DPI 603 Series
Portable Pressure Calibrator
Physical Specification
Weight
4kg nominal.
Dimensions
320mm x 195mm x 125mm.
Pressure Connection
G1/8 female or 1/8’’ NPT (female).
OPTIONS
(A) Pressure/Vacuum Capability
A switch facility enables the integral hand-pump to
generate pressure in the down stroke to a maximum
of 20 bar gauge, or vacuum on the reverse stroke to
-850mbar gauge.
(B) Rechargeable Batteries
Ni-Cad rechargeable cells can be supplied complete
with an external charger unit.
(B1) UK 240V a.c. plug version
(B2) USA 110V a.c. plug version
(B3) European 220V a.c. plug version
(B4) 110V a.c. cable version.
ACCESSORIES
Carrying case, electrical test probes, calibration
certificate and user handbook supplied with
instrument.
ORDERING INFORMATION
Please state the following (where applicable):(1) Type number
(2) Pressure range
(3) Pressure connection
(4) Options
(5) Pressure media
Continuing development sometimes necessitates
specification changes without notice.
RELATED PRODUCTS
GE Druck manufacture a comprehensive range of
pressure indicators, controllers, calibrators,
transducers and transmitters. .
The range of portable calibrators also covers
temperature and electrical parameters.
Please refer to manufacturer for further information
and data sheets.
CALIBRATION STANDARDS
Instruments manufactured by GE Druck are
calibrated against precision pressure calibration
equipment which is traceable to International
Standards.
The cost effective DPI 603, which extends the
GE Druck range of rugged portable battery
powered pressure calibrators, has standard
ranges of -1 to +1 bar, 2 bar, 10 bar or 20 bar
gauge with an accuracy of 0.075%.
Designed for use in remote locations where
a.c. power is not available, the DPI 603
enables pressure instrumentation to be
serviced and calibrated.
The instrument features the capability to
generate test pressures with an integral handpump up to 20 barg. A volume adjuster
provides fine control of the desired test value
and a venting valve enables pressure release
without disconnection of the device under test.
An optional pressure/vacuum switching facility
can also be provided, which enables the handpump to source vacuum down to -850 mbarg.
To facilitate pressure transmitter testing and
calibration, an energising voltage of 24V d.c. is
provided and the ability to monitor the loop
current over the range 0 to 55mA d.c.
For loop verification, the DPI 603 can also
output a single 12mA test value, or provide a
passive sink if required.
The clear LCD display allows readout of
pressure, voltage and current parameters with
the ability to show current or voltage
simultaneously with pressure.
STANDARD SPECIFICATIONS
Pressure Measurement
Specification
Operating Pressure Ranges
Standard full scale ranges can be specified of
-1bar g to: +1 bar, +2 bar, +10 bar or +20 bar
gauge.
Over Pressure
4 x causing negligible calibration change to a
maximum of 25 bar.
Pressure Media
Compatible with most common gases.
Accuracy
Combined non-linearity, hysteresis and
repeatability 0.075% F.S. 1 digit.
Temperature Effects
Over 0 to 40C, the averaged Span
temperature coefficient will not exceed
0.0125% of reading/C.
Digital Indicator
Readout and Display
99999 capability, 13.6mm LCD digits with
additional 16 text characters.
Pressure Scale Units
24 scale units plus special.
Display Overload
120% F.S. nominal.
Resolution
0.005% F.S. maximum.
Response
Three readings per second.
Zero Control
Pressure and electrical. Protected against
accidental use.
Calibration
Front panel key pad calibration.
Position effect
Negligible.
Electrical Specification
Electrical Inputs
Voltage Measurement:Range 50V
Accuracy 0.15% of reading
0.02% F.S.
Temperature Coefficient 0.01% reading/C
Current Measurement:Range 55mA
Accuracy 0.075% of reading
0.005% F.S.
Temperature Coefficient 0.01% reading/C
Electrical Outputs
Voltage Output:24V nominal, floating, 25mA maximum
Current Output:Loop check, fixed 12mA nominal, source or
sink.
Power Supply
Battery powered 4 x 1.5V zinc carbon ‘D’ cells,
50 hours nominal continuous operation with no
load on output terminals. Battery low condition
indicated by display symbol.
Optional rechargeable batteries and external
power adaptor/charger unit available.
Environmental Specification
Temperature
Operating -10 to +50C
Calibrated 0 to 40C
Storage -20 to +60C
Sealing
Sealed to IP65
Safety
Electrical and Mechanical Safety EN61010
EMC Emissions EN50081-1
EMC Immunity EN50082-1
Certification CE Marked.
Druck Limited
Fir Tree Lane, Groby, Leicester, LE6 0FH, UK
+44 (0)116 2317100 Fax:+44 (0)116 2317103
E-Mail: sales@druck.com www.druck.com
Agent:
11/03
GE Druck
g