Citizen CY 104, CX 120, CY 64, CY 54, CX 65 Operating Manual

...
Page 1
Operating Manual
CS-09-05-11/2
Analytical Balances | Carat Balances | Precision Balances | High Precision Industrial Balances
CITIZEN SCALE (I) PVT. LTD.
Parwanoo. Tel. : 022 - 4243 7777 | Fax : 022 - 4243 7900 |Web: www.citizenscales.com |Email: service@citizenscales.com
Subject to technical changes and to the availability of the accessories supplied with the instruments.
g
0.00
g
0.0
g
Page 2
1
2
Contents
1 Introduction ........................................................................................................ 3
1.1 Warning & Safety Precautions ........................................................................... 4
1.2 Getting Started 5...................................................................................................
1.3 Layout 6..................................................................................................................
2 Setting up the Balance 9......................................................................................
2.1 Location 12..............................................................................................................
2.2 Warm up 14..............................................................................................................
2.3 Keys of Balance 15..................................................................................................
3 Power On 17.............................................................................................................
4 Simple Weighing 19.................................................................................................
4.1 External Calibration 23............................................................................................
4.2 Internal Calibration 25............................................................................................
4.3 Calibration Test 27.................................................................................................
4.4 Calibration Test with Actual Calibration .......................................................... 29
5 Overview of Menu 32..............................................................................................
5.1 Operating Instruction ......................................................................................... 33
5.2 Description of Menu .......................................................................................... 34
6 Functions 44..............................................................................................................
6.1 Piece Counting 44...................................................................................................
6.2 Percent Weighing (%) ........................................................................................ 49
6.3 Custom Units 57........................................................................................................
6.4 Animal Weighing 61.................................................................................................
6.5 Formulation 65..........................................................................................................
6.6 Check weighing 69.................................................................................................
6.7 Totalization 73...........................................................................................................
6.8 Density Determination 77.......................................................................................
6.9 Pipette Calibration ............................................................................................. 83
6.10 Statistics ............................................................................................................... 89
7 Parameter Settings 93............................................................................................
7.1 Time & Date Settings 96..........................................................................................
7.2 ID / LID Settings 98.................................................................................................
7.3 Auto Calibration 99.................................................................................................
7.4 Windows Direct Communication 102.....................................................................
8 ISO / GLP - Compliant Printout ......................................................................... 105
9 Data Interface 107.....................................................................................................
9.1 Data Output Format 108.........................................................................................
9.2 Data Input Format 110.............................................................................................
9.3 Cabling Diagram 111...............................................................................................
10 Error Codes 112..........................................................................................................
11 Care & Maintenance 114.........................................................................................
12 Dimensional Drawings 119.......................................................................................
13 Specification 123......................................................................................................
14 Accessories ........................................................................................................ 134
Page 3
The models of the CX, CY, CT, CG and SSH Series are weighing instruments of special and high accuracy designed for the measurement of mass, covering a range from 0.01mg to 100 kg.
CX, CY, CT, CG and SSH models meet the highest requirements on the accuracy and reliability of weighing results through the following features:
– Filtering for unfavorable ambient
conditions, such as vibration, drafts, etc.
– Stable and repeatable weighing
results
– Excellent readability under any
lighting conditions
– Rugged, durable weighing system
These weighing instruments speed up your simple routine applications through following features:
Extremely fast response times
Built-in applications
• Counting
• Percent weighing
• Animal weighing
• Formulation
• Totalization
• Custom Unit
• Check Weighing
• Density Determination
• Pipette Calibration
• Statistics
1. Introduction
3
The balance has been constructed in accordance with the European Directives as well as international regulations and standards for operation of electrical equipment, electromagnetic compatibility, and stipulated safety requirements. Improper use or handling,however, can result in damage and/or injury.
To prevent damage to the equipment, please read these operating instructions carefully before using your balance.
Keep these instructions in a safe place. Follow the instructions below to ensure safe and trouble-free operation of your balance.
Do not use this balance/scale in a hazardous area/location.
If you use electrical equipment in installations and under ambient conditions requiring higher safety standards, you must comply with the provisions as specified in the applicable regulations for installation in your country.
Make sure that the voltage rating printed on the AC adapter is identical to your local line voltage.
Warning when using pre-wired RS-232 connecting cables: The pin assignments in RS-232 cables purchased from other manufacturers may be incompatible with Citzien balances. Be sure to check the pin assignment against the chart on page 111 before connecting the cable.
The only way to switch the power off
completely is to disconnect the AC adapter.
Connect only Citizen accessories
and options, as these are optimally designed for use with your Citizen balances.
– Note on Installation:
The operator shall be responsible for any modifications to Citizen equipment andfor any connections of cables or equipment not supplied
by Citizen and must check and, if necessary, correct these modifications and connections. On request, Citizen will provide information on the minimum operating specifications
Protect the DC adapter and the
weighing instrument from contact with liquids.
When cleaning your balance,
make sure that no liquid enters the balance housing; use only a slightly moistened cloth to clean the balance.
Do not open the balance/scale
housing. If the seal is broken, this will result in forfeiture of all claims under the manufacturer’s warranty.
If you have any problems with your
balance contact your local Citizen office, dealer or service center
1.1 Warnings and Safety precautions
4
– Total ease of operation
– Direct Communication with MS Excel,
MS Word and other windows application.
– ISO/GLP-compliant recording
capability for printouts
– Serial RS-232 port for optional
connection to a PC or Printer.
– Optional USB interface available on
request.
Page 4
CX / CY (0.1 mg) & CT
1. Keypad
2. Display
3. Model plate
4. Weighing pan (90 Fmm)
5. Draft shield (CX, CY & CT)
6. Leveling feet
7. Pan Cover
8. DC adapter socket
9. Provision for anti-theft device
10. Sprit Level
11. RS232C interface
12. Additional Display Sockets.
13. Foot Tare Switch Socket.
14. Calibration Switch (for Verified Balance)
15. Gem Bowl (CT)
Keys, operation and display are identical for all CITIZEN balances.
CY (1 mg with Draft shield)
1. Keypad
2. Display
3. Model plate
4. Weighing pan (128mm x 128mm)
5. Draft shield (CX, CY & CT)
6. Leveling feet
8. DC adapter socket
9. Provision for ant-itheft device
10. Spirit Level
11. RS232C interface
12. Additional Display Sockets.
13. Foot Tare Switch Socket.
14. Calibration Switch ( for Verified Balance)
Keys, operation and display are identical for all CITIZEN balances.
1.3 Layout
6
4
1
2
3
7
5
4
CX / CY (0.1 mg) & CT
CY (1 mg)
910
11
146 12 138
Storage and Shipping Conditions
Do not expose the balance/scale to extreme temperatures, blows, shocks, vibration or moisture.
Unpacking the Equipment
After unpacking the balance/scale, check it immediately for any visible damage as a result of rough handling during shipment
If you see any sign of damage:
It is a good idea to save the box and all parts of the packaging until you have successfully installed your balance. Only the original packaging provides the best protection for shipment. Before packing your balance, unplug all connected cables to prevent damage.
Contact your local Citizen office, dealer or service center
Accessories Supplied
The equipment supplied includes the following :
– Balance with display and control unit – Operating Manual – DC adapter – Gem bowl (CT scales only) – Pan Support (CX / CY / CT) – Weighing pan – Draft shield (CX / CY / CT) – Wind Shield (CX / CY / CT) – Base Plat S.S. (CX / CY / CT) – Pan Cover (CX / CT) – 1 Pair of Corner hole Cover
(CX / CY / CT)
Cautionary notes
Citizen balances may not be operated in hazardous areas.
Before attachment of the DC adapter, check whether the imprinted voltage value matches the local supply voltage. If it does not, contact your local CITIZEN dealers.
CITIZEN balance may only be used indoor in dry environment.
1.2 Getting Started
5
15
Page 5
8
7
CG
1. Keypad
2. Display
3. Model plate
4. Weighing pan (198 mmX 205 mm)
6. Leveling feet
8. DC adapter socket
9. Provision for anti-theft device
10. Spirit Level
11. RS232C interface
12. Additional Display Sockets.
13. Foot Tare Switch Socket.
14. Calibration Switch (for Verified Balance)
Keys, operation and display are identical for all CITIZEN balances.
910
11
146 12 138
CY (1mg)
1
2
3
4
CG
SSH
1. Keypad
2. Display
3. Model plate
4. Weighing pan (400mm x 300mm)
6. Leveling feet
8. DC adapter socket
9. Provision for anti-theft device
10. Spirit Level
11. RS232C interface
12. Additional Display Sockets.
13. Foot Tare Switch Socket.
14. Calibration Switch (for Verified Balance)
Keys, operation and display are identical for all CITIZEN balances.
11
6 12 8
1
2
3
4
10
CY (1 mg with wind shield)
1. Keypad
2. Display
3. Model plate
4. Weighing pan (128mm x 128mm)
5. Wind shield (CX, CY & CT)
6. Leveling feet
8. DC adapter socket
9. Provision for ant-itheft device
10. Spirit Level
11. RS232C interface
12. Additional Display Sockets.
13. Foot Tare Switch Socket.
14. Calibration Switch ( for Verified Balance)
Keys, operation and display are identical for all CITIZEN balances.
5
Page 6
2. Setting Up the balance
9
Balance with Analytical Draft-Shield (or Wind Shield) Chamber
• Position the Draft shield carefully on the balance.
Step 1: Open your New balance and draft shield you will see lock mounted on balance body and lock plate mounted below draft shield .
Step 2: Put your draft shield in same place and direction guided through RED color lines marked on this sticker, with gentle hand push & rotate draft shield clock wise as shown in figure.
: While keeping draft shield on balance ensure you do not damage weight loading point.
Note : Ensure there must not be any gap between balance and draft shield, if incase there is gap repeat again above procedure.
Step 3: Push fix both corner hole cover on bottom body of draft shield as shown in figure.
• The Procedure for setting up the balance with wind shield is same as that of one with draft shield
Weight
Loading
Point
Corner
Hole Cover
Lock Plate
Lock
Place the components listed below inside the chamber in the order given:
(For CX / CY (0.1mg) / CT Series Balances)
– Corner Cover – Base plate – Pan Cover – Weighing pan
Place the components listed below inside the chamber in the order given:
(For CY (1mg) Series Balances)
– Corner Cover – Base plate – Pan Support – Weighing pan
10
Draft shield
Wind shield
Page 7
The optimum location
The correct location makes an important contribution to the accuracy of the weighing results of high-resolution analytical and precision balances.
Hence, ensure a stable, vibration-free position as horizontal as possible.
Avoid
• Direct sunlight
• Excessive temperature fluctuations,
• Drafts (Power ----- Air Conditioning System, Fans can also cause drafts)
The best position is an a stable bench in a corner protected against drafts as far possible from doors, windows, radiators or the ventilation slots of air conditioners.
Anti-theft device
Citizen Balance are equipped with a lug for optional anti-theft device.
The anti-theft device (cable with lock) is suitable for all models. It is available from CITIZEN under order number CAD01.
2.1 Location
12
11
Place the components listed below inside the chamber in the order given: (For CG Series Balance)
– Bush – Weighing pan
Balances with a Rectangular Weighing Pan and a Weighing Capacity over 10 kg – Place the weighing pan on the balance/scale
Under Weighing
A port for a under weighing hanger is located on the bottom of the balance.
• Remove the under weighing cover from the bottom of the balance (as shown alongside)
• Suspend the below-balance weighing hook supplied from the hanger
• If necessary, install a shield for protection against drafts
Page 8
14
13
Leveling the Balance
Citizen balances have a level control and adjustable leveling feet to compensate for slight irregularities in the weighing bench surface. The balance is exactly horizontal when the air bubble is in middle.
Turn the two leveling feet as desire picture in diagram so that air bubble comes in middle. Air bubble at “12 o’clock” Turn both leveling feet counter- clockwise. Air bubble at “3 o’clock” Turn left leveling foot clockwise,
right leveling foot counterclockwise Air bubble at “6 o’clock” Turn both leveling feet clockwise Air bubble at “9 o’clock” Turn left leveling foot clockwise,
right leveling foot counterclockwise
Note : The balance must be re-leveled each time it is moved to a new location.
Leveling Balances with a Weighing Capacity up to 10 kg
Leveling Balances with a Weighing Capacity over 10 kg
• Adjust the leveling feet until the air bubble is centered within the circle of the level indicator
Connecting Electronic Peripheral Devices
Make absolutely sure to unplug the balance from DC power before you connect or disconnect a peripheral device (printer or PC)to or from the interface port.
Warmup Time
To deliver exact results, balance must warm up to operating temperature for as leasted below before the first weighing operation is carried out.
• CX / CY (0.1mg) / CT all Analytical models: at least 60 minutes
• CY (1mg) / CG / SSH all Precision models: at least 30 minutes
Using Verified Balances as Legal Measuring Instruments in the EU* balance must warm up for at least 24 hours after initial connection to DC power.
2.2 Warm Up
Page 9
16
15
2
11
10
12
13
8
9
3
4
17
14
15
16
5
2
7
18
1 6
1 On/Off key: Switches the display on / off 2 Tare key: Press here to tare the weight of any container so
that the readout shows the net weight of samples, also used to store reference settings. This key used to delete the statistics when F StAt mode.
3 Cancel Function : Delete (Clear Function) This key is
generally used to interrupt/cancel functions; for example: – to end an application program – to interrupt calibration/ adjustment routines
4 Toggle Key : Press here to change the Unit, Also used to
increment digit.
5 Cal Menu Function : Press here to start calibration/
adjustment or to enter user menu, Also use to shift flashing digit from left to right. this key is used in the F PiP mode to accept the volume during the calibration procedure.
6 Print Key : Press this key to send displayed values over the
built-in data interface to a DataPrint printer or a PC. 7 Weight Units 8 Weight readout in the selected weight unit
9 Capacity Bar : This indicates the total amount of
weight on the Pan
10 Stability Symbol : This symbol is displayed when the
weight place on the pan achieve stability
11 Asterik Symbol : This Symbol is displayed when the
display is locked
12 Stability Filer : This symbol indicates the chosen stability
filter
13 Symbol indicating that the Auto / Manual
calibration/adjustment function is active 14 Symbol indicating the active program 15 Battery Level Indicator : This symbol is indicates the
Current charge of the battery 16 Symbol indicating that a printout is being generated 17 Symbol indicating that a GLP compliant printout is
being generated 18 Seven segment readout indicating the active
program
2.3 Keys of Balance
Page 10
18
17
Switch on & off
Change Weighing Unit
Switch References
Print Weighing Results
Calibrate (adjust)
Call Menu
Weighing mode (operation)
Press briefly Press & Hold
Confirm Setting
Tare
Confirm Reference Setting
Abort / Cancel and Footer when GLP is ON
On / Off key
Toggling key
Tare keyCancel / Footer
Print key
CAL /
Menu key
Key Functionality in simple weighing mode
Citizen balances have several control levels the following section explains key functionality in simple weighing mode.
Short-form operating Instructions
Press key briefly
Press and hold key until desired display appears
Automatic procedure
Connect DC Adaptor and Power ON the balance.
It will display version number for software
It will display numeric countdown
It will display 88888888
The system initialization process will begin with the display indicating the current
progress. (INIT 1% to INIT 100%)
After the initialization is complete (100%) It will enter stand by mode &
display clock.
Stand by Mode
After Power ON and initial test balance will automatically come in stand by mode.
Press ON / OFF key to come to basic weighing
Press ON / OFF key in basic weighing to come back to Stand by Mode
3. Power ON
Page 11
Purpose
The basic weighing function is always accessible and can be used alone or in combination with an application program (counting, weighing in percent, etc.).
Features
– Taring the balance you can tare the balance within the entire weighing range. – Assigning IDs to weights (as needed) – Printing weights
4. Simple Weighing
19
20
Simple Weighing
4.0.2 Tarring
ð Place empty container on the balance.
ð The weight is displayed.
ð Press <Tare> key briefly, the balance displays zero
ð Add weighing sample to container, the net weight is displayed.
If the container is removed from the balance, the tare weight will be shown as a negative value.
4.0.1 Simple weighing
ð Place weighing sample on the weighing pan.
ð Wait until the stability symbol appears
ð Read the result.
ð Bar Graph will glow according to weight kept ON the PAN.
Tarring
Page 12
22
In case if stability is not achieved & user presses <TARE> key, display will shows “---------------” until stability is achieved.
Display then shows “0.0000 gm”
4.0.3 Switching of weight units
All Citizen balances can display weight values in different weight units.
The factory setting is Unit 1 (gm), Unit 2 (ct), Unit 3 (gm)
Switching between unit 1, unit 2 and unit 3
ðPress <TOGGLE> key the display switches to the 2nd weight unit.
ðPress <TOGGLE> key again, the display switches to the 3rd weight unit. OR
Application unit (If same is selected in menu)
Note : With certified balances the unit selection can be blocked following installation
if required by national legislation.
21
Switching of Weight Units
-----------------------­20-Jul-10 10:35AM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-----------------------­LID: 1111111 + 49.9999 g + 249.9990 ct + 1.763690 oz + 49.9998 g + 249.9990 ct + 1.763690 oz + 49.9999 g
-----------------------­20-Jul-10 10:36AM Name:
................
------------------------
When GLP ON
Print out generated when Unit Toggling is done between Unit1 (g), Unit2 (ct), Unit3 (Oz) in Simple Weighing.
+ 49.9999 g + 249.9990 ct + 1.763690 oz + 49.9998 g + 249.9990 ct + 1.763690 oz + 49.9999 g
When GLP OFF
Print out generated when Unit Toggling is done between Unit1 (g), Unit2 (ct), Unit3 (Oz) in Simple Weighing.
4.0.4 Simple weighing Print out
Note : 1) User cannot enter into the external calibration or menu when GLP is ON & footer has not been printed.
2) To print footer user will have to Press <CANCEL> key.
Page 13
If Balance is connected externally to PC or Data Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated after the completion of Calibration process.
Calibration Report
24
Unsuccessful Calibration
When GLP OFF
When GLP ON
-----------------------------­14-Jul-10 03:46PM Citizen Model CX 220 Ser.no. 9930508 Ver.no. R0.1.04 ID 1234567
-----------------------------­Calibration: External
W-ID ..................
Temperature 32.898'C Set + 20.00g
External Cal Failed
-----------------------------­14-Jul-09 03:46PM
Name: ................
------------------------------
Calibration : External
W-ID ............
Temperature 32.905'C Set + 20.00g
External Cal Failed
Successful Calibration
When GLP OFF
Calibration : External
W-ID ............
Temperature 32.905'C Set + 20.00g Diff. -0.01g External Cal Done Diff. 0.00g
When GLP ON
-----------------------------­14-Jul-10 03:46PM Citizen Model CX220 Ser.no. 9930508 Ver.no. R0.1.04 ID 1234567
-----------------------------­Calibration: External
W-ID ..................
Temperature 32.898'C Set + 20.00g Diff. 0.00g External Cal Done
Diff. 0.00g
-----------------------------­14-Jul-09 03:46PM
Name: ................
------------------------------
To obtain weighing results, the balance must be matched to the acceleration due to gravity at its location.
Calibration is necessary
ð Before the balance is used for the first time. ð At regular intervals during weighing operation. ð After a change in location.
Procedure
To obtain accurate results, the balance must be connected to the power supply and allowed to warm up to the operating temperature as described on Page No
Ensure that the weighing pan is unloaded and close the doors of the draft shield (if used). Balance should be Zero before calibration.
ð Have required calibration weight ready ð Press and hold <CAL> key, display, shows “CAL Et” ð Release <CAL> key now. ð The required calibration weight value will be displayed. ð Place calibration weight in center of pan.
The calibration (adjustment) is finished when “CAL donE” message is displayed. The balance is again in the weighing mode and ready for operation.
Note : With certified balances, the calibration can be disabled after installation if required by the national certification regulations.
The adjustment can be terminated at any time using the <CANCEL> key. The following message appears : ‘Abort’
14
23
Calibration
4.1 External Calibration (adjusting)
Page 14
To obtain weighing results, the balance must be matched to the acceleration due to gravity at its location.
Calibration is necessary
ð Before the balance is used for the first time. ð At regular intervals during weighing operation. ð After a change in location.
Procedure
To obtain accurate results, the balance must be connected to the power supply and allowed to warm up to the operating temperature as described on Page No
Ensure that the weighing pan is unloaded and close the doors of the draft shield (if used). Balance should be Zero before calibration.
ð Press and hold <CAL> key, display, shows “CAL Int” ð Release <CAL> key now.
Internal Calibration process Starts.....
- When the Internal Weight is being loaded “C” will be displayed on display.
- When the Internal Weight is being unloaded “CC” will be displayed on display.
Calibration is finished when ‘Int.done’ is message is displayed.
The adjustment can be terminated at any time using the <CANCEL> key. The following message appears : ‘Abort’
14
If Balance is connected externally to PC or Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated after the completion of Calibration process.
26
25
Internal Calibration
Successful Calibration Unsuccessful Calibration
When GLP OFF
Calibration: Internal
Start: Manual Temperature 29.449'C Diff. + 0.0009g
Internal Cal Done Diff. 0.0000g
When GLP ON
-----------------------­20-Jul-10 10:32AM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-----------------------­Calibration: Internal
Start: Manual Temperature 29.449'C Diff. + 0.0009g
Internal Cal Done Diff. 0.0000g
-----------------------­20-Jul-10 10:32AM Name:
................
------------------------
When GLP OFF
When GLP ON
-------------------------------­20-Jul-10 10:34AM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-----------------------­Calibration: Internal
Start: Manual Temperature 29.495'C
Internal Cal Failed
-----------------------­20-Jul-10 10:34AM Name:
................
------------------------
Calibration: Internal
Start: Manual Temperature 29.495'C
Internal Cal Failed
Internal Calibration Report
4.2 Internal Calibration
Page 15
If Balance is connected externally to PC or Data Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated after the completion of Calibration process.
Calibration test determines the difference between the actual weight and the measured weight Calibration test can be turned ON or OFF from the user menu. When ON, cal test would be performed on external or internal calibration whichever is selected in User Menu.
Procedure
ð Have required calibration weight ready ð Press and hold <CAL> key, display, shows “CAL Et” ð Release <CAL> key now. ð The required calibration weight is shown on the display. ð Place calibration weight in center of pan.
ð After the cal Test procedure is completed the difference between the actual & the
measured weight will be displayed on display.
The adjustment can be terminated at any time using the <CANCEL> key. The following message appears : ‘Abort’
Calibration Test Report
28
27
Calibration (adjusting)
When GLP ON
-------------------------------­14-Jul-10 00:03AM Citizen Model CX220 Ser.no. Ver.no. ID
-------------------------------­Calibration: External
W-ID ..................
Temperature Set Diff. Calibration Test Done
--------------------------------­14-Jul-10 00:03AM
Name: ................
---------------------------------
1111111
r0.1.04.01
860054081
30.710'C +200.0000g + 0.0047g
When GLP OFF
Calibration: External
W-ID ..................
Temperature Set Diff. Calibration Test Done
30.710'C +200.0000g + 0.0047g
4.3 Calibration Test
Page 16
To correct the weighing results, the TARE key need to be pressed when the difference is display upon pressing the TARE key. Actual calibration is performed ‘CAL done’ is displayed and the weighing results are corrected as shown alongside.
Procedure
ð Have required calibration weight ready ð Press and hold <CAL> key, display, shows “CAL Et” ð Release <CAL> key now. ð The required calibration weight is shown on the display. ð Place calibration weight in center of pan. ð The difference between the actual & the measured weight will be displayed. ð Press the Tare key when the difference is displayed. ð Actual Calibration is perform and Cal done is displayed.
The adjustment can be terminated at any time using the <CANCEL> key. The following message appears : ‘Abort’
If Balance is connected externally to PC or Printer via Rs232 Cable, successful or unsuccessful calibration report will be automatically generated after the completion of Calibration process.
Calibration Test Report
30
29
Calibration (adjusting)
Calibration : External
W-ID ............
Temperature 32.905'C Set + 200.0000g Diff. -0.1235g External Cal Done Diff. 0.0000g
-----------------------------------­14-Jul-10 03:46PM Citizen Model CX220 Ser.no. 9930508 Ver.no. R0.1.04 ID 1234567
-----------------------------------­Calibration: External
W-ID ..................
Temperature 32.898'C Set + 200.0000g Diff. -0.1235g External Cal Done
Diff. 0.0000g
-----------------------------------­14-Jul-09 03:46PM
Name: ................
------------------------------------
When GLP OFF
When GLP ON
4.4 Calibration Test with Actual Calibration
Page 17
31
Print User Menu
On / Off
Change Sub Menu Setting
Store and quit menu
Confirm Setting
Press briefly Press & Hold
Menu (called up with )
Increments the value of digit
Change Main menu options
Shifts the digit from left to right
Quit User Menu
Key Functionality in User Menu
The following section explains key functionality in User Menu mode.
On / Off keyOn / Off key
Toggling key
Tare key
Cancel key
Print keyPrint key
CAL /
Menu key
32
In this menu, you can select unit 1, 2, 3 or Application Program, adjust the stability filter, Calibration choice, Auto Zero Tracking, automatic shutdown and print setting.
Entry into menu : Press and hold <CAL> Key until “MENU” appears on the display. Release key, the 1st menu options “F none”
Select menu options Press <CAL> Key briefly. Press key repeatedly to view the current balance settings Main Menu Option.
Modify settings Press <Toggle> Key repeatedly until the desired setting appears.
Confirm settings Press <TARE> key breifly when desired setting appears.
Store settings Press and hold <TARE> Key until “StoreED” appears.
Abort Press <CANCEL> Key briefly, The balance returns to the weighing
mode without storing the changes.
3
4
1
2
lb
2
OZ
2
Ozt
GN
dwt
Weighing Unit 2 Stability Filter Calibration Menu Calibration Test
Unit 2 ct Stb 2
CAL ET CA .
Weighing Unit 1
Unit 1 gm
GLP
G P
ES
Reset
Reset
2
2
2
2
2
2
b
2
m
2
gm
2
mg
2
kg
2
1
2
Stop Bit
A. 1
A. 5
A. 10
Auto Off
A. OFF -
O
Baud Rate
9600
tlh(2)
GNdwt
tlt(2)
mo
b
m
mg
kg
ct
lb
OZ
Ozt
tlh(2)
tls(2)
Weighing Unit 3
Unit 3 gm
3
3
3
3
3
3
3
3
3
3
3
3
3
3
GNdwt
tlt(2)
mo
b
m
mg
kg
ct
lb
OZ
Ozt
tlh(2)
tls(2)
tls(2)
tlt(2)
mo
Auto Zero Tracking
Azt
Print Menu
P
A.
Parity
e
P
P
E E
P
a
P
spac
P
F
Pcs
F
F
edF
PPF
CsF
A
F
A
F
F f
SAF
F
Weighing Mode
F e
W. Application
%
Auto
F f
Auto
F
Auto
4800
19200
57600
5 Overview of Menu
Page 18
ð Press & hold CAL Key until “Menu” appears on the display. ð Release the CAL Key ð Existing preselected function is displayed from among the above mentioned 4
functions. This existing function is displayed with the stability indicator ON
ð Press Toggle Key to Toggle to the desired functions. ð Press Tare Key briefly, once the desired function is displayed.
Stability indicator will be ON for that particular function now.
ð Then Press & hold TARE Key once again till “STORED” is 5displayed on the display.
ð If more then one changes are to be made within the Menu, make all the desired
changed by pressing toggle key to toggle within the SUB-Menus & Pressing Tare
key briefly. Stability indicator will be ON for those changed Submenu Options Then finally, pressing & holding TARE key until “Stored” is displayed, will store all the changes made within different menu options.
Note : If user did not press & hold <TARE> key & “Stored” message appears on the
display then any change made by user will remain unchanged.
Above instructions are to be followed for all menu options except : “Reset”
33
Application Menu :
5.2.1 Special applications and functions
Your balance can do more than just weighing. Built-in applications and functions expand its possibilities and facilitate your daily work. You will learn these applications and functions in the following Sections.
Preselecting a function
In this menu option you can preselect a function which will then be available in the weighing mode (Unit 3) at a keystroke The following functions are available.
Piece counting
Your balance counts the pieces you add to or remove from the weighing container.
Percent weighing
Your balance allows you to weigh in to a preset value or determines percentage weight deviations .
Custom Unit : Your Balance allows you to weigh in any customized unit.
Animal Weighing : Your Balance allows you to weigh animals in motion. You have the
option of Auto and Manual Animal Weighing.
Checkweighing : Your balance allows you to check whether a sample corresponds to a preset target or is within a specific tolerance range.
5.2 Description of Menu
34
5.1 Operating Instruction
Page 19
Formulation : Your Balance allow individual weighing values to be summed to a total.
Totalization : Your balance allows you to weigh, individual weighing in piece which can
be summed to a total.
Density Determination : Your balance allows you to determine density of solids. Purity of gold can also be determined on the basis of density.
Pipette Calibration : Your balance allows you to calibrate the pipette used in laboratories for experimenting with liquids.
Statistics : Your balance allows you to obtain the statistics of the weighing data.
No function preselected You have no function available in the weighing mode (factory setting).
Note :
Above function will replace preset unit 3 automatically.
35
36
5.2.2 Unit 1, 2, 3 - selecting
The following weight units can be selected. With certified balance the unit selection can be blocked following installation if required by national legislation. Unit Conversion factor
g kg lb oz ozt GN dwt ct mg mo m H tl S tl t tl b
gram kilogram pound ounce troy ounce grain pennyweight carat Milligram momme mesghal Hong Kong taels Singapore taels Taiwan taels baht
1 1 kg 1 lb 1 oz 1 ozt 1 GN 1 dwt 1 ct 1 mg 1 mo 1 m 1 Htl 1 S tl 1 t tl 1 b
= = = = = = = = = = = = = =
1000g
453.59237g
28.349523125g
31.1034768g
0.06479891 g
1.555173843g
0.2g
0.001g
3.749999953 g
4.6083162 37,42900 g
37.799366256g
37.499995313g
15.1999998438g
Page 20
37
5.2.3 Adjusting the stability Filter
You can use the stability Filter to match the balance to the ambient conditions.
2 Setting with normal balance surroundings (factory setting) 3 Setting with unstable balance surroundings. The balance operates slower
but is less sensitive to external influences (vibrations, etc.) 4 Setting with extreme unstable balance surroundings. The balance operates even slower but is less sensitive to external influences (vibrations, etc.) 1 Setting with very stable balance surroundings. The balance operates very quickly
but is sensitive to external influences (vibrations, etc.)
5.2.4 Selecting Calibration Option
User can select any of the Two option for Calibration.
CAL ET If the user select this option then the machine will perform External Calibration
when the CAL key is press & hold to display “CAL Et” & at this moment if user release the key, user can enter into the External calibration.
CAL OFF When user press & hold CAL key, directly “Menu” appears on the display
without CAL Et option. Thus user cannot enter into the calibration process.
CAL INT If the user select this option then the machine will perform External Calibration
when the CAL key is press & hold to display “CAL Int” & at this moment if user release the key, user can enter into the Internal calibration.
5.2.5 Calibration Test
User Can select any Two calibration test option.
CAL T ON If the user select this option then the machine will perform Calibration Test
when the CAL key is press & hold Calibration Test will perform on Internal or External which ever is selected in calibration menu.
CAL T OFF Actual Calibration will be performed When the CAL key is press & hold .
5.2.6 Auto Zero Tracking
In this option, user can select whether to enable or disable Auto Zero Tracking (Factory setting is ON)
The auto zero tracking continuously corrects any deviation from the zero point for example which can be caused due to slight contamination (i.e. due to dust particles) on the weighing pan.
38
Page 21
5.2.7 Setting baud rate (data transmission rate)
The data transmission rate (baud rate) determines the speed of the transmission via the serial interface. The unit is baud (1 baud (bd) = 1 bit/second).
The following setting are available. bd 300, bd 600, bd 1200, bd 2400, bd 4800, bd 9600 (default), bd 19200, bd 57600
5.2.8 Setting Parity
This feature determines the parity of the transmitted data. The following settings are available: Parity, None (default), odd, even, mark, space.
5.2.9 Stop Bit Selection
This setting allows the user to select the stop bit for the transmitted data. The following settings are available: stop bit 1 (Default) and stop bit 2.
39
Print Menu
5.2.10 Selecting data transfer mode
In this menu block you tell the balance how a value should be transferred to a peripheral device (e.g. computer).
Prn. req The next possible stable value will be transferred after triggering of the
Print key.
Prn. Con All Values will be continuously transferred regardless of stability.
Prn. oFF Data Transfer mode switched off
Prn. Aut Next Possible stable value will be transfer automatically when the display
weight changes by + 1d.
Prn A.Ld Next possible stable value will be transferred automatically when the
display weight changes by +/- 10d
5.2.11 GLP Menu Setting
GLP oFF If the user select this option then the balance print format are not compliance to ISO/GLP/GMP.
GLP on If the user select this option then the balance print format are compliance to ISO/GLP/GMP.
Note
If user selected GLP ON do ensure that user print footer for entering into next transaction and enter into user Menu or Calibration.
40
Page 22
F none No Function Unit 1 gm Unit 2 ct Unit 3 gm Stb 2 balance environment set to Normal CAL Et CAL External CAL t CAL TEST OFF Azt ON Auto Zero Tracking set ON bd9600 Transmission rate Pr None Parity set to none Stpbt 1 Stop bit one Print Req GLP OFF A. oFF - no automatic standby
5.2.14 Reset of the balance setting
Reset balance setting and functions to factory setting (rESEt)
ð Select “rESEt” and Press <TOGGLE> key breifly, display show “YES” ð Press <TARE> key breifly, display show “stored”
The balance is now reset to the factory setting and returns to the weighing mode._
Factory Setting
42
5.2.12 GLP Menu Setting
GLP off : If the user select this option then the balance print format are not compliance to ISO/GLP/GMP.
GLP on : If the user select this option then the balance print format are compliance to ISO/GLP/GMP.
Note
If user selected GLP ON do ensure that user print footer for entering into next transaction and enter into user Menu or Calibration.
5.2.13 A. Off - Setting automatic standby
The automatic standby appreciably extends the operating life of your Battery (If Install) (Optional)
The balance will enter stand by mode if A-OFF is activated. The display on the balance remains zero for a specific time as selected in the A.OFF menu.
A. Off - : no automatic standby (factory setting) A. Off 1 : automatic standby after 1 minutes A. Off 5 : automatic standby after 5 minutes A. Off 10 : automatic standby after 10 minutes
41
Page 23
USER MENU PRINT OUT
Press the Print Key in the user Menu to Print the current status of user menu.
43
When GLP ON When GLP OFF
-----------------------------------­14-Jul-10 03:46PM Citizen Model Cy 220 Ser.no. 9930508 Ver.no. R0.1.04 ID 1234567
-----------------------------------­App : F Per Unit1 : g Unit2 : ct Unit3 : g Stb : 2 Cal : Ext Cal test : Off Azt : On Baudrate : 9600 Parity : None Stop bit : 1 Print : Request GLP : On Auto Off : Off
-----------------------------------­14-Jul-10 03:46PM
Name: ................
------------------------------------
App : F Per Unit1 : g Unit2 : ct Unit3 : g Stb : 2 Cal : Ext Cal test : Off Azt : On Baudrate : 9600 Parity : None Stop bit : 1 Print : Request GLP : On Auto Off : Off
6.1 Piece counting
Procedure
Piece counting presupposes that you have preselected the “F count” function in the menu
ðPlace the empty container on the pan. ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until “PCS” appears on the display.
Your balance now needs the weight of a reference number.
ðPress and hold the <TOGGLE> key until you are prompted to load the reference
pieces.
ðYour balance suggests the last set reference no. as the reference number.
You can accept this suggestion or select one of the other reference numbers available (5, 10, 20,50, 100 pieces, Free, wref, Update) by briefly pressing the <TOGGLE> key.
Note : We recommend you to choose a reference number as high as possible as the balance determines the average weight per piece and stores it as the reference weight. As it is seldom, that all pieces weigh exactly the same, the larger the reference number selected, the greater the accuracy of the reference weight. This application assumes uniform weight of each piece.
ðWhen you have placed exactly the same number of pieces on the weighing pan as
selected reference pieces press TARE key. As soon as the weighing result is stable, the calculated average piece weight is accepted as the reference.
ðAfter your balance has determined the piece weight, it displays the pieces as per
selected number and is now ready for piece counting. You can use the <TOGGLE> key at any time to switch the display between the piece number display, weighing unit 1 and weighing unit 2.
Note : The current set weight remains stored until it has been redetermined.
44
6. Functions
Page 24
6.1.1 FREE (Reference settings)
The FREE option allows the user to set any reference other than the fixed available reference. (Default value is 001 and maximum possible value is 999)
Procedure
ð Place the empty container on the pan. ð Press the <TARE> key briefly to tare the balance. ð Press the <TOGGLE> key briefly until “pcs” appears on the display.
Your balance now needs the weight of a reference number. ð Press and hold the <TOGGLE> key until you are prompted to load the reference pieces.
ð Your balance suggests the last set reference no. as the reference number. ð Press the <TOGGLE> key until FREE is displayed. ð Press the <TARE> key to enter FREE reference settings. ð Last stored FREE value first digit flashing. Flashing digit indicates that digit value or
place can be changed.
ð Press <TOGGLE> key ( ) to change the value of the Flashing digit. ð Press <CAL> key ( )to shift the flashing digit from Left to Right ð When you have placed exactly the same number of pieces on the weighing pan
as set in the FREE setting, press TARE key. As soon as the weighing result is stable, the calculated average piece weight is accepted as the reference
Now further weighing in PCS will be with respect to the reference calculation based upon the FREE setting.
45
6.1.2 wRef settings
If the piece weight is known, it can be entered directly. To do this, press the TARE key when the system displays wRef in the reference menu. An input field appears, in which the piece weight can be entered. Since the balance does not have to determine a reference by weighing, the result of the piece counting (the number of pieces currently on the weighing pan) is displayed right after the piece weight has been confirmed.
Procedure
ð Place the empty container on the pan. ð Press the <TARE> key briefly to tare the balance. ð Press the <TOGGLE> key briefly until “pcs” appears on the display.
Your balance now needs the weight of a reference number. ð Press and hold the <TOGGLE> key until you are prompted to load the reference pieces.
ð Your balance suggests the last set reference no. as the reference number. ð Press the <TOGGLE> key until wRef is displayed. ð Press the <TARE> key to enter wRef reference settings. ð Last stored wRef value first digit flashing. Flashing digit indicates that digit value or
place can be changed.
ð Press <TOGGLE> key ( ) to change the value of the Flashing digit. ð Press <CAL> key ( ) to shift the flashing digit from Left to Right ð Press tge <TARE> key to store the wRef value.
Now further weighing in PCS will be with respect to the wRef value.
46
Page 25
6.1.3 Updating Settings
The Update feature improves the precision of piece counting results. The average piece weight (reference) is recalculated with each reference optimization. Because the new pieces that have been placed in the weighing pan increase the basis for the calculation, the references, and therefore the result of the piece count, are more exact.
Select the UPDATE feature from the reference menu. The reference can be updated by pressing the CAL key which is confirmed by the displaying 'UPDATED' on the display.
Note
• The number of pieces placed in the weighing pan is larger than the reference piece number shown on the display.
• The number of pieces placed in the weighing pan is not greater than twice the most recently saved reference piece number (e.g. If the display shows 100 pcs the added pieces should not be greater than 200).
• Update feature cannot be selected if the previous selected reference was 'wRef'
47
Printouts generated when Unit Toggling is done between Application Unit (Pcs), Unit1 and Unit2 and Reference Weight is changed
When GLP ON
-----------------------------------­29-Jul-09 03:46PM Citizen Model CX 220 Ser.no. 9930508 Ver.no. r0.1.04 ID 1234567
-----------------------------------­LID: 1000000 NRef 5 pcs WRef 4.00 g Qnt + 5 pcs NRef 50 pcs WRef 0.40 g Qnt + 50 pcs Qnt + 25 pcs NRef 5 pcs WRef 2.00 g Qnt + 5 pcs
-----------------------------------­29-Jul-09 03:47PM
Name: ................
------------------------------------
Printouts generated when Unit Toggling is done between Application Unit (Pcs), Unit1 and Unit2 and Reference Weight is changed
When GLP OFF
nRef 5 pcs wRef 2.00 g Qnt + 5 pcs Qnt + 10 pcs nRef 10 pcs wRef 2.00 g Qnt + 10 pcs Qnt + 5 pcs nRef 50 pcs wRef 0.20 g Qnt + 50 pcs Qnt + 100 pcs
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the reference menu of Piece Counting function & make changes in the reference setting, automatically printout is generated on the Peripheral attached.
In the printout, reference number “nRef” and reference weight “wRef” is printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
48
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
Printout: Counting
nRef + 10 : Reference sample quantity wRef + 21.14 g : Reference weight
i.e. weight of one piece
Qnt + 500 pcs : Calculated quantity
Page 26
The “Percent weighing” function enables you to weigh in to a preset value (1, 10, 20,50, 100%, Free, 100r, 100L, AtroM, AtroD) and to determine deviations from this target value. Percent Weighing (%) presupposes that you have preselected the “F per” function in the menu
Procedure
ðPlace the empty container on the pan. ðPress the <TARE> key to tare the balance. ðPress the <TOGGLE> key briefly until “PER (%)” appears on the display.
Your balance now needs the weight of a reference percent (%).
ðPress and hold the <TOGGLE> key until you are prompted to load the reference
PER (%). ðYour balance suggests the last set reference % as the reference percent (%) You can accept this suggestion or select one of the other reference percent (%)
available (1,10, 20, 50, 100 %, Free, 100r, 100L, AtroM, AtroD) by briefly pressing the
<TOGGLE> key. ðDefault is 1%
The FREE option allows the user to set any reference other than the standard available reference.(Default value is 01.00 % and maximum possible value is 99.99%)
ðNow place reference sample on the pan. ðThen press <TARE> key. Until dashes are displayed, your balance is
calculating the reference
ðAfter your balance has determined the reference weight, it is ready for Percent
Weighing. You can use the <TOGGLE> key at any time to switch the display between the Percent (%) display, weighing unit 1 and weighing unit 2.
Note : The current set weight remains stored until it has been redetermined.
6.2.1 FREE (Reference settings)
The FREE option allows the user to set any reference other than the standard available reference. (Default value is 1.00% and maximum possible value is 99.99%)
Procedure
ðPlace the empty container on the pan. ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until “PER” appears on the display.
Your balance now needs the weight of a reference number. ðPress and hold the <TOGGLE> key until you are prompted to load the reference
pieces.
ðYour balance suggests the last set reference no. as the reference number. ðPress the <TOGGLE> key unit FREE is displayed. ðPress the <TARE> key to enter FREE reference settings. ðLast stored FREE value is displayed. Flashing digit indicates that digit value or
place can be changed.
ðPress <TOGGLE> key ( ) to change the value of the Flashing digit. ðPress <CAL> key ( )to shift the flashing digit from Left to Right ðWhen you have placed exactly the same number of pieces on the weighing pan
as set in the FREE setting, press TARE key.
As soon as the weighing result is stable, the calculated average piece weight is accepted as the reference
50
49
6.2 Percent Weighing (%)
Page 27
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the reference menu of Percent Weighing function & make changes in the reference setting, automatically printout is generated on the Peripheral attached. In the printout, reference percent “pRef” and reference weight “wRef” is printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
Printouts generated when Unit Toggling is done between Application Unit (%), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (%), Unit1 and Unit2 and Reference Weight is changed
When GLP ON
When GLP OFF
-----------------------------------­29-Jul-10 03:46PM Citizen Model CX 220 Ser.no. 9930508 Ver.no. R0.1.04 ID 1234567
-----------------------------------­LID: 1000000 Ref 10.00 % wRef 1.00 g Pct + 10.00 % + 10.0000 g + 50.0000 ct Pct + 20.00 % pRef 1.00 % wRef 20.00 g Pct + 1.00 % Pct + 0.50 %
-----------------------------------­29-Jul-10 03:47PM
Name: ................
------------------------------------
Ref 10.00 % wRef 1.00 g Pct + 10.00 % + 10.0000 g + 50.0000 ct Pct + 20.00 % pRef 1.00 % wRef 20.00 g Pct + 1.00 % Pct + 0.50 %
Printout: Counting
pRef 10% : Reference percentage wRef + 21.14 g : Reference weight Pct + 90.34% : Calculated percentage
52
51
Page 28
6.2.2 Differential Weighing
The Differential Weighing application is used to analyze changes in the weight of one or more samples. The first step is to determine the initial weight of the sample (weighing in). Selected components are then separated from or added to the sample. This includes procedures such as drying, centrifugation, filtering, incineration, vaporization, coating, etc. After the sample has been processed, it is re-weighed (residual weight). The balance then determines the difference between the two weighed values.
100L (Loss)
The moisture content of the sample is displayed (and printed out) as a percentage of the wet weight (= ww = initial weight = 100%). When the results are printed out, the moisture content is designated 100L % " (Loss) (e.g. –11.35 100.00L%) and shown as a negative value.
Dry weight DW – Wet weight WW
100L [0…-100%] = -------------------------------------------- *100%
Wet weight WW
100R (Residue)
The dry content of the sample is displayed (and printed out) as a percentage of the wet weight (= ww = initial weight = 100%). When the results are printed out, the dry content is designated "100R%" (Residue) (e.g. 88.65 100.00R%).
Dry weight DW
100R [100…0%] = -------------------------------------------- *100%
Wet weight WW
AtroM Moisture Content
The moisture content of the sample is displayed (and printed out) as a percentage of the dry weight ( = DW = final weight = 100%). When the results are printed out, the ATRO moisture content is designated “AtroM %" (ATRO Moisture Content) (e.g. –255.33 AtroM %) and shown as a negative value.
Dry weight DW – Wet weight WW
AtroM [0…-1000%] = -------------------------------------------- *100%
Dry weight DW
AtroD Dry Content (Wet weight)
The wet weight of the sample is displayed (and printed out) as a percentage of the dry weight (= DW = final weight = 100%). When the results are printed out, the ATRO dry content is designated “AtroD %" (ATRO Dry Content) (e.g. 312.56 AtroD %).
Wet weight WW
AtroD [100…1000%] = ----------------------------------------- *100%
Dry weight DW
53
100L
100R
AtroM[%]
100L% (Loss Content)
100R% (Residual Content)
%AtroM (Atro Moisture Content)
%AtroD (Atro DryContent)
AtroD [%]
54
Page 29
Percentage Weighing (%) (in 100R / 100L / AtroM / AtroD)
Percent Weighing (%) presupposes that you have preselected the “F per” function in the menu
Procedure
ðPress the <TARE> key to tare the balance. ðPress the <TOGGLE> key briefly until “PER (%)” appears on the display.
Your balance now needs the weight of a reference percent (%).
ðPress and hold the <TOGGLE> key until you are prompted to load the reference
PER (%). ðYour balance suggests the last set reference % as the reference percent (%) Press the <TOGGLE> key until the following option is displayed (100R / 100L / AtroM
/ AtroD)
ðNow place reference sample on the pan which is to be analyzed (Initial Weight). ðThen press <TARE> key. Until horizontal dashes are displayed, your balance is
calculating the reference. ðAfter your balance has determined the reference weight, it is ready for Percent
Weighing in Differential weighing. ðNow treat the sample which includes process like drying, centrifugation, filtering, incineration, vaporization, coating, etc. After the sample has been processed, re-
weigh it (residual weight). The balance then determines the difference between
the two weighed values.
You can use the <TOGGLE> key at any time to switch the display between the Percent (%) display, weighing unit 1 and weighing unit 2.
Note: If the current measured value on display mode is greater or less than the predefined limit value (i.e. greater than 999.99 % or less than –999.99 %) the balance displays Over range (----Or----)
55
100L% (Loss Content)
(Treat the sample& Re-Weigh it)
Printouts generated when Unit Toggling is done between Application Unit (Per %), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (Per %), Unit1 and Unit2 and Reference Weight is changed
When GLP ON When GLP OFF
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
----------------------------------­LID: 1111111 pRef ATROD % wRef 0.5000 g Pct + 100.00 % Pct + 500.00 % + 50.0000 g + 250.0000 ct
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
pRef ATROD % wRef 0.5000 g Pct + 100.00 % Pct + 500.00 % + 50.0000 g + 250.0000 ct
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the reference menu of Percent Weighing function & make changes in the reference setting, automatically printout is generated on the Peripheral attached. In the printout, reference percent “pRef” and reference weight “wRef” is printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
56
Page 30
The custom unit feature enables you to perform weighing in a customized unit i.e. weighing can now be performed in a unit other than standard available 15 units.
Procedure
Custom unit presupposes that you have selected the 'F Cust' in the user menu.
ðPlace the empty container on the pan. ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until “CSt” appears on the display.
Your balance now needs conversion factor, accuracy and LSD to perform weighing in custom unit.
ðPress and hold the <TOGGLE> key to browse through the custom unit setting menu.
Your balance suggests the last stored values for the factor, accuracy and LSD.
ðPress the <TARE> key to enter the specific setting.
Factor Setting
The factor value can be set to any user defined value except for zero.
ðPress the <TARE> key to enter the factor setting. ðPress the <TOGGLE> key ( ) to change the value of the flashing digit. ðPress the <CAL> key ( ) to change the flashing digit from left to right. ðPress the <PRINT> key to shift the decimal position in a cyclic way. ðAfter proper setting of factor press the <TARE> key.
Note: Please refer error conditions for errors occurred in storing the Factor, Accuracy and
LSD settings.
Accuracy Setting
ðPress the <TARE> key when the Acc is displayed. ðPress the <TOGGLE> key to browse through the standard available accuracy. ðYou can select any of the standard accuracy with the help of <TARE> key.
Standard accuracies available are
0.000001, 0.00001, 0.0001, 0.001, 0.01, 0.1, 1, 10, 100, 1000
The stability indicator alongside indicates the selected accuracy. The Accuracy Setting is for display purposes and not for calculation of Custom unit.
LSD Setting
ðPress the <TARE> key when LSD is displayed. ðPress the <TOGGLE> key to browse through the standard available LSD's. ðYou can select any of the standard available LSD (Least significant Digit) with the
help of <TARE> key.
Standard LSD's available are
1, 2, 5, 10, 20, 50, 100
The stability indicator alongside indicates the selected LSD.
To store the Factor, Accuracy and LSD values, press and hold the <TARE> key when the display shows Factor or Acc or LSD
The default settings are
Factor = 1.0000 (i.e. 1 gram) Accuracy = 0.01 LSD = 1
57
58
6.3 Custom Unit
Page 31
E.g. If the settings are as follows, Factor = 1.02356 Accuracy = 0.00 LSD = 50
Now if 50 gm of weight is loaded on the pan the calculation for displayed weight will be as follows, Weight * Factor = 50 * 1.023456 = 51.1728
The displayed weight will be 51.150
The second digit after decimal point will change in multiples of 5 because,
Accuracy * LSD = 50 * 0.001 = 0.05
You can use the <TOGGLE> key at any time to switch the display between custom unit display, weighing unit 1 and weighing unit
2. User can go to Standby mode by pressing on/off key and go to Simple Weighing with unit 1 by Cancel key and balance shows Current weights.
60
Printouts generated when Unit Toggling is done between Application Unit (Cst), Unit1 and Unit2.
Printouts generated when Unit Toggling is done between Application Unit (Cst), Unit1 and Unit2.
When GLP ON When GLP OFF
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-----------------------------------
LID: 1111111
Factor 3.023456 Accuracy 0.1 Lsd 2 + 151.2 cst + 50.0000 g + 250.000 ct Factor 1.023456 Accuracy 0.001 Lsd 50 + 61.400 cst + 51.200 cst
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
Factor 3.023456 Accuracy 0.1 Lsd 2 + 151.2 cst + 50.0000 g + 250.000 ct Factor 1.023456 Accuracy 0.001 Lsd 50 + 61.400 cst + 51.200 cst
If the balance is connected eternally to PC or Printer through RS 232 C then, whenever user enter into the Custom Unit Menu & make changes in the Factor, Accuracy and LSD setting, automatically printout is generated on the Peripheral attached.
In the printout, the new Factor, Accuracy and LSD values are printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
59
Page 32
The animal weighing feature enables you to perform weighing of unstable samples (live animals). The balance calculates the weight as the average of a defined number of individual weighing operations.
You can select from the two available animal weighing modes i.e. Auto animal weighing and manual animal weighing.
The weighing unit for animal weighing will be the same as selected for unit 1.
For Animal Weighing Process to start two conditions should be satisfied, the weight of the animal kept on the pan should be higher than 100 display increment i.e. if the balance capacity is 300 gm and accuracy is 0.0001 gm, then in Animal Weighing Process the weight of the animal should be above 100 * 0.0001 g = 0.001g and When two successive weight measured are within predefined tolerance range.
Number of weighing operations for calculation of an average Cnt can be set before the beginning of each series.
Balance returns to the basic weighing mode when unloaded; i.e., when the load is below the unload threshold
The Unload threshold is 50 display intervals.
6.4.1 Manual Animal Weighing
Manual Animal Weighing presupposes that you have selected the 'F Anl' in the user menu.
Procedure
ðPlace the empty container on the pan. ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until “ ”symbol appears on the display.
Your balance now needs to set the countdown value.
ðPress and hold the <TOGGLE> key to enter countdown options. The entire menu can
be accessed by pressing the <TOGGLE> key. The stability symbol indicates the currently selected countdown value.
ðPress the <TARE> key to select specific countdown value. ðKeep the animal on the pan, press the <CAL> key to start the animal weighing process
when both the condition required for animal weighing are met the countdown process will start, when the countdown time ends the average weight on animal is displayed with the display locked with flashing animal symbol.
Locked display is indicated by the flashing animal and AUTO symbol.
The countdown options available are,
• t-5
• t-10 (Default)
• t-20
• t-50
• t-100
You can use the <TOGGLE> key at any time to switch the display between animal weighing, weighing unit 1 and weighing unit 2. User can go to Standby mode by pressing on/off key and go to Simple Weighing with unit 1 by Cancel key and balance shows Current weights.
61
62
6.4 Animal Weighing
Page 33
6.4.2 Auto Animal Weighing
Auto Animal Weighing presupposes that you have selected the 'F Anl AUTO' in the user menu. Auto animal weighing proves to be beneficial when the balance is used majority for animal weighing and less for simple weighing thus reducing the time required for animal weighing.
Procedure
Place the empty container on the pan. Press the <TARE> key briefly to tare the balance.
Press the <TOGGLE> key briefly until “ ” and “AUTO” symbol appears on the display. Your balance now needs to set the countdown value.
Press and hold the <TOGGLE> key to enter countdown menu. The entire menu can
be accessed by the <TOGGLE> key.
The stability symbol indicates the currently selected countdown value.
Press the <TARE> key to select specific countdown value.
Keep the animal on the pan, when both the condition required for animal weighing
are met the countdown process will start, when the countdown time ends the
average weight on animal is displayed with the display locked. Thus there is no need of pressing a key to start the countdown process in the auto animal weighing mode. Locked display is indicated by the flashing animal and AUTO symbol. The countdown options available are,
t-5 (Auto), t-10 (Auto) Default, t-20 (Auto), t-50 (Auto), t-100(Auto)
You can use the <TOGGLE> key at any time to switch the display between animal weighing, weighing unit 1 and weighing unit 2. User can go to Standby mode by pressing on/off key and go to Simple Weighing with unit 1 by Cancel key and balance shows Current weights.
ð ð ð
ð
ð ð
63
Printouts generated when Unit Toggling is done between Application Unit (Anl), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (Anl), Unit1 and Unit2 and Reference Weight is changed
When GLP ON When GLP OFF
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
----------------------------------­Cnt 20 xNt + 50.0709 g + 50.0715 g + 250.3575 ct xNt + 50.0709 g
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
Cnt 20 xNt + 50.0709 g + 50.0715 g + 250.3575 ct xNt + 50.0709 g
Printout: Counting
Cnt 20 : Number of subweighing
operations
xNt + 50.0709 g : Calculated Average
If the balance is connected eternally to PC or Printer through RS 232 C then, whenever user enter into the Countdown Menu of Animal Weighing & make changes in the countdown time automatically printout is generated on the Peripheral attached.
In the printout, the new countdown value ’mDef’ is printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
64
Page 34
65
6.5.1 Manual Formulation
Manual formulation presupposes that you have selected the 'F Form' in the user menu.
Procedure
Place the empty container on the pan.
ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until Fol is displayed on the display. ðAdd weight on the pan. Weight can be added to summation when ‘ ’ is
displayed on the display i.e. when it is greater than 20d.
ðPress the <CAL> key to store the weight, the balance displays “n-1” indicating that 1st
weight is stored. The weight is tarred automatically and simultaneously print command is given.
E.g. if 9.9968 gm is added the printer output is as follows
N1 : 9.9968 g Tot : +9.9968 g
Further addition of weights will give the following output (addition of 20.0070 gm).
N2 : 20.0070 g Tot : +39.0038 g
ðTo view the total weight, press the <CAL> and <TOGGLE> key together. The print
command is given automatically N2 Tot : + 39.0038 g
66
6.5 Formulation
The formulation feature allows individual weighing values to be summed to a total.
User can select from two available formulation modes i.e. Manual formulation and auto formulation.
Maximum no of weights that can be summed is 99.
Store component weights with
• Display zeroed automatically after value is stored, and
• Automatic printout (print application parameters)
• Of the last component weight (net value) and
• Of the total weight (tare value)
Clear component memory when weighing series is canceled by pressing CANCEL key
Note :
Individual weights can be added into summation only if the weights are greater than 20d, this is indicated by ‘ ’ symbol.
The weighing unit for formulation will be the same as selected.
Page 35
67
Printouts generated when Unit Toggling is done between Application Unit (For), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (Forl), Unit1 and Unit2 and Reference Weight is changed
When GLP ON When GLP OFF
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
----------------------------------­N1 + 20.0000 g Tot + 20.0000 g + 20.0000 g + 100.0000 ct N2 + 50.0000 g Tot + 70.0000 g N 2 Tot + 70.0000 g
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
N1 + 20.0000 g Tot + 20.0000 g + 20.0000 g + 100.0000 ct N2 + 50.0000 g Tot + 70.0000 g N 2 Tot + 70.0000 g
Printout Configuration
N1 + 20.0000 g 1st component and its weight Tot + 20.0000 g Sum of components N2 + 50.0000 g 2nd component and its weight Tot + 70.0000 g Sum of components N 2 Total number of components Tot + 70.0000 g Total formulation weight
If the balance is connected eternally to PC or Printer through RS 232 C then, whenever user adds weight to the formulation procedure automatically printout is generated on the Peripheral attached. In the printout, the component added ‘N x ’along with the total sum ‘Tot’ is printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
6.5.2 Auto Formulation
Auto formulation presupposes that you have selected the 'F Form AUTO' in the user menu.
ðPlace the empty container on the pan. ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until Fol and AUTO is displayed on the display. ðAdd weight on the pan. Weight can be added to summation when “ ” is
displayed on the display i.e. when it is greater than 20d.
ðWhen the weight is stable the weight is stored in the formulation procedure and the
balance displays “n-1” indicating that 1st weight is stored. The weight is tarred automatically and simultaneously print command is given. Thus there is no need to press any to start the auto formulation procedure.
You can use the <TOGGLE> key at any time to switch the display between formulation, weighing unit 1 and weighing unit 2. User can go to Standby mode by pressing on/off key and go to Simple Weighing with unit 1 by CANCEL key and balance shows Current weights.
Auto formulation proves to be beneficial when the balance is used in majority for formulation weighing and less for simple weighing thus reducing the time required for formulation as compared to manual formulation.
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
68
Page 36
69
This feature is used to check whether a sample corresponds to a preset target or is with a specific tolerance range.
The results displayed can also be available on external port, which with the help of electronic processing can be used to control additional devices.
The check weighing procedure will always be carried out in unit 'gm'.
Check weighing presupposes that you have selected the 'F CHw' in the user menu.
Procedure
ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until “CW” is displayed on the display.
Your balance now needs the target and tolerance values. If entering the check weighing for the first time the target and tolerance setting menu will be displayed.
Target and Tolerance Settings
ðPress and hold the <TOGGLE> until the target and tolerance setting menu is
prompted.
ðPress the <TOGGLE> key to browse through the TARGET, HI, and LOW setting. ðPress the <TARE> key to enter specific setting. ðPress the <TOGGLE> key ( ) to change the value of the flashing digit. ðPress the <CAL> key ( ) to change the flashing digit from left to right. ðAfter proper setting of values press the <TARE> key. ðPress and hold the <TARE> key to store the values of target and tolerance values.
Note: To view the weight when the balance displays LL or HH press the <CAL> key press
the <CAL> key again to display LL or HH.
70
6.6 Check Weighing
Page 37
Printouts generated when Unit Toggling is done between Application Unit (Chw), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (Chw), Unit1 and Unit2 and Reference Weight is changed
When GLP ON When GLP OFF
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
----------------------------------­Target 150.0000 g Hi 160.0000 g Lo 140.0000 g
-------LL------- + 99.9979 g + 149.9979 g
-------HH------- + 200.0029 g
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
Target 150.0000 g Hi 160.0000 g Lo 140.0000 g
-------LL------- + 99.9979 g + 149.9979 g
-------HH------- + 200.0029 g
If the balance is connected eternally to PC or Printer through RS 232 C then, whenever user enter into the Target and Tolerance Setting Menu of Check Weighing Menu & make changes in the setting, automatically printout is generated on the Peripheral attached.
In the printout, Target and Tolerance value are printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
Load the weight on the pan and the balance will display the results according to the preset values of target and tolerance.
E.g. If the values are as follows,
TARGET = 150.0000 gm HI = 155.0000 gm LOW = 145.0000 gm
Now if 144.0000 gm is loaded on the pan the display shows LL indicating that the loaded weight is less than the low value set.
If 156.0000 gm is loaded on the pan the display shows HH indicating that the loaded weight is greater than the high value set.
While the balance displays LL or HH the buzzer will beep continuously to indicate that the weight is out of the tolerance band and the corresponding lines on the external will be enabled.
If the weight is in the tolerance range the balance will display the weight similar to simple weighing.
You can use the <TOGGLE> key at any time to switch the display between check weighing, weighing unit 1 and weighing unit 2.
User can go to Standby mode by pressing on/off key and go to Simple Weighing with unit 1 by CANCEL key and balance shows Current weights.
For High Range
For Low Range
71
72
Page 38
73
The totalization procedure allows individual weighing pieces to be summed to a total.
User can select from two available formulation modes i.e. Manual totalization and auto totalization. Maximum no of weights that can be summed is 99.
Individual weights an be added into summation only if the added pieces is greater
than 2, this is indicated by “ ” symbol.
6.7.1 Manual Totalization
Procedure
Manual totalization presupposes that you have selected the 'F tot' in the user menu.
ð Place the empty container on the pan. ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until “tot” is displayed on the display.
Your balance now needs the weight of a reference number.
ðPress and hold the <TOGGLE> key until you are prompted to load the reference pieces.
Store component pieces with
• Display zeroed automatically after value is stored, and
• Automatic printout (print application parameters)
• Of the last added pieces and
• Of the total number of pieces.
Clear component memory when weighing series is canceled by pressing CANCEL key
Note:
-
ðYour balance suggests the last set reference no. as the reference number.
You can accept this suggestion or select one of the other reference numbers available
(5, 10, 20, 50, 100 pieces) by briefly pressing the <TOGGLE> key.
ðNow place the selected number of reference pieces on the pan. ðWhen you have placed exactly the same number of pieces on the weighing pan as
selected reference pieces press TARE key.
As soon as the weighing result is stable, the calculated average piece weight is
accepted as the reference. ðAdd weight on the pan. Weight can be added to summation when “ ” is
displayed on the display i.e. when number of pieces is greater than 2.
st
ðPress the <CAL> key to store the weight, the balance displays “n-1” indicating that 1
weight is stored. The weight is tared automatically and simultaneously print command is
given.
E.g. if 20.0000 gm and the reference is selected as 5, the printer output is as follows
nRef 5 pcs
wRef 4 g
N1 + 5 pcs
Tot + 5 pcs
Further addition of weights will give the following output (addition of 10.0000 gm).
N2 + 2 pcs
Tot + 7 pcs ð To view the total weight, press the CAL and NEXT key together. The print command is
given automatically
N 2
Tot + 7 pcs
74
6.7 Totalization
Page 39
75
Printouts generated when Unit Toggling is done between Application Unit (Tot), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (Tot), Unit1 and Unit2 and Reference Weight is changed
When GLP ON When GLP OFF
------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-------------------------------
------------------------------­28-Jul-10 03:23PM Name:
................
-------------------------------
nRef 10 pcs wRef 2.0000 g N1 + 10 pcs Tot + 10 pcs N2 + 25 pcs Tot + 35 pcs N3 + 10 pcs Tot + 45 pcs N 3 Tot + 45 pcs
nRef 10 pcs wRef 2.0000 g N1 + 10 pcs Tot + 10 pcs N2 + 25 pcs Tot + 35 pcs N3 + 10 pcs Tot + 45 pcs N 3 Tot + 45 pcs
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the reference menu of Totalization function & make changes in the reference setting, automatically printout is generated on the Peripheral attached.
In the printout, reference number “nRef” and reference weight “wRef” is printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
6.7.2 Auto Totalization
Auto totalization presupposes that you have selected the 'F tot AUTO' in the user menu.
totalization totalization
totalization
totalization
Procedure
ðPlace the empty container on the pan. ðPress the <TARE> key briefly to tare the balance. ðPress the <TOGGLE> key briefly until “tot” and AUTO is displayed on the display.
Your balance now needs the weight of a reference number.
ðPress and hold the <TOGGLE> key until you are prompted to load the reference pieces. ðYour balance suggests the last selected reference number.
You can accept this suggestion or select one of the other reference numbers available (5, 10, 20, 50, 100 pieces) by briefly pressing the <TOGGLE> key.
ðNow place the selected number of reference pieces on the pan. ðWhen you have placed exactly the same number of pieces on the weighing pan as
selected reference pieces press TARE key. As soon as the weighing result is stable, the calculated average piece weight is accepted as the reference.
ðAdd weight on the pan. Weight can be added to summation when “ ” is
displayed on the display i.e. when number of pieces is greater than 2.
ðWhen the weight is stable the number of pieces is stored in the totalization procedure
st
and the balance displays “n-1” indicating that 1 weight is stored. The weight is tarred automatically and simultaneously print command is given. Thus there is no need to press any to start the auto totalization procedure.
Note : You can use the <TOGGLE> key at any time to switch the display between totalization, weighing unit 1 and weighing unit 2. User can go to Standby mode by pressing on/off key and go to Simple Weighing with unit 1 by CANCEL key and balance shows Current weights.
Auto proves to be beneficial when the balance is used in majority for weighing and less for simple weighing thus reducing the time required for as compared to manual
Note : 1) User cannot enter into the external calibration or menu when GLP is ON has not been printed.
2) To print footer user will have to Press <CANCEL> key.
& footer
76
Page 40
77
Liquid Setting
You can select from the three available options i.e. water, ethanol and other.
Press the <TOGGLE> key to change the liquid option. After proper selection of the liquid setting the <TARE> key.
The default option is distilled water.
Mode Setting
You can select from three options i.e. compensated, uncompensated and purity of gold.
Press the <TOGGLE> key to change the Mode setting. After proper selection of mode press the <TARE> key .
The default option is compensated.
The results of purity of gold will be shown in carats.
ð ð
ð ð
The density is determined applying the principle of Archimedes, which states that any body immersed in a fluid becomes lighter by an amount equal to the weight of the fluid that it has displaced. Purity of gold can also be determined on the basis of density.
The weighing unit of density determination will be 'grams'.
Density determination presupposes that you have selected the 'F Den' in the user menu.
To calculate the density of sample, the balance should know the type of liquid and its temperature, used to calculate the density of solid.
Procedure
Press and hold the <TOGGLE> until the functionality menu is prompted. Press the <TOGGLE> key to browse through the Temperature, Liquid and Mode settings.
Temperature Setting
Press the <TARE> key when the 'temp' is displayed. Press the <TOGGLE> key ( ) to change the value of the flashing digit. Press the <CAL> key ( ) to change the flashing digit from left to right. After proper setting of values press the <TARE> key. The default value of temperature is 25.0ºC This setting is alterable only when the liquid selected is water or ethanol. If the liquid selected is 'Other' the temperature setting will ‘nA’ i.e. Not applicable.
ð ð
ð ð ð ð ð ð
78
6.8 Density Determination
Page 41
79
Density Determination Procedure
ð ð ð ð
ð
ð
ð
ð
Attach the Density determination Kit to the Balance Press the <TARE> key briefly to TARE the balance. Press the <TOGGLE> key briefly until “Den” appears on the display. Press the <CAL> key to start the density determination procedure. The Balance now asks for weight of the sample in air. (With Wt.Air flashing on the display every 10 seconds) Place the sample on density determination kit, When the weight is stable press the CAL key to accept the weight of sample in air. Now the balance asks for the weight of sample in liquid (With Wt.Liqd flashing on the display every 10 seconds) Now immerse the sample in the liquid, When the weight is stable press the CAL key to accept the weight of sample in liquid.
The balance will show the results upon the selected mode i.e. compensated, uncompensated or gold.
Density Calculations
With compensation for air density
ñ = A * (ñ0 - ñl)
+ ñl
(A - B)
ñ = density of sample A = weight of sample in air B = weight of sample in auxiliary liquid ñ0 = density of auxiliary liquid ñl = air density (0.0012 g/cm3)
e.g.
ñ = (A / (A - B)) (ñ0 - ñl) + ñl = (20.000 / (20.000-18.935)) (0.99689 – 0.0012) + 0.0012 = 18.699 g/cm3
Without compensation for air density
ñ = A
ñ0
*
(A - B)
ñ = ñ0 (A / (A - B)) e.g.
ñ = ñ0 (A / (A - B)) = (20.000 / (20.000-18.935)) (0.99689) = 18.721 g/cm3
Add Wt. in Air
Add Wt. in Liquid
Display for compensated density
Display for uncompensated density
80
Page 42
81
Printouts generated when Unit Toggling is done between Application Unit (Den with Compensated Mode Liquid as Ethanol), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (Den with Uncompensated Mode and Liquid as None), Unit1 and Unit2 and Reference Weight is changed
When GLP ON When GLP ON
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-----------------------------------
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
Mode COMPENSATED Liquid ETHANOL Temperature 28.0 Den + 0.6995g/c3 + 18.9350 g + 94.675 ct
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-----------------------------------
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
Mode UNCOMPENSATED Liquid OTHER Temperature NA Den + 0.6994g/c3 + 18.9350 g + 94.675 ct
Printouts generated when Unit Toggling is done between Application Unit (Den), Unit1 and Unit2 and Reference Weight is changed
Printouts generated when Unit Toggling is done between Application Unit (Den), Unit1 and Unit2 and Reference Weight is changed
When GLP ON When GLP OFF
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
-----------------------------------
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
Mode GOLD Liquid WATER Temperature 25.0 Pur + 23.1 ct + 18.9350 g + 94.675 ct
Mode GOLD Liquid WATER Temperature 25.0 Pur + 23.1 ct + 18.9350 g + 94.675 ct
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the menu of Density function & make changes in the, automatically printout is generated on the Peripheral attached.
In the printout, Temperature, Liquid and Mode are printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the menu of Density function & make changes in the, automatically printout is generated on the Peripheral attached.
In the printout, Temperature, Liquid and Mode are printed. After this user can Press Print Key to Print the reading on the display. Also user can Toggle to other unit through Toggle key & Press Print Key to get the print out of those corresponding units.
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
Note : 1) User cannot enter into the external calibration or menu when GLP is ON
2) To print footer user will have to Press <CANCEL> key.
& footer has not been printed.
82
Page 43
83
Note: To enter the settings below, it is assumed that the user has already performed the above procedure and the scale is in F PiP mode.
Count Cycle Settings
ð Press <TOGGLE> key to enter F PiP settings. ð Press < TARE> to enter the count settings. ð ‘n’ represents the number of calibration cycles the user intends to perform. ‘n’ can have any value between 5 and 15 (including both).
ð Press <TOGGLE> to select the required value of ‘n’. ð Press <TARE> once to select the value of ‘n’ and return to F PiP settings. ð The default count value is 5.
Temperature settings
ð Press <TOGGLE> key to enter F PiP settings. ð Select ‘tEMP’ in F PiP settings by pressing <TOGGLE> key. ð Press < TARE> to enter the ‘tEMP’ settings. ð The user can select from a list of 30 predefined temperatures ranging from 15.0ºC to
30.0ºC with a step interval of 0.5ºC. ð Change the temperature value by pressing <TOGGLE> key and then select it pressing <TARE>. This will bring you back to F PiP settings. ð The default temperature is 25.0ºC
Pressure Settings
ð Press <TOGGLE> key to enter F PiP settings. ð Select ‘PrES’ in F PiP settings by pressing <TOGGLE> key. ð Press < TARE> to enter the ‘PrES’ settings. ð The user can select from a list of 7 predefined pressures ranging from 800 hPa to 1050
hPa with step interval of 50 hPa. ð Change the pressure value by pressing the <TOGGLE> key and then select it pressing <TARE>. This will bring you back to ‘FPiP’ settings. ð The default pressure is 800 hPa.
In laboratories, where pipettes are used for experimenting with liquids, it becomes important to calibrate the pipette. Thus this feature enables the user to calibrate the pipette. The feature “pipette calibration” can be activated by selecting the “F PiP” function in the menu. By pressing the CAL/MENU key for 4 seconds the user can enter the User menu and select the “F PiP” function. Note: The below procedure presupposes that you have selected the ‘F PiP ‘function in the user menu.
Procedure
ð Place the empty container on the pan. ð Press the < TARE> key briefly to tare the balance. ð Press the <TOGGLE> key for 2 seconds to enter the F PiP function.
84
6.9 Pipette Calibration
Page 44
85
Note:
ðIt is assumed that the user has already performed the above procedure and the s c a l e
is in F PiP mode.
ðWhen the pipette calibration procedure process has not started the system will
perform normal weighing with the unit as gm.
Calibration procedure:
ðThe calibration procedure repeats itself 'n' times where 'n' is the count entered in the
count cycle settings (in FPiP settings).
ðOnce in F PiP mode, press the <CAL/MENU> key to start the calibration process. ðOnce the <MENU/CAL> key is pressed, it will ask for the initial volume V
0.
ðFor a calibration procedure of 'n' times, the scale will ask for V 'n' times. Each of these
0
value will be stored as 'PLACE n'.
st
ðThus, for the 1 time, place the initial volume and press <CAL/MENU> key once the
stability is achieved. This is accepted as 'PLACE 1'.
ðFollow the above procedure for 'PLACE 1' to 'PLACE n'. ðRepeat the above procedure for half volume V and full volume V each 'n' times
half full
respectively.
ðIn case of a successful calibration the scale will display 'PiP done'. If not, then it will
ask to repeat the step wrongly performed.
Volume Settings:
Initial volume V
0:
ðPress <TOGGLE> key to enter F PiP settings. ðSelect 'V ' in F PiP settings by pressing <TOGGLE> key.
0
ðPress < TARE> to enter the 'V ' settings.
0
ðThe user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key. ðPressing the <CAL/MENU> key will shift the cursor to the right in a cyclic way and pressing
the <TOGGLE> key will increment the digit.
ðPress the <TARE> key to select the value. This will bring you back to the F PiP menu ðThe default volume is, V = 10.00 uL
0
Half volume V
Half :
ðPress <TOGGLE> key to enter F PiP settings. ðSelect 'V ' in F PiP settings by pressing <TOGGLE> key.
Half
ðPress < TARE> to enter the 'V ' settings.
Half
ðThe user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key. ð Pressing the <CAL/MENU> key will shift the cursor right in a cyclic way and pressing the
<TOGGLE> key will increment the digit.
ðPress the <TARE> key to select the value. This will bring you back to the F PiP menu ðThe default volume is, V = 11.00 uL
Half
Full volume, V
Full. :
ðPress <TOGGLE> key to enter F PiP settings. ðSelect 'V ' in F PiP settings by pressing <TOGGLE> key.
Full.
ðPress < TARE> to enter the 'V ' settings.
Full.
ðThe user can change the volume with the help of <CAL/MENU> key and <TOGGLE>key. ð Pressing the <CAL/MENU> key will shift the cursor right in a cyclic way and pressing the
<TOGGLE> key will increment the digit.
ðPress the <TARE> key to select the value. This will bring you back to the F PiP menu ðThe default volume is, V = 12.00 UL.
Full
86
Page 45
87
88
PIPETTE CAL. RESULTS
Count : 5 Cnt Temp. : 15.0 °C Pressure : 800 hPa Results : Vmin
1 10016.45 uL 2 10016.45 uL 3 10016.35 uL 4 10016.35 uL 5 10016.35 uL
Vmin = 10.00 uL Va = 10016.39 uL Es = 10006.39 uL Es% = 100063.88 % Sr = 0.05 uL CV = 0.00 %
------------------------
Results : V1/2
1 15023.12 uL 2 15023.12 uL 3 15024.42 uL 4 15024.42 uL 5 15024.12 uL
V1/2 = 11.00 uL Va = 15023.84 uL Es = 15012.84 uL Es% = 136480.37 % Sr = 0.67 uL CV = 0.00 %
------------------------
Results : Vmax
1 20032.40 uL 2 20032.30 uL 3 20032.30 uL 4 20032.20 uL 5 20032.20 uL
Vmax = 12.00 uL Va = 20032.28 uL Es = 20020.28 uL Es% = 166835.63 % Sr = 0.08 uL CV = 0.00 %
-----------------------­Legend
Va : Mean Value Es : Systematic Error Es% : Es expressed as % of nominal value Sr : Standard Deviation CV : Coefficient of Variation
------------------------
11-Apr-11 00:18
Citizen
Model 00Ct603 Ser.no. 1012652 Ver.no. r0.1.5.8 ID 1234567
-----------------------­PIPETTE CAL. RESULTS
Count : 5 Cnt Temp. : 15.0 °C Pressure : 800 hPa
Results : Vmin
1 10016.45 uL 2 10016.15 uL 3 10015.75 uL 4 10015.75 uL 5 10015.65 uL
Vmin = 10.00 uL Va = 10015.95 uL Es = 10005.95 uL Es% = 100059.47 % Sr = 0.34 uL CV = 0.00 %
------------------------
Results : V1/2
1 15023.82 uL 2 15023.62 uL 3 15023.62 uL 4 15023.62 uL 5 15023.72 uL
V1/2 = 11.00 uL Va = 15023.68 uL Es = 15012.68 uL Es% = 136478.92 % Sr = 0.09 uL CV = 0.00 %
------------------------
Results : Vmax
1 20031.89 uL 2 20031.89 uL 3 20031.89 uL 4 20031.59 uL 5 20031.59 uL
Vmax = 12.00 uL Va = 20031.77 uL Es = 20019.77 uL Es% = 166831.45 % Sr = 0.16 uL CV = 0.00 %
-----------------------­Legend
Va : Mean Value Es : Systematic Error Es% : Es expressed as % of nominal value Sr : Standard Deviation CV : Coefficient of Variation
-----------------------­11-Apr-11 00:18 Name:
................
------------------------
WITH GLP OFF
WITH GLP ON
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the menu of Pipette Calibration & make changes in it, automatically printout is generated on the Peripheral attached.
The following are the parameters that are Printed along with the readings :
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the menu of Pipette Calibration & make changes in it, automatically printout is generated on the Peripheral attached.
The following are the parameters that are Printed along with the readings :
Page 46
89
With this feature, the user can obtain the statistics of the data stored in the scale. These statistics includes the details such as number of readings 'n', Average, Minimum value, Maximum value, Standard deviation, Difference and Co efficient of Variance. The feature 'F StAt' can be activated by selecting the 'F StAt' function in the menu. This can be done by pressing the <CAL/MENU> key for 4 seconds and changing the feature by <TOGGLE> key. Press <TARE> key to select 'F StAt'. press <TARE> for 2 seconds to store the selection. The below procedure presupposes that you have selected the 'F StAt' function in the user menu.
Procedure
ðPlace the empty container on the pan. ðPress the < TARE> key briefly to tare the balance. ðPress the <TOGGLE> key for entering the 'F StAt' function.
Note: To use the 'F StAt' function, it is assumed that the user has already performed the above procedure and the scale is in 'F StAt' mode.
Statistics procedure
ðThe user can now place weight on the pan and press the <CAL/MENU> key once
the stability is achieved.
ðThe scale should display 'n-1' (where 'n' is the number of the current weight) and will
retain this as first weight.
ð“rmv” will be displayed on screen for 2 seconds to instruct the user to unload the
weight.
ðThe next weight will be taken into statistics only after user has taken off the weight
from the pan, such that the weight on the pan should now be 0.0000g.
ðRepeat the above procedure for rest of the data entries.
90
6.10 Statistics
Page 47
91
Reading the statistics
ðAt any point of time, the user can press and hold the <TOGGLE> key for 2 seconds to
go through the statistics.
st
ðThe screen will display the 1 parameter as number of readings 'n' along with the star
symbol which indicates that the user is in the statistics result mode.
ðThe user can now go through the other parameters such as average, Min value, M a x
value, Standard deviation, Difference, Co efficient of Variance by pressing the <TOGGLE> key.
ðThe user can come out of the statistics mode by pressing the < CANCEL > key.
Deleting statistics:
ðThe user can delete the statistics by pressing the <TARE> key for 2 seconds when in
'StAt' mode.
ðThis will delete the previous data entries and start again. ðThe entire data is will be sent to the terminal or to the printer before it gets deleted.
Note:
ðThe minimum weight on the pan should be 10d (where'd' is the accuracy of the
scale), if less than that, then the weight will not be taken into consideration for calculating statistics.
ðThe statistics will be deleted if the system goes into standby mode (i.e. if the auto-off
feature is active or by power on-off).
ðThe user can also print the statistics by pressing the <PRINT> key when in 'StAt' mode.
----------------------------------­28-Jul-10 03:19PM Citizen Model CX 220 Ser.no. 9223102 Ver.no. r0.1.5.3 ID 1234567
----------------------------------- 1 + 1.5750 g 2 + 2.3500 g 3 + 7.8950 g 4 + 4.1750 g
----------------------------------­n4 max 7.8950 g min 1.5750 g avg 3.9987 std 2.8169 var 7.9349
----------------------------------­28-Jul-10 03:23PM Name:
................
-----------------------------------
----------------------------------- 1 + 1.5750 g 2 + 2.3500 g 3 + 7.8950 g 4 + 4.1750 g
----------------------------------­n4 max 7.8950 g min 1.5750 g avg 3.9987 std 2.8169 var 7.9349
-----------------------------------
WITH GLP ON WITH GLP OFF
92
If the balance is connected externally to PC or Printer through RS 232 C then, whenever user enter into the menu of Statistics & presses in the print key , the user can obtain the list of parameters shown in the below example :
Page 48
93
Key Functionality in parameter settings mode
7. Parameter Settings
The following section explains key functionality in parameter settings mode.
To Change Time Format (AM / PM / 24 hours) in Time Settings
Change Sub Menu Setting
Store and quit menu (Auto Cal Menu)
Confirm Setting
Press briefly Press & Hold
Increments the value of digit
Change Main menu options
Shifts the digit from left to right
Quit the Current Parameter Menu
ID / LID SETTINGS
TIME DATE SETTINGS
TIME
DATE
TIME
0
0.5 C
1
0
C
2
0
C
5
0
C
OFF
AUTOCAL SETTINGS
0.5 Hr
1 Hr
3 Hr
24 Hr
OFF
TEMP.
P CAL ON
P CAL OFF
AUTOCAL OFF
AUTO CAL ON
ID
LID
94
On / Off key
Toggling key
Tare keyCancel / Footer
Print key
CAL /
Menu key
Page 49
Parameter Settings
By accessing the parameter menus the user can change the following settings.
• ID and LID settings.
• Time and Date Settings.
• Auto Calibration and Power On Calibration Settings.
Operating Instructions
These menus can be accessed by pressing the PRINT key or CAL key when all the
characters of the display when coming out of stand by mode or Power On.
95
ID & LID SETTINGS
TIME & DATE SETTINGS
AUTO CAL SETTINGS
& POWER ON
In this menu, User can set the Clock.
Clock setting consist of 2 settings. They are
TIME : In this submenu user can set the time in hours, minutes & seconds
AM, PM & 24 hrs.
DATE : In this submenu user can set the date, Month & Years
Operating Instructions
ð Press the PRINT key for 2 sec when coming out from stand by
or Power on mode.
7.1.1 SET TIME
ð Current Time is displayed with first digit flashing. Flashing digit indicates that digit
value or place can be changed.
ð Press TOGGLE key ( ) to change the value of the Flashing digit.
ð Press CAL key ( ) to shift the flashing digit from Left to Right
ð After proper setting of time in hours, minutes & seconds respectively
for zeroes starting from left, press Tare key
ð Press the PRINT key to Change the format AM, PM & 24hrs.
96
7.1 Time & Date Setting
Page 50
7.1.2 SET DATE
ð Press TARE key, “date” is display
ð Press TARE key, current date is displayed with first digit flashing. Flashing digit indicates
that digit value or place can be changed.
ð Press TOGGLE key ( ) to change the value of the Flashing digit.
ð Press CAL key ( ) to shift the flashing digit from Left to Right
ð After proper setting of date in day, month & year respectively for zeroes starting from left,
ð Press TARE key to set the date
In this menu user can set the identification number & Lot Identification number.
Operating Instructions
ð Press the PRINT key briefly when coming out from stand by or Power on mode.
7.2.1 SET ID
ð Last stored ID is displayed with first digit flashing. Flashing digit indicates that digit
value or place can be changed.
ð Press TOGGLE key ( ) to change the value of the Flashing digit.
ð Press CAL key ( ) to shift the flashing digit from Left to Right
ð Press the TARE key to store ID Value
7.2.2 SET LID
ð Last stored LID is displayed with first digit flashing. Flashing digit indicates that digit
value or place can be changed.
ð Press TOGGLE key ( ) to change the value of the Flashing digit.
ð Press CAL key ( ) to shift the flashing digit from Left to Right
ð Press the TARE key to store LID Value
97
98
7.2 ID / LID Setting
Page 51
7.3.2 Auto CAL Temperature Settings
Select Auto CAL ON and press the <TARE< key, now press the <TOGGLE> key when the
LCM displays TIME, press the <TARE> key to enter temperature settings.
0
User can set temp. value = 0.5 C.
0
User can set temp. value = 1 C.
0
User can set temp. value = 2 C.
0
User can set temp. value = 5 C.
User can set temp. value = OFF
Auto Cal triggered due to temp. change will take place irrespective of CAL test is On
or oFF
Note : The Above setting is available with balances with internal calibration.
100
ð Auto Internal Calibration can be triggered automatically, by the following
• Time
• Temperature
• Power on Calibration
Operating Instruction
ð Press the CAL key briefly when coming out from stand by or Power on mode.
7.3.1 Auto CAL Time Settings
Select Auto CAL ON and press the <TARE> key, now press the <TARE> key when the
Displays shows TIME to enter time settings.
User can set time value = 0.5 hrs.
User can set time value = 1 hrs.
User can set time value = 2 hrs.
User can set time value = 3 hrs.
User can set time value = 24 hrs.
User can set time value = OFF
Auto Cal triggered due to time will take place irrespective of cal test on or OFF
Note : The Above setting is available with balances with internal calibration.
99
7.3 Auto Calibration
Page 52
102
The windows direct communication function enables you to send the data from the balance directly to any windows application program for e.g. Microsoft word, excel etc.
The printer settings in the user menu will be applicable to the windows direct communication also i.e. Data Transfer Mode, Baud rate, Parity, Stop Bit and GLP
The settings attributed to windows direct communication are
• Unit ON or OFF.
• Separator type ENTER or TAB.
To enable windows direct comm­unication, make sure that you have turned it on from the windows side as well.
Enter control panel. Open ACCESSIBILITY OPTIONS from control panel. In the general tab turn on serial key option. Set the baud rate and COM port from the settings option.
ð ð
ð
ð
7.3.3 Power on Calibration
Press the CAL key when the LCM displays Auto CAL On or Off enter Power ON
calibration options.
This setting enables the user to turn on or turn off power on calibration.
Power on calibration will take place every time the balance is powered on.
Power on Cal will take place irrespective of whether CAL Test is On or Off.
Note : The Above setting is available with balances with internal calibration.
101
ð
ð
Click OK to accept the settings for
serial key.
Click APPLY and then OK to save the Accessibility options.
7.4 Windows Direct Communication
Page 53
103
Windows direct communication settings (Balance Side)
Press and hold the <PRINT> key in simple weighing mode until the windows print menu is prompted.
Press the <TOGGLE> key briefly to change the windows option to ON of OFF. The default option is OFF.
Windows Unit settings
Select Windows print option as ON and press the <TARE> key, now press the <TARE> key when the LCM displays UNIT to enter unit settings.
User can set Unit option as ON (Along with the numerical value the unit will also be sent to windows).
User can set Unit option as OFF (Only the numerical value will be sent to windows and not the unit).
ð
ð
Windows Separator settings
Select Windows print option as ON and press the <TARE> key, now press the <TARE> key when the LCM displays 'Separat' to enter Separator settings.
User can set SEPERATOR option as ENTER (After every value printed on the windows side an ENTER command is given so every subsequent data will print on new line, in Excel every new data will be printed in new row).
User can set SEPERATOR option as TAB (After every value printed on the windows side a TAB command is given so every subsequent data will printed with tab, in Excel every new data will be printed in new column).
104
Page 54
Features
You can have the parameters pertaining to the ambient weighing conditions printed before (GLP header) and after (GLP footer) the values of a weighing series. These parameters include:
GLP header: – Date – Time at beginning of measurement – Balance manufacturer – Balance model – Balance serial number – Software version number – Identification number of the current sampling operation
GLP footer: – Date – Time at end of measurement – Field for operator signature
The record is output to a CITIZEN data printer or a computer.
Settings
Set print option to request & GLP ON
Function Keys
Press the Print key to output header and first measured value.
End an Application: Output GLP Footer : Press Cancel Key End an application program Press Cancel key
------------------------ Dotted line 28-Jul-10 03:19PM Date / Time (beginning of measurement) Citizen Balance Manufacturer Model CX 220 Balance Model Ser.no. 9223102 Balance Serial Number Ver.no. r0.1.5.3 Software Version ID 1234567 ID
------------------------ Dotted line LID: 1111111 Lot ID nRef 170 pcs Counting : Reference Sample Quantity wRef 0.2945 g Counting : Reference Weight Qnt + 170 pcs Counting Result + 50.0650 g Weighing Result + 250.3250 ct Weighing Result
------------------------ Dotted line 28-Jul-10 03:23PM Date / Time (end of measurement) Name: Name of Operator
................
------------------------ Dotted line
------------------------ Dotted line 20-Jul-10 10:32AM Date / Time (beginning of measurement) Citizen Balance Manufacturer Model CX 220 Balance Model Ser.no. 9223102 Balance Serial Number Ver.no. r0.1.5.3 Software Version ID 1234567 ID
------------------------ Dotted line Calibration: External Calibration / Adjustment Mode
Blank Line W-ID Temperature Set + 200.0000 g Calibration Weight Diff. + 0.1234 g Diff. After Calibration External Cal Done Confirmation of Completed Calibration
Blank Line Diff. 0.0000g Difference from Nominal Value after Calibration
------------------------ Dotted line 20-Jul-10 10:32AM Date / Time (end of measurement) Name: Name of Operator
................
------------------------ Dotted line
The ISO/GLP-compliant record can contain the following lines:
............ Weight ID
32.905'C Temperature
The ISO/GLP-compliant record can contain the following lines:
105
106
8. ISO/GLP-compliant Printout/Record
Page 55
The following characters can be output, depending on the characters displayed on the balance :
Normal Operation Position 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16 17 18 19 20 21 22 23 24 25 26
****+DDDDDDDDDDDDDDD* UUULFCR
or ****-DDDDDDDDDDDDDDD* UUULFCR
or l l l * + DDDDDDDDDDDDDDD* UUULFCR
* : Space CR : Carriage Return D : Digit or letter LF : Line Feed U : Unit Symbol l : ID code Character
Special Codes Position 1 234567891011121314151617181920212223242526
-------OL------ LFCR
or -------UL------ LFCR
or -------Or------ LFCR
-------LL------ LFCR
-------HH------ LFCR
OL : Space UL : Carriage Return Or : Digit or letter LL : Line Feed HH : Unit Symbol
107
Purpose
Your balance is equipped with an interface port for connection to a computer or other peripheral device. You can use an on-line computer to change, start and/or monitor the functions of the balance and theapplication programs.
Features
• Type of interface: Serial interface
• Operating mode: Full duplex
• Standard: RS-232
• Transmission rates: 300; 600; 1,200; 2,400; 4,800; 9,600; 19,200 baud 57600
• Parity: Mark, space, odd, even, none
• Character format: 1 start bit, 8-bit ASCII, parity,1 or 2 stop bits
• Handshake: None
• Data output format of the balance : 26 characters
Factory settings:
Transmission rate: 1,200 baud (9600)
Parity: Odd (none)
Stop bits: 1 stop bit
Handshake: None
Print manually/automatically: Manual at stability
Preparation
• See “Pin Assignments”
108
9. Data Interface
9.1 Output Format with 26 Characters
Page 56
Data Output Examples + 123.4567 g
Position 1234567891011121314151617181920212223242526
+ 123. 4567 gLFCR
or + 617. 2835 ct LFCR
or N1 + 20. 0000 gLFCR
or Tot + 60. 0000 gLFCR
Position 1 - 4 : ID code Character or Space Position 5 : Sign Position 6 - 20 : Weight with Decimal Point; leading zeros = space Position 21 : Space Position 22 - 24 : Unit Symbol or Space Position 25 : Line Feed Position 26 : Carriage Return
109
110
You can connect a computer to your balance to send commands via the balance interface port to control balance functions and applications. Format for commands
[ Command Code Data ]
•[: it shows start of command frame.
•Command Code: it shows which functionality to be carried out for this command frame.
•Data: This field in frame is optional and it is intended to provide data information between Bi-directional communications.
•]: it shows end of command frame.
Commands
[W] : If host computer send this command to balance then balance will Send weight with current unit.
[T] : If host computer send this command then balance will do taring in balance. If stability is not achieved within 45 second then it comes to that specific feature till that time it shows “-------“on LCM.
9.2 Data Input Format
ID code ID Code
characters Meaning Characters Meaning
nRef Counting: Reference sample quantity N1 Formulation, Net : N1
wRef Piece Counting, Percentage Weighing : N Formulation, Net N
Qnt Piece Counting: Quantity Tot Formulation,
pRef Weighing in percent: Reference percentage Pur density : Purity of Gold
Pct Weighing in percent: Reference percentage Den Density : density of sample
Cnt Animal weighing: No. of sub-weighing operations Pip Pipette Calibration
xNt Animal weighing: Calculated average Sta Statistics : to obtain the statistics of the data
Totalization
Totalization
Reference weight
Totalization : Total Weight
Page 57
• For connecting a computer or other peripheral device to the balance using the RS-232 protocol and cables up to 15m (50 ft.) long.
RS-232
3
TxD RxD
Gnd Gnd
RxD TxD
2
55
23
Balance
DB 9 Female Connector
Peripheral Device
e.g. Computer or Printer
Display
10. Error Codes
Cause
Overload
Under load
Weight set is to low for storing any reference at PCS, %, Custom Unit or Check Weighing.
While calibrating the scale, the load on the pan is more than 10% of the capacity. (During power on of the scale.)
Calibration
• User does not keep any weight on the pan within 60 second.
• Weight load on the pan is not within the tolerance limit.
GLP is ON and user tries to enter in to the User Menu before the footer is printed.
Calibration Display shows any weight other than 0.00 and user tries to Calibrate the balance
Incorrect value of TIME or DATE.
Solution
Remove excess weight from the weighing pan.
• Keep weighing Pan on Weighing Shaft. Check whether weighing pan is positioned properly.
Increase weight on the pan.
Switch OFF the Balance and Switch ON again without any load on the pan.
• Add the calibration weight on the pan when
demanded by the balance
• Calibrate with the exact Calibration Weight.
Print the footer first, by pressing <CANCEL> key, and then access the USER MENU.
Tare the balance or enter Calibration procedure when “0.00 g” is displayed.
Enter proper value of TIME or DATE.
111
112
9.3 Cabling Diagram
Page 58
Display Cause
Last stored PRINT option is AUTO or AUTO LOAD or CONTINUOUS and user tries to set GLP ON from USER MENU.
RTC not operational.
Error 2 Calibration Error + RTC Error.
Error 3 Calibration Error + RTC Error.
Unstable ambient conditions A foreign object is caught between the load plate and the balance/scale frame
The balance has not been calibrated / adjusted. The balance was not zeroed before weighing.
Solution
Change the print option to Print on REQUEST and then turn GLP ON.
Contact Citizen Service center.
Contact Citizen Service center.
Contact Citizen Service center.
Set up the balance/scale in another area Remove the foreign object
Calibrate / Adjust the balance. Tare or Zero the balance before weighing.
The weight readout changes constantly
The weight readout is obviously wrong
Service
Regular servicing by a CITIZEN technician will extend the service life of your balance and ensure its continued weighing accuracy. CITIZEN can offer you service contracts, with your choice of regular maintenance intervals.
The optimum maintenance interval depends on the operating conditions at the place of installation and on the individual tolerance requirements.
Repairs
Repair work must be performed by trained service technicians. Any attempt by untrained persons to perform repairs may lead to hazards for the user.
Cleaning
• Unplug the DC adapter from the wall outlet (mains supply). If you have an interface cable connected to the balance/scale port, unplug it from the port.
Make sure that no liquid enters the balance/scale housing
Do not use any aggressive cleaning agents (solvents or similar agents)
• Clean the balance/scale using a piece of cloth which has been wet with a mild detergent (soap)
• After cleaning, wipe down the balance/scale with a soft, dry cloth
11. Care & Maintenance
113
114
Page 59
Cleaning Stainless Steel Surfaces
• Clean all stainless steel parts regularly. Remove the stainless steel weighing pan and thoroughly clean it separately. Use a damp cloth or sponge to clean any stainless steel parts on the scale. You can use any commercially available household cleaning agent that is suitable for use on stainless steel. Clean stainless steel surfaces by wiping them down. Then clean the weighing pan thoroughly, making sure to remove all residues. Use a damp cloth or sponge to wipe down any stainless steel parts on the scale again. Afterwards, allow the scale to dry. If desired, you can apply oil to the cleaned surfaces as additional protection.
Do not use stainless steel cleaning agents that contain soda lye (caustic), acetic acid, hydrochloric acid, sulfuric acid or citric acid. The use of scrubbing sponges made of steel wool is not permitted. Solvents are permitted for use only on stainless steel parts.
Safety Inspection
If there is any indication that safe operation of the balance/scale with the DC adapter is no longer warranted:
• Turn off the power and disconnect the equipment from DC power immediately
• Lock the equipment in a secure place to ensure that it cannot be used for the time being.
• Safe operation of the balance/scale with the DC adapter is no longer ensured when:
• There is visible damage to the DC adapter.
• The DC adapter no longer functions properly.
• The DC adapter has been stored for a relatively long period under unfavorable conditions.
115
116
Page 60
117
LIMITED WARRANTY
ADDRESS :-
NAME :-
TEL NO. :-
MODEL NO. :-
WARRANTY PERIOD :-
SERIAL NO. :-
PURCHASE DATE :-
BRANCH / DISTRIBUTOR / CONTACT DETAILSDEALER
Owners Signature / Date
WARRANTY REGISTRATION
SEND YOUR WARRANTY CARD DULY FILL TO ABOVE ADDRESS FOR REGISTRATION
STAMP / SIGN
118
CITIZEN SCALES (I) PVT.LTD.
Citizen House, E2, Plot No. 11, WICEL Estate, Opp. Seepz Gate no. 1,
Andheri (E), Mumbai - 400 093. Maharashtra, India
e-mail :- service@citizenscales.com • web.: www.citizenscales.com
Tel. No. :- +91-22-4243 7700 • Fax :- +91-22-4243 7800
CITIZEN products are warranted against defects in materials and workmanship from the date of delivery through the duration of the warranty period.
During the warranty period Citizen will repair, or, at its option, replace any component (s) that proves to be defective at no charge, provided that
the product is returned, freight prepaid, to CITIZEN.
This warranty does not apply if the product has been damaged by accident or misuse, exposed to radioactive or corrosive materials, has foreign
material penetrating to the inside of the product, or as a result of service or modification by other than CITIZEN. In lieu of a properly returned warranty
registration card, the warranty period shall begin on the date of shipment to the authorized dealer. No other express or implied warranty is given by
Citizen Scale (i) Pvt. Ltd.
te and country to country, please contact citizen or your local CITIZEN dealer for further details.
ee of charge subject to terms & condition mentioned below.
5. Citizen Scale (I) Pvt. Ltd. shall not be liable for any consequential damages.
As warranty legislation differs from state to sta
CITIZEN service center will repair the product fr
TERMS & CONDITION
1. It covers only weighing balance purchased from authorized channel and does not cover accessories like Battery, Adaptor, RS232 cable, Pan, Pan support etc
2. It does not cover the product of which model and serial number has been altered, removed or defaced and / or is open by unauthorized person and found void sticker has been tampered.
3. This warranty is non-transferable and applicable only to first end user purchasing the product from authorized dealer.
4. For repair based on this warranty you need to hand over this product or send this product to address mentioned in warranty card in original packing, enclosing copy of this warranty card.‘
ISO 9001 : 2008 CERTIFIED
Page 61
119
1)
212
171.5
204.5
342.5
231
67.5
190.5
189.5
152
97
165
255.5
271
78
169.5
29
Ø90
1)
212
171.5
204.5
342.5
231
67.5
97
325.5
239.5
165
78
341
Ø90
190.5
189.5
152
29
120
Analytical Balances
(CX / CY(0.1mg)Series Balances) Draft Shield High
Carat Balances
(CT Series Balances) Draft Shield Low
1)
212
171.5
204.5
342.5
231
67.5
96.5
78
128 X 128
169.5
29
255.5
190.5
152
271
189.5
165
212
342.5
231
89.5
67.5
210 X 198
171.5
208.5
48
1)
204.55
Precision Balances
(CY Series Balances) Draft Shield Low
Top Loading Precision Balances
(CG Series Balances) No Draft Shield
12. Dimensional Drawing
Page 62
121
122
High Capacity Precision Balances
(SSH Series Balances)
High Capacity Precision Balances
(CY Series Balance with wind shield)
402
419.18
129.95
247
347
268.05
190.5
212
171.5
204.5
189.5
159
81
67.5
231
342.5
101
182
Page 63
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External calibration weight (of at least accuracy class...) Net Weight, approx. Pan size Weighing chamber height Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CX 301
301
0.1 1
-301
0.1
0.3
I
Internal 200 (E2)
CX 220
220
0.1 1
-220
0.1
0.2
I
18º to 30ºC
Internal
100 (E2)
CX 120
120
0.1 1
-120
0.1
0.2
I
18º to 30ºC
Internal
50 (E2)
CX 65
65
0.1 1
-65
0.1
0.2
I
18º to 30ºC
Internal
25 (E2)
CX 24
21
0.1 1
-21
0.1
0.2
II
15º to 30ºC
Internal
10 (E2)
3 sec. 18º to 30ºC
8kg 90 Ø
228.5
342.5 x 212 x 341 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
g mg mg g <=mg <=mg
s
0
C
0
C g kg mm mm mm V~ Hz VA
Analytical Balances
13. Specifications
123
124
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External calibration weight (of at least accuracy class...) Net Weight, approx. Pan size Weighing chamber height Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CY 204
220
0.1 1
-220
0.1
0.2
I
18º to 30ºC
External 100 (E2)
CY 104
120
0.1 1
-120
0.1
0.2
I
18º to 30ºC
External
50 (E2)
CY 64
61
0.1 1
-61
0.1
0.2
I
18º to 30ºC
External
25 (E2)
CY 24
21
0.1 1
-21
0.1
0.2
II
15º to 30ºC
External
10 (E2)
CY 54
51
0.1 1
-51
0.1
0.2
II
15º to 30ºC
External
25 (E2)
3 sec.
8kg 90 Ø
228.5
342.5 x 212 x 341 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
g mg mg g <=mg <=mg
s
0
C
0
C
g kg mm mm mm V~ Hz VA
Analytical Balances
CX 54
51
0.1 1
-51
0.1
0.2
II
15º to 30ºC
Internal
25 (E2)
Page 64
125
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External cal. wt. (of at least accuracy class..) Net Weight, approx. Pan size Weighing chamber height with draftshield Weighing chamber height with Windshield Dimensions (W x D x H) with draftshield Dimensions (W x D x H) with windshield DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CY 120
120
0.001
0.01
-120
0.001
0.002
II
External
50 (F1)
CY 220
220
0.001
0.01
-220
0.001
0.002
II
15º to 30ºC
External 100 (F1)
CY 320
320
0.001
0.01
-320
0.001
0.002
II
15º to 30ºC
External 200 (F1)
CY 360C
360
0.001
0.01
-360
0.001
0.002
II
15º to 30ºC
Internal 200 (F1)
CY 360
360
0.001
0.01
-360
0.001
0.002
II
15º to 30ºC
External 200 (F1)
CY 720
720
0.001
0.01
-720
0.002
0.003
I
18º to 30ºC
External 500 (F1)
CY 510
510
0.001
0.01
-510
0.002
0.003
II
15º to 30ºC
External 300 (F1)
CY 420
420
0.001
0.01
-420
0.001
0.002
II
15º to 30ºC
External 200 (F1)
2 - 3 sec. 15º to 30ºC
7.5kg 128 x 128
158.5 81
342.5 x 212 x 271
342.5 x 212 x 193.5 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
g g g g <=g <=g
s
0
C
0
C g kg mm mm mm mm mm V~ Hz VA
Precision Balances
CY 1003
1000
0.001
0.01
-1000
0.002
0.003
I
18º to 30ºC
External 500 (F1)
126
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External cal. wt. (of at least accuracy class..) Net Weight, approx. Pan size Weighing chamber height with draftshield Weighing chamber height with Windshield Dimensions (W x D x H) with draftshield Dimensions (W x D x H) with windshield DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CY 420C
420
0.001
0.01
-420
0.001
0.002
II
15º to 30ºC
Internal 200 (F1)
CY 510C
510
0.001
0.01
-510
0.002
0.003
II
15º to 30ºC
Internal 300 (F1)
CY 320C
320
0.001
0.01
-320
0.001
0.002
II
15º to 30ºC
Internal 200 (F1)
CY 220C
220
0.001
0.01
-220
0.001
0.002
II
15º to 30ºC
Internal 100 (F1)
CY 120C
120
0.001
0.01
-120
0.001
0.002
II
Internal
50 (F1)
CY 1003C
1000
0.001
0.01
-1000
0.002
0.003
I
18º to 30ºC
Internal
500 (F1)
2 - 3 sec. 15º to 30ºC
7.8kg 128 x 128
158.5 81
342.5 x 212 x 271
342.5 x 212 x 193.5 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
g g g g <=g <=g
s
0
C
0
C g kg mm mm mm mm mm V~ Hz VA
Precision Balances
Page 65
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External cal. wt. (of at least accuracy class..) Net Weight, approx. Pan size Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CG 1202
1200
0.01
0.1
-1200
0.01
0.02
II
15º to 30ºC
External
1000 (F1)
198 x 205
CG 613
610
0.01
0.01
-610
0.01
0.02
II
15º to 30ºC
External
300 (F1)
128 x 128
CG 612
610
0.01
0.1
-610
0.01
0.02
III
15º to 45ºC
External
300 (F1)
198 x 205
CG 312
310
0.01
0.1
-310
0.01
0.02
III
External 200 (F1)
2 - 3 sec. 15º to 45ºC
5.5kg 198 x 205
342.5 x 212 x 89.5 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
g g g g <=g <=g
s
0
C
g kg mm mm V~ Hz VA
Top Loading Precision Balances
127
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External cal. wt. (of at least accuracy class..) Net Weight, approx. Pan size Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
2 - 3 sec. 15º to 30ºC
5.5kg 198 x 205
342.5 x 212 x 89.5 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
g g g g <=g <=g
s
0
C
g kg mm mm V~ Hz VA
Top Loading Precision Balances
128
CG 2202
2200
0.01
0.1
-2200
0.01
0.02
II
External
1000 (F1)
CG 3102
3100
0.01
0.1
-3100
0.01
0.02
II
15º to 30ºC
External
1500 (F1)
198 x 205
CG 4102
4100
0.01
0.1
-4100
0.01
0.02
II
15º to 30ºC
External
2000 (F1)
198 x 205
CG 6102
6100
0.01
0.1
-6100
0.02
0.03
I
18º to 30ºC
External
4000 (F1)
198 x 205
CG 2102DR
200 / 2100
0.001 / 0.01
0.01 / 0.1
-200 / 2100
0.002 / 0.02
0.003 / 0.03
I
18º to 30ºC
External
1000 (F1)
128 x 128
342.5 x 212 x 271
Page 66
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External calibration weight (of at least accuracy class...) Net Weight, approx. Pan size Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
SSH 96
10
0.01
0.1
-10
0.02
0.04
II
External
5 (E2)
SSH 95
15
0.1 1
-15
0.1
0.2
II
15º to 30ºC
External
10 (F1)
SSH 94
20
0.1 1
-20
0.1
0.2
II
15º to 30ºC
External
10 (F1)
SSH 94H
25
0.1 1
-25
0.1
0.2
II
15º to 30ºC
External
15 (F1)
2 - 3 sec. 15º to 30ºC
15kg 400 x 300 517 x 302 x 130 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
Kg g g Kg <=g <=g
s
0
C
Kg kg mm mm V~ Hz VA
High Capacity Precision Balances
130
CG 312C
310
0.01
0.1
-310
0.01
0.02
III
Internal
200 (F1)
CG 612C
610
0.01
0.1
-610
0.01
0.02
III
15º to 45ºC
Internal 300 (F1)
CG1202C
1200
0.01
0.1
-1200
0.01
0.02
II
15º to 30ºC
Internal
1000 (F1)
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External cal. wt. (of at least accuracy class..) Net Weight, approx. Pan size Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CG 2202C
2200
0.01
0.1
-2200
0.01
0.02
II
15º to 30ºC
Internal
1000 (F1)
CG3102C
3100
0.01
0.1
-3100
0.01
0.02
II
15º to 30ºC
Internal
1500 (F1)
CG 4102C
4100
0.01
0.1
-4100
0.01
0.02
II
15º to 30ºC
Internal
2000 (F1)
CG 6102C
6100
0.01
0.1
-6100
0.02
0.03
I
18º to 30ºC
Internal
4000 (F1)
2 - 3 sec. 15º to 45ºC
6kg 198 x 205
342.5 x 212 x 89.5 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
g g g g <=g <=g
s
0
C
g kg mm mm V~ Hz VA
Top Loading Precision Balances
129
Page 67
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External calibration weight (of at least accuracy class...) Net Weight, approx. Pan size Weighing chamber height Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CT 103
101
0.001
0.01
-101
0.001
0.002
II
External
10 (F1)
CT 253
250
0.001
0.01
-250
0.001
0.002
II
15º to 30ºC
External
25 (F1)
CT 603
600
0.001
0.01
-600
0.001
0.002
I
18º to 30ºC
External
50 (F1)
CT 1603
1600
0.001
0.01
-1600
0.002
0.003
I
18º to 30ºC
External 200 (F1)
CT 1103
1100
0.001
0.01
-1100
0.001
0.002
I
18º to 30ºC
External
100 (F1)
3 sec. 15º to 30ºC
7.5 90 Ø
158.5
342.5 x 212 x 271 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
ct ct ct ct <=ct <=ct
s
0
C
kg mm mm mm V~ Hz VA
Carat Balances
132
131
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External calibration weight (of at least accuracy class...) Net Weight, approx. Pan size Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
SSH 93
35
0.1 1
-35
0.1
0.2
II
External
15 (F1)
SSH 92
50
1
10
-50 1 2
III
15º to 45ºC
External
25 (F1)
SSH 91
60
1
10
-60 1 2
III
15º to 45ºC
External
30 (F1)
SSH 90
100
1
10
-100 1 2
III
15º to 45ºC
External
50 (F1)
2 - 3 sec. 15º to 30ºC
15kg 400 x 300 517 x 302 x 130 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
Kg g g Kg <=g <=g
s
0
C
Kg kg mm mm V~ Hz VA
High Capacity Precision Balances
Page 68
14. Accessories (Option)
Statistical Printer “CPR 02”
with Date / Time & Statistics
Remote Display “SRD01”
Calibration Weights
(F1) (ERTL, F2 with certificate) for further details, contact CITIZEN Dealers.
USB Interface
Density Kit “CDK 01”
For determination of solids for determination of liquids with displacement body
Antitheft device
Cable and lock (for all models)
Dust Cover
134
Model
Weighing Capacity Readability (d) Accuracy (e) Tare Range (Subtractive) Repeatability (std. deviation) Linearity Weighing Class Response time (average) Operating temperature range Calibration External calibration weight (of at least accuracy class...) Net Weight, approx. Pan size Weighing chamber height Dimensions (W x D x H) DC power source / Power requirements Frequency Power consumption (average) Selectable weight units
Built-in-interface Format Parity Transmission rates : Handshake mode
CT 103C
101
0.001
0.01
-101
0.001
0.002
II
Internal
10 (F1)
CT 253C
250
0.001
0.01
-250
0.001
0.002
II
15º to 30ºC
Internal
25 (F1)
CT 603C
600
0.001
0.01
-600
0.001
0.002
I
18º to 30ºC
Internal
50 (F1)
CT 1103C
1100
0.001
0.01
-1100
0.001
0.002
I
18º to 30ºC
Internal 100 (F1)
CT 1603C
1600
0.001
0.01
-1600
0.002
0.003
I
18º to 30ºC
Internal 200 (F1)
3 sec. 15º to 30ºC
7.8kg 90 Ø
158.5
342.5 x 212 x 271 DC Adapter, input 100 ~ 240 0.8A output 13V / 1.5A 50 / 60Hz maximum 18; typical 9
RS-232 1 start bit, 8-bit ASCII, parity, 1 or 2 stop bits Mark, Space, Odd, even, none 300; 600; 1200; 2400; 4800; 9600; 19200; 57600 baud None
gram, kilogram, pound, ounce, troy ounce, grain, pennyweight carat, Milligram, momme, mesghal, Hong Kong tales, Singapore taels Taiwan tales, baht
ct ct ct ct <=ct <=ct
s
0
C
kg mm mm mm mm V~ Hz VA
Carat Balances
133
Loading...