidden pages