This manual contains Installation and operating instructions for PACE Pneumatic Pressure
Controllers.
Safety
The manufacturer has designed this equipment to be safe when operated using the procedures
detailed in this manual. Do not use this equipment for any other purpose than that stated, the
protection provided by the equipment may be impaired.
This publication contains operating and safety instructions that must be followed to make sure of
safe operation and to maintain the equipment in a safe condition. The safety instructions are either
warnings or cautions issued to protect the user and the equipment from injury or damage.
Use qualified * technicians and good engineering practice for all procedures in this publication.
Pressure
Do not apply pressures greater than the maximum working pressure to the equipment.
Toxic Materials
There are no known toxic materials used in construction of this equipment.
Maintenance
The equipment must be maintained using the procedures in this publication. Further
manufacturer’s procedures should be carried out by authorized service agents or the
manufacturer’s service departments.
Technical Advice
For technical advice contact the manufacturer.
*A qualified technician must have the necessary technical knowledge, documentation,
special test equipment and tools to carry out the required work on this equipment.
PACE 5000: Input range: 100-240V (50/60Hz) 2A, Installation Catergory II, Fuse
Power Supply
Pressure safetyPressure equipment Directive - class: sound engineering practice (SEP)
T2AH250V
PACE 6000: Input range: 100-120/200-240V (50/60Hz) 5A, Installation Catergory II, Fuse T5AH250V
[EN] EnglishiK0443 Revision A
Environmental Conditions
Operating EnvironmentIndoor use only
Operating temperature10°C to 50°C (50° to 122°F)
Storage temperature-20°C to 70°C (-4° to 158°F)
Ingress protectionIP20 (EN60529)
Operating humidity5% to 95% RH (non-condensing)
VibrationMIL-PRF-28800 Type 2 class 5 Style E/F
Operating altitudeMaximum 2000 metres (6560ft)
Pollution degree2
Abbreviations
The following abbreviations are used in this manual; the abbreviations are the same in the singular and plural.
aAbsoluteminMinute or minimum
a.cAlternating currentmmmillimetre
ALTAltitudemVmillivolts
ASCIIAmerican Standard Code for Information Interchange MWPMaximum working pressure
BSPBritish pipe threadNoNumber
CASCalibrated airspeedNPTNational Pipe Thread
CSK CountersunkPACE
d.c.Direct currentPara.Paragraph
DPIDigital Pressure InstrumentPDCRPressure transducer
etc.And so onPEDPressure equipment directive
e.g.For examplepsiPounds per square inch
Fig.FigurePTXPressure transmitter
ftFootROCRate of climb (vertical speed)
gGaugeRS232Serial communications standard
GPIBGeneral purpose interface busRt CASRate of Calibrated airspeed
HgMercuryRt MACH Rate of MACH
HzHertzRxReceive data
IASIndicated airspeedSCPI
IDOSIntelligent digital output sensor (GE product)SDSSales data sheet
This equipment meets the requirements of all relevant European safety directives.
The equipment carries the CE mark.
This symbol, on the equipment, indicates that the user should read the user manual.
This symbol, on the equipment, indicates a warning and that the user should refer to
the user manual.
Ce symbole, sur l’instrument, indique que l’utilisateur doit consulter le manuel d’utilisation. Ce symbole, dans le manuel, indique une situation dangereuse.
This symbol warns the user of the danger of electric shock.
Ce symbole alerte l’utilisateur sur le danger de choc électrique.
Do not dispose of this product as household waste. Use an approved organisation
that collects and/or recycles waste electrical and electronic equipment. For more
information, contact one of these:
- Our customer service department: www.gemeasurement.com
- Your local government office.
[EN] EnglishiiiK0443 Revision A
WARNINGS
TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE.
REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED
FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY
PROTECTIVE SAFETY GROUND.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE
REAR PANEL.
Convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS 2,
calculate as follows:
VALUE 2=VALUE 1 x FACTOR 1
FACTOR 2
[EN] EnglishvK0443 Revision A
CONTENTS
SectionTitlePage
1Description1-1
1.1Introduction1-1
2Installation2-1
2.1Packaging2-1
2.2Packaging for Storage and Transportation2-1
2.3Preparation for Use2-1
2.4Pneumatic connections2-2
2.5Rack-mount option2-8
2.6Electrical connections2-9
3OPERATION3-1
3.1Preparation3-1
3.2Power-up Sequence3-2
3.3Measure Mode3-3
3.4Control Mode3-6
3.5Operation and Example Procedures3-10
3.6Global Set-up Selections3-14
3.7Barometric Reference Option3-15
3.8Supervisor Set-up3-16
3.9Instrument Status3-17
4MAINTENANCE4-1
4.1Introduction4-1
4.2Visual inspection4-1
K0443 Revision Avi[EN] English
PACE Pressure Controller User Manual
4.3Cleaning4-1
4.4Test4-1
4.5Calibration4-2
4.6Replacement parts4-2
4.7Fuse replacement4-2
4.8Filter replacement4-4
4.9Pressure module replacement4-5
5TESTING AND FAULT FINDING5-1
5.1Introduction5-1
5.2Standard Serviceability Test5-1
5.3Fault Finding5-3
5.4Approved Service Agents5-3
6REFERENCE AND SPECIFICATION6-1
6.1Installation notes6-1
6.2Operational Requirements6-4
6.3Icons6-6
6.4Measure Set-up6-9
6.5Control Set-up6-10
6.6Global Set-up6-11
6.7Supervisor Set-up6-12
6.8Options6-24
6.9Calibration6-41
6.10Communications - Instrument Emulation6-41
6.11Specification6-41
6.12Return Goods/Material Procedure6-41
6.13Packaging Procedure6-42
[EN] EnglishviiK0443 Revision A
6.14Vacuum System Parts6-43
K0443 Revision Aviii[EN] English
PACE Pressure Controller User Manual
1Description
1.1Introduction
The PACE5000 single-channel and PACE6000 single/dual-channel, Pressure Automated
Calibration Equipment measures and controls pneumatic pressures and displays, on a
touch-screen, the pressure measurement and controller status. The touch-screen enables
selections and settings in both measure and control modes. The instrument can be operated
remotely through communication interfaces.
Figure 1-1 PACE5000 General view
Figure 1-2 PACE6000 General view
The rear of the instrument houses all the electrical and pneumatic output and input
connections. The electrical connections provide an a.c. power supply, serial and parallel
communication interfaces, d.c. output and logic input and output. The system pneumatic
controller module contains a positive and negative pressure supply port, an output port,
vent port and reference port.
The instrument can be mounted in a standard 19 inch rack system (rack-mount option).
[EN] English1 - 1K0443 Revision A
1 Description
Figure 1-3 PACE5000 Rear view
Figure 1-4 PACE6000 Rear view
Options available are detailed in the product datasheet.
Information and notes on applications (Ref: Reference and Specification, Section 6) or
www.gemeasurement.com
K0443 Revision A1 - 2[EN] English
PACE Pressure Controller User Manual
2Installation
2.1Packaging
Check the contents of the PACE5000/6000 packaging with the list that follows:
Packaging List
i)PACE5000 or PACE6000 Pressure Controller.
ii)Cable, power supply.
iii)User guide and CD (UD-0001) containing the full documentation suite.
iv)Pneumatic Control Module blanking plate (keep this plate for future use).
CAUTIONS
After removing a control module, use a blanking plate to keep the flow of cooling air.
After unpacking an instrument that has been in cold conditions allow time to stabilise
and any condensation to evaporate.
2.2Packaging for Storage or Transportation
To store or return the instrument for calibration/repair do the procedures that follow:
1.Pack the instrument (Ref: Reference and Specification, Section 6.8).
2.Return the instrument for calibration/repair complete the return goods procedure (Ref:
Reference and Specification, Section 6.8).
Note: The procedure above applies to the pressure control module as a separate item.
2.3Preparation for Use
The instrument can be used as a:
•Free-standing instrument positioned on a horizontal surface.
•Rack-mounted in a standard 19 inch rack using the rack-mount option kit (Ref: Section
2.5, Rack-mount option).
For free-standing instruments, the feet on the front of the base can be used elevate the
instrument to a better viewing angle.
Note: Do not obstruct the air cooling outlet on the underside of the instrument and allow a free
flow of air around the instrument, especially at high ambient temperatures.
[EN] English2 - 1K0443 Revision A
2 Installation
Parallel threads must be used.
Female thread type is parallel thread to
ISO228/1 (DIN ISO228/1, JIS B0202)
G1/8
Tapered threads NOT allowed.
PACE pressure port
PACE rear
panel
9
8
PACE rear
panel
PACE pressure port
2.4Pneumatic connections
WARNINGS
TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE.
REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED
FITTINGS AND EQUIPMENT.
DO NOT EXCEED THE MAXIMUM WORKING PRESSURE OF THE INSTRUMENT.
THIS EQUIPMENT IS NOT RATED FOR OXYGEN USE.
PARALLEL THREADS MUST BE USED. FEMALE THREAD TYPE IS PARALLEL THREAD TO
ISO228/1 (DIN ISO228/1, JIS B0202) G1/8.
TAPERED THREADS NOT ALLOWED.
K0443 Revision A2 - 2[EN] English
PACE Pressure Controller User Manual
2
1
1 Connector
2 Bonded seal
ConnectionPort
Input
Output
Refer to the data sheet for a complete range of adaptors.
1.The pressure supply must be clean, dry gas, nitrogen or air and at the correct pressure,
(Ref: Section 6, Reference and Specification).
2.Make sure the user systems can be isolated and vented.
3.Connect pressure and vacuum supplies to the SUPPLY + and SUPPLY - connection
ports.
4.Connect the Unit Under Test (UUT) to the required output connection port.
supply +
supply Output
Reference
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)
ISO228/1 G 1/8 parallel threads (DIN ISO228/1, JIS B0202)Vent
Figure 2-1, Pneumatic Connections
Note: For systems requiring NPT connections, please order optional NPT adaptors.
[EN] English2 - 3K0443 Revision A
2 Installation
Installation
The safety of any system incorporating the equipment is the responsibility of the assembler
of the system.
The instrument requires a positive pressure supply, instruments operating in an absolute
range or negative pressure range require a vacuum supply.
A vacuum supply should be used for a fast response for instruments operating near
atmospheric pressure.
For dual channel operation two independent pressure and vacuum supplies can be used.
Figure 2-2, Pressure module rear view
Note: When using two pressure modules (Ref: Section 4.9, Pressure module replacement)
make sure that:
•The module with the highest pressure rating is fitted to the right hand side
Module 1 position as viewed from the rear of the product refer to (Ref: Fig 2-2)
•If two modules have the same pressure rating, make sure that the module with
the higher serial number is fitted to the right hand side Module 1 position as
viewed from the rear of the product.
Note: All pneumatic connections must comply with the Pressure Equipment Directive (PED)
or other regional pressure standards.
Note: When connecting the output ports of two pressure modules together make sure both
are either:
•below 70 bar
OR
•between 100 to 210 bar.
To prevent over-pressurisation of pneumatic parts and maintain compliance with the PED do
not mix categories.
K0443 Revision A2 - 4[EN] English
PACE Pressure Controller User Manual
OUTPUT
1
2
3
4
5
6
7
+
5
9
a
-
SUPPLY
VENT REF
8
14
14
Supply equipment
Pneumatic supplies should have isolation valves and, where necessary, conditioning
equipment.
The positive pressure supply should be regulated to between 110% of the full-scale pressure
range and MWP stated on the control module.
To protect the instrument from over-pressure a suitable protection device (such as a relief
valve or bursting disc) must be fitted to prevent over pressurization.
On instruments without a negative supply, the positive pressure discharges from the system
to atmosphere through the negative supply port. Pipe the negative port to a safe discharge
area or fit a diffuser to the negative port.
During system pressure vent operations, the pressure discharges from the system to
atmosphere through the negative and vent ports. Pipe both ports to a safe discharge area or
fit a diffuser to the negative port.
Pneumatic Connection Examples (Ref: Figures 2-3, 2-4 and 2-5)
The examples that follow show a single channel connection detail, using supply equipment
described above.
Figure 2-3, Pneumatic Connections without vacuum supply
1) Pressure source2) Conditioner3) Filter
4) Regulate to between 110% full-scale and MWP5) Diffuser*
6) Unit under test7) Optional reservoir †8) Protection device
Note: The PACE option IO-VAC-SYS Vacuum System Check Valve Kit should be used in the
vacuum line mounted near the PACE CM -ve port to exhaust most HP gas directly to
atmosphere. The vacuum buffer volume needs to be rated at least to the highest
system pressure.
Note: (Ref: Section 6, Reference and Specification) for details of other system components
K0443 Revision A2 - 6[EN] English
PACE Pressure Controller User Manual
1
2
3
4
7
8
5
6
9
13
7
10
12
11
+
-
a
14
14
SUPPLY
OUTPUT
Figure 2-5, Pneumatic Connections with negative gauge pressure generator
1) Pressure source2) Conditioner3) Filter
4) Regulate to between 110% full-scale and MWP5) Diffuser *
6) Unit under test7) Optional reservoir †
9) Optional differential connection
10) Vacuum generator ‡11) Source pressure (regulated compressed air supply)
Notes:(Ref: Section 6 Reference and Specification) for details of other system components.
*High pressure gas exhaust - depending on pressure range.
**Optional vacuum system kit, allows the -ve port gas to be directly discharged to
atmosphere, by-passing the vacuum pump.
†Optimum controller transient response and minimum time to set-point may be
degraded if either the pneumatic supply or vacuum system has restricted flow.
Installing a reservoir volume, which has larger capacity than the load volume, located
in close proximity to the controller supply ports can improve the controller response.
‡Optional negative gauge pressure generator kit.
To protect the control module, for ranges of 70 bar and above from over-pressure a
suitable protection device (such as a relief valve or bursting disc) must be fitted to limit
the applied supply pressure to below the MWP.
Optional differential connection kit.
[EN] English2 - 7K0443 Revision A
2 Installation
2.5Rack-mount option (Ref: Figure 2-6)
General
There must be enough space at the rear of the instrument for all the cables and pipes. The
length of the cables and pipes must allow for the removal and fitment of the instrument. The
cooling air of the instrument must not be obstructed. Allow a free flow of air through the
equipment rack and around the instrument, especially at high ambient temperatures.
Figure 2-6 Rack-mounting
K0443 Revision A2 - 8[EN] English
PACE Pressure Controller User Manual
Procedure
1.Locate bracket in rack assembly
2.Remove the four M3x10mm countersunk screws from each of the instrument side
panels.
3.Locate the two brackets
4.Secure with the four countersunk screws.
5.Support the instrument and connect the cables and pipes.
6.Refer to the electrical connections below before fitting the instrument into the rack.
7.Temporarily locate the two spigots * to each side of the equipment rack.
8.Locate and slide the instrument into the rack.
9.Locate the instrument on the spigots*.
10.Secure the instrument in the equipment rack with two of the screws and washers
(supplied).
11.Remove the two spigots* and replace with the remaining two screws and washers
(supplied).
2.6Electrical connections
WARNINGS
THE GROUND LEAD OF THE INSTRUMENT MUST BE CONNECTED TO THE AC SUPPLY
PROTECTIVE SAFETY GROUND.
on each side of the instrument.
.
ISOLATE THE POWER SUPPLY BEFORE MAKING ANY ELECTRICAL CONNECTIONS TO THE
REAR PANEL.
1.Install an accessible power isolator to use as the disconnecting device in the power
supply circuit.
2.For the power input power supply range, Installation Category and VA (Ref:
Introduction, General specification table).
Note: The power must be supplied by a fused or overload-protected power supply.
3.Connect the power supply to the instrument.
4.Switch the power supply on.
5.Check that the front panel display shows the power-up sequence (Ref: section 3.2,
Power-up sequence).
[EN] English2 - 9K0443 Revision A
2 Installation
3
2
1
1 IEC power connector
2 Fuse carrier
3 Fuse
Figure 2-7 Electrical Connections
Pressure Control Module Input and Output Connectors
24V DC Output @ 100mA maximum
4-way connector:pin “+”= +24 Vdc
pin “-”= 0 Vdc
An integral self-resetting fuse protects this output.
Logic (switch) Input
4-way connector: Input
Output
This facility can be used to trigger the instrument from a pressure switch contact during the
Pressure Switch Task (Ref Section 3.4, Control Mode).
Connections are not polarised and can be connected either way. Integral opto-isolators
protect this input circuit.
This facility can be energised by external SELV compliant equipment.
K0443 Revision A2 - 10[EN] English
PACE Pressure Controller User Manual
1
3
4
5
6
Communication Connections (Ref: Fig 2-8 Communication Connectors)
Connect the applicable connectors into the rear panel communications ports and, if
appropriate, secure with the captive screws.
Note: Refer to the data sheet for a list of optional communication ports.
Refer to the data sheet for a list of standard shipped communication ports.
Set the required parameters in Supervisor Setup/communications menu, (Ref: Section 3.8,
Supervisor set-up).
Figure 2-8, Communication Connectors
1RS2323 IEEE4884USB B
5USB A6Ethernet
[EN] English2 - 11K0443 Revision A
2 Installation
RS232 Interface
When using the RS232 interface, a cable must be connected directly from the instrument to
a suitable port on the computer in a ‘point to point’ link. The pin connections for the 9-pin Dtype, RS232 connector and the relationship between the instrument and the RS232 control
signals, together with device interconnection interface is shown in Table 2-1. The instrument
is configured as Data Circuit Terminating Equipment (DCE).
InstrumentControl LineComputer/Printer
Instrument
Function
RxD (I/P)3TxD32
TxD (O/P)2RxD23
GND5GND57
CTS (I/P)7RTS74
RTS (O/P)8CTS85
Pulled
high
internally
Not
connected
Pulled
high
internally
Equipment
chassis
Connector
9-way
D-type
Pin No.
1
4DTR420
6
Connector
shell
Signal
Direction
Cable Screen-1
RS232
Terminology
RLSD
(DCD)
DSR
DCE Ready
9-way
D-type
Pin No.
18
66
Table 2-1, RS232 Connections
Handshaking connections
Connector Type
25-way
D-type
Pin No.
Software handshaking use:TXD, RXD and GND.
Hardware handshaking use: TXD, RXD, GND, CTS, RTS and DTR.
K0443 Revision A2 - 12[EN] English
PACE Pressure Controller User Manual
2
3
IEEE 488 Interface
The interface complies with IEEE 488 standard.
The IEEE 488 parallel interface connects a computer/controller to one or more PACE
instruments and other instruments.
Up to 30 instruments can be connected through a high-speed data bus to the computer/
controller.
Note: The length of each IEEE 488 cable must be less than 3 metres to comply with the EMC
requirements (Ref: Section 6,
Single Unit Installation (Ref: Figure 2-9)
1.Connect an IEEE 488 connector/cable assembly to the rear panel of the instrument.
2.Connect the other end of the connector/cable assembly to the IEEE 488 connector on
the controller/computer.
3.Change the IEEE 488 communication parameters (Ref: Section 6.7, Supervisor set-up).
Multiple Unit Installation (Ref: Figure 2-9)
Reference and Specification).
To install multiple units use stacking plugs to
1
link the first instrument and second
instrument as follows.
1Connector to rear panel of first
instrument (Ref Illustration).
2Connector from controller/computer
(Ref Illustration).
3Connector to rear panel of second
instrument (Ref Illustration).
4.Connect the IEEE 488 connector on
the controller/computer and the other
connector into the next instrument.
5.Repeat this procedure for all the instruments in the system.
6.Use the Supervisor set-up (communications) menu on each instrument to set-up the
required communication parameters (Ref: Section 3.8, Supervisor set-up).
[EN] English2 - 13K0443 Revision A
2 Installation
IEEE 488
ADDRESS 1
ADDRESS 2
ADDRESS N
(30 maximum)
678
DAV (DATA VALID)
NRFD (NOT READY FOR DATA)
NDAC (NO DATA ACCEPTED)
59101117
EOI (END OF IDENTIFY)
IFC (INTERFACE CLEAR)
SRQ (SERVICE REQUEST)
ATN (ATTENTION)
REN (REMOTE ENABLE)
123413
DIO1
DIO2
DIO3
DIO4
DIO5
141516
DIO6
DIO7
DIO8
1819202122
GND (6)
GND (7)
GND (8)
GND (9)
GND (10)
23
24
GND (11)
GND
12 CHASSIS/FRAMECHASSIS/FRAME
DATA /
STATUS
BAR
0V (GND)
Figure 2-9 - IEEE 488 Connection
K0443 Revision A2 - 14[EN] English
PACE Pressure Controller User Manual
3Operation
This section contains quick reference charts detailing all the available functions and the setup menu.
3.1Preparation
Make sure the electrical cables and pneumatic pipes comply with the installation
requirements (Ref: Section 2, Installation).
Before use do the following:
1.If necessary, do the maintenance task (Ref: Section 4, Maintenance).
2.For bench-top, single instrument operation do the following:
a.Connect the instrument to the electrical supply.
b.Inspect the pneumatic hoses for damage, ingress of dirt and moisture.
3.Before use, the instrument should be tested.
4.Review and become familiar with the procedure before starting a process on a
component or system.
Note: The touch-screen can be permanently damaged by sharp objects.
[EN] English3 - 1K0443 Revision A
3 Operation
1
2
3
4
56
3.2Power-up sequence
The following sequences of operation shows the instrument display.
Note: The following sequence is an example, the values and selections displayed depend on
the range(s) and options enabled in the instrument (Ref: Touch screen areas).
1.Set the power supply to ON:
2.The display shows the power-up sequence.
3.The instrument carries out a self-test.
If the test finds a fault, the display shows an
error (Ref: Fault Finding and Testing, Section
5).
4.If the self-test is successful the system
enables the touch screen and changes to
measure mode.
5.The touch screen shows the measured
pressure in the parameters selected in set-up.
6.The instrument is now ready for use.
Note: The PACE 6000 shows a single display
(default) this is the left hand pressure control
module. Change to dual display in Global Setup/Display.
Touch screen areas
Measure set-up
1
Status (touch to enter control set-up
5
Control/Measure
2
Nudge down
6
3
Nudge up
Set-point entry
4
K0443 Revision A3 - 2[EN] English
PACE Pressure Controller User Manual
PACE5000
PACE6000
2
3
4
56
7
1
2
3
4
56
3.3Measure mode
7
1
Pressure measurement of current selected sensor in current selected pressure measurement units
1
Current enabled functions
2
Nudge up, changed in control set-up
4
Nudge down, changed in control set-up
6
Control/measure selection
3
Current set-point value, change with numeric keys
5
Status area, changed in global set-up
7
Display Icons
Auto zeroControl mode with overshootPercentage
Reference level difference
(gas head correction)
Control mode with no
overshoot
Timed zero
Control mode activeFilter pressure readingTare enabled
Control mode passiveLinear rateZero
Control mode gaugeMaximum rate
[EN] English3 - 3K0443 Revision A
3 Operation
current pressure reading to zero,
offset stored for current range.
allows selection of available pressure
ranges. Ê
PACE5000
PACE6000
-
-
select from list of available
pressure measurement
units.
-
Absolute range selection available when
barometric option installed.
Auto-range only available with two-channel
instruments.
Ê
%
Filter pressure reading
Tare
Pressure zero
Range
Process
Task
Units
Global set-up
Set-up zero
Status
Measure set-up
More
Timed
Auto
Tare
Next page of menu options. Loops
from last page to first page.
Stores settings and exits
set-up.
see 3.6:-
next page:-
Exit set-up.
K0443 Revision A3 - 4[EN] English
PACE Pressure Controller User Manual
AUTO-RANGE (only available with two-channel instruments)
Note: Not all Auto-Range and task functions are available using the remote
communications, this allows more flexibility to the remote programmer.
Controller Off – Increasing Set-point
With both controllers in Measure mode, if a set-point within the range of the lower ranged
controller is entered and Control is then selected the lower ranged controller controls to the
entered set-point.
With both controllers in Measure mode, if a set-point above the range of the lower ranged
controller is entered and Control is then selected, then the range is changed to the higher
ranged controller and this then controls to the entered set-point.
Controller Off – Decreasing Set-point
With both controllers in Measure mode, if a set-point within the range of the higher ranged
controller is entered and Control is then selected the higher ranged controller controls to the
entered set-point.
With both controllers in Measure mode, if a set-point above the range of the lower ranged
controller is entered and Control is then selected, then the higher ranged controller will
control to this set-point. When the controller is in limits then the range is changed to the
lower ranged controller and this then controls to the entered set-point .
Controller On – Increasing Set-point
With the lower ranged controller in Control mode, if a set-point within the range of the lower
ranged controller is entered then the lower ranged controller controls to the entered set-point.
If the set-point is increased to above the lower range but still within the higher range then the
lower range controller switches off and that the higher range controller switches on and controls
to the entered set-point.
Controller On – Decreasing Set-point
With the higher ranged controller in Control mode, if a set-point within the range of the
higher ranged controller is entered then the higher ranged controller controls to the entered
set-point.
If the set-point is decreased to within the lower range then the higher range controller will
control to this set-point. When the controller is in limits then the range is changed to the
lower ranged controller and this then controls to the entered set-point .
[EN] English3 - 5K0443 Revision A
Loading...
+ 68 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.