Datex-Ohmeda Aespire View User manual

Aespire View
User’s Reference Manual
Software Revision 6.X
Datex-Ohmeda, Inc., a General Electric Company, doing business as GE Healthcare.

User Responsibility

This Product will perform in conformity with the description thereof contained in this User’s Reference manual and accompanying labels and/or inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided. This Product must be checked periodically. A defective Product should not be used. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. Should repair or replacement become necessary, Datex-Ohmeda recommends that a telephonic or written request for service advice be made to the nearest Datex-Ohmeda Customer Service Center. This Product or any of its parts should not be repaired other than in accordance with written instructions provided by Datex-Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered without the prior written approval of Datex-Ohmeda. The user of this Product shall have the sole responsibility for any malfunction which results from improper use, faulty maintenance, improper repair, damage, or alteration by anyone other than Datex-Ohmeda.
CAUTION U.S. Federal law restricts this device to sale by or on the
order of a licensed medical practitioner. Outside the U.S.A., check local laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code, the year of manufacture, and a sequential unit number for identification. The serial number can be in one of two formats.
AAAX11111 AAAXX111111AA The X represents an alpha
character indicating the year the product was manufactured;
H = 2004, J = 2005, etc. I and O are not used.
Aespire, 7900 Ventilator, Advanced Breathing System, Link-25, ComWheel, Disposable Multi Absorber, EZchange, Reusable Multi Absorber, PSVPro, SmartVent, Tec 6, and Tec 7 are
registered trademarks of Datex-Ohmeda, Inc.
Other brand names or product names used in this manual are trademarks or registered trademarks of their respective holders.
The XX represents a number indicating the year the product was manufactured; 04 = 2004,
05 = 2005, etc.

1 Introduction

Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Symbols used in the manual or on the equipment . . . . . . . . 1-3
Typeface conventions used . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6

2 System Controls and Menus

Anesthesia system controls . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Advanced breathing system (ABS) components . . . . . . . . . 2-5
Vaporizer controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
ACGO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10

Table of Contents

Optional ABS components . . . . . . . . . . . . . . . . . . . . . . . 2-7
Scavenging the ACGO sample flow . . . . . . . . . . . . . . . 2-11

3 Operation

Scavenging from an auxiliary manual breathing circuit 2-11
Scavenging a gas monitor sample flow . . . . . . . . . . . . 2-11
Ventilator controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Ventilator screen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Using menus . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
Turning on the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Start mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Start manual ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Ventilator setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Using quick keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Changing ventilator modes and settings . . . . . . . . . . . . 3-4
Alarm Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Setting volume alarms . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Alarm limit setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Setup/Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
M1132382 i
Aespire View

4 Preoperative Checkout

Screen and audio setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Cardiac Bypass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Measure circuit compliance . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Pressure waveform . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
EZchange canister (optional) . . . . . . . . . . . . . . . . . . . . . . . 3-10
Condenser (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Passive AGSS (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Connecting passive AGSS . . . . . . . . . . . . . . . . . . . . . . 3-12
Active AGSS (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-13
Connecting active AGSS with a flow indicator . . . . . . . 3-14
Connecting active adjustable AGSS . . . . . . . . . . . . . . 3-15
Total flow sensing (optional) . . . . . . . . . . . . . . . . . . . . . . . 3-15
Every day before your first patient . . . . . . . . . . . . . . . . . . . . 4-2

5 Preoperative Tests

Before every patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
Inspect the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-2
Power failure alarm test . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Pipeline test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-3
Total flow sensing test (if equipped) . . . . . . . . . . . . . . . . . . 5-3
Cylinder test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-4
Flow control test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-5
Vaporizer installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-7
Vaporizer back pressure test . . . . . . . . . . . . . . . . . . . . . . . . 5-8
Low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-9
Negative low-pressure leak test . . . . . . . . . . . . . . . . . . . 5-9
Positive low-pressure leak test . . . . . . . . . . . . . . . . . .5-10
Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-11
Breathing system tests . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Bellows test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-13
Breathing circuit test . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
APL valve test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-14
Monitor and ventilator tests . . . . . . . . . . . . . . . . . . . . . . . . 5-15
ii M1132382

6 Alarms and Troubleshooting

Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
List of alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-4
Alarm ranges . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Alarm tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-10
Breathing system problems . . . . . . . . . . . . . . . . . . . . . . . . 6-12
Electrical problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-13
Pneumatic problems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-14

7 User Maintenance

Table of Contents
Audio . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-2
Latching alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-3
Minimum system shutdown and monitoring alarms . . . . 6-9
Repair policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-2
Maintenance summary and schedule . . . . . . . . . . . . . . . . . 7-2
User maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-3
Datex-Ohmeda approved service . . . . . . . . . . . . . . . . . 7-3
Breathing system maintenance . . . . . . . . . . . . . . . . . . . . . . 7-3
O2 cell replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-4
O2 cell calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
21% O2 cell calibration . . . . . . . . . . . . . . . . . . . . . . . . . 7-5
100% O2 cell calibration . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Zeroing flow sensor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-6
Prevent water buildup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7-7
M1132382 iii
Aespire View

8 Setup and Connections

Canister setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-4
When to change the absorbent . . . . . . . . . . . . . . . . . . . 8-6
Removing a canister . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-6
Removing an EZchange canister . . . . . . . . . . . . . . . . . . 8-7
Reusable Multi Absorber canister filling . . . . . . . . . . . . . 8-7
Electrical connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Outlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-9
Mains inlet . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Serial port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Pneumatic connections . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Pipeline Inlets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-10
Scavenging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8-11
Sample gas return port . . . . . . . . . . . . . . . . . . . . . . . . 8-11

9 Parts

Suction regulator (optional) . . . . . . . . . . . . . . . . . . . . . 8-12
Auxiliary O2 flowmeter (optional) . . . . . . . . . . . . . . . . . 8-12
How to install gas cylinders . . . . . . . . . . . . . . . . . . . . . . . . 8-13
Pin indexed cylinder yokes . . . . . . . . . . . . . . . . . . . . .8-13
DIN cylinder connections . . . . . . . . . . . . . . . . . . . . . . . 8-13
High-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . 8-14
How to attach equipment to the top of the machine . . . . . . 8-14
Flow sensor module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Exhalation valve assembly . . . . . . . . . . . . . . . . . . . . . . . . . . 9-2
Breathing circuit module . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-3
Bellows assembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-4
Absorber canister . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-5
AGSS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-6
EZchange canister system . . . . . . . . . . . . . . . . . . . . . . . . . . 9-7
Condenser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9-8
Test tools and system parts . . . . . . . . . . . . . . . . . . . . . . . . . 9-9
iv M1132382

10 Specifications and Theory of Operation

System pneumatic circuits . . . . . . . . . . . . . . . . . . . . . . . . . 10-2
Gas supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
O2 flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
N2O flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Air flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-4
Mixed gas . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
EZchange canister . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Condenser . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Pneumatic specifications . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Gas supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
ACGO Port relief . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-5
Electrical block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-6
Electrical power . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-7
Table of Contents
Power cord . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
Battery information . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-8
Flow specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
Pneumatic flow . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-9
Breathing system specifications . . . . . . . . . . . . . . . . . . . . 10-10
Gas scavenging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-11
Physical specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-12
Environmental requirements . . . . . . . . . . . . . . . . . . . . . . 10-12
Suction regulators (optional) . . . . . . . . . . . . . . . . . . . . . . 10-13
Auxiliary O2 flowmeter (optional) . . . . . . . . . . . . . . . . . . . 10-13
Ventilator theory . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-14
O2 monitoring theory of operation . . . . . . . . . . . . . . . 10-15
Modes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-15
Ventilator operating specifications . . . . . . . . . . . . . . . . . . 10-23
Pneumatics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-23
Fresh gas compensation . . . . . . . . . . . . . . . . . . . . . . 10-23
Pressure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-23
Volume . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-23
Oxygen . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-23
Ventilator accuracy data . . . . . . . . . . . . . . . . . . . . . . . . . 10-24
M1132382 v
Aespire View

Index

Warranty

Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . 10-26
Guidance and manufacturer’s declaration -
electromagnetic emissions . . . . . . . . . . . . . . . . . . . . . 10-26
Guidance and manufacturer’s declaration -
electromagnetic immunity . . . . . . . . . . . . . . . . . . . . . 10-27
Recommended separation distances . . . . . . . . . . . . . 10-29
Electrical safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-30
IEC 60601-1 Classification . . . . . . . . . . . . . . . . . . . . . . . . 10-31
Standards . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-31
System components . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-32
Integral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10-32
Not integral . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10-32
vi M1132382

1 Introduction

In this section Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Symbols used in the manual or on the equipment . . . . . . . . 1-3
Typeface conventions used . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
Abbreviations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-6
M1132382 1-1
Aespire View

Intended use

The Aespire anesthesia system is a compact, integrated, and intuitive anesthesia delivery system. The 7900 Ventilator provides mechanical ventilation for patients during surgery as well as monitoring and displaying various patient parameters.
The 7900 Ventilator uses a microprocessor-controlled ventilator with internal monitors, electronic PEEP, Volume Mode, and other optional features. A serial interface permits communication to external monitoring.
This anesthesia system is not suitable for use in an MRI environment.
This system must only be operated by medical personnel authorized and trained to use this product. It must be operated according to the instructions in this User’s Reference manual.
Note Configurations available for this product depend on local market and
standards requirements. Illustrations in this manual may not represent all configurations of the product. This manual does not cover the operation of every accessory. Refer to the accessory documentation for further information.
1-2 M1132382

Symbols used in the manual or on the equipment

Symbols replace words on the equipment, on the display, or in manuals.
Warnings and Cautions tell about the dangerous conditions that can occur if the instructions in the manual are not followed.
Warnings tell about a condition that can cause injury to the operator or the patient.
Cautions tell about a condition that can cause damage to the equipment. Read and follow all warnings and cautions.
On (power) Off (power)
1 Introduction
Standby
Type BF equipment Type B equipment
Dangerous voltage Frame or chassis ground
Direct current Alternating current
Caution Attention, refer to product instructions
Refer to product instructions Exhaust
Electrical input Electrical output
Electrical input/output Sample gas inlet to scavenging
O2+
O2 flush button
Pneumatic inlet Pneumatic outlet
SN
M1132382 1-3
Serial number
REF
Stock number
Aespire View
Equipotential Lamp, lighting, illumination
Protective earth ground Earth ground
Variability Variability in steps
Suction bottle outlet Vacuum inlet
Max Vacuum
+
134°C
Plus, positive polarity
Bag position/manual ventilation Mechanical ventilation
Inspiratory flow Expiratory flow
Movement in one direction Movement in two directions
Lock Unlock
Isolation transformer Low pressure leak test
Autoclavable Not autoclavable
-
Minus, negative polarity
Bellows volumes are approximate
APL settings are approximate Anesthetic Gas Scavenging System
O2%
O2 cell connection
1-4 M1132382
Pinch hazard Read to top of float
1 Introduction
EZchange Canister (CO2 bypass) Caution: federal law prohibits dispensing
Open drain (remove liquid) Close drain
Alarm silence touch key Menu touch key
Silence alarm touch key (Tec 6) Volume alarms On/Off touch key
End case touch key Cylinder
Systems with this mark agree with the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their User’s Reference manuals. The xxxx is the certification number of the Notified Body used by GE Healthcare’s Quality Systems.
Manufacturer Date of manufacture
without prescription
Authorized representative in the European Community
Indicates that the waste of electrical and electronic equipment must not be disposed as unsorted municipal waste and must be collected separately. Please contact an authorized representative of Datex-Ohmeda for information concerning the decommissioning of equipment.
This way up
GOST R Russian certification
USB port
Ethernet connection This product consists of devices that
may contain mercury, which must be recycled or disposed of in accordance with local, state, or country laws. (Within this system, the backlight lamps in the monitor display contain mercury.
M1132382 1-5
Aespire View

Typeface conventions used

Menu items are written in bold italic typeface; for example, Main Menu.
Messages that are displayed on the screen are enclosed in single quotes: for example, ‘Total pressure exceeds Pmax.’
When referring to different sections and other documents, the names are written in italic typeface and enclosed in double quotes; for example, “System Controls and Menus.”

Abbreviations

Abbreviation Definition
A
ABS Advanced breathing system
ACGO Auxiliary common gas outlet
AGSS Anesthetic gas scavenging system
APL Adjustable pressure-limiting valve
C
CO2 Carbon dioxide
D
DAC Digital to analog converter
E
EMC Electromagnetic compatibility
ESD Electrostatic discharge
I:E Inspiratory-expiratory ratio
F
FiO2 Fraction of inspired oxygen
M
MIN Minimum
MV Minute volume
MVexp Expired minute volume
O
O2 Oxygen
1-6 M1132382
1 Introduction
P
Paw Patient airway pressure
PCV Pressure controlled ventilation
PCV-VG Pressure controlled ventilation - volume guaranteed
PEEP Positive end expiratory pressure
Pmax Maximum pressure
Pmean Average pressure calculated over the patient breath
Pinsp Target airway pressure
Ppause Positive airway pressure measured at the end of
Tpause
Ppeak Maximum airway pressure measured during patient
breath
Psupport Pressure support
PSVPro Pressure supported ventilation with apnea backup
R
RR Respiratory rate
S
SIMV-PC Synchronized intermittent mandatory ventilation -
pressure controlled
SIMV/PSV Synchronized intermittent mandatory ventilation /
pressure supported ventilation
T
TFS Total flow sensing
Tpause Pause time
Tinsp Inspired tidal volume
TV Tidal volume
TVexp Expired tidal volume
V
VCV Volume controlled ventilation
M1132382 1-7
Aespire View
1-8 M1132382

2 System Controls and Menus

WARNING Explosion Hazard. Do not use this system with flammable
anesthetic agents.
w Do not use antistatic or electrically-conductive breathing
tubes or masks. They can cause burns if used near high frequency surgical equipment.
In this section Anesthesia system controls . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Advanced breathing system (ABS) components . . . . . . . . . 2-5
Vaporizer controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8
ACGO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-10
Ventilator controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12
Ventilator screen. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-13
Using menus. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-14
M1132382 2-1
Aespire View
AC.20.002
1
2
13
11
10
9
8
4
3
5
6
7
12

Anesthesia system controls

1. Light switch 8. Brake
2. Dovetail rails 9. O2 flush button
3. Vaporizer 10. Auxiliary O2 flow control (optional)
4. Pipeline pressure gauge(s) (upper row) 11. Breathing system
5. Cylinder pressure gauge(s) (lower row) 12. Flow controls
6. System switch 13. Ventilator display
7. Integrated suction (optional)
Figure 2-1 • Front view
2-2 M1132382
2 System Controls and Menus
Item, Figure 2-1 Description
6 System switch Set the switch to the On (|) position to permit gas flow and to turn on the system.
7 Integrated suction
(optional)
8 Brake Push down to lock. Lift to release.
9 O2 flush button Push the O2 flush button to supply high flows of O2 to the breathing system.
10 Auxiliary O2 flow
control (optional)
Turn the switch to MAX for full vacuum. Turn the switch to Off (0) for no vacuum. Turn the switch to On (|) for adjustable vacuum. When in adjustable vacuum, turn the knob clockwise to increase the vacuum and counterclockwise to decrease the vacuum.
Turn the knob counterclockwise to increase the flow. Turn the knob clockwise to decrease the flow.
12 Flow controls Turn the knob counterclockwise to increase the flow. Turn the knob clockwise to
decrease the flow. The System switch must be On for gas to flow
M1132382 2-3
.
Aespire View
1
2
4
5
7
8
AC.20.012
6
3
1. Outlet circuit breaker 5. Mains inlet
2. Electrical outlet 6. System circuit breaker
3. Suction items (optional) 7. Cylinder
4. Equipotential stud 8. Pipeline connections
Figure 2-2 • Rear view
2-4 M1132382
2 System Controls and Menus
AC.20.013
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
17
16

Advanced breathing system (ABS) components

1. Expiratory check valve 10. Breathing system release
2. Inspiratory check valve 11. Manual bag port
3. Auxiliary common gas outlet (ACGO) switch 12. Adjustable pressure-limiting (APL) valve
4. ACGO 13. Bag/Vent switch
5. Inspiratory flow sensor 14. Bellows assembly
6. Expiratory flow sensor 15. Sample gas return port
7. Absorber canister 16. Scavenger flow indicator (optional)
8. Absorber canister release 17. Airway pressure gauge
9. Leak test plug
Figure 2-3 • Advanced breathing system
M1132382 2-5
Aespire View
20
30
70
M
I
N
Item, Figure 2-3 Description
3 Auxiliary common gas outlet
(ACGO) switch
5,6 Inspiratory flow sensor and
Expiratory flow sensor
8 Absorber canister release Push to remove the canister. This causes the breathing system to vent to
Sends fresh gas to the ACGO when the switch is activated. The ACGO provides fresh gas to an external manual breathing circuit.
Flow sensors provide volume measurements for some monitoring functions and tidal volume delivery.
the room (unless the EZchange canister option is installed). Be sure to hold the canister by the handle before releasing the canister.
12 Adjustable pressure-limiting
(APL) valve
Adjusts breathing system pressure limit during manual ventilation. The scale shows approximate pressures. Above 30 cmH2O, the knob will click as it turns.
13 Bag/Vent switch Selects between manual ventilation (bag) or mechanical ventilation
(ventilator).
2-6 M1132382
Optional ABS
AB.82.043
AB.74.120
1
2
3
4
5
components
2 System Controls and Menus
1. Bag support arm
2. EZchange canister system (CO2 bypass)
3. EZchange canister release
4. Condenser drain button
5. Condenser
Figure 2-4 • Breathing system options
Item, Figure 2-4 Description
1 Bag support arm Squeeze the button to raise or lower the arm.
3 EZchange canister
release
4 Condenser drain button Push to drain water out of the condenser.
Push to drop the canister to EZchange position. This seals the breathing circuit, permitting continued ventilation and rebreathing of exhaled gases. Be sure to hold the canister by the handle before releasing the canister.
M1132382 2-7
Aespire View
2
1
3
4
5
6
AA43051
AB.80.009
4
3

Vaporizer controls

Refer to the vaporizer User’s Reference manual for more detailed information on the vaporizer.
1. Tec 6 series
2. Tec 7
3. Lock lever
4. Concentration control and release
5. Indicators (Tec 6 series)
6. Silence alarm touch key (Tec 6 series)
Figure 2-5 • Vaporizer controls
2-8 M1132382
2 System Controls and Menus
Åben
Item, Figure 2-5 Description
3 Lock Lever Turn the lever fully clockwise to lock the vaporizer in position.
4 Concentration control and
release
Push the release and turn the concentration control to set the agent concentration. The Tec 6 series concentration control does not turn as long as the warm-up indicator is on.
5 Indicators (Tec 6 series) All indictors come on briefly at the start. The warm-up indicator goes off
after approximately 10 minutes and the operational indicator comes on. Other indicators come on to advise the user of required action.
6 Silence alarm touch key
(Tec 6 series)
Push to silence alarms. Hold for 4 seconds to sound the speaker and light all indicators (alarm test).
M1132382 2-9
Aespire View

ACGO

Fresh gas flow with anesthetic agent is directed through the Auxiliary Common Gas Outlet (ACGO) on the front of the system when the ACGO switch is in the ACGO position. Mechanical ventilation is not available when operating an auxiliary manual breathing circuit with fresh gas from the ACGO. The Bag/Vent switch, APL valve, and bag arm are not part of the external circuit. Volume and pressure monitoring are not available.
O2 monitoring of fresh gas is available automatically when the ACGO is selected. A sample of the fresh gas is diverted to the O2 cell in the breathing system. The sample flow to the O2 cell is dependent on the pressure in the external circuit. The sample flow reduces the fresh gas flow rate to the auxiliary breathing circuit equal to the amount delivered to the O2 cell.
Fresh gas oxygen concentration is displayed on the screen. Set the alarm limits appropriately. Note that fresh gas oxygen concentration may not reflect FiO2 during spontaneous breathing or in rebreathing circuits.
Important Use an external O2 monitor if using a rebreathing circuit on ACGO.
Do not use an external ventilator on the ACGO. Do not use the ACGO to drive external ventilators or for jet ventilation.
See “Scavenging” in the “Setup and Connections” section for more information on connections.
WARNING The maximum pressure at the ACGO can be up to 55 kPa
(8 psi). Use a breathing circuit with pressure relief.
2-10 M1132382
2 System Controls and Menus
Scavenging the
ACGO sample flow
A sample of the fresh gas is diverted to the O2 cell in the breathing system to show the O2 numerics on the monitor display. This sample flow should be scavenged when an auxiliary manual breathing circuit is used with N2O or volatile anesthetics.
If scavenging is not connected, the sample flow is emptied into the room.
To connect the scavenging:
1. Attach a circle breathing circuit to the inspiratory and expiratory ports.
2. Occlude the circle circuit by connecting the Y-piece to the plug located to the rear of the expiratory port.
3. Check for clinically correct settings.
4. Check the position of the Bag/Vent switch:
If the Bag/Vent switch is set to mechanical ventilation mode,
the bellows fills slowly with the sample flow. When the bellows is full, the sample flow goes to the AGSS (Mechanical ventilation does not start when ACGO switch is set to ACGO).
If the Bag/Vent switch is set to bag mode, set the APL valve
to MIN and attach a bag. The bag fills slowly with the sample flow. When the bag is full, the sample flow goes to the AGSS.
Scavenging from an
auxiliary manual breathing circuit
Important Do not use these connectors as an outlet for exhaust flow.
Scavenging a gas
monitor sample flow
Scavenge the exhaust if an auxiliary manual breathing circuit is used with N2O or volatile anesthetics.
An auxiliary inlet is available for active and passive AGSS units. It provides a female connection with 30 mm male - 30 mm male connector (or a 30 mm male - 19 mm male connector) into the auxiliary port under the breathing system.
The auxiliary inlet is a convenience inlet to the air brake of active AGSS units. There is a reservoir to capture exhaust flows higher than the extract flow.
A separate exhaust hose is needed from the auxiliary manual breathing circuit to the disposal point for all AGSS units.
Sample gas from a gas monitor can be scavenged using the sample gas return port or the AGSS. To scavenge from a gas monitor using the sample gas return port, connect the tubing from the monitor to the sample gas return port. To scavenge from a gas monitor using the AGSS, connect tubing from the monitor to the male luer inlet on the bottom of the AGSS underneath the breathing system.
M1132382 2-11
Aespire View
2
3
4
5
6
789
10
11
12
13
14
AC.20.003
1

Ventilator controls

The ventilator controls include touch keys, menu screens, and a control knob (ComWheel). The System switch provides power functions to the ventilator display. The Bag/Vent switch starts and stops mechanical ventilation.
1. High priority alarm indicator 8. PEEP quick key
2. Low or medium priority alarm indicator 9. Pmax or Psupport quick key
3. Alarm silence key 10. I:E or Tinsp quick key
4. Menu key 11. Respiratory rate (RR) quick key
5. End case key 12. Tidal volume (TV) or Pinsp quick key
6. ComWheel 13. Mains power indicator
7. More settings quick key 14. Volume alarms On/Off key
Figure 2-6 • Ventilator controls
2-12 M1132382

Ventilator screen

AC.20.004
Total
TV
ml
Vent On
RR
/min
I:E Pmax
cmH2O
PEEP
cmH2O
More
Settings
VCV - Volume Control
Circle
Fresh Gas Flow
l/min
O2
Air
N2O
Push the knob to confirm the change. Turn the knob to change the setting.
Time (sec)
Paw
O2% low
Ppeak
Pmean
Ppause
O2
%
TVexp
ml
MV
l/min
RR
/min
cmH2O
TVexp low
1
2
5
6
7
8
9
10
11
4
3
2 System Controls and Menus
1. Alarm silence indicator and countdown clock 7. Total flow sensing (optional)
2. Alarm message areas 8. Ventilator settings
3. Waveform area 9. Mechanical ventilation status
4. Alarm limit settings 10. Ventilation mode
5. Measured values area 11. User message area
6. Circuit type
Figure 2-7 • Normal view
M1132382 2-13
Aespire View
TV
ml
Vent On
RR
/min
I:E Pmax
cmH2O
PEEP
cmH2O
More
Settings
VCV - Volume Control
Circle
Time (sec)
Ppeak
Pmean
Ppause
O2
%
TVexp
ml
MV
l/min
RR
/min
cmH2O
Main Menu
Ventilation Mode
Alarm Setup
Setup/Calibration
Screen and Audio Setup
Cardiac Bypass
Normal Screen
VCV
PCV
SIMV/PSV
PSVPro
SIMV-PC
AC.20.009

Using menus

Figure 2-8 • Menu example
1. Push the Menu key to show the Main Menu.
2. Turn the ComWheel counterclockwise to highlight the next menu item. Turn the ComWheel clockwise to highlight the previous menu item.
3. Push the ComWheel to enter the highlighted window or a sub menu.
4. Turn the ComWheel clockwise or counterclockwise to highlight the desired selection.
5. Push the ComWheel to confirm the selection.
6. Push the Menu key to exit the menu and return to the normal monitoring screen.
2-14 M1132382
Loading...
+ 120 hidden pages