A&D AD-4329 User Manual

0 (0)

' a

.-'" '1

i-'ii ® [:|f ;I^IG

DlMinMy(STOi

W e i g h i n g I n d i c a t o r

WM+PD4000173D

A&D Company, Limited

■!<*K!

A This is a hazard aiert mark.

This mark informs you about tha operation of the product

of this manual may be photocopied, reproduced, or translated into another lianguage without the prior writteri consent of A&D Company, Ltd,

Product specifications are subject to change without any obligation on the part of the manufacture.

© 2005 A&D Company Ltd. Ail rights reserved;

Contents

 

Compliance........................... .........................................................................

......... 3

1.1.

к

Compliance with FCC Rules................ ...........................................................

3

1.1.2,

 

Compliance with European Directive.............................................................

3

 

Introduction................................................... ....................................... ............. .

4

 

Installation and Precautions ............. ............... ................................. ......... .........

5

3.1.1.

 

Installation and Precautions..........................................................................

5

3.1.2.

 

The Load Ceil Connections............................................................................

5

3.1.3.

 

Adjustment of the Load Cell Output............................................................

6

3.1.4.

 

Verifying Load Cell Output and Input Sensitivity..........................................

6

3.1.5.

 

Installing the Option Board..........................................................................

7

 

Description of Panels and symbols......................................... ...............................

8

4.1.1.

 

Front Panel Description...............................................................................

8

4.1.2.

 

Rear Panel Description....................................................................................

9

4.1.3.

 

Other Displays and Symbols............ ...........................................................

9

4.1.4.

 

Accessories and Option OP"02.....................................................................

10

 

Calibration.................................... ...................................................................

11

5.1.1.

 

Items of Calibration Mode.............................................................................

11

5.2.

 

Calibration Procedure....................................................................................

12

5.2.1.

 

The Whole Procedure.................................................................................

12

5.2.2.

 

Preparation.......................................... ..........................................................

13

5.2.3.

 

Entering the Calibration Mode.....................................................................

13

5.2.4.

 

Minimum Division (Weighing Interval)........................................................

14

5.2.5.

 

Maximum Capacity............. ......................................................................

14

5.2.6.

 

Zero Calibration.................................................................................... .

15

5.2.7.

 

Span Calibration............................................................................................

16

5.2.8.

 

Exiting the Calibration Mode........................................................................

17

5.3.

 

Weighing Range Function...............................................................................

18

5.3.1.

 

Setting Division and Range............................................. ...... ......................

19

5.4.

 

Digital Linearization Function.........................................................................

22

5.5.

 

Gravity Compensation Function....................................................................

23

5.5.1.

 

The Gravity Acceleration Table....................................................................

24

5.6.

 

Calibration Error Code List...............................................................................

25

6.

Functions............................... .............................................................................

26

6.1.

 

Changing the Function Settings........................................................................

26

6.2.

 

F“Functions....................................................................................................

27

6.3.

 

CF“Functions...................................................................................................

32

AD4329 Weighing Indicator, instruction Manual, A&D Co„ itd. Page 1

ш

I

I

7.

Tare.,...,.,.......... ........*....».......

:............

;.:.....;;....,.л...Л...

:,...» 33

;

7.1,1. Weighing Tare ......

;^

 

......

....................................33

 

7.1.2.

Digital Input.......... .....................................................................................

 

 

33

 

7.1.3.

ClearingTare...............................................................................................

 

 

33

8.

Accumulation....................

.................................................................................

 

 

34

 

8.1.1.

Preparation and Specification..................................................................

 

34

 

8.1.2.

Display and Operation.................................................................................

 

 

35

9.

Comparison.....................

....................................................................................

 

 

36

 

9.1.

Comparator Output Relay (OP-02)..........

.......................................................

36

 

9.2.

Upper/Lower Limit Comparison.......................................................................

 

36

 

9.2.1.

Selecting Upper/Lower Limit Comparison.................................................

37

 

9.2.2.

Condition formula for Comparison........................................................

37

 

9.2.3.

Setting Upper/Lower Limit Values.............................................................

38

 

9.2.4.

Example.....................................................................................................

 

 

 

.38

 

9.3.

Setpoint Comparison...................................................................................

 

 

39

 

9.3.1.

Selecting Setpoint Comparison...................................................................

 

39

 

9.3.2.

Description of Input parameters and Outputs............................................

39

 

9.3.3.

Setting the Parameters of Setpoint Comparison........................................

40

10.

External Control Input...................................

 

..................................................

41

11.

Standard Serial Data Output

 

(SER.OUT).............................................

42

 

11.1.1.

Data Format................................................................................................

 

 

43

12.

RS-232C Interface............................................................................................

 

 

 

44

 

12.1.1.

Specification.............................................................................................

 

 

44

 

12.1.2.

Data Format................................................................................................

 

 

45

 

12.1.3.

Command Format....................................................................................

 

 

46

 

12.1.4.

Command to Request Data.....................................................................

 

46

 

12.1.5.

Commands to Control the Indicator..................... .....................................

47

 

12.1.6.

Commands to Set Parameters.............................................................

 

........48

 

12.1.7.

Related Functions....................................................................................

 

 

49

13.

Specifications...............

.....................................................................................

 

 

50

 

13.1.

Dimensions..................................................................................................

 

 

 

51

Page 2

AD-4329 Weighing indicator

Compliance

1.1.1.Compliance with FCC Rules

Please note thaï this equipment

generates, uses and can radiate radio frequency

energy. This equipment has been

tested and has been found to comply with the

limits of a Class A computing device pursuant to Subpart J of Part 15 of FCC rules.

These rules are designed

to provide

reasonable

protection

against

interference

when this equipment

is

operated in a commercial environment. If

this unit

is operated

in a residential area

it

may

cause some

interference

and under

these circumstances

the user would be required to take, at his own expense, whatever measures are necessary to eliminate the interference.

(FCC = Federal Communications Commission in the U.S.A.)

1.1.2.Compliance with European Directive

C€ This appliance complies with the

statutory

EMC (Electromagnetic

Compatibility)

directive 89/336/EEC and the Low

Voltage

Directive 73/23/EEC

for safety of

electrical equipment designed for certain voltages.

Note: The displayed value may be adversely affected under extreme electromagnetic influences.

AD-4329 Weighing indicator

Page 3

1. Compliance

2.Introduction

The AD-4329 is a compact weighing indicator that amplifies the signal from a load cell, converts it to digital data and displays it as a mass value.

a This indicator has the following performance:

Input sensitivity:.............................

0.2juV/division.

Maximum display:..........................

10000 divisions (or 20000 divisions).

Refresh rate of the display:..............

10 times/second approximately.

Input voltage range:........................

-1 mV ~+15 mV.

There are the following standard functions:

The upper/lower limit comparison to check a mass value.

The setpoint comparison for batching applications.

The accumulation function to totalize these mass values and to count the number of accumulations.

a There are the following standard interfaces: □ The standard serial output for a printer.

Q The RS-232C serial interface to communicate with a computer.

This interface can request weighing data, enter parameters and control the state of the indicator.

The calibration function includes the following functions:

Setting of the minimum division (weighing interval) and the maximum capacity.

Zero and span calibration.

a The weighing range function of the multi-interval weighing instrument (scale).

Digital linearization function.

Gravity compensation function.

There are seven external input terminals controlled by the F function,

o The option OP-02 has three output relays used for the upper/lower limit comparison and setpoint comparison.

2. Introduction

Page 4

AD-4329 Weighing Indicator

3. Installation and Precautions

ЗЛ .1. Instailation and Precautions

Aa

AD

The AD-4329 weighing indicator is a precision electronic instrument Handle the Indicator carefully.

The operating temperature is -10t: to *f40t: (14“F to 104"^).

Do not install the scale in direct sunlight.

Misoperation or other problems may be caused by an unstable power source including momentary power failrue or instantaneous noise. Use a stable power source.

Do not connect the power cord before the installation Is finished.

Please confirm that the local voltage and receptacle type are correct for your scale. The jOperate/Standbvi key @ key) can not cut the power. The load cell and the internal circuit are supplied with power in the standby state.

Use shielded cable for ail connections. Connect the cable shields to the shield terminal or case as an earth terminal.

Earth ground the indicator. Do not Join the earth ground line with other electric power equipment. There is an earth ground terminal at the power cord receptacle.

Do not install the scale in a place where it is af3t to be charged with static electricity, or where the relative humidity is lower than 45%RH. Plastic and isolators are apt to be charged with static electricity.

3.1.2.The Load Cell Connections

Connect the load ceil wires to the connector (receptacle), at the rear panel, using the

accessory load cell plug.

It is possible to connect a 4 wire cable that 1 pin"2pin and 3pin-4pin are shorted, if the distance between the indicator and a load cell is shorter than 5m.

The output voltage of a load ceil is a very sensitive signal. Space the load cell cable away from any noise source.

It is possible to connect eight 350Q load cells.

The load cel! drive is 5VDC ± 5% between EXC+ and EXC-, the maximum current 120mA.

EXC+

Cable

Positive Sense

 

Cable

i\

 

 

 

 

Positive Excitation Input

Load Cell Plug

EXC-

 

Negative Excitation input

 

Negative Sense

 

 

SIG-

 

Negative signal Output

 

 

S1G+

 

Positive signai Outpur

 

 

Load ceii

 

Standard

Available connection for

 

 

 

 

connection

a cable shorter than 5m.

AD-4329 Weighing Indicator

 

Page 5

 

3. Installation and Precaution

3.1.3. Adjustment of the Load Ceil Output

Caution a Use a inetaf film resistor in the range of 50kQ to 500kQ with a good temperature coefficient when adding a resistor to adjust a load cell output. Use as a large resistance value as possible in the range in which zero adjustment is possible. Solder this resistor at a point near the load cell or the indicator.

In Case of Reducing the Output Voltage

When the zero output is too large, add a resistor between EXC+and SIG-.

in Case of Adding an Offset Voltage to the Output When the zero output is too small, add a resistor between EXC+ and S1G+.

3.1.4. Verifying Load Ceil Output and Input Sensitivity

The input sensitivity of the indicator is 0.2pV/division or more. Adapt to the following inequality, when you design a weighing instrument using the indicator and load cell(s).

Caution □ A change in input voltage sensitivity is equivalent to a one division

change of the display. Select as large an Input voltage sensitivity voltage as possible so that the weighing interval becomes stable,

a Consider the leverage If a lever is used.

Weighing

instrument

E^tB^D

using one load ceil.

0.2 £------------

A

, Weighing

Instrument

E^BH^D

using multi-load cell

0.2^——~

A^N

Verification Example

A:Rated capacity of load cell [kg]

B:Rated output [mVA/]

D:Weighing interval [kg]

E:Excitation voltage [mV]

N:Number of load cells

 

Design;

 

 

 

 

Load cell

N=1

6000*3*0.05 , ^

 

 

Rated capacity

A=750 [kg]

„ -j ^ 0.2. Therefore,

 

Rated output

8=3 [mVA/I

750

 

 

 

Excitation voltage

E=5000 [mV]

regard the instrument as a good design.

 

Weighing interval

D=0.05 [kg]

 

 

 

Weighing capacity

300 [kg]

 

 

3. Insteilatìon and Precaution

Page 6

AD-4329 Weighing Indicator

3,1,5, Installing the Option Board

Caution □ Do not remove screws without the following steps.

This is a procedure for the relay output board (OP-02).

Step 1 Remove the power cord and other cables from the indicator.

Step

2

Remove four screws and four boits from the rear panel.

Step

3

Remove two slide-rails, one on each sides.

Step 4

Pull the rear panel and main board from the case. Then the connector between the

 

 

main board and the display board may be disconnected smoothly.

Step

5

Remove two screws and the small blank panel from the rear panel.

Step

6

Attach the option board using two screws.

Step

7

Connect the option cable connector to J5 on the main board.

Step 8 Insert the main board into the both main board guides and insert it as before. Then the connector between the main board and the display board Is re-connected, if the rear panel will not close completely, retry step 8 to insert the main board correctly.

Step 9 Attach both slide-rails. Fix the rear panel using four screws and four bolts as before.

AD-4329 Weighing indicator

Page 7

3. Installation and Precaution

Panels and symbols

4.1.1.Front Panel Description

No.

Name •

Description

1

Standby indicator

The indicator lights at the standby state. This state means

connecting the power cord and turning the display off.

 

 

2

Data display

Weighing data Is displayed.

3

Weighing unit

A unit selected in the CF functions.

 

 

The ^ marks indicate kind of data and weighing state.

 

 

M-f- It lights when there is accumuíation data.

 

 

it is blinking when displaying accumulation data.

 

 

MD The motion detection mark.

4

State indicator

It lights when motion is detected.

It turns off while data is stabilized.

 

 

 

 

GROSS It lights when displaying gross data.

 

 

NET It lights when displaying net data.

 

 

PT It lights while entering a preset tare parameter.

 

 

ZERO it lights when displaying zero.

5

CAL cover

Calibration protection cover.

6

CAL key

Calibration key. The key to adjust the scale so that mass

value is displayed correctly and to enter the CF functions.

 

 

7

Operate/Standby key

The key to turn the display on or off.

8

Preset tare key

The key to enter preset tare setting mode.

9

M+ (Accumulation) key

The key for the accumulation function.

 

 

Refer to "8. Accumulation", CF11, F4 and F5.

10

Mode key

The key for the accumuiation function and calibration.

The key to select a unit of lb or kg.

 

 

11

Net/Gross key

The key to select net or gross.

+/" key

The key to set the polarity of data while in the setting mode.

 

12

Zero key

The key to zero a current display.

>key

The key to select a figure while in the setting mode.

 

13

Tare key

The key to perform tare.

A key

The key to select a value while in the setting mode.

 

14

Print key

The key to output data.

Enter key, key

The key to store current data while in the setting mode.

 

4. Description of Panels and symbols

Page 8

AD-4329 Weighing Indicator

4.1.2. Rear Panel Description

No.

Name

Description

15

Power connector

Use accessory power cord.

Please Confirm that the iocal voltage and receptacle type are

 

 

correct for your scale.

16

Grounding terminal

 

17

RS-232C interface

Refer to "12. RS-232C Interface" for the detail.

18

Blank panel

Option space (for relay output).

19

Fuse holder

Use 0.5A time lag fuse for AC1OOV to AC120V.

Use 0.316A time lag fuse for AC200V to AC240V.

 

 

20

External input

Seven control inputs selected function by the F functions.

Refer to FI 0 to 16.

 

 

21

Serial output

Printer port. Refer to "11. Standard Serial Output" for the

details.

 

 

22

Load cell connector

Connect accessory load cell plug.

4.1.3. Other Displays and Symbols

standby display.

Zero error when turning display on. If the I MODE I key is pressed, a current weighing value may be displayed.

Blank

Decimai point

r r

Over load display. Remove any load from the load cel! immediately. It may cause damage to the indicator.

Example of an error display.

AD-4329 Weighing Indicator

Page 9

4. Description of Panels and symbols

4.1.4. Accessories and Option OP“02

 

Instruction manual

1

 

 

Load cell piuq

1

JM-NJC-207-PF

 

Power cord

(1)

 

Accessories

7pin Din connector

1

JA-TCP0576

Spin Din connector

1

JA-TCP0586

 

 

0.5A or 0.315A time lag fuse

1

FS-EAWK-500MA

 

 

 

FS-EAWK-315MA

 

Rubber foot

4

10-SJ-5023

Caution Please Confirm that the receptacle type and focal voltage is correct for

your indicator (scale).

Comparator relay output

OP-02

AD4329-02

 

Accessory

Cable

1

4. Description of Panels and symbols

Page 10

AD-4329 Weighing Indicator

5. Calibration

This weighing Indicator, converts an input voltage from a load cel! to the "mass*' value, and displays it. Calibration is the adjustment function so that the scale (indicator) can weigh correctly.

5:1,1, items of Calibration Mode_______________________ _____

There are seven items In the calibration function.

Basic Items

 

The minimum division....................

Setting the weighing interval.

The maximum capacity..................

Setting the maximum display.

Zero calibration... .........................

When unloading any weight on the weighing unit

 

( there is nothing on the weighing pan), this function

 

performs adjustment so as to display the zero point

 

mark. Select an adjustment method from the.

 

"weighing input" or "digital input".

Comment This function Is the fundamental starting point to

 

weigh anything, and influences the performance of

 

scale (indicator).

Span calibration..................... ......

The function to measure an input voltage variation

 

by loading to the weighing unit correctly. Select an

 

adjustment method from the "weighing arbitrary

 

mass", "weighing capacity mass" or "digital input".

Optional Items (Sub-functions)

 

Range function...... ....................

The function for a multi-interval scale. Select "dual

 

range scale" or "triple range scale". Each range has

 

parameters of "weighing range" and "division"

pigital linearization function...........

The function to revise a linearity deviation using

 

weighing points, (up to three weighing points)

Gravity compensation function... The function to revise weighing error between the calibration location and another weighing location using gravity acceleration.

Caution □ The maximum display is less than or equal to 10000 divisions. This number

is calculated from the maximum capacity divided by the minimum division.

Check the accuracy of weighing instrument periodically.

Recommended mass, use a mass heavier than 2/3 maximum capacity.

AD-4329 Weighing Indicator

Page 11

5. Calibration

□ Calibrate the scale, if It is moved to another location or the environment

'has changed.

It Is not necessary to set the gravity acceleration correction, when calibrating the scale with a calibration mass at the place where the scale is used.

Enter the stable weighing data while the MD mark is turned off. If unstable data Is used, it may cause a weighing error. Arrange the condition using the FOO filter function.

The span calibration needs the zero calibration data. We recommend that you perform the span calibration Immediately after the zero calibration.

Perform the digital linearization function immediately after the zero calibration. And perform the span calibration continuously.

Select a decimal point and weighing unit at the CF function CFOO, CF 01.

If you use, the dual or triple range function of the multi-interval scale, perform the “Range Function", “Zero Calibration" and "Span Calibration".

5.2.Calibration Procedure

5.2.1.The Whole Procedure

Essential Items

Step A Preparation for calibration.

Step B Minimum division.

Step C Maximum capacity.

Step B Zero calibration.

Step G Span calibration

Optional Items

Step D Weighing range function, i Perform this function just

before zero calibration and span calibration.

Step F Digital linearization function. Perform this function after the zero calibration immediately. And perform the span calibration continuously.

Step H Gravity compensation function.

Perform this function after span calibration.

5. Calibration

Page 12

AD-4329 Weighing Indicator

A&D AD-4329 User Manual
Turn the display on for 10 miniuies
CAL

5.2.2.Preparation

step 1 Keep the following conditions to calibrate the scale (indicator) correctly.

Maintain a constant temperature, stable power and stable input voltage from the load ceil.

Avoid direct sunshine and in the neighborhood of an air conditioner.

Do not install the scale (indicator) where there is a strong magnetic field.

Step 2 Turn the display on during 10 minutes.

5.2.3.Entering the Calibration Mode

Step 3 Press the ICAlj key In the norma! weighing mode. The CHL (O is displayed for

two seconds when entering the calibration mode.

MODE! key.... Parameters are not revised and the next item is displayed.

CALI key........Parameters are stored in the indicator and Snd I is displayed. Press the jOperate/Standbvl key to turn the display off.

Press and hold the ¡NET/GROSSl key and press the |CAU key

....... .............All parameters are not changed and the calibration mode is.

finished. Press the jOperate/Standby key to turn the display off after displaying CBnCEL

m ,n

Zero Calibration^

^

Span Calibration.

5. Calibration

AD-4329 Weighing Indicator

Page 13

To Zero Calibration
AD-4329 Weighing Indicator
Maximum capacity

5.2.4.Minimum Division (Weighing Interval)

Step 4 Check the | d-xx Idisplay. The xx is minimum division (01,02, 05,10, 20,50).

Minimum division

Step 5 Select a minimum division using the following keys,

Akey... The key to select a parameter, key... The key to store a parameter and

proceed to step 6.

MODE! key..The key not to change a parameter and proceed to step 6.

M-H key ..... The key to proceed to "5.6. Gravity Compensation Function".

To Gravity

Compesation Function

To Maximum Capacity

5.2.5.Maximum Capacity

Step 6 Check the [ ixxxx | display. The xxxx is current capacity. Select a new maximum capacity using the foliowing keys.

> key......... . The key to select a figure.

Akey...........The key to select a number. key.......... The key to store a parameter

and proceed to step 7 (of the

 

zero calibration).

 

IMOD^ key..........The key not to change a

To Weighing Range

parameter and proceed to step 7 Punction

(of the zero calibration). PHESETTABg key,,..... The key to porceed to "5.3.

Weighing Range Function".

5. Calibration

Page 14

5.2.6.Zero Calibration

Step 7 Check the

CfiL 0 display.

 

 

 

Zero calibration

 

Select a zero calibration method to adjust the zero point

 

 

 

Weighing input

The adjustment method

 

To step 8

 

 

with nothing on the

 

 

 

(Normal way)

 

 

 

weighing unit.

 

 

 

 

 

 

 

 

 

 

 

 

Digital input

 

The numerical input to

 

To step 9

 

 

 

enter a load cell output

 

 

 

 

 

voltage.

 

 

 

 

Weighing Input

 

 

 

 

 

 

Step 8 Place nothing on the weighing unit. Press

the

key

 

after the MD mark has turned off. Proceed to steplO.

 

 

 

MOPg key....... The key not to change the zero point

 

 

data and proceed to step 10.

Caution Do not press the [ *-■ key while the MD mark is lit (detecting motion). Arrange the condition using the FOO filter function.

Digital Input

Step 9 Pressing the @ key, a stored input voltage parameter

of the zero point is displayed in the unit of mV/V. Adjust the input voltage using the following keys.

key.....The key to select a figure, key.....The key to select a number.

key.....The key to store the zero point parameter and proceed to step 10.

MODE] key..The key not to change parameter and proceed to step 10.

To Span Calibration

AD-4329 Weighing Indicator

Page 15

5. Calibration

Loading...
+ 39 hidden pages