The DPI 605 can also be supplied as an indicator having
wider pressure range capability but no pressure
generation and adjustment facilities. Vent valve for
pressure release fitted as standard.
An external pressure source such as bottled gas can be
utilized to enable an indicator to be configured as a
calibrator.
Operating Pressure Range
Any FS range can be supplied between the ranges listed
below:
• 1 to 5000 psig (70 mbar to 350 bar)
• 5 to 5000 psia (350 mbar to 350 bar)
• Gauge versions available with -14.7 psig (-1 barg)
calibration.
Accuracy
Ranges 1 to 1000 psi (70 mbar to 70 bar)
• ±0.05% of reading between 20% and 100% FS
• ±0.01% of FS between 0 and 20% FS
• For -14.7 psi (-1 bar) calibrations, FS range is defined
as the compound range
Ranges 1001 to 5000 psi (71 to 350 bar)
• ± 0.08% of reading between 20% and 100% FS
• ± 0.016% of FS between 0 and 20% FS
Figures include 90 day stability.
Temperature Effects
1 to 5000 psi (70 mbar to 350 bar) internal sensors,
±0.0015% of reading/°F (±0.003% of reading °C).
Options
(A) Barometric Reference
A precision resonant pressure transducer (RPT)
monitors barometric pressure and, when used in
conjunction with the normal gauge sensors, allows
absolute ranges to be measured.
The RPT provides an accuracy of ±0.0044 inHg
(±0.15 mbar) over the ranges 23 to 33 inHg
(800 to 1150 mbar) absolute and over a temperature
range of 50°F to 85°F (10°C to 30°C). Over extended
15°F to 105°F (-10°C to 40°C) error 0.0088 inHg
(0.3 mbar). Accuracy includes one year stability.
(B) External Transducers
In addition to the internal transducers, the
instrument can be configured with ten external piezo
resistive sensors (one at a time), and one resonant
sensor. Supplied assembled with the 12-pin
connector for direct interfacing with the instrument.
Any FS can be specified between the ranges:
• 0 to 1 psi (0 to 70 mbar) to 0 to 1000 psi
(0 to 70 bar) gauge
• 0 to 1001 psi (0 to 71 bar) to 0 to 10,000 psi
(0 to 700 bar) sealed gauge
• 0 to 5 psi (0 to 350 mbar) to 0 to 10,000 psi
(0 to 700 bar) absolute
• 0 to 2.5 psi (175 mbar) to 0 to 500 psi (0 to 40 bar)
differential
Maximum line pressure 500 psi.(35 bar). Differential
ranges unidirectional calibrations only. For high line
pressures, refer to manufacturer.
Accuracy
For external transducers range 1 to 1000 psi
(70 mbar to 70 bar):
• ±0.05% of reading between 20% and 100% FS
• ±0.01% of FS between 0 and 20% FS
• For ranges 1001 to 10,000 psi (71 to 700 bar)
±0.08% of reading between 20% and 100% of FS
• ±0.016% of FS between 0 and 20% FS
Figure includes 90 day stability.
Temperature Effects
• Option B1: provides normal piezo-resistive
transducer temperature error bands
_
see data sheets
• Option B2: provides enhanced temperature
coefficients as detailed for internal sensors.
Standard cable length 6 ft (2 m).
(C) Temperature Probe
A platinum resistance temperature probe (PT 100
Class A):
• °F or °C
• Range: -60°F to 400°F to (-55°C to 200°C)
• Resolution: 0.01°F or 0.01°C
• Accuracy: ±0.2°F (±0.1°C); includes 90 day stability
• Temperature effects: ±0.0005% reading/°F
±0.0044% °F (±0.0009% reading/°C ±0.0022°C)
PT 100 probe accuracy not included in above figures