Druck PACE5000, PACE6000 Operating Manual

GE
Digital Solutions

Pressure Automated Calibration Equipment

User manual - K0443 Revision A

PACE5000
PACE6000
© 2008 General Electric Company. All Rights Reserved. Specifications are subject to change without notice. GE is a registered trademark of General Electric Company. Other company or product names mentioned in this document may be trademarks or registered trademarks of their respective companies, which are not affiliated with GE.
PACE Pressure Controller User Manual

Introduction

This manual contains Installation and operating instructions for PACE Pneumatic Pressure Controllers.

Safety

The manufacturer has designed this equipment to be safe when operated using the procedures detailed in this manual. Do not use this equipment for any other purpose than that stated, the protection provided by the equipment may be impaired. This publication contains operating and safety instructions that must be followed to make sure of safe operation and to maintain the equipment in a safe condition. The safety instructions are either warnings or cautions issued to protect the user and the equipment from injury or damage. Use qualified * technicians and good engineering practice for all procedures in this publication.

Pressure

Do not apply pressures greater than the maximum working pressure to the equipment.

Toxic Materials

There are no known toxic materials used in construction of this equipment.

Maintenance

The equipment must be maintained using the procedures in this publication. Further manufacturer’s procedures should be carried out by authorized service agents or the manufacturer’s service departments.

Technical Advice

For technical advice contact the manufacturer. * A qualified technician must have the necessary technical knowledge, documentation,
special test equipment and tools to carry out the required work on this equipment.

General Specification

Display LCD: Colour display with touch screen
EMC EN61326
Electrical safety EN 61010-1, UL61010-1, CSA 22.2, No. 61010-1 and IEC 61010-1
PACE 5000: Input range: 100-240V (50/60Hz) 2A, Installation Catergory II, Fuse
Power Supply
Pressure safety Pressure equipment Directive - class: sound engineering practice (SEP)
T2AH250V
PACE 6000: Input range: 100-120/200-240V (50/60Hz) 5A, Installation Cater­gory II, Fuse T5AH250V
[EN] English i K0443 Revision A

Environmental Conditions

Operating Environment Indoor use only Operating temperature 10°C to 50°C (50° to 122°F) Storage temperature -20°C to 70°C (-4° to 158°F) Ingress protection IP20 (EN60529) Operating humidity 5% to 95% RH (non-condensing) Vibration MIL-PRF-28800 Type 2 class 5 Style E/F Operating altitude Maximum 2000 metres (6560ft) Pollution degree 2

Abbreviations

The following abbreviations are used in this manual; the abbreviations are the same in the singular and plural.
a Absolute min Minute or minimum a.c Alternating current mm millimetre ALT Altitude mV millivolts ASCII American Standard Code for Information Interchange MWP Maximum working pressure BSP British pipe thread No Number CAS Calibrated airspeed NPT National Pipe Thread
CSK Countersunk PACE
d.c. Direct current Para. Paragraph DPI Digital Pressure Instrument PDCR Pressure transducer etc. And so on PED Pressure equipment directive e.g. For example psi Pounds per square inch Fig. Figure PTX Pressure transmitter ft Foot ROC Rate of climb (vertical speed) g Gauge RS232 Serial communications standard GPIB General purpose interface bus Rt CAS Rate of Calibrated airspeed Hg Mercury Rt MACH Rate of MACH Hz Hertz Rx Receive data
IAS Indicated airspeed SCPI
IDOS Intelligent digital output sensor (GE product) SDS Sales data sheet
i.e. That is SELV
IEEE 488
in Inch UUT Unit under test kg kilogram V Volts
kts knots +ve Positive m Metre -ve Negative mA milliampere °C Degrees Celsius max Maximum °F Degrees Fahrenheit mbar Millibar
Institute of Electrical and Electronic Engineers standard 488 (for programmable devices with a digital interface)
Tx Transmit data
Pressure automated calibration equipment
Standard commands for programmable instruments
Separated (or Safety) extra low voltage
K0443 Revision A ii [EN] English
PACE Pressure Controller User Manual

Related publications

K0447 PACE 5000/6000 User Guide and Safety Instructions
K0450 PACE Series Calibration Manual
K0476 Pressure Control Module User Guide and Safety Instructions
K0472 Remote Communications Manual
K0469 Heritage Communications Manual - Instrument Emulation

Symbols

This equipment meets the requirements of all relevant European safety directives. The equipment carries the CE mark.
This symbol, on the equipment, indicates that the user should read the user manual.
This symbol, on the equipment, indicates a warning and that the user should refer to the user manual.
Ce symbole, sur l’instrument, indique que l’utilisateur doit consulter le manuel d’utili­sation. Ce symbole, dans le manuel, indique une situation dangereuse.
This symbol warns the user of the danger of electric shock. Ce symbole alerte l’utilisateur sur le danger de choc électrique.
Do not dispose of this product as household waste. Use an approved organisation that collects and/or recycles waste electrical and electronic equipment. For more information, contact one of these:
- Our customer service department: www.gemeasurement.com
- Your local government office.
[EN] English iii K0443 Revision A
WARNINGS
TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE. REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE REAR PANEL.
K0443 Revision A iv [EN] English
PACE Pressure Controller User Manual
Pressure units and conversion factors
Pressure units Factor (hPa) Pressure units Factor (hPa)
mbar 1.0
bar 1000.0
cmH
mH
O @ 20°C
2
O @ 20°C
2
0.978903642
97.8903642
2
Pa (N/m
)
hPa 1.0
0.01
kg/m
kg/cm
2
2
0.0980665
980.665
kPa 10.0 torr 1.333223684
MPa 10000.0 atm 1013.25
mmHg @ 0°C 1.333223874 psi 68.94757293
cmHg @ 0°C 13.33223874
mHg @ 0°C 1333.223874
inHg @ 0°C 33.86388640341
mmH
cmH
mH
mmH
O @ 4°C
2
O @ 4°C
2
O @ 4°C
2
O @ 20°C
2
0.0980665
0.980665
98.0665
0.097890364
inH
inH
inH20 @ 60°F
ftH2O @ 4°C
ftH2O @ 20°C
ftH20 @ 60°F
2
lb/ft
O @ 4°C
2
O @ 20°C
2
0.4788025898
2.4908891
2.486413
2.487641558
29.8906692
29.836983
29.8516987

Unit Conversion

Convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS 2, calculate as follows:
VALUE 2 = VALUE 1 x FACTOR 1
FACTOR 2
[EN] English v K0443 Revision A

CONTENTS

Section Title Page
1 Description 1-1
1.1 Introduction 1-1
2 Installation 2-1
2.1 Packaging 2-1
2.2 Packaging for Storage and Transportation 2-1
2.3 Preparation for Use 2-1
2.4 Pneumatic connections 2-2
2.5 Rack-mount option 2-8
2.6 Electrical connections 2-9
3 OPERATION 3-1
3.1 Preparation 3-1
3.2 Power-up Sequence 3-2
3.3 Measure Mode 3-3
3.4 Control Mode 3-6
3.5 Operation and Example Procedures 3-10
3.6 Global Set-up Selections 3-14
3.7 Barometric Reference Option 3-15
3.8 Supervisor Set-up 3-16
3.9 Instrument Status 3-17
4 MAINTENANCE 4-1
4.1 Introduction 4-1
4.2 Visual inspection 4-1
K0443 Revision A vi [EN] English
PACE Pressure Controller User Manual
4.3 Cleaning 4-1
4.4 Test 4-1
4.5 Calibration 4-2
4.6 Replacement parts 4-2
4.7 Fuse replacement 4-2
4.8 Filter replacement 4-4
4.9 Pressure module replacement 4-5
5 TESTING AND FAULT FINDING 5-1
5.1 Introduction 5-1
5.2 Standard Serviceability Test 5-1
5.3 Fault Finding 5-3
5.4 Approved Service Agents 5-3
6 REFERENCE AND SPECIFICATION 6-1
6.1 Installation notes 6-1
6.2 Operational Requirements 6-4
6.3 Icons 6-6
6.4 Measure Set-up 6-9
6.5 Control Set-up 6-10
6.6 Global Set-up 6-11
6.7 Supervisor Set-up 6-12
6.8 Options 6-24
6.9 Calibration 6-41
6.10 Communications - Instrument Emulation 6-41
6.11 Specification 6-41
6.12 Return Goods/Material Procedure 6-41
6.13 Packaging Procedure 6-42
[EN] English vii K0443 Revision A
6.14 Vacuum System Parts 6-43
K0443 Revision A viii [EN] English
PACE Pressure Controller User Manual

1Description

1.1 Introduction

The PACE5000 single-channel and PACE6000 single/dual-channel, Pressure Automated Calibration Equipment measures and controls pneumatic pressures and displays, on a touch-screen, the pressure measurement and controller status. The touch-screen enables selections and settings in both measure and control modes. The instrument can be operated remotely through communication interfaces.
Figure 1-1 PACE5000 General view
Figure 1-2 PACE6000 General view
The rear of the instrument houses all the electrical and pneumatic output and input connections. The electrical connections provide an a.c. power supply, serial and parallel communication interfaces, d.c. output and logic input and output. The system pneumatic controller module contains a positive and negative pressure supply port, an output port, vent port and reference port.
The instrument can be mounted in a standard 19 inch rack system (rack-mount option).
[EN] English 1 - 1 K0443 Revision A
1 Description
Figure 1-3 PACE5000 Rear view
Figure 1-4 PACE6000 Rear view
Options available are detailed in the product datasheet.
Information and notes on applications (Ref: Reference and Specification, Section 6) or
www.gemeasurement.com
K0443 Revision A 1 - 2 [EN] English
PACE Pressure Controller User Manual

2 Installation

2.1 Packaging

Check the contents of the PACE5000/6000 packaging with the list that follows:

Packaging List

i) PACE5000 or PACE6000 Pressure Controller. ii) Cable, power supply. iii) User guide and CD (UD-0001) containing the full documentation suite. iv) Pneumatic Control Module blanking plate (keep this plate for future use).
CAUTIONS
After removing a control module, use a blanking plate to keep the flow of cooling air.
After unpacking an instrument that has been in cold conditions allow time to stabilise and any condensation to evaporate.

2.2 Packaging for Storage or Transportation

To store or return the instrument for calibration/repair do the procedures that follow:
1. Pack the instrument (Ref: Reference and Specification, Section 6.8).
2. Return the instrument for calibration/repair complete the return goods procedure (Ref:
Reference and Specification, Section 6.8).
Note: The procedure above applies to the pressure control module as a separate item.

2.3 Preparation for Use

The instrument can be used as a:
Free-standing instrument positioned on a horizontal surface.
Rack-mounted in a standard 19 inch rack using the rack-mount option kit (Ref: Section
2.5, Rack-mount option).
For free-standing instruments, the feet on the front of the base can be used elevate the instrument to a better viewing angle.
Note: Do not obstruct the air cooling outlet on the underside of the instrument and allow a free
flow of air around the instrument, especially at high ambient temperatures.
[EN] English 2 - 1 K0443 Revision A
2 Installation
Parallel threads must be used. Female thread type is parallel thread to ISO228/1 (DIN ISO228/1, JIS B0202) G1/8
Tapered threads NOT allowed.
PACE pressure port
PACE rear panel
9
8
PACE rear panel
PACE pressure port

2.4 Pneumatic connections

WARNINGS TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE. REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
PARALLEL THREADS MUST BE USED. FEMALE THREAD TYPE IS PARALLEL THREAD TO ISO228/1 (DIN ISO228/1, JIS B0202) G1/8.
TAPERED THREADS NOT ALLOWED.
K0443 Revision A 2 - 2 [EN] English
PACE Pressure Controller User Manual
2
1
1 Connector 2 Bonded seal
Connection Port
Input
Output
Refer to the data sheet for a complete range of adaptors.

Pressure supply (Ref: Figure 2-1, Pneumatic connections)

1. The pressure supply must be clean, dry gas, nitrogen or air and at the correct pressure,
(Ref: Section 6, Reference and Specification).
2. Make sure the user systems can be isolated and vented.
3. Connect pressure and vacuum supplies to the SUPPLY + and SUPPLY - connection ports.
4. Connect the Unit Under Test (UUT) to the required output connection port.
supply + supply ­Output
Reference
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)Vent
Figure 2-1, Pneumatic Connections
Note: For systems requiring NPT connections, please order optional NPT adaptors.
[EN] English 2 - 3 K0443 Revision A
2 Installation

Installation

The safety of any system incorporating the equipment is the responsibility of the assembler of the system.
The instrument requires a positive pressure supply, instruments operating in an absolute range or negative pressure range require a vacuum supply.
A vacuum supply should be used for a fast response for instruments operating near atmospheric pressure.
For dual channel operation two independent pressure and vacuum supplies can be used.
Figure 2-2, Pressure module rear view
Note: When using two pressure modules (Ref: Section 4.9, Pressure module replacement)
make sure that:
The module with the highest pressure rating is fitted to the right hand side
Module 1 position as viewed from the rear of the product refer to (Ref: Fig 2-2)
If two modules have the same pressure rating, make sure that the module with
the higher serial number is fitted to the right hand side Module 1 position as viewed from the rear of the product.
Note: All pneumatic connections must comply with the Pressure Equipment Directive (PED)
or other regional pressure standards.
Note: When connecting the output ports of two pressure modules together make sure both
are either:
below 70 bar
OR
between 100 to 210 bar.
To prevent over-pressurisation of pneumatic parts and maintain compliance with the PED do not mix categories.
K0443 Revision A 2 - 4 [EN] English
PACE Pressure Controller User Manual
OUTPUT
1
2
3
4
5
6
7
+
5
9
a
-
SUPPLY
VENT REF
8
14
14

Supply equipment

Pneumatic supplies should have isolation valves and, where necessary, conditioning equipment.
The positive pressure supply should be regulated to between 110% of the full-scale pressure range and MWP stated on the control module.
To protect the instrument from over-pressure a suitable protection device (such as a relief valve or bursting disc) must be fitted to prevent over pressurization.
On instruments without a negative supply, the positive pressure discharges from the system to atmosphere through the negative supply port. Pipe the negative port to a safe discharge area or fit a diffuser to the negative port.
During system pressure vent operations, the pressure discharges from the system to atmosphere through the negative and vent ports. Pipe both ports to a safe discharge area or fit a diffuser to the negative port.

Pneumatic Connection Examples (Ref: Figures 2-3, 2-4 and 2-5)

The examples that follow show a single channel connection detail, using supply equipment described above.
Figure 2-3, Pneumatic Connections without vacuum supply
1) Pressure source 2) Conditioner 3) Filter
4) Regulate to between 110% full-scale and MWP 5) Diffuser*
6) Unit under test 7) Optional reservoir 8) Protection device
9) Optional differential connection 14) Manual external vent valves
a) Atmosphere
[EN] English 2 - 5 K0443 Revision A
2 Installation
1
2
3
4
5
6
7
8
9
13
7
10
12
11
a
OUTPUT
+
-
SUPPLY
VENT REF
14
14
Note: (Ref: Section 6, Reference and Specification) for details of other system components.
Figure 2-4, Pneumatic Connections with vacuum supply
1) Pressure source 2) Conditioner 3) Filter
4) Regulate to between 110% full-scale and MWP 5) Diffuser*
6) Unit under test 7) Optional reservoir 8) Protection device
9) Optional differential connection 10) Oil mist trap
11) Vacuum source 12) Normally open electrical release valve 13) Check valve **
14) Manual external vent valves a) Atmosphere
Note: The PACE option IO-VAC-SYS Vacuum System Check Valve Kit should be used in the
vacuum line mounted near the PACE CM -ve port to exhaust most HP gas directly to atmosphere. The vacuum buffer volume needs to be rated at least to the highest system pressure.
Note: (Ref: Section 6, Reference and Specification) for details of other system components
K0443 Revision A 2 - 6 [EN] English
PACE Pressure Controller User Manual
1
2
3
4
7
8
5
6
9
13
7
10
12
11
+
-
a
14
14
SUPPLY
OUTPUT
Figure 2-5, Pneumatic Connections with negative gauge pressure generator
1) Pressure source 2) Conditioner 3) Filter
4) Regulate to between 110% full-scale and MWP 5) Diffuser *
6) Unit under test 7) Optional reservoir
9) Optional differential connection
10) Vacuum generator 11) Source pressure (regulated compressed air supply)
12) Exhaust to atmosphere 13) Check valve ** 14) Manual external vent valves
8) Protection device
a) Atmosphere
VENT REF
Notes: (Ref: Section 6 Reference and Specification) for details of other system components.
* High pressure gas exhaust - depending on pressure range. ** Optional vacuum system kit, allows the -ve port gas to be directly discharged to
atmosphere, by-passing the vacuum pump.
Optimum controller transient response and minimum time to set-point may be
degraded if either the pneumatic supply or vacuum system has restricted flow. Installing a reservoir volume, which has larger capacity than the load volume, located in close proximity to the controller supply ports can improve the controller response.
Optional negative gauge pressure generator kit.
To protect the control module, for ranges of 70 bar and above from over-pressure a
suitable protection device (such as a relief valve or bursting disc) must be fitted to limit the applied supply pressure to below the MWP.
Optional differential connection kit.
[EN] English 2 - 7 K0443 Revision A
2 Installation

2.5 Rack-mount option (Ref: Figure 2-6)

General

There must be enough space at the rear of the instrument for all the cables and pipes. The length of the cables and pipes must allow for the removal and fitment of the instrument. The cooling air of the instrument must not be obstructed. Allow a free flow of air through the equipment rack and around the instrument, especially at high ambient temperatures.
Figure 2-6 Rack-mounting
K0443 Revision A 2 - 8 [EN] English
PACE Pressure Controller User Manual

Procedure

1. Locate bracket in rack assembly
2. Remove the four M3x10mm countersunk screws from each of the instrument side panels.
3. Locate the two brackets
4. Secure with the four countersunk screws.
5. Support the instrument and connect the cables and pipes.
6. Refer to the electrical connections below before fitting the instrument into the rack.
7. Temporarily locate the two spigots * to each side of the equipment rack.
8. Locate and slide the instrument into the rack.
9. Locate the instrument on the spigots*.
10. Secure the instrument in the equipment rack with two of the screws and washers (supplied).
11. Remove the two spigots* and replace with the remaining two screws and washers (supplied).

2.6 Electrical connections

WARNINGS
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY PROTECTIVE SAFETY GROUND.
on each side of the instrument.
.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE REAR PANEL.

Connecting (Ref: Figure 2.7 Electrical connections)

1. Install an accessible power isolator to use as the disconnecting device in the power supply circuit.
2. For the power input power supply range, Installation Category and VA (Ref:
Introduction, General specification table).
Note: The power must be supplied by a fused or overload-protected power supply.
3. Connect the power supply to the instrument.
4. Switch the power supply on.
5. Check that the front panel display shows the power-up sequence (Ref: section 3.2,
Power-up sequence).
[EN] English 2 - 9 K0443 Revision A
2 Installation
3
2
1
1 IEC power connector 2 Fuse carrier 3 Fuse
Figure 2-7 Electrical Connections
Pressure Control Module Input and Output Connectors

24V DC Output @ 100mA maximum

4-way connector: pin “+” = +24 Vdc
pin “-” = 0 Vdc
An integral self-resetting fuse protects this output.

Logic (switch) Input

4-way connector: Input
Output
This facility can be used to trigger the instrument from a pressure switch contact during the Pressure Switch Task (Ref Section 3.4, Control Mode).
Connections are not polarised and can be connected either way. Integral opto-isolators protect this input circuit.
This facility can be energised by external SELV compliant equipment.
K0443 Revision A 2 - 10 [EN] English
PACE Pressure Controller User Manual
1
3
4
5
6
Communication Connections (Ref: Fig 2-8 Communication Connectors)
Connect the applicable connectors into the rear panel communications ports and, if appropriate, secure with the captive screws.
Note: Refer to the data sheet for a list of optional communication ports. Refer to the data sheet for a list of standard shipped communication ports. Set the required parameters in Supervisor Setup/communications menu, (Ref: Section 3.8,
Supervisor set-up).
Figure 2-8, Communication Connectors
1 RS232 3 IEEE488 4 USB B 5 USB A 6 Ethernet
[EN] English 2 - 11 K0443 Revision A
2 Installation
RS232 Interface
When using the RS232 interface, a cable must be connected directly from the instrument to a suitable port on the computer in a ‘point to point’ link. The pin connections for the 9-pin D­type, RS232 connector and the relationship between the instrument and the RS232 control signals, together with device interconnection interface is shown in Table 2-1. The instrument is configured as Data Circuit Terminating Equipment (DCE).
Instrument Control Line Computer/Printer
Instrument
Function
RxD (I/P) 3 TxD 3 2
TxD (O/P) 2 RxD 2 3
GND 5  GND 5 7
CTS (I/P) 7 RTS 7 4
RTS (O/P) 8 CTS 8 5
Pulled
high
internally
Not
connected
Pulled
high
internally
Equipment
chassis
Connector
9-way
D-type
Pin No.
1
4 DTR 4 20
6 
Connector
shell
Signal
Direction
 Cable Screen - 1
RS232
Terminology
RLSD (DCD)
DSR
DCE Ready
9-way D-type Pin No.
18
66
Table 2-1, RS232 Connections

Handshaking connections

Connector Type
25-way
D-type Pin No.
Software handshaking use: TXD, RXD and GND. Hardware handshaking use: TXD, RXD, GND, CTS, RTS and DTR.
K0443 Revision A 2 - 12 [EN] English
PACE Pressure Controller User Manual
2
3

IEEE 488 Interface

The interface complies with IEEE 488 standard. The IEEE 488 parallel interface connects a computer/controller to one or more PACE
instruments and other instruments. Up to 30 instruments can be connected through a high-speed data bus to the computer/
controller.
Note: The length of each IEEE 488 cable must be less than 3 metres to comply with the EMC requirements (Ref: Section 6,

Single Unit Installation (Ref: Figure 2-9)

1. Connect an IEEE 488 connector/cable assembly to the rear panel of the instrument.
2. Connect the other end of the connector/cable assembly to the IEEE 488 connector on the controller/computer.
3. Change the IEEE 488 communication parameters (Ref: Section 6.7, Supervisor set-up).

Multiple Unit Installation (Ref: Figure 2-9)

Reference and Specification).
To install multiple units use stacking plugs to
1
link the first instrument and second instrument as follows.
1 Connector to rear panel of first
instrument (Ref Illustration).
2 Connector from controller/computer
(Ref Illustration).
3 Connector to rear panel of second
instrument (Ref Illustration).
4. Connect the IEEE 488 connector on the controller/computer and the other connector into the next instrument.
5. Repeat this procedure for all the instruments in the system.
6. Use the Supervisor set-up (communications) menu on each instrument to set-up the required communication parameters (Ref: Section 3.8, Supervisor set-up).
[EN] English 2 - 13 K0443 Revision A
2 Installation
IEEE 488
ADDRESS 1
ADDRESS 2
ADDRESS N
(30 maximum)
678
DAV (DATA VALID)
NRFD (NOT READY FOR DATA)
NDAC (NO DATA ACCEPTED)
59101117
EOI (END OF IDENTIFY)
IFC (INTERFACE CLEAR)
SRQ (SERVICE REQUEST)
ATN (ATTENTION)
REN (REMOTE ENABLE)
123413
DIO1
DIO2
DIO3
DIO4
DIO5
141516
DIO6
DIO7
DIO8
1819202122
GND (6)
GND (7)
GND (8)
GND (9)
GND (10)
23
24
GND (11)
GND
12 CHASSIS/FRAMECHASSIS/FRAME
DATA /
STATUS
BAR
0V (GND)
Figure 2-9 - IEEE 488 Connection
K0443 Revision A 2 - 14 [EN] English
PACE Pressure Controller User Manual

3Operation

This section contains quick reference charts detailing all the available functions and the set­up menu.

3.1 Preparation

Make sure the electrical cables and pneumatic pipes comply with the installation requirements (Ref: Section 2, Installation). Before use do the following:
1. If necessary, do the maintenance task (Ref: Section 4, Maintenance).
2. For bench-top, single instrument operation do the following:
a. Connect the instrument to the electrical supply.
b. Inspect the pneumatic hoses for damage, ingress of dirt and moisture.
3. Before use, the instrument should be tested.
4. Review and become familiar with the procedure before starting a process on a component or system.
Note: The touch-screen can be permanently damaged by sharp objects.
[EN] English 3 - 1 K0443 Revision A
3 Operation
1
2
3
4
56

3.2 Power-up sequence

The following sequences of operation shows the instrument display. Note: The following sequence is an example, the values and selections displayed depend on
the range(s) and options enabled in the instrument (Ref: Touch screen areas).
1. Set the power supply to ON:
2. The display shows the power-up sequence.
3. The instrument carries out a self-test. If the test finds a fault, the display shows an
error (Ref: Fault Finding and Testing, Section
5).
4. If the self-test is successful the system enables the touch screen and changes to measure mode.
5. The touch screen shows the measured pressure in the parameters selected in set-up.
6. The instrument is now ready for use.
Note: The PACE 6000 shows a single display
(default) this is the left hand pressure control module. Change to dual display in Global Set­up/Display.
Touch screen areas
Measure set-up
1
Status (touch to enter control set-up
5
Control/Measure
2
Nudge down
6
3
Nudge up
Set-point entry
4
K0443 Revision A 3 - 2 [EN] English
PACE Pressure Controller User Manual
PACE5000
PACE6000
2
3
4
56
7
1
2
3
4
56

3.3 Measure mode

7
1
Pressure measurement of current selected sensor in current selected pressure measurement units
1
Current enabled functions
2
Nudge up, changed in control set-up
4
Nudge down, changed in control set-up
6
Control/measure selection
3
Current set-point value, change with numeric keys
5
Status area, changed in global set-up
7
Display Icons
Auto zero Control mode with overshoot Percentage
Reference level difference (gas head correction)
Control mode with no overshoot
Timed zero
Control mode active Filter pressure reading Tare enabled
Control mode passive Linear rate Zero
Control mode gauge Maximum rate
[EN] English 3 - 3 K0443 Revision A
3 Operation
current pressure reading to zero, offset stored for current range.
allows selection of available pressure ranges. Ê
PACE5000
PACE6000
-
-
select from list of available pressure measurement units.
-
Absolute range selection available when barometric option installed. Auto-range only available with two-channel instruments.
Ê
% Filter pressure reading Tare
Pressure zero
Range
Process
Task
Units
Global set-up
Set-up zero
Status
Measure set-up
More
Timed Auto Tare
Next page of menu options. Loops from last page to first page.
Stores settings and exits set-up.
see 3.6:-
next page:-
Exit set-up.
K0443 Revision A 3 - 4 [EN] English
PACE Pressure Controller User Manual
AUTO-RANGE (only available with two-channel instruments) Note: Not all Auto-Range and task functions are available using the remote
communications, this allows more flexibility to the remote programmer.

Controller Off – Increasing Set-point

With both controllers in Measure mode, if a set-point within the range of the lower ranged controller is entered and Control is then selected the lower ranged controller controls to the entered set-point.
With both controllers in Measure mode, if a set-point above the range of the lower ranged controller is entered and Control is then selected, then the range is changed to the higher ranged controller and this then controls to the entered set-point.

Controller Off – Decreasing Set-point

With both controllers in Measure mode, if a set-point within the range of the higher ranged controller is entered and Control is then selected the higher ranged controller controls to the entered set-point.
With both controllers in Measure mode, if a set-point above the range of the lower ranged controller is entered and Control is then selected, then the higher ranged controller will control to this set-point. When the controller is in limits then the range is changed to the lower ranged controller and this then controls to the entered set-point .

Controller On – Increasing Set-point

With the lower ranged controller in Control mode, if a set-point within the range of the lower ranged controller is entered then the lower ranged controller controls to the entered set-point.
If the set-point is increased to above the lower range but still within the higher range then the lower range controller switches off and that the higher range controller switches on and controls to the entered set-point.

Controller On – Decreasing Set-point

With the higher ranged controller in Control mode, if a set-point within the range of the higher ranged controller is entered then the higher ranged controller controls to the entered set-point.
If the set-point is decreased to within the lower range then the higher range controller will control to this set-point. When the controller is in limits then the range is changed to the lower ranged controller and this then controls to the entered set-point .
[EN] English 3 - 5 K0443 Revision A
Loading...
+ 68 hidden pages