This manual provides maintenance instructions for
the Site Master S820A Antenna and Coax/ Wave
guide Analyzer. It describes the product and provides
performance verification procedures, parts replace
-
ment procedures, and a replaceable parts list.
2. DESCRIPTION
The Site Master (Figure 1) is a hand held SWR/RL
(standing wave ratio/return loss) and Distance-ToFault measurement instrument. It combines a syn
thesized source, VSWR Bridge, and receiver on a
single printed circuit board (PCB). An optional
power monitor is also available. A block diagram is
shown in Figure 2.
3. PERFORMANCE VERIFICATION
Paragraphs 4 through 7 contain tests that can be
used to verify the performance of the Site Master
model S820A having any version of firmware.
3.1. Initial Setup for Testing
1.Press and hold the
then press the
ESCAPE/CLEAR key,
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
2.Release the
ESCAPE/CLEAR key and use
the Up/Down Arrow key to adjust the
contrast to give a readable display.
490 JARVIS DRIVE ¨ MORGAN HILL, CA 95037-2809
P/N: 10580-00031
REVISION D
PRINTED: FEBRUARY 2003
COPYRIGHT 1999-2003 ANRITSU CO.
RF
Source
Local
Oscillator
Multiplier
Wave
Module
Multiplier
CouplerCoupler
Test
Port
Incident
Local
Oscillator
Display
Figure 2. Site Master Block Diagram
4. FREQUENCY ACCURACY
The following test can be used to verify the CW frequency accuracy of the Site Master. Measurement
calibration of the Site Master is not required for this
test.
a. Equipment Required:
Reflected
Synchronous
Detector
A/D Converter
CPU
3.Press the
press the
4.Press the
press the
Serial
Port
F1 soft key, set to 9 GHz, then
ENTER key.
F2 soft key, set to 9 GHz, then
ENTER key.
5.Connect the RF cable from the Site Mas
ter Test Port to the RF Input on the
MS2602A.
-
·
Spectrum Analyzer
Anritsu Model MS2602A
b. Procedure:
6.Set up the Spectrum Analyzer as follows:
(a) Press Preset.
(b) Press Center and enter 1 GHz.
(c) Press the Max Hold button.
1.Press and hold the
then press the
Site Master. (This sets the instrument to
the factory preset state.)
ESCAPE/CLEAR key,
ON/OFF key to turn on the
7.If the Site Master has gone into the hold
mode, press the
RUN/HOLD key to make
the measurement.
8.Use the Spectrum Analyzer marker to
NOTE
Before continuing, allow a five
measure the center of the response. The
frequency should be 1 GHz ±75 kHz.
minute warm up for the internal
circuitry to stabilize.
NOTE
Nominal power is approximately
2.Press the
FREQ soft key.
0 - 10 dBm at 1 GHz.
2Site Master S820A MM
5. RETURN LOSS VERIFICATION
The following test can be used to verify the accuracy
of return loss measurements. Verification is per
formed by measuring the Source Match and Cor
-
rected Directivity. This method eliminates the uncer
tainties of measuring standard offset. Measurement
calibration of the Site Master is required for this
test.
Refer to the RF Measurement Chart on page 12, and
the Return Loss Verification worksheet on page 13.
8.Use the RF Measurement Chart to corre
late the peak-to-peak ripple reading to a
return loss value in dB. The measured
value should be:
£ –18 dB (3.3 - 10.5 GHz)
-
£ –14 dB (10.5 - 20 GHz).
9.Remove the 22KF50 Open / Short.
10.Connect the 29KF50-15 Precision Offset
Termination to the open end of the air
-
line.
11.Press the
AUTOSCALE key.
a. Equipment Required:
19K50 Precision Airline
·
22KF50 Open / Short
·
22K50 Open / Short
·
29KF50-15 Precision Offset Termination
·
· 28K50 Precision Termination
b. Procedure:
1.Press and hold the
then press the
ESCAPE/CLEAR key,
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
NOTE
Before continuing, allow a five
minute warm up for the internal
circuitry to stabilize.
2.Perform an OSL calibration over the fre
quency range of the unit using the 22K50
Open/Short and the 28K50 Precision Ter
mination. Verify that noise spikes are be
low 44 dB and that the trace is smooth.
Repeat as required.
12.Measure and record the worst case
peak-to-peak ripple.
13.Add this value to the mean value of the
29KF50-15 Precision Offset Termination.
14.Use the RF Measurement Chart to corre
late the reading to a return loss value in
dB. The measured value should be:
£ –29 dB (3.3 - 20 GHz).
15.Remove the airline and Precision Offset
Termination.
-
-
-
3.Remove the 28K50 Termination.
4.Connect a 19K50 Precision Airline to the
test port.
5.Connect a 22KF50 Open / Short to the
open end of the airline.
6.Press the
AUTOSCALE key.
7.Measure and record the worst case
peak-to-peak ripple.
Site Master S820A MM3
6. POWER MONITOR VERIFICATION
If the Power Monitor (Option 5) is installed in the
Site Master, the following test can be used to verify
the accuracy of the power measurements. Measure
ment calibration of the Site Master is not required
for this test.
a. Equipment Required:
2.Connect the MA2418A Reference Source
to the input of the 560-7N50B RF detec
-
tor.
-
3.Connect the RF Detector output to the
RF Detector input of the Site Master.
4.Connect the DC power supply to the ap
propriate line voltage to supply power to
the MA2418A Reference Source.
RF Detector, 10 MHz to 20 GHz,
·
Anritsu 560-7N50B
10 dB Attenuator, Weinshel 1R-10
·
30 dB Attenuator, Weinshel 1R-30
·
RF Reference Source, 0.050 GHz,
·
Anritsu MA2418A
DC Power Supply, Anritsu 2000-933
·
b. Procedure
1.Connect the DC power supply to the
MA2418A Reference Source. (Refer to
Figure 3.)
560-7N50B
RF Detector
Interface
Serial
12V
External
Charging
Battery
(45C mA)
12-15V DC
Port
Test
820A
5.Press and hold the
then press the
ESCAPE/CLEAR key,
ON/OFF key to turn on the
Site Master. (This sets the instrument to
the factory preset state.)
6.Press the
MODE soft key.
7.Use the Up/Down Arrow key to highlight
POWER MONITOR, then press ENTER.
8.Press the
ZERO soft key to zero the power
monitor.
When complete,
ZERO ADJ:ON is dis
-
played in the message area.
9.Verify that the power monitor reading is
0.0 dBm ±1 dB.
Attenuators
MA2418A
Reference Source
MA2418AReference Source
W
50MHz, 50
0dBm
Det
RF
20mA
15-24VDC
AUTO
SCALE
1
START
CAL
2
ESCAPE
CLEAR
CAL
3
SAVE
SETUP
4
RECALL
SETUP
5
LIMIT
6
MARKER
7
SAVE
DISPLAY
8
ENTER
RECALL
DISPLAY
9
0
RUN
ON
HOLD
OFF
PRINT
2000-933 DC Power Supply
Site Master w/ Option 5 Installed
Figure 3. Power Monitor Verification
4Site Master S820A MM
10.Connect the output of the MA2418A Ref
erence Source to the two attenuators so
as to add 40 dB of attenuation (Figure 3).
11.Connect the MA2418A Reference Source
and the attenuators to the input of the
560-7N50B RF detector.
-
b. Procedure
1.Connect the test equipment as shown in
Figure 3.
2.Press the Power key on the 541XXA to
On.
12.Verify that the power monitor reading is
now –40.0 dBm ±2 dB.
7. TERMINATION VERIFICATION
This test can be used to verify the accuracy of termi
nations used with the Site Master. The test uses the
precision return loss mode of the 541XXA Scalar
Measurement System. Measurements of termina
tions using this mode provide results that are trace
able to the NIST (National Institute of Standards
and Technology) standards for the precision airline.
a. Equipment Required:
· Scalar Measurement System,
Anritsu 54161A/69A
· Offset SWR Autotester,
Anritsu 560-98KF50-15
· Precision Airline, Anritsu 19KF50
· Open/Short, Anritsu 22K50
· 50 Ohm Termination, Anritsu 28K50
·
Source Adapter, Anritsu K220B
3.Press the System Menu key.
4.Using the Menu up-down keys: Highlight
RESET, then press the Select key.
5.At the RESET MENU display, use the
-
Menu up-down keys to highlight
RESET TO FACTORY DEFAULTS, then
press the Select key.
-
6.Set the signal source for the frequency
range as follows:
(a) Press the Frequency key.
(b) Using the Data Entry Keypad or
Data Entry Knob, set the START fre
quency to 0.01 GHz. Press the Enter
key.
(c) Using the Data Entry Keypad or
Data Entry Knob, set the STOP frequency to 20.0 GHz. Press the Enter
key.
7.Press the Channel 2 Display On/Off key
to Off.
8.Press the Channel 1 Menu key.
Open/Short
50 Ohm
Termination
541XXA Scalar Measurement System
B
A
Source
Adapter
Figure 4. 541XXA Precision Return Loss TestSetup
Precision
Air Line
Termination
under Test
Beadless
End
Offset SWR
Autotester
Connect Dashed Line Connections
When Directed By Procedure
Site Master S820A MM5
Loading...
+ 11 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.