Datex-Ohmeda, Inc., a General Electric Company, doing business as GE Healthcare.
User Responsibility
This Product will perform in conformity with the description thereof
contained in this User’s Reference manual and accompanying labels
and/or inserts, when assembled, operated, maintained, and repaired
in accordance with the instructions provided. This Product must be
checked periodically. A defective Product should not be used. Parts
that are broken, missing, plainly worn, distorted, or contaminated
should be replaced immediately. Should repair or replacement
become necessary, Datex-Ohmeda recommends that a telephonic or
written request for service advice be made to the nearest
Datex-Ohmeda Customer Service Center. This Product or any of its
parts should not be repaired other than in accordance with written
instructions provided by Datex-Ohmeda and by Datex-Ohmeda
trained personnel. The Product must not be altered without the prior
written approval of Datex-Ohmeda. The user of this Product shall
have the sole responsibility for any malfunction which results from
improper use, faulty maintenance, improper repair, damage, or
alteration by anyone other than Datex-Ohmeda.
CAUTIONU.S. Federal law restricts this device to sale by or on the
order of a licensed medical practitioner. Outside the
U.S.A., check local laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic
which indicates a product group code, the year of manufacture, and a
sequential unit number for identification. The serial number can be in
one of two formats.
AAAX11111AAAXX111111AA
The X represents an alpha
character indicating the year
the product was manufactured;
H = 2004, J = 2005, etc. I and
O are not used.
Aespire, 7900 Ventilator, Advanced Breathing System, Link-25,
ComWheel, Disposable Multi Absorber, EZchange, Reusable
Multi Absorber, PSVPro, SmartVent, Tec 6, and Tec 7 are
registered trademarks of Datex-Ohmeda, Inc.
Other brand names or product names used in this manual are
trademarks or registered trademarks of their respective holders.
The XX represents a number
indicating the year the product
was manufactured; 04 = 2004,
The Aespire anesthesia system is a compact, integrated, and intuitive
anesthesia delivery system. The 7900 Ventilator provides mechanical
ventilation for patients during surgery as well as monitoring and
displaying various patient parameters.
The 7900 Ventilator uses a microprocessor-controlled ventilator with
internal monitors, electronic PEEP, Volume Mode, and other optional
features. A serial interface permits communication to external
monitoring.
This anesthesia system is not suitable for use in an MRI environment.
This system must only be operated by medical personnel authorized
and trained to use this product. It must be operated according to the
instructions in this User’s Reference manual.
NoteConfigurations available for this product depend on local market and
standards requirements. Illustrations in this manual may not
represent all configurations of the product. This manual does not
cover the operation of every accessory. Refer to the accessory
documentation for further information.
1-2M1132382
Page 11
Symbols used in the manual or on the equipment
Symbols replace words on the equipment, on the display, or in
manuals.
Warnings and Cautions tell about the dangerous conditions that can
occur if the instructions in the manual are not followed.
Warnings tell about a condition that can cause injury to the operator
or the patient.
Cautions tell about a condition that can cause damage to the
equipment. Read and follow all warnings and cautions.
On (power)Off (power)
1 Introduction
Standby
Type BF equipmentType B equipment
Dangerous voltageFrame or chassis ground
Direct currentAlternating current
CautionAttention, refer to product instructions
Refer to product instructionsExhaust
Electrical inputElectrical output
Electrical input/outputSample gas inlet to scavenging
O2+
O2 flush button
Pneumatic inletPneumatic outlet
SN
M11323821-3
Serial number
REF
Stock number
Page 12
Aespire View
EquipotentialLamp, lighting, illumination
Protective earth groundEarth ground
VariabilityVariability in steps
Suction bottle outletVacuum inlet
MaxVacuum
+
134°C
Plus, positive polarity
Bag position/manual ventilationMechanical ventilation
Inspiratory flowExpiratory flow
Movement in one directionMovement in two directions
LockUnlock
Isolation transformerLow pressure leak test
AutoclavableNot autoclavable
-
Minus, negative polarity
Bellows volumes are approximate
APL settings are approximateAnesthetic Gas Scavenging System
O2%
O2 cell connection
1-4M1132382
Page 13
Pinch hazardRead to top of float
1 Introduction
EZchange Canister (CO2 bypass)Caution: federal law prohibits dispensing
Systems with this mark agree with the
European Council Directive (93/42/EEC) for
Medical Devices when they are used as
specified in their User’s Reference manuals.
The xxxx is the certification number of the
Notified Body used by GE Healthcare’s
Quality Systems.
ManufacturerDate of manufacture
without prescription
Authorized representative in the European
Community
Indicates that the waste of electrical and
electronic equipment must not be disposed
as unsorted municipal waste and must be
collected separately. Please contact an
authorized representative of Datex-Ohmeda
for information concerning the
decommissioning of equipment.
This way up
GOST R Russian certification
USB port
Ethernet connectionThis product consists of devices that
may contain mercury, which must be
recycled or disposed of in accordance
with local, state, or country laws. (Within
this system, the backlight lamps in the
monitor display contain mercury.
M11323821-5
Page 14
Aespire View
Typeface conventions used
Menu items are written in bold italic typeface; for example, Main
Menu.
Messages that are displayed on the screen are enclosed in single
quotes: for example, ‘Total pressure exceeds Pmax.’
When referring to different sections and other documents, the names
are written in italic typeface and enclosed in double quotes; for
example, “System Controls and Menus.”
6System switchSet the switch to the On (|) position to permit gas flow and to turn on the system.
7Integrated suction
(optional)
8BrakePush down to lock. Lift to release.
9O2 flush buttonPush the O2 flush button to supply high flows of O2 to the breathing system.
10Auxiliary O2 flow
control (optional)
Turn the switch to MAX for full vacuum. Turn the switch to Off (0) for no vacuum. Turn
the switch to On (|) for adjustable vacuum. When in adjustable vacuum, turn the knob
clockwise to increase the vacuum and counterclockwise to decrease the vacuum.
Turn the knob counterclockwise to increase the flow. Turn the knob clockwise to
decrease the flow.
12Flow controlsTurn the knob counterclockwise to increase the flow. Turn the knob clockwise to
decrease the flow. The System switch must be On for gas to flow
M11323822-3
.
Page 20
Aespire View
1
2
4
5
7
8
AC.20.012
6
3
1. Outlet circuit breaker5. Mains inlet
2. Electrical outlet6. System circuit breaker
3. Suction items (optional)7. Cylinder
4. Equipotential stud8. Pipeline connections
Figure 2-2 • Rear view
2-4M1132382
Page 21
2 System Controls and Menus
AC.20.013
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
17
16
Advanced breathing system (ABS) components
1. Expiratory check valve10. Breathing system release
2. Inspiratory check valve11. Manual bag port
3. Auxiliary common gas outlet (ACGO) switch12. Adjustable pressure-limiting (APL) valve
4. ACGO13. Bag/Vent switch
5. Inspiratory flow sensor14. Bellows assembly
6. Expiratory flow sensor15. Sample gas return port
8Absorber canister releasePush to remove the canister. This causes the breathing system to vent to
Sends fresh gas to the ACGO when the switch is activated. The ACGO
provides fresh gas to an external manual breathing circuit.
Flow sensors provide volume measurements for some monitoring
functions and tidal volume delivery.
the room (unless the EZchange canister option is installed). Be sure to hold
the canister by the handle before releasing the canister.
12Adjustable pressure-limiting
(APL) valve
Adjusts breathing system pressure limit during manual ventilation. The
scale shows approximate pressures. Above 30 cmH2O, the knob will click
as it turns.
13Bag/Vent switchSelects between manual ventilation (bag) or mechanical ventilation
(ventilator).
2-6M1132382
Page 23
Optional ABS
AB.82.043
AB.74.120
1
2
3
4
5
components
2 System Controls and Menus
1. Bag support arm
2. EZchange canister system (CO2 bypass)
3. EZchange canister release
4. Condenser drain button
5. Condenser
Figure 2-4 • Breathing system options
Item, Figure 2-4Description
1Bag support armSqueeze the button to raise or lower the arm.
3EZchange canister
release
4Condenser drain buttonPush to drain water out of the condenser.
Push to drop the canister to EZchange position. This seals the breathing circuit,
permitting continued ventilation and rebreathing of exhaled gases. Be sure to
hold the canister by the handle before releasing the canister.
M11323822-7
Page 24
Aespire View
2
1
3
4
5
6
AA43051
AB.80.009
4
3
Vaporizer controls
Refer to the vaporizer User’s Reference manual for more detailed
information on the vaporizer.
1. Tec 6 series
2. Tec 7
3. Lock lever
4. Concentration control and release
5. Indicators (Tec 6 series)
6. Silence alarm touch key (Tec 6 series)
Figure 2-5 • Vaporizer controls
2-8M1132382
Page 25
2 System Controls and Menus
Åben
Item, Figure 2-5Description
3Lock LeverTurn the lever fully clockwise to lock the vaporizer in position.
4Concentration control and
release
Push the release and turn the concentration control to set the agent
concentration. The Tec 6 series concentration control does not turn as long
as the warm-up indicator is on.
5Indicators (Tec 6 series)All indictors come on briefly at the start. The warm-up indicator goes off
after approximately 10 minutes and the operational indicator comes on.
Other indicators come on to advise the user of required action.
6Silence alarm touch key
(Tec 6 series)
Push to silence alarms. Hold for 4 seconds to sound the speaker and light
all indicators (alarm test).
M11323822-9
Page 26
Aespire View
ACGO
Fresh gas flow with anesthetic agent is directed through the Auxiliary
Common Gas Outlet (ACGO) on the front of the system when the
ACGO switch is in the ACGO position. Mechanical ventilation is not
available when operating an auxiliary manual breathing circuit with
fresh gas from the ACGO. The Bag/Vent switch, APL valve, and bag
arm are not part of the external circuit. Volume and pressure
monitoring are not available.
O2 monitoring of fresh gas is available automatically when the ACGO
is selected. A sample of the fresh gas is diverted to the O2 cell in the
breathing system. The sample flow to the O2 cell is dependent on the
pressure in the external circuit. The sample flow reduces the fresh
gas flow rate to the auxiliary breathing circuit equal to the amount
delivered to the O2 cell.
Fresh gas oxygen concentration is displayed on the screen. Set the
alarm limits appropriately. Note that fresh gas oxygen concentration
may not reflect FiO2 during spontaneous breathing or in rebreathing
circuits.
ImportantUse an external O2 monitor if using a rebreathing circuit on ACGO.
Do not use an external ventilator on the ACGO. Do not use the ACGO
to drive external ventilators or for jet ventilation.
See “Scavenging” in the “Setup and Connections” section for more
information on connections.
WARNINGThe maximum pressure at the ACGO can be up to 55 kPa
(8 psi). Use a breathing circuit with pressure relief.
2-10M1132382
Page 27
2 System Controls and Menus
Scavenging the
ACGO sample flow
A sample of the fresh gas is diverted to the O2 cell in the breathing
system to show the O2 numerics on the monitor display. This sample
flow should be scavenged when an auxiliary manual breathing circuit
is used with N2O or volatile anesthetics.
If scavenging is not connected, the sample flow is emptied into the
room.
To connect the scavenging:
1. Attach a circle breathing circuit to the inspiratory and expiratory
ports.
2. Occlude the circle circuit by connecting the Y-piece to the plug
located to the rear of the expiratory port.
3. Check for clinically correct settings.
4. Check the position of the Bag/Vent switch:
•If the Bag/Vent switch is set to mechanical ventilation mode,
the bellows fills slowly with the sample flow. When the
bellows is full, the sample flow goes to the AGSS
(Mechanical ventilation does not start when ACGO switch is
set to ACGO).
•If the Bag/Vent switch is set to bag mode, set the APL valve
to MIN and attach a bag. The bag fills slowly with the sample
flow. When the bag is full, the sample flow goes to the AGSS.
Scavenging from an
auxiliary manual
breathing circuit
ImportantDo not use these connectors as an outlet for exhaust flow.
Scavenging a gas
monitor sample flow
Scavenge the exhaust if an auxiliary manual breathing circuit is used
with N2O or volatile anesthetics.
An auxiliary inlet is available for active and passive AGSS units. It
provides a female connection with 30 mm male - 30 mm male
connector (or a 30 mm male - 19 mm male connector) into the
auxiliary port under the breathing system.
The auxiliary inlet is a convenience inlet to the air brake of active
AGSS units. There is a reservoir to capture exhaust flows higher than
the extract flow.
A separate exhaust hose is needed from the auxiliary manual
breathing circuit to the disposal point for all AGSS units.
Sample gas from a gas monitor can be scavenged using the sample
gas return port or the AGSS. To scavenge from a gas monitor using
the sample gas return port, connect the tubing from the monitor to the
sample gas return port. To scavenge from a gas monitor using the
AGSS, connect tubing from the monitor to the male luer inlet on the
bottom of the AGSS underneath the breathing system.
M11323822-11
Page 28
Aespire View
2
3
4
5
6
789
10
11
12
13
14
AC.20.003
1
Ventilator controls
The ventilator controls include touch keys, menu screens, and a
control knob (ComWheel). The System switch provides power
functions to the ventilator display. The Bag/Vent switch starts and
stops mechanical ventilation.
1. High priority alarm indicator8. PEEP quick key
2. Low or medium priority alarm indicator9. Pmax or Psupport quick key
3. Alarm silence key10. I:E or Tinsp quick key
4. Menu key11. Respiratory rate (RR) quick key
5. End case key12. Tidal volume (TV) or Pinsp quick key
6. ComWheel13. Mains power indicator
7. More settings quick key14. Volume alarms On/Off key
Figure 2-6 • Ventilator controls
2-12M1132382
Page 29
Ventilator screen
AC.20.004
Total
TV
ml
Vent On
RR
/min
I:EPmax
cmH2O
PEEP
cmH2O
More
Settings
VCV - Volume Control
Circle
Fresh Gas Flow
l/min
O2
Air
N2O
Push the knob to confirm the change. Turn the
knob to change the setting.
Time (sec)
Paw
O2% low
Ppeak
Pmean
Ppause
O2
%
TVexp
ml
MV
l/min
RR
/min
cmH2O
TVexp low
1
2
5
6
7
8
9
10
11
4
3
2 System Controls and Menus
1. Alarm silence indicator and countdown clock7. Total flow sensing (optional)
2. Alarm message areas8. Ventilator settings
3. Waveform area9. Mechanical ventilation status
4. Alarm limit settings10. Ventilation mode
5. Measured values area11. User message area
6. Circuit type
Figure 2-7 • Normal view
M11323822-13
Page 30
Aespire View
TV
ml
Vent On
RR
/min
I:EPmax
cmH2O
PEEP
cmH2O
More
Settings
VCV - Volume Control
Circle
Time (sec)
Ppeak
Pmean
Ppause
O2
%
TVexp
ml
MV
l/min
RR
/min
cmH2O
Main Menu
Ventilation Mode
Alarm Setup
Setup/Calibration
Screen and Audio Setup
Cardiac Bypass
Normal Screen
VCV
PCV
SIMV/PSV
PSVPro
SIMV-PC
AC.20.009
Using menus
Figure 2-8 • Menu example
1. Push the Menu key to show the Main Menu.
2. Turn the ComWheel counterclockwise to highlight the next menu
item. Turn the ComWheel clockwise to highlight the previous
menu item.
3. Push the ComWheel to enter the highlighted window or a sub
menu.
4. Turn the ComWheel clockwise or counterclockwise to highlight
the desired selection.
5. Push the ComWheel to confirm the selection.
6. Push the Menu key to exit the menu and return to the normal
monitoring screen.
2-14M1132382
Page 31
3 Operation
WARNINGMaintain sufficient fresh gas flow when using sevoflurane.
In this sectionTurning on the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
1. Plug the power cord into an electrical outlet. Make sure the
system circuit breaker is on.
•The mains indicator comes on when AC power is connected.
•The battery is charging if not already fully charged.
2. Check that the breathing system is properly connected.
CAUTIONDo not turn on the system with the right-hand (inspiratory)
port plugged.
3. Turn the System switch to On.
•The display shows the power-up screen.
•The system does a series of automated self tests.
•A status bar is provided to show progress.
Start mechanical ventilation
WARNINGMake sure that the patient circuit is correctly assembled
and that the ventilator settings are clinically appropriate
before starting ventilation.
1. Set the ACGO switch to the circle system position.
2. Set the Bag/Vent switch.
3. Push the O2 flush button to inflate the bellows if needed.
Start manual ventilation
1. Connect a manual breathing circuit.
2. Make sure that the APL valve is set to a clinically appropriate
3. Set the Bag/Vent switch to Bag.
•If Bag/Vent switch is set to Vent, move Bag/Vent switch to
Bag and then back to Vent to start mechanical ventilation.
•If Bag/Vent switch is set to Bag, move Bag/Vent switch to
Vent to start mechanical ventilation.
setting.
3-2M1132382
Page 33
Ventilator setup
3 Operation
The system has the following mechanical ventilation modes:
•Volume Control Ventilation (VCV).
The ventilator delivers the tidal volume to the patient at the set
respiratory rate (RR).
•Pressure Control Ventilation (PCV) (optional).
The airway pressure is controlled to the value of Pinsp in every
breath.
•Synchronized Intermittent Mandatory Ventilation with Pressure
Support Ventilation (SIMV/PSV) (optional).
Sets a minimum number of mechanical breaths using volume
control to be delivered to the patient, but allows the patient to
have spontaneous breaths which can be pressure supported.
•Pressure Support Ventilation (PSVPro) (optional).
This is a spontaneous mode that provides the set amount of
pressure to the patient during each triggered breath or, if the
patient does not have any spontaneous breaths in a set period of
time, the ventilator transitions to SIMV-PC where it delivers a
minimum number of mechanically controlled breaths.
•Synchronized Intermittent Mandatory Ventilation with Pressure
Control (SIMV-PC) (optional).
Sets a minimum number of mechanical breaths using pressure
control to be delivered to the patient, but allows the patient to
have spontaneous breaths which can be pressure supported.
•Pressure Control Ventilation-Volume Guaranteed (PCV-VG)
(optional).
A tidal volume is set and the ventilator delivers that volume using
a decelerating flow and a constant pressure. The ventilator
adjusts the inspiratory pressure needed to deliver the set tidal
volume breath-by-breath so that the lowest pressure is used.
NoteSee the “Specifications and Theory of Operation” section for more
information on ventilation modes.
Using quick keysThe ventilator settings for each mode can be easily changed using
the ventilator quick keys.
1. Push a ventilator quick key to select the corresponding ventilator
setting.
2. Turn the ComWheel to make a change.
3. Push the ComWheel or quick key to activate (confirm) the
change.
M11323823-3
Page 34
Aespire View
Changing ventilator
modes and settings
1. Push the Menu key.
2. Select Ventilation Mode from the Main Menu.
Main Menu
Ventilation ModeVCV
Alarm SetupPCV
Setup/CalibrationSIMV/PSV
Screen and Audio Setup
Cardiac BypassOffSIMV-PC
Normal Screen
PSVPro
PCV-VG
3. Use the ComWheel to highlight the desired setting (PSVPro
shown), and push the ComWheel to confirm the change.
4. Set the values for the selected ventilation mode.
•The value is highlighted while being set.
5. Push the ComWheel or the quick key to activate the change.
•The system returns to the normal monitoring screen.
•The ventilation mode shows on the screen.
3-4M1132382
Page 35
Alarm Setup
3 Operation
Setting volume
alarms
Use the volume alarms key to turn the volume alarms on and off.
When the alarms are off, an X covers the alarm limits. Use this control
during manual ventilation when constant attention is on the patient.
Use the End case key to minimize alarms between cases. The alarms
will reactivate when two or more breaths are detected within 30
seconds.
WARNINGDo not turn off volume alarms for a spontaneously
breathing patient. The system will not alarm for low
volume.
Alarm limit setupTo set or change alarm limits:
1. Push the Menu key.
2. Select Alarm Setup from the Main Menu.
Alarm Setup
O22199
MV3.07.0
TVexp200600
Leak AudioOn
Go to Main Menu
3. To set an alarm limit, use the ComWheel to scroll to the desired
alarm limit. Push the ComWheel to select the limit. The following
alarm limits can be set or changed:
•O2, MV, and TVexp - Alarm limits for both high and low can
be set. The low limit is the left numeric field, and the high limit
is the right numeric field.
•Leak Audio - The patient circuit leak alarm activates during
mechanical ventilation if less than half of the inspired volume
returns through the expiratory flow sensor. Prevent expected
alarms from known circuit leaks by setting the Leak Audio to
Off.
4. Change the value with the ComWheel.
5. Select Go to Main Menu to return to the Main Menu, or push the
Menu key to return to the normal monitoring screen.
M11323823-5
Page 36
Aespire View
Setup/Calibration
NoteSee the “Specifications and Theory of Operation” section for more
To change or enter Setup and Calibration information:
1. Push the Menu key.
2. Select Setup/Calibration from the Main Menu.
Setup/Calibration
More Vent Settings
O2 Sensor Calibration
About Ventilator...
Go to Main Menu
3. Use the ComWheel to scroll to the desired submenu. Push the
ComWheel to confirm the selection. Select More Vent Settings
to set the following ventilator values:
•Pmax.
•Trig Window.
•Trigger.
•End of Breath.
•Backup Mode Active.
•Tpause.
•Rise Rate.
information on individual settings.
4. After selecting a ventilator setting, set it to the desired value by
turning the ComWheel. Confirm the value is correct by pushing
the ComWheel.
5. Select Go to Setup/Calibration Menu to return to the
Setup/Calibration, or push the Menu key to return to the normal
monitoring screen.
3-6M1132382
Page 37
Screen and audio setup
To change the audio and visual appearance of the screen:
1. Push the Menu key.
2. Select Screen and Audio Setup from the Main Menu.
Screen and Audio Setup
Brightness5
Alarm Volume4
Alarm LimitsShow
Units of MeasureShow
Go to Main Menu
3. Use the ComWheel to select the desired setting and push the
3 Operation
ComWheel to confirm the change. The following settings are
available:
•Brightness - Sets the brightness of the screen.
•Alarm Volume - Sets the volume, or loudness of audible
alarms.
•Alarm Limits - Simplifies the screen by hiding the alarm limits.
To do this, select Hide for this option. If the screen is set to
hide the alarm limits, the limits automatically show when the
corresponding alarm occurs.
•Units of Measure - Simplifies the screen by hiding the units of
measure. To do this, select Hide for this option.
4. Select Go to Main Menu to return to the Main Menu, or push the
Menu key to return to the normal monitoring screen.
M11323823-7
Page 38
Aespire View
Cardiac Bypass
WARNINGCardiac Bypass mode should only be used when the
Cardiac Bypass suspends alarms for patients on cardiac bypass
when the ventilator is not mechanically ventilating. Mechanical
ventilation must be off. When mechanical ventilation is turned on,
Cardiac Bypass is automatically turned off, the alarms are enabled,
and monitoring is available.
patient is receiving extra-corporeal oxygenation by means
of a heart-lung machine. This mode of ventilation is not
intended to provide metabolic levels of ventilation to the
patient.
1. Set the Bag/Vent switch to Bag.
2. Push the Menu key.
3. Select Cardiac Bypass.
4. Set Cardiac Bypass to On.
5. Push the ComWheel to activate the change.
•The screens shows ‘Cardiac bypass’ and ‘Apnea alarm off’ in
the alarm area.
6. Select Go to Main Menu to return to the Main Menu, or push the
Menu key to return to the normal monitoring screen.
3-8M1132382
Page 39
Measure circuit compliance
Exhaled tidal volume (TVexp) measures the gas needed to fill the
patient circuit at the measured pressure. The compliance factor can
be used to calculate the approximate gas which goes into expanding
the compliant patient tubing and not delivered into the patient.
1. Set the ventilator to volume control ventilation (VCV) mode.
2. Set the ventilator parameters.
•TV = 25 ml.
•RR = 20.
•I:E = 1:1.
•Pmax = 100 cmH2O.
•PEEP = Off.
3. Occlude the patient Y.
4. Turn on mechanical ventilation.
5. Monitor the exhaled tidal volume (TVexp) and the measured peak
airway pressure (Ppeak).
6. Calculate the tubing compliance factor.
•TVexp/(Ppeak - 2.5* cmH2O) = Compliance factor in
NoteSee the “Specifications and Theory of Operation” section for more
The pressure waveform shows the measured value of the airway
pressure. The waveform automatically adjusts the time and pressure
scales. The time scale changes with the respiratory rate. The
pressure scale changes with the pressure limit.
waveform information on the ventilation modes.
Respiratory Rate (breaths per minute) Time scale (seconds)
0 to 250 to 16
26 to 750 to 8
76 to 1000 to 4
M11323823-9
Page 40
Aespire View
AB.82.042
AB.74p.043
EZchange canister (optional)
Push the absorber canister release to activate the EZchange canister
mode. The canister will swing down to the EZchange position. The
EZchange canister mode seals the breathing circuit when the canister
holder is down. This permits continued ventilation and rebreathing of
exhaled gases while replacing the absorber canister.
Systems with EZchange canister have the following label on the
canister holder. When the system is in EZchange position, the
message ‘No CO2 absorption’ shows on the ventilator display.
To return to absorber mode, reinsert the canister into the holder, push
the canister back up and snap it into absorber position. When the
canister is in the absorber position, the exhaled gas flows through the
absorber, removing CO2.
NoteCheck the absorber canister to ensure it has side rails. If the canister
does not have side rails, it will not work on the EZchange canister
holder.
3-10M1132382
Page 41
Condenser (optional)
12
3
AB.75p080
3 Operation
Visually check the condenser reservoir daily. Drain the reservoir daily.
1. Place a container under the reservoir.
2. Push the drain button to empty any water in the condenser.
1. Drain button
2. Reservoir
3. Condenser
Figure 3-1 • Condenser
M11323823-11
Page 42
Aespire View
Passive AGSS (optional)
WARNINGAlways verify the proper operation of any gas scavenging
system; ensure the scavenging system is not occluded.
The passive AGSS (Anesthesia Gas Scavenging System) contains
both positive and negative pressure relief valves to protect the
breathing system and the patient. The outlet is a 30-mm connector on
the bottom of the receiver.
There is also a connector that may be used for scavenging the
sample from a gas monitor. The male luer inlet connection is located
near the 30-mm connector.
Passive AGSS is intended primarily for use in operating room
environments which have no active gas extraction system for waste
gas disposal. The disposal system generally consists of large
diameter tubing directly linking the passive AGSS with the building
exterior. The tubing should be as large in diameter and as short as
possible for the particular application.
Passive AGSS may also be used with a non-recirculating facility
ventilation system for waste gas disposal. The tubing connection from
passive AGSS to the non-recirculating facility ventilation system
should be an open connection, essentially at atmospheric pressure,
such as to an exhaust grill.
NoteSee “Setup and Connections” for additional scavenging connection
information.
Connecting passive
AGSS
NoteThe tubing connection from the passive AGSS to the non-circulating
NoteIn the event that excess gas accumulates in the scavenging system
To use the optional passive AGSS installed on the system, connect it
as follows:
1. Connect the proper large diameter tubing to the AGSS 30 mm
outlet connector on the bottom of the AGSS underneath the
breathing system.
facility ventilation system should be an open connection, essentially
an atmospheric pressure, such as to an exhaust grill.
2. Connect the free end of the tubing to the building exterior or
outside ventilation system.
3. The passive scavenging system relies on slight positive pressure
of gases, or on slight negative pressure caused by an exhaust
fan, to move gases through the system. For this reason, all
unused ports must be capped to prevent gas from leaking into the
room and to maintain the expected pressures.
and cannot exit the machine properly, pressure could build. If this
pressure reaches 10cmH20, the brass weight of the positive relief
valve will be lifted up allowing the gas to escape into the room. This
prevents the pressure from backing up into the patient circuit.
3-12M1132382
Page 43
Active AGSS (optional)
WARNINGAlways verify the proper operation of any gas scavenging
system; ensure the scavenging system is not occluded.
There are several versions of the optional active AGSS (Anesthesia
Gas Scavenging System) available depending on the hospital’s type
of waste gas disposal system.
Each version has a two-liter reservoir to capture peak exhaust flows
that briefly exceed the extract flow. Active scavenging systems
contain two relief valves to protect the patient from excess gas
pressure, or from the opposite situation, gas being drawn from the
breathing circuit.
The positive relief valve for the scavenging system is mounted on the
exhalation valve assembly. In the event that excess gas accumulates
in the scavenging system and cannot exit the machine properly,
pressure could build. If this pressure reaches 10 cmH2O, the positive
relief valve will be lifted up, allowing gas to escape into the room. This
prevents the pressure from backing up into the patient circuit.
3 Operation
The negative relief valve (which is located in the same place on all
scavenging receivers) serves to protect the patient in the event that
the extract flow exceeds system input flow. That is, it prevents gas
from being pulled out of the breathing circuit. If more than a small
amount of negative pressure (-0.3 cmH2O) is applied to the
scavenging system, the negative relief valve opens, allowing room air
to be drawn into the scavenging system instead of pulling gas from
the patient circuit. Its effectiveness is limited by the extract flow of the
particular active AGSS device:
•The active low flow system is for use with high vacuum disposal
systems. It requires a vacuum system capable of a continuous
nominal flow of 36 l/min and 300 mmHg (12 inHg) or greater
vacuum pressure. A flow indicator on the system indicates when
the unit is in operation.
•The active high flow system is for use with low vacuum (blower
type) disposal systems. This requires a system capable of
providing 50 to 80 l/min extract flow. A flow indicator on the
system indicates when the unit is in operation.
•Another version is the active adjustable flow. It provides the
capability to adjust the flow with a needle valve (located in a
receiver beneath the breathing system) and a visual indicator bag
which should be properly inflated. It requires a vacuum system
capable of a continuous nominal flow of 36 l/min and 300 mmHg
(12 inHg) or greater vacuum pressure. The extract flow is limited
to 36 l/min and 300 mmHg (12 inHg) vacuum.
•The active low flow system with a 12.7 mm hose barb connector
is for use with a low vacuum disposal system. It requires an
external venturi/ejector system with a flowmeter and a minimum
36 l/min extract flow.
M11323823-13
Page 44
Aespire View
•The active low flow system with a 25 mm barb connector is for
use with low vacuum disposal systems. It requires an external
venturi/ejector system with a 40 to 50 l/min extract flow. A flow
indicator on the system indicates when the unit is in operation.
•The active low flow system with 30-mm ISO taper is for use with
low vacuum disposal systems. It requires an external venturi/
ejector system with a 40 to 50 l/min extract flow. A flow indicator
on the system indicates when the unit is in operation.
See “Setup and Connections” for additional scavenging connection
information.
Connecting active
AGSS with a flow
indicator
NoteThe ball in the upper red zone indicates excessively high extraction
To use the optional active AGSS on a system that has a flow
indicator, connect it as follows:
1. Connect the proper hose to the AGSS outlet connector on the
bottom of the AGSS, beneath the breathing system. Attach the
other end to the hospital disposal system.
2. With the AGSS operating, verify that the flow indicator ball on the
flow indicator rises to the green zone, indicating adequate flow.
flow. The ball in the lower red zone indicates extraction flow rate is
too low or the filter is blocked.
3. Complete the tests in the “Preoperative Tests” section of this
manual.
3-14M1132382
Page 45
3 Operation
AB.75p.110
Connecting active
adjustable AGSS
The active adjustable AGSS option flow rate is limited to 30 l/min with
this option.
To use the optional active AGSS installed on the system which uses
the three-liter bag as a visual indicator, connect it as follows:
1. Connect a disposal hose to the DISS connector on the needle
valve on the bottom of the AGSS (beneath). The hose should be
flexible and reinforced to help prevent kinking and crushing.
2. Attach the other end of the hose to the hospital disposal system.
3. Attach the three-liter bag to the 30-mm auxiliary 1 port on the
bottom of the AGSS.
4. Use the needle valve to adjust the flow rate to match the amount
of gas being scavenged. Use the visual indicator bag when
adjusting the flow rate. The bag should remain partially inflated.
5. Complete the tests in the “Preoperative Tests” section of this
manual.
Total flow sensing (optional)
The total flow sensing (TFS) option electronically measures the fresh
gas flow at the mechanical flow tubes. The total fresh gas flow
measurement displays on the bottom right corner of the screen. The
measurement specifies:
•Total gas flow.
•O2 gas flow.
•Air gas flow.
•N2O gas flow.
M11323823-15
Page 46
Aespire View
3-16M1132382
Page 47
4 Preoperative Checkout
WARNINGRead each component’s User’s Reference manual and
understand the following before using this system:
•All system connections.
•All warnings and cautions.
•How to use each system component.
•How to test each system component.
wBefore using this system:
•Complete the preoperative checkout.
•For in-depth test instructions, see the “Preoperative Tests” section.
•Test all other system components.
wIf a test fails, do not use the equipment.
Have a Datex-Ohmeda trained service representative
repair the equipment.
In this sectionEvery day before your first patient . . . . . . . . . . . . . . . . . . . . 4-2
WARNINGDo not exceed the top shelf weight limit of 34 kg (75 lb).
wMake sure that the breathing circuit is correctly connected
wDo not leave gas cylinder valves open if the pipeline
and not damaged. Replace the breathing circuit if it is
damaged.
supply is in use. Cylinder supplies could be depleted,
leaving an insufficient reserve supply in case of pipeline
failure.
Before using the system, make sure that:
•The equipment is not damaged.
•Components are correctly attached.
•The breathing circuit is correctly connected, not damaged, and
the breathing system contains sufficient absorbent.
•The vaporizers are locked in position and contain sufficient agent.
•Pipeline gas supplies are connected and the pressures are
correct.
•Cylinder valves are closed on models with cylinder supplies.
•Models with cylinder supplies have a cylinder wrench attached to
the system.
•Models with cylinder supplies have a reserve supply of O2
connected to the machine during system checkout.
•The necessary emergency equipment is available and in good
condition.
•Equipment for airway maintenance, manual ventilation, tracheal
intubation, and IV administration is available and in good
condition.
•Applicable anesthetic and emergency drugs are available.
•If an auxiliary O2 flowmeter is present, ensure there is adequate
flow.
•If an optional suction regulator is present, ensure proper
connections and that there is adequate suction.
•The brakes are set and prevent movement.
•The power cord is connected to a wall outlet. The mains indicator
comes on when AC power is connected. If the indicator is not on,
the system does not have mains (electrical) power. Use a
different outlet, close the circuit breaker, or replace or connect the
power cable.
•The O2 flush button is in working condition.
5-2M1132382
Page 53
Power failure alarm test
1. With the System switch set to On, unplug the power cord.
2. Make sure that the power failure alarm sounds.
3. Connect the power cord to the electrical outlet.
4. Verify that the alarm tone stops.
Pipeline test
1. Disconnect the pipeline supplies and close all cylinder valves.
2. If the pipeline and cylinder pressure gauges are not at zero.
3. Connect the pipeline supplies.
4. Verify the pipeline pressure is between 280-600 kPa (41-87 psi).
5 Preoperative Tests
•Connect an O2 supply.
•Set the System switch to On.
•Set the flow controls to mid range.
•Make sure that all of the gauges except for O2 go to zero.
•Disconnect the O2 supply.
•Make sure that the O2 gauge goes to zero. As the pressure
decreases, the alarms for O2 supply failure should occur.
Total flow sensing test (if equipped)
1. Connect the pipeline supplies to the anesthesia system.
2. Set the System switch to On.
3. Set the Bag/Vent switch to Bag.
4. Set each gas flow to 0.80 l/min using flow tubes. Refer to “Flow controls” for more information.
5. Verify that the flow readings on the display for each gas read
between 0.60 l/min and 1.0 l/min.
M11323825-3
Page 54
Aespire View
Cylinder test
WARNINGDo not leave gas cylinder valves open if the pipeline
CAUTIONTo prevent damage to the system, open the cylinder
valves slowly and do not force the flow controls.
supply is in use. Cylinder supplies could be depleted,
leaving an insufficient reserve supply in case of pipeline
failure.
1. Disconnect the pipeline supplies and close all cylinder valves.
2. If the pipeline and cylinder pressure gauges are not at zero:
•Connect an O2 supply.
•Set the System switch to On.
•Set the flow controls to mid range.
•Make sure that all of the gauges except for O2 go to zero.
•Disconnect the O2 supply.
•Make sure that the O2 gauge goes to zero. As the pressure
decreases, the alarms for O2 supply failure should occur.
3. Make sure that the cylinders are full.
•Open each cylinder valve.
•Make sure that each cylinder has sufficient pressure.
•If the cylinder does not have sufficient pressure, close the
cylinder valve and install a full cylinder.
4. Set the System switch to Standby.
5. Turn off the auxiliary O2 flowmeter (if equipped).
6. Turn off the suction (if equipped).
7. Test one cylinder at a time for high-pressure leaks.
•Open the cylinder.
•Note the cylinder pressure.
•Close the cylinder valve.
•Wait for one minute and record the cylinder pressure again.
•If the cylinder pressure for Air or O2 decreases more than
690 kPa (100 psi), there is a leak.
•If the cylinder pressure for N2O decreases more than
690 kPa (100 psi), there is a leak.
5-4M1132382
Page 55
Flow control test
WARNINGThe Link-25 system cannot replace an O2 monitor.
5 Preoperative Tests
Sufficient O2 in the fresh gas may not prevent hypoxic
mixtures in the breathing circuit.
wNitrous oxide (N2O), if available, flows through the system
during this test. Use a safe and approved procedure to
collect and remove the N2O.
wIncorrect gas mixtures can cause patient injury. If the
Link-25 system does not supply O2 and N2O in the
correct proportions, do not use the system.
wThis procedure tests for significant malfunction of the
Link-25 system. It does not confirm the proper calibration
of the Link-25 system. Perform periodic calibrations using
an accurate and properly calibrated O2 monitor as
recommended in the “User Maintenance” section.
1. Connect the pipeline supplies or slowly open the cylinder valves.
2. Turn all flow controls fully clockwise for minimal flow.
3. Set the ACGO switch to ABS.
4. Set the System switch to On.
NoteDo not use the system if the Low battery or any ventilator failures
occur.
5. Make sure that the O2 flow tube shows approximately 0.025 to
0.075 l/min.
WARNINGKeep the Link-25 system engaged. Adjust only the test
control for the following steps.
•Test N2O first and then O2.
•The O2 cell must be correctly calibrated.
M11323825-5
Page 56
Aespire View
6. Test the Link-25 system by increasing the N2O flow.
•Slowly turn the N2O flow control counterclockwise.
•Increase the N2O flow as specified in the following table and
make sure the O2 flow is as specified.
N2O flow l/minO2 flow greater than l/min
0.80.2
2 0.5
4 1.0
102.5
7. Test the Link-25 system with O2 flow decreasing.
•Set the N2O flow to 9 l/min.
•Set the O2 flow to 3 l/min or higher.
•Slowly turn the O2 flow control clockwise. Set the N2O flow to
the rates shown in the following table.
N2O flow l/minO2 flow greater than l/min
82
4 1
0.8 0.2
8. Adjust the flow of all gases through the full range and make sure
that the flow tube floats move smoothly.
9. Disconnect the O2 pipeline supply or close the O2 cylinder valve.
10. Make sure that:
•The low O2 supply alarm occurs.
•The N2O and the O2 flows stop. The O2 flow should stop
last.
•The air flow continues (if equipped).
•The gas supply alarms occur on the ventilator if the ventilator
uses O2 as the drive gas.
11. Turn all the flow controls fully clockwise for minimum flow.
12. Reconnect the O2 pipeline supply.
5-6M1132382
Page 57
Vaporizer installation
WARNINGUse only Datex-Ohmeda, Inc. Selectatec series
wDo not use a vaporizer if it lifts off the manifold when the
wDo not use the system if more than one vaporizer can be
wTec 6 series vaporizers need the power cable to go
5 Preoperative Tests
vaporizers Tec 4 or greater.
lock lever is in the locked position.
turned on at the same time.
through the channel on the bottom of the vaporizer for
proper alignment. Do not put the power cable on the top
of the manifold or between the vaporizers.
1. Place the vaporizer straight on the manifold.
2. Make sure that the top of each vaporizer is horizontal.
•If a vaporizer is not horizontal, remove it and reinstall it.
3. Set each vaporizer lock lever to the locked position.
4. Try to lift each vaporizer straight up off the manifold.
•Do not pull forward on the vaporizer.
•Do not rotate the vaporizer on the manifold.
5. If the vaporizer lifts off of the manifold, install it again and repeat
this “Vaporizer installation” procedure. If the vaporizer lifts off a
second time, do not use the system.
6. For a Tec 6 series vaporizer:
•Connect the vaporizer to an electrical outlet.
•Hold the Silence alarm touch key (minimum of 4 seconds).
•Make sure all indicators turn on and that the alarm tone
occurs.
•Release the Silence alarm touch key.
7. Try to turn on more than one vaporizer at the same time.
•Test each possible vaporizer combination.
•If more than one vaporizer turns on at the same time, remove
the vaporizers, install them again, and repeat the “Vaporizer installation” procedure. If the test continues to fail, contact a
Datex-Ohmeda trained service representative.
M11323825-7
Page 58
Aespire View
Vaporizer back pressure test
WARNINGAnesthetic agent comes out of the common gas outlet
during this test. Use a safe, approved procedure to
remove and collect the agent.
CAUTIONTo prevent damage to the vaporizer, turn the flow controls
fully clockwise (minimum flow or off) before turning on the
system.
1. Set the System switch to On. Alarms may occur.
2. Set the O2 flow to 6 l/min.
3. Make sure that the O2 flow stays constant and the float moves
freely.
4. Adjust the vaporizer concentration from 0 to 1%. The O2 flow
must not decrease more than 1 l/min through the full range.
•If the O2 flow decreases more than 1 l/min, install a different
vaporizer and repeat the steps 1 through 4.
•If the O2 flow decreases less than 1 l/min when testing a
different vaporizer, the malfunction is in the vaporizer that
failed the test.
•If the O2 flow decreases more than 1 l/min with the different
vaporizer, the malfunction is most likely in the system. Do not
use the system.
5. Repeat steps 1 through 4 for each vaporizer.
5-8M1132382
Page 59
Low-pressure leak test
WARNINGDo not use a system that has a low-pressure leak.
Anesthetic agent will go into the atmosphere instead of
into the breathing circuit.
Perform either the “Negative low-pressure leak test” or the “Positive low-pressure leak test” depending on local requirements. It is not
necessary to perform both tests.
5 Preoperative Tests
Negative low-
pressure leak test
1. Make sure the System switch is set to Standby.
2. Turn off all vaporizers.
3. Turn the ACGO switch to the ACGO position.
4. Compress and release the bulb until all air is removed from the
bulb.
•Occlude the inlet of the test device. Make sure it is a tight
seal.
•If the bulb of the test device inflates in less than 60 seconds,
use a different test device.
5. Test the system for low-pressure leaks.
•Turn the flow controls one and a half turns counterclockwise.
•Connect the test device to the auxiliary gas outlet.
•Compress and release the bulb until all air is removed from
the bulb.
•The floats will move. If the bulb inflates in 30 seconds or less,
there is a leak in the low-pressure circuit. See “Pneumatic problems” in the “Alarms and Troubleshooting” section.
•Disconnect the test device.
6. Test each vaporizer for low-pressure leaks.
•Turn on one vaporizer.
•Set the vaporizer to 1%.
•Perform step 5.
•Repeat this test with each vaporizer.
•If a low-pressure leak occurs while testing any of the
vaporizers, see “Pneumatic problems” in the “Alarms and Troubleshooting” section.
•Turn off all vaporizers.
7. Turn all flow controls fully clockwise for minimum flow. Do not
overtighten.
M11323825-9
Page 60
Aespire View
WARNINGAgent mixtures from the low-pressure leak test stay in the
system. Clear the system by flowing O2 at 1 l/min for one
minute.
8. Clear the system of agent.
•Set the System switch to On.
•Set the O2 flow to 1 l/min.
•Flow O2 for one minute.
•Turn the O2 flow control fully clockwise for minimum flow.
Positive low-
pressure leak test
NotePush the positive-pressure leak test adapter into the ACGO port
CAUTIONDo a positive low-pressure leak test at the ACGO port
NoteKeep the test device flow tube vertical for accurate results.
CAUTIONIf the needle valve is not fully open, this test can damage
1. Connect the test device to the ACGO port with the positivepressure leak test adapter.
throughout the test to maintain a tight seal.
only.
2. Turn the ACGO switch to the ACGO position.
3. Turn all flow controls fully clockwise for minimum flow.
4. Turn all vaporizers off.
5. Fully open the needle valve on the test device.
the pressure gauge on the test device.
6. Open the O2 flow control and set for a total flow on the test
device of 0.4 l/min.
7. Make sure that the pressure gauge on the test device is at zero,
and make sure that all other flow controls are fully closed.
8. Close the needle valve on the test device until the gauge reads
20 kPa (3 psi) for (BSI) or 3 kPa (.4 psi) for ISO.
9. If the flow through the test device is less than 0.35 l/min ISO or
0.3 l/min (BSI), there is a low pressure leak in the system. See
“Pneumatic problems” in the “Alarms and Troubleshooting”
section.
10. Test each vaporizer.
•Turn on the vaporizer being tested, and set it to 1%.
•Perform steps 5 through 9 of this test for each vaporizer.
11. Make sure all vaporizers are turned off.
5-10M1132382
Page 61
Alarm tests
5 Preoperative Tests
WARNINGAgent mixtures from the low-pressure leak test stay in the
system. Clear the system by flowing O2 at 1 l/min for one
minute.
12. Clear the system of agent.
•Set the O2 flow to 1 l/min.
•Flow O2 for one minute.
•Turn the O2 flow control fully clockwise for minimum flow.
1. Connect a test lung to the patient connection.
2. Set the Bag/Vent switch to Vent.
3. Set the System switch to On.
4. Push the Menu key.
5. Select Ventilation Mode - VCV.
6. Set the ventilator parameters.
•TV 400 ml.
•Rate 12.
•I:E 1:2.
•Pmax 40 cmH2O.
•PEEP Off.
7. Set the O2 flow to the minimum flow.
8. Turn off all other gases.
9. Push the O2 flush button to fill the bellows.
10. Set the Bag/Vent switch to Bag and then to Vent. Make sure that:
•Mechanical ventilation starts.
•A subatmospheric pressure alarm does not occur.
•The ventilator shows the correct data based on settings.
•The bellows inflate and deflate during mechanical ventilation.
11. Set the O2 flow control to 5 l/min. Make sure that:
•The pressure at the end of the breath is approximately
2 cmH2O. This can be seen on the pressure waveform
displayed below.
•The ventilator shows the correct data based on settings.
•The bellows inflate and deflate during mechanical ventilation.
M11323825-11
Page 62
Aespire View
12. Test the O2 monitor and alarms (alarms other than Low O2 and
High O2 may occur).
•Remove the O2 cell, and make sure that the cell measures
approximately 21% O2 in room air.
•Push the Menu key.
•Select Alarm Setup from the Main Menu.
•Set the Low O2 alarm to 50%, and make sure that a Low O2
alarm occurs.
•Set the Low O2 alarm to 21%, and make sure that the Low
O2 alarm stops. This will create a latched alarm,
acknowledge this by pushing the Alarm silence key.
•Put the O2 cell back into the circuit.
•Set the High O2 alarm to 50%.
•Push the O2 flush button to fill the breathing system, and
make sure that the High O2 alarm occurs.
•Set the High O2 alarm to Off, and make sure that the alarm
stops.
•Flow 100% O2 for 2 minutes, and make sure that the O2 cell
measures 100% O2.
13. Test the low minute volume alarm.
•Push the Menu key.
•Select Alarm Setup from the Main Menu.
•Set the alarm limit for low minute volume to 6 l/min.
•Make sure that the low minute volume alarm occurs.
•Set the low minute volume alarm to Off.
14. Test the low airway pressure alarm.
•Remove the test lung from the patient connection.
•Make sure that the low airway pressure alarm occurs (other
alarms may occur).
15. Test the sustained airway pressure alarm.
•Set the APL valve to 70 cmH2O.
•Set the Bag/Vent switch to Bag.
•Occlude the patient connection and push the O2 flush button.
•Make sure that the Ppeak high. Blockage? (sustained airway
pressure) alarm occurs after approximately 15 seconds at the
sustained pressure limit.
5-12M1132382
Page 63
Breathing system tests
1. Make sure that the auxiliary equipment is functioning correctly.
2. Verify that AGSS is functioning correctly.
3. Make sure that the check valves on the breathing circuit module
WARNINGObjects in the breathing system can stop gas flow to the
patient, causing injury or death. Use a test plug that is the
appropriate size so that it will not fall into the breathing
system.
5 Preoperative Tests
•Some breathing systems with active AGSS have a flow
indicator on the side. Make sure that the flow indicator shows
a flow in the green range.
work correctly.
•The expiratory check valve rises during expiration and falls at
the start of inspiration.
•The inspiratory check valve rises during inspiration and falls
at the start of expiration.
After performing the breathing system tests, make sure
that there are no test plugs or other objects caught in the
breathing system.
Bellows test1. Set the System switch to Standby.
2. Set the Bag/Vent switch to Vent.
3. Set all flow controls to minimum flow.
4. Occlude the patient connection.
5. Push the O2 flush button to fill the bellows. Release the O2 flush
button.
6. Make sure that the pressure does not increase to more than
15 cmH2O on the pressure gauge.
7. If the bellows falls lower than the top of the indicator, there is a
leak. See “Breathing system problems” in the “Alarms and Troubleshooting” section.
M11323825-13
Page 64
Aespire View
Breathing circuit test1. Set the System switch to On.
2. Set the Bag/Vent switch to Bag.
3. Occlude the bag port.
4. Set the APL valve to 70 cmH2O.
5. Set the O2 flow to 250 ml/min.
6. Occlude the patient connection.
7. Push the O2 flush button and pressurize the bag to approximately
30 cmH2O.
8. Release the O2 flush button. The pressure must not decrease.
Any pressure decrease shown on the pressure gauge indicates a
leak. Repair any leaks in the breathing circuit.
APL valve test1. Set the System switch to On.
2. Occlude the patient connection.
3. Occlude the bag port.
4. Set the APL valve to 70 cmH2O.
5. Set the O2 flow to 3 l/min. Make sure that the value on the
inspiratory pressure gauge does not exceed 85 cmH2O. Some
pressure fluctuation is normal.
6. Set the APL valve to MIN.
7. Make sure that the value on the inspiratory pressure gauge is
less than approximately 5 cmH2O.
8. Push the O2 flush button. Make sure that the value on the
inspiratory pressure gauge stays near zero.
9. Set the O2 flow to minimum and make sure that the value on the
inspiratory pressure gauge does not decrease below 0 cmH2O.
5-14M1132382
Page 65
Monitor and ventilator tests
1. Connect a test lung to the patient connection.
2. Set the Bag/Vent switch to Bag.
3. Set the System switch to On.
4. Push the Menu key.
5. Select Ventilation Mode - VCV.
6. Set the ventilator parameters.
•TV 400 ml.
•Rate 12.
•I:E 1:2.
•Pmax 40 cmH2O.
•PEEP Off.
7. Set the Bag/Vent switch to Vent.
8. Push the O2 flush button to fill the bellows.
9. Make sure that:
•Mechanical ventilation starts.
•A subatmospheric pressure alarm does not occur.
•The ventilator shows the correct data based on settings.
•The bellows inflate and deflate during mechanical ventilation.
10. Set the O2 flow control to 5 l/min.
5 Preoperative Tests
11. Make sure that:
•The end expiratory pressure is approximately 2 cmH2O.
Positive end expiratory pressure that occurs when PEEP is
Off may indicate that the scavenging system is not removing
enough gas.
•The ventilator shows the correct data based on settings.
•The bellows inflate and deflate during mechanical ventilation.
M11323825-15
Page 66
Aespire View
12. Set the ventilator controls and alarm limits to clinically appropriate
levels.
13. Prepare the system.
•Turn all vaporizers off.
•Set the APL valve to MIN.
•Set the Bag/Vent switch to Bag.
•Set all flow controls to minimum.
•Set sufficient patient suction.
•Make sure that the breathing system is correctly connected
and not damaged.
WARNINGFlush the system with 5 l/min of O2 for at least one minute
to remove any gas mixtures or by-products from the
system.
5-16M1132382
Page 67
6 Alarms and Troubleshooting
WARNINGSet alarm levels appropriately before starting ventilation
(manual or mechanical).
wIf an alarm occurs, safeguard the patient first before
performing troubleshooting or doing repair procedures.
wDo not use malfunctioning equipment. Contact a
Datex-Ohmeda trained representative for service.
wNo repair should ever be attempted by anyone not having
experience in the repair of devices of this nature. See the
“Repair policy” in the “User Maintenance” section.
When an alarm occurs during a case, an alarm tone sounds and the
alarm message is displayed in the alarm message field. The alarm
message area has room for four alarms to be shown at one time. If
more than four alarms occur, the alarms cycle every two seconds.
Alarms have three general causes:
•Malfunctions - Result in reduced system function or prevent
mechanical ventilation.
•Patient monitoring - Are caused by high and low limit settings that
are adjusted by the user.
•Informational - Are caused by control settings or system
conditions that may change system operation.
AudioAlarm priority is dependent on the level of risk to the patient.
A high-priority alarm tone sounds in two bursts of five tones, a 10
second pause, and then repeats. Some high-priority alarms can be
silenced for 120 seconds.
Medium-priority alarm tones sound in three tones with a 25 second
pause, then repeats. Medium priority alarms can be silenced for 120
seconds.
Informational alarms have a single alarm tone and the tone does not
repeat.
Silencing an alarm stops the audible tone for 120 seconds. Pushing
the Alarm silence key when no medium or high-priority alarms are
active suspends audible alarm tones for 90 seconds.
DisplayMessages for alarms are displayed as follows:
•High-priority alarms show in white text on a red background.
•Medium priority alarms show in yellow text on a dark gray
background.
•Informational messages show in white text on a dark gray
background.
6-2M1132382
Page 69
6 Alarms and Troubleshooting
Latching alarmsSome patient parameter alarms continue to be displayed (latch) when
the alarm condition is corrected. When an alarm is latched, the alarm
messages show in white text on a black background.
Latching alarmFlashing parameter
O2% lowO2
No O2 pressureNone
Ppeak highPpeak
PEEP high. Blockage?Ppeak
Negative airway pressureNone
Apnea > 120 sTVexp
Inspiration stoppedNone
The alarm remains in this condition until it is acknowledged by
pushing the Alarm silence key or until the alarm reoccurs. When the
alarm is acknowledged, it is removed from the screen. If an alarm has
latched and the alarm reoccurs before it is acknowledged, the alarm
will revert to an active state.
M11323826-3
Page 70
Aespire View
List of alarms
If corrective action does not resolve the alarm, contact a
Datex-Ohmeda trained service representative.
MessagePriorityCauseAction
+15V Analog outof-range
+6V Analog out-ofrange
-6V Analog out-ofrange
5V Ref out-of-range Minimum shutdown
1.225 V Ref out-ofrange
12 hour test
recommended
A/D converter
failure
Absorber panel
open
ACGO onMedium
Apnea > 120 sHighNo mechanical breaths or
ApneaMediumNo mechanical breaths or
Apnea alarm offInformationalThe Cardiac Bypass option is set to
Apnea alarm
standby
Backup mode
active
Battery charger fail InformationalThe current in the battery charging
Battery chargingInformationalBattery is not fully charged.Leave the system plugged in to charge the
Battery circuit
failure
Battery current high InformationalBattery current greater than 6 amps
Minimum shutdown
(High)
Minimum shutdown
(High)
Minimum shutdown
(High)
(High)
Minimum shutdown
(High)
InformationalSystem is in use for more than 12
Minimum shutdown
(High)
Medium The breathing system is not fully
(Informational after
acknowledge)
InformationalNormal condition after End Case,
InformationalNo spontaneous breaths in set period
InformationalBattery measures less than 7 V or
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
hours without a power-up self test.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
latched.
The outlet selection switch is set to
the auxiliary common gas outlet.
spontaneous breaths greater than 5
ml in last 120 seconds.
spontaneous breaths greater than 5
ml in last 30 seconds.
On in the Main Menu.
power-up, or ACGO change from On
to Off.
of time and 30 seconds have elapsed
since starting PSVPro mode.
circuit is too high.
higher than 16.5 V.
for 10 seconds.
Contact a Datex-Ohmeda trained service
representative.
Contact a Datex-Ohmeda trained service
representative.
Contact a Datex-Ohmeda trained service
representative.
Contact a Datex-Ohmeda trained service
representative.
Contact a Datex-Ohmeda trained service
representative.
At end of the case, move the System switch
from On to Standby to On.
Contact a Datex-Ohmeda trained service
representative.
Fully latch the breathing system.
Connect the patient circuit to the auxiliary outlet.
For mechanical ventilation or ventilation with
monitoring, select the circle system setting.
Check the patient. Bag as needed. Check for
disconnects. If the patient is on a heart lung
machine, select Cardiac Bypass from the Main
menu.
Check the patient. Bag as needed. Check for
disconnects. If the patient is on a heart lung
machine, select Cardiac Bypass on the Main
Menu.
Set Cardiac Bypass to Off in the Main Menu.
Monitoring resumes after first breath
(mechanical) or two breaths within 30 seconds
(non-mechanical).
Select a new ventilation mode or switch to
manual ventilation.
System is operational, but may fail on battery if
mains power is lost. Contact a Datex-Ohmeda
trained service representative.
battery.
System is operational, but may fail on battery if
mains power is lost. Contact a Datex-Ohmeda
trained service representative.
System is operational, but may fail on battery if
mains power is lost. Contact a Datex-Ohmeda
trained service representative.
6-4M1132382
Page 71
6 Alarms and Troubleshooting
MessagePriorityCauseAction
Battery failure high InformationalBattery voltage greater than 16 V for
Battery failure lowInformationalThe battery voltage is too low (less
Calibrate flow
sensors
Calibrate O2
sensor
Cardiac bypassInformationalThe Cardiac Bypass option is set to
Check flow sensors Medium
Circuit leakMediumExhaled volume less than 50% of
Circuit leak audio
off
Connect O2 sensor MediumO2 cell is not installed in the breathing
CPU failureMinimum shutdown
Display failureMediumBacklight input supply is less than
InformationalThe last flow sensor calibration failed. Zero the flow sensors. Look for water in the flow
InformationalCalibration failure or O2% > 110%.Calibrate the O2 sensor. Replace the sensor if
(Informational after
acknowledge)
InformationalLeak Audio is set to Off on the Alarm
(High)
Medium
(Informational after
acknowledge)
Minimum shutdown
(High)
MediumThe low MV limit is set to Off.Turn on the MV low setting.
Minimum shutdown
(High)
Minimum shutdown
(High)
10 seconds.
than 7 V) to supply the system if
power fails.
On. Apnea alarms are off.
System has detected an improper flow
pattern in the breathing circuit.
inspired volume for at least 30
seconds (mechanical ventilation).
Setup menu.
system. The O2 cell is not measuring
gas in the breathing circuit.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
9.35V or greater than 13.65V. This
alarm message may not always be
viewable.
Expired volume is greater than
inspired volume for six breaths with a
circle module.
Ventilator pressure sensor
malfunction.
pressure).
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
System is operational, but may fail on battery if
mains power is lost. Contact a Datex-Ohmeda
trained service representative.
System is operational, but will fail on battery if
mains power is lost. Leave the system plugged
in to charge the battery. If the battery does not
charge in 24 hours, contact a Datex-Ohmeda
trained service representative.
sensor tubes and dry if necessary. Replace
sensor if necessary. Contact a Datex-Ohmeda
trained service representative if calibrating or
replacing the sensor does not correct the
problem.
calibration is unsuccessful. Contact a DatexOhmeda trained service representative if
calibrating or replacing the sensor does not
correct problem.
Set Cardiac Bypass to Off in the Main Menu.
Check if the flow sensors are correctly installed.
Check for water buildup in the flow sensor
tubes. Inspect one way valves (breathing circuit
module.) Check the condition of the flow sensor
and its tubing.
Check breathing circuit and flow sensor
connections. Patient circuit leak audio can be
turned off in the Alarm Setup menu.
Turn Leak Audio to On on the Alarm Setup
menu.
Install or replace the O2 cell.
Contact a Datex-Ohmeda trained service
representative.
Continue to use the system normally. Contact a
Datex-Ohmeda trained service representative.
Check patient condition. Check that the flow
sensors are installed correctly. Check that there
is no water buildup in the flow sensor tubes.
Verify proper check valve operation. Inspect one
way valves (breathing circuit module). Replace
flow sensors.
Ventilate manually. Monitoring is not available.
Contact a Datex-Ohmeda trained service
representative.
Adjust the controls. Check the system for
blockages. Contact a Datex-Ohmeda trained
service representative if problem continues.
Contact a Datex-Ohmeda trained service
representative.
Contact a Datex-Ohmeda trained service
representative.
M11323826-5
Page 72
Aespire View
MessagePriorityCauseAction
Invalid circuit
module
Limit task light useInformationalThe system is operating on battery
Loss of backup
audio
Low battery voltage MediumVoltage is less than 11.65V while
Memory
(EEPROM) failure
Memory (flash)
failure
Memory (redundant
storage) fail
Monitoring onlyMediumA severe malfunction prevents
MVexp highMediumExhaled minute volume is greater
MVexp lowMediumExhaled minute volume less than the
Negative airway
pressure
No CO2 absorption Medium
No exp flow sensor Medium
No insp flow sensor Medium
No O2 pressureHigh (cannot be
O2 flush stuck on? InformationalO2 flush flow is detected for greater
O2% highMediumO2% is greater than the alarm high
InformationalAbsorber switches do not detect a
Medium
(Informational after
acknowledge)
InformationalThe system cannot access some
Minimum shutdown
(High)
Minimum shutdown
(High)
HighSubatmospheric pressure (less than
(Informational after
acknowledged)
(Informational after
acknowledge)
(Informational after
acknowledge)
silenced)
valid breathing circuit module.
power.
Alarm audio malfunction.Continue to use normally. Contact a Datex-
using battery power.
stored values.
Ventilator malfunction.Ventilate manually. Monitoring is not reliable.
Ventilator malfunction.Ventilate manually. Monitoring is still available.
mechanical ventilation. Other alarms
may also occur.
than the high limit setting. This alarm
is suspended for nine breaths or one
minute (whichever is greater) after the
ventilator settings are changed.
low limit alarm setting. This alarm is
suspended for nine breaths or one
minute (whichever is greater) after the
ventilator settings are changed.
-10 cmH2O).
Absorber canister is not latched when
CO2 bypass is in place.
Electrical signals show the flow
sensor is not connected.
Electrical signals show the flow
sensor is not connected.
The O2 supply has failed for 10
seconds.
than or equal to 30 seconds.
limit setting.
Make sure the breathing system is correctly
installed. Continue to use normally. Contact a
Datex-Ohmeda trained service representative.
Manually ventilate the patient to save power.
Make sure power is connected and circuit
breakers are closed.
Ohmeda trained service representative.
Manually ventilate the patient to save power.
Make sure power is connected and circuit
breakers are closed.
Default settings are used. Ventilation is still
possible but service is necessary. Contact a
Datex-Ohmeda trained service representative.
Contact a Datex-Ohmeda trained service
representative.
Contact a Datex-Ohmeda trained service
representative.
Ventilate manually. Cycle system power (OnStandby-On). If the alarm clears, restart
mechanical ventilation. Contact a DatexOhmeda trained service representative.
Check patient for spontaneous breathing. Adjust
control settings.
Check the patient condition for spontaneous
activity. Increase fresh gas flow. Look for high
flow through gas scavenging. Calibrate the flow
sensors. With active scavenging, check the
negative relief valve on the receiver.
Check that the absorber canister is properly
latched. Continue to use the system normally.
Contact a Datex-Ohmeda trained service
representative.
Connect the flow sensors. Make sure the flow
sensor module is on all the way.
Connect the flow sensors. Make sure the flow
sensor module is on all the way.
Air flow will continue. Ventilate manually if
necessary. Connect a pipeline supply or install
an O2 cylinder.
Stop pressing the O2 flush button. If this alarm
occurs when flush is not in use, contact a DatexOhmeda trained service representative.
Verify that the limit it set correctly. Check that
the O2 flow is adequate. Calibrate the O2
sensor. If calibration fails, replace the O2
sensor.
6-6M1132382
Page 73
6 Alarms and Troubleshooting
MessagePriorityCauseAction
O2% lowHighO2% is less than the alarm low limit
PEEP high.
Blockage?
PEEP high.
Blockage?
Plug in power
cable. On battery.
Positive SIB Vref
out-of-range
Ppeak highHighPpeak is greater than Pmax. The
Ppeak low. Leak?MediumPpeak is less than 4 cm above Pmin
Pressure/volume
monitor inactive
Replace exp flow
sensor
Replace insp flow
sensor
Replace O2 sensor InformationalO2% is less than 5%.Make sure the patient receives O2. Use a
Reverse exp flowMedium
Reverse insp flowMedium
Select gas outletMediumWith ACGO On, the flow sensors
Service calibrationInformationalCorrupt or invalid calibration data for
Software errorMinimum
Software/Hardware
Incompatibility
System leak? Medium
TFS module errorInformationalTFS module connected but not
Minimum shutdown
(High)
HighPaw greater than or equal to
Medium
(Informational after
acknowledge)
Minimum shutdown
(High)
Medium
(Informational after
acknowledge)
InformationalThe system cannot read the
InformationalThe system cannot read the
(Informational after
acknowledge)
(Informational after
acknowledge)
Shutdown (High)
Minimum
Shutdown (High)
(Informational after
acknowledge)
setting.
Paw greater than 100 cmH2O for 10
seconds.
sustained pressure limit for 15
1
seconds
The mains supply is not connected or
has failed and the system is using
battery power.
Ventilator malfunction.Ventilate manually. Monitoring is still available.
ventilator cycles to expiration.
for 20 seconds when RR is greater
than or equal to four.
Ppeak is less than 4 cm above Pmin
for 35 seconds when RR is less than
four.
The outlet selection switch is set to
the ACGO.
calibration data stored in the sensor.
calibration data stored in the sensor.
Flow through the expiratory sensor
during inspiration (for six breaths in a
row).
Flow through the inspiratory sensor
during expiration (for six breaths in a
row).
have detected three breaths in patient
circuit during the last 30 seconds.
the flow valve.
Indicates that a software error has
occurred.
Compatibility error between the
software and the hardware.
Leak detected between ventilator and
patient circuit.
operating correctly.
.
Verify that the limit it set correctly. Check that
the O2 flow is adequate. Calibrate the O2
sensor. If calibration fails, replace the O2
sensor. As the cell wears out, the measured %
of O2 decreases.
Check the tubing and breathing system for
blockages. Ventilate manually. Contact a DatexOhmeda trained service representative.
Check tubing for kinks, blockages, disconnects.
Calibrate the flow sensors.
Ventilate manually to save power. At full charge,
the battery permits approximately 90 minutes of
mechanical ventilation. Make sure power is
connected and circuit breakers are closed.
Contact a Datex-Ohmeda trained service
representative.
Verify Pmax and other controls are set correctly.
Look for blockages. Check patient connection.
Verify circuit connections are okay. Look at the
Paw gauge on the absorber. Look for circuit
disconnection.
Connect the patient circuit to the auxiliary outlet.
For mechanical ventilation or ventilation with
monitoring, select the circle system setting.
Operation continues with reduced accuracy.
Replace the flow sensor.
Operation continues with reduced accuracy.
Replace the flow sensor.
different monitor. Calibrate the O2 sensor.
Replace the O2 cell.
Look at the check valves. Check for water
buildup in the flow sensor tubes. Check the flow
sensor condition. Replace the expiratory check
valve.
Look at the check valves. Check for water
buildup in the flow sensor tubes. Check the flow
sensor condition. Replace the expiratory check
valve.
Turn the ACGO off or connect the non-circle
patient circuit to the ACGO. Note: The bag arm
will not ventilate a patient at the auxiliary outlet.
The system is operational. Contact a DatexOhmeda trained service representative.
Ventilate manually. Monitoring is not reliable.
Contact a Datex-Ohmeda trained service
representative.
Ventilate manually. Contact a Datex-Ohmeda
trained service representative.
Look for leaks in the absorber system. Check
the integrity of the flow sensors. Zero the flow
sensors. Inspect for leaks (repair). Inspect or
replace flow sensors.
System is operational, but TFS is not
functioning correctly. Contact a Datex-Ohmeda
trained service representative.
M11323826-7
Page 74
Aespire View
MessagePriorityCauseAction
TFS module error Air sensor
TFS module error N2O sensor
TFS module error O2 sensor
TVexp highMediumExhaled tidal volume is greater than
TVexp lowMediumExhaled tidal volume is less than the
TV not achievedInformationalTidal volume measured by inspiratory
Unable to drive
bellows
Ventilator has no
drive gas
Vol vent only. No
PEEP or PSV.
InformationalTFS module connected but the Air
InformationalTFS module connected but N2O fresh
InformationalTFS module connected but O2 fresh
InformationalManifold pressure is greater than
Medium
(Informational after
acknowledge)
MediumManifold pressure error. Pressure
sensor is reporting an error.
gas sensor is reporting an error.
gas sensor is reporting an error.
the high alarm limit. This alarm is
suspended for nine breaths after a
change in the ventilator settings.
low alarm limit. This alarm is
suspended for nine breaths after a
change in the ventilator settings.
flow sensor is less than the set value
for six breaths in a row after the first
minute of mechanical ventilation.
Paw.
The ventilator does not detect supply
pressure.
control unavailable.
System is operational, but the Air flow sensor is
not functioning correctly. Contact a DatexOhmeda trained service representative.
System is operational, but N2O sensor is not
functioning correctly. Contact a Datex-Ohmeda
trained service representative.
System is operational, but the O2 flow sensor is
not functioning correctly. Contact a DatexOhmeda trained service representative.
Check patient for spontaneous breathing.
Check ventilator and alarm settings.
Adjust controls to supply adequate tidal
volumes. Check I:E, Pmax, and volume
settings. Possible leak. Modify settings or check
for system leaks.
Check the drive gas. Increase fresh gas flow (or
push the O2 flush button) to fill the bellows.
Manually ventilate the patient. Make sure that
the appropriate gas supplies (O2 or Air) are
connected and pressurized.
Continue to use volume control ventilation or
ventilate manually. Shut down system as soon
as possible. Contact a Datex-Ohmeda trained
service representative.
1
The sustained pressure threshold is calculated from the pressure limit setting. The sustained limit is
calculated as follows:
Mechanical
ventilation with
PEEP Off
Mechanical
ventilation with
PEEP On
Mechanical
Ventilation Off
For Pmax less than or equal to 30 cmH2O, the sustained pressure limit is 6 cmH2O.
For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of Pmax.
For Pmax greater than or equal to 60 cmH2O, the sustained pressure limit is 12 cmH2O.
For Pmax less than or equal to 30 cmH2O, the sustained pressure limit is 6 cmH2O plus
“set PEEP” minus 2 cmH2O.
For Pmax between 30 and 60 cmH2O, the sustained pressure limit is 20% of Pmax plus
“set PEEP” minus 2 cmH2O.
For Pmax greater than 60 cmH2O, the sustained pressure max is 12 cmH2O plus “set
PEEP” minus 2 cmH2O.
For Pmax between 12 and 60 cmH2O, the sustained pressure limit is 50% of Pmax.
For Pmax greater than 60 cmH2O, the sustained pressure limit is 30 cmH2O.
6-8M1132382
Page 75
6 Alarms and Troubleshooting
Minimum system
shutdown and
monitoring alarms
There are two special alarm types:
•Minimum system monitoring alarms stop mechanical ventilation.
•Minimum system shutdown alarms stop mechanical ventilation
and monitoring.
The software goes to minimum system monitoring when a nonrecoverable error occurs during bootup or normal operation. The
ventilator shows data and mechanical ventilation stops. ‘Monitoring
only’ appears in the Alarm message area and a specific failure
message appears in the User message area.
The following are minimum system monitoring alarms:
•Manifold pressure sensor failure.
•Pressure limit switch failure.
•Flow valve (DAC) failure.
•Flow valve (current) failure.
•Valve power failure.
A severe malfunction causes minimum system shutdown alarms.
This condition prevents mechanical ventilation and monitoring. If this
condition occurs:
•Ventilate manually.
•Use a stand-alone monitor.
•Cycle system power (On-Standby-On).
If the alarm clears, restart mechanical ventilation. If the alarm does
not clear, contact a Datex-Ohmeda
trained service representative.
M11323826-9
Page 76
Aespire View
Alarm ranges
Alarm tests
AlarmRangeIncrementDefault
O2% highOff, 21-99%1%Off
O2% low18-99%1%21%
MVexp highOff, 0.5-30 l/min0.5 l/min10.0 l/min
MVexp lowOff, 0.1-10 l/min0.1 l/min2.0 l/min
TVexp high20-1600 ml20 ml1000 ml
TVexp lowOff, 5-1500 ml5 ml for less than
20 ml,
otherwise 20 ml
Leak AudioOn, OffN/AOn
Test the system to verify that alarms are functioning:
1. Connect a test lung to the patient connection.
2. Start a case.
3. Set the Bag/Vent switch to Vent.
Off
4. Set the controls.
•Ventilation mode: Volume Control (VC).
•Ventilator.
— Tidal Vol: 400 ml
— Rate: 12
— I:E Ratio: 1:2
— Plimit: 40 cmH2O
— PEEP: Off
•Anesthesia machine.
— O2 flow: minimum flow (25-75 ml/min).
— All other gases: Off
— Push O2 flush button to fill bellows.
5. Set the O2 concentration to 30% and allow the O2 reading to
stabilize.
6-10M1132382
Page 77
6 Alarms and Troubleshooting
6. Test the O2 alarms.
•Remove the O2 cell from the breathing circuit and make sure
it measures approximately 21% O2 in room air.
•Set the O2 low alarm limit to 50%. Make sure the O2% low
alarm occurs.
•Set the O2 low alarm limit back to 21% and make sure that
the O2% low alarm cancels.
•Install the O2 cell in the breathing circuit.
•Set the O2 high alarm limit to 50%.
•Push the O2 flush button to fill the breathing system.
•Make sure the O2% high alarm occurs.
•Set the O2 high alarm limit back to Off. Make sure that the
O2% high alarm cancels.
•After two minutes in pure O2, the O2 cell measures
approximately 100% O2.
7. Test the MVexp low alarm.
•Go to the Alarm Setup menu.
•Set the MV low alarm limit to 6.0 l/min
•Make sure the MVexp low alarm occurs.
•Set the MV low alarm limit to Off.
8. Test the Ppeak high alarm.
•Set Pmax to less than the peak airway pressure.
•Make sure the Ppeak high alarm occurs.
•Set Pmax to the desired level.
9. Test the PEEP high. Blockage? alarm.
•Close the APL valve.
•Set the Bag/Vent switch to Bag. Mechanical ventilation stops.
•Block the patient connection and push the O2 flush button.
•Make sure the PEEP high. Blockage? alarm occurs after
approximately 15 seconds.
10. Test the Ppeak low. Leak? alarm.
•Unblock the patient connection.
•Set the Bag/Vent switch to Vent.
•Set the tidal volume and total flow to minimum.
•Other alarms such as MVexp low can occur.
•Make sure that the Ppeak low. Leak? alarm occurs.
11. Set all alarm limits to approved clinical values.
M11323826-11
Page 78
Aespire View
Breathing system problems
SystemProblemSolution
Gas scavenging flow is
too low or too high.
The bellows fills when the
Bag/Vent switch is set to
Bag or the bag fills when
the switch is set to Vent.
The ventilator does not
read the position of the
Bag/Vent switch.
APL valve does not
operate correctly.
Large breathing system
leak not quickly located.
Bellows falls below top of
indicator during “Bellows
test.”
Scavenging extract flow
problem.
Filter blockage. Active systems
have a flow indicator.
Leak through the Bag/Vent
switch.
Ventilator or absorber
malfunction.
APL valve problem.Replace APL valve seal and
Bag hose not connected
properly.
Absorber canister not installed
correctly.
Leak in the breathing system.Check, clean or reposition the
Use a different scavenging
extraction system. Verify flow is
within specification.
Replace the filter. See
“Remove the AGSS Receiver
Filter” in the “ABS Cleaning
and Sterilization” manual.
Contact a Datex-Ohmeda
trained service representative
to repair the system.
Ventilate manually. Contact a
Datex-Ohmeda trained service
representative to repair the
system.
diaphragm.
Ensure that the bag hose is
connected to the bag port
(below the APL valve).
Reinstall the absorber canister,
ensure both pins are engaged.
pressure relief valve. If the
problem persists, replace the
pressure relief valve, bellows
base, or bellows assembly.
6-12M1132382
Page 79
Electrical problems
WARNINGIf a circuit breaker opens frequently, do not use the
SystemProblemSolution
Mains indicator is not on. The electrical power cable is not
One electrical outlet does
not have power.
A circuit breaker opens
frequently.
Tec 6 series vaporizer
has no power.
6 Alarms and Troubleshooting
system. Have a Datex-Ohmeda trained service
representative repair the system.
Connect the power cable.
connected.
The inlet circuit breaker (switch)
is off.
The power cable is damaged.Replace the power cable.
The electrical socket the power
cable connects to had no power.
An internal fuse is open.Have a Datex-Ohmeda trained
The outlet circuit breaker is off.Turn the circuit breaker on.
Equipment connected to the
outlet uses more current than
the circuit breaker rating.
The equipment connected to the
outlet has a short.
Not plugged into outlet.Connect power cable.
The outlet circuit breaker is offTurn the circuit breaker on.
Turn the circuit breaker on.
Use a different electrical
socket.
service representative repair
the system.
Use a different power supply
for some of the equipment.
Have a Datex-Ohmeda trained
service representative repair
the system.
M11323826-13
Page 80
Aespire View
Pneumatic problems
SystemProblemSolution
High-pressure leak test
fails.
Low-pressure leak test
fails with a vaporizer on.
Low-pressure leak with a
vaporizer off.
Controls are not set correctly.Make sure that no gas is
flowing, turn off the auxiliary
flowmeter, and repeat the test.
Incorrect cylinder connection.Make sure that there is only
The vaporizer is not correctly
installed.
The vaporizer filler is loose
(funnel fill type vaporizer).
Vaporizer port o-rings (external)
are damaged or not installed.
A vaporizer malfunction (the leak
stops if a different vaporizer is
used in the same manifold
position).
A port valve malfunction (the
leak continues if a different
vaporizer in the same manifold
position).
Anesthesia machine problem.Contact a Datex-Ohmeda
one cylinder gasket, the gasket
is in good condition, and the
connection is tight.
Correctly install the vaporizer.
Tighten the filler.
Install new o-rings.
Send the vaporizer to an
approved service center for
repair.
Contact a Datex-Ohmeda
trained service representative
to repair the vaporizer
manifold.
trained service representative.
6-14M1132382
Page 81
7 User Maintenance
WARNINGTo help prevent fires:
•Only use lubricants approved for anesthesia or O2
equipment, such as Krytox.
•Do not use lubricants that contain oil or grease. They
may burn or explode in high O2 concentrations.
•All materials used to cover the system must be made
from antistatic (conductive) materials. Static electricity
can cause fires.
•Desiccated (dehydrated) absorbent material may
produce dangerous chemical reactions when exposed
to inhalation anesthetics. Adequate precautions
should be taken to ensure that absorbent does not dry
out. Turn off all gases when finished using the system.
wObey infection control and safety procedures. Used
equipment may contain blood and body fluids.
wMoveable parts and removable components may present
a pinch or a crush hazard. Use care when moving or
replacing system parts and components.
Do not use malfunctioning equipment. Make all necessary repairs or
have the equipment serviced by a Datex-Ohmeda trained service
representative. After repair, test the equipment to ensure that it is
functioning properly in accordance with Datex-Ohmeda’s published
specifications.
To ensure full reliability, have all repairs and service done by a
Datex-Ohmeda trained service representative.
No repair should ever be attempted by anyone not having training
and experience in the repair of devices of this nature.
Replace damaged parts with components manufactured or sold by
Datex-Ohmeda. Then test the unit to ascertain that it complies with
Datex-Ohmeda’s published specifications.
Contact a Datex-Ohmeda Field-Service Representative for service
assistance.
Maintenance summary and schedule
These schedules indicate the minimum frequency of maintenance
based on typical usage of 2000 hours per year. Equipment should be
serviced more frequently if it is used more than the typical yearly
usage.
For detailed cleaning instructions, refer to the Advanced Breathing
System Cleaning and Sterilization manual which accompanies the
machine.
NoteLocal policies or regulations may require that maintenance be
performed more frequently than stated here.
7-2M1132382
Page 83
User maintenance
7 User Maintenance
Minimum Frequency Maintenance
Daily•Clean the external surfaces.
Two weeksDrain the vaporizers and discard the agent. This is not
Monthly•Perform 100% O2 calibration.
During cleaning and
setup
AnnuallyReplace the external o-rings on the vaporizer ports.
As necessary•Install new cylinder gaskets on cylinder yokes.
•Perform 21% O2 calibration.
•Check the condenser reservoir (if equipped). Drain if
needed.
•Zero the flow sensors.
necessary for the Tec 6 series vaporizers.
•Lubricate all cylinder supply tee handle threads with
Krytox or a lubricant approved for use with 100% O2.
Inspect the parts for damage. Replace or repair as
necessary.
•Empty the water reservoir and replace the absorbent in
the canister.
•Empty the overflow trap on the optional suction
regulator (if equipped).
•Replace the circuit O2 cell (under typical use the cell
meets specifications for 1 year).
•Replace the disposable (plastic) flow sensors (under
typical use the non-offset flow sensors meet
specifications for a minimum of three months and the
offset flow sensors meet specifications for a minimum
of six months).
•Replace the autoclavable (metal) flow sensors (under
typical use the sensors meet specifications for a
minimum of one year).
•Replace the receiver filter (if equipped - active gas
scavenging only).
Datex-Ohmeda
approved service
This is the minimum level of maintenance recommended by
Datex-Ohmeda. Local regulations may contain additional
maintenance requirements. Datex-Ohmeda advocates compliance
with local regulations which meet or exceed this minimum level of
maintenance.
Minimum Frequency Maintenance
12 monthsHave a Datex-Ohmeda trained service representative
Breathing system maintenance
Replace any parts that are visibly cracked, chipped, distorted, or worn
when cleaning the breathing system.
complete the scheduled service maintenance checks, tests,
calibrations, and parts replacement as defined in the
Technical Reference manual.
M11323827-3
Page 84
Aespire View
AB.74p.052
AB.74p.065
AB.74p.083
O2 cell replacement
WARNINGHandle and dispose of O2 cells according to site
biohazard policies. Do not incinerate.
1. Pull the latch to unlock the flow sensor module.
2. Pull the flow sensor module out of the breathing system.
3. Remove the cable connector from the O2 cell, and unscrew the
cell counterclockwise.
ImportantMake sure that the o-ring is on the replacement O2 cell before
installation.
4. Install the replacement O2 cell and reconnect the O2 cell cable.
5. Put the flow sensor module back into the system and push the
latch closed to secure the flow sensor module.
6. Perform “O2 cell calibration” after O2 cell replacement.
7-4M1132382
Page 85
O2 cell calibration
WARNINGDo not perform O2 cell calibration while the system is
ImportantIt may take a new O2 cell 90 minutes to stabilize. If the O2 cell
7 User Maintenance
connected to a patient.
wThe O2 cell must be calibrated at the same environment
pressure at which it will be used to monitor oxygen
delivery in the patient circuit.
wOperation at pressures other than the pressures present
during calibration may result in readings outside of the
stated monitoring accuracy.
calibration fails after a new O2 cell has been installed, wait 90
minutes and repeat the calibration.
21% O2 cell
calibration
Complete a 21% O2 cell calibration before performing the 100% O2
cell calibration.
1. Push the Menu key.
2. Select Setup/Calibration.
3. Select O2 Sensor Calibration.
4. Select 21% O2.
5. Remove the O2 cell from the circuit.
•Pull the latch to unlock the flow sensor module.
•Pull the flow sensor module out of the breathing system.
•Remove the O2 cell by unscrewing the cell counterclockwise.
This exposes the O2 cell to room air.
6. Select Start Calibration. ‘Calibrating . . .’ shows on the screen
while the O2 cell is being calibrated to the room air.
7. ‘Complete’ shows on the screen upon successful calibration:
•Reinstall the O2 cell.
•Select Go to Setup/Calibration Menu.
8. If the screen shows ‘Failure,’ repeat the 21% O2 cell calibration.
9. If the calibration fails after another attempt, perform a 100% O2
cell calibration. Then try the 21% O2 cell calibration again.
10. Replace the O2 cell if repeated failures occur.
M11323827-5
Page 86
Aespire View
100% O2 cell
calibration
Zeroing flow sensor
Complete a 21% O2 cell calibration before performing a 100% O2 cell
calibration.
1. Make sure that the O2 cell is in the circuit.
2. After performing a 21% calibration, select 100% O2.
3. Push the O2 flush button for 5 seconds and set the O2 flow to
5 l/min and set other gases to minimal flow.
4. Select Start Calibration.
5. ‘Complete’ shows on the screen upon successful calibration.
6. If ‘Failure’ shows on the screen, repeat the 100% O2 cell
calibration.
•If the calibration fails after another attempt, decrease the
airway pressure and try the 100% O2 cell calibration again.
•If calibration fails after repeated attempts, perform a 21% O2
cell calibration. Then try a 100% O2 cell calibration again.
•If the 100% O2 cell calibration does not pass, replace the O2
cell.
7. Perform the “Breathing system tests” in the “Preoperative Tests”
section before using the system.
WARNINGDo not perform calibration while system is connected to a
patient.
NoteThe system automatically corrects for zero offset when the flow
sensor connectors are unplugged and the system power is on.
1. Set the Bag/Vent switch to Bag.
2. Remove the flow sensor module.
•Pull the latch to unlock the flow sensor module from the
breathing system.
•Pull the flow sensor module out of the breathing system.
3. ‘No insp flow sensor’ and ‘No exp flow sensor’ show on the
display when the zeroing is complete.
4. Reinstall the flow sensor module.
5. Perform the “Breathing system tests” in the “Preoperative Tests”
section before using the system.”
7-6M1132382
Page 87
Prevent water buildup
NotePooled water in the flow sensor or water in the sensing lines can
7 User Maintenance
Water is created from exhaled gas and a chemical reaction between
CO2 and the absorbent. Water buildup increases when the system is
used at low fresh gas flows. At low flows, more CO2 stays in the
absorber producing water and more moist exhaled gas remains in the
absorber.
cause inaccurate alarms. Small beads of water or a foggy
appearance in the flow sensors is acceptable.
To manage excess water:
•Empty the water reservoir in the absorbent canister when
changing the absorbent.
•Make sure that any water condensing in the breathing circuit
tubes is not allowed to drain into the flow sensors.
•Water condensation in the breathing circuit tubing might be
lessened by using a Heat and Moisture Exchange (HME) filter at
the airway connection.
•Install the condenser or EZchange canister option.
How to install gas cylinders. . . . . . . . . . . . . . . . . . . . . . . . . 8-13
How to attach equipment to the top of the machine . . . . . . 8-14
M11323828-1
Page 90
Aespire View
WARNINGDatex-Ohmeda strongly recommends the use of O2
monitoring with this equipment. Refer to local standards
for mandatory monitoring.
wEuropean, international, and national standards require
the following monitoring be used with this system:
•Exhaled volume monitoring.
•O2 monitoring.
•CO2 monitoring.
•Anesthetic agent monitoring be used when anesthetic
vaporizers are in use.
wAlways make sure that the pipeline supply hoses and the
breathing circuit components are not toxic and will not:
•Cause an allergic reaction in the patient.
•React with the anesthetic gases or agent to produce
dangerous by-products.
wTo prevent incorrect values or equipment malfunction,
use only cables, hoses, and tubing approved by
Datex-Ohmeda.
wThis system operates correctly at the electrical
interference levels of IEC 60601-1-2. Higher levels can
cause nuisance alarms that may stop mechanical
ventilation.
wTo help prevent false alarms from devices with high-
intensity electrical fields:
•Keep the electrosurgical leads away from the
breathing system, flow sensors, and O2 cell.
•Do not allow the electrosurgical leads to contact any
part of the anesthesia system.
•Do not use cell phones near the anesthesia system.
8-2M1132382
Page 91
8 Setup and Connections
wTo protect the patient when electrosurgical equipment is
used:
•Monitor the correct operation of all life support and
monitoring equipment.
•Keep backup manual ventilation available in case the
electrosurgical equipment prevents safe use of the
ventilator.
wDo not use antistatic or electrically-conductive breathing
tubes or masks. They can cause burns if used near high
frequency surgical equipment.
wUse only reservoir bags that comply with EN1820 on this
system.
wUse only breathing tubes that comply with EN12342 on
this system.
wA malfunction of the medical gas central supply system
may cause all connected devices to stop.
M11323828-3
Page 92
Aespire View
7
1
2
3
4
5
6
AB.74p042
AB.74p043
Canister setup
1. Canister support pin
2. Canister handle
3. Disposable Multi Absorber canister
4. Absorbent
5. Expiratory water reservoir
6. Canister release latch
7. Reusable Multi Absorber canister
Figure 8-1 • Canister
8-4M1132382
Page 93
8 Setup and Connections
WARNINGObey applicable safety precautions:
•Do not use the absorber with chloroform or
trichloroethylene.
•The Disposable Multi Absorber is a sealed unit which
should not be opened or refilled.
•The Disposable Multi Absorber cannot be disinfected
and is not autoclavable. Be aware that crosscontamination is possible.
•Avoid skin or eye contact with the contents of the
absorber. In the event of skin or eye contact,
immediately rinse the affected area with water and
seek medical assistance.
•Do not change the absorber during ventilation unless
the EZchange canister system is installed.
•Change absorbent often to prevent the buildup of
non-metabolic gases when the system is not in use.
•Inspect absorbent color at the end of a case. During
non-use, absorbent can go back to the original color.
Refer to the absorbent labeling for more information
about color changes.
•If the absorbent completely dries out, it may give off
carbon monoxide (CO) when exposed to anesthetic
agents. For safety, replace the absorbent.
•Desiccated (dehydrated) absorbent material may
produce dangerous chemical reactions when exposed
to inhalation anesthetics. Adequate precautions
should be taken to ensure that absorbent does not dry
out. Turn off all gases when finished using the system.
The absorber canister is available in two versions:
•Disposable Multi Absorber.
•Reusable Multi Absorber.
Both versions are removed and installed on the breathing system in
the same way.
Each canister holds 800 grams of loose absorbent. Datex-Ohmeda
recommends Medisorb
Both absorber versions should only be used with mixtures of air,
oxygen, nitrous oxide, halothane, enflurane, isoflurane, desflurane
and sevoflurane.
TM
absorbent.
M11323828-5
Page 94
Aespire View
AB.74p058
When to change the
absorbent
ImportantRead the canister instructions completely before using the product.
A gradual color change of the absorbent in the canister indicates
absorption of carbon dioxide. The color change of the absorbent is
only a rough indicator. Use carbon dioxide monitoring to determine
when to change the canister.
Discard the absorbent when it has changed color. If left standing for
several hours, absorbent may regain its original color giving a
misleading indication of activity.
Removing a canister1. Hold the canister by the handle and push on the release latch to
unlock the canister.
2. Remove the canister by tilting it downward and off the two
support pins.
8-6M1132382
Page 95
8 Setup and Connections
AB.75p088
AB.75p089
AC.20p003
Removing an
EZchange canister
1. Hold the canister by the handle and push the canister cradle
release latch to unlock the canister cradle.
2. Slide the canister up and out of the cradle.
Reusable Multi
Absorber canister
1. Turn the canister upside down and, using your thumbs, turn the
cover locking ring counterclockwise to unlock it.
filling
2. Push up to release the seal.
M11323828-7
Page 96
Aespire View
AC.20p003
AC.20p0004
3. Lift off the cover to remove it.
4. Remove and properly discard the foam filters, the absorbent, and
any water in the reservoir.
WARNINGBe careful when draining condensate from the absorber.
The liquid is caustic and may burn skin.
5. To clean and disinfect the canister, see “Absorber canister” in the “Cleaning and Sterilization” manual.
6. Assemble canister.
•Place a new filter in the bottom of the canister.
•Pour absorbent into the canister.
•Place a new filter over the absorbent.
•Align the cover slots with the canister locking tabs and press
the cover down into place.
•Turn the cover locking ring clockwise to lock the cover in
place.
•Ensure cover is properly sealed to prevent leaks and spillage.
•Wipe off any absorbent dust.
NoteAlignment of the arrows helps to indicate correct assembly.
8-8M1132382
Page 97
8 Setup and Connections
AB.82p001
AB.74p049
WARNINGThe filters must be in place to help prevent dust and
particles from entering the breathing circuit.
Electrical connections
OutletsLabels show outlet voltage ratings and circuit breaker amp ratings.
WARNINGEquipment connected to electrical outlets that are not
7. When replacing the canister, make sure that it is seated properly
on the support pins or in the EZchange canister module before
latching it into place.
These are isolated outlets. Regularly test the leakage current.
isolated outlets can increase the leakage current.
Regularly test the leakage current.
M11323828-9
Page 98
Aespire View
AB.74p048
AB.91p040
Mains inletArrow shows the mains power inlet and cord.
Serial portThe system has an RS-232C electrical interface. The RS-232C
connector allows serial input/output of commands and data. The 15pin connector is located on the back of the display unit.
The 15-pin female D connector - Data Communications Equipment
(DCE) configuration:
•Pin 1 - Monitor On/Standby.
•Pin 5 - Signal ground.
•Pin 6 - Receive data.
•Pin 9 - Monitor On/Standby Return.
•Pin 13 - Transmit data.
Pneumatic connections
CAUTIONUse only medical grade gas supplies. Other types of gas
supplies may contain water, oil, or other contaminants
which could affect the operation of the pneumatic system.
The gas supplies provide gas to these optional devices through
internal connections:
•Venturi suction regulator (optional).
•Auxiliary O2 flowmeter (optional).
Pipeline Inlets
8-10M1132382
Page 99
8 Setup and Connections
AB.91p045
AC.82p001
ScavengingThe scavenging assembly is located below the bellows on the
breathing system. Adapters may be necessary to interface to the
scavenging connector.
See the “Operation” section for more scavenging information.
Sample gas return
port
Connect the sample gas exhaust tube from the airway module to the
gas return port. Exhaust gas will be directed to the scavenging
system.
M11323828-11
Page 100
Aespire View
1
2
3
4
5
AB.74p032
AB.74p189
1
2
AB.91p039
Suction regulator
(optional)
Auxiliary O2
flowmeter (optional)
Venturi regulators use the system air or O2 supply. Vacuum
regulators must be connected to an external vacuum supply.
1. External vacuum (non-venturi) connection
2. Venturi muffler
3. Collection bottle connection
4. Splash guard
5. Overflow safety trap
1. Auxiliary O2 outlet
2. Auxiliary O2 flow control
8-12M1132382
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.