Agilent Technologies G6600-90006 User Manual

Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors
Operation and Maintenance Manual
Agilent Technologies
Notices
© Agilent Technologies, Inc. 2007
No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or transla­tion into a foreign language) without prior agreement and written consent from Agi­lent Technologies, Inc. as governed by United States and international copyright laws.
Acknowledgements
Torx® is a registered trademark of Textron Inc.; Teflon® is a registered trademark of E.I. du Pont de Nemours and Company. Kimwipe® is a registered trademark of the Kimberly-Clark Corporation.
Manual Part Number
G6600-90006
Edition
First edition, June 2007 Replaces G6600-90002
Printed in USA
Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA
Warranty
The material contained in this docu­ment is provided “as is,” and is sub­ject to being changed, without notice, in future editions. Further, to the max­imum extent permitted by applicable law, Agilent disclaims all warranties, either express or implied, with regard to this manual and any information contained herein, including but not limited to the implied warranties of merchantability and fitness for a par­ticular purpose. Agilent shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or perfor­mance of this document or of any information contained herein. Should Agilent and the user have a separate written agreement with warranty terms covering the material in this document that conflict with these terms, the warranty terms in the sep­arate agreement shall control.
Technology Licenses
The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license.
Restricted Rights Legend
defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR
52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Agilent Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Gov­ernment will receive no greater than Restricted Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S. Govern­ment users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a haz­ard. It calls attention to an operat­ing procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met.
WARNING
A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly per­formed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated condi­tions are fully understood and met.
If software is for use in the performance of a U.S. Government prime contract or sub­contract, Software is delivered and licensed as “Commercial computer soft­ware” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as
Warnings
English
WARNING
WARNING
WARNING
WARNING
This symbol on the instrument indicates that the user should refer to the man­ual for operating instructions.
Any operation requiring access to the inside of the equipment, could result in injury. To avoid potentially dangerous shock, disconnect from power supply before opening the equipment.
For continued protection against fire hazard replace fuse with same type and rating.
This symbol indicates that to comply with European Union Directive 2002/96/EC for waste electrical and electronic equipment (WEEE), the Ana­lyzer should be disposed of separately from standard waste.
WARNING
WARNING
WARNING
Operation and Maintenance Manual 3
This is a safety Class I product. It must be wired to a mains supply with a protective earthing ground incorporated into the power cord. Any interruption of the protective conductor, inside or outside the equipment, is likely to make the instrument dangerous. Intentional interruption is prohibited.
If this instrument is used in a manner not specified by Agilent, the protection provided by the instrument may be impaired.
High voltage is present in the instrument when the power cord is connected, even if the main power switch is in the standby mode. To avoid potentially dangerous shock, discon­nect the power cord before removing the side panels.
WARNING
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized by using the instrument in a well-ventilated area and by venting the exhaust of the vac­uum pump to a fume hood. The ozone generator should be turned off when the instrument is not in use.
WARNING
WARNING
WARNING
WARNING
Burner temperature Is extremely hot. Do not touch. Allow to cool before servicing.
Hydrogen is an extremely flammable gas. Use appropriate care when handling. Inspect all connections with a suitable leak detector.
Oxygen rich environments can promote combustion and even result in spontaneous com­bustion under conditions of high pressure and exposure to contamination. Use only oxy­gen rated components and ensure that components are oxygen clean prior to use with pure oxygen.
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxi­dant sensors or burst their connective tubing.
4 Operation and Maintenance Manual
Español
WARNING
WARNING
WARNING
WARNING
Cualquier operación que requiera acceso al interior del equipo, puede causar una lesión. Para evitar peligros potenciales, desconectarlo de la alimentación a red antes de abrir el equipo.
Para protección contínua contra el peligro de fuego, sustituir el fusible por uno del mismo tipo y características.
Este símbolo en el instrumento indica que el usuario debería referirse al man­ual para instrucciones de funcionamiento.
Esto es un producto con clase I de seguridad. Debe conectarse a una red que disponga de tierra protectora en el cable de red. Cualquier interrupción del conductor protector, dentro o fuera del equipo, puede ser peligroso. Se prohibe la interrupción intencionada.
WARNING
Si este instrumento se usa de una forma no especificada por Agilent, puede desactivarse la protección suministrada por el instrumento.
Operation and Maintenance Manual 5
Français
WARNING
WARNING
WARNING
WARNING
Chaque opération à l'intérieur de l'appareil, peut causer du préjudice. Afin d'éviter un shock qui pourrait être dangereux, disconnectez l'appareil du réseau avant de l'ouvrir.
Afin de protéger l'appareil continuellement contre l'incendie, échangez le fus­ible par un fusible du même type et valeur.
Le symbol indique que l'utilisateur doit consulter le manuel d'instructions.
Ceci est un produit de Classe de sécurité I. L'instrument doit être branché sur l'alimentation secteur par un fil de secteur prévu d'une prise de masse. Chaque interruption du conducteur protégeant, à l'intérieur ou á l'extérieur de l'appareil peut rendre l'instrument dangereux. Interruption intentionnelle est interdite.
WARNING
Si l'instrument n'est pas utilisé suivant les instructions de Agilent, les dispositions de sécurité de l'appareil ne sont plus valables.
6 Operation and Maintenance Manual
Deutsch
WARNING
WARNING
WARNING
WARNING
Vor dem Öffnen des Gerätes Netzstecker ziehen!
Für kontinuierlichen Schutz gegen Brandgefahr dürfen bei Sicherungswech­sel nur Sicherungen der gleichen Stärke verwendet werden!
Dieses Symbol auf dem Gerät weist darauf hin, dass der Anwender zuerst das entsprechende Kapitel in der Bedienungsanleitung lesen sollte.
Dies ist ein Gerät der Sicherheitsklasse I und darf nur mit einem Netzkabel mit Schutzleiter betrieben werden. Jede Unterbrechung des Schutzleiters au erhalb oder innerhalb des Gerätes kann das Gerät elektrisch gefährlich machen. Absichtliches Unterbrechen des Schutzleiters ist ausdrücklich verboten.
WARNING
Wenn das Gerät nicht wie durch die Firma Agilent, vorgeschrieben und im Handbuch beschrieben betrieben wird, können die im Gerät eingebauten Schutzvorrichtungen beeinträchtigt werden.
Operation and Maintenance Manual 7
Italiano
WARNING
WARNING
WARNING
WARNING
Qualsiasi intervento debba essere effettuato sullo strumento può essere potenzialmente pericoloso a causa della corrente elettrica. Il cavo di alimen­tazione deve essere staccato dallo strumento prima della sua apertura.
Per la protezione da rischi da incendio in seguito a corto circuito, sostituire I fusibili di protezione con quelli dello stesso tipo e caratteristiche.
Il simbolo sullo strumento avverte l'utilizzatore di consultare il Manuale di Istruzioni alla sezione specifica.
Questo strumento è conforme alle specifiche per I prodotti in Classe I - Il cavo di alimentazione dalla rete deve essere munito di "terra". Qualsiasi interruzione del cavo di terra all'interno ed all'esterno dello strumento potrebbe risultare pericolòsa. Sono proibite interruzioni intenzionali.
WARNING
Se questo strumento viene utilizzato in maniera non conforme alle specifiche di Agilent, le protezioni di cui esso è dotato potrebbero essere alterate.
8 Operation and Maintenance Manual
Dutch
WARNING
WARNING
WARNING
WARNING
Iedere handeling binnenin het toestel kan beschadiging veroorzaken. Om ied­ere mogelijk gevaarlijke shock te vermijden moet de aansluiting met het net verbroken worden, vóór het openen van het toestel.
Voor een continue bescherming tegen brandgevaar, vervang de zekering door een zekering van hetzelfde type en waarde.
Het symbool geeft aan dat de gebruiker de instructies in de handleiding moet raadplegen.
Dit is een produkt van veiligheidsklasse I. Het toestel moet aangesloten worden op het net via een geaard netsnoer. Bij onderbreking van de beschermende geleider, aan de binnenzijde of aan de buitenzijde van het toestel, kan gebruik het toestel gevaarlijk maken. Opzettelijke onderbreking is verboden.
WARNING
Indien het toestel niet gebruikt wordt volgens de richtlijnen van Agilent, gelden de veiligheidsvoorzieningen niet meer.
Operation and Maintenance Manual 9
The following symbols are used on the equipment:
Caution - Refer to manual for operating instructions
Atención - Ver documentación pertinente.
Attention - Consultez les ocuments d'accomagnement.
Vorsicht - Siehe beiliegende Unterlagen.
Pericolo - Vedi documentazione allegata.
Voorzichtig - Raadpleeg di bijehorende documentatie.
Caution - Risk of electrical shock.
Atención - Riesgo de sacudidas eléctricas.
Attention - Risque de choc électrique.
Vorsicht - Risiko eines Elektroschocks.
Pericolo - Pericolo di scosse elettriche.
Voorzichtig - Hoge spanning, levensgevaar.
Caution - Hot surface.
Atención - Superficie caliente.
Attention - Surface brûlante.
Vorsicht - Heisse Oberfläche.
Pericolo - Superficie rovente.
Voorzichtig - Heet oppervlak.
10 Operation and Maintenance Manual

Contents

1 Introduction
2 System Description
Specifications 20
Technical Information — 355 SCD 20 Technical Information — 255 NCD 20 Physical Specifications 21
Theory of Operation 23
Sulfur Chemiluminescence Detector 23 Nitrogen Chemiluminescence Detector 23 Dual Plasma Controller 24
Description of Major Components 25
Dual Plasma Burner 25 Figure 1. Cross-Section of the Dual Plasma Burner for the 355 SCD 26 Figure 2. Cross-Section of the Dual Plasma Burner for the 255 NCD 27 Dual Plasma Controller 27 Ozone Generator 28 Chemiluminescence Reaction Cell and Photomultiplier Tube 28 Figure 3. 355 SCD Left Side 29 Figure 4. 355 SCD Right Side 30 Figure 5. 255 NCD Left Side View 31 Figure 6. 255 NCD Right Side View 32 Pressure Transducer 32 Vacuum Pump 32 Chemical Trap 33 Oil Coalescing Filter 33 FID Adapter (Optional) 33 NCD Reaction Cell 33 NCD Photomultiplier Tube and Cooler 34 Figure 7. Schematic for 355 SCD 35 Figure 8. Schematic for 255 NCD, in Nitrogen Mode 36 Figure 9. Schematic for 255 NCD, in Nitrosamine Mode 37
Operation and Maintenance Manual 11
3 Installation
Overview 40
Step 1: Selecting a Location 41
Power Requirements 41 Figure 10. Drawing of the Detector with Dual Plasma Burner and Controller 42 Environmental Considerations 42 Combustion Gas Requirements 42
Step 2: Unpack and Inspect the Instrument 45
Required Installation Tools 45
Step 3: Set Up the Vacuum Pump 46
Installing the Edwards RV5 Pump Oil-Sealed Vacuum Pump 46 Figure 11. RV5 Oil-Sealed Vacuum Pump and Associated Traps (Front Side) 49 Figure 12. RV5 Oil-Sealed Vacuum Pump and Associated Traps (Back Side) 49 Figure 13. RV5 Oil- Sealed Vacuum Pump and Associated Traps (Top) 50 Figure 14. RV-5 Oil-S ealed Vacuum Pump Exhaust Line 51 Installing the Welch Dry Piston Vacuum Pump 52 Figure 15. The Welch Dry Piston Pump 54 Figure 16. Oil Drain Kit with Ballast Control 54
Step 4: Connect the Power Cord 56
Figure 17. SCD Rear Panel Diagram 56
Step 5: Install the Dry Compressed Air or O
Supply 57
2
Step 6: Install the Signal Output Cables 58
Standard Cable Connection 58 HP 3390 Series Integrator Cable Connection 58 HP 3396 Integrator Cable Connection 58 HP 5890 GC Analog Input Board 58 Agilent 6890 GC Analog Input Board 58
Step 7: Install the Dual Plasma Burner 59
Figure 18. Dimensions of GC Liner Cut-Outs 59
Step 8: Install the Dual Plasma Controller 60
Figure 19. Dual Plasma Controller Rear Panel 60
Step 9: Install Column Connections 61
Capillary Columns 61 Figure 20. Measuring Column Insertion 61
12 Operation and Maintenance Manual
Packed Columns and Columns with an Outside Diameter > 0.8 mm 61
Step 10: Install the Transfer Line 62
4 Front Panel Controls and Initial Startup
Detector Front Panel Controls 64
Figure 21. Front Panel Controls 64 Power Controls 64 Display Output Controls 65 Signal Controls 65 Dual Plasma Controller Controls 66 Figure 22. Dual Plasma Controller Front Panel 66 Initial Startup 67
Detector Interface Setup 69
Initial Checkout 69 Monitoring Oxidizer and Hydrogen Flow with the Dual Plasma Controller 70
5Operation
Start-Up Procedure 72
Detector Operation 73
Detector Stability and Response 73 Column Bleed 73 Coking 73 Hydrogen Poisoning 73 Contaminated Gases 74 Fluctuating Pressures 74
Typical Operating Conditions 75
Table 1. Typical Operating Conditions 75
Detection Limits 76
Table 2. Expected Detection Limits for Chromatographic Conditions 76
Instrument Shut-Down 77
Daily Shutdown 77 Complete Shutdown 77
Special Operating Modes 78
Using the 255 NCD in Nitrosamine Mode 78 Using the SCD in High Sensitivity Mode for Nonhydrocarbon Gaseous Samples 78
Operation and Maintenance Manual 13
6Maintenance
Pump Maintenance 82
Cleaning the Detector 83
Changing the Oil Mist Filter (RV5) 84
Reaction Cell Cleaning 85
Flow Sensor Calibration 88
Detector Sensitivity 89
Assembling the Dual Plasma Burner for Component Replacement with the SCD 90
Table 3. Operating Life of Components for Edwards RV5 Vacuum Pump 82
Figure 23. Reaction Cell, PMT Housing and PMT Socket 86 Figure 24. Reaction Cell 86
Figure 25. Ferrule Placement on Lower Burner Tube 90 Figure 26. Proper Ferrule Orientation to the Large Ceramic Tube 91 Figure 27. Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly 91 Figure 28. Orientation of the Double Taper Ferrule 92 Figure 29. Positioning the Upper Tube in the Union Fitting 92 Figure 30. Tightening the Heater Swivel Nut 93 Figure 31. Proper Alignment of the Burner 93
Assembling the Dual Plasma Burner for Component Replacement with the NCD 95
Figure 32. Ferrule Placement on Lower Burner Tube 95 Figure 33. Proper Ferrule Orientation to the Large Quartz Tube 96 Figure 34. The Quartz Tube Properly Inserted into the Quartz Heater Assembly 96 Figure 35. Burner Assembly Detail 97 Figure 36. Burner Assembly Alignment 97
Tube Replacement for the SCD 98
Figure 37. Orientation of the Double Taper Ferrule 98 Figure 38. Positioning the Upper Tube in the Union Fitting 99 Figure 39. Proper Ferrule Orientation to the Large Ceramic Tube 99 Figure 40. Large Ceramic Tube Properly Inserted into the Quartz Heater Assembly 100 Figure 41. Tightening the Heater Swivel Nut 100 Figure 42. Proper Alignment of the Burner 101
Tube Replacement for the NCD 102
Figure 43. Proper Ferrule Orientation to the Large Quartz Tube 102 Figure 44. Large Quartz Tube Properly Inserted into the Quartz Heater Assembly 103 Figure 45. NCD Tube Replacement Detail 104 Figure 46. NCD Tube Replacement Detail 104
14 Operation and Maintenance Manual
7 Troubleshooting
Solving Detector Problems 106
Power Problems 107
Detector Fuse 107 Vacuum Pump Fuse 107 Dual Plasma Controller Fuses 107 Table 4. Fuses for 100 V, 120 V and 230 V Versions of 355 SCD and 255 NCD 108 Figure 47. Fuse Positions on the Power Supply Board 108
Ozone Generation Problems 109
Ozone Generator 109 High Voltage Transformer 109 Plugged Restrictor Lines 110
Response Problems 111
Temperature Reading Problems 112
Diagnosing General Problems 113
Table 5. Troubleshooting Detector Issues 113 Table 6. Troubleshooting Pump Issues 115 Table 7. Troubleshooting Burner Issues 116
Index
Operation and Maintenance Manual 15
16 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
1 Introduction
This manual will guide you in the installation, operation, and troubleshooting of the Agilent 355 Sulfur Chemiluminescence Detector (SCD) and the Agilent 255 Nitrogen Chemiluminescence Detector (NCD). This manual is intended for use with the Agilent 355 SCD or 255 NCD with the Dual Plasma Burner and Controller.
This operation and service manual has some important conventions, such as the use of boxed warnings. This information is deliberately set out from the text for emphasis and should be followed to assure operator safety and proper instrumental operation.
If you are installing the 355 SCD or 255 NCD yourself, follow the installation procedures described in the following sections. If your instrument is already installed, turn to the Operation section to begin.
Agilent Technologies
17
18 Operation and Maintenance Manual
Agilent 355 Sulfur and 255 Nitrogen Chemiluminescence Detectors Operation and Maintenance Manual
2 System Description
Specifications 20
Theory of Operation 23 Description of Major Components 25
Agilent Technologies
19

Specifications

Technical Information — 355 SCD
Sensitivity
Typical Selectivity g S/g C > 2 x 10
Linearity >10
Precision and Stability
*
Typical < 0.5 pg S/second (signal to noise 3.3:1)
7
4
<2% RSD 2 hours
<5% RSD 72 hours
Ozone flow through the Post Ozone Restrictor 20-30 mL/min at 3-6 psig
Typical reaction cell pressure 4 - 8 Torr RV5 Oil Sealed Pump
6 - 12 Torr Dry Piston Pump
Typical Burner Pressure 250-400 Torr operating
Typical Burner Temperature 800 °C
Typical Air Flow Rate 65 SCCM recommended
3-10 SCCM recommended with FID adapter
Typical Hydrogen Flow Rate 40 SCCM recommended
Signal Output Ranges 0-1V, 0-10V
Typical time to reach 800 °C from room
10 min typical(120 VAC, 60 Hz)
temperature
Typical safety shroud outside temperature <65 °C at 800 °C Burner temperature typical
Recorder output 0-1 V or 0-10 V
* Burner temperature 800 °C, 80 SCCM air, and 60 SCCM hyrdrogen, test compound dimethyl sulfide in toluene.
† Based on thiopene in benzene at 1 ppm mass sulfur, 1 µL injection split 1:10, 30 m, 0.32 mm ID, 1 µm thick CP
wax (n=10 for 2.5 hours; n=42 for 72 hours).
Subject to change without notice.
Technical Information — 255 NCD
Sensitivity <3 pg N/second (signal to noise 3:1) in both N
and nitrosamine modes
Selectivity g N/g C > 2 to 10
nitrosamine mode is matrix-dependent)
Linearity >10
4
20 Operation and Maintenance Manual
7
in N mode (selectivity in
Repeatability
Gas flow through Ozone Generator 20-30 mL/min at 3-6 psig (inlet pressure)
Typical reaction cell pressure 4 - 8 Torr RV5 Oil Sealed Pump
Typical Burner Pressure 130 - 150 torr operating
Typical Burner Temperature 900 °C
Typical Hydrogen Flow Rate 6-10 SCCM
Typical Oxygen Flow Rate 8 - 15 SCCM
Signal Output Ranges 0-1V, 0-10V, 0 - 10V
* Burner temperature 950 °C, 11 SCCM oxygen, and 6 SCCM hydrogen; 25 ppm N as nitrobenzene in toluene;
0.2 µL injection on column (HP 19095-121Z), n=7 for 3 hours; n=13 for 18 hours and n=10 n-dipropylnitrosamine in toluene at 4 µg/mL, 0.2 µL injection on column (HP 19095-121Z).
*

Physical Specifications

<1.5% RSD 8 hours (~ the same in N and nitrosamine)
<2% RSD 18 hours (~3% RSD in nitrosamine mode over 21 hours)
6 - 12 Torr Dry Piston Pump
Power requirements
355 SCD Detector 115 VAC, 50/60 Hz, 1400 W
100 VAC, 50/60 Hz, 1400 W 220-240 VAC, 50/60 Hz, 650W
255 NCD Detector 115 VAC, 50/60 Hz, 1400 W
100 VAC, 50/60 Hz, 1400 W 220-240 VAC, 50/60 Hz, 650W
Dual Plasma Controller 100-120 VAC, 50/60 Hz, 200 W
220-240 VAC, 50/60 Hz, 200 W
Dimensions and weight
Detector Height: 16.0 in (40.6 cm)
Width: 9.2 in (23.4 cm) Depth: 21.8 in (55.3 cm) 355 SCD Weight: 34.0 lbs (15.0 kg) 255 NCD Weight: 37.5 lbs (17.0 kg)
Operation and Maintenance Manual 21
Dual Plasma Controller Height: 5.0 in (12.7 cm)
Width: 9.5 in (24.1 cm) Depth: 12.5 in (31.8 cm) Weight: 9.9 lbs (4.5 kg)
Dual Plasma Burner Height: 12.3 in (31.2 cm)
Diameter: 4.0 in (10.2 cm) Weight: 1.9 lbs (0.9 kg)
Oil Sealed Vacuum Pump (RV5) Height: 10.3 in (26.1 cm)
Width: 6.0 in (15.2 cm) Depth: 16.9 in (43.0 cm) Weight: 47.3 lbs (21.5 kg)
or
Oil Free Dry Piston Pump Height: 12.0 in (30 cm)
Width: 9.0 in (22.9 cm) Depth: 14.0 in (35.6 cm) Weight: 29.9 lbs (13.6 kg)
Installation Category II
Pollution Degree 2
Ambient Temperature 50 - 104 °F (10-40 °C)
Relative Humidity Up to 95%, noncondensing
Normal Operating Environment Intended for indoor use only
Maximum Altitude 2,000 m (6,562 ft)
Mains Supply Voltage Fluctuation not to exceed 10% of nominal
voltage
22 Operation and Maintenance Manual

Theory of Operation

-

Sulfur Chemiluminescence Detector

The Agilent model 355 Sulfur Chemiluminescence Detector (SCD) is a sulfur-selective detector for gas chromatography. Operation of the SCD is based on the chemiluminescence (light-producing reaction) from the reaction of ozone with sulfur monoxide (SO) produced from combustion of the analyte:
Sulfur compound (analyte)
SO + O
SO2 + O2 + hη (<300–400 nm)
3
SO + H
A vacuum pump pulls the combustion products into a reaction cell at low pressure, where excess ozone is added. Light produced from the subsequent reaction is optically filtered and detected with a blue-sensitive photomultiplier tube and the signal is amplified for display or output to a data system. Figure 1 is a pneumatic flow diagram that illustrates how the components of the system are integrated.
The Detector has an enclosed, dedicated (Dual Plasma) Burner designed to enhance production of the SO intermediate. This Dual Plasma Burner mounts in the detector port of the GC. A Dual Plasma Controller provides temperature control and gas-flow regulation to operate the Dual Plasma Burner.
The Agilent model 355 SCD provides high sensitivity (<0.5 pg S/sec), with linear and equimolar response over four orders of magnitude (per Sulfur atom) while maintaining high selectivity over common solvents. The Agilent SCD is compatible with most commercially available gas chromatographs.

Nitrogen Chemiluminescence Detector

Operation of the Agilent model 255 Nitrogen Chemiluminescence Detector is based on the chemiluminescence or light-producing reaction of ozone with nitric oxide formed from combustion. Reacting nitric oxide with ozone results in the formation of electronically excited nitrogen dioxide. The excited nitrogen dioxide emits light, a chemiluminescence reaction, in the red and infrared region of the spectrum. The light emitted is directly proportionally to the amount of nitrogen in the sample,
O + other products
2
NO + O
NO2 NO2 + hη (>800 nm)
3
The light (hη) emitted by the chemical reaction is optically filtered and detected by a photomultiplier tube. The signal from the photomultiplier tube is amplified for display or output to a data system. Organic compounds containing nitrogen react to form nitric oxide, carbon dioxide, and water.
H
2/O2
--------------
R-N + O
Operation and Maintenance Manual 23
→ NO + CO
2
Δ
+ H2O
2

Dual Plasma Controller

The Dual Plasma Controller provides all operational parameters of the Dual Plasma Burner except for the Detector base temperature. The Detector base temperature is controlled by circuitry in the GC. Parameters monitored or regulated by the Controller include Burner temperature, Burner temperature set-point, hydrogen and oxidant flow rates, and Burner pressure. The temperature set-point, actual pressure, oxidant and hydrogen flow rates are displayed by rotation of a 4-position switch. Power, valve operation, temperature within set-point range and fault conditions are indicated with LED illumination on the front display panel.
The Dual Plasma Controller incorporates several safety features. The safety circuitry detects faults such as power loss, vacuum loss, thermocouple failure, heater element failure, broken ceramic tube, or high temperature. When a fault is detected, the Fault LED illuminates and hydrogen and oxidant flow is stopped by normally-closed solenoid valves.
Dual Plasma Burner with the 355 SCD
The Dual Plasma Burner is based on the same chemistry and basic principles of earlier SCD Burner designs. A key difference, however, is the addition of a second “flame” or “plasma,” the lower is oxygen-rich and the upper is hydrogen-rich. The Burner consists of a tower assembly that contains an outer sheath for burn protection, a heating element, thermocouple, and combustion tubes. Conversion of sulfur containing compounds to SO occurs within the ceramic reaction chamber housed in the Burner assembly. Agilent also provides a Flame Ionization Detector (FID) option for the simultaneous collection of hydrocarbon and sulfur chromatograms for some GCs.
Dual Plasma Burner with the 255 NCD
Compounds eluted from the GC column are combusted in the Dual Plasma Burner first by an oxygen rich flame (plasma) followed by catalytic combustion on a Noble metal screen. For hydrocarbons, this two stage combustion technique results in complete conversion of the matrix to products, such as carbon dioxide and water, which do not chemiluminesce with ozone. Nitrogen atoms in a compound are converted into nitric oxide and potentially other nitrogen oxide species. In the second stage, the catalyst is used to convert other nitrogen oxide species to nitric oxide, resulting in a high efficiency of conversion.
24 Operation and Maintenance Manual

Description of Major Components

Dual Plasma Burner

The Dual Plasma Burner consists of a tower assembly that contains an outer sheath for burn protection, a heating element, thermocouple, and combustion tubes. In the SCD, conversion of sulfur containing compounds to SO occurs within the ceramic reaction chamber housed in the Burner assembly and potentially interfering hydrocarbons are oxidized to CO
the oxidant. In the NCD, oxygen is used as the oxidant.
A fitting is located on top of the Burner. The vacuum line from the Detector box is connected directly to the top of this fitting and H2 is input to the longer side of this fitting. The air inlet is connected to the base of the Burner.
The Burner is mounted onto the GC by a model-specific mounting kit (see www.Agilent.com/chem or contact Agilent for the most current information). The GC column is inserted into the Burner using a 1/32" knurled nut and fused silica adapter ferrule.
A cross-section illustration of the Dual Plasma Burner for the 355 SCD is shown in Figure 1 and for the 255 NCD is shown in Figure 2.
and H2O, with air as
2
Operation and Maintenance Manual 25
Figure 1 Cross-Section of the Dual Plasma Burner for the 355 SCD
26 Operation and Maintenance Manual
Figure 2 Cross-Section of the Dual Plasma Burner for the 255 NCD

Dual Plasma Controller

The Dual Plasma Controller powers the Dual Plasma Burner and supplies its gases. Hydrogen and oxidant should be provided at 25 psig (1.72 bar) and the Controller is plugged into an appropriate AC electrical outlet.
WARNING
Operation and Maintenance Manual 27
Exceeding the gas inlet pressure of 25 psig (1.72 bar) may damage the hydrogen and oxi­dant sensors or burst their connective tubing.
The parameters monitored or regulated by the Controller include Burner
temperature, hydrogen and oxidant f low rates, and Burner pressure. The temperature, actual pressure, oxidant and hydrogen flow rates are selected for display by rotation of a 4-position control knob. Power, valve operation, temperature within set-point range and fault conditions are indicated with LED illumination on the front display panel.
The Dual Plasma Controller incorporates several safety features. The safety circuitry detects faults such as power loss, vacuum loss, thermocouple failure, heater element failure, broken ceramic tube, or high temperature. When a fault is detected, the Fault LED illuminates and hydrogen and oxidant flow is stopped by normally-closed solenoid valves.

Ozone Generator

The SCD and NCD produce ozone by corona discharge using a clean, pressurized air or oxygen source. Use of oxygen should increase ozone production and, hence, Detector response. High voltage to the ozone generator is applied only when the reaction cell pressure is less than 100 torr in the SCD and less than 200 torr in the NCD. Gas flow through the ozone generator is controlled by a pressure regulator and flow restrictors.
WARNING
Ozone is a hazardous gas and a strong oxidant. Exposure to ozone should be minimized by using the instrument in a well ventilated area, changing the chemical trap regularly, and venting the exhaust of the vacuum pump. The ozone generator should be turned off when the instrument is not in use to reduce maintenance requirements.

Chemiluminescence Reaction Cell and Photomultiplier Tube

Sulfur monoxide (formed in the Burner) and ozone (produced in the ozone generator located in the Detector) are mixed in the reaction cell. The cell is designed such that the reaction between SO and O
the photomultiplier tube (PMT). A UV band pass filter (300 - 400 nm) located between the reaction cell and the PMT selectively transmits the light emitted by the SO/O
the UV band pass filter eliminates interference from non-sulfur containing analytes (e.g. nitric oxide, olefins, etc.) which also undergo chemiluminescent reactions with ozone. A background signal is typically present as a result of ozone-wall interactions and low level sulfur contamination of Detector gases. This background signal can be used as a troubleshooting aid (see Section 10).
reaction. Efficient combustion in the ceramic tubes coupled with
3
occurs directly in front of
3
28 Operation and Maintenance Manual
7 9
10
6
8
1. Ozone Generator
2. High Voltage Transformer
3. Photomultiplier Tube Socket
4. Photomultiplier Housing
5. Reaction Cell
Figure 3 355 SCD Left Side
3
4
5
2
6
1
6. Vacuum Line
7. Particulate Filter
8. Pre-Ozone Restrictor
9. Post-Ozone Restrictor
10. Pressure Transducer
Operation and Maintenance Manual 29
P4 P5
WHT LOAD NEUT
P2 P3
TP3
TP8
RV1
HV
TP9 TP11
TP8 TP10 TP3
TP12
AMP 1 I
ALCO
11 12 15
18
16
ELECTRONICS
OZONE
PUMP
17
TP4
TP5 TP1
TP2
TP12
HV/100
TP11
TP13
GND
TP14
TP9 TP10
TP15 TP6
TP7 TP16
LOAD NEUT
PUMP
OUT
P6 P7
OZONE
GEN
WHT
LINE NEUT
AC
WHT
IN OFST
OUT OFST
RV1
RV2
TP2
TP7 TP6 TP4 TP 5
TP13
OUT
13
TO FRONT
TP1
PANEL
SIEVERS RESEARCH
USA
14
Figure 4 355 SCD Right Side
11. Amplifier Cable
12. HV Cable
13. PMT Amplifier
14. EMI Filter
15. Fuses
16. Pressure Regulator
17. Transfer line
18. Front panel display
30 Operation and Maintenance Manual
Loading...
+ 88 hidden pages