Endress+Hauser FMD78 Operating Manual

BA00274P/00/EN/18.16 71336359
valid from Software version:
02.30.zz
Products Solutions Services
Description of Instrument Functions
Cerabar S PMC71, PMP71, PMP75 Deltabar S FMD77 / 78, PMD75 Deltapilot S FMB70
Process pressure / Differential pressure, Flow / Hydrostatic
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Order code:
Ext. ord. cd.:
Ser. no.:
www.endress.com/deviceviewer
Endress+Hauser Operations App
XXXXXXXXXXXX
XXXXX-XXXXXX
XXX.XXXX.XX
Serial number
1.
3.
2.
A0023555
Make sure the document is stored in a safe place such that it is always available when working on or with the device. To avoid danger to individuals or the facility, read the "Basic safety instructions" section carefully, as well as all other safety instructions in the document that are specific to working procedures. The manufacturer reserves the right to modify technical data without prior notice. Your Endress+Hauser Sales Center will supply you with current information and updates to these Instructions.
2 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Table of contents
1 Document information . . . . . . . . . . . . . . 4
1.1 Symbols used . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
2 Basic safety instructions . . . . . . . . . . . . . 4
3 Notes on use . . . . . . . . . . . . . . . . . . . . . . . 4
4 Pressure measurement . . . . . . . . . . . . . . 5
4.1 Calibration with reference pressure . . . . . . . . . . . . 5
4.2 Calibration without reference pressure . . . . . . . . . 6
5 Level measurement . . . . . . . . . . . . . . . . . 8
5.1 Overview of level measurement . . . . . . . . . . . . . . 8
5.2 "Level Easy Pressure" level selection . . . . . . . . . . . . 9
5.3 "Level Easy Height" level selection . . . . . . . . . . . . 13
5.4 "Level Standard" level selection,
"Linear" level type . . . . . . . . . . . . . . . . . . . . . . . . . 17
5.5 "Level Standard" level selection,
"Pressure Linearized" level type . . . . . . . . . . . . . . 21
5.6 "Level Standard" level selection,
"Height Linearized" level type . . . . . . . . . . . . . . . 26
6 Flow measurement . . . . . . . . . . . . . . . .33
6.1 Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
6.2 Totalizers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
7 On-site display operating menu . . . . .36
8 FieldCare operating menu . . . . . . . . . .43
9 Description of parameters . . . . . . . . . .62
10 Troubleshooting. . . . . . . . . . . . . . . . . .131
10.1 Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131
10.2 Response of outputs to errors . . . . . . . . . . . . . . 140
10.3 Confirming messages . . . . . . . . . . . . . . . . . . . . . 141
Index. . . . . . . . . . . . . . . . . . . . . . . . . . . .142
Endress+Hauser 3
Document information Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
DANGER
WARNING
CAUTION
NOTICE

1 Document information

1.1 Symbols used

1.1.1 Safety symbols
Symbol Meaning
DANGER!
A0011189-DE
A0011190-DE
A0011191-DE
A0011192-DE
This symbol alerts you to a dangerous situation. Failure to avoid this situation will result in seriousor fatal injury.
WARNING!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in seriousor fatal injury.
CAUTION!
This symbol alerts you to a dangerous situation. Failure to avoid this situation can result in minoror medium injury.
NOTICE!
This symbol contains information on procedures and other facts which do not result in personalinjury.
1.1.2 Symbols for certain types of information
Symbol Meaning
Tip
Indicates additional information.
A0011193

2 Basic safety instructions

Siehe Betriebsanleitung: Deltabar S BA00270P Cerabar S BA00271P Deltapilot S BA00332P

3 Notes on use

Typical examples of configuration see chapter 4 to 6
Operating menu of the on-site display see Chapter 7
FieldCare operating menu see Chapter 8
Parameter description see Chapter 9
Finding parameter description using parameter names (index) see Page 142
4 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Pressure measurement
WARNING
!

4 Pressure measurement

4.1 Calibration with reference pressure

Example:
In this example, a device with a 500 mbar (7.5 psi) sensor is configured for the 0...+300 mbar (4.5 psi)measuring range, i.e. 0 mbar and 300 mbar (4.5 psi) are assigned to the 4 mA value and 20 mA value respectively.
Prerequisite:
• The pressure values 0 mbar and 300 mbar (4.5 psi) can be specified. The device is already installed, for example.
• See also Operating Instructions Deltabar S (BA00270P), Section "Differential pressure measurement", Cerabar S (BA00271P), Section "Pressure measurement" or Deltapilot S (BA00332P), Section "Pressure measurement".
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 69, Table 7: BASIC SETUP.
• For a description of further relevant parameters, see – Page 100, Table 15: EXTENDED SETUP – Page 120, Table 25: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Endress+Hauser 5
Pressure measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
+
20
4
I
[mA]
0 300
p
[mbar]
Description
1 Deltabar S: before configuring the device for your
application, the pressure piping must be cleaned and filled with fluid. See Operating Instructions BA00270P.
2 Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3 If necessary, select the "Pressure" measuring mode
via the MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
4On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU SETTINGS BASIC SETUP
5 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
6 The pressure for the lower range value (4 mA value)
is present at the device, here 0 mbar for example.
Select GET LRV parameter.
Confirm value present. The pressure value present is assigned to the lower current value (4 mA).
7 The pressure for the upper range value (20 mA
value) is present at the device, here 300 mbar (4.5 psi) for example.
Select GET URV parameter.
Confirm value present. The pressure value present is assigned to the upper current value (20 mA).
8Result:
The measuring range is set for 0...+300 mbar (4.5 psi).
Fig. 1: Calibration with reference pressure
1 See table, step 6. 2 See table, step 7.
P01-PMD75xxx-19-xx-xx-xx-000
P01-xxxxxxxx-05-xx-xx-xx-010
You can also specify a customer-specific unit. See parameter description for PRESS. ENG. UNIT (Page 69).

4.2 Calibration without reference pressure

Example:
In this example, a device with a 400 mbar (6 psi) sensor is configured for the 0...+300 mbar (4.5 psi) measuring range, i.e. 0 mbar and 300 mbar (4.5 psi) are assigned to the 4 mA value and 20 mA value respectively.
Prerequisite:
• This is a theoretical calibration, i.e. the pressure values for the lower range and upper
6 Endress+Hauser
range value are known.
• See also Operating Instructions Deltabar S (BA00270P), Section "Differential pressure measurement", Cerabar S (BA00271P), Section "Pressure measurement" or Deltapilot S (BA00332P), Section "Pressure measurement".
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when the container is empty, the MEASURED VALUE parameter does not display zero. To perform a position adjustment see also Page 68, Table 6: Position adjustment.
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Pressure measurement
WARNING
!
20
4
I
[mA]
0 300
p
[mbar]
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 69, Table 7: BASIC SETUP.
• For a description of further relevant parameters, see – Page 100, Table 15: EXTENDED SETUP – Page 120, Table 27: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 If necessary, select the "Pressure" measuring mode
via the MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
2 On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU SETTINGS BASIC SETUP
3 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
4 Select SET LRV parameter.
Enter value, here 0 mbar, for the SET LRV parameter and confirm. This pressure value is assigned to the lower current value (4 mA).
5 Select SET URV parameter.
Enter value, here 300 mbar (4.5 psi), for the SET URV parameter and confirm. This pressure value is assigned to the upper current value (20 mA).
6 Result:
The measuring range is set for 0...+300 mbar (4.5 psi).
Fig. 2: Calibration without reference pressure
1 See table, step 4. 2 See table, step 5.
P01-PMP71xxx-19-xx-xx-xx-000
P01-xxxxxxxx-05-xx-xx-xx-010
Endress+Hauser 7
• You can also perform calibration without reference pressure by means of the QUICK SETUP menu. See Page 64 ff, Table 3: QUICK SETUP menu.
• You can also specify a customer-specific unit. See parameter description for PRESS. ENG. UNIT (Page 69).
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART

5 Level measurement

5.1 Overview of level measurement

Measuring task LEVEL SELECTION/
The measured variable is in direct proportion to the measured pressure. Calibration is performed by entering two pressure-level value pairs.
The measured variable is in direct proportion to the measured pressure. Calibration is performed by entering the density and two height-level value pairs.
The measured variable is in direct proportion to the measured pressure.
The measured variable is not in direct proportion to the measured pressure as, for example, with containers with a conical outlet. A linearisation table must be entered for the calibration.
– Two measured variables
are required or
– The container shape is
given by value pairs, such as height and volume.
The 1st measured variable %-height or height must be in direct proportion to the measured pressure. The 2nd measured variable volume, mass or % must not be in direct proportion to the measured pressure. A linearisation table must be entered for the 2nd measured variable. The 2nd measured variable is assigned to the 1st measured variable by means of this table.
LEVEL MODE
LEVEL SELECTION: Level Easy Pressure
LEVEL SELECTION: Level Easy Height
LEVEL SELECTION: Level standard/ LEVEL MODE: Linear
LEVEL SELECTION: Level standard/ LEVEL MODE: Pressure Linearized
LEVEL SELECTION: Level standard/ LEVEL MODE: Height Linearized
Measured variable options
Via OUTPUT UNIT parameter: %, level, volume or mass units.
Via OUTPUT UNIT parameter: %, level, volume or mass units.
Via LIN. MEASURAND parameter: –% (Level) – Level –Volume –Mass
Via LINd MEASURAND parameter: – Pressure + % – Pressure +
Volume
– Pressure + Mass
Via COMB. MEASURAND parameter: –Height + Volume –Height + Mass –Height + % –%-Height +
Volume –%-Height + Mass –%-Height + %
Description Comment Measured value
–Calibration with
reference pressure – wet calibration, see Page 9, Section 5.2.1
– Calibration without
reference pressure – dry calibration, see Page 11, Section 5.2.2
–Calibration with
reference pressure – wet calibration, see Page 13, Section 5.3.1
– Calibration without
reference pressure – dry calibration, see Page 15, Section 5.3.2
–Calibration with
reference pressure – wet calibration, see Page 17, Section 5.4.1
– Calibration without
reference pressure – dry calibration, see Page 19, Section 5.4.2
–Calibration with
reference pressure: semiautomatic entry of linearisation table, see Page 21, Section
5.5.1
– Calibration without
reference pressure: manual entry of linearisation table, see Page 24, Section 5.5.2
–Calibration with
reference pressure: wet calibration and semiautomatic entry of linearisation table, see Page 26, Section
5.6.1
– Calibration without
reference pressure: dry calibration and manual entry of linearisation table, see Page 30, Section 5.6.2
– Incorrect entries are
possible – SIL mode possible – Customised units are
not possible
– Incorrect entries are
possible – SIL mode not possible – Customised units are
not possible
– Incorrect entries are
rejected by the device – SIL mode not possible – Customised level,
volume and mass
units are possible
– Incorrect entries are
rejected by the device – SIL mode not possible – Customised level,
volume and mass
units are possible
– Incorrect entries are
rejected by the device – SIL mode not possible – Customised level,
volume and mass
units are possible
display
The measured value display and the LEVEL BEFORE LIN parameter show the measured value.
The measured value display and the LEVEL BEFORE LIN parameter show the measured value.
The measured value display and the LEVEL BEFORE LIN parameter show the measured value.
The measured value display and the TANK CONTENT parameter show the measured value.
The measured value display and the TANK CONTENT parameter show the 2nd measured value (volume, mass or %).
The LEVEL BEFORE LIN parameter displays the 1st measured value (%-height or height).
8 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
WARNING
!
0 mbar
300 mbar
3 m
0 m

5.2 "Level Easy Pressure" level selection

5.2.1 Calibration with reference pressure – wet calibration
Example:
In this example, the level in a tank should be measured in m. The maximum level is 3 m (9.8 ft). The pressure range is set to 0 to 300 mbar (4.5 psi).
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled or emptied.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB. and SET LRV/SET URV must have a minimum interval of 1% for the "Level Easy Pressure" level mode. The value will be rejected with a warning message if the values are too close together. Further limit values are not checked; i.e. the values entered must be appropriate for the sensor and the measuring task so that the measuring device can measure correctly.
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 70, Table 8: LEVEL SELECTION "Level Easy Pressure"
• For a description of further relevant parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Deltabar S: Before you configure the device for your
application, the pressure piping must be cleaned and filled with medium. See Operating Instructions BA00270P.
2 Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3 If necessary, select the "Level" measuring mode via
the MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
4 If necessary, select "Level Easy Pressure" level mode
using the LEVEL SELECTION parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: See Page 63
Fig. 3: Calibration with reference pressure –
wet calibration
1 See Table, Step 9. 2 See Table, Step 10.
P01-PMP75xxx-19-xx-xx-xx-008
Endress+Hauser 9
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
3
0
h
[m]
0 300
p
[mbar]
20
4
I
[mA]
0
3
h
[m]
Description
5On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
6 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7 Select a level unit via the OUTPUT UNIT parameter,
here m for example.
8 Select the "Wet" option by means of the
CALIBRATION MODE parameter.
9 Hydrostatic pressure for the lower calibration point
is present at the device, here 0 mbar for example.
Select EMPTY CALIB. parameter.
Enter the level value, here 0 m for example. Confirm the value to assign the pressure value present to the lower level value. To accept the value displayed you must first switch to the Edit mode (see the "Editing values" section) and then press the button to save the value.
10 Hydrostatic pressure for the upper calibration point
is present at the device, here 300 mbar (4.5 psi) for example.
Select FULL CALIB. parameter.
Enter the level value, here 3 m (9.8 ft) for example. Confirm the value to assign the pressure value present to the upper level value.
To accept the value displayed you must first switch to the Edit mode (see the "Editing values" section) and then press the button to save the value.
11 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
12 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
13 Result:
The measuring range is set for 0 to 3 m (9.8 ft).
Fig. 4: Calibration with reference pressure –
1 See Table, Step 9. 2 See Table, Step 10. 3 See Table, Step 11. 4 See Table, Step 12.
wet calibration
P01-xxxxxxxx-05-xx-xx-xx-011
P01-xxxxxxxx-05-xx-xx-xx-014
1. You can also perform calibration with reference pressure by means of the QUICK SETUP menu. See Page 65 ff, Table 4: QUICK SETUP menu.
2. For this level mode, the measured variables %, level, volume and mass are available.  See also parameter description for OUTPUT UNIT, Page 71.
3. For operation using the on-site display, the parameters EMPTY CALIB. (Page 72) and FULL CALIB. (Page 73) also show the respective pressure present at the device. For operation using Digital communication, the pressure present at the device is displayed in the PROCESS VALUES group (menu path: OPERATING MENUPROCESSINFO PROCESS VALUES).
10 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
WARNING
!
1000 l
50 mbar
450 mba
0 l
ρ = 1
kg
dm
3
1
2
5.2.2 Calibration without reference pressure – dry calibration
Example:
In this example, the volume in a tank should be measured in litres.The maximum volume of 1000 litres (264 US gal) corresponds to a pressure of 450 mbar (6.75 psi). The minimum volume of 0 litres corresponds to a pressure of 50 mbar (0.75 psi), as the device is mounted below the level lower range value. The device is mounted below the level lower range value.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration i.e. the pressure and volume values for the lower and upper calibration point must be known.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB. and SET LRV/SET URV must have a minimum interval of 1% for the "Level Easy Pressure" level mode. The value will be rejected with a warning message if the values are too close together. Further limit values are not checked; i.e. the values entered must be appropriate for the sensor and the measuring task so that the measuring device can measure correctly.
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when the container is empty, the MEASURED VALUE parameter does not display zero. To perform a position adjustment see also Page 68, Table 6: Position adjustment.
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 70, Table 8: LEVEL SELECTION "Level Easy Pressure"
• For a description of further relevant parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
2 If necessary, select "Level Easy Pressure" level mode
using the LEVEL SELECTION parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: See Page 63
3 On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
Fig. 5: Calibration without reference pressure –
1 See Table, Steps 7 and 8. 2 See Table, Steps 9 and 10.
dry calibration
P01-PMC71xxx-19-xx-xx-xx-000
Endress+Hauser 11
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
1000
0
V [l]
50 450
p
[mbar]
20
4
I
[mA]
0 1000
V [l]
Description
4 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
5 Select a volume unit via the OUTPUT UNIT
parameter, here l (litres) for example..
6 Select the "Dry" option by means of the
CALIBRATION MODE parameter.
7 Enter the volume value for the lower calibration
point via the EMPTY CALIB. parameter, here 0 l for example.
8 Enter the pressure value for the lower calibration
point via the EMPTY PRESSURE parameter, here 50 mbar (0.75 psi) for example.
9 Enter the volume value for the upper calibration
point via the FULL CALIB. parameter, here 1000 l (264 gal) for example.
10 Enter the pressure value for the upper calibration
point via the FULL PRESSURE parameter, here 450 mbar (6.75 psi) for example.
11 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
12 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
13 Result:
The measuring range is set for 0 to 1000 l (264 gal).
P01-xxxxxxxx-05-xx-xx-xx-026
P01-xxxxxxxx-05-xx-xx-xx-028
Fig. 6: Calibration with reference pressure –
1 See Table, Step 7. 2 See Table, Step 8. 3 See Table, Step 9. 4 See Table, Step 10. 5 See Table, Step 11. 6 See Table, Step 12.
wet calibration
For this level mode, the measured variables %, level, volume and mass are available. See also parameter description for OUTPUT UNIT, Page 71.
12 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
WARNING
!
1000 l
0,5 m
4,5 m
0 l
ρ = 1
kg
dm
3
1
3
2

5.3 "Level Easy Height" level selection

5.3.1 Calibration with reference pressure – wet calibration
Example:
In this example, the volume in a tank should be measured in litres. The maximum volume of 1000 litres (264 US gal) corresponds to a level of 4.5 m (15 ft). The minimum volume of 0 litres corresponds to a level of 0.5 m (1.6 ft), as the device is mounted below the level lower range value. The density of the medium is 1 kg/dm
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled or emptied.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB., EMPTY PRESSURE/FULL PRESSURE, EMPTY HEIGHT/FULL HEIGHT and SET LRV/SET URV must have a minimum interval of 1% for the "Level Easy Height" level mode. The value will be rejected with a warning message if the values are too close together. Further limit values are not checked; i.e. the values entered must be appropriate for the sensor and the measuring task so that the measuring device can measure correctly.
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 73, Table 9: LEVEL SELECTION "Level Easy Height"
• For a description of further relevant parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
3
.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Deltabar S: Before you configure the device for your
application, the pressure piping must be cleaned and filled with medium. See Operating Instructions BA00270P.
2 Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3 Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
4 If necessary, select the "Level Easy Height" level mode
using the LEVEL SELECTION parameter.
Endress+Hauser 13
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: See Page 63
5 On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
Fig. 7: Calibration with reference pressure –
1 See Table, Steps 10 and 11. 2 See Table, Step 12. 3 See Table, Step 13.
wet calibration
P01-PMC71xxx-19-xx-xx-xx-001
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
4.5
0.5
h
[m]
49 441
p
[mbar]
h =
p
r · g
r = 1
g
cm
3
1000
0
V [l]
0.5 4.5
h =
p
ρ · g
h
[m]
20
4
I
[mA]
0 1000
V [l]
Description
6 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7 Select a volume unit via the OUTPUT UNIT
parameter, here l (litres) for example..
8 Select a height unit via the HEIGHT UNIT parameter,
here m for example.
9 Select the "Wet" option via the CALIBRATION MODE
parameter.
10 Select a density unit via the DENSITY UNIT
parameter, here kg/dm
11 Enter the density of the fluid using the ADJUST
DENSITY parameter, here kg/dm
12 Enter the volume value for the lower calibration
point via the EMPTY CALIB. parameter, here 0 l for example. (The currently measured hydrostatic pressure is displayed as height, here 0.5 m (1.6 ft) for example.)
To accept the value displayed you must first switch to the Edit mode (see the "Editing values" section) and then press the button to save the value.
13 Enter the volume value for the upper calibration
point via the FULL CALIB. parameter, here 1000 l (264 US gal) for example. (The currently measured hydrostatic pressure is displayed as height, here 4.5 m (15 ft) for example.)
3
for example.
3
for example.
P01-xxxxxxxx-05-xx-xx-xx-029
To accept the value displayed you must first switch to the Edit mode (see the "Editing values" section) and then press the button to save the value.
14 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
15 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
16 Result:
The measuring range is set for 0 to 1000 l (264 US gal).
Fig. 8: Calibration with reference pressure –
1 See Table, Steps 10 and 11. 2 See Table, Step 12. 3 See Table, Step 13. 4 See Table, Step 14. 5 See Table, Step 15.
wet calibration
P01-xxxxxxxx-05-xx-xx-xx-030
P01-xxxxxxxx-05-xx-xx-xx-031
For this level mode, the measured variables %, level, volume and mass are available. See also parameter description for OUTPUT UNIT, Page 75.
14 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
WARNING
!
1000 l
0,5 m
4,5 m
0 l
ρ = 1
kg
dm
3
1
3
2
5.3.2 Calibration without reference pressure – dry calibration
Example:
In this example, the volume in a tank should be measured in litres. The maximum volume is 1000 l (264 US gal), and the maximum height is 4.5 m (15 ft). The minimum volume of 0 litres corresponds to a level of 0.5 m (1.6 ft), as the device is mounted below the level lower range value. The density of the fluid is 1 kg/dm
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration i.e. the height and volume values for the lower and upper calibration point must be known.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• The values entered for EMPTY CALIB./FULL CALIB., EMPTY PRESSURE/FULL PRESSURE, EMPTY HEIGHT/FULL HEIGHT and SET LRV/SET URV must have a minimum interval of 1% for the "Level Easy Height" level mode. The value will be rejected with a warning message if the values are too close together. Further limit values are not checked; i.e. the values entered must be appropriate for the sensor and the measuring task so that the measuring device can measure correctly.
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when the container is empty, the MEASURED VALUE parameter does not display zero. To perform a position adjustment see also Page , Table 6: Position adjustment.
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 73, Table 9: LEVEL SELECTION "Level Easy Height"
• For a description of further relevant parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
3
.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
2 If necessary, select "Level Easy Height" level mode
using the LEVEL SELECTION parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: See Page 63
3 On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
Fig. 9: Calibration without reference pressure –
1 See Table, Steps 8 and 9. 2 See Table, Steps 10 and 11. 3 See Table, Steps 12 and 13.
dry calibration
P01-PMC71xxx-19-xx-xx-xx-007
Endress+Hauser 15
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
4.5
0.5
h
[m]
49 441
p
[mbar]
h =
p
r · g
r = 1
g
cm
3
1000
0
V [l]
0.5 4.5
h =
p
ρ · g
h
[m]
20
4
I
[mA]
0 1000
V [l]
Description
4 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
5 Select a volume unit via the OUTPUT UNIT
parameter, here l (litres) for example.
6 Select a height unit via the HEIGHT UNIT parameter,
here m for example.
7 Select the "Dry" option via the CALIBRATION MODE
parameter.
8 Select a density unit via the DENSITY UNIT
parameter, here kg/dm
9 Enter the density of the fluid using the ADJUST
DENSITY parameter, here kg/dm
10 Enter the volume value for the lower calibration
point via the EMPTY CALIB. parameter, here 0 l for example.
11 Enter the height value for the lower calibration point
via the EMPTY HEIGHT parameter, here 0.5 m (1.6 ft) for example.
12 Enter the volume value for the upper calibration
point via the FULL CALIB. parameter, here 1000 l (litres) (264 US gal) for example.
13 Enter the height value for the upper calibration point
via the FULL HEIGHT parameter, here 4.5 m (15 ft) for example.
14 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
15 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
16 Result:
The measuring range is set for 0 to 1000 l (litres) (264 US gal).
3
for example.
3
for example.
P01-xxxxxxxx-05-xx-xx-xx-029
P01-xxxxxxxx-05-xx-xx-xx-032
P01-xxxxxxxx-05-xx-xx-xx-033
Fig. 10: Calibration with reference pressure –
1 See Table, Steps 8 and 9. 2 See Table, Step 10. 3 See Table, Step 11. 4 See Table, Step 12. 5 See Table, Step 13. 6 See Table, Step 14. 7 See Table, Step 15.
wet calibration
For this level mode, the measured variables %, level, volume and mass are available. See also parameter description for OUTPUT UNIT, Page 75.
16 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
WARNING
!
0 mbar
300 mbar
3 m
0 m
5.4 "Level Standard" level selection,
"Linear" level type
5.4.1 Calibration with reference pressure – wet calibration
Example:
In this example, the level in a tank should be measured in m. The maximum level is 3 m (9.8 ft). The pressure range is set to 0 to 300 mbar (4.5 psi).
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• The tank can be filled or emptied.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 77, Table 10: BASIC SETUP – Page 79, Table 11: BASIC SETUP – "Linear" level type.
• For a description of further relevant parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Deltabar S: before configuring the device for your
application, the pressure piping must be cleaned and the device filled with fluid. See Operating Instructions BA00270P.
2 Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
3 If necessary, select the "Level" measuring mode via
the MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
4 If necessary, select "Level Standard" level mode using
the LEVEL SELECTION parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: See Page 63
Fig. 11: Calibration with reference pressure –
1 See table, step 11. 2 See table, step 12.
wet calibration
P01-PMP75xxx-19-xx-xx-xx-008
Endress+Hauser 17
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
3
0
h
[m]
0 300
p
[mbar]
20
4
I
[mA]
0
3
h
[m]
Description
5On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
6 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7 Select the "Linear" option by means of the LEVEL
MODE parameter.
8 Select the "Level" option by means of the LIN.
MEASURAND parameter.
9 Select a level unit via the HEIGHT UNIT parameter,
here m for example.
10 Select the "Wet" option by means of the
CALIBRATION MODE parameter.
11 The pressure for the lower calibration point is
present at the device, here 0 mbar for example.
Select EMPTY CALIB. parameter.
Enter the level value, here 0 m for example. Confirm the value to assign the pressure value present to the lower level value.
12 The pressure for the upper calibration point is
present at the device, here 450 mbar (6.75 psi) for example.
Select FULL CALIB. parameter.
Enter the level value, here 3 m (9.8 ft) for example. Confirm the value to assign the pressure value present to the upper level value.
13 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
14 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
15 Result:
The measuring range is set for 0...3 m (9.8 ft).
Fig. 12: Calibration with reference pressure –
1 See table, step 11. 2 See table, step 12. 3 See table, step 13. 4 See table, step 14.
wet calibration
P01-xxxxxxxx-05-xx-xx-xx-034
P01-xxxxxxxx-05-xx-xx-xx-014
1. You can also perform calibration with reference pressure by means of the QUICK SETUP menu. See Page 65 ff, Table 4: QUICK SETUP menu.
2. You can also specify customer-specific units. See parameter description for PRESS. ENG. UNIT (Page 77), HEIGHT UNIT (Page 79), UNIT VOLUME (Page 80) and MASS UNIT (Page 81).
3. For this level type, the measured variables %, level, volume and mass are available. See Page 79 ff.
4. The EMPTY PRESSURE (Page 83) and FULL PRESSURE (Page 83) parameters display the pressure values belonging to the EMPTY CALIB. and FULL CALIB. parameters.
18 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
WARNING
!
4 m
–0.5 m
r = 1
kg
dm
3
V = 5 m
3
5.4.2 Calibration without reference pressure – dry calibration
Example:
In this example, the volume in a tank should be measured in m
3
m
and the maximum height 4 m (13 ft). The density of the fluid is 1 kg/dm3. The device is
3
. The maximum volume is 5
mounted below the level lower range value.
Prerequisite:
• The measured variable is in direct proportion to the pressure.
• This is a theoretical calibration, i.e. the tank volume, tank height and density of the fluid are known.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when the container is empty, the MEASURED VALUE parameter does not display zero. To perform a position adjustment see also Page 68, Table 6: Position adjustment.
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 77, Table 10: BASIC SETUP – Page 79, Table 11: BASIC SETUP – "Linear" level type.
• For a description of further relevant parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 26: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Select the "Level" measuring mode via the
MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
2 If necessary, select "Level Standard" level mode using
the LEVEL SELECTION parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: See Page 63
3 On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
P01-PMP75xxx-19-xx-xx-xx-003
Fig. 13: Calibration without reference pressure –
1 See table, step 9. 2 See table, step 10. 3 See table, step 11. 4 See table, step 12.
dry calibration
Endress+Hauser 19
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
20
4
I
[mA]
05
V
[m ]
3
Description
4 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
5 Select the "Linear" option by means of the LEVEL
MODE parameter.
6 Select the "Volume" option by means of the LIN.
MEASURAND parameter.
7 Select a volume unit via the UNIT VOLUME
parameter, here m
8 Select the "Dry" option by means of the
CALIBRATION MODE parameter. See also the following note, point 3.
9 Enter the value for density via the ADJUST DENSITY
parameter, here 1 kg/dm
10 Enter the tank volume via the TANK VOLUME
parameter, here 5 m
11 Enter the tank height via the TANK HEIGHT
parameter, here 4 m (13 ft) for example.
12 Enter the level offset via the ZERO POSITION
parameter, here –0.5 m (-1,6 ft) for example.
13 Set the value for the lower current value (4 mA) by
means of the SET LRV parameter.
14 Set the value for the upper current value (20 mA) by
means of the SET URV parameter.
15 Result:
The measuring range is set for 0...5 m
3
for example.
3
for example.
3
for example.
P01-xxxx xxxx-19-xx-xx-xx-012
Fig. 14: Current output calibration
5 See table, step 13. 6 See table, step 14.
3
.
1. For this level type, the measured variables %, level, volume and mass are available. See Page 79 ff.
2. You can also specify customer-specific units. See parameter description for PRESS. ENG. UNIT (Page 77), HEIGHT UNIT (Page 79), UNIT VOLUME (Page 80) and MASS UNIT (Page 81).
3. A level value is assigned to the lower and upper current value by means of the SET LRV (Page 86) and SET URV (Page 86) parameters respectively. Once you have selected the "Dry" calibration mode, the error message A711 "LRV or URV out of edit limits" can appear. The error message goes out as soon as level values which are within the editing limits are entered for the SET LRV and SET URV parameters. By means of the ENTER RESET CODE parameter ( Page 124), you can use the code 2710 to automatically set the SET LRV and SET URV parameters to level values which are within the editing limits.
20 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
WARNING
!
3.5
0
0 350
V
[m ]
3
p
[mbar]

5.5 "Level Standard" level selection, "Pressure Linearized" level type

5.5.1 Semiautomatic entry of the linearisation table
Example:
In this example, the volume in a tank with a conical outlet should be measured in m
3
.
Prerequisite:
• The tank can be filled. The linearisation characteristic must rise continuously.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans
for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P) or
Deltapilot S (BA00332P).
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 77, Table 10: BASIC SETUP – Page 87, Table 11: BASIC SETUP – "Pressure Linearized" level type – Page 104, Table 18: LINEARISATION – on-site operation – Page 107, Table 19: LINEARISATION – Digital communication.
• For a description of further relevant parameters, see
– Page 100, Table 16: EXTENDED SETUP – Page 121, Table 26: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Deltabar S: before configuring the device for your
application, the pressure piping must be cleaned and filled with fluid. See Operating Instructions BA00270P.
2 Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
Carry out basic setup:
3 If necessary, select the "Level" measuring mode via
the MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE
Digital communication: See Page 62
4 If necessary, select "Level Standard" level mode using
the LEVEL SELECTION parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: See Page 63
P01-PMP75xxx-19-xx-xx-xx-002
Endress+Hauser 21
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
Description
5On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
6 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7 Select the "Pressure Linearized" option by means of
the LEVEL MODE parameter. See also the following note, point 3.
8 Select the "Volume" option by means of the LINd.
MEASURAND parameter.
9 Select a volume unit via the UNIT VOLUME
parameter, here m
10 Select HYDR. PRESS MIN. parameter.
Enter the minimum hydrostatic pressure to be expected, here 0 mbar for example.
11 Select HYDR. PRESS MAX
Enter the maximum hydrostatic pressure to be expected.
Carry out linearisation:
12 Change the function group:
Menu path: (GROUP SELECTION ) OPERATING MENU SETTINGS LINEARISATION
13 Select TANK CONTENT MIN parameter.
Specify the minimum tank contents to be expected, here 0 m
3
14 Select TANK CONTENT MAX parameter.
Specify the maximum tank contents to be expected, here 3.5 m
15 On-site display:
Select the "Editor table" option by means of the TABLE SELECTION parameter.
16 Select the "Semiautomatic" option by means of the
LIN. EDIT MODE parameter.
17 Select the "New table" option by means of the EDITOR
TABLE parameter.
3
for example.
for example.
3
for example.
.
22 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
3.5
0
0 350
V
[m ]
3
p
[mbar]
20
4
I
[mA]
0 3.5
V
[m ]
3
Description
18 Enter linearisation table (min. 2 points, max.
32 points).
Fill the tank to the height of the 1st point.
LINE-NUMB: confirm value displayed.
X-VAL.: the hydrostatic pressure present is displayed.
On-site display, Digital communication: The X-VAL. displayed is saved by confirming the Y­value. See following line, Y-VAL.
HART handheld terminal: Confirm X-VAL. displayed.
Y-VAL.: enter the volume value, here 0 m example, and confirm the value.
19 On-site display:
If you want to enter another point for the linearisation table, select the "Next point" option and enter the point as described in step 18. If you want to finish entering the values and activate the linearisation table, select the "Accept input table" option.
Digital communication: You can enter further points for the linearisation table as explained in step 18. Once all the points have been entered, the table must be activated by means of the TAB. ACTIVATE parameter.
19 Result:
The linearisation table has been entered.
3
for
P01-xxxxxxxx-05-xx-xx-xx-015
P01-xxxxxxxx-05-xx-xx-xx-016
Fig. 15: Semiautomatic entry of the linearisation table
1 See table, step 10. 2 See table, step 11. 3 See table, step 13. 4 See table, step 14. 5 See table, steps 15 – 19. 6 See the following note, point 4. 7 See the following note, point 4
Endress+Hauser 23
1. For this level type, the measured variables %, volume and mass are available.
See Page 87 ff.
2. You can also specify customer-specific units. See parameter description for PRESS. ENG.
UNIT (Page 77), HEIGHT UNIT (Page 87), UNIT VOLUME (Page 87) and MASS UNIT (Page 88).
3. Once you have selected the "Pressure Linearized" level type, the warning message "W710
Set span too small. Not allowed." can appear. At this stage the linearisation table already consists of two points as standard. It could be the case that the 2nd value, and thus the highest X-VAL. of the linearisation table, is smaller than the minimum span permitted ( MINIMUM SPAN, Page 120). The message goes out as soon as the highest X-VAL. is larger than the minimum span.
4. A level value is assigned to both the lower and upper current value with the SET LRV
(Page 102) and SET URV (Page 102) parameters. If you enter values for TANK CONTENT MIN (Page 104 or 107) and TANK CONTENT MAX (Page 104 or 107), the SET LRV and SET URV parameters are also changed. If you want to assign values other than those for TANK CONTENT MIN and TANK CONTENT MAX to the lower and upper current values, the desired values must be entered for SET LRV and SET URV.
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
WARNING
!
3.5
0
0 350
V
[m ]
3
p
[mbar]
5.5.2 Manual entry of the linearisation table
Example:
In this example, the volume in a tank with a conical outlet should be measured in m
3
.
Prerequisite:
• This is a theoretical calibration, i.e. the points for the linearisation table are known.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 77, Table 10: BASIC SETUP – Page 87, Table 12: BASIC SETUP – "Pressure Linearized" level type – Page 104, Table 18: LINEARISATION – on-site operation – Page 107, Table 19: LINEARISATION – Digital communication.
• For a description of further relevant parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Perform basic setup as per Section 5.3.1, steps 2 to
11.
Carry out linearisation:
2 Change the function group:
Menu path: (GROUP SELECTION ) OPERATING MENU SETTINGS LINEARISATION
3 Select TANK CONTENT MIN parameter .
Specify the minimum tank contents to be expected, here 0 m
4 Select TANK CONTENT MAX parameter .
Specify the maximum tank contents to be expected, here 3.5 m
3
for example.
3
for example.
P01-PMP75xxx-19-xx-xx-xx-002
24 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
3.5
0
0 350
V
[m ]
3
p
[mbar]
20
4
I
[mA]
0 3.5
V
[m ]
3
Description
5 On-site display:
Select the "Editor table" option by means of the TABLE SELECTION parameter.
6 Select the "Manual" option by means of the LIN. EDIT
MODE parameter.
7 Select the "New table" option by means of the
EDITOR TABLE parameter.
8 Enter linearisation table (min. 2 points, max.
32 points).
LINE-NUMB: confirm value displayed.
X-VAL.: enter the pressure value and confirm.
Y-VAL.: enter the volume value, here 0 m example, and confirm.
9 On-site display
If you want to enter another point for the linearisation table, select the "Next point" option and enter the point as described in step 8. If you want to finish entering the values and activate the linearisation table, select the "Accept input table" option.
Digital communication: You can enter further points for the linearisation table as explained in step 8. Once all the points have been entered, the table must be activated by means of the TAB. ACTIVATE parameter.
10 Result:
The linearisation table has been entered.
3
for
P01-xxxxxxxx-05-xx-xx-xx-015
P01-xxxxxxxx-05-xx-xx-xx-016
Fig. 16: Manual entry of the linearisation table
1 See Section 5.3.1, table, step 9. 2 See Section 5.3.1, table, step 10. 3 See table, step 3. 4 See table, step 4. 5 See table, steps 5 – 9. 6 See the following note, point 4. 7 See the following note, point 4.
Endress+Hauser 25
1. For this level type, the measured variables %, volume and mass are available. See Page 87 ff.
2. You can also specify customer-specific units. See parameter description for PRESS. ENG. UNIT (Page 77), HEIGHT UNIT (Page 87), UNIT VOLUME (Page 87) and MASS UNIT (Page 88).
3. Once you have selected the "Pressure Linearized" level type, the warning message "W710 Set span too small. Not allowed." can appear. At this stage the linearisation table already consists of two points as standard. It could be the case that the 2nd value, and thus the highest X-VAL. of the linearisation table, is smaller than the minimum span permitted ( MINIMUM SPAN, Page 120). The message goes out as soon as the highest X-VAL. is larger than the minimum span.
4. A level value is assigned to both the lower and upper current value with the SET LRV (Page 102) and SET URV (Page 102) parameters. If you enter values for TANK CONTENT MIN (Page 104 or 107) and TANK CONTENT MAX (Page 104 or 107), the SET LRV and SET URV parameters are also changed. If you want to assign values other than those for TANK CONTENT MIN and TANK CONTENT MAX to the lower and upper current values, the desired values must be entered for SET LRV and SET URV.
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
WARNING
!

5.6 "Level Standard" level selection, "Height Linearized" level type

5.6.1 Wet calibration and semiautomatic entry of the linearisation table
Example:
In this example, the height and the volume should be measured at the same time.
Prerequisite:
• The tank can be filled. The linearisation characteristic must rise continuously.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans
for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P) or
Deltapilot S (BA00332P).
• For a description of the parameters mentioned, see
– Page 62, Table 2: MEASURING MODE – Page 68, Table 6: POSITION ADJUSTMENT – Page 77, Table 10: BASIC SETUP – Page 89, Table 13: BASIC SETUP – "Height Linearized" level type – Page 104, Table 18: LINEARISATION – on-site operation – Page 107, Table 19: LINEARISATION – Digital communication.
• For a description of further parameters, see
– Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
1 Deltabar S: before configuring the device for your
application, the pressure piping must be cleaned and filled with fluid. See Operating Instructions BA00270P.
2 Carry out position adjustment if necessary. See Page
68, Table 6: POSITION ADJUSTMENT.
Perform calibration for the 1st measured variable:
3 If necessary, select the "Level" measuring mode via
the MEASURING MODE parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE See Page 62
26 Endress+Hauser
p
[mbar]
➁➃
3
0
h
[m]
0 300
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
Description
4 If necessary, select "Level Standard" level mode using
the LEVEL SELECTION parameter.
On-site display: Menu path: GROUP SELECTION MEASURING MODE "Level" LEVEL SELECTION
Digital communication: Menu path: OPERATING MENU SETTINGS BASIC SETUP MEASURING MODE "Level"  LEVEL SELECTION
5 On-site display:
Select BASIC SETUP function group. Menu path: GROUP SELECTION OPERATING MENU  SETTINGS BASIC SETUP
6 Select a pressure unit via the PRESS. ENG. UNIT
parameter, here mbar for example.
7 Select the "Height Linearized" option by means of the
LEVEL MODE parameter.
8 Select the "Height + Volume" option by means of the
COMB. MEASURAND parameter.
9 Select the unit for the 1st measured value via the
HEIGHT UNIT parameter, here m for example.
10 Select the unit for the 2nd measured variable via the
UNIT VOLUME parameter, here m3 for example.
11 Select LEVEL MIN parameter.
Enter the minimum level to be expected, here 0 m for example.
12 Select LEVEL MAX parameter.
Enter the maximum level to be expected, here 3 m (9.8 ft) for example.
13 Select the "Wet" option via the CALIBRATION MODE
parameter (calibration mode for the 1st measured variable).
14 The pressure for the lower calibration point is
present at the device, here 0 mbar for example.
Select EMPTY CALIB. parameter.
Enter the level value, here 0 m for example. Confirm the value to assign the pressure value present to the lower level value.
15 The pressure for the upper calibration point is
present at the device, here 300 mbar (4.5 psi) for example.
Select FULL CALIB. parameter.
Enter the level value, here 3 m (9.8 ft) for example. Confirm the value to assign the pressure value present to the upper level value.
16 Result:
The calibration for the 1st measured variable is carried out.
Fig. 17: Calibrating the 1st measured variable
1 See table, step 11. 2 See table, step 12. 3 See table, step 14. 4 See Table, step 15.
P01-PMP75xxx-19-xx-xx-xx-004
P01-xxxxxxxx-05-xx-xx-xx-017
Endress+Hauser 27
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
5
0
03
V
[m ]
3
h
[m]
5
0
03
V
[m ]
3
h
[m]
20
4
I
[mA]
05
V
[m ]
3
Description
Perform linearisation (calibration for the 2nd measured variable)
17 Change the function group.
Menu path: (GROUP SELECTION ) OPERATING MENU SETTINGS LINEARISATION
18 Select TANK CONTENT MIN parameter.
Specify the minimum tank contents to be expected, here 0 m
19 Select TANK CONTENT MAX parameter.
Specify the maximum tank contents to be expected, here 5 m
20 On-site display:
Select the "Editor table" option by means of the TABLE SELECTION parameter.
21 Select the "Semiautomatic" option by means of the
LIN. EDIT MODE parameter.
22 Select the "New table" option by means of the EDITOR
TABLE parameter.
23 Enter linearisation table (min. 2 points, max.
32 points).
Fill the tank to the height of the 1st point.
LINE-NUMB: confirm value displayed.
X-VAL.: the hydrostatic pressure present is measured and converted to the corresponding level and displayed.
On-site display, Digital communication: The X-VAL. displayed is saved by confirming the Y-value. See following line, Y-VAL.
HART handheld terminal: Confirm X-VAL. displayed.
Y-VAL.: enter the volume value, here 0 m example, and confirm the value.
24 On-site display
If you want to enter another point for the linearisation table, select the "Next point" option and enter the point as described in step 23. If you want to finish entering the values and activate the linearisation table, select the "Accept input table" option.
Digital communication: You can enter further points for the linearisation table as explained in step 23. Once all the points have been entered, the table must be activated by means of the TAB. ACTIVATE parameter.
25 Result:
– The linearisation table has been entered. – The measured value display and the
– The LEVEL BEFORE LIN parameter displays the
3
for example.
3
for example.
3
for
TANK CONTENT parameter display the 2nd measured value (here the volume).
1st measured value (here the height). See also the following note, point 5.
P01-PMP75xxx-19-xx-xx-xx-005
P01-xxxxxxxx-05-xx-xx-xx-018
P01-xxxxxxxx-05-xx-xx-xx-019
Fig. 18: Calibrating the 2nd measured variable
5 See table, step 18. 6 See table, step 19. 7 See table, steps 20 – 24. 8 See the following note, point 4. 9 See the following note, point 4.
28 Endress+Hauser
Cerabar S/Deltabar S/Deltapilot S 4...20mA HART Level measurement
1. For this level type, the measured variables "Height + %", "Height + Volume", "Height + Mass", "%-Height + %", "%-Height + Volume" and "%-Height + Mass" are available. See Page 87 ff.
2. You can also specify customer-specific units. See parameter description for PRESS. ENG. UNIT (Page 77), HEIGHT UNIT (Page 90), UNIT VOLUME (Page 91) and MASS UNIT (Page 92).
3. Once you have selected the "Pressure Linearized" level type, the warning message "W710 Set span too small. Not allowed." can appear. At this stage the linearisation table already consists of two points as standard. It could be the case that the 2nd value, and thus the highest X-VAL. of the linearisation table, is smaller than the minimum span permitted ( MINIMUM SPAN, Page 120). The message goes out as soon as the highest X-VAL. is larger than the minimum span.
4. A level value is assigned to both the lower and upper current value with the SET LRV (Page 102) and SET URV (Page 102) parameters. You can use the ASSIGN CURRENT parameter (Page 114) to specify whether the current output should depict the 1st or 2nd measured variable. Depending on the setting of the ASSIGN CURRENT parameter, enter the following values for SET LRV and SET URV: – ASSIGN CURRENT = tank content (factory setting) %-value, volume value or
mass value
– ASSIGNMENT = height level value
The following applies for the setting ASSIGN CURRENT "Tank content": If you enter values for TANK CONTENT MIN (Page 104 or 107) and TANK CONTENT MAX (Page 104 or 107), the SET LRV and SET URV parameters are also changed. If you want to assign values other than those for TANK CONTENT MIN and TANK CONTENT MAX to the lower and upper current values, the desired values must be entered for SET LRV and SET URV.
The following applies for the setting ASSIGN CURRENT "Height": If you enter values for LEVEL MIN (Page 93) and LEVEL MAX (Page 93), the SET LRV and SET URV parameters are also changed. If you want to assign values other than those for LEVEL MIN and LEVEL MAX to the lower and upper current values, the desired values must be entered for SET LRV and SET URV.
5. You can use the MENU DESCRIPTOR parameter (Page 111) to specify which measured value should be displayed on the on-site display.
Endress+Hauser 29
Level measurement Cerabar S/Deltabar S/Deltapilot S 4...20mA HART
WARNING
!
5
0
03
V
[m ]
3
h
[m]
5.6.2 Dry calibration and manual entry of the linearisation table
Example:
In this example, the height and the volume should be measured at the same time.
Prerequisite:
• This is a theoretical calibration, i.e. the points for the linearisation table are known.
• A minimum gap of 0.5 % of the distance between two points must be maintained. Spans for the "Pressure linearized" option: HYDR. PRESS MAX. – HYDR. PRESS MIN.; TANK CONTENT MAX. – TANK CONTENT MIN. Spans for the "Height linearized" option: LEVEL MAX – LEVEL MIN; TANK CONTENT MAX. – TANK CONTENT MIN.
• See also Operating Instructions for Deltabar S (BA00270P) or Cerabar S (BA00271P), Section "Level measurement" or Deltapilot S (BA00332P), Section "Level measurement".
• Due to the orientation of the device, there may be a shift in the measured value, i.e. when the container is empty, the MEASURED VALUE parameter does not display zero. To perform a position adjustment see also Page 68, Table 6: Position adjustment.
• For a description of the parameters mentioned, see – Page 62, Table 2: MEASURING MODE – Page 77, Table 10: BASIC SETUP – Page 89, Table 12: BASIC SETUP – "Height Linearized" level type – Page 104, Table 18: LINEARISATION – on-site operation – Page 107, Table 19: LINEARISATION – Digital communication.
• For a description of further parameters, see – Page 100, Table 16: EXTENDED SETUP – Page 121, Table 28: PROCESS VALUES.
Changing the measuring mode can affect the adjustment data!
This situation can result in product overflow.
Check calibration data when the measuring mode is changed.
Description
Perform calibration for the 1st measured variable:
1 Perform calibration as per Section 5.4.2, steps 3 to
12.
2 Select the "Dry" option via the CALIBRATION MODE
parameter (calibration mode for the 1st measured variable).
3 Enter the density of the fluid via the ADJUST
DENSITY parameter, here 1 kg/dm
4 If necessary, enter a level offset via the ZERO
POSITION parameter, here 0 m for example.
5Result:
The calibration for the 1st measured variable is carried out.
3
for example.
P01-PMP75xxx-19-xx-xx-xx-005
30 Endress+Hauser
Loading...
+ 118 hidden pages