Zeiss Stemi 305 Operating Manual

Page 1
Operating Manual
Stereo microscope Stemi 305
Page 2
Page 3
You have acquired a product made by ZEISS. Please read this operating manual before using your instrument for the first time to make sure that its quality is maintained and you can utilize your instrument reliably for a long time.
Page 4
COPYRIGHT Stereo microscope ZEISS Stemi 305
We reserve the right to perform modifications in view of technical progress; this operating manual is not subject to revision.
© Unless expressly authorized, forwarding and reproduction of this document as well as utilization
and communication of its contents are not permitted. Violations will entail an obligation to pay compensation.
All rights reserved in the event of granting of patents or registration of a utility model.
Issued by: Carl Zeiss Microscopy GmbH
Carl-Zeiss-Promenade 10 D-07745 Jena, Germany
microscopy@zeiss.com www.zeiss.com/microscopy
Carl Zeiss Microscopy GmbH Königsallee 9-21 D-37081 Göttingen, Germany
SAP no.: 435063-7044-001
Date of issue (Version 6): 01/28/2015
4 435063-7044-001 01/2015
Page 5
Stereo microscope CONTENTS Stemi 305 ZEISS
CONTENTS
Page
1 INTRODUCTION ........................................................................................................... 7
1.1 General information ....................................................................................................... 7
1.2 Notes on instrument safety ............................................................................................ 8
1.3 Warranty notes ............................................................................................................ 12
2 DESCRIPTION ............................................................................................................. 13
2.1 Indication for use ......................................................................................................... 13
2.2 Microscope system ....................................................................................................... 13
2.3 Microscope sets and fields of application ...................................................................... 14
2.4 Interfaces on the microscope body Stemi 305 ............................................................... 15
2.5 System overview .......................................................................................................... 16
2.6 Technical data.............................................................................................................. 18
3 INSTALLATION ........................................................................................................... 21
3.1 General information ..................................................................................................... 21
3.2 Installing the stereo microscope .................................................................................... 21
3.3 Mounting optional components ................................................................................... 23
3.3.1 Mounting additional reflected-light illuminator ............................................................. 23
3.4 Connecting stereo microscope to the power supply ...................................................... 26
3.5 Switching stereo microscope ON or OFF ....................................................................... 26
4 OPERATION ............................................................................................................... 27
4.1 Adjusting the stereo microscope ................................................................................... 27
4.2 Setting reflected-light illumination ................................................................................ 28
4.2.1 Integrated vertical illuminator ....................................................................................... 28
4.2.2 Spot illuminator ........................................................................................................... 29
4.2.3 Double spot illuminator ................................................................................................ 29
4.2.4 Ring illuminator ........................................................................................................... 30
4.2.5 Controller K LED .......................................................................................................... 31
4.3 Transmitted-light illuminators ....................................................................................... 32
4.3.1 Transmitted-light module in stand K EDU...................................................................... 32
4.3.2 Transmitted-light unit in stand K LAB ............................................................................ 33
5 CARE, MAINTENANCE AND SERVICE ........................................................................ 34
5.1 Care ............................................................................................................................ 34
5.2 Maintenance ............................................................................................................... 35
5.2.1 Replacing the power unit ............................................................................................. 35
5.2.2 Opening the stand cover plate, stand K EDU ................................................................. 36
5.2.3 Opening the stand cover plate, stand K LAB ................................................................. 37
5.2.4
Wearing parts .............................................................................................................. 38
5.3 Service ......................................................................................................................... 38
5.4 Product disposal ........................................................................................................... 39
01/2015 435063-7044-001 5
Page 6
CONTENTS Stereo microscope ZEISS Stemi 305
6 APPENDIX ................................................................................................................... 40
6.1 List of abbreviations ...................................................................................................... 40
6.2 Troubleshooting ........................................................................................................... 41
6.3 List of illustrations ......................................................................................................... 42
6.4 Index ............................................................................................................................ 43
6 435063-7044-001 01/2015
Page 7
Stereo microscope INTRODUCTION Stemi 305 General information ZEISS

1 INTRODUCTION

1.1 General information

The stereo microscope Stemi 305 has been designed, produced and tested in compliance with the standards DIN EN 61010-1 (IEC 61010-1) and IEC 61010-2-101 "Safety Requirements for Electrical Measuring, Control and Laboratory Equipment".
The device meets the requirements of the EC Directive 98/79/EC Annex 1 regarding ivd products and the EC RoHS Directive 2011/65/EC and carries the mark.
This operating manual contains information and warnings to be observed by the operator.
The devices must be disposed of in compliance with the WEEE Directive 2012/19/EC.
In this operating manual, the following warning and information symbols are used:
CAUTION This symbol indicates a possible hazard to the user.
CAUTION LED risk group 2 acc. to DIN EN 62471:2009 Optical radiation is emitted. Do not look into the beam. This may damage your eyes.
CAUTION: High-energy UV radiation!
Risk of damage to eyes and skin!
CAUTION
Hot surface!
CAUTION
Remove the power plug before any intervention in the instrument!
ATTENTION
This symbol indicates a possible hazard to the instrument or system.
NOTICE This symbol indicates information which deserves special attention.
01/2015 435063-7044-001 7
Page 8
are no longer effective, the device must be
taken out of service and secured against being used again unintentionally. Please contact a
000 as well as the
Do not operate the devices included in the scope of supplies in a potentially explosive atmosphere, in the presence of volatile anaesthetics or flammable solvents such as alcohol,
structed persons who have been informed about the
cation in
equipped with special equipment protecting them from corrosive,
light illuminators have been
energy light), never look
INTRODUCTION Stereo microscope ZEISS Notes on instrument safety Stemi 305

1.2 Notes on instrument safety

If it becomes apparent that the safety measures
ZEISS service agency or the Carl Zeiss Microscopy Service to have the instrument repaired.
The Stemi 305 Information regarding instrument safety and installation conditions 435063­6044-008 and the Stemi 305 short operating instructions 435063-8044­operating instructions regarding the light sources also must be observed.
gasoline or similar substances.
The devices may only be operated by in possible risks in conjunction with the use of microscopes and the fields of appli question. The microscope is a high-precision device that may be impaired in its performance or even destroyed if handled improperly.
Before commissioning the instrument, make sure that the existing power supply is suitable for the instrument.
Make sure that the mains plug is removed from the socket before the instrument is opened.
The devices are not potentially infectious, toxic and radioactive or other samples that may be hazardous to health. If you handle such samples, be sure to observe all legal requirements, in particular the relevant national accident prevention regulations.
The LED reflected-light illuminators and the LED transmitted­classified as belonging to LED risk group 2 acc. to DIN EN 62471:2009. Avoid looking directly into the LED light.
If the device is operated with an external cold light source (high­directly into the fibre optic output of the cold light source. Otherwise, there is a risk of dazzling and blindness.
Never cover the open light guide socket or the fibre optic output. There is a fire hazard!
In all cases avoid covering the open light guide socket or the fibre optic output with your hand or other parts of your body. There is a risk of burns!
8 435063-7044-001 01/2015
Page 9
Do not replace detachable mains cables by inadequately designed mains cables. Only the
The microscope and the controller K LED are equipped with a desktop power unit each
The desktop power unit meets the requirements of protection class II (with protective
Changing the lamps on the cold light source is only admissible in accordance with the manufacturer's operating manual. If the appropriate instructions are not complied with, there is
hold waste; dispose of them in
Samples must also be disposed of appropriately in accordance with the applicable statutory
305 incl. its original accessories may only be used for the
ating manual. The manufacturer cannot assume any liability
or single components. Changes to and maintenance on this device and on devices which are operated together with the microscope may only be performed by our service or by authorized persons. The
r damage due to unauthorized
Dirt and dust may impair the performance of the device. The device must therefore be
Considerable temperature fluctuations, direct exposure to sunlight and vibrations should be
lation louvers free and do not
Stereo microscope INTRODUCTION Stemi 305 Notes on instrument safety ZEISS
Safe disconnection from the power supply is ensured exclusively by removing the mains plug. The switch on the microscope only switches into standby mode.
specified power supply lines may be used.
allowing line voltages in the range from 100 to 240 V ±10 %, 50 / 60 Hz, without the need for changing the voltage setting on the device.
insulation). If its casing is damaged, put the power unit out of operation. The microscope may only be operated with the desktop power unit supplied with the instrument.
a risk of burns and explosion when changing the lamp.
Defective devices should not be disposed of with house compliance with the statutory provisions.
provisions and internal work instructions.
The stereo microscope Stemi applications described in this oper for other applications, including those of individual modules
manufacturer of the devices does not accept any liability fo interventions in the device. Moreover, such interventions invalidate all guarantee / warranty claims.
protected from such influences to the greatest possible extent and covered with the dust cover when not in use. Before covering the device, always make sure that it is definitely switched OFF.
avoided.
Blocking or covering ventilation louvers may result in heat accumulation which may damage the device and in extreme cases even cause fire. Always keep venti insert or allow any objects to fall into them. All electrical components and parts must be located at a distance of min. 15 cm from flammable objects and walls.
01/2015 435063-7044-001 9
Page 10
refer to the appropriate operating instructions incl. the safety provisions contained in them
For transport over an extended distance, the device must be dismantled partially and
305. Only the connection cables supplied
305 is
INTRODUCTION Stereo microscope ZEISS Notes on instrument safety Stemi 305
When operating a Stemi 305 cam, please refer to the appropriate operating instructions incl. the safety provisions contained in them before commissioning the device.
If it is intended to operate the Stemi 305 with an external fibre optical cold light source, please
before commissioning the light source.
The desktop power unit must not get into contact with moisture.
transported in its original packaging.
To transport the device over an extended distance, use the original packaging or the transport case Stemi 305/508.
To connect the LED illuminators of the Stemi 305 system, six-pin modular connectors are used which are known from telecommunications (type: RJ 12).
Only the LED illuminators pertaining to the Stemi 305 may be connected to the RJ 12 sockets in the microscope body and in the stands of the Stemi with the device may be used.
The use of commercially available cables with RJ 12 connectors or the connection of fax units, modems or other telecommunication equipment to the RJ 12 sockets of the Stemi prohibited.
10 435063-7044-001 01/2015
Page 11
Stereo microscope INTRODUCTION Stemi 305 Warranty notes ZEISS
Fig. 1 Warning and information labels on the device
1 LED aperture, transmitted-light illuminators
2 LED aperture, spot illuminator K LED 3 LED aperture, ring illuminator K LED, segmentable 4 LED aperture, double spot illuminator K LED 5 LED aperture, integrated vertical illuminator
Fig. 2 Warning labels and apertures for LED radiation
01/2015 435063-7044-001 11
Page 12
INTRODUCTION Stereo microscope ZEISS Warranty notes Stemi 305

1.3 Warranty notes

The manufacturer guarantees that the device is free from material or manufacturing defects when delivered. Any defects which may have occurred must be reported to us immediately and steps be taken to minimize damage. If notified of such a defect, the device manufacturer is obligated to rectify it at his discretion, either by repairing the instrument or by delivering an intact replacement. No guarantee is provided for defects caused by natural wear (wearing parts in particular) and improper use.
The instrument manufacturer shall not be liable for damage caused by faulty operation, negligence or any other tampering with the device, particularly the removal or replacement of instrument components, or the use of accessories from other manufacturers. Such actions will render all warranty claims invalid.
No maintenance or repair work may be performed on the microscopes which exceed the activities specified in this operating manual. Repair may only be performed by ZEISS Service or by persons specifically authorized by it. Should any malfunctions occur on the device, please first contact the Carl Zeiss Microscopy Service or, abroad, the ZEISS representative in your area.
12 435063-7044-001 01/2015
Page 13
existing!
Stereo microscope DESCRIPTION Stemi 305 Indication for use ZEISS

2 DESCRIPTION

2.1 Indication for use

The stereo microscopes Stemi 305 are microscopes for magnifying spatial observation of small objects. They have been designed and built for training purposes in schools, universities and natural science facilities. They are used, moreover, in biological and medical laboratories, in industrial manufacture and quality assurance.
Stemi 305 is envisaged for applications in biology and in medicine for the analysis of blood and/or tissue samples from the human body. Applications in the field of diagnostic medicine are explicitly excluded, except for the field of medical research.

2.2 Microscope system

1 Eyepiece in eyepiece tube (eyepiece 10x/23)
2 Microscope body (trinocular) with installed
vertical illuminator
3 Interface for cameras with C-mount connector 4 Zoom button to adjust the magnification 5 Handle 6 Stand (model K EDU) 7 Focusing drive to focus the specimen
8 Rotary/push-button for switching ON and OFF the illuminators or the additional reflected-light
illumination and for adjusting the illumination intensity
9 Rotary/push-button for switching ON and OFF the transmitted-light illumination and for adjusting the illumination intensity 10 Reflected-light illumination (spot illuminator K LED) 11 Lever for adjusting the transmitted-light illuminator – bright field or dark field on the stand K EDU 12 Insert plate for placing the specimen
Fig. 3 Microscope system Stemi 305
The illustrated microscope equipment shows an example and may differ from the ones actually
01/2015 435063-7044-001 13
Page 14
Fig. 6 Stemi 305 MAT microscope set
DESCRIPTION Stereo microscope ZEISS Microscope sets and fields of application Stemi 305

2.3 Microscope sets and fields of application

Stemi 305 EDU:
Stemi 305 body in stand K EDU
integrated controllable vertical illuminator
reflected-light spot illuminator with height and
angle adjustment, zoomable
flat transmitted-light unit for bright and dark
field
for training in tutorial rooms and training
Fig. 4 Stemi 305 EDU microscope set
institutes
Stemi 305 LAB:
Fig. 5 Stemi 305 LAB microscope set
Stemi 305 body in stand K LAB
integrated controllable vertical illuminator
double spot illuminator with height and angle
adjustment
reflected-light module for bright and dark field
and relief contrast
hand rest (optional)
for separation and documentation in the
laboratory
Stemi 305 MAT:
Stemi 305 ESD body in stand K MAT
integrated controllable vertical illuminator
segmentable ring illuminator for reflected light
for industrial quality assurance and manufacture
ESD-capable thanks to its antistatic surface
14 435063-7044-001 01/2015
Page 15
1
Optional mount for eyepiece plates
2
3
Integrated camera adapter 0.5x with
4
Internal thread M52x1.0 to accommodate
5
6
7
8
Guide for mounting a reflected-light spot
9
M8 thread to mount retaining arms (only
Stereo microscope DESCRIPTION Stemi 305 Interfaces on the microscope body Stemi 305 ZEISS

2.4 Interfaces on the microscope body Stemi 305

26 mm
Eyepiece tube ∅ 30 mm to replace the eyepieces
C-mount connector for cameras up to 2/3"(only for Stemi 305 trino)
front optics and other accessories
Mount ∅ 66 mm for ring illuminator
Mount ∅ 76 mm for microscope body
Mount ∅ 84 mm for tables, insert plates
illuminator
stand K)
Thanks to the interface ∅ 76 mm, the Stemi 305 can also be used in conjunction with other stands of the modular kit of the Stereo system – and other ZEISS stereo microscopes such as e.g. the Stemi 508, may also be inserted in the stands K.
Front optics, Stemi 305
External thread M52x1.0 for fastening in the
zoom body Stemi 305
Internal thread M49x0.75 to accommodate
optical filters and analyzer for pol.
Fig. 7 Interfaces on the Stemi 305
with stand K LED (schematic diagram)
01/2015 435063-7044-001 15
Fig. 8 Front optics, Stemi 305 (example)
Page 16
DESCRIPTION Stereo microscope ZEISS System overview Stemi 305

2.5 System overview

16 435063-7044-001 01/2015
Page 17
Stereo microscope DESCRIPTION Stemi 305 System overview ZEISS
01/2015 435063-7044-001 17
Page 18
Microscope system Stemi 305 EDU
Weight
Microscope system Stemi 305 LAB
Weight
DESCRIPTION Stereo microscope ZEISS Technical data Stemi 305

2.6 Technical data

Dimensions
Stemi 305 in stand K EDU ................................. 4.6 kg
Dimensions
Stemi 305 in stand K LAB ................................. 6.2 kg
18 435063-7044-001 01/2015
Page 19
Stereo microscope DESCRIPTION Stemi 305 Technical data ZEISS
Ambient conditions
Storage (in packaging)
Admissible ambient temperature ...................................................................................... +10 to +40 °C
Admissible relative humidity ........................................................ max. 75 % at +35 °C (non-condensing)
Transport (in packaging)
Admissible ambient temperature ....................................................................................... -40 to +70 °C
Operation
Admissible ambient temperature ...................................................................................... +10 to +40 °C
Admissible relative humidity .................................................................................................. max. 75 %
Atmospheric pressure .............................................................................................800 hPa to 1060 hPa
Pollution degree ................................................................................................................................... 2
Field of application ..............................................................................................................closed rooms
Height of field of application ............................................................................................... max. 2000 m
Operating data – desktop power units, microscope and controller K LED
Safety Class .......................................................................................................................................... II
Type of enclosure ........................................................................................................................... IP 20
Electrical safety .............................................................................. acc. to DIN EN 61010-1 (IEC 61010-1)
taking CSA and UL regulations into account
Pollution degree ................................................................................................................................... 2
Overvoltage category ............................................................................................................................ 2
Line voltage ............................................................................................................. 100 to 240 V ±10 %
Due to the fact that the instrument features a multi-voltage power supply unit, change-over of the device voltage not required!
Line frequency .................................................................................................................. 50 Hz 60 Hz
Power consumption: Desktop power unit with connected microscope ..................................... max. 40 VA
Output desktop power unit for microscope
and for Controller K LED ............................................................................................ 12 V DC, max. 2 A
Input desktop power unit for microscope
and for Controller K LED .................................................................... 100-240 V, 50-60 Hz, max. 0.55 A
01/2015 435063-7044-001 19
Page 20
DESCRIPTION Stereo microscope ZEISS Technical data Stemi 305
Optical risk group classification acc. to DIN EN 62471:2009
Overall device ....................................................................... LED risk group 2 acc. to DIN EN 62471:2009
Integrated vertical illumination .............................................. LED risk group 2 acc. to DIN EN 62471:2009
Spot illuminator K LED (reflected illumination) ....................... LED risk group 2 acc. to DIN EN 62471:2009
Double spot illuminator K LED (reflected illumination) ............ LED risk group 2 acc. to DIN EN 62471:2009
Transmitted-light unit in stand K LAB .................................... LED risk group 2 acc. to DIN EN 62471:2009
Transmitted-light unit in stand K EDU ................................... LED risk group 2 acc. to DIN EN 62471:2009
LED reflected illumination, peak at 460 nm, .......................... LED risk group 2 acc. to DIN EN 62471:2009
LED transmitted illumination, peak at 465 nm, ...................... LED risk group 2 acc. to DIN EN 62471:2009
20 435063-7044-001 01/2015
Page 21
Notes on Instrument
For installation and/or transport of the microscope or the stand, exclusively use the handle
Stereo microscope INSTALLATION Stemi 305 General information ZEISS

3 INSTALLATION

3.1 General information

Before installing and commissioning the device, make sure to read the Safety carefully (see Section 1.2, Page 8).
The Stemi 305 with the necessary tools and optional equipment is delivered in several standard packages.
Take all units out of the packaging and check them for completeness according to the delivery note.
(Fig. 3/5) provided to this effect.
Remove transport securing devices (adhesive tapes etc.).
Keep original packaging for a possibly extended period of storage or return of the device to the
manufacturer or dispose of it appropriately.

3.2 Installing the stereo microscope

Place stand on a level, firm support.
If the Stemi 305 body (Fig. 10/2) has not yet
been mounted, insert it in the Stemi mount (Fig. 10/4), align it and secure it using the clamping screw (Fig. 10/6). The appropriate Allen key (Fig. 10/9) is located in the support on the back of the device.
If the two eyepieces (Fig. 10/1) have not yet
been mounted, insert them both to their stop in the eyepiece tube.
Remove the two protective caps from the
eyepieces (Fig. 9/1) and remove the protective cap (Fig. 9/2) from the objective by turning.
Fig. 9 Removing caps
01/2015 435063-7044-001 21
Page 22
tor has been described in
focusing drive can be set as required
case basis by adjusting
INSTALLATION Stereo microscope ZEISS Installing the stereo microscope Stemi 305
Insert the connection cable (Fig. 10/3) supplied
with the device into the socket on the Stemi body and in the socket on the Stemi mount.
Insert glass plate or B/W plastic plate (Fig. 10/7)
in the mount in the base of the stand. To this effect, press the glass plate or the B/W plastic plate against the spring (Fig. 10/8) and make it lock in.
Installation of an additional reflected­light illumina Section 3.3.1 on page 23.
Fig. 10 Installing the stereo microscope
If necessary, the ease of motion of the
on a case-by­the two focusing buttons (Fig. 11/1 and 2) in opposite directions.
The ease of motion must not be set so loosely that the drive moves down­wards by itself. This might damage the microscope or the specimen.
Fig. 11 Setting the ease of motion of the
focusing drive
22 435063-7044-001 01/2015
Page 23
Stereo microscope INSTALLATION Stemi 305 Mounting optional components ZEISS

3.3 Mounting optional components

3.3.1 Mounting additional reflected-light illuminator

Various reflected-light illuminators can be used, depending on the case of application.
For fastening, defined interfaces have been provided on the microscope.
3.3.1.1 Mounting spot illuminator K LED or
double spot illuminator K LED
Move the microscope body (Fig. 12/1) into its
uppermost position via the focusing knob (Fig. 12/2).
Insert the spot illuminator (Fig. 12/7) in the
stand by sliding the holder (Fig. 12/5) into the stand guide (Fig. 12/4) from below and push it upwards into the required position.
Plug the connection cable (Fig. 12/6) of the spot
illuminator into the lower socket of the Stemi mount (Fig. 12/3).
The double spot illuminator is mounted in an analogical manner.
Fig. 12 Mounting spot illuminator
01/2015 435063-7044-001 23
Page 24
The best clamping effect is achieved if
) of the adapter and
) of the
INSTALLATION Stereo microscope ZEISS Mounting optional components Stemi 305
3.3.1.2 Mounting ring illuminator K LED (segmentable)
Mounting ring illuminator to microscope body
Insert the connection cable (Fig. 13/1) of the
ring illuminator (Fig. 13/4) into the lower socket (Fig. 13/2) of the Stemi mount and into the socket of the ring illuminator.
Push ring illuminator onto the microscope body
(Fig. 13/3) from below, and secure using the knurled screw (Fig. 13/5).
Fig. 13 Mounting ring illuminator on
microscope body
Mounting ring illuminator to front optics 3
Put the slotted adapter (Fig. 14/4) supplied with
the device (OD 66 mm, ID 58 mm) onto the front optics (Fig. 14/1) from below. The circumferential bead (Fig. 14/3) inside the adapter must lock into the outer groove (Fig. 14/2) of the front optics.
Push ring illuminator (Fig. 14/6) onto the AD of
the adapter and secure it by means of clamping screw (Fig. 14/7).
the slot (Fig. 14/5 the clamping screw (Fig. 14/7 illuminator are offset by approx. 90°.
Fig. 14 Mounting ring illuminator to front
optics 3
24 435063-7044-001 01/2015
Page 25
To this effect, refer also to the
ating instructions
s are inserted by the
customer, it must be ensured that the
Stereo microscope INSTALLATION Stemi 305 Mounting optional components ZEISS
3.3.1.3 External fiber-optic illumination
Switch on external fiber-optic illumination
(Fig. 15/1), adjust illumination intensity and set illumination by bending the two goose necks (Fig. 15/2) so that the specimen is optimally lit.
separate oper regarding the cold light source.
3.3.1.4 Placing reversible eyepiece cup onto the eyepiece
The eyepieces feature rubber protection rings for spectacles to avoid the spectacles being scratched. These rings may be replaced at choice by reversible eye cups (Fig. 16/1).
Fig. 15 Mounting external fiber-optic
illumination
To this effect remove the protection rings for spectacles (Fig. 16/2) from the eyepieces and position
the eye cups (Fig. 16/1).
3.3.1.5 Insert the eyepiece plate into the eyepiece.
The adjustable eyepieces are intended for use with eyepiece plates.
Unscrew stop (Fig. 16/6) out of the eyepiece
(Fig. 16/3).
Pull locking ring (Fig. 16/4) out.
Insert eyepiece plate (Fig. 16/5).
Insert locking ring and re-fasten stop.
If eyepiece plate
writing is mirror-inverted before inser­tion into the eyepiece, and true to side after insertion.
Fine cotton gloves should be worn to insert or replace eyepiece plates.
01/2015 435063-7044-001 25
Fig. 16 Inserting the eyepiece plate
Page 26
Safe disconnection from the power supply is ensured exclusively by removing the mains plug.
Do not replace detachable mains cables by inadequately designed mains cables. Only the
INSTALLATION Stereo microscope ZEISS Connecting stereo microscope to the power supply Stemi 305
The slight image offset which is due to the additional glass distance is taken into account on the diopter scale by the zero position not being indicated by the white dot (Fig. 16/W), but by a red dot (Fig. 16/R).
Eyepieces with pre-inserted eyepiece plates are available directly from ZEISS.

3.4 Connecting stereo microscope to the power supply

Insert the mains cable (Fig. 17/2) in the mains
supply socket (Fig. 17/1) of the stand.
Connect the mains cable (Fig. 17/2) to a mains
supply socket outlet.
Fig. 17 Connecting the stereo microscope
The switch on the stereo microscope only switches into standby mode.
specified power supply lines may be used.

3.5 Switching stereo microscope ON or OFF

Use the power switch (Fig. 17/3) to switch the
stereo microscope ON or OFF.
26 435063-7044-001 01/2015
Page 27
The functionality of the knob for the
light illuminator has been
"0" on white dot (without eyepiece plate). "0" on red dot (with eyepiece
Check that the eyepieces are inserted
To this effect, a distance of approx.
between the
Stereo microscope OPERATION Stemi 305 Adjusting the stereo microscope ZEISS

4 OPERATION

4.1 Adjusting the stereo microscope

The stereo microscope has been connected and switched ON.
1. Place specimen (Fig. 18/7) centrally on the glass or plastic plate (Fig. 18/6) and illuminate it (Fig. 18/5).
reflected­described in Section 4.2 on page 28.
2. Set diopter compensation on the adjustable eyepieces (Fig. 18/1) to "0".
plate), see Fig. 19.
into the tube to their stop.
3. Set individual eye distance by shifting the
Fig. 18 Adjusting the stereo microscope
eyepiece tube laterally (Fig. 18/2), until only one complete light circle (specimen field) is
visible when looking into the eyepieces with both eyes.
2 cm must be maintained eye and the eyepiece.
4. Initially, set the zoom (Fig. 18/3) to the smallest magnification.
5. Focus on a small distinct detail in the center of the specimen (Fig. 18/4).
Fig. 19 Adjusting the eyepieces
6. Adjust maximum zoom value (Fig. 18/3). Magnification will probably make the distinct detail appear
blurred and no longer in the image center.
7. Sharpen the specimen by focusing, then shift the specimen to re-search for the detail and focus "sharply" again (Fig. 18/4).
8. Afterwards, reset to the lowest zoom value and correct any image blurring for each eye separately by diopter compensation on the adjustable eyepieces (not on the focusing knob, Fig. 18/4) (compensation of visual defect).
After adjustment of the stereo microscope, the image focus remains unchanged over the entire zoom range.
After change of the observer, repeat steps 3. to 8.
01/2015 435063-7044-001 27
Page 28
light illuminators, the intensity of the additionally installed
spot illuminator, ring illuminator, etc.) is
OPERATION Stereo microscope ZEISS Setting reflected-light illumination Stemi 305

4.2 Setting reflected-light illumination

4.2.1 Integrated vertical illuminator

Each Stemi 305 body features a vertical illuminator of its own.
If no second reflected-light illumination has been installed, the vertical illuminator is operated as described below:
Press the knob (Fig. 20/1) once to switch the
vertical illuminator (Fig. 20/2) alternately ON or OFF.
Turning of the knob serves to adjust the
illumination intensity.
If a second reflected-light illumination has been installed, the reflected-light illuminators are operated as described below:
Fig. 20 Setting the vertical illuminator
When mixing both reflected­reflected-light illuminator (spot illuminator, double­adopted. In this case, only the vertical illumination is adjusted by turning the knob. Now, the required intensity of the vertical light is blended in with the reflected-light illuminator.
Press the knob successively to set four
illumination states:
1. Press: to switch the vertical illumination ON.
2. Press: to switch the second reflected-light illuminator ON (vertical illumination OFF).
3. Press: to switch both reflected-light illuminators ON.
4. Press: to switch both reflected-light illuminators OFF.
By turning the knob, the illumination intensity
of the switched-ON reflected-light illuminator can be adjusted.
28 435063-7044-001 01/2015
Page 29
In the lower positions, an oblique light
For switching the illumination intensity ON and OFF and adjusting the
For switching the illumination intensity ON and OFF and adjusting the
Here again, an oblique light effect can be produced in the lower positions with
The spot illuminator and the double spot illuminator have a thread M24 for placing
By turning the spot polarizers, overexposure reflexes on the specimen can be reduced.
Stereo microscope OPERATION Stemi 305 Setting reflected-light illumination ZEISS

4.2.2 Spot illuminator

Push spot illuminator in the guide to the re-
quired height (Fig. 21/2).
effect is created to produce cast sha­dows for enhancing surface structures.
Set the angle (Fig. 21/1) via the pivot of the
spot illuminator so that the object is optimally illuminated.
Set size and brightness of the spot by axially
shifting the focusing unit (Fig. 21/3) (illumination zoom).
illumination intensity refer to Section
4.2.1, page 28.
Fig. 21 Setting the spot illuminator

4.2.3 Double spot illuminator

Move the microscope body to its uppermost
position via the focusing knob (Fig. 22/1). Push double spot illuminator (Fig. 22/2) to the required level within the guide.
Bend the two goose necks (Fig. 22/3) to adjust
the spots so that the specimen is optimally illuminated.
illumination intensity refer to Section
4.2.1, page 28.
the goose neck set horizontally.
Fig. 22 Double spot illuminator
polarization filters in front. The appropriate analyzer is screw-fastened in the M52x1.0 thread of the microscope body (Fig. 7/4).
01/2015 435063-7044-001 29
Page 30
1. Press: full circle
OPERATION Stereo microscope ZEISS Setting reflected-light illumination Stemi 305

4.2.4 Ring illuminator

By pressing the front button (Fig. 23/2)
successively, the user can change over between four illumination states of the ring illuminator (Fig. 23/3):
2. Press: semi-circle
3. Press: quadrant
4. Press: two quadrants (opposite each other)
When switching the semi-circle or quadrant
illumination ON, the specimen is first illuminated from the rear (from the direction of the stand column).
By pressing the rear button once briefly
(Fig. 23/1), the adjusted segment is turned
Fig. 23 Ring illuminator
forward clockwise by one quarter turn in each case.
By pressing and keeping the rear button
(Fig. 23/1) pressed for over 2 seconds, the adjusted segment is set into continuous rotation (rotation steps of 1/8 circle, 1 second per turn in each case).
By pressing the rear button once more (Fig. 23/1), rotation is stopped and the specimen is backlit
again.
For switching the illumination intensity ON and OFF and adjusting the illumination intensity refer to Section 4.2.1, page 28.
30 435063-7044-001 01/2015
Page 31
Stereo microscope OPERATION Stemi 305 Setting reflected-light illumination ZEISS

4.2.5 Controller K LED

The controller K (Fig. 24/4) is used to activate the vertical illumination of the Stemi 305 body or the ring illuminator in case of applications with boom stands A or U or additional stands.
Mount the stands A and U including the
fastening of the Stemi 305 body to the Stemi mount as described in the operating instructions for the boom stands.
Screw-fasten front lens system 0.5x/185 mm
(Fig. 24/9) to the microscope body.
When using the ring illuminator, adapt the
latter to the front lens system (Fig. 24/9) using the adapter supplied with the device (ID 58 mm, OD 66 mm).
Secure the adhesive side of the magnetic plate
(Fig. 24/7) to the swivel arm stand.
Plug the connection cable supplied with the
device (1 m long) (Fig. 24/2) into the socket (Fig. 24/8) on the microscope body or the ring illuminator and into the socket (Fig. 24/3) on the back of the Controller K (Fig. 24/4).
Fig. 24 Controller K LED
Place Controller K (Fig. 24/4) next to the
magnetic plate (Fig. 24/7).
Connect Controller K to the power supply using the power unit (Fig. 24/1).
Align swivel arm stand (Fig. 24/6) to enable an ergonomic working position.
Turn the knob (Fig. 24/5) on the Controller K to switch the vertical illuminator (or the ring illuminator)
ON and adjust illumination intensity by turning the knob.
Place specimen in the illuminated specimen field.
Set the stereo microscope to its initial setting, see Section 4.1 on page 27.
01/2015 435063-7044-001 31
Page 32
Press the upper knob (Fig. 25/1) to blend
OPERATION Stereo microscope ZEISS Transmitted-light illuminators Stemi 305

4.3 Transmitted-light illuminators

4.3.1 Transmitted-light module in stand K EDU

Switch transmitted-light illumination ON by
pressing the lower knob (Fig. 25/2).
With the transmitted light ON, turn this knob to
adjust the illumination intensity.
Change over by adjusting the lever (Fig. 25/3)
between the bright field and the dark field illumination.
Front position: Dark field (DF)
Rear position: Bright field (BF)
in reflected light (see Section 4.2 on page 28).
Fig. 25 Transmitted-light module in
stand K EDU
32 435063-7044-001 01/2015
Page 33
Clear mirror for high-contrast bright
By pushing the sliding knob (Fig. 26/2) into a central position and by tilting the mirror, oblique light effects (RC) are created which are very well suited to
Stereo microscope OPERATION Stemi 305 Transmitted-light illuminators ZEISS

4.3.2 Transmitted-light unit in stand K LAB

Switch transmitted-light illumination ON by
pressing the lower knob (Fig. 26/1).
With the transmitted light ON, turn this knob to
adjust the illumination intensity.
Adjust the bright field (BF) by pushing the
sliding knob (Fig. 26/2) into its rear position (BF).
Turn the sliding knob (Fig. 26/2) to swing the
clear or diffuse mirror into position or to adjust the inclination angle.
field, diffuse mirror for homogenous bright field.
Change over to dark field illumination (DF) by
pushing the sliding knob (Fig. 26/2) into its front position.
Turn the mirror until the specimen structures
are brightly lit against the dark background.
Fig. 26 Transmitted-light unit in stand K LAB
display non-coloured specimens.
01/2015 435063-7044-001 33
Page 34
The devices are not equipped with special equipment protecting them from corrosive,
ments, in particular the relevant
rom the mains power before cleaning. Make sure, that no cleaning
CARE, MAINTENANCE AND SERVICE Stereo microscope ZEISS Care Stemi 305

5 CARE, MAINTENANCE AND SERVICE

5.1 Care

Care of the devices is restricted to the following operations:
potentially infectious, toxic and radioactive or other samples that may be hazardous to health. If you handle such samples, be sure to observe all legal require national accident prevention regulations.
Remove contamination in accordance with the national accident prevention regulations.
Disconnect the devices from the mains power after use. Protect the devices from dust and moisture by
using a suitable cover (protective device cover).
Never expose the devices to inadmissible climate conditions (high humidity and temperature).
Disconnect the devices f fluids penetrate into the device.
Tenacious impurities on glass surfaces such as fingerprints and traces of grease, are best removed using a cotton wool bud wound around a round wooden stick and very little distilled water or non-aggressive solvent:
Distilled water: Clean glass surface with the slightly moistened cotton wool bud from the center to the edge, using circular movements.
Cleaning solution for optics, consisting of 15 % isopropanol and 85 % surgical spirit (gasoline): Clean glass surface by the slightly moistened cotton wool bud from the center to the edge, using circular movements.
Remove dust from optical surfaces using a brush with natural bristles or blow using rubber bellows (air blower).
Clean plastic parts using commercially available cleaning products (no solvents!). Tenacious impurities can be treated carefully with benzine or white spirit.
All labels on the components and on the desktop power unit may only be cleaned using a dry cotton cloth.
34 435063-7044-001 01/2015
Page 35
Only the desktop power unit specified
plug is
removed from the socket before the
Stereo microscope CARE, MAINTENANCE AND SERVICE Stemi 305 Maintenance ZEISS

5.2 Maintenance

5.2.1 Replacing the power unit

by the manufacturer may be used.
Make sure that the mains
instrument is opened.
Remove the mains plug of the power cable
from the socket.
Remove the power cable from the back of the
device.
Use an Allen key to release the screw (Fig. 27/1)
on the underside of the housing cover.
Pull off housing cover (Fig. 27/2) slightly at the
bottom, then press it upwards.
Release the plug (Fig. 27/3) and remove the
faulty desktop power unit (Fig. 27/4).
Insert the new desktop power unit and connect
the plug.
Press the upper part of the housing cover into
place and push housing cover down.
Secure housing cover by means of the screw.
Insert the power cable on the back of the
device.
Insert the mains plug into a mains supply
socket.
Fig. 27 Changing the power unit
01/2015 435063-7044-001 35
Page 36
To clean the LED board, use a soft
CARE, MAINTENANCE AND SERVICE Stereo microscope ZEISS Maintenance Stemi 305

5.2.2 Opening the stand cover plate, stand K EDU

The stand cover plate (Fig. 28/1) may be removed for cleaning or to remove any foreign matter which might fallen in inadvertently.
To this effect, release the three recessed head
screws (Fig. 28/2), lift off the stand cover plate (Fig. 28/1) carefully and then put it aside.
Make sure that the cable is not disconnected
from the stand cover plate (Fig. 28/1).
lintfree cloth, a cotton wool bud wound around a round wooden stick or a brush with natural bristles, possibly moistened slightly in distilled water.
Fig. 28 Opening the stand cover plate,
stand K EDU
36 435063-7044-001 01/2015
Page 37
Make sure that the cable (Fig. 29/4) is
To clean the LED, the optical elements
free cloth, a cotton wool bud wound around a round wooden stick or a natural hair brush, possibly moistened slightly in
Stereo microscope CARE, MAINTENANCE AND SERVICE Stemi 305 Maintenance ZEISS

5.2.3 Opening the stand cover plate, stand K LAB

The transmitted-light base (Fig. 29/1) can be removed from the stand cover plate (Fig. 29/3) for cleaning or for removal of any foreign matter which might have fallen in inadvertently.
Release the three screws (Fig. 29/2) from the
stand cover plate (Fig. 29/1) using an Allen key SW 2.5 and put them aside.
Carefully lift the stand cover plate (Fig. 29/3) off
the transmitted-light base (Fig. 29/1) and put the stand aside.
not disconnected from the transmitted­light base (Fig. 29/1).
and the mirror, use a soft lint
distilled water.
01/2015 435063-7044-001 37
Fig. 29 Opening the stand cover plate,
stand K LAB
Page 38
CARE, MAINTENANCE AND SERVICE Stereo microscope ZEISS Service Stemi 305

5.2.4 Wearing parts

The following wearing parts can be ordered directly from ZEISS:
Description Order No. Remark
Transparent glass plate, d = 84 mm 435425-9310-000
S/W plastic plate, d = 84 mm 435425-9320-000
Eyepiece PL 16x/16 Br foc. 444054-9000-000
Eyepiece cup 444801-0000-000 required 2x
Eyepiece 16x/14 Br foc. 435063-9160-000
Eyepiece W 25x/10 foc. 455046-9901-000
Set of small components consisting of:
- 2x spiral cables RJ 12
- 1x cable RJ 12, length 1 m
- 2x caps, eyepiece
- 1x cap, microscope body
- 1x Allen key, SW 3
- 1x M6 clamping screw, microscope mount
- 2x M6 clamping screws for C mount
- 2x rubber rings, eyepiece
Protective dust cover, stand K 415500-1800-000
Cover, C-mount Stemi 305 trino

5.3 Service

All interventions in mechanical, optical and electronic internal stereo microscope components may only be performed by Carl Zeiss Service or by specifically authorized qualified staff.
To ensure your stereo microscope is optimally set and also operates correctly over an extended period of time, we recommend that you conclude a service/maintenance contract with ZEISS.
For reordering or in case of service, contact your nearest regional ZEISS representative.
38 435063-7044-001 01/2015
Page 39
Stereo microscope CARE, MAINTENANCE AND SERVICE Stemi 305 Product disposal ZEISS

5.4 Product disposal

The product was developed, tested and produced in accordance with the valid regulations and guidelines for environmental law of the European Union.
The product and its accessories comply with the EU-regulations 2011/65/EC (RoHS) and 2012/19/EC (WEEE), and the German Electrical and Electronic Equipment Act (ElektroG).
The product contains electronic components which must be disposed of according to the requirements of the 2002/19/EC WEEE Directive and not as domestic waste. In addition, the national regulations must be complied with.
For details on disposal and recycling please refer to your relevant ZEISS sales or service organization.
01/2015 435063-7044-001 39
Page 40
APPENDIX Stereo microscope ZEISS List of abbreviations Stemi 305

6 APPENDIX

6.1 List of abbreviations

BF Bright field CL Cold Light DIN Deutsche Industrie-Norm (German Industrial Standard) DF Dark field EN Europäische Norm (European Standard) IEC International Electrotechnical Commission IP Internal Protection (Type of enclosure) IvD In-vitro diagnostics LED Light Emitting Diode RC Relief contrast (oblique light) S/W Schwarz/weiß (black / white) UV Ultra-violet
40 435063-7044-001 01/2015
Page 41
Stereo microscope APPENDIX Stemi 305 Troubleshooting ZEISS

6.2 Troubleshooting

Description of fault Cause of error Troubleshooting
Stereo microscope illumination cannot be switched ON.
Vertical reflected-light illumination is not lit.
Additional reflected-light illuminators are not lit.
Transmitted-light illuminator does not light up
Power supply interrupted. Power switch not switched ON
Power unit faulty. Replacing the power unit, see
Power supply cable RJ 12 not plugged in.
Power supply cable RJ 12 not plugged in.
Reflected light not switched ON or dimmed.
Transmitted light not switched ON or dimmed.
Transmitted light not connected. Open stand K EDU or K LAB and
Check or establish connection to power supply. Switch power switch on back of stereo microscope ON.
Section 5.2 on page 35
Insert RJ plug correctly, see Section
3.2 or 3.3 on pages 21 and 23
Insert RJ plug correctly, see Section
3.2 or 3.3 on pages 21 and 23
Press and turn knob Fig several times to switch reflected light ON and to increase its intensity; see Section 4.2 on page 28
Press and turn rotary / push-button
. 25/2 to switch illumination ON
Fig and to increase its intensity; see Section 4.3 on page 32
according to Section 5.2.3 on page 37
connect the plug-in connectors of the LED illumination.
. 20/1
Transmitted-light illuminator dirty or control lever cannot be actuated.
Microscope mount moves down by itself.
Liquids or foreign matter in the transmitted-light unit.
Focusing drive too easy to move. Set ease of motion of the focusing
Open and clean stand K EDU or K LAB and eliminate foreign matter according to Section 5.2.3 on page 37
drive, see Section 3.2 on page 21
Should any other faults occur or should you have queries regarding troubleshooting, please contact your regional ZEISS representative.
01/2015 435063-7044-001 41
Page 42
APPENDIX Stereo microscope ZEISS List of illustrations Stemi 305

6.3 List of illustrations

Fig. 1 Warning and information labels on the device................................................................... 11
Fig. 2 Warning labels and apertures for LED radiation .................................................................. 11
Fig. 3 Microscope system Stemi 305 ............................................................................................ 13
Fig. 4 Stemi 305 EDU microscope set .......................................................................................... 14
Fig. 5 Stemi 305 LAB microscope set ........................................................................................... 14
Fig. 6 Stemi 305 MAT microscope set ......................................................................................... 14
Fig. 7 Interfaces on the Stemi 305 with stand K LED (schematic diagram) ..................................... 15
Fig. 8 Front optics, Stemi 305 (example) ...................................................................................... 15
Fig. 9 Removing caps .................................................................................................................. 21
Fig. 10 Installing the stereo microscope ......................................................................................... 22
Fig. 11 Setting the ease of motion of the focusing drive ................................................................ 22
Fig. 12 Mounting spot illuminator ................................................................................................. 23
Fig. 13 Mounting ring illuminator on microscope body .................................................................. 24
Fig. 14 Mounting ring illuminator to front optics 3 ........................................................................ 24
Fig. 15 Mounting external fiber-optic illumination ......................................................................... 25
Fig. 16 Inserting the eyepiece plate ............................................................................................... 25
Fig. 17 Connecting the stereo microscope ..................................................................................... 26
Fig. 18 Adjusting the stereo microscope ........................................................................................ 27
Fig. 19 Adjusting the eyepieces ..................................................................................................... 27
Fig. 20 Setting the vertical illuminator ........................................................................................... 28
Fig. 21 Setting the spot illuminator ............................................................................................... 29
Fig. 22 Double spot illuminator ..................................................................................................... 29
Fig. 23 Ring illuminator ................................................................................................................. 30
Fig. 24 Controller K LED ............................................................................................................... 31
Fig. 25 Transmitted-light module in stand K EDU .......................................................................... 32
Fig. 26 Transmitted-light unit in stand K LAB ................................................................................. 33
Fig. 27 Changing the power unit .................................................................................................. 35
Fig. 28 Opening the stand cover plate, stand K EDU ..................................................................... 36
Fig. 29 Opening the stand cover plate, stand K LAB ....................................................................... 37
42 435063-7044-001 01/2015
Page 43
Stereo microscope APPENDIX Stemi 305 Index ZEISS

6.4 Index

A
Adjusting the stereo microscope .................... 27
Ambient Conditions ....................................... 19
Ambient temperature .................................... 19
Apertures for LED radiation ............................ 11
B
Bright field .................................................... 32
C
Care .............................................................. 34
Compensation of visual defect ....................... 27
Connecting ................................................... 26
Contents ......................................................... 5
Controller K LED ............................................ 31
Copyright ........................................................ 4
D
Dark field ...................................................... 32
Description .................................................... 13
Dimensions .................................................... 18
Diopter compensation ................................... 27
Double spot illuminator.................................. 23
E
Ease of motion of the focusing drive .............. 22
Eyepiece ........................................................ 27
Eyepiece cup ................................................. 25
Eyepiece plate ............................................... 25
F
M
Maintenance ................................................. 35
Microscope system ........................................ 13
Mixed light .................................................... 28
O
Operation ..................................................... 27
P
Product disposal ............................................ 39
Protection ring for spectacles ......................... 25
R
Reflected-light illumination ............................ 23
Ring illuminator ............................................. 24
S
Safety ............................................................. 8
Service .......................................................... 38
Setting
Reflected-light illumination ......................... 28
Stereo microscope ..................................... 27
Transmitted-light illumination ..................... 32
Setting reflected-light illumination ................. 28
Setting transmitted-light illumination ............. 32
Spot illuminator ............................................. 23
Switching OFF ............................................... 26
Switching ON ................................................ 26
Swivel arm stand
........................................... 31
System overview ............................................ 16
Fiber-optic illumination .................................. 25
Front optics 3 ................................................ 24
I
Illumination intensity ................................ 29, 30
Illumination state of the ring illuminator ......... 30
Index ............................................................. 43
Indication for use ........................................... 13
Information labels .......................................... 11
Information, general .................................. 7, 21
Installation..................................................... 21
Instrument safety ............................................. 8
Interfaces ...................................................... 15
Introduction .................................................... 7
L
LED class ......................................................... 8
T
Technical data ............................................... 18
Transmitted-light module in stand K EDU ....... 32
Transmitted-light unit in stand K LAB ............. 33
Troubleshooting ............................................ 41
U
Use ............................................................... 13
V
Vertical illumination ....................................... 28
W
Warning labels .............................................. 11
Warranty notes ............................................. 12
Wearing parts ............................................... 38
Weight .......................................................... 18
01/2015 435063-7044-001 43
Page 44
Loading...