All details contained in this operating manual are valid data
at the time of going to press. However, SCHOTT may add
details for both technical and commercial reasons, or in order to fulfil legal requirements of different countries. This
does not affect the characteristics described.
Warranty
declaration
The designated instrument is covered by a warranty of three
years from the date of purchase.
The instrument warranty extends to manufacturing faults
that are determined within the period of warranty.
The warranty excludes components that are replaced during
maintenance such as batteries, etc.
The warranty claim extends to restoring the instrument to
readiness for use but not, however, to any further claim for
damages. Improper handling or unauthorized opening of the
instrument invalidates any warranty claim.
To ascertain the warranty liability, return the instrument and
proof of purchase together with the date of purchase freight
paid or prepaid.
The compact
pH measurements rapidly and reliably. The
vides the highest degree of operating comfort, reliability and
measuring safety for all applications.
The proven calibration procedures and special
AutoRead
CG 843
function support your work with the pH meter.
precision pH meter lets you perform
CG 843
3
2
1
pro-
1Keypad
2Display
3Sockets
Note
The measuring instrument can also be delivered as part of
a set.
Information on this and other accessories is available in the
SCHOTT general catalog.
73
Overview
1.1 Keypad
Switch measuring instrument on or off
<on/off>
Select measuring mode
<pH>
Call up calibration procedure
<CAL>
Activate/deactivate AutoRead
<auto read>
Store measured value
<STO>
Display/transmit measured values
<RCL>
Reduce values, scroll
▼>
<
Confirm inputs, start AutoRead
<run/enter>
74
Increase values, scroll
▲>
<
1.2 Display
1.3Sockets
Overview
Status line
Measured value display
Function and
temperature displ a y
12345
1Socket according to DIN 19262/BNC
2PT 1000/NTC30 connection for pH combined electro-
de with temperature probe
3Reference electrode connection
4RS 232 interface/analog output
5Connection for plug-in power supply (optional)
Caution
Only connect probes to the instrument that cannot feed excessive voltages or currents (> SELV and > circuit with current limiter).
Nearly all commercial electrodes – especially SCHOTT
electrodes – meet these requirements.
75
Overview
1.4 Declaration of conformity
29.03.99
SCHOTT Geräte GmbH
Im Langgewann 5
D 65719 Hofheim am Taunus
Deutschland, Germany, Allemagne
29. März, March 29rd, 29, Mars, 1999
AGQSF0000-A058-00/990329
76
Ambient
temperature
Overview
1.5 Technical data
Storage temperature- 25 °C ... + 65 °C
Operating temp.0 °C ... + 55 °C
batteries
RuntimeApprox. 3000 operating hours
Mains power
supply (option)
Plug-in power supply unit:
Connection max. overvoltage category II
Plug-in power supply (Euro plug):
Type no.: Z851
Order no.: 28 520 4897
FRIWO FW3288, 11.8134
Friwo Part No. 1816492
Input: 230V ~ / 50 Hz / 23 VA
Output: 6 V = / 1,8 A /10,8 VA
Plug-in power supply (US plug):
Type no.: Z852
Order no.: 28 520 4901
FRIWO FW3288, 11.8451
Friwo Part No. 1816493
Input: 120V ~ / 60 Hz / 21,5 VA
Output: 6 V = / 1,8 A
Plug-in power supply (UK plug):
Type no.: Z849
Order no.: 28 520 4975
FRIWO FW3288, 11.8453
Friwo Part No. 1770896
Input: 230V ~ / 50 Hz / 23 VA
Output: 6 V = / 1,8 A
Data storageRing store for 200 value pairs, pH/mV, temperature
Temperature
compensation
Automatic with Pt
1000/NTC (30 k
-5 ... 99.9 °C
W)
Manual input-20 ... 130 °C resolution 1K
79
Overview
Connectors
Input amplifier
Calibration modes
ElectrodesSocket acc. to DIN 19 262 or BNC
4-mm socket for reference electode
Temp. probe4-mm socket for temperature probe
Interface/
analog output
Bidirectional RS 232 interface or ana-
log output with autom. recognition of
the PC, printer, or recorder connected
Plug-in power
2 pole special Friwo
supply (option)
12
10
-12
W
A
Input impedance 10
Offset current
AutoCal TECWith Technical buffers, pH values
2.00; 4.00; 7.00; 10.00 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C
AutoCal DINWith standard buffers according to
DIN 19 266/NIST, pH values:
1.68; 4.01; 6.87; 9.18 at 25 °C;
buffers are temperature compensa-
ted in the range 0 ... 90 °C
ConCalWith any buffers; one and two-point
calibration, manual acceptance of
measured values and temp. input
80
Safety
2Safety
This operating manual contains basic instructions that you
must follow during the commissioning, operation and maintenance of the pH meter. Consequently, all responsible personnel must read this operating manual before working with
the instrument.
The operating manual must always be available within the
vicinity of the instrument.
Target groupThis measuring instrument was developed for use in the lab-
oratory.
Thus, we assume that, as a result of their professional training and experience, the operators will know the necessary
safety precautions to take when handling chemicals.
Symbols used
Caution
indicates instructions that have to be followed to prevent
damage to your instrument.
Warning
indicates instructions that have to be followed to protect
yourself and the instrument from dangerous electrical voltage.
Note
Indicates notes that draw your attention to special features.
Note
Indicates cross-references to other documents, e.g. application reports, operating manuals of electrodes, etc.
81
Safety
2.1Authorized use
This instrument is authorized exclusively for pH and Redox
measurements in the laboratory.
The technical specifications as given in the section 1.5
T
ECHNICALDATA must be observed. Only the operation and
running of the measuring instrument according to the instructions given in this operating manual is authorized.
Any other use is considered unauthorized.
2.2General safety instructions
This instrument is constructed and tested in compliance with
the EN 61010-1 safety regulations for electronic measuring
instruments.
It left the factory in a safe and secure technical condition.
Function and opera-
tional safety
The smooth functioning and operational safety of the instrument can only be guaranteed if the generally applicable
safety measures and the specific safety instructions in this
operating manual are followed.
The smooth functioning and operational safety of the instrument can only be guaranteed under the climatic conditions
specified in the section 1.5 T
ECHNICALDATA.
If the instrument was transported from a cold environment to
a warm environment, the formation of condensate can lead
to the faulty functioning of the instrument. In this event, wait
until the temperature of the instrument reaches room temperature before putting the instrument back into operation.
Caution
The instrument is only allowed to be opened by personnel
authorized by SCHOTT.
82
Safety
Safe operationIf safe operation is no longer possible, the instrument must
be taken out of service and secured against inadvertent operation.
Safe operation is no longer possible if:
l the instrument has been damaged in transport
l the instrument has been stored under adverse conditions
for a lengthy period of time
l the instrument is visibly damaged
l the instrument no longer operates as described in this
manual
If you are in doubt contact the supplier of the instrument.
Obligations of the
operator
The operator of this measuring instrument must ensure that
the following laws and guidelines are observed when using
dangerous substances:
l EEC directives for protective labor legislation
l National protective labor legislation
l Safety regulations
l Safety datasheets of the chemical manufacturer.
83
Safety
84
Commissioning
3Commissioning
Perform the following activities for the initial commissioning:
l Set the date and time
l Connect the plug-in power supply.
Setting the date
and time
1Press and hold down the <pH> key.
2Press the <on/off> key.
The
display test
appears briefly on the display.
The measuring instrument then switches automatically to the setting of the baud rate.
3Press the <run/enter> key repeatedly until the date
flashes on the display.
4Set today’s date by pressing <
▲> <▼>.
5Confirm with <run/enter>.
The date (month) flashes on the display.
6Set the current month by pressing <
▲> <▼>.
7Confirm with <run/enter>.
The year appears on the display.
8Set the current year by pressing <
▲> <▼>.
9Confirm with <run/enter>.
The hour field flashes on the display.
10Set the current time by pressing <
▲> <▼>
11Confirm with <run/enter>.
The minutes field flashes on the display.
12Set the current time by pressing <
▲> <▼>
13Confirm with <run/enter>.
The measuring instrument then switches to the
pH measuring mode automatically.
85
Commissioning
Connecting the
plug-in power
supply (optional)
The measuring instrument works battery-powered. It can,
however, also be supplied by the plug-in power supply
which is available as an accessory.
Warning
The line voltage on site must lie within the input voltage
range of the original plug-in power supply unit (see section
ECHNICALDATA).
1.5 T
Caution
Use original plug-in power supplies only (see section 1.5
T
ECHNICALDATA).
2
1
3
1Insert the plug (1) into the socket (2) of the pH meter.
2Connect the original plug-in power supply (3) to an
easily accessible mains socket.
Note
You can also perform measurements without a plug-in power supply.
86
Operation
4Operation
4.1Switch on the instrument
1Place the instrument on a flat surface and protect it
against intense light and heat.
2Press the <on/off> key.
display test
The
The instrument then switches automatically to the
previously selected measuring mode.
Note
The instrument has an energy saving feature to avoid
unnecessary battery depletion.
The energy saving feature switches the instrument off if no
key has been pressed for an hour.
The energy saving feature is not active:
appears briefly on the display.
l if the power is supplied by the plug-in power supply
l if the
l if the communication cable is connected
l if the recorder cable is connected
l if the printer cable is connected
AutoStore
function is active
(for external printers)
87
Operation
4.2Measuring
Preparatory
activities
Perform the following preparatory activities when you want
to measure:
1Connect the electrode to the instrument.
2Adjust the temperature of the buffer or test solutions
or measure the current temperature if the measurement is made without a temperature probe.
3Calibrate or check the instrument with the electrode
according to section 4.3 C
ALIBRATING.
4Select the measuring mode by pressing <pH>.
Note
Incorrect calibration of the pH electrode will result in incorrect measured values. Therefore, regularly perform calibration before measuring.
Caution
When connecting an earthed PC/printer, measurements
cannot be performed in earthed media as incorrect values
would result.
The RS232 interface is not galvanically isolated.
88
Operation
Temperature probeMeasurements can be performed with and without a temper-
ature probe. A connected temperature probe is indicated by
TP on the display.
Note
The pH meter automatically recognizes the type of the temperature probe used. As a result, you can connect electrodes with the NTC30 or Pt1000.
The temperature measurement is absolutely essential for a
reproducible pH measurement. If the measurement is made
without a temperature probe, proceed as follows:
1Determine the current temperature using an external
thermometer.
2Set up the temperature by pressing <
▲> <▼>.
Note
When calibrating without a temperature probe, set up the
current temperature of the respective buffer solution manually by pressing the <
▲> <▼> keys.
89
Operation
4.2.1 Measuring the pH value
1Perform the preparatory activities according to sec-
tion 4.2.
2Immerse the pH electrode into the test sample.
AutoRead AR
(Drift control)
3Press the <pH> key until
pH
appears in the status
display. The pH value appears on the display.
The
AutoRead
function (drift control) checks the stability of
the measurement signal. The stability has a considerable
effect on the reproducibility of the measured values.
For identical measurement conditions, the following criteria
apply:
l pH value: better than 0.02 (response time: > 30 s)
1Call up the pH measuring mode by pressing <pH>.
90
2Activate the AutoRead function by pressing
<auto read>.
The current measured value is frozen (Hold function).
3Start the AutoRead function by pressing <run/
enter>.
AR flashes on the display until a stable measured value is reached. This measured value is transmitted to
the RS interface.
4If necessary, start the next AutoRead measurement
by pressing <run/enter>.
5To cancel the AutoRead function:
Press the <auto read> key.
Operation
Note
The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
<run/enter>.
4.2.2 Measuring the Redox voltage
The pH meter can measure the Redox voltage (mV) of a
solution when connected with a Redox electrode, e.g. BlueLine 31 Rx.
1Perform the preparatory activities according to sec-
tion 4.2.
2Immerse the Redox electrode into the test sample.
3Press the <pH> key repeatedly until U appears in the
status line. The Redox voltage (mV) of the sample
appears on the display.
4Wait for a stable measured value.
Note
Redox electrodes are not calibrated. However, you can
check Redox electrodes using a test solution.
91
Operation
4.2.3 Transmitting measured values
You can transmit measured values (data records) in 3 ways:
l Switch on the data transmission (Int 2) (see 114)
– After expiry of the selected interval, the current data
record is sent to the RS interface.
l Switch on AutoStore (Int 1) (see section 4.4.2 SWITCHING
ON AUTOSTORE (INT 1) 108)
– After expiry of the selected interval, the current data
record is sent to the RS interface and in addition is
stored in the data store of the instrument
– AutoStore (Int 1) covers the
data transmission
interval
(Int 2).
l Press the <run/enter> key
This manually triggers a transmitting of the current measured values at any time - independently of the selected
intervals.
Note
If you connect a recorder (analog output), both the output to
the digital output are switched off.
92
Operation
4.3Calibrating
Why calibrate?pH electrodes age. This changes the asymmetry (zero
point) and slope of the pH electrode. As a result, an inexact
measured value is displayed. Calibration determines the
current values of the asymmetry and slope of the electrode
and they are stored in the instrument.
Thus, you should calibrate at regular intervals.
When to calibrate?
AutoCal TECis specially adapted to the Technical buffer solutions as a
AutoCal DINis specially adapted to permanently programmed buffer
l After connecting another electrode
l When the sensor symbol flashes:
– after expiry of the calibration interval
– after a voltage interruption, e.g. battery change
You can choose between 3 calibration procedures:
fully automatic two or three-point calibration. The buffer
solutions are automatically recognized by the instrument.
The following pH values are available:
2.00, 4.00, 7.00, 10.00.
solutions according to DIN 19266 as a fully automatic two or
three-point calibration. The buffer solutions are automatically recognized by the instrument.
The following pH values are available:
1.68, 4.01, 6.87, 9.18.
ConCalis the conventional two-point calibration with 2 freely
selectable buffer solutions or single-point calibration as the
rapid method.
AutoReadIn calibration using AutoCal TEC and AutoCal DIN, the
AutoRead
function is automatically activated.
The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
<run/enter>.
93
Operation
Calibration protocolThe calibration protocol contains the cal ibr ati on dat a of the
current calibration. You can call up the calibration protocol
by outputting the data storage (section 4.4.3).
Note
You can automatically print out a calibration protocol after
the calibration. To do so, connect a printer according to section section 4.5.3 before the calibration. After a valid calibration, the protocol is printed.
After the calibration, the instrument automatically evaluates
the current status. The asymmetry and slope are separately
evaluated. The worst evaluation appears on the display.
DisplayAsymmetry
[mV]
Slope
[mV/pH]
-15 ... +1558 ... 60.5
-20 ... +2057 ... 58
-25 ... +2556 ... 57
or
60.5 ... 61
-30 ... +3056 ... 50
or
61 ... 62
Preparatory
activities
Clean the electrode according to the electrode operating manual
Clear the fault according to
chapter 6 W
HATTODOIF...
< -30 or
> 30
< 50 or
> 62
1Switch on the instrument by pressing <on/off>.
2Connect the pH electrode to the instrument.
3Keep the buffer solutions ready.
4Adjust the temperature of the solutions and measure
the current temperature if the measurement is performed without a temperature probe.
95
Operation
4.3.1 Calibration interval (Int 3)
The flashing sensor symbol reminds you to calibrate regularly. After the selected calibration interval (Int 3) expires,
the sensor symbol flashes. Measurements can continue.
Note
To ensure the high measuring precision of the measuring
system, perform a calibration after the calibration interval
expires.
Setting the calibra-
tion interval
The calibration interval (Int 3) is set to 7 days in the factory.
The interval can be changed (1 ... 999 days):
1Switch off the instrument.
2Press <pH> and hold down the key.
3Press the <on/off> key.
The
display test
appears briefly on the display.
The instrument then switches automatically to the
configuration level.
Int 3
4Press the <run/enter> key until
appears on the
display.
96
5Press <▲> <▼> to set the required time interval until
the next calibration.
6Confirm with <run/enter>.
7Change to the measuring mode by pressing <pH>.
Operation
4.3.2 AutoCal TEC
Use any two or three of the Technical buffer solutions for this
procedure in increasing or decreasing order.
Note
Steps 2, 6 and 13 are not required if you use a temperature
probe.
1Press the <CAL> key repeatedly until
AutoCal TEC
function display appears.
Ct1
and the
2If necessary, set the temperature of the buffer solu-
tion by pressing <
▲> <▼>.
3Submerse the pH electrode in the first buffer solution
4Press the <run/enter> key.
AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,
Ct2
ap-
pears.
.
5Thoroughly rinse the electrode with distilled water.
97
Operation
6If necessary, set the temperature of the second buffer
solution by pressing <
▲> <▼>.
7Submerse the electrode in the second buffer solution.
8Press the <run/enter> key.
AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,
AR
disappears.
The sensor symbol shows the electrode evaluation
after the two-point calibration.
The value of the slope (mV/pH) appears on the display.
9Press the <run/enter> key.
The value of the asymmetry (mV) appears on the display.
10To return to the measuring mode: Press the <pH>
key or continue with three-point calibration.
98
Operation
Three-point
calibration
11Press the <run/enter> key.
Ct3
appears on the display.
12Thoroughly rinse the electrode with distilled water.
13If necessary, set the temperature of the third buffer
solution by pressing <
▲> <▼>.
14Submerse the electrode in the third buffer solution.
15Press the <run/enter> key.
AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,
AR
disappears.
The sensor symbol shows the electrode evaluation
after the three-point calibration.
The value of the slope (mV/pH) appears on the display.
16Press the <run/enter> key.
The value of the asymmetry (mV) appears on the display.
17To return to the measuring mode: Press the <pH>
key.
Note
You can also prematurely terminate the three-point calibration by pressing <pH>. The values of the two-point calibration for the slope and asymmetry are then stored.
99
Operation
4.3.3 AutoCal DIN
Use two or three different DIN buffer solutions for this procedure in increasing or decreasing order.
DIN buffer solutionpH
A1.68
C4.01
D6.87
F9.18
Note
Steps 2, 6 and 13 are not required if you use a temperature
probe.
1Press the <CAL> key repeatedly until
AutoCal DIN
function display appear.
Cd1
and the
2If necessary, set the temperature of the buffer solu-
tion by pressing <
▲> <▼>.
3Submerse the pH electrode in the first buffer solution.
4Press the <run/enter> key.
AR
flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,
Cd2
ap-
pears.
100
Operation
5Thoroughly rinse the electrode with distilled water.
6If necessary, set the temperature of the second buffer
solution by pressing <
▲> <▼>.
7Submerse the electrode in the second buffer solution.
8Press the <run/enter> key.
AR flashes on the display.
The electrode voltage (mV) appears on the display.
As soon as a stable value is recognized,
AR
disappears.
The sensor symbol shows the electrode evaluation
after the two-point calibration.
The value of the slope (mV/pH) appears on the display.
9Press the <run/enter> key.
The value of the asymmetry (mV) appears on the display.
10To return to the measuring mode: Press the <pH>
key or continue with the three-point calibration.
101
Operation
Three-point
calibration
11Press the <run/enter> key.
Cd3
appears on the display.
12Thoroughly rinse the electrode with distilled water.
13If necessary, set the temperature of the third buffer
solution by pressing <
▲> <▼>.
14Submerse the electrode in the third buffer solution.
15Press the <run/enter> key.
AR
flashes on the display.
The electrode voltage (mV) appears on the display.
AR
As soon as a stable value is recognized,
disappears.
The sensor symbol shows the electrode evaluation
after the three-point calibration.
The value of the slope (mV/pH) appears on the display.
16Press the <run/enter> key.
The value of the asymmetry appears on the display
(mV).
17To return to the measuring mode: Press the <pH>
key.
Note
You can also prematurely terminate the three-point calibration by pressing <pH>. The values of the two-point calibration for slope and asymmetry are then retained.
102
4.3.4 ConCal
Operation
Two-point
calibration
Use two buffer solutions for this procedure:
l pH 7.0 ± 0.5
l any other buffer solution
Note
Steps 2 and 9 are not required if you use a temperature
probe.
1Press the <CAL> key repeatedly until
ConCal
function display appear.
ASY
and the
2If necessary, set the temperature of the buffer solu-
tion by pressing <
▲> <▼>.
3Submerse the pH electrode in the first buffer solution
pH 7.0 ± 0.5.
4Press the <run/enter> key.
The measured pH value appears on the display.
5Set the nominal pH value of the buffer solution (at the
current temperature) by pressing the <
▲> <▼> keys.
6Press the <run/enter> key.
The value of the asymmetry (mV) and the sensor
symbol appear on the display.
7Press the <run/enter> key.
SLO
(pe) appears on the display.
103
Operation
8Thoroughly rinse the electrode with distilled water.
9If necessary, set the temperature of the second buffer
solution by pressing <
▲> <▼>.
10Submerse the electrode in the second buffer solution.
11Press the <run/enter> key.
The second measured pH value appears on the display.
12Set the nominal pH value of the second buffer solu-
tion (at the current temperature).
13Press the <run/enter> key.
The value of the slope (mV/pH) appears on the display.
The sensor symbol shows the evaluation of the electrode after the two-point calibration.
14Press the <run/enter> key.
The value of the asymmetry (mV) appears on the display again.
15To return to the measuring mode: Press the <pH>
key.
104
Operation
Single-point
calibration
Use a buffer solution in the range pH = 7.0 ± 0.5 for this procedure.
Note
Only the electrode asymmetry is determined in single-point
calibration. The slope of the last two-point calibration is
retained.
Note
Step 2 is not required if you use a temperature probe. The
TP message indicates an active temperature measurement.
1Press the <CAL> key repeatedly until
ConCal
function display appear.
ASY
and the
2Set the temperature of the buffer solution by pressing
▲> <▼>.
<
3Submerse the pH electrode in the buffer solution.
4Press the <run/enter> key.
The measured pH value appears on the display.
5Set the nominal pH value of the buffer solution (at the
current temperature) by pressing the <
▲> <▼> keys.
6Press the <run/enter> key. The value of the asym-
metry (mV) and the sensor symbol for the evaluation
of the electrode appears on the display.
7To return to the measuring mode: Press the <pH>
key.
105
Operation
4.4Storing
The pH meter has an internal data storage device. Up to 200
data records can be stored in it.
A complete data record consists of:
l Memory location
l Date
l Time
l Measured value
l Temperature
l Temperature measurement procedure
l I.D. number
You can transmit measured values (data records) to the
data storage in 2 ways:
l Manual storage
l Switching on the AutoStore function (Int 1).
4.4.1 Manual storage
You can transmit a measured value to the data storage as
follows:
1Press the <STO> key.
The current number of the next free memory location
appears on the display.
106
Operation
2Confirm with <run/enter>.
The display changes to the input of the I.D. number.
3Enter the required I.D. number (1 ... 999) by pressing
▲> <▼>.
<
4Confirm with <run/enter>.
The instrument changes to the measuring mode.
*(+,, messageThis message appears if all 200 memory locations are full.
You have the following options:
Store the current measured value.
The oldest measured value (memory
Press
<run/enter>
location 1) is overwritten by this
Return to the measuring mode without
Press any key
storing
Output the data storage See section
4.4.3
Delete the data storage See section
4.4.4
107
Operation
4.4.2 Switching on AutoStore (Int 1)
The storage interval (Int 1) determines the time interval
between automatic storage processes.
After the time interval expires, the current data record is
transmitted to the data storage and to the RS interface.
Setting the storage
interval
The storage interval (Int 1) is set to OFF in the factory.
Thus, the
AutoStore
function is switched off.
To switch the function on, set up a time interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):
1Press and hold down the <run/enter> key.
Int.1
2Press the <STO> key.
appears on the display.
3Press the <▲> <▼> keys to set up the required time
interval between storage processes.
108
4Confirm with <run/enter>.
The number of free memory locations appears on the
display.
Operation
5As soon as all 200 memory locations are full, the
AutoStore function is terminated (Int 1 = OFF).
If too few storage locations are available for your
measurements:
– backup the data storage (see 110) and
– clear the data stored (see 113).
6Confirm with <run/enter>.
The prompt for the I.D. number appears on the display.
Switching off the
AutoStore
7Press <▲> <▼> to set the required I.D. number.
8Confirm with <run/enter>.
The instrument changes to the pH measuring mode
and starts the measuring and storage procedure.
AutoStore
flashes on the display.
Note
The
AutoStore
function is interrupted if you perform other
functions, e.g. output data storage.
After completing the other function, the
AutoStore
function
continues. However, as a result, gaps can occur in the
recording of the measured values.
Switch off the
l Setting the storage interval (Int 1) to OFF or
l Switch the pH meter off and on again.
AutoStore
function by:
109
Operation
4.4.3 Outputting the data storage
The contents of the data storage can be output to the:
l display
l RS interface
Outputting to the
display
1Press the <RCL> key repeatedly until
pears on the display.
2Press the <run/enter> key.
A measured value appears on the display.
The memory location of the data store appears for
approx. 2 s. This is followed by the corresponding
temperature.
You can perform the following activities:
Sto disp
ap-
110
Display further parameters of the data
record (I.D. no., date, time, memory loca-
Press
<run/enter>
tion)
Advance one data record (memory location) Press <
Go back one data record (memory location)Press <
▲>
▼>
Note
If you want to find a specific parameter (e.g. date), proceed
as follows:
Operation
1Select the parameter (e.g. date) by pressing <run/
enter>.
Outputting to
RS interface
2Press <
▲> or <▼> repeatedly until the required date
appears on the display.
After approx. 2 s, the temperature of the displayed
measured value appears.
1Press the <RCL> key repeatedly until
Sto SEr
pears on the display.
ap-
2Press the <run/enter> key.
Sto CAL
appears on the display. The calibration protocol is transmitted to the RS interface.
Following the calibration protocol, the complete contents of the storage is transmitted to the RS interface.
Note
You can the cancel the transmission by pressing <pH> or
<run/enter>.
09.03.99 17:11
pH 10.01 24,7 °C
Tauto AR
Ident : 1
No. 3:
09.03.99 17:12
305 mV
Tauto
Ident : 13
The printout contains:
l calibration protocol
l storage contents
112
Operation
4.4.4 Clearing the storage
This function can erase the stored data records. 200 memory locations will then become available again.
Note
Clear store
The
function only appears if data records have
already been stored in the storage. Otherwise, the pH meter
automatically changes to the measuring mode.
In order to delete all the data records, proceed as follows:
1 Switch off the instrument.
2Press and hold down the <STO> key.
3Press the <on/off> key.
The
display test
appears briefly on the display.
4Confirm the clearing process by pressing <run/
enter>.
Pressing any other key stops the clearing process
and the data records remain in the storage.
Note
The calibration data remain in the storage and can be called
up via the calibration protocol.
113
Operation
4.5Data transmission
You can use the following options to transmit data:
l One of the following options:
– The
AutoStore
storage interval) save measured values internally
1
function (108) is used to periodically (
Int
and output them on the RS interface.
– The
data transmission interval(Int 2)
function is used to
periodically output measured values to a RS interface
(see below).
l The
Output data store
function (110) is used to output calibration data and stored measured values to the RS interface.
l The analog recorder output (116) is used to output mea-
sured values as voltages.
4.5.1 Data transmission interval (Int 2)
The interval for the data transmission (Int 2) is determined
by the time interval between automatic data transmissions.
After the time interval expires, the current data record is
transmitted to the RS interface.
114
Note
The setting of the interval (Int 2) only has an effect when the
storage interval (
AutoStore
function) is switched off.
Operation
Setting the
data transmission
interval
The interval is set to OFF in the factory.
To start the data transmission, set up an interval (5 s, 10 s,
30 s, 1 min, 5 min, 10 min, 15 min, 30 min, 60 min):
1Press and hold down the <run/enter> key.
2Press the <auto read> key.
Int 2
appears on the display.
3Press <▲> <▼> to set up the required time interval
between storage processes.
4Confirm with <run/enter>.
The instrument changes automatically to the pH
measuring mode.
Note
If the
AutoStore
function is active, the data transmission is
performed according to the setting of the storage interval
(Int1). Set the storage interval (Int 1) to OFF to activate the
data transmission
interval (Int 2).
Note
You can also set the
Configuration
menu (see 118).
data transmission
interval (Int 2) in the
115
Operation
4
4.5.2 Recorder (analog output)
You can transmit the data to a recorder via the analog output. Connect the analog output to the recorder via the Z394
RS interface cable.
The data output switches automatically to
recorder output
.
Socket assignment
1 Free
1
RS 232
REC
2
3
2 Plug coding
3 Ground
4 Analog output
low impedance electrode signal
corresponds to 50 ... 62 mV/pH
(internal resistance < 5 Ohm)
Note
Activate the analog output by connecting 2 and 3.
116
Operation
4
4.5.3 PC/external printer (RS232 interface)
You can transmit data to a PC or an external printer via the
RS232 interface.
Connect the RS interface to the instrument via the Z395
cable (PC) or Z391 cable (external printer).
The data output switches automatically to
RS232
.
Note
The RS232 interface is not galvanically isolated.
If it is connected to an earthed PC/printer, measurements
cannot be made in earthed media as this would give incorrect results!
Set up the following transmission data on the PC/printer:
Socket assignment
Baud rateSelectable between:
1200, 2400, 4800, 9600
HandshakeRTS/CTS + Xon/Xoff
PC only:
ParityNone
Data bits8
Stop bits1
1 CTS
1
RS 232
REC
2
3
2 RxD
3 Ground
4TxD
117
Operation
4.6Configuration
You can adapt the pH meter to your individual requirements.
To do this, the following parameters can be changed (the
status on delivery is marked in bold):
When the measuring parameters are reset, the calibration
data are lost. After the parameters have been reset, calibrate!
Configuration
parameters
The following configuration parameters (InI) are reset to the
values they had on delivery:
Baud rate4800
Interval 1
(automatic storing)OFF
Interval 2
(for data transmission)OFF
122
Resetting measur-
ing parameters
Operation
1Press and hold down the <run/enter> key.
2Press the <CAL> key.
3Use <▲> <▼> to toggle between no and yes.
yes: reset measuring parameters.
no: retain settings.
Resetting configura-
tion parameters
4Confirm with <run/enter>.
The instrument changes to the configuration param-
eters.
5Toggle between no and yes by pressing <▲> <▼>.
yes: reset configuration parameters.
no: retain settings.
6Confirm with <run/enter>.
The instrument changes automatically to the pH
measuring mode.
123
Operation
124
Maintenance, cleaning, disposal
5Maintenance, cleaning, disposal
5.1Maintenance
The measuring instrument is almost maintenance-free. The
only maintenance task is replacing the batteries:
1Open the battery compartment (1) on the underside
of the instrument.
2Remove the batteries from the battery compartment
3Insert four new batteries (Type Mignon AA) into the
battery compartment.
4Close the battery compartment (1).
The date (day) flashes on the display.
5Set up date/time accordi ng chapter 3 COMMISSIONING.
1
.
Caution
Make sure that the poles of the batteries are the right way
round. The
spond to the
Only use leakproof alkaline manganese batteries.
Note
See the relevant operating manual of the electrode for instructions on maintenance.
± signs in the battery compartment must corre-
± signs on the batteries.
125
Maintenance, cleaning, disposal
5.2Cleaning
Occasionally wipe the outside of the measuring instrument
with a damp, lint-free cloth. Disinfect the housing with isopropanol as required.
Caution
The housing is made of synthetic material (ABS). Thus,
avoid contact with acetone or similar detergents that contain
solvents. Remove any splashes immediately.
5.3Disposal
PackingThe measuring instrument is sent out in a protective trans-
port packing.
We recommend: Keep the packing material. It protects the
instrument against damage during transport.
BatteriesThis note refers to the battery regulation that applies in the
Federal Republic of Germany. We would ask end-consumers in other countries to follow their local statutory provisions.
Note
In compliance with §14 of the B
ATTERY REGULATION, we
would like to point out that this instrument contains batteries.
Batteries that have been removed must only be disposed of
at the recycling facility set up for this purpose or via the retail
outlet.
It is illegal to dispose of them in household refuse.
Measuring
instrument
Dispose of the measuring instrument as electronic waste at
an appropriate collection point. It is illegal to dispose of them
in household refuse.
126
Error message,
What to do if...
6What to do if...
CauseRemedy
I/
Error message,
pH electrode:
– Not connected– Connect electrode
– Air bubbles in front of the
Electrode + sample:
– Conductivity too low– Use suitable electrode
– Temperature too high– Use suitable electrode
– Organic liquids– Use suitable electrode
CauseRemedy
– Calibrating interval
expired
– Newly calibrate measuring
system
128
CauseRemedy
What to do if...
LoBat display
*( display
Obviously incorrect
measured values
– Batteries almost
depleted
– Replace batteries
(see section 5.1
MAINTENANCE)
CauseRemedy
– Timeout of the interface– Checkout connected
instrument
CauseRemedy
– pH electrode unsuitable– Use suitable electrode
– Temperature difference
between buffer and
– Adjust temperature of
buffers or samples
sample too large
– Measuring procedure not
– Follow special procedure
suitable
Instrument does not
react to keystroke
*(+,, message
CauseRemedy
– Operating state
undefined or EMC
electric stress unallowed
– Processor reset:
Press the <auto read>
key and switch on
instrument
CauseRemedy
– All 200 memory locations
are full
– Output data store and
clear data store
129
What to do if...
130
Lists
7Lists
This chapter provides additional information and orientation
aids.
AbbreviationsThe list of abbreviations explains abbreviations that appear
on the display or when dealing with the instrument.
Specialist termsThe glossary briefly explains the meaning of the specialist
terms. However, terms that should already be familiar to the
target group are not described here.
IndexThe index helps you find the topics that you are looking for.
131
Lists
Abbreviations
ARAutoRead (drift control)
ARngAutomatic range switching
Measuring instrument measures with high-
est resolution
ASYAsymmetry
AutoCal DINAutomatic calibration with DIN buffer solu-
tions
AutoCal TECAutomatic calibration with Technical buffer
solutions
CalCalibration
Cd...Calibration with DIN buffer solutions (acc.
to DIN 19266)
ConCalConventional one / tw o po in t ca l ibration
Ct...Calibration with Technical buffer solutions
dispDisplay
Displays the data storage on the screen
E3Error message (see chapter 6 W
IF...)
HATTODO
InIInitialization
Resets individual basic functions to
the status they had on delivery
LoBatLow Battery
Batteries are almost empty
mVVoltage unit
mV/pHUnit of the electrode slope
132
OFLOverflow
Display range exceeded
pHpH value
RCLRecall
memory dump
SSlope
SELVSafety Extra Low Voltage
SErSerial interface
Output of the data storage on the RS 232
or on the internal printer
SLOSlope
Slope setting on calibration
StoStore
Memory
TPTemperature probe
Temperature measurement active
Lists
U
ASYAsymmetry potential
°CTemperature unit, °Celsius
°FTemperature unit, Fahrenheit
133
Lists
Glossary
AsymmetryZero point of a pH electrode.
ResolutionNumber of decimal places that appear for a measured
value.
AutoReadMonitors the electrode drift and releases the measured val-
ue only after the stability criterion has been reached. In this
way, this procedure ensures the highest degree of precision
and reproducibility.
Baud rateTransmission rate in bits/s.
DiaphragmContact point between the reference electrolytic solution
and the sample.
Drift controlSee A
UTOREAD.
Test sampleSample to be measured (can be liquid or solid).
Test solutionStable solution with a precisely known Redox voltage.
Buffer solutionStable solution with a precisely known pH value.
Redox voltagePotentiometric quantity.
SlopeSpecifies the voltage change per pH unit.
134
Index
Lists
A
analog output 116
asymmetry 93
authorized use 82
AutoCal DIN 93, 100
AutoCal TEC 93, 97
AutoRead 90
B
battery compartment 125
baud rate
setting
119
C
calibrating 93
calibration
evaluation
interval 96
procedures 93
protocol 94
ConCal 93, 103
condition on delivery 122
95
D
data record 106
data transmission (interval) 115
date
setting
display 75
drift control 90
85, 120
calibration
data transmission 114
storing 108
96
K
keys 74
L
LoBat 129
M
measuring precision 96
O
operational safety 82
P
place of the instrument 87
plug-in power supply 86
power supply 79, 86
printing
calibration protocol
measured values 92
printing measured values 92
94
R
Redox electrode 91
Redox voltage 91
replacing the batteries 125
reset 122
RS232 interface 117