Datex-Ohmeda Aestiva/5, Aestiva Series, 2 Vap. Trolley, 3 Vap. Trolley, Pendant Operation Manual

...
Page 1
Aestiva/5
Operation Manual - Part 1
Software Revision 4.X
System Controls, Operation, Checkout
Page 2
User Responsibility
w CAUTION U.S. Federal law restricts this device to sale by or on the order of a
licensed medical practitioner. Outside the U.S.A., check local laws for any restriction that may apply.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a product group code, the year of manufacture, and a sequential unit number for identification.
AAA F 12345
This alpha character indicates the year of product manufacture and when the serial number was assigned; “D” = 2000, “E” = 2001, “F” = 2002, etc. “I” and “O” are not used.
Aestiva, S/5, PSVPro, Tec 5, Tec 6 and Tec 7 are registered trademarks of
Datex-Ohmeda Inc. Other brand names or product names used in this manual are trademarks or
registered trademarks of their respective holders.
Page 3
i
Table of Contents
1/Introduction
i
How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
What is an Aestiva? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Ventilators and monitors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3
Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . . 1-5
2/System Controls and Menus
Anesthesia system controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Breathing system controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Vaporizer controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
3/Operation and Tutorial
Ventilator controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Optional features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
Control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-11
How to set controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-14
How to use the menu . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-15
Menu map . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-16
More about menu functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-17
How to change menu settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-18
Optional flowmeter and suction regulators . . . . . . . . . . . . . . . . . . . . . . . . .2-19
Suction regulator controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-20
External flowmeter controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-20
Optional CO
Turn On the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
Set the alarm loudness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
Bypass mode operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-21
2
1006-0938-000
Show or hide alarm limits and units. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5
Adjust patient data for Heliox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
Turn the volume alarms on or off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-8
Set alarm limits . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-9
Set an audible alarm for circuit leaks . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-11
Set Cardiac Bypass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-12
Page 4
Aestiva
Start mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
ii
Stop mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Set the ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Set ventilator controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Optional features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Ventilator controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-19
Volume Control mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
Pressure Control mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
SIMV mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20
PSVPro mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21
4/Preoperative Checklist
Set inspiratory pause (volume mode) . . . . . . . . . . . . . . . . . . . . . . . . 3-22
Set SIMV and PSVPro controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Silence alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Reading the pressure waveform (Paw) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Scales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Measure circuit compliance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
Show the service settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Optional Passive AGSS operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
Optional Active AGSS operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
Connecting Active AGSS with a flow indicator . . . . . . . . . . . . . . . . . . 3-35
Connecting Active AGSS without a flow indicator . . . . . . . . . . . . . . . 3-36
Every day before the first patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-2
Every time a different clinician uses the system . . . . . . . . . . . . . . . . . .4-3
Before every patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4-3
Appendix - Preoperative Tests
ii
Test Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-2
Every day before the first patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Inspect the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-3
Minimize alarms (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-5
1006-0938-000
Page 5
Table of Contents
Pipeline and cylinder tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-5
Flow control tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7
Vaporizer back pressure test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-8
Power failure test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9
Precase steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-11
Every time a different clinician uses the system . . . . . . . . . . . . . . . . . . . . .A-12
Low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-12
Before every patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-16
Inspect the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-16
Minimize alarms (optional) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-18
iii
Index
Warranty
Breathing system tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .A-19
Monitor and ventilator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-21
1006-0938-000
iii
Page 6
Aestiva
iv
iv
1006-0938-000
Page 7
1 Introduction
1-1
In this section How to use this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1-2
What is an Aestiva?. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
Symbols used in the manual or on the equipment . . . . . . . . . . . . . . . . . . . .1-5
1006-0938-000 1-1
Page 8
1-2
Aestiva
How to use this manual
The Aestiva comes with several user manuals. This manual describes the controls and how to use them.
Section 1 shows the different models and supplies information about the symbols used on the equipment.
Section 2 shows control locations.
Section 3 tells you how to use the controls.
Section 4 is a two-page, preoperative checklist.
The appendix provides complete instructions for the preoperative tests shown on the checklist.
The Aestiva/5 7900 can be equipped with several optional ventilation functions. References made in this manual to Heliox mode, and SIMV and PSVPro modes, are only applicable to systems equipped with these functions.
Use this manual together with Part 2, which includes setup, troubleshooting, calibration, and maintenance procedures.
What is an Aestiva?
The Aestiva is a flexible, accessible and intuitive anesthesia delivery system. A wide selection of frames, gases, and vaporizers give you full control of the system configuration.
Options include pendant mounted systems, extra gas cylinders or vaporizers, and left or right-hand breathing systems.
Model
Number of vaporizers 2 3 2
Number of gases 2 or 3 2, 3, or 4 2 or 3
Optional gases (Heliox and CO cylinder only)
Breathing system and ventilator display mounting
Total number of cylinders (maximum 2 per gas)
are
2
2 Vap. Trolley 3 Vap. Trolley Pendant
Air or Heliox; CO
Left or right side Left or right side Left or right side
Up to 4 Up to 5 Up to 2
Air, Heliox, CO2 (up to two) Air or Heliox; CO
2
2
1-2 1006-0938-000
Page 9
1 Introduction
Ventilators and monitors
The system uses a microprocessor-controlled ventilator with internal monitors, electronic PEEP, multiple modes of ventilation, and a pressure waveform dis­play. Built-in connectors and communication software permit optional cardiovascular and respiratory gas monitoring.
1-3
1006-0938-000 1-3
Page 10
Aestiva
1-4
Two vaporizer positions, left-hand configuration with basic display mount.
AA.96.025
Two vaporizer positions, left-hand configuration with arm display mount.
AA.96.016
AA.96.019
Three vaporizer positions, left-hand configuration with arm display mount.
Three vaporizer positions, right-hand configuration with arm display mount.
Figure 1-1 • Aestiva systems
AA.96.028
Pendant with arm display mount.
AA.96.102
1-4 1006-0938-000
Page 11
Symbols used in the manual or on the equipment
Warnings and wCautions tell you about dangerous conditions that can
w
occur if you do not follow all instructions in this manual.
Warnings tell about a condition that can cause injury to the operator or the patient.
Cautions tell about a condition that can cause damage to the equipment. Read and follow all warnings and cautions.
Other symbols replace words on the equipment or in Datex-Ohmeda manuals. No one device or manual uses all of the symbols. These symbols include:
1 Introduction
1-5
l
O o q p œ †
~
x
On (power)
Not autoclavable
Í
Off (power)
Type B equipment
m
Standby
Type BF equipment
µ
Standby or preparatory state for part of the equipment
“ON” only for part of the equipment
H
Type CF equipment
Caution, ISO 7000-0434
w
“OFF” only for part of the equipment
wW
Direct current
N
N
Alternating current Dangerous Voltage
Protective earth ground
Attention, refer to product instructions, IEC 601-1
This way up
Earth ground
y
Electrical input Electrical output
Pneumatic inlet Pneumatic outlet
1006-0938-000 1-5
Page 12
Aestiva
1-6
r å Y
t T
+
Frame or chassis ground
Alarm silence button
Equipotential Systems with this mark agree with
Variability Read top of float
Variability in steps Vacuum inlet
Plus, positive polarity Suction bottle outlet
REF
SN
Stock Number
Serial Number
the European Council Directive (93/42/EEC) for Medical Devices when they are used as specified in their Operation and Maintenance Manuals. The xxxx is the certification number of the Notified Body used by Datex­Ohmeda’s Quality Systems.
-
P
N ˆ
z Z
Minus, negative polarity
Lamp, lighting, illumination Cylinder
Movement in one direction Isolation transformer
Movement in two directions Linkage system
Lock Risk of Explosion
Unlock Low pressure leak test
O2+
O2 Flush button
1-6 1006-0938-000
Page 13
1 Introduction
134°C
R u q t
Autoclavable
Bag position/ manual ventilation
Mechanical ventilation
r
Open drain (remove liquid)
Close drain
U
Inspiratory flow
Expiratory flow
Q
O2 sensor connection End case
The primary regulator is set to pressure less than 345 kPa.
European Union Representative
The primary regulator is set to pressure less than 414 kPa.
1-7
1006-0938-000 1-7
Page 14
1-8
Aestiva
1-8 1006-0938-000
Page 15
2
2 System Controls
and Menus
In this section
Anesthesia system controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
Breathing system controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
Vaporizer controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9
Ventilator controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-11
Optional flowmeter and suction regulators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-19
Optional CO
Bypass mode operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2-21
2
1006-0938-000
2-1
Page 16
Aestiva
Anesthesia system controls
w
WWWWAAAARRRRNNNNIIIINNNNGG
GG
Explosion Hazard. Do not use Aestiva systems with flammable anesthetic agents.
w
WWWWAAAARRRRNNNNIIIINNNNGG
GG
Do not use antistatic breathing tubes or masks. They can cause burns if you use them near high frequency surgical equipment.
4
AA.96.054
3
2
1
5
6
7
10
12
9
11
1. Breathing system (Figure 2-3)
2. Flow controls
3. Ventilator/monitoring display (Figure 2-5)
4. Light switch and Gooseneck lamp connector (some models)
5. Light (some models)
6. Dovetail rails
7. Vaporizers (Figure 2-4)
8. Gauge (cylinder pressure)
9. Gauge (pipeline pressure)
10. System switch
11. Flush button
12. Brake
8
2-2
Figure 2-1 • Aestiva (front view)
1006-0938-000
Page 17
2
2 System Controls and Menus
Figure 2-1 shows these controls on the front of the Aestiva.
Item Description
System switch Set the switch to on to permit gas flow and to turn on the
monitoring.
On
Standby
AA.96.103
Flow controls Turn the control counterclockwise to increase the flow and
clockwise to decrease. The system switch must be on.
Decrease
to the breathing
2
AA.96.105
O
flush Push O
2
Increase
Flush to supply high flows of O
2
system.
AA.96.106
1006-0938-000
Light switch Turns the light on and off.
Off On
AA.96.107
Brake Push down to lock. Lift to release.
AA.96.100
2-3
Page 18
Aestiva
The circuit breakers are on the rear panel of the Aestiva.
Item Description (Figure 2-2)
Circuit Breakers
System circuit breaker
Open
(No Power)
AA.96.108
Closed
(Power)
Outlet circuit breakers
Open
(No Power)
Closed (Power)
3
AA.96.109
4
2-4
5
2 1
6 7
8
AA.96.053
1. Circuit Breaker for Electrical Outlet
2. Electrical Outlet
3. Circuit Breaker for Total Outlet Current
4. Circuit Breaker for Mains Inlet
5. Mains Inlet
6. Ventilator fuse
7. Pneumatic Outlet
8. Pipeline Connection
Figure 2-2 • Aestiva rear view
1006-0938-000
Page 19
2
Breathing system controls
8
2 System Controls and Menus
9
10
7
6
5a
4
3
2
1
11
12
5b
13
AB.23.003
1. Canister release
2. Auxiliary common gas outlet (optional)
3. Outlet switch (Auxiliary Common Gas Outlet)
4. Door
5. Flow sensor /patient connection (circuit connections) a.Inspiratory (Circle circuit module) or to-fro connection (Mapleson/Bain circuit module) b.Expiratory (Circle circuit module) or fresh gas connection (Mapleson/Bain circuit module)
6. Breathing circuit module (Circle)
7. Bag arm
8. Bag/Vent switch
9. APL valve
10.Bellows
11.Pressure gauge
12.Check valves
13. O
sensor
2
Figure 2-3 • Breathing system parts
1006-0938-000
2-5
Page 20
Aestiva
.
Item Description
,
Bag/Vent switch When you first turn on the system, mechanical ventilation is always off. To start mechanical
ventilation, move the switch from Bag to Vent.
Mechanical ventilation On (gas to bellows)
AB.23.059
Mechanical ventilation Off (gas to bag arm)
AB.23.060
APL valve Limits breathing system pressure during manual ventilation. The scale shows approximate
pressures. Above 30 cmH
30
70
~0 cmH2O
O, you will feel clicks as the knob turns.
2
Increase
20
5
~20 cmH2O
70
5
30
20
AB.23.064
Bag arm One arm is adjustable (push in and turn). The other is not.
Adjustable
Not adjustable
2-6
AB.23.023
AB.23.024
1006-0938-000
Page 21
2
2 System Controls and Menus
Item Description
Canister release Open to replace soda lime or remove the canisters. Note, this also opens the breathing
system.
Pull
Turn
Open
Turn
Push
Close
CO
bypass (optional) Pulling and turning the canister release opens the canisters and activates the CO
2
mode. The CO continued ventilation and rebreathing of exhaled gases.
bypass seals the breathing circuit when the canisters are open. This permits
2
AB.23.061
bypass
2
Drain plug Unscrew the plug to remove condensate.
1006-0938-000
AB.23.179
AB.23.062
2-7
Page 22
Aestiva
Item Description
Outlet switch (optional)
Sends fresh gas to the selected outlet when the system has an auxiliary common gas outlet. Monitoring and ventilation turn off when the auxiliary outlet is selected.
Breathing system selected
Auxiliary outlet selected
AB.23.063
2-8
1006-0938-000
Page 23
2
Vaporizer controls
2 System Controls and Menus
Refer to the description in this section and the vaporizer operation and maintenance manual for more detailed information on the vaporizer.
Figure 2-4 shows these controls.
Item Description
Lock lever Turn the lever fully clockwise to lock the vaporizer in
position.
AA.43.062
Concentration control and release
Alarm Silence Switch (Tec 6 only)
Push the release and turn the concentration control to set the agent concentration. The Tec 6 concentration control does not turn as long as the warm-up indicator is on.
ben
Å
1 6
14
1 2
10
9
8
7
6
5
1
4
23
D
6
c
e
T
Push to silence alarms. Hold for four seconds to start the speaker and light all indicators (alarm test).
> 4 sec = Alarm test < 4 sec = Alarm silence
AA.43.064
AA.43.063
1006-0938-000
2-9
Page 24
Aestiva
Item Description
Indicators (Tec 6 only)
All indicators come on at the start. The warm-up indicator goes off after approximately ten minutes when the operational indicator comes on. Other indicators come on during malfunctions.
5
4
3
!
Keep upright at all times
Lock
before
use
5
4
3
2
1
AA.43.015
AA.43.005
8
a
W
See manual for
r
Operational Instructions.
g
in
n
Factors affecting performance.
Service and maintena
recommendations
2
nce
e t
a d e
c i
v r e
s
t x e N
1
.4
.6
.8
234
k
re
oc
L
fo
e
b
se
u
Of
0%
.2
Tec 5
ce.
e
n
a
nc
na
e
ns
s.
t
a
erform
d
nte
tio
on
i
e c
p
i v
da
cti
g
r
es
ma
e
ru
s
en
d
t x
ctin
m
e
an
N
ll tim
Inst
l
om
affe
ice
for
ec
l
rs
r
rv
iona
e
to
ua
S
at
ac
ight at a
an
er
F
pr
m
Op
e
p u
Se
Warning
Kee
!
1
8
.
2
3
4
5
AA.13.032
e
n a
r
lu
f
o
s
I
ly
n
O
e
s
U
6
nly
rane
se O
U
nflu
E
6
Tec 6
6
7
1. Power Cable (Only the Tec 6 Vaporizer)
2. Concentration Control
3. Concentration Control Release
4. Lock lever
5. Alarm Silence Switch (Tec 6 only)
6. Filler Port Controls
7. Indicators (Tec 6 only)
8. Battery Cover (Tec 6 only)
Figure 2-4 • Vaporizer controls
2-10
1006-0938-000
Page 25
2
Ventilator controls
2 System Controls and Menus
Optional Features
Control panel
13
The Aestiva 7900 can be equipped with several optional ventilation functions. References made in this manual to Heliox mode, and SIMV and PSVPro modes, are only applicable to systems equipped with these functions.
Ventilator controls include:
• Touch keys
• Menu screens
•A control knob
12
3
AB.90.025
4
5
1006-0938-000
12
1. Alarm silence (key)
2. Alarm message (display)
3. Volume alarms On/Off (key)
4. Menu key
5. Breathing circuit module (display)
6. Control knob
7. Control setting
8. Selection key
9. Ventilation mode (display)
10. Ventilator status (On or Off)
11. Mains indicator
12. End Case (key)
13. Measured values
10
11
Figure 2-5 • SmartVent controls and monitored data
9
78
6
2-11
Page 26
Aestiva
All but two of the controls for the ventilator are located on the Ventilation/ Monitoring display. The two controls are:
• The system switch, which powers the ventilator.
• The Bag/Vent switch, which starts and stops mechanical ventilation.
Item Description
Menu key Shows the main menu.
AB.90.036
Alarm silence key and indicator
Volume alarm key and status
Silences most alarms for 120 seconds.
Pushing the key when no alarm is active pre-silences low and medium priority alarms, except Minimum Monitoring, for 90 seconds.
The "No O2 Pressure" alarm cannot be silenced.
120
Remaining
silence time
Tur ns volume alarms on and off.
Vol Alarms On Vol Alarms Off
Vol Alarms Off
AB.29.004
2-12
1006-0938-000
Page 27
2
2 System Controls and Menus
Item Description
End Case key End Case helps to prevent false alarms when no
patient is connected. It:
• Puts the apnea and volume alarms into Standby.
• Returns user selections to the most common settings: Cardiac bypass off; Alarm limits shown; and Heliox mode off.
• Sets the PEEP to 0 cmH
• Sets Plimit to one of two values: facility default or
00
444400
cmH
O.
2
• Forces the circuit Leak Audio to On.
Mechanical ventilation must be off (set the Bag/Vent switch to Bag or select the auxiliary common gas outlet).
O (default value).
2
AB.29.006
1006-0938-000
2-13
Page 28
Aestiva
How to set controls
Step 1
Push the selection key below the setting.
Step 2
Turn the knob to change the setting.
Step 3
The bottom of the screen shows control settings.
Notes:
• The ventilator will not allow the setting of values it cannot supply. A reject tone will sound or a message will appear on the screen.
• If the incorrect key is pushed, wait ten seconds or push the correct key.
• If the new setting is not saved, the ventilator continues to use the old setting.
AB.29.010
AB.29.002
Push the knob or the key to save the change.
AB.29.011
2-14
1006-0938-000
Page 29
2
2 System Controls and Menus
How to use the menu
Step 1
Push the Menu key to see the main menu.
Step 2
Turn the knob to select an option (highlight).
Screens go back to the normal display 25 seconds after the last action. During a calibration or other procedure, the screen shows the instructions.
AB29.013
AB.90.036
Step 3
Push the knob to show the next screen.
AB90.037
AB.90.045
1006-0938-000
2-15
Page 30
Aestiva
AB.90.036
Menu map
Figure 2-6 shows the menu map. The table tells you about some of the options.
SE
AB.90.041
AB.90.054
::
NNNNooootttteeee::
If the Alarm Settings page shows E Auto Limits during mechanical ventilation,
the system automatically calculates alarm limits.
Figure 2-6 • Menu map
AB.90.045
AB.90.063
2-16
1006-0938-000
Page 31
2
More about menu functions
Menu Option Function
2 System Controls and Menus
Main Cardiac Bypass
(In Progress/No)
Alarm Settings Circuit Leak
(Audio On/Off)
Setup/
SIMV/PSVPro Setup Shows additional ventilation settings for SIMV and PSVPro modes.
Calibration
2 Sensor Cal Shows menu for O
O
Inspiratory Pause Adds an inspiratory pause time to volume control breaths. Heliox mode
(On/Off) About Ventilator... Shows service level settings: software version; if facility defaults or the
Screen and Audio
Alarm Limits (Show/Hide)
Tur ns off volume and apnea alarms when these are not appropriate (e.g., during heart lung bypass).
Tur ns off the alarm tone for circuit leaks. You must set the low ◊E alarm first. Select Audio off if the circuit has a known leak (e.g., an uncuffed endotracheal tube).
sensor calibration.
2
Tells the ventilator if heliox is in use.
control settings from the previous case are used when the system is first turned on; altitude; and drive gas (O2 or Air).
’Show’ displays alarm limits next to the data on the screen.
1006-0938-000
Units of Measure (Show/Hide)
Show Limits Hide Limits
’Show’ displays units under the data on the screen.
Show Units
Hide Units
2-17
Page 32
Aestiva
How to change menu
settings
Step 1
Select the desired menu.
Step 2
Tur n, then push the knob to select an option.
This example changes alarm limits. The screen goes back to the normal display 25 seconds after the last action.
AB.90.038AB.90.054AB.90.055
Step 3
Turn the knob to change the setting.
75
82
Step 4
Push the knob to save the change.
2-18 1006-0938-000
82
82
Page 33
2
Optional flowmeter and suction regulators
These options are available:
2 System Controls and Menus
• An external O
flowmeter
2
• Continuous suction regulators (3 Mode models)
•A venturi suction regulator (2 Mode models)
AA.96.0168.f0499
AA.96.168
2
3
9
1
467*8
5
* Only non-venturi regulators have this connection.
Venturi regulators generate vacuum from system gas supplies.
1. Flowmeter outlet
2. External O2 flowmeter
3. Suction regulator
4. Suction mode switch (On/Off/MAX)
5. Vacuum gauge
6. Suction adjustment
7. Vacuum inlet (non-venturi models only)
8. Collection bottle connection
9. Overflow safety trap
Figure 2-7 • Controls and connectors for flowmeters and suction regulator
1006-0938-000 2-19
Page 34
Aestiva
Suction regulator
controls
Item Description
Mode switch: With a continuous suction regulator (3 Modes):
• MAX: for maximum suction, set the switch to MAX.
• On (|): for adjustable suction, set the switch to On (|) or l.
• Off (O): to turn Off suction, set the switch to O.
With a venturi regulator (2 Modes):
• On (|): for adjustable suction, set the switch to On (|) or l.
AA.96.169
• Off (O): to turn Off suction, set the switch to O.
External flowmeter
controls
With the mode switch set to “llll”, turn the control clockwise to increase suction and counterclockwise to decrease it.
Increase
AA.96.170
Decrease
Turn the control counterclockwise to increase the O flow and clockwise to decrease.
AA.96.171
Increase
Decrease
2
2-20 1006-0938-000
Page 35
2
Optional CO2 Bypass mode operation
Pulling and turning the canister release opens the canisters and activates the optional CO2 bypass mode. The CO2 bypass seals the breathing circuit when the canisters are open. This permits continued ventilation and rebreathing of exhaled gases.
When the canisters are open, the medium priority alarm message “No CO2 Absorption” is displayed. The alarm is downgraded to low priority after silenc­ing.
Close the absorber canisters to activate the Absorber mode. Closing the can­isters directs exhaled gas flow through the absorber, removing CO2. Do the breathing system tests as soon as possible after returning to Absorber mode.
When operating in CO2 Bypass mode, water condensate may collect in the bypass assembly. The water drains automatically when the canisters are closed.
Datex-Ohmeda strongly recommends using CO2 monitoring when using the CO2 Bypass device.
2 System Controls and Menus
Systems which have the optional CO2 Bypass mode installed have the follow­ing label at the canister release.
AB.23.179
1006-0938-000 2-21
Page 36
Aestiva
2-22 1006-0938-000
Page 37
3

3 Operation and Tutorial

w WWWWAAAARRRRNNNNIIIINNNNGG
In this section
GG
In addition to volume apnea and low airway pressure alarms, other ventilator
alarms are included to indicate potential hazard conditions. All alarms that occur should be investigated to ensure adequate patient safety.
This section describes specific tasks. Use it as a step-by-step guide or a training tool.
Turn On the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-2
Set the alarm loudness . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-3
Show or hide alarm limits and units . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-5
Adjust patient data for Heliox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-7
Turn the volume alarms on or off . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-8
Set alarm limits. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3-9
Set an audible alarm for circuit leaks . . . . . . . . . . . . . . . . . . . . . . . . . . 3-11
Set Cardiac Bypass . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-12
Start mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14
Stop mechanical ventilation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15
Set the ventilation mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-16
Set ventilator controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18
Set SIMV and PSVPro controls. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-24
Silence alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-26
Reading the pressure waveform (Paw) . . . . . . . . . . . . . . . . . . . . . . . . . 3-27
Measure circuit compliance. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30
Show the service settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-31
Optional Passive AGSS operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-33
Optional Active AGSS operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-34
1006-0938-000
3-1
Page 38
Aestiva
Turn On the system
Step 1
Connect the power cord to a wall outlet. The mains indicator comes on when AC Power is connected.
Step 2
Set the system switch to On (|).
AB.29.007
On
Step 3
The display shows the power-up screen, and the system does a series of self tests.
Step 4
When the self tests pass, the display shows the normal screen.
AA.96.104
AB.90.018
If a test fails, the screen shows an alarm. Refer to the troubleshooting section.
3-2
AB.90.021
1006-0938-000
Page 39
Set the alarm loudness
Step 1
Select the Screen and Audio Setup menu.
• Push the menu key.
• Turn, then push the knob to select Screen and Audio.
AB.29.013
AB.90.036
3 Operation and Tutorial
Step 2
Turn, then push the knob to select alarm loudness.
AB.90.039
AB.90.063
AB.90.043
1006-0938-000
3-3
Page 40
Aestiva
Step 3
Turn, then push the knob to adjust the volume.
• The volume range is 1 to 5 (loudest).
• As the volume is changed, the system sounds test tones.
• Push the knob to save the change.
2
Then
3-4
1006-0938-000
Page 41
Show or hide alarm limits and units
To simplify the displays, hide alarm limits and units of measurement. If the alarm limits are hidden, the screen automatically shows the limits when:
• An alarm occurs.
• Volume alarms are off or the auxiliary common gas outlet is selected (monitoring off).
• An individual alarm limit is set to Off.
When the system is set to Standby, alarm limits go back to Show.
3 Operation and Tutorial
Step 1
Select the Screen and Audio Setup menu.
• Push the menu key.
• Turn, then push the knob to select Screen and Audio Setup.
Show All
AB90.036
Hide Units Hide Alarm Limits
AB.90.033
1006-0938-000
AB90.039
3-5
Page 42
Aestiva
Step 2
Turn, then push the knob to select Alarm Limits or Units of Measure.
AB90.063
AB90.044
Step 3
Turn, then push the knob to select Show or Hide. Push the knob to save the change.
Show
Hide
Then
3-6
1006-0938-000
Page 43
Adjust patient data for Heliox
The Aestiva 7900 can be equipped with several optional ventilation functions. References made in this manual to Heliox mode, and SIMV and PSVPro modes, are only applicable to systems equipped with these functions.
When the Heliox mode is selected, the system automatically corrects measurements for the lower density of Heliox (compared to air).
GG
WWWWAAAARRRRNNNNIIIINNNNGG
w
Step 1
Select the Setup/Calibration menu.
• Push the menu key.
• Turn, then push the knob to select Setup/Calibration.
The Heliox mode must be set correctly for accurate volume monitoring and
delivery.
3 Operation and Tutorial
AB.29.013
AB.90.036
AB.90.037
1006-0938-000
3-7
Page 44
Aestiva
Step 2
Turn, then push the knob to select Heliox Mode.
AB.90.045
AB.90.046
Step 3
Turn, then push the knob to select on or off. You must push the knob to save the change
On
Turn the volume alarms on or off
w
WWWWAAAARRRRNNNNIIIINNNNGG
GG
Do not turn volume alarms off with a spontaneously breathing patient, the
system will not alarm for low volume.
The volume alarm key ( alarms are off, a large X covers the limits.
Use this control to prevent false alarms if you switch to manual ventilation at lower tidal volumes.
Vol Alarms On
Then
E
/V
TE
) turns volume alarms on and off. When the
AB.29.014
3-8
Use the End Case key (on control panel) to prevent apnea alarms between patients.
1006-0938-000
Page 45
Set alarm limits
AB
90
036
AB
90
038
AB
90
055
3 Operation and Tutorial
Step 1
Select the Alarm Settings menu.
• Push the menu key.
• Turn, then push the knob to select Alarm Settings.
Note: If the Alarm Settings page shows
E
Auto Limits during mechanical
ventilation, the system is set to automatically calculate
AB.29.013
. .
. .
E
limits.
Step 2
Turn, then push the knob to select a limit.
AB.90.054
. .
1006-0938-000
3-9
Page 46
Aestiva
Step 3
Turn, then push the knob to change the limit.
• Push the knob to save the change.
• The screen returns to the normal display 25 seconds after the last change.
75
TTTThhhheeeenn
nn
3-10
1006-0938-000
Page 47
3
Set an audible alarm for circuit leaks
The patient circuit leak alarm is activated if less than half of the inspired volume returns through the expiratory flow sensor during mechanical ventilation. To prevent nuisance alarms from a known leak (e.g., an un-cuffed endotracheal tube), set the audio to Off. Normal volume and apnea monitoring does not change.
::
NNNNooootttteeee::
This alarm is the first stage in detecting a circuit disconnect. The audible leak alarm cannot be turned off unless volume alarms are on, and the low limit is set to a value other than off.
3 Operation and Tutorial
E
Step 1
Select the Alarm Settings menu.
• Push the menu key.
• Turn, then push the knob to select Alarm Settings.
AB.29.013
AB.90.036
AB.90.038
1006-0938-000
3-11
Page 48
Aestiva
Step 2
Turn, then push the knob to select circuit leak audio.
AB.90.054AB.90.056
Step 3
Turn, then push the knob to change the setting. You must push the knob to save the change.
Set Cardiac Bypass
Audio On
Audio Off
Set Cardiac Bypass to In Progress to prevent volume and apnea alarms when the patient is on cardio-pulmonary bypass.
When Cardiac Bypass In Progress is selected, the display shows:
• Cardiac Bypass
• Apnea Alarm Off
• Vol Alarms Off ::
Note ::
The mechanical ventilation must be off. When the mechanical ventilation is turned back on, Cardiac Bypass returns to the No Bypass setting and alarms become active.
Then
3-12
1006-0938-000
Page 49
3
Step 1
Select the Cardiac Bypass menu item:
• Push the menu key.
• Turn, then push the knob to select Cardiac Bypass In Progress.
AB.90.036
AB.29.013
3 Operation and Tutorial
Step 2
Push the knob again to return to No.
AB.90.040
AB.90.068AB.90.040
1006-0938-000
3-13
Page 50
Aestiva
Start mechanical ventilation
w
WWWWAAAARRRRNNNNIIIINNNNGG
GG
Make sure the patient circuit is correctly assembled and the control
settings are correct before you start or stop ventilation.
Step 1
Make sure the control settings are OK.
Step 2
Turn off the auxiliary common gas outlet (some models).
Mechanical ventilation is off when you first turn on the system.
OK?
AB.90.025
AB.23.065
Step 3
Set the Bag/Vent switch back to Vent.
• If mechanical ventilation is not available, a message tells you what to do. For example “To start mech vent set the Bag/Vent switch to Bag and back to Vent.”
AA.96.110
3-14
1006-0938-000
Page 51
3
Stop mechanical ventilation
3 Operation and Tutorial
w WWWWAAAARRRRNNNNIIIINNNNGG
GG
Make sure the patient circuit is correctly assembled and the control
settings are correct before you start or stop ventilation.
Step 1
Make sure control settings are OK.
Step 2
Set the Bag/Vent switch to Bag.
OK?
70
5
30
20
AB.23.066
1006-0938-000
AA.96.111
3-15
Page 52
Aestiva
Set the ventilation mode
Step 1
Text below the waveform shows the ventilation mode:
• Pressure controlled modes supply a set pressure during inspiration.
• Volume controlled modes supply a set tidal volume.
AB.90.092
Select the Ventilation Mode.
• Push the menu key.
• Push the knob to select Ventilation Mode.
Step 2
Turn, then push the knob to change the mode. Push the knob to save the change.
AB.29.013
AB.90.036
Volume Control
Pressure Cntrl
SIMV Mode
Then
PSVPro Mode
Note PSVPro is pressure supported ventilation with apnea backup.
3-16 1006-0938-000
Page 53
3
Step 3
Set the highlighted control parameter. Each mode has one parameter that must be set (V
Pinspired for Pressure, and Psupport for PSVPro).
• Turn, then push the knob to set the value.
• Until a value is set, the ventilator shows “---”. If any other key is pressed at this time, a tone will sound.
T
for Volume and SIMV,
AB.90.074
AB.90.075
TTTThhhheeeenn
3 Operation and Tutorial
nn
1006-0938-000 3-17
Page 54
Aestiva
Set ventilator controls
Optional features
Step 1
Push the selection key.
Step 2
A tone sounds and a box flashes around the setting.
The Aestiva 7900 can be equipped with several optional ventilation functions. References made in this manual to Heliox mode, and SIMV and PSVPro modes, are only applicable to systems equipped with these functions.
The ventilator controls present are based on the ventilation mode.
Messages appear on the screen if:
• You try to set a value the system cannot supply.
• You change a setting but do not save it: "Push knob to confirm change Turn knob to change setting".
AB.90.076
AB.90.077
Step 3
Turn the knob to set the value.
AB.90.078
Step 4
Push the knob to save the setting.
•A tone sounds.
• The flashing stops.
3-18 1006-0938-000
AB.90.079
Page 55
3 Operation and Tutorial
3
Ventilator controls
Flow Trigger Level This parameter sets the minimum flow detected by the ventilator which triggers
the ventilator to deliver a mechanical breath to a spontaneously breathing patient. Only active in SIMV and PSVPro modes.
I:E This control sets the inspiratory to expiratory ratio of mechanical breaths
supplied to the patient.
Insp. Termination Level This parameter sets the percentage of the peak inspiratory flow where the
ventilator stops a pressure supported breath. Only active in SIMV and PSVPro modes.
PEEP This control sets the positive end expiratory pressure. This is only available
during mechanical ventilation, but the control can be set at any time.
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
Do not use a separate mechanical PEEP valve; incorrect operation and
patient injury can result.
Pinspired This control sets the amount of pressure delivered to the patient in each
pressure controlled breath.
Plimit This control sets the maximum (and sustained) airway pressures tolerated in
the patient’s breathing system.
• If the high airway pressure limit is reached, inspiration stops and exhalation starts.
• The limit is an absolute value. There is no offset for PEEP pressure.
Note: Pmax is the peak sensed airway pressure; Plimit is the airway pressure limit set with front panel controls.
Psupport This control sets the delivered pressure during pressure support ventilation.
Only active in SIMV and PSVPro modes.
Rate This control permits you to set the frequency of mechanical breaths delivered
to the patient. It also establishes the apnea delay time in the PSVPro mode.
Tinspired This control sets the time in seconds for each timed inspiration. Only active in
SIMV and PSVPro modes.
1006-0938-000 3-19
Page 56
Aestiva
Trigger Window This control sets the range as a percent of the exhalation phase within which the
patient may trigger the next mechanical breath. Only active in SIMV and PSVPro modes.
VT This control sets the tidal volume delivered to the patient in the Volume Control
and SIMV modes.
Volume Control mode
Pressure Control mode
The figure and table below show Volume Control settings.
AB.90.025
Settings VT Rate I:E Plimit PEEP Range 20 - 1500 4 - 100 2:1 - 1:8 12 - 100 Off, 4 - 30 Increments varies
Increments of 1mL from 20 to 50, 5 mL from 50 to 100, 10 mL from 100 to 300, 25 mL from 300 to
*
1000, 50 mL from 1000 to 1500
1 bpm 0.5 1 cmH2O 1 cmH2O
*
The figure and table below show Pressure Control settings.
AB.90.064
Settings Pinspired Rate I:E Plimit PEEP Range 5 - 60 4 -1 00 2:1 - 1:8 12 - 100 Off, 4 - 30 Increments 1 cmH2O 1 bpm 0.5 1 cmH2O 1 cmH2O
SIMV mode
3-20 1006-0938-000
The figure and table below show SIMV settings.
AB.90.071
Page 57
3 Operation and Tutorial
3
PSVPro mode
Settings V
T Rate Tinspired Psupport PEEP
Range 20 -1500 2 - 60 0.2 - 5.0 Off, 2-40 Off, 4 - 30 Increments varies
Increments of 1mL from 20 to 50, 5 mL from 50 to 100, 10 mL from 100 to 300, 25 mL from 300 to
*
1000, 50 mL from 1000 to 1500
1 bpm 0.1 s 1 cmH2O 1 cmH2O
*
PSVPro is pressure supported ventilation with apnea backup. The figure below shows PSVPro settings. In this mode the Pinspired, Rate, and
Tinspired parameters are not active, but they may be changed.
AB.90.087
During backup ventilation, the ventilator will ventilate the patient using the SIMV-PC + PSV mode. All parameters shown are active in this mode.
AB.90.086
Settings Pinspired Rate Tinspired Psupport PEEP Range 5 - 60 2 - 60 0.2 - 5.0 Off, 2-40 Off, 4 - 30 Increments 1 cmH2O 1 bpm 0.1 s 1 cmH2O 1 cmH2O
1006-0938-000 3-21
Page 58
Aestiva
Set inspiratory pause
(volume modes)
You can only use inspiratory pause in Volume Control or SIMV modes. In pressure modes, Inspiratory Pause displays a message “No Pause w”.
When Pause is on, the inspiratory volume stays in the patient’s lungs for the set pause time at the end of inspiration.
Pause can be set from Off to 60 percent of inspiratory time in increments of five percent. Minimum pause time is 400 ms.
Pause
AB.90.092
Exp.Insp.
Step 1
Select the Setup/Calibration menu.
• Push the menu key.
• Turn, then push the knob to select the Setup/Calibration menu.
AB.29.013AB.90.036
AB.90.037
3-22 1006-0938-000
Page 59
3
Step 2
Tur n, then push the knob to select Inspiratory Pause.
Step 3
Tur n, then push the knob to change the setting. You must push the knob to save the change.
AB.90.083
20% of TI
3 Operation and Tutorial
Then
1006-0938-000 3-23
Page 60
Aestiva
Set SIMV and PSVPro controls
The SIMV and PSVPro modes allow the user to set additional ventilator controls. The Pinspired, Rate, Tinspired, Psupport and PEEP controls can be set using the selection keys. The Plimit, Trigger Window, Flow Trigger Level, and Inspiratory Termination Level may be set through the Setup/Calibration menu.
Step 1
Select the SIMV/PSVPro Setup menu.
• Push the menu key.
• Turn, then push the knob to select Setup/Calibration.
Step 2
Push the knob to select SIMV/PSVPro Setup.
AB.29.013AB.90.036
AB.90.037
AB.90.045
wWARNING Most anesthetic agents will cause patients to have reduced ventilatory
responses to carbon dioxide and to hypoxemia. Therefore, triggered modes of ventilation may not produce adequate ventilation.
wWARNING The use of neuromuscular blocking agents will reduce the patient’s
breathing response, which will interfere with triggering.
3-24 1006-0938-000
Page 61
3
Step 3
Tur n, then push the knob to select a setting.
Step 4
3 Operation and Tutorial
AB.90.073
Tur n, then push the knob to change the setting. You must push the knob to save the change.
40
Then
1006-0938-000 3-25
Page 62
Aestiva
Silence alarms
The alarm silence key silences current alarms for 120 seconds. When no alarm is active, holding down the alarm silence key for one second pre-silences low or medium priority alarms for 90 seconds. Minimum monitoring cannot be pre-silenced.
• The screen shows the time remaining in the silence period.
• Pushing the alarm silence key while an alarm is silenced will reset the countdown time to 120 seconds.
• High priority alarms always cause an audible tone and must be silenced individually.
120
AB.29.004
Alarm tones identify the alarm priority:
• High Priority: 10 tones, 10 seconds pause, (repeat)....
• Medium Priority: 3 tones, 25 seconds pause, (repeat)....
• Low Priority: single tone.
Alarms appear at the top of the screen. The highest priority alarm will be shown in Area 1, the next highest priority alarm in area 2. If all areas are used, the lowest priority alarms cycle in area 4.
AB.90.072
Note: Error mode messages may appear. Refer to ’Alarms and Troubleshooting’ in Part 2 of the Operation Manual for more information.
• Minimum Monitoring: Monitoring data is available but a failure prevents mechanical ventilation.
• Minimum System Shutdown: Monitoring and ventilation are not possible.
3-26 1006-0938-000
Page 63
3
Reading the pressure waveform (Paw)
Different points on the waveform are instantaneous values for measured pressures. The horizontal axis indicates the time scale for the rate, I:E ratio, and inspiratory pause (volume control setting). The vertical axis indicates the pressure.
3 Operation and Tutorial
Volume Control Mode
Scales
Pmean
Pmax
AB.90.060
PEEP
Pressure Control Mode
Pinsp
Pmax
AB.90.057
The display automatically adjusts time and pressure scales to fit the control settings.
The time scale changes with the set Rate:
• 25 or less breaths per minute - time scale is 0 to 16 seconds
• 26 to 75 breaths per minutes - time scale is 0 to 8 seconds
• 76 or more breaths per minutes - time scale is 0 to 4 seconds
• On a change, existing pressure data is erased and new waveform data starts at time = 0
PEEP
The pressure scale changes with the Plimit setting:
• 12 to 40 Plimit, y-axis range is -5 to 40
• 41 to 60 Plimit, y-axis range is -5 to 60
• 61 to 100 Plimit, y-axis range is -5 to 100
• When the pressure scale changes, existing pressure data is erased and new waveform data starts at time = 0
1006-0938-000 3-27
Page 64
Aestiva
Example Volume Control • PEEP: Off
• Maximum sensed inspiratory pressure (Pmax): 25
• Mean positive airway pressure (Pmean): 11
AB.90.004
Figure 3-1 • Paw waveform in Volume Control mode
Example Pressure Control • Pmax: 34
• Upper pressure limit (Plimit): 40
• PEEP: 10
Figure 3-2 • Paw waveform in Pressure Control mode
AB.90.002
3-28 1006-0938-000
Page 65
3
Example SIMV • Plateau pressure (Ppl): 15
• Inspiratory time (Tinspired): 1.5
• Inspiratory Pause: 60
Figure 3-3 • Paw waveform in SIMV mode.
3 Operation and Tutorial
AB.90.028AB.90.087
Example PSVPro • PEEP: 5
• Pressure support: 10
• Mean positive airway pressure (Pmean): 7
Figure 3-4 • Paw waveform in PSVPro mode.
1006-0938-000 3-29
Page 66
Aestiva
Measure circuit compliance
To measure compressible volume in patient tubes:
1. Set the ventilator to volume control mode.
2. Set a tidal volume (V
3. Set a rate of 10 breaths/min.
4. Set an I:E ratio of 1: 1
5. Set the Plimit control to 30 cmH
6. Occlude the patient connection of the Y piece. Do not contaminate a clean patient connection.
7. Turn on mechanical ventilation.
8. Monitor the exhaled tidal volume V pressure).
The V pressure.
The example shows how tubing compliance factor can be calculated. V
/(P
TE
Example: P
= 21 cmH2O
max
V
= 24 mL
TE
24/(21–2.5) = 1.3 mL/cmH This factor can be used to calculate the approximate gas compression in
patient tubes. For example, if the patient is requiring 30 cmH
30 X 1.3 = 39 mL of gas is compressed in the tubes each breath. This gas (39 mL) is part of the set tidal volume but it does not reach the patient.
1
Force of the Bellows.
) of 500 mL
T
O.
2
and P
TE
measures the gas needed to fill the patient circuit at the measured
TE
–2.51 cmH2O) = Compliance factor in mL per cmH2O
max
O
2
(measured peak airway
max
O to ventilate,
2
3-30 1006-0938-000
Page 67
3
Show the service settings
The About Ventilator screen shows ventilator settings that can only be changed by an approved service representative.
3 Operation and Tutorial
Item Use
Software Version If you call for service, a representative may ask for
this.
Step 1
Select the Setup/Calibration menu.
• Push the menu key.
• Turn, then push the knob to select Setup/Calibration.
Facility defaults or last settings
Tells you if the system saves the current settings when you turn it off or goes back to facility defaults.
Altitude Used for gas calculations. If the altitude is not
Drive gas (O
correct, O
or Air) Tells you which gas the ventilator uses to drive the
2
calibration can fail.
2
bellows. This gas comes from the same supply (pipeline or cylinder) that the anesthesia machine uses. If this gas comes from a cylinder, the cylinder empties faster than you would expect from the flowmeter settings.
AB.29.013
AB.90.036
AB.90.037
1006-0938-000 3-31
Page 68
Aestiva
Step 2
Turn, then push the knob to select About Ventilator.
AB.90.045
AB.90.047
AB.90.051
3-32 1006-0938-000
Page 69
3
Optional Passive AGSS operation
a
AB.23.155
The Aestiva/5 7100 Passive AGSS consists of a two-liter reservoir to minimize rapid pressure changes. It contains both positive and negative pressure relief valves to protect the breathing system. The outlet is a 30 mm male taper swivel connector (a) at the rear of the breathing system.
The connector also has a capped hose barb that may be used for scavenging the sample from a gas monitor.
Passive AGSS (Anesthesia Gas Scavenging System) is intended primarily for use in operating room environments which have no dedicated vacuum system for waste gas disposal. The disposal system generally consists of large diameter tubing and/or duct directly linking the passive AGSS with the building exterior. The tubing should be as large in diameter and as short as needed for the particular application.
3 Operation and Tutorial
Passive AGSS may also be used with a non-recirculating ventilation system for waste gas disposal. The tubing connection from passive AGSS to the non­recirculating ventilation system should be an open connection, essentially at atmospheric pressure, such as to an exhaust grill.
Passive AGSS may also be used as a protective interface to an externally mounted active AGSS such as the Datex-Ohmeda adjustable Waste Gas Scavenging Interface Valve Assembly.
1006-0938-000 3-33
Page 70
Aestiva
Optional Active AGSS operation
Versions There are at least three versions of the optional active AGSS (Anesthesia Gas
Scavenging System) available for Aestiva depending on the hospital’s type of waste gas disposal system.
The low flow system is for use with high vacuum disposal systems. It requires a minimum vacuum pressure at the wall of 300 mmHg (12 inHg). The extract flow is restricted to a nominal of 36 L/min or slightly higher at higher vacuum pressures. A flow indicator on the side of the breathing system indicates when the unit is in operation.
The high flow system is for use with low vacuum (blower type) disposal systems with nominal extract flow in the range of 50 to 80 L/min. A flow indicator on the side of the breathing system indicates when the unit is in operation.
The third type is for hospital supplied venturi/ejector systems having their own extract flowmeter. This will operate satisfactorily with a minimum extract flow of 30 L/min to 100 L/min, but there is no built-in flow indicator.
All three versions are constant extract flow, air brake designs with a two-liter reserve volume to capture peak exhaust flows that briefly exceed the extract flow. The disposal system normally entrains room air through the air brake, located underneath the breathing system and AGSS reservoir, but will spill from this port during extended periods of high exhaust flow.
Datex-Ohmeda AGSR assembly The Datex-Ohmeda adjustable Anesthesia Gas Scavenging Receiver (AGSR)
assembly may be externally mounted on the dovetail rail behind the breathing system, or on the back of the articulating arm. The AGSR is an active AGSS with adjustable extract flow. It may be useful with a low flow, high vacuum disposal system without the capacity for a constant 36 L/min extract flow. Refer to the instructions provided with the device. It is for use only with the Aestiva/5 internal passive AGSS.
3-34 1006-0938-000
Page 71
3 Operation and Tutorial
3
Connecting Active AGSS
with a flow indicator
To use the optional active AGSS on a system which has a flow indicator (on either the left or right side of the breathing system), connect it as follows.
Step 1
Connect the proper hose to the AGSS outlet connector (f) on the rear of the base of the breathing system.
Step 2
Attach the other end to the hospital disposal system. Note: To scavenge gas from a gas monitor, connect
tubing from the monitor to the 3.18 mm (1/8 inch) hose barb (g) in the breathing system base.
Step 3
AB.23.174, AB.23.165
b
a
c
d
AB.23.162
AB.23.155
e
With the AGSS operating, verify that the flow indicator ball (d) on the flow indicator (a) rises to the green zone (c), indicating adequate flow.
Note: The ball in the upper red zone (b) indicates excessively high extraction flow or a blocked filter. The ball in the lower red zone (e) indicates extraction flow rate is too low.
Step 4
Before you use the system, complete the Preoperative Test procedure. Refer to the Appendix "Preoperative Tests".
g
f
1006-0938-000 3-35
Page 72
Aestiva
Connecting Active AGSS
without a flow indicator
The active AGSS option without a flow indicator is for use only with hospital disposal systems having their own visual indicator of disposal flow rate.
The recommended nominal disposal system flow rate is 36 L/min but will operate satisfactorily between 30 and 100 L/min.
To use the optional active AGSS installed on the system which does not have a flow indicator, connect it as follows.
Step 1
Connect a nominal 1/2 inch ID hose to the AGSS outlet hose connector (a) on the rear of the breathing system base.
• The hose should be flexible and reinforced to help prevent kinking and crushing.
Step 2
Attach the other end of the hose to the hospital disposal system.
a
AB.23.155
Step 3
Before you use the system, complete the Preoperative Test procedure. Refer to the Appendix "Preoperative Tests".
3-36 1006-0938-000
Page 73
-

4 Preoperative Checklist

In this section
w WWWWAAAARRRRNNNNIIIINNNNGG
w WWWWAAAARRRRNNNNIIIINNNNGG
This section is a checklist of the necessary preoperative tests under different conditions. For step-by-step instructions, refer to the appendix “Preoperative Tests.”
GG
Do not use this system unless you have read each component’s operation
and maintenance manual and understand:
• The system connections
• The warnings and cautions
• How to use each system component
• How to test each system component
GG
Before you use this system:
• Complete the tests in this section
• Test all other system components
4
1006-0938-000
w WWWWAAAARRRRNNNNIIIINNNNGG
GG
If a test failure occurs, do not use the system. Have an approved service
representative repair the system.
Every day before the first patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
Every time a different clinician uses the system . . . . . . . . . . . . . . . . . . . . . . 4-3
Before every patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
4-1
Page 74
Aestiva
Every day
before the first patient
Inspect the system. Look for damage, necessary drugs and equipment, correct breathing circuit setup, and hazardous conditions.
Turn on the system.
Set the ventilator controls to decrease alarms.
Do the pipeline and cylinder tests. Look for sufficient pressures and no
high pressure leaks (cylinders).
Do the flow control tests:
•Minimum flows: O
•Link system: Increase N drive down N
•O
supply failure alarm. Alarm operates when O
2
below set limit. Air flow continues. All other gases stop.
Do the vaporizer back pressure tests:
•Set the O
•Turn On one vaporizer at a time.
•Make sure that the O
Do a low-pressure leak test.
Do the alarm tests:
•Make sure all monitors operate correctly.
•Make sure the O 21% O
•Make sure these ventilator alarms operate correctly: high and low O low minute volume; high airway pressure; apnea and low airway pressure; sustained airway pressure.
flow to 6 L/min.
2
in room air and 100% O
2
25-75 mL/min, all other gases no flow.
2
O flow to drive up O
2
O flow. The O
2
sensor operates correctly. It shows approximately
2
2
flow stays above 5 L/min.
2
≥ nominal 25%.
flow is
after two min in pure O
2
flow. Decrease O
2
pressure is decreased
2
flow to
2
.
2
;
2
4-2
1006-0938-000
Page 75
4
-
Every time a different
clinician uses the system
Before every patient
Do a low-pressure leak test.
Look for damage, necessary drugs and equipment, correct breathing
circuit setup, and hazardous conditions.
Check vaporizer installation:
•Make sure the top of each vaporizer is horizontal (not on crooked).
•Make sure each vaporizer is locked and cannot be removed.
•Make sure the alarms and indicators operate correctly (Tec 6 vaporizer).
•Make sure you cannot turn on more than one vaporizer at the same time.
Do the breathing system tests:
•Make sure the one way valves and auxiliary equipment (humidifier, etc.)
operate correctly.
•With a circle breathing-circuit module, push the drain button for
to remove condensate.
•Ventilator circuit leak test.
•Bag/Manual circuit leak test.
• Bag/Manual circuit APL valve test.
•Circuit leak test.
Set the appropriate controls and alarm limits for the case.
Preoperative Checklist
4
≥ 10 sec
1006-0938-000
4-3
Page 76
Aestiva
4-4
1006-0938-000
Page 77

Appendix - Preoperative Tests

In this section
Test Intervals . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2
Every day before the first patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-3
Precase steps. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-11
Every time a different clinician uses the system . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
Low-pressure leak test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12
Before every patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-16
Breathing system tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-19
Monitor and ventilator tests . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-21
1006-0938-000
A-1
Page 78
Aestiva
Test Intervals
The preoperative tests are done at one of three intervals:
1. Every day before the first patient
2. Every time a different clinician uses the system
3. Before each patient
w
WWWWAAAARRRRNNNNIIIINNNNGG
GG
Do not use this system unless you have read each component’s
operation and maintenance manual and understand:
• All system connections
• All of the warnings and cautions
• How to use each system component
• How to test each system component
Before you use this system:
• Complete all of the tests in this section
• Test all other system components
If a test failure occurs, do not use the equipment. Have an
approved service representative repair the equipment.
A-2
1006-0938-000
Page 79
Every day before the first patient
Inspect the System
Appendix - Preoperative Tests
w
w
WWWWAAAARRRRNNNNIIIINNNNGG
CCCCAAAAUUUUTTTTIIIIOOOONN
GG
Make sure that the breathing circuit is correctly connected and not damaged.
NN
The total weight on each accessory shelf must be less than 23 kg.
Systems without accessory shelves have a weight limit on the top surface of 23 kg.
Make sure that:
1. The equipment is not damaged.
2. All components are correctly attached.
3. The breathing circuit is correctly connected, not damaged, and contains sufficient absorbent.
4. The vaporizers are locked in position and contain sufficient agent.
5. Pipeline gas supplies are connected and the pressures are correct.
6. Cylinder valves are closed on models with cylinder supplies.
1006-0938-000
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGG
GG
Do not leave gas cylinder valves open if the pipeline supply is in use. Cylinder supplies could be depleted, leaving an insufficient reserve supply in case of pipeline failure.
7. Models with cylinder supplies have a cylinder wrench attached to the system.
8. The necessary emergency equipment is available and in good condition.
9. Equipment for airway maintenance, tracheal intubation, and IV administration is available and in good condition.
10. Applicable anesthetic and emergency drugs are available.
11. With the optional O On and make sure that:
•The O
•The suction regulator provides adequate suction.
flowmeter provides sufficient flow.
2
Note: system gas supplies must be connected. Non-venturi suction
regulators must be connected to a vacuum source.
flowmeter and suction regulators, turn the devices
2
A-3
Page 80
Aestiva
12. On trolley model, make sure the casters are not loose and the brake is set and prevents movement.
AA.96.100AB.29.007
13. Connect the power cord to a wall outlet. The mains indicator comes On when AC Power is connected.
•If the indicator is not on, the system does not have mains (electrical) power. Use a different outlet. Close the circuit breaker or replace or connect the power cable. Refer to Figure 2-2.
14. Set the system switch to On.
AA.96.104
A-4
1006-0938-000
Page 81
Appendix - Preoperative Tests
Minimize alarms
(optional)
Pipeline and cylinder
tests
w
CCCCAAAAUUUUTTTTIIIIOOOONN
Set the ventilator controls to decrease the number of alarms:
1. Control Keys:
•Volume alarms: Off
•Plimit: 100 cmH
O
2
2. Alarm menu:
•Low O
•High O
: 21%
2
: Off
2
•Bag/Vent switch: Bag
NN
To prevent damage:
• Open the cylinder valves slowly.
• Do not force the flow controls.
If your system does not use cylinder supplies, do not do steps 2 and 3.
1. Disconnect the pipeline supplies and close all cylinder valves. If the pipeline and the cylinder pressure gauges are not at zero:
•Connect an O
supply.
2
•Turn On the system, if it is not already on.
•Set the flow controls to mid range.
•Make sure that all gauges but O
•Disconnect the O
•Make sure that the O alarms for low O
supply.
2
gauge goes to zero. As pressure decreases,
2
supply pressure should occur.
2
go to zero.
2
2. Make sure that the cylinders are full:
•Open each cylinder valve.
•Make sure that each cylinder has sufficient pressure. If not, close the applicable cylinder valve and install a full cylinder.
3. Test one cylinder at a time for high-pressure leaks:
•Set the system switch to Standby, which stops the O
flow.
2
•Disconnect all accessories from the pneumatic outlets.
•Turn Off the auxiliary O
flowmeter.
2
•Open the cylinder.
•Record the cylinder pressure.
1006-0938-000
A-5
Page 82
Aestiva
•Close the cylinder valve.
•Record the cylinder pressure after one minute. If the pressure decreases more than 690 kPa (100 psig) there is a leak:
•Install a new cylinder gasket and tighten the tee handle as shown in the Setup section of the Setup, maintenance and troubleshooting manual.
•Do this step again. If the leak continues, do not use the system.
•Repeat step 3 for all cylinders.
•Close cylinder valves.
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGG
GG
Do not leave gas cylinder valves open if the pipeline supply is in
use. Cylinder supplies could be depleted, leaving an insufficient reserve supply in case of pipeline failure.
4. Connect the pipeline supplies.
5. Turn on the system.
6. Use the chart below to check pipeline pressure:
ANSI (USA and Intl.), Australian, Canadian, French, Japanese
ISO, Italian, Scandinavian, South African, Spanish, Swiss
Austrian, German 500 kPa (75 psig)
7. Connect equipment to the pneumatic outlets as appropriate.
345 kPa (50 psig)
414 kPa (60 psig)
A-6
1006-0938-000
Page 83
Flow control tests
WWWWAAAARRRRNNNNIIIINNNNGG
w
Appendix - Preoperative Tests
monitor. Sufficient O
2
GG
The Link system cannot replace an O fresh gas may not prevent hypoxic mixtures in the breathing circuit.
Nitrous oxide (N
O) flows through the system during this test. Use a
2
safe and approved procedure to collect and remove it.
Incorrect gas mixtures can cause patient injury. If the Link system does not supply O
and N
2
O in the correct proportions, do not use
2
the system.
To perform the flow control tests:
1. Connect the pipeline supplies or slowly open the cylinder valves.
2. Turn all flow controls fully clockwise (minimum flow).
in the
2
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGG
3. Turn on the system.
4. Do not use the system if low battery or other ventilator failure alarms occur.
5. Make sure the O other flowtubes must show no gas flow.
GG
Keep the Link system engaged during steps 6 and 7:
• Adjust only the test control (N
flowtube shows approximately 25 to 75 mL/min. The
2
O in step 6 and O
2
2
in
step 7).
• Test the flows in sequence (N
O then O
2
).
2
• If you adjust the test control too far, set the flow controls to their
initial positions and do the step again.
• The O
sensor used in steps 6 and 7 must be correctly
2
calibrated.
1006-0938-000
6. Test the N
•Turn the N
•Slowly turn the N
•Make sure that the O The measured O
2
O flow control:
O and O
2
flow controls fully clockwise (minimum flow).
2
O flow control counterclockwise.
2
2
concentration must be
2
flow increases.
≥ 21% through the full range.
A-7
Page 84
Aestiva
7. Test the O
•Set the N
•Set the O
•Slowly turn the O
•Make sure that the N The measured O
flow control:
2
O flow to 9.0 L/min.
2
flow to 3 L/min or higher.
2
flow control clockwise.
2
O flow decreases.
2
concentration must be
2
≥ 21% through the full range.
8. Set the flow controls to mid range and make sure that the flowtube floats move smoothly.
9. Stop the O
supply. Disconnect the pipeline supply or close the cylinder
2
valve.
10. Make sure that:
•The low O
•N
O, CO
2
supply alarm occurs.
2
, Heliox, and O
2
2
flows stop. The O
flow stops last.
2
•Air flow continues.
•Gas supply alarms occur on the ventilator if the ventilator uses O
as the
2
drive gas.
11. Turn all of the flow controls fully clockwise (minimum flow).
Vaporizer back
pressure test
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGG
ww
ww
CCCCAAAAUUUUTTTTIIIIOOOONN
12. Reconnect the pipeline supplies.
GG
Anesthetic agent comes out of the common gas outlet during this
test. Use a safe, approved procedure to remove and collect the agent.
NN
To prevent damage, turn the flow controls fully clockwise
(minimum flow or Off) before you turn On the system.
1. Turn the system On. Alarms can occur.
2. Set the O
3. Make sure that the O
4. Adjust the vaporizer concentration from 0 to 1% one click at a time. The O
2
flow to 6 L/min.
2
flow stays constant and the float moves freely.
2
flow must not decrease more than 1 L/min through the full range.
A-8
If the O
flow decreases more than 1 L/min:
2
•Install a different vaporizer and try this step again.
•If the O
flow decreases less than 1 L/min with a different vaporizer, the
2
malfunction is in the first vaporizer.
1006-0938-000
Page 85
Appendix - Preoperative Tests
•If the O
flow also decreases more than 1 L/min with a different
2
vaporizer, the malfunction is in the Aestiva. Do not use the Aestiva system until it is serviced.
5. Complete steps 3 and 4 for each vaporizer.
Concentration Control
AA.43.005
ock
L
befo
AA.13.032
re
use
Lock
g
. e c
.
n
s
a
n
arnin
tio
orm
c
ce
W
rf
r
an
stru
pe
l fo
n
!
g
a
In
te
u
in
al
in
ct
an
s
a
n
fe
Keep upright at all times
f
m
tion
m
e
d
tio
a
ra
rs
Se
da
an
pe
e
to
e
O
en
ic
dat
ac
m
ce
F
i
rv e S
com
re
Next serv
5
4
0%
3
2
1
.2
.4
.6
.8
before
. e c
e
n
c
a
n
m
a
r
s
.
n
o
e
n
s
f
t
e
r
a
t
n
io
d
e
t
o
in
e
i
a
t
p
ic
a
s
v
d
c
g
r
e
m
e
n
u
s
r
in
e
m
d
t
t
t
i
x
c
s
t
n
m
e
e
l
n
N
a
f
l
m
f
I
r
e
l
a
o
c
o
a
a
c
i
t
s
n
v
l f
r
re
a
r
a
o
io
t
e
t
t
u
h
S
c
a
n
r
g
a
i
a
e
F
r
p
p
m
O
u
e e
p
S
Warning
e e
K
!
use
f
O
Use Only Enflurane
1
.8
2
3
4
5
e n
ra lu f
o
Is ly n
O
e
s
U
Power failure test
1. Unplug the power cord with the system turned On.
2. Make sure that the power failure alarm comes on.
3. Connect the power cable again.
4. Make sure the alarm cancels.
1006-0938-000
A-9
Page 86
Aestiva
Alarm tests 1. Connect a test lung to the patient connection.
2. Set the Bag/Vent switch to Vent.
3. Set the controls:
•Ventilation Mode: Volume control (Select from main menu)
•Ventilator: Tidal Vol: 400 ml Rate: 12 I:E Ratio: 1:2 Plimit: 40 cmH PEEP: Off
•Anesthesia Machine O
flow: minimum flow (25-75 mL/min)
2
All other gases: Off Push flush to fill the bellows.
4. Make sure that:
O
2
•Mechanical ventilation starts.
•A subatmospheric pressure alarm does not occur.
•The ventilator displays the correct data.
•The bellows inflate and deflate during mechanical ventilation.
5. Set the O
flow control to 5 L/min.
2
6. Make sure that:
•The end expiratory pressure is approximately 0 cmH
•The ventilator displays the correct data.
•The bellows inflate and deflate during mechanical ventilation.
7. Test the O
monitor and alarms:
2
•Make sure the sensor measures approximately 21% O
•Set the low O
•Set the low O
•Put the O
•Set the High O
alarm to 50%. Make sure a low O2 alarm occurs.
2
alarm back to 21% and make sure that alarm cancels.
2
sensor back in the circuit.
2
alarm to 50%.
2
•Push the flush button to fill the breathing system.
•Make sure the high O
•Set the high O
2
•After 2 min. in pure O
alarm comes On.
2
alarm back to 100% and make sure that alarm cancels.
, the sensor measures approximately 100% O2.
2
8. Test the low minute volume alarm:
O.
2
in room air.
2
A-10
•Go to the alarms menu.
•Set the alarm limit for low minute volume to 6.0 L/min.
1006-0938-000
Page 87
Appendix - Preoperative Tests
•Make sure that a low minute volume alarm occurs.
•Go to the alarms menu.
•Set the low minute volume alarm to Off.
9. Test the high airway pressure alarm:
•Set Plimit to less than the peak airway pressure.
•Make sure that the high airway pressure alarm occurs.
•Set Plimit to the correct level.
10. Test the apnea and low airway pressure alarms:
•Remove the test lung from the patient connection.
•Other alarms such as low minute volume can occur.
•Make sure that the low airway pressure and apnea alarms occur. The apnea alarm occurs after 30 sec.
11. Test the sustained airway pressure alarm:
•Set the controls:
Precase steps
APL Valve Closed Bag/Vent switch: Bag
•Mechanical ventilation stops when the Bag/Vent switch is set to Bag.
•Close the patient connection and push the O
Flush button.
2
•Make sure that the sustained pressure alarm occurs after approximately 15 seconds at the sustained pressure limit (6-30 cmH
O varies with
2
pressure limit).
1. Before the first case, do a low-pressure leak test. Refer to “Every time a different clinician uses the system.”
2. Do the last two sections of tests in “Before every patient”:
•Breathing system tests
•Monitor and ventilator tests
1006-0938-000 A-11
Page 88
Aestiva
Every time a different clinician uses the system
The following tests must be performed every time a different clinician uses the anesthesia system.
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
Low-pressure leak test
Negative low-pressure leak test 1. Turn on the auxiliary common gas outlet (AUX some models).
Do not use a system with a low-pressure leak. Anesthetic gas will
go into the atmosphere, not into the breathing circuit.
2. Or, access the common gas outlet (CGO).
CGO
3. Test the leak test device:
AB.23.063
AUX
AB.23.099
>60 sec
A-12 1006-0938-000
AB.23.063
Page 89
Appendix - Preoperative Tests
•Put your hand on the inlet of the leak test device. Push hard for a good seal.
•Remove all air from the bulb.
•If the bulb inflates in less than 60 seconds, replace the leak test device.
4. Set the system switch to Standby.
5. Turn off all vaporizers.
6. Test the anesthesia machine for low-pressure leaks:
•Turn the flow controls one and a half turns counterclockwise.
•Connect the test device to the common or auxiliary gas outlet.
•Compress and release the bulb until it is empty.
•The vacuum causes the floats to move. This is usual. If the bulb inflates in 30 seconds or less, there is a leak in the low-pressure circuit. Refer to the Troubleshooting table (Part 2 of this manual).
•Disconnect the test device.
7. Test each vaporizer for low-pressure leaks:
•Set the vaporizer to 1%
•Repeat step 6. If there is a low-pressure leak, refer to the Setup, Maintenance and Troubleshooting Manual.
8. Keep the test device with the system.
9. Turn all flow controls fully clockwise (minimum flow). Do not over tighten.
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
Agent mixtures from the low-pressure leak test stay in the system.
O
Always flush the system with
after the low-pressure leak test
2
(1 L/min for one minute).
w WWWWAAAARRRRNNNNIIIINNNNGGGGTurn off all vaporizers at the end of the low-pressure leak test.
10. Remove all condensate from the breathing circuit module.
11. Assemble the breathing system.
12. Flush the system with O
•Turn on the system.
•Set the O
•Continue the O
•Turn the O
flow to 1 L/min.
2
2
flow control fully clockwise (minimum flow).
2
•Set the system switch to Standby.
:
2
flow for one minute.
1006-0938-000 A-13
Page 90
Aestiva
ISO 5358 or BSI
low-pressure leak test
wCCCCAAAAUUUUTTTTIIIIOOOONNNN::
::
You can only do a positive pressure test at the common gas outlet.
1. Access the common gas outlet (CGO).
CGO
2. Connect the leak test device to the common gas outlet with a section of tubing.
AB.23.099
AB.23.101
•Turn off the auxiliary common gas outlet.
•Set the Outlet switch to common gas outlet.
•Keep the flow tube vertical for accurate results.
3. Fully close all flow controls.
4. Fully open the needle valve on the test device.
A-14 1006-0938-000
Page 91
w CC
CCAAAAUUUUTTTTIIIIOOOONNNN
Appendix - Preoperative Tests
If the needle valve is not fully open, this test can damage the
pressure gauge on the test device.
5. Open the Air or N2O flow control and set a total flow of 0.4 L/min through the flowmeter on the test device.
6. Make sure that the pressure gauge on the test device reads zero and that all other flow controls are fully closed.
7. Close the needle valve on the test device until the test gauge reads:
BSI 20 kPa ISO 5358 3 kPa
8. If the flow through the test device is less than 0.35 L/min (ISO) or
0.3 L/min (BSI), there is a low-pressure leak in the anesthesia machine. Refer to the Setup, Maintenance and Troubleshooting Manual.
9. Repeat this low-pressure leak test for each vaporizer:
WWAAAARRRRNNNNIIIINNNNGGGGSSSS
w WW
•Set the applicable vaporizer to 1% and do steps 2 through 8.
•Fully open the needle valve on the test device to decrease the back pressure.
•Turn the vaporizer off.
Agent mixtures from the low-pressure leak test stay in the system. Always flush the system with O
after the low-pressure leak test
2
(1 L/min for one minute).
Turn all vaporizers Off at the end of the low-pressure leak test.
10. Remove all condensate from the breathing circuit module.
11. Assemble the breathing system.
12. Flush the system with O
•Turn the system On.
•Set the O
•Continue the O
•Turn the O
•Set the system switch to Standby.
flow to 1 L/min.
2
2
flow control fully clockwise (minimum flow).
2
:
2
flow for one minute.
1006-0938-000 A-15
Page 92
Aestiva
Before every patient
Inspect the system
WWAAAARRRRNNNNIIIINNNNGGGG
w WW
w
CCCCAAAAUUUUTTTTIIIIOOOONNNNThe total weight on each accessory shelf must be less than 23 kg.
Before each case, perform the following anesthesia system checks.
Make sure that the breathing circuit is correctly connected and not damaged.
Systems without accessory shelves have a weight limit on the top surface of 23 kg.
Make sure that:
1. The equipment is not damaged.
2. All components are correctly attached.
3. The breathing circuit is correctly connected, not damaged, and contains sufficient absorbent.
4. The vaporizers are locked in position and contain sufficient agent.
5. Pipeline gas supplies are connected and the pressures are correct.
6. Cylinder valves are closed on models with cylinder supplies.
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGGGGDo not leave gas cylinder valves open if the pipeline supply is in
use. Cylinder supplies could be depleted, leaving an insufficient reserve supply in case of pipeline failure.
7. Models with cylinder supplies have a cylinder wrench attached to the system.
8. The necessary emergency equipment is available and in good condition.
9. Equipment for airway maintenance, tracheal intubation, and IV administration is available and in good condition.
10. Applicable anesthetic and emergency drugs are available.
A-16 1006-0938-000
Page 93
Appendix - Preoperative Tests
11. On trolley model, make sure the casters are not loose and the brake is set and prevents movement.
AA.96.100AB.29.007
12. Connect the power cord to a wall outlet. The mains indicator comes on when AC Power is connected.
•If the indicator is not on, the system does not have mains (electrical) power. Use a different outlet. Close the circuit breaker or replace or connect the power cable. Refer to Figure 2-2.
13. Set the system switch to On.
AA.96.104
1006-0938-000 A-17
Page 94
Aestiva
Minimize alarms
(optional)
Vaporizer installation
WWAAAARRRRNNNNIIIINNNNGGGG
wWW
Set the ventilator controls to decrease the number of alarms:
1. Control Keys:
•Volume alarms: Off
•Plimit: 100 cmH
2. Alarm menu:
•Low O
•High O
3. Bag/Vent switch: Bag
After you have finished testing, set the Plimit back to a clinically appropriate value.
: 21%
2
: Off
2
O
2
Use only the Tec 5, Tec 6, and Tec 7 vaporizers.
Do not use a vaporizer that lifts off the manifold when the lock lever is in the locked position.
Do not use this anesthesia system if you can turn on more than one vaporizer at the same time.
Tec 6 vaporizers will not align correctly unless the power cable goes through the channel on the bottom of the vaporizer. Do not put the power cable on top of the manifold or between vaporizers.
The vaporizer storage bracket is not part of the vaporizer manifold. You cannot use a vaporizer while it is on the bracket.
1. If the top of a vaporizer is not horizontal, remove the vaporizer and reinstall it.
2. Set each vaporizer lock lever to the locked position.
3. Try to lift each vaporizer off of the manifold.
4. If a vaporizer lifts off the manifold, install it again and complete steps 1, 2, and 3. If the vaporizer lifts off a second time, do not use the system.
A-18 1006-0938-000
Page 95
Appendix - Preoperative Tests
5. With a Tec 6 vaporizer:
•Make sure that the vaporizer is under the Tec 6 electrical outlet.
•Make sure that the vaporizer is connected to an electrical outlet.
•Hold down the alarm silence switch for a minimum of four seconds.
•Make sure all indicators come on and the alarm speaker starts.
•Release the alarm silence switch.
•Do not continue until the operational indicator comes on. The concentration control will not turn if the operational indicator is off.
6. Try to turn on more than one vaporizer at the same time:
•Test each possible combination.
•If more than one vaporizer turns on at the same time, remove the vaporizers, install them again, and complete steps 1 through 6.
Breathing system tests
Refer to the applicable operation and maintenance manuals.
At a minimum:
1. Make sure that the auxiliary equipment (humidifier, etc.) operates correctly.
2. Absorbers with active scavenging have a flow tube on the side. Make sure that it shows a flow in the green (normal) region.
3. With a circle breathing module, push the drain button for 10 seconds or more to drain condensate into the absorber.
AB.23.100
1006-0938-000 A-19
Page 96
Aestiva
4. Make sure that the one-way valves (breathing circuit module) work correctly.
• The inspiratory check valve rises during inspiration and falls at the start of expiration.
• The expiratory check valve rises during expiration and fall at the start of inspiration.
Note: The Bain/Mapleson D circuit module does not have one-way valves.
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGGGGObjects in the breathing system can stop gas flow to the patient.
This can cause injury or death:
• On Circle systems, press the drain button for a minimum of 10 seconds to remove moisture buildup.
• Do not use a test plug that is small enough to fall into the breathing system.
ww
ww
WWWWAAAARRRRNNNNIIIINNNNGGGGMake sure that there are no test plugs or other objects caught in the
breathing system.
5. Test the ventilator circuit for leaks:
•Set the Bag/Vent switch to Vent.
•Set all flow controls to minimum.
•Set the system switch to Standby.
•Close the breathing circuit at the patient connection. Use your hand or
an approved test plug located in the handle of the breathing system.
•Push flush to fill the bellows.
•The pressure must not increase to more than 15 cmH
•If the bellows falls more than 100 mL/min, it has a leak. Refer to the
troubleshooting procedure in the Setup, Maintenance and Troubleshooting manual.
•Turn on the system.
6. Test the Bag circuit for leaks:
•Set the Bag/Vent switch to Bag.
•Close the APL valve.
•Set the O
•Close the patient connection (hand or test plug) and inflate the bag
(flush) to 30 cmH
flow to 250 mL/min.
2
O.
2
O on the gauge.
2
A-20 1006-0938-000
Page 97
Appendix - Preoperative Tests
•Release the flush button. The pressure must not decrease. A pressure decrease large enough to see on the gauge indicates a leak. Look for and repair the leak (loose drain plug, open canister, breathing circuit assembly not pushed on completely).
•If your system has CO
bypass, move the absorber canister release to
2
the open position and do this test again to look for leaks in the bypass.
Note: If the message window shows “Close absorber canister”, you DO
NOT have a CO2 bypass. Close the canisters and do step 6 again.
7. Test the APL valve:
•Fully close the APL valve.
•Set the total fresh gas flow to 3.0 L/min and make sure that the value on the inspiratory pressure gauge is less than approximately 82 cmH
O.
2
•Fully open the APL valve.
•Make sure that the value on the inspiratory pressure gauge decreases to approximately zero.
•Push the flush button and make sure that the value on the inspiratory pressure gauge stays near zero.
•Set the O inspiratory pressure gauge does not decrease below 0 cmH
flow to minimum and make sure that the value on the
2
2
O.
8. Remove your hand (or the test plug - step 5) from the patient connection.
Monitor and ventilator
tests
1. Connect a test lung to the patient connection.
2. Set the Bag/Vent switch to Vent.
3. Set the controls:
•Ventilation Mode: Volume control (Select from main menu)
•Ventilator: Tidal Vol: 400 ml Rate: 12 I:E Ratio: 1:2 Plimit: 40 cmH
O
2
PEEP: Off
•Anesthesia Machine
flow: minimum flow (25-75 mL/min)
O
2
All other gases: Off Push flush to fill the bellows.
4. Make sure that:
•Mechanical ventilation starts.
•A subatmospheric pressure alarm does not occur. Note: With active gas scavenging, too much scavenging flow can cause subatmospheric alarms.
1006-0938-000 A-21
Page 98
Aestiva
•The ventilator displays the correct data.
•The bellows inflate and deflate during mechanical ventilation.
5. Set the O
6. Make sure that:
flow control to 5 L/min.
2
•The end expiratory pressure is approximately 0 cmH
Note: Positive end expiratory pressure when PEEP is Off, may indicate
that the scavenging system is not removing enough gas.
•The ventilator displays the correct data.
•The bellows inflate and deflate during mechanical ventilation.
7. Set the ventilator controls and alarm limits to clinically appropriate levels.
8. If the system will not be used immediately, set the system switch to Standby and close all cylinder valves.
9. Make sure that you have:
•Equipment for airway maintenance, manual ventilation, tracheal intubation, and IV administration.
•Applicable anesthetic and emergency drugs.
10. Prepare the system:
•Turn all vaporizers off.
•Open the APL valve.
•Set the Bag/Vent switch to Bag.
•Set all flow controls to minimum.
•Set sufficient patient suction.
•Make sure that the breathing system is correctly connected and not damaged.
O.
2
wwwwWWWWAAAARRRRNNNNIIIINNNNGGGGSSSSMake sure that the breathing circuit is correctly connected and not
damaged.
Before you connect a patient, flush the anesthesia machine with 5 L/min of O
for at least one minute. This removes unwanted
2
mixtures and by-products from the system.
A-22 1006-0938-000
Page 99

Index

A
About ventilator
show service settings About Ventilator... AGSS
connecting
operation, active Alarm
tone indicates priority Alarm limits
how to set
show or hide option
show/hide Alarm settings
menu function Alarm Settings menu
cardiac bypass Alarm silence
function
location Alarm volume
adjust Alarms
silence Alarms messages
display area for Altitude APL valve
Apnea alarm
Audible alarm for circuit leaks Auxiliary gas outlet
Auxiliary outlet
3-31
function
location
off during cardiac bypass
location
On/off
Stops ventilation
switch function
2-12 2-11
3-3
3-26
2-6 2-5
2-5
2-5
3-34
3-9
3-5
3-32
,
3-33
2-17
3-9
3-13
2-11
3-14
2-8
2-17
3-35
3-26
2-17
,
B
Bag arm
adjust position
location Bag/Vent switch
function
how to stop mech vent
start mechanical ventilation
2-5
2-6
2-6
3-11
3-12 3-11
3-15
3-14
Bag/vent switch
location
Bellows
location
Brake
function of
Breathing circuit Module
location
Breathing circuit module
display
Breathing system
controls
Breathing system tests
2-5
2-5
2-5
2-11
2-5
2-3
C
Canister release
function location
Cardiac bypass
function how to use
Check valves
location
Circuit breakers
location
Circuit compliance
how to measure
Circuit leak audio
turn on or off
CO2 bypass
function CO2 Bypass operation Control knob
location Control settings
display
2-7
2-5
2-17
2-5
2-4
2-7
2-11
2-11
3-12
3-11 2-17
3-30
D
Default settings Drain plug
function Drive gas
3-31
2-7
3-31
E
Electrical outlet
location
2-4
A-19
2-21
1006-0938-000
I-1
Page 100
Aestiva
End case key
function location
2-13 2-11
F
2-2
2-3
2-5
2-20 2-19
A-7
Flow control tests Flow controls
function of
Flow Sensor
location
Flowmeter (external O2)
controls location
G
Gas machine
controls
Gauge
airway pressure cylinder pressure pipeline pressure
Gooseneck lamp
location
2-2
2-2
2-5
H
Heliox
adjust patient data for heliox
Heliox mode
function
2-17
I
2-2 2-2
3-7
Mains inlet
location Measured values
display Mechanical ventilation
how to start
how to stop Menu
how to change menu settings
how to change settings
how to use
menu map
what different functions do Menu key
function
location Minimum system Modes
Pressure Control
PSVPro
SIMV
Volume Control Monitor and ventilator tests
2-11
3-21
3-20
2-4
3-1 3-15
2-15
2-16
2-12
2-11
,
3-26
3-20
3-20
3-14
O
O2 flush
function of O2 sensor
location O2 sensor cal
function Options
2-3
2-5
2-17
1-2
2-18
A-21
2-18
2-17
I
E ratio
3-19
Inspiratory pause
function how to set
2-17
3-22
L
Light
switch
2-2
turn on and off
Low pressure leak test
2-3
M
Mains indicator
location
I-2
2-11
A-12
P
PEEP 3-19, 3-21 Pinspired Pipeline and cylinder supplies Pipeline connection
Plimit Pneumatic outlet
Power failure alarm Power-up screen Preoperative checklist Preoperative test
3-19
location
2-4
3-19
location
2-4
A-9
3-2
power failure test
4-1
A-9
A-5
1006-0938-000
Loading...