A-Dec 500 Service guide

A-dec 500® Service Guide
Including ICV® and Clinical Products
Copyright
Regulatory Information
©2012 A-dec Inc. All rights reserved.
No part of this document may be copied, reproduced, altered, or transmitted in any form or by any means, electronic or mechanical, including photocopying, recording, or by any information storage and retrieval system, without prior written permission from A-dec Inc.
Trademarks and Additional Intellectual Property Rights
A-dec, the A-dec logo, A-dec 500, A-dec 300, Cascade, Cascade Master Series, Century Plus, Continental, Decade, ICX, ICV, Performer, Preference, Preference Collection, Preference ICC, and Radius are trademarks of A-dec Inc. and are registered in the U.S. and other countries. A-dec 200, Preference Slimline, and reliablecreativesolutions are also trademarks of A-dec Inc. None of the trademarks or trade names in this document may be reproduced, copied, or manipulated in any manner without the express, written approval of the trademarkowner.
Regulatory information is provided with A-dec equipment as mandated by agency requirements. This information is delivered in the equipment’s Instructions for Use or the separate Regulatory Information and Specifications document. If you need this information, please go to the Document Library at
www.a-dec.com
.
Product Service
For service information, contact your local authorized A-dec dealer. To find your local dealer, go to www.a-dec.com
.
Service Reference
This document is a companion to the A-dec 500 Service Reference (p/n 86.0329.00). The Service Reference contains tubing, flow diagrams, and illustrated parts breakdown content. Circuit board component information is available in both this document and the Service Reference.
Certain touchpad symbols are proprietary to A-dec Inc. Any use of these symbols, in whole or in part, without the express written consent of A-dec Inc. is strictly prohibited.
A-DEC 500
S
ERVICE GUIDE
CONTENTS
1 INTRODUCTION........................................................1
Inside This Guide....................................................................1
Document Conventions.............................................................1
Get Support .....................................................................2
Customer Service....................................................................2
Other Sources of Information ...............................................2
Genuine A-dec Parts Catalog......................................................2
A-dec Dental Furniture Technical Packet........................................2
A-dec Illustrated Parts Breakdown ...............................................2
Electronic Documentation .........................................................2
Serial, Model Numbers, Color Identification .............................3
Serial and Model Numbers .........................................................3
Service Tools ....................................................................4
2 DENTAL CHAIR ........................................................7
Product Overview..............................................................8
A-dec 511 Chair......................................................................8
A-dec 511 Chair Specifications....................................................8
Load Capacity ........................................................................... 8
Power On/Off Button .................................................................. 8
Limp-Along Feature .................................................................... 8
Chair Power Supply .................................................................... 9
Service, Usage, and Adjustments ......................................... 10
Chair Covers ....................................................................... 10
Factory Default Routine..................... ....... ........ ....... ........ ....... 11
Chair Circuit Board Components................................................ 12
LED Identification .....................................................................14
The Hydraulic System ............................................................ 15
Hydraulic Fluid Reservoir ............................................................15
Hydraulic Cylinders ...................................................................16
Motor Driven Hydraulic Pump .......................................................16
Capacitor ..............................................................................17
Solenoids ...............................................................................18
Test Solenoids .........................................................................18
Magnetic pull ...........................................................................18
Coil resistance .........................................................................18
Potentiometer......................................................................19
Chair Stop Plate ...................................................................20
Chair Bump-Up Feature ..............................................................20
Swivel Brake Adjustment.........................................................21
Tension Adjustment ...............................................................21
Headrest Adjustment .............................................................22
3 PROGRAMMING ...................................................... 23
Status Icon .....................................................................24
Chair Positioning..............................................................24
Position Buttons....................................................................24
Program Chair Preset Positions ......................................................25
Customize the X-Ray/Rinse Button .................................................25
Cuspidor Functions...........................................................26
Cupfill ...............................................................................26
Customize Cupfill and Bowl Rinse...............................................26
Dental Light....................................................................27
Dental Light Auto Feature........................................................27
A and B Buttons....................................................................27
Electric Handpiece Settings (Deluxe Touchpad Only) .................28
Standard Mode Programming ....................................................28
Standard Mode Display Settings .................................................29
Forward/Reverse Button .............................................................29
Endodontics Mode .................................................................30
Endodontics Mode Settings.......................................................31
Technician Setup Options...................................................32
Using Touchpad Buttons for Navigation ........................................32
Holder Setup .......................................................................32
Technician Help Messages........................................................36
Touchpad Circuit Board Components.....................................37
Standard Touchpad................................................................37
86.0348.00 Rev A i
A-dec 500 and Clinical Products Service Guide Table of Contents ii
Deluxe Touchpad .................................................................. 38
4 DELIVERY SYSTEMS ................................................. 39
Delivery Systems Overview ...................................................... 39
Chair-Mounted Delivery Systems (532/533) ............................ 40
Product Overview (532/533) .............................................. 41
Service, Usage, and Adjustments (532/533) ........................... 42
Covers............................................................................... 42
Delivery System Cover ................................................................42
Delivery System Front Cover — Continental .......................................42
Whip Assembly..................................................................... 43
Continental Tray Holder (533 only)............................................. 43
Adjustments (532/533)........................................................... 44
Level Adjustment .....................................................................44
Tension Adjustment ...................................................................44
Front-to-Back Leveling ........................................................... 45
Side-to-Side Leveling ............................................................. 45
Side-Mounted Delivery Systems (542) ................................... 46
Product Overview (532/533/542 Common Feature)......................... 47
Service, Usage, and Adjustments (542) ................................. 48
Flexarm Counterbalance......................................................... 48
Remove covers..................................................................... 49
Rotational Tension Adjustment ................................................. 49
12 O’Clock Delivery Systems (541 and 545)............................ 50
Product Overview (541 and 545) ......................................... 51
A-dec 541 12 O’Clock Duo Delivery System ................................... 51
A-dec 541 Internal Components .....................................................51
Flex-Holder......................................................................... 52
Control Center Covers............................................................ 52
Doctor’s Holder.................................................................... 53
A-dec 500 Delivery Systems Common Features ........................57
Product Overview (All Delivery Systems)................................58
A-dec Tubing ...........................................................................58
Quick Connect Fittings ...............................................................58
Connect Tubing ........................................................................58
Disconnect Tubing .....................................................................58
Data Communication System ....................................................59
Circuit Board Components .......................................................60
Brake Handle.......................................................................61
Accessory Holder ..................................................................61
Master On/Off Toggle .............................................................62
Flush Toggle ........................................................................62
Traditional Tray Holder .............................................. ........ .....63
Foot Control........................................................................63
Self-Contained Water Bottle.....................................................64
Standard Syringe...................................................................65
Warm Water Syringe...............................................................65
Syringe Temperature Selection Input ...............................................65
Warm Water Syringe Specifications ................................................. 66
Tooth Dryer.........................................................................67
Service, Usage, and Adjustments (Common Features) ...............68
Control Block...................... ....... ....... ........ ....... ....... ........ .....68
Air Coolant .............................................................................68
Water Coolant ..........................................................................69
Drive Air ................................................................................69
Remove the Control Block........................................................70
532/533/542 Delivery Systems ......................................................70
541 Delivery System ...................................................................70
Tray Holder .........................................................................71
Adjust Tray Holder Tension ..........................................................71
Level Delivery System Tray ..........................................................71
Intraoral Light Source Voltage...................................................72
Intraoral Light Source Length and Voltage ........................................72
Service, Usage, and Adjustments (541 and 545) ...................... 54
Level the Arm Assembly and Worksurface..................................... 5 4
Level the Arm Assembly Front To Back ............................................54
Level the Arm Assembly Side To Side ..............................................54
Level Round Worksurface ............................................................55
Worksurface Height............................................................... 55
Instrumentation Arm Positioning................................................ 56
Position Holders .......................................................................56
5 ASSISTANT’S INSTRUMENTATION (551 AND 545) ............. 73
Product Overview (551).....................................................74
Holders..............................................................................75
Standard Holder .......................................................................75
Electric Holder .........................................................................76
Assistant’s Touchpad ..............................................................77
Solids Collector ......................................... ....... ....... ........ .....78
Tip-Up Feature .................................................................... 78
Adjustments (551) ........................................................... 79
Left or Right Conversion ......................................................... 79
Product Overview (545) .................................................... 80
Solids Collector.................................................................... 81
Remove Solids Collector and Replace Screen .....................................81
Leveling (545)....................................................... ........ ....... 82
Arm Assembly...................................................................... 82
Front-To-Back .......................................................................... 82
Side To Side ............................................................................82
Worksurface........................................................................ 83
Level Round Worksurface ............................................................83
Adjustments (545) ........................................................... 84
Worksurface Height............................................................... 84
Worksurface and Instrumentation Arm Height ....................................84
Instrumentation Arm Positioning................................................ 85
Assistant’s Arm ........................................................................85
Bump-Up Feature ......................................................................94
Adjustments/Maintenance ..................................................95
Leveling.......................................... ...................................95
Support Link and Lower Support Arm Leveling ...............................96
9 CLINICAL PRODUCTS ............................................... 97
EA-50LT and EA-51LT Electric Micromotor.....................................98
O-ring Replacement...............................................................98
Micromotor Tubing Terminal O-ring Replacement ................................98
Micromotor O-ring Replacement ....................................................98
Bushing Replacement .............................................................98
Micromotor Tubing Insert Bushing Replacement ..................................98
EA-50LT Bulb Replacement.......................................................99
Replace Bulb ...........................................................................99
EA-51LT Bulb Replacement ..... ....... ....... ........ ....... ....... ........ ... 100
Remove Bulb ......................................................................... 100
Replace Bulb ......................................................................... 100
6 CUSPIDOR ............................................................ 87
Product Overview............................................................ 88
Cuspidor Circuit Board Components............................................ 88
Service, Usage, and Adjustments ......................................... 89
Cupfill Functions .................................................................. 89
Bowl Rinse Functions ............................................................. 89
Cupfill and Bowl Rinse Timing................................................... 89
7 FLOOR BOX .......................................................... 91
Product Overview............................................................ 91
Air and Water Manual Shutoff Valves........................................... 91
Gauge and Pre-Regulator......................................................... 91
Service, Usage, and Adjustments ......................................... 92
Filter Element .................................................................... 92
8 SUPPORT SIDE FEATURES ......................................... 93
Product Overview............................................................ 94
Components........................................................................ 94
Limit Switch Operation........................................................... 94
10 ICV ................................................................. 101
Product Overview (ICV) ................................................... 102
Vacuum Requirements .......................................................... 102
ICV Display Pane................................................................. 103
Service, Usage, and Adjustments (ICV) ................................ 104
ICV Control Circuit Board....................................................... 104
ICV Display Circuit Board....................................................... 105
Adjustments...................................................................... 106
Cycle Time Test.................................................................. 106
Cycle Time Adjustment......................................................... 106
11 TROUBLESHOOTING............................................. 107
Dental Chair ................................................................. 108
Delivery Systems ........................................................... 114
Circuit Board Troubleshooting and LED Diagnostics ................. 122
Circuit Board Replacement..................................................... 122
Circuit Board Troubleshooting ..................................................... 122
Touchpad Diagnostics ...................................................... 123
500 Deluxe Touchpad Help Messages................................... 126
86.0348.00 Rev A Table of Contents iii
A-dec 500 and Clinical Products Service Guide Table of Contents iv
Cuspidor ..................................................................... 139
Floor Box .................................................................... 145
Clinical Products ........................................................... 146
Warm Water Syringes ...........................................................146
Tooth Dryer........................................................................148
A-dec Intraoral Camera .........................................................149
Sopro Intraoral Cameras (717 and 595).......................................150
SP Newtron........................................................................152
MiniLED ............................................................................154
EA-50/51LT Electric Micromotor ...............................................155
ICV ............................................................................ 156

INTRODUCTION

1

Inside This Guide

This guide is intended for newly trained and seasoned service technicians responsible for installing and maintaining A-dec products. The technician should understand dental equipment operation, how to use flow diagrams, and how to perform basic maintenance on dental or medical equipment.
Inside this guide you will find the tools, maintenance, adjustments, and troubleshooting information for A-dec 500 products.
Document Conventions
NOTE Notes indicate additional information, and when it is
important that instructions are followed.
CAUTION Caution indicates when failure to follow instructions could
result in damage to product or minor injury.
TIP Tips indicate tips or tricks to make installation, use, or
maintenance easier.
IMPORTANT Important indicates areas in which to refer to or use
specific instructions.
BIOHAZARD Biohazard indicates potential infection if instructions are
not properly followed.
DANGER Danger indicates warnings of dangerous voltage and of
certain electrical shock.
WARNING Warning indicates potential severe injury or death if
instructions are not followed properly.
86.0348.00 Rev A 1
A-dec 500 and Clinical Products Service Guide Introduction 2

Get Support

Customer Service

For questions not addressed in this document, contact A-dec Customer Service using contact information for your region.
U.S. and Canada
A-dec Inc. 2601 Crestview Drive Newberg, OR 97132 USA Tel: 1.800.547.1883 within US and Canada Tel: 1.503.538.7478 outside US and Canada www.a-dec.com www.a-dec.biz
United Kingdom
A-dec United Kingdom Austin House, 11 Liberty Way Nuneaton, Warwickshire CV11 6RZ England Tel: 0800 ADECUK (233285) within UK Tel: 44 24 7635 0901 outside UK www.a-dec.co.uk
Australia

Other Sources of Information

Genuine A-dec Parts Catalog

The Genuine A-dec Service Parts Catalog (p/n 85.5000.00), provides part number and ordering information for A-dec serviceable parts. This catalog details service parts for current products and products which are no longer manufactured, but still supported. Refer to this catalog for additional details on parts found in the service guide addendum.

A-dec Dental Furniture Technical Packet

The A-dec Dental Furniture Technical Packet, P/N 86.0142.00, contains information specifically related to dental furniture. The content is intended to assist you in specifying plumbing, utilities, framing and construction requirements for installation of dental furniture.

A-dec Illustrated Parts Breakdown

The A-dec Illustrated Parts Breakdown (IPB), P/N 85.0851.00, contains illustrated, exploded views of assemblies with part numbers and descriptions for associated parts for products produced before A-dec 500.

Electronic Documentation

Electronic versions (PDF files) of our documentation (installation instructions, service guides, technical information) can be viewed or downloaded from the A-dec website (www.a-dec.com). Check this location for current detail on products and technical information.
A-dec Australia Unit 8, 5-9 Ricketty St. Mascot, NSW 2020 Australia Tel: (02) 8332 4000
1.800.225.010 www.a-dec.com.au
Web Contact
Partner Resources websites: www.a-dec.biz

Serial, Model Numbers, Color Identification

Light Location
Chair (Under Toeboard)
Cuspidor (Inside Support Center)
Power supply (inside motor pump area)
Assistant’s Instrumentation Location
Delivery System Location
Monitor Mount Location
Products Manufactured
before June 2011
Serial
Number
Model
Number
Products Manufactured
June 2011 and after
511
11F28293
DENTAL CHAIR

Serial and Model Numbers

Product serial and model number information can be found on the serial/ model number labels. When contacting customer service, the serial number helps identify the product and when it was manufactured.
The first letter of the serial number indicates the month the product was manufactured. For products manufactured before June 2011, the first digit of the serial number indicates the year of manufacture (for example, L0 = December 2010). For products manufactured June 2011 and after, the first two digits indicate the year of manufacture (for example, 11F = June 2011).
Use Table 1 and Figure 2 to reference how to identify serial/model number information.
Figure 1. Serial/Model Number Locations on A-dec 500
Tab l e 1 . M ont h Ide n tif i c at i o n Tab l e
Letter Month Letter Month
A January G July B February H August C March I September DApril JOctober EMay KNovember F June L December
Figure 2. Serial Number Label Examples
86.0348.00 Rev A Introduction 3
A-dec 500 and Clinical Products Service Guide Introduction 4

Service Tools

Table 3 lists the types of tools available from A-dec for servicing A-dec equipment and their recommended use:
Figure 3. Recommended Tools
Tool Tas k Part Illustration Part Number
Drive air pressure gauge Adjusting handpiece drive air pressure, 0-60 psi (4 .13 b ar). This
gauge does not fit the Borden 3-hole coupler
Hemostat Troubleshooting or repairing a unit to stop air or water flow
through tubing
Hex wrench or key set Servicing or installing A-dec equipment (plastic case included) 009.018.00
®
Loctite
O-ring tools Replacing O-rings during quick field repairs (fits the four
Installing threaded fasteners to prevent loosening 060.001.00 (Red 271)
smallest O-ring sizes)
50.0271.00
009.008.00
060.002.00 (Blue 242)
009.013.00
Panel mount gauge Checking air/water pressure
Can also be used as an inline pressure gauge for testing purposes
026.118.00
Tool Tas k Part Illustration Part Number
Silicone lubricant Lubrication of internal moving parts such as O-rings, oral
evacuator valves, and bushings
Sleeve tool Aid in securing 1/4" tubing sleeves and 1/8" uni-clamps 98.0072.00
98.0090.01
Snap ring tool Installation and removal of internal and external snap rings (fits
all snap rings used in A-dec equipment)
Tubing stripper Separation of the extruded air and water lines in vinyl tubing 009.035.00
Umbilical stringer Route additional tubing or wiring through existing umbilical
assemblies (12’ [3.66 mm] stringer with threading holes on both ends)
Valve test syringe Quick tests of pilot operated valves; used to apply a static
pressure of 5-75 psi (.34-5.17 bar, package of 4)
009.007.00
009.015.00
98.0050.01
86.0348.00 Rev A Introduction 5
A-dec 500 and Clinical Products Service Guide Introduction 6

DENTAL CHAIR

2
This section provides information related to servicing, maintenance, and adjustments of the A-dec 511 chair. For information on service parts, see the Genuine A-dec Service Parts Catalog (p/n 85.5000.00) or contact A-dec customer service.
Contents
• Product Overview, page 8
• Service, Usage, and Adjustments, page 10
Figure 4. A-dec 511 Dental Chair
86.0348.00 Rev A 7
A-dec 500 and Clinical Products Service Guide Dental Chair 8
Hydraulic Motor Pump Assembly
Baseplate
Armrest
Seat
Headrest
Foot Switch
Back
On/Off Button

Product Overview

A-dec 511 Chair

The A-dec 511 chair provides a range of movements to position the patient for dental treatment. The chair offers seat and back support, adjustable head support, and movable arm support, and a footswitch and/or touchpad(s) control the chair movement.
Figure 5. A-dec 511 Chair Features

A-dec 511 Chair Specifications

Load Capacity
Patient Load: 400 lb. (181 kg) maximum
Accessory Load: 250 lb. (113 kg) maximum
Specifications are subject to change without notice.
NOTE Ensure the chair is bolted to the floor after installation.
Power On/Off Button
The power on/off button is located on the base of the chair, and is the main disconnect that completely shuts down the electrical systems. When the button is pressed in, power is on. When the button is out, power is off.
Figure 6. On/Off Button
Limp-Along Feature
If there is a problem or malfunction, the limp-along feature allows the operator to move the chair in the up direction for one second intervals by pushing the manual control buttons on the touchpad or footswitch.
Chair Power Supply
CB 4 ­Assistant’s Arm, Control Head, Chair Circuit Board.
CB 3 - Support side arm
CB 1 and CB2 - Mains
CB 5 - Dental Light
The 300-watt power supply comes standard with the A-dec 511 chair. It is located in the motor pump area of the chair. The total available auxiliary load is a maximum of 4 Amps.
NOTE The electric switch connects the power supply to pilot air.
Figure 7. Chair Power Supply Circuit Breaker Identification
86.0348.00 Rev A Dental Chair 9
A-dec 500 and Clinical Products Service Guide Dental Chair 10
Stop Plate Cover
Lift Arm Cover
Motor Pump Cover
Motor Pump Cover Plug

Service, Usage, and Adjustments

Chair Covers

The A-dec 511 chair motor pump, lift arm and stop plate covers are removed in the following order:
1. Motor Pump Cover
To remove: Remove screw from each side and lift up.
To replace: Replace cover, and attach with two screws.
2. Lift Arm Cover
To remove: Pull one side of the cover until it releases from the lift arm.
To replace: Align one side of the cover with the lift arm and snap into
place. Ensure both sides are firmly attached.
3. Stop Plate
To remove: Pull one side of the cover until it releases from the lift arm.
To replace: Slide one side of the cover over the post on the lift arm and
attach.
Figure 8. A-dec 511 Chair Covers

Factory Default Routine

When a new circuit board is installed in the chair, the circuit board needs to run the factory default routine to learn the range of motion of the chair. The routine:
• sets the base and back upper limits
• calculates new presets based on actual range of motion of the chair
• verifies that the potentiometers work
To start the factory default routine, place the “spare” jumper in the factory default position on the P3 test points of the chair circuit board.
When running the factory default routine the chair:
1. Moves base down
2. Moves base up
3. Moves back down
4. Moves back up
5. Moves base and back to Position 0
6. Beeps three times
NOTE The jumper must remain in the factory default position to
complete the factory default routine. The status LEDs on the standard and deluxe touchpads and the chair circuit board double blink while the factory default routine is running and after the routine is complete.
86.0348.00 Rev A Dental Chair 11
A-dec 500 and Clinical Products Service Guide Dental Chair 12

Chair Circuit Board Components

Part No: 90.1072.00
Item Description
1 P7, P8, P9 - Data line ports 2 DS4 - Stop switch LED (limit switch) and P10 connector 3DS5 - Back potentiometer LED and P1 connector 4 DS6 - Base potentiometer LED and P2 connector 5 P5 - Footswitch connector 6P3 - Test points 7DS12 - Base down LED and relay K5 8 DS11 - Base up LED and relay K4
9 DS10 - Back down LED and relay K3 10 DS9 - Back up LED and relay K2 11 DS1 - AC power LED 12 DS2 - Stat us LED 13 DS3 - Data LED 14 DS13 - Chair lockout LED and terminal strip J1 15 J2 - Ø VAC terminal strip (output) 16 J2 - 24VAC terminal strip (outp ut) 17 P4 - Input power and dental light connector 18 J3 - Vacuum relay K7 and output terminal strip 19 P11 - Pump motor and solenoid connector 20 DS8 - Dental light LED and relay K1 21 DS7 - Dental light LED and relay K6
Figure 9. A-dec 511 Chair Circuit Board Components
1
11
12
13
14
4
3
5
67
8
9
10
21
18
16
15
17
20
2
19
86.0348.00 Rev A Dental Chair 13
A-dec 500 and Clinical Products Service Guide Dental Chair 14
LED Identification
This table describes the LEDs on the chair circuit board.
Figure 10. LED Identification Table
LED Status Description
DS1 - AC power LED Off No 24 VAC power, tripped circuit breaker, power supply turned off, no line voltage
Green, steady 24VAC at the terminal strip
DS2 - Status LED Off System is not functioning, no power or circuit board has failed
Green, steady Normal operation
DS3 - Data LED Off No DCS communication, not connected to the DCS, or DCS has failed
Green, steady Detects active DCS Green, blinking Valid DCS Message
DS4 - Chair limit switch Off Closed, (normal)
Red Open, (activated)
DS13 - Chair lockout Off Open, (normal)
Red Closed, (activated)
DS5 + DS6 - Chair potentiometers Off Potentiometer:
• Not connected or bad connection
• Moving in wrong direction
• Limited range of mot io n, or
•Cable is not on wheel
Yellow, steady Normal operation Yellow, fast blink Upper end of travel
DS9, DS10, DS11, DS12 - Chair relay LEDs Off Relay is off
On Relay is on
DS7, DS8 - Dental light relay LEDs Off Relay is off
On Relay is on
DS14 - Vacuum relay LED Off Relay is off
On Relay is on

The Hydraulic System

Max Level Min Level
The hydraulic system deactivates automatically at the upper and lower extremes of travel. The system is leak-free during transportation, storage, and operation. The hydraulic system consists of:
• Hydraulic fluid reservoir
•Hydraulic cylinders
• Motor-driven hydraulic pump with solenoids
Hydraulic Fluid Reservoir
The hydraulic fluid reservoir is located in the lift arm of the chair under the stop plate cover. You can see the fluid level in the reservoir through the sides of the reservoir. A top fill cap allows you to add fluid. The hydraulic system holds 40 ounces (2.5 pints [1.18 l]) of hydraulic fluid. To fill the reservoir:
1. Place the chair in the full base and back up position.
2. Fill to the green Max line. See Figure 11.
CAUTION Do not overfill.
3. Cycle the chair after the reservoir is filled.
Figure 11. Hydraulic Fluid Reservoir
86.0348.00 Rev A Dental Chair 15
A-dec 500 and Clinical Products Service Guide Dental Chair 16
Hydraulic Cylinders
The hydraulic cylinders operate during the Base Up and Back Up functions. Springs and gravity retract the piston during Base Down and Back Down functions.
The chair seat travels vertically from a low point of 13.5" (343 mm) to a high point of 31.5" (800 mm) above the floor. See Figure 12.
Motor Driven Hydraulic Pump
The hydraulic pump takes hydraulic fluid from the reservoir and pressurizes it to extend the chair lift and tilt hydraulic cylinders for back and base up functions. The bi-directional pump rotates one direction for Base Up and the opposite direction for Back Up.
The solenoids mounted to the pump assembly gate hydraulic fluid from the two cylinders. Depending on the chair Down function, the controller selects which solenoid-actuated manifold valves are open or closed. The 100-120 VAC pump and 220-240 VAC pump are equipped with an automatic reset 110°C (230°F) thermal limiter. There are no serviceable parts on the hydraulic pump other than the solenoids.
NOTE You cannot adjust the speed of the chair.
Figure 12. Hydraulic Cylinder Operation
Capacitor
Capacitor
Part Number: 041.642.00, 100 VAC, 041.643.00, 110 - 120 VAC, 041.644.00, 220 - 240 VAC
The capacitor is energized during chair Base Up or Back Up functions.
Figure 13. A-dec 511 Chair Capacitor
86.0348.00 Rev A Dental Chair 17
A-dec 500 and Clinical Products Service Guide Dental Chair 18
Solenoids
Part No: 90.1070.00, 110 - 120 VAC, 90.1071.00, 220 - 240 VAC
Figure 14. A-dec 511 Chair Solenoids
Test Solenoids
A solenoid is entergized during Base Down and Back Down functions. To check for a failed solenoid, test the solenoids using a volt/ohm meter or
Back Down Solenoid
magnetic pull test:
Base Down Solenoid
Magnetic Pull
1. Hold a paper clip loosely in your hand.
2. Activate the solenoid by pressing Base Down or Back Down on the
footswitch or touchpad.
3. If there is a pull on the paper clip, the solenoid is being entergized.
Coil Resistance
Disconnect the solenoid power at the 2-position connector. Place on Ohm meter probe on each solenoid connector terminals.
• 100 - 120 VAC = 177 Ohms ± 18 Ohms
• 220 - 240 VAC = 845 Ohms ± 85 Ohms
NOTE If the solenoid is hot, then the resistance reads higher.
NOTE When replacing a solenoid, wipe up any oil, and replace
existing O-rings on the solenoid base.
WARNING The solenoid coils are powered by line voltage (100, 120, or
240 VAC). Failure to unplug the chair may result in serious injury from electrical shock.
WARNING You must depressurize the base or back system prior to
removing the solenoid.
Motor Pump Assembly
1. Depressurize base or back system.
2. Remove the failed solenoid coil.
3. Replace with the operating solenoid coil.
4. Lower the chair base and back.

Potentiometer

Base Up Potentiomete
Back Up Potentiometer
The potentiometer and cable assembly is a simple, accurate unit, which eliminates position float. “Float” is a slight change or variation in the pre-programmed positions. The chair uses the same potentiometer assembly for both lift and tilt requirements. If a potentiometer should fail, the limp-along feature allows the operator to position the chair for one second intervals by pushing the manual control buttons on the touchpad or footswitch.
Figure 15. Location of Chair Base Up and Back Up Potentiometers
86.0348.00 Rev A Dental Chair 19
A-dec 500 and Clinical Products Service Guide Dental Chair 20
Stop Switch Assembly

Chair Stop Plate

The chair stop switch stops chair movement when you press the stop plate. Should anything inadvertently become lodged under the chair, press Base Up on the touchpad or footswitch to raise the chair so you can remove the object. As long as you apply pressure to the stop plate, the chair does not move down.
The stop plate has only one switch. The switch and all other parts snap into place for easy removal or replacement. No tools are required.
WARNING Be sure to power off the chair and disconnect it from its
power source before replacing the stop switch.
CAUTION Cable tie the wires to the lift arm to prevent kinking and
pinching.
Chair Bump-Up Feature
The chair stop plate and the assistant’s arm trigger the chair to move upwards if it was moving down when the stop plate switch was activated.
Figure 16. A-dec 511 Chair Stop Plate

Swivel Brake Adjustment

The chair can rotate to any position within 30° either side of center. The chair swivel brake keeps the chair from moving. To engage the brake, push the brake lever firmly to the left. To release the swivel brake, push the brake lever to the right.
Figure 17. Chair Swivel
Figure 18. Swivel Brake Tension Adjustment

Tension Adjustment

If the chair swivels left or right with the brake engaged or if it is difficult to move with the brake disengaged, adjust the swivel brake tension. To adjust the tension, use a 5/32 hex key and turn the tension adjustment screw;
• Clockwise to increase brake friction.
• Counterclockwise to decrease brake friction.
If you cannot obtain proper adjustment through rotation of the hex key, replace the brass brake pad by removing the brake handle and using a hex key to disengage the pad. Remove the old pad and replace with new one. Replace the brake handle and handle retainer.
NOTE To disable the swivel feature, reinstall the shipping pin.
Tension Adjustment Screw
Engaged
86.0348.00 Rev A Dental Chair 21
Released
A-dec 500 and Clinical Products Service Guide Dental Chair 22
Glide Bar
Headrest Adjustment Lever
Glide Bar Tension Adjustment

Headrest Adjustment

The headrest adjustment lever allows you to use one hand to adjust the headrest. When the lever is released, the headrest holds its position.
If the headrest drifts downward, or if it is difficult to move up or down, adjust the glide bar tension. To adjust the tension, use a 1/8 hex key and turn the tension adjustment screw clockwise to increase friction or counterclockwise to decrease friction.
Figure 19. Headrest Adjustments

PROGRAMMING

Standard Surf
Standard White
Deluxe Surf
Deluxe White
Note: Touchpad symbols are proprietary to A-dec Inc.
3
The A-dec 500 touchpad centralizes treatment room controls into one touch surface. The standard touchpad controls the chair, cuspidor, and dental light functions. The deluxe touchpad adds controls for the handpieces, electric motors, and other options.
For information on service parts, see the Genuine A-dec Service Parts Catalog (p/n 85.5000.00) or contact A-dec customer service.
Programming Contents
• Status Icon, page 24
• Chair Positioning, page 24
•Cuspidor Functions, page 26
• Dental Light, page 27
• Electric Handpiece Settings (Deluxe Touchpad Only), page 28
• Technician Setup Options, page 32
• Touchpad Circuit Board Components, page 37
Figure 20. Standard and Deluxe Touchpads
86.0348.00 Rev A 23
A-dec 500 and Clinical Products Service Guide Programming 24

Status Icon

Or
Or
Or
Or
Or
Or
Or
Or
Status Icon
The A-dec logo indicates the system status.
• Solid blue—normal operation and power is on.
• Blinking— chair stop plate or cuspidor limit switch is active. The icon returns to solid blue once you remove any obstructions.
• Double blink —jumper is on the factory default position on the chair circuit board. The icon returns to solid blue once the jumper is removed.
Figure 21. Status Icon

Chair Positioning

Your touchpad provides manual and programmed controls for A-dec chair positioning. The direction arrows on the touchpad allow you to manually move the chair base and back up or down.
Table 2. Manual Chair Buttons

Position Buttons

Chair position buttons are factory preset to automatically move the chair. Four programmable chair buttons, which can be easily customized, automatically move the chair into a factory preset position.
Table 3. Programmable Chair Buttons /Factory Presets
Button Position Description
Entry/Exit: Automatically positions chair for entry/exit and
turns off dental light.
Treat m e n t 1 : Automatically positions the chair base and back down and turns on the dental light.
Treat m e n t 2 : (Standard touchpad only): Automatically positions the chair base and back and turns on the dental light.
X-ray/Rinse: Automatically positions the chair for either x-ray or rinse. Toggles between the x-ray/rinse and the last manual position, and turns off the dental light.
Icon Action
Back down
Base down
Back up
Base up
Program Chair Preset Positions
A
B
C
Standard Surf
Standard White
To program Entry/Exit, Treatment 1, and Treatment 2 buttons:
1. Use the manual buttons to move the chair into the desired position.
Customize the X-Ray/Rinse Button
The X-ray/Rinse button functions either as x-ray/rinse or as another preset position (Treatment 3). To change the function:
2. Press and release the program button ( or ). One beep indicates
programming mode.
3. Press the chair position you want to program (for example or for
entry/exit). Three beeps indicate the button has been set.
Figure 22. Touchpad Chair Position Buttons
(A) Manual Chair Buttons; (B) Program Buttons; (C) Programmable Chair Buttons
1. Press and hold the program button ( or ) and or at the
same time for three seconds.
One beep indicates the button has been configured as Treatment 3.
Three beeps indicate that the X-ray/Rinse button has been configured as
the x-ray/rinse function (toggles between the x-ray/rinse and the previous position).
2. Program the preset position as instructed under "Program Chair Preset
Positions."
NOTE If the X-Ray/Rinse button is changed to a preset position, it
operates the same as Treatment buttons 1 and 2.
86.0348.00 Rev A Programming 25
A-dec 500 and Clinical Products Service Guide Programming 26
or
or

Cuspidor Functions

Cupfill

The cupfill function controls water flow from the cuspidor into a cup.
Figure 23. Cupfill and Bowl Rinse Buttons
Button Description
Cupfill Button: Controls water flow from the cuspidor into a cup.
• Press the cupfill button for a timed operation. The factory preset is a 2.5 second fill.
• Press and hold the cupfill butt on for manual operation.
Bowl Rinse Button: Provides rinse water for the cuspidor bowl.
• Press the bowl rinse button for a timed operation. The factory preset is a 30 second rinse.
• Press and hold the bowl rinse button for manual operation.
TIP Press the bowl rinse button twice in less than two seconds to
activate the continuous operation mode. Press the button once to end the continuous bowl rinse mode.

Customize Cupfill and Bowl Rinse

To program the cupfill and bowl rinse timing:
1. Press and release the Program button ( or ), or at the cuspidor
press and hold both the cupfill and bow rinse buttons located near the spout. One beep indicates programming mode.
2. Press and hold the cupfill or bowl rinse button for the desired time.
3. Release the button. Three beeps confirm the setting.
NOTE Cupfill and bowl rinse functions are not available on standard
touchpads that use auxiliary buttons ( or ) along with
an A-dec relay module. In this case, use the cuspidor-mounted buttons or the deluxe touchpad to control automated cuspidor functions.

Dental Light

The dental light button on the touchpad allows you to turn the dental light on or off, and change intensity. You can turn the dental light on or off from either the touchpad or the dental light.
Figure 24. Dental Light Button
Button Description
Dental Light Button: Functions as a three-way on/off switch and
toggles between intensity settings.
Press to toggle between high and composite or medium and composite.
When the dental light is in composite mode, the indicator light next to the
button blinks. See Figure 25. Press and hold for one second to turn the
dental light off.
Figure 25. Dental Light Composite Mode
Indicator Light

Dental Light Auto Feature

The dental light has an auto on/off feature. When you use a programmed chair position, the dental light turns on when the chair reaches operating position.
Press Entry/Exit ( or ) or X-ray/rinse ( or ) and the dental light
automatically turns off.
NOTE If the X-ray/Rinse button is changed to a preset position, the
dental light auto feature operates the same as Treatment buttons 1 and 2.
To activate/deactivate:
Press and hold the light ( ) and the program ( or ) buttons at the
same time for three seconds. One beep confirms the dental light auto feature is off.

A and B Buttons

Deluxe touchpads include an A/B button ( ) that offers two separate sets of programmable settings. The currently active set is indicated by the A or B in the
touchpad display. See Figure 26.
Figure 26. A/B Selection
86.0348.00 Rev A Programming 27
A-dec 500 and Clinical Products Service Guide Programming 28

Electric Handpiece Settings (Deluxe Touchpad Only)

Standard Mode Programming

Activate the electric motor by lifting the handpiece from the holder. The settings that appear are the ones last used for that handpiece position. This table lists the factory presets for electric handpieces:
Table 4. Electric Motor Presets (Standard Mode)
Memory Button Preset Speed Air Coolant Water Coolant
M1 2,000 RPM on on M2 10,000 RPM on on M3 20,000 RPM on on M4 36,000 RPM on on
The A-dec Deluxe touchpad allows you to program four memory buttons with your specific RPM setting. The total range is 300 - 36,000 RPM. With the A/B button, you can program two settings for each memory button, for a total of 8 customized settings per handpiece (the endodontic mode offers an additional 8
memory settings).
NOTE If two operators use the delivery system, be sure to select the
correct operator set (A or B) before changing any memory settings. The currently active set is indicated by the A or B in the touchpad display. See Figure 27 on page 29.
To program the handpiece setting:
1. Lift the handpiece from the holder.
2. If the touchpad screen does not display standard mode, press the
Endodontics Mode Toggle button.
3. Press the minus (-) and plus (+) buttons to adjust the RPM. The RPM values
are displayed in the display screen. See Figure 27.
4. If desired, use the toggle buttons on the touchpad to change air and water
settings.
5. To place the setting into memory, press the program button ( or ).
One beep sounds.
6. Select the desired memory setting (m1, m2, m3, or m4). Three beeps confirm
the setting.

Standard Mode Display Settings

A
H
J
K
L
M
B
C
E
G
D
F
Figure 27. Program Handpiece Standard Mode Settings
Item Description Item Description
A Memory Setting Indicator G Water Coolant Button B Water Coolant Indicator H Air Coolant Button CAir Coolant Indicator J Speed Limit Setting D A/B Operator Indicator K Forward/Reverse Indicator E Memory Buttons (m1 - m4) L Forward/Reverse Toggle Button F Program Button M Adjustment Buttons
Forward/Reverse Button
The Forward/Reverse toggle button changes the electric motor’s direction. The system defaults to the forward position when you return the handpiece to the holder or turn off the system. In reverse mode, the screen indicator flashes
continuously.
NOTE You can also use the foot control as a forward/reverse toggle.
When the motor has stopped, tap the accessory (chip/air) button to change the direction.
86.0348.00 Rev A Programming 29
A-dec 500 and Clinical Products Service Guide Programming 30

Endodontics Mode

Endodontics mode allows you to change a number of settings based on the specific file and desired handpiece behavior.
NOTE For more information regarding speed limit and torque limit
for a specific file, consult the file manufacturer.
Figure 28. Endodontics Mode Touchpad Screen (Deluxe Surf Shown)
A B C L
D
E
G
H J
K M
N
To program the endodontics memory settings:
1. Lift the handpiece from the holder.
2. If the touchpad window is not in endodontics mode, press the Endodontics
Mode Toggle button. The endodontics screen is displayed.
3. Press the plus (+) or minus (-) button to activate the endodontics change
mode. A white reverse video box appears.
4. Use the chair positioning buttons to move from setting to setting in the
touchpad window.
5. Use the + and - buttons to change the setting as desired.
6. To place the speed limit, torque limit and ratio into memory (optional),
press the program button ( or ), then the memory button you want to set. Three beeps confirm the setting.
F
Item Description Item Description
A Memory Setting Indicator J Handpiece Light Indicator B Water Coolant Indicator K File Torque Unit Indicator C Air Coolant Indicator L File Torque Limit D Operator Status Indicator M Forward/Reverse Indicator E Memory Button N Torque Mode Indicator F Endodontics Mode Toggle Button O Handpiece Ratio Setting G File Speed Setting P Forward/Reverse Toggle Button H Warning Beep Indicator
O P

Endodontics Mode Settings

This table lists and defines the touchpad window icons for endodontics mode.
Figure 29. Endodontics Mode Settings
Icon Setting Description
Speed Setpoint for file speed limit. For more information, consult your
file manufacturer.
Torque Setpoint for file torque limit. For more information, consult your
file manufacturer.
Torque Units
Ratio Sets the handpiece ratio. For more information, consult your
Air Coolant
Toggles between Ncm (Newton centimeters) and gcm (Gram centimeters). Adjusting this setting for one handpiece changes it for all handpiece settings. 1 Ncm=102 gcm
handpiece manufacturer. On/Off—when active, supplies air coolant to the handpiece.
NOTE The A-dec/W&H WD-79M endodontics attachment has a
special feature due to its ball-bearing design. Its life-long efficiency factor is stable and known, therefore the A-dec endodontics system is able to control and display file torque very accurately. All other handpieces have unknown life-long efficiency factors and therefore stated torque values are approximate.
Water Coolant
Light Source
Torque Mode
Auto off
Auto reverse
Auto forward
Beeper On/Off—when active, warning beep sounds when you approach
86.0348.00 Rev A Programming 31
On/Off—when active, supplies water coolant to the handpiece.
Enables/disables endodontics hand piece light source.
Adjusting this setting for one handpiece changes it for all handpiece settings. This icon appears with the forward/reverse indicator. Auto-off—the motor stops when t he file speed reac hes the torque limit. Auto-reverse—the motor stops and reverses direction when the file reaches the torque limit. Auto-forward—when the file reaches the torque limit, the motor stops, reverses 3 turns, then changes back to forward again. Note: If the file is stuck, this cycle repeats three times before the motor stops.
torque limit and double beeps when the file auto-reverses. Adjusting this setting for one handpiece changes it for all handpiece positions.
A-dec 500 and Clinical Products Service Guide Programming 32

Technician Setup Options

The deluxe touchpad allows service technician access to adjust handpiece and touchpad settings for user preferences.

Using Touchpad Buttons for Navigation

Chair button functions become navigation buttons while you are in setup mode. You will use the back up ( down (
) buttons to navigate the setup screens.
) back down () base up () and base

Holder Setup

You can set up how handpieces are configured for each handpiece holder.
To set up handpiece holders:
1. From the deluxe touchpad main screen, press and hold the A/B ( ) and
Program ( or ) buttons at the same time for three seconds; then
to begin. You see the System Setup screen.
press
2. Press the
SYSTEM SETUP
Handpieces
Touchpad
= BACK = SELECT = MOVE
or to select Handpieces, and press .
5. Press
. Three beeps confirm the setup for the handpiece is complete.
6. Return the handpiece to the holder.
7. Repeat steps 4 through 7 to set up each handpiece.
8. When you are finished setting up handpieces, press
A-dec logo.
until you see the
3. From the Handpiece Setup screen, press
Setup, and press
HANDPIECE SETUP
Holder Setup Electric Setup Ultrasonic Setup Intraoral Lt Source
= BACK = SELECT = MOVE
.
4. Lift the desired handpiece from the holder.
or to select to select Holder
Electric Options Setup
HANDPIECE SETUP
= BACK = SELECT = MOVE
Holder Setup Electric Setup
Intraoral Lt Source
Ultrasonic Setup
You can change display information and how the functions operate with the electric motor. The following options are available:
Torque Units: Select how to display the units, either Ncm (Newtons per centimeters) or gcm (gram per centimeters).
Endo Handpiece Auto Mode: Configure how the electric motor reacts when the torque limit is reached.
Auto Forward: Motor will stop, reverse three turns, and return forward
again.
Auto Reverse: Motor will reverse.
Auto Stop: Motor will stop.
3. From the Handpiece Setup screen, press
Setup, and press
4. From the Electric Setup screen, press minus (-) or plus (+) to highlight an
option, such as Torque Units, and press
.
or to select to select Electric
.
Auto Reverse Beep: Enable to hear two beeps when the motor exceeds its torque threshold and automatically reverses direction.
Torque Warning: Enable to hear a continuous beep when the motor exceeds 75% of the torque threshold.
To set up electric options:
1. From the deluxe touchpad main screen, press and hold the A/B ( ) and
Program ( or ) buttons at the same time for three seconds; then press
to begin. You see the System Setup screen.
2. Press the
or to select Handpieces, and press .
SYSTEM SETUP
Handpieces
Touchpad
= BACK = SELECT
= MOVE
ELECTRIC SETUP
Torque Units Torque Warn Beep Auto Rev Beep Auto Mode
= BACK -/+ = CHANGE
= MOVE
5. Specify other electric handpiece options by pressing minus (-), plus (+), and
to move through the screens. Once the setup is complete, three beeps
confirm the setting for all handpieces.
6. When all settings are complete, press
= SELECT
gcm
ON ON
Stop
until you see the A-dec logo.
86.0348.00 Rev A Programming 33
A-dec 500 and Clinical Products Service Guide Programming 34
ULTRASONIC SETUP
= BACK +/- = CHANGE = MOVE
Acteon Colors OFF
= SAVE
Ultrasonic Setup
You can select whether to turn on or off the ultrasonic colors.
To set up ultrasonic colors:
1. From the deluxe touchpad main screen, press and hold the A/B ( ) and
Program ( or ) buttons at the same time for three seconds; then press
to begin. You see the System Setup screen.
2. Press the
or to select Handpieces, and press .
SYSTEM SETUP
Handpieces
Touchpad
= BACK = SELECT
= MOVE
4. Press minus (-) or plus (+) to display On or Off, and press
confirm the setting has been programmed.
5. Press
until you see the A-dec logo.
. Three beeps
3. From the Handpiece Setup screen, press Setup, and press
HANDPIECE SETUP
Holder Setup Electric Setup Ultrasonic Setup
Intraoral Lt Source
= BACK = SELECT = MOVE
or to select to select Ultrasonic
.
Light Source Setup
HANDPIECE SETUP
= BACK = SELECT = MOVE
Holder Setup Electric Setup
Intraoral Lt Source
Ultrasonic Setup
LT SOURCE SETUP
On When Sel
Press -/+ to change <Back
Select>
The following light source options are available:
On When Selected: Specify whether the intraoral light source turns on or remains off when the handpiece is removed from the holder.
Auto Off Delay: Determine how long the light remains on when the foot control is released. This time is reset when you use the drive air again or replace and pick up the handpiece. The factory setting is five seconds.
On in Endo: Specify whether the intraoral light source turns on or off when in endodontics mode. Because most endodontics handpieces do not have light pipes, it is recommended that Off is selected to reduce heat and to extend bulb life.
To set up the light source:
1. From the deluxe touchpad main screen, press and hold the A/B ( ) and
Program ( or ) buttons at the same time for three seconds; then press to begin. You see the System Setup screen.
2. Press the
or to select Handpieces, and press .
SYSTEM SETUP
Handpieces
Touchpad
4. Lift the desired handpiece from the holder.
5. From the Light Source Setup screen, withdraw the desired handpiece from
the holder and press
6. From the Light Source Setup screen, press minus (-) or plus (+) to select the
desired option, for example, On When Selected, and press
.
.
LIGHT SOURCE SETUP
On When Selected ON On In Endo OFF
Auto O Delay 15
= BACK -/+ = CHANGE = MOVE = SAVE
7. Specify other light source options for the handpiece by pressing minus (-),
plus (+), and three beeps confirm the setting.
8. Return the handpiece to the holder.
9. Repeat steps 3 through 7 to configure each handpiece.
10.When you are finished setting up handpieces, press
A-dec logo.
to move through the screens. Once the setup is complete,
until you see the
= BACK = SELECT
= MOVE
3. From the Handpiece Setup screen, press Lt Source, and press
86.0348.00 Rev A Programming 35
or to select to select Intraoral
.
A-dec 500 and Clinical Products Service Guide Programming 36
TOUCHPAD SETUP
= BACK = SELECT = MOVE
Contrast Adjust
Help Messages
CONTRAST ADJUST
= BACK -/+ = CHANGE
21
of 63
= SAVE
HELP MESSAGES
= BACK -/+ = CHANGE
OFF
= SAVE
Touchpad Setup
Use the touchpad setup to change the contrast of the touchpad screen display. You can also specify whether to display help messages in the technical mode help screens.
The help messages are listed in the Troubleshooting section of this document. Refer to 500 Deluxe Touchpad Help Messages, page 126, for definitions of the messages and any related troubleshooting procedures.
NOTE Help messages should only be set to On for
technician use.
To set up touchpad options:
1. From the deluxe touchpad main screen, press and hold the A/B ( ) and
Program ( or ) buttons at the same time for three seconds; then press
to begin. You see the System Setup screen.
2. Press the
or to select Touchpad, and press .
Select Contrast Adjust to adjust the contrast on the screen. Press minus (-
) or plus (+) to adjust the contrast.
Select Help Messages and press minus (-) or plus (+) to turn on or off the
technician help messages.
SYSTEM SETUP
Handpieces
Touchpad
= BACK = SELECT
= MOVE
3. From the Touchpad Setup screen, press minus (-) or plus (+) to highlight Contrast Adjust or Help Messages:
4. Press
. Three beeps confirm the setting has been programmed.

Technician Help Messages

The 500 deluxe touchpad allows technicians to view help messages. For the message descriptions, see page 126 in the Troubleshooting section.

Touchpad Circuit Board Components

7
4
5
1
3
2
9
10
8
6
8
2
9
10

Standard Touchpad

The standard touchpad has two LEDs for communicating status (AC Power and Data). The touchpad circuit board’s Status LED is the A-dec icon, visible on the touchpad. Check the chair circuit board LEDs, as well as the touchpad, when troubleshooting.
Item Description
1DS1 - AC power LED 2 DS2 - Status LEDs 3DS3 - Data LED 4 J1 - Ø VAC terminal strip 5 J1 - 24VAC terminal strip 6 P1 - Data line port (DCS) 7 J2 - Electric holder terminal strip 8 DS7 - Auxiliary 1 and bowl rinse LED 9 DS8 - Dental light LED
10 DS9 - Auxiliary 2 and cupfill LED
Figure 30. Standard Touchpad Circuit Board Components
86.0348.00 Rev A Programming 37
A-dec 500 and Clinical Products Service Guide Programming 38
1
2
13
10
3
4 5
8
6
7
9
11
12
14
11
9
10
2
13

Deluxe Touchpad

Figure 31. Deluxe Touchpad Circuit Board Components
Item Description
1DS1 - AC power LED 2 DS2 - Status LEDs 3DS3 - Data LED 4 J1 - Ø VAC terminal strip 5 J1 - 24VAC terminal strip 6 P1 - Data line port (DCS) 7 P2 - LCD display connector 8 P3 - LCD back light power connector
9 DS6 - Auxiliary 1 LED 10 DS7 - Auxiliary 2 LED 11 DS8 - Bowl rinse LED 12 DS9 - Cupfill LED 13 DS10 - Dental light LED 14 LCD display

DELIVERY SYSTEMS

4
A-dec 500 delivery systems regulate the air and water used to operate handpieces, syringes, and accessories. Delivery system options include Front­mounted delivery, both Traditional (model 532) and Continental (model 533); side mounted (model 542) and 12 O’clock, duo delivery (model 541) and assistant’s Instrumentation (model 545).

Delivery Systems Overview

Front (Chair) Mounted: The doctor’s side of the A-dec 500 system includes both
®
the Traditional (model 532) and Continental front-mount monitor (model 531), and self-contained water system. The A-dec 532 and 533 delivery systems have been designed to mount only to the A-dec 511 chair and communicate with the entire A-dec 500 product line through the Data Communications System (DCS).
Side Mounted: The A-dec 542 side delivery system is designed to mount to a
®
wall or to a Preference Collection optional tray holder can be installed. When installed, the tray holder positions over the control head.
12 O’clock, duo delivery and assistants: The A-dec 12 O’clock system features two models. The A-dec models 541 and 545 are floor-mounted delivery systems
that install with a variety of Preference Collection cabinets.
or Preference Slimline cabinet. An
(model 533) delivery systems, the
®
and Preference Slimline
Figure 32. 500 Delivery Systems
Contents
• Chair-Mounted Delivery Systems (532/533), page 40
• Side-Mounted Delivery Systems (542), page 46
• 12 O’Clock Delivery Systems (541 and 545), page 50
• A-dec 500 Delivery Systems Common Features, page 57
86.0348.00 Rev A 39
A-dec 500 and Clinical Products Service Guide Delivery Systems 40
A-dec 300 Radius Chair Mounted Delivery System
(Refer to the A-dec 300 Service Supplement, p/n 86.0273.00, for information about this option.)
A-dec 500 Chair Mounted Delivery Systems

Chair-Mounted Delivery Systems (532/533)

The chair-mounted system features two models, the A-dec Traditional (model
532) and Continental (model 533). Both models are chair-mounted delivery systems that install on the A-dec 511 chair. The standard configuration includes:
• Balanced flexarm with air brake
• Four handpiece control block positions
• Control head with room to house integrated accessories
• Autoclavable saliva ejector
• Autoclavable syringe (option of warm water syringe)
• 2-liter self-contained water system with quick-disconnect water bottle
• Standard multi-function touchpad (optional) or deluxe multi-function touchpad (optional)
Figure 33. Chair-Mounted Delivery Systems

Product Overview (532/533)

Cover
Front Mount Arm
Touchpad
Front Cover (533 only)
Flexarm Hub
Flexarm
Brake Handle
Front Mount Casting
The A-dec 500 delivery systems and front-mount monitor both mount to the A-dec 511 chair. This mounting structure provides left/right capabilities for both the delivery system and the monitor.
Figure 34. Delivery system and front-mount arm components
86.0348.00 Rev A Delivery Systems 41
A-dec 500 and Clinical Products Service Guide Delivery Systems 42
Screw
Screw
532/533
533 only
532 only
(Refer to the A-dec 300 Supplement p/n 86.0 273.00 for information about the front mounted A-dec 300 Radius delivery system option.)

Service, Usage, and Adjustments (532/533)

Covers

The delivery system covers are designed for easy access to control components.
Delivery System Cover
To open the delivery system cover, lift up on the center tab on the back of the delivery system cover and then, lift up the tabs on both sides of the delivery system cover. To remove the cover, pinch the hinge brackets at the cover base.
Delivery System Front Cover — Continental
To open the delivery system cover, move the whips forward and remove the two screws holding the front cover in place. Open the front cover carefully, until the lanyard is taut.
CAUTION Remove handpieces from the delivery system before
opening the delivery system front cover.
Figure 35. Open the Delivery System Cover and Front Covers

Whip Assembly

Attach spring here for greater tension
Attach spring here for lesser tension
Part Number: 77.0291.01/77.0291.02

Continental Tray Holder (533 only)

Part Number: 77.0294.01
Adjust the whip return tension by placing the bail in two different locations. You can remove the whip cover for easy handpiece tubing replacement. To reinstall the whip cover, fully extend the whip and attach the back starting at the ball end and work toward the handpiece end.
Figure 36. Whip Assembly with Adjustable Bail Position
Continental tray holders can be mounted on the left or right side below the delivery system.
NOTE Standard delivery systems use the Traditional tray holders. For
information on Traditional tray holders, page 63.
Figure 37. Continental Tray Holder Components
86.0348.00 Rev A Delivery Systems 43
A-dec 500 and Clinical Products Service Guide Delivery Systems 44
Leveling Cam
Stabilizing Screw, 1/4-20” x 2-1/4"
Flanged Nuts
Leveling Bolts
Bubble Level
Rotation Bolt

Adjustments (532/533)

Level Adjustment
1. Position all front-mount modules, (e.g., delivery system, flexarm, tray
holder assemblies) to align with the centerline of the chair.
2. Use the bubble level (if applicable) to determine when the correct front mount adjustments have been made.
3. Loosen the stabilizing screws and flanged nuts.
4. Adjust the leveling cam for side-to-side leveling.
5. Use one of the leveling bolts to adjust the front-to-back leveling. Assure
both bolts are touching the casting when the delivery system is level front­to-back.
6. Tighten the leveling cam securely.
7. Tighten the stabilizing screws until the screws make contact with the casting
when all of the front mount leveling adjustments have been made. Do not overtighten.
NOTE You may need to lower the back to improve access.
8. Securely tighten the flanged nuts.
Tension Adjustment
If the front-mount arm drifts, adjust the front-mount tension. To adjust the front mount tension, tighten or loosen the rotation bolt connecting the front­mount arm to the front-mount casting. Turning the rotation bolt right increases the tension on the arm and left decreases the tension.
Figure 38. Leveling and Adjusting the Front Mount Arm

Front-to-Back Leveling

Adjustment Screws
Front Knuckle Covers
1. Remove the flexarm front knuckle covers.
2. Position the delivery system in line with the flexarm.
3. Tighten or loosen the adjustment screw on the underside of the front
knuckle until the delivery system is level front-to-back.
4. Replace the covers.

Side-to-Side Leveling

Alternately tighten and loosen the two leveling screws on the underside of the delivery system until it is level side-to-side. Tighten both screws when level.
Figure 39. Level the Delivery System in Two Locations
86.0348.00 Rev A Delivery Systems 45
A-dec 500 and Clinical Products Service Guide Delivery Systems 46
Cover
Touchpad (Optional)
Front Cover
Brake Handle
Rigid Arm
Flexarm
Mounting bracket

Side-Mounted Delivery Systems (542)

The side-mounted system features the A-dec 542 delivery system. This delivery system can be used with a variety of Preference and Slimline cabinets as well as a wall mount option. The standard configuration includes:
• Balanced flexarm with air brake
• Four handpiece control block positions
• Control head with room to house integrated accessories
• Autoclavable saliva ejector
• Autoclavable syringe (option of warm water syringe)
• 2-liter self-contained water system with quick-disconnect water bottle
• Standard multi-function touchpad (optional) or deluxe multi-function touchpad (optional)
• An optional tray holder (not shown) can be installed.
Figure 40. Chair/Side Systems
Figure 41. Side-mounted 542 Delivery System
Figure 42. Delivery System and Mount Components - 542

Product Overview (532/533/542 Common Feature)

Touchpad
Delivery System Circuit Board
Control Block
Curing Light Control Module
Ultrasonic Instrument Control Module
Electric Micromotor Control Module
Accessory Holder
The A-dec 532/533/542 delivery systems feature a structural platform in the base of the control head. This metal mounting grid allows easy attachment of component parts and extra control modules required by accessories.
Figure 43. Internal Control Components Mounted on the Structural Platform
86.0348.00 Rev A Delivery Systems 47
A-dec 500 and Clinical Products Service Guide Delivery Systems 48
Adjustment Screw
Flexarm
Rigid Arm
Adjustment screw

Service, Usage, and Adjustments (542)

Flexarm Counterbalance

Adjust the flexarm counterbalance if the control head on the A-dec 542 Side Delivery System drifts up against the cabinet or drops when the master switch is turned off.
NOTE If the delivery system is not mounted under a cabinet, adjust
the counterbalance to the midpoint of the operating position. Load the tray holder with a typical load before adjusting the flex arm counterbalance.
1. Load the control head for normal use attaching handpieces and placing a loaded tray on the tray holder.
2. Move the control head to the lowest position.
3. Turn the master on/off toggle off. If the control head drifts up against the
cabinet, the counterbalance needs adjustment.
4. Insert a hex key into the adjustment screw.
Turn the hex key to the right to increase the upward drift.
Turn the hex key to the left to reduce the upward drift.
Figure 44. Flexarm Counterbalance
Remove the Covers
1. Lift up on the tabs at the back (center) and on both sides of the delivery
system cover and open the cover. See Figure 45.
2. Pinch the hinge brackets at the cover base to remove the cover.
3. Remove the two screws holding the delivery system front cover in place.
See Figure 45.
4. Slide the front cover forward and lift up.
CAUTION Remove handpieces from the delivery system before
opening the delivery system front cover.

Figure 45. Remove Covers

1) Lift on tabs and open cover.
Delivery System Cover

Rotational Tension Adjustment

Tighten or loosen the friction adjustment screw located in the center of the delivery system structural platform to adjust the rotational tension.
Figure 46. Adjusting the Delivery System Rotation Tension
Tab
2) Remove screws, slide front cover forward and lift up.
Screw
Delivery System Front Cover
86.0348.00 Rev A Delivery Systems 49
Tab
A-dec 500 and Clinical Products Service Guide Delivery Systems 50

12 O’Clock Delivery Systems (541 and 545)

The 12 O’clock system features two models, the A-dec 541 duo delivery and the A-dec 545 assistant’s instrumentation. Both models are floor-mounted delivery
®
systems that install with a variety of Preference Collection Slimline
• Height-adjustable round worksurface
• Multi-position assistant’s instrument holder
• Autoclavable saliva ejector
• Autoclavable syringe (option of warm water syringe)
• Autoclavable HVE (choice of single/dual)
• 2-liter self-contained water system with quick-disconnect water bottle
•Solids collector
• No touchpad (optional), standard multi-function touchpad (optional) or
In addition to the assistant’s instrumentation features, the 541 duo delivery system includes:
• Four handpiece control block positions
• Control center with room to house integrated ancillary equipment
• Multi-voltage intraoral light source
• Wet/dry foot control with chip blower/accessory button
• Height-adjustable, multi-position doctor’s instrument holder
cabinets. The standard configuration includes
deluxe multi-function touchpad (optional)
and Preference
Figure 47. 12 O’clock Delivery Systems
Optional tray and instrumentation holders are available. Both mount on the round worksurface.

Product Overview (541 and 545)

Round Worksurface
Doctor’s To uchpad (Deluxe Touchpad)
Doctors Instrumentation
Assistant’s Holder (Standard, or Electric)
Hub Mount
Control Center
Assistant’s Touchpad
Tray Holder Mount
Control Block
Flush Toggle
Control Center Circuit Board
Oil Collector

A-dec 541 12 O’Clock Duo Delivery System

Figure 48. Delivery System Components - 541
A-dec 541 Internal Components
NOTE For information on the master on/off toggle, refer to Master
On/Off Toggle, page 62.
Figure 49. Internal Components
86.0348.00 Rev A Delivery Systems 51
A-dec 500 and Clinical Products Service Guide Delivery Systems 52
Holder Assembly
Clamp
Knob
Opening for Cover Removal
Doctor’s Cover Assembly
Doctor’s Cover Assembly

Flex-Holder

Part Number: 99.0705.00/99.0705.01 (Syringe/Saliva Ejector),
99.0706.00/99.0706.01 (Standard HVE), 99.0707.00/99.0707.01
(15 mm HVE)
The flex-holder can mount to a variety of worksurfaces. The mounting location versatility allows you to locate an additional holder where you need it.
Figure 50. Flex-Holder

Control Center Covers

Some adjustments to the 541 duo delivery require the removal of covers. To remove covers, locate the hole directly under the control center and pull the covers apart.
To replace covers, position a cover over the control center and snap into place, then position the other cover and snap the two covers together.
NOTE Make sure the umbilical tubing is contained inside the covers
and not pinched.
Figure 51. Control Center 541 Cover Removal

Doctor’s Holder

541 Arm Assembly
End Cap
Deluxe Touchpad Option
541 Holder Assembly
No Touchpad Option
Standard Touchpad Option
T
o
r
e
m
o
v
e
co
v
e
r
,
s
l
i
d
e
i
n
t
h
i
s
d
i
r
e
ct
i
o
n
.
The doctor’s holder is a configurable design. The holder configurations available from A-dec are:
• Two on the left side of the rotation hub and three on the right, syringe at left-most position
• Three on the left side of the rotation hub and two on the right, syringe at right-most position
The holders use a pneumatic A-dec 500 microvalve for handpiece activation. The microvalve located in the syringe position is unplumbed and can be used as a replacement valve. A maximum of four valves connect to the holdback air system on the control block. These valves can be locked out or activated as needed.
The top cover of the 541 arm assembly is easily removed by sliding it backwards and lifting off.
CAUTION Do not pinch any wires or tubings when reinstalling the
cover.
Figure 52. Doctor’s Holder Assembly
86.0348.00 Rev A Delivery Systems 53
A-dec 500 and Clinical Products Service Guide Delivery Systems 54
Tighten this screw
Balance point
Loosen this screw
To raise this end

Service, Usage, and Adjustments (541 and 545)

Level the Arm Assembly and Worksurface

Level the Arm Assembly Front To Back
1. Position the arm perpendicular to the cabinet. See Figure 53.
NOTE Make sure the arm assembly is in line with the worksurface
and perpendicular to the cabinet before leveling
2. Place the level on the hub, parallel to the arm.
3. Loosen and tighten the mounting screws as necessary.
TIP The support arm balances on a pivot. To raise or lower the long
side of the arm, adjust the screw nearest the cabinet. See Figure 55. Once the arm is level, tighten the second screw to hold the arm in place.
Level the Arm Assembly Side To Side
Position the arm parallel to the cabinet. See Figure 54.
4. Place level on the hub, parallel to the arm.
5. Loosen and tighten the mounting screws as necessary. See Figure 55.
Figure 54. Adjustment - Arm Assembly Leveling Side to Side
Figure 55. Level the Hub
Figure 53. Leveling - Arm Assembly Leveling Front to Back
Level Round Worksurface
Setscrews
Level
Worksurface
Worksurface Screws
Master On/ Off Toggle
Vertical Post of the Worksurface
Height Adjustment Ring for Worksurface
Vertical Post of the Arm
2-1/2” (63.5 mm) Adjustment Range
3" (76.2 mm) Adjustment Range
Height Adjustment Ring for Arm
Vertical Post of the Worksurface
Height Adjustment Ring for Worksurface
To level the round worksurface, adjust the two set screws in the worksurface support housing until the surface is level. See Figure 56.
NOTE Ensure that the arm assembly is level before leveling the
worksurface.

Worksurface Height

To adjust the height of the worksurface or the instrumentation arm:
1. Lift the upper part of the vertical post.
2. Slide the height adjustment ring to the desired position.
3. Lower the vertical post onto the ring.
1. Place the worksurface arm in a typical working position and center a level
on the round worksurface over the support housing.
2. Loosen the worksurface screws on the side nearest the master on/off toggle.
3. Use a 1/8" hex key to adjust the two setscrews in the worksurface support
housing until the surface is level.
4. Tighten the worksurface screws until the worksurface is secure.
NOTE Do not overtighten the worksurface screws or the level may
change.
Figure 56. Level Work Surface
NOTE The height adjustment range for the worksurface is 2-1/2"
(63.5 mm) and the doctor’s instrumentation arm is 3" (76.2 mm).
Figure 57. Worksurface and Doctor’s Instruments Height Adjustment - 541
Figure 58. Worksurface Height Adjustment - 545
86.0348.00 Rev A Delivery Systems 55
A-dec 500 and Clinical Products Service Guide Delivery Systems 56

Instrumentation Arm Positioning

The A-dec 500 instrumentation holders offer horizontal and vertical positioning. Each holder rotates for independent angle adjustment.
Position Holders
You can customize the position of each holder on the holder assembly arm:
1. Pull holder slightly away from the adjacent holder.
2. Twist to the desired position and release See Figure 60.
NOTE Doctor’s and assistant’s holders adjust in the same manner.
Figure 59. Instrumentation Arm Positioning
Figure 60. Instrument Holder Positioning
Holder positions rotate independently

A-dec 500 Delivery Systems Common Features

A-dec 500 delivery systems have features in common, including:
•Tubing
• Quick-connect fittings
• Data communication system
• Control components
• Master on/off toggle
• System accessories
Figure 61. A-dec 500 Delivery System Common Features
86.0348.00 Rev A Delivery Systems 57
A-dec 500 and Clinical Products Service Guide Delivery Systems 58
Barb Retainer Air Coolant Tubing
Water Coolant Tubing
Drive Air Tubing

Product Overview (All Delivery Systems)

A-dec Tubing

A-dec products use four sizes of tubing: 1/8" OD, 1/4" OD, 3’8" OD and 5/16" OD. A-dec 500 products use the 5/16” OD for all high flow air and water applications.
Starting with A-dec 500, products use silicone handpiece tubing. The silicone handpiece tubing uses a European color code for air (blue) and water (green) that differs from the current U.S. standard.
Figure 62. Handpiece Tubing Connections

Quick Connect Fittings

A-dec 500 products use quick-connect fittings for some tubing connections. These fittings provide fast, secure, push-on installation of tubing. These quick­connects also make removal of tubing during servicing easy.

Connect Tubing

To connect tubing using the quick-connects:
1. Cut the tubing square, to ensure a secure connection.
2. Push the tubing into the connector until it can go no further.
3. Pull gently on the tubing to verify grip action.

Disconnect Tubing

1. Depressurize the unit.
2. Push and hold the release ring on the fitting connector.
3. Remove the tubing from the connector.
Figure 63. Make a Square Cut
Figure 64. Push Tubing Into the Fitting
Figure 65. Remove Tubing From the Fitting Connector

Data Communication System

DCS Lines
Data Line Connectors
A-dec 500 equipment uses an electronic communication tool, the Data Communication System (DCS). The DCS allows each module to communicate with the other modules and automatically detects when modules are added. For example, when connecting a cuspidor to the system, the DCS automatically recognizes the cuspidor module without any programming or setting changes.
You can plug any module into the data line connector on the circuit board. The circuit board recognizes which module has been plugged in and allows operation of that module to begin. Should a module malfunction or fail to work, the DCS maintains service to the rest of the modules. Other modules continue to function in spite of the one that has failed.
Figure 66. DCS Line and Connector
86.0348.00 Rev A Delivery Systems 59
A-dec 500 and Clinical Products Service Guide Delivery Systems 60
8
1
2
18
3
4
5
6
7
9
10
11
12
13
14
15
16
17
19
20
22
21
23

Circuit Board Components

Part Number: 90.1076.00
Table 5. Delivery System Circuit Board Components
Item Description
1 P1, P2, P3 - Data line ports (DCS) 2 J6 - ultrasonic instrument power terminal strip 3 J4, J5 - Intraoral light source strip and LEDs DS7, DS8, DS9, DS10 4 J2 - 24 VAC terminal strip 5 J2 - Ø VAC terminal strip 6 DS6 - Air coolant solenoid LED and connector 7 DS5 - Water coolant solenoid LED and connector 8 X8 - Water coolant signal transducer (foot control toggle) 9 X7 - Air coolant signal transducer 10 X6 - Accessory button (foot control) signal transducer 11 X5 - Drive air signal transducer 12 X4 - Handpiece 4, holdback transducer 13 X3 - Handpiece 3, holdback transducer 14 X2 - Handpiece 2, holdback transducer 15 X1 - Handpiece 1, holdback transducer 16 J3 - Handpiece 5, terminal strip 17 DS11, DS12 - Drive air LED display 18 DS1 - AC Power LED 19 DS2 - Status LED 20 DS3 - Data LED 21 S2 - Increase intraoral light source voltage adjustment 22 S1 - Decrease intraoral light source voltage adjustment 23 DS13 - ultrasonic instrument relay LED
Figure 67. Delivery System Circuit Board

Brake Handle

Brake Button
Brake Handle
Hub
Syringe/Saliva Holder Assembly
Part Number: 77.0104.03 (White) 77.0104.01 (Surf)
The brake handle activates a replaceable microvalve, releasing pressure on the brake inside the flexarm.
Figure 68. Brake Handle Components

Accessory Holder

Part Number: 99.0681.01 (White) 99.0681.00 (Surf); Standard,
99.0684.01 (White) 99.0684.00 (Surf) Auto-electric
The accessory holder is either an active or non-active position on the delivery system. It can be mounted on either side of the delivery system, opposite the touchpad. Auto-electric holders have a power source at the terminal for handpiece position 5 on the delivery system circuit board.
Figure 69. Accessory Holder Components
86.0348.00 Rev A Delivery Systems 61
A-dec 500 and Clinical Products Service Guide Delivery Systems 62

Master On/Off Toggle

Master On/Off Toggle

Flush Toggle

Flush Toggle
Master On/Off Toggle
Part Number: 33.0175.01 (White) 33.0048.05 (Surf)
The master on/off toggle is located on the worksurface support of the 541 delivery system and on the side of the 532, 533 and 542 delivery systems. The toggle operates with unregulated air and supplies air to the control block for holdback air. When the master toggle is off, the delivery system does not have power.
Figure 70. Master On/Off Toggle — 541
Figure 71. Master On/Off Toggle — 532/533/542
Flush Toggle
Part Number: 33.0168.01 (White) 33.0168.00 (Surf)
The flush toggle is located on side of the 541 12 O’clock duo delivery system and on the back of the 532, 533 and 542 delivery systems. The flush toggle uses holdback air to operate handpiece flush functions. When the momentary valve is on, the holdback air is exhausted and air signal is sent toward the water cartridge.
• If the handpiece is in the holder (holdback active), the flush is not activated.
• If the handpiece is out of the holder (holdback exhausted), the flush operates.
Figure 72. Flush Toggle — 541
s
Figure 73. Flush Toggle — 532/533/542

Traditional Tray Holder

Piston
Signal Relay Valve
Diaphragm
Part Number: 77.1024.00 (White) 77.0294.00 (Surf); 77.1023.00 (White) 77.0295.00 (Surf)

Foot Control

Part Number: 38.1829.01 (White) 38.1805.01 (Surf); 38.1829.00 (White) 38.1805.00 (Surf)
The traditional tray holder can be mounted on the left or right side below the delivery system.
Figure 74. Traditional Tray Holder Components
The A-dec 500 foot control operates handpieces by applying pressure on the foot control. As pressure is applied, the black piston exhaust vent seals against the poppet, then pushes the poppet away from the inlet seat, and when unseated, regulated air flows through the valve and out to the delivery system as drive air.
Regulated air is sent through the relay block to the delivery system as air coolant signal. When pressure is released from the foot control, the poppet reseals the inlet and pressure is exhausted from the outlet side of the valve and up through the piston.
Figure 75. A-dec Foot Control
86.0348.00 Rev A Delivery Systems 63
A-dec 500 and Clinical Products Service Guide Delivery Systems 64

Self-Contained Water Bottle

Part Number: 14.0468.01 (White)/14.0468.00 (Surf)
The A-dec self-contained water system is a closed system, which isolates treatment water from the municipal water supply. The A-dec 500 water bottle is designed to prevent cross-contamination during refilling. The pickup tube is located inside the bottle assembly and is not exposed to outside contaminates. Air pressure forces water from the water bottle into the pick-up tube. Water is distributed to the delivery system, syringe, handpieces and assistant’s syringe.
WARNING Use only A-dec water bottles. Do not use other brands, or
damaged bottles. They can pose a serious safety hazard if broken while pressurized. The A-dec plastic water bottles cannot withstand heat sterilization. Attempting to do so damages the bottle and the sterilizer.
CAUTION Use caution when using the self-contained water system
with any dental units equipped with components that might fail when the water supply is interrupted. Some types of ultrasonic instruments and water heaters are examples of components that can be permanently damaged if operated without a continuous water source. A-dec does not recommend using saline solutions, mouth rinses, or any chemical solutions, not specified by A-dec, in the A-dec self­contained water system. These may damage the water system components and cause the failure of the dental unit.
Figure 76. Self-Contained Water Bottle

Standard Syringe

Continental Syringe
Continental Wa rm Water Syringe
Traditional Warm Water Syringe
Traditional Syringe
Syringe Tip
Syringe Nut, Smooth
Traditional Syringe Head
Button Valve
Syringe handle
Continental Syringe Head
Continental Wa rm Water Syringe
T raditional W arm Water Syringe
Button Valve
Syringe Tip
Syringe Nut, Smooth
Continental and Traditional autoclavable syringes have serviceable air and water button valves. The tip retaining nuts have replaceable O-rings. See the A-dec 500 Service Reference (p/n 86.0329.00) for the warm water syringes’ illustrated parts breakdown. See the Illustrated Parts Breakdown (p/n 85.0851.00) for Continental and Traditional syringe parts.
WARNING Turn off and bleed the system of air and water before
servicing the syringe. The use of disposable syringe tips in A-dec syringe tip nuts is not recommended.
Figure 77. Autoclavable Syringes

Warm Water Syringe

The Continental and Traditional warm water syringe share many of the same components. The syringe head is the only component specific to each model of syringe.
Figure 78. Warm Water Syringe
86.0348.00 Rev A Delivery Systems 65
Syringe Temperature Selection Input
The warm water syringe temperature selection input allows the technician to select the estimated patient cheek temperature between a minimum temperature of 82.4ºF (28ºC) and a maximum temperature of 100.4ºF (38ºC).
A-dec 500 and Clinical Products Service Guide Delivery Systems 66
2
4
5
6
7
8
9
10
11
1
3
Warm Water Syringe Specifications
There are five positions ranging from minimum to maximum for temperature variations.
Table 6. Warm Water Syringe Temperature Selections
Jumper Selection Estimated Temperature at Patient’s Cheek
5 (maximum) 100.4ºF (38ºC) 4 96.8ºF (36ºC) 3 93.2ºF (34ºC) 2 89ºF (32ºC) 1 (minimum) 86.0ºF (30ºC) No jumper 82.4ºF (28ºC)
NOTE If two jumpers are installed, the micro controller averages the
two selections together. If three or more are installed, only the lowest two jumpers are recognized.
Table 7. Warm Water Syringe Circuit Board
Item Description
1 DS1—AC power LED 2 DS2—Status LED 3 DS3—Data LED - NOT USED (reserved for future use) 4 DS4—Heater output LED 5 DS5—Thermistors (temperature sensors) LED 6 J1—0 VAC terminal strip 7 J1—24 VAC terminal strip 8 J2—Heater output terminal strip
9 P1—Temperature selection header 10 P2—Data line port (DCS) - NOT USED - (reserved for future use) 11 P3—Thermistors (temperature sensors) connector
Figure 79. Warm Water Syringe Circuit Board

Tooth Dryer

Red
Blue
Green
The A-dec warm air tooth dryer needs a dedicated four-hole handpiece tubing and a minimum of 60 psi compressed air to operate properly. The tooth dryer is only plumbed to the drive air on the control block. Remove the barb from the exhaust tubing (red) and cut the tubing at an angle to reduce noise. Anchor, but do not pinch, the exhaust tubing to the structural platform using a cable tie. There are no moving parts in the tooth dryer, only cleaning is needed.
Figure 80. Connecting the Tooth Dryer to the Control Block Drive Air
86.0348.00 Rev A Delivery Systems 67
A-dec 500 and Clinical Products Service Guide Delivery Systems 68

Service, Usage, and Adjustments (Common Features)

Control Block

Part Number: 38.1776.00
All A-dec 500 delivery systems use the same control block assembly for adjusting handpiece air and water coolant flow. The adjustment controls for air and water coolant are located on the side of the delivery system.
Figure 81. Control Block — 541
Figure 82. Control Block — 532/533/542
Air Coolant
Adjusting the air coolant for one handpiece sets it for all of the positions.
1. Flip the wet/dry toggle on the foot control to the on position.
2. Install a bur into a handpiece and activate the position.
3. Fully depress the foot control.
4. Adjust the air coolant adjustment key to create a cooling mist at the bur.
Water Coolant
Drive Air Adjustment Stem
Water Coolant Adjustment Stem
Air Coolant Adjustment Stem
Increase Flow
Decrease Flow
Drive Air
1. Flip the wet/dry toggle on the foot control to the ON position.
2. Install a bur into the handpiece and activate the position.
3. Fully depress the foot control.
4. Turn the water coolant adjustment key to adjust the flow to fit the operator’s
needs.
5. Repeat steps two through four for each wet handpiece.
Figure 83. Location of the Air and Water Coolant Adjustments
1. Use a handpiece pressure gauge to adjust the drive air for each handpiece
tubing.
2. Install a bur into the handpiece and activate the position.
3. Fully depress the foot control.
4. Adjust the drive air flow (controls are on the top of the control block).
Air turbine handpieces should be set to 40 psi. EA-50LT electric micromotors (controller) should be set to a minimum of 60 psi.
NOTE The digital gauge inside the delivery system reads drive air
pressure only at the control block. For accurate handpiece pressure settings, use a pressure gauge at the end of the handpiece tubing.
Figure 84. Adjusting Flow
86.0348.00 Rev A Delivery Systems 69
A-dec 500 and Clinical Products Service Guide Delivery Systems 70

Remove the Control Block

532/533/542 Delivery Systems
1. Loosen the hex screw holding the master on/off toggle mounting bracket to
the delivery system platform.
2. Slide the bracket toward the back of the delivery system.
3. Lift the control block from the back, unhooking the front of the block.
Figure 85. Remove Control Block on 532/533/542 Delivery Systems
US Pat. 5,201,899
541 Delivery System
1. Loosen and remove the hex screw and washer that secures the control block
to the control center frame.
2. Slide the control block up so that the frame tab is no longer connected in the
control block mounting slot.
Figure 86. Remove Control Block on 541 Delivery System

Tra y Ho lde r

Leveling Screw
Mounting Screw­DO NOT ADJUST
Leveling Screw
Adjust Tray Holder Tension
• To tighten the rotational tension of the tray holder, turn the hex bolt under the tray to the right.
• To loosen the rotational tension, turn it to the left.
• To adjust the rotational tension of the tray arm, tighten or loosen the hex bolt at the delivery system end of the arm.
Level Delivery System Tray
1. Use the adjustment screw to level side-to-side. See Figure 87.
2. Use the adjustment screws to level front-to-back. See Figure 88.
Figure 87. Side-To-Side Tray Level
Figure 88. Front-To-Back Tray Level
86.0348.00 Rev A Delivery Systems 71
A-dec 500 and Clinical Products Service Guide Delivery Systems 72

Intraoral Light Source Voltage

The intraoral light source voltage adjustment on the A-dec 500 doctor’s delivery system is located on the delivery system circuit board. The voltage is preset to
3.2 volts at the lamp terminals when the lamp is on.
Figure 89. Intraoral Light Source Voltage
Intraoral Light Source Length and Voltage
Table 8. Length and Voltage Table
Wire length in A-dec tubing Voltage at terminal strip A-dec/W&H,
Bien Air or other bulbs rated at 3.5V (On) (in) (cm) VDC +/- .02 (in) (cm) VDC +/- .02 48 122 3.40 108 274 3.69 54 137 3.43 114 290 3.72 60 152 3.46 120 305 3.75 66 168 3.49 126 320 3.78 72 183 3.52 132 335 3.81 78 198 3.55 138 351 3.84 84 213 3.58 144 366 3.87 90 229 3.61 150 381 3.90 96 244 3.64 156 396 3.93 102 259 3.67
Wire length in A-dec tubing Voltage at terminal strip A-dec/W&H, Bien Air
or other bulbs rated at 3.5V (On)
NOTE Table 8 pertains to fiber-optics powered with 26AWG wires, 750mA loads, and a desired bulb voltage of 3.2VDC. For fiber-optics powered
with 26AWG wires and other ratings, use the equation:
T = (Z x 0.006 x Y) + X where: T: Terminal strip voltage (VDC) X: Desired voltage at lamp (VDC)
Y: Rated lamp/load current (in Amps) Z: Length of 26AWG wire (inches) from terminal trip to lamp

ASSISTANT’S INSTRUMENTATION (551 AND 545)

5
The A-dec 551 assistant’s instrumentation is equipped with either a short or long assistant’s arm for easy positioning of instrumentation. Both arms are equipped with a touchpad and a holder assembly with independent positioning features. The solids collector, located at the base of the arm, is also a part of the assistant’s instrumentation.
The A-dec 545 assistant’s instrumentation is floor-mounted and installs with a variety of Preference
standard configuration for 545 has a height-adjustable round worksurface, multi-position assistant’s instrument holder, autoclavable saliva ejector, autoclavable syringe (option of warm water syringe), autoclavable HVE (choice of single/dual), 2-liter self-contained water system with quick-disconnect water bottle, solids collector, standard multi-function touchpad (optional) and no touchpad (optional).
Collection
®
and Preference Slimline cabinets. The
Figure 90. Assistant’s Instrumentation
86.0348.00 Rev A 73
A-dec 500 and Clinical Products Service Guide Assistant’s Instrumentation (551 and 545) 74
Vacuum Pipe Kit
Red O-ring, Package of 10
1 *
Buttonhead Socket Screw, 10-32 x 1/4"
Vacuum Connector Kit

Product Overview (551)

Figure 91. 551 Assistant’s Long Arm Assembly with a Single 3 Position Holder Assembly
* Item 1:
• Long arm with single 3-position or 4-position standard holder, or dual 2-position standard holder
• Short arm with single 3-position or 4 position standard holder, or dual 2-position standard holder

Holders

Turret
Turret Insert
*
• Holder, SE/syringe kit
•HVE standard kit
End Cap Assembly
Axle (Included with Items *)
Assistant’ s arm assembly
End Cap
Standard Touchpad
Assistant’s Holder Assembly
There are two types of assistant’s holders, standard and electric.
Standard Holder
The touchpad does not need to be removed when disassembling the standard holder.
Figure 92. Standard Assistant’s Holder Components
Figure 93. Standard Holder Assembly
86.0348.00 Rev A Assistant’s Instrumentation (551 and 545) 75
A-dec 500 and Clinical Products Service Guide Assistant’s Instrumentation (551 and 545) 76
Turret
Turret Insert
Inner Hub
End Cap Assembly
Bearing
Stop
Spacer
Head Mount
Knuckle Cover
Axle
Bearing
• Holder, SE Electric Holder Kit
• Holder, HVE Electric Holder Kit, 11mm, 15mm
Electric Holder
Each handpiece in the holder assembly automatically activates when you lift it from its holder. The electric holders provide vacuum on/off switching for users whose vacuum system requires this functionality. The vacuum pump activates when the high volume evacuator (HVE) or saliva ejector (SE) are lifted from their holders, and turns off when replaced into their holders.
Figure 94. Electric Assistant’s Holder Components
Figure 95. Single Electric Holder Assembly
Figure 96. Dual Electric Holder Assembly
CAUTION When reassembling the electric holder, rotate all holders,
holder assemblies and the touchpad to ensure there is enough service loop in the wires to prevent them from breaking.

Assistant’s Touchpad

24 VAC
Data Line
24 VAC
Data Line
Electric Holder
Standard Holder Connections
Standard Holder Connections
1
3 4
5
6
7 2
8
7 9 10
The assistant’s instrumentation uses a standard touchpad that allows for control of the chair, dental light, and cuspidor or A-dec relay module(s). The touchpad can rotate 340° for access and visibility.
Item Part Number Description
1 90.1078.00 Standard touchpad with circuit board 2 77.0335.01
77.0335.00 3 025.002.01 Cable tie, package of 10 4 041.663.00 Cable bushing 5 77.0123.00 Bolt 6 99.0651.00 Spline 7 004.173.00 Washer, Nylatron 8 003.078.00 9 004.237.00 Washer, wave
10 004.060.00 Washer
Touchpad base, white Touchpad base, gray
Screw, socket head, 4-40 x 1/4
"
Figure 98. Electric Assistant’s Touchpad Components
Figure 97. Assistant’s Touchpad
86.0348.00 Rev A Assistant’s Instrumentation (551 and 545) 77
A-dec 500 and Clinical Products Service Guide Assistant’s Instrumentation (551 and 545) 78
Vacuum T ower Cap
O-ring
Screen
O-ring
Connector, triple vacuum
Vacuum Tower Base with Funnel
Red O-ring
Cap
Pivot Joint

Solids Collector

The solids collector screen prevents solids from entering the central vacuum system when the HVE or SE are used. If the vacuum system performance is less than optimal, verify that the screen has been replaced.
CAUTION Use appropriate gloves when handling contaminated parts.
1. Lift off the screen and cap, then twist off the triple vacuum connector.
2. Turn the vacuum tower base to face the back of the chair and lift up.
Figure 99. Solids Collector Components

Tip-Up Feature

The pivot joint, located close to the support link, allows the entire arm to tip up, when it contacts an object.
Figure 100. 551 Assistant’s Arm Tip-Up Feature

Adjustments (551)

Locking Knob

Left or Right Conversion

The assistant’s instrumentation can be easily positioned to the opposite side of the chair.To reposition the assistant’s instrumentation:
1. Loosen the support link locking knob located under the swivel brake
handle.
2. Unsnap the support link cover, and reposition the arm to the opposite side.
If the support side system includes a support center, tip up the assistant’s arm so the support center can swing past it when repositioning the assistant’s arm.
3. Reattach the support link cover; tighten the support link locking knob.
Figure 101. Reposition the Assistant’s Arm
86.0348.00 Rev A Assistant’s Instrumentation (551 and 545) 79
A-dec 500 and Clinical Products Service Guide Assistant’s Instrumentation (551 and 545) 80
Arm Assembly
Round Worksurface
Hub Mount
Assistant’s Holder (Standard)
Assistant’s Holder (Electric)
Master On/Off Toggle (Under Worksurface)

Product Overview (545)

Figure 102. 545 Assistant’s Instrumentation

Solids Collector

Solids Collector Cap
Solids Collector Screen
Vacuum Canister Base
Connector Cap
Screw
The solids collector screen prevents solids from entering the central vacuum system when the HVE or SE is used. A regularly maintained solids collector is necessary for optimal performance of the vacuum system. If the vacuum system performance is less than optimal, verify that the screen has been replaced.
CAUTION Use appropriate gloves when handling contaminated
parts.
Remove Solids Collector and Replace Screen
1. Turn off vacuum or open the HVE control valve.
2. Remove the solids collector cap.
3. Remove the solids collector screen.
4. Discard the screen according to local regulations.
5. Insert new screen in the collector and replace the cap.
Figure 103. Solids Collector
86.0348.00 Rev A Assistant’s Instrumentation (551 and 545) 81
A-dec 500 and Clinical Products Service Guide Assistant’s Instrumentation (551 and 545) 82
Tighten Here
Balance Point
Loosen this Screw
To Raise Here

Leveling (545)

Arm Assembly

Front-To-Back
1. Position the arm perpendicular to the cabinet. See Figure 104.
NOTE Make sure the arm assembly is in line with the worksurface
and perpendicular to the cabinet before leveling
2. Place the level on the hub, parallel to the arm.
3. Loosen and tighten the mounting screws as necessary.
TIP The support arm balances on a pivot. To raise or lower the long
side of the arm, adjust the screw nearest the cabinet. See Figure 106. Once the arm is level, tighten the second screw to hold the arm in place.
Figure 104. Leveling - Arm Assembly Leveling Front to Back
Side To Side
1. Position the arm parallel to the cabinet. See Figure 105.
2. Place level on the hub, parallel to the arm.
3. Loosen and tighten the mounting screws as necessary. See Figure 106
Figure 105. Adjustment - Arm Assembly Leveling Side to Side
Figure 106. Level the Hub
.

Worksurface

Setscrew
Level
Worksurface
Worksurface Screws
Master On/Off Toggle
Level Round Worksurface
To level the round worksurface, adjust the two set screws in the worksurface support housing until the surface is level. See Figure 107.
NOTE Ensure that the arm assembly is level before leveling the
worksurface.
1. Place the worksurface arm in a typical working position and center a level
on the round worksurface over the support housing. See Figure 107.
2. Loosen the worksurface screws on the side nearest the master on/off toggle.
3. Adjust the two setscrews in the worksurface support housing until the
surface is level. Use a 1/8” hex key. See Figure 107.
4. Tighten the worksurface screws until the worksurface is secure.
NOTE Do not overtighten the worksurface screws, or the level may
change.
Figure 107. Level Work Surface
86.0348.00 Rev A Assistant’s Instrumentation (551 and 545) 83
A-dec 500 and Clinical Products Service Guide Assistant’s Instrumentation (551 and 545) 84
Vertical Post
Height Adjustment Ring
2-1/2" (63.5 mm) Adjustment Range
Height Adjustment Ring
Vertical Post

Adjustments (545)

Worksurface Height

Worksurface and Instrumentation Arm Height
The A-dec 12 O’clock system has an arm you can adjust for efficient operation and comfort. To adjust the height of the worksurface or the instrumentation arm:
1. Lift the upper part of the vertical post.
2. Slide the height adjustment ring to the desired position.
3. Lower the vertical post onto the ring.
NOTE The height adjustment range for the worksurface is 2-1/2" (63.5
mm); for the doctor’s instrumentation arm, 3" (76.2 mm).
Figure 108. Worksurface and Doctor’s Instruments Height Adjustment - 545
Figure 109. Worksurface Height Adjustment - 545

Instrumentation Arm Positioning

Holder positions rotate independently
The A-dec 500 instrumentation holders offer horizontal and vertical positioning. Each holder rotates for independent angle adjustment See Figure 110.
Assistant’s Arm
To adjust the assistant’s arm tension:
1. Tighten or loosen the screw located under the assistant’s arm.
2. Check the adjustment. See Figure 111.
To position individual holders: You can customize the position of each holder on the holder assembly arm:
1. Pull holder slightly away from the adjacent holder.
2. Twist to the desired position and release. See Figure 112.
Figure 110. Instrumentation Arm Positioning
Figure 111. Assistant’s Arm Adjustment
Figure 112. Instrument Holder Positioning
86.0348.00 Rev A Assistant’s Instrumentation (551 and 545) 85
A-dec 500 and Clinical Products Service Guide Assistant’s Instrumentation (551 and 545) 86

CUSPIDOR

Support Center
Cuspidor
6
The cuspidor is installed on the lower support arm. The support side-mounted water bottle and the cuspidor circuit board are inside the support center. This section provides information related to servicing, maintenance and adjustments for the cuspidor.
Figure 113. 561 Cuspidor and Support Center
86.0348.00 Rev A 87
A-dec 500 and Clinical Products Service Guide Cuspidor 88
Relay outputs
N.O.
COM
N.C.
1
2
3
10
14
4
5 6
9
8
11
12 13
7
Relay jumper
15 16

Product Overview

The support center houses the cuspidor’s fill/rinse manifold, cuspidor circuit board, and support side water bottle. The cuspidor is preset with an automatic timed cupfill and bowl rinse. You can program both the cupfill and the bowl rinse functions.

Cuspidor Circuit Board Components

Part Number: 90.1079.00
The cuspidor circuit board controls the bowl rinse and cupfill solenoids. The LEDs on the circuit board indicate the status of the cuspidor and can be used for troubleshooting. The circuit board is also equipped with a built-in A-dec relay. The built-in relay functions exactly like the A-dec relay module.
Item Description
1P5 - DCS terminals 2J1 - Ø VAC terminal strip 3 J1 - 24VAC terminal strip 4 P4 - Cuspidor limit switch connector 5P2 - Bowl rinse switch connector 6 P3 - Cupfill switch connector 7P1 - A-dec relay selection header 8 DS5 - Bowl rinse relay LED
9 DS6 - cupfill relay LED 10 DS4 - Auxiliary relay LED 11 DS1 - AC power LED 12 DS2 - Status LED 13 DS3 - Data LED 14 DS7 - Limit switch LED 15 P6 - Bowl rinse solenoid connector 16 P7 - cupfill solenoid connector
Figure 114. Cuspidor Circuit Board Components

Service, Usage, and Adjustments

Cupfill
Bowl Rinse
Cupfill Spout
Bowl Rinse Button Cupfill Button
Bowl Screen
Bowl Rinse Spout

Cupfill Functions

• A quick press of the cupfill button activates a timed operation.
• A long press activates a manual operation.
NOTE The cupfill will only run a maximum of two minutes in the
manual operation mode.

Bowl Rinse Functions

• A quick press of the bowl rinse button activates a timed operation.
• A long press activates the manual operation.
• Two presses in less than two seconds, activates the continuous operation. Press the bowl rinse button once to stop the continuous operation.
NOTE There is no maximum time limit for this function.
Figure 115. Cuspidor Buttons for Cupfill and Bowl Rinse

Cupfill and Bowl Rinse Timing

Cupfill and bowl rinse functions can be reprogrammed using a touchpad or footswitch program button and the appropriate button on the cuspidor.
1. Press the Program button. One beep indicates programming mode is ready.
2. Press and hold the Cupfill or Bowl Rinse button for the desired amount of
time.
3. Release the button. Three beeps indicate the programmed time has been
changed.
Figure 116. Cuspidor Programming Features
86.0348.00 Rev A Cuspidor 89
A-dec 500 and Clinical Products Service Guide Cuspidor 90

FLOOR BOX

B
A
F
C
D
E
G
H
7
This section provides information related to servicing, maintenance and adjustments. For information on service parts, see the Genuine A-dec Service Parts Catalog (p/n 85.5000.00) or contact A-dec customer service.

Product Overview

The floor box contains the air and water manual shutoff valves, filters, air and water regulators, pressure pre-regulator, vacuum and gravity drains.

Air and Water Manual Shutoff Valves

Shutoff valves supply the air and water to the unit. To prevent leaks, these valves should remain fully open (turned counterclockwise) except while the unit is being serviced.

Gauge and Pre-Regulator

The pre-regulator controls the air and water pressure in the unit. The gauge displays the unit air pressure.
Figure 117. Identification of the Floor Box Utilities
86.0348.00 Rev A 91
Item Description Item Description
A Electrical Duplex Outlet E Moisture Separator (optional) B Water Filter Regulator (cuspidor
only) C Air Filter Regulator G Cuspidor Drain D Gauge and Pre-regulator
Assembly
F Vacuum Drain
H Manual Shutoff Valves
A-dec 500 and Clinical Products Service Guide Floor Box 92
Filter Housing
New Filter
Beveled Edge
O-ring

Service, Usage, and Adjustments

Filter Element

A clogged filter reduces the amount of air and water pressure available to the unit. To replace:
1. Move the master on/off toggle to the off position.
2. Close the manual shutoff valves.
3. Bleed the system of air and water pressure by operating the syringe buttons
until air and water no longer flow.
4. Remove the filter housing from the air and/or water regulator assembly
with a flat blade screwdriver.
5. Replace the filter if it is visibly clogged or discolored.
6. Install the new filter with the beveled edge toward the manifold.
CAUTION Be sure to install the new filter with the beveled side
toward the manifold. The unit may not work properly if installed incorrectly.
NOTE Turn the pre-regulator knob clockwise to increase air or water
pressure. Turn the pre-regulator know counterclockwise to decrease air or water pressure. Read the pressure gauge while adjusting. Water pressure will increase/decrease by half gauge indication.
Figure 118. Replacing the Filter
Loading...