This manual contains Installation and operating instructions for PACE Pressure Indicators.
Safety
The manufacturer has designed this equipment to be safe when operated using the
procedures detailed in this manual. Do not use this equipment for any other purpose than
that stated, the protection provided by the equipment may be impaired.
This publication contains operating and safety instructions that must be followed to make
sure of safe operation and to maintain the equipment in a safe condition. The safety
instructions are either warnings or cautions issued to protect the user and the equipment
from injury or damage.
Use qualified * technicians and good engineering practice for all procedures in this
publication.
Pressure
Do not apply pressures greater than the maximum working pressure to the equipment.
Toxic Materials
There are no known toxic materials used in construction of this equipment.
Maintenance
The equipment must be maintained using the procedures in this publication. Further
manufacturer’s procedures should be done by an authorized service agents or the
manufacturer’s service departments.
Technical Advice
For technical advice contact the manufacturer.
* A qualified technician must have the necessary technical knowledge, documentation,
special test equipment and tools to carry out the required work on this equipment.
General Specification
DisplayLCD: Colour display with touch-screen
EMCEN 61326
Electrical safetyEN 61010-1, UL61010-1, CSA 22.2, No. 61010-1 and IEC61010-1
Power adaptorInput range: 100 - 240VAC, 50 to 60Hz, 700mA. Installation category II
Pressure safetyPressure Equipment Directive - class: sound engineering practice (SEP)
K0470 Issue No. 2i
Environmental conditions
For indoor use only
Operating temperature10°C to 50°C (50° to 122°F)
Storage temperature-20°C to 70°C (-4° to 158°F)
Ingress protectionIP20 (EN60529)
Operating humidity5% to 95% RH (non-condensing)
VibrationMIL-PRF-28800 Type 2 class 5 style E/F
Operating altitudeMaximum 2000 metres (6560ft)
Pollution degree2
Abbreviations
The following abbreviations are used in this manual; abbreviations are the same in the singular and plural.
aAbsoluteminMinute or minimum
a.cAlternating currentmmmillimetre
ALTAltitudemVmillivolts
ASCIIAmerican Standard Code for Information Interchange MWPMaximum working pressure
BSPBritish pipe threadNoNumber
CASCalibrated airspeedNPTNational Pipe Thread
CSK CountersunkPACE
d.c.Direct currentPara.Paragraph
DPIDigital Pressure InstrumentPDCRPressure transducer
etc.And so onPEDPressure equipment directive
e.g.For examplepsiPounds per square inch
Fig.FigurePTXPressure transmitter
ftFootROCRate of climb (vertical speed)
gGaugeRS232Serial communications standard
GPIBGeneral purpose interface busRt CASRate of Calibrated airspeed
HgMercuryRt MACH Rate of MACH
HzHertzRxReceive data
IASIndicated airspeedSCPI
IDOSIntelligent digital output sensor (GE product)SDSSales data sheet
Institute of Electrical and Electronic Engineers standard
488 (for programmable devices with a digital interface)
TxTransmit data
Pressure automated calibration
equipment
Standard commands for
programmable instruments
Separated (or Safety) extra low
voltage
K0470 Issue No. 2ii
Related publications
K0467 User Guide and Safety Instructions
K0469 PACE Heritage Communications Manual
K0450 PACE Series Calibration Manual
K0472 PACE Series SCPI Manual
Symbols
The equipment contains the following symbols to identify hazards.
This equipment meets the requirements of all relevant European safety directives.
The equipment carries the CE mark.
This symbol, on the instrument, indicates that the user should refer to the user
manual. This symbol, in this manual, indicates a hazardous operation.
Ce symbole, sur l’instrument, indique que l’utilisateur doit consulter le manuel
d’utilisation. Ce symbole, dans le manuel, indique une situation dangereuse.
This symbol, on the instrument, indicates do not throw-away in domestic bin,
hazardous material, dispose correctly in accordance with local regulations.
Convert FROM pressure VALUE 1 in pressure UNITS 1 TO pressure VALUE 2 in pressure UNITS
2, calculate as follows:
VALUE 2=VALUE 1 x FACTOR 1
FACTOR 2
K0470 Issue No. 2iv
CONTENTS
SectionTitlePage
1Description1-1
1.1Introduction1-1
2Installation2-1
2.1Packaging2-1
2.2Packaging for Storage and Transportation2-1
2.3Preparation for Use2-1
2.4Connecting the Instrument2-2
2.5Mounting kits2-5
2.6Electrical connections2-9
3OPERATION3-1
3.1Preparation3-1
3.2Power-up Sequence3-1
3.3Measure Mode3-2
3.4Operation and Example Procedures3-3
3.5Global Set-up Selections3-5
3.6Supervisor Set-up3-7
3.7Instrument Status3-8
4MAINTENANCE4-1
4.1Introduction4-1
4.2Visual inspection4-1
4.3Cleaning4-1
K0470 Issue No. 2v
4.4Test4-1
4.5Calibration4-1
5TESTING AND FAULT FINDING5-1
5.1Introduction5-1
5.2Standard Serviceability Test5-1
5.3Fault Finding5-1
5.4Approved Service Agents5-1
6REFERENCE AND SPECIFICATION6-1
6.1Installation notes6-1
6.2Reference port6-2
6.3Icons.6-3
6.4Measure Set-up6-5
6.5Status6-6
6.6Global Set-up6-6
6.7Supervisor Set-up6-7
6.8Calibration6-9
6.9Specification6-10
6.10Options6-10
6.11Installation and Ancillary Equipment6-10
6.12Return Goods/Material Procedure6-11
6-13Packaging Procedure6-12
K0470 Issue No. 2vi
PACE1000 Pressure Indicator User Manual
1Description
1.1Introduction
The PACE Pressure Indicator measures both pneumatic and hydraulic pressures and
displays, on a colour touch-screen, the measured pressure and instrument status. The
touch-screen enables selections and settings in measuring modes. The instrument can be
operated remotely through communication interfaces.
Figure 1-1 PACE1000 General view
The rear of the instrument houses all the electrical and pressure input connections. The
electrical connections provide a power supply, serial and parallel communication interfaces,
pressure ports and option ports.
The instrument can be used as follows:
• Free-standing instrument positioned on a horizontal surface.
• Rack-mounted in a standard 19 inch rack using the rack-mount option kit.
• Panel mount using the panel-mount option kit.
K0470 Issue No. 21 - 1
1 Description
Options available for the PACE1000 refer to the SDS.
Information and notes on applications (Ref: Reference and Specification, Section 6) or
www.gesensinginspection.com.
K0470 Issue No. 21 - 2
PACE1000 Pressure Indicator User Manual
2Installation
2.1Packaging
Check the contents of the PACE1000 packaging with the list that follows:
Packaging List - PACE1000
i)PACE1000 Pressure Indicator.
ii)Adaptor, power supply (GE part number 191-370).
iii)User guide and safety instructions, and CD containing the full documentation suite.
iv)Calibration certificate.
2.2Packaging for Storage or Transportation
To store or return the instrument for calibration/repair do the procedures that follow:
1.Pack the instrument (Ref: Reference and Specification, Section 6.13).
2.Return the instrument for calibration/repair complete the return goods procedure (Ref:
Reference and Specification, Section 6.12).
2.3Preparation for Use
The instrument can be used as a:
•Free-standing instrument positioned on a horizontal surface
•Panel-mounted using the panel-mount option kit (Ref: Section 2.5)
•Rack-mounted in a standard 19 inch rack using the rack-mount option kit (Ref: Section
2.5).
For free-standing instruments, the feet on the front of the base can be used elevate the instrument
to a better viewing angle.
Note: Allow a free flow of air around the instrument, especially at high ambient temperatures.
K0470 Issue No. 22 - 1
2 Installation
2.4Connectioning the Instrument
WARNINGS
TURN OFF THE SOURCE PRESSURE(S) AND CAREFULLY VENT THE PRESSURE LINES BEFORE
DISCONNECTING OR CONNECTING THE PRESSURE LINES. PROCEED WITH CARE.
ONLY USE EQUIPMENT WITH THE CORRECT PRESSURE RATING.
BEFORE APPLYING PRESSURE, EXAMINE ALL FITTINGS AND EQUIPMENT FOR DAMAGE.
REPLACE ALL DAMAGED FITTINGS AND EQUIPMENT. DO NOT USE ANY DAMAGED
FITTINGS AND EQUIPMENT.
Pneumatic Pressure (Figure 2-1)
1.Refer to the SDS for the correct pressure mediums to be used.
2.Connect the Unit Under Test (UUT) to the required connection port.
Note: For instruments with NPT connections, use applicable pressure sealing.
recommended method alternative method below 100 bar
Figure 2-1, Sealing Pneumatic Connections
Pneumatic connections
Connection
Input
Reference
G 1/8
G 1/8
For examples of adaptors (Ref: page 2 - 4).
Input Pressure (Figure 2)
1.Make sure the user systems can be isolated and vented.
2.Connect the Unit Under Test (UUT) to the output connection port.
Note: For instruments with NPT connections, use applicable pressure sealing.
K0470 Issue No. 22 - 2
PACE1000 Pressure Indicator User Manual
2
1
2) Bonded seal
1) G 1/8 connector
Figure 2-2, Pneumatic Connections
Installation
The instrument connects to the Unit Under Test.
Input Pressure and Equipment
The pressure should not exceed 1.25 x full-scale or MWP stated on the rear panel of the
instrument.
To protect the instrument from over-pressure a suitable protection device (such as a relief valve or
bursting disc) must be fitted to prevent over pressurization.
Pneumatic Connection
WARNING
PRESSURE RANGES >210 bar (3000 psi) ARE ONLY RATED FOR HYDRAULIC USE.
Cautions
Do not exceed the maximum pressures stated in the appropriate Component Manual for
the unit under test.
Reduce pressure at a controlled rate when venting to atmosphere.
Carefully de-pressurize all pipes to atmospheric pressure before disconnecting and
connecting to the unit under test.
Connections
1. Switch off the power supply before connecting or disconnecting the instrument.
2.Use the appropriate sealing method for all pressure connections.
K0470 Issue No. 22 - 3
2 Installation
recommended method
Method of connection G1/8
Adaptors
Refer to the data sheet SDS0014 for the range of adaptors.
K0470 Issue No. 22 - 4
Loading...
+ 42 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.