Datex-Ohmeda, Inc., a General Electric Company, doing business as GE Healthcare.
User Responsibility
This Product will perform in conformity with the description thereof
contained in this User’s Reference manual and accompanying labels
and/or inserts, when assembled, operated, maintained, and repaired
in accordance with the instructions provided. This Product must be
checked periodically. A defective Product should not be used. Parts
that are broken, missing, plainly worn, distorted, or contaminated
should be replaced immediately. Should repair or replacement
become necessary, Datex-Ohmeda recommends that a telephonic or
written request for service advice be made to the nearest
Datex-Ohmeda Customer Service Center. This Product or any of its
parts should not be repaired other than in accordance with written
instructions provided by Datex-Ohmeda and by Datex-Ohmeda
trained personnel. The Product must not be altered without the prior
written approval of Datex-Ohmeda. The user of this Product shall
have the sole responsibility for any malfunction which results from
improper use, faulty maintenance, improper repair, damage, or
alteration by anyone other than Datex-Ohmeda.
CAUTIONU.S. Federal law restricts this device to sale by or on the
order of a licensed medical practitioner. Outside the
U.S.A., check local laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic
which indicates a product group code, the year of manufacture, and a
sequential unit number for identification. The serial number can be in
one of two formats.
AAAX11111AAAXX111111AA
The X represents an alpha
character indicating the year
the product was manufactured;
H = 2004, J = 2005, etc. I and
O are not used.
Aespire, 7900 Ventilator, Advanced Breathing System, Link-25,
ComWheel, Disposable Multi Absorber, EZchange, Reusable
Multi Absorber, PSVPro, SmartVent, Tec 6, and Tec 7 are
registered trademarks of Datex-Ohmeda, Inc.
Other brand names or product names used in this manual are
trademarks or registered trademarks of their respective holders.
The XX represents a number
indicating the year the product
was manufactured; 04 = 2004,
The Aespire anesthesia system is a compact, integrated, and intuitive
anesthesia delivery system. The 7900 Ventilator provides mechanical
ventilation for patients during surgery as well as monitoring and
displaying various patient parameters.
The 7900 Ventilator uses a microprocessor-controlled ventilator with
internal monitors, electronic PEEP, Volume Mode, and other optional
features. A serial interface permits communication to external
monitoring.
This anesthesia system is not suitable for use in an MRI environment.
This system must only be operated by medical personnel authorized
and trained to use this product. It must be operated according to the
instructions in this User’s Reference manual.
NoteConfigurations available for this product depend on local market and
standards requirements. Illustrations in this manual may not
represent all configurations of the product. This manual does not
cover the operation of every accessory. Refer to the accessory
documentation for further information.
1-2M1132382
Symbols used in the manual or on the equipment
Symbols replace words on the equipment, on the display, or in
manuals.
Warnings and Cautions tell about the dangerous conditions that can
occur if the instructions in the manual are not followed.
Warnings tell about a condition that can cause injury to the operator
or the patient.
Cautions tell about a condition that can cause damage to the
equipment. Read and follow all warnings and cautions.
On (power)Off (power)
1 Introduction
Standby
Type BF equipmentType B equipment
Dangerous voltageFrame or chassis ground
Direct currentAlternating current
CautionAttention, refer to product instructions
Refer to product instructionsExhaust
Electrical inputElectrical output
Electrical input/outputSample gas inlet to scavenging
O2+
O2 flush button
Pneumatic inletPneumatic outlet
SN
M11323821-3
Serial number
REF
Stock number
Aespire View
EquipotentialLamp, lighting, illumination
Protective earth groundEarth ground
VariabilityVariability in steps
Suction bottle outletVacuum inlet
MaxVacuum
+
134°C
Plus, positive polarity
Bag position/manual ventilationMechanical ventilation
Inspiratory flowExpiratory flow
Movement in one directionMovement in two directions
LockUnlock
Isolation transformerLow pressure leak test
AutoclavableNot autoclavable
-
Minus, negative polarity
Bellows volumes are approximate
APL settings are approximateAnesthetic Gas Scavenging System
O2%
O2 cell connection
1-4M1132382
Pinch hazardRead to top of float
1 Introduction
EZchange Canister (CO2 bypass)Caution: federal law prohibits dispensing
Systems with this mark agree with the
European Council Directive (93/42/EEC) for
Medical Devices when they are used as
specified in their User’s Reference manuals.
The xxxx is the certification number of the
Notified Body used by GE Healthcare’s
Quality Systems.
ManufacturerDate of manufacture
without prescription
Authorized representative in the European
Community
Indicates that the waste of electrical and
electronic equipment must not be disposed
as unsorted municipal waste and must be
collected separately. Please contact an
authorized representative of Datex-Ohmeda
for information concerning the
decommissioning of equipment.
This way up
GOST R Russian certification
USB port
Ethernet connectionThis product consists of devices that
may contain mercury, which must be
recycled or disposed of in accordance
with local, state, or country laws. (Within
this system, the backlight lamps in the
monitor display contain mercury.
M11323821-5
Aespire View
Typeface conventions used
Menu items are written in bold italic typeface; for example, Main
Menu.
Messages that are displayed on the screen are enclosed in single
quotes: for example, ‘Total pressure exceeds Pmax.’
When referring to different sections and other documents, the names
are written in italic typeface and enclosed in double quotes; for
example, “System Controls and Menus.”
6System switchSet the switch to the On (|) position to permit gas flow and to turn on the system.
7Integrated suction
(optional)
8BrakePush down to lock. Lift to release.
9O2 flush buttonPush the O2 flush button to supply high flows of O2 to the breathing system.
10Auxiliary O2 flow
control (optional)
Turn the switch to MAX for full vacuum. Turn the switch to Off (0) for no vacuum. Turn
the switch to On (|) for adjustable vacuum. When in adjustable vacuum, turn the knob
clockwise to increase the vacuum and counterclockwise to decrease the vacuum.
Turn the knob counterclockwise to increase the flow. Turn the knob clockwise to
decrease the flow.
12Flow controlsTurn the knob counterclockwise to increase the flow. Turn the knob clockwise to
decrease the flow. The System switch must be On for gas to flow
M11323822-3
.
Aespire View
1
2
4
5
7
8
AC.20.012
6
3
1. Outlet circuit breaker5. Mains inlet
2. Electrical outlet6. System circuit breaker
3. Suction items (optional)7. Cylinder
4. Equipotential stud8. Pipeline connections
Figure 2-2 • Rear view
2-4M1132382
2 System Controls and Menus
AC.20.013
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
17
16
Advanced breathing system (ABS) components
1. Expiratory check valve10. Breathing system release
2. Inspiratory check valve11. Manual bag port
3. Auxiliary common gas outlet (ACGO) switch12. Adjustable pressure-limiting (APL) valve
4. ACGO13. Bag/Vent switch
5. Inspiratory flow sensor14. Bellows assembly
6. Expiratory flow sensor15. Sample gas return port
8Absorber canister releasePush to remove the canister. This causes the breathing system to vent to
Sends fresh gas to the ACGO when the switch is activated. The ACGO
provides fresh gas to an external manual breathing circuit.
Flow sensors provide volume measurements for some monitoring
functions and tidal volume delivery.
the room (unless the EZchange canister option is installed). Be sure to hold
the canister by the handle before releasing the canister.
12Adjustable pressure-limiting
(APL) valve
Adjusts breathing system pressure limit during manual ventilation. The
scale shows approximate pressures. Above 30 cmH2O, the knob will click
as it turns.
13Bag/Vent switchSelects between manual ventilation (bag) or mechanical ventilation
(ventilator).
2-6M1132382
Optional ABS
AB.82.043
AB.74.120
1
2
3
4
5
components
2 System Controls and Menus
1. Bag support arm
2. EZchange canister system (CO2 bypass)
3. EZchange canister release
4. Condenser drain button
5. Condenser
Figure 2-4 • Breathing system options
Item, Figure 2-4Description
1Bag support armSqueeze the button to raise or lower the arm.
3EZchange canister
release
4Condenser drain buttonPush to drain water out of the condenser.
Push to drop the canister to EZchange position. This seals the breathing circuit,
permitting continued ventilation and rebreathing of exhaled gases. Be sure to
hold the canister by the handle before releasing the canister.
M11323822-7
Aespire View
2
1
3
4
5
6
AA43051
AB.80.009
4
3
Vaporizer controls
Refer to the vaporizer User’s Reference manual for more detailed
information on the vaporizer.
1. Tec 6 series
2. Tec 7
3. Lock lever
4. Concentration control and release
5. Indicators (Tec 6 series)
6. Silence alarm touch key (Tec 6 series)
Figure 2-5 • Vaporizer controls
2-8M1132382
2 System Controls and Menus
Åben
Item, Figure 2-5Description
3Lock LeverTurn the lever fully clockwise to lock the vaporizer in position.
4Concentration control and
release
Push the release and turn the concentration control to set the agent
concentration. The Tec 6 series concentration control does not turn as long
as the warm-up indicator is on.
5Indicators (Tec 6 series)All indictors come on briefly at the start. The warm-up indicator goes off
after approximately 10 minutes and the operational indicator comes on.
Other indicators come on to advise the user of required action.
6Silence alarm touch key
(Tec 6 series)
Push to silence alarms. Hold for 4 seconds to sound the speaker and light
all indicators (alarm test).
M11323822-9
Aespire View
ACGO
Fresh gas flow with anesthetic agent is directed through the Auxiliary
Common Gas Outlet (ACGO) on the front of the system when the
ACGO switch is in the ACGO position. Mechanical ventilation is not
available when operating an auxiliary manual breathing circuit with
fresh gas from the ACGO. The Bag/Vent switch, APL valve, and bag
arm are not part of the external circuit. Volume and pressure
monitoring are not available.
O2 monitoring of fresh gas is available automatically when the ACGO
is selected. A sample of the fresh gas is diverted to the O2 cell in the
breathing system. The sample flow to the O2 cell is dependent on the
pressure in the external circuit. The sample flow reduces the fresh
gas flow rate to the auxiliary breathing circuit equal to the amount
delivered to the O2 cell.
Fresh gas oxygen concentration is displayed on the screen. Set the
alarm limits appropriately. Note that fresh gas oxygen concentration
may not reflect FiO2 during spontaneous breathing or in rebreathing
circuits.
ImportantUse an external O2 monitor if using a rebreathing circuit on ACGO.
Do not use an external ventilator on the ACGO. Do not use the ACGO
to drive external ventilators or for jet ventilation.
See “Scavenging” in the “Setup and Connections” section for more
information on connections.
WARNINGThe maximum pressure at the ACGO can be up to 55 kPa
(8 psi). Use a breathing circuit with pressure relief.
2-10M1132382
2 System Controls and Menus
Scavenging the
ACGO sample flow
A sample of the fresh gas is diverted to the O2 cell in the breathing
system to show the O2 numerics on the monitor display. This sample
flow should be scavenged when an auxiliary manual breathing circuit
is used with N2O or volatile anesthetics.
If scavenging is not connected, the sample flow is emptied into the
room.
To connect the scavenging:
1. Attach a circle breathing circuit to the inspiratory and expiratory
ports.
2. Occlude the circle circuit by connecting the Y-piece to the plug
located to the rear of the expiratory port.
3. Check for clinically correct settings.
4. Check the position of the Bag/Vent switch:
•If the Bag/Vent switch is set to mechanical ventilation mode,
the bellows fills slowly with the sample flow. When the
bellows is full, the sample flow goes to the AGSS
(Mechanical ventilation does not start when ACGO switch is
set to ACGO).
•If the Bag/Vent switch is set to bag mode, set the APL valve
to MIN and attach a bag. The bag fills slowly with the sample
flow. When the bag is full, the sample flow goes to the AGSS.
Scavenging from an
auxiliary manual
breathing circuit
ImportantDo not use these connectors as an outlet for exhaust flow.
Scavenging a gas
monitor sample flow
Scavenge the exhaust if an auxiliary manual breathing circuit is used
with N2O or volatile anesthetics.
An auxiliary inlet is available for active and passive AGSS units. It
provides a female connection with 30 mm male - 30 mm male
connector (or a 30 mm male - 19 mm male connector) into the
auxiliary port under the breathing system.
The auxiliary inlet is a convenience inlet to the air brake of active
AGSS units. There is a reservoir to capture exhaust flows higher than
the extract flow.
A separate exhaust hose is needed from the auxiliary manual
breathing circuit to the disposal point for all AGSS units.
Sample gas from a gas monitor can be scavenged using the sample
gas return port or the AGSS. To scavenge from a gas monitor using
the sample gas return port, connect the tubing from the monitor to the
sample gas return port. To scavenge from a gas monitor using the
AGSS, connect tubing from the monitor to the male luer inlet on the
bottom of the AGSS underneath the breathing system.
M11323822-11
Aespire View
2
3
4
5
6
789
10
11
12
13
14
AC.20.003
1
Ventilator controls
The ventilator controls include touch keys, menu screens, and a
control knob (ComWheel). The System switch provides power
functions to the ventilator display. The Bag/Vent switch starts and
stops mechanical ventilation.
1. High priority alarm indicator8. PEEP quick key
2. Low or medium priority alarm indicator9. Pmax or Psupport quick key
3. Alarm silence key10. I:E or Tinsp quick key
4. Menu key11. Respiratory rate (RR) quick key
5. End case key12. Tidal volume (TV) or Pinsp quick key
6. ComWheel13. Mains power indicator
7. More settings quick key14. Volume alarms On/Off key
Figure 2-6 • Ventilator controls
2-12M1132382
Ventilator screen
AC.20.004
Total
TV
ml
Vent On
RR
/min
I:EPmax
cmH2O
PEEP
cmH2O
More
Settings
VCV - Volume Control
Circle
Fresh Gas Flow
l/min
O2
Air
N2O
Push the knob to confirm the change. Turn the
knob to change the setting.
Time (sec)
Paw
O2% low
Ppeak
Pmean
Ppause
O2
%
TVexp
ml
MV
l/min
RR
/min
cmH2O
TVexp low
1
2
5
6
7
8
9
10
11
4
3
2 System Controls and Menus
1. Alarm silence indicator and countdown clock7. Total flow sensing (optional)
2. Alarm message areas8. Ventilator settings
3. Waveform area9. Mechanical ventilation status
4. Alarm limit settings10. Ventilation mode
5. Measured values area11. User message area
6. Circuit type
Figure 2-7 • Normal view
M11323822-13
Aespire View
TV
ml
Vent On
RR
/min
I:EPmax
cmH2O
PEEP
cmH2O
More
Settings
VCV - Volume Control
Circle
Time (sec)
Ppeak
Pmean
Ppause
O2
%
TVexp
ml
MV
l/min
RR
/min
cmH2O
Main Menu
Ventilation Mode
Alarm Setup
Setup/Calibration
Screen and Audio Setup
Cardiac Bypass
Normal Screen
VCV
PCV
SIMV/PSV
PSVPro
SIMV-PC
AC.20.009
Using menus
Figure 2-8 • Menu example
1. Push the Menu key to show the Main Menu.
2. Turn the ComWheel counterclockwise to highlight the next menu
item. Turn the ComWheel clockwise to highlight the previous
menu item.
3. Push the ComWheel to enter the highlighted window or a sub
menu.
4. Turn the ComWheel clockwise or counterclockwise to highlight
the desired selection.
5. Push the ComWheel to confirm the selection.
6. Push the Menu key to exit the menu and return to the normal
monitoring screen.
2-14M1132382
Loading...
+ 120 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.