No p art o f this manu al may be re produce d in
any form or by any means (including electronic storage and retrieval or translation
into a foreign language) without prior agreement and written consent from Agilent
Technologies, Inc. as governed by United
States and international copyright laws.
Manual Part Number
54684-97024
Edition
Eighth Edition, January 2008
Printed in Malaysia
Agilent Technologies, Inc.
395 Page Mill Road
Palo Alto, CA 94303 USA
A newer version of this manual
may be available at
www.agilent.com/find/mso6000
Software Revision
This guide was written for version 05.00 of
the Agilent InfiniiVision 6000 Series Oscilloscope software.
Trademark Acknowledgments
Java is a U.S. trademark of Sun Microsystems, Inc.
Sun, Sun Microsystems, and the Sun Logo
are trademarks or registered trademarks of
Sun Microsystems, Inc. in the U.S. and other
countries.
Windows and MS Windows are U.S. registered trademarks of Microsoft Corporation.
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a license
and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
If software is for use in the performance of a
U.S. Government prime contract or subcontract, Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June 1995),
or as a “commercial item” as defined in FAR
2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June
1987) or any equivalent agency regulation or
contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license
terms, and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted Rights as
defined in FAR 52.227-19(c)(1-2) (June
1987). U.S. Government users will receive
no greater than Limited Rights as defined in
FAR 52.227-14 (June 1987) or DFAR
252.227-7015 (b)(2) (November 1995), as
applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and
met.
6000 Series Oscilloscope User’s Guide
In This User’s Guide…
1Getting Started
2Front-Panel Controls
3Viewing and Measuring Digital Signals
4Triggering the Oscilloscope
5Making Measurements
6Displaying Data
This guide shows you how to use the Agilent InfiniiVision
6000 Series oscilloscopes. It contains the following chapters and
topics:
Unpacking and setting up your oscilloscope, using Quick Help.
A quick overview of the front-panel controls.
How to connect and use the digital channels of a mixed-signal
oscilloscope (MSO).
Trigger modes, coupling, noise rejection, holdoff, external
trigger and more. Edge, pulse width, and pattern triggering.
CAN, duration, I
TV/video, and USB triggering modes.
XY mode, FFTs, math functions, using cursors, automatic
measurements.
Using pan and zoom; normal, average, peak detect, and high
resolution (smoothing) modes; noise rejection modes, glitch
capture, and more.
2
C, Nth Edge Burst, LIN, sequence, SPI,
7Saving and Printing Data
Printing waveforms, saving setups and data, and using the file
explorer.
8Reference
Upgrading a DSO to an MSO, adding memory, software updates,
I/O, synchronizing instruments with the 10 MHz reference
clock, warranty status, digital signal probing, and more.
9Power and Environmental Conditions
Specifications and characteristics of the oscilloscope.
6000 Series Oscilloscope User’s Guide3
The Agilent InfiniiVision 6000 Series oscilloscopes deliver
powerful features and high performance:
The following options can be easily installed without returning
the oscilloscope to a Service Center. These upgrades are
licensed.
Tab l e 4Upgrade Options
Licensed OptionOrder
Mixed Signal Oscilloscope
(MSO)
I2C/SPI serial decode
option (for 4 channel or
4+16 channel models only)
Order N2914A or N2915A (see data sheet). You
can easily install this option yourself. The logic
cable kit is supplied with the MSO license.
Order N5423A after purchase (Option LSS at time
of purchase). You can easily install this option
yourself.
CAN/LIN automotive
triggering and decode (for 4
channel or 4+16 channel
models only)
N5406A FPGA dynamic
probe for Xilinx
N5434A FPGA dynamic
probe for Altera
Order N5424A after purchase (Option AMS at
time of purchase). You can easily install this
option yourself.
N5406A with Option 001 (Oscilloscope-locked
license) or Option 002 (PC-locked license).
N5434A with Option 001 (Oscilloscope-locked
license) or Option 002 (PC-locked license).
66000 Series Oscilloscope User’s Guide
The following option cannot be installed after time of purchase.
Tab l e 5Order-Only Options
Licensed OptionOrder
Battery Operation (Option BAT)Available at time of purchase. Option
can not be added after purchase.
Visit www.agilent.com/find/mso6000 to view the 6000A Series and 6000L Series
data sheets.
6000 Series Oscilloscope User’s Guide7
Built-in Quick Help
A Quick Help system is built into the oscilloscope. Instructions for using the
quick help system are given on page 58.
Digital Channels
Because all of the oscilloscopes in the Agilent 6000 Series have analog channels,
the analog channel topics in this book apply to all instruments. Whenever a topic
discusses the digital channels, that information applies only to Mixed-Signal
Oscilloscope (MSO) models or DSO models that have been upgraded to an MSO.
Using this book with the 6000L Series oscilloscopes
The 6000L Series oscilloscopes do not have a built-in display or front panel
control keys. If you are using a 6000L Series oscilloscope, and this book refers to
using front panel controls, you can use the built-in Web control feature described
on page 44 to complete the instructions.
Abbreviated instructions for pressing a series of keys
Instructions for pressing a series of keys are written in an abbreviated manner.
Instructions for pressing Key1, then pressing Key2, then pressing Key3 are
abbreviated as follows:
Press Key1 & Key2 & Key3.
The keys may be front panel keys, or softkeys, which are located directly below
the oscilloscope display.
86000 Series Oscilloscope User’s Guide
Contents
1Getting Started19
To inspect package contents22
To adjust the 6000A Series handle 28
To mount the oscilloscope in a rack29
To mount the 6000A Series oscilloscope in a rack29
To mount the 6000L Series oscilloscope in a rack29
To establish a LAN connection (6000A Series)39
To establish a LAN connection (6000L Series)40
To establish a point-to-point LAN connection42
To use the Web interface43
Controlling the oscilloscope using a Web browser44
Setting a password46
Scrolling and Monitor Resolution49
Identify Function49
Printing the oscilloscope’s display from a web browser50
To connect the oscilloscope probes51
Maximum input voltage in 50 Ω mode 51
6000 Series Oscilloscope User’s Guide9
Contents
Maximum input voltage for analog inputs 52
To verify basic oscilloscope operation52
To compensate the oscilloscope probes54
To calibrate the probes55
Passive Probes Supported55
Active Probes Supported56
By 300 MHz, 500 MHz, and 1 GHz Bandwidth Models56
By 100 MHz Bandwidth Models57
Using Quick Help58
Quick Help Languages59
2Front-Panel Controls61
6000L Series Oscilloscope Controls62
Front and Rear Panel Controls and Connectors63
6000A Series Oscilloscope Front-Panel Controls66
Conventions67
Graphic Symbols in Softkey Menus67
4-Channel 6000A Series Oscilloscope Front Panel68
Front Panel Controls69
2-Channel 6000A Series Oscilloscope Front Panel (differences
only)74
Interpreting the display75
106000 Series Oscilloscope User’s Guide
6000A Series Front-Panel Operation76
To adjust the waveform intensity76
To adjust the display grid (graticule) intensity76
To start and stop an acquisition77
To make a single acquisition78
To pan and zoom79
Choosing Auto trigger mode or Normal trigger mode80
Using AutoScale80
To set the probe attenuation factor81
Using the analog channels83
To set up the Horizontal time base88
To make cursor measurements95
To make automatic measurements96
Using Labels97
To print the display101
To s et the cl oc k102
To set up the screen saver103
To set the waveform expansion reference point104
To perform service functions105
User Calibration105
Self Test108
About Oscilloscope108
To restore the oscilloscope to its default configuration110
Contents
3Viewing and Measuring Digital Signals111
To connect the digital probes to the circuit under test112
Use only Agilent digital probe cable 112
Acquiring waveforms using the digital channels115
To display digital channels using AutoScale116
Example116
6000 Series Oscilloscope User’s Guide11
Contents
Interpreting the digital waveform display118
To switch all digital channels on or off119
To switch groups of channels on or off119
To switch a single channel on or off119
To change the displayed size of the digital channels120
To reposition a digital channel120
To change the logic threshold for digital channels121
To display digital channels as a bus122
4Triggering the Oscilloscope127
Selecting Trigger Modes and Conditions130
To select the Mode and Coupling menu130
Trigger modes: Normal and Auto131
To select trigger Coupling133
To select trigger Noise Rejection and HF rejection133
To set Holdoff134
The External Trigger input136
2-Channel oscilloscope External Trigger input136
Maximum input voltage for external trigger (2-channel oscilloscopes) 137
4-Channel oscilloscope External Trigger input138
Maximum input voltage for external trigger (4-channel oscilloscopes) 138
Trig ge r Typ es139
To use Edge triggering140
Trigger level adjustment141
126000 Series Oscilloscope User’s Guide
To use Pulse Width triggering142
< qualifier time set softkey144
> qualifier time set softkey144
To use Pattern triggering145
Hex Bus Pattern Triggering147
To use CAN triggering148
To use Duration triggering152
< qualifier time set softkey154
> qualifier time set softkey154
To use FlexRay triggering155
Modes of VPT1000 Control/Operation155
Setting Up the Oscilloscope and the VPT1000156
Triggering on FlexRay Frames, Times, or Errors160
To use I2C triggering164
To use Nth Edge Burst triggering170
Contents
To use LIN triggering172
To use Sequence triggering175
Define the Find: stage177
Define the Trigger on: stage178
Define the optional Reset on: stage180
Adjust the trigger level181
To use SPI triggering182
Assign source channels to the clock, data, and frame
signals184
Set up the number of bits in the serial data string and set values
for those data bits187
Resetting all bits in the serial data string to one value187
6000 Series Oscilloscope User’s Guide13
Contents
To use TV triggering187
Example exercises191
To trigger on a specific line of video191
To trigger on all sync pulses193
To trigger on a specific field of the video signal194
To trigger on all fields of the video signal195
To trigger on odd or even fields196
To use UART/RS232 triggering199
To use USB triggering204
The Trigger Out connector206
Tr ig g er s206
Source frequency206
Source frequency/8206
5Making Measurements207
146000 Series Oscilloscope User’s Guide
To use the XY horizontal mode208
Math Functions213
Math scale and offset214
Multiply215
Subtract217
Differentiate219
Integrate221
FFT Measurement223
FFT Operation225
Square Root230
Cursor Measurements232
To make cursor measurements232
Cursor Examples236
Automatic Measurements239
To make an automatic measurement240
To set measurement thresholds241
Time Measurements243
Delay and Phase Measurements247
Voltage Measurements249
Overshoot and Preshoot Measurements255
6Displaying Data257
Pan and Zoom258
To pan and zoom a waveform259
To set the waveform expansion reference point259
Antialiasing260
Using the XGA video output260
Display Settings261
Infinite persistence261
Grid intensity262
Vectors (connect the dots)262
Contents
Varying the intensity to view signal detail263
Acquisition Modes265
At Slower Sweep Speeds265
Selecting the Acquisition mode265
Normal Mode266
Peak Detect Mode266
High Resolution Mode266
Averaging Mode267
Realtime Sampling Option269
6000 Series Oscilloscope User’s Guide15
Contents
Using Serial Decode271
To d ec ode I
To decode SPI data276
To decode CAN data281
CAN Totalizer286
To decode LIN data288
To decode FlexRay data294
FlexRay Totalizer298
To decode UART/RS232 data300
UART/RS232 Totalizer306
To reduce the random noise on a signal307
HF Reject307
LF Reject308
Noise rejection308
To capture glitches or narrow pulses with peak detect and infinite
persistence309
Using peak detect mode to find a glitch310
2
C data272
How AutoScale Works312
Undo AutoScale312
Specifying the Channels Displayed After AutoScale313
Preserving the Acquisition Mode During AutoScale313
Selecting a file name322
Waveform Trace and Oscilloscope Setup324
Display Image and Waveform Data Files324
Choosing save settings325
To save a waveform and/or setup to a USB device327
To save a waveform and/or setup to the oscilloscope’s internal
memory327
To recall waveform trace and/or oscilloscope setup328
File explorer328
To use the file explorer330
Software and firmware updates334
To set up the I/O port335
Using the 10 MHz reference clock336
Sample clock and frequency counter accuracy336
Supplying an external timebase reference336
To supply a sample clock to the oscilloscope336
To synchronize the timebase of two or more instruments338
To check warranty and extended services status338
To return the instrument339
To clean the oscilloscope339
6000 Series Oscilloscope User’s Guide17
Contents
Digital channel signal fidelity: Probe impedance and
grounding340
Input Impedance340
Probe Grounding342
Best Probing Practices344
To replace digital probe leads345
Binary Data (.bin)346
Binary Data in MATLAB346
Binary Header Format346
Example Program for Reading Binary Data350
Examples of Binary Files351
To inspect package contents 22
To adjust the 6000A Series handle 28
To mount the oscilloscope in a rack 29
To power-on the oscilloscope 33
Ventilation requirements 32
The remote interface 38
To establish a LAN connection (6000A Series) 39
To establish a LAN connection (6000L Series) 40
To establish a point-to-point LAN connection 42
To use the Web interface 43
Printing the oscilloscope’s display from a web browser 50
To connect the oscilloscope probes 51
To verify basic oscilloscope operation 52
To compensate the oscilloscope probes 54
To calibrate the probes 55
Passive Probes Supported 55
Active Probes Supported 56
Using Quick Help 58
To get started using the oscilloscope:
✔ Unpack the oscilloscope and verify the contents.
✔ Adjust the 6000A Series oscilloscope’s handle position.
✔ Apply power to the oscilloscope.
✔ Connect the probes to the oscilloscope.
A
19
1Getting Started
✔ Verify basic oscilloscope operation and compensate the
probes.
206000 Series Oscilloscope User’s Guide
Getting Started1
Built-in Quick Help
A Quick Help system is built into the oscilloscope. Instructions for using the
quick help system are given on page 58.
Digital Channels
Because all of the oscilloscopes in the Agilent 6000 Series have analog channels,
the analog channel topics in this book apply to all instruments. Whenever a topic
discusses the digital channels, that information applies only to Mixed-Signal
Oscilloscope (MSO) models or DSO models that have been upgraded to an MSO.
Using this book with the 6000L Series oscilloscopes
The 6000L Series oscilloscopes do not have a built-in display or front panel
control keys. If you are using a 6000L Series oscilloscope, and this book refers to
using front panel controls, you can use the built-in Web control feature described
on page 44 to complete the instructions.
Abbreviated instructions for pressing a series of keys
Instructions for pressing a series of keys are written in an abbreviated manner.
Instructions for pressing Key1, then pressing Key2, then pressing Key3 are
abbreviated as follows:
Press Key1 & Key2 & Key3.
The keys may be front panel keys, or softkeys, which are located directly below
the oscilloscope display.
6000 Series Oscilloscope User’s Guide21
1Getting Started
To inspect package contents
✔ Inspect the shipping container for damage.
If your shipping container appears to be damaged, keep the
shipping container or cushioning material until you have
inspected the contents of the shipment for completeness and
have checked the oscilloscope mechanically and electrically.
✔ Verify that you received the following items and any optional
accessories you may have ordered:
• InfiniiVision 6000 Series Oscilloscope
• Front-panel cover (6000A Series only)
• Power cord (see Table 8 on page 37)
• Models with Option BAT only: Power Supply (P/N
0950-4866)
• LAN Crossover Cable 5061-0701 (6000L Series only)
• GPIB cable extender P/N 5183-0803 (6000L Series only)
Figure 3Package contents for 6000L Series oscilloscopes
6000 Series Oscilloscope User’s Guide25
1Getting Started
Tab l e 6Accessories available
ModelDescription
N2918A6000 Series Oscilloscope Evaluation Kit
1180CZTestmobile oscilloscope cart (requires N2919A adapter kit)
N2919ATestmobile Adapter Kit
N2916A6000A Rackmount Kit
54684-44101Front-panel cover
N2605A-097USB cable
10833AGPIB cable, 1 m long
10073CPassive probe, 10:1, 500 MHz, 1.5 m
10074CPassive probe, 10:1, 100 MHz, 1.5 m
54620-68701Digital probe kit
54684-42301Digital probe cable guide (cable tray)
0960-030150-Ohm Feedthrough
01650-6160716:16 logic cable and terminator (use with header on target sys.)
54620-6870116:2 x 8 logic input probe assembly (standard with MSO models)
10070CPassive probe, 1:1 20 MHz, 1.5 m
10074CPassive probe, 10:1, 150 MHz, 1.5 m
10073CPassive probe, 10:1, 500 MHz, 1.5 m
1165APassive probe, 10:1, 600 MHz, 1.5 m
10076APassive probe, 100:1, 4 kV, 250 MHz
N2771APassive probe, 1000:1, 30 kV, 50 MHz
1156AActive probe, 1.5 GHz AutoProbe interface
1144AActive probe, 800 MHz (requires 1142A – power supply)
†
1145A
†
1130A
N2772AActive differential probe, 20 MHz, 1.2 kVDC + peak AC max
1141AActive differential probe, 200 MHz, 200 VDC + peak AC max
1146ACurrent probe, 100 kHz, 100 A, AC/DC
†
1147A
N2780ACurrent probe, 2 MHz, 500 A, AC/DC (use with N2779A power
N2781ACurrent probe, 10 MHz, 150 A, AC/DC (use with N2779A power
Active probe, 750 MHz 2-ch (requires 1142A – power supply)
For active differential probes: 1.5 GHz InfiniiMax amplifier with
AutoProbe interface (requires one or more InfiniiMax probe head
– E2675A, E2668A, E2669A).
(requires N2773A power supply)
(requires 1142A power supply)
Current probe, 50 MHz, 30 A, AC/DC with AutoProbe interface
supply)
supply)
266000 Series Oscilloscope User’s Guide
Getting Started1
ModelDescription
N2782ACurrent probe, 50 MHz, 30 A, AC/DC (use with N2779A power
supply)
N2783ACurrent probe, 100 MHz, 30 A, AC/DC (use with N2779A power
supply)
10072AFine-pitch probe kit
10075A0.5 mm IC clip kit
10076A100:1, 4 kV 250 MHz probe
E2613B0.5 mm Wedge probe adapter, 3-signal, qty 2
E2614A0.5 mm Wedge probe adapter, 8-signal, qty 1
E2615B0.65 mm Wedge probe adapter, 3-signal, qty 2
E2616A0.65 mm Wedge probe adapter, 8-signal, qty 1
E2643A0.5 mm Wedge probe adapter, 16-signal, qty 1
E2644A0.65 mm Wedge probe adapter, 16-signal, qty 1
†
Indicates a maximum of two of this model probe can be connected to each
oscilloscope due to AutoProbe interface current supply limitation. See also
“Passive Probes Supported” on page 55 and
“Active Probes Supported” on page 56.
You can find these items at www.agilent.com or at www.parts.agilent.com.
For information on more probes and accessories see
“5989-6162EN Probes and Accessories Selection Guide” and
“5968-8153EN 5000 and 6000 Series Oscilloscope Probes and
Accessories data sheet, available at www.agilent.com.
6000 Series Oscilloscope User’s Guide27
1Getting Started
To adjust the 6000A Series handle
You can use the oscilloscope’s handle for carrying the
instrument, or you can use it as a stand to tilt the instrument up
for easier viewing of its display.
1 Grasp the handle hubs on each side of the instrument and
pull the hubs out until they stop.
286000 Series Oscilloscope User’s Guide
2 Without releasing the hubs, rotate the handle to the desired
position. Then release the hubs. Continue rotating the handle
until it clicks into a set position.
To mount the oscilloscope in a rack
The InfiniiVision 6000 Series oscilloscopes can be mounted into
Electronic Industries Association (EIA) standard 19-inch
(487-mm) rack cabinets.
To mount the 6000A Series oscilloscope in a rack
Purchase and install the N2916A rack mount kit. Instructions
are included in the kit.
To mount the 6000L Series oscilloscope in a rack
The 6000L Series oscilloscope is supplied with all necessary
hardware for installation into a standard EIA 19-inch rack.
Tab l e 7Rack Mount Hardware Supplied
Getting Started1
Quantity DescriptionAgilent Part
2Front Extender SupportD6104-01201
2Rear Extender SupportD6104-01202
4Rear Extender Screw (M3 x 6 mm)0515-0430
4Dress Screw (10-32 x 0.0625)0570-1577
8Rail Screw (10-32 x 0.375)2680-0281
12Clip-nut (10-32)0590-0804
Number
Tools required (not supplied)
• #2 Phillips screwdriver
• T20 Torx driver
• T10 Torx driver
6000 Series Oscilloscope User’s Guide29
1Getting Started
Step 4
If needed
Step 1,
step 5
Step 3
If needed
1 Loosely attach the Front Extender Supports to the Rear
Extender Supports with four (4) clip-nuts and four (4) of the
10-32 x 0.375 Rail Screws. (The screws require a Torx T20
driver.) Choose the correct set of slots in the supports such
that their overall length is approximately correct for the
depth of your cabinet.
306000 Series Oscilloscope User’s Guide
Step 2
Getting Started1
2 Fasten the Rack Mount Extenders to the oscilloscope chassis
with the four (4) M3 x 6 mm screws, using a Torx T10 driver
as follows:
NOTE
The sets of holes in the Rack Mount Extenders are slightly offset. This was
done to ensure that the Rack Mount Extenders are attached to the
oscilloscope at the correct points so that the oscilloscope’s ventilation
area is not obscured. The holes in the Rack Mount Extenders will align
with the correct holes in the oscilloscope and the screws will go in easily.
Do not force the screws into the wrong holes.
a Attach a Rack Mount Extender to the left side of the
oscilloscope using two (2) of the M3 x 6 mm screws in the
inner set of holes on the Rack Mount Extender.
Use inner
holes in
extender
Use outer
holes in
extender
b Attach the other Rack Mount Extender to the right side of
the oscilloscope using two (2) of the M3 x 6 mm screws in
the outer set of holes on the rack mount extender.
3 Place the instrument in the rack. Install the four (4)
10-32 x 0.625 Dress Screws in the chassis front ears to secure
the front of the instrument to the rack. Use the Phillips
screwdriver.
4 Align the ears in the Rear Mount Extenders with the correct
set of holes in the rear of the rack and secure the Rack Mount
6000 Series Oscilloscope User’s Guide31
1Getting Started
Ventilation requirements
Extenders to the rack using the four (4) remaining
10-32 x 0.375 Rail Screws. Use the Torx T20 driver.
5 Securely attach the Rear Extender Supports to the Front
Extender Supports by tightening the four (4) 10-32 x 0.375
Rail Screws screws that you loosely attached in step 1.
The air intake and exhaust areas must be free from
obstructions. Unrestricted air flow is required for proper
cooling.
6000A Ventilation Requirements
The fan draws air in from underneath the oscilloscope and
pushes it out behind the oscilloscope. Always ensure that the
air intake and exhaust areas are free from obstructions.
When using the oscilloscope in a bench-top setting, provide at
least 4" (100 mm) clearance behind and above the oscilloscope
for proper cooling.
6000L Ventilation Requirements
The fan draws air from the left and pushes it to the right.
Ensure that air flow is not obstructed.
326000 Series Oscilloscope User’s Guide
To power-on the oscilloscope
Digital
Cursors
AutoScale
AC-Powered 6000 Series
1 Connect the power cord to the rear of the oscilloscope, then
to a suitable AC voltage source.
The oscilloscope automatically adjusts for input line voltages
in the range 100 to 240 VAC. Ensure that you have the correct
line cord. See Table 8 on page 37. The line cord provided is
matched to the country of origin.
Getting Started1
WARNING
Always use a grounded power cord. Do not defeat the power cord
ground.
2 Press the power switch.
The power switch is located on the lower left corner of the
front panel. Some front panel lights will come on and the
oscilloscope will be operational in a few seconds.
Battery-Powered 6000A Series
54684A
Oscilloscope
1 GHz
4 GSa/s
Select
D15
Thru
D0
Position
Main
Delayed
Quick
Acquire
Meas
Quick
Save
Print
Recall
Math
1
2
AC
AC
BW
BW
50
50
1234
XYZ
1 M
300 V RMS
50
Run
Single
Stop
Mode
Edge
Coupling
Display
Utility
Analog
Label
AC
BW
50
~
14pF
~
CAT
I
5 V RMS
More
Pulse
Pattern
Width
3
4
AC
BW
50
54684e82
Figure 4Additional indicators on battery-powered 6000A Series
6000 Series Oscilloscope User’s Guide33
1Getting Started
The 6000A Series Option BAT oscilloscopes are battery
powered. They have additional LED indicators on the front
panel:
Caution indicator. Lights (amber) when running
on the internal battery. See “Operating with the
Internal Battery” below.
Battery power indicator.Turns from green to red
when there is 15 to 20 minutes of battery power
remaining.
Charging indicator. Lights when the battery is
charging. Turns off when the battery is fully
charged or when the charger is disconnected.
Battery life is approximately 1.75 hours, depending on the
oscilloscope configuration.
WARNING
346000 Series Oscilloscope User’s Guide
Operating with the Internal Battery
When operating with the internal battery, the operating
temperature should be in the range from -10°C to 50°C, ≤ 80%
relative humidity.
1 Make sure battery is charged before use. See “Charging the
Battery” below.
2 Connect the supplied ground wire from the ground post on
the back of the instrument to earth ground.
Maintain oscilloscope ground connection
Do not negate the protective action of the ground connection to the
oscilloscope. If the circuit under test has voltages greater than 30
Vrms, the oscilloscope must be grounded through its ground
terminal to prevent electric shock to the user.
Ground Post
Getting Started1
Figure 5Ground post on rear panel
3 Press the power switch.
The power switch is located on the lower left corner of the
front panel. Some front panel lights will come on and the
oscilloscope will be operational in a few seconds.
6000 Series Oscilloscope User’s Guide35
1Getting Started
Charging the Battery
When charging the battery, the operating temperature should be
in the range from 0°C to 45°C, ≤ 80% relative humidity.
1 Connect the supplied AC adapter to the back of the
oscilloscope, and connect the adapter’s power cord to a
suitable AC voltage source.
CAUTION
NOTE
Use only the supplied AC adapter. Damage to the equipment could
result if an improper AC adapter is used.
You can use the oscilloscope while the battery is being charged.
If the battery charging indicator does not light
If the batteries are discharged enough, there may not be enough charge for
internal circuitry to cause the battery charging indicator LED to light. In
this case, it can take up to about 20 minutes of charging before the
indicator will light.
Operating with the Automotive Power Adapter Cable
The N5429A 12V DC automotive adapter cable is for charging
and operating 6000A Series Option BAT battery-powered
oscilloscopes.
1 Connect the N5429A adapter cable to the back of the
oscilloscope, and connect the other end of the cable to a 12V
DC automotive power source.
Replacing the Battery
The oscilloscope battery is not user replaceable. You must
return the oscilloscope to Agilent for battery replacement.
Contact Agilent for further instructions (see
www.agilent.com/find/contactus).
366000 Series Oscilloscope User’s Guide
Getting Started1
Tab l e 8Power Cords
Plug TypeCable Part NumberPlug TypeCable Part Number
You can communicate with all InfiniiVision 6000 Series
oscilloscopes via LAN, USB, or GPIB. 6000A Series
oscilloscopes can also be controlled using the front panel keys.
The 6000L Series oscilloscopes do not have a keyboard or
display, so communication must be established via LAN, USB,
or GPIB.
The Automation Ready CD-ROM provided with the oscilloscope
contains connectivity software to enable communication over
these interfaces. Refer to the instructions provided on the
CD-ROM to install this software on your PC.
Remote commands can be issued via LAN, USB, or GPIB. These
commands are generally used when the oscilloscope is under
program control for automated test and data acquisition.
Information about controlling the oscilloscope through remote
commands is contained in the Programmer’s Quick Start Guide, which is included in the documentation CD-ROM
supplied with this oscilloscope. You can also access this
document online: direct your web browser to
www.agilent.com/find/mso6000 and select Technical Support,
then select Manuals.
All 6000 Series oscilloscopes feature a built-in Web server
(requires software version 4.0 or greater; see page 334 for
software updating instructions). Using the Web browser you can
set up measurements, monitor waveforms, capture screen
images and operate the oscilloscope remotely.
Detailed Connectivity Information
For detailed connectivity information, refer to the Agilent Technologies
USB/LAN/GPIB Interfaces Connectivity Guide. For a printable electronic copy of
the Connectivity Guide, direct your Web browser to www.agilent.com and search
for Connectivity Guide.
386000 Series Oscilloscope User’s Guide
To establish a LAN connection (6000A Series)
1 If the controller PC isn’t already connected to the local area
network (LAN), do that first.
2 Get the oscilloscope’s network parameters (hostname,
domain, IP address, subnet mask, gateway IP, DNS IP, etc.)
from your network administrator.
3 Connect the oscilloscope to the local area network (LAN) by
inserting the LAN cable into the “LAN” port on the rear
panel of the oscilloscope.
4 On the oscilloscope, ensure the controller interface is
enabled:
a Press the Utility key.
b Using the softkeys, press I/O and Control.
c Use the Entry knob to select “LAN”; then, press the Control
softkey again.
5 Configure the oscilloscope’s LAN interface:
a Press the Configure softkey until “LAN” is selected.
b Press the LAN Settings softkey.
c Press the Addresses softkey. Use the IP Options softkey and
the Entry knob to select DHCP, AutoIP, or netBIOS. Use
the Modify softkey (and the other softkeys and the Entry
knob) to enter the IP Address, Subnet Mask, Gateway IP,
and DNS IP values. When you are done, press the return
(up arrow) softkey.
d Press the Domain softkey. Use the Modify softkey (and the
other softkeys and the Entry knob) to enter the Host name
and the Domain name. When you are done, press the
return (up arrow) softkey.
Getting Started1
NOTE
6000 Series Oscilloscope User’s Guide39
When you connect the oscilloscope to a LAN it is a good practice to limit
access to the oscilloscope by setting a password. By default, the
oscilloscope is not password protected. See page 46 to set a password.
1Getting Started
To establish a LAN connection (6000L Series)
1 Connect a cable from your site Local Area Network (LAN) to
the LAN port on the rear panel of the oscilloscope.
2 Switch on the oscilloscope power. When the LAN indicator
illuminates green, the LAN is connected as configured. This
may take a few minutes. The oscilloscope will attempt to
connect to the LAN using DHCP, AutoIP, then Manual, in that
order (when enabled). These connection methods can be
enabled or disabled later using the Configure Network tab in
the web interface.
3 Open a web browser and enter the oscilloscope’s hostname in
the address field. The oscilloscope’s Web page will be
displayed.
Hostname
If you do not know the oscilloscope’s hostname you can reset
the hostname as follows:
1 Press and release the Reset button (see page 62, 63) on the
front panel of the 6000L Series oscilloscope. This will reset
the hostname to that which is printed on the label near the
oscilloscope’s power cord connection. The following LAN
parameters will be reset:
• Hostname
• IP address
• Domain name
• VISA address
• Password (see page 46.)
NOTE
406000 Series Oscilloscope User’s Guide
When you connect the oscilloscope to a LAN it is a good practice to limit
access to the oscilloscope by setting a password. By default, the
oscilloscope is not password protected. See page 46 to set a password.
Getting Started1
NOTE
Any time you modify the oscilloscope’s hostname it will break the
connection between the oscilloscope and the LAN. You will need to
re-establish communication to the oscilloscope using the new hostname.
For more information about connecting to the oscilloscope,
refer to the Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide. For a printable electronic copy of the
Connectivity Guide, direct your Web browser to
www.agilent.com and search for Connectivity Guide.
6000 Series Oscilloscope User’s Guide41
1Getting Started
To establish a point-to-point LAN connection
Stand-alone connection to a PC
The following procedure describes how to establish a
point-to-point (stand alone) connection to the oscilloscope. This
is useful if you want to control the oscilloscope using a laptop
computer or a stand-alone computer.
1 Install Agilent I/O Libraries Suite from the CD that was
supplied with the oscilloscope. If you do not have the CD you
can download the I/O Libraries Suite from
www.agilent.com/find/iolib.
2 Connect your PC to the oscilloscope using a cross-over LAN
cable. (The 5061-0701 LAN crossover cable is supplied with
6000L Series oscilloscopes.)
3 Switch on the oscilloscope power. Wait until the LAN
connection is configured:
• On 6000A Series oscilloscopes, press Utility & I/O and wait
until the LAN status shows “configured”.
• On 6000L Series oscilloscopes, the LAN indicator
illuminates green when the LAN connection is configured.
This may take a few minutes.
4 Start the Agilent Connection Expert application from the
Agilent I/O Libraries Suite program group.
5 When the Agilent Connection Expert application is
displayed, select Refresh All.
6 Right Click LAN and select Add Instrument.
7 In the Add Instrument window, the LAN line should be
highlighted; select OK.
8 In the LAN Instrument window, select Find Instruments…
9 In the Search for instruments on the LAN window, LAN and
Look up hostnames should be checked.
10 Select the Find Now key. (NOTE: It may take up to three
minutes before the instrument is found. If the instrument is
not found the first time, wait about one minute and try
again.)
426000 Series Oscilloscope User’s Guide
To use the Web interface
Getting Started1
11 When the instrument is found, select OK and OK to close the
Add Instrument windows.
Now the instrument is connected and the instrument’s Web
interface may be used.
All InfiniiVision 6000 Series oscilloscopes include a built-in Web
server (requires software version 4.0 and above). See page 334
for information about updating your oscilloscope’s software.
When you connect to the oscilloscope using a computer and web
browser, you can:
• Control the oscilloscope using the Remote Front Panel
function.
• Activate the Identify function (see page 49) to identify a
particular instrument by causing it’s front panel light to
blink.
• View information about the oscilloscope like its model
number, serial number, host name, IP address, and VISA
(address) connect string.
• View oscilloscope firmware version information and upload
new firmware into the oscilloscope.
• View and modify the oscilloscope’s network configuration
and status information.
6000 Series Oscilloscope User’s Guide43
1Getting Started
Controlling the oscilloscope using a Web browser
A built-in Web server allows communication and control via a
Java™-enabled Web browser. Measurements can be set up,
waveforms can be monitored, screen images can be captured,
and the oscilloscope can be operated remotely. Also, SCPI
(Standard Commands for Programmable Instrumentation)
commands can be sent over the LAN.
Microsoft Internet Explorer 6 is the recommended Web browser
for communication and control of the oscilloscope. Other Web
browsers may work but are not guaranteed to work with the
oscilloscope. The Web browser must be Java-enabled with Sun
Microsystems™ Java Plug-in.
Operating the oscilloscope using a Web browser
1 Connect the oscilloscope to your LAN (see page 39 or 40), or
establish a point-to-point connection (see page 42). It is
446000 Series Oscilloscope User’s Guide
Getting Started1
possible to use a point-to-point connection (see page 42), but
using a LAN is the preferred method.
2 Type the oscilloscope’s hostname or IP address in the web
browser.
3 When the oscilloscope’s Web page is displayed, select
Browser Web Control, then select Remote Front Panel. After a few
seconds the Remote Front Panel appears.
NOTE
If Java is not installed on your PC, you will be prompted to install the
Sun Microsystems Java Plug-in. This plug-in must be installed on the
controlling PC for Remote Front Panel operation.
4 Use the Main Menu and the Function Keys to control the
oscilloscope. This is a manual way to control an oscilloscope
which is normally controlled by a remote program.
6000 Series Oscilloscope User’s Guide45
1Getting Started
Setting a password
Step 1
Whenever you connect the oscilloscope to a LAN, it is good
practice to set a password to prevent unauthorized access to the
oscilloscope via Web browser.
1 Select the Configure Network tab from the instrument’s
Wel come page.
Step 2
466000 Series Oscilloscope User’s Guide
2 Select the Modify Configuration button.
Getting Started1
Step 3
3 Enter your desired password, and click Apply Changes.
4 To reset the password:
• 6000A Series oscilloscopes - Press Utility & I/O & LAN Reset.
• 6000L Series oscilloscopes - Press the
LAN RESET button on
the front panel.
6000 Series Oscilloscope User’s Guide47
1Getting Started
Main Menu
Function Keys
Hint appears
when you
roll over with
mouse
Softkeys
(Left-click to
select, Rightclick for
Quick Help
486000 Series Oscilloscope User’s Guide
Scrolling and Monitor Resolution
When using a monitor resolution of 1024 x 768 or less on the
remote computer, you need to scroll to access the full remote
front panel. To display the remote front panel without scroll
bars, use a monitor resolution greater than 1024 x 768 on your
computer’s display.
Identify Function
Select Identification on button on the oscilloscope’s web page.
On a 6000L Series oscilloscope, the LAN indicator on the front
of the instrument will flash green to identify the oscilloscope
until you select Identification off. On a 6000A Series
oscilloscope, an “Identify” message is displayed; you can either
select Identification off or press the OK Softkey on the
oscilloscope to continue. This feature is useful when trying to
locate a specific instrument in a rack of equipment.
Getting Started1
Identification Option
6000 Series Oscilloscope User’s Guide49
1Getting Started
Printing the oscilloscope’s display from a web browser
To print the oscilloscope’s display from a web browser:
1 Establish a connection to the oscilloscope as described in
this chapter and go to the oscilloscope’s Welcome page.
2 Select the Get Image tab from the left side of the Welcome
screen. After a delay of several seconds, the oscilloscope’s
screen image will be displayed.
3 Right-click on the image and select “Save Picture As...”.
4 Select a storage location for the image file and click Save.
For more information about connecting the oscilloscope to a
LAN see the Agilent Technologies USB/LAN/GPIB Interfaces Connectivity Guide. For a printable electronic copy of the
Connectivity Guide, direct your Web browser to
www.agilent.com and search for Connectivity Guide.
506000 Series Oscilloscope User’s Guide
To connect the oscilloscope probes
The analog input impedance of the 100 MHz oscilloscopes is
fixed at 1 MΩ. The 1 MΩ mode is for use with many passive
probes and for general purpose measurements. The high
impedance minimizes the loading effect of the oscilloscope on
the circuit under test. If a 50 Ω input impedance is required,
attach a 50 ohm feedthrough termination adapter such as
Agilent part number 0960-0301 to the oscilloscope’s channel
input BNC connector.
The analog input impedance of the 300 MHz, 500 MHz, and
1 GHz oscilloscopes can be set to either 50 Ω or 1 MΩ . The 50 Ω
mode matches 50 Ω cables and some active probes commonly
used in making high frequency measurements. This impedance
matching gives you the most accurate measurements since
reflections are minimized along the signal path.
1 Connect the supplied oscilloscope probe to an oscilloscope
channel BNC connector on the front panel of the
oscilloscope.
2 Connect the retractable hook tip on the probe tip to the
circuit point of interest. Be sure to connect the probe ground
lead to a ground point on the circuit.
Getting Started1
CAUTION
Maximum input voltage in 50 Ω mode
Do not exceed 5 Vrms at the BNC in 50 Ω mode on the Agilent 6000
Series oscilloscopes. Input protection is enabled in 50 Ω mode and the
50 Ω load will disconnect if greater than 5 Vrms is detected. However
the inputs could still be damaged, depending on the time constant of
the signal. The 50 Ω input protection mode on the Agilent 6000 Series
oscilloscopes only functions when the oscilloscope is powered on.
CAUTION
The probe ground lead is connected to the oscilloscope chassis and
the ground wire in the power cord. If you need to measure between
two live points, use a differential probe. Defeating the ground
connection and “floating” the oscilloscope chassis will probably result
in inaccurate measurements.
6000 Series Oscilloscope User’s Guide51
1Getting Started
WARNING
CAUTION
Do not negate the protective action of the ground connection to the
oscilloscope. The oscilloscope must remain grounded through its
power cord. Defeating the ground creates an electric shock hazard.
Maximum input voltage for analog inputs
CAT I 300 Vrms, 400 Vpk; transient overvoltage 1.6 kVpk
CAT II 100 Vrms, 400 Vpk
with 10073C or 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
To verify basic oscilloscope operation
If you have a 6000L Series oscilloscope, you will need to start a
Web control session, as described on page 44.
1 Press the Save/Recall key on the front panel, then press the
Default Setup softkey. (The softkeys are located directly below
the display on the front panel.) The oscilloscope is now
configured to its default settings.
2 Connect an oscilloscope probe from channel 1 to the Probe
Comp signal terminal on the front panel.
3 Connect the probe’s ground lead to the ground terminal that
is next to the Probe Comp terminal.
4 Press AutoScale.
526000 Series Oscilloscope User’s Guide
Getting Started1
5 You should see a waveform on the oscilloscope’s display
similar to this:
If you see the waveform, but the square wave is not shaped
correctly as shown above, perform the procedure “To
compensate the oscilloscope probes” on page 54.
If you do not see the waveform, ensure your power source is
adequate, the oscilloscope is properly powered-on, and the
probe is connected securely to the front-panel oscilloscope
channel input BNC and to the Probe Comp terminal.
6000 Series Oscilloscope User’s Guide53
1Getting Started
To compensate the oscilloscope probes
You should compensate your oscilloscope probes to match their
characteristics to the oscilloscope’s channels. A poorly
compensated probe can introduce measurement errors.
1 Perform the procedure “To verify basic oscilloscope
operation” on page 52.
2 Use a nonmetallic tool to adjust the trimmer capacitor on the
probe for the flattest pulse possible. The trimmer capacitor is
located on the probe BNC connector.
Perfectly compensated
Over compensated
Under compensated
3 Connect probes to all other oscilloscope channels (channel 2
of a 2-channel oscilloscope, or channels 2, 3, and 4 of a
4-channel oscilloscope). Repeat the procedure for each
channel. This matches each probe to each channel.
comp.cdr
The process of compensating the probes serves as a basic test to
verify that the oscilloscope is functional.
546000 Series Oscilloscope User’s Guide
To calibrate the probes
Getting Started1
The oscilloscope can accurately calibrate its analog oscilloscope
channels to certain active probes, such as InfiniiMax probes.
Other probes, such as the 10073C and 10074C passive probes,
do not require calibration. The Calibrate Probe softkey will be
grayed-out (displayed in faint text) when a connected probe
does not require calibration.
When you connect a probe that can be calibrated (such as an
InfiniiMax probe), the Calibrate Probe softkey in the channel’s
menu will become active. Connect the probe to the Probe Comp
terminal, and the probe ground to the Probe Comp ground
terminal. Press the Calibrate Probe softkey and follow the
instructions on the display.
NOTE
Passive Probes Supported
When calibrating a differential probe, connect the positive lead to the
Probe Comp terminal and the negative lead to the Probe Comp ground
terminal. You may need to connect an alligator clip to the ground lug to
allow a differential probe to span between the Probe Comp test point and
ground. A good ground connection ensures the most accurate probe
calibration.
The following passive probes can be used with the 6000 Series
oscilloscopes. Any combination of passive probes can be used.
Tab l e 9Passive Probes
Passive ProbesQuantity Supported
10070C4
10073C4
10074C4
10076A4
6000 Series Oscilloscope User’s Guide55
1Getting Started
Active Probes Supported
By 300 MHz, 500 MHz, and 1 GHz Bandwidth Models
Active probes that do not have their own external power supply
require substantial power from the AutoProbe interface. (The
AutoProbe interface is present on the 300 MHz, 500 MHz, and
1 GHz bandwidth models.) “Quantity Supported” indicates the
maximum number of each type of active probe that can be
connected to the oscilloscope. If too much current is drawn
from the AutoProbe interface, an error message will be
displayed, indicating that you must momentarily disconnect all
probes to reset the AutoProbe interface.
Tab l e 1 0Active Probes for 300 MHz, 500 MHz, and 1 GHz bandwidth
models
Active ProbesQuantity Supported
1130A2
1131A2
1132A2
1134A2
1141A with 1142A power supply4
1144A with 1142A power supply4
1145A with 1142A power supply2
1147A2
1156A4
1157A4
1158A4
N2772A with N2773A power supply4
N2774A with N2775A power supply4
566000 Series Oscilloscope User’s Guide
By 100 MHz Bandwidth Models
The following active probes use their own power supply.
Therefore, they can be used on all 6000 Series oscilloscopes,
including the 100 MHz bandwidth models.
Many active probes have a 50 Ω output impedance. The input
impedance of 6000 Series 100 MHz bandwidth models is fixed at
1MΩ. When connecting these probes to 6000 Series 100 MHz
bandwidth models, a 50 Ω feedthrough terminator (such as
Agilent part number 0960-0301 is required).
Tab l e 1 1Active Probes for All 6000 Series Oscilloscopes
Active ProbesQuantity Supported
1141A with 1142A power supply4
1144A with 1142A power supply4
1145A with 1142A power supply2
N2772A with N2773A power supply4
Getting Started1
N2774A with N2775A power supply4
6000 Series Oscilloscope User’s Guide57
1Getting Started
Using Quick Help
To view Quick Help on 6000L Series oscilloscopes:
Start a Web browser control session as described on page 44
and select Remote Front Panel. To view Quick Help information,
right-click on the softkey. Help is not available for front panel
keys; only softkeys.
To view Quick Help on 6000A Series oscilloscopes
1 Press and hold down the key or softkey for which you would
like to view help.
Quick Help Message
Press and Hold 6000A Front Panel Key or Softkey
or Right-Click Softkey when using Web browser control
You can set Quick Help to close when you release the key (this is
the default mode) or to remain on the screen until another key
is pressed or a knob is turned. To select this mode, press the
Utility key, then press the Language softkey, then press the Help
Remain/Help Close softkey.
586000 Series Oscilloscope User’s Guide
Quick Help Languages
At the time this manual was published, Quick Help was
available in English, Simplified Chinese, Japanese, German,
French, and Russian.
To choose a Quick Help language in the oscilloscope:
1 Press Utility, then press the Language softkey.
2 Repeatedly press and release the Language softkey or rotate
the Entry knob until the desired language is selected.
If Quick Help updates become available, you can download the
updated Quick Help language file and load it into the
oscilloscope.
To download the InfiniiVision 6000 Series Oscilloscope Quick
Help Language Support file:
1 Direct your web browser to www.agilent.com/find/mso6000.
2 On the resulting page, select Technical Support, then select
Software Downloads & Utilities.
Getting Started1
6000 Series Oscilloscope User’s Guide59
1Getting Started
606000 Series Oscilloscope User’s Guide
Agilent 6000 Series Oscilloscope
User’s Guide
2
Front-Panel Controls
6000L Series Oscilloscope Controls 62
6000A Series Oscilloscope Front-Panel Controls 66
6000A Series Front-Panel Operation 76
Agilent Technologies
61
2Front-Panel Controls
6000L Series Oscilloscope Controls
The 6000L Series oscilloscope is designed to be remotely
controlled. Therefore the front panel layout is simple.
Front Panel
1
Power
Switch
10
External
Trigger
Input
2
Power
Indicator
11
Trigger
Output
3
LAN
Status
Indicator
12
Calibration
Protect
Switch
4
Reset
Switch
13
10 MHz
Reference
Output
5
Intensity
Control
Rear Panel
XGA Video
Output
14
6
USB
Port
Channels
Terminals
15
Digital
Input
7
Probe
Comp
16
USB Host
Port
8
AutoProbe
Interface
17
USB
Device
Port
9
Channel
Inputs
18
LAN
Connector
19
GPIB
Connector
626000 Series Oscilloscope User’s Guide
Front and Rear Panel Controls and Connectors
1. Power SwitchPress once to switch power on; press again to
switch power off. See page 33.
2. Power IndicatorIlluminates green when power is on.
3. LAN Status IndicatorThis indicator illuminates green when a
LAN connection has been detected and is connected as
configured. The LAN status indicator illuminates red when the
following LAN faults occur:
• failure to acquire a valid IP address
• detection of a duplicate IP address
• failure to renew an already acquired DHCP lease
The LAN status indicator flashes green when the identify
function is activated (see page 49).
4. Reset SwitchWhen power is on, press and release this
recessed pushbutton to default LAN parameters. See page 40.
Front-Panel Controls2
5. Intensity ControlRotate clockwise to increase analog
waveform intensity; counterclockwise to decrease. You can vary
the intensity control to bring out signal detail, much like an
analog oscilloscope. Digital channel waveform intensity is not
adjustable. More details about using the Intensity control to
view signal detail are on page 263.
6. USB Host PortConnect a USB compliant mass storage device
to store or recall oscilloscope setup files or waveforms. You can
also use the USB port to update the oscilloscope’s system
software or Quick Help language files if updates are available.
You do not need to take special precautions before removing the
USB mass storage device from the oscilloscope (you do not need
to “eject” it). Simply unplug the USB mass storage device from
the oscilloscope when the file operation is complete. More
information on using the USB port is given in Chapter 7, “Saving
and Printing Data,” starting on page 315.
6000 Series Oscilloscope User’s Guide63
2Front-Panel Controls
CAUTION
Only connect USB devices to the USB host port. Do not attempt to
connect a host computer to this port to control the oscilloscope. Use
the USB device port if you want to connect a host (see the InfiniiVision 6000 Series Oscilloscope Programmer’s Quick Start Guide for details).
7. Probe Compensation TerminalsUse the signal at these
terminals to match each probe’s characteristics to the
oscilloscope channel to which it is connected. See page 54.
8. AutoProbe Interface(Not available on 100 MHz bandwidth
models.) When you connect a probe to the oscilloscope, the
AutoProbe Interface attempts to determine the type of probe
and set its parameters in the Probe menu accordingly. See
page 81. Note: Although the 100 MHz models lack the AutoProbe
interface, they do have a probe sense ring around the BNC.
Therefore, the probe attenuation factor will be automatically set
when you connect a compatible probe such as the 10073C or
10074C.
9. Channel Input BNC ConnectorAttach the oscilloscope probe
or BNC cable to the BNC connector. This is the channel’s input
connector.
10. External Trigger InputAllows you to trigger the oscilloscope
using an external signal. See page 136.
11. Trigger OutputThis output is related to the oscilloscope’s
internal trigger signal. See page 206.
12. Calibration Protect SwitchSet this switch to “Protect” to
prevent unwanted re-calibration. See page 105.
13. 10 MHz Reference Input/OutputYou can input a signal,
synchronizing multiple instruments, or you can output this
instrument’s reference signal. See page 336.
14. XGA Video OutputYou can connect an external display to
the oscilloscope for viewing waveforms. See page 260.
646000 Series Oscilloscope User’s Guide
Front-Panel Controls2
15. Digital Channels InputIf you purchased the MSO upgrade
you can view and trigger on digital signals. See page 111.
16. USB Host PortConnect a USB compliant mass storage
device to store or recall oscilloscope setup files or waveforms.
You can also use the USB port to update the oscilloscope’s
system software or Quick Help language files if updates are
available. You do not need to take special precautions before
removing the USB mass storage device from the oscilloscope
(you do not need to “eject” it). Simply unplug the USB mass
storage device from the oscilloscope when the file operation is
complete. More information on using the USB port is given in
Chapter 7, “Saving and Printing Data,” starting on page 315.
CAUTION
Only connect USB devices to the USB host port. Do not attempt to
connect a host computer to this port to control the oscilloscope. Use
the USB device port if you want to connect a host (see the InfiniiVision 6000 Series Oscilloscope Programmer’s Quick Start Guide for details).
17. USB Device PortYou can use this port for remote control of
the oscilloscope by connecting to a PC host computer. See the
InfiniiVision 6000 Series Oscilloscope Programmer’s Quick
Start Guide for details.
18. LAN ConnectorYou can use this port to connect to the
oscilloscope and control it via your LAN. This standard LAN
port is not Auto-MDIX compliant. See page 39 or page 40.
19. GPIB ConnectorThis connector is for connecting the
oscilloscope to a General Purpose Interface Bus. See the
InfiniiVision 6000 Series Oscilloscope Programmer’s Quick
Start Guide for details.
6000 Series Oscilloscope User’s Guide65
2Front-Panel Controls
6000A Series Oscilloscope Front-Panel Controls
This is an introduction to the front-panel controls of the Agilent
6000A Series oscilloscope. Generally, you set up the front-panel
controls and then make a measurement.
The keys on the front panel bring up softkey menus on the
display that provide access to oscilloscope features. Many
softkeys use the Entry knobto select values.
Six softkeys are located below the display. To understand the
symbols used in the softkey menus and throughout this guide,
see “Conventions” on page 67.
Note that the Digital Select key, not the Entry knob, is used to
select digital channels when you want to reposition a digital
channel waveform.
NOTE
The simplest way to set up the oscilloscope is to connect it to the signals
of interest and press the AutoScale key.
666000 Series Oscilloscope User’s Guide
Conventions
Throughout this book, the front-panel keys and softkeys are
denoted by a change in font. For example, the Cursors key is in
the Measure section of the front panel. The Acq Mode softkey is
the left-most softkey when the Acquire menu is displayed.
In this manual, instructions for pressing a series of keys are
written in an abbreviated manner. Pressing the Utility key, then
the I/O softkey, then the Show I/O Config softkey is abbreviated
as follows:
Press Utility & I/O & Show I/O Config.
Graphic Symbols in Softkey Menus
The following graphic symbols appear in the oscilloscope’s
softkey menus. The softkey menus appear at the bottom of the
display, just above the six softkeys.
Use the Entry knob to adjust the parameter. The Entry knob
is located on the front panel. The symbol above the knob is
illuminated when this control is active.
Front-Panel Controls2
Press the softkey to display a pop up with a list of choices.
Repeatedly press the softkey until your choice is selected.
Use the Entry knob labeledand/or press the softkey to
adjust the parameter.
Press the softkey to display a pop-up menu. Press the softkey
or rotate and push the Entry knob to make an item active.
Option is selected but not active.
Option is selected and active.
Feature is on. Press the softkey again to turn the feature off.
Feature is off. Press the softkey again to turn the feature on.
Press the softkey to view the menu.
Press the softkey to return to the previous menu.
6000 Series Oscilloscope User’s Guide67
2Front-Panel Controls
4-Channel 6000A Series Oscilloscope Front Panel
The following diagram shows the front panel of the 6000A
Series 4-channel oscilloscopes. The controls of the 2-channel
oscilloscopes are very similar. For a diagram showing the
differences of the 2-channel oscilloscope, see page 74.
22
Display
1
Power
switch
Measure
2
Intensity
Control
21
Keys
AutoScale
USB
Port
Sweep Speed
23
Entry
Knob
24
Key
25
Softkeys
3
20
Horizontal
Control
4
Probe
Compensation
Terminals
19
Horizontal
Menu/Zoom
Key
5
Digital
Channel
Controls
Horizontal
Position
Control
6
Vertical
Position
Control
18
7
Channel
On/Off
Key
17
Waveform
Keys
8
Math
Key
16
Run
Controls
15
Trigger
Controls
14
Utility
Key
13
File Keys
12
Label Key
11
Channel
Input BNC
10
AutoProbe
Interface
9
Vertical
Sensitivity
Control
Figure 66000A Series 4-Channel Oscilloscope Front Panel
686000 Series Oscilloscope User’s Guide
Front Panel Controls
1. Power SwitchPress once to switch power on; press again to
switch power off. See page 33.
2. Intensity ControlRotate clockwise to increase analog
waveform intensity; counterclockwise to decrease. You can vary
the intensity control to bring out signal detail, much like an
analog oscilloscope. Digital channel waveform intensity is not
adjustable. More details about using the Intensity control to
view signal detail are on page 263.
3. USB Host PortConnect a USB compliant mass storage device
to store or recall oscilloscope setup files or waveforms. You can
also use the USB port to update the oscilloscope’s system
software or Quick Help language files if updates are available.
You do not need to take special precautions before removing the
USB mass storage device from the oscilloscope (you do not need
to “eject” it). Simply unplug the USB mass storage device from
the oscilloscope when the file operation is complete. More
information on using the USB port is given in Chapter 7, “Saving
and Printing Data,” starting on page 315.
Front-Panel Controls2
CAUTION
Only connect USB devices to the USB host port. Do not attempt to
connect a host computer to this port to control the oscilloscope. Use
the USB device port if you want to connect a host (See the InfiniiVision 6000 Series Oscilloscope Programmer’s Quick Start Guide for details).
4. Probe Compensation TerminalsUse the signal at these
terminals to match each probe’s characteristics to the
oscilloscope channel to which it is connected. See page 54.
5. Digital Channel ControlsThese controls switch the digital
channels on/off, and can be used to select a digital channel for
repositioning on the display. You can press the D15 Thru D0 key
to show the Digital Channel Menu above the softkeys. See
Chapter 3, “Viewing and Measuring Digital Signals,” starting on
page 111.
6000 Series Oscilloscope User’s Guide69
2Front-Panel Controls
6. Vertical Position Control Use this knob to change the
channel’s vertical position on the display. There is one Vertical
Position control for each channel. See “Using the analog
channels” on page 83.
7. Channel On/Off KeyUse this key to switch the channel on or
off, or to access the channel’s menu in the softkeys. There is one
Channel On/Off key for each channel. See “Using the analog
channels” on page 83.
8. Math KeyThe Math key provides access to FFT (Fast Fourier
Transform), multiply, subtract, differentiate, and integrate
functions. See “Math Functions” on page 213.
9. Vertical Sensitivity Use this knob to change the vertical
sensitivity (gain) of the channel. See “Using the analog
channels” on page 83.
10. AutoProbe Interface(Not available on 100 MHz bandwidth
models.) When you connect a probe to the oscilloscope, the
AutoProbe Interface attempts to determine the type of probe
and set its parameters in the Probe menu accordingly. See
page 81. Note: Although the 100 MHz models lack the AutoProbe
interface, they do have a probe sense ring around the BNC.
Therefore, the probe attenuation factor will be automatically set
when you connect a compatible probe such as the 10073C or
10074C.
11. Channel Input BNC ConnectorAttach the oscilloscope probe
or BNC cable to the BNC connector. This is the channel’s input
connector.
12. Label KeyPress this key to access the Label menu, which
lets you enter labels to identify each trace on the oscilloscope
display. See page 97.
13. File KeysPress the File key to access file functions such as
save or recall a waveform or setup. Or press the Print key
(labeled Quick Print on older models) to print the waveform
from the display. See “Saving and Printing Data” on page 315.
706000 Series Oscilloscope User’s Guide
Front-Panel Controls2
14. Utility KeyPress this key to access the Utility menu, which
lets you configure the oscilloscope’s I/O settings, printer
configuration, file explorer, service menu, and other options.
15. Trigger ControlsThese controls determine how the
oscilloscope triggers to capture data. See “Choosing Auto trigger
mode or Normal trigger mode” on page 80 and Chapter 4,
“Triggering the Oscilloscope,” starting on page 127.
16. Run ControlsPress Run/Stop to make the oscilloscope begin
looking for a trigger. The Run/Stop key will illuminate in green.
If the trigger mode is set to “Normal,” the display will not
update until a trigger is found. If the trigger mode is set to
“Auto,” the oscilloscope looks for a trigger, and if none is found,
it will automatically trigger, and the display will immediately
show the input signals. In this case, the Auto indicator at the top
of the display will change to Auto? and the background will
flash, indicating that the oscilloscope is forcing triggers.
Press Run/Stop again to stop acquiring data. The key will
illuminate in red. Now you can pan across and zoom-in on the
acquired data.
Press Single to make a single acquisition of data. The key will
illuminate in yellow until the oscilloscope triggers. See “To start
and stop an acquisition” on page 77.
17. Waveform KeysThe Acquire key lets you set the
oscilloscope to acquire in Normal, Peak Detect, Averaging, or
High Resolution modes (see “Acquisition Modes” on page 265),
and lets you turn Realtime sampling off or on (see page 269).
The Display key lets you access the menu where you can select
infinite persistence (see page 261), switch vectors on or off (see
page 262), and adjust the display grid (graticule) intensity (see
page 262).
18. Horizontal Position ControlTurn the knob marked to pan
through the waveform data horizontally. You can see the
captured waveform before the trigger (turn the knob clockwise)
or after the trigger (turn the knob counterclockwise). If you pan
through the waveform when the oscilloscope is stopped (not in
6000 Series Oscilloscope User’s Guide71
2Front-Panel Controls
Run mode) then you are looking at the waveform data from the
last acquisition taken. See “To set up the Horizontal time
base” on page 88.
19. Horizontal Menu/Zoom Key(Labeled Main/Delayed on older
models.) Press this key to access the menu where you can split
the oscilloscope display into Main and Zoom sections, and
where you can select XY and Roll modes. You can also select
horizontal time/division vernier and select the trigger time
reference point on this menu. See “To set up the Horizontal time
base” on page 88.
20. Horizontal Sweep Speed ControlTurn the knob in the
Horizontal section that is marked to adjust the sweep
speed. The symbols under the knob indicate that this control
has the effect of spreading out or zooming in on the waveform
using the horizontal scale. The horizontal sweep speed control
changes the time per horizontal division on the display. See “To
set up the Horizontal time base” on page 88.
21. Measure KeysPress the Cursors key to switch on cursors
that you can use for making measurements. Press the Quick
Meas key to access a set of predefined measurements. See
Chapter 5, “Making Measurements,” starting on page 207.
22. DisplayThe display shows captured waveforms using a
different color for each channel. For more information about
display modes see Chapter 6, “Displaying Data,” starting on
page 257. Signal detail is displayed using 256 levels of intensity.
For more information about viewing signal detail see “Varying
the intensity to view signal detail” on page 263.
23. Entry KnobThe entry knob is used to select items from
menus and to change values. Its function changes based upon
which menu is displayed. Note that the curved arrow
symbol above the entry knob illuminates whenever the
entry knob can be used to select a value. Use the entry knob to
select among the choices that are shown on the softkeys.
726000 Series Oscilloscope User’s Guide
Front-Panel Controls2
24. AutoScale KeyWhen you press the AutoScale key the
oscilloscope will quickly determine which channels have
activity, and it will turn these channels on and scale them to
display the input signals. See “How AutoScale Works” on
page 312
25. SoftkeysThe functions of these keys change based upon
the menus shown on the display directly above the keys.
6000 Series Oscilloscope User’s Guide73
2Front-Panel Controls
2-Channel 6000A Series Oscilloscope Front Panel (differences only)
External
Trigger
Input
Figure 76000A Series 2-Channel Oscilloscope Front Panel
The differences between the front panel of the 4-channel
oscilloscopes and the 2-channel oscilloscopes are:
• The 2-channel oscilloscope has two sets of channel controls
• The 2-channel oscilloscope’s external trigger input is on the
front panel instead of the rear panel. Some trigger features
differ. See “The External Trigger input” on page 136.
746000 Series Oscilloscope User’s Guide
Interpreting the display
The oscilloscope display contains acquired waveforms, setup
information, measurement results, and softkeys for setting up
parameters.
Front-Panel Controls2
Analog
channel
sensitivity
Status line
Analog
channels and
ground levels
Trigger Level
Digital
channels
Measurement
line
Softkeys
Trigger point,
time reference
Delay
time
Sweep
speed
Run/Stop
mode
Trigger
type
Trigger
source
Trigger level
or digital
threshold
Cursor
markers
defining
measurement
Figure 8Interpreting the display
Status lineThe top line of the display contains vertical,
horizontal, and trigger setup information.
Display areaThe display area contains the waveform
acquisitions, channel identifiers, and analog trigger and ground
level indicators. Each analog channel’s information appears in a
different color.
Measurement lineThis line normally contains automatic
measurement and cursor results, but can also display advanced
trigger setup data and menu information.
SoftkeysThe softkeys let you set up additional parameters for
the selected mode or menu.
6000 Series Oscilloscope User’s Guide75
2Front-Panel Controls
6000A Series Front-Panel Operation
This section provides a brief overview of operating the
front-panel controls. Detailed oscilloscope operating
instructions are provided in later chapters.
Digital Channels
Because all of the oscilloscopes in the 6000 Series have analog channels, the
analog channel topics in this book apply to all instruments. Whenever a topic
discusses the digital channels, that information applies only to Mixed-Signal
Oscilloscope (MSO) models or DSO models that have been upgraded to an MSO.
To adjust the waveform intensity
The Intensity control is at the lower left corner of the front
panel near the power switch.
• To decrease waveform intensity, rotate the Intensity control
counter-clockwise.
• To increase waveform intensity, rotate the Intensity control
clockwise.
The intensity control does not affect digital channel intensity.
Digital channel intensity is fixed.
To adjust the display grid (graticule) intensity
1 Press the Display key.
2 Turn the Entry knob to change the intensity of the
displayed grid. The intensity level is shown in the Grid softkey
and is adjustable from 0 to 100%.
Each major vertical division in the grid corresponds to the
vertical sensitivity shown in the status line at the top of the
display.
Each major horizontal division in the grid corresponds to the
sweep speed time shown in the status line at the top of the
display.
766000 Series Oscilloscope User’s Guide
Front-Panel Controls2
NOTE
To change the analog waveform intensity, turn the Intensity knob at the
lower-left corner of the front panel. The digital waveform intensity is not
adjustable.
To start and stop an acquisition
• When you press the Run/Stop key, it illuminates in green and
the oscilloscope is in continuous running mode.
The oscilloscope examines the input voltage at each probe,
and updates the display whenever the trigger conditions are
met. Trigger processing and screen update rate are optimized
based on the oscilloscope settings. The oscilloscope displays
multiple acquisitions of the same signal similar to the way an
analog oscilloscope displays waveforms.
• When you press the Run/Stop key again, it illuminates in red,
and the oscilloscope is stopped.
“Stop” is displayed in the trigger mode position in the status
line at the top of the display. You can pan and zoom the
stored waveform by turning the horizontal and vertical
control knobs.
• To manually run and stop 6000L models start the Remote
Front Panel (see page 44) and select Run Control from the
Main Menu or press ctrl+R for run/stop or ctrl+S for single.
When the oscilloscope is running and you press the Run/Stop
key, it will flash until the current acquisition is completed. If the
acquisition completes immediately, the Run/Stop key will not
flash.
At slower sweep speeds, you may not want to wait for the
acquisition to finish. Just press Run/Stop again. The acquisition
will stop immediately and a partial waveform will be displayed.
You can display the results of multiple acquisitions using
infinite persistence. See “Infinite persistence” on page 261.
6000 Series Oscilloscope User’s Guide77
2Front-Panel Controls
To make a single acquisition
Memory Depth/Record Length
Run/Stop versus Single
When the oscilloscope is running, the trigger processing and update rate are
optimized over the memory depth.
Single
Single acquisitions always use the maximum memory available—at least twice as
much memory as acquisitions captured in Run mode—and the oscilloscope
stores at least twice as many samples. At slow sweep speeds, the oscilloscope
operates at a higher sample rate when Single is used to capture an acquisition
due to the increased memory available. To acquire data with the longest possible
record length, press the Single key.
Running
When running, versus taking a single acquisition, the memory is divided in half.
This allows the acquisition system to acquire one record while processing the
previous acquisition, dramatically improving the number of waveforms per second
processed by the oscilloscope. While running, maximizing the rate at which
waveforms are drawn on the display provides the best picture of your input signal.
When you press the Single key it illuminates in yellow and the
oscilloscope starts the acquisition system, searching for the
trigger condition. When the trigger condition is met, the
captured waveform is displayed, the Single key is extinguished
and the Run/Stop key illuminates in red.
• Use the Single key to view single-shot events without
subsequent waveform data overwriting the display.
• To manually make a single acquisition on 6000L models start
the Remote Front Panel (see page 44) and select Run Control
from the Main Menu or press ctrl+S for single.
Use Single when you want the maximum sample rate and the
maximum memory depth for pan and zoom. (See “To pan and
zoom” on page 79)
1 Set the trigger mode to Normal (see “Trigger modes: Normal
and Auto” on page 131 for instructions).
This keeps the oscilloscope from automatically triggering
immediately.
786000 Series Oscilloscope User’s Guide
Front-Panel Controls2
2 If you are triggering on analog channel events, turn the
Trigger Level knob to the trigger threshold to a level that your
waveform will cross.
3 To begin a single acquisition, press the Single key.
When you press Single, the display is cleared, the trigger
circuitry is armed, the Single key is illuminated yellow, and
the oscilloscope will wait until a trigger condition occurs
before it displays a waveform.
When the oscilloscope triggers, the single acquisition is
displayed and the oscilloscope is stopped (the Run/Stop key is
illuminated in red).
4 To acquire another waveform, press Single again.
Auto Single
Auto-trigger will generate a trigger for you if one is not found in
the predetermined time (about 40 ms) after you press Single. If
you want to make a single acquisition and you are not
particularly interested in triggering the acquisition (for
example, if you are probing a DC level), set the trigger mode to
Auto (see page 131) and press the Single key. If a trigger
condition occurs, it will be used; if a trigger doesn’t happen, an
untriggered acquisition will be taken.
To pan and zoom
You can pan across and zoom-in on a waveform even when the
acquisition system is stopped.
1 Press the Run/Stop key to stop acquisitions (or press the
Single key and allow the oscilloscope to acquire the waveform
and stop). The Run/Stop key is illuminated red when the
oscilloscope is stopped.
2 Turn the sweep speed knob to zoom horizontally and turn
the volts/division knob to zoom vertically.
The ∇ symbol at the top of the display indicates the time
reference point where the zoom-in/zoom-out is referenced.
6000 Series Oscilloscope User’s Guide79
2Front-Panel Controls
Choosing Auto trigger mode or Normal trigger mode
3 Turn the Horizontal Position knob () to pan horizontally
and turn the channel’s vertical position knob ( ) to pan
vertically.
The stopped display may contain several triggers worth of
information, but only the last trigger acquisition is available for
pan and zoom.
For more information about Pan and Zoom see page 258.
In Auto trigger mode, the oscilloscope automatically triggers
and captures waveforms when you press Run.
If you press Run when the oscilloscope is in Normal trigger
mode, a trigger must be detected before an acquisition can
complete.
In many cases, a triggered display is not needed to check signal
levels or activity. For these applications, use Auto trigger mode
(which is the default setting). If you only want to acquire
specific events as specified by the trigger settings, use Normal
trigger mode.
You can select the trigger mode by pressing the Mode/Coupling
key, then the Mode softkey.
For more detailed discussion of Auto trigger mode and Normal
trigger mode, see “Trigger modes: Normal and Auto” on page 131.
Using AutoScale
To configure the oscilloscope quickly, press the AutoScale key to
display the connected signal(s) that are active.
To undo the effects of AutoScale, press the Undo AutoScale
softkey before pressing any other key. This is useful if you have
unintentionally pressed the AutoScale key or do not like the
settings AutoScale has selected and want to return to your
previous settings.
806000 Series Oscilloscope User’s Guide
Front-Panel Controls2
To make the oscilloscope remain in the acquisition mode you
have chosen, press the AutoScale Acq Mode softkey and select
Preserve Acquisition Mode. Otherwise, the acquisition mode will
default to Normal whenever the AutoScale key is pressed.
ExampleConnect the oscilloscope probes for channels 1 and 2 to the
Probe Comp output on the front panel of the instrument. Be
sure to connect the probe ground leads to the ground lug beside
the Probe Comp output. Set the instrument to the factory
default configuration by pressing the Save/Recall key, then the
Default Setup softkey. Then press the AutoScale key. You should
see a display similar to the one shown below.
Figure 9AutoScale of oscilloscope channels 1 and 2
For an explanation of how AutoScale works, see “How
AutoScale Works” on page 312.
To set the probe attenuation factor
Passive Probes
All 6000 Series oscilloscopes including the 100 MHz models
recognize passive probes such as the 10073C and 10074C. These
probes have a pin on their connector that connects to the ring
6000 Series Oscilloscope User’s Guide81
2Front-Panel Controls
around the oscilloscope’s BNC connector. Therefore, the
oscilloscope will automatically set the attenuation factor for
recognized Agilent passive probes.
Passive probes that do not have a pin that connects to the ring
around the BNC connector will not be recognized by the
oscilloscope, and you must set the probe attenuation factor
manually.
Active Probes
All 6000 Series oscilloscopes except the 100 MHz bandwidth
models (MSO/DSO6012A and MSO/DSO6014A/L) have an
AutoProbe interface. Most Agilent active probes are compatible
with the AutoProbe interface. The AutoProbe interface uses a
series of contacts directly below the channel’s BNC connector to
transfer information between the oscilloscope and the probe.
When you connect a compatible probe to the oscilloscope, the
AutoProbe interface determines the type of probe and sets the
oscilloscope’s parameters (units, offset, attenuation, coupling,
and impedance) accordingly.
Manually Setting the Probe Attenuation Factor
If you connect a probe that the oscilloscope does not
automatically identify, you can set the attenuation factor as
follows:
1 Press the channel key
2 Press the Probe softkey
3 Turn the Entry knobto set the attenuation factor for the
connected probe.
The attenuation factor can be set from 0.1:1 to 1000:1 in a 1-2-5
sequence for measuring voltage. The attenuation factor can be
set from 10 V/A to .001V/A to for measuring current values with
a current probe. The probe attenuation factor must be set
properly for measurements to be made correctly.
826000 Series Oscilloscope User’s Guide
Channel 1
ground level
Using the analog channels
Connect the oscilloscope probes for channels 1 and 2 to the
Probe Comp output on the front panel of the instrument.
1 Press the 1 key in the Analog section of the oscilloscope’s
front panel to display the Channel 1 menu.
Channel,
Volts/div
Front-Panel Controls2
Trigger
source
Trigger level
or threshold
Channel 2
ground level
Channel
coupling
Input
impedance
Bandwidth
limit
Channel
vernier
Invert
channel
Probe
menu
Pressing an analog channel key displays the channel’s menu
and turns the display of the channel on or off. The channel is
displayed when the key is illuminated.
On 6000L models, select Analog from the Main Menu and use
the menu.
6000 Series Oscilloscope User’s Guide83
2Front-Panel Controls
Turning channels off
You must be viewing the menu for a channel before you can turn it off. For
example, if channel 1 and channel 2 are turned on and the menu for channel 2 is
being displayed, to turn channel 1 off, press 1 to display channel 1 menu, then
press 1 again to turn channel 1 off.
Vertical sensitivityTurn the large knob above the channel key
to set the sensitivity (volts/division) for the channel. The
vertical sensitivity knob changes the analog channel sensitivity
in a 1-2-5 step sequence (with a 1:1 probe attached). The analog
channel Volts/Div value is displayed in the status line.
VernierPress the Vernier softkey to turn on vernier for the
selected channel. When Vernier is selected, you can change the
channel’s vertical sensitivity in smaller increments. The
channel sensitivity remains fully calibrated when Vernier is on.
The vernier value is displayed in the status line at the top of the
display.
When Vernier is turned off, turning the volts/division knob
changes the channel sensitivity in a 1-2-5 step sequence.
Vertical expansionThe default mode for expanding the signal
when you turn the volts/division knob is vertical expansion
about the ground level of the channel. To set the expansion
mode to expand about the center of the screen instead, press
Expand in the Utility&Options&Preferences&Expand menu and
select Center.
Ground levelThe ground level of the signal for each displayed
analog channel is identified by the position of the icon at the
far-left side of the display.
Vertical posit ionTurn the small vertical position knob ( ) to
move the channel’s waveform up or down on the display. The
voltage value momentarily displayed in the upper right portion
of the display represents the voltage difference between the
vertical center of the display and the ground level () icon. It
also represents the voltage at the vertical center of the display if
vertical expansion is set to expand about ground.
846000 Series Oscilloscope User’s Guide
Front-Panel Controls2
Measurement Hints
If the channel is DC coupled, you can quickly measure the DC component of the
signal by simply noting its distance from the ground symbol.
If the channel is AC coupled, the DC component of the signal is removed, allowing
you to use greater sensitivity to display the AC component of the signal.
2 Press the channel’s on/off key, then press the Coupling softkey
to select the input channel coupling.
Coupling changes the channel's input coupling to either AC
(alternating current) or DC (direct current). AC coupling
places a 3.5 Hz high-pass filter in series with the input
waveform that removes any DC offset voltage from the
waveform. When AC is selected, “AC” is illuminated on the
front panel next to the channel position knob ( ).
• DC coupling is useful for viewing wavefor ms as low as 0 Hz
that do not have large DC offsets.
• AC coupling is useful for viewing waveforms with large DC
offsets. When AC coupling is chosen, you cannot select
50Ω mode. This is done to prevent damage to the
oscilloscope.
Note that Channel Coupling is independent of Trigger Coupling.
To change Trigger coupling see page 133.
3 Press the Imped (impedance) softkey.
NOTE
The input impedance of the MSO/DSO6012A and MSO/DSO6014A/L
(100 MHz bandwidth) oscilloscope is fixed at 1 MΩ. Therefore, the
impedance selection is not available on these models.
When you connect an AutoProbe, self-sensing probe, or a compatible
InfiniiMax probe, the oscilloscope will automatically configure the
oscilloscope to the correct impedance.
The analog channel input impedance can be set to either
1M Ohm or 50 Ohm by pressing the Imped softkey.
6000 Series Oscilloscope User’s Guide85
2Front-Panel Controls
• 50 Ohm mode matches 50-ohm cables commonly used in
making high frequency measurements, and 50-ohm active
probes. This impedance matching gives you the most
accurate measurements since reflections are minimized
along the signal path. When 50 Ohm is selected, “50Ω” is
illuminated on the front panel next to the channel position
knob. The oscilloscope automatically switches to 1 M Ohm
mode to prevent possible damage if AC coupling is
selected.
• 1M Ohm mode is for use with many passive probes and for
general-purpose measurements. The higher impedance
minimizes the loading effect of the oscilloscope on the
circuit under test.
4 Press the BW Limit softkey to turn on bandwidth limiting.
Pressing the BW Limit softkey turns the bandwidth limit on or
off for the selected channel. When bandwidth limit is on, the
maximum bandwidth for the channel is approximately
25 MHz on the 300 MHz, 500 MHz, and 1 GHz bandwidth
oscilloscope models and approximately 20 MHz for the
100 MHz bandwidth oscilloscope models. For waveforms
with frequencies below this, turning bandwidth limit on
removes unwanted high frequency noise from the waveform.
The bandwidth limit also limits the trigger signal path of any
channel that has BW Limit turned on.
When BW Limit is selected, “BW” is illuminated on the front
panel next to the channel position knob ( ).
5 Press the Invert softkey to invert the selected channel.
When Invert is selected, the voltage values of the displayed
waveform are inverted. Invert affects how a channel is
displayed, but does not affect triggering. If the oscilloscope is
set to trigger on a rising edge, it remains set to trigger on a
same edge (triggers at the same point on the waveform) after
the channel is inverted.
Inverting a channel will also change the result of any
function selected in the Math menu or any measurement.
866000 Series Oscilloscope User’s Guide
Front-Panel Controls2
6 Press the Probe softkey to display the channel probe menu.
This menu lets you select additional probe parameters such
as attenuation factor and units of measurement for the
connected probe.
Probe calibration not needed, not available when this softkey is ghosted.
Measurement
units
Attenuation
factor
Skew adjust
Calibrate
probe
Return to
previous menu
• Probe Units — Press the Units softkey to select the proper
measurement unit for the connected probe. Select Volts for
a voltage probe and select Amps for a current probe.
Channel sensitivity, trigger level, measurement results,
and math functions will reflect the measurement units you
have selected.
• Probe Attenuation Factor— See “To set the probe attenuation
factor” on page 81.
• Skew — When measuring time intervals in the ns range,
small differences in cable length can affect the
measurement. Use Skew to remove cable-delay errors
between any two channels.
Probe the same point with both probes, then press Skew
and turn the Entry knob to enter the amount of skew
between channels. Each analog channel can be adjusted
±100 ns in 10 ps increments for a total of 200 ns
difference.
The skew setting is reset to zero when you press
Save/Recall&Default Setup.
• Probe Units — Press the Units softkey to select the proper
measurement unit for the connected probe. Select Volts for
a voltage probe and select Amps for a current probe.
Channel sensitivity, trigger level, measurement results,
and math functions will reflect the measurement units you
have selected.
6000 Series Oscilloscope User’s Guide87
2Front-Panel Controls
To set up the Horizontal time base
• Calibrate Probe — See “To calibrate the probes” on page 55.
1 Press the Menu/Zoom key (labeled Main/Delayed on older
models) in the Horizontal section of the front panel.
To manually control the horizontal time base on 6000L models
start the Remote Front Panel (see page 44) and select Horizontal
from the Main Menu or press ctrl+N.
Current
sample rate
Main sweep
mode
The Horizontal menu lets you select the horizontal mode (Main,
Zoom, Roll, or XY), and set the time base vernier and time
reference.
The current sample rate is displayed above the Vernier and Time
Ref softkeys.
Main mode
1 Press the Main softkey to select Main horizontal mode.
Trigger
point
Zoom sweep
mode
Time
reference
Roll
mode
Delay
time
XY
mode
Sweep
speed
Trigger
source
Time base
vernier
Trigger level
or threshold
Time
reference
Main horizontal mode is the normal viewing mode for the
oscilloscope.When the oscilloscope is stopped, you can use
the Horizontal knobs to pan and zoom the waveform.
886000 Series Oscilloscope User’s Guide
Front-Panel Controls2
2 Turn the large knob (time/division) in the Horizontal section
and notice the change it makes to the status line.
When the oscilloscope is running in Main mode, use the large
Horizontal knob to change sweep speed and use the small
knob () to set the horizontal position. When the
oscilloscope is stopped, use these knobs to pan and zoom
your waveform. The sweep speed (seconds/division) value is
displayed in the status line.
3 Press the Vernie r softkey to turn on the time base vernier.
The Ver nier softkey lets you change the sweep speed in
smaller increments with the time/division knob. The sweep
speed remains fully calibrated when Vernier is on. The value
is displayed in the status line at the top of the display.
When Vernier is turned off, the Horizontal sweep speed knob
changes the time base sweep speed in a 1-2-5 step sequence.
4 Note the setting of the Time Ref (time reference) softkey.
Time reference is the reference point on the display for
horizontal position (also called delay time). Time reference
can be set to one major division from the left or right edge, or
to the center of the display.
A small hollow triangle (∇
) at the top of the display grid
marks the position of the time reference. When horizontal
position is set to zero, the trigger point indicator ()
overlays the time reference indicator.
Turning the Horizontal sweep speed knob will expand or
contract the waveform about the time reference point (∇).
Turning the horizontal position () knob in Main mode will
move the trigger point indicator () to the left or right of the
time reference point (∇).
The time reference position sets the initial position of the
trigger event within acquisition memory and on the display,
with delay (horizontal position) set to 0. The horizontal
position setting sets the specific location of the trigger event
with respect to the time reference position. The time
reference setting affects the Zoom sweep as described in the
following ways:
6000 Series Oscilloscope User’s Guide89
2Front-Panel Controls
• When the horizontal mode is set to Main, the horizontal
position knob positions the main sweep relative to the
trigger. This delay is a fixed number. Changing this delay
value does not affect the sweep speed.
• When the horizontal mode is set to Zoom, the horizontal
position knob controls the position of the Zoom sweep
window inside the main sweep display. This delay value is
independent of sampling interval and sweep speed.
Changing the delay does not affect the position of the main
window.
5 Turn the horizontal position knob () and notice that its
value is displayed in the status line.
The horizontal position knob moves the main sweep
horizontally, and it pauses at 0.00 s, mimicking a mechanical
detent. Changing the horizontal position moves the sweep
horizontally and indicates how far the trigger point (solid
inverted triangle) is from the time reference point (hollow
inverted triangle ∇). These reference points are indicated
along the top of the display grid. The previous figure shows
the trigger point with the delay time (horizontal position) set
to 400 µs. The delay time number tells you how far the time
reference point is located from the trigger point. When delay
time (horizontal position) is set to zero, the delay time
indicator overlays the time reference indicator.
All events displayed left of the trigger point happened before
the trigger occurred, and these events are called pre-trigger
information. You will find this feature very useful because
you can now see the events that led up to the trigger point.
Everything to the right of the trigger point is called
post-trigger information. The amount of delay range
(pre-trigger and post-trigger information) available depends
on the sweep speed selected and memory depth.
906000 Series Oscilloscope User’s Guide
Front-Panel Controls2
Zoom mode
Zoom sweep is an expanded version of main sweep. When Zoom
mode is selected, the display divides in half and the Zoom
sweep icon displays in the middle of the line at the top of the
display. The top half of the display shows the main sweep and
the bottom half displays the Zoom sweep.
Zoom sweep is a magnified portion of the main sweep. You can
use Zoom sweep to locate and horizontally expand part of the
main sweep for a more detailed (higher-resolution) analysis of
signals.
The following steps show you how to use Zoom sweep. Notice
that the steps are very similar to operating the “delayed sweep”
in analog oscilloscopes.
1 Connect a signal to the oscilloscope and obtain a stable
display.
2 Press Menu/Zoom.
3 Press the Zoom softkey.
To change the sweep speed for the Zoom sweep window, turn
the sweep speed knob. As you turn the knob, the sweep speed is
highlighted in the status line above the waveform display area.
The area of the main display that is expanded is intensified and
marked on each end with a vertical marker. These markers show
what portion of the main sweep is expanded in the lower half.
The Horizontal knobs control the size and position of the Zoom
sweep. The delay value is momentarily displayed in the
upper-right portion of the display when the horizontal position
() knob is turned.
To change the sweep speed for the main sweep window, press
the Main softkey, then turn the sweep speed knob.
6000 Series Oscilloscope User’s Guide91
2Front-Panel Controls
Main sweep
window
Zoom sweep
window
These markers define the
beginning and end of the
Zoom sweep window
Time/div for
Zoom sweep
Time/div for
main sweep
Delay time
momentarily displays
when the horizontal
position knob is turned
Select main or
Zoom sweep
The area of the main display that is expanded is intensified and
marked on each end with a vertical marker. These markers show
what portion of the main sweep is expanded in the lower half.
The Horizontal knobs control the size and position of the Zoom
sweep. The delay value is momentarily displayed in the
upper-right portion of the display when the horizontal position
() knob is turned.
926000 Series Oscilloscope User’s Guide
Front-Panel Controls2
To change the sweep speed for the Zoom sweep window, turn
the sweep speed knob. As you turn the knob, the sweep speed is
highlighted in the status line above the waveform display area.
The time reference position sets the initial position of the
trigger event within acquisition memory and on the display,
with delay (horizontal position) set to 0. The delay setting sets
the specific location of the trigger event with respect to the time
reference position. The time reference setting affects the Zoom
sweep in the following ways.
When the horizontal mode is set to Main, the horizontal position
knob positions the main sweep relative to the trigger. This delay
is a fixed number. Changing this delay value does not affect the
sweep speed. When the horizontal mode is set to Zoom, the
delay controls the position of the Zoom sweep window inside
the main sweep display. This delay value is independent of
sampling interval and sweep speed.
To change the sweep speed for the main sweep window, press
the Main softkey, then turn the sweep speed knob.
For information about using Zoom mode for measurements,
refer to Chapter 5, “Making Measurements,” starting on page
207.
Roll mode
• Press the Menu/Zoom key, then press the Roll softkey.
• Roll mode causes the waveform to move slowly across the
screen from right to left. It only operates on time base
settings of 500 ms/div and slower. If the current time base
setting is faster than the 500 ms/div limit, it will be set to
500 ms/div when Roll mode is entered.
• In Normal horizontal mode, signal events occurring before
the trigger are plotted to the left of the trigger point (t) and
signal events after the trigger plotted to the right of the
trigger point.
6000 Series Oscilloscope User’s Guide93
2Front-Panel Controls
• In Roll mode there is no trigger. The fixed reference point on
the screen is the right edge of the screen and refers to the
current moment in time. Events that have occurred are
scrolled to the left of the reference point. Since there is no
trigger, no pre-trigger information is available.
If you would like to pause the display in Roll mode press the
Single key. To clear the display and restart an acquisition in Roll
mode, press the Single key again.
Use Roll mode on low-frequency waveforms to yield a display
much like a strip chart recorder. It allows the waveform to roll
across the display.
XY mode
XY mode changes the display from a volts-versus-time display to
a volts-versus-volts display. The time base is turned off. Channel
1 amplitude is plotted on the X-axis and Channel 2 amplitude is
plotted on the Y-axis.
You can use XY mode to compare frequency and phase
relationships between two signals. XY mode can also be used
with transducers to display strain versus displacement, flow
versus pressure, volts versus current, or voltage versus
frequency.
Use the cursors to make measurements on XY mode waveforms.
For more information about using XY mode for measurements,
refer to “To use the XY horizontal mode” on page 208.
Z-Axis Input in XY Display Mode (Blanking)
When you select the XY display mode, the time base is turned off. Channel 1 is the
X-axis input, channel 2 is the Y-axis input, and channel 4 (or the External trigger on
2-channel models) is the Z-axis input. If you only want to see portions of the Y
versus X display, use the Z-axis input. Z-axis turns the trace on and off (analog
oscilloscopes called this Z-axis blanking because it turned the beam on and off).
When Z is low (<1.4 V), Y versus X is displayed; when Z is high (>1.4 V), the trace
is turned off.
946000 Series Oscilloscope User’s Guide
To make cursor measurements
You can use the cursors to make custom voltage or time
measurements on oscilloscope signals, and timing
measurements on digital channels.
• To manually control Cursors on 6000L models start the
Remote Front Panel (see page 44) and select Measure from
the Main Menu or press ctrl+C to access the oscilloscope’s
Cursors Menu.
1 Connect a signal to the oscilloscope and obtain a stable
display.
2 Press the Cursors key. View the cursor functions in the
softkey menu:
• Mode — Set the cursors to measure voltage and time
(Normal), or display the binary or hexadecimal logic value
of the displayed waveforms.
• Source — selects a channel or math function for the cursor
measurements.
• X Y — Select either the X cursors or the Y cursors for
adjustment with the Entry knob.
• X1 and X2 — adjust horizontally and normally measure
time.
• Y1 and Y2 — adjust vertically and normally measure
voltage.
• X1 X2and Y1 Y2 — move the cursors together when turning
the Entry knob.
Front-Panel Controls2
NOTE
6000 Series Oscilloscope User’s Guide95
If you intend to make cursor measurements on a trace that you recall from
memory, be sure to recall both setup and trace. See “To recall waveform
trace and/or oscilloscope setup” on page 328.
For more information about using the cursors, see “To make
cursor measurements” on page 232.
2Front-Panel Controls
To make automatic measurements
You can use automatic measurements on any channel source or
any running math function. Cursors are turned on to focus on
the most recently selected measurement (right-most on the
measurement line above the softkeys on the display).
1 Press the Quick Meas key to display the automatic
measurement menu.
To manually control Quick Measurements on 6000L models
start the Remote Front Panel (see page 44) and select
Measure from the Main Menu or press ctrl+M to access the
oscilloscope’s Quick Measurements Menu.
2 Press the Source softkey to select the channel or running
math function to be measured.
Only channels or math functions that are displayed are
available for measurements. If you choose an invalid source
channel for a measurement, the measurement will default to
the nearest in the list that makes the source valid.
If a portion of the waveform required for a measurement is
not displayed or does not display enough resolution to make
the measurement, the result will be displayed with a message
such as greater than a value, less than a value, not enough
edges, not enough amplitude, incomplete, or waveform is
clipped to indicate that the measurement may not be reliable.
3 Choose a type of measurement by pressing the Select softkey,
then turn the Entry knobto select the desired
measurement from the popup list.
4 Press the Measure softkey to make the selected measurement.
5 To stop making measurements and erase the measurement
results from the measurement line above the softkeys, press
the Clear Meas softkey.
For more information about automatic measurements, see
“Automatic Measurements” on page 239.
966000 Series Oscilloscope User’s Guide
Using Labels
Front-Panel Controls2
You can define labels and assign them to each analog input
channel, or you can turn labels off to increase the waveform
display area. Labels can also be applied to digital channels on
MSO models.
To turn the label display on or off
1 Press the Label key on the front panel.
This turns on display labels for the analog and digital
channels. When the Label key is illuminated, the labels for
displayed channels are displayed at the left edge of the
displayed traces. The figure below shows an example of the
default labels assigned when you turn the label display on.
The channel labels default to their channel number.
2 To turn the labels off, press the Label key until it is not
illuminated.
6000 Series Oscilloscope User’s Guide97
2Front-Panel Controls
To assign a predefined label to a channel
1 Press the Label key.
2 Press the Channel softkey, then turn the Entry knob or
successively press the Channel softkey to select a channel for
label assignment.
The figure above shows the list of channels and their default
labels. The channel does not have to be turned on to have a
label assigned to it.
3 Press the Library softkey, then turn the Entry knob or
successively press the Library softkey to select a predefined
label from the library.
4 Press the Apply New Label softkey to assign the label to your
selected channel.
5 Repeat the above procedure for each predefined label you
want to assign to a channel.
986000 Series Oscilloscope User’s Guide
Front-Panel Controls2
To define a new label
1 Press the Label key.
2 Press the Channel softkey, then turn the Entry knob or
successively press the softkey to select a channel for label
assignment.
The channel does not have to be turned on to have a label
assigned to it. If the channel is turned on, its current label
will be highlighted.
3 Press the Spell softkey, then turn the Entr y knob to select the
first character in the new label.
Turning the Entry knob selects a character to enter into the
highlighted position shown in the "New label =" line above
the softkeys and in the Spell softkey. Labels can be up to six
characters in length.
4 Press the Enter softkey to enter the selected character and to
go to the next character position.
You may position the highlight on any character in the label
name by successively pressing the Enter softkey.
5 To delete a character from the label, press the Enter softkey
until the letter you want to delete is highlighted, then press
the Delete Character softkey.
6 When you are done entering characters for the label, press
the Apply New Label softkey to assign the label to the selected
channel.
When you define a new label, it is added to the nonvolatile label
list.
Label Assignment Auto-Increment Features
When you assign a label ending in a digit, such as ADDR0 or DATA0, the
oscilloscope automatically increments the digit and displays the modified label in
the “New label” field after you press the Apply New Label softkey. Therefore, you
only need to select a new channel and press the Apply New Label softkey again
to assign the label to the channel. Only the original label is saved in the label list.
This feature makes it easier to assign successive labels to numbered control lines
and data bus lines.
6000 Series Oscilloscope User’s Guide99
2Front-Panel Controls
To load a list of labels from a text file you create
It may be convenient to create a list of labels using a text editor,
then load the label list into the oscilloscope. This lets you type
on a keyboard rather than edit the label list using the
oscilloscope’s controls.
You can create a list of up to 75 labels and load it into the
oscilloscope. Labels are added to the beginning of the list. If
more than 75 labels are loaded, only the first 75 are stored.
To load labels from a text file into the oscilloscope:
1 Use a text editor to create each label. Each label can be up to
six characters in length. Separate each label with a line feed.
2 Name the file labellist.txt and save it on a USB mass storage
device such as a thumb drive.
3 Load the list into the oscilloscope using the File Explorer
(press Utility→ File Explorer).
Label List Management
When you press the Library softkey, you will see a list of the last 75 labels used.
The list does not save duplicate labels. Labels can end in any number of trailing
digits. As long as the base string is the same as an existing label in the library, the
new label will not be put in the library. For example, if label A0 is in the library and
you make a new label called A12345, the new label is not added to the library.
When you save a new user-defined label, the new label will replace the oldest
label in the list. Oldest is defined as the longest time since the label was last
assigned to a channel. Any time you assign any label to a channel, that label will
move to the newest in the list. Thus, after you use the label list for a while, your
labels will predominate, making it easier to customize the instrument display for
your needs.
When you reset the label library list (see next topic), all of your custom labels will
be deleted, and the label list will be returned to its factory configuration.
1006000 Series Oscilloscope User’s Guide
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.