No part of this manual may be reproduced
in any form or by any means (including
electronic storage and retrieval or translation into a foreign language) without prior
agreement and written consent from Agilent Technologies, Inc. as governed by
United States and international copyright
laws.
Acknowledgements
Torx® is a registered trademark of Textron
Inc.; Teflon® is a registered trademark of
E.I. du Pont de Nemours and Company.
Kimwipe® is a registered trademark of the
Kimberly-Clark Corporation.
Manual Part Number
G6600-90006
Edition
First edition, June 2007
Replaces G6600-90002
Printed in USA
Agilent Technologies, Inc.
2850 Centerville Road
Wilmington, DE 19808-1610 USA
Warranty
The material contained in this document is provided “as is,” and is subject to being changed, without notice,
in future editions. Further, to the maximum extent permitted by applicable
law, Agilent disclaims all warranties,
either express or implied, with regard
to this manual and any information
contained herein, including but not
limited to the implied warranties of
merchantability and fitness for a particular purpose. Agilent shall not be
liable for errors or for incidental or
consequential damages in connection
with the furnishing, use, or performance of this document or of any
information contained herein. Should
Agilent and the user have a separate
written agreement with warranty
terms covering the material in this
document that conflict with these
terms, the warranty terms in the separate agreement shall control.
Technology Licenses
The hardware and/or software described in
this document are furnished under a
license and may be used or copied only in
accordance with the terms of such license.
Restricted Rights Legend
defined in FAR 2.101(a) or as “Restricted
computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent
agency regulation or contract clause. Use,
duplication or disclosure of Software is
subject to Agilent Technologies’ standard
commercial license terms, and non-DOD
Departments and Agencies of the U.S. Government will receive no greater than
Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than
Limited Rights as defined in FAR 52.227-14
(June 1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in any
technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like
that, if not correctly performed or
adhered to, could result in damage
to the product or loss of important
data. Do not proceed beyond a
CAUTION notice until the indicated
conditions are fully understood and
met.
WARNING
A WARNING notice denotes a
hazard. It calls attention to an
operating procedure, practice, or
the like that, if not correctly performed or adhered to, could result
in personal injury or death. Do not
proceed beyond a WARNING
notice until the indicated conditions are fully understood and met.
If software is for use in the performance of
a U.S. Government prime contract or subcontract, Software is delivered and
licensed as “Commercial computer software” as defined in DFAR 252.227-7014
(June 1995), or as a “commercial item” as
Warnings
English
WARNING
WARNING
WARNING
WARNING
This symbol on the instrument indicates that the user should refer to the manual for operating instructions.
Any operation requiring access to the inside of the equipment, could result in
injury. To avoid potentially dangerous shock, disconnect from power supply
before opening the equipment.
For continued protection against fire hazard replace fuse with same type and
rating.
This symbol indicates that to comply with European Union Directive
2002/96/EC for waste electrical and electronic equipment (WEEE), the Analyzer should be disposed of separately from standard waste.
WARNING
WARNING
WARNING
Operation and Maintenance Manual3
This is a safety Class I product. It must be wired to a mains supply with a protective
earthing ground incorporated into the power cord. Any interruption of the protective
conductor, inside or outside the equipment, is likely to make the instrument dangerous.
Intentional interruption is prohibited.
If this instrument is used in a manner not specified by Agilent, the protection provided by
the instrument may be impaired.
High voltage is present in the instrument when the power cord is connected, even if the
main power switch is in the standby mode. To avoid potentially dangerous shock, disconnect the power cord before removing the side panels.
WARNING
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized
by using the instrument in a well-ventilated area and by venting the exhaust of the vacuum pump to a fume hood. The ozone generator should be turned off when the instrument
is not in use.
WARNING
WARNING
WARNING
WARNING
Burner temperature Is extremely hot. Do not touch. Allow to cool before servicing.
Hydrogen is an extremely flammable gas. Use appropriate care when handling. Inspect
all connections with a suitable leak detector.
Oxygen rich environments can promote combustion and even result in spontaneous combustion under conditions of high pressure and exposure to contamination. Use only oxygen rated components and ensure that components are oxygen clean prior to use with
pure oxygen.
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxidant sensors or burst their connective tubing.
4Operation and Maintenance Manual
Español
WARNING
WARNING
WARNING
WARNING
Cualquier operación que requiera acceso al interior del equipo, puede causar
una lesión. Para evitar peligros potenciales, desconectarlo de la alimentación
a red antes de abrir el equipo.
Para protección contínua contra el peligro de fuego, sustituir el fusible por
uno del mismo tipo y características.
Este símbolo en el instrumento indica que el usuario debería referirse al manual para instrucciones de funcionamiento.
Esto es un producto con clase I de seguridad. Debe conectarse a una red que disponga de
tierra protectora en el cable de red. Cualquier interrupción del conductor protector,
dentro o fuera del equipo, puede ser peligroso. Se prohibe la interrupción intencionada.
WARNING
Si este instrumento se usa de una forma no especificada por Agilent, puede desactivarse
la protección suministrada por el instrumento.
Operation and Maintenance Manual5
Français
WARNING
WARNING
WARNING
WARNING
Chaque opération à l'intérieur de l'appareil, peut causer du préjudice. Afin
d'éviter un shock qui pourrait être dangereux, disconnectez l'appareil du
réseau avant de l'ouvrir.
Afin de protéger l'appareil continuellement contre l'incendie, échangez le fusible par un fusible du même type et valeur.
Le symbol indique que l'utilisateur doit consulter le manuel d'instructions.
Ceci est un produit de Classe de sécurité I. L'instrument doit être branché sur
l'alimentation secteur par un fil de secteur prévu d'une prise de masse. Chaque
interruption du conducteur protégeant, à l'intérieur ou á l'extérieur de l'appareil peut
rendre l'instrument dangereux. Interruption intentionnelle est interdite.
WARNING
Si l'instrument n'est pas utilisé suivant les instructions de Agilent, les dispositions de
sécurité de l'appareil ne sont plus valables.
6Operation and Maintenance Manual
Deutsch
WARNING
WARNING
WARNING
WARNING
Vor dem Öffnen des Gerätes Netzstecker ziehen!
Für kontinuierlichen Schutz gegen Brandgefahr dürfen bei Sicherungswechsel nur Sicherungen der gleichen Stärke verwendet werden!
Dieses Symbol auf dem Gerät weist darauf hin, dass der Anwender zuerst das
entsprechende Kapitel in der Bedienungsanleitung lesen sollte.
Dies ist ein Gerät der Sicherheitsklasse I und darf nur mit einem Netzkabel mit
Schutzleiter betrieben werden. Jede Unterbrechung des Schutzleiters au erhalb oder
innerhalb des Gerätes kann das Gerät elektrisch gefährlich machen. Absichtliches
Unterbrechen des Schutzleiters ist ausdrücklich verboten.
WARNING
Wenn das Gerät nicht wie durch die Firma Agilent, vorgeschrieben und im Handbuch
beschrieben betrieben wird, können die im Gerät eingebauten Schutzvorrichtungen
beeinträchtigt werden.
Operation and Maintenance Manual7
Italiano
WARNING
WARNING
WARNING
WARNING
Qualsiasi intervento debba essere effettuato sullo strumento può essere
potenzialmente pericoloso a causa della corrente elettrica. Il cavo di alimentazione deve essere staccato dallo strumento prima della sua apertura.
Per la protezione da rischi da incendio in seguito a corto circuito, sostituire I
fusibili di protezione con quelli dello stesso tipo e caratteristiche.
Il simbolo sullo strumento avverte l'utilizzatore di consultare il Manuale di
Istruzioni alla sezione specifica.
Questo strumento è conforme alle specifiche per I prodotti in Classe I - Il cavo di
alimentazione dalla rete deve essere munito di "terra". Qualsiasi interruzione del cavo di
terra all'interno ed all'esterno dello strumento potrebbe risultare pericolòsa. Sono
proibite interruzioni intenzionali.
WARNING
Se questo strumento viene utilizzato in maniera non conforme alle specifiche di Agilent,
le protezioni di cui esso è dotato potrebbero essere alterate.
8Operation and Maintenance Manual
Dutch
WARNING
WARNING
WARNING
WARNING
Iedere handeling binnenin het toestel kan beschadiging veroorzaken. Om iedere mogelijk gevaarlijke shock te vermijden moet de aansluiting met het net
verbroken worden, vóór het openen van het toestel.
Voor een continue bescherming tegen brandgevaar, vervang de zekering door
een zekering van hetzelfde type en waarde.
Het symbool geeft aan dat de gebruiker de instructies in de handleiding moet
raadplegen.
Dit is een produkt van veiligheidsklasse I. Het toestel moet aangesloten worden op het
net via een geaard netsnoer. Bij onderbreking van de beschermende geleider, aan de
binnenzijde of aan de buitenzijde van het toestel, kan gebruik het toestel gevaarlijk
maken. Opzettelijke onderbreking is verboden.
WARNING
Indien het toestel niet gebruikt wordt volgens de richtlijnen van Agilent, gelden de
veiligheidsvoorzieningen niet meer.
Operation and Maintenance Manual9
The following symbols are used on the equipment:
Caution - Refer to manual for
operating instructions
Atención - Ver documentación
pertinente.
Attention - Consultez les
ocuments d'accomagnement.
Vorsicht - Siehe beiliegende
Unterlagen.
Pericolo - Vedi
documentazione allegata.
Voorzichtig - Raadpleeg di
bijehorende documentatie.
Caution - Risk of electrical
shock.
Atención - Riesgo de sacudidas
eléctricas.
Attention - Risque de choc
électrique.
Vorsicht - Risiko eines
Elektroschocks.
Pericolo - Pericolo di scosse
elettriche.
Voorzichtig - Hoge spanning,
levensgevaar.
Caution - Hot surface.
Atención - Superficie caliente.
Attention - Surface brûlante.
Vorsicht - Heisse Oberfläche.
Pericolo - Superficie rovente.
Voorzichtig - Heet oppervlak.
10Operation and Maintenance Manual
Contents
1 Introduction
2 System Description
Specifications 20
Technical Information — 355 SCD 20
Technical Information — 255 NCD 20
Physical Specifications 21
Dual Plasma Burner 25
Figure 1. Cross-Section of the Dual Plasma Burner for the 355 SCD 26
Figure 2. Cross-Section of the Dual Plasma Burner for the 255 NCD 27
Dual Plasma Controller 27
Ozone Generator 28
Chemiluminescence Reaction Cell and Photomultiplier Tube 28
Figure 3. 355 SCD Left Side 29
Figure 4. 355 SCD Right Side 30
Figure 5. 255 NCD Left Side View 31
Figure 6. 255 NCD Right Side View 32
Pressure Transducer 32
Vacuum Pump 32
Chemical Trap 33
Oil Coalescing Filter 33
FID Adapter (Optional) 33
NCD Reaction Cell 33
NCD Photomultiplier Tube and Cooler 34
Figure 7. Schematic for 355 SCD 35
Figure 8. Schematic for 255 NCD, in Nitrogen Mode 36
Figure 9. Schematic for 255 NCD, in Nitrosamine Mode 37
Operation and Maintenance Manual11
3 Installation
Overview 40
Step 1: Selecting a Location 41
Power Requirements 41
Figure 10. Drawing of the Detector with Dual Plasma Burner and Controller 42
Environmental Considerations 42
Combustion Gas Requirements 42
Step 2: Unpack and Inspect the Instrument 45
Required Installation Tools 45
Step 3: Set Up the Vacuum Pump 46
Installing the Edwards RV5 Pump Oil-Sealed Vacuum Pump 46
Figure 11. RV5 Oil-Sealed Vacuum Pump and Associated Traps (Front Side) 49
Figure 12. RV5 Oil-Sealed Vacuum Pump and Associated Traps (Back Side) 49
Figure 13. RV5 Oil- Sealed Vacuum Pump and Associated Traps (Top) 50
Figure 14. RV-5 Oil-S ealed Vacuum Pump Exhaust Line 51
Installing the Welch Dry Piston Vacuum Pump 52
Figure 15. The Welch Dry Piston Pump 54
Figure 16. Oil Drain Kit with Ballast Control 54
Step 4: Connect the Power Cord 56
Figure 17. SCD Rear Panel Diagram 56
Step 5: Install the Dry Compressed Air or O
Supply 57
2
Step 6: Install the Signal Output Cables 58
Standard Cable Connection 58
HP 3390 Series Integrator Cable Connection 58
HP 3396 Integrator Cable Connection 58
HP 5890 GC Analog Input Board 58
Agilent 6890 GC Analog Input Board 58
Figure 25. Ferrule Placement on Lower Burner Tube 90
Figure 26. Proper Ferrule Orientation to the Large Ceramic Tube 91
Figure 27. Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly 91
Figure 28. Orientation of the Double Taper Ferrule 92
Figure 29. Positioning the Upper Tube in the Union Fitting 92
Figure 30. Tightening the Heater Swivel Nut 93
Figure 31. Proper Alignment of the Burner 93
Assembling the Dual Plasma Burner for Component Replacement with the NCD 95
Figure 32. Ferrule Placement on Lower Burner Tube 95
Figure 33. Proper Ferrule Orientation to the Large Quartz Tube 96
Figure 34. The Quartz Tube Properly Inserted into the Quartz Heater Assembly 96
Figure 35. Burner Assembly Detail 97
Figure 36. Burner Assembly Alignment 97
Tube Replacement for the SCD 98
Figure 37. Orientation of the Double Taper Ferrule 98
Figure 38. Positioning the Upper Tube in the Union Fitting 99
Figure 39. Proper Ferrule Orientation to the Large Ceramic Tube 99
Figure 40. Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly 100
Figure 41. Tightening the Heater Swivel Nut 100
Figure 42. Proper Alignment of the Burner 101
Tube Replacement for the NCD 102
Figure 43. Proper Ferrule Orientation to the Large Quartz Tube 102
Figure 44. Large Quartz Tube Properly Inserted into the Quartz Heater Assembly 103
Figure 45. NCD Tube Replacement Detail 104
Figure 46. NCD Tube Replacement Detail 104
14Operation and Maintenance Manual
7 Troubleshooting
Solving Detector Problems 106
Power Problems 107
Detector Fuse 107
Vacuum Pump Fuse 107
Dual Plasma Controller Fuses 107
Table 4. Fuses for 100 V, 120 V and 230 V Versions of 355 SCD and 255 NCD 108
Figure 47. Fuse Positions on the Power Supply Board 108
Ozone Generation Problems 109
Ozone Generator 109
High Voltage Transformer 109
Plugged Restrictor Lines 110
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors
Operation and Maintenance Manual
1
Introduction
This manual will guide you in the installation, operation, and troubleshooting
of the Agilent 355 Sulfur Chemiluminescence Detector (SCD) and the Agilent
255 Nitrogen Chemiluminescence Detector (NCD). This manual is intended for
use with the Agilent 355 SCD or 255 NCD with the Dual Plasma Burner and
Controller.
This operation and service manual has some important conventions, such as
the use of boxed warnings. This information is deliberately set out from the
text for emphasis and should be followed to assure operator safety and proper
instrumental operation.
If you are installing the 355 SCD or 255 NCD yourself, follow the installation
procedures described in the following sections. If your instrument is already
installed, turn to the Operation section to begin.
Agilent Technologies
17
18Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors
Operation and Maintenance Manual
2
System Description
Specifications 20
Theory of Operation 23
Description of Major Components 25
Agilent Technologies
19
Specifications
Technical Information — 355 SCD
Sensitivity
Typical Selectivityg S/g C > 2 x 10
Linearity>10
Precision and Stability
*
†
Typical < 0.5 pg S/second (signal to noise 3.3:1)
7
4
<2% RSD 2 hours
<5% RSD 72 hours
Ozone flow through the Post Ozone Restrictor20-30 mL/min at 3-6 psig
* Burner temperature 950 °C, 11 SCCM oxygen, and 6 SCCM hydrogen; 25 ppm N as nitrobenzene in toluene;
0.2 µL injection on column (HP 19095-121Z), n=7 for 3 hours; n=13 for 18 hours and n=10
n-dipropylnitrosamine in toluene at 4 µg/mL, 0.2 µL injection on column (HP 19095-121Z).
*
Physical Specifications
<1.5% RSD 8 hours (~ the same in N and
nitrosamine)
<2% RSD 18 hours (~3% RSD in nitrosamine
mode over 21 hours)
Dual Plasma Controller100-120 VAC, 50/60 Hz, 200 W
220-240 VAC, 50/60 Hz, 200 W
Dimensions and weight
DetectorHeight: 16.0 in (40.6 cm)
Width: 9.2 in (23.4 cm)
Depth: 21.8 in (55.3 cm)
355 SCD Weight: 34.0 lbs (15.0 kg)
255 NCD Weight: 37.5 lbs (17.0 kg)
Operation and Maintenance Manual21
Dual Plasma ControllerHeight: 5.0 in (12.7 cm)
Width: 9.5 in (24.1 cm)
Depth: 12.5 in (31.8 cm)
Weight: 9.9 lbs (4.5 kg)
Dual Plasma BurnerHeight: 12.3 in (31.2 cm)
Diameter: 4.0 in (10.2 cm)
Weight: 1.9 lbs (0.9 kg)
Oil Sealed Vacuum Pump (RV5)Height: 10.3 in (26.1 cm)
Width: 6.0 in (15.2 cm)
Depth: 16.9 in (43.0 cm)
Weight: 47.3 lbs (21.5 kg)
or
Oil Free Dry Piston PumpHeight: 12.0 in (30 cm)
Width: 9.0 in (22.9 cm)
Depth: 14.0 in (35.6 cm)
Weight: 29.9 lbs (13.6 kg)
Installation Category II
Pollution Degree2
Ambient Temperature50 - 104 °F (10-40 °C)
Relative HumidityUp to 95%, noncondensing
Normal Operating EnvironmentIntended for indoor use only
Maximum Altitude2,000 m (6,562 ft)
Mains Supply VoltageFluctuation not to exceed 10% of nominal
voltage
22Operation and Maintenance Manual
Theory of Operation
-
Sulfur Chemiluminescence Detector
The Agilent model 355 Sulfur Chemiluminescence Detector (SCD) is a
sulfur-selective detector for gas chromatography. Operation of the SCD is
based on the chemiluminescence (light-producing reaction) from the reaction
of ozone with sulfur monoxide (SO) produced from combustion of the analyte:
Sulfur compound (analyte)
SO + O
→ SO2 + O2 + hη (<300–400 nm)
3
→ SO + H
A vacuum pump pulls the combustion products into a reaction cell at low
pressure, where excess ozone is added. Light produced from the subsequent
reaction is optically filtered and detected with a blue-sensitive photomultiplier
tube and the signal is amplified for display or output to a data system. Figure 1
is a pneumatic flow diagram that illustrates how the components of the system
are integrated.
The Detector has an enclosed, dedicated (Dual Plasma) Burner designed to
enhance production of the SO intermediate. This Dual Plasma Burner mounts
in the detector port of the GC. A Dual Plasma Controller provides temperature
control and gas-flow regulation to operate the Dual Plasma Burner.
The Agilent model 355 SCD provides high sensitivity (<0.5 pg S/sec), with
linear and equimolar response over four orders of magnitude (per Sulfur
atom) while maintaining high selectivity over common solvents. The Agilent
SCD is compatible with most commercially available gas chromatographs.
Nitrogen Chemiluminescence Detector
Operation of the Agilent model 255 Nitrogen Chemiluminescence Detector is
based on the chemiluminescence or light-producing reaction of ozone with
nitric oxide formed from combustion. Reacting nitric oxide with ozone results
in the formation of electronically excited nitrogen dioxide. The excited
nitrogen dioxide emits light, a chemiluminescence reaction, in the red and
infrared region of the spectrum. The light emitted is directly proportionally to
the amount of nitrogen in the sample,
O + other products
2
NO + O
NO2 NO2 + hη (>800 nm)
3
The light (hη) emitted by the chemical reaction is optically filtered and
detected by a photomultiplier tube. The signal from the photomultiplier tube is
amplified for display or output to a data system. Organic compounds
containing nitrogen react to form nitric oxide, carbon dioxide, and water.
H
2/O2
--------------
R-N + O
Operation and Maintenance Manual23
→ NO + CO
2
Δ
+ H2O
2
Dual Plasma Controller
The Dual Plasma Controller provides all operational parameters of the Dual
Plasma Burner except for the Detector base temperature. The Detector base
temperature is controlled by circuitry in the GC. Parameters monitored or
regulated by the Controller include Burner temperature, Burner temperature
set-point, hydrogen and oxidant flow rates, and Burner pressure. The
temperature set-point, actual pressure, oxidant and hydrogen flow rates are
displayed by rotation of a 4-position switch. Power, valve operation,
temperature within set-point range and fault conditions are indicated with
LED illumination on the front display panel.
The Dual Plasma Controller incorporates several safety features. The safety
circuitry detects faults such as power loss, vacuum loss, thermocouple failure,
heater element failure, broken ceramic tube, or high temperature. When a fault
is detected, the Fault LED illuminates and hydrogen and oxidant flow is
stopped by normally-closed solenoid valves.
Dual Plasma Burner with the 355 SCD
The Dual Plasma Burner is based on the same chemistry and basic principles
of earlier SCD Burner designs. A key difference, however, is the addition of a
second “flame” or “plasma,” the lower is oxygen-rich and the upper is
hydrogen-rich. The Burner consists of a tower assembly that contains an outer
sheath for burn protection, a heating element, thermocouple, and combustion
tubes. Conversion of sulfur containing compounds to SO occurs within the
ceramic reaction chamber housed in the Burner assembly. Agilent also
provides a Flame Ionization Detector (FID) option for the simultaneous
collection of hydrocarbon and sulfur chromatograms for some GCs.
Dual Plasma Burner with the 255 NCD
Compounds eluted from the GC column are combusted in the Dual Plasma
Burner first by an oxygen rich flame (plasma) followed by catalytic combustion
on a Noble metal screen. For hydrocarbons, this two stage combustion
technique results in complete conversion of the matrix to products, such as
carbon dioxide and water, which do not chemiluminesce with ozone. Nitrogen
atoms in a compound are converted into nitric oxide and potentially other
nitrogen oxide species. In the second stage, the catalyst is used to convert
other nitrogen oxide species to nitric oxide, resulting in a high efficiency of
conversion.
24Operation and Maintenance Manual
Description of Major Components
Dual Plasma Burner
The Dual Plasma Burner consists of a tower assembly that contains an outer
sheath for burn protection, a heating element, thermocouple, and combustion
tubes. In the SCD, conversion of sulfur containing compounds to SO occurs
within the ceramic reaction chamber housed in the Burner assembly and
potentially interfering hydrocarbons are oxidized to CO
the oxidant. In the NCD, oxygen is used as the oxidant.
A fitting is located on top of the Burner. The vacuum line from the Detector
box is connected directly to the top of this fitting and H2 is input to the longer
side of this fitting. The air inlet is connected to the base of the Burner.
The Burner is mounted onto the GC by a model-specific mounting kit (see
www.Agilent.com/chem or contact Agilent for the most current information).
The GC column is inserted into the Burner using a 1/32" knurled nut and fused
silica adapter ferrule.
A cross-section illustration of the Dual Plasma Burner for the 355 SCD is
shown in Figure 1 and for the 255 NCD is shown in Figure 2.
and H2O, with air as
2
Operation and Maintenance Manual25
Figure 1 Cross-Section of the Dual Plasma Burner for the 355 SCD
26Operation and Maintenance Manual
Figure 2 Cross-Section of the Dual Plasma Burner for the 255 NCD
Dual Plasma Controller
The Dual Plasma Controller powers the Dual Plasma Burner and supplies its
gases. Hydrogen and oxidant should be provided at 25 psig (1.72 bar) and the
Controller is plugged into an appropriate AC electrical outlet.
WARNING
Operation and Maintenance Manual27
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxidant sensors or burst their connective tubing.
The parameters monitored or regulated by the Controller include Burner
temperature, hydrogen and oxidant f low rates, and Burner pressure. The
temperature, actual pressure, oxidant and hydrogen flow rates are selected for
display by rotation of a 4-position control knob. Power, valve operation,
temperature within set-point range and fault conditions are indicated with
LED illumination on the front display panel.
The Dual Plasma Controller incorporates several safety features. The safety
circuitry detects faults such as power loss, vacuum loss, thermocouple failure,
heater element failure, broken ceramic tube, or high temperature. When a fault
is detected, the Fault LED illuminates and hydrogen and oxidant flow is
stopped by normally-closed solenoid valves.
Ozone Generator
The SCD and NCD produce ozone by corona discharge using a clean,
pressurized air or oxygen source. Use of oxygen should increase ozone
production and, hence, Detector response. High voltage to the ozone generator
is applied only when the reaction cell pressure is less than 100 torr in the SCD
and less than 200 torr in the NCD. Gas flow through the ozone generator is
controlled by a pressure regulator and flow restrictors.
WARNING
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized by
using the instrument in a well ventilated area, changing the chemical trap regularly, and
venting the exhaust of the vacuum pump. The ozone generator should be turned off when
the instrument is not in use to reduce maintenance requirements.
Chemiluminescence Reaction Cell and Photomultiplier Tube
Sulfur monoxide (formed in the Burner) and ozone (produced in the ozone
generator located in the Detector) are mixed in the reaction cell. The cell is
designed such that the reaction between SO and O
the photomultiplier tube (PMT). A UV band pass filter (300 - 400 nm) located
between the reaction cell and the PMT selectively transmits the light emitted
by the SO/O
the UV band pass filter eliminates interference from non-sulfur containing
analytes (e.g. nitric oxide, olefins, etc.) which also undergo chemiluminescent
reactions with ozone. A background signal is typically present as a result of
ozone-wall interactions and low level sulfur contamination of Detector gases.
This background signal can be used as a troubleshooting aid (see Section 10).
reaction. Efficient combustion in the ceramic tubes coupled with
3
occurs directly in front of
3
28Operation and Maintenance Manual
7 9
10
6
8
1. Ozone Generator
2. High Voltage Transformer
3. Photomultiplier Tube Socket
4. Photomultiplier Housing
5. Reaction Cell
Figure 3 355 SCD Left Side
3
4
5
2
6
1
6. Vacuum Line
7. Particulate Filter
8. Pre-Ozone Restrictor
9. Post-Ozone Restrictor
10. Pressure Transducer
Operation and Maintenance Manual29
P4 P5
WHT LOAD NEUT
P2 P3
TP3
TP8
RV1
HV
TP9 TP11
TP8 TP10 TP3
TP12
AMP 1 I
ALCO
11 12 15
18
16
ELECTRONICS
OZONE
PUMP
17
TP4
TP5
TP1
TP2
TP12
HV/100
TP11
TP13
GND
TP14
TP9
TP10
TP15 TP6
TP7 TP16
LOAD NEUT
PUMP
OUT
P6 P7
OZONE
GEN
WHT
LINE NEUT
AC
WHT
IN OFST
OUT OFST
RV1
RV2
TP2
TP7 TP6 TP4 TP 5
TP13
OUT
13
TO
FRONT
TP1
PANEL
SIEVERS RESEARCH
USA
14
Figure 4 355 SCD Right Side
11. Amplifier Cable
12. HV Cable
13. PMT Amplifier
14. EMI Filter
15. Fuses
16. Pressure Regulator
17. Transfer line
18. Front panel display
30Operation and Maintenance Manual
Loading...
+ 88 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.