Bio-Rad Bio-Scale Mini Bio-Gel P-6 Desalting Cartridges User Manual

Page 1
Bio-Scale™Mini Bio-Gel®P-6 Desalting Cartridges, 10 and 50 ml
Catalog # 732-5304, 732-5312, 732-5314
Page 2
Page 3
Table of Contents
Section 1...Introduction ........................................1
Section 2 Connecting to Bio-Rad’s
Low-Pressure Chromatography
Instruments ......................................24
Section 3 Connecting to Other Liquid
Chromatography Systems..................9
3.1 Connecting to BioLogic DuoFlow
Systems .................9
3.2 HPLC Systems ......................10
3.3 FPLC Systems .......................11
Section 4 Preparing a Cartridge for Use...........12
4.1 Sample Preparation ...............13
Section 5...Desalting and Buffer Exchange.........14
5.1 With a Liquid Chromatography
System ..................................15
Section 6 Technical Assistance ........................16
Page 4
Section 7 Ordering Information.........................17
Section 8 References .......................................19
Page 5
Section 1 Introduction
Bio-Scale Mini cartridges have a patent-pending double-wall design that provides extra durability and allows easy, reliable runs with aqueous buffers commonly used for protein purification. The polypropylene luer fittings and internal sealing surfaces assure leak-free operation at pressures up to 45 psi. Bio-Scale Mini cartridges are convenient, disposable, and supplied ready for use. They are easy-to-use and prepacked for fast, reproducible chromatographic separations. Cartridges are available for a variety of chromatographic techniques, including desalting, ion exchange and affinity chromatography. The patent-pending design of Bio-Scale Mini cartridges offers:
• Supplied ready for use; simply equilibrate the cartridge in the buffer of choice
1
Page 6
• Luer fittings for convenient connection to any chromatography system
Bio-Scale Mini Bio-Gel P-6 cartridges are packed with Bio-Gel
®
P-6 gel. This gel consists of hydrophilic spherical polyacrylamide beads designed for the purification of proteins, nucleic acids, viruses, plasmids, and other macromolecules. The Bio-Gel P-6 beads are designed to provide high-efficiency desalting or buffer exchange at low backpressure. Detailed product information is given in Table 1. See Ordering Information for a listing of the complete Bio-Scale Mini cartridge product line.
2
Page 7
Table 1. Bio-Scale Mini Bio-Gel P-6 cartridge specifications.
Size 10 and 50 ml bed volume
Dimensions 10 ml: 80 mm length x 12.6 mm inner
diameter
50 ml: 80 mm length x 28.2 mm inner
diameter
Maximum pressure tolerance 10 ml: 45 psi
50 ml: 40 psi
Recommended flow rates: at ambient temperature (18–24°C) 10 ml: 0.5–4.0 ml/min (24–192 cm/hr)
50 ml: 2.5–20.0 ml/min (24–192 cm/hr)
at cold room temperature (4–8°C) 10 ml: 0.5–2.0 ml/min (24–96 cm/hr)
50 ml: 2.5–10.0 ml/min (24–96 cm/hr)
Fittings Female luer-lock fitting inlet and
male luer fitting oulet
Column material Polypropylene
Frit material Polyethylene (HDPE)
Shipping conditions 20 mM Bis-Tris, pH 6.5 + 0.05% azide
Storage recommendation 20 mM Bis-Tris, pH 6.5 + 0.05% azide
Autoclavability Not autoclavable
3
Page 8
Section 2 Connecting to Bio-Rad’s Low-Pressure Chromatography Instruments
Bio-Scale Mini cartridges are ideal for use with Bio-Rad's BioLogic
LP chromatography system,
Econo
gradient pump, the patented* Model EP-1
Econo
pump, and all low-pressure chromatography instruments. Bio-Scale Mini cartridges can be con­veniently connected directly to the system using the luer fittings on the cartridge.
1. Install 1.6 mm inner diameter (ID) tubing in the
pumphead. Adjust platen pressure screw (on pumphead) – using a screwdriver or coin, turn the screw counterclockwise as far as it will go, then turn clockwise three full turns. Assemble with fittings and lock rings as shown in Figure 1.
*US Patent 5,135,658
4
Page 9
(Use orange lock rings and medium size barb fittings with 1.6 mm tubing.)
Fig. 1. BioLogic LP system setup.
5
Platen pressure
screw
See detail
Lock-ring
Tubing
Luer fitting
Platen pressure
screw
See detail
23
1
9
456
7 8
0
C .
Alarm
Lock-ring
Tubing
Luer fittimg
Page 10
2. To apply samples efficiently, install 1.6 mm ID
tubing from the pump to the MV-6 sample inject valve (if available). If using the MV-6 sample inject valve, turn the knob counterclockwise as far as it will go so it will now correspond to the printed diagram on the valve (see Figure 2).
Fig. 2. Connecting to an MV-6 valve.
6
SAMPLE LOOP
INJECT PORT
MV-6
BIOLOGIC LP
SYSTEM OR ECONO
GRADIENT PUMP
TO
INJECT
VALVE
FILL
TO
COLUMN
WASTE
SYSTEM OR ECONO
GRADIENT PUMP
BIOLOGIC LP
SAMPLE LOOP
INJECT PORT
FILL
MV-6
INJECT
TO
VALVE
"INJECT" POSITION"FILL" POSITION
TO
COLUMN
WASTE
Page 11
3. Connect the inlet of the cartridge to the male
luer fitting on the MV-6 sample inject valve (see Figure 2). If not using the MV-6 sample inject valve, connect a barb to male luer fitting on the
1.6 mm ID tubing, then connect to the top of the female luer on the Bio-Scale Mini cartridge. For optimum performance, a cartridge should be mounted vertically with the arrow on the cartridge pointing downward (see Figure 3).
4. Connect the cartridge outlet to the 1.6 mm ID
tubing leading to the BioLogic LP system optics module or to the Model EM-1 Econo
UV monitor. It is recommended to use the shortest length (approximately 10 cm) of 1.6 mm ID tubing. Connect a barb to female luer to the
1.6 mm ID tubing, then connect to the bottom of the male luer on the Bio-Scale Mini cartridge.
7
Page 12
Fig. 3. Cartridge and fittings.
8
Page 13
Section 3 Connecting to Other Liquid Chromatography Systems
Bio-Scale Mini cartridges can be connected to any liquid chromotography system, provieded that the maximum pressure limits for the 10 ml (3 bar, 45 psi, or 300 kPa) or for the 50 ml (2.6 bar, 40 psi, or 260 kPa) cartridges are not exceeded. It is recommended that the system pressure limit be set according to the cartridge pressure limit. Pressures in excess of 3 bar are usually caused by restrictions in tubing or detector cells downstream from the cartridge. Bio-Rad offers two fittings kits for easy connection of a Bio-Scale Mini cartridge to a BioLogic DuoFlow, HPLC, or FPLC-type system.
3.1 Connecting to a BioLogic DuoFlow System
The luer to BioLogic system fittings kit (catalog #732-0113) includes 1/4-28 female to male luer and 1/4-28 female to female luer to connect one Bio-Scale Mini cartridge to the BioLogic DuoFlow system, see Figure 4.
9
Page 14
Fig. 4. Luer to 1/4-28 adaptor.
3.2 HPLC Systems
The luer to 10-32 adaptor fittings kit (catalog #732-0112), provides fittings necessary to connect the Bio-Scale Mini cartridge to nut and ferrule type fittings found on most HPLC systems. Alternatively, the cartridge can be connected to HPLC systems via a low dead-volume 1/16 inch union with a new piece of stainless steel tubing attached to the union. Simply slip a short length of the 0.8 mm ID tubing over 1/16 inch OD stainless-steel tubing to a distance of 1 cm.
10
Page 15
3.3 FPLC Systems
The luer to M6 adaptor fittings kit (catalog #732-0111) provides fittings necessary to connect the Bio-Scale Mini cartridge to the M6 fittings found on FPLC or related systems. Alternatively, connection can be made by using one GE Healthcare Union luer lock female to M6 female fitting (GE 18-1027-12) and one female luer lock to M6 male fitting (Upchurch P-686 or GE 18-1027-62). To prevent tubing or cartridge failure, do not exceed the maximum recommended flow rate of the cartridge.
11
Page 16
Section 4 Preparing a Cartridge For Use
Bio-Scale Mini Bio-Gel P-6 cartridges contain 20 mM Bis-Tris, pH 6.5 + 0.05% sodium azide as the storage solution. The fully hydrated support is ready to use after equilibrating the cartridge in the buffer of choice. To perform a buffer exchange, connect the cartridge to a liquid chromatography system or peristaltic pump and condition it as instructed below:
1. Set the pump flow rate for the 10 ml cartridge to 4.0 ml/min (192 cm/hr) for room temperature use or 2.0 ml/min (96 cm/hr) for cold room temperature use.
2. Set the pump flow rate for the 50 ml cartridge to
20.0 ml/min (192 cm/hr) for room temperature use or 10.0 ml/min (96 cm/hr) for cold room temperature use.
12
Page 17
3. Equilibrate the cartridge with 2.2 column volumes of degassed equilibration buffer: 6 minutes at the room temperature flow rate setting or 12 minutes at the cold room temperature setting.
4. Reduce the flow rate to that which will be used in the buffer exchange or desalting cycle.
4.1 Sample Preparation
If the sample contains insoluble material, filter it with a 0.45 µm filter before applying it to the cartridge.
13
Page 18
Section 5 Desalting and Buffer Exchange
Samples can be desalted or the sample buffer exchanged by connecting the cartridge to a liquid chromatography system. The void volume for the 10 ml Bio-Scale Mini Bio-Gel P-6 Desalting cartridge is
3.6 ml and for the 50 ml Bio-Scale Mini Bio-Gel P-6
Desalting cartridge is 18 ml. With the 10 ml cartridge the desalted sample, the sample, or the sample in its new buffer is contained within the fraction between 3 ml and 9 ml with the highest concentration of sample in the 3.6 to 6.0 ml fraction. With the 50 ml cartridge the desalted sample or the sample or the sample in its new buffer is contained within the fraction between 15 ml and 45 ml with the highest concentration of sample in the 18 to 30 ml fraction.
14
Page 19
5.1 With a Liquid Chromatography System
1. Set the flow rate to the desired flow. a. 10 ml cartridge: 0.5–4.0 ml/min at room
temperature, 0.5–2.0 ml/min at cold room temperature.
b. 50 ml cartridge: 2.5–20 ml/min at room
temperature, 2.5–10 ml/min at cold room temperature.
2. Wash the prepared cartridge with degassed buffer as indicated below. a. 10 ml cartridge: 10 ml b. 50 ml cartridge: 50 ml
3. Inject up to the amount of sample indicated below, discarding the first amount of effluent indicated below. a. 10 ml cartridge: 6 ml of sample, discarding
the first 3 ml of effluent
b. 50 ml cartridge: 30 ml of sample, discarding
first 15 ml of effluent
4. Monitor the cartridge effluent with a UV monitor or conductivity monitor.
15
Page 20
Section 6 Technical Assistance
For additional information and technical assistance, contact your local Bio-Rad representative as listed on the back cover of our catalog, or in the US, call Technical Support at 1-800-4BIORAD.
16
Page 21
Section 7 Ordering Information
Bio-Scale Mini Cartridges*
Catalog #s
Description 5 x 1 ml 1 x 5 ml 5 x 5 ml
UNOsphere
Q Support 732-4100 731-4102 731-4104 UNOsphere S Support 732-4110 731-4112 731-4114 Macro-Prep
High Q Support 732-4120 732-4122 732-4124 Macro-Prep High S Support 732-4130 732-4132 732-4134 Macro-Prep DEAE Support 732-4140 732-4142 732-4144 Bio-Gel P-6 Support 732-4502 732-4504 Affi-Prep
®
Protein A Support 732-4600 732-4602
Profinity
IMAC Support 732-4610 732-4612 732-4614 Profinity GST Support 732-4620 732-4622 732-4624 Affi-Gel
®
DEAE Blue Support 732-4632 732-4634
Affi-Gel Blue Support 732-4642 732-4644
1x 5x
Bio-Gel P-6 Desalting, 10 ml 732-5304 Bio-Gel P-6 Desalting, 50 ml 732-5312 732-5314 * Visit www.bio-rad.com/cartridges/ for current information on
prepacked cartridges.
• Larger package sizes of media are available for process-scale
chromatography. Inquire with your local Bio-Rad representative.
17
Page 22
Fittings Kits
18
Catalog # Description
732-0111 Luer to M6 Adaptor Fittings Kit, includes luer to
732-0112 Luer to 10-32 Adaptor Fittings Kit, includes luer
732-0113 Luer BioLogic System Fittings Kit, includes
M6 fitting to connect to an FPLC system
to polypropylene/PTFE 10-32 fittings to connect 1 cartridge to an HPLC system
1/4-28 female to male luer and 1/4-28 female to female luer to connect 1 cartridge to the BioLogic DuoFlow system
Page 23
Section 8
19
References
Gagnon P, Avoiding Instrument Associated Variation in Purification Scale-up and Scale-down, BioPharm, 10 (3) 42–45, 1997
Harris ELV and Angal S, Protein Purification Methods, A Practical Approach, IRL Press, Oxford, 1989
Scopes RK, Protein Purification, Principles and Practice, (2nd Edition), Springer-Verlag, New York, 1987
Snyder LR and Kirkland JJ, Introduction to Modern Liquid Chromatography, (2nd Edition), John Wiley & Sons, Inc., New York, 1979
FPLC is trademark of GE Healthcare. Luer-Lok is trademark of Becton, Dickinson and Co. Triton is a trademark of Union Carbide.