7800 VentilatorInternational Operation and Maintenance Manual
Software Revision 4.XX
Page 2
User Responsibility
This Product will perform in conformity with the description thereof contained in this
operating manual and accompanying labels and/or inserts, when assembled, operated,
maintained and repaired in accordance with the instructions provided. This Product
must be checked periodically. A defective Product should not be used. Parts that are
broken, missing, plainly worn, distorted or contaminated should be replaced
immediately. Should such repair or replacement become necessary, Datex-Ohmeda
recommends that a telephonic or written request for service advice be made to the
nearest Datex-Ohmeda Field Service Support Center. This Product or any of its parts
should not be repaired other than in accordance with written instructions provided by
Ohmeda and by Datex-Ohmeda trained personnel. The Product must not be altered
without the prior written approval of Datex-Ohmeda’s Quality Assurance Department.
The user of this Product shall have the sole responsibility for any malfunction which
results from improper use, faulty maintenance, improper repair, damage, or alteration
by anyone other than Datex-Ohmeda.
Datex-Ohmeda products have unit serial numbers with coded logic which indicates a
product group code, the year of manufacture and a sequential unit number for
identification.
AAA A 12345
This alpha character indicates the year of product manufacture
and when the serial number was assigned; “Y” = 1995, “Z” = 1996,
“A” = 1997, etc. “I” and “O” are not used.
Page 3
Table of Contents
Table of Contents
1/Introduction1-1
How to use this manual 1-2
2/Getting Started2-1
General 2-1
Unpacking 2-2
Checking the ~ voltage 2-2
Setting the reverse flow alarm, sigh, contrast, audio volume 2-2
Adjusting the altitude or changing the language 2-4
Checking the supply gas 2-4
Matching the ventilator to anesthesia systems 2-5
Setting up the control module when the ventilator is used as a stand-
alone device 2-8
When the bellows assembly is mounted on an Ohmeda GMS Ab-
sorber 2-9
Setting up the control module with an Ohmeda Excel Anesthesia Sys-
tem 2-10
Connect sensor interface 2-10
Connect sensor interface cable 2-10
The control module’s location in an Ohmeda Excel Anesthesia Sys-
tem 2-12
When the bellows assembly is remotely located on an arm 2-13
Making the monitoring connections 2-15
Connecting the pressure sensing tube 2-15
Connecting the volume sensor 2-17
Connecting the oxygen sensor 2-19
Charging the battery 2-22
Long Term Ventilator Storage 2-22
3/General Information3-1
The controls, connectors, and display 3-1
The ventilator control module’s front panel 3-1
The ventilator control module’s rear panel 3-5
The ventilator’s modes 3-7
The sigh function 3-7
The alarm system 3-8
Alarm quick reference charts 3-9
Alarm definitions 3-11
Theory of operation 3-16
The ventilation cycle 3-16
Volume monitoring 3-18
Airway pressure monitoring 3-18
Oxygen monitoring 3-18
Control range computation 3-18
Tidal volume compensation 3-20
i1500-0134-00002/01/93
i
Page 4
Table of Contents
4/Preoperative Checkout Procedure4-1
Checking the ventilator connections 4-1
Checking the monitoring connections 4-1
Testing the bellows assembly 4-4
Testing the ventilator alarms 4-5
Testing the low and high oxygen alarms 4-5
Testing the low minute volume, reverse flow, and apnea alarms 4-6
Testing the high, low, and sustained pressure alarms 4-8
5/Operating the Ventilator5-1
Using the setup page 5-1
Setting the alarm limits 5-3
Setting the ventilation parameters, beginning ventilation 5-5
Responding to alarms 5-8
6/Maintaining the Ventilator6-1
Maintenance schedule 6-1
Long Term Ventilator Storage 6-1
Cleaning and sterilizing 6-2
Cleaning the control module 6-2
Cleaning and sterilizing the bellows assembly 6-2
Cleaning and sterilizing the volume sensor clip assembly 6-3
Cleaning and sterilizing the volume sensor cartridge 6-3
Checking the volume sensor 6-4
O2 sensor maintenance 6-5
Maintenance schedule 5
Installing a cartridge or disassembling the O2 sensor for cleaning 6-5
Cleaning and sterilization 6-7
100% O2 calibration 6-8
Cleaning the supply gas filter 6-9
Operation 6-9
Filter servicing 6-9
7/Service Procedures7-1
Repair policy 7-1
Troubleshooting guide 7-2
Ventilator problems 7-2
Ventilator problems, continued 7-3
Troubleshooting ventilator failure messages 7-4
Ventilator failure messages 7-5
Ventilator failure messages, continued 7-6
ii1500-0134-00002/01/93
ii
Page 5
Table of Contents
8/Autoclavable Bellows Assembly8-1
Introduction 8-1
Getting started 8-1
Ventilator Connections 8-2
Post Assembly Test 8-6
Cleaning and Sterilization 8-7
Cleaning 8-8
Sterilization 8-8
Periodic maintenance 8-9
Visual inspection 8-9
Pressure leak test. 8-9
Illustrated Parts List 8-11
9/Appendix9-1
Specifications 9-1
Electrical 9-1
Controls 9-2
Monitoring 9-2
Performance characteristics 9-4
Physical characteristics 9-5
Accessories 9-6
Ventilator mounting kits 9-6
Replaceable parts 9-7
Format for data in compressed mode 9-10
Analog outputs 9-14
Using a Bain circuit 9-14
Non-autoclavable bellows assembly, cleaning and sterilizing 9-15
Disassembling the bellows assembly 9-15
Cleaning the bellows assembly 9-17
Sterilizing the bellows assembly 9-18
Reassembling the bellows assembly 9-19
iii1500-0134-00002/01/93
iii
Page 6
Notes
Table of Contents
iv1500-0134-00002/01/93
iv
Page 7
1/Introduction
1/Introduction
This instrument combines an electronically-controlled, pneumaticallydriven ventilator with built-in monitoring for exhaled volume, inspiredoxygen concentration, and airway pressure. The ventilator also features
controls with clinically significant ranges, selectable inspiratory pause,
and an adjustable inspiratory pressure limit control. The Ohmeda 7800
Ventilator is designed to be used as a stand-alone ventilator; or as an
integrated part of the Ohmeda Excel Anesthesia System.
The following symbols are used on Ohmeda products and technical
manuals. No one product or manual has every symbol listed. Refer to this
listing concerning symbols found on various products and manuals.
øOn (power)
OOff (power
oStandby
qStandby or preparatory state
for a part of the equipment
p“ON” only for part of the equipment
œ“OFF” only for part of the equipment
†Direct Current
∏Alternating Current
xProtective earth ground
yEarth Ground
rFrame or chassis ground
åAlarm silence button
YEquipotential
PLamp, lighting, illumination
NMovement in one direction
ˆMovement in both directions
zLock
ZUnlock
ÍNon-autoclavable
mType B equipment
µType BF equipment
HType CF equipment
∑Warning ISO 7000-0085
wCaution, ISO 7000-0434
wWAttention, consult accompany-
ing documents, IEC 601-1
ÊThis way up
tVariability
TVariability in steps
+Plus, positive polarity
ËMinus, negative polarity
What warning and
caution statements in
this manual indicate
1-11500-0134-00002/01/93
No matter which part of the manual you are using, you should always be
familiar with the CAUTIONS and WARNINGS that appear throughout
this manual. WARNINGS alert you to conditions or actions that may
cause harm to humans. CAUTIONS alert you to conditions or actions that
may result in damage to equipment. Read the user responsibility statement; it describes what is expected of you to maintain the ventilator.
Read the warranty; it describes Ohmeda’s responsibility in case of a
functional defect.
πDangerous Voltage
≈Input
ÙOutput
1-1
Page 8
1/Introduction
Keep this manual with the system for answering questions that arise
about the ventilator’s operation, maintenance or, if necessary, repair.
WARNING: Before using the Ohmeda 7800 Ventilator, familiarize
∑
How to use this manual
yourself with it by reading through this entire manual. As with all
medical equipment, attempting to use this device without a thorough
understanding of its operation may result in injury to the patient.
This manual is designed both as a guide for you to follow when you are
learning to operate the Ohmeda 7800 Ventilator, and as a reference tool
for you to use once you are familiar with the system.
The Ohmeda 7800 Ventilator consists of two basic units: the bellows
assembly, which contains the bellows and bellows housing, and the
control module, which contains the ventilator’s control valves, processing
circuits, controls, and display screen.
If you are setting up the system for the first time, thoroughly read all of
the manual sections. The Ohmeda 7800 is available in two configurations: for use as a stand-alone device, or as an integrated component of
an Ohmeda Anesthesia System. Refer to the specific section in “Getting
Started” that tells you how to make the basic connections for your
configuration.
If the system is already in place, but you haven’t used it before, pay
particular attention to all of the sections starting with Section Three:
“General Information.”
If you have used the Ohmeda 7800 Ventilator before, but need reminding
about details of using the instrument, refer to Sections Four: “Preoperative Setup Procedures” and Section Five: “Operating the Ventilator.”
Section Six: “Maintaining the System” and Section Seven: “Service
Procedures,” are included to inform you about routine maintenance of
the ventilator and to help you solve problems that might occur with the
instrument.
Throughout this manual we have provided step-by-step instructions to
simplify the ventilator’s operation. To further clarify the instructions, we
have used a special typeface to identify messages that appear on the
ventilator’s screen. Messages from the ventilator are represented by a
dot-matrix typeface that simulates the messages’ actual appearance. A
low minute volume alarm message looks like this:
LOW MINUTE VOL!
In addition, the system’s alarm silence key is represented in a typeface
similar to the one printed on the key itself. An instruction to push the key
looks like this:
Press: å
1-21500-0134-00002/01/93
1-2
Page 9
1/Introduction
What the manual’s
symbols mean
What we mean by
“powering ON” the
control module
A number of Warnings∑ and Cautionsw are used throughout this
manual to draw attention to the possible hazards and/or adverse conditions which may occur if the information and instructions provided are
not strictly observed
Warnings are used to draw attention to a condition which can endanger
either the patient or operator. Cautions are used to draw attention to a
condition which can result in damage to the equipment. Special attention
must be paid to each Warning and Caution as it appears in the manual.
We have also used—both in the manual and on the device itself—symbols to represent some common terms. These symbols include:
T
IP
25% TIEnable inspiratory pause at 25% of inspiratory time
v
E
V
T
inspiratory pause
minute volume
tidal volume
ƒfrequency
I:EInspiratory to Expiratory ratio
EExpiratory time
IInspiratory time
CCompliance
PIPPeak Inspiratory Pressure
In this manual, when we say “power ON” the control module, use the
control module’s power switch or the anesthesia system’s master switch,
depending on your ventilator’s configuration.
1-31500-0134-00002/01/93
1-3
Page 10
1/Introduction
Notes
1-41500-0134-00002/01/93
1-4
Page 11
2/Getting Started
2/Getting Started
General 2-1
Unpacking 2-2
Checking the ~ voltage 2-2
Setting the reverse flow alarm, sigh, contrast, audio volume 2-2
Adjusting the altitude or changing the language 2-4
Checking the supply gas 2-4
Matching the ventilator to anesthesia systems 2-5
Setting up the control module when the ventilator is used as a stand-
alone device 2-8
When the bellows assembly is mounted on an Ohmeda GMS Ab-
sorber 2-9
Setting up the control module with an Ohmeda Excel Anesthesia Sys-
tem 2-10
Connect sensor interface 2-10
Connect sensor interface cable 2-10
The control module’s location in an Ohmeda Excel Anesthesia Sys-
tem 2-12
When the bellows assembly is remotely located on an arm 2-13
Making the monitoring connections 2-15
Connecting the pressure sensing tube 2-15
Connecting the volume sensor 2-17
Connecting the oxygen sensor 2-19
General
∑
Charging the battery 2-22
Long Term Ventilator Storage 2-22
Many of the steps in the following sections will be performed when the
ventilator is installed. However, during use, maintenance, or sterilization,
ventilator components may be left disconnected or may be reconnected
incorrectly. Read through the steps in each section to confirm that the
system is set up properly. Perform any steps necessary to correctly
connect your system’s components.
WARNING: To avoid explosion hazard, flammable anesthetic agents
such as ether and cyclopropane must not be used in this machine. Only
anesthetic agents which comply with requirements for non-flammable
anesthetic agents in the IEC Standard, Particular requirements for
safety of anesthetic machines, are suitable for use in this machine.
WARNING: As this machine is not suitable for use with flammable
anesthetic agents such as ether and cyclopropane, the use of anti-static
breathing tubes and face masks is not necessary. The use of anti-static
or electrically conductive breathing tubes when utilizing high frequency electric surgery equipment may cause burns and is therefore
not recommended in any application of this machine.
The following sections tell you how to set the ~ voltage, how to install
the ventilator’s bellows assembly and control module, how to install the
monitoring sensors, and how to charge the battery. Although these steps
are straightforward, they should be performed only by someone experienced in working with anesthesia and monitoring equipment.
2-11500-0134-00002/01/93
2-1
Page 12
2/Getting Started
Unpacking
Upon delivery, inspect the ventilator and its accessories for damage that
may have occurred during shipment. If you detect any damage, immediately notify the transportation company and file a damage claim. Save
the original shipping container and materials.
The Ohmeda 7800 Ventilator’s functions should be completely checked
as soon as possible. Follow the instructions in this section to install the
instrument. Then, after you have used this manual to familiarize yourself
with the ventilator, confirm that it is working correctly by performing the
preoperative checkout procedures described in “4/Preoperative Checkout
Procedures.”
Checking the ~ voltage
CAUTION: The Ohmeda 7800 Ventilator can be set to operate on 100,
w
120, 220, or 240 volts ~. Either 50 or 60 Hertz supplies are acceptable
and do not have to be set manually. Make sure the ventilator is set for
the voltage used at your location.
To change the
ventilator’s operating
voltage
Figure 2-1
Changing the control
module’s operating
voltage with the
voltage selector switch
1. Turn the control module upside down (disconnect any cables that
interfere).
2. Pull out the preoperative checklist. This exposes the voltage selector,
which you should be able to see through the hole labeled “Mains
Voltage Selector.”
3. Use a large, flat-blade screwdriver to align the arrow on the voltage
selector to the correct voltage.
4. Make sure an appropriate power connector is installed.
120
220
240
100
Mains
Voltage
Selector
AA.10.113
Setting the reverse flow alarm, sigh, contrast, audio
volume
Parameters set in the setup page are saved when the ventilator is turned
OFF and then back ON. The ventilator uses those parameters established
in the setup page when the power was cycled OFF.
1. Move the mechanical ventilation switch to OFF.
2. Power ON the control module. With the Excel, move the system
master switch to ON.
2-21500-0134-00002/01/93
2-2
Page 13
2/Getting Started
3. Press and continue to hold down the alarm silence button, å, then
press in the inspiratory pause button. Release both buttons. The
ventilator displays:
321
7800 REV 4.XX /O
ENGLISH 1300 m
45
1.Ventilator Model
2.Software Version
3.Ventilator Supply Gas (A=Air; O=O
4.Language
5.Altitude
WARNING: Pay attention to the information on the setup page. If
∑
the model number or supply gas is incorrect, have a trained
Ohmeda service representative service the ventilator.
Note:The ventilator stores the original inspiratory pause setting (ON or
OFF) in memory when the inspiratory pause button is used to access
setup pages. At the end of the setup page, the ventilator will use the
stored setting.
1 meter (3.28 feet)
)
2
4. Press å. The ventilator displays:
FLOW KNOB TO SET
REV FLOW ALM ON or OFF
Turn the flow control to switch the alarm selection ON or OFF. If the
volume sensor is at the proximal end of the Y, select OFF to disable
the alarm. If the volume sensor is at the expiratory port of the absorber, select ON to enable the alarm.
5. Press å. The ventilator displays:
FLOW KNOB TO SET
SIGH ON or OFF
Turn the flow control to switch sigh breaths ON or OFF. When sigh is
ON, the ventilator delivers one and a half times the tidal volume (up
to a maximum 1500 mL) once every 64 breaths.
6. Press å. The ventilator displays:
FLOW KNOB TO SET
CONTRAST: XX
Turn the flow control to adjust the ventilator display contrast (XX)
from 1 (lowest contrast) to 10 (highest).
7. Press å. A tone sounds and the ventilator displays:
FLOW KNOB TO SET
AUDIO VOLUME: XX
2-31500-0134-00002/01/93
2-3
Page 14
2/Getting Started
Turn the flow control to adjust the ventilator alarm volume (XX) from
1 (lowest) to 10 (highest). Tone volume changes to the selected level.
8. Press å. The ventilator beeps once and displays:
CHECK SETTINGS!
To exit the setup pages at any step, repeatedly press å, set the me-
chanical ventilation switch to ON, or do not adjust a control for 30 seconds. All previous changes will be saved in the ventilator memory.
Adjusting the altitude or changing the language
1. Set the mechanical ventilation switch and the control module power
to OFF. With the Excel, set the system master switch to OFF.
2. Hold down the inspiratory pause button and power ON the control
module. With the Excel, move the system master switch to ON. Turn
the flow control to set the altitude (meters).
FLOW KNOB TO SET
ALTITUDE: 1300 m
Normally the altitude compensation needs to be set only when the
system is first installed.
3. Press å to display the language page. Turn the Flow control to
select language.
FLOW KNOB TO SET
ENGLISH
Checking the supply gas
CAUTION: If the supply gas displayed is other than the supply gas you
w
How to determine your
ventilator’s drive-gas
setting
are using (“/O” for oxygen or “/A” for air), have an Ohmeda trained
service representative reset the ventilator. Using a supply gas that
does not match the displayed supply gas will result in operational
errors.
Either oxygen or medical-grade air can be used to power the ventilator.
Before changing from one supply gas to another, however, qualified
service personnel must set up the ventilator to operate correctly with the
new supply gas.
On the first line of ventilator’s setup page a character is displayed that
indicates the current supply-gas setting. “O” indicates oxygen and “A”
indicates medical-grade air.
Note:To enter the setup page: make sure the mechanical ventilation switch is
off, press and continue to hold down the alarm silence å button, press
the inspiratory pause button, then release both buttons.
To return to normal operations press the alarm silence å button to
move through the menu or, leave the Setup Page display on without any
parameter changes for 30 seconds.
2-41500-0134-00002/01/93
2-4
Page 15
2/Getting Started
Matching the ventilator to anesthesia systems
This chart describes how certain components of the Ohmeda 7800
Ventilator’s two configurations are installed, connected, or controlled.
Some of these attributes, such as the placement of the bellows assembly,
are optional, so more than one can apply to a configuration. Attributes
are marked with figure numbers that refer to relevant illustrations that
follow this chart. (The Ohmeda Excel configuration is listed twice because the control module can be installed in two different positions on
the anesthesia machine.)
Figure 2-2
Possible Configurations
of the Ohmeda 7800
Ventilator
Ohmeda ExcelOhmeda Excel
(control module(control module
Attributes ofmounted onhung from
configurationsStand-aloneoptional arm)shelf)
Bellows assemblyFigure 2-3Figure 2-4
on control module
Bellows assemblyFigure 2-4Figure 2-5 or 2-6
on optional arm
Connect monitorFigure 2-7Figure 2-7
device to sensor
interface panel on
anesthesia machine
Connect monitorFigure 2-8
devices to control
module
Use master ON/Figure 2-9Figure 2-9
OFF to power ON
ventilator
Use control moduleFigure 2-8
power switch to
power ON ventilator
Figure 2-3
Bellows assembly
mounted on control
module
2-51500-0134-00002/01/93
300
600
900
1200
1500
~
W V (mL)
z Z
Ohmeda7800 Ventilator
V
•
V
(L/min) (%)
Rate
(mL)
TE
300
200
100
50
20
500
10
1000
2
1500
mL
B/min
T
L
T
IP
O
2
Low v
+ + + + + +
60 21 00
.
- -- -
L/min % %
40
60
80
100
L/min
E
(25% T )
AA.32.130
I
Low O
High O
2
2
60
40
208010010100
cm H O
2
ø
o
2-5
Page 16
2/Getting Started
Figure 2-4
Bellows assembly
mounted on control
module that is attached
to an optional arm on
Ohmeda Excel
Anesthesia System
Figure 2-5
Bellows Assembly on
optional arm on
Ohmeda Excel
Anesthesia System
AA.32.013
Figure 2-6
Bellows assembly on a
GMS absorber Ohmeda
Excel Anesthesia
System
AA.32.015AA.32.158
APL
2-61500-0134-00002/01/93
2-6
Page 17
2/Getting Started
Figure 2-7
Sensor interface panel
on Ohmeda Excel
Anesthesia System
Figure 2-8
Sensor interface panel
on stand-alone Ohmeda
7800 Ventilator
Vol
O
2
Monitors
Ohmeda7800 Ventilator L
(25% T )
IP
Low O2 High O
E
.
I
T
Low v
+ ++ ++ +
60 21 00
- -- -- -
L/min % %
AA.32.004
2
Figure 2-9
System master switch
on Ohmeda Excel
Anesthesia System
300
200
500
100
50
1500
mL
1 2345
1.Sensor interface
2.Mains indicator
3.Volume monitor port
4.Oxygen monitor port
5.Power switch
1000
20
40
B/min
60
80
100
L/min
10
2
60
PUSH
TO
TURN
cm H O
80
10010100
2
40
20
p
q
AA.32.033
2-71500-0134-00002/01/93
AA.11.007
2-7
Page 18
2/Getting Started
Setting up the control module when the ventilator is
used as a stand-alone device
When the bellows assembly is mounted on the control
module
1. Connect the correct supply gas hose to the control module’s connector that is labeled either “Use Only Oxygen” or “Use Only Medical
Grade Air.”
2. A 23-cm long, drive-gas tube carries gas from the control module to
the bellows assembly. Connect one end of this tube to the connector
labeled “connect to bellows ass’y inlet” on the control module’s rear
panel.
3. Connect the free end of the drive-gas tube to the 17-mm inlet on the
bellows assembly.
Figure 2-10
Connecting the drivegas tube to a control
module that has a
bellows mounted on
top
300600
90012001500
M
W
Benutzen Sie nur
Keimfreie Luft
1
(241-517 kPa)
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
Verbinden Sie mit der
Inspirations - seite des
Beatmungssystems
W
Verbinden Sie mit
dem Einlaß an der
Balgeinheit des
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
y
Serieller Ventilatoranschluß
M
AA.32.007
Y
1.Supply gas connection
2.Supply gas filter, ventilators with air supply gas only
3.Drive-gas tube
2-81500-0134-00002/01/93
23
2-8
Page 19
2/Getting Started
When the bellows assembly is mounted on an Ohmeda
GMS Absorber
Figure 2-11
Connecting the drivegas tube to a remote
control module and
absorber interface
manifold
1
2
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
Ohmeda 7800 Ventilator
W
Verbinden Sie mit der
Inspirations - seite des
M
Beatmungssystems
W
Verbinden Sie mit
dem Einlaß an der
Balgeinheit des
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
Ohmeda
Madison WI 53707 7550
A Division of BOC Health Care Inc
BOC Health Care
µ
Made in USA
AA.32.006AA.32.129
34
576
1.Supply gas connection
2.Drive-gas connection to 17-mm inlet port
3.GMS Absorber
4.Interface manifold
5.Bellows assembly
6.17-mm inlet port
7.19-mm or 30-mm exhaust port
2-91500-0134-00002/01/93
2-9
Page 20
2/Getting Started
Setting up the control module with an Ohmeda Excel
Anesthesia System
For the ventilator to function as an integrated component of an Ohmeda
Excel Anesthesia System, the anesthesia system must contain internal
wiring that is designed to accommodate the 7800 Ventilator. All Excel
systems built since November 1989 contain this wiring. Older Excel
systems must be modified by trained service personnel to work with the
7800 Ventilator.
Your machine may be equipped with a sensor interface panel. Check for
this panel under the table top of your Excel machine. If the panel is not
present, contact your Ohmeda service representative.
Included with the ventilator configuration for Ohmeda Excel Anesthesia
Systems are two additional components that you, the user, can install.
These components, which connect to the internal wiring, are the interface cable and the sensor interface.
Connect sensor interface
Connecting the sensor
interface
Connecting the interface cable
Sensors for the oxygen and volume sensors connect to the sensor interface, which is a small box that attaches to a “D” connector that is under
the anesthesia system’s table. You need a small screwdriver to install
this component.
1. Check the left side of the anesthesia machine, under the table, to see
if the sensor interface is already installed. The sensor interface is a
small box that includes two connectors labeled “vol” and “O2.” If the
sensor interface is already in place, move to “installing the interface
cable.”
2. A plastic cover protects the 15-pin, female “D” connector that is on
the left side of the anesthesia machine, under the table. Remove this
cover.
3. Align the 15-pin, male “D” connector that is on the interface box with
the 15-pin, female connector that is on the anesthesia machine. Make
sure the two captive bolts in the box align with the corresponding
threaded posts in the anesthesia machine. Press up gently.
4. Use a small screwdriver to tighten the two screws.
Connect sensor interface cable
Signals from the sensor interface panel and the system master switch are
routed to a 25-pin male “D” connector at the rear of the anesthesia
system. This connector is labeled “ventilator/monitor pod interface.” The
Excel interface cable carries these signals from the 25-pin connector to
the 15-pin connector that is on the ventilator’s Excel interface panel.
1. Route the interface cable from the rear of the ventilator to the lower
right of the anesthesia machine’s back.
2. Insert the 15-pin male “D” connector that is on the cable into the 15pin female “D” connector that is labeled “Excel interface” on the
ventilator.
2-101500-0134-00002/01/93
2-10
Page 21
2/Getting Started
3. Tighten the two screws that hold the connector in place.
4. A plastic cover protects the 25-pin, male “D” connector (labeled
“ventilator/monitor pod interface”) that is mounted vertically at the
lower right of the anesthesia machine’s back. Remove this cover.
5. Insert the 25-pin, female “D” connector that is on the cable into the
25-pin, male “D” connector that is on the anesthesia machine.
6. Tighten the two screws that hold the connector in place.
Figure 2-12
Installing the Excel
sensor interface panel
Oxygen
Power
Outlet
Monitors
Vol
2
O
Figure 2-13
Excel interface cable
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
Benutzen Sie nur
Keimfreie Luft
(241-517 kPa)
Use Only Oxygen35-75 psig (241-517-kPa)
W
Inspirations - seite des
Beatmungssystems
W
dem Einlaß an der
Balgeinheit des
Ventilators
∑Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
y
Serieller Ventilatoranschluß
Verbinden Sie mit der
Verbinden Sie mit
AA.11.009
M
M
Y
2-111500-0134-00002/01/93
AA.32.017
2-11
Page 22
2/Getting Started
The control module’s location in an Ohmeda Excel
Anesthesia System
On all Ohmeda Excel Anesthesia Systems, the ventilator’s control module
can be mounted on a shelf or on an optional arm that is installed in the
dovetail mounting groove.
Figure 2-14
Control module
mounted from shelf.
∑
Figure 2-15
Control module
mounted on arm on
dovetail groove track
AA.32.015
WARNING: When the ventilator is mounted on a dovetail groove on an
Ohmeda Excel Anesthesia Machine, extra weight must be added to the
machine’s base to reduce the possibility of the machine tipping over.
2-121500-0134-00002/01/93
AA.32.013
2-12
Page 23
2/Getting Started
When the bellows assembly is remotely located on
an arm
To make these connections you’ll need one of the optional kits that are
listed in “Accessories” in the appendix.
1. Connect the supply gas hose to the control module’s connector that
is labeled either “use only oxygen” or “use only medical grade air.”
Figure 2-16
Connecting the drivegas tube to a bellows
assembly that is
mounted on an arm on
an Ohmeda Excel
Anesthesia System
M
W
Benutzen Sie nur
Keimfreie Luft
4
(241-517 kPa)
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
3
Verbinden Sie mit der
Inspirations - seite des
Beatmungssystems
W
Verbinden Sie mit
dem Einlaß an der
Balgeinheit des
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
M
AA.32.018
ToAnesthesiaMachine
ExhauseInlet
1.17-mm Inlet
2.Drive-gas tube
3.Supply gas filter, ventilators set up for air supply gas only
4.Supply gas connection
2. An optional 100-cm long, drive-gas tube carries drive-gas from the
control module to the bellows assembly. Connect one end of this
tube to the optional 90-degree adapter that connects to the control
module. The 90-degree adapter has an internal O-ring; make sure it is
in place.
1
2
3. Use the set screws on the 90-degree adapter to attach the adapter to
the connector labeled “connect to bellows ass’y inlet” on the control
module’s rear panel.
4. Connect the other end of the tubing to the far right 17-mm port on
the rear of the bellows assembly.
2-131500-0134-00002/01/93
2-13
Page 24
2/Getting Started
Figure 2-17
Connections between
an absorber and a
bellows assembly
1
M
W
Verbinden Sie mit der
Inspirations - seite des
Beatmungssystems
W
Verbinden Sie mit
M
dem Einlaß an der
Balgeinheit des
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
7
6
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
2
5
3
AA.32.024
1.Drive-gas tube
2.30-mm/19-mm port
3.17-mm port
4.To waste gas scavenging system
5.22-mm port
6.Supply gas
7.Supply gas filter, ventilators set up for air supply gas only
2-141500-0134-00002/01/93
4
2-14
Page 25
2/Getting Started
Making the monitoring connections
Where monitors
connect
How you make the monitoring connections depends upon the type of
anesthesia system and breathing system you use. The Ohmeda Excel
Anesthesia System includes a sensor interface panel that connects the
ventilator’s oxygen- and volume-sensor cables to the control module. On
other systems, all of which use the stand-alone configuration, these
cables connect directly to the front of the control module. At the sensor
end of the connections, the Ohmeda GMS Absorber provides ports that
connect the oxygen sensor and pressure-sensing tube to the inspiratory
side of the breathing system. To provide these ports for other types of
breathing systems you must use the adapters that are provided with the
ventilator.
Connecting the pressure sensing tube
The airway-pressure sensor is housed in the control module. A clear, 3mm (1/8-inch) tube connects between the control module and the distal
sensing tee, which must be in the inspiratory limb of the breathing
system.
For pressure monitoring to function correctly, the distal-sensing tee must
connect to the inspiratory side of the breathing system.
1. A barbed connector on either the absorber system or the patientcircuit adapter provides the distal-sensing tee for the ventilator’s
pressure sensor. Install one end of the sensing tube onto the barbed
fitting.
∑
2. Install the tube’s free end onto the barbed connector labeled “connect to inspiratory limb of breathing system” on the control module’s
rear panel.
WARNING: Position the pressure-sensing tube so that the absorber
arm cannot pinch the tube. If the tube is pinched, the system’s
pressure monitoring will not function correctly
WARNING: When used, the oxygen-sensor adapter and the pressure-sensor’s patient-circuit adapter must be connected to the
inspiratory side of the patient breathing system. If these devices
are not correctly connected to the inspiratory side of the patient
breathing system, oxygen and pressure monitoring and related
alarms will not function properly.
2-151500-0134-00002/01/93
2-15
Page 26
2/Getting Started
Figure 2-18
Connecting the
pressure-sensing tube
to the Ohmeda GMS
Absorber or the patient
circuit pressure-sensing
tee
Figure 2-19
Connecting the
pressure-sensing tube
to the ventilator
2-161500-0134-00002/01/93
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
Ohmeda 7800 Ventilator
W
Verbinden Sie mit der
Inspirations - seite des
M
Beatmungssystems
W
Verbinden Sie mit
dem Einlaß an der
Balgeinheit des
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
Ohmeda
Madison WI 53707 7550
A Division of BOC Health Care Inc
BOC Health Care
AA.32.006AA.15.007; AA.33.005,6,9
2-16
Page 27
2/Getting Started
Connecting the volume sensor
The volume sensor cartridge must be placed in the expiratory limb of the
breathing system, either in the distal or proximal position. Placing the
cartridge at the distal position in the expiratory limb lets the system
detect reverse flow and generate reverse flow alarms. You may also
place the volume sensor cartridge at the proximal end of the “Y” connector; however, you then must use the setup page to disable the reverse
flow alarms that would otherwise be generated when the patient inhales.
WARNING: When the volume sensor is in the distal position of the
∑
breathing system, confirm that the reverse-flow alarm is enabled. Do
not use the system with the reverse-flow alarm disabled if the volume
sensor is in the distal position.
To install the volume
sensor assembly
Figure 2-20
Connecting the volume
sensor to the Excel
machine’s sensor
interface panel
Figure 2-21
Connecting the volume
sensor to the control
module on a standalone ventilator.
1. Insert the sensor cable plug into the volume monitor receptacle on
either the anesthesia machine’s sensor interface panel or the
ventilator’s control module.
Vol
O
2
Monitors
(25% T )
Ohmeda7800 Ventilator L
300
200
100
500
50
1500
mL
1000
20
40
10
60
80
2
100
B/min
T
Low v
+ + + ++ +
60 21 00
L/min % %
L/min
IP
Low O2 High O
E
.
- -- -- -
I
2
60
PUSH
TO
TURN
cm H O
80
10010100
2
40
20
AA.32.008
2-171500-0134-00002/01/93
p
q
AA.32.034
2-17
Page 28
2/Getting Started
2. The volume sensor cartridge must be placed in the expiratory limb of
the breathing system, either in the distal or proximal position.
If you are using a Bain
circuit
Figure 2-22
Inserting the volume
sensor into the distal
location in the
breathing system
Figure 2-23
Inserting the volume
sensor into the
proximal location in the
breathing system
If you connect a Bain circuit and Bain circuit adapter to the Ohmeda 7800
Ventilator, you must place the volume sensor assembly in the proximal
Bain circuit position, between the end of the Bain circuit and the patient
connector to the ET (Endotracheal Tube) or mask. (See “Using a Bain
circuit” in the appendix.)
AA.33.007
∑
AA.33.002
3. Clip the sensor over the cartridge. The arrows on the sensor must
point in the direction of gas flow during expiration; the arrows must
point away from the patient.
WARNING: Take care not to crack or break the volume sensor
cartridge. When you are placing the cartridge on the absorber, be
certain to obtain a secure fit, but do not force the cartridge in place
as tightly as possible. Avoid striking the cartridge. A broken or
cracked cartridge could cause a circuit disconnection and a break
in the breathing system.
WARNING: The volume cartridge and sensor must be correctly
installed at either the distal location in the patient circuit’s expiratory limb or the proximal end of the “Y” connector. If the cartridge
and sensor are installed incorrectly, volume data will be inaccurate
and associated alarms, including the apnea and low-minute-volume alarms, will not function properly.
AA.33.002
2-181500-0134-00002/01/93
2-18
Page 29
2/Getting Started
WARNING: Destroy old or malfunctioning volume sensor cartridges to
∑
Figure 2-24
Attaching the volume
clip to the volume
sensor cartridge
prevent inadvertent reuse.
WARNING: Position the volume sensor’s cable so that the absorber arm
cannot pinch the cable. If the cable is pinched, the system’s volume
monitoring may not function correctly.
MD.10.016
Figure 2-25
Volume clip correctly
installed on volume
sensor cartridge
∑
w
MD.10.017
Connecting the oxygen sensor
General
WARNING: Disconnecting the O2 sensor without removing and short-
ing the sensor cartridge can cause false high O2 readings that may take
hours to stabilize.
CAUTION: The cable on the O2 sensor must point up to help keep the
contacts and the front of the cartridge free of condensate.
A newly installed O2 sensor cartridge needs five minutes of connection
time to stabilize. This waiting period has nothing to do with the sensor
response time which is much faster.
2-191500-0134-00002/01/93
2-19
Page 30
2/Getting Started
See “6/Maintaining the Ventilator.” for information on sensor housing
disassembly and reassembly as well as maintaining and replacing the
oxygen sensor cartridge.
1. Insert the sensor cable plug into the oxygen monitor receptacle
marked either on the anesthesia machine’s sensor interface panel or
on the ventilator’s control module. To reduce cartridge wear, avoid
leaving the sensor unplugged.
Figure 2-26
Oxygen sensor,
exploded view
MD.02.032
Figure 2-27
Connecting the oxygen
sensor to an Excel
machine’s sensor
interface panel
123 4
1.Sensor Housing
2.Oxygen sensing cartridge
3.Inner O-ring
4.Outer O-ring
Vol
O
2
Monitors
2-201500-0134-00002/01/93
AA.11.272
2-20
Page 31
2/Getting Started
Figure 2-28
Connecting the oxygen
sensor to the
ventilator’s interface
panel on a stand alone
ventilator.
(25% T )
Ohmeda7800 Ventilator
300
200
100
50
mL
20
500
10
1000
2
1500
B/min
L
40
60
80
100
L/min
I
T
IP
Low v
Low O
High O
E
2
+ + + + + +
60 21 00
.
- -- -- -
L/min % %
40
208010010100
p
q
60
PUSH
TO
TURN
cm H O
2
2
2. Insert the sensor into either the absorber’s sensor port or the port
provided by the oxygen-sensor tee in the inspiratory limb.
3. Calibrate the oxygen sensor as described in “6/Maintaining the
Ventilator.”
AA.32.035AA.33.008AA.15.027
Figure 2-29
Oxygen sensor
locations, see
“Accessories and
Replaceable parts for
oxygen sensor tee part
number
If any part of the sensor assembly is damaged or malfunctions, replace
the entire assembly. In addition, the oxygen cartridges wear out and
must be periodically replaced. See “6/Maintaining the Ventilator.” for
information about maintaining and replacing the oxygen sensor cartridge.
2-211500-0134-00002/01/93
2-21
Page 32
2/Getting Started
If, during operation, the sensor’s temperature is lower than or equal to
the breathing gas dew point temperature, water vapor will condense on
the sensor’s sensor screen. This condensate may reduce the amount of
oxygen reaching the sensor’s screen, which causes the ventilator to
display lower than actual oxygen-concentration values.
WARNING: Position the oxygen sensor’s cable so that the absorber arm
∑
w
Charging the battery
cannot pinch the cable. If the cable is pinched, the system’s oxygen
monitoring may not function correctly.
WARNING: When used, the oxygen-sensor adapter and the pressuresensor’s patient-circuit adapter must be connected to the inspiratory
side of the patient breathing system. If these devices are not correctly
connected to the inspiratory side of the patient breathing system,
oxygen and pressure monitoring and related alarms will not function
properly.
CAUTION: When in use, the oxygen sensor should always point down
to reduce condensation collecting on the sensor surfaces.
A built-in rechargeable battery provides backup power to the ventilator.
This battery is shipped uncharged. Before using the ventilator on a
patient, plug the control module into an energized ~ outlet for at least 24
hours to charge the battery.
w
CAUTION: If the integrity of the protective earth ground is in question,
the ventilator may be run on its fully charged battery. However, this
internal battery is for backup and not intended to operate the ventilator unless there is an emergency situation.
Long Term Ventilator Storage
It is not necessary to disconnect the rechargeable batteries before long
term storage.
If the ventilator is to be stored for an extended period of time, the batteries will eventually discharge. This is not destructive to the ventilator or
the batteries, however, the ventilator must be plugged into a wall outlet
for at least 24 hours prior to use to fully recharge the batteries.
2-221500-0134-00002/01/93
2-22
Page 33
3/General Information
3/General Information
The controls, connectors, and display 3-1
The ventilator control module’s front panel 3-1
The ventilator control module’s rear panel 3-5
The ventilator’s modes 3-7
The sigh function 3-7
The alarm system 3-8
Alarm quick reference charts 3-9
Alarm definitions 3-11
Theory of operation 3-16
The ventilation cycle 3-16
Volume monitoring 3-18
Airway pressure monitoring 3-18
Oxygen monitoring 3-18
Control range computation 3-18
Tidal volume compensation 3-20
The controls, connectors, and display
The Ohmeda 7800 Ventilator consists of two basic units: the bellows
assembly, which contains the bellows and bellows housing, and the
control module, which contains the ventilator’s control valves, processing
circuits, controls, and display screen.
The control module serves three functions: it controls mechanical ventilation; it contains the ventilator’s integrated monitors, which provide
oxygen, airway-pressure, and exhaled-volume monitoring; and it supplies
the ventilator’s alarm system. By using the control module’s front panel
knobs, pushwheels, and display, you can set and view the ventilator’s
operating parameters and alarm limits, view output from the integrated
monitors, and initiate mechanical ventilation. Switching ON the
ventilator’s power (with either the control module’s power switch or the
anesthesia system’s master switch, depending on your system’s configuration) enables the ventilator’s monitors and alarm system, even if the
mechanical ventilation switch is OFF.
The ventilator control module’s front panel
Display screenThe ventilator’s Liquid Crystal Display screen serves three functions: on
its top line it provides numeric readouts for expired tidal volume, breath
rate, expired minute volume and inspired oxygen concentration; on its
bottom line it displays messages such as alarms and control settings. For
certain functions, such as the setup page, the ventilator will display
instructions on both lines of the screen.
3-11500-0134-00002/01/93
3-1
Page 34
3/General Information
Figure 3-1
Front panel controls for
a stand-alone
ventilator.
Excel system ventilators have the same
controls.
Ohmeda7800 Ventilator
10
300
200
9
100
8
1.Inspiratory pause button
2.Alarm set pushwheels
3.Inspiratory pressure limit knob
4.Oxygen calibration knob
5.Inspiratory flow knob
6.Breath rate knob
7.Tidal volume knob
8.Alarm silence button
9.Mechanical ventilation switch
10.Screen
500
1000
50
1500
mL
å
(25% T )
L
20
40
B/min
60
80
100
10
2
T
Low v
+ ++ ++ +
60 21 00
- -- -- -
L/min % %
L/min
I
IP
Low O
High O
E
2
2
.
60
PUSH
TO
TURN
cm H O
80
10010100
2
40
20
1
2
3
4
p
q
AA.32.033
567
Tidal volume knobThe tidal volume knob lets you set the tidal volume at levels from 50 mL
to 1500 mL. As you turn the knob , the ventilator displays the tidal
volume setting as well as the resulting I:E ratio. The resolution varies
within four ranges, depending on the tidal volume knob ’s position. In the
range from 50 mL to 100 mL, the tidal volume can be set in 2-milliliter
increments. In the range from 100 mL to 250 mL, the tidal volume can be
set in 5-milliliter increments. In the range of 250 mL to 1000 mL, the tidal
volume can be set in 10-milliliter increments. And in the range of 1000
mL and up, the tidal volume resolution is 20-mL.
To check the tidal volume setting without changing its value, just touch
the front of the knob ; the ventilator will then display the current tidal
volume setting and I:E ratio.
Breath rate knobTurning the rate knob changes the breath rate used for mechanical
ventilation. The ventilator displays the rate as it changes as well as the
resulting I:E ratio. The rate is adjustable from 2 breaths per minute to
100 breaths per minute in whole number increments. Touching the rate
knob will display the current rate and I:E ratio setting on the screen.
Inspiratory flow knobThe inspiratory flow knob lets you set the inspiratory flow rate, which is
continuously variable from 10 liters per minute to 100 liters per minute in
increments of 1. Whenever you adjust or just touch the inspiratory flow
knob , the ventilator will display the current I:E ratio, which it calculates
based on the set inspiratory flow, tidal volume, breath rate, and the
inspiratory pause status. Because the inspiratory flow is continuously
variable within its range, the ventilator’s actual I:E ratios are continuously variable from 1:0.5 to 1:999. Rather than display I:E ratios in non-
3-21500-0134-00002/01/93
3-2
Page 35
3/General Information
standard increments, such as 1:2.13 or 1:1.97, the ventilator displays the
I:E ratio rounded to the nearest 0.5. For example, when the ventilator
uses a ratio of 1:2.13, it displays 1:2. And when it uses 1:1.97, it displays
1:2.
Inspiratory pressure
limit knob
Both the maximum inspiratory pressure and sustained-pressure alarm
limits are set by the inspiratory pressure limit knob , which must be
pushed in while turning to change the settings; the ventilator sets the
sustained pressure limit to correspond to the inspiratory pressure limit
knob setting. The maximum inspiratory pressure limit range is 20 to 100
cm H2O with a resolution of 1 cm. For inspiratory pressure limits of 20 cm
H2O to 60 cm H2O, the ventilator sets the sustained pressure limit to one
half the inspiratory pressure limit. Any inspiratory pressure limit setting
higher than 60 cm H2O will result in a sustained pressure limit of 30 cm
H2O.
As you push and turn the inspiratory pressure limit knob , the ventilator
will display both the maximum-pressure-limit and sustained-pressurelimit settings. However, unlike the other three control knobs, just touching this knob will not generate a display.
During mechanical ventilation, the maximum inspiratory pressure limit
you set is used by the ventilator’s electronically-controlled, automatic,
high-pressure-relief system to manage excessive airway pressure. If,
while the mechanical ventilation switch is ON, the ventilator detects
airway pressure higher than the limit you set, it will generate a high
pressure alarm and terminate the inspiratory cycle.
The ventilator also displays a message if you set the inspired pressure
limit to more than 60 cm H2O. This message is displayed in the nonmechanical ventilation mode only; during mechanical ventilation this
message is not displayed.
Inspiratory pause
button
Mechanical ventilation
ON/OFF switch
Pressing the inspiratory pause button adds an inspiratory pause—an
inflation hold—to the inspiratory cycle. When the inspiratory pause
function is active, the ventilator adds an inspiratory pause equal to 25
percent of the set inspiratory time. The ventilator, to maintain the original breath rate, then decreases the expiratory time by the same amount
that the inspiratory time is increased; pressing the inspiratory pause
button alters the I:E ratio.
After you press the inspiratory pause button, the ventilator displays the
new I:E ratio and lights the green indicator on the button to indicate that
the inspiratory pause function is active. To disable the inspiratory pause,
press the button again; the ventilator will switch OFF the indicator light
and display the I:E ratio, which is calculated from the other front-panel
control settings.
You can continue to adjust the ventilator’s front panel controls even
when the inspiratory pause function is active. If, while the function is
ON, you adjust a front panel control, the instrument takes the inspiratory
pause formula into account when it calculates and displays a new I:E
ratio.
The mechanical ventilation switch controls mechanical ventilation only.
When the switch is OFF, the monitors still function and the alarm system
is still active, although certain alarms are enabled only during mechanical
ventilation. When you want to start mechanical ventilation, move the
switch to ON.
Always switch ON the ventilator (using either the power switch or the
anesthesia system’s master switch, depending on your system’s configu-
3-31500-0134-00002/01/93
3-3
Page 36
3/General Information
ration) and set the ventilator’s controls before switching ON mechanical
ventilation.
Alarm set pushwheelsUse the three alarm-set pushwheels to change the low-minute-volume,
low-oxygen and high-oxygen alarms’ set points. To increase the value of
an alarm set point, push the button directly over the digit you want to
change. To decrease the value, push the button under the digit you are
changing.
Anytime you change the value of an alarm set point, the ventilator will
display, for a few seconds, that alarm’s value. Although all of the digits
can be set to zero, the ventilator will not accept certain oxygen alarm
settings, and will generate warning messages for others.
The system will not accept low oxygen alarm limits of less than 18
percent. The system also warns you if you set a high oxygen alarm limit
below or equal to the low oxygen alarm limit. Setting the high oxygen
alarm limit to 00 disables the high O2 alarm.
The oxygen alarm limits are 18 percent to 99 percent in one percent
increments. The low minute volume alarm limits are zero liters per
minute to 9.9 liters per minute in 0.1 liters per minute increments.
Oxygen calibration
thumbwheel
Alarm silence buttonTo silence an audible alarm, press the alarm silence å button. If that
The oxygen calibration thumbwheel is used to calibrate the oxygen
monitor’s sensor. Use this thumbwheel during the oxygen monitor
calibration procedure only.
alarm condition continues, the alarm will sound again in 30 seconds. If,
however, a new alarm condition occurs, its audible alarm will sound
immediately. Certain alarms can be silenced permanently, even if the
alarm conditions continue. These permanently silenceable alarms include
power failure, oxygen sensor failure, low battery, ventilator failure,
oxygen calibration error, and volume sensor failure.
When the mechanical ventilation switch is OFF—when the ventilator is
in its non-mechanical ventilation mode—pressing the alarm silence
button cancels and resets the apnea and low minute volume alarms; the
“VOL MON STANDBY” message will be displayed and the two alarms will
not sound again even if the alarm conditions continue. However, if the
ventilator senses another breath, the alarm timers and sensor circuits
will again be activated and any alarm condition that occurs will trigger
an appropriate alarm.
The alarm silence button å — combined with the inspiratory pause
button — also provides a way to enter and step through the setup page
mode. To enter the setup page, move the mechanical ventilation switch
to OFF, hold down the alarm silence button, then press the inspiratory
pause button. Once the setup page is displayed, press the alarm silence
button again to move from step to step.
Alarm indicator LEDsThe two light emitting diodes (LEDs) embedded in the alarm silence å
button indicate the status of alarms. When an alarm condition first
occurs, a message will appear on the screen, a tone will sound, and an
LED will flash. Once the alarm silence button is pressed, the ventilator
will light the LED continuously to remind you that the alarm condition
still exists. The red LED is lighted during alarm conditions that require
immediate operator response. The yellow LED indicates alarm conditions
that require prompt operator response or operator awareness.
3-41500-0134-00002/01/93
3-4
Page 37
3/General Information
Stand-alone power
switch
∑
Figure 3-2
Stand-alone
configuration power
switch and sensor
connectors
The ventilator’s control module has two power modes: ON and
“STANDBY.” The power switch determines whether the ventilator is ON
or in STANDBY. For the ventilator to operate, the control module must be
ON (in standby electrical power is delivered to the module but all of the
module’s functions are disabled).
WARNING: Switching ON the ventilator before setting the controls
may result in inappropriate ventilation of the patient and may activate
alarms.
(25% T )
Ohmeda7800 Ventilator L
300
200
100
500
1000
50
1500
mL
20
40
B/min
60
80
100
10
2
T
Low v
+ ++ ++ +
60 21 00
- -- -- -
L/min % %
L/min
I
IP
Low O
High O
E
2
2
.
60
PUSH
TO
TURN
cm H O
80
10010100
2
40
20
p
q
AA.32.033
1
2
1.Sensor connectors
2.Power switch
Powering ON configurations other than the
stand-alone ventilator
How you power ON the control module will depend on your system’s
configuration. If your ventilator is a stand-alone device, you will use the
power switch on the lower panel of the control module to power ON the
ventilator. If, however, your ventilator has been installed in an Ohmeda
Excel Anesthesia System, your control module doesn’t have a power
switch; instead you will use your anesthesia system’s master switch to
power ON the module. When you turn the master switch to ON, the
control module will also be powered ON.
WARNING: Switching ON mechanical ventilation before setting the
∑
controls may result in inappropriate ventilation of the patient and may
trip alarms that relate to mechanical ventilation.
The ventilator control module’s rear panel
Supply gas inputThe supply gas is introduced to the ventilator via a connector located on
the rear panel of the control module and is labeled for the type of gas and
pressure to be used.
Either oxygen or medical-grade air can be used as a supply gas to the
ventilator. Before changing from one supply gas to another, however,
qualified service personnel must set the ventilator’s software to operate
3-51500-0134-00002/01/93
3-5
Page 38
3/General Information
correctly with the new supply gas. (See “Setting the supply gas” in “2/
Getting Started.”)
Drive gas outputThe ventilator bellows’ drive gas is delivered from the connector labeled
“Connect to Bellows Ass’y Inlet.”
Figure 3-3
The ventilator’s rear
panel
1
M
W
Benutzen Sie nur
Keimfreie Luft
(241-517 kPa)
7
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
6
Verbinden Sie mit der
Inspirations - seite des
Beatmungssystems
W
Verbinden Sie mit
dem Einlaß an der
Balgeinheit des
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
M
2
AA.32.021
345
1.Pressure sensing input
2.Drive gas output
3.Excel interface connection
4.RS232 serial interface connection
5.Electrical power connection
6.Supply gas filter, ventilators set up for air supply gas only
7.Supply gas inlet
Pressure sensing inputA 3-mm (1/8-inch) inside diameter tube connects to the distal-sensing tee
located in the inspiratory limb of the breathing system to the connector
marked “Connect to Inspiratory Limb of Breathing System” on the
control module’s rear panel.
Serial interface
connector
For remote recording, a 25-pin female “D” type connector that is labeled
“Ventilator Serial Interface” provides access to the ventilator’s RS232C
serial port, which conforms to the Ohmeda standard communications
protocol (see the appendix).
WARNING: When specific DIP switches are set, writing to the
∑
ventilator’s RS232 port can alter the operation of the ventilator’s
software, which may result in unpredictable performance. Do not alter
the ventilator’s hardware or software.
The Excel interface connector
The Excel interface, which is a 15-pin female “D” connector, appears in
only the Excel configuration. Mounted on the back of the control
module’s lower panel, the interface connects to the anesthesia machine’s
power switch and sensor interface panel, which provides inputs for the
volume and oxygen sensors.
3-61500-0134-00002/01/93
3-6
Page 39
3/General Information
The ventilator’s modes
The Ohmeda 7800 Ventilator uses three basic modes:
•setup page mode
The setup page groups parameters not normally adjusted during a
case, such as the screen contrast and the alarm volume.
•mechanical ventilation mode
In the mechanical ventilation mode—when the mechanical ventila-
tion switch is ON—patient monitoring and alarms are active, and the
control module is driving the bellows assembly.
•non-mechanical ventilation mode
In the non-mechanical ventilation mode—when the mechanical
ventilation switch is OFF—mechanical ventilation is OFF but all
patient monitors still function and the alarm system is still active,
although certain alarms are enabled only during mechanical ventilation.
When you first enter the setup page, the ventilator displays the ventilator model, the software version number it is using, the supply gas setting, the selected language and selected altitude compensation . The
ventilator then lets you enable or disable the reverse flow alarm and sigh
function, and adjust the screen contrast and audio alarm volume. These
parameter settings are stored in the ventilator’s memory even when the
system’s power is disconnected; if you are satisfied with these parameter
settings, you can skip past the setup page.
In both the non-mechanical ventilation mode and the mechanical ventilation mode, the patient monitoring and the alarms systems are active.
The sigh function
When the sigh function is selected, the ventilator delivers 150 percent of
the set tidal volume once every 64th breath. The ventilator accomplishes
this by adding 50 percent to the inspiratory and expiratory times while
maintaining the set inspiratory flow and I:E ratio. The maximum sigh
breath is limited to 1.5 liters. For example, if the set tidal volume is 1
liter, the sigh breath will be 1.5 L (1 L plus 50 percent). But if the set tidal
volume is 1.4 L, the sigh breath will still be 1.5 L , because the inspiratory
flow decreases to maintain the I:E and limits the VT to 1.5 L.
To turn ON the sigh function, select “sigh on” from the setup page. Once
the sigh function is ON, the ventilator will display “sigh on” on the
bottom line of its screen. When the ventilator actually delivers the sigh
breath, the ventilator will display “sigh breath.” These messages will
alternate with other messages the ventilator displays.
When using the sigh function, pay particular attention both to the high
pressure alarm setting and to apnea alarms. Because circuit pressure is
higher during the sigh breath than during normal cycles, you must set
the inspiratory pressure limit knob to compensate for the sigh breath.
Also, at low rates the sigh function can cause apnea alarms. The apnea
alarm will be triggered if the ventilator does not sense a complete breath
in a 30-second period. At a breath rate of two, the sigh function increases
the breath period from 30 seconds to 45 seconds, triggering the apnea
alarm.
3-71500-0134-00002/01/93
3-7
Page 40
3/General Information
The alarm system
When the ventilator senses an alarm condition, it will display an appropriate message, which is updated every one-and-a-half seconds until the
condition is resolved (apnea alarms are updated at one-second intervals).
If a second alarm condition occurs before the first is resolved, the ventilator will alternate the messages for each condition. The ventilator will
alternate, at one-and-a-half-second intervals, the alarm messages for as
many alarm conditions as exist at one time.
Audible alarmsAlthough when more than one alarm is active the ventilator displays the
messages for all of the alarms (by alternating the messages), it sounds
the audible alarm for only the highest priority alarm. If that condition is
resolved, the ventilator will then sound the alarm for the next highest
priority alarm that occurred. The priority of the audible alarms, from high
to low, is:
•Warble
•Intermittent
•Continuous
•Single beep
What the LEDs indicateThe two light emitting diodes (LEDs) embedded in the alarm silence å
button indicate the status of alarms. When an alarm condition first
occurs, a message will appear on the screen, a tone will sound, and an
LED will flash. Once the alarm silence button is pressed, the ventilator
will light the LED continuously to remind you that the alarm condition
still exists. The red LED is lighted during alarm conditions that require
immediate operator response. The yellow LED indicates alarm conditions
that require prompt operator response or operator awareness.
Silencing alarmsTo silence an audible alarm, press the alarm silence button. If that alarm
condition continues, the alarm will sound again in 30 seconds. If, however, a new alarm condition occurs, its audible alarm will sound immediately. Certain alarms can be silenced permanently, even if the alarm
conditions continue. These permanently silenceable alarms include
power failure, oxygen sensor failure, low battery, ventilator failure,
oxygen calibration error, and volume sensor failure.
The ventilator’s built-in computer must be functioning correctly for the
alarm system to work correctly. If the ventilator’s computer fails, the
screen may flash erratically and an intermittent tone will sound; the
alarm silence button will not terminate the alarm if the ventilator’s
computer fails. Do not attempt to continue using the ventilator if this
failure occurs.
Responding to alarmsSee section 5, “Responding to alarms” for more information.
ApneaIf the ventilator doesn’t detect a sufficient breath for 30 seconds, an
apnea alarm will be generated. The alarm message will indicate the
number of seconds that have passed since the last sufficient breath was
detected. The tone will warble once at 30 seconds, twice at 60 seconds,
and three times at 90 seconds. At 120 seconds the tone will warble
continuously and the alarm message will be “apnea **” only.
The ventilator uses the volume-sensing circuits to activate the apnea
alarm. When the ventilator is first powered ON it displays the “vol mon
standby” message to indicate that a set threshold of flow is not being
sensed and the apnea alarm is not activated. Then, once the ventilator
senses a sufficient volume level, it removes the “vol mon standby”
message and starts a timer that controls the apnea alarm. Each time the
ventilator senses sufficient volume, it resets this apnea timer.
The actual volume threshold required to start or reset the apnea timer
varies depending on the level you set on the tidal volume knob. For tidal
volume settings between 180 mL and 400 mL, the threshold varies
linearly from 20 mL to 100 mL. For tidal volume settings below 180 mL,
the threshold is always 20 mL. And for tidal volume settings above 400
mL, the threshold is always 100 mL. In the manual mode, however, once
the apnea timer starts, the volume threshold required to reset the timer is
20-mL.
For example, if the tidal volume knob is set to 250 mL, and mechanical
ventilation is switched ON, the timer starts and is reset when the ventilator senses a breath of at least 46 mL. If you then increase the tidal volume
knob to 320 mL, the ventilator will increase the threshold to 71 mL. In
either case, if the timer reaches 30 seconds (because the ventilator didn’t
sense enough volume to indicate a breath), the first in the series of four
apnea alarms will sound.
Apnea alarm offBecause very low flow levels that are not sufficient to trigger the volume-
sensing circuits may occur in spontaneously breathing patients, the
ventilator disables the apnea alarm when the tidal volume knob is set to
less than 300 mL and mechanical ventilation is switched OFF.
Whenever tidal volume is set to less than 300 mL and the mechanical
ventilation switch is OFF, the ventilator, once it has detected a breath,
will display “APNEA ALARM OFF.” To enable the apnea alarm while the
patient is breathing spontaneously, increase the tidal volume knob to 300
mL or more. Although you should always carefully set the low minute
volume pushwheel to enable the low minute volume alarm at an appropriate level, the low minute volume alarm is especially important when the
tidal volume knob is set to less than 300 mL, disabling the apnea alarm.
Check O2 sensor/check
gas supply
If the ventilator detects less than five percent oxygen, it assumes that
either the oxygen sensor has failed or that insufficient oxygen is in the
breathing system, and it generates an alarm. A check O2 sensor/check
gas supply alarm will also be generated if the sensor isn’t connected
correctly, if the sensor is broken, or if no oxygen is in the area of the
sensor.
Check settingsIn the setup page the front panel controls are used to set parameters not
normally adjusted during a case, such as the screen contrast and the
alarm volume. Before you use the ventilator on a patient, you must
readjust any controls you used in the setup page. To remind you to check
your control settings, when you exit the setup page the ventilator displays the “check settings” message.
3-111500-0134-00002/01/93
3-11
Page 44
3/General Information
Drive circuit openOne type of ventilator failure—exhalation valve failure—does not display
a numbered message; instead DRIVE CKT OPEN is displayed. This message can also appear if, during mechanical ventilation, the absorber’s
Bag/APL-ventilation switch is in the “BAG/APL” position. During this
alarm, the ventilator will attempt to continue monitoring and mechanical
ventilation.
1. Check the patient.
2. If mechanical ventilation is ON, make sure the absorber’s Bag/APLventilation switch is in the “ventilator” position, if applicable.
Hardware errorThis display (error A, B, and C) and alarm should not occur unless there
is a problem with the ventilator control module hardware. The alarm is
silenceable with the alarm silence å button. Mechanical ventilation
does not turn OFF.
∑
High oxygenIf the ventilator detects an oxygen concentration equal to or higher than
High pressureIf, during mechanical ventilation or while in the non-mechanical ventila-
Limit setting errorIf you attempt to set the high oxygen alarm limit for a level below or
Low batteryTwo sources can power the control module: the ventilator’s power
WARNING: Do not use the ventilator if this display and alarm occur.
the one you set using the high-O2 pushwheel, the ventilator will generate
a high oxygen alarm.
tion mode, the ventilator detects airway pressure higher than the limit
you set using the inspiratory pressure limit control, the ventilator will
generate a high pressure alarm. In addition, during mechanical ventilation only, the ventilator uses automatic, high-pressure relief to manage
excessive airway pressure. If, while the mechanical ventilation switch is
ON, the airway pressure rises to a level that causes a high pressure
alarm, the ventilator will release the remaining drive gas into the atmosphere and end inspiration.
equal to the low oxygen limit, the ventilator will generate a limit setting
error alarm. This alarm will also be generated if you attempt to set the
low oxygen alarm limit for less than 18 percent.
supply and the ventilator’s backup battery. If the power supply fails,
either because of an electronic failure or because a-c power is lost, the
backup battery takes over.
If the battery’s voltage discharges below a set threshold, the ventilator
generates a “low battery” alarm. Pressing the alarm silence å button
permanently silences the alarm tone. To remind you of the condition the
LED remains lighted and the alarm message will still appear.
Low minute volumeIf the ventilator detects that the minute volume is less than the level you
set using the low minute volume pushwheel, the ventilator will generate
an alarm. To reduce nuisance alarms that can be generated when control
settings are changed, whenever you adjust the tidal volume knob or the
rate knob, move the mechanical ventilation switch to ON, or exit the
volume monitor standby condition, the ventilator will disable the low
minute volume alarm for 40 seconds.
WARNING: Always correctly set the low minute volume alarm and use
∑
3-121500-0134-00002/01/93
CO2 monitoring to aid in the detection of breathing system disconnections.
3-12
Page 45
3/General Information
Low oxygenIf the ventilator detects an oxygen concentration lower than the one you
set using the low-O2 pushwheel, the ventilator will generate a low
oxygen alarm.
Low pressureThe ventilator generates a low pressure alarm if, for at least 20 seconds,
the airway pressure fails to change by a value that varies proportionally
to the setting of the inspiratory flow knob. To determine the change in
airway pressure, the system compares the airway pressure at a point 50
milliseconds after the peak pressure to a point at the end of patient
exhalation. The amount of change required to prevent an alarm from
triggering will vary between 4 cm H2O to 9 cm H2O to correspond to the
inspiratory-flow range of 10 liters per minute to 100 liters per minute. For
example, if the inspiratory flow is set to 30 liters per minute, the low
pressure alarm will activate if the pressure doesn’t change by at least 5.1
cm H2O. But if the inspiratory flow is set to 80 liters per minute, the
change must be at least 7.9 cm H2O to keep the alarm from activating.
Unlike other alarms, the low pressure alarm is active only when mechanical ventilation is switched ON.
Low supply pressureIf the ventilator’s internal, regulated supply gas pressure is less than
152 kPa (22 psig), the ventilator will generate a low supply pressure
alarm.
Maximum pressure limitAnytime you adjust the inspiratory pressure limit knob, the ventilator
will briefly display the pressure limit in centimeters of water. However, if
the ventilator is in the non-mechanical ventilation mode and the inspiratory pressure limit is set to more than 60 cm H2O, the ventilator will also
light the yellow alarm LED and will continuously display the “max
pres=xxx cm” reminder. The pressure limit reminder is displayed in the
non-mechanical ventilation mode only; during mechanical ventilation this
reminder is disabled.
Oxygen calibration errorSome oxygen sensors may deliver a signal that is out of the ventilator’s
range. If the ventilator senses too large of a sensor signal, it will display
the “O2 cal error” message. To remove the message, continue to turn the
O2 calibration knob.
Power failureIf the a-c power to the ventilator is interrupted, the ventilator will gener-
ate a power failure alarm.
Reverse flowIf the ventilator senses reverse flow of an unacceptable volume, it will
generate an alarm. The volume that will trigger a reverse flow alarm
depends on the tidal volume you set. If you set the tidal volume knob for
less than 300 mL, then 20 mL or more will trigger an alarm. However, if
you set the tidal volume knob for 300 mL or more, the ventilator will
allow up to 100 mL before triggering an alarm. Once the alarm has been
triggered, two forward breaths must be sensed before the ventilator will
automatically switch OFF the alarm.
The reverse flow alarm is intended to warn you of inadvertent reverse
flow conditions, such as those caused by a defective or sticking exhalation check valve. These kinds of conditions will be detected only when
the volume sensor is correctly placed at the distal position of the expiratory limb of the breathing system. If you place the volume sensor at the
proximal end of the “Y” connector, volume monitoring will still function.
However, when the volume sensor is located at the proximal position, a
reverse flow condition will exist each time the patient inhales and the
reverse flow alarm will activate for each breath. To disable the reverse
flow alarm, select “rev flow alm off” from the setup page (see “Using the
setup page” in “5/Operating the Ventilator”).
3-131500-0134-00002/01/93
3-13
Page 46
3/General Information
WARNING: The reverse flow alarm will function correctly only if the
∑
Reverse flow alarm offTo advise you that the reverse flow alarm is disabled, in the non-me-
Sigh alarmsSigh breath replaces sigh on until the next breath is delivered
volume sensor cartridge is working correctly and is properly placed in
the distal position of the expiratory limb of the breathing system.
WARNING: When the volume sensor is in the distal position of the
breathing system, check the ventilator’s setup page to confirm that the
reverse flow alarm is enabled. Do not use the system with the reverse
flow alarm disabled if the volume sensor is in the distal position.
chanical ventilation mode the ventilator will repeatedly flash this message. If you power ON the control module with the alarm disabled and
mechanical ventilation ON (although you should always have the mechanical ventilation switch set to OFF when you power ON), the message will be displayed once.
Placing the volume sensor cartridge at the distal position in the expiratory limb lets the system detect reverse flow and generate reverse flow
alarms. You may also place the volume sensor cartridge at the proximal
end of the “Y” connector; however, you then must use the setup page to
disable the reverse flow alarms that would otherwise be generated when
the patient inhales.
Software errorThis display (error A through F) and alarm should not occur unless there
is a problem with the ventilator software programs. The alarm is
silenceable with the alarm silence å button. Mechanical ventilation
does not turn OFF.
∑
Sub-atmospheric
pressure
Sustained pressureThe ventilator sets the sustained pressure limit to correspond to the
Ventilator failureVentilator failure messages can indicate anything from a defective IC
WARNING: Do not use the ventilator if this display and alarm occur.
If the ventilator detects airway pressure of less than -10 cm H2O, it will
generate a sub atmospheric pressure alarm. For example, airway pressure of -12 cm H2O will cause an alarm.
inspiratory pressure limit knob’s setting. For maximum inspiratory
pressure limits of 20 cm H2O to 60 cm H2O, the ventilator sets the sustained pressure limit to one half the inspiratory pressure limit. For example, if the inspiratory pressure limit is 42 cm H2O, the sustained
pressure limit will be 21 cm H2O. However, any inspiratory pressure limit
setting of more than 60 cm H2O will result in a sustained pressure limit of
30 cm H2O. For example, inspiratory pressure limits of 65 cm H2O and 80
cm H2O will both result in a sustained pressure limit of 30 cm H2O.
Anytime the airway pressure exceeds—for 15 seconds or more—the
sustained pressure limit set by the inspiratory pressure limit knob, the
ventilator will generate a sustained pressure alarm.
chip to excessive pressure in the ventilator’s gas supply. However, some
ventilator problems may not generate any ventilator failure message,
even though the ventilator may not be functioning correctly.
3-141500-0134-00002/01/93
3-14
Page 47
3/General Information
WARNING: Do not attempt to use the ventilator while a ventilator
∑
failure message is displayed.
And, even if no ventilator failure message is displayed, do not use the
ventilator if you suspect a malfunction has occurred.
IMPORTANT
If the ventilator experiences extreme electrical interference, it may
interrupt mechanical ventilation. If this interruption occurs, the ventilator
generates an internal reset function and resumes normal operation after
two (2) seconds. For situations where continuous electrical interference
is experienced by the ventilator, causing a continuous interruption, the
ventilator’s internal reset repeats until the interference ceases.
If the electrical interference is continuously present and mechanical
ventilation is interrupted for approximately 30 seconds, the ventilator
produces a continuous beeping audio alarm. Manual ventilation of the
patient must be performed while the mechanical ventilation is interrupted. When the electrical interference ceases, the continuous beeping
audio alarm can be silenced only by turning, as applicable, the ventilator
or anesthesia machine power switch OFF and after five seconds back
ON.
WARNING: Manual ventilation must be performed when electrical
∑
∑
Ventilator setting errorIf you attempt to set a combination of the inspiratory-flow, tidal-volume,
interference causes interruption of ventilator delivered mechanical
ventilation. Manual ventilation must be continued until the ventilator
resumes normal operation or an alternate ventilator/anesthesia system can be used.
See the “7/Service Procedures” for descriptions of specific ventilator
failure messages.
WARNING: The use of electrosurgical units or other devices that
radiate high-intensity electrical fields can affect the operation of the
ventilator and monitors attached to the patient. Maintain as much
distance as possible between the electrosurgical leads and the cables
to the flow and oxygen sensors. Do not drape the electrosurgical leads
across the absorber or the anesthesia machine. Do not let the
electrosurgical leads rest on any surface of the anesthesia system.
Constant surveillance of all monitoring and life support equipment is
mandatory whenever electrosurgical devices are in operation, on, or in
the vicinity of the patient.
rate, and inspiratory-pause controls that results in a level the ventilator is
not designed to deliver, the ventilator will continue to use the most
recent acceptable settings, and will generate a “VENT SET ERROR”
alarm until the control combination is corrected. For example, a tidal
volume of 500 mL, combined with a rate of 20 B/min., and an inspiratory
flow of 10 liters per minute will trigger a ventilator setting error. To
remove the error, either decrease the tidal volume setting, or decrease
the rate setting, or Increase the flow setting.
3-151500-0134-00002/01/93
3-15
Page 48
3/General Information
WARNING: Do not use the ventilator while the “vent set error” mes-
∑
Volume monitor standbyWhen the ventilator is first powered ON, or when the ventilator is
Volume sensor failureThis message will be displayed if the volume sensor’s heater voltage is
sage is displayed; when this message is displayed, the control settings
do not reflect the settings the ventilator is using. If the “vent set
error” message is displayed during mechanical ventilation, the system
will use the most recent acceptable settings. If, however, the ventilator
is powered ON in the “vent set error” condition, moving the ventilation switch to ON will not start mechanical ventilation until the controls are moved to an acceptable setting.
switched from mechanical ventilation to the non-mechanical ventilation
mode, or when—while the ventilator is in the non-mechanical ventilation
mode—the alarm silence button is pressed, the ventilator displays the
“VOL MON STANDBY” message to indicate that the alarm system is
waiting for a sufficient breath. Once the ventilator senses a sufficient
volume level, it removes the “VOL MON STANDBY” message and starts
the timers that control the apnea alarms.
too low, which can happen if the volume sensor clip is broken or disconnected.
Dashes displayed in
place of readings
Question marks
displayed in place of
readings
If the system doesn’t measure any volume during mechanical ventilation,
it will display dashes in place of the volume and rate data.
Certain combinations of ventilator front panel control settings can result
in ventilation conditions the ventilator can deliver but the volume sensor
cannot measure. If you set a control combination the ventilator can
deliver, but the volume monitor cannot measure, or if—in the nonmechanical ventilation mode—breathing occurs that the monitor cannot
measure, the ventilator will display question marks in place of the VT, VE,
and rate readings.
Theory of operation
The ventilation cycle
The bellows assembly is the interface between the control module’s
driving-gas circuit and the patient breathing system. During inspiration,
driving gas from the control module compresses the bellows; during
expiration, breathing system gas fills the bellows, forcing it to rise. As
the ventilator cycles from inspiration to expiration, a set of valves controls the pressure in the two circuits.
3-161500-0134-00002/01/93
3-16
Page 49
3/General Information
Figure 3-4
Ventilation cycle
2
3
4
AA.04.039
1
1.Exhalation valve
2.Driving gas
3.Patient circuit gas
4.Pressure relief valve
5.To patient circuit
Inspiration starts when the
ventilator’s control module
closes the exhalation valve (1)
and delivers driving-gas (2) to
the bellows housing. As the
driving-gas pressure increases,
the pressure relief valve (4)
closes and the bellows is
compressed. This bellows
compression forces gas out of
the bellows, into the patientcircuit (5), breathing system,
and into the patient’s lungs. The
control module computes the
volume, rate, and timing of
driving-gas needed, and
delivers drive gas until it
reaches the calculated gas
volume; then flow stops. If the
ventilator detects airway
pressure higher than the limit
you set using the inspiratory
pressure limit control, the
ventilator generates a high
pressure alarm, opens the
exhalation valve, and ends
inspiration.
5
7
8
6
6.Driving gas
7.From patient circuit
At the start of expiration, the
exhalation valve opens and the
gas-flow direction in driving-gas
(6) and the patient-circuit (7)
breathing system reverses.
Fresh gas from the anesthesia
machine and exhaled gases
from the breathing system enter
the bellows’ interior, forcing the
bellows to expand; the
extending bellows displaces the
drive-gas (6), which is released
to the atmosphere.
8.Excess patient circuit gas
If the pressure inside the
bellows exceeds 2.5 cm H
during the expiratory cycle
(when the bellows has
extended completely), the
pressure relief valve opens and
releases any excess breathing
system gas (8)through the
bellows assembly exhaust port.
O
2
3-171500-0134-00002/01/93
3-17
Page 50
3/General Information
Volume monitoring
Two volume measurements—tidal volume (VT) and expired minute
volume (VE)—and the breath rate (Rate) are displayed on the ventilator’s
front panel. The ventilator measures all of these displayed values directly, based on the readings of a single volume sensor in the breathing
system. Because of compliance losses and fresh gas gains in the breathing system, these measured and displayed values may be different than
the values you set using the ventilator’s front panel controls.
To measure the exhaled patient volume, the ventilator uses a volume
sensor cartridge which contains a vane that is forced to rotate by gas
traveling through the breathing system. A clip-on, optically coupled
sensor translates the direction and speed of the vane’s rotation into
electrical pulses for the ventilator’s microprocessor to analyze. The
sensor clip also contains a heater, which is used to help prevent condensation in the transducer cartridge. Anytime the control module is ON, the
heater is ON.
Airway pressure monitoring
The Ohmeda 7800 Ventilator continuously monitors airway pressures in
the patient breathing system, and then uses this information to generate
alarms and manage airway pressure. This airway pressure monitoring
information is used only internally by the ventilator; the ventilator does
not display this information.
The ventilator’s airway pressure monitoring uses a transducer located inside the ventilator control module. A flexible tube fastened to a sensing
port in the breathing system connects to this transducer in the control
module.
Oxygen monitoring
The Ohmeda 7800 Ventilator uses a galvanic fuel cell to measure the
concentration of inspired oxygen. In addition to displaying the oxygen
concentration, the ventilator uses this information to generate highoxygen and low-oxygen alarms based on the levels set using the front
panel pushwheels.
Oxidation gradually consumes the electrode inside the oxygen sensor, so
it must be calibrated periodically and occasionally replaced (See Section
6, “Maintaining the Oxygen Sensor”).
Control range computation
The control module establishes four operating parameters directly from
the settings of the four front panel controls:
•The tidal-volume knob sets (VT), the volume of each breath in mL.
•The rate knob sets (R), the number of breaths per minute.
•The inspiratory flow knob sets (F), the instantaneous gas flow in
liters per minute. However, the ventilator doesn’t display this flow;
instead it calculates and displays the I:E ratio.
•The inspiratory pause button determines the status of the inspiratory
pause function, which adds an inspiratory pause—an inflation hold—
to the inspiratory cycle.
3-181500-0134-00002/01/93
3-18
Page 51
3/General Information
Then, based on the settings of the tidal volume, rate, and inspiratory
flow knobs, the control module calculates three more operating parameters (these calculations assume that the sigh and inspiratory pause
functions are OFF):
I =
E =
I:E =
V x 60
T
F x 1000
60 - I
R
I
E
•Inspiratory time (I), in seconds, is derived from the tidal-volume and
flow settings.
•Expiratory time (E), in seconds, is derived from the inspiratory time
and the rate setting.
•I:E ratio (I:E), is the ratio of the inspiratory time to the expiratory
time. The inspiratory side of the ratio is always expressed as “1.”
Whenever the inspiratory flow knob is touched, the control module
displays the approximated I:E ratio. Because the inspiratory flow is
continuously variable within its range, the ventilator’s actual I:E
ratios are continuously variable from 1:0.5 to 1:999. However, the
ventilator displays the I:E ratio rounded to the nearest 0.5.
Although all of the front panel knobs can be set independently to their
full-scale limits, certain combinations of the tidal-volume, rate, and
inspiratory-flow knobs will result in I:E ratios the ventilator is not designed to deliver. The control module will not accept I:E levels less than
1:0.5. Instead, the ventilator continues to use the most recent acceptable
settings, and displays the ventilator setting error message until the I:E
ratio is corrected.
For example, if the tidal volume is set to 250 mL, the rate is set to 40
BPM, and the flow is set to 10 liters per minute, the I:E ratio will be only
1:0.33. To increase the I:E ratio to 1:0.5 or more, you must either decrease
the tidal volume, decrease the rate, or increase the inspiratory flow. Once
the ventilator senses an acceptable control combination, it removes the
“vent set error” message and implements the new settings.
Figure 3-5
Diagram of the
ventilator’s range
100
Flow
L/min
"B"
10
"C"
2 4 10 20 30 40 50 60 70 80 90 100
Rate B/minTV mL
1500 1250 1000 750 500 250 066 50
"A"
"D"
This diagram illustrates the ranges of front-panel control combinations
the ventilator is designed to deliver. Control combinations that result in
I:E ratios the ventilator can deliver are represented as the solid area at
the back of the cube; the cutaway area at the cube’s front represents I:E
ratios that are out of the ventilator’s range. A control combination—as in
AA.32.019
3-191500-0134-00002/01/93
3-19
Page 52
3/General Information
the example above—of 250 mL tidal volume, 60 BPM rate, and 10 liters
per minute flow sets the I:E ratio at point “A,” which is in the cutaway
area. Increasing the flow moves the point up and out of the cutaway area
to point “B.” Decreasing the rate moves the point to the left and out of
the cutaway area to point “C.” And, decreasing the tidal volume moves
the point to the right and out of the cutaway area to point “D.”
Tidal volume compensation
You may notice that the exhaled tidal volume (VT) the ventilator measures and displays usually does not match the setting on the tidal volume knob. In most cases this is normal: the ventilator measures the
patient’s actual exhaled volume, which—because of a number of factors—will usually be different than the set tidal volume; use the measured volume as a guide when setting tidal volume. Factors contributing
to differences between the set tidal volume and the measured tidal
volume include breathing system compliance, fresh gas flow, breathing
system leakage, the location of the volume sensor within the breathing
system, and airway resistance.
ComplianceBecause of the compressibility of gases and the expansion of some
breathing system components under pressure, not all of the gas delivered from the ventilator enters the patient’s lungs. Instead of reaching
the patient, some of the gas the ventilator delivers is needed to raise the
breathing system pressure to the peak inspiratory pressure. Higher peak
inspiratory pressures results in greater tidal-volume losses.
Fresh gas flowAny fresh gas flow the anesthesia machine introduces to the breathing
system during inspiration will be delivered to the patient in addition to
the gas the ventilator delivers. Higher fresh gas flows result in greater
tidal volumes.
LeakageBreathing system leakage during inspiration reduces the delivered tidal
volume. In a properly maintained breathing system, leakage is usually
small enough to ignore when calculating tidal volume compensation.
Location of the volume
sensor
When the volume sensor is in the proximal location—on the patient side of
the “Y” connector—the ventilator will measure only the patient’s exhaled
tidal volume. If the volume sensor is placed in the distal location—at the
absorber’s exhalation check-valve port—the ventilator measures the
patient’s exhaled tidal volume plus the compliance volume in a portion of
the patient circuit that is between the absorber’s inhalation and exhalation check valves and the patient; this compliance volume is not delivered
to the patient.
Tidal volumes measured distally will always be artificially higher than
those measured proximally; the difference between the measurements
will usually be small (about 2 to 3 mL/cm H2O) when standard, 75 to 100
cm (30-to-40-inch) long, patient-circuit tubing is between the absorber and
the patient. Adding volume to the circuit, for example by connecting a
humidifier in the inspiratory limb, increases the differences between
distally and proximally-measured tidal volumes.
Airway resistanceHigh airway resistance, such as caused by a small endotracheal tube or
other airway obstruction, may reduce the tidal volume the ventilator
delivers to the patient. A tidal volume delivered at a high inspiratory flow
may be partially restricted from reaching the lungs, causing a larger-thannormal portion of that tidal volume to remain in the breathing system. You
can determine if airway resistance is a factor in your system: if reducing
the inspiratory flow or enabling the inspiratory pause function increases
the measured tidal volume, then high airway resistance is a factor.
3-201500-0134-00002/01/93
3-20
Page 53
3/General Information
These breathing system factors may cause the measured tidal volume
indicated on the screen to be different than the level you set on the tidal
volume knob. During operation compensate for these factors by adjusting
the ventilator controls so the measured-and-displayed tidal volume
indicates the ventilation level you want to use. Occasionally, however,
you may want to calculate an expected tidal volume.
V = V + V - V
Tsfgfc
V =
fgf
Vc = C x PIP
C = compliance factor
in mL/cm H
FGF
R 1 +
( )
O.
2
E
I
The exhaled tidal volume you would expect to measure (VT) equals the
tidal volume set on the ventilator (Vs) plus the tidal volume fresh gas
flow adds (V
) minus the tidal volume lost to breathing system compli-
fgf
ance (Vc).
The equation above doesn’t account for leakage or high airway resis-
tance. You can compensate for high airway resistance by reducing
inspiratory flow or using the inspiratory pause function.
Step One, calculating V
, the total volume of fresh gas delivered during
fgf
inspiration.
•FGF = total fresh gas flow from the anesthesia system, in mL per
minute.
•R = breathing rate from the ventilator, in breaths/minute
•f(E,I) = inverse I:E ratio from the ventilator
Step Two, calculating Vc, the volume lost to breathing system compli-
ance.
When the volume sensor is in the distal position, the compliance factor
(C) for the Ohmeda GMS Absorber is about 8 mL/cm H2O with adult
bellows and about 6 mL/cm H2O with pediatric bellows; because the
volume sensor is located distally, however, actual patient volume will be
somewhat less than the tidal volume the ventilator measures and displays. When the volume sensor is in the proximal position, the compliance factor is about 10 mL/cm H2O for the Ohmeda GMS Absorber with
adult bellows, 150 cm (60-inch) long, disposable, patient-circuit tubes,
and no humidifier. If your system includes components different from
these, see step three to calculate your system’s compliance factor.
PIP = peak inspiratory
pressure, as shown on
the breathing system’s
pressure gauge, in cm
O.
H
2
For example, to verify the volume reading of a system connected to
either a patient or a test lung: assume that the ventilator settings are V
= 750 mL (on the tidal volume knob), R = 10/min. (on the rate knob), and
the inspiratory flow is set so that I:E = 1:3. Assume also that fresh gas
flows total 3 liters per minute (3000 mL per minute), that PIP = 30 cm H2O
(peak inspiratory pressure as shown on the breathing system’s pressure
gauge), and that the volume sensor is located in the distal position.
V =
fgf
V = C x PIP
c
V = V + V - V
Tsfgfc
FGF
R 1 +
( )
E
I
so
so
so
3000 mL per minute
V = = 75 mL
fgf
10/min. 1 +
V = (8 mL/cm H O) x (30 cm H O) = 240 mL
c22
V = 750 mL + 75 mL - 240 mL = 585 mL
T
The expected tidal volume (VT) is 585 mL. Because of the number of
variables in the equations above, measured tidal volumes within about
15 percent of the calculated value represent a reasonable correlation to
the set tidal volume of 750 mL. Leakage will further reduce the measured
value.
3-211500-0134-00002/01/93
3
( )
1
s
3-21
Page 54
3/General Information
Step Three, determining the compliance factor. The calculations above
use an approximate compliance factor. Occasionally, you may want to
determine the compliance for a specific breathing system.
Determining the
breathing system
compliance factor when
the volume sensor is in
the proximal position
∑
1. Connect the ventilator’s control module and bellows assembly to the
breathing system as if ready for use.
2. If applicable, ensure that the absorber is full of absorbent as if ready
for use.
3. If applicable, ensure that the absorber’s Bag/APL-Ventilator switch is
in the “ventilator” position.
4. Connect all of the breathing system components—such as a humidifier, if included in the system, and patient-circuit tubing—as if ready
to use.
5. Occlude the patient end—at the “Y” connector—of the patient
circuit.
WARNING: When occluding the breathing system for test purposes,
do not use any object small enough to slip completely into the
system. Objects in the breathing system can interrupt or disrupt
the delivery of breathing-system gases, possibly resulting in injury
to the patient. Before using the breathing system on a patient,
always check the breathing system components for foreign objects.
6. Move the ventilator’s mechanical ventilation switch to OFF.
7. Turn all of the fresh-gas flow controls fully clockwise.
8. Power the ventilator ON.
9. Set the tidal volume to 200 mL.
10. Set the rate to 10 B/min.
11. Turn the inspiratory flow knob completely counter-clockwise to 10
liters per minute.
12. Inflate bellows using O2 flush button.
13. Activate the inspiratory pause function.
14. Move the ventilator’s mechanical ventilation switch to ON.
15. Adjust the tidal volume knob until the peak inspiratory pressure—
shown on the system’s pressure gauge—reads exactly 30 cm H2O.
16. Move the ventilator’s mechanical ventilation switch to OFF.
17. Touch the tidal volume knob to display the tidal volume that was
required to achieve a peak inspiratory pressure of 30 cm H2O. Note
the displayed value.
18. C = Vs ÷ PIP. Divide the tidal volume value you noted by 30 cm H2O
to calculate the compliance of your system.
19. Power the ventilator OFF.
20. Remove the occlusion from the patient circuit.
3-221500-0134-00002/01/93
3-22
Page 55
4/Preoperative Checkout Procedure
4/Preoperative Checkout Procedure
Checking the ventilator connections 4-1
Checking the monitoring connections 4-1
Testing the bellows assembly 4-4
Testing the ventilator alarms 4-5
Testing the low and high oxygen alarms 4-5
Testing the low minute volume, reverse flow, and apnea alarms 4-6
Testing the high, low, and sustained pressure alarms 4-8
WARNING: Always perform the preoperative checkout procedures
∑
before using the system. Failure to ensure proper setup and operation
prior to use may result in patient injury.
WARNING: Ensure that all hoses, tubing, and other circuit connections
are made properly before using any anesthesia system. Failure to do so
may result in patient injury. Refer to the operation manuals for these
devices.
Checking the ventilator connections
Before you use the ventilator on a patient, check that all of the ventilator
connections are correct and secure.
1. Verify that the proper hose connections have been made between
the bellows assembly and the control module.
2. Verify that the proper hose connections have been made between
the bellows assembly and the patient breathing circuit.
3. Verify that the correct supply gas is securely connected to the control
module.
4. On ventilators with air supply gas, check the supply gas filter bowl, if
water has accumulated, drain as described in section 6.
5. Verify that the electrical power cord is plugged into a properly
grounded outlet.
6. Verify that a properly functioning scavenging system is connected to
the bellows assembly’s 30-mm/19-mm exhaust port . Do not connect
the bellows exhaust directly to a vacuum source.
Checking the monitoring connections
Before you use the ventilator on a patient, check that all of the monitoring connections are correct and secure.
4-11500-0134-00002/01/93
4-1
Page 56
4/Preoperative Checkout Procedure
Figure 4-1
The volume sensor
cartridge must be
correctly inserted in the
breathing system. The
cartridge must be clear
and free of any
obstructions.
~
Figure 4-2
The volume sensor clip
must be attached to the
volume sensor
cartridge. The arrows
on the clip must point
in the direction of gas
flow in the breathing
system.
N
AA.15.046AA.15.045M D.10.017
Figure 4-3
The volume sensor
Ohmeda 7800 Ventilator
clip’s electrical cord
must be plugged into
the receptacle marked
“volume monitor” on
the sensor interface
panel
300
200
500
100
50
10
1000
1500
mL
4-21500-0134-00002/01/93
(25% T )
T
IP
Low v
Low O2 High O
E
+ + + + + +
60 21 00
.
- -- -- -
L/min % %
L/min
I
2
60
40
PUSH
TO
TURN
208010010100
cm H O
2
p
q
Vol
O
2
Monitors
AA.32.036
L
20
40
60
80
2
100
B/min
4-2
Page 57
4/Preoperative Checkout Procedure
Figure 4-4
One end of the clear, 3
mm (1/8 -inch) pressure
sensor tube must
connect to the barbed
connector labeled
“connect to inspiratory
limb of breathing
system” on the control
module’s rear panel.
The other end of the
pressure sensor tube
must connect to the
inspiratory side of the
breathing system. This
tube must be free of
obstructions and kinks.
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
Ohmeda 7800 Ventilator
W
Verbinden Sie mit der
Inspirations - seite des
M
Beatmungssystems
W
Verbinden Sie mit
dem Einlaß an der
Balgeinheit des
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
Ohmeda
Madison WI 53707 7550
A Division of BOC Health Care Inc
BOC Health Care
µ
Made in USA
AA.15.007; AA.32.006; AA.33.009
Figure 4-5
The oxygen sensor’s
electrical cord must be
connected to the
receptacle labeled
“oxygen monitor” on
the sensor interface
panel. If the sensor has
been left unplugged,
replace the oxygen
sensor cartridge.
The oxygen sensor
must be placed the
inspiratory side of the
breathing system.
Vol
O
Monitors
2
AA.15.043AA.32.032
4-31500-0134-00002/01/93
4-3
Page 58
4/Preoperative Checkout Procedure
Testing the bellows assembly
Before each case, perform the preoperative checkout procedure that is
recommended for your breathing system (check the operation and
maintenance manual for your breathing system). If applicable zero the
pressure gauge before each case. The breathing system must pass its
leak test before the bellows assembly can be tested.
For more detailed maintenance, testing and exploded view with parts
listed, see “Autoclavable bellows assembly,” section 8 of this O&M
manual.
1. Occlude the patient end—at the “Y” connector—of the patient
circuit.
Figure 4-6
Occluding the
breathing system at the
“Y” connector
AA.12.031
∑
WARNING: When occluding the breathing system for test purposes, do
not use any object small enough to slip completely into the system.
Objects in the breathing system can interrupt or disrupt the delivery of
breathing-system gases, possibly resulting in injury to the patient.
Before using the breathing system on a patient, always check the
breathing system components for foreign objects.
2. If applicable, turn the breathing system’s Bag/APL-ventilator switch
to the “ventilator” position.
3. Watch the breathing system’s pressure gauge. Press the oxygen
flush button to fill the bellows. When full the bellows may swell, but
it must remained installed on the bellows base. The pressure gauge
reading must not exceed 15 cm H2O.
4. Examine the bellows and confirm that it is undamaged.
5. Release the oxygen flush button.
6. Watch the pressure gauge and adjust the oxygen flow from 200 mL
per minute to 10 liters per minute and back. The pressure reading
must stay within the range of 1.0 cm H2O to 5.0 cm H2O. This tests
the bellows assembly’s pressure relief valve.
7. Shut off fresh gas flow, switch OFF the anesthesia machine. The
bellows must not drop more than 100 mL per minute. If the bellows
drops more than 100 mL per minute, either the bellows is leaking or
the pressure relief valve is not functioning properly.
8. If any leak cannot be corrected, do not use the ventilator; have a
qualified service representative make repairs.
9. Remove any occlusions you have added to the circuit.
4-41500-0134-00002/01/93
4-4
Page 59
4/Preoperative Checkout Procedure
Testing the ventilator alarms
Testing the low and high oxygen alarms
1. Make sure the mechanical ventilation switch is set to OFF.
2. Power ON the ventilator.
3. Remove the oxygen sensor from the sensing port. Let the sensor sit
in room air at least three minutes before you move to the next step.
4Adjust the O2 calibration knob until the O2 (%) display reads 21
percent. (If you can not calibrate to 21 percent replace the cartridge,
see section 6/Maintaining the ventilator.)
5. Use the low-O2 pushwheel to set the low O2 alarm limit to 18 percent.
The ventilator displays: LOW O2 LIMIT = 18%
6. Use the high-O2 pushwheel to set the high O2 alarm limit to 40
percent.
Figure 4-7
Removing the oxygen
sensor from the
absorber
The ventilator displays: HI O2 LIMIT = 40%
AA.15.043
7. Use the low-O2 pushwheel to readjust the low O2 alarm limit to 22
percent.
The ventilator displays: LOW O2 LIMIT = 22%
Within four seconds the ventilator sounds the warble tone, flashes
the red LED, and displays the “low oxygen” message.
The ventilator displays: LOW OXYGEN!
4-51500-0134-00002/01/93
4-5
Page 60
4/Preoperative Checkout Procedure
8. Now use the low-O2 pushwheel to readjust the low O2 alarm limit to
18 percent.
The ventilator displays: LOW O2 LIMIT = 18%
Because you have resolved the alarm condition, the ventilator will
cancel the low O2 alarm within four seconds.
9. Use the high-O2 pushwheel to readjust the high oxygen alarm limit to
20 percent.
The ventilator displays: HI O2 LIMIT = 20%
Within four seconds the ventilator sounds the intermittent tone,
flashes the yellow LED, and displays the “high oxygen” message.
The ventilator displays: HIGH OXYGEN!
10. Use the high-O2 pushwheel to readjust the high oxygen alarm limit to
40 percent.
The ventilator displays: HI O2 LIMIT = 40%
The ventilator cancels the high oxygen alarm within four seconds.
11. Return the oxygen sensor to the sensing port.
12. At least once a month perform the 100 percent calibration procedure
as described in “6/Maintaining the Ventilator.” If you don‘t know
when the sensor was last calibrated, do it now.
Testing the low minute volume, reverse flow, and
apnea alarms
1. Add a breathing bag to the patient circuit at the patient end of the
“Y” connector.
2. Make sure that inspiratory pause is OFF.
3. Set the tidal volume to 500 mL.
The ventilator displays: 500 mL I:E 1:X.X
4. Set the rate to 10 breaths per minute.
The ventilator displays: 10/min I:E 1:X.X
Where X.X represents the current I:E depending upon the present
flow setting.
5. Set the inspiratory flow to 30 liters per minute to change the I:E ratio
to 1:5.0.
The ventilator displays: I:E = 1:5.0
4-61500-0134-00002/01/93
4-6
Page 61
4/Preoperative Checkout Procedure
Figure 4-8
Adding a breathing bag
to the patient circuit
AA.15.093
6. Set the inspiratory limit to 40 cm H2O.
The ventilator displays: PMAX=40SUST=20
7. Use the low VE pushwheel to set the low VE alarm limit to 0.0 liters
per minute.
The ventilator displays: VE LIMIT = 0.0 L/min
8. Set the anesthesia system’s oxygen flow to 2 liters per minute.
9. Ensure that the volume sensor cartridge is in the expiratory limb of
the breathing system. Make sure that the arrows on the sensor clip
point in the direction of exhaled gas flow.
10. Move the absorber’s bag/ventilator switch to “ventilator, if applicable.”
11. Use the anesthesia machine’s oxygen flush button to fill the bellows.
12. Switch ON mechanical ventilation, then wait 40 seconds.
13. Use the low VE pushwheel to readjust the low VE alarm limit to 9.9
liters per minute.
The ventilator displays: VE LIMIT = 9.9 L
Then the ventilator sounds the intermittent tone, flashes the yellow
LED and displays: LOW MINUTE VOL!
14. Use the low VE pushwheel to set the low VE alarm limit to 0.0 liters
per minute.
The ventilator cancels the low minute volume alarm.
15. Remove the volume-sensor clip from the volume-sensor cartridge.
4-71500-0134-00002/01/93
4-7
Page 62
4/Preoperative Checkout Procedure
16. After 30 seconds the ventilator flashes the yellow LED, sounds the
warble tone once, and displays: APNEA 31 SEC.
From this point on the ventilator displays the number of seconds
since apnea was detected.
17. In 30 more seconds the tone warbles twice and the ventilator displays: APNEA 60 SEC.
18. Thirty seconds later the tone warbles three times and the ventilator
displays: APNEA 90 SEC.
19. Wait until the ventilator displays: APNEA 120 SEC.
The tone warbles continuously, the ventilator flashes the red LED,
and the ventilator displays: APNEA **
20. Push the alarm silence button å. The ventilator silences the audio
alarm. The red LED continues to flash
21. Put the volume sensor clip back on the volume cartridge. Make sure
the arrows point in the direction of exhaled gas flow.
22. After one more ventilation cycle the red LED stops flashing and the
ventilator clears the apnea alarm.
23. Use the setup page to verify the reverse flow detection is enabled. If
reverse flow detection is OFF, switch it ON.
24. Remove the volume sensor clip from the volume cartridge.
25. Reinstall the volume sensor clip backwards (so the arrows point in
the opposite direction of the exhaled gas flow) on the volume cartridge.
26. Switch ON mechanical ventilation.
The ventilator sounds a continuous tone, flashes the yellow LED,
and displays: REVERSE FLOW!
27. Correctly reinstall the volume sensor clip back on the volume cartridge. Make sure the arrows point in the direction of exhaled gas
flow.
After two ventilation cycles the ventilator clears the reverse flow
alarm.
Testing the high, low, and sustained pressure
alarms
1. Make sure that inspiratory pause is OFF.
2. Set the tidal volume to 500 mL.
The ventilator displays: 500 mL I:E 1:XX
3. Set the Rate to 10 breaths per minute.
The ventilator displays: 10/min I:E 1:XX
4-81500-0134-00002/01/93
4-8
Page 63
4/Preoperative Checkout Procedure
4. Set the inspiratory flow to 30 liters per minute to change the I:E ratio
to 1:5.0.
The ventilator displays: I:E = 1:5.0
5. Set the inspiratory pressure limit to 40 cm H2O.
The ventilator displays: PMAX=40 SUST=20
6. Set the anesthesia machine’s oxygen flow to 2 liters per minute.
7. Make sure the pressure sensing tube is securely connected between
the control module’s connector marked “Connect to Inspiratory Limb
of Breathing System” and the distal-sensing tee on the inspiratory
side of the breathing system.
8. Move the absorber’s Bag/APL-ventilator switch to “ventilator”, if
applicable.
9. Connect the common gas outlet to the breathing system.
10. Open the breathing system at the “Y” connector.
11. Move the mechanical ventilation switch to ON. After 20 seconds of
mechanical ventilation, the ventilator sounds the warble tone once,
flashes the red LED, and displays the “low pressure” message. The
ventilator displays: LOW PRESSURE!
Figure 4-9
Opening the patient
circuit at the “Y”
connector
∑
Figure 4-10
Occluding the patient
circuit at the “Y”
connector
AA.15.037
12. Move the mechanical ventilation switch to OFF. Within five seconds
the ventilator cancels the low pressure alarm.
13. Occlude the patient end of the “Y” connector.
WARNING: When occluding the breathing system for test purposes,
do not use any object small enough to slip completely into the
system. Objects in the breathing system can interrupt or disrupt
the delivery of breathing-system gases, possibly resulting in injury
to the patient. Before using the breathing system on a patient,
always check the breathing system components for foreign objects.
4-91500-0134-00002/01/93
AA.12.031
4-9
Page 64
4/Preoperative Checkout Procedure
14. Wait for the bellows to inflate. When the bellows are completely
extended, move the mechanical ventilation switch to ON. Within two
seconds the ventilator warbles once, flashes the red LED, and displays: HIGH PRESSURE!
15. Move the mechanical ventilation switch to OFF. Within ten seconds
the ventilator clears the high pressure alarm.
16. Perform the APL valve test, if applicable to your system.
a. Maintain the occlusion in the breathing circuit.
b. Turn the APL valve fully clockwise.
c. Connect a three-liter bag to the bag port.
d. Move the Bag/APL-Ventilator switch to “Bag/APL.” (Maintain
the occlusion in the breathing system.)
e. As you press the anesthesia system’s oxygen flush button, watch
the absorber’s pressure gauge until the system pressure reaches
at least 20 cm H2O. Release the flush button and wait 15 seconds
more. Because the pressure still exceeds 20 cm H2O, the ventilator now sounds the warble tone continuously, lights the red LED,
and displays: SUSTAINED PRES!
∑
f.Slowly open the APL valve to release the pressure in the system.
The ventilator silences the sustained pressure alarm when the
pressure falls below 20 cm H2O and extinguishes the red LED.
17. Remove the occlusion from the “Y” connector.
18. If applicable, set the “Bag/APL-Ventilator” switch to the position you
plan to use.
WARNING: After completing the preoperative checkout procedures,
and before starting any procedure during which this device will be
used, confirm that all hoses, tubing, and other circuit components
are connected properly. Failure to do so may result in patient
injury. Refer to the operation manuals for all devices in the system
to confirm that they are setup and connected correctly
4-101500-0134-00002/01/93
4-10
Page 65
5/Operating the Ventilator
5/Operating the Ventilator
Using the setup page 5-1
Setting the alarm limits 5-3
Setting the ventilation parameters, beginning ventilation 5-5
Responding to alarms 5-8
Using the setup page
In the setup page mode the front panel controls are used to set parameters normally not adjusted during a case, such as
•screen contrast,
•alarm volume.
•reverse flow detection
•sigh on/off
When you first enter the setup page, the ventilator displays
•ventilator model,
•software version number it is using,
•selected supply gas,
•and selected language and current altitude setting in meters.
The ventilator then lets you enable or disable the reverse flow alarm and
sigh function, and adjust the screen contrast and audio alarm volume.
These parameter settings are stored in the ventilator’s memory even
when the system’s power is OFF. To adjust the setup parameters, turn
the flow knob. To move from one parameter display to the next, press
å alarm silence button.
To exit the setup page, repeatedly press the alarm silence button to
move through all the steps; make no changes to any parameters for 30
seconds; or, before you reach the final setup page step, move the mechanical ventilation switch to ON. When the ventilator exits the setup
page, it stores any changes you’ve made, exits the setup page and
begins normal operation. As the ventilator exits the setup page it beeps
once and displays “CHECK SETTINGS.” Although you must press the
inspiratory pause button to begin using the setup page, the ventilator
retains the inspiratory pause function’s setting; you do not have to
manually reset the inspiratory pause function when you are finished
using the setup page.
To use the setup page1. Move the mechanical ventilation switch to OFF.
2. Power ON the control module.
5-11500-0134-00002/01/93
5-1
Page 66
5/Operating the Ventilator
3. Press and continue to hold down the alarm silence button å, then
press the inspiratory pause button. Release both buttons.
The ventilator displays:
321
7800 REV 4.XX /O
ENGLISH 1300 m
∑
45
1.Ventilator Model
2.Software Version
3.Ventilator Drive Gas (A=Air; O=O
4.Language
5.Altitude
WARNING: Pay attention to the model, software revision number,
ventilator supply gas, language and altitude setting shown on the
setup page. If the language displayed is other than the appropriate
language, or if the altitude setting is incorrect, refer to section 2,
“Getting Started, Checking and setting altitude compensation.” If
model is other than “7800,” or if the drive gas abbreviation is
incorrect (“/O” indicates oxygen, “/A” indicates air), have an
Ohmeda-trained service representative reset the ventilator. Other
languages, models, and drive gases have associated operating
parameters that are not described in this manual.
4. Press: å
The ventilator displays:
FLOW KNOB TO SET
REV FLOW ALM ON (or) OFF
1 meter (3.28 feet)
)
2
5. If you are using the volume sensor at the proximal end of the “Y”
connector in the patient breathing system, you may want to disable
the reverse flow alarm, which, in that position, can be triggered by
normal breathing. Do not disable the reverse flow alarm if the sensor
is mounted in the distal position of the expiratory limb; in this position the alarm provides valuable information about possible breathing-system malfunctions. (See “Connecting the volume sensor” in “2/
Getting Started” for information about installing the volume sensor).
To change the status of the reverse flow alarm, turn the flow knob
until the display changes to either ON or OFF. When you are ready to
move to the next step,
Press: å
The ventilator displays:
FLOW KNOB TO SET
SIGH ON (or) OFF
5-21500-0134-00002/01/93
5-2
Page 67
5/Operating the Ventilator
6. When the sigh function is selected, the ventilator delivers 150 percent of the set tidal volume once every 64th breath.
Note:The ventilator can only deliver a maximum 1500 mL per breath.
To change the status of the sigh function, turn the flow knob until
the display changes to either ON or OFF. When you are ready to
move to the next step.
7. Press: å
The ventilator displays:
FLOW KNOB TO SET
CONTRAST: xx
If you want to adjust the LCD screen’s contrast, turn the flow knob.
As you turn the knob, the ventilator will show a one-to-ten scale that
indicates the screen contrast that will be used during normal operation. For best results, set the room lighting to the level that will be
used while you are operating the ventilator. When you are ready to
move to the next step,
8. Press: å
The audio alarm sounds continuously and the ventilator displays:
FLOW KNOB TO SET
AUDIO VOLUME: xx
If you want to adjust the audio alarm’s volume level, turn the flow
knob. The displayed number represents, on a scale of one to ten, the
current volume level, which is also sounding. As you turn the knob,
the ventilator will change the tone’s volume and the displayed
number to indicate the level that will be used during normal operation. When you are ready to move to the next step,
9. Press: å
The ventilator silences the alarm, beeps once, clears the top display
row and displays:
CHECK SETTINGS!
Any changes you selected are now saved and will be implemented
(the inspiratory pause button remains in the state it was in when you
selected the setup page).
Setting the alarm limits
Use the three alarm-set pushwheels to change the low-minute-volume,
low-oxygen and high-oxygen alarms’ set points. To increase the value of
an alarm set point, push the button directly over the digit you want to
change. To decrease the value, push the button under the digit you are
changing. Use the inspiratory pressure limit knob to set the inspiratory
pressure limit and the sustained pressure limit. To adjust these limits,
press the inspiratory pressure limit knob as you turn it.
5-31500-0134-00002/01/93
5-3
Page 68
5/Operating the Ventilator
Anytime you change the value of an alarm set point, the ventilator will
display that alarm’s value on the screen.
To set the alarm limits1. Move the mechanical ventilation switch to OFF before powering ON
the ventilator.
2. Power ON the ventilator, if it is not ON already.
Because the monitoring is active whenever the ventilator’s power is
ON, some alarms may sound.
3. Use the low-VE pushwheel to set the low minute volume alarm limit.
The low minute volume alarm limits are 0.1 liters per minute to 9.9
liters per minute in 0.1 liter per minute increments. Setting 0.0 on the
pushwheel disables low minute volume alarms.
WARNING: Always correctly set the low minute volume alarm and
∑
use CO2 monitoring to aid in the detection of breathing system
disconnections.
4. Use the low-O2 pushwheel to set the low oxygen alarm limit. The low
oxygen alarm limits are 18 percent to 99 percent in one percent
increments.
Figure 5-1
Alarm push wheels and
inspiratory pressure
limit knob on a standalone ventilator.
Controls are the same
for an Excel system
ventilator.
Although all the digits can be physically set to zero, the ventilator will
not accept low oxygen alarm limits of less than 18 percent.
(25% T )
Ohmeda7800 Ventilator
300
200
500
100
50
1000
1500
mL
1.Alarm pushwheels
2.Inspiratory limit knob
L
20
40
B/min
60
80
100
10
2
T
Low v
+ ++ ++ +
60 21 00
- -- -- -
L/min % %
L/min
I
IP
Low O
High O
E
2
2
.
60
PUSH
TO
TURN
cm H O
80
10010100
2
40
20
p
q
1
2
AA.32.033
If you set the low oxygen alarm pushwheel to less than 18 percent, the
ventilator will continue to use 18 percent for the low oxygen alarm’s set
point; and the ventilator will display a
“LIMIT SET ERROR” message.
5. Use the high-O2 pushwheel to set the high oxygen alarm’s limit. The
high oxygen alarm’s limits are 18 percent to 99 percent in one percent increments.
5-41500-0134-00002/01/93
5-4
Page 69
5/Operating the Ventilator
If you set the high-O2 pushwheel to a level lower than or equal to the
limit set by the low-O2 pushwheel, the ventilator will display a
“LIMIT SET ERROR” message.
To disable the high oxygen alarm, set the high-O2 pushwheel to “00”;
setting the high-O2 pushwheel to “00” will not generate a
“LIMIT SET ERROR” message.
6. Use the inspiratory pressure limit knob to set the inspiratory pressure
limit and the sustained pressure limit. If the sigh function is enabled,
be sure to compensate for the additional pressure that will occur
during the sigh breath.
Both the maximum inspiratory pressure and the sustained pressure
alarm limits and pressure-release points are set by the inspiratory pressure limit knob, which must be pushed in while turned to change the
settings; the ventilator sets the sustained pressure limit to correspond to
the inspiratory pressure limit knob setting. For inspiratory pressure limits
of 20 cm H2O to 60 cm H2O, the ventilator sets the sustained pressure
limit to one half the inspiratory pressure limit. For example, if the inspiratory pressure limit is 42 cm H2O, the sustained pressure limit will be
21 cm H2O. However, any inspiratory pressure limit setting higher than
60 cm H2O will result in a sustained pressure limit of 30 cm H2O. For
example, inspiratory pressure limits of 65 cm H2O and 80 cm H2O will
both result in a sustained pressure limit of 30 cm H2O.
As you push and turn the inspiratory pressure limit knob, the ventilator
will display both the maximum pressure limit and the sustained pressure
limit settings. However, unlike the other three control knobs, just touching this knob will not generate a display.
If while the mechanical ventilation switch is OFF you set the inspired
pressure limit for more than 60 cm H2O, the ventilator will beep, light the
yellow LED, and continually display the maximum pressure setting. This
pressure limit message is displayed in the non-mechanical ventilation
mode only; during mechanical ventilation this reminder is disabled.
In addition to the three alarms set by the pushwheels and the two
alarms set by the inspiratory pressure limit knob, the ventilator also sets
trigger points for certain alarms based on the positions of other front
panel controls. These alarms include the low pressure alarm, the apnea
alarm, and the reverse flow alarm.
WARNING: The reverse flow alarm only works when the volume
∑
sensor cartridge is mounted at the distal end of the expiratory limb,
reverse flow is ON and the volume sensor is working properly.
Setting the ventilation parameters, beginning
ventilation
Set the ventilation parameters before moving the mechanical ventilation
switch to ON. Because we recommend that you set the ventilator’s
controls starting on the left and moving to the right, these instructions
describe setting the ventilator’s controls from left to right. If you adjust
either the tidal volume knob or rate knob after you have set the inspiratory flow knob, you may need to readjust the inspiratory flow to maintain
the desired I:E ratio.
5-51500-0134-00002/01/93
5-5
Page 70
5/Operating the Ventilator
It is possible to adjust the controls for a combination of settings that will
result in a level the ventilator cannot deliver. If the combination of
settings results in a level the ventilator is not designed to deliver, a “vent
set error” alarm will be displayed. (See “Control range computation” in
“3/General Information” for a description of the range of control settings.)
The measured tidal volume indicated on the screen may be different than
the level you set on the tidal volume knob. Under pressure gases compress and certain breathing system components expand, which results in
some loss of tidal volume in the breathing system. Also, any fresh gas
introduced to the system will be measured by the volume sensor in
addition to the gas the bellows assembly delivers. For instructions on
calculating these breathing system losses and gains, see “Tidal volume
compensation” in “3/General Information.”
You don’t, however, have to manually calculate the compliance effect to
compensate for compliance losses. With the volume sensor correctly
connected in the breathing system, adjust the tidal volume knob until the
tidal-volume reading on the ventilator’s screen indicates the level you
want to use.
To set the ventilation
parameters and begin
mechanical ventilation
1. Before connecting the ventilator to a patient, perform the preoperative checkout procedures described in “4/Preoperative Checkout
Procedures.”
2. Move the mechanical ventilation switch to OFF.
3. Power ON the control module, if it is not ON already.
The ventilator enters the non-mechanical ventilation mode. Because
the monitoring is active whenever the ventilator’s power is ON, some
alarms may sound. To silence any alarms, press å.
4. Set the alarm limits. Be sure to set the ventilation controls before
moving the mechanical ventilation switch to ON.
5. Use the tidal volume knob to set the tidal volume.
The tidal volume knob lets you set the tidal volume from 50 mL to
1500 mL. As you turn the knob, the ventilator will display the tidal
volume setting. To check the tidal volume setting without changing
its value, just touch the knob; the ventilator will then display the
current tidal volume setting. Since changing the tidal volume may
change the I:E ratio, the I:E ratio is displayed along with the tidal
volume as a reminder. If the resulting I:E ratio is not as desired, be
sure to readjust the inspiratory flow to maintain the desired I:E ratio.
Until the ventilator senses sufficient volume to indicate a breath, it
will display “VOL MON STANDBY.” Once the ventilator senses a
sufficient volume level, it will remove the “VOL MON STANDBY”
message and start the timer that controls the apnea alarm.
If, however, while the mechanical ventilation switch is OFF, you set
the tidal volume to less than 300 mL, the ventilator will disable the
apnea alarm. Once the ventilator removes the “VOL MON STANDBY”
message—after an activation breath—the ventilator will display the
“APNEA ALARM OFF” message. If you want the apnea alarm enabled
when the ventilator is in the non-mechanical ventilation mode , set
the tidal volume knob to 300 mL or higher.
5-61500-0134-00002/01/93
5-6
Page 71
5/Operating the Ventilator
Note:When switching from mechanical to non-mechanical ventilation mode
and the VOL MON STANDBY is displayed, the apnea timer does not
begin until an initial breath has been sensed.
6. Use the rate knob to set the mechanical breath rate.
Turning the rate knob changes the breath rate used for mechanical
ventilation and displays the rate. The rate’s range is 2 breaths per
minute to 100 breaths per minute in whole number increments.
Touching the rate knob will display the current rate on the screen.
Since changing the breath rate may change the I:E ratio, the I:E ratio
is displayed along with the breath rate as a reminder. If the resulting
I:E ratio is not as desired, be sure to readjust the inspiratory flow to
maintain the desired I:E ratio.
7. If you want to enable the inspiratory pause function, press the
inspiratory pause key.
8. Set the desired I:E ratio by adjusting the inspiratory flow rate.
The inspiratory flow knob lets you set the inspiratory flow rate,
which is continuously variable from 10 liters per minute to 100 liters
per minute. Whenever you adjust or just touch the inspiratory flow
knob, the ventilator will display the current I:E ratio, which the
ventilator calculates based on the set inspiratory flow, tidal volume,
inspiratory pause status, and breath rate. Adjusting any of these
parameters will change the I:E ratio. However, you should use the
inspiratory flow knob to adjust the I:E ratio once the tidal volume and
breath rate have been correctly set. To check the current I:E ratio,
touch the inspiratory flow knob.
9. Move the absorber’s Bag/APL-Ventilator switch to “ventilator”, if
applicable.
10. Use the anesthesia system’s oxygen flush valve to fill the bellows.
Set the oxygen flow to a level that keeps the bellows fully extended.
11. To start mechanical ventilation, move the mechanical ventilation
switch to ON.
The mechanical ventilation switch controls mechanical ventilation
only. When the switch is OFF, the monitors still function and the
alarm system is still active. When you want to start mechanical
ventilation, move the switch to ON.
Always turn ON the anesthesia system and set the control module’s
front panel controls before switching ON ventilation. Starting mechanical ventilation before setting the controls may result in inappropriate ventilation of the patient and may trip alarms that relate to
mechanical ventilation.
12. Once the ventilator is mechanically ventilating the patient, check the
tidal volume. (See “3/General Information” for information about tidal
volume compensation.) If necessary, adjust the front panel controls to
modify the ventilator’s operating parameters. You can adjust any of
the front panel controls while the ventilator is ON, see the preceding
parameter setting steps.
5-71500-0134-00002/01/93
5-7
Page 72
5/Operating the Ventilator
Responding to alarms
For detailed descriptions of the ventilator’s alarms, see “The alarm
system” in “3/General Information.”
WARNING: Always respond to alarms promptly. Failure to respond to
∑
APNEA xxx SECIf for 30 seconds the ventilator doesn’t detect enough exhaled volume in
∑
Note:If a sigh breath occurs when a breath rate setting of two breaths per
alarms may result in injury to the patient.
the breathing system, an apnea alarm will be generated. The alarm
message will indicate the number of seconds that have passed since
sufficient flow was last detected.
WARNING: If you remove the sensor clip from the volume sensor
cartridge before switching ON the anesthesia system, the apnea
alarms will be inoperative. Do not use the ventilator without the
sensor clip properly attached to the volume sensor cartridge.
minute is set, the apnea alarm will be triggered during the sigh breath.
1. Check the patient.
2. Check for disconnections in the patient breathing system.
3. Check for excessive moisture in the volume sensor cartridge.
4. If so equipped, check for excessive moisture in the absorber’s check
valves.
5. Make sure the volume sensor clip is connected securely to the volume sensor cartridge.
6. Make sure the arrows on the volume sensor clip point toward the
direction of exhaled gas flow.
7. Make sure the volume sensor clip is plugged into the sensor interface.
8. Replace the volume sensor cartridge. It may be worn out if its rotor is
turning too slowly or not at all. (See “Checking the volume sensor” in
“6/Maintaining the Ventilator.”)
9. Replace the volume sensor clip.
APNEA ALARM OFF!Whenever tidal volume is set to less than 300 mL and the mechanical
ventilator switch is OFF, the ventilator, once it has detected a breath,
will display “APNEA ALARM OFF.” To enable the apnea alarm system
while the patient is breathing spontaneously, increase the tidal volume
knob to 300 mL or more.
1. The “APNEA ALARM OFF” message is normal if the tidal volume knob
is set below 300 mL and the mechanical ventilation switch is set to
OFF.
CHECK O2 SENSOR!/
CHECK GAS SUPPLY
5-81500-0134-00002/01/93
If the ventilator doesn’t detect at least five percent oxygen, it will assume that the oxygen sensor has failed, and generate an alarm. An alarm
will also be generated if the sensr isn’t connected correctly, if the sensor
is broken, or if no oxygen is in the area of the sensor.
5-8
Page 73
5/Operating the Ventilator
1. Check the oxygen supply.
2. Make sure the oxygen sensor is plugged into the anesthesia system’s
sensor interface.
3. Check the oxygen-sensor cartridge’s surface for excessive moisture.
4. Has the oxygen sensor been left unplugged from the sensor interface? If it has, see “Maintaining, Replacing and Calibrating the
Oxygen Sensor” in 6/Maintaining the Ventilator.
5. Replace the oxygen-sensor cartridge. It may be worn out. (See “Replacing the oxygen sensor” in “6/Maintaining the Ventilator.”)
DRIVE CKT OPEN!One type of ventilator failure—exhalation valve failure—does not display
a numbered message; instead “DRIVE CKT OPEN” is displayed. This
message can also appear if, during mechanical ventilation, the absorber’s
Bag/APL-Ventilation switch is in the “Bag/APL” position. During this
alarm, the ventilator will attempt to continue monitoring and mechanical
ventilation.
1. Check the patient.
2. If mechanical ventilation is ON, make sure the absorber’s Bag/APLVentilation switch is in the “ventilator” position, if applicable.
HARDWARE ERROR XThis display (error A, B, and C) and alarm should not occur unless there
is a problem with the ventilator control module hardware. The alarm is
silenceable with the alarm silence å button. Mechanical ventilation
does not turn OFF.
∑
HIGH OXYGEN!If the ventilator detects an oxygen concentration equal to or higher than
HIGH PRESSURE!If the ventilator detects airway pressure higher than the limit you set
WARNING: Do not use the ventilator if this display and alarm occur.
the one you set using the high-O2 pushwheel, the ventilator will generate
a high oxygen alarm.
1. Check the high-O2 alarm limit. Is it set correctly?
2. Check the anesthesia system settings.
3. Has the oxygen sensor been left unplugged from the sensor interface? If it has, see “Maintaining, Replacing and Calibrating the
Oxygen Sensor” in 6/Maintaining the Ventilator.
using the inspiratory pressure limit knob, the ventilator will generate a
high pressure alarm. And, during mechanical ventilation only, the ventilator will also terminate inspiration.
1. Check the patient.
2. Check for a blockage in the patient breathing system.
3. Check the inspiratory pressure limit knob. Is it set correctly?
4. Check for moisture in the pressure sensing tube that connects the
breathing system to the ventilator’s control module.
5. Check for kinks in the pressure sensing tube.
5-91500-0134-00002/01/93
5-9
Page 74
5/Operating the Ventilator
LIMIT SET ERROR!If you attempt to set the high oxygen alarm limit for a level below or
equal to the low oxygen limit, the ventilator will generate a limit setting
error alarm. This alarm will also be generated if you attempt to set the
low oxygen alarm limit for less than 18 percent.
Reset the alarm limits to acceptable values.
LOW BATTERY!If the battery’s voltage drops to an unacceptable level, the ventilator will
generate a “Low Battery” alarm.
Leave the ventilator powered ON and connected to ~ power for at least
24 hours to recharge the battery.
LOW MINUTE VOL!If the ventilator senses that the minute volume is less than the level you
set using the low-VE pushwheel, the ventilator will generate an alarm.
1. Check the patient.
2. Check the low-VE alarm limit. Is it set correctly?
3. Check for breathing tube disconnections.
4. Check for excessive moisture in the volume sensor cartridge.
5. Check for excessive moisture in the absorber’s check valves.
6. Make sure the volume sensor clip is connected securely to the volume sensor cartridge.
7. Make sure the arrows on the volume sensor clip point in the direction
of exhaled gas flow.
8. Make sure the volume sensor clip is plugged into the sensor interface.
9. Replace the volume sensor cartridge. It may be worn out if its rotor is
turning too slowly or not at all. (See “Checking the volume sensor” in
“6/Maintaining the Ventilator.”)
10. Replace the volume sensor clip.
LOW OXYGEN!If the ventilator detects an oxygen concentration lower than the one you
set using the low-O2 pushwheel, the ventilator will generate a low
oxygen alarm.
1. Check the patient.
2. Check the anesthesia system’s flowmeter settings. Are they set
correctly?
3. Check the anesthesia system’s pressure gauges.
4. Check the low-O2 alarm limit. Is it set correctly?
5. Check the oxygen supply.
6. Check the oxygen sensor assembly for damage.
7. Make sure the oxygen sensor is inserted securely into the oxygen
sensing port.
5-101500-0134-00002/01/93
5-10
Page 75
5/Operating the Ventilator
8. Make sure the oxygen sensor is plugged into the sensor interface.
9. Check the oxygen-sensor cartridge’s surface for excessive moisture.
10. Has the oxygen sensor been left unplugged from the sensor interface? If it has, see “Maintaining, Replacing and Calibrating the
Oxygen Sensor” in 6/Maintaining the Ventilator.
11. Replace the oxygen-sensor cartridge. It may be worn out. (See “Replacing the oxygen sensor cartridge” in “6/Maintaining the Ventilator.”)
LOW PRESSURE!The ventilator will generate a low pressure alarm if, for at least 20
seconds, the airway pressure fails to change by a value that varies
proportionally to the setting of the inspiratory flow knob. The low pressure alarm is active only when mechanical ventilation is switched ON.
1. Check the patient.
2. Check the breathing system for leaks or disconnections.
3. Check for moisture in the pressure sensing tube that connects the
breathing system to the ventilator’s control module.
4. Check for kinks in the pressure sensing tube.
5. Is the patient breathing spontaneously? During mechanical ventilation spontaneous breathing may trip this alarm.
LOW SUPPLY PRES!If the ventilator’s regulated supply gas pressure is less than 22 psig, (155
kPa) the ventilator will generate a low supply pressure alarm. Low
supply pressure will reduce delivered tidal volumes during mechanical
ventilation.
1. Check the supply pressure.
2. Switch to cylinder use, if applicable.
3. Switch to manual ventilation, if necessary.
MAX PRES =xxx cmAnytime you adjust the inspiratory pressure limit knob, the ventilator
will briefly display the pressure limit in centimeters of water. If, however,
the ventilator is in the non-mechanical ventilation mode and the inspiratory pressure limit is set to 60 cm or more, the ventilator will also light
the yellow alarm LED and will continually display the “MAX PRES=XXX
CM” message. This pressure limit reminder is displayed in the non-
mechanical ventilation mode only.
O2 CAL ERROR!If the oxygen sensor delivers a signal that is out of the ventilator’s range,
the ventilator will generate an oxygen-calibration-error alarm.
•Calibrate the oxygen sensor as described in “Calibrating the oxygen
sensor” in “6/Maintaining the Ventilator.”
POWER FAIL!If ~ power to the ventilator is lost, the ventilator will generate a power-
failure alarm.
1. Continue normal operation; a fully charged backup battery will allow
you to continue for about 20 minutes with mechanical ventilation. To
extend the monitoring time, discontinue mechanical ventilation, and
manually ventilate the patient.
5-111500-0134-00002/01/93
5-11
Page 76
5/Operating the Ventilator
2. Make sure the system power cord has not been disconnected.
3. Resolve the cause of the power failure.
REVERSE FLOW!If the ventilator senses reverse flow of an unacceptable volume, it will
generate a “Reverse Flow” alarm. For tidal volumes less than 300 mL,
the reverse flow limit is 20 mL. For tidal volumes greater than 300 mL,
this limit is 100 mL.
1. Make sure the volume sensor assembly is in the expiratory limb of
the patient breathing system. If the sensor is in the expiratory limb,
check the exhalation valve; its disk may be sticking.
2. Make sure the arrows on the volume sensor clip point in the direction
of exhaled gas flow.
REV FLOW ALM OFFTo advise you that the reverse flow alarm is disabled, in the non-me-
chanical ventilation mode the ventilator will repeatedly flash this message. If you power ON the control module with the alarm disabled and
mechanical ventilation ON (although you should always have the mechanical ventilation switch set to OFF when you power ON), the message will be displayed once.
•If you want the reverse flow alarm enabled, use the setup page to set
the reverse flow alarm status to ON.
SOFTWARE ERROR XThis display (error A through F) and alarm should not occur unless there
is a problem with the ventilator software programs. The alarm is
silenceable with the alarm silence å button. Mechanical ventilation
does not turn OFF.
∑
SUB-ATMOS PRES!If the ventilator detects airway pressure of less than minus (-)10 centime-
WARNING: Do not use the ventilator if this display and alarm occur.
ters of water, it will generate a sub atmospheric pressure alarm.
1. Check patient.
2. Check for inadvertent vacuum hook-ups to the patient breathing
system.
3. Check for kinks or occlusions in the breathing system.
4. The inspiratory check valve in the absorber may be stuck. Check the
inspiratory check valve.
5. Is the patient breathing spontaneously? Spontaneous breathing may
trip this alarm.
6. Check the gas-scavenging system for excessive vacuum.
7. Check for moisture in the pressure sensing tube that connects the
breathing system to the ventilator’s control module.
8. Check for kinks in the pressure sensing tube.
SUSTAINED PRES!Anytime the sustained airway pressure exceeds—for 15 seconds or
more—the sustained pressure limit set by the inspiratory pressure limit
knob, the ventilator will generate a sustained pressure alarm.
5-121500-0134-00002/01/93
5-12
Page 77
5/Operating the Ventilator
1. Check the patient.
2. Check for kinks or blockages in the breathing system.
3. Check to make sure the absorber’s Bag/APL-Ventilator valve is in the
correct position.
4. Check for moisture in the pressure sensing tube that connects the
breathing system to the ventilator’s control module.
5. Check for kinks in the pressure sensing tube.
VENT FAIL xx!The number in the ventilator failure message, such as “VENT FAIL 8,”
corresponds to the specific type of failure occurring. Vent fail messages
require that you take the ventilator out of service and call a qualified,
trained Ohmeda Service Representative. If a ventilator failure alarm
occurs, pressing the alarm silence button permanently silences the alarm
tone, although the yellow LED and alarm message remains ON, and the
ventilator may not function.
∑
∑
WARNING: Do not use the ventilator if this display and alarm occur.
Some ventilator failure alarm conditions may be caused by transitory
electrical interference that devices such as electrocautery instruments
can generate. Although the ventilator will disable mechanical ventilation
during most ventilator failure alarms, during certain of these alarms,
whose causes may be transitory, the ventilator will attempt to maintain
mechanical ventilation.
See the “7/Service Procedures” for descriptions of specific ventilator
failure messages.
WARNING: The use of electrosurgical units or other devices that
radiate high-intensity electrical fields can affect the operation of the
ventilator and monitors attached to the patient. Maintain as much
distance as possible between the electrosurgical leads and the cables
to the flow and oxygen sensors. Do not drape the electrosurgical leads
across the absorber or the anesthesia machine. Do not let the
electrosurgical leads rest on any surface of the anesthesia system.
Constant surveillance of all monitoring and life support equipment is
mandatory whenever electrosurgical devices are in operation, on, or in
the vicinity of the patient.
WARNING: Ventilator failure messages indicate that a problem exists
in the ventilator. Do not attempt to use the ventilator while a ventilator failure message is displayed.
IMPORTANT
If the ventilator experiences extreme electrical interference, it may
interrupt mechanical ventilation. If this interruption occurs, the ventilator
generates an internal reset function and resumes normal operation after
two (2) seconds. For situations where continuous electrical interference
is experienced by the ventilator, causing a continuous interruption, the
ventilator’s internal reset repeats until the interference ceases.
If the electrical interference is continuously present and mechanical
ventilation is interrupted for approximately 30 seconds, the ventilator
produces a continuous beeping audio alarm. Manual ventilation of the
patient must be performed while the mechanical ventilation is interrupted. When the electrical interference ceases, the continuous beeping
5-131500-0134-00002/01/93
5-13
Page 78
5/Operating the Ventilator
audio alarm can be silenced only by turning, as applicable, the ventilator
or anesthesia machine power switch OFF and after five seconds back
ON.
WARNING: Manual ventilation must be performed when electrical
∑
VENT SET ERROR!If you attempt to set a combination of the inspiratory-flow, tidal-volume,
interference causes interruption of ventilator delivered mechanical
ventilation. Manual ventilation must be continued until the ventilator
resumes normal operation or an alternate ventilator/anesthesia system can be used.
1. Manually ventilate the patient.
2. Have the ventilator checked by qualified service personnel.
Refer to “7/Service Procedures” for more information about ventilator
failure messages.
inspiratory pause, and rate controls that results in a level the ventilator is
not designed to deliver, the ventilator will continue to use the most
recent acceptable settings, and will generate a ventilator setting error
alarm until the control combination is corrected.
1. Readjust the ventilator’s controls within the ventilator’s operating
limits.
WARNING: Do not use the ventilator while the “VENT SET ER-
∑
VOL MON STANDBY!Volume monitor standby indicates that, during mechanical or non-
VOL SENSOR FAIL!A volume sensor failure alarm will be displayed if the volume sensor’s
ROR” message is displayed; when this message is displayed, the
control settings do not reflect the settings the ventilator is using. If
the “VENT SET ERROR” message is displayed during mechanical
ventilation, the system will use the most recent acceptable settings. If, however, the ventilator is powered ON in the “VENT SET
ERROR” condition, moving the ventilation switch to ON will not
start mechanical ventilation until the controls are moved to an
acceptable setting.
mechanical ventilation, either the ventilator has not sensed sufficient
breaths to check for apnea and volume alarm conditions, or that, while
the mechanical ventilation switch is OFF, the alarm silence button has
been used to cancel and reset the apnea and low minute volume alarms.
If these two alarms have been canceled, they will not sound again even if
these alarm conditions continue. However, if the ventilator senses
another breath, the alarm timers and sensor circuits will again be activated and any alarm condition that occurs will trigger an appropriate
alarm.
heater voltage is too low, which can happen if the volume sensor clip is
broken or disconnected.
1. Check the connection between the volume sensor cartridge and the
sensor clip.
2. Make sure the sensor clip is plugged into the sensor interface.
3. Replace the volume sensor clip.
- - - - - - - -displayed in
place of readings
5-141500-0134-00002/01/93
If the system doesn’t measure any volume during mechanical ventilation,
it will display dashes in place of the volume and rate data.
5-14
Page 79
5/Operating the Ventilator
1. Check the patient.
2. Check the breathing system for disconnections.
3. Check for obstructions in the volume-sensor cartridge that may be
preventing the cartridge vanes from spinning.
4. Replace the volume sensor cartridge.
5. Replace the volume sensor clip.
???????? is displayed in place of VE ,
, or rate readings
V
T
If the alarm silence
button will not silence
the alarm
∑
Certain combinations of ventilator front panel control settings can result
in ventilation conditions the ventilator can deliver but the volume sensor
cannot measure. If you set a control combination the ventilator can
deliver, but the volume monitor cannot measure, or if—in the nonmechanical ventilation mode —breathing occurs that the monitor cannot
measure, the ventilator will display question marks in place of the VT, VE,
and rate readings.
Certain electronic failures may be so significant as to cause the system to
lose the ability to generate ventilator failure messages even though the
ventilator has failed. If such a serious failure occurs, the alarm silence
button will not silence the alarm. Do not attempt to use the ventilator if
this type of failure occurs.
WARNING: Do not attempt to use the ventilator if the alarm silence
button å will not silence alarms.
1. Ventilate the patient manually.
2. Switch to a functioning system.
5-151500-0134-00002/01/93
5-15
Page 80
Notes
5/Operating the Ventilator
5-161500-0134-00002/01/93
5-16
Page 81
6/Maintaining the Ventilator
6/Maintaining the Ventilator
Maintenance schedule 6-1
Long Term Ventilator Storage 6-1
Cleaning and sterilizing 6-2
Cleaning the control module 6-2
Cleaning and sterilizing the bellows assembly 6-2
Cleaning and sterilizing the volume sensor clip assembly 6-3
Cleaning and sterilizing the volume sensor cartridge 6-3
Checking the volume sensor 6-4
O2 sensor maintenance 6-5
Maintenance schedule 6-5
Installing a cartridge or disassembling the O2 sensor for cleaning 6-5
Cleaning and sterilization 6-7
100% O2 calibration 6-8
Cleaning the supply gas filter 6-9
Operation 6-9
Filter servicing 6-9
Maintenance schedule
w
The following schedule is a recommended minimum standard that is
based on normal usage and environmental conditions. Higher frequencies
of use or unusual environments may dictate more frequent maintenance.
CAUTION: No repair should ever be undertaken or attempted by anyone
not having proper qualifications and equipment.
Replace damaged parts with components manufactured or sold by
Ohmeda. Then test the unit to ascertain that it complies with the
manufacturer’s published specifications.
Full checkout and maintenance byEvery three months
trained service personnel
Calibrate oxygen sensorDaily
Replace oxygen sensorAs required at minimum yearly
Clean bellows assemblyAs required
Clean or replace pressure sensing tubeAs required
Replace volume sensor cartridgeAs required at minimum
monthly
Replace air supply gas filter elementEvery six months or as
required by filter discoloration
and flow resistance
Long Term Ventilator Storage
It is not necessary to disconnect the rechargeable batteries before long
term storage.
If the ventilator is to be stored for an extended period of time, the batteries will eventually discharge. This is not destructive to the ventilator or
the batteries, however, the ventilator must be plugged into a wall outlet
for at least 24 hours prior to use to fully recharge the batteries.
6-11500-0134-00002/01/93
6-1
Page 82
6/Maintaining the Ventilator
Cleaning and sterilizing
This chart is provided for quick reference once you are familiar with the
cleaning and sterilization of the system. Refer to the following sections
for cleaning-and-sterilization details. Use a cleaning and sterilization
schedule that conforms to your institution’s sterilization and risk-management policies.
ItemTo cleanTo sterilize
Control modulemild detergentn/a
Bellows assembly see section 8 see section 8 or 9
Air supply gas filterPlastic bowl warmn/a
water only
Other parts warmn/a
water and soap
Volume sensordamp clothdisinfectant
assembly
Volume sensordamp clothethylene oxide or liquid
cartridgesterilizing agent
(pH less than 9),oxide
cold germicidal alkali
detergent
Cleaning the control module
Clean the exterior surfaces of the control module with a soft, lint-free
cloth lightly moistened in a solution of mild liquid detergent.
CAUTION: Use cleaning solution sparingly. Do not saturate system
components. Excessive solution can damage internal devices.
CAUTION: Do not cover the system with any type of fabric or plastic
covering. These types of coverings can generate static charges that
may damage the equipment.
Cleaning and sterilizing the bellows assembly
See section 8 or 9 for disassembly, cleaning, sterilization, reassembly and
checkout of the Ohmeda Bellows Assembly
6-21500-0134-00002/01/93
6-2
Page 83
6/Maintaining the Ventilator
Cleaning and sterilizing the volume sensor clip
assembly
Because no part of the sensor assembly—which includes the clip, cable
and plug—is exposed to the breathing system, usually no sterilization is
required. If the clip, cable and plug do need cleaning:
1. Unplug the sensor from the sensor interface panel.
2. Remove the sensor clip from the volume cartridge.
3. Wipe the clip, cable and plug with a cloth moistened in disinfectant
(liquid sterilizing).
CAUTION: Never immerse any part of the volume sensor assembly
w
Figure 6-1
Disconnecting the
volume sensor
cartridge from its clip
in cleaning solution. Immersion will destroy the clip’s electrical
contacts.
Cleaning and sterilizing the volume sensor
cartridge
Replace the sensor cartridge at least every thirty days, or when the
volume sensor checkout (see “Checking the volume sensor”) indicates
the sensor has become inaccurate. If cleaning is required between
replacements:
Note:Be very careful while you are handling the volume cartridge. The car-
tridge is a precision device containing jeweled bearings. Do not drop the
cartridge. Do not allow any contaminant, such as hair or dust, to enter
the cartridge.
1. Remove the cartridge from the breathing system.
2. Unsnap the sensor clip from the cartridge.
MD.10.016
3. Use an accepted gas or liquid sterilization technique to sterilize the
sensor cartridge.
CAUTION: Never insert cleaning brushes or other foreign objects
w
6-31500-0134-00002/01/93
through the cartridge vanes. Contacting the sensor’s moving vane
may damage its precision movement.
6-3
Page 84
6/Maintaining the Ventilator
CAUTION: Following ethylene oxide sterilization, quarantine the
w
Checking the volume sensor
To check volume sensing1. Add a bag to the patient circuit at the “Y” connector.
equipment in a well-ventilated area to allow dissipation of absorbed
ethylene-oxide gas. In some cases, aeration periods of seven days or
more may be required. Aeration time can be decreased when special
aeration devices are used. Follow the sterilizer manufacturer’s recommendations for aeration periods required.
CAUTION: Always perform the preoperative checkout procedures for
volume sensing functions after cleaning or replacing the volume
sensor cartridge.
This volume-sensing checkout procedure, which you should perform
before each case, tests both the sensor cartridge and the sensor assembly. If the checkout fails, replace the volume cartridge, then repeat the
procedure. If the checkout still fails after you replace the cartridge,
replace the sensor assembly and repeat the test again.
2. Set the tidal volume to 500 mL.
•The ventilator displays: 500 mL I:E 1:XX
3. Set the rate to 10 breaths per minute.
•The ventilator displays: 10/min I:E 1:XX
4. Using the inspiratory flow knob, set the I:E ratio to 5.0. This corresponds to a flow of 30 liters per minute.
•The ventilator displays: I:E=1:5.0
5. Set the inspiratory pressure limit to 40 cm H2O.
•The ventilator displays: PMAX=40 SUST=20
6. Make sure that the ventilator’s inspiratory pause function is OFF.
7. Set the low VE alarm limit to 0.0 liters per minute.
8. Set the anesthesia system’s oxygen flow to 2 liters per minute.
9. Ensure that the volume sensor cartridge is on the exhalation side of
the breathing system. Make sure that the arrows on the sensor clip
point in the direction of exhaled gas flow.
10. Move the absorber’s bag-to-ventilator switch to the ventilator position, if applicable.
11. Use the anesthesia machine’s O2 flush button to fill the bellows.
12. Switch ON mechanical ventilation and wait 40 seconds. The displayed VE should be between 3.3 liters and 4.3 liters assuming that
your system:
•includes an adult bellows
•includes 60-inch patient tubes
6-41500-0134-00002/01/93
6-4
Page 85
6/Maintaining the Ventilator
•does not include a humidifier
•has a peak inspiratory pressure of 15 cm H2O
(If your system’s components and peak inspiratory pressure are
different than this, see “Tidal volume compensation” in “3/General
Information” to calculate the compliance factor of your system. Then
use this factor to calculate the expected minute volume.)
Figure 6-2
Adding a breathing bag
at the patient circuit’s
“Y” connector
AA.15.093
O2 sensor maintenance
Maintenance schedule
Before each use•Preoperative checkout procedure (includes 21% O
Monthly•100% O
Annually•Replace the O
at 50% O2 and 24°C (77°F) . Different operating conditions (higher
concentrations, high temperature, and elevated CO2 concentrations)
can shorten cartridge life expectancy. Freezing can destroy the
sensor cartridge
calibration
2
sensor cartridge. Cartridge life expectancy is one year
2
calibration
2
Installing a cartridge or disassembling the O2 sensor for
cleaning
WARNING: Use protective gloves and eye-wear when you open the O
∑
w
sensor in case the cartridge is leaking. The sensor cartridge contains
potassium hydroxide (caustic).
CAUTION: Handle the cartridge with care to avoid damaging it. Always perform the calibration and preoperative checkout procedures for
oxygen-sensing functions after replacing a new sensor cartridge or a
recently cleaned and sterilized oxygen sensor.
2
6-51500-0134-00002/01/93
6-5
Page 86
6/Maintaining the Ventilator
Figure 6-3
Open the sensor
housing and remove
the old cartridge
1.Sensor screen
2.Probe section of housing
Figure 6-4
Remove the metal disk
or clip
1
12
MD.02.055,56; OT.01.022
2
Figure 6-5
Install a new cartridge
with circular contacts
toward the cable end of
the housing, screen
facing out
MD.12.011
1.Shorting disk
2.Shorting clip
MD.02.032
123456
1.Cable section
2.Contact rings
3.Sensor cartridge
4.Inner O-ring
5.Outer O-ring
6.Probe section
6-61500-0134-00002/01/93
6-6
Page 87
6/Maintaining the Ventilator
Figure 6-6
Finger tighten the
housing to a gas tight
seal
MD.02.055AA.32.036
Figure 6-7
Immediately connect
sensor
the O
2
∑
(25% T )
Ohmeda7800 Ventilator
300
200
100
50
mL
20
500
10
1000
2
1500
B/min
L
40
60
80
100
T
IP
Low v
Low O
E
+ + + + + +
60 21 00
.
- -- -- -
L/min % %
L/min
I
High O
2
2
60
40
PUSH
TO
TURN
208010010100
cm H O
2
p
q
Vol
O
2
Monitors
Cleaning and sterilization
WARNING: Use protective gloves and eye-wear when you open the O
sensor in case the cartridge is leaking. The sensor cartridge contains
potassium hydroxide (caustic).
2
WARNING: Do not inhale any fumes generated by the oxygen-sensor
cleaning procedure. Such fumes can cause respiratory system or skin
damage. This material is caustic.
CAUTION: Following ethylene oxide sterilization, quarantine the
w
equipment in a well-ventilated area to allow dissipation of absorbed
ethylene-oxide gas. In some cases, aeration periods of seven days or
more may be required. Aeration time can be decreased when special
aeration devices are used. Follow the sterilizer manufacturer’s recommendations for aeration periods required.
6-71500-0134-00002/01/93
6-7
Page 88
6/Maintaining the Ventilator
Figure 6-8
Cleaning and
sterilization methods
123
1.Wipe with a damp cloth (liquid disinfectant, mild detergent, isopropyl
alcohol); do not immerse, gas sterilize, or autoclave; remove oxide from
leaked electrolyte under a fume hood with vinegar using eye and skin
protection
2.Gas sterilize with ethylene oxide (ambient temperature, no vacuum or
pressure) or wipe with a damp cloth (water, white vinegar, liquid disinfectant): do not immerse or autoclave; do not use alcohol
3.Gas sterilize with ethylene oxide or clean with liquid disinfectant, mild
detergent solution or isopropyl alcohol
MD.02.032MD.07.009
3
100% O2 calibration
At least once a month and following sensor cartridge replacement,
calibrate the oxygen sensor using 100% O2:
1. Adjust the high O2 alarm to 00% (alarm disabled).
2. Expose the O2 sensor to pure oxygen and allow the display to stabilize for approximately two minutes as oxygen fills the patient circuit.
3. Adjust the calibration control until the display reads 99%.
4. Expose the O2 sensor to room air and allow the display to stabilize for
approximately two minutes as room air fills the sensor housing. If the
final reading is outside the allowed range 21 ± 3% (18 to 24%), the
sensor cartridge is no longer linear and must be replaced. Refer to
“Installing a cartridge or disassembling the O2.sensor for cleaning “
in this section.
5. If the system will not be used immediately, switch the monitor or
system OFF.
6-81500-0134-00002/01/93
6-8
Page 89
6/Maintaining the Ventilator
Cleaning the supply gas filter
Operation
Ventilators that are set up to use air as a supply gas have a 5 micron
filter encased in a plastic sediment bowl. The bowl has a drain at the
bottom which operates similar to the valve stem on an automobile tire.
To release any collected sediment and/or moisture, push in the valve
stem and the supply gas pressure should clear the bowl. See figure 6-9.
WARNING: The liquid and sediment in the plastic bowl are under
∑
ventilator supply gas pressure. Wear protective eye gear while draining. Use a container to catch the liquid.
A manual drain valve (7) allows the accumulation to be periodically
removed from the bowl. The liquid level should never be allowed to reach
the level of the filter mounting stud (4) or the liquid may be carried
downstream damaging the internal workings of the ventilator.
Filter servicing
∑
This filter has a manual drain and must be drained as liquid and sediment build up in the bowl. To operate the drain valve, simply use a blunt
object to push upward on the exposed plunger.
Ohmeda strongly recommends that the filters be replaced every six
months and more frequently as the white filter element discolors or
supply gas flow appears to be impeded. See section 8 “Replaceable
Parts” for service kit and filter replacement. A dirt laden filter will impede the supply gas flow and may cause the minute volume alarm to
occur.
WARNING: Clean the plastic bowl with warm water only. Plastic
deteriorates when subjected to certain solvents, strong alkalis and
compressor oils.
Disassembly, see figure 6-9
1. Shut OFF inlet supply gas pressure and push the drain valve stem (7)
upward to reduce the pressure in the bowl (6) to zero.
2. Unscrew the clear filter bowl (6) from the supply gas block.
3. Unscrew the mounting stud (4). Remove the gasket (1), louver (2) and
filter element (3).
4. Discard the O-ring (5), gasket (1) and filter (3).
5. Clean the louver and mounting stud in warm water and a mild soap
solution. Clean the filter bowl (6) in warm water only.
6-91500-0134-00002/01/93
6-9
Page 90
6/Maintaining the Ventilator
Figure 6-9
Supply gas filter
assembly, ventilators
setup for air supply gas
only
Benutzen Sie nurKeimfreie Luft (241-517 kPa)
1
2
3
4
Benutzen Sie nur
Keimfreie Luft
(241-517 kPa)
M
W
Verbinden Sie mit der
Inspirations - seite des
Beatmungssystems
W
Verbinden Sie mit
dem Einlaß an der
Balgeinheit des
Ventilators
∑
Warnung:Verschließen
Sie nicht die GasAuslaßöffnungen. Dies
hätte erhöhten Druck im
Patientenkreislauf zur Folge.
M
AA.32.039AA.32.038
5
6
7
1.Gasket
2.Louver
3.Filter element
4.Mounting stud
5.O-ring
6.Bowl
7.Drain valve
6. Install the new filter (3), stepped end toward the louver, louver (2),
and gasket (1) onto the mounting stud (4).
CAUTION: Do not cross-thread the mounting stud. Carefully en-
w
gage the threads and turn slowly. The filter must be mounted
correctly to help prevent possible damage to the ventilator from
particles in the supply gas.
7. Screw the mounting stud (4) into the gas supply block, finger tight.
6-101500-0134-00002/01/93
6-10
Page 91
6/Maintaining the Ventilator
8. Lubricate the filter bowl threads and the O-ring with an oxygen
approved lubricant (KRYTOX®).
9. Slip the O-ring (5) over the lip on the filter bowl and carefully screw
the filter bowl into the gas block hand tight. Do not cross-thread the
filter bowl.
10. Connect the supply gas.
WARNING: Check for supply gas leaks around the filter bowl. A
leaking filter bowl can deplete reserve cylinders, and a reserve gas
supply may not be available for emergency use and patient injury
could result.
11. Check around the filter bowl for leaks:
•Listen for a hiss sound
•Test the area around the top of the filter bowl with a leak check
fluid
•Periodically check the cylinder pressure gauge
® KRYTOX Trademark E.I. Du Pont de Nemours Co. (Inc.)
Do not use malfunctioning equipment. Make all necessary repairs, or
have the equipment serviced by an authorized Ohmeda representative.
After repair, test the equipment to ensure that it is functioning properly,
in accordance with the manufacturer’s published specifications.
To help ensure full reliability, have all repairs and service done by an
authorized Ohmeda representative. If this cannot be done, replacement
and maintenance of those parts listed in this manual may be undertaken
by a competent, trained individual having experience in the repair of the
Ohmeda 7800 Ventilator and having appropriate test and calibration
equipment.
CAUTION: No repair should ever be undertaken or attempted by
w
anyone not having proper qualifications and equipment.
Replace damaged parts with components manufactured or sold by
Ohmeda. Then test the unit to ascertain that it complies with the
manufacturer’s published specifications.
In some cases, special diagnostic equipment may be required to properly
service components of the Ohmeda 7800 Ventilator. The components
must then be sent to the nearest Ohmeda Service Center.
Contact the nearest Ohmeda Service Center for service assistance. If you
send any unit to an Ohmeda Service Center, package it securely in the
original shipping container, if possible, and ship it prepaid. Enclose a
letter with the unit describing in detail any difficulties experienced and
the repairs felt necessary. In all cases, other than where Ohmeda’s
warranty is applicable, repairs will be made at Ohmeda’s current list
price for the replacement part(s) plus a reasonable labor charge.
7-11500-0134-00002/01/93
7-1
Page 94
7/Service Procedures
Troubleshooting guide
This guide is divided into two sections: ventilator problems and ventilator failure messages.
Ventilator problems
SymptomPossible causeRecommended action
No display on the~ power has failed andPlug control monitor into
ventilator screen. Nobackup battery isworking ~ source and
alarms sounding.completely discharged.power it ON for 24-hours to
recharge battery.
Excel to ventilator cableReattach and use screws to
disconnectedhold connector
Bellows does not expandLeak in the breathing system.Check breathing system
during ventilation or tendshoses and connections.
to collapse.
Bellows not installed properly.Check bellows to base
attachment.
Tear or leak in bellows.Check the entire surface of
the bellows. Pay close
attention to the angles in the
convolutions
Insufficient fresh gas flow.Check that settings on
flowmeter are adequate.
Bellows does not descendExhaust valve failedCheck exhaust valve
during inspiration
Drive gas hose disconnectedReconnect drive gas hose
Ventilator switch in APL/BagPlace switch in Vent position
position
Bellow is bad, distended,Incorrect scavenging systemCheck the scavenging system
or slips off the base.pressure.for vacuum or high pressure.
Transient apnea alarm isBreath rate is set to two (2),This is normal. No action is
triggered by sigh breath.sigh function is enabled, andrequired.
sigh breath occurring.
System sounds alarms atLiquid in pressure sensing tubeDrain the sensing tube.
incorrect pressures.
Tube disconnected.Reconnect the tube.
Kink in tube.Replace the tube.
7-21500-0134-00002/01/93
7-2
Page 95
7/Service Procedures
Ventilator problems, continued
SymptomPossible causeRecommended action
Volume readings areLow oxygen supply pressureCheck, and repair, the oxygen
consistently low.supply.
Failed volume sensor cartridge.Replace the volume sensor
cartridge.
Breathing system leak.Find and repair leaks.
Control module’s altitudeReset the altitude compensa-
compensation set incorrectly.tion as described in “Setting
the altitude compensation” in
section “2/Getting Started.”
Volume readings are high.Control module’s altitudeReset the altitude consis-
compensation set incorrectlytently compensation as
described in “Setting the
altitude compensation” in
section “2/Getting Started.”
Alarms sound withoutCertain electronic failures mayDo not use the ventilator.
apparent cause and cannotbe so significant as to cause
be silenced.the system to lose the ability to
generate ventilator failure
messages even though the
ventilator has failed. If such a
serious failure occurs, the alarm
silence button will not silence
the alarm.
Reverse flow alarm isExpiratory check valve onReplace check valve disk.
activated. (While theabsorber is functioning
volume sensor is in theincorrectly.
distal position of the
expiratory limb of the
breathing circuit.)
Reverse flow alarm isIf the volume sensor is locatedEither locate sensor in the
activated during everyin the proximal end of the “Y”distal position of the expirabreath.connector in the patient circuit,tory limb, (see “2/Getting
the alarm may sound for eachStarted”) or use the setup
breath.page to disable the reverse
flow alarm. (see section
“5/Operating the Ventilator”)
Volume sensor cartridge isCorrectly connect the clip to
connected backwards to sensorthe volume cartridge.(see
clip.“2/Getting Started”)
7-31500-0134-00002/01/93
7-3
Page 96
7/Service Procedures
Troubleshooting ventilator failure messages
Ventilator failure messages can indicate anything from a defective
electronic chip to excessive pressure in the ventilator’s driving gas
supply. Do not attempt to use the ventilator while a ventilator failure
message is displayed. And, even if no ventilator failure message is
displayed, do not use the ventilator if you suspect a malfunction has
occurred.
IMPORTANT
If the ventilator experiences extreme electrical interference, it may
interrupt mechanical ventilation. If this interruption occurs, the ventilator
generates an internal reset function and resumes normal operation after
two (2) seconds. For situations where continuous electrical interference
is experienced by the ventilator, causing a continuous interruption, the
ventilator’s internal reset repeats until the interference ceases.
If the electrical interference is continuously present and mechanical
ventilation is interrupted for approximately 30 seconds, the ventilator
produces a continuous beeping audio alarm. Manual ventilation of the
patient must be performed while the mechanical ventilation is interrupted. When the electrical interference ceases, the continuous beeping
audio alarm can be silenced only by turning, as applicable, the ventilator
or anesthesia machine power switch OFF and after five seconds back
ON.
∑
WARNING: Manual ventilation must be performed when electrical
interference causes interruption of ventilator delivered mechanical
ventilation. Manual ventilation must be continued until the ventilator
resumes normal operation or an alternate ventilator/anesthesia system can be used.
WARNING: The use of electrosurgical units or other devices that
radiate high-intensity electrical fields can affect the operation of the
ventilator and monitors attached to the patient. Maintain as much
distance as possible between the electrosurgical leads and the cables
to the flow and oxygen sensors. Do not drape the electrosurgical leads
across the absorber or the anesthesia machine. Do not let the
electrosurgical leads rest on any surface of the anesthesia system.
Constant surveillance of all monitoring and life support equipment is
mandatory whenever electrosurgical devices are in operation, on, or in
the vicinity of the patient.
If your ventilator displays a ventilator failure message, please note the
failure number, any other symptoms, and any corrective actions you took,
then call trained service personnel.
7-41500-0134-00002/01/93
7-4
Page 97
7/Service Procedures
Ventilator failure messages
Vent fail messagePossible cause
VENT FAIL 0 !Electronic failure
VENT FAIL 1 !Electronic failure
VENT FAIL 2 !Electronic failure
VENT FAIL 3 !Electronic failure
VENT FAIL 4 !Regulated gas pressure more than 30 psig (210 kPa)
VENT FAIL 5 !~ power failure and backup-battery voltage low
VENT FAIL 6 !Flow valve circuit failure
DRIVE CKT OPEN !Exhalation valve circuit failure or Bag/APL switch in wrong position
VENT FAIL 8 !Gas inlet circuit failure
VENT FAIL 9 !Electronic failure
VENT FAIL 10 !Electronic failure
VENT FAIL 11 !Electronic failure
VENT FAIL 12 !Electronic failure
VENT FAIL 13 !Electronic failure
VENT FAIL 14 !Electronic failure
VENT FAIL 15 !Auxiliary port selected on Excel dual port system, but French language
Introduction 8-1
Getting started 8-1
Ventilator Connections 8-2
Post Assembly Test 8-6
Cleaning and Sterilization 8-7
Cleaning 8-8
Sterilization 8-8
Periodic maintenance 8-9
Visual inspection 8-9
Pressure leak test. 8-9
Illustrated Parts List 8-11
Introduction
The Ohmeda Autoclavable Bellows Assembly (ABA) was specifically
intended for use with Ohmeda 7000, 7800 ventilators; Modulus II Plus/
7810 and Modulus CD/7850 Anesthesia systems. For detailed system
information or information on setting up, theory of operation and preoperative checkout procedures, see the appropriate section of this O&M
manual.
Getting started
∑
Special note: The ventilator manuals, anesthesia system manuals and
the appendix of this manual may contain information that pertains to
other versions of Ohmeda’s bellows assemblies. The information contained in this section pertains only to the Ohmeda ABA and DOES NOT
APPLY TO OTHER BELLOWS ASSEMBLY VERSIONS.
WARNING: Disassembly, reassembly, cleaning and sterilization of the
ABA should never be undertaken by any person who has not read this
manual thoroughly and clearly understands the text. Failure to be
totally familiar with the disassembly and reassembly of the ABA can
result in equipment malfunction and injury to the patient.
This unit is not sterile as it is shipped from the factory.
The ABA mounting plate mounts on any Ohmeda 7000 or 7800 series
ventilator control module, or in a remote mounting location using the four
mounting screws provided. There are several possible orientations for
the mounting plate and connection ports. See the exploded view illustrated parts list of the ABA at the end of this section.
8-11500-0134-00002/01/93
8-1
Page 100
8/Autoclavable Bellows Assembly
Ventilator Connections
WARNING: Do not connect the 30-mm exhaust port directly to a high
∑
Figure 8-1
Port identification and
typical adapter
vacuum source. The vacuum may remove required gases from the
breathing circuit
∑
123
1.22-mm port to breathing system
2.30-mm port waste gas scavenging
system
3.17-mm port drive gas
4.30-mm to 19-mm Adapter
(see section 9.2)
WARNING: Disassembly, reassembly cleaning and sterilizing of the
ABA should never be undertaken by any person who has not read this
section thoroughly and clearly understood the text. Failure to be
totally familiar with disassembly and reassembly of the ABA can
result in equipment malfunction.
WARNING: Always perform the preoperative checkout procedures
before using the system. Failure to ensure proper assembly, setup and
operation prior to use may result in injury to the patient.
Note: Territorial standards may
dictate gas scavenging ports
other than 30-mm (see section
9.2). Adapters may have been
included with your system.
Contact your local Ohmeda
representative for port conversion information.
4
AA.51.002AA.51.001
Disassembly
The sequence of the following illustrated procedure is for disassembly of
the bellows assembly — reassembly is the reverse of this sequence.
8-21500-0134-00002/01/93
8-2
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.