Xylem CG 853P User Manual

Laboratory
conductivity meter
CG 853P
Operating Manual
ba12246e_2 09/99
Conductivity meter with integrated printer
77
Accuracy when
All details contained in this operating manual are valid data at the time of going to press. However, SCHOTT may add details for both technical and commercial reasons, or in or­der to fulfil legal requirements of different countries. This does not affect the characteristics described.
Warranty
declaration
The designated instrument is covered by a warranty of three years from the date of purchase. The instrument warranty extends to manufacturing faults that are determined within the period of warranty. The warranty excludes components that are replaced during maintenance such as batteries, etc.
The warranty claim extends to restoring the instrument to readiness for use but not, however, to any further claim for damages. Improper handling or unauthorized opening of the instrument invalidates any warranty claim.
To ascertain the warranty liability, return the instrument and proof of purchase together with the date of purchase freight paid or prepaid.
78
Copyright
© Hofheim SCHOTT Geräte GmbH 1999
Reprinting - even as excerpts - is only allowed with the explicit written authorization of SCHOTT Geräte GmbH, Hofheim. Printed in Germany.
List of contents
1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 81
1.1 Keyboard . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .82
1.2 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
1.3 Sockets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .84
1.4 Declaration of Conformity . . . . . . . . . . . . . . . . . . .85
1.5 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . .86
2 Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
2.1 Authorized use . . . . . . . . . . . . . . . . . . . . . . . . . . .92
2.2 General safety instructions . . . . . . . . . . . . . . . . . .92
3 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . 95
4 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
4.1 Switch on the instrument . . . . . . . . . . . . . . . . . . . .97
4.2 Measuring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .99
4.2.1 Conductivity . . . . . . . . . . . . . . . . . . . . . . .101
4.2.2 Salinity . . . . . . . . . . . . . . . . . . . . . . . . . . .102
4.2.3 TDS (total dissolved solids) . . . . . . . . . . .103
4.2.4 Printing/transmitting measured values . .105
4.3 Determining/setting up the cell constant [C] . . . .106
4.4 Storing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117
4.4.1 Manual storage . . . . . . . . . . . . . . . . . . . . .117
4.4.2 Switching on AutoStore (Int 1) . . . . . . . . .119
4.4.3 Outputting the data storage . . . . . . . . . . .121
4.4.4 Clearing the storage . . . . . . . . . . . . . . . . .127
4.5 Data transmission . . . . . . . . . . . . . . . . . . . . . . . .128
4.5.1 Data transmission interval (Int 2) . . . . . . .128
4.5.2 Recorder (analog output) . . . . . . . . . . . . .130
4.5.3 PC/external printer (RS232 interface) . . .131
4.6 Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . .132
4.7 Reset . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .137
79
List of contents
5 Maintenance, cleaning, disposal . . . . . . . . . . . 139
5.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
5.1.1 Changing the batteries . . . . . . . . . . . . . . .140
5.1.2 Replacing the roll of printer paper . . . . . .141
5.2 Cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .142
5.3 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .142
6 What to do if... . . . . . . . . . . . . . . . . . . . . . . . . . . 143
7 Lists . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
80

1Overview

Overview
The compact perform conductivity measurements rapidly and reliably. The
CG 853P
fort, reliability and measuring safety for all applications. The integrated printer enables the measurements to be
documented in compliance with GLP. The proven procedures to determine or set up the cell con­stant support your work with the conductivity meter.
4
CG 853P
provides the highest degree of operating com-
precision conductivity meter lets you
3
2
1
1 Keypad 2 Display 3 Sockets 4 Printer
Note
The measuring instrument can also be delivered as part of a set. Information on this and other accessories is available in the SCHOTT general catalog.
81
Overview

1.1 Keyboard

Measuring instrument ON/OFF
<on/off>
Select measuring mode
ì>
<
Set up or determine cell constant; select temperature compensation <CAL>
Activate/deactivate AutoRead function
<auto read>
Store measured value
<STO>
Display/transmit measured values
<RCL>
Reduce values, scroll
>
<
Confirm inputs, start AutoRead
<run/enter>
82
Increase values, scroll
>
<
Overview
Print LED Red: printer off (offline) or malfunction Green: printer ready for operation (online)
Printer ON/OFF
<print on/off>
Paper feed/line feed (Printer off - Print LED lights up red)
<feed>
83
Overview

1.2 Display

Status line
Measured value display
Function and temperature displ a y

1.3 Sockets

123
Connections:
1 8 pole socket with pin contacts 2 RS 232 interface/analog output 3 Plug-in power supply
Caution
Only connect probes to the instrument that cannot feed excessive voltages or currents (> SELV and > circuit with current limiter). Almost all commercial electrodes - especially SCHOTT electrodes - meet these requirements.
84

1.4 Declaration of Conformity

Overview
29.03.99
SCHOTT Geräte GmbH
Im Langgewann 5
D 65719 Hofheim am Taunus
Deutschland, Germany, Allemagne
29. März, March 29rd, 29, Mars, 1999 AGQSF0000-A060-00/990329
85
Overview

1.5 Technical data

Ambient
temperature
Measuring ranges
and resolution
Storage tempera­ture
Operating temper­ature
Allowable relative humidity
[µS/cm]
[mS/cm]
- 25 °C ... + 65 °C
0 °C ... + 55 °C
Annual mean: < 75 % 30 days/year: 95 % Other days: 85 %
0.000 ... 1.999 (only with 0.01 cm
-1
cell constant )
0.00 ... 19.99
-1
(only with 0.01 cm
0.1 cm
-1
cell constant )
constant and
0.0 ... 199.9 0 ... 1999
0.00 ... 19.99
0.0 ... 199.9 0 ... 500
SAL 0.0 ... 70.0 according to the IOT
table
TDS [mg/l] 0 ... 1999
Factor can be set in the range
0.40 ... 1.00 T [°C] − 5.0 ... + 100.0 T [°F] + 23.0 ... + 212.0
86
Overview
Measuring accuracy
(± 1 digit)
Given as measuring uncertainty at a confidence level of 95%. All accuracies additionally depend on the accuracies of the measuring probes.
No compensation:
Accuracy± 0.5 %
Non-linear compensation : AccuracyTest sample temperature ± 0.5 %0 °C ... 35 °C according to EN 27 888; ± 0.5 % 35 °C ... 50 °C extended nLF function according to SCHOTT measurements
Linear compensation : AccuracyTest sample temperature ± 0.5 % 10 °C ... 75 °C (the accuracy percentage always refers to the measured value!)
SAL Range 0.0 ... 42.0
AccuracyTest sample temperature ± 0.1at 5 °C ... 25 °C
± 0.2 at 25 °C ... 30 °C TDS [mg/l] 1 T [°C] NTC 30:
Accuracy± 0.1
PT 1000:
AccuracyOperating temperature
± 0.5 0 °C ... 15 °C
± 0.1 15 °C ... 35 °C
± 1 35 °C ... 55 °C
87
Overview
Measuring accuracy
(± 1 digit)
(Continuation)
Cell constant,
to be set
Cell constant,
calibrated
Reference tempera-
ture, selectable
T [°F] NTC 30:
Accuracy± 0.2 PT 1000:
AccuracyOperating temperature ± 0.9 at 32 °F ... 59 °F ± 0.2at 59 °F ... 95 °F ± 1.8 at 95 °F ... 131 °F
C [cm
-1
] 0.01
0.1
0.250 ... 2.500
C [cm
-1
] 0.450 ... 0.500
0.800 ... 1.200
T
REF [°C] 20
25
Temperature input
Dimensions and
weight
Manual [°C] - 5 ... +100
Length [mm] 240 Width [mm] 285 Height [mm] 85 Weight [kg] Approx. 1.3
(without plug-in power supply unit)
88
Overview
Power supply
Batteries 4 x 1.5 V AA type alkaline manga-
nese batteries Runtime Approx. 3000 operating hours Mains power supply Connection max. overvoltage cat-
egory II:
Plug-in power supply (Euro plug):
Type no.: Z851
Order no.: 28 520 4897
FRIWO FW3288, 11.8134
Friwo Part No. 1816492
Input: 230V ~ / 50 Hz / 23 VA
Output: 6 V = / 1,8 A /10,8 VA
Plug-in power supply (US plug):
Type no.: Z852
Order no.: 28 520 4901
FRIWO FW3288, 11.8451
Friwo Part No. 1816493
Input: 120V ~ / 60 Hz / 21,5 VA
Output: 6 V = / 1,8 A
EMC and
VDE norms
Plug-in power supply (UK plug):
Type no.: Z849
Order no.: 28 520 4975
FRIWO FW3288, 11.8453
Friwo Part No. 1770896
Input: 230V ~ / 50 Hz / 23 VA
Output: 6 V = / 1,8 A
Interference emission (generic standard)
Interference immunity
EN 50081-1
FCC class A
EN 50082-1 (generic standard)
Protective class 3, EN 61010-1 Climatic class 2, VDI/VDE 3540
89
Overview
Test marks TÜV GS, UL/CUL, CE
Display Multifunctional LCD Keypad Foil keypad (Polyester)
Data storage Ring store for 200 value pairs, conductivity/salinity/TDS,
temperature
Temperature
compensation
Connectors
Printer
Automatic with Pt 1000/NTC (30 k
W)
-5 ... 100.0 °C
Manual input -5 ... 100 °C resolution 1K
Measuring cell 8 pole socket with pin contacts RS interface/
analog output
Bidirectional RS 232 interface or analog output with autom. recogni­tion of the PC, printer, or recorder connected
Plug-in power supply 2 pole special Friwo
Print principle Matrix thermoprint Signs per line 26 Printing width 46.1 mm
90
Width of printing paper 58 mm Printing speed 40 signs/sec
Safety

2 Safety

This operating manual contains basic instructions that you must follow during the commissioning, operation and main­tenance of the instrument. Consequently, all responsible personnel must read this operating manual before working with the instrument. The operating manual must always be available within the vicinity of the instrument.
Target group This measuring instrument was developed for use in the lab-
oratory. Thus, we assume that, as a result of their professional train­ing and experience, the operators will know the necessary safety precautions to take when handling chemicals.
Symbols used
Caution
indicates instructions that have to be followed to prevent damage to your instrument.
Warning
indicates instructions that have to be followed to protect yourself and the instrument from dangerous electrical volt­age.
Note
Indicates notes that draw your attention to special features.
Note
Indicates cross-references to other documents, e.g. appli­cation reports, operating manuals of measuring cells, etc.
91
Safety

2.1 Authorized use

This instrument is authorized exclusively for measuring the conductivity, salinity, temperature and TDS (total dissolved solids) in the laboratory. The technical specifications as given in the section
ECHNICAL DATA, must be observed. Only the operation
1.5 T and running of the measuring instrument according to the instructions given in this operating manual is authorized. Any other use is considered unauthorized.

2.2 General safety instructions

This instrument is constructed and tested in compliance with the EN 61010-1 safety regulations for electronic measuring instruments. It left the factory in a safe and secure technical condition.
Function and
operational safety
The smooth functioning and operational safety of the instru­ment can only be guaranteed if the generally applicable safety measures and the specific safety instructions in this operating manual are followed.
The smooth functioning and operational safety of the instru­ment can only be guaranteed under the climatic conditions specified in the section 1.5 T
ECHNICAL DATA.
If the instrument was transported from a cold environment to a warm environment, the formation of condensate can lead to the faulty functioning of the instrument. In this event, wait until the temperature of the instrument reaches room tem­perature before putting the instrument back into operation.
Caution
The instrument is only allowed to be opened by personnel authorized by SCHOTT.
92
Safety
Safe operation If safe operation is no longer possible, the instrument must
be taken out of service and secured against inadvertent operation. Safe operation is no longer possible if:
l the instrument has been damaged in transport l the instrument has been stored under adverse conditions
for a lengthy period of time
l the instrument is visibly damaged l the instrument no longer operates as described in this
manual
If you are in doubt contact the supplier of the instrument.
Obligations of the
operator
The operator of this measuring instrument must ensure that the following laws and guidelines are observed when using dangerous substances:
l EEC directives for protective labor legislation l National protective labor legislation l Safety regulations l Safety datasheets of the chemical manufacturer.
93
Safety
94
Commissioning

3 Commissioning

Perform the following activities for initial comissioning:
l Set the date and time l Connect the plug-in power supply (for printer functions).
Setting the date
and time
1 Press and hold down the <ì> key. 2 Press the <on/off> key.
The
display test
appears briefly on the display. The measuring instrument then switches automati­cally to the setting of the baud rate.
3 Press the <run/enter> key repeatedly until the date
flashes on the display.
4 Set today’s date by pressing <
> <>.
5 Confirm with <run/enter>.
The date (month) flashes on the display.
6 Set the current month by pressing <
> <>.
7 Confirm with <run/enter>.
The year appears on the display.
8 Set the current year by pressing <
> <>.
9 Confirm with <run/enter>.
The hour field flashes on the display.
10 Set the current time by pressing <
> <>.
11 Confirm with <run/enter>.
The minutes field flashes on the display.
12 Set the current time by pressing <
> <>.
13 Confirm with <run/enter>. The measuring instrument
then switches to the measuring mode.
14 Switch off the instrument by pressing <on/off>.
95
Commissioning
Connecting the
plug-in power
supply
The plug-in power supply supplies the pH meter with low voltage (7.5 V ... 12 V DC).
Warning
The line voltage on site must lie within the input voltage range of the original plug-in power supply unit (see section
1.5 T
ECHNICAL DATA).
Caution
Use original plug-in power supplies only (see section 1.5 T
ECHNICAL DATA).
2 1
3
1 Insert the plug (1) into the socket (2) of the conductiv-
ity meter.
2 Connect the original SCHOTT plug-in power supply
(3) to an easily accessible mains socket.
Note
You can also perform measurements without a plug-in power supply. However, the functions of the integrated printer can only be used with connected plug-in power sup­ply.
96
Operation

4Operation

4.1 Switch on the instrument

1 Place the instrument on a flat surface and protect it
against intense light and heat.
2 Connect the conductivity measuring cell to the instru-
ment.
3 Press the <on/off> key.
display test
The The instrument then switches automatically to the previously selected measuring mode.
4 Check or determine the cell constant [C]. The proce-
dure is described in section 4.3, on page 106.
appears briefly on the display.
Note
The instrument has an energy saving feature to avoid unnecessary battery depletion. The energy saving feature switches the instrument off if no key has been pressed for an hour. The energy saving feature is not active:
l if the power is supplied by the plug-in power supply l if the l if a PC is connected l if the recorder cable is connected l if the printer cable is connected
AutoStore
function is active
(for external printers)
97
Operation
Checking the cell
constant
1 Press the <CAL> key repeatedly until LF CELL is dis-
played.
2 Press the <run/enter> key.
The previously selected cell constant ist displayed, e. g. 0.475 cm
-1
with a calibrated cell constant.
3 To return to the measuring mode: press the <ì> key
when the correct cell constant ist displayed.
4 If a different cell constant is to be set, proceed ac-
cording to section 4.3 D
CELL CONSTANT [C].
ETERMINING/SETTING UP THE
98

4.2 Measuring

Operation
Preparatory
activities
Temperature probe Measurements can be performed with and without a temper-
Perform the following activities when you want to measure:
1 Connect the measuring cell to the instrument. 2 Adjust the temperature of the test solutions or mea-
sure the current temperature if the measurement is made without a temperature probe.
3 Calibrate the instrument with the measuring cell or
check the cell constant set up.
4 Select the measuring mode by pressing <
Caution
When connecting an earthed PC/printer, measurements cannot be performed in earthed media as incorrect values would result. The RS232 interface is not galvanically isolated.
ature probe. A connected temperature probe is indicated by TP on the display. If you want to use a SCHOTT conductivity measuring cell without a temperature probe, you have to connect it with an adapter (available at SCHOTT).
ì>.
Note
The conductivity meter automatically recognizes the type of the temperature probe used. As a result, you can connect electrodes with the NTC30 or Pt1000.
The temperature measurement is absolutely essential for a reproducible conductivity measurement. If the measure­ment is made without a temperature probe, proceed as fol­lows:
1 Determine the current temperature using a thermom-
eter.
2 Set up the temperature by pressing <
> <>.
99
Operation
AutoRead AR (Drift control)
The
AutoRead
function (drift control) checks the stability of the measurement signal. The stability has a considerable effect on the reproducibility of the measured value.
1 Call up the measuring mode by pressing <ì>. 2 Activate the AutoRead function by pressing
<auto read>. The current measured value is frozen (Hold function).
3 Start the AutoRead function by pressing
<run/enter>. AR flashes on the display until a stable measured val­ue is reached. This measured value is transmitted to the interface.
.
100
4 If necessary, start the next AutoRead measurement
by pressing <run/enter>.
5 To terminate the AutoRead function: Press the
<auto read> key.
Note
The current AutoRead measurement (with acceptance of the current value) can be terminated at any time by pressing
<run/enter>.
Operation
4.2.1 Conductivity
To measure the conductivity, proceed as follows:
1 Perform the preparatory activities according to sec-
tion 4.2.
2 Immerse the conductivity measuring cell into the test
sample.
3 Press the <ì> key until appears in the status dis-
play. The conductivity value appears on the display.
4 If necessary, set the temperature by pressing
<
><> (see section 4.2, page 99).
5 Perform an AutoRead measurement according to
section 4.2.
101
Operation
4.2.2 Salinity
To measure the salinity, proceed as follows:
1 Perform the preparatory activities according to sec-
tion 4.2.
2 Immerse the conductivity measuring cell into the test
sample.
3 Press the <ì> key repeatedly until Sal appears in the
status line. The salinity value of the sample appears on the display.
4 If necessary, set the temperature by pressing
><> (see section 4.2, page 99).
<
5 Wait for a stable measured value.
102
6 Perform an AutoRead measurement according to
section 4.2.
Operation
4.2.3 TDS (total dissolved solids)
To measure the TDS, proceed as follows:
1 Perform the preparatory activities according to sec-
tion 4.2.
2 Immerse the conductivity measuring cell into the test
sample.
3 When measuring with an integrated temperature
probe continue with step 4. When measuring without temperature probe:
– Determine the temperature of the test sample using
a thermometer
– Press the <
ì> key repeatedly until appears in
the status line
– Enter the temperature using <
4 Press the <
ì> key repeatedly until TDS appears in
> <>.
the status line. The TDS value of the sample appears on the display.
5 Set up the TDS factor (0.40 ... 1.00) using <> <>.
6 Wait for a stable measured value.
103
Operation
7 Perform an AutoRead measurement according to
section 4.2.
104
Operation
4.2.4 Printing/transmitting measured values
Measured values (data records) can be:
l printed on the integrated printer or l transmitted to the interface.
Note
To print, you must switch on the printer using the Print key <print on/off> (Print LED lights up green). To transmit to the interface, you must connect the interface cable. The integrated printer is then switched off automati­cally.
You can print measured values (data records) in 3 ways:
l Switch on the data transmission (Int 2) (see page 128)
– After expiry of the selected interval, the current data
record is sent to the printer/interface.
l Switch on AutoStore (Int 1) (see page 119)
– After expiry of the selected interval, the current data
record is sent to the printer/interface and in addition is stored in the data store of the instrument
– AutoStore (Int 1) covers the
data transmission
interval
(Int 2).
l Press the <run/enter> key
This manually triggers a printout of the current measured values at any time - independently of the selected inter­vals.
Note
If you connect a recorder (analog output), both the output to the integrated printer and the digital output are switched off.
105
Operation

4.3 Determining/setting up the cell constant [C]

Why determine/set
up the cell
constant?
Determining the cell
constant
(calibration in
control standard)
Due to ageing, the cell constant slightly changes. As a result, an inexact measured value is displayed. Calibration determines the current value of the cell constant and stores this value in the instrument. Thus, you should calibrate at regular intervals.
You can determine the cell constant of the conductivity mea­suring cell in the range 0.450 ... 0.500 cm
-1
cm
by calibrating in the control standard or set it up manu-
ally in the range 0.250... 2.500 cm
-1
select one of the fixed cell constants 0.1 cm
-1
or 0.800 ... 1.200
. Additionally, you can
-1
or 0.01 cm-1.
Determine the cell constant as follows:
1 Press the <CAL> key repeatedly until LF CELL ap-
pears on the display.
106
2 Press the <run/enter> key repeatedly to go to the
"Determining/ setting up the cell constant" function.
3 Press the <CAL> key repeatedly until the following is
displayed.
Operation
4 Immerse the measuring cell into the 0.01 mol/l KCL
control standard.
5 Press the <run/enter> key.
– If no temperature probe is connected, enter the
current temperature of the solution using <
> <>
and confirm with <run/enter>
– If a temperature probe is connected, the AR
measurement to determine the cell constant starts.
flashes until a stable signal is achieved. The determined cell constant ist displayed; the measuring instrument automatically stores the cell constant.
Note
If error message  appears see chapter 6 W
HAT TO DO IF...
AutoRead During calibrating, the
activated. The AR display flashes. The calibration procedure is finished when the AR display stops flashing.
AutoRead
function is automatically
107
Operation
Note
This method of automatically determining the cell constant by calibrating in the 0.01 mol/l KCL control standard can only be used for measuring cells with a cell constant in the
-1
range 0.450 ... 0.500 cm
or 0.800 ... 1.200 cm-1.
Calibration
evaluation
After calibrating, the instrument automatically evaluates the current condition of the calibration. The evaluation appears on the display.
Display Cell constant [cm-1]
0.450 ... 0.500 cm
0.800 ... 1.200 cm
-1
-1
outside of the ranges

0.450 ... 0.500 cm
-1
or
Perform error elimination
0.800 ... 1.200 cm
-1
according to chapter 6W
HAT TO DO IF...
Calibration protocol The calibration protoc ol con tai ns the cal ibr ati on dat a of the
current calibration.
108
Note
You can automatically print out a calibration protocol after the calibration. To do so, switch on the printer before the cal­ibration (Print LED lights up green). After a valid calibration, the protocol is printed.
CALIBRATION PROTOCOL
14.04.99 11:37 Device No.: 99990000 CALIBRATION CONDUCTIVITY Cal Time: 14.04.99 / 11:37 Cal Interval: 180d Cal Std.: 0.01 mol/l KCL
40.0 °C Conduct./Tref25: 1413µS/cm Cell Const : 0.975 1/cm Probe : +++
Operation
Calibration interval
(Int 3)
Setting the
calibration interval
The flashing sensor symbol reminds you to calibrate regu­larly. After the selected calibration interval (Int 3) expires, the sensor symbol flashes. Measurements can continue.
Note
To ensure the high measuring precision of the measuring system, perform a calibration after the calibration interval expires.
The calibration interval (Int 3) is set to 180 days in the fac­tory. The interval can be changed (1 ... 999 days):
1 Switch off the instrument. 2 Press < 3 Press the <on/off> key.
The The instrument then switches automatically to the configuration level.
ì> and hold down the key.
display test
appears briefly on the display.
4 Press the <run/enter> key until Int 3 appears on the
display.
5 Press <> < > to set the required time interval until
the next calibration. 6 Confirm with <run/enter>. 7 Change to the measuring mode by pressing <ì>.
109
Operation
Setting up the cell
constant manually
To set up the cell constant manually, proceed as follows:
1 Press the <CAL> key repeatedly until LF CELL ap-
pears on the display.
2 Press the <run/enter> key. 3 Press the <CAL> key repeatedly until the cell con-
stant to be set, e. g. 0.475 cm
-1
is displayed:
110
4 Set up the cell constant to be used by pressing
><>, e. g. 0.614 cm
<
-1
.
5 To return to the measuring mode: press the <ì> key.
Operation
Note
The cell constant to be set up must either be taken from the operating manual of the measuring cell or is printed on the measuring cell.
Selecting the
0.1 cm
-1
cell constant
-1
To select the 0.1 cm
cell constant proceed as follows:
1 Press the <CAL> key repeatedly until LF CELL ap-
pears on the display.
2 Press the <run/enter> key. 3 Press the <CAL> key repeatedly until the 0.100 cm
cell constant appears on the display.
-1
4 To return to the measuring mode: press the <ì> key.
111
Operation
Selecting the
0.01 cm
-1
cell constant
To select the 0.01 cm
-1
cell constant proceed as follows:
1 Press the <CAL> key repeatedly until LF CELL ap-
pears on the display.
2 Press the run/enter> key. 3 Press the <CAL> key repeatedly until the 0.010 cm
cell constant appears on the display.
-1
112
4 To return to the measuring mode: press the <ì> key.
Operation
Setting up the
temperature
compensation TC
ips Select the following temperature compensations to work
The calculation of the temperature compensation is based on the preset reference temperature, Tref 20 or Tref 25 (see chapter 4.6 C
ONFIGURATION).
You can select one of the following temperature compensa­tions:
l Non-linear temperature compensation "nLF"
according to DIN 38404 or EN 27 888
l linear temperature compensation "Lin" with a
coefficient that can be set in the range 0.001 ... 3.000 %/K
l no temperature compensation
Note
with the test samples given in the table:
Application
notes
Test sample Temperature
compensation TC
Natural water (ground water, surface water,
nLF according to DIN 38404 EN 27 888
drinking water) Ultrapure water nLF
according to DIN 38404 EN 27 888
Other aqueous solutions
Set linear temperature coefficient
0.001 ... 3.000 %/K
Salinity (seawater)
Automatically nLF accord­ing to IOT
Display indicator
Sal,
113
Operation
Selecting the non-
linear temperature
compensation
To select the non-linear temperature compensation proceed as follows:
1 Press the <CAL> key repeatedly until LF tc appears
on the display.
2 Press the <run/enter> key. 3 Press the <CAL> key repeatedly until nLF appears
on the display.
114
4 To return to the measuring mode: press the <ì> key.
Operation
Selecting the linear
temperature
compensation
To select the linear temperature compensation proceed as follows:
1 Press the <CAL> key repeatedly until LF tc appears
on the display.
2 Press the <run/enter> key. 3 Press the <CAL> key repeatedly until the adjustable
linear temperature coefficient appears on the display.
4 Set up the temperature coefficient, e. g. 1.880 %/K
using <
> <>.
5 To return to the measuring mode: press the <ì> key.
115
Operation
Switching off the
temperature
compensation
To switch off the temperature compensation proceed as fol­lows:
1 Press the <CAL> key repeatedly until LF tc appears
on the display.
2 Press the <run/enter> key. 3 Press the <CAL> key repeatedly until the following
display appears.
116
4 The temperature compensation has been switched
off.
5 To return to the measuring mode: press the <
ì> key.
Operation

4.4 Storing

The conductivity meter has an internal data storage device. Up to 200 data records can be stored in it. A complete data record consists of:
l Memory location l Date l Time l Measured value l Temperature l I.D. number
You can transmit measured values (data records) to the data storage in 2 ways:
l Manual storage l Switching on the AutoStore function (Int 1), see page 119
4.4.1 Manual storage
You can transmit a measured value to the data storage as follows:
1 Press the <STO> key.
The current number of the next free memory location
appears on the display.
117
Operation
2 Confirm with <run/enter>.
The display changes to the input of the I.D. number.
3 Enter the required I.D. number (1 ... 999) by pressing
<
> <>.
4 Confirm with <run/enter>.
The instrument changes to the measuring mode.
*(+,, message This message appears if all 200 memory locations are full.
You have the following options: Store the current measured value.
The oldest measured value (memory
Press <run/enter>
location 1) is overwritten by this Return to the measuring mode without
Press any key
storing Output the data storage See section 4.4.3 Delete the data storage See section 4.4.4
118
Operation
4.4.2 Switching on AutoStore (Int 1)
The storage interval (Int 1) determines the time interval between automatic storage processes. After the time interval expires, the current data record is transmitted to the data storage and to the interface.
The storage interval (Int 1) is set to OFF in the factory. Thus, the
AutoStore
function is switched off. To switch the function on, set up a time interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min).
Note
If the
AutoStore
function is active, the setting of the data
transmission interval (int 2) is ineffective (see page 128).
Setting the storage
interval
1 Press and hold down the <run/enter> key. 2 Press the <STO> key. Int 1 appears on the display.
3 Press the <> <> keys to set up the required time
interval between storage processes.
4 Confirm with <run/enter>.
The number of free memory locations appears on the display.
119
Operation
5 As soon as all 200 memory locations are full, the Au-
toStore function is terminated (Int 1 = OFF). If too few storage locations are available for your measurements:
– backup the data storage (see page 121) and – clear the data stored (see page 127).
6 Confirm with <run/enter>.
The prompt for the I.D. number appears on the dis­play.
7 Press <> <> to set the required I.D. number. 8 Confirm with <run/enter>.
The instrument changes to the measuring mode and starts the measuring and storage procedure.
AutoStore
flashes on the display.
120
Note
The
AutoStore
function is interrupted if you perform other functions, e.g. output data storage. After completing the other function, the
AutoStore
function continues. However, as a result, gaps can occur in the recording of the measured values.
Operation
Switching off the
AutoStore
Outputting to the
display
Switch off the
l Setting the storage interval (Int 1) to OFF or l Switch the conductivity meter off and on again.
AutoStore
function by:
4.4.3 Outputting the data storage
The contents of the data storage can be output to the:
l display l interface
1 Press the <RCL> key repeatedly until
Sto disp
ap-
pears on the display.
2 Press the <run/enter> key.
A measured value appears on the display. The I. D. number of the data record appears for ap­prox. 2 s.
121
Operation
3 After 2 s the respective temperature of the data
record appears on the display. Stored data records are displayed together with the RCL display indica­tor.
You can perform the following activities: Display further parameters of the data
record (I.D. no., date, time, memory loca-
Press <run/enter>
tion) Advance one data record (memory location) Press < Go back one data record (memory location) Press <
> >
122
Operation
Note
If you want to find a specific parameter (e.g. date), proceed as follows:
1 Press the <RCL> key repeatedly until
Sto disp
pears on the display.
2 Press the <run/enter> key.
A measured value appears on the display. The I. D. number of the data record appears for ap­prox. 2 s.
3 After 2 s the respective temperature of the data
record appears on the display.
ap-
4 Select the parameter (e.g. date) by pressing
<run/enter>.
5 Press <
> or <> repeatedly until the required date
appears on the display. After approx. 2 s, the temperature of the displayed measured value appears.
123
Operation
Outputting to the
interface
1 Press the <RCL> key repeatedly until
pears on the display.
2 Press the <run/enter> key.
The protocol of the last calibration is transmitted to the printer/RS interface. In the meantime, appears on the display. Following the calibration protocol, the complete con­tents of the storage is transmitted to the interface.
Sto SEr
Sto CAL
ap-
Note You can cancel the transmission by pressing < <run/enter>.
ì> or
124
Operation
Sample printout
CALIBRATION PROTOCOL
14.04.99 11:37 Device No.: 99990000 CALIBRATION CONDUCTIVITY Cal Time: 14.04.99 / 11:37 Cal Interval: 180d Cal Std.: 0.01 mol/l KCL
23.0 °C Conduct./Tref25: 1413 µS/cm Cell Const : 0.975 1/cm Probe: +++
No. 1:
01.01.99 00:04
2.40 mS/cm 25 °C Tman nLF Tref25 C = 0.475 1/cm Ident : 1
No. 2:
10.01.99 10:09
2.40 mS/cm 25.3 °C Tauto nLF Tref25 C = 0.475 1/cm Ident : 1
No. 3:
12.01.99 01:48
2.40 mS/cm 21.6 °C Tauto nLF Tref25 C = 0.475 1/cm Ident : 1
The printout
contains:
l Calibration protocol:
– Date/time of the printout – Instrument number (Device No.) – Date/time of the calibration (Cal Time) – Calibration interval (Cal Interval) – Calibration standard 0.01 mol/l KCL (Cal Std.) [°C/F] – Conductivity of the calibration standard at 25 °C
(Conduct./Tref25) – Cell constant (Cell Const) – Probe evaluation (Probe)
125
Operation
l Contents of the measuring storage:
– Number of the storage location (No.) – Date/time of measurement – Measured value/unit (mS/cm) [°C/F] – Automatic/manual temperature measurement
(Tauto/Tman)
– Temperature compensation (nLF) – Tref – Cell constant (1/cm) – AutoRead function (AR) – I.D. number (Ident)
126
Operation
4.4.4 Clearing the storage
This function can erase the stored data records. 200 mem­ory locations will then become available again.
Note
Clear store
The
function only appears if data records have already been stored in the storage. Otherwise, the conduc­tivity meter automatically changes to the measuring mode.
In order to delete all the data records, proceed as follows:
1 Switch off the instrument. 2 Press and hold down the <STO> key. 3 Press the <on/off> key.
The
display test
appears briefly on the display.
4 Confirm the clearing process by pressing
<run/enter>. Pressing any other key stops the clearing process and the data records remain in the storage.
Note
The calibration data remain in the storage and can be called up via the calibration protocol.
127
Operation

4.5 Data transmission

You can use the following options to transmit data:
l One of the following options:
– The
AutoStore
periodically (
function (page 119) is used to
Int 1
storage interval) save measured
values internally and output them on the interface.
– The
data transmission interval (Int 2)
function is used to periodically output measured values to the interface (see below).
l The
Output data store
function (page 121) is used to out­put calibration data and stored measured values to the in­terface.
l The analog recorder output (page 130) is used to output
measured values as voltages.
4.5.1 Data transmission interval (Int 2)
The interval for the data transmission (Int 2) determines the time interval between automatic data transmissions. After the time interval expires, the current data record is transmit­ted to the printer/RS interface.
128
Note
The setting of the interval (Int 2) only has an effect when the storage interval (
AutoStore
function) is switched off.
Operation
Setting the
data transmission
interval
The interval is set to OFF in the factory. To start the data transmission, set up an interval (5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):
1 Press and hold down the <run/enter> key. 2 Press the <auto read> key.
Int 2 appears on the display.
3 Press <> <> to set up the required time interval
between storage processes.
4 Confirm with <run/enter>.
The instrument changes automatically to the measur­ing mode.
Note
If the
AutoStore
function is active, the data transmission is performed according to the setting of the storage interval (Int1). Set the storage interval (Int1) to OFF to activate the
data transmission
interval (Int2).
Note
You can also set the
Configuration
menu (see page 128).
data transmission
interval (Int2) in the
129
Operation
/
4.5.2 Recorder (analog output)
You can transmit the data to a recorder via the analog out­put. Connect the analog output to the recorder via the Z394 interface cable. The data output switches automatically to
recorder output
.
Socket assignment
Conductivity
resistance
1 Free 2 Plug coding 3 Ground 4 Analog output
(Internal resistance < 5 Ohm)
Note
Activate the analog output by connecting 2 and 3 or use the original cable.
The signal range of the analog output depends on the mea­sured parameter and the measuring range:
Measuring range Voltage Resolution
0.000 ... 1.999 µS/cm 0 ... 1999 mV 1 mV
0.00 ... 19.99 µS/cm 0 ... 1999 mV 1 mV
0.0 ... 199.9 µS/cm 0 ... 1999 mV 1 mV
130
Salinity
TDS
0 ... 1999 µS/cm 0 ... 1999 mV 1 mV
0.00 ... 19.99 mS/cm 0 ... 1999 mV 1 mV
0.0 ... 199.9 mS/cm 0 ... 1999 mV 1 mV 0 ... 500 mS/cm 0 ... 500 mV 1 mV
Measuring range Voltage Resolution
0 ... 70.0 0 ... 700 mV 1 mV
Measuring range Voltage Resolution
0 ... 1999 mg/l 0 ... 1999 mV 1 mV
Operation
4.5.3 PC/external printer (RS232 interface)
You can transmit data to a PC or an external printer via the RS232 interface. Connect the interface to the instrument via the Z395 cable (PC) or Z391 cable (external printer). The data output switches automatically to
RS232
.
Note
The RS232 interface is not galvanically isolated. If it is connected to an earthed PC/printer, measurements cannot be made in earthed media as this would give incor­rect results!
Socket assignment
Set up the following transmission data on the PC/printer:
Baud rate Selectable between:
1200, 2400, 4800, 9600 Handshake RTS/CTS + Xon/Xoff PC only: Parity None Data bits 8 Stop bits 1
1 CTS 2 RxD 3 Ground 4 TxD
131
Operation

4.6 Configuration

You can adapt the conductivity meter to your individual requirements. To do this, the following parameters can be called up/changed (the status on delivery is marked in bold):
Baud rate 1200, 2400, 4800. 9600 Data transmission inter-
val (Int 2)
Calibration interval (Int 3) 1 ... 180 ... 999 d AutoRange ARng yes, no TREF 25/TREF 20 t25 or t20 Measured values given
as conductivity or resi­stance values
Temperature unit °C or °F
Date/time as required
Note
You can leave the configuration menu at any time. Parameters that have already been changed are stored. To do this, press the <
OFF, 5 s, 10 s, 30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min
S/cm or M
ì> key.
132
1 Switch off the instrument. 2 Press and hold down the < 3 Press the <on/off> key.
The display test
The instrument then switches automatically to the setting of the baud rate.
appears briefly on the display.
ì> key.
Baud rate
Data transmission
interval
Operation
4 Set up the required baud rate by pressing <> <>. 5 Confirm with <run/enter>.
Int 2 appears on the display.
6 Set up the required time interval by pressing
<
><▼>.
7 Confirm with <run/enter>.
Int.3
appears on the display.
133
Operation
Calibration
interval
ARng (AutoRange)
8 Set up the required time interval by pressing
><>.
<
9 Confirm with <run/enter>.
ARng appears on the display.
10 With <> <> select between YES and no. 11 Confirm with <run/enter>.
t25 appears on the display.
134
Reference
temperature
Conductivity
Operation
12 With <> <> select between t25 and t20
.
13 Confirm with <run/enter>. LF appears on the display.
Temperature unit
14 With <> <> select between S/cm and MΩ. 15 Confirm with <run/enter>.
USE °C appears on the display.
16 Select between °C and °F by pressing <> <>. 17 Confirm with <run/enter>.
The date flashes on the display.
135
Operation
Date and time
18 Set today’s date by pressing <> <>. 19 Confirm with <run/enter>.
The date (month) flashes on the display. 20 Set the current month by pressing <> <>. 21 Confirm with <run/enter>.
The year appears on the display.
136
22 Set the current year by pressing <> <>. 23 Confirm with <run/enter>.
The hours flash on the display. 24 Set the current time by pressing <> <>. 25 Confirm with <run/enter>.
The minutes flash on the display. 26 Set the current time by pressing <> <>. 27 Confirm with <run/enter>.
The instrument changes automatically to the conduc-
tivity measuring mode.
Operation

4.7 Reset

You can reset (initialize) measuring and configuration parameters separately from one another.
Measuring
parameters
The following measuring parameters (Cond InI) are reset to the values they had on delivery:
Measuring mode Cell constant 0.475 cm
0.475 cm
Temperature
nLF
-1
(calibrated)
-1
(set up)
compensation Reference temperature Temperature coefficient of
2.000 %/K the linear temperature com­pensation
TDS factor 1.00
Note
When the measuring parameters are reset, the calibration data are lost. After the parameters have been reset, cali­brate!
Configuration
parameters
The following configuration parameters (InI) are reset to the values they had on delivery:
Baud rate 4800 Interval 1
OFF
(automatic storing) Interval 2
OFF
(for data transmission)
137
Operation
Resetting
measuring
parameters
1 Press and hold down the <run/enter> key. 2 Press the <CAL> key.
3 Use <> <> to toggle between no and yes.
yes: reset measuring parameters. no: retain settings.
Resetting
configuration
parameters
4 Confirm with <run/enter>.
The instrument changes to the configuration para­meters.
5 Toggle between no and yes by pressing <> <>.
yes: reset configuration parameters. no: retain settings.
6 Confirm with <run/enter>.
The instrument changes automatically to the conduc­tivity measuring mode.
138
Maintenance, cleaning, disposal

5 Maintenance, cleaning, disposal

5.1 Maintenance

The maintenance tasks are restricted to the following activi­ties:
l replacing the batteries and l replacing the roll of printer paper.
Note
See the relevant operating manual of the electrode for instructions on maintenance.
139
Maintenance, cleaning, disposal
5.1.1 Changing the batteries
1 Open the battery compartment (1) on the underside
2 Remove the four batteries from the battery compart-
3 Insert four new batteries (Type Mignon AA) into the
4 Close the battery compartment (1).
5 Set up the date and time according to chapter 3 COM-
of the instrument.
ment.
battery compartment.
The date (day) flashes on the display.
MISSIONING.
1
Caution
Make sure that the poles of the batteries are the right way round.
± signs in the battery compartment must correspond to
The
± signs on the batteries.
the Only use leakproof alkaline manganese batteries.
140
Maintenance, cleaning, disposal
5.1.2 Replacing the roll of printer paper
1 Switch off the printer (Print LED lights up red). 2 Open the lid of the printer (1). 3 Remove the empty roll of paper (2). 4 Insert the new roll of paper:
– Fold the leading edge of the paper inwards and
thread it into the printer (3) from underneath.
– Press the <feed> Print key (paper feed).
5 Close the lid of the printer (1).
1
2
3
Note
Only use original SCHOTT HQ thermal paper. Information on this is given in the SCHOTT general catalog.
141
Maintenance, cleaning, disposal

5.2 Cleaning

Occasionally wipe the outside of the measuring instrument with a damp, lint-free cloth. Disinfect the housing with iso­propanol as required.
Caution
The housing is made of synthetic material (ABS). Thus, avoid contact with acetone or similar detergents that contain solvents. Remove any splashes immediately.

5.3 Disposal

Packing The measuring instrument is sent out in a protective trans-
port packing. We recommend: Keep the packing material. It protects the instrument against damage during transport.
Batteries This note refers to the battery regulation that applies in the
Federal Republic of Germany. We would ask end-consum­ers in other countries to follow their local statutory provi­sions.
Note
In compliance with §14 of the B
ATTERY REGULATION, we
would like to point out that this instrument contains batteries. Batteries that have been removed must only be disposed of at the recycling facility set up for this purpose or via the retail outlet. It is illegal to dispose of them in household refuse.
Measuring
instrument
Dispose of the measuring instrument as electronic waste at an appropriate collection point. It is illegal to dispose of them in household refuse.
142
Error message,
What to do if...

6 What to do if...

Cause Remedy
I/
Error message,

LoBat
Instrument does not
react to keystroke
– Measuring cell not
connected
– Connect measuring
cell
– Cable broken – Replace electrode
Cause Remedy
– Measuring cell
contaminated
– Unsuitable calibration
solution
– Clean measuring cell;
if necessary, replace it
– Check calibration
solutions
Cause Remedy
– Batteries almost
depleted
– Replace batteries
(see section 5.1 M
AINTENANCE)
Cause Remedy
– Operating state
undefined or EMC electric stress unallowed
– Processor reset:
Press the <auto read> key and switch on instrument
*(
display
Cause Remedy
– Timeout of the interface – Check connected
instrument
143
What to do if...
Integrated printer
does not print
Cause Remedy
– Printer switched off – Switch on printer
Integrated printer
prints automatically
Print key <feed>
does not react
– No mains power supply
connected
– Interface cable
– Connect mains power
supply
– Unplug cable
connected
– No paper available – Replace roll of paper
AutoStore
switched on with a longer
function is
– Switch off
AutoStore
function (see page 121)
interval duration (Int 1)
Cause Remedy
AutoStore
(Int 1) or
transmission
function
Data
(Int 2) is
– Switch off functions
(see page 121/page 128)
switched on
Cause Remedy
– Printer is switched on – Switch off printer
Printer operating -
paper not
being printed
*(+,,
message
144
– Interface cable
– Unplug cable
connected
Cause Remedy
– Paper inserted with
wrong side upwards
– Turn the roll of paper
around and insert it with the other side upwards
Cause Remedy
– All 200 memory
locations are full
– Output data storage and
clear data storage (see page 127)
Lists

7Lists

This chapter provides additional information and orientation aids.
Abbreviations The list of abbreviations explains abbreviations that appear
on the display or when dealing with the instrument.
Specialist terms The glossary briefly explains the meaning of the specialist
terms. However, terms that should already be familiar to the target group are not described here.
Index The index helps you find the topics that you are looking for.
145
Lists
Abkürzungsverzeichnis
Conductivity value ABS Plastic housing AR AutoRead (drift control) ARng Automatic range switching
Measuring instrument measures with high-
est resolution AutoStore Automatic storing Baud Baud rate C Cell constant cm
-1
Cal Calibration CELL Cell constant disp Display
Output of the data storage on the display E3 Error message (see chapter 6 W
IF...)
Ident I. D. number InI Initialization
Resets individual basic functions to
the status they had on delivery Int Interval Lin Linear temperature compensation LoBat Low Battery
Batteries are almost empty
HAT TO DO
146
nLF Non-linear temperature compensation OFL Overflow
Display range exceeded Sal Salinity SELV Safety Extra Low Voltage
Sto CIr Clear storage Sto disp Output of the data storage on the display Sto Full Memory full Sto SEr Output of the data storage to the printer/
interface Tauto Automatic temperature measurement TC Temperature coefficient TDS Total dissolved solids TP Temperature Probe
Temperature measurement active Tref 20/T20 Reference temperature 20 °C
Lists
Tref 25/T25 Reference temperature 25 °C °C Temperature unit °Celsius °F Temperature unit Fahrenheit
147
Lists
Glossary
AutoRead Monitors the electrode drift and releases the measured
value only after the stability criterion has been reached. In this way, this procedure ensures the highest degree of pre­cision and reproducibility.
Calibration The cell constant is determined through calibration. To do
so, the conductivity measuring cell is immersed into a series of aqueous salt solutions with exactly known electric con­ductiviy. The relevant conductivity values are determined using the conductivity measuring instrument.
Cell constant Linear factor describing the geometrical dimensions of a
measuring cell and its electrodes.
l With the value of the cell constant you can roughly de-
scribe the application range of a conductivity measuring cell.
l The cell constant is determined through calibration in one
or several calibration solutions.
Conductivity The conductivity value is a sum parameter for the ion con-
centration of a test sample.
Control standard
solution
Solution with a known conductivity to determine or check the conductivity.
Drift control See AUTOREAD.
148
Reference
temperature
In order to compare values measured at different tempera­tures, the values have to be converted to a fixed tempera-
ture. This temperature is 25 °C, or, as an exception, 20 °C.
Lists
Resistance All substances (solids, liquids, or gases) with mobile charge
carriers like for example electrons or ions have a finite ohmic resistance, which means they have an electric conductance that can be measured or an electric conducitivtiy.
Salinity The salinity is a sum parameter especially for seawater;
it gives the salt content of seawater.
Temperature
compensation (TC)
Test sample The substance to be measured. It can be liquid or solid.
Total dissolved
solids (TDS)
The temperature has a very strong impact on the electrical conductivity. To be able to compare measured values, it is necessary to temper or convert the test sample to a refer­ence temperature.
Mass that remains of the substances dissolved in an aque­ous solution after a fixed filtering and drying procedure, as far as these substances are not volatile under the conditions of this procedure. The total dissolved solids refer to the vol­ume of the filtered aqueous sample used und are given in mg/l.
149
Lists
150
Index
Lists
A
analog output 130 authorized use 92 AutoRange 134 AutoRead 100 AutoStore 119
B
batteries
replacing 139 battery compartment 140 baud rate, setting 133
C
calibration 106
interval 109
protocol 108 calibration evaluation 108 cell constant 98, 106 changing the batteries 140 cleaning 142 commissioning 95 condition on delivery 137 conductivity 135 conductivity measurement 101 conductivity/resistance 130
D
data record 117 data transmission (interval) 129 date, setting 136 display 84 disposal 142 drift control 100
E
energy saving feature 97 error messages 143
I
initialize 137 interval
calibration (Int 3) 109 storing (Int 1) 119
K
keys 82
L
linear temperature compensation 113, 115 LoBat 143
M
maintenance 139 manual storing 117 measuring 99 measuring precision 109
N
no temperature compensation 113 non-linear temperature compensation 113, 114
O
operation 97 operational safety 92 output data storage 121
P
paper roll
replacing 139 place of the instrument 97 plug-in power supply
connecting 96 printing
calibration protocol 108
measured values 105
151
Lists
R
recorder output 130 reference temperature 135 replacing the roll of printer paper 141 reset 137 RS232 interface 131
S
safety 91 safety precautions 91 salinity 130 salinity measurement 102 sockets 84 storing 117
T
TDS 130 TDS (total dissolved solids) 103 temperature compensation 113
linear 115 non-linear 114 setting 113
switch off 116 temperature probe 99 temperature unit 135 time
setting 95 time, setting 136 total dissolved solids measurement 103 transmit data 128
W
What to do if... 143
152
Loading...