This device performs as described in this manual, and in accompanying labels
and inserts, when assembled, operated, maintained, and repaired in accordance
with the instructions provided.
Attention! Consult the accompanying instructions before using this
device.
The safety, reliability, and performance of this device can be assured only under
the following conditions:
• If it is used according to the accompanying operating instructions.
• If fittings, extensions, readjustments, changes, or repairs are carried out by
agents authorized by Datex-Ohmeda.
•If it is used in buildings having ground equalization wiring that complies with
relevant local standards and regulations.
CAUTION: US Federal law restricts this device to sale by or on the order of a licensed medical
practitioner. Outside the USA, check local laws for any restriction that may apply.
This device must be cleaned and checked periodically. Do not use a defective
device. Parts that are broken, missing, plainly worn, distorted, or contaminated
should be replaced immediately. If repair or replacement becomes necessary,
request service advice from Datex-Ohmeda (information is listed on the back
cover). Do not repair this device or any of its parts other than in accordance with
written instructions provided by Datex-Ohmeda.
The user of this device shall have the sole responsibility for any malfunction that
results from improper use, faulty maintenance, improper repair, unauthorized
service, damage, or alteration by anyone other than Datex-Ohmeda.
Trademarks
Datex®, Ohmeda®, OxyTip®, PerfTrak®, TruTrak®, and PIr™ are the property of
Instrumentarium Corp. or its subsidiaries.
Cidex is a registered trademark of Johnson & Johnson.
Microsoft Windows Terminal is a trademark of Microsoft Corporation.
ProComm is a trademark of DataStorm Technologies.
All other product and company names are the property of their respective owners.
Principles of operation..................................................................................................................1-4
Front panel.......................................................................................................................................................1- 6
Powering the oximeter...............................................................................................................................2-1
Signal and data validity............................................................................................................................2- 8
High priority........................................................................................................................................3-4
Medium priority................................................................................................................................3-4
Recharging the battery..............................................................................................................................4-2
Replacing the battery.................................................................................................................................4-2
Replacing the fuses.....................................................................................................................................4-3
Repair policy and procedure.................................................................................................................4-4
Packaging and return procedure............................................................................................4-4
Parts list ..............................................................................................................................................................4-5
ii
Table of Contents
A/Compliance and Specifications
Compliance with standards...................................................................................................................A-1
General safety requirements.....................................................................................................A-1
• Introduces the product, including the principles of its operation.
• Describes the oximeter’s controls and features.
• Lists the precautions you must take when using the oximeter.
Product description
The Datex-Ohmeda Model 3800 pulse oximeter with TruTrak
features two easy-to-read displays that present patient data and status information.
• The numeric display shows the SpO2 and pulse rate values.
• The graphic display shows the plethysmographic waveform, messages, the
Relative Perfusion Index (PIr™) pulsatile value, and the high and low alarm
limit settings for SpO2 and pulse rate.
Intended use
The 3800 pulse oximeter with TruTrak+ technology is indicated for spot-checking
and continuous monitoring of functional oxygen saturation and pulse rate,
including monitoring during conditions of clinical patient motion.1 This device is
intended for use with adult, pediatric, and neonatal patients in both hospital and
non-hospital environments.
Important: Only Datex-Ohmeda OxyTip
TruTrak+ technology
TruTrak+ technology improves pulse oximetry performance during conditions of
clinical patient motion. In the clinical environment, oximetry readings are affected
by several types of patient motion. The types of motion include clenching,
pressing, and rubbing as well as extending, flexing, and kicking. Unlike motion
technologies that use only a single method to correct for motion, TruTrak+ selects
one of many proprietary motion-correction algorithms, depending on the type and
intensity of the motion.
®
+ technology
®
+ sensors can be used with this monitor.
TruTrak+ technology employs a patented five-step process that consists of
1) high-speed data sampling; 2) motion identification, quantification, and
1
Anesthesia & Analgesia. 2002;94,1S, S54-S60
1-1
3800 User’s Manual
PIr pulsatile value
correction; 3) calculation of the SpO2 value; 4) weighting and averaging of the
SpO2 value; and 5) the display of an improved SpO2 value. The result of this
process is a more accurate and stable displayed SpO2 value, with fewer false
alarms or dashed displays.
Important: For TruTrak+ performance, the averaging mode must be set to Long.
See Setup in chapter 2.
The PIr pulsatile value indicates the strength of the pulse signal at the sensor site:
the higher the PIr value, the stronger the pulse signal. A strong pulse signal
increases the validity of SpO2 and pulse rate data.
PIr is a relative value that varies from patient to patient. Clinicians can use the PI
value to compare the strength of the pulse signal at different sites on a patient in
order to locate the best site for the sensor (the site with the strongest pulse signal).
You can choose to display or not display the PIr value (see Setup in chapter 2).
Other features
• PerfTrak® waveform display, an automatic scale of the plethysmographic
• Large SpO2 digital display for clear differentiation from the pulse rate value.
• Backlit display and contrast control for excellent visibility in subdued lighting
• Direct access to user-selectable high and low alarm limits for SpO2 and pulse
• An audible pulse indicator with an adjustable volume; the automatic pitch
• Visual and audible (adjustable volume) alarms.
• An alarm-silence feature that silences audible alarms for 120 seconds.
• An all-mute feature that silences audible alarms until deactivated.
• Automatic tiered alarm messages.
• Short, medium, or long SpO2 response averaging modes.
• Adult or neonatal patient modes for default pulse rate alarm settings.
• Automatic storage of up to 12 hours of SpO2, pulse rate, and alarm limit
• An automatic self-test and calibration check at start-up. After start up, the
• Rechargeable, sealed, lead-acid battery operation, including battery status
r
waveform to provide a relative indication of the sensor site perfusion level.
conditions; adjustable viewing angle, using the pull-down feet under the
monitor.
rate.
modulation reflects changing SpO2 level.
violations data in trend memory, which can be output through the RS-232
serial connector.
The 3800 oximeter uses the following key electrical component elements to
determine SpO2, pulse rate, and PIr pulsatile values:
• The sensor
• Sensor-signal processing
• Microprocessor calculations
The sensor consists of
• The light source—red and infrared light-emitting diodes (LEDs)
• The photodetector—an electronic device that produces an electrical current
proportional to incident light intensity
Timing
Control
1/Overview
Sensor
Figure 1-1. Signal processing block diagram
The two light wavelengths generated by the LEDs pass through the tissue at the
sensor site. The photodetector collects this light (partially absorbed and modulated) and converts it into an electronic signal that is sent to the oximeter for
further processing.
The electronic circuitry receives the photodetector’s electronic signal, processes it,
and passes it on to the microprocessor for calculation of the SpO2, pulse rate, and
PIr pulsatile value.
Analog
Processing
A/D
Converter
Digital
Processing
Input/
Output
1-3
3800 User’s Manual
T
Principles of operation
The 3800 pulse oximeter uses a two-wavelength pulsatile system—red and infrared
light—to distinguish between oxygenated (O2Hb) and reduced (HHb) hemoglobin,
each of which absorbs different amounts of light emitted from the oximeter
sensor. The system then calculates the relative percentage of these two
constituents and displays SpO
.
2
Variable absorption
(due to arterial pulse)
Arterial blood absorption
Venous blood absorption
Absorption
Other tissue absorption
ime
Figure 1-2. Comparative light absorption
Arterial blood pulsation at the test site modulates transmission of the oximeter
sensor’s light. Since other fluids and tissues present generally don’t pulsate, they
don’t modulate the light passing through that location. The attenuation of light
energy due to arterial blood flow is detected and isolated by using the pulsatile
portion of the incoming signal. PIr pulsatile value is a measure of the relative size
of this portion of the signal.
1
(Red)
660 nm
0
(Infrared)
940 nm
Extinction (10x)
O2Hb
-1
HHb
-2
6007008009001000
Wavelength (nm)
Figure 1-3. Extinction versus wavelength graph
The sensor’s photodetector converts the light, which is partially absorbed and
modulated as it passes through the tissue sample, into an electronic signal. Since
O2Hb and HHb allow different amounts of light to reach the photodetector at the
selected wavelengths, the electronic signal varies according to the light source that
1-4
1/Overview
is “on” and the oxygenation of the arterial hemoglobin. Analog and digital signal
processing then converts the light-intensity information into SpO2, pulse rate, and
PIr pulsatile values for display on the monitor.
Calibration
A CO-oximeter typically uses four or more wavelengths of light and calculates
reduced hemoglobin (HHb), oxyhemoglobin (O2Hb), carboxyhemoglobin (COHb),
and methemoglobin (MetHb). Datex-Ohmeda pulse oximeters use two
wavelengths ranges, 650 nm - 665 nm and 930 nm - 950 nm, both with an average
power of less than 1 mW. These wavelengths are used to calculate the presence of
O2Hb and reduced HHb. Because of this, pulse oximetry readings will be
different than CO-oximetry readings in situations where a patient’s COHb or
MetHb are increased.
Two different methods of calibration are currently used by manufacturers of pulse
oximeters: fractional and functional.
Important: This pulse oximeter uses the functional calibration method. The user
cannot change the calibration method to fractional.
•Fractional saturation is represented mathematically as the percentage of the
total amount of hemoglobin carrying oxygen. It is determined by dividing the
oxyhemoglobin by the total hemoglobin.
Fractional SpO2 =x 100 =x 100
•Functional saturation is represented mathematically as the percentage of
hemoglobin capable of carrying oxygen that is carrying oxygen.
Functional SpO2 =x 100 =x 100
The calculation of SpO2 assumes 1.6% carboxyhemoglobin (COHb), 0.4%
methemoglobin (MetHb), and no other pigments. Appreciable variation from these
values will influence SpO2 accuracy. These values are based on the DatexOhmeda Pulse Oximeter Empirical Calibration Study.
•120-second alarm silence—activated by a single press.
•Continuous all mute—activated by three quick presses (if the all-mute
feature is enabled). Press once to deactivate.
Alarm silence
When an active alarm condition exists, press this button to silence the audible
portion of the alarm for 120 seconds. The flashing red or yellow alarm light
becomes a steady light. If an alarm condition still exists after 120 seconds, the
audible tone and flashing light resume.
Exceptions: Both NO SENSOR and SENSOR OFF audible alarms will not be
activated until after the unit obtains a valid signal. The same conditions apply
to an active audible alarm for NO SENSOR, SENSOR OFF, or SENSOR
FAILURE that has been silenced; i.e., once the sensor alarm condition is
acknowledged by silencing the audible alarm, a new audible alarm will not
sound until the condition has been cleared and the unit obtains a valid signal.
NOTE: Pressing the alarm silence button produces 120 seconds of silence,
regardless of other alarm conditions that may occur during this 120-second
interval, except for the SYSTEM FAILURE, CONNECT UNIT TO LINE
POWER, and BUTTON STUCK alarms.
All mute
To continuously silence any alarm that can be silenced, press the alarm
silence button three times within three seconds. After you have activated all
mute, the all mute icon flashes between the SpO2 and pulse rate alarm limit
settings on the right side of the screen display. When an alarm condition
occurs, the alarm button light flashes and the alarm message appears on the
waveform display but no audible alarm sounds.
When all mute is active, press the alarm silence button once to deactivate this
feature and enable all audible alarms.
2 Numeric display
SpO2 numeric area—calculated SpO
Pulse rate numeric area—calculated pulse rate
2
1-7
3800 User’s Manual
3Graphic display
21
85
3
PI
r
4
1Plethysmographic waveform
2Message area
3SpO2and pulse rate alarm limits
1.25
The PerfTrak waveform display appears after the monitor has detected
data from the sensor. It represents the blood volume change of the
hemodynamic system, assuming no other factors (e.g., motion artifact) are
present. This waveform scales automatically to the perfusion level or
strength of the signal being received at the patient monitoring site.
Status and alarm messages appear above the waveform area on the
waveform display (the height of the waveform is reduced while messages
are displayed). Status messages give you information about the oximeter's
operating condition. Alarm messages alert you to conditions that need
your attention. See chapter 3 for complete alarm and status message
information.
The high and low alarm limit settings appear here. If a limit is set to OFF,
three dashes appear in the location for that limit.
When an SpO2 or pulse rate limit is violated, the LED on the numeric
display and the LCD’s limit value flashes for that alarm.
130
40
1-8
SpO
4PIr pulsatile value
Dashes (- - -) appear if the following conditions exist: no sensor is
connected to the unit, the sensor is not attached to the patient, the sensor
has failed, there is insufficient light penetrating the tissue site, or there is
too much ambient light.
4SpO2 alarm limits, high and low
2
+
–
The top button sets the high alarm limit and the bottom button sets the low
alarm limit. For either limit, press the + side of the button to raise the value or
the – side to lower it. As you press one of these buttons, the values do not
cycle through the available settings; e.g., when you reach 100, the value does
not cycle (or wrap) to 50 or OFF.
♥
y
1/Overview
5Pulse rate alarm limits, high and low
+
–
The top button sets the high alarm limit, the bottom button sets the low alarm
limit. For either limit, press the + side to raise the value or the – side to lower
it. As you press one of these buttons, the values do not cycle through the
available settings; e.g., when you reach 235, the value does not cycle (or wrap)
to 30 or OFF.
6Display contrast adjuster
Use this sliding lever to adjust the vertical viewing angle of the graphic
display Slide the lever to the left to reduce the contrast and to the right to
increase it.
7Power/Standby button/AC power light
This button toggles between On (operational mode) and Off (standby mode).
The battery recharges as long as the unit is plugged into the AC power supply.
No displays are visible in the Off/Standby mode.
The green light to the right of the button is lit when the unit is connected to
an AC power supply.
85
130
40
Battery operation
The oximeter runs for at least five and one-half hours on a new, fully charged
battery at normal operating temperatures. LOW BATTERY appears when
between 5 and 15 minutes of battery operation time remain. Plug the
monitor into AC power to continue monitoring and recharge the battery. The
unit will operate with a dead or defective battery when it is connected to the
AC power supply.
When the CONNECT UNIT TO LINE POWER message appears, you must
immediately plug the oximeter into the AC power supply or the unit turns
itself off after 10 seconds.
When operating on battery power, an icon appears between the two pairs of
alarm limit values on the right side of the LCD. This icon indicates the batter
condition as follows:
Charged/not low
Low
If the all mute condition exists, the display of this icon alternates with the
display of the all mute icon.
This icon appears on the status screen:
Depleted, not installed, or defective
8Carrying handle
The lower front portion of the oximeter’s case is designed to be a carrying
handle for ease of moving the unit from one place to another.
1-9
3800 User’s Manual
9Sensor connector
The sensors for this oximeter plug into this nine-contact connector. Use only
Datex-Ohmeda sensors compatible with this oximeter (see Parts list in
chapter 4).
♥
+
–
+
–
10 Pulse beep volume button
This button adjusts the volume level for the pulse indicator in incremental
steps from OFF to level 5 (default is 2). Press the + side of the button to
increase the volume or the – side to decrease it; you will hear the volume
level as you press the button. As you press one of these buttons, the values do
not cycle through the available settings; e.g., when you reach 5, the value does
not cycle (or wrap) to OFF.
As you adjust the volume, the volume setting is shown in the message area
above the waveform.
NOTE: The pitch of the pulse tone changes as the SpO2 value increases or
decreases—the higher the SpO2value, the higher the pitch of the pulse tone.
11 Alarm volume button
This button adjusts the audible alarm volume level in incremental steps from
1 to 5 (default is 3). You cannot set the alarm volume to OFF. Press the + side
of the button to increase the alarm volume or the – side to decrease it; you
will hear the volume level as you press the button. As you press one of these
buttons, the values do not cycle through the available settings; e.g., when you
reach 5, the value does not cycle (or wrap) to 1.
As you adjust the volume, the volume setting is shown in the message area
above the waveform.
1-10
Loading...
+ 37 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.