Note: Drawings and Schematics are
not available in this revision.
FDT Visual Field Instrument
SERVICE MANUAL
Part Number 112120
Revision B October 1997
Includes CE Model
Copyright 1997
Welch Allyn Inc.
Part NoRev
DescriptionECN #DateApproved
112120 ANew Release of FDT Service Manual5-361808/22/97RJS
112120B
Added CE version Section 9 , Calibration
Limits and Error Messages.
5-3657410/21/97RJS
DRAWINGS AND / OR ILLUSTRATIONS AND / OR PART NUMBERS
CONTAINED IN THIS DOCUMENT ARE FOR REFERENCE PURPOSES
ONLY! FOR CURRENT REVISIONS CALL HUMPHREY INSTRUMENTS
CUSTOMER SERVICE
ii HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Contents
Section 1:General Information
About The 710 Series and the CE Model 1
Help Information CUSTOMER SERVICE PHONE NUMBER 2
Important Safety Warnings and Service Notes 3
Main Menu Architecture Ver.2.105
Hardware System Wiring Block Diagram6
System Function Electrical Architecture Block Diagram7
Section 2:Service
Incoming Inspection1
Tools Required for Service and Factory Calibration and Software Uploading2
Factory Calibration for Software Version 1.20 or Higher3
Loading Software for FDT 5
Final Testing6
Hypot Ground Continuity, and Leakage Current Testing11
Important Shipping Information12
Calibration Limits Information13
Product Updates15
Section 3:Troubleshooting
Chief Complaint, Cause, Corrective Action1
Power On Self Test (P.O.S.T.) Status Table3
FDT Terminology4
FDT Error Messages5
Section 4:Disassembly and Repair
4.1Removal of Rear Housing/LCD Assembly1
4.2Replacement of Rear Housing/LCD Assembly3
4.3Disassembly of Optics3
4.4Assembly of Optics11
4.5Removal of Printer15
4.6Installing Printer16
4.7Removal of MCU/ PCB17
4.8Installing MCU/PCB 19
4.9Installing a Replacement Chassis/Monitor Assembly20
4.10Removal of Transformer20
4.11Installing Transformer21
4.12Removal of PRB Jack21
4.13Replacement of PRB Jack21
4.14Removal of Power Entry Module21
4.15Replacement of Power Entry Module22
4.16Removal of Visor Position Sensor (VPS)22
4.17Replacement of VPS22
4.18Removal of Internal Photodetector (IP)23
4.19Replacement of IP / PCB23
4.20Removal of LCD Module/Keypad23
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT iii
Section 4:Disassembly and RepairCont’d
4.21Replacement of LCD Module/Keypad 24
4.22Removal of Visor Pad24
4.23Installing Visor Pad25
4.24Removal of Acrylic Lenses and Optics Tube Window25
4.25Installing Acrylic Lenses and Optics Tube Window26
FDT Assembly Drawing 112200 International (CE) 4 sheets
Section 8:Schematics 112106, 112091 , 112037
Section 9:CE Model Disassembly and Repair
9.1Removal of Rear Housing/LCD Assembly1
9.2Replacement of Rear Housing/LCD Assembly3
9.3Disassembly of Optics3
9.4Assembly of Optics10
9.5Removal of Printer14
9.6Installing Printer15
9.7Removal of MCU/ PCB16
9.8Installing MCU/PCB 19
9.9Installing a Replacement Chassis/Monitor Assembly20
9.10Removal of Transformer20
9.11Installing Transformer21
9.12Removal of PRB Jack21
9.13Replacement of PRB Jack21
9.14Removal of Power Entry Module21
9.15Replacement of Power Entry Module22
9.16Removal of Visor Position Sensor Board (VPS)22
9.17Replacement of VPS22
9.18Removal of Internal Photodetector (IP)23
9.19Replacement of IP23
9.20Removal of LCD Module/Keypad23
9.21Replacement of LCD Module/Keypad 24
9.22Removal of Visor Pad24
9.23Installing Visor Pad25
9.24Removal of Acrylic Lenses and Optics Tube Window25
9.25Installing Acrylic Lenses and Optics Tube Window26
iv HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 1 - General Information
The 710 Series FDT Visual Field Instrument
utilizes Frequency Doubling Technology (FDT).
Various preprogrammed patterns of flickering black
and white vertical bars are presented to the patient on a video screen. Each patient is tested on
the degree of contrast between the black and
white bars which enables them to see each pattern.
The Frequency Doubling stimulus taps into a
select subset of retinal ganglion cells that have
been observed to die in early glaucoma patients.
Because the stimulus is selective only to this
subset of cells, the test is accurate and rapid.
About The 710 Series FDT
Visual Field Instrument
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 1
Section 1 - General Information
Help Information:
All service and repairs must be performed by authorized
Humphrey Instruments personnel or agents, using genuine
replacement parts. Failure to do so will invalidate the prod-
uct warranty.
Read and understand all safety warnings and service
notes printed in this Service Manual and the User’s
Guide. If in doubt about any precaution or procedure,
call Humphrey Instruments Customer Service phone
number below.
For Phone Help, call Humphrey Instruments Customer Service at: 800-341-6968
The part and serial numbers are located on the bottom of
the unit.
Troubleshooting assistance is contained in Section 3 of this
manual. Troubleshooting is limited to board level replacement and not board component replacement.
Please read and understand FDT User’s Guide #112098,
a copy of which may be obtained by calling Humphrey Instruments Customer Service Department at the toll free 800
number above.
2 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 1 - General Information
Read Important Safety Information in User’s Guide.
EXERCISE EXTREME CAUTION WHEN
SERVICING THE FDT! THE MONITOR ASSEMBLY
OPERATES ON APPROXIMATELY 10,000 VOLTS!
Do not remove the plug from the end of the CRT
exposing sharp glass process tube.This can
cause cuts! (Photo 1 at right.)
Disconnect FDT from Mains when disassembling.
Clean as recommended in Users Guide.
Follow ESD precautions during service.
Test repaired FDT units at 115v and 230v prior to
returning them to customers.
Support the FDT by holding under the base to protect
the upper optics section from stress.
Important Safety Warnings
and Service Notes
Photo 1
Please DO NOT adjust pots on the monitor board.
The video monitor is a non-standard design and
because of this, there are no field monitor or video
service adjustments. Video adjustments can only be
performed by the monitor manufacturer with special
equipment. Unauthorized adjustments may result
in clinically invalid results without obvious
indications of any malfunction. (Photo 2 at right)
Tighten fasteners carefully. Refer to drawing
#112000 in this manual for recommended fastener
torque specifications.
Put a calibration cap (see bubble number 90 on
drawing 112000) on the optics tube assembly during
servicing to protect optics tube window.
The RS232 connector housing of the MCU PCB
projects below the bottom of the MCU PCB (Photo 3
at right). Do not stand the chassis monitor assembly
on this connector. Put the chassis monitor assembly
on its back between repair steps.
Photo 2
Photo 3
Always install latest revision software when installing
a new MCU/PCB. Call Humphrey Customer Service.
See Section 2 Pg. 4 for details.
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 3
Section 1 - General Information
Important Safety Warnings
and Service Notes
Photo 4
Copper ESD shields may have sharp edges.
Exercise caution to avoid cuts!
4 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 1 - General Information
FDT MAIN
ENTER
PATIENT
AGE
UTILITIES
MENU
**ADVANCE
PAPER
RUN DEMO.
RUN
PATIENT
TESTS
SET-UP
INSTRUMENT
MENU
RUN SELF
CAL.
INSTRUMENT
TESTS MENU
ABOUT FDT
RUN
SCREENING
C-20 TEST
RUN FULL
THRESHOLD
C-20 TEST
RUN FULL
THRESHOLD
N-30 TEST
SKIP EYE
SET CLOCK
MENU
SET-UP
OPTIONS
MENU
PRINT
SET-UP
REPORT
TEST
INPUTS
TEST
OUTPUTS
TEST
SERIAL
PORT
SELECT
LANGUAGE
OPTIONS
RESET TO
DEFAULTS
TEST LCD
TEST VIDEO
MENU
TEST
SPEAKER
TEST
PRINTER
MENU
AMBIENT
LIGHT
CHECK
CALIBRATION
CHECK
FACTORY
CALIBRATION
RUN
PRINTOUT
TEST
ADVANCE
PAPER
TEST VIDEO
SYSTEM
TEST A/D
CIRCUITRY
+10 years
-10 years
+ 1 year
ACCEPT SETTING
BUTTONS
PRB
VIEWPORT
POS
CHECKERBOARD
BLANK SCREEN
FULL SCREEN
** AFTER TEST PERFORMED, WILL READ
LAST PATIENT RESULTS
DIXIE PATTERN
LCD MEMORY
BACKLIGHT (ON/OFF)
Printer
Status
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 5
Section 1 - General Information
6 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 1 - General Information
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 7
Section 2 - Service
When an FDT unit is returned, check:
1) Physical condition of packaging.
2) Physical condition of unit.
3) List any accessories with unit, for example:
power cord
Patient Response Button (PRB)
calibration cap
dust cover
computer interface cable
Incoming Inspection
4) Perform “PRINT SET-UP REPORT” if
possible, and keep for reference.
1/4” square to hex adapter
Torque Wrench - Adjustable 2 - 14 in/lbs range
Torque Handle: Mountz 06-8002 TB1H 12-21
in/lb or equivalent.
Torque Handle Box Wrench accessory:
Mountz 068127 Box Head - 11/32”
Photo 1b
T15145 - for PRB connector
T15104 - Calibrator and Software Floppy Disc
(see Photo 1 at left)
T15810- Serial Port Tester
(can be made with a standard DB 9 female with
these connections: Pins 1,4,6,9 common
Pins 2,3 common
Pins 6,8 common
Digital volt meter ( three plus range)
Needle nose pliers
Photo 2
2 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 2 - Service
Upgrade recommendation: Upgrade all re-
turned FDT’s with Software Version 2.10. Perform
Factory Calibration Process.
Start Factory Calibration Process:
1 Turn FDT unit OFF.
2 Connect the wire from the T-15102 Factory
Calibrator to connector board as shown in Photo 3.
Wires must exit from top of connector as shown.
3 Remove Visor/Browrest (Sect 4 or 9 for CE FDT)
Factory Calibration
for Software Version.
1.20 or higher
4 Tilt FDT on its side to remove label over Calibration/
Software loading port J2.
5 Align threaded studs of connector board with the two
holes near port J2 as shown in Photo 4.
6 Insert T 15102 into port J2 as shown in Photo 5.
Hookup should be as shown in Figure 1 below.
Fig. 1
connector board
T 15102
7 Position the FDT in upright position and place
calibrator onto the slide bracket . See Photo 6
Note: Place calibrator on viewport tightly to prevent
light from entering the calibrator housing.
Photo 3
Photo 4
8 Turn the FDT ON and view the menu on the LCD:
a) Press button next to UTILITIES MENU
b) Press button next to INSTRUMENT TEST MENU
c) Press button next to TEST OUTPUTS
d) Press button next to TEST VIDEO MENU
e) Press button next to FACTORY CALIBRATION
NOTE: Unit can be calibrated with PC utility Calibrat
through the FDT computer interface. Refer to the Calibrate utility for more information.
9 Press any button to indicate that you have placed
the factory calibrator on the view port.
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 3
Photo 5
Photo 6
Section 2 - Service
Factory Calibration
for Software Version.
1.20 or higher
(continued)
10 Press any button to start the Factory Calibration
calibration process. Duration : less than 1minute.
NOTE: Self-Calibrate is automatically performed
during Factory Calibration in Version 2.10
11 Turn the FDT OFF and remove calibrator from
slide bracket.
12 Remove T-15102 connector board from Port J2 on
bottom of FDT. Install a new label part # 112066
over J2 port.
13 Turn FDT ON and view the menu on the FDT LCD:
a) Press the green button until the UTILITIES
Menu is reached.
b) Put the black calibration cap on the viewport.
c) Press the button next to RUN SELF
CALIBRATION. (Process time: 2 - 3 min.)
Note: A progress indicator will display on the LCD. It
will indicate what percentage of the calibration is
complete. The unit will indicate when calibration is
complete and return to UTILITIES MENU.
14 Now perform a Calibration check:
a) Press button next to INSTRUMENT TEST
MENU.
b) Press button next to TEST OUTPUTS
c) Press button next to TEST VIDEO MENU
d) Press button next to CALIBRATION CHECK
Note: The meter will indicate that the FDT is fully
calibrated.
15 Press any button to continue.
16 Press the green button 4 times or until FDT MAIN
MENU is reached.
17 Install the visor/browrest with the 4 black hex socket
head cap screws as shown in Section 4.
Note:See Calibration Limits/Parameters Tables
4 HUMPHREY FDT VISUAL FIELD INSTRUMENT
in this section.
(End of Factory Calibration Process)
SERVICE MANUAL 112120 REV. B
Section 2 - Service
Start Software Loading Process:
1Connect the ribbon cable to the ICD Cable
connector as shown in Photo 7 at right.
Note: a) 2 Pins at right are not used
b) Align the red wire of the ribbon cable
with the red arrow on ICD connector.
2Connect the ribbon cable to the connector board as
shown in Photo 8.
3Connect ICD connector to the PC Parallel Port .
See Figure 2 below.
Loading Software for FDT
Note: FDT software is not
Mac compatible.
Photo 7
Figure 2
4Insert T 15102 into port J2 as shown in Photo 5
and follow software loading instructions displayed
on the PC when running FDT Software Installation
Program.
Note: Whenever installing a new MCU/PCB and
loading highest revision software, verify proper
hardware (marked on MCU/PCB) version using
“ABOUT FDT”. LCD Screen will display
software/hardware data. See Photo 9
Note: Follow this final testing process to verify
correct operation after the FDT has been serviced.
1. Calibration
Follow instructions in the calibration section of
manual.
2.230 Volt testing
Remove fuse block. Replace the 115v fuses
(.315A) with those recommended for
230v.(.160A)
Place fuse block back into unit so the 230v
shows. Plug unit into a 230v line and turn unit
on. After approximately 10-15 seconds the unit
should go through a series of beeps and the
FDT Main menu should appear. After this test is
done, turn off unit, remove power and replace
the fuses with the 115V (.315A) ones. Be sure
that the115V shows in the fuse drawer.
3. Operational testing
3.1Plug in PRB.
3.2Turn unit on. After approximately 10-15
seconds the unit will complete self
diagnostics and emit two double beeps.
The FDT main menu will appear.
3.3Press the button next to UTILITIES
MENU.
3.3.1 Check the intensity adjustment.
a)Press the button pointing
b)Press the button pointing
c)Use both buttons as needed
downward. The LCD
contrast will decrease.
upward. The LCD contrast
will increase.
to return to usable LCD
contrast.
6 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 2 - Service
3.3.2 Press the button next to ABOUT
3.3.3 Press the green button on the
3.4Press the button to the left of
INSTRUMENT TESTS MENU.
3.4.1 Press the button next to
FDT. The screen will indicate that
the unit is fully calibrated and show
the software revision. Make
sure that it has the latest revision
software.
bottom left. This should bring you
back to the UTILITIES MENU.
TEST INPUTS.
a)Press each of the 4 upper
left buttons. When each
buttons is pressed, the
word PRESS to the right of
the button will change to a
PRESSED highlighted.
Release the button. PRESS
will reappear unhighlighted.
b)Move the visor to the three
different positions - left,
middle and right (patients
side). The Eye test posi-tion on the right hand side
of the display will change
state to indicate the current
eye test position.
c)The Patient button will
indicate NOT PRESSED.
Press the PRB and it
will report a highlighted
PRESSED.
d)Unplug the PRB plug from
the PRB jack. The Patient
Button status will indicate:
Not connected. Plug the
PRB plug into the PRB jack.
Final Testing
(page 2 of 6)
3.4.2 Press the green button on the
bottom left to return to
INSTRUMENT TESTS MENU.
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 7
Section 2 - Service
Final Testing
(page 3 of 6)
3.5 Press the button next to TEST OUTPUTS.
3.5.1 Press the button next to TEST
LCD.
a)Observe highlighted test
option on screen.
b)Press button next to RUN
TEST. The LCD will reflect
state of the highlighted test
option.
c)Press any button to exit
present LCD test.
d)Press button next to NEXT
CHOICE to change test
option.
e)Perform the above steps for
the CHECKERBOARD,
BLANK SCREEN, FULL
SCREEN, AND DIXIE
PATTERN options.
f)Press button next to NEXT
CHOICE to highlight LCD
MEMORY TEST.
g)Press button next to RUN
TEST. The display will be;
Testing LCD memory andchange to LCD memory
passed. Press any button
to exit test.
h)Press button next to NEXT
CHOICE to highlight the
BACKLIGHT option.
i)Press button next to RUN
TEST a number of times
and observe the LCD
screen. The backlight
should toggle on and off.
Make sure the Backlight is
on before exiting this test.
j)Press the green button to
return to OUTPUT TESTS.
8 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 2 - Service
3.5.2 Press the button next to TEST
3.5.3 Press the button next to AMBIENT
VIDEO MENU.
a)Press the button next to
TEST VIDEO SYSTEM.
b)Look into the patient viewing
area. A large flashing circle
exhibiting the Model 710
Seriesgrating pattern in the
center of the viewing area
will be visible.
c)Press the Green button to
return to TEST VIDEO
MENU.
LIGHT CHECK.
a)There may be a short time
for monitor warm up.
b)With the calibration cap on,
the pointer will be in the
dark range.
c)Remove the calibration cap.
The pointer will rise to
the light or bright range,
depending upon the intensity of the room lighting.
d)Put the calibration cap back
on and press the green
button to return to TEST
VIDEO MENU.
Final Testing
(page 4 of 6)
3.5.4 Put Calibration cap on. Press
button next to CALIBRATION
CHECK.
NOTE:Calibration will be denied if cap is off and room
light level is too high.
a)The pointer should be in the
fully calibrated range.
b)Do not press the factory
calibration option. If this
option is accidentally
chosen, the unit will detect that a calibrator is not
attached and warn user.
c)Press the green button twice
to return to OUTPUT
TESTS MENU.
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 9
Section 2 - Service
Final Testing
(page 5 of 6)
Final Testing
3.6 Press the button next to TEST SPEAKER.
3.6.1 There will be increasing tones.
3.7 Press the button next to TEST PRINTER
MENU.
3.7.1 The TEST PRINTER MENU will be
displayed. Press the button next to
RUN PRINTOUT TEST. The FDT
will print out a series of patterns
and instrument data.
3.7.2 Press the button next to ADVANCEPAPER. The paper will
advance about 1-2 inches beyond
the slot in the paper door.
3.7.3 Press the button next to PRINTER
STATUS. “PRINTER OK”
message will appear.
Photo 10
3.7.4 Press the green button 3 times to
return to INSTRUMENT TESTS
MENU.
3.8 Press the button next to TEST SERIAL
PORT.
3.8.1 Press any of top four buttons to
execute test.
3.8.2 Both the data transfer and hand
shaking test should fail.
3.8.3 Press the green button to return to
INSTRUMENT TESTS MENU.
3.8.4 Place T-15810 into the serial port
on bottom of unit. See Photo 10.
3.8.5 Press the button next to TESTSERIAL PORT.
a)Press any of top four but-
tons to execute tests.
b)LCD will display:
DATA TRANFER PASSED
HANDSHAKING PASSED
10 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 2 - Service
3.8.6 Press the green button twice to
3.8.7 Remove T-15810.
3.9 Press the button next to A/D CIRCUITRY.
3.9.1 A reading will come up for
3.10 Perform a screening test to check unit
operation.
3.11 Turn unit off.
4. Agency testing
4.1 Hypot
Testing at 1500VAC for one second.
Apply the voltage between primary wiring
(supply terminals connected together)
and accessible dead metal parts without
breakdown indication.
4.2 Ground continuity
Total resistance of the product shall not
exceed 0.2 ohms (2.0VDC). The test will
be conducted using a 10 amp DC current
source with a no load voltage no greater
than 6.0VDC for five seconds. Measure
ment will be taken from the grounding pin
of the power supply cord to the accessible
metal part of the product.
4.3 Leakage current
The leakage current measured from any
exposed grounded metal parts to the
grounding pin of the power supply cord
shall not exceed 100 microamps, AC
RMS (95.5MVAC RMS) with input voltage
set to 120VAC, 60HZ or 240VAC, 50HZ.
return to INSTRUMENT TESTS
MENU.
Mean gray and printhead tem perature. If the unit fails this test
a malfunction error will appear.
Final Testing
(Page 6 jof 6)
Hypot, Ground Continuity,
and Leakage Current
Testing
(For In-House use only.)
If a FTD fails any test in this section, set it aside for
troubleshooting and repair. When the FDT is repaired and passes all the tests, place a label over
the calibration port.
SERVICE MANUAL 112120 REV. B
HUMPHREY FDT VISUAL FIELD INSTRUMENT 11
Section 2 - Service
Important Shipping
Information
Paper roll foam prevents
un-winding during ship-
ping
Note:Install a new roll of paper.
Install the paper roll foam before
packing the unit for shipping. See
photo at left. This foam strip (called
foam shipping wedge in User’s Guide)
prevents the paper roll from unwinding
during shipping.
Return all of the customer’s
accessories.
12 HUMPHREY FDT VISUAL FIELD INSTRUMENT
SERVICE MANUAL 112120 REV. B
Section 2 - Service
Calibration Limits Information
Table A:Calibration Parameters Printed on the System Status Report
ParameterNominalSW
Limit
Expected
Min
Min
SCAN LINE
(internal)
GAMMA*2.351.0*2.2*2.5<3.0
REFERENCE
POT
INTERNAL
TARGET
DC OFFSET300*10*501023
INTERNAL
PEAK
*906*60*110235
NA0
*22500*1500*30004095
Internal
Target0
*15
(new CRT)
Internal
Target - 40
Expected
Max
*60
(new CRT)
Internal
Target + 40
SW
Limit
Max
255
4095
Comments/DescriptionF.C.**S.F.***
Reflects mechanical alignment of
photodetector to CRT.
CRT response correction
exponent.
Adjusts for CRT aging.
( *150 expected max on old CRT)
Absolute value read from Internal
photodetector ADC when CRT is
200 c/m^2.
DC offset on internal
photodetector signal.
Absolute value read from Internal
photodetector ADC when CRT is
as close to 200 c/m^2 as the
reference pot can adjust.
X
X
X
* These numbers are approximate and have not been verified by statistical tracking.
** Adjusted during Factory Calibration
***Adjusted during Self Calibration
X
X
Table B:Calibration Parameters to LCD and Serial Port: Factory Calibration
MessageExpected LimitsComment
INTERNAL PEAKNAThis is a transeint message that reports DAC settings.
External
Photo Sensor Stop
Line
External Photosensor3000+/- 20
DC OFFSETSame as in table (A) above.Same as in table (A) above.
Internal Photosensor
Stop Line
Internal Photosensor
* These numbers are approximate and have not been verified by statistical tracking.
SERVICE MANUAL 112120 REV. B
*40 min, *70 nom, *100 maxTwo passes: Second pass should be within 16 of first.
External target 3000 = 200c/m^2. This is how close the
reference pot gets to 200 c/m^2.
Same as SCAN LINE in table
(A) above.
Same as INTERNAL TARGET
in table (A) above.
Same as in table (A) above.
Same as in table (A) above.
HUMPHREY FDT VISUAL FIELD INSTRUMENT 13
Section 2 - Service
Calibration Limits Information
Table C:Calibration Parameters to LCD and Serial Port: Self Calibration
MessageExpected LimitsComment
Internal Photosensor
Ambient Light
REFERENCE POT
Dynamic Range82 min allowed by SW (50:1)Full screen brightness to zero screen brightness ratio.
Gamma Factor
Linearity
Same as INTERNAL TARGET
in table (A) above.
82 max allowed by SW
(2% of full brightness)
Same as REFERENCE POT in
table (A) above.
Same as in GAMMA
CORRECTION in table (A) above.
Second Column is limited to 120
(3% deviation from straight line)
Same as in table (A) above.
Keep unit as dark as possible. If this fails in dark, unit
is broken, or DC offset is bad - Factory calibration
needed.
Same as in table (A) above.
Same as in table (A) above.
First column is linear ramp target value written to
DAC. Second column is photodetector reading
deviation from the linear ramp. Scale is 127 - 4095.
* These numbers are approximate and have not been verified by statistical tracking.
14 HUMPHREY FDT VISUAL FIELD INSTRUMENTSERVICE MANUAL 112120 REV. B
Section 2 - Service
Product Updates
EffectivityS/N Cut-In
7/97S#9700200
8/97S#9701000
8/29/97~S#9701145 Color changes of housings:cream to gray.
9/11/97S#9701199
9/15/97~S#9701210
10/3/97S#9701294SW Version 2.10 (Language translations)
DOMESTIC FDT UNITS :
Description of Changes for 00710-11
Sales Demo Units (all S#>9700199 and
9701000) . Monitor needs rework at some time
to extend calibration life.
SW Version 8.30.
HW Version 50300 (MCU PCB 112037-503)
Initial Production units for end users (all
S#>9701000). SW Version 1.20.
HW Version 50300 (MCU PCB 112037-503)
HW version 50400 (MCU PCB 112037-504 for
CE)
Calibration Caps (pn112078-1) Started shipping
with instruments.
EffectivityS/N Cut-In
9/97S#9731000
S#9731025
(also
10/3/97
S#9731019
and
S#9731023
are Ver. 2.10
SERVICE MANUAL 112120 REV. B
CE INTERNATIONAL FDT UNITS :
Description of Changes for 00710-X1 (X=2-9)
Note:Domestic S#'s have a 0 in the 3rd digit,
International S#'s have a 3)
Initial International Production units (all
S#'s>9731000)
SW Version 1.20.
HW version 50400 (MCU PCB 112037-504 for CE)
SW Version 2.10 (language translations)
HUMPHREY FDT VISUAL FIELD INSTRUMENT 15
Loading...
+ 68 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.