Druck PACE1001 Data sheet

GE
Measurement & Control
PACE1001 Barometer
Precision Barometric Indicator and Recorder
A high precision Druck indicator and recorder, designed for barometric
monitoring in laboratories, airelds and other applications.
Features
Utilises GE’s unique range of Resonant Pressure Sensor
• Pressure ranges 750-1150 mbar (10.9-16.7 psi, 75-115
Kpa), absolute
• Choice of precision up to 0.025 mbar/0.0003625 psi
Data Logging as standard with on screen replay
Selectable numeric or graphic display
High resolution touch screen operation
Intuitive icon driven task menu structure
• Aireld Task as standard: Display QFE, QFF or QNH in
pressure units or as altitude in feet or meters
• Leak Test Option
• Switch Test/Analogue Output option
• Switch Test/Voltage Free Contact option
• RS232, IEEE connectivity, Ethernet and USB as standard
• Min/Max/Average display
Compatible with software packages
• 28 selectable pressure units plus 4 user dened units
• Various service support options available
PACE 1001 Precision Barometric Indicator and Recorder
The new PACE 1001 precision barometric indicator brings together the latest measurement technology from GE to oer an elegant, fast, exible and economical solution to barometric indication & monitoring.
PACE 1001 Barometer employs digitally characterized pressure sensors which oer the quality, stability and precision associated with this latest generation of resonant devices.
The PACE1001 Barometer oers three levels of precision to accommodate specication and budget requirements, optional external IDOS universal pressure modules can be connected to add more pressure measurement capability.
The colour touch screen display of the PACE 1001 Barometer can be congured by a user to indicate pressure measurement simultaneously from up to three sensors either graphically or numerically.
The measurements displayed can be logged to an internal memory over a customized time period, sample rate and trigger ready for replay on the display, can also be saved to a USB storage device or exported to a PC.
Pressure sensor measurements can be re-transmitted via one of the communication ports.
The Aireld task is supplied as standard with the PACE1001 Barometer and enables the user to display QFE, QFF or QNH in pressure units or as altitude in feet or meters.
PACE 1001 Barometer Options
Leak Test
Leak Test measures leak rate over the measure dwell time.
At the start of the test, the instrument measures the applied
test pressure of the user system. The instrument records the pressure change during measure dwell time. On completion the
display shows the leak rate results, with leak rate per second or
per minute in the current pressure units selected.
Switch Test – standard with the Analogue Output or Voltage Free Contacts option
Switch Test automates the testing of pressure switch devices. Following the test, displayed is the pressure at which contacts open and close and the switch hysteresis. Switch Test task can also be set to capture switch toggle max, min and average values.
Analogue Output
The Analogue Output can be programmed via the setup menu
screen to output a signal proportional to the instrument range
selected. This allows the instrument to interface with PC or PLC I/O cards, remote displays, chart recorders or other data logging equipment.
Users can select outputs of 0 to 10 V, 0 to 5 V, -5 to 5 V and 0/4 to 20 mA. Precision with respect to host instrument measured pressure 0.05% FS over the host instrument operating temperature range, variable update rate to 80 readings per second. The option is programmable between minimum and FS pressure for proportional output against pressure.
Volt Free Contacts
Volt Free Contacts enable control of peripheral devices such as vacuum pumps, ovens, etc. Each VFC option has three independent volt-free NO/NC relay contacts. A number of
conditions can be set within a PACE1001 instrument to trigger
a relay toggling its contacts.
2
Specications
Pressure measurement
Barometer pressure range 750-1150 mbar absolute, 10.9-16.7 psi absolute, 75-115 kPa absolute
Over range indication: 10% above mbar/bar full scale pressure range.
Pressure media: Dry, oil free, non-corrosive gas, air
Display
Panel ¼ VGA wide format 4.3 inch colour graphics LCD c/w integral touch screen
Comms update rate 8 times per second
Display update rate readout
Pressure units mbar, bar, Pa(N/m2), hPa, kPa, MPa, mmHg @ 0°C, cmHg @ 0°C, mHg @ 0°C,, inHg @ 0°C,
2 times per second
± 9999999
mmH2O @ 4°C, cmH2O @ 4°C, mH2O @ 4°C, mmH2O @ 20°C, cmH2O @20°C, mH2O @ 20°C,
kg/m2, kg/cm2, torr, atm, psi, lb/ft2, inH2O @ 4°C, inH2O @ 20°C, inH2O @ 60°F, ftH2O @ 4°C,
ftH2O @ 20°C, ftH2O @ 60°F, User Dened 1, User Dened 2, User Dened 3, User Dened 4
(Feet & Meters in Aireld task)
Performance over the calibrated temperature range
PACE1001 Barometer Standard Precision
PACE1001 Barometer High Precision
PACE1001 Barometer Premium Precision
PACE1001 Barometer long term stability
Standard precision is 0.10 mbar or 0.001450 psi. Includes non-linearity, hysteresis, repeatability and temperature eects between 15°C (59°F) and 45°C (113°F).
High precision is 0.05 mbar or 0.000725 psi. Includes non-linearity, hysteresis, repeatability and temperature eects between 15°C (59°F) and 45°C (113°F).
Premium precision is 0.025 mbar or 0.0003625 psi. Includes non-linearity, hysteresis, repeatability and temperature eects between 15°C (59°F) and 45°C (113°F)
0.1 mbar, 0.001450 psi/annum
Electrical
Power Supply 90 VAC to 130 VAC @ 47 to 63 Hz & 180 VAC to 260 VAC @ 47 to 63 Hz. 15 VA
Communications
Communication RS232, USB and IEEE-488, SCPI, DPI141, DPI142 and DPI150 emulation. LabVIEW drivers
Ethernet (VXI-II & Sockets using SCPI).
Data log
Data log Display screen shot stored in CSV format, onto memory card or external USB storage
device. User dened update rate from 1 second
3