Agilent Technologies 54624A, 54621D, 54622A, 54622D, 54621A User Manual

Service Guide
Publication Number 54622-97037 December 2002
For Safety Information, Warranties, and Regulatory information, see the pages at the end of this book.
© Copyright Agilent Technologies 2000, 2002
All Rights Reserved
Agilent 54621A/22A/24A
Agilent 54621D/22D
Mixed-Signal Oscilloscope

The Oscilloscope At a Glance

Choose from a variety of oscilloscopes for capturing long, non-repeating signals
with 200 MSa/s sample rate and 2 MBytes of MegaZoom deep memory per channel.
Agilent 54621A - 2-channel, 60-MHz bandwidth
Agilent 54621D - 2-channel +16 logic channels, 60­MHz bandwidth
Agilent 54622A - 2-channel, 100-MHz bandwidth
Agilent 54622D - 2-channel +16 logic channels, 100-MHz bandwidth
Agilent 54624A - 4-channel, 100-MHz bandwidth
Display shows current input signals
• All analog and digital (54621D/22D) channels displayed in main and delayed mode
• Indicators for channel, time base, digital (54621D/ 22D) channel activity, trigger and acquisition status
• Softkey labels
• Measurement results
Digital channel controls select, position, and label inputs (54621D/22D)
• Turn channels on or off individually or in groups of 8
• Rearrange order of channels to group related signals
• Create and display labels to identify channels
Run control keys begin and end data acquisition
• Run/Stop starts and stops continuous acquisitions
• Single performs one acquisition
• Infinite persistence accumulates and displays the results of multiple acquisitions
Horizontal Controls select sweep speed and delay parameters
• Sweep speeds from 5 ns/div to 50 ns/div
• Delay control moves waveform display to point of interest
• Delayed mod e and delay allow zoom ing in to show a portion of waveform in detail (split screen)
Trigger keys define what data the oscilloscope will trigger on
• Source key allows conventional oscilloscope triggering
• Modes include Edge, Pulse Width, Pattern, CAN, Duration, I triggering
General controls measure, save and restore results, and configure the oscilloscope
• Waveform math including FFT, subtract, multiply, integrate, and differentiate
• Use Quick Meas to make automatic measurements Integrated counter included with Quick Meas.
• Use cursors to make manual measurements
• Save or recall measurement configurations or previous results
• Autoscale performs simple one-button setup of the oscilloscope
Softkeys extend the functionality of command keys
Select measurement types, operating modes, trigger specifications, label data, and more
Digital channel inputs through a flexible probing system (54621D/22D)
• Sixteen channels through a dual 8-channel cable with micro-clips
• Set logic levels as TTL, CMOS, ECL, or to a user­definable voltage
Utilities
• Dedicated parallel printer port, controller operation, floppy disk storage
Built in Quick Help system
• Press and hold any key front-panel key or softkey to get help in 11 languages.
2
C, LIN, Sequence, SPI, TV, and USB
ii

In This Book

This book provides the service information for the Agilent 54621A/22A/24A Oscilloscope and the Agilent 54621D/22D Mixed-Signal Oscilloscope.
This manual is divided into these chapters:
Chapter 1 provides general information and specifications. Chapter 2 shows you how to prepare the oscilloscope for use. Chapter 3 gives performance tests. Chapter 4 covers calibration and adjustment procedures. Chapter 5 provides troubleshooting information. Chapter 6 gives the procedures and techniques for replacing assemblies and
other parts. Chapter 7 includes a list of replaceable parts, part ordering information, and
shipping information.
At the back of the book you will find Safety information, Warranties, and Regulatory information.
iii

Contents

1 General Information
To inspect package contents 1-3 To inspect options and accessories 1-6 Performance Characteristics 1-9
2 Preparing the Oscilloscope for Use
Setting up the Oscilloscope 2-3
To adjust the handle 2-4 To power-on the oscilloscope 2-5 To adjust the display intensity 2-6 To connect the oscilloscope probes 2-7 To compensate your probe 2-7 To use the digital probes (mixed-signal oscilloscope only) 2-8 To connect a printer 2-12 To connect an RS-232 cable 2-12 To verify basic oscilloscope operation 2-13
Getting started using the oscilloscope interface 2-14 Using Quick Help 2-16
Selecting a language for Quick Help when the oscilloscope starts up 2-16 Selecting a language for Quick Help after you have been operating the oscilloscope 2-17 Loading a language from floppy disk 2-18
Cleaning the oscilloscope 2-19
3 Testing Performance
List of Test Equipment 3-3 To construct the test connector 3-4 To test the 54621D/22D Oscilloscope digital channels 3-5 To verify threshold accuracy 3-6 To verify voltage measurement accuracy 3-10 To verify bandwidth 3-13 To verify horizontal Dt and 1/Dt accuracy 3-15 To verify trigger sensitivity 3-17 Agilent 54622A/22D/24A Performance Test Record 3-20 Agilent 54621A/21D Performance Test Record 3-21
Contents-1
Contents
4 Calibrating and Adjusting
To adjust the power supply 4-4 To perform User Cal 4-7 To adjust the oscilloscope display 4-8
5 Troubleshooting
Solving General Problems with the Oscilloscopes 5-3
If there is no trace display 5-3 If the trace display is unusual or unexpected 5-4 If you cannot see a channel 5-5
Troubleshooting the Oscilloscope 5-6
To construct your own dummy load 5-7 To check out the oscilloscope 5-8 To check the Low Voltage Power Supply 5-11 To run the internal self-tests 5-13
6 Replacing Assemblies
To remove the cabinet 6-4 To remove the fan 6-5 To remove the floppy drive 6-6 To remove the front panel 6-7 To remove the display 6-8 To remove the system board 6-10 To remove the power supply 6-12 To remove the keyboard assembly 6-15 To remove the handle 6-16
7 Replaceable Parts
To order a replacement part 7-3
Contents-2
1

General Information

General Information
This chapter lists general information for the Agilent 54620-series Oscilloscopes. It also includes performance characteristics and specifications for the oscilloscopes.
1-2
General Information

To inspect package contents

Inspect the shipping container for damage.
If your shipping container appears to be damaged, keep the shipping container or cushioning material until you have inspected the contents of the shipment for completeness and have checked the oscilloscope mechanically and electrically.
Verify that you received the following items and any optional accessories in
the oscilloscope packaging (see figure following).
• 54620-Series Oscilloscope (54621A, 21D, 22A, 22D, or 24A)
• 10074C 10:1 150 MHz passive probes with ID: (2) for 54621A, 21D, 22A, or 22D oscilloscopes (4) for 54624A oscilloscope
• 54620-68701 digital probe kit (for 54621D or 22D)
• Accessory pouch and front-panel cover (standard for 54622A, 22D, and 24A) (optional on 54621A and 21D; order N2726A)
• Power cord (see table 1-3)
• IntuiLink for 54600-series Oscilloscopes software and RS-232 cable (for
54622A, 22D, or 24A). IntuiLink is a Windows application that makes it very easy for you to
download images, waveform data, or oscilloscope setups from the oscilloscope to your pc using either Microsoft Word or Microsoft Excel. After installation of IntuiLink, a tool bar in these Microsoft applications will make connection and data transfer from the oscilloscope very simple.
IntuiLink software is available free on the web at: www.agilent.com/find/5462xsw RS-232 cable may be ordered separately, part number 34398A
If anything is missing, contact your nearest Agilent Sales Office. If the shipment was d amaged, contact th e carrier, then cont act the ne arest Agil ent Sale s Office.
1-3
General Information
• Agilent IntuiLink Data Capture (for 54622A, 22D, or 24A) IntuiLink Data Capture is a standalone program for downloading waveform
data from the oscilloscopes to your PC via GPIB or RS-232 interface. It provides the capability to transfer deep memory data out of the oscilloscope, allowing up to 4MB (scope channels) and 8MB (logic channels). The IntuiLink for 54600-Series limits the size of acquisition data available to a maximum of 2,000 points regardless of actual number of acquisition points on the screen. With the IntuiLink Data Capture, the amount of points transferred will be the actual number of acquisition points currently displayed or you may select the number of points to download. It provides the following functionality:
• Download waveform data and display the data as a simple chart
• Save the data as binary or text files
• Copy the chart and a selected portion of the data to the clipboard. The
maximum data saved to the clipboard is 50,000 point
• Load saved waveform data back into the application IntuiLink Data Capture software is available free on the web at:
www.agilent.com/find/5462xsw RS-232 cable may be ordered separately, part number 34398A
Inspect the oscilloscope
• If there is mechanical damage or a defect, or if the oscilloscope does not operate properly or does not pass the performance tests listed in the Service Guide, notify your Agilent Sales Office.
• If the shipping container is damaged, or the cushioning materials show signs of stress, notify the carrier and your Agilent Sales Office. Keep the shipping materials for the carrier’s inspection. The Agilent Sales Office will arrange for repair or replacement at Agilent’s option, without waiting for claim settlement.
1-4
Figure 1-1
General Information
54620-Series Oscilloscope
Accessories pouch and front-panel cover**
Power cord
54620-68701 digital probe kit*
54620-61801 16-channel cable***
IntuiLink for 54600-series software, Data Capture software and serial cable**
5959-9334 2 Probe ground lead (qty 5)
5090-4833 Grabber (qty 20)
10074C Probes
s
s1
* 54621D /22D only ** 54622A/22D/24A only ** The following additional replacement parts (not included) are available for the digital cable:
5959-9333 replacement probe leads (qty 5) 5959-9335 replacement pod grounds (qty 5) 01650-94309 package of probe labels
Package contents for 54620-Series Oscilloscopes
1-5
General Information

To inspect options and accessories

Verify that you received the options and accessories you ordered and that none
were damaged.
If anything is missing, contact your nearest Agilent Sales Office. If the shipment was damaged, or the cushioning materials show signs of stress, notify the carrier and your Agilent Sales Office.
Some of the options and accessories available for the 54620-Series Oscilloscopes are listed in tables 1-1 am 1-2. Contact your Agilent Sales Office for a complete list of options and accessories.
Table 1-1
Options available
Option Description
003 Shielding Option for use in severe environments or with sensitive devices under
test–shields both ways (in and out):
RS-03 magnetic interface shielding added to CRT, and
RE-02 display shield added to CRT to reduce radiated interference. 1CM Rackmount kit (same as Agilent 1186A) A6J ANSI Z540 compliant calibration service
See table 1-3 for power cord options
1-6
General Information
Table 1-2
Accessories available
Model Description
01650-61607 16:16 logic cable and terminator (for use with 54621D/22D) 54620-68701 16:2 x 8 logic input probe assembly (shipped standard with 54621D/22D) 1146A 100 kHz current probe, ac/dc 1183A Testmobile scope cart 1185A Carrying case 1186A Rackmount kit 10070C 1:1 passive probe with ID 10072A Fine-pitch probe kit 10073C 10:1 500 MHz probe with ID 10075A 0.5 mm IC clip kit 10076A 100:1, 4 kV 250 MHz probe with ID 10100C 50 10833A GPIB cable, 1 m long 34398A RS-232 cable (standard with 100 MHz models) E2613B 0.5 mm Wedge probe adapter, 3-signal, qty 2 E2614A 0.5 mm Wedge probe adapter, 8-signal, qty 1 E2615B 0.65 mm Wedge probe adapter, 3-signal, qty 2 E2616A 0.65 mm Wedge probe adapter, 8-signal, qty 1 E2643A 0.5 mm Wedge probe adapter, 16-signal, qty 1 E2644A 0.65 mm Wedge probe adapter, 16-signal, qty 1 N2726A Accessory pouch and front-panel cover (standard with 100 MHz models) N2727A Thermal printer and pouch N2728A 10 rolls of thermal printer paper N2757A GPIB Interface Module N2758A CAN Trigger Module N2771A 1000:1, 15 kV, 50 MHz high voltage probe N2772A 20 MHz differential probe N2773A Differential probe power supply N2774A 50 MHz current probe, ac/dc N2775A Power supply for N2774A
termination
1-7
General Information
Table 1-3. Power Cords
Plug Type Cable Part Number Plug Type Cable Part Number
Opt 900 (U.K.) 8120-1703 Opt 918 (Japan) 8120-4754
Opt 901 (Australia) 8120-0696 Opt 919 (Israel) 8120-6799
Opt 902 (Europe) 8120-1692 Opt 920 (Argentina) 8120-6871
Opt 903 (U.S.A.) 8120-1521 Opt 921 (Chile) 8120-6979
Opt 906 (Switzerland) 8120-2296 Opt 922 (China) 8120-8377
Opt 912 (Denmark) 8120-2957 Opt 927 (Thailand) 8120-8871
Opt 917 (Africa) 8120-4600
1-8
General Information
Acquisition: Analog Channels

Performance Characteristics

* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Acquisition: Analog Channels
Max Sample rate 200 MSa/s Max Memory Depth 4 MB interleaved, 2 MB each channel Vertical Resolution 8 bits Peak Detection 5 ns Averages selectable from 2, 4, 8, 16, 32, 64 ...to 16383 High Resolution Mode 12 bits of resolution when Filter: Sinx/x interpolation (single shot BW = sample rate/4 or bandwidth of scope, whichever
is less) with vectors on.
500 us/div, average mode with average = 1
Acquisition: Digital Channels (on 54621D and 54622D only)
Max Sample Rate 400 MSa/s Max Memory Depth 8 MB Vertical Resolution 1 bit Glitch Detection (min pulse width) 5 ns
1-9
General Information
Vertical System: Analog Channels
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Vertical System: Analog Channels
Analog channels 54621A/21D, 54622A/22D: Ch1 and 2 simultaneous acquisition
Bandwidth (-3dB)* 54621A/22D: dc to 60 MHz
ac coupled 54621A/21D: 3.5 Hz to 60 MHz
Calculated rise time
(= 0.35/bandwidth)
Single Shot Bandwidth 50 MHz
1
Range Maximum Input CAT I 300 Vrms, 400 Vpk
Offset Range ±5 V on ranges <10 mV/div
Dynamic Range Lesser of ±8 div or ±32 V Input Resistance 1 M Input Capacitance ~ 14 pF Coupling ac, dc, ground BW Limit ~ 20 MHz selectable Channel-to-Channel Isolation
(with channels at same V/div) Probes 10:1 10074C shipped standard for each analog channel Probe ID (Agilent/HP &
Tek Compatible)
1
1 mV/div is a magnification of 2 mV/div setting. For vertical accuracy calculations, use full scale of 16 mV for 1 mV/div sensitivity setting.
54624A: Ch 1, 2, 3, and 4 simultaneous acquisition
54622A/22D/24A: dc to 100 MHz
54622A/22D/24A: 3.5 Hz to 100 MHz 54621A/22D: ~5.8 ns
54622A/22D/24A: ~3.5 ns
1 mV/div to 5 V/div
CAT II 100 Vrms, 400 Vpk with 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
±25 V on ranges 10 mV/div to 199 mV/div ±100 V on ranges
200 mV/div
±1%
dc to 20 MHz > 40 dB 20 MHz to max bandwidth > 30 dB
Auto probe sense
1-10
General Information
Vertical System: Analog Channels (continued)
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Vertical System: Analog Channels (continued)
ESD Tolerance ±2 kV Noise Peak-to-Peak 2% full scale or 1 mV, whichever is greater Common Mode Rejection Ratio 20 dB @ 50 MHz DC Vertical Gain Accuracy*
1
±2.0% full scale
DC Vertical Offset Accuracy < 200 mV/div: ±0.1 div ±1.0 mV ±0.5% offset
200 mV/div: ±0.1 div ±1.0 mV ±1.5% offset value
Single Cursor Accuracy
Dual Cursor Accuracy*
1
1 mV/div is a magnifica tion of 2 mV/div setting. For vertical a ccuracy calculations, use full scale of 1 6 mV for 1 mV/div sensitivity s etting.
1
1
±{DC Vertical Gain Accuracy + DC Vertical Offset Accuracy + 0.2% full scale (~1/2 LSB) }
Example: For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset, accuracy = ±{2.0%(80mV) + 0.1 (10 mV) + 1.0 mV + 0.5% (5 mV) + 0.2%(80 mV)} = ± 3.78 mV
±{DC Vertical Gain Accuracy + 0.4% full scale (~1 LSB)} Example: For 50 mV signal, scope set to 10 mV/div (80 mV full scale), 5 mV offset,
accuracy = ±{2.0%(80 mV) + 0.4%(80 mV)} = ±1.92 mV
Vertical System: Digital Channels (54621D and 54622D only)
Number of Channels 16 Digital – labeled D15 – D0 Threshold Groupings Pod 1: D7 - D0
Threshold Selections TTL, CMOS, ECL, user-definable (selectable by pod) User-Defined Threshold Range ±8.0 V in 10 mV increments Maximum Input
Voltage Threshold Accuracy* ±(100 mV + 3% of threshold setting) Input Dynamic Range ±10 V about threshold Minimum Input Voltage Swing 500 mV peak-to-peak Input Capacitance ~ 8 pF Input Resistance 100 k Channel-to-Channel Skew 2 ns typical, 3 ns maximum
Pod 2: D15 - D8
±40 V peak CAT I
±2% at probe tip
1-11
General Information
Horizontal
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Horizontal
Range 5 ns/div to 50 s/div Resolution 25 ps Vernier 1-2-5 increments when off, 25 minor increments between major settings when on Reference Positions Left, Center, Right Delay Range
Pre-trigger (negative delay)
Post-trigger (positive delay)
Analog Delta-t Accuracy
Same Channel*
Greater of 1 screen width or 10 ms 500 seconds
±0.01% reading ±0.1% screen width ±40 ps Example: for signal with pulse width of 10 us, scope set to 5 us/div (50 us screen width), delta-t accuracy = ±{.01%(10 us) + 0.1% (50 us) + 40 ps} = 51.04 ns
Channel-to-Channel Digital Delta-t Accuracy
Same Channel
Channel-to-Channel
Delay Jitter 10 ppm RMS Jitter 0.025% screen width + 30 ps Modes Main, Delayed, Roll, XY XY
Z blanking
Bandwidth
Phase error @ 1 MHz
±0.01% reading ±0.1% screen width ±80 ps (non-Vernier settings)
±0.01% reading ±0.1% screen width ±(1 digital sample period, 2.5 or 5 ns based on sample rate of 200/400 MSa/s) Example: for signal with pulse width of 10 us, scope set to 5 us/div (50 us screen width), and single pod active (400 MSa/s), delta-t accuracy = ±{.01%(10 us) + 0.1% (50 us) + 2.5 ns} = 53.5 ns
±0.01% reading ±0.1% screen width ±(1 digital sample period, 2.5 or 5 ns) ±chan-to-chan skew (2 ns typical, 3 ns maximum)
1.4 V blanks trace (use External trigger) Max bandwidth
1.8 degrees
1-12
General Information
Trigger System
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Trigger System
Sources: 54621A/22A: Ch 1, 2, line, ext
Modes Auto, Auto level, Triggered (normal), Single Holdoff Time ~60 ns to 10 seconds Selections Edge, Pattern, Pulse Width, CAN, Duration, I
Edge Trigger on a rising or falling edge of any source. Pattern Trigger on a pattern of high, low, and dont care levels and a rising or falling edge
Pulse Width Trigger when a positive- or negative-going pulse is less than, greater than, or within a
CAN Trigger on CAN (Controller Area Network) version 2.0A and 2.0B signals. It can trigger
Duration Trigger on a multi-channel pattern whose time duration is less than a value, greater
2
C Trigger on I2C (Inter-IC bus) serial protocol at a start/stop condition, a restart, a missing
I
LIN Trigger on LIN (L ocal Interconnect Network) sync br eak at beginning of message frame. Sequence Find event A, trigger on event B, with option to reset on event C or time delay. SPI Trigger on SPI (Serial Peripheral Interface) a data pattern during a specific framing
USB Trigger on USB (Universal Serial Bus) Start of Packet, End of Packet, Reset Complete,
TV Trigger on any analog channel for NTSC, PAL, PAL-M, or SECAM broadcast standards
Autoscale Finds and displays all active analog and digital (for 54621D/54622D) cha nnels, sets edge
54621D/22D: Ch 1, 2, line, ext, D15 - D0 54624A: Ch 1, 2, 3, 4, line, ext
2
C, Sequence, SPI, TV, USB
established ac ross any of the sources . The analog channels high or low level is defi ned by that channels trigger level.
specified range on any of the source channels.
Minimum pulse width setting: 5 ns Maximum pulse width setting: 10 s
on the Start of Fra me bit of a data frame, a remote trans fer request frame, or an overlo ad frame.
than a value, greater than a time value with a timeout value, or inside or outside of a set of time values.
Minimum duration setting: 5 ns Maximum duration setting: 10 s
acknowledge, or user defined frame with address and/or data values. Also trigger on Missing Acknowledge, Restart, EEPROM read, and 10-bit write.
period. Supp ort positive and negative Chip Select framing ad well as clock Idle framing and user-specified number of bits per frame.
Enter Suspend, or Exit Suspend on the differential USB data lines. USB low speed and high speed are supported.
on either positive or negative composite video signals. Modes supported include Field 1, Field 2, or both, all lines, or any line within a field. Also supports triggering on non­interlaced fields. TV trigger sensitivity: 0.5 division of synch signal.
trigger mode on h ighest numbered channel, se ts vertical sensitivity on analog channels and thresholds on di gital channels, time base to displ ay ~1.8 periods. Requires minimum voltage >10 mVpp, 0.5% duty cycle and minimum frequency >50 Hz.
1-13
General Information
Analog Channel Triggering
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Analog Channel Triggering
Range (Internal) ±6 div from center screen Sensitivity* Greater of 0.35 div or 2.5 mV Coupling AC (~3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz)
Digital (D15 - D0) Channel Triggering (54621D and 5462 2D)
Threshold Range (user-defined) ±8.0 V in 10 mV increments Threshold Accuracy* ±(100 mV + 3% of threshold setting) Predefined Thresholds TTL = 1.4 V, CMOS = 2.5 V, ECL = -1.3 V
External (EXT) Triggering
Input Resistance 1 M ±3% Input Impedance ~ 14 pF Maximum Input CAT I 300 Vrms, 400 Vpk
CAT II 100 Vrms, 400 Vpk
with 10074C 10:1 probe:CAT I 500 Vpk, CAT II 400 Vpk Range ±10 V Sensitivity dc to 25 MHz, < 75 mV
Coupling AC (~ 3.5 Hz), DC, noise reject, HF reject and LF reject (~ 50 kHz) Probe ID (Agilent/HP & Tek
Compatible)
25 MHz to max bandwidth, < 150 mV
Auto probe sense for 54621A/22A
1-14
General Information
Display System
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
Display System
Display 7-inch raster monochrome CRT Throughput of Analog Channels 25 million gray scale vectors/sec per channel Resolution 255 vertical by 1000 horizontal points (waveform area)
Controls Waveform intensity on front panel
Built-in Help System Key-specific help in 11 languages displayed by pressing and holding key or softkey of
Real Time Clock Time and date (user setable)
32 levels of gray scale
Vectors on/off; infinite persistence on/off 8 x 10 grid with continuous intensity control
interest
Measurement Features
Automatic Measurements Measurements are continuously updated
Voltage (analog channels only) Peak-to-Peak, Maximum, Minimum, Average, Amplitude, Top, Base, Overshoot,
Time Frequency, Period, + Width, - Width, and Duty Cycle on any channels.
Counter Built-in 5-digit frequency counter on any channel. Counts up to 125 MHz Threshold Definition Variable by percent and absolute value; 10%, 50%, 90% default for time measurements Cursors Manually or automatically placed readout of Horizontal (X,
Waveform Math 1-2, 1*2, FFT, differentiate, integrate.
Cursors track current measurement
Preshoot, RMS (DC)
Rise time, Fall time, X at Max (Time at max volts), X at Min (Time at min volts), Delay, and Phase on analog channels only.
X, 1/X) and
Vertical (Y, hex values
Source of FFT: differentiate, integrate, analog channels 1 or 2 (or 3 or 4 for 54624A), 1­2, 1+2, 1*2
Y). Additionally digital or analog channels can be displayed as binary or
1-15
General Information
FFT
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
FFT
Points Fixed at 2048 points Source of FFT Analog channels 1 or 2 (or 3 or 4 for 54624A), 1+2, 1-2, 1*2 Window Rectangular, Flattop, Hanning Noise Floor -70 to -100 dB depending on averaging Amplitude Display In dBV Frequency Resolution: 0.097656/(time per div) Maximum Frequency 102.4/(time per div)
Storage
Save/Recall (non-volatile) 3 setups and traces can be saved and recalled internally Floppy Disk
Image formats Data formats Trace/setup formats
3.5 1.44 MB double density
TIF, BMP
X and Y (time/voltage) values in CSV format
Recalled
I/O
RS-232 (serial) standard port 1 port; XON or DTR; 8 data bits; 1 stop bits; parity=none; 9600, 19200, 38400, 57600 baud
Parallel standard port Printer support Printer Compatibility HP DeskJet, HP LaserJet with HP PCL 3 or greater compatibility
Optional GPIB Interface Module Fully programmable with IEEE488.2 compliance
rates
Compatibility– black and white @150x150 dpi
gray scale @ 600x600 dpi Epson–black and white @180x180 dpi Seiko–DPU-414 black and white
Typical GPIB throughput of 20 measurements or twenty 2000-point records per second.
1-16
General Information
General Characteristics
* Denotes Warranted Specifications, all others are typical. Specifications are valid after a 30-minute warm-up period and ±10 °C from firmware calibration temperature.
General Characteristics
Physical:
Size
Weight Calibrator Output Frequency ~1.2 kHz; Amplitude ~5 V Trigger Out 0 to 5 V with 50 Printer Power 7.2 to 9.2 V, 1 A Kensington lock Connection on rear panel for security
32.26 cm wide x 17.27 cm high x 31.75 cm deep (without handle)
6.35 kgs (14 lbs)
source impedance; delay ~ 55 ns
Power Requirements
Line Voltage Range 100 - 240 VAC ±10%, CAT II, automatic selection Line Frequency 47 to 440 Hz Power Usage 100 W max
Environmental Characteristics
Ambient Temperature Operating -10 °C to +55 °C
Humidity Operating 95% RH at 40 °C for 24 hr
Altitude Operating to 4,570 m (15,000 ft)
Vibration HP/Agilent class B1 and MIL-PRF-28800F Class 3 random Shock HP/Agilent class B1 and MIL-PRF-28800F (operating 30 g, 1/2 sine, 11-ms duration, 3
Pollution degree2 Normally only dry non-conductive pollution occurs. Occasionally a temporary
Indoor use only This instrument is rated for indoor use only
Non-operating -51 °C to +71 °C
Non-operating 90% RH at 65 °C for 24 hr
Non-operating to 15,244 m (50,000 ft)
shocks/axis along major axis. Total of 18 shocks)
conductivity caused by condensation must be expected.
1-17
2

Preparing the Oscilloscope for Use

Preparing the Oscilloscope for Use
To prepare your oscilloscope for use, you need to do the following tasks. A ft er y ou ha ve co mp l et ed th em , y ou wi ll be re a dy to us e t he o sc il lo sc op e.
In the following topics you will:
• adjust the handle
• power-on the oscilloscope
• adjust the display intensity
• connect the oscilloscope probes
• connect the digital probes (with 54621D and 54622D)
• connect a printer
• connect a RS-232 cable
• verify basic oscilloscope operation
• get started using the oscilloscope interface
• learn how to use Quick Help
This chapter also tells you how to:
• clean the oscilloscope
2-2

Setting up the Oscilloscope

After you have done a few basic tasks, you will connect probes to the oscilloscope. The number of probes, and the type of probes that you will use depends on the oscilloscope model that you have.
• When using the Agilent 54621A and 54622A 2-channel Oscilloscopes, and the Agilent 54624A 4-channel Oscilloscope, you will connect and use analog probes to examine analog signals.
Analog channels
(2 or 4, depending
on the oscilloscope
model)
• When using the Agilent 54621D and 54622D Mixed-Signal Oscilloscopes, you will connect and use both analog and digital probes to examine analog and digital signals.
Analog channels (2)
Digital channels (16)
2-3
Preparing the Oscilloscope for Use

To adjust the handle

To adjust the handle
1 Grasp the handle pivot points on each side of the instrument and pull
the pivot out until it stops.
54622D
Agilent
MIXED SIGNAL OS CILLOSCOPE
CHANNEL
Select
Time/Div
150
5 ns1 s
INPUTS
2 Without releasing the pivots, swivel the handle to the desired position.
Then release the pivots. Continue pivoting the handle until it clicks into a set position.
2-4
Preparing the Oscilloscope for Use

To power-on the oscilloscope

To power-on the oscilloscope
1 Connect the power cord to the rear of the oscilloscope, then to a suitable
ac voltage source.
The oscilloscope power supply automatically adjusts for input line voltages in the range 100 to 240 VAC. Therefore, you do not need to adjust the input line voltage setting. The line cord provided is matched to the country of origin. Ensure that you have the correct line cord. See table 1-3
2 Press the power switch.
Trigger out
~5V
Some front panel key lights will come on and the oscilloscope will be operational in about 5 seconds.
2-5
Preparing the Oscilloscope for Use

To adjust the display intensity

To adjust the display intensity
The Intensity control is at the lower left corner of the front panel.
• To decrease display intensity, rotate the Intensity control counter­clockwise.
• To increase display intensity, rotate the Intensity control clockwise.
Dim Bright
Intensity control
The grid or graticule intensity on the display can be adjusted by pressing the Display key, then turn the Entry knob (labeled on the front panel) to adjust the Grid control.
2-6
Preparing the Oscilloscope for Use

To connect the oscilloscope probes

To connect the oscilloscope probes
1 Connect the Agilent 10074C 1.5-meter, 10:1 oscilloscope probe to the
analog channel 1 or 2 BNC connector input on the oscilloscope, or channel 1 through channel 4 on the 54624A.
Maximum input voltage for analog inputs:
CAT I 300 Vrms, 400 Vpk CAT II 100 Vrms, 400 Vpk with 10074C 10:1 probe: CAT I 500 Vpk, CAT II 400 Vpk
2 Connect the retractable hook tip on the probe tip to the circuit point of
interest. Be sure to connect the probe ground lead to a ground point on the circuit.
The probe ground lead is connected to the oscilloscope chassis and the ground wire in the power cord. If you need to connect the ground lead to a point in the circuit that cannot be grounded to power ground, consider using a differential probe.
Perfectly compensated
Over compensated
Under compensated

To compensate your probe

You should compensate you probes to match their characteristics to the oscilloscope. A poorly compenstated probe can introduce measurement errors. To compensate a probe, follow these steps:
1 Connect the probe from channel 1 to the Probe Comp signal on the lower-
right corner of the front panel. 2Press 3 Use a nonmetallic tool to adjust the trimmer capacitor on the probe for
Autoscale.
the flattest pulse possible.
comp.cdr
2-7
Preparing the Oscilloscope for Use

To use the digital probes (mixed-signal oscilloscope only)

To use the digital probes (mixed-signal oscilloscope only)
1 If you feel it’s necessary, turn off the power supply to the circuit under
test.
Off
Turning off power to the circuit under test would only prevent damage that might occur if you accidentally short two lines together while connecting probes. You can leave the oscilloscope powered on because no voltage appears at the probes.
2 Connect the digital probe cable to D15 - D0 connector on the front panel
of the mixed-signal oscilloscope. The digital probe cable is indexed so you can connect it only one way. You do not need to power-off the oscilloscope.
Use only the Agilent part number 54620-68701 digital probe kit supplied with the mixed-signal oscilloscope.
2-8
Preparing the Oscilloscope for Use
To use the digital probes (mixed-signal oscilloscope only)
3 Connect a grabber to one of the probe leads. Be sure to connect the
ground lead. (Other probe leads are omitted from the figure for clarity.)
Grabber
4 Connect the grabber to a node in the circuit you want to test.
2-9
Preparing the Oscilloscope for Use
To use the digital probes (mixed-signal oscilloscope only)
5 For high-speed signals, connect a ground lead to the probe lead, connect
a grabber to the ground lead, and attach the grabber to ground in the circuit under test.
Signal Lead
Ground Lead
Grabber
6 Connect the ground lead on each set of channels, using a probe grabber.
The ground lead improves signal fidelity to the instrument, ensuring accurate measurements.
Channel Pod Ground
Circuit Ground
2-10
Preparing the Oscilloscope for Use
To use the digital probes (mixed-signal oscilloscope only)
7 Repeat steps 3 through 6 until you have connected all points of interest.
Signals
Ground
8 If you need to remove a probe lead from the cable, insert a paper clip
or other small pointed object into the side of the cable assembly, and
push to release the latch while pulling out the probe lead.
Replacement parts are available. See the “Replaceable Parts” chapter for details.
2-11
Preparing the Oscilloscope for Use

To connect a printer

To connect a printer
The oscilloscope connects to a parallel printer through the Parallel output connector on the rear of the oscilloscope. You will need a parallel printer cable to connect to the printer.
1 Attach the 25-pin small “D” connector to the Parallel output connector
on the rear of the oscilloscope. Tighten the thumbscrews on the cable
connector to secure the cable. 2 Attach the larger 36-pin “D” connector to the printer. 3 Set up the printer configuration on the oscilloscope.
a Press the Utility key, then press the Print Confg softkey.
b Press the Print to: softkey and set the interface to Parallel.
c Press the Format softkey and select your printer format from the list.
For more information on printer configuration, refer to the “Utilities” chapter
in the User’s Guide.

To connect an RS-232 cable

The oscilloscope can be connected to a controller or a PC through the RS-232
connector on the rear of the oscilloscope. An RS-232 cable is shipped with each
54622A/22D/24A oscilloscope and may be purchased for the 54621A/21D
oscilloscopes.
1 Attach the 9-pin “D” connector on the RS-232 cable to the RS-232
connector on the rear of the oscilloscope. Tighten the thumbscrews on
the cable connector to secure the cable 2 Attach the other end of the cable to your controller or pc. 3 Set up the RS-232 configuration on the oscilloscope.
a Press the Utility key, then press the I/O softkey.
b Press the Controller softkey and select RS-232.
c Press the Baud softkey and set the baud rate to match your controller or pc.
d Press the XON DTR softkey and set the handshake to match your controller
or pc.
For more information on RS-232 configuration, refer to the “Utilities” chapter
in the User’s Guide.
2-12
Preparing the Oscilloscope for Use

To verify basic oscilloscope operation

To verify basic oscilloscope operation
1 Connect an oscilloscope probe to channel 1. 2 Attach the probe to the Probe Comp output on the lower-right side of
the front panel of the oscilloscope.
Use a probe retractable hook tip so you do not need to hold the probe.
3Press the Save/Recall key on the front panel, then press the Default Setup
softkey under the display.
The oscilloscope is now configured to its default settings.
4Press the Autoscale key on the front panel.
You should then see a square wave with peak-to-peak amplitude of about 5 divisions and a period of about 4 divisions as shown below. If you do not see the waveform, ensure your power source is adequate, the oscilloscope is properly powered-on, and the probe is connected securely to the front-panel channel input BNC and to the Probe Comp calibration output.
Verifying Basic Oscilloscope Operation
2-13

Getting started using the oscilloscope interface

When the oscilloscope is first turned on, it performs a self-test, then
momentarily shows a startup screen as shown below.
This menu is only accessible when the oscilloscope first starts up.
2-14
Preparing the Oscilloscope for Use
To verify basic oscilloscope operation
•Press the Getting Started softkey to view the symbols used in the oscilloscope softkey menus.
Use the Entry knob labeled to adjust the parameter.
Press the softkey to display a pop up with a list of choices. Repeatedly press the softkey until your choice is selected.
Use the Entry knob labeled or press the softkey to adjust the parameter.
Option is selected and operational.
Feature is on. Press the softkey again to turn the feature off.
Feature is off. Press the softkey again to turn the feature on.
Press the softkey to view the menu.
Press the softkey to return to the previous menu.
Links you to another menu.
2-15

Using Quick Help

The oscilloscope has a Quick Help system that provides user help for each front-panel key and softkey on the oscilloscope. To view Quick Help information:
1Press 2 Release the key after reading the message. Releasing the key returns
and hold down the key for which you would like to view help.
the oscilloscope to the previous state.

Selecting a language for Quick Help when the oscilloscope starts up

When the oscilloscope first powers up, you can press the Language softkey to select a language for viewing Quick Help. Successive press the Language s of tk ey until the desired language in the list selected.
You can also select and load a language later from the Utility Language menu.
2-16
Preparing the Oscilloscope for Use

Selecting a language for Quick Help after you have been operating the oscilloscope

Selecting a language for Quick Help after you have been operating the oscilloscope
1Press the Utility key, then press the Language softkey to display the
Language menu.
2Press the
Language softkey until the desired language in the list selected.
If the language you want to load is grayed-out in the list, you will need to load the language from floppy disk. The language file can be downloaded from www.agilent.com/find/5462xsw disk for your instrument
or call an A gilent cen ter an d req uest a lan guage
2-17
Preparing the Oscilloscope for Use

Loading a language from floppy disk

Loading a language from floppy disk
Language files can be downloaded from www.agilent.com/find/5462xsw or call an Agilent center and request a language disk for your instrument.
1 Insert the floppy disk with a language file into the floppy disk drive on
the oscilloscope.
2Press the
Language menu.
3Press the
the oscilloscope.
4Press the
For more information about loading and deleting languages, refer to the “Utilities” chapter in the User’s Guide.
Utility key, then press the Language softkey to display the
Load Languages softkey to load the updated language file into
Language softkey and select the language to be viewed.
2-18

Cleaning the oscilloscope

1 Disconnect power from the instrument.
CAUTION Avoid Damage to Sensitive Electronic Components!
Do not use too much liquid in cleaning the oscilloscope. Water can enter the front-panel keyboard, control knobs, or floppy disk damaging sensitive electronic components.
2 Clean the oscilloscope with a soft cloth dampened with a mild soap and
water solution.
3 Make sure that the instrument is completely dry before reconnecting to
a power source.
2-19
3

Testing Performance

Testing Performance
This chapter explains how to verify correct oscilloscope operation and perform tests to ensure that the oscilloscope meets the performance specifications.
To completely test and troubleshoot the mixed-signal oscilloscope, you will create and use a test connector accessory, as described in this chapter.
• The test connector makes it easy for you to connect the oscilloscope probes to function generators and measurement equipment with minimum electrical distortion.
• The connector is used in the digital channel threshold accuracy test.
Let the Equipment Warm Up Before Testing For accurate test results, let the test equipment and the oscilloscope warm up 30 minutes before testing.
Verifying Test Results During the tests, record the readings in the Performance Test Record at the end of this chapter for your oscilloscope. To verify whether a test passes, verify that the reading is within the limits in the Performance Test Record.
If a performance test fails
If a performance test fails, first perform the User Cal procedure given in Chapter 4. If the User Cal procedure does not correct the problem , refer to Chapter 4, Calibrating and Adjusting.
3-2
Testing Performance

List of Test Equipment

List of Test Equipment
Below is a list of test and equipment and accessories required to perform the performance test verification procedures.
Equipment Critical Specifications Recommended Model/
Test connector, 8-by-2 Not required if 01660-63801 Test Fixture is available
See To construct the test connector
later in this chapter Test fixture For testing digital channel threshold accuracy Agilent 01660-63801 Digital Multimeter 0.1 mV resolution, 0.005% accuracy Agilent 34401A Power Splitter Outputs differ by 0.15 dB Agilent 11667B Power Supply DC offset voltage of -5.5 V to 35.5 V, 0.1 mV resolution Agilent 3245A Oscilloscope Calibrator 25 MHz100 MHz sine wave, 5 ppm Fluke 5820A BNC banana cable Agilent 11001-60001 BNC cable (qty 3) Agilent 10503A Probe cable Agilent 01650-61607 Shorting Cap BNC Agilent 1250-0774 Adapter BNC(f) to banana(m) Agilent 1251-2277 Feedthrough 50 Adapter BNC Tee (m) (f) (f) Agilent 1250-0781 Blocking capacitor Agilent 10240-60001 Adapter (qty 3) N(m) to BNC(f) Agilent 1250-0780
, BNC (m) and (f) Agilent 11048C
Part Number
n/a
3-3
Testing Performance

To construct the test connector

To construct the test connector
The Agilent 54621D/22D Mixed-Signal Oscilloscope has digital channels that you will need to connect to test equipment during testing. To easily connect the digital channels, you will construct a test connector only if the 01660-63801 Test Fixture is not available.
Construct Test Connector only if Test Fixture is not available
The test connector is not required if 01660-63801 Test Fixture is available.
Table 3-1
Figure 3-1
Materials Required to Construct the Test Connectors
Description Recommended Part Qty
BNC (f) Connector Agilent 1250-1032 1 Berg Strip, 8-by-2 1 Jumper wire
1
Obtain a BNC connector and an 8-by-2 section of Berg strip.
2 On one side of the Berg strip, solder a jumper wire to all of the pins. 3 On the other side of the Berg strip, solder another jumper wire to all of
the pins.
4 Solder the center of the BNC connector to a center pin on one of the
rows on the Berg strip.
5 Solder the ground tab of the BNC connector to a center pin on the other
row on the Berg strip.
Constructing the 8-by-2 Connector
3-4
Testing Performance

To test the 54621D/22D Oscilloscope digital channels

To test the 54621D/22D Oscilloscope digital channels
Only the 54621D/22D Mixed-Signal Oscilloscope has Digital Channels
You need to perform these instructions only if you will be testing the digital channels on the Agilent 54621D/22D Mixed-Signal Oscilloscope.
The acquisition system testing provides confidence that the acquisition system is functioning correctly. It does not, however, check a particular specification.
1
Disconnect all probes from the circuit under test and from any other input source.
2 Using probe leads and grabbers, connect digital channels D0, D1, D2,
and D3 to the calibration point on the 54621D/22D front panel.
3 Press the Autoscale key.
If four square waves appear, the acquisition system is functioning correctly. If the square waves do not appear, go to the “Troubleshooting” chapter. Then
return here to finish testing the digital channels.
4
Disconnect the digital channels from the calibration point.
5 Use steps 2 and 3 to test the following sets of digital channels. After you
test one set of digital channels, remove them before connecting the next set.
• D4, D5, D6, D7
• D8, D9, D10, D11
• D12, D13, D14, D15
3-5
Testing Performance

To verify digital channel threshold accuracy

To verify digital channel threshold accuracy
This test verifies the digital channel threshold accuracy specification of the Agilent 54621D/22D Mixed-Signal Oscilloscope.
Test Threshold Accuracy only on the 54621D/22D Mixed-Signal Oscilloscope
You need to perform these instructions only if you will be testing the Agilent 54621D/22D Mixed-Signal Oscilloscope.
Threshold accuracy test limits= ±(100 mV + 3% of threshold setting)
When to Test You should perform this test every 24 months or after 4000 hours of operation, whichever comes first.
What to Test Use these instructions to test the threshold settings of digital channels D7-D0. Then, use the same instructions to test digital channels D15-D8.
Table 3-2
Verifying Test Results After each threshold test, record the voltage reading in the Performance Test Record at the end of this chapter. To verify whether a t e st p a s se s , v er i fy t h at t he v o l ta g e r e a d in g i s wi t h i n th e l im i ts i n t he P e r fo r ma n c e Test Rec ord.
Equipment Required to Test Threshold Accuracy
Equipment Critical Specifications Recommended Model/Part
Digital Multimeter 0.1 mV resolution, 0.005%
accuracy Oscilloscope Calibrator DC offset voltage 6.3 V Fluke 5820A BNC-Banana Cable Agilent 11001-60001 BNC Tee Agilent 1250-0781 BNC Cable Fluke 50 BNC Test Connector, 8-by-2 User-built (See Obtain a BNC
Test Fixture PV test fixture Agilent 01660-63801 Probe Cable Agilent 01650-61607
Agilent 34401A
cable, P/N 686318
connector a nd an 8-by-2 section of Berg strip. on page 3-4.)
3-6
Figure 3-2
Save/Recall
Mixed Signal Oscilloscope
Trigg er ou t
!
!
Testing Performance
To verify digital channel threshold accuracy
1
Turn on the test equipment and the oscilloscope. Let them warm up for 30 minutes before starting the test.
2 Set up the oscilloscope calibrator.
a Set the oscilloscope calibrator to provide a DC offset voltage at the
Channel 1 output.
b Use the multimeter to monitor the oscilloscope calibrator DC output
voltage.
3
Use either method 1 or method 2, described in the following, to connect the digital channels for testing.
a Method 1 — Using the Test Connector
Use the 8-b y- 2 t est co nn ec to r an d th e B NC cab le as sem bl y to co nn ec t di gi tal channels D0-D7 to one side of the BNC Tee. Then connect the D0-D7 ground lead to the ground side of the 8-by-2 connector. See figure 3-2.
Oscilloscope Calibrator
Digital Multimeter
54620A
16 CHANNEL 500 MSa/s
HP 34401A
Channels 8 - 15
BNC-Banana cable
Test Connector
Setting Up Equipment and Test Connector for the Threshold Test
Ext trigge r in
STORAGE
Entry
TRIGGERHORIZONTAL
Delay
Time/Div
INPU TS
!
Measure time
CHANNEL
Select
Position
Line
Channels 0 - 7
hrh.r
3-7
Figure 3-3
Save/Recall
Mixed Signal Oscillos cope
STORAGE
TRIGGER
Trigger out
!
Delay
!
!
Digital Multimeter
Oscilloscope Calibrator
Testing Performance
To verify digital channel threshold accuracy
b Method 2 — Using the Test Fixture
Use the Agilent 01660-63801 Performance Verification Test Fixture and the Agilent 01650-61607 cable, BNC Tee, and BNC cable to connect the digital channels D0 - D15 to the oscilloscope calibrator. See figure 3-3.
$
16 CHANNEL 500 MSa/s
HP 34401A
01650-61607 Cable
Measure time
Entry
HORIZONTAL
CHANNEL
Time/Div
Select
INPUTS
Posit io n
Ext trig ger in
Line
01660-63801
BNC-Banana
Test Fixture
cable
Setting Up Equipment and Test Fixture for the Threshold Test
4 Use a BNC-banana cable to connect the multimeter to the other side of
the BNC Tee.
5 Connect the BNC Tee to the Channel 1 output of the calibrator as shown
in figure 3-2 and figure 3-3.
6 On the oscilloscope, press the D7 Thru D0 key, then press the Threshold
softkey.
3-8
To verify digital channel threshold accuracy
7
Press the oscilloscope User softkey, then turn the Entry knob () on
the front panel
on the oscilloscope to set the threshold test settings as
shown in Table 3-3.
Table 3-3 Threshold Accuracy Voltage Test Settings
Testing Performance
Threshold voltage setting (in oscilloscpe User softkey)
+5.00 V +5.250 V ±1 mV dc Lower limit = +4.750 V
–5.00 V –4.750 V ±1 mV dc Lower limit = –5.250 V
0.00 V +100m V ±1 mV dc Upper limt = +100 mV
8
Do the following steps for each of the threshold voltage levels shown
DC offset voltage setting (on oscilloscope calibrator)
Limits
Upper limit = +5.250 V
Upper limit = –4.750 V
Lower limit = –100 mV
in Table 3-3.
a Set the threshold voltage shown in the User softkey using the Entry knob on
the oscilloscope.
b Enter the corresponding DC offset voltage on the oscilloscope calibrator
front panel. Then use the multimeter to verify the voltage.
Digital channel activity indicators are displayed on the status line at the top of the oscilloscope display. The activity indicators for D7-D0 should show all of the channels at digital high levels.
c Use the knob on the oscilloscope calibrator to decrease the offset voltage,
in increments of 10 mV, until the activity indicators for digital channels D7-D0 are all at digital low levels. Record the oscilloscope calibrator voltage in the performance test record.
d Use the knob on the oscilloscope calibrator to increase the offset voltage,
in increments of 10 mV, until the activity indicators for digital channels D7-D0 are all at digital high levels. Record the oscilloscope calibrator voltage in the performance test record.
Before proceeding to the next step, make sure that you have recorded the oscilloscope calibrator voltage levels for each of the threshold settings shown in Table 3-3.
Use the 8-by-2 test connector or the Agilent 01660-63801 Test Fixture
9
to connect digital channels D15-D8 to the output of the oscilloscope calibrator. Then connect the D15-D8 ground lead to the ground side of the 8-by-2 connector.
10 Repeat this procedure for digital channels D15-D8 to verify threshold
accuracy and record the threshold levels in the Performance Test Record.
3-9
Testing Performance

To verify voltage measurement accuracy

To verify voltage measurement accuracy
This test verifies the voltage measurement accuracy. In this test, you will measure the output of a power supply using dual cursors on the oscilloscope, and compare the results with the multimeter reading.
Test limits: ±2.0% of full scale ±1 LSB*
• Full scale is defined as 16 mV on the 1 mV/div range.
• Full scale on all other ranges is defined as 8 divisions times the V/div setting.
*1 LSB = 0.4% of full scale
Table 3-4
Equipment Required to Verify Voltage Measurement Accuracy
Equipment Critical Specifications Recommended Model/Part
Power supply 14 mV to 35 Vdc,
0.1 mV resolution Digital multimeter Better than 0.01% accuracy Agilent 34401A Cable BNC, Qty 2 Agilent 10503A Shorting cap BNC Agilent 1250-0774 Adapter BNC (f) to banana (m) Agilent 1251-2277 Adapter BNC tee (m) (f) (f) Agilent 1250-0781 Blocking capacitor Agilent 10240B
Fluke 5820A or Agilent 3245A
Do this procedure first for Channel 1. Then repeat the procedure for Channel 2.
1
Set up the oscilloscope.
a Adjust the channel 1 position knob to place the baseline at approximately
0.5 division from the bottom of the display.
3-10
Testing Performance
To verify voltage measurement accuracy
b Set the Volts/Div setting to the value in the first line in Table 3-5.
Table 3-5
Settings Used to Verify Voltage Measurement Accuracy
Volts/Div Setting Power Supply Setting Test Limits
5 V/Div 35 V 34.04 V to 35.96 V 2 V/Div 14 V 13.616 V to 14.384 V 1 V/Div 7 V 6.808 V to 7.192 V
0.5 V/Div 3.5 V 3.404 V to 3.596 V
0.2 V/Div 1.4 V 1.3616 V to 1.4384 V
0.1 V/Div 700 mV 680.8 mV to 719.2 mV 50 mV/Div 350 mV 340.4 mV to 359.6 mV 20 mV/Div 140 mV 136.16 mV to 143.84 mV 10 mV/Div 70 mV 68.08 mV to 71.92 mV 5 mV/Div 35 mV 34.04 mV to 35.96 mV 2 mV/Div 14 mV 13.616 mV to 14.384 mV 1 mV/Div* 7 mV 6.616 mV to 7.384 mV
*Full scale is defined as 16 mV on the 1 mV/div range.
Full scale on all other ranges is defined as 8 divisions times the V/div setting.
c Press the Acquire key. Then press the Averaging softkey and set #Avgs to 64.
Wait a few seconds for the measurement to settle.
2
Press the Cursors key, set the Mode softkey to Normal, then press the XY softkey and select (labeled
Y. Press the Y1 softkey, then use the Entry knob
on the front panel) to set the Y1 cursor on the baseline of
the signal.
3-11
Testing Performance
To verify voltage measurement accuracy
3
Use the BNC tee and cables to connect the oscilloscope calibrator /power supply to both the oscilloscope and the multimeter.
4 Adjust the output so that the multimeter reading displays the first
Volts/div supply setting value in Table 3-5.
Wait a few seconds for the measurement to settle.
5 Press the Y2 softkey, then position the Y2 cursor to the center of the
voltage trace using the Entry knob.
The Y value on the lower line of the display should be within the test limits of Table 3-5. If a result is not within the test limits, see the “Troubleshooting” chapter. Then return here.
6
Continue to check the voltage measurement accuracy with the remaining Volts/div setting values in Table 3-5.
7 When you are finished checking all of the power supply setting values,
disconnect the power supply from the oscilloscope.
8 Repeat this procedure for Channels 2, 3, and 4, if applicable on your
oscilloscope model.
Use a Blocking Capacitor to Reduce Noise
On the more sensitive ranges, such as 1 mV/div, 2 mV/div, and 5 mV/div, noise may be a factor. To eliminate the noise, use a BNC Tee, blocking capacitor, and BNC shorting cap to shunt the noise to ground. See figure 3-4.
Figure 3-4
To Power Supply or Calibrator
To oscilloscope input
Using a Blocking Capacitor to Reduce Noise
3-12
Blocking Capacitor
BNC shorting cap
Testing Performance

To verify bandwidth

To verify bandwidth
This test verifies bandwidth. In this test you will use an oscilloscope calibrator with a level sinewave output.
Yo u wi l l u s e t h e p e ak - to - pe a k v o lt a ge b ot h at 1 M Hz an d at ba n dw i dt h fr e qu e nc y to verify the bandwidth response of the oscilloscope.
54622A, 54622D, and 54624A
Test limits at 1 mV/div to 5 V/div:
• All channels (±3 dB)
• dc to 100 MHz
• ac coupled 10 Hz to 100 MHz
54621A and 54621D
Test limits at 1 mV/div to 5 V/div:
• All channels (±3 dB)
• dc to 60 MHz
• ac coupled 10 Hz to 60 MHz
Table 3-6
Equipment Required to Verify Bandwidth
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator Fluke 5820A Cable * Type N (m), 24-inch Agilent 11500B Feedthrough 50 * The oscilloscope calibrator is supplied with 2 or more coaxial cables N (m), BNC (m),
1 meter long, Fluke P/N 686318.
, BNC (m) and (f) Agilent 11048C
3-13
Testing Performance
To verify bandwidth
1
Connect the oscilloscope calibrator output through a 50 feedthrough to the oscilloscope channel 1 input.
2 Set up the oscilloscope.
a Set the time base to 500 ns/div. b Set the Volts/Div for channel 1 to 200 mV/div. c Press the Acquire key, then press the Averaging softkey. d Turn the Entry knob to set #Avgs to 8 averages.
Set the calibrator to “Level Sine” and OPR/STBY to “OPR”.
3 4 Set the calibrator for 1 MHz and six divisions of amplitude. 5 Press Autoscale on the oscilloscope. 6 Press the Quick Meas key, then press the Peak-Peak softkey.
Wait a few seconds for the measurement to settle (averaging is then complete). View the Pk-Pk reading at the bottom of the display.
Record the reading: V
7
Change the frequency of the signal generator to the value shown below
= _______ V.
p-p
for your instrument.
Table 3-7 Signal Generator Frequency Setting
Selected Channel 54621A/21D 54622A/22D 54624A Channel 1 60 MHz 100 MHz 100 MHz Channel 2 60 MHz 100 MHz 100 MHz Channel 3 ——100 MHz Channel 4 ——100 MHz
8
Change the time base to 5 ns/div.
a Wait a few seconds for the measurement to settle. b View the Pk-Pk reading at the bottom of the display. c Record the reading: V
9
Calculate the response using this formula:
Step8Result
20 10log
------------ ------------- -------
Step6Result
= ______ mV.
p-p
If the result is not ±3 dB, see the “Troubleshooting” chapter. Then return here.
10 Repeat this procedure (steps 1 to 9) for channel 2, 3, and 4, as applicable
to your oscilloscope model.
Proceed to the next step after you have completed the procedure for channels 2, 3, and 4, as applicable.
3-14
Testing Performance

To verify horizontal Dt and 1/Dt accuracy

To verify horizontal t and 1/t accuracy
This test verifies the horizontal t and 1/t accuracy. In this test, you will use the oscilloscope to measure the output of a time mark generator.
Test limits: ±0.01% of reading ±0.1% of full scale ±40 ps (same channel)
Table 3-8
Equipment Required to Verify Horizontal t and 1/t Accuracy
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator Stability 5 ppm after 1/2 hour Fluke 5820A Cable BNC, 3 feet Agilent 10503A Feedthrough 50
1
Connect the oscilloscope calibrator to channel 1 using the 50 feedthrough at the oscilloscope input. Then, select calibrator for 100
2 Set up the oscilloscope.
a Press the Display key, then set the Vectors softkey to off. b Press the Autoscale key. c Set the time base to 20 µs/div. d Press the Main/Delayed key, then set the Time Ref softkey to Left. e Adjust the Trigger Level knob to obtain a stable display.
Press the Quick Meas softkey, set the Source softkey to 1, then select and
3
measure
Frequency 10 kHz — The test limits are 9.98 kHz to 10.02 kHz. Period 100 µs — The test limits are 99.79 µs to 100.2 µs.
If the measurements are not within the test limits, see the “Troubleshooting” chapter. Then return here.
Frequency and Period. Measure the following:
, BNC connectors (m) and (f) Agilent 11048C
Marker and set the
µs markers.
3-15
54622A/22D/24A only
54621A and 54621D only
Testing Performance
To verify horizontal Dt and 1/Dt accuracy
4
Change the calibrator to 1-µs markers. Change the time base to 200 ns/div. Adjust the trigger level to obtain a stable display.
5 Measure the following:
Frequency 1 MHz — The test limits are 997.9 kHz to 1.002 MHz. Period 1 µs — The test limits are 997.9 ns to 1.002 µs.
If the measurements are not within the test limits, see the “Troubleshooting” chapter. Then return here.
Change the calibrator to 10-ns markers. Change the time base to
6
5 ns/div. Adjust the trigger level to obtain a stable display.
7 Measure the following:
Frequency 100 MHz — The test limits are 99.10 MHz to 100.9 MHz. Period 10 ns — The test limits are 9.91 ns to 10.09 ns.
If the measurements are not within the test limits, see the “Troubleshooting” chapter. Then return here.
8
Change the calibrator to 20 ns markers. Change the time base to 5 ns/div. Adjust the trigger level to obtain a stable display.
9 Measure the following:
Frequency 50 MHz — The test limits are 49.77 MHz to 50.23 MHz. Period 20 ns — The test limits are 19.91 ns to 20.09 ns.
If the measurements are not within the test limits, see the “Troubleshooting” chapter. Then return here.
3-16

To verify trigger sensitivity

Th is test verifie s the tr igger s ensiti vity. I n this test, you will apply 25 MHz to the oscilloscope. You will then decrease the amplitude of the signal to the specified levels, and check to see if the oscilloscope is still triggered. You will then repeat the process at the upper bandwidth limit.
Test limits for the Internal trigger:
1 mV to 5 V/div (dc to max bandwidth): greater of 0.35 div or 2.5 mV
Test limits for the External trigger:
dc to 25 MHz: <75 mV 25 MHz to max bandwidth: <150 mV
Table 3-9 Equipment Required to Verify Trigger Sensitivity
p-p
p-p
Testing Performance
To verify trigger sensitivity
p-p
Equipment Critical Specifications Recommended Model/Part
Oscilloscope Calibrator 25-MHz, 60-MHz and 100-MHz sine
waves Power splitter Outputs differ < 0.15 dB Agilent 11667B Cable * BNC, Qty 3 Agilent 10503A Adapter N (m) to BNC (f), Qty 3 Agilent 1250-0780 Feedthrough 50 * The oscilloscope calibrator is supplied with 2 or more coaxial cables N (m), BNC (m),
1 meter long, Fluke P/N 686318.
, BNC connectors (m) and (f) Agilent 11048C (2 required)
Fluke 5820A
3-17
Testing Performance
To verify trigger sensitivity
Test Internal Trigger Sensitivity
1
Press the Save/Recall key, then press the Default Setup softkey.
2 Connect the calibrator to channel 1 using a 50 feedthrough at the
oscilloscope input.
3 Verify the trigger sensitivity at 25 MHz and 0.35 divisions.
a Set the output of the calibrator to 25 MHz, and set the amplitude to about
100 mV
b Press the Autoscale key. c Set the time base to 50 ns/div. d Set channel 1 to 100 mV/div. e Decrease the output of the calibrator until 0.35 vertical divisions of the
p-p
.
signal are displayed.
The trigger should be stable. If the trigger is not stable, try adjusting the trigger level. If adjusting the trigger level makes the trigger stable, the test still passes. If adjusting the trigger does not help, see the “Troubleshooting” chapter. Then return here.
f Record the result as Pass or Fail in the Performance Test Record.
Verify the trigger sensitivity at maximum bandwidth and 0.35 division.
4
a Change the output of the calibrator to 100 MHz for the 54622A/22D/24A or
60 MHz or the 54621A/21D, and set the amplitude to about 100 mV
b Set the time base to 10 ns/div. c Decrease the output of the calibrator until 0.35 vertical divisions of the
p-p
.
signal is displayed.
The trigger should be stable. If the trigger is not stable, try adjusting the trigger level. If adjusting the trigger level makes the trigger stable, the test still passes. If adjusting the trigger does not help, see the “Troubleshooting” chapter. Then return here.
d Record the result as Pass or Fail in the Performance Test Record.
Repeat this procedure for channels 2, 3, and 4, as applicable to your
5
oscilloscope model.
3-18
Testing Performance
To verify trigger sensitivity
Test External Trigger Sensitivity
Verify the external trigger sensitivity at these settings:
100 MHz (54622A/22D/24A), <150 mV 60 MHz (54621A/21D), <150 mV 25 MHz (All models), <75 mV
p-p
p-p
p-p
1 Use the power splitter to connect the calibrator to both the channel 1
input and the external trigger input. The Ext Trigger input is on the rear panel of the mixed-signal oscilloscope and the 4-channel oscilloscope. Connect 50
2 Change the output of the calibrator to 100 MHz for the 54622A/22D/24A
or 60 MHz for the 54621A/21D, and set the amplitude to 106 mV (300 mV
The power splitter divides the 300 mV of the oscilloscope inputs.
3
Press the Autoscale key.
4 Press the Trigger Edge key, then press the Ext softkey to set the trigger
source to
5 Check for stable triggering, and adjust the trigger level if necessary. 6 Record the results as Pass or Fail in the Performance Test Record.
If the test fails, see the “Troubleshooting” chapter. Then return here.
7 Change the output of the calibrator to 25 MHz and set the amplitude to
25.74 mV
8 Check for stable triggering, and adjust the trigger level if necessary. 9 Record the results as Pass or Fail in the Performance Test Record.
If the test fails, see the “Troubleshooting” chapter. Then return here.
feedthroughs to the oscilloscope inputs.
).
p-p
so that 150 mV
p-p
external trigger.
rms
(75 mV
p-p
).
rms
is applied to each
p-p
3-19

Agilent 54622A/22D/24A Performance Test Record

Agilent 54622A/54622D/54624A
Serial No. ______________________________________ Test by _____________________________ Test Interval ____________________________________ Work Order No. ______________________ Recommended Next Testing ________________________ Temperature ____________
Threshold Specification Limits Ch D7-D0 Ch D15-D8 Accuracy Test
(100 mV + 3% of threshold setting)
Voltage Measurement Accuracy Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3 Channel 4
5 V/Div 35 V 34.04 V to 35.96 V ________ ________ ________ ________ 2 V/Div 14 V 13.616 V to 14.384 V ________ ________ ________ ________ 1 V/Div 7 V 6.808 V to 7.192 V ________ ________ ________ ________ 500 mV/Div 3.5 V 3.404 V to 3.596 V ________ ________ ________ ________ 200 mV/Div 1.4 V 1.3616 V to 1.4384 V ________ ________ ________ ________ 100 mV/Div 700 mV 680.8 mV to 719.2 mV ________ ________ ________ ________ 50 mV/Div 350 mV 340.4 mV to 359.6 mV ________ ________ ________ ________ 20 mV/Div 140 mV 136.16 mV to 143.84 mV ________ ________ ________ ________ 10 mV/Div 70 mV 68.08 mV to 71.92 mV ________ ________ ________ ________ 5 mV/Div 35 mV 34.04 mV to 35.96 mV ________ ________ ________ ________ 2 mV/Div 14 mV 13.616 mV to 14.384 mV ________ ________ ________ ________ 1 mV/Div 7 mV 6.616 mV to 7.384 mV ________ ________ ________ ________
5 V - 250 mV 4.750 V ________ ________ 5 V + 250 mV 5.250 V ________ ________
-5 V - 250 mV -5.250 V ________ ________
-5 V + 250 mV -4.750 V ________ ________ 0 V - 100 mV -100 mV ________ ________ 0 V + 100 mV 100 mV ________ ________
Bandwidth Test Limits Channel 1 Channel 2 Channel 3 Channel 4
Horizontal t and 1/t Accuracy
Frequency 10 kHz 9.98 kHz to 10.02 kHz ________ Period 100 Frequency 1 MHz 997.9 kHz to 1.002 MHz ________ Period 1 Frequency 100 MHz 99.10 MHz to 100.9 MHz ________ Period 10 ns 9.91 ns to 10.09 ns ________
Trigger Sensitivity Test Limits Channel 1 Channel 2 Channel 3 Channel 4
Generator Setting Test Limits Results
µs 99.79 µs to 100.2 µs ________
µs 997.9 ns to 1.002 µs ________
Internal trigger 25 MHz at 0.35 division ________ ________ ________ ________
External trigger 25 MHz at <75 mV
3 dB at 100 MHz ________ ________ ________ ________
100 MHz at 0.35 division ________ ________ ________ _______
Ext
________
p-p
100 MHz at <150 mV
p-p
________
3-20

Agilent 54621A/21D Performance Test Record

Agilent 54621A/54621D
Serial No. ______________________________________ Test by _____________________________ Test Interval ____________________________________ Work Order No. ______________________ Recommended Next Testing ________________________ Temperature ____________
Threshold Specification Limits Ch D7-D0 Ch D15-D8 Accuracy Test (100 mV + 3% of threshold setting)
Voltage Measurement Accuracy Range Power Supply Setting Test Limits Channel 1 Channel 2 Channel 3 Channel 4
5 V/Div 35 V 34.04 V to 35.96 V ________ ________ ________ ________ 2 V/Div 14 V 13.616 V to 14.384 V ________ ________ ________ ________ 1 V/Div 7 V 6.808 V to 7.192 V ________ ________ ________ ________ 500 mV/Div 3.5 V 3.404 V to 3.596 V ________ ________ ________ ________ 200 mV/Div 1.4 V 1.3616 V to 1.4384 V ________ ________ ________ ________ 100 mV/Div 700 mV 680.8 mV to 719.2 mV ________ ________ ________ ________ 50 mV/Div 350 mV 340.4 mV to 359.6 mV ________ ________ ________ ________ 20 mV/Div 140 mV 136.16 mV to 143.84 mV ________ ________ ________ ________ 10 mV/Div 70 mV 68.08 mV to 71.92 mV ________ ________ ________ ________ 5 mV/Div 35 mV 34.04 mV to 35.96 mV ________ ________ ________ ________ 2 mV/Div 14 mV 13.616 mV to 14.384 mV ________ ________ ________ ________ 1 mV/Div 7 mV 6.616 mV to 7.384 mV ________ ________ ________ ________
Bandwidth Test Limits Channel 1 Channel 2
Horizontal t and 1/t Accuracy
Frequency 10 kHz 9.98 kHz to 10.02 kHz ________ Period 100 Frequency 1 MHz 997.9 kHz to 1.002 MHz ________ Period 1 Frequency 50 MHz 49.77 MHz to 50.23 MHz ________ Period 20 ns 19.91 ns to 20.09 ns ________
5 V - 250 mV 4.750 V ________ ________ 5 V + 250 mV 5.250 V ________ ________
-5 V - 250 mV -5.250 V ________ ________
-5 V + 250 mV -4.750 V ________ ________ 0 V - 100 mV -100 mV ________ ________ 0 V + 100 mV 100 mV ________ ________
3 dB at 60 MHz ________ ________
Generator Setting Test Limits Results
µs 99.79 µs to 100.2 µs ________
µs 997.9 ns to 1.002 µs ________
Trigger Sensitivity Test Limits Channel 1 Channel 2
Internal trigger 25 MHz at 0.35 division ________ ________
60 MHz at 0.35 division ________ ________
External trigger 25 MHz at <75 mV
60 MHZ at <150 mV
p-p
p-p
Ext
________ ________
3-21
4

Calibrating and Adjusting

Calibrating and Adjusting
This chapter explains how to adjust the oscilloscope for optimum operating performance. You should perform the hardware adjustments and self-calibration according to the following recommendations.
• Perform hardware adjustments every 12 months or after 2,000 hours of operation.
• Perform self-calibration:
• every 6 months or after 1000 hours of operation
• if the ambient temperature is >10 °C from the calibration temperature
• if you want to maximize the measurement accuracy
The amount of use, environmental conditions, and experience with other instruments help determine if you need shorter adjustment intervals.
In this chapter, you will:
• Adjust the power supply
• Perform self-calibration
• Adjust the display
Let the Equipment Warm Up Before Adjusting
Before you start the adjustments, let the oscilloscope and test equipment warm up for at least 30 minutes.
Read All Cautions and Warnings
Read the following cautions and warning before making adjustments or performing self-calibration.
4-2
Calibrating and Adjusting
WARNING HAZARDOUS VOLTAGES !
Read the safety summary at the back of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with the protective covers removed. Only trained service personnel who are aware of the hazards involved should perform the maintenance. Whenever possible, perform the procedures with the power cord removed from the oscilloscope.
CAUTION REMOVE POWER TO AVOID DAMAGE !
Do not disconnect any cables or remove any assemblies with power applied to the oscilloscope. Otherwise, damage to the oscilloscope can occur.
CAUTION USE EXTERNAL FAN TO REDUCE TEMPERATURE !
When you must operate the oscilloscope with its cover removed, use an external fan to provide continuous air flow over the samplers. Air flow over the s am pl er s i s r ed uc ed wh en th e c ov er is re mo ve d, w hi ch le ad s t o h ig he r t ha n normal operating temperatures. Have the fan blow air across the system PC board where the heat sinks are located.
CAUTION AVOID DAMAGE TO ELECTRONIC COMPONENTS !
Electrostatic discharge (ESD) can damage electronic components. When you use any of the procedures in this chapter, use proper ESD precautions. As a minimum, place the oscilloscope on a properly grounded ESD mat and wear a properly grounded ESD strap.
4-3
Calibrating and Adjusting

To adjust the power supply

To adjust the power supply
The oscilloscope power supply has both a +3.3 V adjustment, and a +5.1 V / -5.2 V balance adjustment. Other oscilloscope voltages are based on the +3.3 V adjustment.
In this procedure, you will use a digital multimeter to measure the +3.3 V and +5.1 V / -5.2 V test points, and adjust the power supplies to be within tolerance, if necessary.
Table 4-1
Equipment Required to Adjust the Power Supply
Equipment Critical Specifications Recommended Model/Part
Digital multimeter 0.1 mV resolution, accuracy ±0.05% Agilent 34401A
1
Prepare the oscilloscope for the voltage adjustment.
a Turn off the oscilloscope and disconnect the power cable. b Remove the oscilloscope cover. For a list of parts, see the “Replaceable
Parts” chapter, then return here.
c Place the oscilloscope on its side. d Connect the negative lead of the digital multimeter to a ground point on the
oscilloscope chassis.
e Reconnect the power cable. f Turn on the oscilloscope.
4-4
Figure 4-1
Calibrating and Adjusting
To adjust the power supply
2
Locate the power supply voltages at E204, E205, E207, and E208 on the system board.
The power supply voltages are not labeled on the system board. See Figure 4-1 for the locations.
(
9
(
9
- -
(
9
9
(
-
Low Voltage Power Supply Voltages (on the bottom of the oscilloscope)
3 Make sure that the voltage measurements are within the tolerances
listed in Table 4-2.
Table 4-2 Power Supply Voltage Tolerances
Supply Voltage Tolerance
+5.1 V ±250 mV (4.85 V to 5.35 V)
-5.2 V ±156 mV (-5.04 V to -5.36 V) +15.75 V +1.260 V, -787 mV (+14.963 V to +17.010 V) +3.3 V ±100 mV (+3.20 V to +3.40 V) +8.2 V ±82 mV (+8.12 V to +8.28 V)
measure at J4 on the power supply board
-
4-5
Figure 4-2
Calibrating and Adjusting
To adjust the power supply
4
If the +5.1 V measurement is out of tolerance, adjust the +5.1 V / -5.2 V balance adjustment on the power supply. See Figure 4-2.
5 If the -5.2 V measurement is out of tolerance, adjust the
+5.1 V / -5.2 V balance adjustment on the power supply. See Figure 4-2.
The +15.75 V supply is not adjustable, and is dependent upon the +5.1 V supply.
+3.3V Adjustment
E2 and E3 +5.1 / -5.2 V Balance Adjustment
Fan Connector Printer Power
8.5 V DC
Low Voltage Power Supply Adjustments (on the top of the oscilloscope)
6 If adjusting the power supply does not bring all the voltages within
tolerance, go to the “Troubleshooting” chapter.
4-6
Calibrating and Adjusting

To perform User Cal

To perform User Cal
Perform self-calibration:
• every 6 months or after 1000 hours of operation.
• if the ambient temperature is >10 °C from the calibration temperature.
• if you want to maximize the measurement accuracy. The amount of use, environmental conditions, and experience with other
instruments help determine if you need shorter adjustment intervals.
Use r Cal perfor ms an i nternal self- alignm ent routi ne to optimi ze the sign al pat h in the oscilloscope. The routine uses internally generated signals to optimize circuits that affect channel sensitivity, offset, and trigger parameters. Disconnect all inputs and allow the oscilloscope to warm up before performing this procedure.
User Cal should be performed at least once a year, any time the ambient temperature of the oscilloscope has changed more than 10 °C since the last User Cal, or after any repair.
Performing User Cal will not invalidate your Certificate of Calibration. Successful completion of User Cal does not certify this oscilloscope with a National Institute of Standards and Technology (NIST) calibration.
1
Disconnect all inputs signals from the oscilloscope
2 Set the rear-panel CALIBRATION switch to UNPROTECTED. 3 Press the Utility key, then press the Service softkey. 4 Begin the Self Cal by pressing the User Cal softkey. 5 When the User Cal is completed, set the rear-panel CALIBRATION sw it c h
PROTECTED.
to
4-7
Calibrating and Adjusting

To adjust the oscilloscope display

To adjust the oscilloscope display
When to Adjust the Oscilloscope Display
The oscilloscope display normally does not require adjustment. However, when it is obvious that the display is out of adjustment, use this optional procedure.
In this procedure, you will make adjustments using two display patterns.
Table 4-3
Equipment Required to Adjust the Oscilloscope Display
Equipment Critical Specifications Recommended Model/Part
Digital multimeter Accuracy ±0.05,% 1 mV resolution Agilent 34401A
1
Connect the digital multimeter to the end of R901, located closest to the fuse, as shown in Figure 4-3.
2 Adjust +B.ADJ for +14.00 V.
In the remainder of this procedure, you will make adjustments to:
V.HOLD (vertical hold) INTENSITY (intensity control on oscilloscope front panel) H.HOLD (horizontal hold) FOCUS V.LIN (vertical linearity) V.SIZE (vertical size)
3
Adjust V.HOLD (vertical hold) for vertical synchronization.
4 Set the INTENSITY control (lower-left on the front panel) to
approximately 50%.
Do not adjust Sub Bri or Cont controls
Do not adjust the SUB BRT (sub bright) or CONTRAST Controls. These controls are preset at the factory to assure the correct number of display grayscales.
4-8
Figure 4-3
68%%57
)2&86
$'-
Calibrating and Adjusting
To adjust the oscilloscope display
5
Adjust H.HOLD (horizontal hold) to center the display horizontally.
6 Adjust FOCUS to achieve the best focus. 7 Adjust V.L IN (vertical linearity) to position all four corners of the half-
bright borders equally.
8 Adjust V.S IZ E (vertical size) to center the display vertically at the
maximum allowable size without losing the half-bright borders.
Because V.LIN and V.SIZE interact with each other, you may need to re-adjust the sizing and vertical centering of the pattern on the display.
++2/'
9+ 2 /'
96 , =(
9 / ,1
5
%
Oscilloscope Display Board Adjustments
)XVH
E
4-9
5

Troubleshooting

Troubleshooting
This chapter begins with suggestions for solving general problems that you may encounter with the oscilloscope. It tells you what to do in these cases:
• If there is no trace display
• If the trace display is unusual or unexpected
• If you cannot see a channel
• If you cannot get any response from the oscilloscope
Procedures for troubleshooting the oscilloscope follow the problem solving suggestions. The troubleshooting section shows you how to:
• Check out the oscilloscope
• Clear error messages from the display
• Check the low voltage power supply
• Run internal self-tests
Using the Dummy Load
Before you begin troubleshooting, you must either already have the dummy load that is mentioned, or you must construct one, as described in this chapter. You will use this dummy load while troubleshooting.
Read All Cautions and Warnings
Before you begin any troubleshooting, read all Warning and Cautions in the Troubleshooting section.
5-2

Solving General Problems with the Oscilloscopes

This section describes how to solve general problems that you may encounter while using the 54621A/22A/24A Oscilloscopes and the 54621D/22D Mixed-Signal Oscilloscopes to make measurements.
After troubleshooting the oscilloscope, if you need to replace parts, refer to the “Replaceable Parts” chapter.

If there is no trace display

Check that the power cord is connected to the oscilloscope and to a live power
source.
Check that the front-panel power switch is set to 1 (on).Check that the display is illuminated and that the
correctly.
If there is no display, go to the troubleshooting procedures in this chapter.Check that the oscilloscope probe lead wires are securely inserted into the
connector assembly and that the probe clips make good contact with the probe lead wires.
Check that the probe clips are securely connected to points in the circuit under
test, and that the ground is connected.
Check that the circuit under test is powered on.Press the Autoscale key.Reset the oscilloscope.
• To reset the entire oscilloscope, press and hold any softkey and switch the oscilloscope power on. Release the softkey when the display appears.
Obtain service from Agilent Technologies, if necessary.
INTENSITY knob is adjusted
5-3
Troubleshooting

If the trace display is unusual or unexpected

If the trace display is unusual or unexpected
Check that the Horizontal time/division setting is correct for the expected
frequency range of the input signals.
The sampling speed of the oscilloscope depends on the time/division setting.
It may be that when the time/division setting is slower than 1 ms/div, the oscilloscope is sampling too slowly to capture all of the transitions on the waveform.
Check that all oscilloscope probes are connected to the correct signals in the
circuit under test.
Check to see that the ground lead on the cable is securely connected to ground
in the circuit under test. For high-speed measurements, each probe’s individual ground lead should also be connected to a ground point closest to the signal point in the circuit under test.
Use chapter 2 for information on probing considerations.Check that the trigger setup is correct.A correct trigger setup is the most important factor in helping you capture the
data you desire. See the User’s Guide for information about triggering.
Check that infinite persistence in the Display menu is turned off, then press
the Clear Display softkey.
Press the Autoscale key.
5-4
Troubleshooting

If you cannot see a channel

If you cannot see a channel
Check that the oscilloscope probe cable is securely connected to the input
connector.
Check that the oscilloscope probe lead wires are securely inserted into the
connector assembly and that the probe clips make good contact with the probe lead wires.
Check that the probe clips are securely connected to points in the circuit under
test.
Check that the circuit under test is powered on.
You may have pressed the Autoscale key before an input signal was available. Performing the checks listed here ensures that the signals from the circuit under
test will be seen by the oscilloscope. Perform the remaining checks in this topic to make sure the oscilloscope channels are on, and to obtain an automatic setup.
Check that the desired oscilloscope channels are turned on.
a Press the analog channel key until it is illuminated. b Press the digital channel group key (D15 Thru D8 or D7 Thru D0 keys) until it
is illuminated. Use the Entry knob or Channel Select knob to scroll through the digital channels to make sure that the desired channels are turned on.
Press the Autoscale key to automatically set up all channels.
5-5

Troubleshooting the Oscilloscope

The service policy for the 54621A/22A/24A Oscilloscopes and the 54621D/22D Mixed-Signal Oscilloscopes is instrument replacement. Please call (877) 447-7278 to return the defective instrument, and Agilent Technologies will expedite a refurbished instrument to you under the Agilent Express Exchange program.
WARNING HAZARDOUS VOLTAGES EXIST — REMOVE POWER FIRST !
The maintenance described in this section is performed with power supplied to the oscilloscope and with the protective covers removed. Only trained service personnel who are aware of the hazards involved should perform the maintenance. Whenever possible, perform the procedures with the power cord removed from the oscilloscope. Read the safety summary at the back of this book before proceeding.
CAUTION REMOVE POWER TO AVOID DAMAGE !
Do not disconnect any cables or remove any assemblies while power is applied to the oscilloscope, or damage to the oscilloscope can occur.
CAUTION AVOID ESD DAMAGE TO COMPONENTS !
ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. Use proper ESD precautions when doing any of the procedures in this chapter. As a minimum, place the oscilloscope on a properly grounded ESD mat and wear a properly grounded ESD strap.
Equipment required for troubleshooting
The equipment listed in Table 5-1 is required to troubleshoot the oscilloscope.
Table 5-1
Equipment Required to Troubleshoot the Oscilloscope
Equipment Critical Specifications Recommended Model/Part
Digital multimeter Accuracy ±0.05,% 1 mV resolution Agilent 34401A Oscilloscope 100 MHz, 1 M Dummy load Compatible with power supply See To construct your own
5-6
input R Agilent 54622A
dummy load on page 5-7.
Figure 5-1
Troubleshooting

To construct your own dummy load

To construct your own dummy load
1 Obtain a connector that is compatible with the connector J3 on the Low
Voltage Power Supply.
2 Connect load resistors to the connector as follows:
Connector Load (Amps) Resistor Watts Connect to Pin Ground to Pin
+5.1 V 4.4 A 1.2 22 W 10, 11, 12 13, 14
-5.2 V 1.6 A 3.25 8 W 4, 5 6 +15.75 V 1.2 A 13.1 20 W 1 2 +3.3 V 4 A 0.825 13 W 15, 16, 17, 18 19, 20
+3.3V Adjustment
+5.1 / -5.2 V Balance Adjustment
Location of the Low Voltage Power Supply Connectors
Fan Connector Printer Power
E2 and E3
8.5 V DC
5-7
Troubleshooting

To check out the oscilloscope

To check out the oscilloscope
1 Is an interface module connected to the rear panel of the oscilloscope?
If the answer is yes, follow these steps. Otherwise, go to step 2.
a Turn off the oscilloscope. b Remove the interface module. c Turn on the oscilloscope, then check for the failing symptom. d If the failing symptom is not present, replace the interface module. e If the failing symptom is still present, go to step 2.
Disconnect any external cables from the front panel.
2 3 Disconnect the power cord, then remove the cabinet.
CAUTION USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
If you operate the oscilloscope with the cabinet removed, you must use an external fan to blow air across the system board. This external air flow is necessary to cool the heat sinks on the system board. Otherwise, damage to the components can occur.
4 Connect the power cord, then turn on the oscilloscope.
• If the oscilloscope display comes on after a few seconds, you will need to che ck the lo w volta ge power su pply. Go to “To check t he Low Volt age Pow er Supply” on page 5-11. Then return here.
• If, after checking the low voltage power supply, the voltages are within the test limits, go to step 8. If they are not within the test limits, go to step 6.
If the display did not come on, follow these steps:
a Disconnect the power cord. b Check to verify that all of the cable connections are secure. c Check the low voltage power supply. Go to “To check the Low Voltage
Power Supply” on page 5-11.
If the volt age s are wit hin the limits g o to st ep 5. If the vol tages are not within the limits, go to step 6.
5-8
To check out the oscilloscope
5
Disconnect the display ribbon cable from the system board. Then check the signals on the system board listed in Table 5-2.
Table 5-2 Display Signals on the System Board – All Oscilloscopes
Signal Name Frequency Pulse Width Voltage
J900 Pin 14 Hsync 19.72 kHz 3.0 ms 4.5 V J900 Pin 13 Vsync 60.00 Hz 253.5 ms 4.5 V
If the signals are good, replace the display assembly. If the signals are not good, replace the system board.
6
Disconnect the ribbon cable from the display board.
7 Check the low voltage power supply. Go to “To check the Low Voltage
Power Supply” on page 5-11.
If the voltages are within the test limits, replace the display assembly. If the voltages are not within the test limits, follow these steps:
a Disconnect the power cord. b Disconnect the ribbon cable from the power supply. c Connect the dummy load to the power supply connector. d Connect the power cord. Then measure the power supply voltages again.
See the new tolerances in Table 5-3.
Troubleshooting
p-p
p-p
Table 5-3
Low Voltage Power Supply Tolerances
Supply Voltage Tolerance
+5.1 V ±153 mV (+4.947 V to +5.253 V)
-5.2 V ±156 mV (-5.04 V to -5.36 V) +15.75 V +1.260 V, -787 mV (+14.963 V to +17.010 V) +3.3 V ±100 mV (+3.20 V to +3.40 V)
If the voltages are within the test limits now, replace the system board. If the voltages are still not within the test limits, replace the power supply.
5-9
Figure 5-2
Troubleshooting
To check out the oscilloscope
8
Is the fan running?
The low voltage power supply has a thermal cut-out circuit. If the fan is defective, the low voltage power supply will shut down when it gets too hot for safe operation.
If the fan is running, perform the internal self-tests. Go to “To run the internal self-tests” on page 5-13.
If the fan is not running, it may be defective. Follow these steps:
a Disconnect the fan cable from the power supply. b Measure the fan voltage at the connector on the power supply.
See Figure 5-2 for the location of the fan connector.
c If the fan voltage is approximately +8 Vdc, replace the fan. If the fan voltage
is not approximately +8 Vdc, replace the power supply.
+5.1 / -5.2 V Balance Adjustment
Location of the Fan Connector
+3.3V Adjustment
Fan Connector Printer Power
5-10
Troubleshooting

To check the Low Voltage Power Supply

To check the Low Voltage Power Supply
1 Disconnect the power cord from the oscilloscope. Then remove the
oscilloscope cover and set the oscilloscope on its side.
CAUTION USE AN EXTERNAL FAN TO AVOID OVERHEATING COMPONENTS !
If you operate the oscilloscope with the cabinet removed, you must use an external fan to blow air across the system board. This external air flow is necessary to cool the heat sinks on the system board. Otherwise, damage to the components can occur.
2 Connect the negative lead of the multimeter to a ground point on the
oscilloscope. Connect the power cord and turn on the oscilloscope.
3 Measure the power supply voltages at E1 through E6 on the system
board. See Figure 5-3 and Table 5-4.
Figure 5-3
- -
(
(
(
9
9
9
9
(
-
Low Voltage Power Supply Test Points
-
5-11
Troubleshooting
To check the Low Voltage Power Supply
Table 5-4 Low Voltage Power Supply Tolerances
Supply Voltage Tolerance
+5.1 V ±153 mV (+4.947 V to +5.253 V)
-5.2 V ±156 mV (-5.04 V to -5.36 V)
+15.75 V +1.260 V, -787 mV (+14.963 V to +17.010 V) +3.3 V ±100 mV (+3.20 V to +3.40 V)
4
If the +5.1 V measurement is out of tolerance, adjust the +5.1 V adjustment on the power supply.
5 If the -5.2 V measurement is out of tolerance, adjust the -5.2 V
adjustment on the power supply.
The +15.75 V supply is not adjustable, and is dependent upon the +5.1 V supply.
If the Fuse is Blown
If the power supply fuse is blown, the power supply is defective, and you must replace it. See the Replaceable Parts chapter for information about removing the power supply.
CAUTION BEWARE OF HAZARDOUS VOLTAGES !
Be careful when performing component-level repair. Voltages greater than 300 Vdc exist, and can cause injury.
5-12
Troubleshooting

To run the internal self-tests

To run the internal self-tests
Self Test performs a series of internal procedures to verify that the oscilloscope and any attached module are operating properly.
It is recommended that you run the Self Test:
• after experiencing abnormal operation
• for additional information to better describe an oscilloscope failure
• to verify proper operation after the oscilloscope has been repaired Successfully passing Self Test does not guarantee 100% of the oscilloscope’s
functionality. Self Test is designed to provide an 80% confidence level that the oscilloscope is operating properly.
Press the Utility key, then press the Service softkey.
1 2 Begin the internal self tests by pressing the Self Test softkey.
a The oscilloscope will display the message “Running self tests” on the screen. b The oscilloscope will then display one of the following messages on the
screen indicating the status of the self-tests:
• Self Tests Passed.
• Self Tests Failed.
5-13
6

Replacing Assemblies

Replacing Assemblies
This chapter describes how to remove assemblies from the oscilloscope. After you have removed an assembly, to install the replacement assembly, follow the instructions in reverse order.
The removable assemblies include:
•Cabinet
•Fan
• Floppy Drive
•Front panel
•Display
•System board
• Power supply
•Keyboard
• Handle
Tools Used for Disassembly
Use these tools to remove and replace the oscilloscope assemblies:
• T8, T10, T15, and T20 TORX drivers
• 9/16-inch socket driver.
See how the Oscilloscope Parts Fit Together
An exploded view of the oscilloscope is included in the “Replaceable Parts” chapter. It shows the individual part numbers used in the assemblies, and shows you how the parts fit together.
Read All Warnings and Cautions
Read the following warnings and cautions before removing and replacing any assemblies in the oscilloscope.
6-2
WARNING HAZARDOUS VOLTAGES !
Read the safety summary at the back of this book before proceeding.
Maintenance is performed with power supplied to the oscilloscope and with the protective covers removed. Only trained service personnel who are aware of the hazards involved should perform the maintenance. Whenever possible, perform the procedures with the power cord removed from the oscilloscope.
WARNING AVOID ELECTRICAL SHOCK !
Hazardous voltages exist on the CRT, power supply, and display sweep board. To avoid electrical shock:
1
Disconnect the power cord from the oscilloscope.
2 Wait at least three minutes for the capacitors in the oscilloscope to
discharge before you begin disassembly.
Read the Safety Summary at the back of this manual before you begin.
Replacing Assemblies
CAUTION REMOVE POWER TO AVOID DAMAGE !
Remove power before you begin to remove and replace assemblies. Do not remove or replace assemblies while the oscilloscope is turned on, or damage to the components can occur.
CAUTION AVOID DAMAGE TO ELECTRONIC COMPONENTS !
ELECTROSTATIC DISCHARGE (ESD) can damage electronic components. When doing any of the procedures in this chapter, use proper ESD precautions. As a minimum, you should place the instrument on a properly grounded ESD mat and wear a properly grounded ESD strap.
6-3
Figure 6-1
Replacing Assemblies

To remove the cabinet

To remove the cabinet
1 Turn off the oscilloscope and disconnect the power cable. 2 Rotate the handle up as shown below. 3 If a module is installed, remove it from the oscilloscope. 4 Using the T20 TORX driver, remove the two screws from the rear of the
cabinet.
5 Using your thumbs, gently push on the rear-panel connectors to slide
the oscilloscope out of the cabinet.
Removing the cabinet
6-4
Figure 6-2
Replacing Assemblies

To remove the fan

To remove the fan
1 Disconnect the fan cable from the power supply board. 2 Using the T20 TORX driver, remove the three screws that hold the fan
to the deck.
Removing the fan
6-5
Figure 6-3
Replacing Assemblies

To remove the floppy drive

To remove the floppy drive
1 Using the T8 TORX driver, loosen the two screws that hold the drive to
the deck.
2 Lift the drive away from the system board. 3 Disconnect the ribbon cable from the connector on the rear of the drive.
Removing the floppy drive
6-6
Figure 6-4
Replacing Assemblies

To remove the front panel

To remove the front panel
1 Remove the intensity knob and shaft by grasping the knob with one
hand and the grasping the shaft behind the front panel with the other hand and pulling the shaft straight out from the front panel.
2 Disconnect the keyboard ribbon cable W2 from the keyboard. 3 Use a screwdriver to release retainer tab A, and your finger to release
retainer tab B. See the tab locations in Figure 6-4.
Removing the intensity shaft and front panel
4 Swing the front panel out until the bottom clears the deck assembly,
then lift it up to free the hooks on top and pull it away from the deck.
When ins tal li ng th e fr ont pan el, mak e sur e th at th e power swit ch shaf t is alig ne d with the connection hole in the front panel to make a proper connection.
Before engaging the retainer tabs, make sure that all of the hooks on top of the front panel are fully engaged with their connection holes in the sheet metal. Then swing the front panel in to engage the two retainer tabs.
6-7
Figure 6-5
Replacing Assemblies

To remove the display

To remove the display
1 Remove the front panel from the oscilloscope. 2 Disconnect the ribbon cable from the display board. 3 Using the T20 TORX driver, remove the four screws that hold the display
to the deck. Set these screws aside.
Removing the display
CAUTION USE CORRECT SCREWS TO AVOID SHORTING SYSTEM BOARD !
When you re-install the display, you must re-insert the screws you removed in this step. If you re-insert screws that are longer, they can short the system board to ground.
6-8
Figure 6-6
4
Use a flat-blade screwdriver to gently unhook the latch that holds the white power-actuator shaft to the power switch, then disconnect the shaft from the power switch. After you disconnect the shaft, make sure you position it in the recess along the side of the display bracket.
Gently apply outward pressure at this corner of the slot
(
Replacing Assemblies
To remove the display
Unhooking the power switch shaft
5 Lift the display and remove it from the deck. 6 When reinserting the power shaft, push the shaft into the power switch
until the shaft seats.
6-9
Figure 6-7
Replacing Assemblies

To remove the system board

To remove the system board
1 Remove the floppy drive. 2 Using the T10 TORX driver, remove the six screws that hold the system
board to the deck.
3 Remove the two hex nuts and washers from the rear BNCs using the
9/16-inch socket driver.
The 54621D , 54622D, and 54624A have two rear BNCs, the 54621A and 54622A have one rear BNC.
4
Using the T8 TORX driver, remove two screws from module connector and remove the rear RFI gasket.
5 Remove the intensity knob and shaft by grasping the knob with one
hand and the grasping the shaft behind the front panel with the other hand and pulling the shaft straight out from the front panel.
6 Disconnect the three ribbon cables and the calibration cable.
Preparing to remove the system board
6-10
Figure 6-8
Replacing Assemblies
To remove the system board
7
Lift the back of the board to clear the main deck and then pull the board straight out.
Removing the system board
6-11
Figure 6-9
Replacing Assemblies

To remove the power supply

To remove the power supply
1 Remove the fan. 2 Disconnect the ground wire (green wire with the yellow stripe) from
the deck.
3 Disconnect the ribbon cable from the power supply board. 4 Remove the printer power connector by sliding the connector out of
the chassis.
Disconnecting ground wire and ribbon cable
6-12
Figure 6-10
5
Use a flat-blade screwdriver to gently unhook the latch that holds the white power-actuator shaft to the power switch, then disconnect the shaft from the power switch. After you disconnect the shaft, make sure you position it in the recess along the side of the display bracket.
Gently apply outward pressure at this corner of the slot
(
Replacing Assemblies
To remove the power supply
Unhooking the power switch shaft
6-13
Figure 6-11
Replacing Assemblies
To remove the power supply
6
Using the T20 TORX driver, remove the screw holding the power supply board to the deck.
Removing the power supply
7 Slide the power supply board toward the front panel about 1/2 inch. Slip
the power supply board keyhole slots off of the pins on the deck.
6-14
Figure 6-12
Replacing Assemblies

To remove the keyboard assembly

To remove the keyboard assembly
1 Remove the front panel. 2 Remove all of the knobs by pulling them straight out. 3 Using the T10 TORX driver, remove the four screws from the keyboard
shield.
4 Lift the ribbon cable end of the small keyboard and simultaneously slide
the small keyboard toward the main keyboard to remove.
Removing the keyboard assembly
CAUTION USE CORRECT SCREWS TO AVOID DAMAGING FRONT PANEL
LABEL!
When you re-install the keyboard, you must re-insert the screws you removed in this step. If you re-insert screws that are longer, they can damage the front panel label.
5 Lift the keyboard assembly straight out.
6-15
Replacing Assemblies

To remove the handle

To remove the handle
• Rotate the handle downward until it just passes the last detente position; this is about 1/2 inch before the handle touches the bottom of the oscilloscope.
• Pull the sides of the handle out of the cabinet.
6-16