Bio-Rad Mini-PROTEAN II Multiscreen Apparatus User Manual

Page 1
Mini-PR OTEAN®II
Multiscreen Apparatus
Instruction Manual
Catalog number
170-4017
Page 2
Table of Contents
Page
Section 1 Introduction .............................................................................................1
1.1 Specifications ........................................................................................................1
Section 2 Equipment and Reagents........................................................................1
Section 3 Special Handling Features......................................................................3
3.1 Chemical Stability.................................................................................................3
Section 4 Multiscreen Assembly Procedure..........................................................3
4.1 Preparation for the Immunoassay in the Multiscreen Apparatus.........................3
4.2 Assembly of the Multiscreen Apparatus...............................................................3
4.3 Sample Loading and Washing ..............................................................................5
4.4 Color Development of Enzyme Conjugates.........................................................7
4.5 Detection with Colloidal Gold Conjugates...........................................................8
Section 5 Solution for Immunoassay Applications...............................................8
Section 6 References.................................................................................................9
Page 3
Section 1 Introduction
With the Mini-PROTEAN II multiscreen apparatus you can quickly and easily screen up to 40 different antibody or serum samples on Western blots, without having to cut the membranes into strips. Only 600 µl of sample is used per channel, eliminating waste of precious antibody. Two indi­vidual and independent sample templates allow screening of either one or two mini blots.
The multiscreen apparatus is simple to operate. Electrophorese an antigen sample on an SDS­PAGE mini gel using the Mini-PROTEAN II cell, and blot it onto nitrocellulose or Zeta-Probe
®
membrane with the Mini Trans-Blot®, Trans-Blot, or Trans-Blot SD cell. After blocking the unre­acted sites, clamp the membrane between the gasket and sample template. The assembly is held together with four screws, and the rubber sealing gasket prevents any well-to-well leakage. Pipet serum or antibody samples into each of the channels for incubation with the anitgen. Wash solu­tions can be easily introduced with the Eppendorf®Repeater™pipet, and are rapidly removed by vacuum aspiration. The Multiscreen apparatus is compatible with all common Western blotting procedures.
1.1 Specifications
Materials
Multiscreen apparatus Acrylic plastic Multiscreen gasket Silicone rubber Shipping weight 1.4 kg Overall size 11 x 27 x 6 cm (W x L x H) Membrane size 7 x 8.4 cm Channel dimensions 2.5 mm x 5.2 cm x 5 mm (W x L x H) Channels per sample template 20
Section 2 Equipment and Reagents
Catalog Number Product Description
170-4017 Mini-PROTEAN II Multiscreen Apparatus, includes Multiscreen
sample templates, 2 Gaskets and Base Plate
170-4018 Mini-PROTEAN II Multiscreen Gaskets, 2
Mini-Protean II Cell
165-2940 Mini-PROTEAN II Cell, includes 10 well combs, 0.75 mm
spacers (4), Electrode Core with Gaskets, lower buffer chamber, lid with cables, 3 sets Glass Plates, Clamp Assemblies (2), Casting Stand with Gaskets, leveling bubble, and instructions
Transfer Cells
170-3930 Mini Trans-Blot Electrophoretic Transfer Cell 170-3935 Mini Trans-Blot Module 170-3946 Trans-Blot Electrophoretic Transfer Cell, with plate electrodes 170-3910 Trans-Blot Electrophoretic Transfer Cell, with standard electrodes 170-3940 Trans-Blot SD Semi-Dry Electrophoretic Transfer Cell
1
Page 4
Catalog Number Product Description
Power Supplies
165-5052 PowerPac 200 Power Supply, 100/120 V 165-5053 PowerPac 200 Power Supply, 220/240 V
Related Instruments
170-6545 Bio-Dot
®
Microfiltration Apparatus
170-6542 Bio-Dot SF Microfiltration Apparatus
Blotting Media Nitrocellulose Membrane (0.45 micron)
162-0115 Nitrocellulose Membrane, roll, 33 cm x 3 m, 1 162-0113 Nitrocellulose Membrane, sheets, 20 x 20 cm, 5 162-0116 Nitrocellulose Membrane, sheets, 15 x 15 cm, 10 162-0114 Nitrocellulose Membrane, sheets, 15 x 9.2 cm, 10 162-0117 Nitrocellulose Membrane, sheets, 9 x 12 cm, 10 162-0145 Nitrocellulose Membrane, sheets, 7 x 8.4 cm, 10
Nitrocellulose Membrane (0.2 micron)
162-0112 Nitrocellulose Membrane, roll, 33 cm x 3 cm, 1 162-0146 Nitrocellulose Membrane, sheets, 7 x 8.4 cm, 10 162-0147 Nitrocellulose Membrane, sheets, 13.5 x 16.5 cm, 10
Immun-Blot
®
Assay Kits
Bio-Rad's Immun-Blot assay kits contain the necessary components and instructions for performing immune detection assays on blotted membranes.
170-6460 Immun-Blot Assay Kit - Goat Anti-Rabbit IgG (H+L)
AP Conjugate
170-6461 Immun-Blot Assay Kit - Goat Anti-Mouse IgG (H+L)
AP Conjugate
170-6462 Immun-Blot Assay Kit - Goat Anti-Human IgG (H+L)
AP Conjugate
170-6463 Immun-Blot Assay Kit - Goat Anti-Rabbit IgG (H+L)
HRP Conjugate
170-6464 Immun-Blot Assay Kit - Goat Anti-Mouse IgG (H+L)
HRP Conjugate
170-6465 Immun-Blot Assay Kit - Goat Anti-Human IgG (H+L)
HRP Conjugate 170-6466 Immun-Blot Assay Kit - Protein A HRP 170-6467 Immun-Blot Assay Kit - Protein G HRP
Total Protein Detection Kits
170-6512 Biotin-Blot Protein Detection Kit 170-6517 Enhanced Colloidal Gold Total Protein Detection Kit
Blotting Standards
161-0305 Prestained SDS-PAGE Standards, Low Range 161-0309 Prestained SDS-PAGE Standards, High Range 161-0318 Prestained SDS-PAGE Standards, Broad Range 161-0307 Biotinylated SDS-PAGE Standards Kit, Low Range, HRP 161-0308 Biotinylated SDS-PAGE Standards Kit, Low Range, AP 161-0312 Biotinylated SDS-PAGE Standards Kit, High Range, HRP 161-0313 Biotinylated SDS-PAGE Standards Kit, High Range, AP 161-0321 Biotinylated SDS-PAGE Standards Kit, Broad Range, HRP 161-0322 Biotinylated SDS-PAGE Standards Kit, Broad Range, AP
2
Page 5
Section 3 Special Handling Features
The multiscreen apparatus can be cleaned with a mild, non-abrasive detergent but not be autoclaved. Do not subject the unit to temperatures greater than 50 °C, as this will warp acrylic plates. If the unit becomes warped, it will no longer provide a proper seal. Heating the appa­ratus to >50 °C voids all warranties.
3.1 Chemical Stability
Chemicals compatible with acrylic plastic:
hydrochloric acid < 50% ethanol sodium hydroxide < 50% ethanol
Chemicals that will attack acrylic plastic:
all polar aromatic solvents or chlorinated hydrocarbons, esters, and ketones glacial acetic acid chromic acid trichloroacetic acid > 50% ethanol > 50% methanol
Section 4 Multiscreen Operating Procedure
4.1 Preparation for the Immunoassay in the Multiscreen Apparatus
1. Electrophorese the antigen sample into a mini gel following the instructions provided
with the Mini-PROTEAN II cell. The maximum length of the separating gel should not
exceed 5.2 cm, the length of the channels on the multiscreen sample template. The stack-
ing gel should be cast with the Mini-PROTEAN II preparative comb. This comb con-
tains one large sample well and one reference well.
2. Blot the gel to nitrocellulose or Zeta-Probe membrane, using a 7 x 8.4 cm membrane
size. Refer to the Mini Trans-Blot, Trans-Blot or Trans-Blot SD cell instruction manual
for electrophoretic transfer procedures.
3. Following the transfer, mark the outline of the gel on the membrane using a pen or pen-
cil. This will aid in aligning the blot with the sample template. Block the unreacted sites
on the blot with a blocking solution, 1 hour at room temperature for nitrocellulose, 2
hours at room temperature for Zeta-Probe membrane. Rinse the blocked blot in TBS
before applying to the multiscreen apparatus. See Section 5 for buffer formulations.
Note: Always use forceps or wear gloves when handling membranes.
4.2 Assembly of the Multiscreen Apparatus
1. Clean and dry the multiscreen apparatus and gaskets prior to assembly.
Note: Do not heat the apparatus to temperatures greater than 50 °C. This will cause the
unit to warp.
2. Place the sealing gasket onto the base plate with the raised surface down, using the guide
pins to help align the gasket.
3
Page 6
Fig. 1. Assembly of the multiscreen apparatus.
3. Lay the blocked blot on the gasket with the antigen side facing up. Center the membrane
so that the channels of the sample template cover the length of the blotted sample.
4. Place the sample template on top of the membrane. The guide pins insure that the template
will be properly aligned. Finger tighten the four screws. When tightening the screws, use
a diagonal crossing pattern to insure even pressure on the membrane surface (see Figure
2). The multiscreen apparatus is ready for sample application.
Note: Use of excessive force when tightening the screws is not necessary to prevent well-
to-well leakage. Finger tightening is sufficient to obtain a good seal. Overtightening can
cause the channels to cut into the membrane.
4
2
4
6
8
10
12
14
16
18
20
2
4
6
8
10
12
14
16
18
20
MINI PROTEAN II
MULTISCREEN
MINI PROTEAN II
MULTISCREEN
Page 7
Fig. 2. Diagonal crossing pattern for tightening screws in the multiscreen apparatus.
4.3 Sample Loading and Washing
Detailed instructions for performing immunoassays, including a comprehensive trou­bleshooting guide, can be found in any of the Immun-Blot assay instruction manuals. See page 2 for a complete listing of Immun-Blot assay kits available from Bio-Rad.
1. To load an antibody or serum sample, tilt the multiscreen apparatus toward you so that the
back of the unit is tilted up ~30° (see Figure 3). Using a syringe or Eppendorf pipet, load
the solution into the bottom unmarked holes of the channels. Slow, careful delivery of
sample is necessary to avoid trapping bubbles inside the channels. Titling the apparatus
helps the bubbles rise to the top, towards the numbered holes of the channels. Fill the
channel with 600 µl antibody solution.
Note: Antibody buffers containing BSA or BLOTTO* are recommended for use with
the multiscreen apparatus. Do not use antibody buffers with gelatin, as this may cause
coagulation of gelatin within the channels of the unit.
* BLOTTO is an acronym for Bovine Lacto Transfer Technique Optimizer, and refers
to non-fat dry milk.
1,2
5
13
42
Page 8
Fig. 3. Tilt the multiscreen apparatus toward you during sample application.
2. Wash solutions can be applied in the same manner as the antibody samples, or with the
Eppendorf repeater pipet. Use 600 µl per channel. The number and stringency of wash-
es may vary and should be determined separately for each experiment. However, a min-
imum of three washes with a buffer containing a detergent such as Tween-20 is
recommended after each antibody incubation. See Section 5 for buffer formulations.
6
Page 9
Fig. 4. Vacuum aspiration of sample and wash solutions.
3. Antibody samples and wash solutions can be rapidly removed by vacuum aspiration. The
tubing from the vacuum source should be attached to the bottom, unmarked row of holes.
Move the tubing back and forth along the row of holes until all the channels are dry (see
Figure 4). To save an antibody or serum sample after incubation, remove the solution
individually from the channel with a pipet or syringe.
7
Page 10
4.4 Color Development of Enzyme Conjugates
Incubation of the blot with antibodies conjugated to enzymes, such as horseradish per­oxidase or alkaline phosphatase, can be conducted either in the multiscreen apparatus or in a separate vessel. If this step is carried out in the unit, wash the blot in the apparatus after the second antibody incubation as outlined in Section 4.3.
1. Color development of enzyme conjugated antibodies should be performed in a separate
container to prevent permanent discoloration of the multiscreen apparatus. Remove the
membrane by loosening the four screws. Lift out the sample template and move the mem-
brane to a color development vessel.
2. Wash the membrane once with TBS for 5 minutes. After the color development solution
has been prepared, incubate the membrane in the solution. Gently agitate the solution
until development is complete. Remove the solution and rinse the membrane several
times in distilled water to stop the reaction. Air dry the blot on filter paper.
4.5 Detection with Colloidal Gold Conjugates
Incubation of the blot with colloidal gold conjugated antibodies, protein A, or protein G, should be conducted in a separate vessel to prevent discoloration of the multiscreen apparatus.
1. Remove the blot from the multiscreen apparatus after washing to remove excess first
antibody. Place the membrane in a color development vessel.
2. Wash the membrane once with TBS for 5 minutes. Add the gold solution to the vessel until
the membrane is completely covered. Gently agitate the solution. Red bands identifying
antigen will appear on the membrane surface within 10–15 minutes at the sites of high-
est antigen concentrations. Allow the incubation to continue until the desired sensitivity
is achieved.
Section 5 Solutions for Immunoassay Applications
Tris Buffered Saline, 1 x TBS, 2 L
20 mM Tris-HCl, 500 mM NaCl, pH 7.5
Dissolve 4.84 g Tris, 58.48 g NaCl in ~1.5 L distilled, deionized H2O. Adjust to pH 7.5
with HCl. Adjust the volume to 2 L with dd H2O.
Note: Bio-Rad's Premixed Tris-Buffered Saline (catalog number 170-6430) eliminates
weighing of buffer components. One bottle produces 1 L of 10 x TBS.
Tween-20 Wash Solution, 1X TTBS, 1 L
20 mM Tris-HCl, 500 mM NaCl, 0.05% Tween-20, pH 7.5
Add 0.5 ml Tween-20 to 1 L of TBS.
Blocking Solution, 100 ml
Both of the following blocking solutions can be used with nitrocellulose. The solutions
containing BLOTTO should be used with Zeta-Probe membrane. Incubate nitrocellulose
blots for 1 hour at room temperature. Zeta-Probe membrane should be blocked for 2 hours
at room temperature.
3% Gelatin - TBS
Add 3.0 g gelatin to 100 ml TBS. Heat to 50 °C, stirring until dissolved. A microwave
oven will quickly solubilize the gelatin, but do not heat above 65 °C.
8
Page 11
OR:
5% BLOTTO in 100 ml of TBS
Add 5% g of BLOTTO to 100 ml of TBS.
Antibody Buffer, 200 ml
1% BSA - TTBS
Add 2 g BSA to 200 ml TTBS. Stir to dissolve.
OR
1% BLOTTO in TTBS.
Add 2 g of BLOTTO to 200 ml of TTBS.
First Antibody Solution
Dilute antigen specific primary antibody to the appropriate titer in antibody buffer.
Second Antibody Solution, 100 ml
Dilute species specific Bio-Rad second antibody enzyme conjugate, 1:3,000 by adding 33 µl of conjugate to 100 ml of antibody buffer.
Consult the Immun-Blot assay kit instruction manual for dilution protocols of the col­loidal gold conjugates.
Color Development Solution
The specific chemicals and buffers are dependent on the enzyme conjugate being used.
See the Immun-Blot assay kit instruction manual for details on how to make the appro-
priate solution.
Section 6 References
1. Jerome, J. F. and Jaehning, J. A., Mol. and Cell Bio., 6, 1633 (1986).
2. Johnson, D. A., et. al., Gene Anal. Tech., 1, 3 (1984).
Eppendorf is a trademark of Eppendorf Geratebau Netheler & Hinz, GmbH
9
Page 12
Bio-Rad Laboratories
Australia,
Bio-Rad Laboratories Pty Limited, Block Y Unit 1, Regents Park Industrial Estate, 391 Park Road, Regents Park, NSW 2143 • Phone 02-9414-2800 • Fax 02-9914-2888
Austria,
Bio-Rad Laboratories Ges.m.b.H., Auhofstrasse 78D, 1130 Wien • Phone (1) 877 89 01 • Fax (1) 876 56 29
Belgium,
Bio-Rad Laboratories S.A./N.V., Begoniastraat 5, 9810 Nazareth Eke • Phone 09-385 55 11 • Fax 09-385 65 54
Canada,
Bio-Rad Laboratories (Canada) Ltd., 5671 McAdam Road, Mississauga, Ontario L4Z 1N9 • Phone (905) 712-2771 • Fax (905) 712-2990
China,
Bio-Rad Laboratories, 14, Zhi Chun Road, Hai Dian District, Beijing 100088 • Phone (01) 2046622 • Fax (01) 2051876
Denmark,
Bio-Rad Laboratories, Symbion Science Park, Fruebjergvej 3, DK-2100 Copenhagen • Phone 39 17 9947 • Fax 39 27 1698
Finland,
Bio-Rad Laboratories, Business Center Länsikeskus, Pihatörmä 1A SF-02240, Espoo, • Phone 90 804 2200 • Fax 90 804 1100
France,
Bio-Rad S.A., 94/96 rue Victor Hugo, B.P. 220, 94 203 Ivry Sur Seine Cedex • Phone (1) 49 60 68 34 • Fax (1) 46 71 24 67
Germany,
Bio-Rad Laboratories GmbH, Heidemannstraße 164, D-80939 München/Postfach 450133, D-80901 München • Phone 089 31884-0 • Fax 089 31884-100
India,
Bio-Rad Laboratories, C-248 Defence Colony, New Delhi 110 024 • Phone 91-11-461-0103 • Fax 91-11-461-0765
Italy,
Bio-Rad Laboratories S.r.l.,Via Cellini, 18/A, 20090 Segrate Milano • Phone 02-21609 1 • Fax 02-21609-399
Japan,
Nippon Bio-Rad Laboratories, 7-18, Higashi-Nippori 5-Chome, Arakawa-ku, Tokyo 116 • Phone 03-5811-6270 • Fax 03-5811-6272
The Netherlands,
Bio-Rad Laboratories B. V., Fokkerstraat 10, 3905 KV Veenendaal • Phone 0318-540666 • Fax 0318-542216
New Zealand,
Bio-Rad Laboratories Pty Ltd., P. O. Box 100-051, North Shore Mail Centre, Auckland 10 • Phone 09-443 3099 • Fax 09-443 3097
Pacific,
Bio-Rad Laboratories, Unit 1111, 11/F., New Kowloon Plaza, 38, Tai Kok Tsui Road, Tai Kok Tsui, Kowloon, Hong Kong • Phone 7893300 • Fax 7891257
Singapore,
Bio-Rad Laboratories (Singapore) Ltd., 221 Henderson Rd #05-19, Henderson Building, Singapore 0315 • Phone (65) 272-9877 • Fax (65) 273-4835
Spain,
Bio-Rad Laboratories, S. A. Avda Valdelaparra 3, Pol. Ind. Alcobendas, E-28100 Alcobendas, Madrid • Phone (91) 661 70 85 • Fax (91) 661 96 98
Sweden,
Bio-Rad Laboratories AB, Gärdsvägen 7D, Box 1276, S-171 24 Solna • Phone 46-(0)8-735 83 00 • Fax 46-(0)8-735 54 60
Switzerland,
Bio-Rad Laboratories AG, Kanalstrasse 17, Postfach, CH-8152 Glattbrugg • Phone 01-809 55 55 • Fax 01-809 55 00
United Kingdom,
Bio-Rad Laboratories Ltd., Bio-Rad House, Maylands Avenue, Hemel Hempstead, Herts HP2 7TD • Free Phone 0800 181134 • Fax 01442 259118
Life Science Group
2000 Alfred Nobel Drive Hercules, California 94547 Telephone (510) 741-1000 Fax: (510) 741-5800
SIG 020996 Printed in USA
M1704017 Rev B
Loading...