AC-1
85.0812.00, 2003-02
Accessories Overview
This section provides descriptions, service, maintenance and adjustment detail on the
following accessories:
• Dual voltage intra-oral light source
•Single voltage intra-oral light source
• Cascade scaler
•Tooth dryer
• Self-contained water system
• Low voltage water heater/heated syringe system
• Curing light.
AC-2
85.0812.00, 2003-02
Accessories General Information
Dual/Single Voltage
Intra-Oral Light Source
Tooth Dryer
The A-dec Intra-oral light sources provide electrical power to illuminate handpiece light bulbs. The
dual voltage control has two potentiometers to allow operation of two different bulb requirements.
The single voltage light source has a single potentiometer to adjust output voltage. Both units have a
low and a bright output that must be adjusted when in bright mode. Refer to the following table
for specifications.
A-dec’s warm air tooth dryer provides warm air, for tooth preparation. It is sterilizable, has no
moving or electrical parts, and functions by routing 60 psi of air pressure through its vortex tube. The
warm air flows out of the tip at 125°F/ 51.7°C and 135°F/57°C while the cool air is exhausted. The
tooth dryer should be connected to a tooth dryer end cap or relay and a dedicated tooth dryer tubing
for optimum performance.
Identifying the Accessories
Intra-Oral Light Source Specifications
Single Dual
Output 2.9-4.25 VAC 3.0-5.6 VAC
at 0.8 amps at 1.4 amp
Input 24 VAC 24 VAC
Power 17 watts 17 watts
Consumption
AC-3
85.0812.00, 2003-02
Accessories General Information
Self-Contained
Water System
Low Voltage Water Heater/
Heated Syringe System
The self-contained water system provides a closed water supply system separate from the municipal
system. When supplied with 40 psi of air pressure, it provides treatment water to the control block
system and syringe. It also allows for water line asepsis and air purging of the control system.
The low voltage water heater/heated syringe tubing system provides instant heated water
(90°F/32.2°C) to the unit handpiece control and syringe.
Specifications
Low Voltage Water Heater 24 VAC
Syringe Tubing 6 VAC
AC-4
85.0812.00, 2003-02
Accessories Wire and Plumbing Diagram
Dual Voltage Intra-Oral Light Source
Cascade delivery systems: from post box
Radius delivery systems: from power supply
Terminal strip
Terminal strip wiring voltage
(after April 1998)
Wire Color Voltage
Black/White 0
Black/White 0
Red 6
Violet 17
Gray 24
Gray 24
Gray 24
Gray 24
To handpiece
holder valve
(holdback air)
NOTE: Confirm that the bulb requires no more
than 1.3 amps before connecting any
lighted handpiece or coupler to the
A-dec dual voltage intra-oral light
source. When additional lighted
handpieces are connected to the control,
an additional handpiece select switch
and shuttle valve (026.074.01) will be
installed for each additional handpiece.
After April 1998
Item # Part Number Description
1 90.0380.00 Intra-oral light
source, dual voltage
2 044.159.00 Air-electric switch
(replace as a
complete assembly)
3 75.0911.01 Switch diaphragm
4 75.0909.00 Intra-oral light
source switch (replace as a
complete assembly)
1
2
4
3
24VAC
INPUT PWR
P/N 90-0380-00
A-DEC
INTRA-ORAL
DUAL VOLTAGE CONTROL
LIGHT SOURCE
MAXMAX
B
A
WIRES BEFORE ADJUSTING
OUTPUT ADJUST
DISCONNECT DIMMER SWITCH
AB
DIMMER
SWITCH
POWER
0061724242424
VOLTS AC
A
W
D
2
D
1
AC-5
85.0812.00, 2003-02
Accessories Wire and Plumbing Diagram
Cascade delivery systems: to post box
Radius delivery system: to power supply
Ground
Terminal
strip
Terminal strip wiring voltage
(before May 1998)
Position Wire Color Voltage
1Black 0
2Brown 0
3Red 6
4Orange 24
5Yellow 24
6Green & Yellow Ground
7 Blue 24
8Violet 17
Before May 1998
NOTE: Confirm that the bulb requires no more than 1.3 amps
before connecting any lighted handpiece or coupler to
the A-dec dual voltage intra-oral light source. When
additional lighted handpieces are connected to the
control, an additional handpiece select switch and
shuttle valve (026.074.01) will be installed for each
additional handpiece.
To handpiece
holder valve
(holdback air)
Item # Part Number Description
1 90.0380.00 Intra-oral light source,
dual voltage
2 044.159.00 Air-electric switch
(replace as a complete
assembly)
3 75.0911.01 Switch diaphragm
4 75.0909.00 Intra-oral light source
switch (replace as a
complete assembly)
1
2
4
3
NOTE: Voltages should be adjusted while the foot control is
being stepped on. This ensures the DIOLS is in bright
mode. Do not measure voltage at the end of the
tubing. It is necessary to have a bulb installed and
illuminated for an accurate reading.
Dual Voltage Intra-Oral Light Source
A-DEC
INTRA-ORAL
DUAL VOLTAGE CONTROL
LIGHT SOURCE
INPUT PWR
P/N 90-0380-00
MAXMAX
B
24VAC
A
WIRES BEFORE ADJUSTING
OUTPUT ADJUST
DISCONNECT DIMMER SWITCH
AB
DIMMER
SWITCH
POWER
12345678
A
W
D
2
D
1
AC-6
85.0812.00, 2003-02
Accessories
Adjustments
Adjusting the Dual Voltage
Intra-Oral Light Source
(DIOLS)
*Voltage is measured at output terminal of IOLS with bulb lit. (Unit must be in
bright mode when adjusting the output voltage if the function is used.
Disconnect one of the bright/dim switch wires temporarily to enable the bright
mode. Reconnect the wire after any adjustments are made.)
DIMMER
SWITCH
AB
POWER
OUTPUT ADJUST
DISCONNECT DIMMER SWITCH
WIRES BEFORE ADJUSTING
A
B
A-DEC
INTRA-ORAL
LIGHT SOURCE
MAXMAX
INPUT PWR
24VAC
DUAL VOLTAGE CONTROL
P/N 90-0380-00
Terminal
(output A and B)
Bright/dim switch
Adjustment
potentiometer
(in) (cm) VDC +/- .02
48 122 3.51
54 137 3.54
60 152 3.56
66 168 3.59
72 183 3.62
78 198 3.65
84 213 3.67
90 229 3.69
96 244 3.71
102 259 3.74
108 274 3.76
114 290 3.79
120 305 3.82
126 320 3.85
132 335 3.87
138 351 3.90
144 366 3.93
150 381 3.96
156 396 3.99
Length and Voltage Table
Wire length in
A-dec tubing
A-dec/W&H, Bien Air,
or other bulbs rated at 3.5V
Voltage at terminal strip
*
Set to
DC volts
Digital voltmeter
Black
(common)
Red
(V Ω)
NOTE: Increase the terminal voltage by rotating the adjustment screw clockwise.
Decrease the terminal voltage by rotating the adjustment screw counterclockwise.
Use potentiometer A for “output A” and potentiometer B for “output B”
DIMMER
SWITCH
POWER
AB
OUTPUT
DISCONNECT
WIRES
AC-7
85.0812.00, 2003-02
Accessories
Adjustments
Adjusting the Single
Voltage Intra-Oral
Light Source
(SIOLS)
Task Description
1 Determine the handpiece wire length and the bulb type. (Wire length and bulb type should be
the same for each lighted handpiece position.)
2Find the corresponding (wire length/bulb type)
terminal voltage in the
“Length/Voltage Table”
on page AC-6.
3 Remove a lighted
handpiece from its holder.
4Move the wet/dry toggle
on the foot control to the
OFF position, away from
the blue dot. Step on the
foot control.
5Using an adjustment
screwdriver, adjust the
brightness potentiometer
until the voltmeter displays the voltage set from
the Length/Voltage Table
on page AC-6.
6 Replace the handpiece in
its holder. All lighted
handpieces have
been adjusted.
Black
Voltmeter (set to DC volts)
Red
To adjust:
Rotate adjustment screw clockwise to increase
output voltage, or counterclockwise to
decrease output voltage.
From 24VAC
source
From handpiece control block.
Holdback air for lighted handpieces.
From air coolant
signal
Bright/dim switch
Handpiece
select switch
NOTE: Voltages should be adjusted
while the foot control is being
stepped on. This ensures the
SIOLS is in bright mode. If
measuring voltage at the end
of the tubing, use A-dec/W&H
tools. It is necessary to have a
bulb installed and illuminated
for an accurate reading.
NOTE: For handpiece select switches, unlighted positions must be connected to pilot air.
Adjust the SIOLS by following these steps.
PN 34-0064-00
DANGER
2.9 - 4.25VDC .8AMP
POWER
24V
/DC
AC
INTRA ORAL
LIGHT SOURCE
ADJUST ONLY
WHEN BRIGHT
BRIGHT
AC-8
85.0812.00, 2003-02
Accessories Electric Handpiece Plumbing and Wire Diagram
52.0068.00
7' electric handpiece cordset
Electric handpiece controller
52.0027.00
Handpiece forward, neutral, reverse
switch (replace as a complete assembly)
75.0909.00
Handpiece select
(replace as a complete assembly)
90.0380.00
Intra-oral light
source, dual voltage
(replace as a
complete assembly)
Cascade delivery systems:
from post box
Radius delivery systems:
from power supply
Terminal and
manifold
Terminal strip wiring voltage
(after April 1998)
Wire color Voltage
Black/White 0
Black/White 0
Red 6
Violet 17
Gray 24
Gray 24
Gray 24
Gray 24
After April 1998
52.0020.00
Electric handpiece kit
(replace as a complete assembly)
Speed
control
T1
T2
A-DEC
INTRA-ORAL
AB
DIMMER
SWITCH
+
M
40
20
24V AC
24VAC
INPUT PWR
P/N 90-0380-00
DUAL VOLTAGE CONTROL
LIGHT SOURCE
MAXMAX
B
A
WIRES BEFORE ADJUSTING
OUTPUT ADJUST
DISCONNECT DIMMER SWITCH
POWER
2a1 3 4 5 6 78910
1234
0061724 24 24 24
VOLTS AC
W
A
AC-9
85.0812.00, 2003-02
Accessories Electric Handpiece Plumbing and Wire Diagram
Ground
Terminal strip wiring voltage
(before May 1998)
Position Wire color Voltage
1 Black 0
2 Brown 0
3 Red 6
4 Orange 24
5 Yellow 24
6 Green and Yellow Ground
7 Blue 24
8 Violet 17
52.0068.00
7' electric handpiece cordset
Electric handpiece controller
52.0027.00 Handpiece forward, neutral,
reverse switch. (replace as a complete assembly)
75.0909.00
Handpiece select
(replace as a complete assembly)
90.0380.00
Intra-oral light source,
dual voltage (replace as
a complete assembly)
Cascade delivery systems: from post box
Radius delivery systems: from power supply
Terminal and
manifold
Speed control
Before May 1998
T1
T2
A-DEC
INTRA-ORAL
AB
DIMMER
SWITCH
+
M
40
20
24V AC
24VAC
INPUT PWR
P/N 90-0380-00
DUAL VOLTAGE CONTROL
LIGHT SOURCE
MAXMAX
B
A
WIRES BEFORE ADJUSTING
OUTPUT ADJUST
DISCONNECT DIMMER SWITCH
POWER
2a1 3 4 5 6 78910
1234
12345678
W
A
AC-10
85.0812.00, 2003-02
Accessories
A-dec/W&H Handpiece Voltage Adjustments
Adjusting Handpieces
Measurement can be done at the handpiece or the Roto-Quick with the use of special tools. When
measuring at this point, the voltage should be 3.2 V. Tool #C709 is a RA-24 bulb with arms used to
attach the volt meter probes. Tool # C679 is a cutout sleeve for the Synea L handpieces. This can be
exchanged for the standard sleeve, connected to the Roto-Quick and allows access for the probes onto
the contacts.
*The air pressure adjustment screw is located under the sleeve opposite the bulb. It is factory set
at 2.2 bar (representing 32 psi). If air pressure needs to reach 45 psi, adjust the screw to 3.0 bar to
compensate for higher pressure.
NOTE: Voltages should be adjusted while the foot control is being stepped on. This ensures the SIOLS
is in bright mode. If measuring voltage at the end of the tubing, use A-dec/W&H tools. It is
necessary to have a bulb installed and illuminated for an accurate reading.
A-dec/W&H Handpiece Drive Air and Light Voltage Settings
Handpiece Model Voltage Setting (DC) Drive Air Pressure (psi) Factory Setting
at Bulb Pins
898 RM or 898 RM 3.2 32 N/A
898LE 3.2 45 3.0 bar
896 3.2 32* 2.2 bar*
All Synea models 3.2 45 3.0 bar
Low-speed motors 3.2 45 N/A
Electric motors 3.2 55-60 N/A
Tooth dryer N/A 60 N/A
AC-11
85.0812.00, 2003-02
Accessories
Handpiece Maintenance
Cleaning
Lubricating
Run
Step
Action
Follow these points to properly maintain handpieces.
With water switched off, run handpiece 20 to 30 seconds to blow all water out of spray tubes using
the foot control. If the spray tubes are not dry, they may become clogged with calcium deposits
during heat sterilization.
After removing the handpiece from the dental unit, remove the bur and thoroughly clean
external surfaces with a soft brush and alcohol or soap and water. Use of disinfectant may have a
harmful effect on the finish of the handpiece.
Install the proper spray cap onto the A-dec/W&H spray oil can. Shake the can before use. Spray for
approximately one second into the drive air port of the handpiece or the back end of the handpiece.
While spraying, visible debris may be expelled from the handpiece head. If this occurs, repeat the
spraying in one second intervals until no visible debris is expelled.
After lubrication, the handpiece should be attached to a handpiece tubing and run for 30
seconds to remove all excess oil. Excess oil will be discharged from the handpiece during this
running. Wipe excess oil off with a soft cloth.
CAUTION
Do not immerse handpieces under water or in any
cleaning solutions. Do not ultrasonically clean handpieces.
Maintaining
Handpieces
The information in the following charts assists in maintaining handpieces properly.