Cerabar S PMC71, PMP71, PMP75
Deltabar S FMD77 / 78, PMD75
Deltapilot S FMB70
Process pressure / Differential pressure, Flow /
Hydrostatic
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Order code:
Ext. ord. cd.:
Ser. no.:
www.endress.com/deviceviewer
Endress+Hauser
Operations App
XXXXXXXXXXXX
XXXXX-XXXXXX
XXX.XXXX.XX
Serial number
1.
3.
2.
A0023555
Make sure the document is stored in a safe place such that it is always available when
working on or with the device.
To avoid danger to individuals or the facility, read the "Basic safety instructions" section
carefully, as well as all other safety instructions in the document that are specific to
working procedures.
The manufacturer reserves the right to modify technical data without prior notice. Your
Endress+Hauser Sales Center will supply you with current information and updates to these
Instructions.
Document informationCerabar S/Deltabar S/Deltapilot S 4...20mA HART
DANGER
WARNING
CAUTION
NOTICE
1Document information
1.1Symbols used
1.1.1Safety symbols
SymbolMeaning
DANGER!
A0011189-DE
A0011190-DE
A0011191-DE
A0011192-DE
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in
seriousor fatal injury.
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
seriousor fatal injury.
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in
minoror medium injury.
NOTICE!
This symbol contains information on procedures and other facts which do not result in
personalinjury.
1.1.2Symbols for certain types of information
Symbol Meaning
Tip
Indicates additional information.
A0011193
2Basic safety instructions
Siehe Betriebsanleitung:
Deltabar S BA00270P
Cerabar S BA00271P
Deltapilot S BA00332P
3Notes on use
Typical examples of configurationsee chapter 4 to 6
Operating menu of the on-site displaysee Chapter 7
FieldCare operating menusee Chapter 8
Parameter descriptionsee Chapter 9
Finding parameter description using parameter names (index)see Page 142
4Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTPressure measurement
WARNING
!
4Pressure measurement
4.1Calibration with reference pressure
Example:
In this example, a device with a 500 mbar (7.5 psi) sensor is configured for the 0...+300
mbar (4.5 psi)measuring range, i.e. 0 mbar and 300 mbar (4.5 psi) are assigned to the 4 mA
value and 20 mA value respectively.
Prerequisite:
• The pressure values 0 mbar and 300 mbar (4.5 psi) can be specified. The device is already
installed, for example.
• See also Operating Instructions Deltabar S (BA00270P), Section "Differential pressure
measurement", Cerabar S (BA00271P), Section "Pressure measurement" or Deltapilot S
(BA00332P), Section "Pressure measurement".
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 68, Table 6: POSITION ADJUSTMENT
– Page 69, Table 7: BASIC SETUP.
• For a description of further relevant parameters, see
– Page 100, Table 15: EXTENDED SETUP
– Page 120, Table 25: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Endress+Hauser5
Pressure measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
+
–
20
4
I
[mA]
0300
p
[mbar]
➀
➁
Description
1Deltabar S: before configuring the device for your
application, the pressure piping must be cleaned and
filled with fluid. See Operating Instructions
BA00270P.
2Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3If necessary, select the "Pressure" measuring mode
via the MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
Digital communication:
See Page 62
4On-site display:
Select BASIC SETUP function group.
Menu path: GROUP SELECTION OPERATING
MENU SETTINGS BASIC SETUP
5Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
6The pressure for the lower range value (4 mA value)
is present at the device, here 0 mbar for example.
Select GET LRV parameter.
Confirm value present. The pressure value present is
assigned to the lower current value (4 mA).
7The pressure for the upper range value (20 mA
value) is present at the device, here 300 mbar (4.5
psi) for example.
Select GET URV parameter.
Confirm value present. The pressure value present is
assigned to the upper current value (20 mA).
8Result:
The measuring range is set for 0...+300 mbar (4.5
psi).
Fig. 1: Calibration with reference pressure
1See table, step 6.
2See table, step 7.
P01-PMD75xxx-19-xx-xx-xx-000
P01-xxxxxxxx-05-xx-xx-xx-010
You can also specify a customer-specific unit. See parameter description for PRESS. ENG.
UNIT (Page 69).
4.2Calibration without reference pressure
Example:
In this example, a device with a 400 mbar (6 psi) sensor is configured for the 0...+300 mbar
(4.5 psi) measuring range, i.e. 0 mbar and 300 mbar (4.5 psi) are assigned to the 4 mA value
and 20 mA value respectively.
Prerequisite:
• This is a theoretical calibration, i.e. the pressure values for the lower range and upper
6Endress+Hauser
range value are known.
• See also Operating Instructions Deltabar S (BA00270P), Section "Differential pressure
measurement", Cerabar S (BA00271P), Section "Pressure measurement" or Deltapilot S
(BA00332P), Section "Pressure measurement".
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when
the container is empty, the MEASURED VALUE parameter does not display zero. To
perform a position adjustment see also Page 68, Table 6: Position adjustment.
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTPressure measurement
WARNING
!
20
4
I
[mA]
0300
p
[mbar]
➀
➁
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 68, Table 6: POSITION ADJUSTMENT
– Page 69, Table 7: BASIC SETUP.
• For a description of further relevant parameters, see
– Page 100, Table 15: EXTENDED SETUP
– Page 120, Table 27: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1If necessary, select the "Pressure" measuring mode
via the MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
Digital communication:
See Page 62
2On-site display:
Select BASIC SETUP function group.
Menu path: GROUP SELECTION OPERATING
MENU SETTINGS BASIC SETUP
3Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
4Select SET LRV parameter.
Enter value, here 0 mbar, for the SET LRV parameter
and confirm. This pressure value is assigned to the
lower current value (4 mA).
5Select SET URV parameter.
Enter value, here 300 mbar (4.5 psi), for the SET
URV parameter and confirm. This pressure value is
assigned to the upper current value (20 mA).
6Result:
The measuring range is set for 0...+300 mbar (4.5
psi).
Fig. 2: Calibration without reference pressure
1See table, step 4.
2See table, step 5.
P01-PMP71xxx-19-xx-xx-xx-000
P01-xxxxxxxx-05-xx-xx-xx-010
Endress+Hauser7
• You can also perform calibration without reference pressure by means of the QUICK SETUP
menu. See Page 64 ff, Table 3: QUICK SETUP menu.
• You can also specify a customer-specific unit. See parameter description for PRESS. ENG.
UNIT (Page 69).
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
5Level measurement
5.1 Overview of level measurement
Measuring taskLEVEL SELECTION/
The measured variable is in
direct proportion to the
measured pressure.
Calibration is performed by
entering two pressure-level
value pairs.
The measured variable is in
direct proportion to the
measured pressure.
Calibration is performed by
entering the density and
two height-level value
pairs.
The measured variable is in
direct proportion to the
measured pressure.
The measured variable is
not in direct proportion to
the measured pressure as,
for example, with
containers with a conical
outlet. A
linearisation table must be
entered for the calibration.
– Two measured variables
are required or
– The container shape is
given by value pairs, such
as height and volume.
The 1st measured variable
%-height or height must be
in direct proportion to the
measured pressure. The
2nd measured variable
volume, mass or % must
not be in direct proportion
to the measured pressure.
A linearisation table must
be entered for the 2nd
measured variable. The 2nd
measured variable is
assigned to the 1st
measured variable by
means of this table.
LEVEL MODE
LEVEL SELECTION:
Level Easy Pressure
LEVEL SELECTION:
Level Easy Height
LEVEL SELECTION:
Level standard/
LEVEL MODE:
Linear
Via COMB.
MEASURAND
parameter:
–Height + Volume
–Height + Mass
–Height + %
–%-Height +
Volume
–%-Height + Mass
–%-Height + %
DescriptionCommentMeasured value
–Calibration with
reference pressure –
wet calibration, see
Page 9, Section 5.2.1
– Calibration without
reference pressure –
dry calibration, see
Page 11, Section 5.2.2
–Calibration with
reference pressure –
wet calibration, see
Page 13, Section 5.3.1
– Calibration without
reference pressure –
dry calibration, see
Page 15, Section 5.3.2
–Calibration with
reference pressure –
wet calibration, see
Page 17, Section 5.4.1
– Calibration without
reference pressure –
dry calibration, see
Page 19, Section 5.4.2
–Calibration with
reference pressure:
semiautomatic entry
of linearisation table,
see Page 21, Section
5.5.1
– Calibration without
reference pressure:
manual entry of
linearisation table, see
Page 24, Section 5.5.2
–Calibration with
reference pressure:
wet calibration and
semiautomatic entry
of linearisation table,
see Page 26, Section
5.6.1
– Calibration without
reference pressure:
dry calibration and
manual entry of
linearisation table, see
Page 30, Section 5.6.2
– Incorrect entries are
possible
– SIL mode possible
– Customised units are
not possible
– Incorrect entries are
possible
– SIL mode not possible
– Customised units are
not possible
– Incorrect entries are
rejected by the device
– SIL mode not possible
– Customised level,
volume and mass
units are possible
– Incorrect entries are
rejected by the device
– SIL mode not possible
– Customised level,
volume and mass
units are possible
– Incorrect entries are
rejected by the device
– SIL mode not possible
– Customised level,
volume and mass
units are possible
display
The measured value
display and the LEVEL
BEFORE LIN parameter
show the measured
value.
The measured value
display and the LEVEL
BEFORE LIN parameter
show the measured
value.
The measured value
display and the LEVEL
BEFORE LIN parameter
show the measured
value.
The measured value
display and the TANK
CONTENT parameter
show the measured
value.
The measured value
display and the TANK
CONTENT parameter
show the 2nd
measured value
(volume, mass or %).
The LEVEL BEFORE
LIN parameter displays
the 1st measured value
(%-height or height).
8Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
WARNING
!
➀
➁
0 mbar
300 mbar
3 m
0 m
5.2"Level Easy Pressure" level selection
5.2.1Calibration with reference pressure – wet calibration
Example:
In this example, the level in a tank should be measured in m. The maximum level is 3 m (9.8
ft). The pressure range is set to 0 to 300 mbar (4.5 psi).
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled or emptied.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB. and SET LRV/SET URV must have a
minimum interval of 1% for the "Level Easy Pressure" level mode. The value will be rejected
with a warning message if the values are too close together. Further limit values are not
checked; i.e. the values entered must be appropriate for the sensor and the measuring task
so that the measuring device can measure correctly.
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 68, Table 6: POSITION ADJUSTMENT
– Page 70, Table 8: LEVEL SELECTION "Level Easy Pressure"
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Deltabar S: Before you configure the device for your
application, the pressure piping must be cleaned and
filled with medium. See Operating Instructions
BA00270P.
2Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3If necessary, select the "Level" measuring mode via
the MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE "Level" LEVEL SELECTION
Digital communication:
See Page 63
Fig. 3: Calibration with reference pressure –
wet calibration
1See Table, Step 9.
2See Table, Step 10.
P01-PMP75xxx-19-xx-xx-xx-008
Endress+Hauser9
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
3
0
h
[m]
0300
p
[mbar]
➀
➁
20
4
I
[mA]
0
3
h
[m]
➂
➃
Description
5On-site display:
Select BASIC SETUP function group. Menu path:
GROUP SELECTION OPERATING MENU
SETTINGS BASIC SETUP
6Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7Select a level unit via the OUTPUT UNIT parameter,
here m for example.
8Select the "Wet" option by means of the
CALIBRATION MODE parameter.
9Hydrostatic pressure for the lower calibration point
is present at the device, here 0 mbar for example.
Select EMPTY CALIB. parameter.
Enter the level value, here 0 m for example. Confirm
the value to assign the pressure value present to the
lower level value.
To accept the value displayed you must first switch
to the Edit mode (see the "Editing values" section)
and then press the button to save the value.
10 Hydrostatic pressure for the upper calibration point
is present at the device, here 300 mbar (4.5 psi) for
example.
Select FULL CALIB. parameter.
Enter the level value, here 3 m (9.8 ft) for example.
Confirm the value to assign the pressure value
present to the upper level value.
To accept the value displayed you must first switch
to the Edit mode (see the "Editing values" section)
and then press the button to save the value.
11 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
12 Set the value for the upper current value (20 mA) by
1.You can also perform calibration with reference pressure by means of the QUICK SETUP
menu. See Page 65 ff, Table 4: QUICK SETUP menu.
2.For this level mode, the measured variables %, level, volume and mass are available.
See also parameter description for OUTPUT UNIT, Page 71.
3.For operation using the on-site display, the parameters EMPTY CALIB. (Page 72) and
FULL CALIB. (Page 73) also show the respective pressure present at the device. For
operation using Digital communication, the pressure present at the device is displayed
in the PROCESS VALUES group (menu path: OPERATING MENUPROCESSINFO
PROCESS VALUES).
10Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
WARNING
!
1000 l
50 mbar
450 mba
0 l
ρ = 1
kg
dm
3
1
2
5.2.2Calibration without reference pressure – dry calibration
Example:
In this example, the volume in a tank should be measured in litres.The maximum volume of
1000 litres (264 US gal) corresponds to a pressure of 450 mbar (6.75 psi). The minimum
volume of 0 litres corresponds to a pressure of 50 mbar (0.75 psi), as the device is mounted
below the level lower range value. The device is mounted below the level lower range value.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration i.e. the pressure and volume values for the lower and upper
calibration point must be known.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB. and SET LRV/SET URV must have a
minimum interval of 1% for the "Level Easy Pressure" level mode. The value will be rejected
with a warning message if the values are too close together. Further limit values are not
checked; i.e. the values entered must be appropriate for the sensor and the measuring task
so that the measuring device can measure correctly.
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when
the container is empty, the MEASURED VALUE parameter does not display zero. To
perform a position adjustment see also Page 68, Table 6: Position adjustment.
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 70, Table 8: LEVEL SELECTION "Level Easy Pressure"
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
For this level mode, the measured variables %, level, volume and mass are available. See
also parameter description for OUTPUT UNIT, Page 71.
12Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
WARNING
!
1000 l
0,5 m
4,5 m
0 l
ρ = 1
kg
dm
3
1
3
2
5.3"Level Easy Height" level selection
5.3.1Calibration with reference pressure – wet calibration
Example:
In this example, the volume in a tank should be measured in litres. The maximum volume of
1000 litres (264 US gal) corresponds to a level of 4.5 m (15 ft). The minimum volume of 0
litres corresponds to a level of 0.5 m (1.6 ft), as the device is mounted below the level lower
range value. The density of the medium is 1 kg/dm
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled or emptied.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB., EMPTY PRESSURE/FULL PRESSURE,
EMPTY HEIGHT/FULL HEIGHT and SET LRV/SET URV must have a minimum interval of
1% for the "Level Easy Height" level mode. The value will be rejected with a warning
message if the values are too close together. Further limit values are not checked; i.e. the
values entered must be appropriate for the sensor and the measuring task so that the
measuring device can measure correctly.
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 68, Table 6: POSITION ADJUSTMENT
– Page 73, Table 9: LEVEL SELECTION "Level Easy Height"
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 28: PROCESS VALUES.
3
.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Deltabar S: Before you configure the device for your
application, the pressure piping must be cleaned and
filled with medium. See Operating Instructions
BA00270P.
2Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
Digital communication:
See Page 62
4If necessary, select the "Level Easy Height" level mode
using the LEVEL SELECTION parameter.
Endress+Hauser13
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE "Level" LEVEL SELECTION
Digital communication:
See Page 63
5On-site display:
Select BASIC SETUP function group. Menu path:
GROUP SELECTION OPERATING MENU
SETTINGS BASIC SETUP
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
4.5
0.5
h
[m]
49441
p
[mbar]
h =
p
r · g
r = 1
g
cm
3
➀
1000
0
V
[l]
0.54.5
➁
➂
h =
p
ρ · g
h
[m]
20
4
I
[mA]
01000
V
[l]
➃
➄
Description
6Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7Select a volume unit via the OUTPUT UNIT
parameter, here l (litres) for example..
8Select a height unit via the HEIGHT UNIT parameter,
here m for example.
9Select the "Wet" option via the CALIBRATION MODE
parameter.
10 Select a density unit via the DENSITY UNIT
parameter, here kg/dm
11 Enter the density of the fluid using the ADJUST
DENSITY parameter, here kg/dm
12 Enter the volume value for the lower calibration
point via the EMPTY CALIB. parameter, here 0 l for
example. (The currently measured hydrostatic
pressure is displayed as height, here 0.5 m (1.6 ft)
for example.)
To accept the value displayed you must first switch
to the Edit mode (see the "Editing values" section)
and then press the button to save the value.
13 Enter the volume value for the upper calibration
point via the FULL CALIB. parameter, here 1000 l
(264 US gal) for example. (The currently measured
hydrostatic pressure is displayed as height, here 4.5
m (15 ft) for example.)
3
for example.
3
for example.
P01-xxxxxxxx-05-xx-xx-xx-029
To accept the value displayed you must first switch
to the Edit mode (see the "Editing values" section)
and then press the button to save the value.
14 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
15 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
16 Result:
The measuring range is set for 0 to 1000 l (264 US
gal).
For this level mode, the measured variables %, level, volume and mass are available. See
also parameter description for OUTPUT UNIT, Page 75.
14Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
WARNING
!
1000 l
0,5 m
4,5 m
0 l
ρ = 1
kg
dm
3
1
3
2
5.3.2Calibration without reference pressure – dry calibration
Example:
In this example, the volume in a tank should be measured in litres. The maximum volume is
1000 l (264 US gal), and the maximum height is 4.5 m (15 ft). The minimum volume of 0
litres corresponds to a level of 0.5 m (1.6 ft), as the device is mounted below the level lower
range value. The density of the fluid is 1 kg/dm
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration i.e. the height and volume values for the lower and upper
calibration point must be known.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB., EMPTY PRESSURE/FULL PRESSURE,
EMPTY HEIGHT/FULL HEIGHT and SET LRV/SET URV must have a minimum interval of
1% for the "Level Easy Height" level mode. The value will be rejected with a warning
message if the values are too close together. Further limit values are not checked; i.e. the
values entered must be appropriate for the sensor and the measuring task so that the
measuring device can measure correctly.
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when
the container is empty, the MEASURED VALUE parameter does not display zero. To
perform a position adjustment see also Page , Table 6: Position adjustment.
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 73, Table 9: LEVEL SELECTION "Level Easy Height"
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 28: PROCESS VALUES.
3
.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
For this level mode, the measured variables %, level, volume and mass are available. See
also parameter description for OUTPUT UNIT, Page 75.
16Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
WARNING
!
➀
➁
0 mbar
300 mbar
3 m
0 m
5.4"Level Standard" level selection,
"Linear" level type
5.4.1Calibration with reference pressure – wet calibration
Example:
In this example, the level in a tank should be measured in m. The maximum level is 3 m (9.8
ft). The pressure range is set to 0 to 300 mbar (4.5 psi).
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled or emptied.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 68, Table 6: POSITION ADJUSTMENT
– Page 77, Table 10: BASIC SETUP
– Page 79, Table 11: BASIC SETUP – "Linear" level type.
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Deltabar S: before configuring the device for your
application, the pressure piping must be cleaned and
the device filled with fluid. See Operating
Instructions BA00270P.
2Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3If necessary, select the "Level" measuring mode via
the MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
Digital communication:
See Page 62
4If necessary, select "Level Standard" level mode using
the LEVEL SELECTION parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE "Level" LEVEL SELECTION
Digital communication:
See Page 63
Fig. 11: Calibration with reference pressure –
1See table, step 11.
2See table, step 12.
wet calibration
P01-PMP75xxx-19-xx-xx-xx-008
Endress+Hauser17
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
3
0
h
[m]
0300
p
[mbar]
➀
➁
20
4
I
[mA]
0
3
h
[m]
➂
➃
Description
5On-site display:
Select BASIC SETUP function group. Menu path:
GROUP SELECTION OPERATING MENU
SETTINGS BASIC SETUP
6Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7Select the "Linear" option by means of the LEVEL
MODE parameter.
8Select the "Level" option by means of the LIN.
MEASURAND parameter.
9Select a level unit via the HEIGHT UNIT parameter,
here m for example.
10 Select the "Wet" option by means of the
CALIBRATION MODE parameter.
11 The pressure for the lower calibration point is
present at the device, here 0 mbar for example.
Select EMPTY CALIB. parameter.
Enter the level value, here 0 m for example. Confirm
the value to assign the pressure value present to the
lower level value.
12 The pressure for the upper calibration point is
present at the device, here 450 mbar (6.75 psi) for
example.
Select FULL CALIB. parameter.
Enter the level value, here 3 m (9.8 ft) for example.
Confirm the value to assign the pressure value
present to the upper level value.
13 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
14 Set the value for the upper current value (20 mA) by
1.You can also perform calibration with reference pressure by means of the QUICK SETUP
menu. See Page 65 ff, Table 4: QUICK SETUP menu.
2.You can also specify customer-specific units. See parameter description for PRESS. ENG.
UNIT (Page 77), HEIGHT UNIT ( Page 79), UNIT VOLUME (Page 80) and MASS
UNIT (Page 81).
3.For this level type, the measured variables %, level, volume and mass are available.
See Page 79 ff.
4.The EMPTY PRESSURE (Page 83) and FULL PRESSURE (Page 83) parameters
display the pressure values belonging to the EMPTY CALIB. and FULL CALIB.
parameters.
18Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
WARNING
!
4 m
–0.5 m
➂
➃
➁
r = 1
kg
dm
3
V = 5 m
3
➀
5.4.2Calibration without reference pressure – dry calibration
Example:
In this example, the volume in a tank should be measured in m
3
m
and the maximum height 4 m (13 ft). The density of the fluid is 1 kg/dm3. The device is
3
. The maximum volume is 5
mounted below the level lower range value.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration, i.e. the tank volume, tank height and density of the fluid
are known.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when
the container is empty, the MEASURED VALUE parameter does not display zero. To
perform a position adjustment see also Page 68, Table 6: Position adjustment.
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 77, Table 10: BASIC SETUP
– Page 79, Table 11: BASIC SETUP – "Linear" level type.
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 26: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
Digital communication:
See Page 62
2If necessary, select "Level Standard" level mode using
the LEVEL SELECTION parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE "Level" LEVEL SELECTION
Digital communication:
See Page 63
3On-site display:
Select BASIC SETUP function group. Menu path:
GROUP SELECTION OPERATING MENU
SETTINGS BASIC SETUP
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
20
4
I
[mA]
05
V
[m ]
3
➀
➁
Description
4Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
5Select the "Linear" option by means of the LEVEL
MODE parameter.
6Select the "Volume" option by means of the LIN.
MEASURAND parameter.
7Select a volume unit via the UNIT VOLUME
parameter, here m
8Select the "Dry" option by means of the
CALIBRATION MODE parameter. See also the
following note, point 3.
9Enter the value for density via the ADJUST DENSITY
parameter, here 1 kg/dm
10 Enter the tank volume via the TANK VOLUME
parameter, here 5 m
11 Enter the tank height via the TANK HEIGHT
parameter, here 4 m (13 ft) for example.
12 Enter the level offset via the ZERO POSITION
parameter, here –0.5 m (-1,6 ft) for example.
13 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
14 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
15 Result:
The measuring range is set for 0...5 m
3
for example.
3
for example.
3
for example.
P01-xxxx xxxx-19-xx-xx-xx-012
Fig. 14: Current output calibration
5See table, step 13.
6See table, step 14.
3
.
1.For this level type, the measured variables %, level, volume and mass are available.
See Page 79 ff.
2.You can also specify customer-specific units. See parameter description for PRESS. ENG.
UNIT (Page 77), HEIGHT UNIT ( Page 79), UNIT VOLUME (Page 80) and MASS
UNIT (Page 81).
3.A level value is assigned to the lower and upper current value by means of the SET LRV
( Page 86) and SET URV (Page 86) parameters respectively. Once you have
selected the "Dry" calibration mode, the error message A711 "LRV or URV out of edit
limits" can appear. The error message goes out as soon as level values which are within
the editing limits are entered for the SET LRV and SET URV parameters.
By means of the ENTER RESET CODE parameter ( Page 124), you can use the code
2710 to automatically set the SET LRV and SET URV parameters to level values which
are within the editing limits.
20Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
WARNING
!
3.5
0
0350
V
[m ]
3
p
[mbar]
5.5"Level Standard" level selection,
"Pressure Linearized" level type
5.5.1Semiautomatic entry of the linearisation table
Example:
In this example, the volume in a tank with a conical outlet should be measured in m
3
.
Prerequisite:
• The tank can be filled. The linearisation characteristic must rise continuously.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans
for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK
CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL
MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P) or
Deltapilot S (BA00332P).
• For a description of the parameters mentioned, see
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Deltabar S: before configuring the device for your
application, the pressure piping must be cleaned and
filled with fluid. See Operating Instructions
BA00270P.
2Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
Carry out basic setup:
3If necessary, select the "Level" measuring mode via
the MEASURING MODE parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE
Digital communication:
See Page 62
4If necessary, select "Level Standard" level mode using
the LEVEL SELECTION parameter.
On-site display:
Menu path: GROUP SELECTION MEASURING
MODE "Level" LEVEL SELECTION
Digital communication:
See Page 63
P01-PMP75xxx-19-xx-xx-xx-002
Endress+Hauser21
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
Description
5On-site display:
Select BASIC SETUP function group. Menu path:
GROUP SELECTION OPERATING MENU
SETTINGS BASIC SETUP
6Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7Select the "Pressure Linearized" option by means of
the LEVEL MODE parameter. See also the following
note, point 3.
8Select the "Volume" option by means of the LINd.
MEASURAND parameter.
9Select a volume unit via the UNIT VOLUME
parameter, here m
10 Select HYDR. PRESS MIN. parameter.
Enter the minimum hydrostatic pressure to be
expected, here 0 mbar for example.
11 Select HYDR. PRESS MAX
Enter the maximum hydrostatic pressure to be
expected.
Carry out linearisation:
12 Change the function group:
Menu path: (GROUP SELECTION ) OPERATING
MENU SETTINGS LINEARISATION
13 Select TANK CONTENT MIN parameter.
Specify the minimum tank contents to be expected,
here 0 m
3
14 Select TANK CONTENT MAX parameter.
Specify the maximum tank contents to be expected,
here 3.5 m
15 On-site display:
Select the "Editor table" option by means of the
TABLE SELECTION parameter.
16 Select the "Semiautomatic" option by means of the
LIN. EDIT MODE parameter.
17 Select the "New table" option by means of the EDITOR
TABLE parameter.
3
for example.
for example.
3
for example.
.
22Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
➀
3.5
0
0350
V
[m ]
3
p
[mbar]
➂
➁
➃
➄
20
4
I
[mA]
03.5
V
[m ]
3
➆
➅
Description
18 Enter linearisation table (min. 2 points, max.
32 points).
Fill the tank to the height of the 1st point.
LINE-NUMB: confirm value displayed.
X-VAL.: the hydrostatic pressure present is
displayed.
On-site display, Digital communication:
The X-VAL. displayed is saved by confirming the Yvalue. See following line, Y-VAL.
HART handheld terminal:
Confirm X-VAL. displayed.
Y-VAL.: enter the volume value, here 0 m
example, and confirm the value.
19 On-site display:
If you want to enter another point for the
linearisation table, select the "Next point" option and
enter the point as described in step 18.
If you want to finish entering the values and activate
the linearisation table, select the "Accept input table"
option.
Digital communication:
You can enter further points for the linearisation
table as explained in step 18. Once all the points
have been entered, the table must be activated by
means of the TAB. ACTIVATE parameter.
19 Result:
The linearisation table has been entered.
3
for
P01-xxxxxxxx-05-xx-xx-xx-015
P01-xxxxxxxx-05-xx-xx-xx-016
Fig. 15: Semiautomatic entry of the linearisation table
1See table, step 10.
2See table, step 11.
3See table, step 13.
4See table, step 14.
5See table, steps 15 – 19.
6See the following note, point 4.
7See the following note, point 4
Endress+Hauser23
1.For this level type, the measured variables %, volume and mass are available.
See Page 87 ff.
2.You can also specify customer-specific units. See parameter description for PRESS. ENG.
UNIT (Page 77), HEIGHT UNIT ( Page 87), UNIT VOLUME (Page 87) and MASS
UNIT ( Page 88).
3.Once you have selected the "Pressure Linearized" level type, the warning message "W710
Set span too small. Not allowed." can appear. At this stage the linearisation table already
consists of two points as standard. It could be the case that the 2nd value, and thus the
highest X-VAL. of the linearisation table, is smaller than the minimum span permitted
( MINIMUM SPAN, Page 120). The message goes out as soon as the highest X-VAL.
is larger than the minimum span.
4.A level value is assigned to both the lower and upper current value with the SET LRV
( Page 102) and SET URV (Page 102) parameters. If you enter values for TANK
CONTENT MIN (Page 104 or 107) and TANK CONTENT MAX (Page 104 or 107),
the SET LRV and SET URV parameters are also changed. If you want to assign values
other than those for TANK CONTENT MIN and TANK CONTENT MAX to the lower and
upper current values, the desired values must be entered for SET LRV and SET URV.
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
WARNING
!
3.5
0
0350
V
[m ]
3
p
[mbar]
5.5.2Manual entry of the linearisation table
Example:
In this example, the volume in a tank with a conical outlet should be measured in m
3
.
Prerequisite:
• This is a theoretical calibration, i.e. the points for the linearisation table are known.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans
for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK
CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL
MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 68, Table 6: POSITION ADJUSTMENT
– Page 77, Table 10: BASIC SETUP
– Page 87, Table 12: BASIC SETUP – "Pressure Linearized" level type
– Page 104, Table 18: LINEARISATION – on-site operation
– Page 107, Table 19: LINEARISATION – Digital communication.
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
1Perform basic setup as per Section 5.3.1, steps 2 to
11.
Carry out linearisation:
2Change the function group:
Menu path: (GROUP SELECTION ) OPERATING
MENU SETTINGS LINEARISATION
3Select TANK CONTENT MIN parameter .
Specify the minimum tank contents to be expected,
here 0 m
4Select TANK CONTENT MAX parameter .
Specify the maximum tank contents to be expected,
here 3.5 m
3
for example.
3
for example.
P01-PMP75xxx-19-xx-xx-xx-002
24Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
➀
3.5
0
0350
V
[m ]
3
p
[mbar]
➂
➁
➃
➄
20
4
I
[mA]
03.5
V
[m ]
3
➆
➅
Description
5On-site display:
Select the "Editor table" option by means of the
TABLE SELECTION parameter.
6Select the "Manual" option by means of the LIN. EDIT
MODE parameter.
7 Select the "New table" option by means of the
EDITOR TABLE parameter.
8Enter linearisation table (min. 2 points, max.
32 points).
LINE-NUMB: confirm value displayed.
X-VAL.: enter the pressure value and confirm.
Y-VAL.: enter the volume value, here 0 m
example, and confirm.
9On-site display
If you want to enter another point for the
linearisation table, select the "Next point" option and
enter the point as described in step 8.
If you want to finish entering the values and activate
the linearisation table, select the "Accept input table"
option.
Digital communication:
You can enter further points for the linearisation
table as explained in step 8. Once all the points have
been entered, the table must be activated by means
of the TAB. ACTIVATE parameter.
10 Result:
The linearisation table has been entered.
3
for
P01-xxxxxxxx-05-xx-xx-xx-015
P01-xxxxxxxx-05-xx-xx-xx-016
Fig. 16: Manual entry of the linearisation table
1See Section 5.3.1, table, step 9.
2See Section 5.3.1, table, step 10.
3See table, step 3.
4See table, step 4.
5See table, steps 5 – 9.
6See the following note, point 4.
7See the following note, point 4.
Endress+Hauser25
1.For this level type, the measured variables %, volume and mass are available.
See Page 87 ff.
2.You can also specify customer-specific units. See parameter description for PRESS. ENG.
UNIT (Page 77), HEIGHT UNIT ( Page 87), UNIT VOLUME (Page 87) and MASS
UNIT ( Page 88).
3.Once you have selected the "Pressure Linearized" level type, the warning message "W710
Set span too small. Not allowed." can appear. At this stage the linearisation table already
consists of two points as standard. It could be the case that the 2nd value, and thus the
highest X-VAL. of the linearisation table, is smaller than the minimum span permitted
( MINIMUM SPAN, Page 120). The message goes out as soon as the highest X-VAL.
is larger than the minimum span.
4.A level value is assigned to both the lower and upper current value with the SET LRV
( Page 102) and SET URV (Page 102) parameters. If you enter values for TANK
CONTENT MIN (Page 104 or 107) and TANK CONTENT MAX (Page 104 or 107),
the SET LRV and SET URV parameters are also changed. If you want to assign values
other than those for TANK CONTENT MIN and TANK CONTENT MAX to the lower and
upper current values, the desired values must be entered for SET LRV and SET URV.
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
WARNING
!
5.6"Level Standard" level selection,
"Height Linearized" level type
5.6.1Wet calibration and semiautomatic entry of the linearisation
table
Example:
In this example, the height and the volume should be measured at the same time.
Prerequisite:
• The tank can be filled. The linearisation characteristic must rise continuously.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans
for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK
CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL
MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P) or
Deltapilot S (BA00332P).
• For a description of the parameters mentioned, see
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
5
0
03
V
[m ]
3
h
[m]
5
0
03
V
[m ]
3
h
[m]
➄
➆
➅
20
4
I
[mA]
05
V
[m ]
3
➈
➇
Description
Perform linearisation (calibration for the 2nd
measured variable)
17 Change the function group.
Menu path: (GROUP SELECTION ) OPERATING
MENU SETTINGS LINEARISATION
18 Select TANK CONTENT MIN parameter.
Specify the minimum tank contents to be expected,
here 0 m
19 Select TANK CONTENT MAX parameter.
Specify the maximum tank contents to be expected,
here 5 m
20 On-site display:
Select the "Editor table" option by means of the
TABLE SELECTION parameter.
21 Select the "Semiautomatic" option by means of the
LIN. EDIT MODE parameter.
22 Select the "New table" option by means of the EDITOR
TABLE parameter.
23 Enter linearisation table (min. 2 points, max.
32 points).
Fill the tank to the height of the 1st point.
LINE-NUMB: confirm value displayed.
X-VAL.: the hydrostatic pressure present is
measured and converted to the corresponding level
and displayed.
On-site display, Digital communication:
The X-VAL. displayed is saved by confirming the
Y-value. See following line, Y-VAL.
HART handheld terminal:
Confirm X-VAL. displayed.
Y-VAL.: enter the volume value, here 0 m
example, and confirm the value.
24 On-site display
If you want to enter another point for the
linearisation table, select the "Next point" option and
enter the point as described in step 23.
If you want to finish entering the values and activate
the linearisation table, select the "Accept input table"
option.
Digital communication:
You can enter further points for the linearisation
table as explained in step 23. Once all the points
have been entered, the table must be activated by
means of the TAB. ACTIVATE parameter.
25 Result:
– The linearisation table has been entered.
– The measured value display and the
– The LEVEL BEFORE LIN parameter displays the
3
for example.
3
for example.
3
for
TANK CONTENT parameter display the 2nd
measured value (here the volume).
1st measured value (here the height). See also the
following note, point 5.
P01-PMP75xxx-19-xx-xx-xx-005
P01-xxxxxxxx-05-xx-xx-xx-018
P01-xxxxxxxx-05-xx-xx-xx-019
Fig. 18: Calibrating the 2nd measured variable
5See table, step 18.
6See table, step 19.
7See table, steps 20 – 24.
8See the following note, point 4.
9See the following note, point 4.
28Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HARTLevel measurement
1.For this level type, the measured variables "Height + %", "Height + Volume", "Height +
Mass", "%-Height + %", "%-Height + Volume" and "%-Height + Mass" are available. See
Page 87 ff.
2.You can also specify customer-specific units. See parameter description for PRESS. ENG.
UNIT (Page 77), HEIGHT UNIT ( Page 90), UNIT VOLUME (Page 91) and MASS
UNIT ( Page 92).
3.Once you have selected the "Pressure Linearized" level type, the warning message "W710
Set span too small. Not allowed." can appear. At this stage the linearisation table already
consists of two points as standard. It could be the case that the 2nd value, and thus the
highest X-VAL. of the linearisation table, is smaller than the minimum span permitted
( MINIMUM SPAN, Page 120). The message goes out as soon as the highest X-VAL.
is larger than the minimum span.
4.A level value is assigned to both the lower and upper current value with the SET LRV
( Page 102) and SET URV (Page 102) parameters.
You can use the ASSIGN CURRENT parameter (Page 114) to specify whether the
current output should depict the 1st or 2nd measured variable. Depending on the
setting of the ASSIGN CURRENT parameter, enter the following values for SET LRV and
SET URV:
– ASSIGN CURRENT = tank content (factory setting) %-value, volume value or
mass value
– ASSIGNMENT = height level value
The following applies for the setting ASSIGN CURRENT "Tank content":
If you enter values for TANK CONTENT MIN (Page 104 or 107) and TANK CONTENT
MAX ( Page 104 or 107), the SET LRV and SET URV parameters are also changed. If
you want to assign values other than those for TANK CONTENT MIN and TANK
CONTENT MAX to the lower and upper current values, the desired values must be
entered for SET LRV and SET URV.
The following applies for the setting ASSIGN CURRENT "Height":
If you enter values for LEVEL MIN (Page 93) and LEVEL MAX (Page 93), the SET
LRV and SET URV parameters are also changed. If you want to assign values other than
those for LEVEL MIN and LEVEL MAX to the lower and upper current values, the
desired values must be entered for SET LRV and SET URV.
5.You can use the MENU DESCRIPTOR parameter (Page 111) to specify which
measured value should be displayed on the on-site display.
Endress+Hauser29
Level measurementCerabar S/Deltabar S/Deltapilot S 4...20mA HART
WARNING
!
5
0
03
V
[m ]
3
h
[m]
5.6.2Dry calibration and manual entry of the linearisation table
Example:
In this example, the height and the volume should be measured at the same time.
Prerequisite:
• This is a theoretical calibration, i.e. the points for the linearisation table are known.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans
for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK
CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL
MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P),
Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when
the container is empty, the MEASURED VALUE parameter does not display zero. To
perform a position adjustment see also Page 68, Table 6: Position adjustment.
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE
– Page 77, Table 10: BASIC SETUP
– Page 89, Table 12: BASIC SETUP – "Height Linearized" level type
– Page 104, Table 18: LINEARISATION – on-site operation
– Page 107, Table 19: LINEARISATION – Digital communication.
• For a description of further parameters, see
– Page 100, Table 16: EXTENDED SETUP
– Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
‣ Check calibration data when the measuring mode is changed.
Description
Perform calibration for the 1st measured
variable:
1Perform calibration as per Section 5.4.2, steps 3 to
12.
2Select the "Dry" option via the CALIBRATION MODE
parameter (calibration mode for the 1st measured
variable).
3Enter the density of the fluid via the ADJUST
DENSITY parameter, here 1 kg/dm
4If necessary, enter a level offset via the ZERO
POSITION parameter, here 0 m for example.
5Result:
The calibration for the 1st measured variable is
carried out.
3
for example.
P01-PMP75xxx-19-xx-xx-xx-005
30Endress+Hauser
Loading...
+ 118 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.