Anritsu S820A User Manual

MAINTENANCE MANUAL
Site Master™ S820A
Antenna and Coax/Waveguide Analyzer
Interface
Serial
Battery
Charging
12V External
(45C mA)
12-15V DC
Port
Test
Det
820A
1
START CAL
3
SAVE SETUP
5
LIMIT
7
SAVE DISPLAY
9
ON OFF
AUTO SCALE
CAL
RECALL SETUP
MARKER
RECALL DISPLAY
PRINT
RF
2
ESCAPE CLEAR
4 6 8
ENTER
0
RUN
HOLD
Figure 1. Site Master S820A
1. INTRODUCTION
This manual provides maintenance instructions for the Site Master S820A Antenna and Coax/ Wave
­guide Analyzer. It describes the product and provides performance verification procedures, parts replace
-
ment procedures, and a replaceable parts list.
2. DESCRIPTION
The Site Master (Figure 1) is a hand held SWR/RL (standing wave ratio/return loss) and Distance-To­Fault measurement instrument. It combines a syn
­thesized source, VSWR Bridge, and receiver on a single printed circuit board (PCB). An optional power monitor is also available. A block diagram is shown in Figure 2.
3. PERFORMANCE VERIFICATION
Paragraphs 4 through 7 contain tests that can be used to verify the performance of the Site Master model S820A having any version of firmware.
3.1. Initial Setup for Testing
1. Press and hold the then press the
ESCAPE/CLEAR key,
ON/OFF key to turn on the
Site Master. (This sets the instrument to the factory preset state.)
2. Release the
ESCAPE/CLEAR key and use
the Up/Down Arrow key to adjust the contrast to give a readable display.
490 JARVIS DRIVE ¨ MORGAN HILL, CA 95037-2809
P/N: 10580-00031
REVISION D
PRINTED: FEBRUARY 2003
COPYRIGHT 1999-2003 ANRITSU CO.
RF
Source
Local
Oscillator
Multiplier
Wave Module
Multiplier
Coupler Coupler
Test
Port
Incident
Local
Oscillator
Display
Figure 2. Site Master Block Diagram
4. FREQUENCY ACCURACY
The following test can be used to verify the CW fre­quency accuracy of the Site Master. Measurement calibration of the Site Master is not required for this test.
a. Equipment Required:
Reflected
Synchronous
Detector
A/D Converter
CPU
3. Press the press the
4. Press the press the
Serial
Port
F1 soft key, set to 9 GHz, then
ENTER key.
F2 soft key, set to 9 GHz, then
ENTER key.
5. Connect the RF cable from the Site Mas ter Test Port to the RF Input on the MS2602A.
-
·
Spectrum Analyzer
Anritsu Model MS2602A
b. Procedure:
6. Set up the Spectrum Analyzer as follows: (a) Press Preset. (b) Press Center and enter 1 GHz. (c) Press the Max Hold button.
1. Press and hold the then press the Site Master. (This sets the instrument to the factory preset state.)
ESCAPE/CLEAR key,
ON/OFF key to turn on the
7. If the Site Master has gone into the hold mode, press the
RUN/HOLD key to make
the measurement.
8. Use the Spectrum Analyzer marker to
NOTE
Before continuing, allow a five
measure the center of the response. The frequency should be 1 GHz ±75 kHz.
minute warm up for the internal circuitry to stabilize.
NOTE
Nominal power is approximately
2. Press the
FREQ soft key.
0 - 10 dBm at 1 GHz.
2 Site Master S820A MM
5. RETURN LOSS VERIFICATION
The following test can be used to verify the accuracy of return loss measurements. Verification is per formed by measuring the Source Match and Cor
-
­rected Directivity. This method eliminates the uncer tainties of measuring standard offset. Measurement calibration of the Site Master is required for this test.
Refer to the RF Measurement Chart on page 12, and the Return Loss Verification worksheet on page 13.
8. Use the RF Measurement Chart to corre
­late the peak-to-peak ripple reading to a return loss value in dB. The measured value should be: £ –18 dB (3.3 - 10.5 GHz)
-
£ –14 dB (10.5 - 20 GHz).
9. Remove the 22KF50 Open / Short.
10. Connect the 29KF50-15 Precision Offset Termination to the open end of the air
-
line.
11. Press the
AUTOSCALE key.
a. Equipment Required:
19K50 Precision Airline
·
22KF50 Open / Short
·
22K50 Open / Short
·
29KF50-15 Precision Offset Termination
·
· 28K50 Precision Termination
b. Procedure:
1. Press and hold the then press the
ESCAPE/CLEAR key,
ON/OFF key to turn on the
Site Master. (This sets the instrument to the factory preset state.)
NOTE
Before continuing, allow a five minute warm up for the internal circuitry to stabilize.
2. Perform an OSL calibration over the fre quency range of the unit using the 22K50 Open/Short and the 28K50 Precision Ter mination. Verify that noise spikes are be low 44 dB and that the trace is smooth. Repeat as required.
12. Measure and record the worst case peak-to-peak ripple.
13. Add this value to the mean value of the 29KF50-15 Precision Offset Termination.
14. Use the RF Measurement Chart to corre
­late the reading to a return loss value in dB. The measured value should be: £ –29 dB (3.3 - 20 GHz).
15. Remove the airline and Precision Offset Termination.
-
-
-
3. Remove the 28K50 Termination.
4. Connect a 19K50 Precision Airline to the test port.
5. Connect a 22KF50 Open / Short to the open end of the airline.
6. Press the
AUTOSCALE key.
7. Measure and record the worst case peak-to-peak ripple.
Site Master S820A MM 3
6. POWER MONITOR VERIFICATION
If the Power Monitor (Option 5) is installed in the Site Master, the following test can be used to verify the accuracy of the power measurements. Measure ment calibration of the Site Master is not required for this test.
a. Equipment Required:
2. Connect the MA2418A Reference Source to the input of the 560-7N50B RF detec
-
tor.
-
3. Connect the RF Detector output to the RF Detector input of the Site Master.
4. Connect the DC power supply to the ap
­propriate line voltage to supply power to the MA2418A Reference Source.
RF Detector, 10 MHz to 20 GHz,
·
Anritsu 560-7N50B
10 dB Attenuator, Weinshel 1R-10
·
30 dB Attenuator, Weinshel 1R-30
·
RF Reference Source, 0.050 GHz,
·
Anritsu MA2418A
DC Power Supply, Anritsu 2000-933
·
b. Procedure
1. Connect the DC power supply to the MA2418A Reference Source. (Refer to Figure 3.)
560-7N50B
RF Detector
Interface
Serial
12V External
Charging
Battery
(45C mA)
12-15V DC
Port
Test
820A
5. Press and hold the then press the
ESCAPE/CLEAR key,
ON/OFF key to turn on the
Site Master. (This sets the instrument to the factory preset state.)
6. Press the
MODE soft key.
7. Use the Up/Down Arrow key to highlight
POWER MONITOR, then press ENTER.
8. Press the
ZERO soft key to zero the power
monitor. When complete,
ZERO ADJ:ON is dis
-
played in the message area.
9. Verify that the power monitor reading is
0.0 dBm ±1 dB.
Attenuators
MA2418A Reference Source
MA2418AReference Source
W
50MHz, 50 0dBm
Det
RF
20mA
15-24VDC
AUTO SCALE
1
START CAL
2
ESCAPE CLEAR
CAL
3
SAVE SETUP
4
RECALL SETUP
5
LIMIT
6
MARKER
7
SAVE DISPLAY
8
ENTER
RECALL DISPLAY
9
0
RUN
ON
HOLD
OFF
PRINT
2000-933 DC Power Supply
Site Master w/ Option 5 Installed
Figure 3. Power Monitor Verification
4 Site Master S820A MM
10. Connect the output of the MA2418A Ref erence Source to the two attenuators so as to add 40 dB of attenuation (Figure 3).
11. Connect the MA2418A Reference Source and the attenuators to the input of the 560-7N50B RF detector.
-
b. Procedure
1. Connect the test equipment as shown in Figure 3.
2. Press the Power key on the 541XXA to On.
12. Verify that the power monitor reading is now –40.0 dBm ±2 dB.
7. TERMINATION VERIFICATION
This test can be used to verify the accuracy of termi nations used with the Site Master. The test uses the precision return loss mode of the 541XXA Scalar Measurement System. Measurements of termina
­tions using this mode provide results that are trace able to the NIST (National Institute of Standards and Technology) standards for the precision airline.
a. Equipment Required:
· Scalar Measurement System, Anritsu 54161A/69A
· Offset SWR Autotester, Anritsu 560-98KF50-15
· Precision Airline, Anritsu 19KF50
· Open/Short, Anritsu 22K50
· 50 Ohm Termination, Anritsu 28K50
·
Source Adapter, Anritsu K220B
3. Press the System Menu key.
4. Using the Menu up-down keys: Highlight RESET, then press the Select key.
5. At the RESET MENU display, use the
-
Menu up-down keys to highlight RESET TO FACTORY DEFAULTS, then press the Select key.
-
6. Set the signal source for the frequency range as follows:
(a) Press the Frequency key. (b) Using the Data Entry Keypad or
Data Entry Knob, set the START fre
­quency to 0.01 GHz. Press the Enter key.
(c) Using the Data Entry Keypad or
Data Entry Knob, set the STOP fre­quency to 20.0 GHz. Press the Enter key.
7. Press the Channel 2 Display On/Off key to Off.
8. Press the Channel 1 Menu key.
Open/Short
50 Ohm
Termination
541XXA Scalar Measurement System
B
A
Source Adapter
Figure 4. 541XXA Precision Return Loss TestSetup
Precision
Air Line
Termination
under Test
Beadless End
Offset SWR
Autotester
Connect Dashed Line Connections When Directed By Procedure
Site Master S820A MM 5
Loading...
+ 11 hidden pages