Datex-Ohmeda Aestiva 7900 SmartVent Technical Reference Manual

Aestiva/5
7900 Anesthesia Ventilator
Technical Reference Manual
Aestiva 7900 Anesthesia Ventilator
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code, the year of manufacture and a sequential unit number for identification.
AAA A 12345
This alpha character indicates the year of product manufacture and when the serial number was assigned; “D” = 2000, “E” = 2001, “F” = 2002, etc. “I” and “O” are not used.
Aestiva
and
SmartVent
are registered trademarks of Datex-Ohmeda Inc.
Other brand names or product names used in this manual are trademarks or registered trademarks of their respective holders.
05/04 1006-0453-000
Technical Reference Manual
Aestiva 7900 Anesthesia Ventilator Software Revisions 1.X, 3.X, and 4.X
This document is not to be reproduced in any manner, nor are the contents to be disclosed to anyone, without the express authorization of the product service department, Datex-Ohmeda, Ohmeda Drive, PO Box 7550, Madison, Wisconsin, 53707.
©
2004 Datex-Ohmeda Inc.
1006-0453-000 05/04
i
Aestiva 7900 Anesthesia Ventilator

Important

The information contained in this service manual pertains only to those models of products which are marketed by Datex-Ohmeda as of the effective date of this manual or the latest revision thereof. This service manual was prepared for exclusive use by Datex-Ohmeda service personnel in light of their training and experience as well as the availability to them of parts, proper tools and test equipment. Consequently, Datex-Ohmeda provides this service manual to its customers purely as a business convenience and for the customer's general information only without warranty of the results with respect to any application of such information. Furthermore, because of the wide variety of circumstances under which maintenance and repair activities may be performed and the unique nature of each individual's own experience, capacity, and qualifications, the fact that customer has received such information from Datex-Ohmeda does not imply in anyway that Datex-Ohmeda deems said individual to be qualified to perform any such maintenance or repair service. Moreover, it should not be assumed that every acceptable test and safety procedure or method, precaution, tool, equipment or device is referred to within, or that abnormal or unusual circumstances, may not warrant or suggest different or additional procedures or requirements.
This manual is subject to periodic review, update and revision. Customers are cautioned to obtain and consult the latest revision before undertaking any service of the equipment. Comments and suggestions on this manual are invited from our customers. Send your comments and suggestions to the Manager of Technical Communications, Datex-Ohmeda, Ohmeda Drive, PO Box 7550, Madison, Wisconsin 53707.
wwww CAUTION
Servicing of this product in accordance with this service manual should never be undertaken in the absence of proper tools, test equipment and the most recent revision to this service manual which is clearly and thoroughly understood.

Technical Competence

The procedures described in this service manual should be performed by trained and authorized personnel only. Maintenance should only be undertaken by competent individuals who have a general knowledge of and experience with devices of this nature. No repairs should ever be undertaken or attempted by anyone not having such qualifications.
Datex-Ohmeda strongly recommends using only genuine replacement parts, manufactured or sold by Datex-Ohmeda for all repair parts replacements.
Read completely through each step in every procedure before starting the procedure; any exceptions may result in a failure to properly and safely complete the attempted procedure.
ii
05/04 1006-0453-000

Table of Contents

Important . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii
Technical Competence . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .ii

1 Introduction

1.1 What this manual includes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.1.1 Software versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2 Standard service procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.1 Operation manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.2 Service manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.3 Ventilator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3 Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1009-0357-000 11/03 iii
Table of Contents

2 Theory of Operation

2.1 General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2 Aestiva 7900 Ventilator features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.1 Safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3 Aestiva 7900 Ventilator components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.1 Ventilator control electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.3.2 Control panel and display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.3.3 Sensor Interface Board (SIB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
2.3.4 Pneumatic Engine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
2.4 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.4.1 Electrical (original) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.4.2 Electrical (integrated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.4.3 Power supply (original) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2.4.4 Power supply (integrated CPU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2.4.5 Sealed lead acid battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.4.6 CPU assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2.4.7 Sensor Interface Board (SIB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
2.5 Mechanical subsystems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23

3 Post-Service Checkout

2.5.1 Supply gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2.5.2 Gas Inlet Valve (GIV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2.5.3 Flow control valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2.5.4 Drive Gas Check Valve (DGCV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2.5.5 Bellows Pressure Relief Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2.5.6 Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.5.7 Mechanical Overpressure Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.5.8 Bleed resistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.5.9 Free breathing valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2.5.10 Breathing circuit flow sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
3.1 Post-service checkout . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.1.1 Test the Aestiva 7900 Ventilator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
3.1.2 Test the Aestiva Anesthesia Machine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
iv 11/03 1009-0357-000
Table of Contents

4a Tests and Calibration — Software Revision 4.X

4a.1 Self tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4a-3
4a.2 Service Mode Confirmation menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4a-4
4a.3 Main Menu - Service Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4a-5
4a.3.1 Alarm Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4a-6
4a.3.2 Error Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4a-7
4a.3.3 System Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4a-8
4a.3.4 User Select Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-12
4a.3.5 Test CPU and Memory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-14
4a.3.6 Test EEPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-15
4a.3.7 Test GIV . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-16
4a.3.8 Test Flow Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-17
4a.3.9 Test Drive Pressure Limit Switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-18
4a.3.10 Test 5V Fail Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-19
4a.3.11 Test Serial Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-20
4a.3.12 Breathing System Leak Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-21
4a.3.13 Display A/D Channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-22
4a.3.14 Display Discrete I/O Signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-24
4a.3.15 Display Battery Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-25
4a.3.16 Test Panel Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-26
4a.3.17 Flow Valve Test Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-27
4a.3.18 Adjust Drive Gas Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-28
4a.3.19 O2 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-29
4a.3.20 Calibrate Flow Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-30
4a.3.21 Pressure Sensitivity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-31
4a.3.22 Calibrate Flow Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-32
4a.3.23 Bleed Resistor Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-33
4a.3.24 Service Calibrations Required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4a-34

4b Tests and Calibration — Software Revisions 1.X and 3.X

4b.1 Self tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-3
4b.2 Service Mode Confirmation menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-4
4b.2.1 Set the altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-5
4b.2.2 Set the language . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-5
4b.2.3 Set the serial connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-5
4b.3 Main Menu - Service Calibration Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-6
1009-0357-000 11/03 v
Table of Contents
4b.4 Diagnostic Tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-7
4b.4.1 Test CPU . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-7
4b.4.2 Test External RAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-7
4b.4.3 Test Display RAM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-8
4b.4.4 Test Flash ROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-8
4b.4.5 Test EEPROM . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-9
4b.4.6 Test Panel Switches . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-9
4b.4.7 Test Serial Ports . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-10
4b.4.8 Test Flow Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-11
4b.4.9 Test GIV (Gas Inlet Valve) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-12
4b.4.10 Test DPL (Drive Pressure Limit) switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-13
4b.4.11 Test 5V Fail Alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-14
4b.5 Diagnostic Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-15
4b.5.1 Display A/D channels . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-15
4b.5.2 Display I/O signals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-17
4b.5.3 Battery Charge Status . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-18
4b.5.4 System Error Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-19
4b.5.5 Alarm log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-19
4b.6 Flow Valve Test Tool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-20
4b.7 Test Breathing System For Leak . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-21
4b.8 Adjust Drive Gas Regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-22
4b.9 Calibrations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-23
4b.9.1 Calibrate O2 Sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-23
4b.9.2 Calibrate Flow Sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-24
4b.9.3 Pressure Sensitivity Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-25
4b.9.4 Calibrate Flow Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-26
4b.9.5 Calibrate Bleed Resistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-27
4b.10 Schedule Service Calibration — Software 3.X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-28
4b.11 Sensor(s) cal due — Software 1.X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-29
4b.12 User Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-30
4b.12.1 Select Altitude . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-30
4b.12.2 Select Drive Gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-30
4b.12.3 Adjust Brightness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-31
4b.12.4 Select Heliox Mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-31
4b.12.5 VE Alarm Limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-32
4b.12.6 User Select Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4b-32
vi 11/03 1009-0357-000
Table of Contents

5 Troubleshooting

5.1 Troubleshooting instructions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-2
5.2 System Error Log . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-3
5.2.1 Error messages for Software Revision 4.X . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.2.2 Error codes for Software Revisions 1.X and 3.X . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
5.3 Alarm messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5-7
5.4 Mechanical/electrical troubleshooting guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
5.5 Troubleshooting Flowcharts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
5.5.1 Ventilator assessment process . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-26
5.5.2 No display troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-27
5.5.3 Inaccurate volume ventilation troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . 5-28
5.5.4 No ventilation troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-29
5.5.5 High intrinsic PEEP troubleshooting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-30
5.6 Power supply test points . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31

6 Maintenance

7 Repair Procedures

5.6.1 Power supply board (original CPU). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-31
5.6.2 Original CPU. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-32
5.6.3 Power supply (Integrated CPU). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-33
6.1 Supply gas inlet filter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-3
6.2 Free breathing valve maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-4
6.3 MOPV differential relief valve test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-5
6.4 MOPV pressure relief valve test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6-6
7.1 Control panel assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-2
7.2 Keyboard and EL display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-3
7.3 Encoder switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-4
7.4 Alarm speaker . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-5
7.5 Access to electrical enclosure components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-6
7.5.1 CPU Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7-8
7.5.2 Firmware replacement procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-9
7.5.3 Power supply board (for original CPU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-10
7.5.4 Power supply (for Integrated CPU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-11
7.5.5 Toroid (original CPU only) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-12
7.5.6 Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-13
7.6 Vent Engine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-14
7.7 Non-relieving regulator . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-16
1009-0357-000 11/03 vii
Table of Contents

8 Illustrated Parts

7.8 Flow control valve 7-17
7.9 Gas inlet valve 7-18
7.10 Mechanical Overpressure Valve (MOPV assembly) 7-20
7.10.1 To service the original MOPV assembly: 7-20
7.10.2 To service the MOPV assembly with the molded housing: 7-22
7.11 Drive gas check valve assembly 7-23
8.1 Special instructions 8-2
8.2 Service tools 8-2
8.3 Ventilator Harnesses 8-3
8.4 Electrical enclosure parts (original CPU) 8-4
8.5 Electrical enclosure parts (integrated CPU) 8-6
8.6 Display Module 8-8
8.6.1 Rear housing parts 8-8
8.6.2 Front housing parts 8-9
8.7 Aestiva 7900 Vent Engine 8-10
8.7.1 Gas Inlet Valve 8-11
8.7.2 Mechanical Over Pressure Valve (MOPV) 8-12
8.7.3 Inlet filter 8-13
8.7.4 Free Breathing Valve 8-14
8.7.5 Manifold 8-15
8.7.6 Tube Assembly 8-16
8.7.7 Twin Tube Lifter 8-17
8.8 Vent Engine mounting bracket 8-18
8.9 Sensor Interface Board (SIB) 8-19
viii 11/03 1009-0357-000
1 Introduction
In this section
1006-0453-000 05/04 1-1
1.1 What this manual includes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.1.1 Software versions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
1.2 Standard service procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.1 Operation manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.2 Service manuals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.2.3 Ventilator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-3
1.3 Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-4
1 Introduction
1.1 What this manual includes
This manual covers the service information for the Aestiva 7900 SmartVent Anesthesia Ventilator which is an integral component in the Aestiva Anesthesia Machine.
The Aestiva Anesthesia Machine has its own service manual (Stock Number 1006-0452-000).
Special notice
1.1.1 Software versions
Some information in this manual can possibly point the reader to electronic troubleshooting and component/repair replacement level of service. This information, when supplied, is only supplied to add clarity to service or troubleshooting statements. Datex-Ohmeda Service Personnel are mandated by Company Policy to service electronic equipment to a board replacement level only.
Read completely through each step in every procedure before starting the procedure; any exceptions can result in a failure to properly and safely complete the attempted procedure.
Unless otherwise specified, values in this manual are nominal.
Sections in this manual begin on odd numbered or right-hand pages. If there is no text on the preceding, backup even numbered page, it is labeled “Notes” for your use if you wish.
Figures that require more than one page have the title and main text on the left (even numbered) page. Additional figure information is on the facing (odd numbered) page.
The revision level is displayed on the ventilator start-up menu. This manual includes test and calibration procedures for Revision 1.X, 3.X, and 4.X software.
1-2 05/04 1006-0453-000
1.2 Standard service procedures
1 Introduction
1.2.1 Operation manuals
1.2.2 Service manuals
1.2.3 Ventilator tests
wwwwWARNING
You must have, and be familiar with, the Operation manuals for this product. Study the Aestiva Operation manuals if you need further information about the operation of the system.
You must determine where a problem is located before you can determine which service manual to use:
Use this manual for 7900 Ventilator related problems.
Use the Anesthesia Machine service manual (1006-0452-000) for all other components of the Aestiva Anesthesia Machine.
Service calibration functions let Datex-Ohmeda trained users and Datex­Ohmeda service personnel perform ventilator setup functions, tests, calibration and measurements from the front panel display.
Normal operational tests, calibration, and troubleshooting can be performed on your Aestiva 7900 Ventilator without removing components from the system. Repair may require removing the ventilator components from the anesthesia machine.
Section 4, whenever you access any internal component of the Ventilator to verify that all critical parts of the Ventilator are still operational and within specification.
“Service Mode Tests and Calibration”
must be performed
wwwwWARNING
wwwwWARNING
After the Ventilator has been serviced, you must perform
Checkout”
before the system can be returned to clinical use.
Do not perform testing or maintenance on this instrument while it is being used to ventilate a patient, possible injury may result.
to verify the entire Anesthesia System is properly functioning
“Post -Service
1006-0453-000 05/04 1-3
1 Introduction
1.3 Symbols used in the manual or on the equipment
w
Warnings and
w
can occur if you do not follow all instructions in this manual.
Warnings tell about a condition that can cause injury to the operator or the patient.
Cautions tell about a condition that can cause damage to the equipment. Read and follow all warnings and cautions.
Other symbols replace words on the equipment or in Datex-Ohmeda manuals. No one device or manual uses all of the symbols. These symbols include:
Cautions tell you about dangerous conditions that
m
L
l
n M
N
p
On (power)
Off (power)
Standby
Standby or preparatory state for part of the equipment
“ON” only for part of the equipment
“OFF” only for part of the equipment
Direct current
Alternating current
A
j
J
D
w
wW
O
Alarm silence button
Alarm silence touch key (Tec 6).
Type B equipment
Type BF equipment
Type CF equipment
Caution, ISO 7000-0434
Attention, refer to product instructions, IEC 601-1
Dangerous Voltage
x
Protective earth ground
Earth ground
y
Frame or chassis ground
P
Equipotential
Y
Plus, positive polarity
+
Minus, negative polarity
-
1-4 05/04 1006-0453-000
h
G
k
E
REF
SN
Input
Output
Movement in one direction
Movement in two directions
Stock Number
Serial Number
1 Introduction
t
T
g
o
z
Z
U
Variability Read top of float.
Variability in steps Vacuum inlet
This way up Suction bottle outlet
Lamp, lighting, illumination
Cylinder
Lock
Isolation transformer
Unlock
Close drain
Linkage system
Risk of Explosion.
u
134°C
Í
q
t
Open drain (remove liquid)
Autoclavable
Low pressure leak test
Mechanical ventilation
r
Not autoclavable
Bag position/ manual ventilation
R
Inspiratory flow
Q
O2+
O2 sensor connection.
Alarm silence touch key. Volume alarms On/Off touch key.
Expiratory flow
O2 Flush button
End case touch key Menu touch key.
1006-0453-000 05/04 1-5
1 Introduction
< 345 kPa
< 414 kPa
Absorber on.
The primary regulator is set to pressure less than 345 kPa (50 psi).
Absorber off (CO
Bypass active). The primary regulator is set to
2
European Union Representative.
pressure less than 414 kPa (60 psi).
Systems with this mark agree with the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their Operation and Maintenance Manuals. The xxxx is the certification number of the Notified Body used by Datex­Ohmeda’s Quality Systems.
1-6 05/04 1006-0453-000
2 Theory of Operation
In this section This section provides functional descriptions and theory of operation for the major
components of the Aestiva 7900 Ventilator.
2.1 General description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2 Aestiva 7900 Ventilator features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-2
2.2.1 Safety features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3 Aestiva 7900 Ventilator components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-3
2.3.1 Ventilator control electronics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-4
2.3.2 Control panel and display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-6
2.3.3 Sensor Interface Board (SIB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
2.3.4 Pneumatic Engine . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-7
2.4 Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.4.1 Electrical (original) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-8
2.4.2 Electrical (integrated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-9
2.4.3 Power supply (original) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
2.4.4 Power supply (integrated CPU) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
2.4.5 Sealed lead acid battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
2.4.6 CPU assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
2.4.7 Sensor Interface Board (SIB) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-20
2.5 Mechanical subsystems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2.5.1 Supply gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23
2.5.2 Gas Inlet Valve (GIV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2.5.3 Flow control valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-24
2.5.4 Drive Gas Check Valve (DGCV) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2.5.5 Bellows Pressure Relief Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-25
2.5.6 Exhalation valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.5.7 Mechanical Overpressure Valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.5.8 Bleed resistor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-26
2.5.9 Free breathing valve . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
2.5.10 Breathing circuit flow sensors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-27
1006-0453-000 05/04 2-1
2 Theory of Operation
2.1 General description
The Aestiva 7900 Ventilator is a microprocessor based, electronically­controlled, pneumatically-driven ventilator with built in monitoring systems for inspired oxygen, airway pressure and exhaled volume. The ventilator is designed to be used as a medical device assisting in the delivery of anesthesia and is part of the Aestiva Anesthesia Machine.
2.2 Aestiva 7900 Ventilator features
• Sensors in the breathing circuit are used to control and monitor patient
ventilation and measure inspired oxygen concentration. This lets the ventilator compensate for compression losses, fresh gas contribution, valve and regulator drift and small leakages in the breathing absorber, bellows and system.
• Positive End Expiratory Pressure (PEEP) is regulated electronically. During
mechanical ventilation the software maintains the set airway pressure. PEEP is not active when mechanical ventilation is off.
• User settings and microprocessor calculations control breathing patterns.
User interface settings are kept in non-volatile memory.
•Mechanical ventilation is started with the Bag/Vent switch on the breathing
system.
• The Aestiva 7900 Ventilator reads the status of the Bag/Vent switch and the
breathing circuit type (Circle, Bain-Mapleson D). The operator does not have to set the breathing circuit type from a menu.
• The Aestiva 7900 Ventilator has an operator-selectable Heliox mode to
permit gas composition compensation when Heliox gas is used.
• All Aestiva 7900 Ventilators have minimum monitoring and alarms
managed on the ventilator panel (there is no other panel for safety relevant alarm management, etc.).
• Ventilator hardware is regularly monitored by software tests.
• An RS-232 serial digital communications port connects to and
communicates with external devices.
• An exhalation valve modulates flow in the pressure mode.
• Pressure and volume modes are selectable by the operator.
• All pneumatic components are located on one manifold.
• Exhausted drive gas and bellows pressure relief valve gases are mixed and
go through the ventilator exhalation valve.
• Exhalation valve block is autoclavable.
• Excess fresh gas released from the bellows and ventilator drive gas is
transferred from the exhalation valve to the Anesthesia Gas Scavenging System (AGSS).
• Optimized for service with a low number of components.
2-2 05/04 1006-0453-000
2.2.1 Safety features • Dual redundant airway overpressure protection, linked to Pmax
setting.
• Volume over-delivery limits and protection.
• Proprietary hose connections and fixed manifolds.
• 10 VA electrical power limiting to potential oxygen enriched environment.
• 150 psi burst overpressure protection.
2.3 Aestiva 7900 Ventilator components
Components of the ventilator are found in different locations on the Aestiva Anesthesia Machine. The ventilator package consists of:
1. Ventilator control electronics
2. Control panel and display (two mounting options shown)
3. Sensor interface board (SIB)
4. Pneumatic engine
2 Theory of Operation
Note For the Aestiva/5 MRI machine, the display is centrally mounted above the
flowhead.
1
2
3
4
1
2
3
4
Figure 2-1 • Aestiva Anesthesia Machine with hinged and outboard arm options
1006-0453-000 05/04 2-3
2 Theory of Operation
2.3.1 Ventilator control electronics
100-120 VAC 220-240 VAC
50/60 Hz
AC Inlet Module
-Fuses
-Voltage Selector
-Filter
The ventilator control electronics is found in the electrical enclosure of the Aestiva machine behind the AC Inlet module.
In the original Aestiva machines, the power supply and the CPU are on separate boards, as shown in Figure 2-2.
The ventilator control electronics includes the power supply and the CPU. The power supply receives AC power from the machine’s AC Inlet module. All the power necessary to operate the ventilator and light package comes from the power supply. The CPU controls operation of the ventilator.
Tec6 Outlet
Toroid
AB.29.087
Battery
Connector Board
Front Panel
Assembly
Power Supply
Board
CPU Board
Connector Board
Pneumatic
Engine
Task Light
Proprietary RS232C Primary RS232C On/StandBy
Vent On/StandBy Switch O2 Supply Switch and O2 Flush Switch
BCID
Sensor
Interface Board
Patient Interface Harness ACGO Oxygen Sensor
Figure 2-2 • Aestiva 7900 Ventilator functional block diagram (Original CPU)
2-4 05/04 1006-0453-000
100-120 VAC 220-240 VAC
50/60 Hz
2 Theory of Operation
In current Aestiva machines, the regulated power supplies and the CPU are included on a single board (Integrated CPU Board), as shown in Figure 2-3.
A universal power supply is used to convert AC to DC that feeds into the power supply circuits of the integrated CPU board. The universal power supply eliminates the need for the toroidal transformer.
Tec6 Outlet
AC Inlet Module
-Fuses
-Voltage Selector
-Filter
Battery
Power Circuits
Integrated
CPU Board
Digital Circuits
Universal
Power Supply
Task Light
Proprietary RS232C Primary RS232C On/StandBy
Vent On/StandBy Switch O2 Supply Switch and O2 Flush Switch
Connector Board Connector Board
Front Panel
Assembly
AB.29.082
Pneumatic
Engine
Sensor
Interface Board
BCID Patient Interface Harness
ACGO Oxygen Sensor
Figure 2-3 • Aestiva 7900 Ventilator functional block diagram (Integrated CPU)
1006-0453-000 05/04 2-5
2 Theory of Operation
2.3.2 Control panel and display
The control panel on the Aestiva 7900 Ventilator is either outboard (on an arm) or on a hinge to fold flat against the machine (depending on the option ordered) and is made to permit you to grip the panel and push the buttons with your thumbs. The ventilator control panel position is easily adjusted.
For the Aestiva/5 MRI machine, the display is centrally mounted above the flowhead.
The front panel assembly has four submodules:
• The electroluminescent (EL) display
• The keyboard front panel
•A rotary encoder
•A speaker
Figure 2-4 • Ventilator control panel with software version 4.X
The keyboard and rotary encoder are used to control the operations of the ventilator. The front panel uses a three step — selection, change, and approve — setting scheme to prevent unwanted selections. The speaker supplies audio input to the operator.
Logical layout of primary controls is left to right in the sequence: tidal volume or inspiratory pressure level, respiratory frequency, I:E ratio, maximum inspiratory pressure, and PEEP (positive end expiratory pressure).
Primary Ventilator interfaces include:
• Display
• Soft keys
• Rotary encoder (control wheel)
• Knob button (control wheel)
• LED
• Audio
2-6 05/04 1006-0453-000
2 Theory of Operation
2.3.3 Sensor Interface Board (SIB)
Figure 2-5 • Sensor Interface Board (SIB)
A Sensor Interface Board (SIB) serves as the interface between the ventilator CPU board and the breathing circuit sensors.
The SIB processes signals from the:
• Inspiratory and expiratory flow transducers sensor
•O
2
• Auxiliary Common Gas Outlet (ACGO) switch (optional)
•Module ID board
• Canister release switch
• Absorber CO
Bypass switch
2
• Bag/Vent switch
• Control panel switch
• Drive Pressure Limit (DPL) switch
•Manifold pressure transducer
• Patient airway pressure transducer
The SIB for the Aestiva 7900 Ventilator is located in the cable duct under the rear subfloor of the breathing circuit chassis. The SIB uses a 50-conductor cable that supplies the link between the Aestiva 7900 Ventilator CPU and the breathing system signals.
2.3.4 Pneumatic Engine The pneumatic engine enclosure is located in the back chamber of the
breathing system and is shielded to contain EMI emissions. The enclosure includes the Pneumatic Engine and a Pneumatic Engine Connector Board.
The Pneumatic Engine comprises the hardware that drives the ventilator bellows. It includes:
•a 5-micron inlet filter
•a gas inlet valve (GIV)
•a pressure regulator
•a flow control valve
•a drive gas check valve
•a mechanical over-pressure relief valve
•a differential pressure relief valve
•a free-breathing check valve
Figure 2-6 • Pneumatic Engine
1006-0453-000 05/04 2-7
2 Theory of Operation
2.4 Electrical
Power
Cord
2.4.1 Electrical (original)
AC Inlet
Line Filter Fuse Voltage Selector
The original (non-integrated) Aestiva 7900 Ventilator electronic/electrical subassemblies or modules include:
• Power entry toroid
• Power supply board
• CPU board
• Display and front panel assembly
• Sensor interface board
• Sealed lead acid battery
Inside Machine
Power Supply Board
EL voltage = +10--14.5V 10.5W max
2
Toroid
+12V light output @ 15W max +15V @ 170mA max
-15V @ 100mA max +5V @ 1.8A max +5.5 @ 1.4A max
50
12 Volt Battery
2
& Harness
CMS
RS232
CPU Board
68340 Microcontroller Memory and I/O Decoding Flash, SRAM & EEPROM SCR Circuitry D/A Converter Watchdog System Inlet Valve Control & Drive (10 VA limited) Flow Valve Control & Drive (10 VA limited) A/D Converter 15 VDC Supply (10VA limited) DC Supply Monitoring
2
2
O2 supply
O2 Flush
4
Machine/Ventilator On/Standby switch
Shielded Cable
50
Shielded Cable
50
Shielded Cable
7
Pneumatic Engine
Connector board Flow valve Inlet valve
Breathing System
EL Display &
Membrane Switch
Key Components EL Display Membrane Switch Connector Board 50 Pin Connector Series 3 50 Pin Ribbon Cable Header Optical Encoder 50 Pin Ribbon Cable 20 Pin Ribbon Cable Speaker
Inspiratory & Expiratory flow transducers Drive Pressure Limit Switch Manifold Pressure Transducer Patient Airway Pressure Transducer
O2 Sensor
ACGO Switch
SIB
Flow Sensors
Bag/Vent Switch Control Panel Switch Canister Release Switch Absorber Bypass Switch Breathing Circuit ID
Figure 2-7 • Electronic functional block diagram (Original CPU)
2-8 05/04 1006-0453-000
2 Theory of Operation
Pow er
Cord
CMS
RS232
2.4.2 Electrical
(integrated)
AC Inlet
Line Filter Fuse Voltage Selector
EL voltage = +10 –14.5V 10.5W max +12V light output @ 15W max +15V @ 170mA max
-15V @ 100mA max +5V @ 1.8A max +5.5 @ 1.4A max
Integrated CPU Board
68340 Microcontroller Memory and I/O Decoding Flash, SRAM & EEPROM SCR Circuitry D/A Converter Watchdog System Inlet Valve Control & Drive (10 VA limited) Flow Valve Control & Drive (10 VA limited) A/D Converter 15 VDC Supply (10VA limited) DC Supply Monitoring
The integrated Aestiva 7900 Ventilator electronic/electrical subassemblies or modules include:
• Universal power supply (AC to DC converter)
• CPU board (with power supply and digital circuits)
• Display and front panel assembly
• Sensor interface board
• Sealed lead acid battery
Inside Machine
AC to DC Power Supply
2
5
2
12 Volt Battery
& Harness
7
VCHGR = 16.0 V Typical VMAIN = 12.5 V Typical
50
Shielded Cable
Shielded Cable
50
Shielded Cable
Pneumatic Engine
Connector board Flow valve Inlet valve
EL Display &
Membrane Switch
Key Components EL Display Membrane Switch Connector Board 50 Pin Connector Series 3 50 Pin Ribbon Cable Header Optical Encoder 50 Pin Ribbon Cable 20 Pin Ribbon Cable Speaker
Inspiratory & Expiratory flow transducers Drive Pressure Limit Switch Manifold Pressure Transducer Patient Airway Pressure Transducer
SIB
O2 supply
2
O2 Flush
2
4
Machine/Ventilator On/Standby switch
Breathing System
O2 Sensor
ACGO Switch
Flow Sensors
Bag/Vent Switch Control Panel Switch Canister Release Switch Absorber Bypass Switch Breathing Circuit ID
Figure 2-8 • Electronic functional block diagram (Integrated CPU)
1006-0453-000 05/04 2-9
AB.29.083
2 Theory of Operation
2.4.3 Power supply
Power
30-60 VDC
VBOOT
Linear
Regulator
+16.6V
Primary
DC to DC
Converter
VBUS
(original)
Cord
0.25A
The power supply performs seven functions:
• AC to DC converter
• DC to DC step-down converter
• Battery charger
•Multiple output DC regulator
• Battery charge/discharge current monitor
• Battery voltage monitor
• Task light power supply 12V
Entry Module
- EMI Filter
- Fuse
- Voltage Selector
VSW
Linear
Regulator
VB
Toroid
Isolation
Transformer
Secondary
DC to DC
Converter #1
Secondary DC to DC
Converter #2
Linear
Regulator
24-45 VAC
Rectifier
& Filter
Bridge Rectifier
& Filter Caps
ON/Standby
Regulator
Switch
Linear
AC_LED
VSW (+5V VDD Fail Buzzer
Rem_On
+12V (Task Lights)
+10-14.5V VH_EL
-15V Analog
AB.29.097
Battery Charge
Controller
Thermal Breaker
12 V Battery
Battery Voltage &
Current Sense
Transformer
Step-Down
Controller
+5.8V
Rectifier
& Filter
OVP &
Current Limit
Linear
Regulator
Linear
Regulator
+15V Analog
+5.5V Valves Supply
+5V VDD Digital
IBatt,VBatt
Charger Disconnect
Figure 2-9 • Ventilator power supply (Original CPU)
2-10 05/04 1006-0453-000
2 Theory of Operation
2.4.4 Power supply (integrated CPU)
11
11 88
88 00
00
..
.. 99
99 22
22
..
.. BB
BB AA
AA
AB.29.081
CHARGER and TRICKLE
DISABLE
BATTERY CHARGER
BULK CHARGE = 400mA
12V BATTERY
THERMAL AUTO
RESETABLE FUSE
Aestiva machines with an integrated CPU board use a universal power supply for AC to DC conversion. The remainder of the power supply functions are derived in the power supply circuits on the integrated CPU:
• DC to DC step-down converter
• Battery charger
•Multiple output DC regulator
• Battery charge/discharge current monitor
• Battery voltage monitor
• Task light power supply 12V
VOLTAGE and CURRENT
MONITOR
VBatt, IBatt
TASK LIGHTS LDO
TYPICALLY 11.718V
VHEL DISPLAY LDO
TYPICALLY 11.718V
AC TO DC POWER SUPPLY
85 -- 264 VAC
VCHGR = 16.0 TYPICAL
VMAIN = 12.5V TYPICAL
AC INLET
POWER CORD
DIODE OR
VMAIN
VCHGR
VBAT
MOSFET
SWITCHES
SELECT VMAIN OR VBAT
AS OUTPUT TO VBUS
VBUZZER
OUTPUT TO VDD FAIL
BUZZER
VSWITCH
ON / STANDBY SWITCH CIRCUIT
VBUS
SIMPLE SWITCHER
DUAL OUTPUT
SWITCHER
POSITIVE OUTPUT
NEGATIVE OUTPUT
VALVES SUPPLY
5.8V OUTPUT and 10VA
VDD DIGITAL OUTPUT LDO
5V +/- 2.5%
SIB_12V OUTPUT
and 10VA LDO
+15V ANALOG
SUPPLY LDO
-15V ANALOG SUPPLY LDO
Rem_On
AC_LED
Figure 2-10 • Ventilator power supply (Integrated CPU)
1006-0453-000 05/04 2-11
2 Theory of Operation
2.4.5 Sealed lead acid battery
A sealed lead acid battery supplies battery backup for the Aestiva 7900 Ventilator. The Aestiva 7900 Ventilator is not a portable unit. Batteries for the ventilator are used as back up power in case of a power failure. Thus the battery is in a float charge state most of the time. Batteries meet the following:
• Capacity to operate unit for 30 minutes.
• Long float charge life.
• Battery pack is internally fused - in line replaceable.
• Battery terminals and connecting wires are protected against short circuits.
tt
IIIInnnnppppuuuutt
Nominally 13.7 VDC at 25
tt
OOOOuuuuttttppppuuuutt
+10V to +14.8VDC during discharge
o
C during float charge.
2-12 05/04 1006-0453-000
2 Theory of Operation
2.4.6 CPU assembly The CPU assembly contains all of the major circuit functions necessary to
control ventilator operation. In the original Aestiva 7900 machines, these functions are on a separate CPU board. For current machines, they are part of the digital circuits section of the integrated CPU board.
Production Connector
DRIVER
RS232C
Proprietary RS232C
Primary RS232C
Connector
Connector
FLASH MEM ( 512K X 16 )
STATIC RAM ( 128K X 16 )
EEPROM
( 2K X 8 )
VIDEO SRAM
( 32K X 8 )
VA0-VA11
VD8-VD15
SED1351F
VIDEO
CONTROLLER
FERRITE
BEADS
Front Panel Assembly
AUDIO AMP
+5VVH_EL
9
Connector
A0-A18, D0-D15
3
A0-A18, D0-D15
3
A0-A10, D8-D15
3
A0-A11, D8-D15
4
LM4860M
MICROCONTROLLER
SYSTEM DATA BUS & ADDRESS BUS
EPLD-EPM7064
WATCHDOG
LOGIC
ADDRESS
DECODING
YM2413
D0-D15
AUDIO
CHIP
1
2
O2 Supply Connector
MC68340
A0-A20 D0-D15
A20, A0, A1 A14-A17, D8-D10
FREQUENCY
DIVIDER
Connector
14
14
ADC
SEQUENCER
VENTILATION
CONTROL
1
O2 Flush
DS1232
WATCHDOG
& RESET
1
INTERFACE
D8-D15
2
2
SERIAL
EEPROM
12-BIT A/D
D0-D15
CONVERTER
6
D0-D15
1
12-BIT D/A
CONVERTER
4
D0-D15
1
12
7
SIB_+12V
10VA LIMIT
5
SIB Connector
DIGITAL INPUT 0
DIGITAL INPUT 1
11
+15V
+15V
1
-15V P12V_LIGHT
INLET VALVE
DRIVE
FLOW VALVE
DRIVE
+5V LOSS
ALARM
VOLTAGE TEST
POINTS
+5V +5.8V +15V
Pneumatic Engine
Connector
2
4
Buzzer
-15V
24-CHANNEL MUTIPLEXER
6
2
5
+5.8V
VSW
1.225V VH_EL
SIB_12V
1
1
2
V
ANALOG SIGNALS
1
10
DD
MONITOR
VH_EL
Power Supply Connector
Figure 2-11 • Ventilator CPU block diagram
Motorola 68340 processor core The CPU core consists of the following internal functions:
• Bus access control signals for all memory and peripheral devices
• Interrupt handling
• Clocks and timers for the system
• Background software development mode
• Two RS232C serial I/O ports
• Baud rate generator for serial ports
• Hard (power-up) and soft (watchdog error) reset generation
• Data bus buffers
•Memory and I/O decoding
• Program memory
• Safety Relevant Computing (SRC)
•Watchdog system
• Data acquisition
• Flow valve control
• Inlet valve drive
• Front panel interface
• Audio alarm
• Regulator output/manifold pressures
1006-0453-000 05/04 2-13
2 Theory of Operation
System clock
An external 32.768 kHz crystal is used with the internal clock synthesizer to generate a 24.12 MHz system clock.
Periodic interrupt timer
The periodic interrupt timer is the time base for the Real Time Operating System.
Software watchdog timer
The software controlled internal watchdog timer guards against program execution going astray.
External bus interface
The external bus interface handles the transfer of information between the CPU32, external memory and peripherals.
Serial communications module Timing reference
An external 3.6864 MHz clock oscillator is used as the timing reference for the Baud Rate Generator.
Configuration
Both of the serial communication channels are configured as full-duplex asynchronous RS232C ports. The internal Baud Rate Generator establishes the communication baud rate, with a higher limit of 19.2k Baud.
Special operating modes
The serial channels are capable of operating in various looping modes for self testing as well as for remote testing of serial communications. These tests include automatic echo, local loop-back, and remote loop-back.
Memory and I/O decoding Microprocessor Chip Selects
The four programmable chip selects from Motorola 68340 access external memory and peripheral circuits, providing handshaking and timing signals as well as a wait state generation, watchdog logic and ventilation control signals.
I/O lines
Spare I/O lines are used for digital control and/or sense lines.
Timer modules Counter/timer #1
The first counter/timer module is used to monitor the MC68340 system clock frequency. The external 3.6864 MHz baud rate clock is the time base for this measurement.
Counter/timer #2
The second counter/timer module can be used as desired by the application software. The time base for this timer is the internal 24.12 MHz system clock.
2-14 05/04 1006-0453-000
Loading...
+ 174 hidden pages