Agilent Technologies G6600-90006 User Manual

Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors
Operation and Maintenance Manual
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2007
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Agi­lent Technologies, Inc. as governed by United States and international copyright laws.
Acknowledgements
Torx® is a registered trademark of Textron Inc.; Teflon® is a registered trademark of E.I. du Pont de Nemours and Company. Kimwipe® is a registered trademark of the Kimberly-Clark Corporation.
Manual Part Number
G6600-90006
Edition
First edition, June 2007 Replaces G6600-90002
Printed in USA
Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Gov­ernment will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Govern­ment users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
If software is for use in the performance of a U.S. Government prime contract or sub­contract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as
Warnings
English
WARNING
WARNING
WARNING
WARNING
This symbol on the instrument indicates that the user should refer to the man­ual for operating instructions.
Any operation requiring access to the inside of the equipment, could result in injury. To avoid potentially dangerous shock, disconnect from power supply before opening the equipment.
For continued protection against fire hazard replace fuse with same type and rating.
This symbol indicates that to comply with European Union Directive 2002/96/EC for waste electrical and electronic equipment (WEEE), the Ana­lyzer should be disposed of separately from standard waste.
WARNING
WARNING
WARNING
Operation and Maintenance Manual 3
This is a safety Class I product. It must be wired to a mains supply with a protective earthing ground incorporated into the power cord. Any interruption of the protective conductor, inside or outside the equipment, is likely to make the instrument dangerous. Intentional interruption is prohibited.
If this instrument is used in a manner not specified by Agilent, the protection provided by the instrument may be impaired.
High voltage is present in the instrument when the power cord is connected, even if the main power switch is in the standby mode. To avoid potentially dangerous shock, discon­nect the power cord before removing the side panels.
WARNING
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized by using the instrument in a well-ventilated area and by venting the exhaust of the vac­uum pump to a fume hood. The ozone generator should be turned off when the instrument is not in use.
WARNING
WARNING
WARNING
WARNING
Burner temperature Is extremely hot. Do not touch. Allow to cool before servicing.
Hydrogen is an extremely flammable gas. Use appropriate care when handling. Inspect all connections with a suitable leak detector.
Oxygen rich environments can promote combustion and even result in spontaneous com­bustion under conditions of high pressure and exposure to contamination. Use only oxy­gen rated components and ensure that components are oxygen clean prior to use with pure oxygen.
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxi­dant sensors or burst their connective tubing.
4 Operation and Maintenance Manual
Español
WARNING
WARNING
WARNING
WARNING
Cualquier operación que requiera acceso al interior del equipo, puede causar una lesión. Para evitar peligros potenciales, desconectarlo de la alimentación a red antes de abrir el equipo.
Para protección contínua contra el peligro de fuego, sustituir el fusible por uno del mismo tipo y características.
Este símbolo en el instrumento indica que el usuario debería referirse al man­ual para instrucciones de funcionamiento.
Esto es un producto con clase I de seguridad. Debe conectarse a una red que disponga de tierra protectora en el cable de red. Cualquier interrupción del conductor protector, dentro o fuera del equipo, puede ser peligroso. Se prohibe la interrupción intencionada.
WARNING
Si este instrumento se usa de una forma no especificada por Agilent, puede desactivarse la protección suministrada por el instrumento.
Operation and Maintenance Manual 5
Français
WARNING
WARNING
WARNING
WARNING
Chaque opération à l'intérieur de l'appareil, peut causer du préjudice. Afin d'éviter un shock qui pourrait être dangereux, disconnectez l'appareil du réseau avant de l'ouvrir.
Afin de protéger l'appareil continuellement contre l'incendie, échangez le fus­ible par un fusible du même type et valeur.
Le symbol indique que l'utilisateur doit consulter le manuel d'instructions.
Ceci est un produit de Classe de sécurité I. L'instrument doit être branché sur l'alimentation secteur par un fil de secteur prévu d'une prise de masse. Chaque interruption du conducteur protégeant, à l'intérieur ou á l'extérieur de l'appareil peut rendre l'instrument dangereux. Interruption intentionnelle est interdite.
WARNING
Si l'instrument n'est pas utilisé suivant les instructions de Agilent, les dispositions de sécurité de l'appareil ne sont plus valables.
6 Operation and Maintenance Manual
Deutsch
WARNING
WARNING
WARNING
WARNING
Vor dem Öffnen des Gerätes Netzstecker ziehen!
Für kontinuierlichen Schutz gegen Brandgefahr dürfen bei Sicherungswech­sel nur Sicherungen der gleichen Stärke verwendet werden!
Dieses Symbol auf dem Gerät weist darauf hin, dass der Anwender zuerst das entsprechende Kapitel in der Bedienungsanleitung lesen sollte.
Dies ist ein Gerät der Sicherheitsklasse I und darf nur mit einem Netzkabel mit Schutzleiter betrieben werden. Jede Unterbrechung des Schutzleiters au erhalb oder innerhalb des Gerätes kann das Gerät elektrisch gefährlich machen. Absichtliches Unterbrechen des Schutzleiters ist ausdrücklich verboten.
WARNING
Wenn das Gerät nicht wie durch die Firma Agilent, vorgeschrieben und im Handbuch beschrieben betrieben wird, können die im Gerät eingebauten Schutzvorrichtungen beeinträchtigt werden.
Operation and Maintenance Manual 7
Italiano
WARNING
WARNING
WARNING
WARNING
Qualsiasi intervento debba essere effettuato sullo strumento può essere potenzialmente pericoloso a causa della corrente elettrica. Il cavo di alimen­tazione deve essere staccato dallo strumento prima della sua apertura.
Per la protezione da rischi da incendio in seguito a corto circuito, sostituire I fusibili di protezione con quelli dello stesso tipo e caratteristiche.
Il simbolo sullo strumento avverte l'utilizzatore di consultare il Manuale di Istruzioni alla sezione specifica.
Questo strumento è conforme alle specifiche per I prodotti in Classe I - Il cavo di alimentazione dalla rete deve essere munito di "terra". Qualsiasi interruzione del cavo di terra all'interno ed all'esterno dello strumento potrebbe risultare pericolòsa. Sono proibite interruzioni intenzionali.
WARNING
Se questo strumento viene utilizzato in maniera non conforme alle specifiche di Agilent, le protezioni di cui esso è dotato potrebbero essere alterate.
8 Operation and Maintenance Manual
Dutch
WARNING
WARNING
WARNING
WARNING
Iedere handeling binnenin het toestel kan beschadiging veroorzaken. Om ied­ere mogelijk gevaarlijke shock te vermijden moet de aansluiting met het net verbroken worden, vóór het openen van het toestel.
Voor een continue bescherming tegen brandgevaar, vervang de zekering door een zekering van hetzelfde type en waarde.
Het symbool geeft aan dat de gebruiker de instructies in de handleiding moet raadplegen.
Dit is een produkt van veiligheidsklasse I. Het toestel moet aangesloten worden op het net via een geaard netsnoer. Bij onderbreking van de beschermende geleider, aan de binnenzijde of aan de buitenzijde van het toestel, kan gebruik het toestel gevaarlijk maken. Opzettelijke onderbreking is verboden.
WARNING
Indien het toestel niet gebruikt wordt volgens de richtlijnen van Agilent, gelden de veiligheidsvoorzieningen niet meer.
Operation and Maintenance Manual 9
The following symbols are used on the equipment:
Caution - Refer to manual for operating instructions
Atención - Ver documentación pertinente.
Attention - Consultez les ocuments d'accomagnement.
Vorsicht - Siehe beiliegende Unterlagen.
Pericolo - Vedi documentazione allegata.
Voorzichtig - Raadpleeg di bijehorende documentatie.
Caution - Risk of electrical shock.
Atención - Riesgo de sacudidas eléctricas.
Attention - Risque de choc électrique.
Vorsicht - Risiko eines Elektroschocks.
Pericolo - Pericolo di scosse elettriche.
Voorzichtig - Hoge spanning, levensgevaar.
Caution - Hot surface.
Atención - Superficie caliente.
Attention - Surface brûlante.
Vorsicht - Heisse Oberfläche.
Pericolo - Superficie rovente.
Voorzichtig - Heet oppervlak.
10 Operation and Maintenance Manual

Contents

1 Introduction
2 System Description
Specifications 20
Technical Information — 355 SCD 20 Technical Information — 255 NCD 20 Physical Specifications 21
Theory of Operation 23
Sulfur Chemiluminescence Detector 23 Nitrogen Chemiluminescence Detector 23 Dual Plasma Controller 24
Description of Major Components 25
Dual Plasma Burner 25 Figure 1. Cross-Section of the Dual Plasma Burner for the 355 SCD 26 Figure 2. Cross-Section of the Dual Plasma Burner for the 255 NCD 27 Dual Plasma Controller 27 Ozone Generator 28 Chemiluminescence Reaction Cell and Photomultiplier Tube 28 Figure 3. 355 SCD Left Side 29 Figure 4. 355 SCD Right Side 30 Figure 5. 255 NCD Left Side View 31 Figure 6. 255 NCD Right Side View 32 Pressure Transducer 32 Vacuum Pump 32 Chemical Trap 33 Oil Coalescing Filter 33 FID Adapter (Optional) 33 NCD Reaction Cell 33 NCD Photomultiplier Tube and Cooler 34 Figure 7. Schematic for 355 SCD 35 Figure 8. Schematic for 255 NCD, in Nitrogen Mode 36 Figure 9. Schematic for 255 NCD, in Nitrosamine Mode 37
Operation and Maintenance Manual 11
3 Installation
Overview 40
Step 1: Selecting a Location 41
Power Requirements 41 Figure 10. Drawing of the Detector with Dual Plasma Burner and Controller 42 Environmental Considerations 42 Combustion Gas Requirements 42
Step 2: Unpack and Inspect the Instrument 45
Required Installation Tools 45
Step 3: Set Up the Vacuum Pump 46
Installing the Edwards RV5 Pump Oil-Sealed Vacuum Pump 46 Figure 11. RV5 Oil-Sealed Vacuum Pump and Associated Traps (Front Side) 49 Figure 12. RV5 Oil-Sealed Vacuum Pump and Associated Traps (Back Side) 49 Figure 13. RV5 Oil- Sealed Vacuum Pump and Associated Traps (Top) 50 Figure 14. RV-5 Oil-S ealed Vacuum Pump Exhaust Line 51 Installing the Welch Dry Piston Vacuum Pump 52 Figure 15. The Welch Dry Piston Pump 54 Figure 16. Oil Drain Kit with Ballast Control 54
Step 4: Connect the Power Cord 56
Figure 17. SCD Rear Panel Diagram 56
Step 5: Install the Dry Compressed Air or O
Supply 57
2
Step 6: Install the Signal Output Cables 58
Standard Cable Connection 58 HP 3390 Series Integrator Cable Connection 58 HP 3396 Integrator Cable Connection 58 HP 5890 GC Analog Input Board 58 Agilent 6890 GC Analog Input Board 58
Step 7: Install the Dual Plasma Burner 59
Figure 18. Dimensions of GC Liner Cut-Outs 59
Step 8: Install the Dual Plasma Controller 60
Figure 19. Dual Plasma Controller Rear Panel 60
Step 9: Install Column Connections 61
Capillary Columns 61 Figure 20. Measuring Column Insertion 61
12 Operation and Maintenance Manual
Packed Columns and Columns with an Outside Diameter > 0.8 mm 61
Step 10: Install the Transfer Line 62
4 Front Panel Controls and Initial Startup
Detector Front Panel Controls 64
Figure 21. Front Panel Controls 64 Power Controls 64 Display Output Controls 65 Signal Controls 65 Dual Plasma Controller Controls 66 Figure 22. Dual Plasma Controller Front Panel 66 Initial Startup 67
Detector Interface Setup 69
Initial Checkout 69 Monitoring Oxidizer and Hydrogen Flow with the Dual Plasma Controller 70
5Operation
Start-Up Procedure 72
Detector Operation 73
Detector Stability and Response 73 Column Bleed 73 Coking 73 Hydrogen Poisoning 73 Contaminated Gases 74 Fluctuating Pressures 74
Typical Operating Conditions 75
Table 1. Typical Operating Conditions 75
Detection Limits 76
Table 2. Expected Detection Limits for Chromatographic Conditions 76
Instrument Shut-Down 77
Daily Shutdown 77 Complete Shutdown 77
Special Operating Modes 78
Using the 255 NCD in Nitrosamine Mode 78 Using the SCD in High Sensitivity Mode for Nonhydrocarbon Gaseous Samples 78
Operation and Maintenance Manual 13
6Maintenance
Pump Maintenance 82
Cleaning the Detector 83
Changing the Oil Mist Filter (RV5) 84
Reaction Cell Cleaning 85
Flow Sensor Calibration 88
Detector Sensitivity 89
Assembling the Dual Plasma Burner for Component Replacement with the SCD 90
Table 3. Operating Life of Components for Edwards RV5 Vacuum Pump 82
Figure 23. Reaction Cell, PMT Housing and PMT Socket 86 Figure 24. Reaction Cell 86
Figure 25. Ferrule Placement on Lower Burner Tube 90 Figure 26. Proper Ferrule Orientation to the Large Ceramic Tube 91 Figure 27. Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly 91 Figure 28. Orientation of the Double Taper Ferrule 92 Figure 29. Positioning the Upper Tube in the Union Fitting 92 Figure 30. Tightening the Heater Swivel Nut 93 Figure 31. Proper Alignment of the Burner 93
Assembling the Dual Plasma Burner for Component Replacement with the NCD 95
Figure 32. Ferrule Placement on Lower Burner Tube 95 Figure 33. Proper Ferrule Orientation to the Large Quartz Tube 96 Figure 34. The Quartz Tube Properly Inserted into the Quartz Heater Assembly 96 Figure 35. Burner Assembly Detail 97 Figure 36. Burner Assembly Alignment 97
Tube Replacement for the SCD 98
Figure 37. Orientation of the Double Taper Ferrule 98 Figure 38. Positioning the Upper Tube in the Union Fitting 99 Figure 39. Proper Ferrule Orientation to the Large Ceramic Tube 99 Figure 40. Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly 100 Figure 41. Tightening the Heater Swivel Nut 100 Figure 42. Proper Alignment of the Burner 101
Tube Replacement for the NCD 102
Figure 43. Proper Ferrule Orientation to the Large Quartz Tube 102 Figure 44. Large Quartz Tube Properly Inserted into the Quartz Heater Assembly 103 Figure 45. NCD Tube Replacement Detail 104 Figure 46. NCD Tube Replacement Detail 104
14 Operation and Maintenance Manual
7 Troubleshooting
Solving Detector Problems 106
Power Problems 107
Detector Fuse 107 Vacuum Pump Fuse 107 Dual Plasma Controller Fuses 107 Table 4. Fuses for 100 V, 120 V and 230 V Versions of 355 SCD and 255 NCD 108 Figure 47. Fuse Positions on the Power Supply Board 108
Ozone Generation Problems 109
Ozone Generator 109 High Voltage Transformer 109 Plugged Restrictor Lines 110
Response Problems 111
Temperature Reading Problems 112
Diagnosing General Problems 113
Table 5. Troubleshooting Detector Issues 113 Table 6. Troubleshooting Pump Issues 115 Table 7. Troubleshooting Burner Issues 116
Index
Operation and Maintenance Manual 15
16 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
1 Introduction
This manual will guide you in the installation, operation, and troubleshooting of the Agilent 355 Sulfur Chemiluminescence Detector (SCD) and the Agilent 255 Nitrogen Chemiluminescence Detector (NCD). This manual is intended for use with the Agilent 355 SCD or 255 NCD with the Dual Plasma Burner and Controller.
This operation and service manual has some important conventions, such as the use of boxed warnings. This information is deliberately set out from the text for emphasis and should be followed to assure operator safety and proper instrumental operation.
If you are installing the 355 SCD or 255 NCD yourself, follow the installation procedures described in the following sections. If your instrument is already installed, turn to the Operation section to begin.
Agilent Technologies
17
18 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
2 System Description
Specifications 20
Theory of Operation 23 Description of Major Components 25
Agilent Technologies
19

Specifications

Technical Information — 355 SCD
Sensitivity
Typical Selectivity g S/g C > 2 x 10
Linearity >10
Precision and Stability
*
Typical < 0.5 pg S/second (signal to noise 3.3:1)
7
4
<2% RSD 2 hours
<5% RSD 72 hours
Ozone flow through the Post Ozone Restrictor 20-30 mL/min at 3-6 psig
Typical reaction cell pressure 4 - 8 Torr RV5 Oil Sealed Pump
6 - 12 Torr Dry Piston Pump
Typical Burner Pressure 250-400 Torr operating
Typical Burner Temperature 800 °C
Typical Air Flow Rate 65 SCCM recommended
3-10 SCCM recommended with FID adapter
Typical Hydrogen Flow Rate 40 SCCM recommended
Signal Output Ranges 0-1V, 0-10V
Typical time to reach 800 °C from room
10 min typical(120 VAC, 60 Hz)
temperature
Typical safety shroud outside temperature <65 °C at 800 °C Burner temperature typical
Recorder output 0-1 V or 0-10 V
* Burner temperature 800 °C, 80 SCCM air, and 60 SCCM hyrdrogen, test compound dimethyl sulfide in toluene.
† Based on thiopene in benzene at 1 ppm mass sulfur, 1 µL injection split 1:10, 30 m, 0.32 mm ID, 1 µm thick CP
wax (n=10 for 2.5 hours; n=42 for 72 hours).
Subject to change without notice.
Technical Information — 255 NCD
Sensitivity <3 pg N/second (signal to noise 3:1) in both N
and nitrosamine modes
Selectivity g N/g C > 2 to 10
nitrosamine mode is matrix-dependent)
Linearity >10
4
20 Operation and Maintenance Manual
7
in N mode (selectivity in
Repeatability
Gas flow through Ozone Generator 20-30 mL/min at 3-6 psig (inlet pressure)
Typical reaction cell pressure 4 - 8 Torr RV5 Oil Sealed Pump
Typical Burner Pressure 130 - 150 torr operating
Typical Burner Temperature 900 °C
Typical Hydrogen Flow Rate 6-10 SCCM
Typical Oxygen Flow Rate 8 - 15 SCCM
Signal Output Ranges 0-1V, 0-10V, 0 - 10V
* Burner temperature 950 °C, 11 SCCM oxygen, and 6 SCCM hydrogen; 25 ppm N as nitrobenzene in toluene;
0.2 µL injection on column (HP 19095-121Z), n=7 for 3 hours; n=13 for 18 hours and n=10 n-dipropylnitrosamine in toluene at 4 µg/mL, 0.2 µL injection on column (HP 19095-121Z).
*

Physical Specifications

<1.5% RSD 8 hours (~ the same in N and nitrosamine)
<2% RSD 18 hours (~3% RSD in nitrosamine mode over 21 hours)
6 - 12 Torr Dry Piston Pump
Power requirements
355 SCD Detector 115 VAC, 50/60 Hz, 1400 W
100 VAC, 50/60 Hz, 1400 W 220-240 VAC, 50/60 Hz, 650W
255 NCD Detector 115 VAC, 50/60 Hz, 1400 W
100 VAC, 50/60 Hz, 1400 W 220-240 VAC, 50/60 Hz, 650W
Dual Plasma Controller 100-120 VAC, 50/60 Hz, 200 W
220-240 VAC, 50/60 Hz, 200 W
Dimensions and weight
Detector Height: 16.0 in (40.6 cm)
Width: 9.2 in (23.4 cm) Depth: 21.8 in (55.3 cm) 355 SCD Weight: 34.0 lbs (15.0 kg) 255 NCD Weight: 37.5 lbs (17.0 kg)
Operation and Maintenance Manual 21
Dual Plasma Controller Height: 5.0 in (12.7 cm)
Width: 9.5 in (24.1 cm) Depth: 12.5 in (31.8 cm) Weight: 9.9 lbs (4.5 kg)
Dual Plasma Burner Height: 12.3 in (31.2 cm)
Diameter: 4.0 in (10.2 cm) Weight: 1.9 lbs (0.9 kg)
Oil Sealed Vacuum Pump (RV5) Height: 10.3 in (26.1 cm)
Width: 6.0 in (15.2 cm) Depth: 16.9 in (43.0 cm) Weight: 47.3 lbs (21.5 kg)
or
Oil Free Dry Piston Pump Height: 12.0 in (30 cm)
Width: 9.0 in (22.9 cm) Depth: 14.0 in (35.6 cm) Weight: 29.9 lbs (13.6 kg)
Installation Category II
Pollution Degree 2
Ambient Temperature 50 - 104 °F (10-40 °C)
Relative Humidity Up to 95%, noncondensing
Normal Operating Environment Intended for indoor use only
Maximum Altitude 2,000 m (6,562 ft)
Mains Supply Voltage Fluctuation not to exceed 10% of nominal
voltage
22 Operation and Maintenance Manual

Theory of Operation

-

Sulfur Chemiluminescence Detector

The Agilent model 355 Sulfur Chemiluminescence Detector (SCD) is a sulfur-selective detector for gas chromatography. Operation of the SCD is based on the chemiluminescence (light-producing reaction) from the reaction of ozone with sulfur monoxide (SO) produced from combustion of the analyte:
Sulfur compound (analyte)
SO + O
SO2 + O2 + hη (<300–400 nm)
3
SO + H
A vacuum pump pulls the combustion products into a reaction cell at low pressure, where excess ozone is added. Light produced from the subsequent reaction is optically filtered and detected with a blue-sensitive photomultiplier tube and the signal is amplified for display or output to a data system. Figure 1 is a pneumatic flow diagram that illustrates how the components of the system are integrated.
The Detector has an enclosed, dedicated (Dual Plasma) Burner designed to enhance production of the SO intermediate. This Dual Plasma Burner mounts in the detector port of the GC. A Dual Plasma Controller provides temperature control and gas-flow regulation to operate the Dual Plasma Burner.
The Agilent model 355 SCD provides high sensitivity (<0.5 pg S/sec), with linear and equimolar response over four orders of magnitude (per Sulfur atom) while maintaining high selectivity over common solvents. The Agilent SCD is compatible with most commercially available gas chromatographs.

Nitrogen Chemiluminescence Detector

Operation of the Agilent model 255 Nitrogen Chemiluminescence Detector is based on the chemiluminescence or light-producing reaction of ozone with nitric oxide formed from combustion. Reacting nitric oxide with ozone results in the formation of electronically excited nitrogen dioxide. The excited nitrogen dioxide emits light, a chemiluminescence reaction, in the red and infrared region of the spectrum. The light emitted is directly proportionally to the amount of nitrogen in the sample,
O + other products
2
NO + O
NO2 NO2 + hη (>800 nm)
3
The light (hη) emitted by the chemical reaction is optically filtered and detected by a photomultiplier tube. The signal from the photomultiplier tube is amplified for display or output to a data system. Organic compounds containing nitrogen react to form nitric oxide, carbon dioxide, and water.
H
2/O2
--------------
R-N + O
Operation and Maintenance Manual 23
→ NO + CO
2
Δ
+ H2O
2

Dual Plasma Controller

The Dual Plasma Controller provides all operational parameters of the Dual Plasma Burner except for the Detector base temperature. The Detector base temperature is controlled by circuitry in the GC. Parameters monitored or regulated by the Controller include Burner temperature, Burner temperature set-point, hydrogen and oxidant flow rates, and Burner pressure. The temperature set-point, actual pressure, oxidant and hydrogen flow rates are displayed by rotation of a 4-position switch. Power, valve operation, temperature within set-point range and fault conditions are indicated with LED illumination on the front display panel.
The Dual Plasma Controller incorporates several safety features. The safety circuitry detects faults such as power loss, vacuum loss, thermocouple failure, heater element failure, broken ceramic tube, or high temperature. When a fault is detected, the Fault LED illuminates and hydrogen and oxidant flow is stopped by normally-closed solenoid valves.
Dual Plasma Burner with the 355 SCD
The Dual Plasma Burner is based on the same chemistry and basic principles of earlier SCD Burner designs. A key difference, however, is the addition of a second “flame” or “plasma,” the lower is oxygen-rich and the upper is hydrogen-rich. The Burner consists of a tower assembly that contains an outer sheath for burn protection, a heating element, thermocouple, and combustion tubes. Conversion of sulfur containing compounds to SO occurs within the ceramic reaction chamber housed in the Burner assembly. Agilent also provides a Flame Ionization Detector (FID) option for the simultaneous collection of hydrocarbon and sulfur chromatograms for some GCs.
Dual Plasma Burner with the 255 NCD
Compounds eluted from the GC column are combusted in the Dual Plasma Burner first by an oxygen rich flame (plasma) followed by catalytic combustion on a Noble metal screen. For hydrocarbons, this two stage combustion technique results in complete conversion of the matrix to products, such as carbon dioxide and water, which do not chemiluminesce with ozone. Nitrogen atoms in a compound are converted into nitric oxide and potentially other nitrogen oxide species. In the second stage, the catalyst is used to convert other nitrogen oxide species to nitric oxide, resulting in a high efficiency of conversion.
24 Operation and Maintenance Manual

Description of Major Components

Dual Plasma Burner

The Dual Plasma Burner consists of a tower assembly that contains an outer sheath for burn protection, a heating element, thermocouple, and combustion tubes. In the SCD, conversion of sulfur containing compounds to SO occurs within the ceramic reaction chamber housed in the Burner assembly and potentially interfering hydrocarbons are oxidized to CO
the oxidant. In the NCD, oxygen is used as the oxidant.
A fitting is located on top of the Burner. The vacuum line from the Detector box is connected directly to the top of this fitting and H2 is input to the longer side of this fitting. The air inlet is connected to the base of the Burner.
The Burner is mounted onto the GC by a model-specific mounting kit (see www.Agilent.com/chem or contact Agilent for the most current information). The GC column is inserted into the Burner using a 1/32" knurled nut and fused silica adapter ferrule.
A cross-section illustration of the Dual Plasma Burner for the 355 SCD is shown in Figure 1 and for the 255 NCD is shown in Figure 2.
and H2O, with air as
2
Operation and Maintenance Manual 25
Figure 1 Cross-Section of the Dual Plasma Burner for the 355 SCD
26 Operation and Maintenance Manual
Figure 2 Cross-Section of the Dual Plasma Burner for the 255 NCD

Dual Plasma Controller

The Dual Plasma Controller powers the Dual Plasma Burner and supplies its gases. Hydrogen and oxidant should be provided at 25 psig (1.72 bar) and the Controller is plugged into an appropriate AC electrical outlet.
WARNING
Operation and Maintenance Manual 27
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxi­dant sensors or burst their connective tubing.
The parameters monitored or regulated by the Controller include Burner
temperature, hydrogen and oxidant f low rates, and Burner pressure. The temperature, actual pressure, oxidant and hydrogen flow rates are selected for display by rotation of a 4-position control knob. Power, valve operation, temperature within set-point range and fault conditions are indicated with LED illumination on the front display panel.
The Dual Plasma Controller incorporates several safety features. The safety circuitry detects faults such as power loss, vacuum loss, thermocouple failure, heater element failure, broken ceramic tube, or high temperature. When a fault is detected, the Fault LED illuminates and hydrogen and oxidant flow is stopped by normally-closed solenoid valves.

Ozone Generator

The SCD and NCD produce ozone by corona discharge using a clean, pressurized air or oxygen source. Use of oxygen should increase ozone production and, hence, Detector response. High voltage to the ozone generator is applied only when the reaction cell pressure is less than 100 torr in the SCD and less than 200 torr in the NCD. Gas flow through the ozone generator is controlled by a pressure regulator and flow restrictors.
WARNING
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized by using the instrument in a well ventilated area, changing the chemical trap regularly, and venting the exhaust of the vacuum pump. The ozone generator should be turned off when the instrument is not in use to reduce maintenance requirements.

Chemiluminescence Reaction Cell and Photomultiplier Tube

Sulfur monoxide (formed in the Burner) and ozone (produced in the ozone generator located in the Detector) are mixed in the reaction cell. The cell is designed such that the reaction between SO and O
the photomultiplier tube (PMT). A UV band pass filter (300 - 400 nm) located between the reaction cell and the PMT selectively transmits the light emitted by the SO/O
the UV band pass filter eliminates interference from non-sulfur containing analytes (e.g. nitric oxide, olefins, etc.) which also undergo chemiluminescent reactions with ozone. A background signal is typically present as a result of ozone-wall interactions and low level sulfur contamination of Detector gases. This background signal can be used as a troubleshooting aid (see Section 10).
reaction. Efficient combustion in the ceramic tubes coupled with
3
occurs directly in front of
3
28 Operation and Maintenance Manual
7 9
10
6
8
1. Ozone Generator
2. High Voltage Transformer
3. Photomultiplier Tube Socket
4. Photomultiplier Housing
5. Reaction Cell
Figure 3 355 SCD Left Side
3
4
5
2
6
1
6. Vacuum Line
7. Particulate Filter
8. Pre-Ozone Restrictor
9. Post-Ozone Restrictor
10. Pressure Transducer
Operation and Maintenance Manual 29
P4 P5
WHT LOAD NEUT
P2 P3
TP3
TP8
RV1
HV
TP9 TP11
TP8 TP10 TP3
TP12
AMP 1 I
ALCO
11 12 15
18
16
ELECTRONICS
OZONE
PUMP
17
TP4
TP5 TP1
TP2
TP12
HV/100
TP11
TP13
GND
TP14
TP9 TP10
TP15 TP6
TP7 TP16
LOAD NEUT
PUMP
OUT
P6 P7
OZONE
GEN
WHT
LINE NEUT
AC
WHT
IN OFST
OUT OFST
RV1
RV2
TP2
TP7 TP6 TP4 TP 5
TP13
OUT
13
TO FRONT
TP1
PANEL
SIEVERS RESEARCH
USA
14
Figure 4 355 SCD Right Side
11. Amplifier Cable
12. HV Cable
13. PMT Amplifier
14. EMI Filter
15. Fuses
16. Pressure Regulator
17. Transfer line
18. Front panel display
30 Operation and Maintenance Manual
Figure 5 255 NCD Left Side View
Operation and Maintenance Manual 31
Figure 6 255 NCD Right Side View

Pressure Transducer

Vacuum in the reaction cell is measured using a pressure transducer. The pressure of the reaction cell can be monitored from the front panel and will typically range from 5 to 10 torr depending on the type of vacuum pump used, condition of the chemical (ozone) trap, ceramic tube position and the condition of the combustion chamber. The range of response is from 0 to 760 torr.

Vacuum Pump

There are two choices of vacuum pumps for the 355 SCD and 255 NCD. A two-stage, oil-sealed rotary vacuum pump is used to produce an operating pressure between 3 and 10 torr in the reaction cell. The oil-free Dry Piston pump produces a vacuum between 5 and 12 torr. All vacuum pumps serve the same purpose:
32 Operation and Maintenance Manual
Collection and transfer of the combustion gases from the Burner to the reaction cell.
Transfer of ozone from the ozone generator to the reaction cell.
Reduction of non-radiative collisional quenching of the emitting species in
the chemiluminescent reaction cell.
The higher vacuum produced by the oil-sealed rotary pump facilitates a shorter residence time in the reaction chamber, and therefore reduces the incidence of collisional relaxation of the excited SO
lower detection limits using the oil-sealed pump as compared to the oil-free pump.

Chemical Trap

To protect the vacuum pump from exposure to ozone, a chemical trap is located at the inlet of the vacuum pump. This trap contains a catalyst that converts ozone to diatomic oxygen. This trap is a consumable part and should be replaced every 90 days of continuous Detector usage.

Oil Coalescing Filter

The oil-sealed rotary vacuum pump is operated with the gas ballast partially open to aid in the elimination of water produced in the Burner and transferred to the pump. As a result of the open gas ballast and the high flow rates of gases, oil vaporized in the pump can escape through the pump exhaust. To minimize oil loss, an oil coalescing filter is placed on the pump exhaust to trap vaporized oil and to return this oil to the vacuum pump oil reservoir. This is not necessary for use with the oil-free pump. A replaceable element in the filter is a consumable part and should be replaced every 90 days of continuous use.
. This results in slightly
2

FID Adapter (Optional)

The Agilent 355 Sulfur Chemiluminescence Detector is designed to mount into most GC detector ports and operate as a stand-alone sulfur detector. For some applications it is desirable to monitor both sulfur and hydrocarbon components using a single column without splitting. For this reason, Agilent offers a few adapter kits to mount the Dual Plasma Burner onto a Flame Ionization Detector for the simultaneous collection of FID and SCD chromatograms. During dedicated SCD operation, 100% of the column effluent passes through the Burner to the Detector. During simultaneous detection approximately 10-20% of the FID exhaust gases are drawn into the Burner through a restrictor, which reduces sulfur sensitivity to approximately 1/10 of the signal observed in a dedicated SCD Burner.

NCD Reaction Cell

The model 255 NCD reacts nitric oxide from the Burner and ozone from the Ozone Generator in the Chemiluminescence Reaction Cell. The reaction occurs
Operation and Maintenance Manual 33
directly in front of the photomultiplier tube (PMT). A red cut-off filter between the reaction cell and the PMT selectivity transmits the light from the nitric oxide and ozone reaction.
Efficient combustion in the ceramic tubes and the red cut-off filter eliminate interference from non-nitrogen containing analytes (sulfur dioxide, alkenes, olefins) that have chemiluminescence reactions with ozone.
A background signal is typically present as a result of ozone-wall interactions and low level nitrogen contamination from carrier and detector gases. The background signal is a useful troubleshooting aid.

NCD Photomultiplier Tube and Cooler

In the NCD, a red-sensitive PMT detects emissions from the nitric oxide and ozone reaction. A thermoelectric cooler cools the PMT 40 °C below ambient temperature to approximately -15 to -20 °C to reduce background noise from the PMT. The cooler operates whenever the Detector is connected to AC power.
34 Operation and Maintenance Manual
nt
25 psig max
P
Dual Plasma Controller
Dual
Plasma
Burner
Column
Gas Chromatograph
Mass
flow
Mass
flow
F.C.
F.C.
S
NC
S
NC
Transfer
Line
25 psig max
Air H
355 SCD
Ozone
Reaction
Cell
2
P
PMT
Ve
Trap
C-Filter
Figure 7 Schematic for 355 SCD
Operation and Maintenance Manual 35
25 psig max
P
Dual
Plasma
Catalyst
Burner
Column
Gas Chromatograph
Mass
flow
Mass
flow
F.C.
F.C.
S
NC
S
NC
Dual Plasma Controller
Transfer
Line
25 psig max
O
2
255 NCD
Reaction
Cell
Ozone
P
PMT
H
2
Vent
Trap
C-Filter
Figure 8 Schematic for 255 NCD, in Nitrogen Mode
36 Operation and Maintenance Manual
Catalyst
Dual
Plasma
Burner
Mass
flow
F.C.
Mass
flow
P
F.C.
Dual Plasma Controller
Transfer
NC
NC
Line
S
S
25 psig max
O2*
*Or O
, He, N
2
2
255 NCD
Column
Gas Chromatograph
Figure 9 Schematic for 255 NCD, in Nitrosamine Mode
Ozone
Reaction
Cell
Trap
P
PMT
Vent
C-Filter
Vacuum Pump
Operation and Maintenance Manual 37
38 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
3 Installation
Overview 40
Step 1: Selecting a Location 41 Step 2: Unpack and Inspect the Instrument 45 Step 3: Set Up the Vacuum Pump 46
Step 4: Connect the Power Cord 56 Step 5: Install the Dry Compressed Air or O2 Supply 57 Step 6: Install the Signal Output Cables 58
Step 7: Install the Dual Plasma Burner 59 Step 8: Install the Dual Plasma Controller 60 Step 9: Install Column Connections 61
Step 10: Install the Transfer Line 62
Agilent Technologies
39

Overview

Installation and start-up of the Agilent 355 SCD or 255 NCD by a qualified Agilent Service technician is recommended. If you choose to install the detector yourself, carefully read all of this chapter prior to installation of the instrument.
Although every reasonable safeguard against shipping damage has been taken, product damage may still occur due to excessive mishandling. If obvious damage has occurred during shipment, contact Agilent. Shipping materials for the 355 SCD or 255 NCD should be saved. If the instrument must be returned to the factory, it must be packed in the original carton to reduce the chance for damage during shipment. Replacement shipping containers can be purchased from Agilent.
WARNING
WARNING
Substituting parts or performing unauthorized modification to the instrument may result in a safety hazard.
Hydrogen is a flammable gas. Perform periodic leak tests to verify there are no leaks in the hydrogen lines and connections. Before making any connections, shut off the hydro­gen supply. Connect or cap all fittings at all times when using hydrogen.
40 Operation and Maintenance Manual

Step 1: Selecting a Location

The instrument should be placed on a clean, unobstructed surface approximately 22" (55 cm) deep by 10" (24 cm) wide that can support 44 pounds (19.9 kg) in addition to existing equipment. Figure 10 illustrates the relationship between the major system components. To facilitate proper heat dissipation, an additional 1-2" (2.5-5.0 cm) should be available at the rear and on both sides of the instrument. Distance between the SCD and the GC is limited by the length of gas and heater lines.
Position the Controller on top of the Detector box or in another convenient location near both the Detector and GC. The Controller requires supplies of hydrogen and oxidant (air for the SCD and oxygen for the NCD). The length of the gas inlet lines limit the distance between the Controller and the Burner to approximately 1 meter.
There are two pump options for the Detector, each with different size requirements. The Edwards RV5 oil-sealed vacuum pump requires a 6.3" (16cm) by 18.5" (47cm) area on a shelf or a nearby floor and a minimum height clearance of 22" (58 cm). The pump weight is 52 lbs (24 kg). The Dry Piston Pump (oil-free) requires a 9" (22 cm) by 14" (35 cm) area on a shelf or a nearby floor and a minimum height clearance of 12" (30 cm). The pump weight is 30 lbs (13.6 kg). The distance to the 355 Detector for both pump options is limited by the power cord connection—6 ft (2 m) for the oil-free piston pump or 8 ft (2.4 m) for the oil-sealed vacuum pump.
NOTE
Consider placing the oil-sealed pump over a plastic or metal container to capture any oil leaks or spills.

Power Requirements

See Chapter 2 for voltage and power requirements. Grounded outlets are required.
Operation and Maintenance Manual 41
Figure 10 Drawing of the Detector with Dual Plasma Burner and Controller

Environmental Considerations

The instrument should be operated in an environment which is comfortable for human habitation with reasonably constant temperature and humidity. Operation of the instrument at elevated temperatures (>30 °C) may result in an increased background noise from the photomultiplier tube.

Combustion Gas Requirements

The Agilent 355 SCD and 255 NCD both require a hydrogen source; the 355 SCD requires an air source, and the 255 NCD requires an oxygen source. Oxygen may be used for the ozone generator in either Detector to obtain a modest increase in sensitivity; the Dual Plasma Burner was designed for use
42 Operation and Maintenance Manual
with air as the oxidant in the SCD and oxygen as the oxidant in the NCD.
For the SCD, detector gases must be sulfur free (<1ppb) for proper Detector operation. In general, bottled air is preferred to “house” air from a compressor because compressors tend to generate lower quality air. In addition, pressure fluctuations induced by the air generator may be detrimental to Detector performance.
For the NCD and GC, use gases that are low in nitrogen and total hydrocarbons. Use gases with an "instrument" or "chromatographic" purity rating. Agilent recommends gases with a purity range of 99.995% to 99.9995%. Do not use oil-pumped supplied air since it contains large concentrations of hydrocarbons.
Use of appropriate traps on carrier, hydrogen and oxidant sources to improve support gas quality, such as a sulfur trap gas purifier, is recommended. The oxidant used to supply the ozone generator should be dry to prevent internal corrosion. Use of a moisture trap or dryer is recommended.
Two-stage pressure regulators
The combustion gases must be supplied to the Controller at a pressure of 25 psig or less. Use a two-stage pressure regulator rather than a single-stage regulator to eliminate pressure surges. High-quality, stainless-steel diaphragm-type regulators are recommended. On/Off valves at the regulators are useful but are not essential. Mount the valves at the outlet fitting of the pressure regulators.
WARNING
If you use gases with a two-stage regulator up-stream, it will be sufficient to use single-stage regulators at the Detector.
Particle Filtration
Up-stream, inline particle filtration of 10 micron or better is required to prevent damage to Controller and Burner components.
Ozone is a hazardous gas and a strong oxidant. Minimize exposure to ozone by using the instrument in a well-ventilated area and venting the exhaust of the vacuum pump to a fume hood. Turn off the ozone generator when the instrument is not in use.
Supply Tubing for Combustion Gases
Use preconditioned and cleaned copper or stainless steel tubing to supply gases to the 355 SCD and 255 NCD. Do not use ordinary copper tubing since it contains oils and other contaminants. Plastic tubing is not recommended since it is permeable to oxygen and other contaminants that can damage columns and detectors or cause elevated background. Teflon tubing may be acceptable in some clean environments.
Operation and Maintenance Manual 43
WARNING
Secure all gas cylinders to an immovable structure or permanent wall. Store compressed gases in accordance with all safety codes.
WARNING
Wear eye protection when using compressed gas to avoid possible eye injury.
Venting Gases
During normal operation of the GC with the SCD, NCD, FID, other detectors and split/splitless inlet purge, some of the carrier gas and sample vents outside the GC. In addition, the vacuum pump will vent a small amount of ozone and other combustion products. If the components of the sample are toxic or noxious, vent the exhaust from the GC outlets to a fume hood. Also, vent the exhaust from the vacuum pump to a fume hood.
44 Operation and Maintenance Manual

Step 2: Unpack and Inspect the Instrument

Before unpacking boxes, inspect them for signs of physical damage. If damage is observed, photographs should be taken in order to make a claim with the carrier should any equipment damage be found. Check contents of boxes against the shipping documents. Contact Agilent as soon as possible should a discrepancy be found.

Required Installation Tools

The following tools are recommended for installation of both the 355 SCD and 255 NCD:
One adjustable wrench, with adjustment up to 1 inch (2.5 cm)
One 1/4"-5/16" open-end wrench
Two 3/8"-7/16" open-end wrenches
Two 3/4"-9/16" open-end wrenches
One 5/8" open-end wrench
One 1/4" Nutdriver with hollow shaft (for FID restrictor)
One Phillips head screwdriver
One pair cotton gloves (for hand protection)
One small flat-head screwdriver
Operation and Maintenance Manual 45

Step 3: Set Up the Vacuum Pump

Initial connections require access to the rear of the Detector and the vacuum pump. Follow the installation instructions for either the RV5 Edwards pump or the oil-free dry piston pump, depending on your configuration.
WARNING
Exhaust gases from the pump should be vented to a fume hood to eliminate any potential hazard.

Installing the Edwards RV5 Pump Oil-Sealed Vacuum Pump

WARNING
WARNING
The normal surface temperature of the pump body at ultimate vacuum (operation) at ambient temperature of 20 °C is 50 °C to 70 °C. If you use the pump at a high ambient tem­perature, the temperature of the pump body may exceed 70 °C, and you must fit suitable guards to prevent contact with hot surfaces. For more information refer to the pump oper­ating manual.
Do not operate the vacuum pump with the oil level below the minimum oil level mark or above the maximum oil level mark.
1 Install the chemical trap mounting bracket to the top of the pump with the
screws and provided wrench. The chemical trap is a white plastic cylinder approximately 1 foot (30 cm) in length by 1.5 inches in (3.8 cm) diameter fitted with a barbed fitting on both ends (see Figure 11).
2 Remove the protective cap from the pump inlet port.
3 Install the metal conical screen and black centering o-ring into the pump
inlet port (see Figure 13).
4 Attach the aluminum 1/2-inch barbed adapter to the pump inlet port with
the metal clamping ring.
5 Slide the 2 inches Tygon vacuum hose over the pump inlet barbed adapter
and secure with a hose clamp.
6 Slide another hose clamp over the 2 inches Tygon vacuum hose.
7 Remove the plastic caps from the barb fittings on the ends of the chemical
trap.
8 Press the trap into the mounting bracket and the 2 inches of Tygon vacuum
hose (see Figure 11 and Figure 12).
9 Tighten the hose clamp.
10 Locate the Oil Return Line Kit.
46 Operation and Maintenance Manual
11
Remove the drain plug and the bonded seal from the oil mist filter. The bonded seal looks like a metal washer with a black inner o-ring.
12 Install the bonded seal to the oil mist filter drain adapter. The drain adapter
looks like a drain plug with a small plastic nozzle.
13 Screw the black drain adapter and bonded seal into the oil mist filter.
14 Remove the plastic protective cover from the pump exhaust port. Place the
centering o-ring on the pump exhaust port (see Figure 13).
15 Place the oil mist filter onto the o-ring on the pump exhaust port. Install the
filter so that the gray half is above the white half of the filter (see Figure 11 and Figure 12).
16 Position the oil mist filter so the drain adapter points toward the gas ballast
inlet (see Figure 11 and Figure 12).
17 Fit the clamping ring onto the adapter and oil mist filter and hand tighten.
18 Turn the gray plastic gas ballast knob counterclockwise to position II. Press
the knob down against the spring and continue to turn counterclockwise until the knob is free. Remove knob from the pump. Confirm that the spring is still in place.
19 Locate the tall aluminum gas ballast control assembly, and install the small
o-ring into the groove on the shaft.
WARNING
20 Insert the gas ballast control assembly into the pump, pressing down
against the spring, and turn it clockwise until the nozzle on the assembly is directly over the mark on the top of the pump.
21 Cut approximately 3/4 of the black silicone oil return line and insert the
steel restrictor approximately half-way into the line. If needed, use a small screwdriver or other small tool to aid in positioning the restrictor.
22 Fit one end of the oil return line to the drain adapter on the oil mist filter.
Fit the other end of the line to the nozzle on the gas ballast adapter. Ensure that the tubing is not tight and has no tight bends. Secure the line at each end, using the black hose clips provided.
23 Add oil to the pump via either of the two oil fill caps. The oil level should be
between one-third and one-half when viewed in the oil sight glass. Replace the oil fill cap prior to operation of the pump.
Do not operate the vacuum pump with the oil level below the minimum oil level mark or above the maximum oil level mark.
24 Place a hose clamp over the black heat shrink end of the 6’ Tygon vacuum
hose, and connect the hose to the barbed fitting labeled Exhaust on the back of the Detector. Tighten the hose clamp securely.
25 The SCD should be placed near the GC and be accessible from the rear in
order to connect the electrical power and the recorder signal cable.
Operation and Maintenance Manual 47
26
The vacuum pump should be placed within approximately 3 feet of the Detector (elevation not important), in order to connect the vacuum hose from the back of the Detector to the chemical trap. A hose clamp should be placed over the hose, and the hose should be connected to the straight end of the chemical trap. Tighten the hose clamp securely.
WARNING
WARNING
Vent the exhaust gas from the vacuum pump to fume hood to eliminate any potential haz­ard.
27 Place the vacuum pump in an adequately ventilated area or connect an
exhaust line (not provided) to the outlet located at the top of the mist filter. Attach an aluminum 1/2-inch barbed adapter to the outlet with the metal clamping ring and another centering o-ring. Secure the exhaust line with a hose clamp (not provided) to the aluminum adapter. Route the exhaust line to a fume hood or other suitable discharge location.
28 Shape the exhaust line to capture excess water (see Figure 14). Do not allow
water to condense in the exhaust line and drip back into the coalescing filter. Alternatively, install a water trap using a vacuum flask after the coalescing filter to capture excess water and prevent water from dripping back into the coalescing filter.
It is imperative that water does not condense in the exhaust line and fall back into the mist filter. It is recommended that transparent 3/8 inch (0.95 cm) id tubing be used as an exhaust line.
48 Operation and Maintenance Manual
t
tor
Chem ical
nlet
O F
Ch em ical Trap
r
Trap
Ballast control
O il return line
M ounting Bracket
Po wer sw itc h
EDWA RDS
Edwards
Figure 11 RV5 Oil-Sealed Vacuum Pump and Associated Traps (Front Side)
il Mist
ilter
B a llast Contro l
O il Retu rn Line
Barbed V acuum I
Oil mis filter
M ode selec
EDW ARDS
5
Powe Plug
Figure 12 RV5 Oil-Sealed Vacuum Pump and Associated Traps (Back Side)
Operation and Maintenance Manual 49
Pum p exhaust
Gas ballast
Pump inlet
O il f ill c aps
Inlet & valve
R e tractable hand le
Chemical
Trap
Figure 13 RV5 Oil- Sealed Vacuum Pump and Associated Traps (Top)
NOTE
Sometimes water condensation and accumulation are visible in the exhaust line. This is normal. However, do not allow water to continue to accumulate after approximately one week of operation. Significant water accumulation may indicate improper pump operation, an improperly vented exhaust line, or a Burner leak. Water accumulation in the exhaust line can cause damage to the pump, especially if allowed to fall back into the pump. Contact Agilent for advice if water accumulation continues to occur.
50 Operation and Maintenance Manual
Figure 14 RV-5 Oil-S ealed Vacuum Pump Exhaust Line
Setting the Gas Ballast Position (RV5)
Set the mode selector halfway between the High Vacuum mode, the small 6 symbol (see Figure 14) and the High Throughput mode. Do not set the mode selector to the High Throughput mode, the large
The RV5 vacuum pump and the Oil Drain Kit with ballast flow control ensure the vacuum pump operates continuously with a gas ballast flow. The purpose of the ballast control is to sweep ambient air into the pump oil. The air purges the water (created from the combustion of air and hydrogen in the Burner) and the oil (vaporized by the pump) into the oil coalescing filter. The filter separates the oil from the water, vents the air and water, and returns the oil to the vacuum pump.
The Oil Drain Kit with the ballast control continuously returns trapped oil in the oil mist filter to the vacuum pump. This feature reduces oil loss from the pump and minimizes the need to refill the pump with oil.
The Oil Drain Kit with the ballast control supplied is configured so that the gas ballast flow rate is equivalent to that with the gas ballast control on the pump in position II. For most applications, there is no need to change the gas ballast flow rate. If required, the gas ballast flow can be adjusted using the following procedure.
6 symbol.
1 The restrictor plate on top of the aluminum ballast control has three
screws. Remove the three screws that secure the restrictor plate. Do not dismantle the assembly (see Figure 16).
2 The restrictor plate has circular indentations. The position of the
indentations with respect to the indentation on the side of the oil return assembly identifies the gas ballast flow setting. Turn the restrictor to the required position:
Operation and Maintenance Manual 51
To select no gas ballast flow (not recommended), turn the restrictor
plate so that none of the indentations are aligned with the indentation on the side of the oil return assembly.
To select low gas ballast flow, turn the restrictor plate so that the single
indentation on the restrictor plate is aligned with the indentation on the side of the oil return assembly.
To select high gas ballast flow (the setting the ballast control is shipped
from the factory), turn the restrictor plate so that the two indentations are aligned with the indentation on the side of the oil return assembly.

Installing the Welch Dry Piston Vacuum Pump

The Welch Dry Piston Pump may be used as a direct replacement for an oil-sealed vacuum pump for the NCD or SCD. This pump produces all the advantages of an oil-free pump with little or no loss in instrument performance, however, operating Detector and Burner pressures will typically be a few Torr higher than those obtained with the oil-sealed pump.
To install the dry piston vacuum pump, follow these steps:
1 Remove the pump from its packaging and place on an accessible work
surface with the power cord side nearest to you and the pump "INLET" port to your right. Refer to Figure 15 as a visual aid to these instructions.
2 Verify that the voltage of the pump matches that of the Detector.
3 Open the plastic bag containing the Dry Piston Pump Kit. Locate the brass
elbow and brass barb (3/8" NPT). Remove their plastic protective caps and wrap three or four turns of Teflon tape (not supplied) onto the threaded connections. Screw the barb fitting into the elbow and tighten using wrenches. Similarly, remove the plastic cap from the pump's inlet fitting and screw elbow into it. Tighten the elbow so that the barb ends up pointed parallel to the floor and pointed toward the power cord side of the pump.
4 Locate the mounting brackets, 4 small screws and washers and spring clips.
Use 2 screws and washers to attach the spring clips, one to each bracket.
5 Using the supplied Torx wrench, remove two Torx screws from the pump,
the right corner screw located closest to you and the closest screw to you located immediately to the left of the pump's handle.
6 Attach the brackets to these locations using the Torx screws as showing in
Figure 15.
7 Attach the short piece of ½" ID hose tubing supplied in the Dry Piston
Pump Kit to the brass barb and use one of the supplied hose clamps to tighten it onto the barb.
8 Place a second hose clamp over the other end of the plastic tube. Insert the
barb on the elbow end of a Chemical Trap into the open end of the short plastic hose that was just attached to the brass barb on the pump.
9 Rotate the Chemical Trap so that it is retained by the spring clips.
52 Operation and Maintenance Manual
10
Tighten the hose clamp located on the plastic tube and the elbow barb end of the Chemical Trap.
11 Remove the plastic cap from the inlet end of the Chemical Trap. Taking the
clear end of the vacuum hose and another hose clamp attach the vacuum hose to the inlet end of the Chemical Trap and tighten the hose clamp.
12 Attach the black end of the vacuum hose to the barb fitting located on the
back of the Detector box and tighten the last hose clamp onto it to make this connection.
13 Using three or four wraps of Teflon® tape around NPT fitting of the
supplied muff ler or user supplied exhaust fitting to the pump's "EXHAUST" port. Any exhaust tubing attached to the pump exhaust should be positioned to avoid the accumulation of water from condensation of water vapor.
WARNING
Vent the exhaust from the vacuum pump to a fume hood or other exhaust system to elimi­nate any potential hazard.
14 Place the vacuum pump where it will be located when the pump is
operated, within the length of the power cord to be attached to the female connector on the back of the Detector.
15 Make sure that the On/Off switch on the pump is in the on position.
Operation Notes (Welch Pump)
1 The pump is designed to start against atmospheric pressure. Therefore, if
the pump is turned off inadvertently, it may be necessary to allow the Detector pressure to rise before restarting the pump.
2 The only user maintenance required is the periodic replacement of the
piston sleeves and seals, e.g., after 8 months of continuous operation. Consult the pump’s manual for further details.
3 Although, the pump is not damaged by exposure to ozone, use of the
Chemical Trap is highly recommended. Refer to the pump manual and seal replacement kit instructions for more detailed information.
4 High line voltage will cause the pump to overheat and trip its thermal fuse.
5 When new, or after changing seals, the pump may require a short break-in
period (2-4 hours).
Operation and Maintenance Manual 53
Figure 15 The Welch Dry Piston Pump
R
Screws
estrictor Plate
Indentation
(Alignment Guide)
Indentation (Low Ballast)
Indentation (High Ballast)
Figure 16 Oil Drain Kit with Ballast Control
Turn the switch on the vacuum pump to the On position.
6
54 Operation and Maintenance Manual
Note the position of the oil level in the window after operating the pump for several hours. For the next several days of operation, recheck the oil level daily. If the oil level is increasing, water is accumulating in the oil reservoir. Confirm that the water vapor is properly being expelled from the oil mist filter. If the oil level is dropping, excess air flow through the pump is forcing oil out of the vacuum pump. Turn the restrictor plate so that the single indentation on the restrictor plate is aligned with the indentation on the side of the oil return assembly. After adjustment of the gas ballast, allow the system to operate for an additional day and check the oil level again.
It is normal for the oil to appear foamy from air in the oil when viewing the oil in the oil level window. The purpose of the ballast control is to sweep ambient air into the pump oil. Ensure the oil level when the pump is operating is not above the “Full” mark on the pump.
Operation and Maintenance Manual 55

Step 4: Connect the Power Cord

Connect the pump power cord to the female socket on the back of the Detector (see Figure 17). The pump has an On/Off switch located on the electrical motor and this switch should be turned On. Do not connect to the AC power supply at this point in the installation procedure.
AIR INLET
RECORDER
OUTPUT
FUSE 250 VOLTS
10 AMPS
Voltage Label
10V
1V
100mV
VACUUM
PUMP
EXHAUST
Figure 17 SCD (230 V Unit) Rear Panel Diagram
56 Operation and Maintenance Manual

Step 5: Install the Dry Compressed Air or O2 Supply

Connect a 1/8" OD Teflon (PFA) line fitted with a 1/8" brass Swagelok nut from the AIR INLET at the rear of the Detector (see Figure 17) to a supply of dry compressed air or oxygen. The air regulator located inside the front door of the Detector should be set to approximately 3-6 psi.
Operation and Maintenance Manual 57

Step 6: Install the Signal Output Cables

Signal output cables are available from Agilent as standard equipment and can be used with most data systems. Confirm that the output cable supplied is correct for your system. A standard cable fitted with two crimp lug connectors is supplied for use with most integrators, recorders or data systems. Attach the BNC connector end of the recorder cable to the matching output connector, labeled RECORDER OUTPUT, on the back of the SCD (see
Figure 17).

Standard Cable Connection

The standard recorder cable is connected to the integrator by installing the red crimp-lug connector to the signal terminal (+) and the black crimp-lug connector to the negative terminal (-). No additional ground connection is required.

HP 3390 Series Integrator Cable Connection

The keyed-edge connector on the HP 3390 series cable is attached to the signal input of HP 3390 series integrators. Note that the connector can only be installed one way.

HP 3396 Integrator Cable Connection

The jack plug on the HP 3396 cable is attached to the analog signal input connector at the rear of the integrator. This cable also works with the Agilent 35900 controller.

HP 5890 GC Analog Input Board

A cable made specifically for this input board is available. This board is used when the SCD signal is input to ChemStation.

Agilent 6890 GC Analog Input Board

A cable made specifically for this input board is available. This board is used when the SCD signal is input to ChemStation.
58 Operation and Maintenance Manual

Step 7: Install the Dual Plasma Burner

Remove the cover plates from the Detector area of the GC to expose the hole into the oven through which a Detector is normally mounted. If the GC has more than one available Detector position, pick the most convenient one.
Prepare the GC by cutting the inside liner and top liner per Figure 17. Note that mounting fastener patterns will vary by GC manufacturer. Make sure the notch in the inside liner is on the right, in order to accommodate the geometry of the Burner as it sits in the shroud and mounting plate.
Figure 18 Dimensions of GC Liner Cut-Outs
Align the Burner mounting plate with the mounting screw holes on the GC. Clear the hole into the GC oven of interfering insulation, and then secure the mounting plate onto the oven with the screws provided. Attach the Burner's heated base connector to the GC's temperature control circuit. Consult your GC's operation or service manual to confirm proper connection of the 100 ohm RTD sensor and the cartridge heaters.
NOTE
Occasionally geometric design changes will occur within or between GC models. If the mounting plate provided does not match up with the top of the GC, contact Agilent for additional information.
Position the Dual Plasma Burner (column end down) into the tapered fitting of the heated base, with the lower hydrogen line and pin aligned with the slot in the heated base. The Burner should be secure when properly positioned.
Operation and Maintenance Manual 59

Step 8: Install the Dual Plasma Controller

Position the Controller such that the gas lines from the Burner can be easily attached to the back of the Controller. Connect the Controller to both a hydrogen source and an oxidant source, per Step 1. Connect the gas supplies to the 1/8" bulkhead unions marked "Oxidizer Inlet" and "Hydrogen Inlet." Clean copper tubing (1/8" OD) is recommended.
Connect the two 1/16" gas delivery lines (provided) from the Burner to the appropriate fitting on the rear of the Controller marked Oxidizer or Hydrogen in the Outlets (to Burner) area. Use a 1/4" open-end wrench to tighten the nut onto the bulkhead unions, using a 9/16" wrench to back-up the union.
Connect the yellow thermocouple plug from the Burner to the thermocouple jack at the rear of the Controller. The thermocouple connector will only fit one way, with the iron terminal (the one with the two grooves) down. Connect the Burner heater line to the two pin locking connector on the back of the Controller.
The Controller can be used for either 100/115 or 230 VAC. The correct voltage for your Controller is set at the factory; however, this should be verified upon installation. The voltage selection is shown through a small window on the power entry module (see Figure 19). If the voltage setting is incorrect, check to make sure that the proper fuses are installed. Contact Agilent for a replacement fuse set.
WARNING
Do not change the voltage selector without changing the main power fuse. Operating the Controller box without the proper fuse may damage the electronics.
Figure 19 Dual Plasma Controller Rear Panel
60 Operation and Maintenance Manual

Step 9: Install Column Connections

The Burner operates under reduced pressure and there will be a slight vacuum on the end of the column. If a higher outlet pressure for the column outlet is desired, fused silica capillary restrictors may be attached to the end of the analytical column (both capillary and packed) prior to making the Detector connection.

Capillary Columns

Place the column nut over the end of the capillary column. Place the appropriate fused silica adapter ferrule onto column. Remove a few centimeters from the end to remove any particles that may have entered the column. Insert the GC column into the Burner by 108-109 mm from the upper end of the nut (114-115 mm if measured from the flat bottom of the nut). Do not force the column. The pathway is narrow and may take several tries to seat it correctly. Tighten the column nut finger tight, or until sealed using a 7/16" open-end wrench to back-up the hydrogen inlet fitting to prevent it from slipping.
108-109 mm
114-115 mm
Figure 20 Measuring Column Insertion

Packed Columns and Columns with an Outside Diameter > 0.8 mm

Connect a short piece (0.5 m or less) of deactivated fused silica tube, for example 0.32 mm internal diameter, to the Detector end of the column. Follow the procedure for capillary column connection as above.
Operation and Maintenance Manual 61

Step 10: Install the Transfer Line

Connect the black transfer line (extending from the side of the Detector) to the top connector on the Burner and tighten with a 3/8" open-end and 7/16" open-end wrench (backing up the union on top of the Burner to prevent its position form slipping).
62 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
4 Front Panel Controls and Initial Startup
Detector Front Panel Controls 64
Detector Interface Setup 69
Agilent Technologies
63

Detector Front Panel Controls

As illustrated in Figure 21, the front panel is divided into three sections: signal control, display output control, and power control. Each section is described below. All the front panel LEDs are red in the ON mode and darkened in the OFF mode except for POWER. This LED toggles green for ON and red for STANDBY. In STANDBY mode, the front panel display and high voltage to the photomultiplier tube (PMT) are turned off. The red LED in the STANDBY mode serves as a reminder that power is still supplied to the instrument.
Figure 21 Front Panel Controls

Power Controls

ON: Turns on front panel displays and high voltage to the PMT. STANDBY: Darkens panel and turns off high voltage to the PMT. Does not shut off main power.
Turns pump ON (red) and OFF (dark). Functions in both ON and STANDBY modes.
64 Operation and Maintenance Manual

Display Output Controls

Turns ozone generator ON or OFF. Functions in both ON and STANDBY modes if the pump is producing sufficient vacuum.
Controls the range of the front panel display between high sensitivity (0.1-200mV) and low sensitivity (0-2V). Does not affect recorder output signal.
Displays the output signal of the SCD in millivolts. (Note: Changing the output range of the SCD by adjusting the recorder switch on the back panel does not change the range of the display.)

Signal Controls

The SCD and NCD use analog amplifier circuitry for the measurement of the current produced by the photomultiplier tube. Two Front Panel Controls are used to adjust the output signal of the Detector.
Displays the pressure of the chemiluminescence reaction chamber in Torr (mm Hg)
Changes the gain of the amplifier by a factor of
100. For high sensitivity measurement of sulfur or nitrogen compounds (measurement of low levels of sulfur or nitrogen compounds), the attenuation should be operated in the 1 position. For samples containing high levels (ppm to percent) of sulfur or nitrogen compounds, use the 100 position.
Operation and Maintenance Manual 65

Dual Plasma Controller Controls

The Agilent Dual Plasma Controller provides easy access to basic settings.
The baseline signal from the SCD and NCD can be adjusted from 0 to ±1% of the full scale recorder output using the offset control. (e.g., +10 mV to
-10 mV for a 1 V full scale setting). This control can be used to offset the background signal.
Figure 22 Dual Plasma Controller Front Panel
Oxidizer and Hydrogen Control Knobs
Temperature Control Knob
Selector Control Knob Changes the display to show the current value or
66 Operation and Maintenance Manual
Allow you to adjust the oxidizer and hydrogen flow rate.
Allows you to adjust the Burner temperature.
set point for each setting (temperature in °C, pressure in Torr, oxidizer in sccm, or hydrogen in sccm).
Controller Status Lights

Initial Startup

Vacuum Test
With the 1/8" Valco cap on the black PFA transfer line still in place and the air supply to the ozone generator off, set the internal air regulator to 0 psi (fully counterclockwise), and power on the SCD or NCD by plugging the power cord into the house power supply. The SCD and NCD will power up in STANDBY mode, with the pump and the ozone generator OFF. For this test, the desired mode is power ON, vacuum pump ON, ozone generator OFF, display range in the low sensitivity mode and PRESSURE signal displayed. Press the following controls in the order shown in order to start the vacuum pump and to view the reaction cell pressure.
Illuminate to reflect current status. The Power LED indicates power is on; the Heater LED indicates temperature set point has been obtained; the Va lves LED indicates the solenoid valves open; and the Fault LED indicates a drop in pressure or excessive temperature (see Chapter 7, “Troubleshooting” for details).
Press until green ON LED is lit
Press until red LED is lit
Press to display reaction cell pressure in torr
After the pressure has stabilized ( Typical pressure in the reaction cell should be 1-2 torr with the oil-sealed
pump, or 5-8 torr for the oil-free pump.
5 minutes), record the pressure below.
~
Operation and Maintenance Manual 67
Reaction cell pressure, with air to ozone generator OFF (0 psi setting on internal regulator) and transfer line capped: ___________________ torr
Tighten connections if necessary and check to make sure pressure stabilizes in the expected region. If proper pressure is not obtained contact Agilent for assistance. If the reaction cell pressure is within the expected range, record the value, reset the internal air regulator to 3-6 psi, turn the pump OFF and proceed with the recorder test.
Recorder Test
The standard 355 SCD and 255 NCD recorder output configuration is 1 volt full scale. In addition, an output range can be set with the recorder output selection switch located on the back panel (Figure 11). Use a small screwdriver to adjust the switch to the proper position for your integrator or data system.
To check that the recorder cable has been properly connected to the integrator or data system, set the integrator to a high sensitivity setting (e.g. attenuation of x 1) and plot the background signal. Use the Output Offset to decrease the Detector baseline. If the recorder cable is connected correctly, the baseline will shift in response to changes in the Output Offset. If the polarity of the connection is incorrect, a negative response will be observed when the baseline is increased, and vice versa. Switch the polarity of the wires to correct this problem. If no response is observed, re-check the signal connections and repeat the test. Contact Agilent if there is no data system response after completion of this test.
68 Operation and Maintenance Manual

Detector Interface Setup

Initial Checkout

Careful attention to eliminating leaks in the Detector interface will lead to better Detector sensitivity and easier troubleshooting if problems develop.
1 Check that the gas connections have been made correctly and that they are
tight.
2 Plug the 3-prong connector for power on the Controller into a 100-volt,
50/60 Hz, 115-volt, 50/60-Hz or 230-volt, 50/60 Hz AC outlet.
3 Connect the oxidant delivery line to the oxidant inlet nut on the Burner.
Leave the heater cable disconnected and the thermocouple plugged in.
4 Connect the black PFA transfer line from the Detector to the top of the
Burner (see Figure 1 and/or Figure 2).
5 Turn on the Controller, the Detector power, and the vacuum pump.
WARNING
6 Let the system pump down to a stable pressure (
7 Read the pressure from the LCD display on the Controller. It should be <30
5 minutes).
~
torr, with higher pressures observed for packed or megabore columns due to the higher column flow. Record this pressure for reference.
Pressure of Burner with column flow but without hydrogen and oxidant flow to Burner:_________ torr
The reaction cell pressure on the Detector should be less than 5 torr (10 with oil-free pump).
8 If pressure readings are not within the above specifications, then there are
leaks in the system. Leak check all positive pressure points (swage nuts, tees, crosses) with SNOOP or a leak Detector. Tighten fittings if necessary and recheck the pressure. Confirm that the Burner fittings (oxidant inlet nut, tee, and hydrogen nut) are strongly hand-tightened. If necessary, tighten the gas delivery lines and transfer line. If pressure readings are still high, contact Agilent for service.
Do not overtighten! Overtightening the fittings on the ceramic tubes can cause the tubes to break. Vespel ferrules that are greatly distorted by overtightening cannot be used again.
9 Turn off the Controller and the vacuum pump.
10 Reconnect and/or turn on the hydrogen and oxidant supplies to the
Controller.
Adjust the hydrogen and oxidant inlet pressures to 25 psig (1.7 bar).
Refer to Table 1 on page 75 for typical operating conditions.
Operation and Maintenance Manual 69

Monitoring Oxidizer and Hydrogen Flow with the Dual Plasma Controller

Hydrogen will flow to the Burner only if the temperature is above 325 °C and the pressure of the Burner is <575 torr.
Turn ON the vacuum pump at the Detector and the power to the Dual Plasma Controller. The power LED should illuminate. When the temperature reaches approximately 325 °C, the valve’s LED should illuminate, indicating hydrogen and oxidant flow.
Use the Controller display knobs to select display of oxidizer or hydrogen flow.
WARNING
WARNING
Hydrogen gas is explosive and must be handled carefully. Keep away from sources of ignition.
The Burner is hot. Do not touch the Burner. Let the Burner cool before performing any operations involving the Burner.
70 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
5 Operation
Start-Up Procedure 72
Detector Operation 73 Typical Operating Conditions 75 Detection Limits 76
Instrument Shut-Down 77 Special Operating Modes 78
Agilent Technologies
71

Start-Up Procedure

1 Turn on the GC and set the carrier flow rate.
2 Turn on the hydrogen and oxygen to the Dual Plasma Controller.
3 Turn the Detector from STANDBY to ON.
4 Press the PUMP button until the red LED illuminates. Make sure the
vacuum pump is running.
5 Turn on the oxygen to the ozone generator and set the regulator inside the
front door of the Detector to between 3 and 6 psig.
6 Allow the system to evacuate for approximately 1 minute before turning on
the ozone at the front panel. A properly functioning system will show a background signal deflection between the ozone on and ozone off.
7 Turn ON the Dual Plasma Controller using the ON/OFF switch on the rear
panel of the Controller. As the Burner heats, the SCD or NCD background will increase and then slowly decrease. The hydrogen and oxidant flow will automatically start when the temperature of the Burner rises above approximately 325 °C.
8 Check the LCD display of Dual Plasma Controller to confirm that all
pressure, temperature, and signal responses are within the desired specifications.
WARNING
The Burner is hot. Do not touch the Burner. Let the Burner cool before performing any operations involving the Burner.
72 Operation and Maintenance Manual

Detector Operation

Detector Stability and Response

The time required for system stabilization varies depending on the application, system cleanliness, presence of active sites and other factors. Useful results could be generated within 30 minutes of start-up, especially with a previously operated system. A longer stabilization time is likely to be required upon changing critical system components, such as the combustion tubes or the GC column. In addition, gas flow rates may drift initially as thermal equilibrium is reached due to changes in gas viscosity with temperature. Therefore, it is good practice to monitor gas flow rates and adjust them accordingly.
Even though a system may not be fully stabilized, sample injections can be made within minutes of instrument start-up. Whether the results are useful largely depends on application. Typically, an elevated baseline will initially be observed, which will diminish upon successive programmed runs.
After stabilization has been reached, the system should exhibit good short-term and long-term precision. Of course this also depends on the application and concentration of components being measured. As an example, analysis of thiophene in benzene at the 1 ppm sulfur level yielded 1.4% RSD (n=10) over about 2 hours and 3.6% RSD (n=42) over about 96 hours. As expected, carbon disulfide at a lower concentration of 90 ppb sulfur yielded
2.6% and 10.4% RSD, respectively.

Column Bleed

Accumulation of column bleed causes silicon dioxide to build up in the Burner. This silicon dioxide creates active sites that are detrimental to performance. In many cases, the choice of column can be optimized for a particular application. Column bleed can be minimized by the use of oxygen traps on the carrier gas, low-bleed columns, and lowest possible maximum oven temperature.

Coking

Contamination from some sample matrices can reduce sensitivity. Crude oils containing volatile metal complexes may contaminate ceramic tubes. The incomplete combustion of certain hydrocarbon-containing compounds leaves behind coke deposits on the tubes. Coke deposits may be removed from the Burner by reducing the hydrogen flow rate. The Dual Plasma Burner is much less susceptible to coke formation than other designs.

Hydrogen Poisoning

Hydrogen poisoning of the ceramic tubes occurs when there is no oxidizer flow through the ceramic tubes. The result is extremely reduced, or no response. Hydrogen poisoned tubes can not be reconditioned and should be discarded.
Operation and Maintenance Manual 73

Contaminated Gases

The use of clean gases for the 355 SCD is essential for optimal performance. High purity gases (99.999% pure or better) are advised. Sulfur and other contaminants from gases may accumulate in the column and bleed out over time desensitizing the tubes and causing elevated baselines. The use of sulfur traps is highly recommended for all gases.

Fluctuating Pressures

Fluctuations in pressure, especially from gas generators, will affect Detector response. It is therefore recommended that only bottled gases equipped with a dual stage regulator or appropriate steps to ensure stable pressure supplies are used.
74 Operation and Maintenance Manual

Typical Operating Conditions

The Controller is calibrated at the factory for flow rates to deliver gas in sccm units. The following table summarizes the typical operating conditions:
Table 1 Typical Operating Conditions
Condition SCD NCD
Detector Pressure (Torr) 4-8 (6-12
*
)
4-10
Dual Plasma Controller Pressure (Torr)
Burner Temperature (°C) 800 900-950
Hydrogen Flow Rate (sccm) 40-50 4-6
Oxidant Flow Rate (sccm) 60-65 (air) 8-12 (oxygen)
Background Noise (mV) 0.3-2.0 0.3-1.0
* Oil-free pump
Thermocouple lifetime at 950 °C is diminished.
300-400 100-250
The recommended conditions should yield satisfactory results for most applications and should be used as a typical starting point for any method development. Like any detector, however, there are optimum conditions which may very somewhat from the recommended conditions. In optimization of conditions for the Dual Plasma Burner and Controller the following guidelines should be considered:
High flow rates of hydrogen and oxidant can release enough heat at high temperature to vaporize combustion tubes and cause blockages downstream where the materials condense. High flow rates will eventually cause the pressure in the Burner to exceed its fault cut-off limit of about 600 Torr. For this reason, do not exceed the recommended flow rate by more than about 25%.
A higher hydrogen to oxidant ratio may initially show higher response but later yield a reduced response because of the accumulation of contaminants, such as soot or other active species, that reduce the Detector response.
Operating the Burner at higher temperatures will place more demand on the heater, thermocouple and seal materials, effectively shortening their lifetime.
In general, when making any parameter change, keep in mind that the system may require time to reach equilibrium.
Operation and Maintenance Manual 75

Detection Limits

The following table lists the detection limits which can be expected for typical chromatographic conditions, assuming proper operation of the Detector and chromatographic systems.
Table 2 Expected Detection Limits for Chromatographic Conditions
Type of Injection Volume Column Detection Limit Per
Liquid, Split 1:10 1 µL Capillary 0.32 mm ID 50 parts per billion
Liquid, Split 1:100 1 µL Capillary 0.32 mm ID 0.5 parts per million
Liquid, On-column or splitless 1 µL Capillary 0.32 mm ID 5 parts per billion
Liquid, Splitless 10 µL Capillary 0.32 mm ID 0.5 parts per billion
Gas, Split 1:10 1 cm
Gas, Direct on-column 1 cm
Gas, Direct on-column 1 cm
Gas, Direct on-column 10 cm
Compound as Sulfur
3
3
3
3
Capillary 0.32 mm ID 50 parts per billion
0.53 mm ID 5 parts per billion
Packed 5 parts per billion
Packed 0.5 parts per billion
76 Operation and Maintenance Manual

Instrument Shut-Down

Daily Shutdown

1 Toggle off the ozone generator.
2 Turn off the air regulator (counter-clockwise), located inside Detector door.
3 Toggle power to "stand-by."
4 Leave the vacuum pump and Dual Plasma Controller operating at all times.

Complete Shutdown

1 Toggle off the ozone generator.
2 Turn off the air regulator (counter-clockwise), located inside Detector door.
3 Toggle power to STANDBY.
4 Turn off power to the Dual Plasma Controller.
5 After 15 minutes, toggle off the vacuum pump, so the Burner cools and
moisture is removed from the system.
6 Turn off gases to unit.
Operation and Maintenance Manual 77

Special Operating Modes

Using the 255 NCD in Nitrosamine Mode

By default, the 255 NCD is configured to detect nitrogen. To change from nitrogen to nitrosamine mode, first turn off and unplug the Controller. Remove the cover, find jumper P6 and the positions labelled High Setpoint and Low Setpoint located on the printed circuit board near the left front of the Controller. Move the jumper position from the High Setpoint to the Low Setpoint position. This changes the temperature control range to 350-500 °C to be used for nitrosamine analysis. Refer to see Figure 9 on page 37 for a schematic drawing of the 255 NCD in nitrosamine mode.
Hydrogen is not used in the nitrosamine mode. Turn off and disconnect the hydrogen inlet gas. Using the supplied tee, couple the oxidant outlet with the two Burner inlets. Use of oxygen is recommended at 5-10 mL/min because the Burner can easily be cleaned by raising the Burner temperature to about 900-1000 °C in flowing oxygen. Alternatively, helium or argon can be used, however, these gases will not permit in situ cleaning at elevated temperature.
The presence of catalyst in the quartz combustion tube will generally yield the highest sensitivity for nitrosamines. However, for some sample types that contain potentially interfering nitrogen compounds, such as nicotine, it may be desirable to remove the catalyst to obtain better selectivity. This is achieved by removing the quartz combustion tube and using a straight 1/16" rod or tube to push the catalyst out of the tube (save the catalyst for reuse or recovery as it contains 90+% platinum). Replace the combustion tube and optimize the burner temperature as desired (in general it is desirable to increase the pyrolysis temperature by 50-100 °C when the catalyst is not used). Refer to the
“Tube Replacement for the NCD” on page 102 for additional instruments with
regard to tube removal and replacement.

Using the SCD in High Sensitivity Mode for Nonhydrocarbon Gaseous Samples

There are circumstances in which it may be desirable to operate the SCD in a non-typical manner. These could include the analysis of very low levels of sulfur species in a nonhydrocarbon gas matrix, such as helium, carbon dioxide, or even hydrogen. It should be possible to measure low ppb or high ppt levels of sulfur species.
In the case of a hydrogen matrix, the sample matrix itself can suffice as the lower source of hydrogen for Dual Plasma operation in a non-chromatographic mode. For instance, a flow rate of nominally 20 SCCM of the sample, a hydrogen calibration gas and a “clean” hydrogen source can be alternately introduced to the Burner. The sample and calibration gas would be introduced into the normal column connection. The side port of the splitter fitting would be plugged and the clean hydrogen source would be plumbed to the lower hydrogen inlet port. The upper hydrogen and air flow rates would be adjusted to nominally 30 and 65 SCCM, respectively.
In the case of nonhydrocarbon gaseous samples, the potential for coking is not a concern, so a Dual Plasma is not necessary. Some improvement in sensitivity
78 Operation and Maintenance Manual
can be achieved by eliminating the lower plasma. This is accomplished by plugging the side port of the splitter fitting and teeing the air into the lower hydrogen line, much like the configuration used for Nitrosamine analysis as shown in Figure 9. A slightly lower air flow rate, such as 40 SCCM and a hydrogen flow rate of around 60 SCCM is used. Sensitivity of around 0.1 pg S/sec or less should be readily achievable. It is not possible to cover all potential applications, so some optimization may be desirable.
Operation and Maintenance Manual 79
80 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
6 Maintenance
Pump Maintenance 82
Cleaning the Detector 83 Changing the Oil Mist Filter (RV5) 84 Reaction Cell Cleaning 85
Flow Sensor Calibration 88 Detector Sensitivity 89 Assembling the Dual Plasma Burner for Component Replacement with the
SCD 90
Assembling the Dual Plasma Burner for Component Replacement with the
NCD 95
Tube Replacement for the SCD 98 Tube Replacement for the NCD 102
Agilent Technologies
81

Pump Maintenance

To maintain optimum performance of the Agilent 355 SCD and 255 NCD, routine replacement of the chemical trap (for ozone destruction), oil coalescing filter and oil (Edwards oil-sealed pump only) is necessary. Refer to
Table 3 for the expected life span of each replacement part or material.
It is beneficial to keep a maintenance log that tracks when maintenance is performed and any instrument or operational changes that might impact performance. Also keep track of detector flow rates (oxidizer and hydrogen), pressures (burner controller and reaction cell in detector), and background signal (the difference between ozone “on” and ozone “off”).
Table 3 Operating Life of Components for Edwards RV5 Vacuum Pump
Component/Material Operating life
Chemical Trap (RV5) ~ 3 month
Oil Coalescing Filter (RV5) ~ 3 months
Pump oil
* The operating life is based on the total time logged during operation of the Detector with the Burner and the
† Pump oil can be purchased from a supplier or directly from Agilent: SAE 10W-30, Multiviscosity Synthetic
ozone generator ON.
Motor Oil such as, MOBIL 1 or AMSOIL.
~ 3 months
*
82 Operation and Maintenance Manual

Cleaning the Detector

You can clean the external housing of the Detector with a damp cloth using water or non-abrasive cleaners. Turn off power to the Detector and disconnect it from main power prior to cleaning. Do not spray liquids directly on the Detector. Wipe dry with a clean, soft cloth.
No cleaning agents which could cause a hazard as a result of reaction with the Burner unit are to be used in cleaning the instrument.
Contact Agilent to address concerns about the compatibility of specific cleaning agents with the Burner unit.
Operation and Maintenance Manual 83

Changing the Oil Mist Filter (RV5)

The oil mist filter on the RV5 pump has two components: the charcoal odor filter and the oil coalescing filter element. To replace the filters, disassemble the oil mist filter assembly with the 4 mm long-handled allen wrench (provided). The smaller charcoal odor filter sits on top of the larger oil coalescing filter element. It is recommended to replace the oil coalescing filter element after 90 days of continuous use, however replacement of the charcoal odor filter is optional. After replacing the filter, re-assemble the filter assembly and attach it to the pump flange.
84 Operation and Maintenance Manual

Reaction Cell Cleaning

Over time, the reaction cell and UV pass filter (SCD) or IR pass filter (NCD) will develop a build-up of material which should be removed for optimum sensitivity. The cleaning schedule depends upon Detector use and the nature of the analyses; however, it is recommended the cell should be cleaned annually. The cleaning procedure requires removal of the photomultiplier tube (PMT) from the Detector, and special precautions are required to prevent damage to the PMT.
Exposure of a photomultiplier tube to bright light can result in damage to the PMT, even when the high voltage is off. To avoid potential damage, minimize light exposure. The black PMT cover included in the Detector accessories package is recommended for this procedure. Carefully read all the instructions below before attempting to remove the PMT. If you have any questions regarding this procedure please contact Agilent.
To clean the cell and pass filter:
1 Disconnect the main power cord (if the vacuum pump is operated from an
independent power outlet, unplug the power to the pump as well).
WARNING
Any operation requiring access to the inside of the equipment, could result in injury. To avoid potentially dangerous shock, disconnect from power supply before opening the equipment.
2 Remove the right side panel from the Detector and disconnect the high
voltage and coaxial signal cables from the rear of the PMT housing (See
Figure 3 and Figure 15).
3 Turn off all room lights and minimize outside light sources.
4 Unscrew the PMT socket connector assembly. Carefully pull the socket
assembly and the PMT out toward the rear of the instrument at a slight angle to remove the PMT/socket assembly from the housing.
5 Immediately place the black PMT cap over the PMT, being careful to avoid
placing fingerprints on the PMT window. Do not remove the PMT from the socket assembly. After the PMT cap is in place, carefully place the PMT and socket assembly on a soft surface inside a drawer or other dark location to minimize exposure of the PMT to light. The room lights may be turned on at this point.
Operation and Maintenance Manual 85
PMT SOCKET
PMT HOUSING
FILTER
REACTION
CELL
Figure 23 Reaction Cell, PMT Housing and PMT Socket
PRESSURE
SENSOR
O-RINGS
TRANSFER
LINE
PMT
HIGH
VOLTAGE
to power supply
SIGNAL
to amplifier
PMT SOCKET
REDUCING
UNION
FROM OZONE
GENERATOR
Figure 24 Reaction Cell
6
Use a 7/64" Allen wrench to remove the three mounting screws from the reaction cell. Slowly pull the reaction cell back from the PMT housing. The optical filter is located between the reaction cell and the PMT housing. Remove the optical filter from the housing by tipping up the back of the Detector, if necessary, and allow the filter to fall onto a soft cloth.
7 Inspect the o-rings and replace them if they show any wear.
86 Operation and Maintenance Manual
8
Clean any deposits on the optical filter using a soft cloth or Kimwipe dampened with methanol or deionized (DI) water. Do not leave fingerprints or fibers on the cleaned filter. Deposits inside the reaction cell can be cleaned in the same manner, however, care must be taken to avoid bending the ozone inlet tube that extends into the cell.
9 Reseat the pass filter into the PMT housing. Confirm that the two o-rings
around the reaction cell are properly seated in the O-ring grooves (see
Figure 24). Carefully re-align the reaction cell to the PMT housing, making
sure the o-rings remain properly seated, and secure the cell to the housing with the three screws. (The two o-rings must be correctly seated in order to obtain a vacuum-tight and light-tight seal.)
10 To re-install the PMT, minimize all light sources. Remove the PMT cap,
being careful not to touch the PMT window. Insert the PMT and the socket assembly into the PMT housing and screw in the socket assembly until it is seated tightly against the housing. The room lights may now be turned on.
11 Reconnect the PMT high voltage cable to the MHV connector (longer
connector) and the PMT signal cable to the BNC connector (shorter connector) at the back of the socket assembly (see Figure 23).
WARNING
The high voltage and signal cables must be attached to the proper connectors on the pmt socket; damage to the pmt will occur if the cables are not properly connected.
12 While the side panel is still removed, reconnect the AC power cord to the
Detector and switch the front panel power to ON. To ensure that the o-rings have been properly sealed, monitor the Detector baseline at "attenuation 1" for several minutes (with the room light ON), and after noting the baseline signal, turn the room lights OFF. If the baseline signal significantly decreases, the o-rings are incorrectly positioned and the PMT must be removed and the cell re-installed.
13 If no change in the baseline is observed, check the system for vacuum leaks
by turning on the vacuum pump and monitoring the pressure as described in Section 7. If the pressure readings are comparable to start-up values, the o-rings are properly positioned. If a significantly higher pressure is observed, the PMT must be removed and the cell re-installed.
Operation and Maintenance Manual 87

Flow Sensor Calibration

The hydrogen and oxidant flow sensors installed in the Dual Plasma Controller have very good repeatability, but significant non-linearity. They have each been factory calibrated at the midrange flow rate (50 SCCM) using an NIST traceable flow meter. Over the typical operating range for the SCD, the flow sensors should produce accuracy of better than ±10% of reading. However, due to sensor non-linearity at low or high flow extremes, error greater than this could be observed.
If greater accuracy is desired at a particular flow rate range, it is possible to re-calibrate the flow sensors. A reference flow meter, appropriate gases, and a trim-pot adjustment tool are required for this procedure. With the controller on and no gases flowing, the zero calibration points are set with RP6 and RP4 for the hydrogen and oxidant flow streams, respectively. Connect a reference flow meter to the gas outlet(s) on the back of the controller. Connect pressurized gas line(s) to the inlets on the back of the controller. Connect jumper (JP1) to bypass the pressure fault circuitry. Allow the burner temperature to exceed 325 °C in order for the hydrogen and oxidant valves to be open. The spans of the sensors are then adjusted for the particular flow stream to the desired flow rate as measured with the reference flow meter. The span adjustments are set with RP3 and RP2, for hydrogen and oxidant respectively, so that the display on the controller matches the value obtained with the reference flow meter. Upon completion, remove the reference flow meter and return the bypass jumper (JP1) to its original position.
88 Operation and Maintenance Manual

Detector Sensitivity

)
Detector sensitivity is an indicator reflecting the performance characteristics of a given system, and is a useful tool to determine when Detector maintenance is warranted. Sensitivity is typically reported as a minimum detection limit (MDL) as calculated from the following formula:
()/'(66.0
MDL
=
Where Am't N/S (amount of nitrogen or sulfur) is the mass of nitrogen or sulfur in picograms that reaches the Detector, PktoPkNoise (peak-to-peak noise) is the measure of the noise (e.g. in mV), Signal is the height of the peak in the same units, and Wd1/2ht is the width of the peak at half height in seconds. The constant 0.66 is used in the calculation assuming the MDL S/N =
3.29.
Before it is released from the factory, each 355 SCD must pass an MDL level of <0.5 pg Sulfur/second and each 255 NCD must pass an MDL level of <3.0 pg Nitrogen/second. The response from individual detection systems may vary by a factor of 2 or 3; however, it is typical for Detectors to perform in the 0.1-0.3 pg S/second range for the 355 SCD and in the 1-2 pg N/second for the 255 NCD when tested at Agilent.
××
SignalWd
×
ht1/2
NoisePktoPkSNtAm
)()(
Operation and Maintenance Manual 89

Assembling the Dual Plasma Burner for Component Replacement with the SCD

The following procedure can be used to assemble the Dual Plasma Burner for use with the SCD or for replacement of Burner components, such as ceramic tubes. Refer to Figure 1 on page 26 for proper part nomenclature.
1 Slide the 0.066" internal diameter (I.D.) double taper ferrule onto the lower
Burner tube. The tube should extend approximately 2 mm past the end of the ferrule.
2 Insert the lower Burner tube and double taper ferrule into the Burner inlet
fitting.
3 Slide the 1/4" Burner adapter over the top of the lower Burner tube all the
way down to the Burner inlet fitting and screw it onto the fitting finger-tight.
Figure 25 Ferrule Placement on Lower Burner Tube
4 Slide a 1/4" Swagelok nut over the Burner adapter and then slide a 1/4"
ferrule over the Burner adapter and position it into the 1/4" nut. Note: If a graphite ferrule is used, a small amount of shavings may be created and some graphite will be left on the tube surface; this is normal. Avoid allowing any shavings to fall inside a tube.
5 Center the lower Burner tube so that it will slide into the tapered union.
Insert the lower end of the tapered union fitting into the 1/4" Swagelok nut and screw it on finger-tight. If necessary, the brazed H bent out of the way, however, be careful not to stress the brazed (welded) connection.
6 Insert the large ceramic tube into the quartz heater assembly. Position a
1/4" ferrule (flat end butted up against the top of the swivel nut) onto the large ceramic tube. With the ferrule positioned against the swivel nut, approximately 0.5 cm of the large ceramic tube should extend outside of the nut. Insert the lower Burner tube into the center of the large ceramic tube and finger tighten the heater swivel nut onto the tapered union fitting.
line can be gently
2
90 Operation and Maintenance Manual
Figure 26 Proper Ferrule Orientation to the Large Ceramic Tube
Figure 27 Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly
7
Position the upper ceramic tube into the long axis of the splitter fitting so that about 4 mm of it extends past the top of the fitting. Slide the 0.054" ID double tapered ferrule over the upper ceramic tube. Gently holding these parts so that neither the ferrule nor the upper ceramic tube slips out of position, finger-tighten the union fitting onto the splitter fitting.
Operation and Maintenance Manual 91
Figure 28 Orientation of the Double Taper Ferrule
Figure 29 Positioning the Upper Tube in the Union Fitting
Approximately 1.5 cm of the large ceramic tube should extend above the top
8
of the quartz heater assembly. Slide a 1/4" Swagelok nut over the large ceramic tube and then also slide a 1/4" ferrule over the tube (flat side on back of the nut).
9 Holding the splitter fitting, gently insert the upper ceramic tube into the
large ceramic tube coaxially, to avoid placing stress on the fragile upper ceramic tube. Lower the splitter fitting into place to engage the threads of the 1/4" Swagelok nut. Tighten finger-tight.
10 To begin the final alignment and tightening, use a 7/16" wrench and 5/16"
wrench to tighten the 1/4" Burner adapter one-quarter turn past finger-tight.
11 Using a 5/8" wrench on the heater swivel nut and a 1/2" wrench on one of
the flats of the tapered union fitting, tighten the heater swivel nut one-quarter turn past finger-tight. Using a 5/16" wrench on the 1/4" Burner
92 Operation and Maintenance Manual
adapter, rotate this fitting so that the brazed H2 line is aligned 180 ° (opposite) from the oxidizer Inlet port.
Figure 30 Tightening the Heater Swivel Nut
12 Making sure that the Burner inlet fitting does not loosen, use a 1/2" wrench
on a flat of the tapered union fitting and 9/16" wrench on the 1/4" Swagelok nut of the Burner adapter to tighten the tapered union fitting 1/4" past finger-tight.
13 Rotate the quartz heater assembly so that the thermocouple and heater
leads are in the same plane and pointed in the same direction as the peg on the Burner inlet fitting. Turn the splitter fitting so that H2 inlet port is also aligned with the peg on the Burner inlet fitting.
Figure 31 Proper Alignment of the Burner
14 Tighten the nut on the splitter fitting one-quarter turn past finger-tight
using a 9/16" wrench on the 1/4" nut and a 7/16" wrench on the flats of the splitter fitting.
Operation and Maintenance Manual 93
15
Carefully bend the H2 line into position so that the 1/16" Valco nut and ferrule can be screwed into the side port of the splitter fitting. Tighten the connection of the H
line to the splitter fitting using a 3/8" wrench on the
2
vertical flat of the splitter fitting and 1/4" wrench on the Valco nut.
16 Make sure that no other connections have loosened or moved out of
alignment, if so, reposition or retighten the fittings as needed.
17 The assembled Burner is now ready for re-installation on the GC.
94 Operation and Maintenance Manual

Assembling the Dual Plasma Burner for Component Replacement with the NCD

The following procedure can be used to assemble the Dual Plasma Burner for use with the NCD or for replacement of Burner components, such as ceramic tubes. Refer to Figure 2 on page 27 for proper par t nomenclature. Note that the lower section of the NCD Burner is identical to the lower section of the SCD Burner.
NOTE
While the use of 1/4" ferrules is optional with the SCD, the quartz tube used in the NCD Burner is more fragile and the use of 1/4" graphite ferrules is highly recommended.
1 Slide the 0.066" internal diameter (I.D.) double taper ferrule onto the lower
Burner tube. The tube should extend approximately 2 mm past the end of the ferrule.
2 Insert the lower Burner tube and double taper ferrule into the Burner inlet
fitting.
3 Slide the 1/4" Burner adapter over the top of the lower Burner tube all the
way down to the Burner inlet fitting and screw it onto the fitting finger-tight.
Figure 32 Ferrule Placement on Lower Burner Tube
4 Slide a 1/4" Swagelok nut over the Burner adapter and then slide a 1/4"
ferrule over the Burner adapter and position it into the 1/4" nut.
5 Center the lower Burner tube so that it will slide into the tapered union.
Insert the lower end of the tapered union fitting into the 1/4" Swagelok nut and screw it on finger-tight. If necessary, the brazed H bent out of the way. Be careful not to stress the brazed (welded) connection.
6 Find the quartz combustion tube, 1/4" ferrule and quartz heater assembly.
7 Insert the quartz tube into the quartz heater assembly. Position a 1/4"
ferrule (flat end butted up against the top of the swivel nut) onto the quartz combustion tube. With the ferrule positioned against the swivel nut, approximately 0.5 cm of the quartz tube should extend outside the nut.
Operation and Maintenance Manual 95
line can be gently
2
Figure 33 Proper Ferrule Orientation to the Large Quartz Tube
Figure 34 The Quartz Tube Properly Inserted into the Quartz Heater Assembly
Insert the lower burner tube into the center of the quartz tube and finger
8
tighten the heater swivel nut onto the tapered union fitting, then tighten an additional 1/4 turn making sure not to break the quartz tube.
9 To begin the final tightening, use a 7/16" wrench and 5/16" wrench to
tighten the 1/4" Burner adapter one-quarter turn past finger-tight.
96 Operation and Maintenance Manual
10
Using a 5/8" wrench on the heater swivel nut and a 1/2" wrench on one of the flats of the tapered union fitting, tighten the heater swivel nut one-quarter turn past finger-tight. Using a 5/16" wrench on the ¼" Burner adapter, rotate this fitting so that the brazed H
line is aligned 180 °
2
(opposite) from the oxidizer Inlet port.
11 Making sure that the Burner inlet fitting does not loosen, use a 1/2" wrench
on a flat of the tapered union fitting and 9/16" wrench on the 1/4" Swagelok nut of the Burner adapter to tighten the tapered union fitting 1/4" past finger-tight.
12 Rotate the quartz heater assembly so that the thermocouple and heater
leads are in the same plane and pointed in the same direction as the peg on the Burner inlet fitting. Turn the splitter fitting so that H
inlet port is also
2
aligned with the peg on the Burner inlet fitting.
Figure 35 Burner Assembly Detail
Figure 36 Burner Assembly Alignment
13 The assembled Burner is now ready for re-installation on the GC.
Operation and Maintenance Manual 97

Tube Replacement for the SCD

The Burner has two combustion tubes that require replacement: the upper ceramic tube and the Large Ceramic Tube. Generally the tubes should be replaced only if sensitivity decreases. The “Troubleshooting” chapter provides additional information to assist in determining whether tube replacement may be necessary. Refer to Figure 1 on page 26 for proper part nomenclature. Follow the instructions below for tube replacement.
1 Turn off power to the GC and the Controller and let the system cool down
under vacuum.
2 Turn off power to the vacuum pump.
3 Lift the Burner out of the shroud. It is recommended to remove the coil,
noting the position of the ferrule. In some instances, if the coil can be uncoiled, it may be convenient to leave it attached to the Burner.
4 Disconnect the hydrogen and oxidant lines.
5 Disconnect the power connector that leads to the GC, if necessary.
6 Tilt the Burner at an angle, so that when loosening the union fitting the
upper ceramic tube does not slide down into the large ceramic tube.
7 Loosen and disconnect the union fitting, and pull the splitter fitting and
upper ceramic tube out of the Burner.
8 Slide the upper ceramic tube out of the splitter fitting.
9 Slide the upper ceramic tube into the splitter fitting, so that approximately
4 mm of the tube extends beyond the top of the fitting. Then, slide the double taper ferrule onto the tube (see Figure 37 and Figure 38 for proper positioning). Gently holding these parts so that neither the ferrule nor the upper ceramic tube slip out of position, finger-tighten the union fitting onto the splitter fitting.
Figure 37 Orientation of the Double Taper Ferrule
98 Operation and Maintenance Manual
Figure 38 Positioning the Upper Tube in the Union Fitting
Holding the splitter fitting, gently insert the upper ceramic tube into the
10
large ceramic tube coaxially, to avoid placing stress on the fragile upper ceramic tube. Lower the splitter fitting into place to engage the threads of the 1/4" Swagelok nut. Tighten finger-tight.
11 If you do not need to replace the Large Ceramic Tube, proceed to step 19.
12 Remove the Tapered Union Fitting from the bottom of the Burner.
13 Slide the Large Ceramic Tube out of the Burner and remove it from the
Quartz Heater Assembly.
14 Insert the new large ceramic tube into the quartz heater assembly. Position
a 1/4" ferrule (flat end butted up against the top of the swivel nut) onto the large ceramic tube. With the ferrule positioned against the swivel nut, approximately 0.5 cm of the large ceramic tube should extend outside of the nut. Insert the lower Burner tube into the center of the large ceramic tube and finger tighten the heater swivel nut onto the tapered union fitting.
Figure 39 Proper Ferrule Orientation to the Large Ceramic Tube
Operation and Maintenance Manual 99
Figure 40 Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly
15
Approximately 1.5 cm of the large ceramic tube should extend above the top of the quartz heater assembly. Slide a 1/4" Swagelok nut over the large ceramic tube and then also slide a 1/4" ferrule over the tube (flat side on back of the nut).
16 To begin the final alignment and tightening, use a 7/16" wrench and 5/16"
wrench to tighten the 1/4" Burner adapter one-quarter turn past finger-tight.
17 Using a 5/8" wrench on the heater swivel nut and a 1/2" wrench on one of
the flats of the tapered union fitting, tighten the heater swivel nut one-quarter turn past finger-tight. Using a 5/16" wrench on the 1/4" Burner adapter, rotate this fitting so that the brazed H
line is aligned 180 °
2
(opposite) from the oxidizer Inlet port.
Figure 41 Tightening the Heater Swivel Nut
18 Making sure that the Burner inlet fitting does not loosen, use a 1/2" wrench
on a flat of the tapered union fitting and 9/16" wrench on the 1/4" Swagelok
100 Operation and Maintenance Manual
Loading...