Druck DPI 603 Data sheet

Accuracy 0.075% standard
Integral pressure/vacuum generation
Standard ranges up to 20 barg
5 digit LCD display, multiple scales
Dual pressure and electrical readout
Transmitter supply and measurement
GE Druck
DPI 603
Portable Pressure Calibrator
DPI 603 Series
Portable Pressure Calibrator
Physical Specification
Weight
4kg nominal.
Dimensions
320mm x 195mm x 125mm.
Pressure Connection
G1/8 female or 1/8’’ NPT (female).
OPTIONS
(A) Pressure/Vacuum Capability
A switch facility enables the integral hand-pump to generate pressure in the down stroke to a maximum of 20 bar gauge, or vacuum on the reverse stroke to
-850mbar gauge.
(B) Rechargeable Batteries
Ni-Cad rechargeable cells can be supplied complete with an external charger unit.
(B1) UK 240V a.c. plug version (B2) USA 110V a.c. plug version (B3) European 220V a.c. plug version (B4) 110V a.c. cable version.
ACCESSORIES
Carrying case, electrical test probes, calibration certificate and user handbook supplied with instrument.
ORDERING INFORMATION
Please state the following (where applicable):­(1) Type number (2) Pressure range (3) Pressure connection (4) Options (5) Pressure media
Continuing development sometimes necessitates specification changes without notice.
RELATED PRODUCTS
GE Druck manufacture a comprehensive range of pressure indicators, controllers, calibrators, transducers and transmitters. . The range of portable calibrators also covers temperature and electrical parameters. Please refer to manufacturer for further information and data sheets.
CALIBRATION STANDARDS
Instruments manufactured by GE Druck are calibrated against precision pressure calibration equipment which is traceable to International Standards.
The cost effective DPI 603, which extends the GE Druck range of rugged portable battery powered pressure calibrators, has standard ranges of -1 to +1 bar, 2 bar, 10 bar or 20 bar gauge with an accuracy of 0.075%.
Designed for use in remote locations where a.c. power is not available, the DPI 603 enables pressure instrumentation to be serviced and calibrated.
The instrument features the capability to generate test pressures with an integral hand­pump up to 20 barg. A volume adjuster provides fine control of the desired test value and a venting valve enables pressure release without disconnection of the device under test.
An optional pressure/vacuum switching facility can also be provided, which enables the hand­pump to source vacuum down to -850 mbarg.
To facilitate pressure transmitter testing and calibration, an energising voltage of 24V d.c. is provided and the ability to monitor the loop current over the range 0 to 55mA d.c.
For loop verification, the DPI 603 can also output a single 12mA test value, or provide a passive sink if required.
The clear LCD display allows readout of pressure, voltage and current parameters with the ability to show current or voltage simultaneously with pressure.
STANDARD SPECIFICATIONS
Pressure Measurement Specification
Operating Pressure Ranges
Standard full scale ranges can be specified of
-1bar g to: +1 bar, +2 bar, +10 bar or +20 bar gauge.
Over Pressure
4 x causing negligible calibration change to a maximum of 25 bar.
Pressure Media
Compatible with most common gases.
Accuracy
Combined non-linearity, hysteresis and repeatability 0.075% F.S. 1 digit.
Temperature Effects
Over 0to 40C, the averaged Span temperature coefficient will not exceed
0.0125% of reading/C.
Digital Indicator
Readout and Display
99999 capability, 13.6mm LCD digits with additional 16 text characters.
Pressure Scale Units
24 scale units plus special.
Display Overload
120% F.S. nominal.
Resolution
0.005% F.S. maximum.
Response
Three readings per second.
Zero Control
Pressure and electrical. Protected against accidental use.
Calibration
Front panel key pad calibration.
Position effect
Negligible.
Electrical Specification
Electrical Inputs
Voltage Measurement:­Range 50V Accuracy 0.15% of reading
0.02% F.S. Temperature Coefficient 0.01% reading/C Current Measurement:­Range 55mA Accuracy 0.075% of reading
0.005% F.S. Temperature Coefficient 0.01% reading/C
Electrical Outputs
Voltage Output:­24V nominal, floating, 25mA maximum Current Output:­Loop check, fixed 12mA nominal, source or sink.
Power Supply
Battery powered 4 x 1.5V zinc carbon ‘D’ cells, 50 hours nominal continuous operation with no load on output terminals. Battery low condition indicated by display symbol.
Optional rechargeable batteries and external power adaptor/charger unit available.
Environmental Specification
Temperature
Operating -10to +50C Calibrated 0to 40C Storage -20to +60C
Sealing
Sealed to IP65
Safety
Electrical and Mechanical Safety EN61010 EMC Emissions EN50081-1 EMC Immunity EN50082-1 Certification CE Marked.
Druck Limited Fir Tree Lane, Groby, Leicester, LE6 0FH, UK +44 (0)116 2317100 Fax:+44 (0)116 2317103 E-Mail: sales@druck.com www.druck.com
Agent:
11/03
GE Druck
g
Loading...