Datex-Ohmeda 3770, 3775 User manual

Page 1
3770/3775 Pulse Oximeter
Service Manual
%SpO
2
2
2798
2798
Not for continuous monitoring
6050-0004-915
Printed in USA
Page 2
Page 3
3770/3775 Pulse Oximeter
Service Manual
Page 4
Important
This manual is subject to periodic review, update, and revision. Customers are cautioned to verify that the information in this manual applies to the software and hardware present in the equipment.
Attention! Consult the accompanying instructions before using this device.
This device performs as described in this manual, and in accompanying labels and inserts, when assembled, operated, maintained, and repaired in accordance with the instructions provided.
This device must be cleaned and checked periodically. Do not use a defective device. Parts that are broken, missing, plainly worn, distorted, or contaminated should be replaced immediately. If repair or replacement become necessary, request service advice from Datex-Ohmeda (information is listed on the back cover). Do not repair this device or any of its parts other than in accordance with written instructions provided by Datex-Ohmeda.
The user of this device shall have the sole responsibility for any malfunction that results from improper use, faulty maintenance, improper repair, unauthorized service, damage, or alteration by anyone other than Datex-Ohmeda.
The safety, reliability, and performance of this device can be assured only under the following conditions:
If it is used according to the accompanying operating instructions.
If fittings, extensions, readjustments, changes, or repairs are carried out by
agents authorized by Datex-Ohmeda.
If it is used in buildings having ground equalization wiring that complies with
relevant local standards and regulations.
Trademarks
Datex® and Ohmeda® are the property of Instrumentarium Corp. or its
subsidiaries. FlexConnect is a trademark of Datex-Ohmeda, Inc. Cidex is a registered trademark of Johnson & Johnson Duracell is a trademark of Duracell, Inc. Eveready is a trademark of Eveready Battery Company, Inc. Formula 409 is a registered trademark of Clorox Company Hewlet Packard is a registered trademark of Hewlett-Packard Co. Kodak is a trademark of Eastman Kodak Company Panasonic is a trademark of Panasonic Industries, Co. Rayovac and Renewal are registered trademarks and Reusable Alkaline is a
trademark of Rayovac Corporation Top Crest is a trademark of Safeway, Inc. Windex is a registered trademark of The Drackett Company All other product and company names are the property of their respective owners.
Text revised: May 2000
© 1997, 1999, 2000 Datex-Ohmeda, Inc. All rights reserved.
Page 5
1/Overview
Related information...........................................................................................................1-1
Technical competence.....................................................................................................1-1
General description.....................................................................................................................................1 -2
Theory of operation.....................................................................................................................................1-2
Calibration............................................................................................................................................1-3
Functional components.............................................................................................................................1 -4
Digital board........................................................................................................................................1-5
80-pin board-to-board connector...................................................................1-5
Analog board.......................................................................................................................................1-6
Specifications..................................................................................................................................................1 -7
Electromagnetic effects.................................................................................................................1-7
Environment..........................................................................................................................1-7
EMC performance..............................................................................................................1-7
International Electrotechnical Commission (IEC) classifications.........................1-7
Compliance with standards.......................................................................................................1-8
Oximeter................................................................................................................................................1-8
Default settings.....................................................................................................................1-8
General......................................................................................................................................1-8
Visual displays and indicators.......................................................................1-8
Dimensions................................................................................................................1-8
Power............................................................................................................................1-9
Circuitry......................................................................................................................1-9
Sensor emitter wavelength ranges..............................................................1-9
(3775 only) Audio indicators...........................................................................1-9
SpO2...........................................................................................................................................1-9
Range............................................................................................................................1-9
Accuracy.....................................................................................................................1-9
Calibration.................................................................................................................1-9
Interfering substances.........................................................................................1-9
(3775 only) Alarm limits....................................................................................1-9
Pulse rate .................................................................................................................................1-9
Range............................................................................................................................1-9
Accuracy.....................................................................................................................1-9
Environmental................................................................................................................... 1-10
Operating................................................................................................................ 1-10
Storage...................................................................................................................... 1-10
Battery packs.................................................................................................................................... 1-10
AA battery pack ................................................................................................................ 1-10
Recommended brands.................................................................................... 1-11
NiCd rechargeable battery pack............................................................................. 1-11
Environmental—NiCd and AA................................................................................. 1-11
Operating................................................................................................................ 1-11
Storage...................................................................................................................... 1-11
Precautions....................................................................................................................................................1-12
Warnings............................................................................................................................................ 1-12
Handling............................................................................................................................... 1-12
Failure of operation........................................................................................................ 1-12
Explosion hazard.............................................................................................................. 1-12
Fire hazard.......................................................................................................................... 1-12
Contents
i
Page 6
Contents
2/ Troubleshooting
3/Service Policy and Repair Procedures
Electrical shock and flammability hazard ........................................................ 1-12
Electrical shock hazard ................................................................................................ 1-13
Patient safety....................................................................................................................... 1-13
Operator safety.................................................................................................................. 1-13
Sensors................................................................................................................................... 1-13
Data validity........................................................................................................................ 1-13
Cautions.............................................................................................................................................. 1-14
Static sensitivity................................................................................................................ 1-14
Battery packs....................................................................................................................... 1-14
Cleaning................................................................................................................................ 1-14
HP printer............................................................................................................................ 1-14
Disposal.................................................................................................................................. 1-14
3770/3775 pulse oximeter.....................................................................................................................2-2
Printer..................................................................................................................................................................2-5
Battery charger monitoring station.....................................................................................................2-6
Battery charger (single)..............................................................................................................................2 -8
Service policy..................................................................................................................................................3 -1
Ordering parts and obtaining service..................................................................................3-1
Packaging and returning equipment....................................................................................3-2
Safety guidelines...........................................................................................................................................3 -2
Cleaning procedures...................................................................................................................................3 -3
Oximeter, battery packs, and battery chargers ...............................................................3-3
Sensors....................................................................................................................................................3-3
Inspecting the oximeter and replacing parts...............................................................................3-4
Disassembling the oximeter .......................................................................................................3-4
Replacing parts..................................................................................................................................3-5
Assembling the oximeter .............................................................................................................3-5
Checking normal operation....................................................................................................................3-7
4/Illustrated Parts
Service kits—3770 and 3775 oximeters.........................................................................................4-1
3775 oximeter assembly..........................................................................................................................4 -2
3775 assembly parts........................................................................................................................4-2
3775 assembly illustration...........................................................................................................4-3
3770 oximeter assembly..........................................................................................................................4 -4
3770 assembly parts........................................................................................................................4-4
3770 assembly illustration...........................................................................................................4-5
ii
Page 7
1/Overview
This manual contains instructions for servicing the Datex-Ohmeda 3770 and 3775 Pulse Oximeters.
This chapter contains:
A general description of the oximeter and its functional components.
Oximeter and battery pack specifications.
Precautions, including specific warnings and cautions, you must follow when
Related information
The content of this manual assumes you are familiar with how the unit operates. For a detailed description of the unit’s components, key functions, and general operating guidelines, see the Datex-Ohmeda 3770/3775P Pulse Oximeter User’s Manual. General maintenance procedures contained in that manual, such as how to replace batteries in the AA battery pack, are not repeated in this manual.
If you need to reference printed circuit board schematics and component lists, purchase and refer to the information contained in the 3770/3775 PCA Drawings Service Kit, REF 6050-0005-558.
You’ll find detailed instructions for servicing and repairing the battery charger monitoring station in the Battery Charger Monitoring Station Service Manual (REF 6050-0003-861). If you need to replace fuses or change the voltage selection for the monitoring station, refer to that manual.
servicing the oximeter.
Note: The NiCd and AA battery packs, the single-station battery charger, and the Hewlett-Packard® (HP) printer are not serviceable accessories.
For information on sensor application and cleaning (reusable sensors only), see the instructions for the sensor.
Technical competence
CAUTION: Only qualified service personnel should perform the procedures described in this manual.
Only Datex-Ohmeda service personnel or competent individuals who are experienced with servicing medical devices of this nature should perform the procedures described in this manual.
1-1
Page 8
3770/3775 Pulse Oximeter Service Manual
General description
Two models of the oximeter are available: the 3770 and the 3775. Both oximeters operate on power supplied by a battery pack. In addition, either oximeter, when a nickel-cadmium (NiCd) battery pack is attached, can be positioned in the battery charger monitoring station and operated using the station’s DC (from AC mains) power supply.
The 3770 oximeter is designed specifically for spot checking SpO
rate. It has no alarms or user-definable parameters and is not intended for continuous monitoring.
The 3775 oximeter is a full-featured model designed for spot checking in
addition to short- and long-term continuous SpO2 and pulse rate monitoring. This model can print trend data, through an infrared link, to the optional Hewlett-Packard® (HP) printer.
Note: Throughout this manual, components and features available on both oximeter models and those found only on the 3775 oximeter are discussed. If you are working on the 3770, references to components specific to the 3775 do not apply.
Theory of operation
The oximeter uses a two-wavelength pulsatile system—red and infrared light—to distinguish between oxygenated (O2Hb) and reduced (HHb) hemoglobin, each of which absorbs different amounts of light emitted from the oximeter sensor. The system then calculates the relative percentage of these two constituents and SpO2.
Variable absorption (Due to arterial pulse)
Arterial blood absorption
Venous blood absorption
and pulse
2
1-2
Absorption
Other tissue absorption
Figure 1-1. Signal composite
Arterial blood pulsation at the test site modulates the transmission of the oximeter sensor’s light. Since other fluids and tissues present generally don’t pulsate, they don’t modulate the light passing through that location. The pulsatile portion of the incoming signal is used to detect and isolate the reduction of light energy due to arterial blood flow.
Time
Page 9
1/Overview
1
(Red)
660 nm
)
0
x
Extinction (10
-1
-2 600 700 800 900 1000
Wavelength (nm)
Figure 1-2. Extinction vs. wavelength
The sensor’s photodetector converts the light, which is partially absorbed and modulated as it passes through the tissue sample, into an electronic signal. Since O2Hb and HHb allow different amounts of light to reach the photodetector at the selected wavelengths, the electronic signal varies according to the light source that is “on” and the oxygenation of the arterial hemoglobin. Analog and digital signal processing then convert the light-intensity information into SpO2 and pulse rate values for display on the monitor.
(Infrared)
940 nm
O2Hb
HHb
Calibration
Datex-Ohmeda pulse oximeters use two wavelength ranges, 650 nm to 665 nm and 930 nm to 950 nm, both with an average power of less than 1 mW. These wavelengths are used to calculate the presence of oxyhemoglobin (O2Hb) and reduced hemoglobin (HHb). A CO-oximeter typically uses four or more wavelengths of light and calculates reduced hemoglobin (HHb), oxyhemoglobin (O2Hb), carboxyhemoglobin (COHb), and methemoglobin (MetHb). Because of this, pulse oximetry readings and CO-oximetry readings will differ in situations where a patient’s COHb or MetHb are increased.
Two different methods of calibration are currently used by oximeter manufacturers: fractional and functional.
The fractional saturation is determined by dividing the oxyhemoglobin by the
total hemoglobin, represented mathematically as
Fractional SpO2 = x 100 = x 100
O2Hb
Hb
TOTAL
O2Hb + HHb + COHb + MetHb
O2Hb
i.e., the percentage of the total amount of hemoglobin carrying oxygen.
The functional saturation is represented mathematically as
Functional SpO2 = x 100 = x 100
Hb
O2Hb
- COHb - MetHb O2Hb + HHb
TOTAL
O2Hb
i.e., the percentage of hemoglobin capable of carrying oxygen that is carrying oxygen.
1-3
Page 10
3770/3775 Pulse Oximeter Service Manual
The calculation of SpO2 assumes 1.6% carboxyhemoglobin (COHb), 0.4% methemoglobin (MetHb), and no other pigments. Appreciable variation from these values will influence SpO2 accuracy. These values are based on the Datex­Ohmeda Pulse Oximeter Empirical Calibration Study.
The 3700/3775 uses the fractional calibration method.
Functional components
Figure 1-3 illustrates the relationship between each functional component.
Battery pack/
AC adapter
Power
Return
Temperature
Battery monitor
board
Analog boardDigital board
Power supplyOn/Off
Signal pathDigital core
Figure 1-3. System block diagram
A complete oximetry system requires digital and analog processing. The digital board provides signal acquisition, numerical processing, and user interface for the oximeter. The analog board amplifies and conditions the biological signal. The digital board then samples the conditioned signal, performs the oximetry algorithm, and displays the results on numeric and graphic LED displays.
The user interface features an eight-character dot-addressable LED display, a 24-segment LED driver, a variable frequency audio driver, and (for the printer option) an infrared wireless transmitter.
Sensor
1-4
Page 11
Digital board
1/Overview
The core processing of the digital board is provided by a 16-bit processor, nonvolatile FLASH memory, semivolatile SRAM, and a Real Time Clock. For signal acquisition, it has an A-to-D converter (which is internal to the processor), I/O ports for timing and control, and an 8-bit D-to-A converter.
On/Off
Power control
(Reset control)
Ambient light
detection
Rotary encoder
and
pushbuttons
Code memory
Data memory
Address
decode
Processor
Analog
Signal
Path
Timing
Analog board
80-pin
connector
Programmable
interval
timer (PIT)
Real-time
clock
Alphanumeric
LED display
LED driver
(bargraph, low
battery, alarm bar)
Peripheral
I/O (PPI)
Audio tone
generator
I.R.
wireless
transmitter
Digital-to-analog
converter (DAC)
Figure 1-4. Digital system block diagram.
80-pin board-to-board connector
The 80-pin board-to-board connector interfaces with the analog board and contains signals shared by both boards: power supplies, rotary encoder signals, oximetry timing and control, and many ground connections.
1-5
Page 12
3770/3775 Pulse Oximeter Service Manual
Analog board
The analog board provides the oximetry analog signal processing, sensor LED drive, detection of interfering signals, housekeeping and fault monitoring, standby and operational power, power switching, and battery interface. It also provides structural support for the digital board and the battery contacts. It contains attachments for the sensor connector, rotary encoder, and serial port. It also provides the grounding system for the case shield with EMC capacitors to shunt high-frequency currents.
VCC
UPS
-V
-PROBE
D
I
G
I
T
A
L
B O A R D
>
INTDET-
AVG DRV
VREF
LEDDRV
LOBAT
AMUX
INTF
MUX 0-3 , MUX 3-
ACA - ACC
VERN
RED
IR
RLT
IRLT
DARK
DLT
PETE
DISABLE
+VSW
REXPULSE
REXCHA
REXCHB
REXPB
OPERATE / TEST
TXD
RXD
GROUND*
A N A L O G
5
B O A R D
<
>
<
GROUND
<
DETECTOR
DET RETURN*
IR CATHODS
RED CATHODE
ANODES*
PROBE ID
ID RETURN
VBATT
BATMON
BATTERY RETURN*
GROUND* GROUND*
REXPULSE
REXCHA REXCHB
REXPB
SERIAL PORT
Sensor
Battery
assembly
Rotary
encoder
*Essential grounds
1-6
Figure 1-5. Analog board and digital board system
Note: The rotary encoder is installed on the 3775 analog board only; in all other respects the 3770 and 3775 analog boards and their components are identical.
Page 13
Specifications
Electromagnetic effects
1/Overview
Unless otherwise indicated, all specifications are nominal and are subject to change without notice. Unless otherwise indicated, the specifications below apply to both oximeters.
The 3770 or the 3775 when used with a LR6 (AA) battery pack, a NiCd battery pack, or a NiCd battery pack in a battery charger monitoring station are referred to as the “system” for EMC purposes.
Indications that the system is experiencing electromagnetic interference
include:
Variations in the display (pleth bar does not correlate to physiological signals).
Sudden increases or decreases in the signal strength indicator that do not
correlate to physiological condition of the patient.
PROBE? messages that are not resolved by the instructions found in the
3770/3775 Pulse Oximeter User’s
T
he display of “rolling dots” when a valid physiological signal is present.
Manua.
This interference may be intermittent and careful correlation between the effect and its possible source is important. The system will not display any of these indications if it is used within its intended electromagnetic environment.
Environment
Suitable for use in the environment described in IEC/EN 60601-1-2 (1993).
EMC performance
The system complies with the requirements of IEC/EN 60601-1-2 (Electromagnetic compatibility - Requirements and tests). The following basic EMC standards were applied to verify conformance.
Emissions: IEC/EN 55011 Group I, Class B (1997)
Immunity: IEC/EN 61000-4-2 (1995), 8 kV air, 3kV contact
IEC/EN 61000-4-3 (1995), 3 V/m IEC/EN 61000-4-4 (1995), 2 kV power, 1kV I/O IEC/EN 61000-4-5(1995), 2 kV line to earth, 1 kV line to line
International Electrotechnical Commission (IEC) classifications
The oximeter complies with the requirements of IEC/EN 60601-1 (1988) Part 1: General requirements for safety of medical electrical equipment:
Degree of protection against electric shock: Type BF applied part
IPX3
Type of protection against electric shock: Internally powered equipment/Class II
Mode of operation: Continuous
Degree of protection against ingress of liquids: Sprayproof (IEC/EN 60529)
1-7
Page 14
3770/3775 Pulse Oximeter Service Manual
Compliance with standards
The presence on the device of any symbol described below indicates compliance with the standard represented by that symbol.
Medical Device Directive 93/42/EEC of the European Union for a
0197
class I (with a measuring function), IIa, IIb, or III device.
Medical electrical equipment classified with respect to electric shock, fire, and mechanical hazards only, in accordance with the Canadian Standards Association CAN/CSA C22.2 No. 601.1 (1990).
Medical electrical equipment classified with respect to electric shock, fire, and mechanical hazards only, in accordance with Underwriters Laboratories Inc. UL 2601-1 (1994).
Oximeter
Default settings
Each time you power on the oximeter the following system settings are in effect:
Low-SpO
High-SpO
Alarm volume = On
Pulse beat volume = Off
SpO
Settings for time, date, and language are retained when the system is powered off as long as a battery is in place.
alarm limit = 85
2
alarm limit = On
2
averaging interval = 12 seconds
2
1-8
General
Visual displays and indicators
The displayed SpO2 and pulse rate values are updated every second. The
plethysmographic waveform sweep is updated every 4 seconds.
Main display: 8-character alpha-numeric; green; variable intensity dependent on
ambient light conditions Waveform display: 8 green LEDs Signal strength indicator bar: 4 LEDs (1 dark green, 1 yellow green, 1 yellow, and
1 orange) Low battery indicator: 1 orange LED Alarm bar (3775 only): red Display update interval: 0.5 seconds Display hold interval: 12 seconds
Dimensions
Height: 18.54 cm (7.3 inches; 9.5 inches including FlexConnect™) Width: 7.62 cm (3.0 in.) Depth: 5.33 cm (2.1 in.) Weight: 0.33 kg (9.6 oz.) without battery pack
Page 15
Power
Typical consumption: 1.6 W Typical current: 180 mAmps
Circuitry
Microprocessor controlled Automatic self-test at power on Automatic setting of default parameters (3775 only) Automatic alarm messages Automatic/continuous system diagnostics
Sensor emitter wavelength ranges
Red LED peak wavelength range: 650 to 665 nm Infrared (IR) LED peak wavelength range: 930 to 950 nm Average power: ≤ 1 mW
(3775 only) Audio indicators
Alarm volume: On (default) and off Pulse rate volume: On and off (default) Pitch modulation to reflect changing SpO2 levels Alarm silence (120 seconds) Low- and high-SpO2 out-of-limits alarm Sensor condition alarms System failure alarms
1/Overview
S p O
2
Range: 0 to 100%
Accuracy (±1 standard deviation)
80 to 100% ± 2% 60 to 79% ± 3% Below 60% unspecified
Calibration: fractional
Interfering substances
Carboxyhemoglobin may erroneously increase readings. The level of increase is approximately equal to the amount of carboxyhemoglobin present. Dyes, or substances containing dyes, that change usual arterial pigmentation may cause erroneous readings.
(3775 only) Alarm limits
Low = 60 to 99% High = 61 to 100% and OFF
Pulse rate
Range 40 to 235 beats per minute
Accuracy ± 1.7% of current reading
(accuracy calculations assume a constant pulse rate)
1-9
Page 16
3770/3775 Pulse Oximeter Service Manual
Environmental
Operating
Temperature: 0 to 50 ºC (32 to 122 ºF) Humidity: 0 to 95% RH, noncondensing Atmospheric pressure: 550 to 1,060 hPa (8 to 15.4 psia) Vibration: Meets or exceeds:
Drop/shock: Meets or exceeds:
Storage
Temperature: –40 to 70 ºC (–40 to 158 ºF) Humidity: 0 to 95% RH, noncondensing Atmospheric pressure: 496 to 1,060 hPa (7.2 to 15.4 psia) Vibration: Meets or exceeds:
Drop/shock: Meets or exceeds:
ASTM 4728, (Method C) MIL-STD-810E, Method 514.4, Section I.3.4.3 IEC/EN 60068-2-37
IEC/EN 60068-2-32 (Procedure 1) MIL-STD 810E, Method 516.4, Section I.3.6
ASTM 4728, (Method C) MIL-STD-810E, Method 514.4, Section I.3.4.3
IEC/EN 60068-2-32 (Procedure 1) MIL-STD 810E, Method 516.4, Section I-3.6
Battery packs
The operation time depends on the monitoring technizue (long monitoring cycles or short, spot-checks).
AA battery pack
Weight: 0.12 kg (4.4 oz.) without cells; 0.26 kg (9.3 oz.) with cells.
Capacity, operation time, charge time, and life cycle depend on the battery cell.
Figure 1-6. AA battery life cycles
Battery type Life expectancy in hours
05101520
Alkaline (8 to 15 hours)
Rechargeable alkaline (2.5 to 8 hours)
Lithium (10 to 20 hours)
Carbon (2 to 4.5 hours)
1-10
Page 17
Low battery indicator LED lights depending on the battery cell, as follows:
Alkaline <2.5 hours
Alkaline rechargeable <1 hour
Carbon <2 hours
Lithium <1 hour
Recommended brands
Alkaline Lithium Carbon
Duracell ™ Eveready Eveready Eveready™ Rayovac Kodak™ Panasonic™ Rayovac®, including
rechargeable "Renewal®"
Top Crest™
NiCd rechargeable battery pack
Weight: 0.43 kg (12.5 oz.) Power: 9.6 V, sealed nickel-cadmium (8 cells) Capacity: 1.4 A-hours Operation time: 8 hours Charge time: 2 hours Life cycle: 1 year (500 cycles)
Low battery indicator LED illuminates when battery usage time is at or below
15 minutes.
~
1/Overview
Environmental—NiCd and AA
Operating
Temperature: 0 to 50º C (32 to 122 ºF) – NiCd
0 to 40º C (32 to 104 ºF) – AA Humidity: 0 to 95% RH, noncondensing Atmospheric pressure: 550 to 1,060 hPa (8 to 15.4 psia) Vibration: Meets or exceeds:
ASTM 4728 (Method C)
MIL-STD-810E, Method 514.4, Section I.3.4.3
IEC/EN 60068-2-37 Drop/shock: Meets or exceeds:
IEC/EN 60068-2-32 (Procedure 1)
(NiCd) MIL-STD 810E, Method 516.4, Section I-3.6
Storage
Temperature: –40 to 50 ºC (–40 to +122 ºF) Humidity: 0 to 65% RH (±20%) noncondensing Atmospheric pressure: 496 to 1,060 hPa (7.2 to 15.4 psia) Vibration: Meets or exceeds:
ASTM 4728 (Method C)
MIL-STD-810E, Method 514.4, Section I.3.4.3 Drop: Meets or exceeds:
IEC/EN 60068-2-32 (Procedure 1)
1-11
Page 18
3770/3775 Pulse Oximeter Service Manual
Precautions
Two types of precautions appear in this manual: warnings and cautions.
A WARNING indicates a potentially harmful situation that may cause injury to a
patient or operator.
A CAUTION indicates a condition that may lead to equipment damage or
malfunction.
Read this section fully and carefully before using or servicing the oximeter and its accessories.
Warnings
Handling
Handle the oximeter and its accessories with care. Improper handling can cause damage or inaccurate operation of these devices.
Failure of operation
If the oximeter fails to respond as described, do not use it until qualified personnel have corrected the situation or the unit has been serviced.
Explosion hazard
Do not use the oximeter, printer, or battery charger in the presence of flammable anesthetics or other flammable substances in combination with air, an oxygen­enriched environment, or nitrous oxide.
Use only a 3770/3775 battery pack as part of the oximeter.
Cadmium is a hazardous substance. Do not incinerate or burn a NiCd battery pack. Dispose of a battery pack through an approved hazardous material disposal facility or by returning it to Datex-Ohmeda for reclamation.
Use only a Datex-Ohmeda battery charger to recharge or condition a Datex-Ohmeda NiCd battery pack. Do not use a Datex-Ohmeda battery charger to recharge or condition a NiCd battery pack that is not made by Datex-Ohmeda.
Do not expose a battery pack to any temperature that is hotter than you can tolerate to touch.
Fire hazard
To protect against fire hazard, replace the fuses in the battery charger only with fuses of the same type and local line voltage rating.
Electrical shock and flammability hazard
Before cleaning or servicing the oximeter, always turn it off.
1-12
Before cleaning or servicing the printer, always turn it off and disconnect the power cord from the AC mains power supply.
Before cleaning or servicing the battery charger, always disconnect the power cord from the AC mains power supply.
Page 19
1/Overview
Electrical shock hazard
To protect yourself from possible electric discharge when the cover of the unit is removed and the unit is receiving power, avoid contact with internal components.
Do not attempt to repair the printer, battery pack, or single-station battery charger. They are not serviceable parts.
The battery charger must be properly grounded.
Insert the three-connector plug on the power cord into a three-wire, grounded
hospital-grade receptacle only. If a three-wire receptacle is not available, a qualified electrician must install one in accordance with the governing electrical code.
Never remove the grounding connector from the power plug. The power cord
and plug must be intact and undamaged.
Do not use extension cords or adapters.
Patient safety
Never test or perform maintenance on equipment while it is being used to monitor a patient.
Operator safety
Do not handle hot or leaking battery packs or batteries.
Sensors
Discard a damaged sensor immediately. Do not repair a damaged sensor or use a sensor repaired by others.
To prevent injury or equipment damage, use only Datex-Ohmeda oximeter sensors approved for use with this oximeter. For complete information about the safe and appropriate use of a sensor, consult the instructions for that sensor.
Data validity
To prevent erroneous readings, do not use an inflated blood pressure cuff or arterial blood pressure measurement device on the same limb as the oximeter sensor.
Interfering substances, excessive ambient light, excessive motion, low perfusion, low signal strength, incorrect sensor placement, or electrical interference at the sensor site can result in the display of inaccurate SpO2 data.
1-13
Page 20
3770/3775 Pulse Oximeter Service Manual
Cautions
US Federal and Canadian laws restrict this device to sale by or on the order of a licensed medical practitioner.
Only qualified service personnel should perform the procedures in this manual.
Static sensitivity
Internal electronic components are susceptible to damage by electrostatic discharge. To avoid damage when disassembling the monitor, observe the standard precautions and procedures for handling static-sensitive components.
Battery packs
To prevent battery pack damage,
Never attempt to take a battery pack apart. Access only the battery
compartment for replacing batteries in the AA battery pack.
Never attempt to replace a cell in the NiCd battery pack. The set of eight cells
is matched for capacity and improper operation could occur.
Do not put a battery pack where it can be short-circuited by contact with
metal objects, such as in a pocket with keys.
Use only batteries recommended by Datex-Ohmeda in the LR6(AA) battery pack.
Service diagnostic mode is to be used by Datex-Ohmeda service personnel only.
Cleaning
Do not autoclave, pressure sterilize, or gas sterilize the oximeter, printer, battery pack, or battery charger.
Use cleaning solution sparingly. Do not immerse the oximeter, printer, battery pack, or battery charger in liquid. Excessive solution can flow into the device and cause damage to internal components.
When cleaning the display lens, do not use abrasive cleaning compounds or other materials that could damage the lens.
Do not use petroleum-based solutions or solutions containing acetone, ethanol, freon, trichloroethylene, or harsh solvents to clean the oximeter, printer, battery pack, battery charger, or the protective covering on these devices. These substances attack the devices’ materials and device failure may result.
When cleaning reusable oximeter sensors, follow all warnings and cautions stated in the sensor user instructions.
Disposable sensors are intended for single patient use only.
HP printer
To avoid damage to the printer, use only an AC adapter recommended by and available from Datex-Ohmeda.
1-14
Disposal
When the oximeter has reached the end of its useful life, dispose of it in accordance with local procedures and regulations.
Page 21
2/Troubleshooting
This chapter contains:
Tables containing messages that may be displayed on the oximeter and
conditions that may occur while using the oximeter.
A list of the possible causes for the message or condition.
Recommended actions for responding to the message or to correct the
condition. These actions include troubleshooting procedures to help you determine what needs to be done to correct the condition.
The tables in this chapter include the instruction to swap boards. This means you should replace one board at a time with a like board you know is functioning correctly to determine which board has failed. In most cases you are instructed to start with a specific board.
Also included in the tables are references to possible failure points. This information suggests component areas on a specific board that may have caused the malfunction. To repair the unit, always replace the board. Do not attempt to repair the board at the component level.
2-1
Page 22
3770/3775 Pulse Oximeter Service Manual
3770/3775 pulse oximeter
Message Cause(s) Recommended action(s)
BOOTCODE The unit failed the power-on
sequence of diagnostics.
PRINT <XX
(3775 only)
PROBE? The sensor is not attached or not
Trend is printing to the printer. XX represents the amount of time remaining (in minutes) to complete printing.
properly applied.
Replace the digital board. See chapter 3.
No action required.
Attach the sensor.
SERVICE
(alternates repeatedly with an error code/type message)
The sensor is not connected to the oximeter.
The sensor is not compatible with the oximeter.
The sensor has failed.
The system or one of its components has failed.
Check the sensor site; make sure the sensor is properly applied. Connect the sensor cable to the sensor connector on the oximeter.
Use a Datex-Ohmeda sensor that is compatible with the 3770/3775 oximeters.
Use another sensor.
If the PROBE? message is continuous,
check the wiring contacts between the sensor connector and the analog board.
If the condition persists, swap boards, starting with the analog board. Replace the failed board. See chapter 3.
Possible failure points:
Analog board: Analog path Digital board: A/D converter
Note: If this message occurs during printing, power the oximeter off and
then on. If you can’t restart printing and the message appears again, see below.
Swap boards and replace the failed board. See chapter 3.
2-2
Page 23
3770/3775 pulse oximeter (continued)
Condition Cause(s) Recommended action(s)
No display. Unit not receiving power.
Make sure a battery pack is correctly inserted in the oximeter and press the on/off button on the front panel of the oximeter.
2/Troubleshooting
Oximeter alpha­numeric display switches on/off continually.
Displayed SpO2 value flashes.
Low battery light flashes every three seconds.
NiCd battery pack needs recharging or AA pack needs new batteries.
If using the battery charger to power the oximeter, the NiCd battery pack’s contacts are not in contact with the power contacts in the battery charger.
Faulty battery pack.
Defective components on analog or digital board.
The low SpO2 or high SpO alarm limit has been violated.
This alert indicates that not much full-power time remains on the batteries.
2
Insert a charged battery pack.
Make sure the oximeter is inserted correctly in the battery charger. A beep sounds when correct contact is made.
If the condition persists, swap boards, starting with the analog board. Replace the failed board. See chapter 3.
Install a known functioning, fully charged battery pack. If the new battery pack corrects the condition, discard and recycle the failed pack.
If the condition persists, swap boards, starting with the analog board. Replace the failed board. See chapter 3.
Possible failure points:
Analog board: Power supplies
Digital board: Power on reset (POR)
Flashing ceases when the alarm condition is cleared. Make sure the alarm limit is within the high and low limits.
Remove the battery pack and replace it with a new, fully charged NiCd pack or AA pack with new batteries.
NiCd pack only—Recharge the battery pack by itself or by placing it, together with the oximeter, into the battery charger.
Waveform bar LEDs and signal strength indicator darken, then “rolling dots” appear in the main display.
Loss of signal quality has occurred. Low perfusion, electrosurgery device, or other interference has been detected.
Check sensor site.
Increase perfusion or relocate sensor.
Remove any device that may be causing signal interference.
2-3
Page 24
3770/3775 Pulse Oximeter Service Manual
3770/3775 pulse oximeter (continued)
Condition Cause(s) Recommended action(s)
Oximeter sounds a constant tone.
(3775 only) Can’t access menus.
Oximeter is not functioning at all.
Oximeter component failure. Swap boards, starting with the digital
board. Replace the failed board. See chapter 3.
Possible failure points:
Digital board:
Microprocessor Speaker External guard dog
Analog board:
Power
Faulty component or area on the analog board.
Faulty battery pack. Install a known functioning, fully
Replace the analog board. See chapter 3.
Possible failure point:
Analog board:
Rotary encoder assembly
charged battery pack. If the new battery pack corrects the condition, discard and recycle the failed pack.
Possible failure points:
Analog board:
PETE Power supplies
Digital board:
On/Off circuit; Power on reset (POR)
2-4
Page 25
Printer
Condition Cause(s) Recommended action(s)
Printer won’t print. Printer not powered on.
Make sure the printer AC adapter is properly connected to the printer and to the AC power supply. Slide the |/O printer button to | (on).
2/Troubleshooting
Printer prints the symbol that represents lost information:
Prints repeatedly over a single line.
Print head scans but no characters are printed.
Print head does not move while unit is powered on.
Red LED behind printer IR window is not lit when power is on.
No data in the trend buffer.
Printer and oximeter are not properly aligned; transmission is not being received at the printer.
Printer malfunction.
Printer cannot print fast enough to keep up with the data transfer rate of the oximeter. The printer battery power is running low.
Paper jam. Pull paper out and reload.
Printer malfunction: defective print head assembly thermal control.
Printer malfunction: jammed print head assembly.
Batteries are dead.
No AC power.
No action required. No data to print.
Make sure the printer and oximeter are properly positioned for infrared transmission. See the 3770/3775 Pulse
Oximeter User’s Manual.
Replace the printer. If the new printer corrects the condition, discard the failed printer. The printer is not a serviceable part.
If the condition persists, the oximeter is malfunctioning. Swap boards, starting with the digital board. Replace the failed board. See chapter 3.
Install new batteries. Always use AC power whenever possible.
Replace and dispose of the malfunctioning printer. The printer is not a serviceable part.
Replace and dispose of the malfunctioning printer. The printer is not a serviceable part.
Install new batteries.
Ensure that the AC adapter is plugged into the printer and into the AC power source.
Malfunctioning AC power adapter.
Printer malfunction.
Check/replace the AC power adapter.
Replace and dispose of the malfunctioning printer. The printer is not a serviceable part.
2-5
Page 26
3770/3775 Pulse Oximeter Service Manual
Battery charger monitoring station
Condition Cause(s) Recommended action(s)
Battery charge status indicator is a steady red light after a battery pack is inserted.
Battery charge status indicator is not lit after battery pack is inserted.
The NiCd battery pack is above or below the allowable temperature charging range.
Oximeter and/or NiCd battery pack is not inserted into the charger correctly.
The charger will automatically start charging with a flashing green light when the battery pack reaches the proper range.
Reinsert the oximeter and/or NiCd battery pack, aligning it so you can hear a beep from the charger station and the light is lit.
NiCd battery pack is not in place on the oximeter and/or in its slot in the monitoring station.
Voltage selected for the charger does not match the available local voltage.
NiCd battery pack is completely dead.
Battery charge status indicator is a steady yellow light after a battery pack is inserted.
Battery charge status indicator is a flashing yellow light after a battery pack is inserted.
Green power light is on but there is a continuous beep.
Charger is “chirping.” Voltage selection drum in the
The NiCd battery pack is above or below the allowable voltage charging range.
The battery pack may be fully charged already.
Charger is in conditioning mode.
Faulty conditioning switch or malfunctioning circuitry.
Voltage selection drum in the power module is not set for correct voltage.
power module is not set for correct voltage.
Install a NiCd battery pack on the oximeter and reinsert it.
Remove voltage selection drum and set correctly. Refer to the Battery Charger
Monitoring Station Service Manual.
Replace NiCd battery pack with a new one.
The charger will automatically start charging with a flashing green light when the battery pack reaches the proper range.
Remove the battery pack from the charger.
The charger will automatically start charging with a flashing green light when the discharge cycle is complete.
The monitoring station requires service. Refer to the Battery Charger Monitoring
Station Service Manual.
Remove voltage selection drum and set correctly. Refer to the Battery Charger
Monitoring Station Service Manual.
Remove voltage selection drum and set correctly. Refer to the Battery Charger
Monitoring Station Service Manual.
2-6
Page 27
Battery charger monitoring station (continued)
Condition Cause(s) Recommended action(s)
Green LED power indicator is not lit. Unit is not functioning.
Monitoring station is not connected to AC power.
Ensure that the power cord is plugged into the monitoring station and the AC power source.
2/Troubleshooting
Charger is functioning but slot not charging or conditioning as expected.
Fuse(s) are blown.
Voltage selection drum in the power module is not set for correct voltage.
Damaged contacts.
Damaged contacts.
If the LED stays at steady yellow when charging batteries, the battery pack may already be fully charged.
Replace the fuse(s) in the power inlet module. Refer to the Battery Charger
Monitoring Station Service Manual.
Remove the voltage selection drum and set correctly. Refer to the Battery
Charger Monitoring Station Service Manual.
Inspect the contacts.
If the fuses, voltage selection setting, and contacts are not causing the condition, open the charger and troubleshoot from inside. Refer to the Battery Charger
Monitoring Station Service Manual.
Inspect the contacts. All four prongs should protrude equally. If not, replace the contacts. Refer to the Battery
Charger Monitoring Station Service Manual.
No action is required.
Charger is not functioning at all.
When conditioning batteries and the LEDs don’t go to flashing yellow, the conditioning switch may be faulty.
Fuse(s) are blown or contacts are damaged.
Voltage selection drum in the power module is not set for correct voltage.
Replace the conditioning switch. Refer to the Battery Charger Monitoring
Station Service Manual.
Inspect the contacts. Replace the fuse(s) in the power inlet module. Refer to the
Battery Charger Monitoring Station Service Manual.
Remove the voltage selection drum and set correctly. Refer to the Battery
Charger Monitoring Station Service Manual.
If the condition persists, open the charger and troubleshoot from inside. Refer to the Battery Charger Monitoring
Station Service Manual.
2-7
Page 28
3770/3775 Pulse Oximeter Service Manual
Battery charger (single)
Condition Cause(s) Recommended action(s)
Battery charger status indicator is not lit after a battery pack is inserted.
NiCd battery pack is not inserted correctly.
Damaged contacts.
Reinsert NiCd battery pack.
Inspect contacts. All three prongs should protrude equally. If contacts are broken, discard charger. The charger is not a serviceable part.
Battery charger status indicator is flashing slowly after a battery pack is installed.
Battery charger status indicator is flashing rapidly after a battery pack is inserted.
Green power indicator is not lit.
Green power indicator is flashing.
Yellow LED is burned out.
Internal failure.
The NiCd battery pack is above or below the allowable temperature or voltage charging range.
The battery pack may be fully charged already.
Defective battery pack.
The battery pack may be fully charged already.
Charger is not connected to AC power.
Fuse(s) are blown.
The battery pack is defective. Remove and replace the battery pack.
Discard charger.
Discard charger.
The charger automatically starts charging with a steady yellow light when the battery pack reaches the proper range.
Remove the battery pack from the charger.
If the indicator continues to flash slowly, replace the battery pack.
Remove the battery pack.
Ensure that the power cord is plugged into the battery charger and the AC power source.
Discard charger. The charger is not a serviceable part.
2-8
Page 29
3/Service Policy and Repair Procedures
This chapter contains:
The Datex-Ohmeda service policy and the procedures to follow for obtaining
service.
Safety procedures you must follow when handling equipment that may be
contaminated and when making repairs.
Instructions for cleaning the oximeter, durable sensors, battery packs, and
battery chargers.
Procedures for inspecting the oximeter and replacing these parts: Alarm bar Analog board Bezels Digital board IR lens Menu select wheel and label
Instructions for checking the normal operation of the oximeter.
Service policy
Warranty repair and service must be performed by Datex-Ohmeda. When the warranty is not applicable, repairs are made by Datex-Ohmeda at the current list price for replacement parts plus a reasonable labor charge.
CAUTION: Only qualified service personnel should perform the procedures described in this manual.
Do not use malfunctioning equipment. Make all necessary repairs or have the monitor repaired by a Datex-Ohmeda service representative. For all repairs, use only genuine replacement parts and service kits sold by Datex-Ohmeda.
Ordering parts and obtaining service
Contact Datex-Ohmeda or your authorized service representative (see the back cover of this manual) to order parts or for assistance.
3-1
Page 30
3770/3775 Pulse Oximeter Service Manual
Packaging and returning equipment
If you are instructed to ship the monitor to Datex-Ohmeda or an authorized service office for repair, follow these steps:
1. Clean the monitor. See Cleaning procedures later in this chapter.
Make sure the monitor is completely dry before you pack it for shipment.
2. Package the monitor carefully for shipment (in the original shipping container,
if possible). Enclose these items:
3. You may be required to include the following items (when you call for
assistance, verify the shipping requirements):
A letter describing the problem in detail.
Person (name, telephone or fax number, and country) to contact for
questions about necessary repairs.
Ship to and Bill to information.
Warranty information (a copy of the invoice or other applicable
documentation).
Purchase order number to cover repairs (if out of warranty) or for tracking
purposes.
4. Ship the monitor as directed by your service office.
Safety guidelines
Read and follow each step of all test and repair procedures to ensure their proper and safe completion. Give special attention to all WARNINGS and CAUTIONS.
WARNING: Patient safety. Never test or perform maintenance on the equipment while it is being used to monitor a patient.
WARNING: Electrical shock and flammability hazard. Before cleaning or servicing the oximeter, always turn it off.
Before you start any procedure that involves disassembly of the oximeter, be sure to
Power off and disconnect the unit from any power supply.
Disconnect the sensor from the unit.
Clean the unit (see below).
After repairs are complete, verify that the device is functioning correctly. See Checking normal operation later in this chapter.
3-2
Page 31
Cleaning procedures
Oximeter, battery packs, and battery chargers
CAUTIONS:
Do not autoclave, pressure sterilize, or gas sterilize the oximeter, printer, battery pack, or
battery charger.
Use cleaning solution sparingly. Do not immerse the oximeter, printer, battery pack, or battery
charger in liquid. Excessive solution can flow into the device and cause damage to internal components.
When cleaning the display lens, do not use abrasive cleaning compounds or other materials
that could damage the lens.
Do not use petroleum-based solutions or solutions containing acetone, ethanol, freon,
trichloroethylene, or harsh solvents to clean the oximeter, printer, battery pack, battery charger, or the protective covering on these devices. These substances attack the devices’ materials and device failure may result.
Cleaning agents
Warm water Hydrogen peroxide solution Cidex Liquid soap/mild detergent Gluteraldehyde (4% or less) Windex Mild chlorine bleach solution Isopropyl alcohol Formula 409
3/Service Policy and Repair Procedures
®
®
®
Sensors
Never Use:
Acetone * Butyl alcohol * Denatured ethanol * Freon * Trichloroethylene
WARNING: Electrical shock and flammability hazard. Before cleaning or servicing the oximeter, always turn it off.
To clean the oximeter, power it off. Then, wipe it gently with a soft cloth dampened with any of the cleaning agents listed above.
WARNING: Electric shock and flammability hazard. Before cleaning the battery charger, always disconnect the power cord from the AC mains power supply.
Before cleaning the battery pack or battery charger, disconnect it from any power supply. Then wipe it gently with a soft cloth dampened with any of the cleaning agents listed above (do not spray the agent directly on any device).
To clean Datex-Ohmeda reusable oximeter sensors, follow the instructions provided with the sensor.
CAUTION: When cleaning oximeter sensors, follow all warnings and cautions stated in the sensor user instructions.
CAUTION: Disposable sensors are intended for single patient use only.
After cleaning a sensor, verify that it is functioning correctly. See Checking normal operation later in this chapter.
3-3
Page 32
3770/3775 Pulse Oximeter Service Manual
Inspecting the oximeter and replacing parts
Refer to chapter 4 for illustrations.
Follow the procedures in this section to inspect the oximeter for damage and replace defective parts.
CAUTION: Static sensitivity. Internal electronic components are susceptible to damage by electrostatic discharge. To avoid damage when disassembling the monitor, observe the standard precautions and procedures for handling static-sensitive components.
Tools and equipment
#1 Phillips screwdriver
Clear silicone rubber adhesive sealant (Dow Corning RTV 3145)
Disassembling the oximeter
Important: As you disassemble the oximeter, carefully set aside each part you plan to reinstall when you assemble the oximeter.
1. Turn off the oximeter and remove the battery pack.
2. Turn the oximeter face down and remove the four screws that secure the
bottom bezel to the top bezel.
3. Remove the bottom bezel.
4. Inspect the top and bottom bezels. If either bezel is damaged, replace it
during reassembly.
5. Inspect the interior for loose or broken parts and for foreign materials.
Remove foreign material and note those parts that need replacement.
6. Grasp the FlexConnect™ sensor connector and lift up to loosen it from the top
bezel.
7. 3770 oximeter: Lift out the board set.
8. 3775 oximeter:
Remove the IR lens (plastic window). Note: If the lens is held in place with sealant, remove the lens only if you
are replacing the lens and/or the top bezel or if the sealant is not intact.
Turn the oximeter face up and gently pry off the menu select wheel.
Remove the rotary encoder nut.
Lift the top bezel off the board set.
9. To separate the analog and digital boards from each other, grasp the board set
near the horizontal center and pull gently but firmly.
3775 oximeter: Avoid stressing the ribbon cable and its connection to the analog board. Do not flex the ribbon cable any more than absolutely necessary.
3-4
10. If you plan to replace the alarm bar or the top bezel, turn the bezel face up
and push the alarm bar through the bezel.
Page 33
Replacing parts
When the oximeter is disassembled, you can replace any part listed below with a new part as you assemble the oximeter:
Alarm bar
Analog board
Bezels
Digital board
IR lens
Menu select wheel and label
Assembling the oximeter
Important: During the manufacture of some 3770/3775 oximeters, a bead of silicone sealant was applied to the IR window and sensor connector cutouts on the top and bottom bezels (see Figures 3-1 and 3-2). The sealant was applied to ensure the waterproof integrity of these oximeters.
You may need to reapply sealant during assembly if disassembly of the oximeter results in loss of the original sealant or if the sealant is no longer intact. Follow the instructions in this section to check the oximeter for the presence of sealant and, if necessary, reapply sealant.
3/Service Policy and Repair Procedures
1. Make sure the rotary encoder, FlexConnect wires, the power on/off switch, the
alarm bar switch, all display and alphanumeric LEDs, and all other component parts are securely in place on their respective boards.
2. If you separated the alarm bar and top bezel during disassembly, position the
alarm bar in the underside of the bezel, and press it into place.
3. Gently but firmly connect the analog and digital boards to each other.
4. Turn the top bezel face down. Place the board set into the top bezel with the
digital board facing down.
5. 3775 oximeter:
Turn the oximeter face up.
Place the rotary encoder nut on the rotary encoder and tighten it.
Place the menu select wheel firmly on the encoder’s stem.
Turn the oximeter face down.
6. Push the FlexConnect sensor connector firmly into position in the top bezel.
3-5
Page 34
3770/3775 Pulse Oximeter Service Manual
7. Check the oximeter for the presence of sealant. If sealant is not present, go to
the next step.
If the sealant on the top bezel is not intact, apply new clear silicone
rubber adhesive sealant (Dow Corning RTV 3145) to the top bezel above the sensor connector. See Figure 3-1.
Figure 3-1. Applying sealant to the top bezel
3775 oximeter: Add sealant in the IR lens cutout on the top bezel, if necessary.
If the sealant on the bottom bezel is not intact, add sealant to the sensor
connector cutouts. See Figure 3-2.
3-6
Figure 3-2. Applying sealant to the bottom bezel
3775 oximeter: Add sealant to the IR lens cutouts on the bottom bezel, if
necessary.
8. 3775 oximeter: Slide the IR lens into its slot on the top bezel.
9. Position the FlexConnect wires so that they will not be crushed between the
ground ring on the analog board and the conductive boss that extends from the bezel.
10. Make sure there are no loose parts inside the top bezel. Then, place the
bottom bezel on the top bezel and install the screws that connect the bezels:
For the 3770, use four 4-20 x 0.5L screws.
For the 3775, use four 4-20 x 5/8L screws.
11. Check the operation of the oximeter. See Checking normal operation on the
next page.
Page 35
Checking normal operation
1. Install a battery pack on the oximeter. Power on the oximeter.
3775 oximeter: No alarms sound for the first two minutes after powering on the
unit.
2. Verify that the low-battery light is not lit.
3. Plug the sensor cable into the sensor cable connector on the oximeter.
4. Place the sensor on a finger and wait until the SpO2 and pulse rate values
appear in the main display.
5. Unplug the sensor cable from the oximeter.
3770 oximeter: Verify that dashes (- - -) are displayed. 3775 oximeter: Verify that the PROBE? message appears on the main display.
6. Plug the sensor cable back into the oximeter and wait for monitoring data to
reappear on the main display.
7. Remove the sensor from the finger.
3770 oximeter: Verify that dashes (- - -) are displayed. 3775 oximeter: Verify that the PROBE? message appears on the main display.
8. Reattach the sensor to the finger.
9. Verify that the signal strength is good to adequate (green or yellow signal
strength indicator is illuminated).
For more information, see Data validity and signal strength in the 3770/3775
Pulse Oximeter User’s Manual.
3/Service Policy and Repair Procedures
3-7
Page 36
3770/3775 Pulse Oximeter Service Manual
3-8
Page 37
4/Illustrated Parts
This chapter contains a list of service kits, a list of parts, and assembly illustrations for each oximeter.
Service kits—3770 and 3775 oximeters
Kit, PCA drawings..........................................................................................6050-0005-558
Includes board drawing, list of components, and schematics for the analog and digital boards
Kit, “Not for continuous monitoring” labels....................................6050-0003-466
Danish, Finnish, French, German, Italian, Spanish, Swedish
Kit, analog board, tested, 3775...............................................................6050-0003-667
PCA, analog board, 3775 Spacer, foam 0.25 x 0.125 x 0.75 in. (2) Spacer, foam 0.50 x 0.125 x 1.0 in. Spacer, foam 0.75 x 0.25 x 1.0 in.
Kit, analog board, tested, 3770...............................................................6050-0003-486
PCA, analog board, 3770 Spacer, foam 0.25 x 0.125 x 0.75 in. Spacer, foam 0.50 x 0.125 x 1.0 in. Spacer, foam 0.75 x 0.25 x 1.0 in.
Kit, digital board, tested, 3775................................................................6050-0003-480
PCA, digital board, 3775
Kit, digital board, tested, 3770................................................................6050-0003-481
PCA, digital board, 3770
Kit, bezel, bottom, 3770/3775................................................................6050-0004-078
Kit, bezel, top, 3775......................................................................................6050-0002-889
Note: This kit does not include the alarm bar,
on/off label, or wheel assembly.
Kit, bezel, top, 3770......................................................................................6050-0002-886
Assembly, alarm bar, 3775.......................................................................6050-0002-600
Assembly, nut, bearing, 3775..................................................................6050-0002-888
On/Off Label, 3775......................................................................................6050-0002-593
Kit, wheel/label, 3775.................................................................................6050-0003-666
Assy, wheel Wheel/select label
Kit, hardware, 3770/3775........................................................................6050-0003-780
Screw, tri-rnd, 4-20 x 0.5/8L (4) Screw, tri-rnd, 4-20 x 0.5L (4) Lens, IR
4-1
Page 38
3770/3775 Pulse Oximeter Service Manual
3775 oximeter assembly (REF 6050-0002-601, Rev. P)
3775 assembly parts
Each item number listed below identifies a component in the assembly illustration that follows.
Item Description
1 Assy, wheel, 3775 2 Assy, nut, bearing 3 Assy, top bezel, 3775 4 Assy, alarm bar 5 PCA, analog board, 3775
Includes: Rotary encoder switch 7 Lens, IR, 3775 9 Bezel, bottom
10 Label, serial port (not shown) 11 PCA, digital board, 3775
Includes: Speaker, 5 V, 0.354 inch diameter
13 Screw, tri-rnd, #4-20 x 5/8L, plastite (4) 14 Label, wheel/select 15 Label, on/off 16 Label, serial REF, 2.75 x 1.68 in. (not shown) 18 Spacer, foam, 0.25 x 0.125 x 0.75 in. (2) 19 Spacer, foam, 0.50 x 0.125 x 1.0 in. 20 Spacer, foam, 0.75 x 0.25 x 1.0 in. 21 O-ring seal 0.354 in. ID , .040 in. wall 25 Label, warranty void, 3770/3775
4-2
Page 39
3775 assembly illustration
3770/3775 Pulse Oximeter Service Manual
14
1
SEE VIEW A
2
ALIGN SHAFT FLATS WITH
FLATS IN MOLDED WHEEL
VIEW A
SEE VIEW B
11
3
Datex-Ohmeda
15
7
4
18 (2X)
19
20
ITEM 4
BOTTOM VIEW
Datex-Ohmeda
TOP VIEW
VIEW B
ITEM 3
21
5
13 (4X)
9
SEE VIEW C
10 REF 16 REF 25 REF
VIEW C
4-3
Page 40
3770/3775 Pulse Oximeter Service Manual
3770 oximeter assembly (REF 6050-0002-632, Rev. H)
3770 assembly parts
Each item number listed below identifies a component in the assembly illustration that follows.
Item Description
1 Label, On/Off 2 Assy, top bezel, 3770 3 Spacer, foam, 0.25 x 0.125 x 0.75 4 PCA, digital, 3770 5 PCA, analog, 3770 6 Bezel, bottom 8 Screw, tri-rnd, #4-20 x 0.5L, plastite (4)
10 Label, serial port 16 Label, serial REF, 2.75 x 1.68 in. (not shown) 18 Spacer, foam, 0.50 x 0.125 x 1.0 in. 19 Spacer, foam, 0.75 x 0.25 x 1.0 in. 20 Label, use warning, 3770 24 Label, warranty void, 3770/3775
4-4
Page 41
3770 assembly illustration
3770/3775 Pulse Oximeter Service Manual
24 REF
16 REF
10 REF
20
2
Datex-Ohmeda
4
5
1
3
18
19
VIEW A
8 (4X)
6
SEE VIEW A
4-5
Page 42
3770/3775 Pulse Oximeter Service Manual
4-6
Page 43
1315 West Century Drive
Louisville CO 80027-9560 USA
1 303 666 7001
Corporate Office
Datex-Ohmeda Division Instrumentarium Corp. PO Box 900 FIN-00031 Helsinki, Finland Tel 358 10 394 11 Fax 358 9 146 3310
Asia/Pacific
China
Datex-Ohmeda Pte. Ltd. Room B416, COFCO Plaza 8 Jianguomennei Avenue Beijing 100005, PR China Tel 86 10 6526 9773 Fax 86 10 6526 0653
Datex-Ohmeda Pte. Ltd. Room 1708, Yunlong Mansion No. 122 Luoguo Street Chengdu 610017, PR China Tel 86 28 661 4424 Fax 86 28 676 2703
Datex-Ohmeda Pte. Ltd. Room 1602, GIE Tower 403 Huan Shi Dong Road Guangzhou 510095, PR China Tel 86 20 8732 2521 Fax 86 20 8732 2518
Datex-Ohmeda Pte. Ltd. Room 2509, Lippo Plaza No. 222 Huaihai Road (M) Shanghai 200021, PR China Tel 86 21 5382 5657 Fax 86 21 5382 1619
Datex-Ohmeda Pte. Ltd. Room 809, Truroll Plaza Wusheng Road Wuhan 430033, PR China Tel 86 27 8571 2536 Fax 86 27 8571 2655
India
Datex-Ohmeda (India) Pvt. Ltd. Block EP & GP, Sector V Plot XI-16, Salt Lake City Calcutta 700091, India Tel 91 33 357 4002 Fax 91 33 357 4001
Indonesia
Datex-Ohmeda Pte. Ltd. Wisma Danamon Aetna Life 19th Fl. Jin. Jend Sudirman Kav. 45-46 Jakarta 12930, Indonesia Tel 62 21 575 0864 Fax 62 21 575 0865
Japan
Datex-Ohmeda K. K. TRC Annex 9F 6-1-1 Heiwajima, Ohta-ku Tokyo 143-0006, Japan Tel 81 3 5763 6801 Fax 81 3 5763 6838
Datex-Ohmeda K. K. Technical Center TRC A Bldg. AE 4-8 6-1-1 Heiwajima, Ohta-ku Tokyo 143-0006, Japan Tel 81 3 5763 6850 Fax 81 3 5763 6852
Korea
Datex-Ohmeda Pte. Ltd. 10th Floor, Sam Sung Building 36–1, Yoido-Dong, Youngdeungpo-Ku Seoul, Korea Tel 82 2 786 7421 Fax 82 2 786 7420
Malaysia
Datex-Ohmeda Pte. Ltd. Level 2 Bangunan O’Connor 13 Jalan 223 46100 Petaling Jaya Selangor, W. Malaysia Tel 60 3 754 7872 Fax 60 3 757 6948
Singapore
Datex-Ohmeda Pte. Ltd. 152 Beach Road #12–05/07 Gateway East Singapore 189721 Tel 65 391 8618 Fax 65 291 6618
Taiwan & Philippines
Datex-Ohmeda Pte. Ltd. 2nd Floor, No. 85 Chien-Kuo North Road, Sec. 2 Taipei, Taiwan Republic of China Tel 886-2 2515 0457 Fax 886-2 2501 9136
Thailand
Datex-Ohmeda Pte. Ltd. 12th Floor (Unit F) Grand Amarin Tower 1550 New Petchburi Road Makasan, Rajathevi Bangkok 10320, Thailand Tel 66 2 207 1012/13 Fax 66 2 207 1014
Vietnam
Datex-Ohmeda Pte. Ltd. 522G Nguyen Tri Phuong St. Ho Chi Minh City, Dist. 10 Vietnam Tel 848 865 5875 Fax 848 862 5501
Australia
Datex-Ohmeda Pty. Ltd. Units 1 & 2 149 Arthur Street PO Box 356 Homebush NSW 2140 Australia Tel 61 132 229 Fax 61 297 461796
Europe
France
Datex-Ohmeda S. A. S Parc de Pissaloup, BP 10 8 Avenue Jean d’Alembert F-78191 Trappes-Cedex, France Tel 33 1 30 68 60 00 Fax 33 1 30 68 60 01
Datex-Ohmeda S. A. S 17 rue Jean-Elysée Dupuy F-69410 Champagne au
Mont d’Or, France Tel 33 1 30 68 60 00 Fax 33 4 78 43 26 58
Germany
Datex-Ohmeda GmbH Auf der Hohe 49 Gewerbehof 49 D-47059 Duisburg, Germany Tel 49 2065 691 247 Fax 49 2065 691 255
Italy
Datex-Ohmeda S. p. A. Via Cassanese 100 20090 Segrate Milan, Italy Tel 39 2 21693431 Fax 39 2 26926226
Netherlands
Datex-Ohmeda B. V. Kantemarsweg 18 Post Box 22 3870 CA Hoevelaken Netherlands Tel 31 33 253 5404 Fax 31 33 253 7223
Spain
Datex-Ohmeda S. L. C/Manuel Tovar 26 28034 Madrid, Spain Tel 34 1 334 26 00 Fax 34 1 358 12 84
United Kingdom
Datex-Ohmeda Ltd. Ohmeda House 71 Great North Road Hatfield Hertfordshire AL9 5EN England Tel 44 1707 263570 Fax 44 1707 260191
Latin America, Caribbean
Datex-Ohmeda Latin America 10685 N. Kendall Drive Miami FL 33176, USA Tel 1 305 273 9940 Fax 1 305 273 4382
Middle East
Datex-Ohmeda Middle East Operations PO Box 5527 Dubai, United Arab Emirates Tel 97 14 822653 Fax 97 14 822659
North America
Canada
Datex-Ohmeda (Canada) Inc. 1093 Meyerside Drive, Unit 2 Mississauga, Ontario L5T 1J6 , Canada Tel 1 800 268 1472 Tel 1 905 565 8572 Fax 1 905 565 8592
United States
Customer Service
Datex-Ohmeda, Inc. Ohmeda Drive PO Box 7550 Madison WI 53707-7550, USA Tel 1 800 345 2700 Fax 1 608 221 4384
Technical Support
Datex-Ohmeda, Inc. Three Highwood Drive Tewksbury MA 01876, USA Tel 1 800 345 2755
Equipment Service Center
Datex-Ohmeda, Inc. 1315 West Century Drive Louisville Co 80027-9560, USA Tel 1 800 345 2755
Sales and Service
Datex-Ohmeda, Inc. Three Highwood Drive Tewksbury MA 01876, USA Tel 1 800 635-6099 Fax 1 978 640 0469
1/00 (388/F)
Loading...