Welch Allyn Spot Vital Signs LXi Service manual

Welch Allyn Spot Vital Signs LXi
SYS
mmg
DIA
mmg
TEMP
SpO2
Patient ID 0 2 4
Weight Height Respiration
4 WEST
7
21
7
9
8
AM
Reading
01
3
F
5
PULSE
/min
%
62
Pain Clear Send/Next Reading
Enter
Service Manual
BP
SYS
mmg
DIA
mmg
TEMP
SpO2
PatientID 0 2 4
Weight Height Respiration
4 WEST
7
21
7
9
8
AM
Reading
01
3
F
5
PULSE
/min
%
62
Pain Clear Send/Next Reading
Enter
BRAUn
Welch Allyn Spot Vital Signs LXi
Service Manual
iv Welch Allyn Spot Vital Signs LXi
Copyright 2012 Welch Allyn. All rights are reserved. To support the intended use of the product described in this publication, the purchaser of the product is permitted to copy this publication, for internal distribution only, from the media provided by Welch Allyn. No other use, reproduction, or distribution of this publication, or any part of it, is permitted without written permission from Welch Allyn.Welch Allyn assumes no responsibility for any injury to anyone, or for any illegal or improper use of the product, that may result from failure to use this product in accordance with the instructions, cautions, warnings, or statement of intended use published in this manual.
Welch Allyn, Spot Vital Signs, SureBP Technology, and SureTemp are registered trademarks of Welch Allyn.
Braun ThermoScan is a registered trademark of the Braun Company.
LNCS is a trademark of, and SET, LNOP, and Masimo are registered trademarks of, Masimo Corporation. Possession or purchase of a Masimo SpO2-equipped device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone or in combination with this device, fall within the scope of one or more of the patents relating to this device.
Nellcor and Oxi-Max are registered trademarks of Nellcor Puritan Bennett Inc.
Health o meter is a registered trademark of Sunbeam Products, Inc. used under license.
Software in this product is Copyright Welch Allyn or its vendors. All rights are reserved. The software is protected by United States of America copyright laws and international treaty provisions applicable worldwide. Under such laws, the licensee is entitled to use the copy of the software incorporated with this instrument as intended in the operation of the product in which it is embedded. The software may not be copied, decompiled, reverse-engineered, disassembled, or otherwise reduced to human-perceivable form. This is not a sale of the software or any copy of the software; all right, title, and ownership of the software remain with Welch Allyn or its vendors.
For information about any Welch Allyn product, call Welch Allyn Technical Support:
USA +1 800 289 2501 Australia +61 2 9638 3000
Canada +1 800 561 8797 China +86 21 6327 9631
European Call Center +353 46 90 67790 France +33 15 5695 849
Germany +49 695 098 5132 Japan +81 3 6383 0852
Latin America +1 305 669 9003 Netherlands +31 202 061 360
Singapore +65 6419 8100 South Africa +27 11 777 7555
United Kingdom +44 207 365 6780 Sweden +46 85 853 6551
Reorder #4500-89
Manual Material No. 718440 Ver. A
Welch Allyn 4341 State Street Road Skaneateles Falls, NY 13153 USA
Printed in USA
Regulatory Affairs Representative Welch Allyn Limited Navan Business Park Dublin Road Navan, County Meath, Republic of Ireland
Service Manual v
1 - Safety Summary .........................................1
Introduction......................................................1
Warnings and Cautions.............................................1
General Warnings..............................................1
Blood Pressure Warnings........................................3
Temperature Warnings..........................................4
SpO2Warnings................................................5
General Cautions ..............................................6
Blood Pressure Cautions ........................................7
Temperature Cautions ..........................................7
SpO2Cautions ................................................7
Electrostatic Discharge (ESD) ........................................8
Symbols ........................................................9
Safety Symbols ...............................................9
Button Symbols ...............................................9
Connection Symbols ...........................................9
Agency Symbols .............................................10
2 - Overview ..............................................11
Purpose and Scope...............................................11
Other Applicable Documents .......................................11
Miscellaneous Mounting Accessories ................................12
Wall Mount Kit ...............................................12
Mobile Stand Kit .................................................14
Radio Mounting Accessory.........................................16
Barcode Reader Mounting Accessory ................................17
Warranty .......................................................18
Spot Vital Signs LXi ...........................................18
Accessories .................................................18
Service ........................................................19
User Serviceable Parts .........................................19
Technical Assistance ..........................................19
Field Replacement Units .......................................19
Service Loaners ..............................................19
Service Intervals ..............................................19
Spot Vital Signs LXi Configurations...................................20
Controls .......................................................20
Display Window .................................................22
Connections ....................................................24
Braun ThermoScan PRO 4000 Lock...............................25
Blood Pressure Hose and Cuff...................................25
Thermometer ................................................26
SpO2Sensor ................................................27
Radio ......................................................27
Quick Reference Card .........................................27
AC Power Transformer.........................................28
Printer......................................................28
Battery .....................................................29
vi Welch Allyn Spot Vital Signs LXi
3 - Functional Overview .....................................31
Self Test .......................................................31
Internal Configuration Mode ........................................32
4 - Spot Vital Signs LXi Technical Overview ....................37
System Description ..............................................37
Battery System ..............................................37
Battery Charger ..............................................37
“CPU I/O” Power Supply .......................................37
Standby Mode ...............................................38
CPU Core Power Supply .......................................38
+5V Supply ..................................................38
Clock/Calendar Power .........................................38
Mod F NIBP Power ...........................................38
SpO2and Thermometer Power ..................................38
LCDPower..................................................38
External Load (RS-232) Power ...................................39
SureTemp Plus Thermometer Interface............................39
Braun ThermoScan PRO 4000 Thermometer Interface ................39
SpO2Description ................................................39
Radio Description ................................................40
Bar Code Scanner Description ......................................40
Interface ....................................................40
Printer Description ...............................................41
Interconnect Diagram .............................................42
5 - Maintenance and Service .................................43
Cleaning .......................................................43
Spot Vital Signs LXi ...........................................43
Blood Pressure Cuff ...........................................43
Blood Pressure Hose and Cable..................................43
SureTemp Plus Thermometer ...................................44
Braun ThermoScan PRO 4000 Thermometer .......................44
SpO2Sensors................................................45
Radio ......................................................45
Printer......................................................45
Battery Replacement .............................................46
Spot Vital Signs LXi ...........................................46
Braun ThermoScan PRO 4000 ...................................46
Printer......................................................47
Spot Vital Signs LXi Firmware Upgrade CD ............................48
Spot Vital Signs LXi Radio Configuration CD ...........................48
Product Disposal.................................................48
6 - Calibration and Verification ...............................49
Terms .........................................................49
Required Test Equipment ..........................................50
Pneumatic test assembly setup .....................................50
Power Supply Test Setup ..........................................52
Service Manual vii
Battery Removal / Power Shutdown .................................53
Voltage Calibration ...............................................55
Pressure verification test ..........................................57
Blood pressure calibration .........................................62
NIBP board Initialization ........................................62
Functional tests .................................................67
Temperature Functional Check......................................77
Temperature test setup ........................................77
Changing the Calibration Set Point Temperature on the 9600 Plus Calibration
Tester ......................................................78
Welch Allyn SureTemp Plus Thermometers............................78
Braun ThermoScan Pro 4000 ....................................82
Changing the Calibration Set Point Temperature .....................85
7 - Specifications ..........................................89
Performance ....................................................89
Blood Pressure Accuracy .......................................89
Temperature Specifications .....................................90
SpO2Specifications ..............................................91
Masimo Sensor Accuracy Guide .................................91
Masimo Patents ..............................................91
Nellcor Sensor Accuracy Guide ..................................92
Nellcor Patents ...............................................92
Mechanical .....................................................93
Electrical .......................................................93
Patents ........................................................93
Batteries .......................................................93
Environmental...................................................93
Radio..........................................................94
Mechanical ..................................................94
Configuration Options .........................................94
Guidance and Manufacturer’s Declaration For Spot Vital Signs LXi Radio .....95
8 - Troubleshooting ........................................99
Error Codes.....................................................99
Radio Test.....................................................110
Service Work Checklist...........................................111
Spot LXi Radio Configuration Tool ..................................112
9 - Supplies and Accessories................................113
Blood Pressure .................................................113
Temperature ...................................................114
Pulse Oximetry .................................................115
Masimo Accessories .........................................115
Nellcor Accessories ..........................................117
Miscellaneous..................................................118
Service Contracts ...............................................119
viii Welch Allyn Spot Vital Signs LXi
10 - Field Replaceable Units ................................121
11 - Disassembly and Repair ................................127
Tool list .......................................................127
Key Pad Disassembly ............................................132
SpO2Circuit Board Disassembly ...................................133
Masimo ...................................................133
Nellcor ....................................................134
Blood Pressure Circuit Board Disassembly ...........................136
LCD Inverter Ballast Board Disassembly .............................137
LCD Disassembly ...............................................137
Pump Disassembly ..............................................138
RS232-Communication Circuit Board Disassembly .....................138
Fan and Power Circuit Board Disassembly............................139
Thermometry Circuit Board Disassembly .............................139
SureTemp Plus ..............................................139
Braun ThermoScan PRO 4000 ..................................141
12 - Repair Test Specifications ..............................143
General Unit Test ...............................................143
A-D Noise Test ..............................................143
Leak Test ..................................................143
Inflation Test ...............................................143
Dump Test .................................................143
Pneumatic Accuracy Test......................................143
Valve Control Test ...........................................144
Voltage Calibration ...........................................144
Blank Mode Current Test ......................................144
Back Light (Idle) Current Test...................................144
Valve/Pump Mode Current Test .................................144
Interface Test ...............................................144
Temperature Option Requirements .................................144
Accuracy Test...............................................144
Temperature Probe Test ......................................145
SpO2Option Requirements .......................................145
SpO2Functional Test .........................................145
SpO2Mode Current Test ......................................145
Fail Safe Test ..................................................145
Over Pressure Test ..........................................145
Over 15 mmHg .............................................145
Interconnect Diagram 1 ..........................................147
Interconnect Diagram 2 ..........................................148
1
1

Safety Summary

Introduction

All users of the Spot Vital Signs LXi must read and understand the safety summary, and all additional specific warnings and cautions located throughout the documentation.

Warnings and Cautions

Familiarize all operating personnel with the general safety information in this summary. Specific warnings and cautions are also found throughout this manual.

General Warnings

A warning statement in this manual identifies a condition or practice, which if not corrected or discontinued immediately, could lead to patient injury, illness, or death.
These warnings pertain entirely to Spot Vital Signs LXi.
WARNING The information in this manual is a comprehensive guide to the operation of Spot Vital Signs LXi. For best results, read this manual thoroughly before using the device.
WARNING Spot Vital Signs LXi is designed for medical clinician use. Although this manual may illustrate medical spot-check techniques, only a trained clinician who knows how to take and interpret a patient’s vital signs should use this device.
WARNING Spot Vital Signs LXi is not intended for use in environments that are without health care practitioner supervision.
WARNING Spot Vital Signs LXi is not intended for continuous monitoring. Do not leave the device unattended while taking measurements on a patient.
WARNING To ensure data integrity, save readings and clear the Spot Vital Signs
LXi display between patients.
WARNING The Spot Vital Signs LXi is not defibrillator proof.
WARNING Spot Vital Signs LXi is not intended for use during patient transport.
WARNING This device is not suitable for use in the presence of a flammable
anesthetic mixture with air or oxygen or nitrous oxide. An explosion may result.
2 Safety Summary Welch Allyn Spot Vital Signs LXi
WARNING To ensure patient safety, use only accessories and supplies (i.e., cuffs, hoses, temperature probes, SpO2sensors, etc.) recommended for or supplied with Spot Vital Signs LXi. Using unapproved accessories with Spot Vital Signs LXi can affect patient and/or operator safety.
WARNING Prevent fluid from entering any connectors on the device. Should this occur, dry the connectors with warm air. Check the accuracy of all operating functions.
WARNING Every three months, inspect the blood pressure cuff, SpO2cable, and other accessories for fraying or other damage. Replace as necessary.
WARNING Do not use Spot Vital Signs LXi on patients who are on heart/lung machines.
WARNING Electric shock hazard. There are no user-serviceable parts inside Spot Vital Signs LXi other than battery replacement (see “Battery Replacement” on page 46). Only an individual specifically trained and approved for the repair and/or verification of this specific Welch Allyn product may perform maintenance procedures specifically described in this manual. For service, refer the device to an Authorized Service Center.
WARNING This device is not intended for hand-held use during operation.
WARNING Do not autoclave.
WARNING This device complies with current required standards for
electromagnetic interference and should not present problems to other equipment or be affected by other devices. As a precaution, avoid using this device in close proximity to other equipment.
WARNING Welch Allyn is not responsible for the integrity of any mounting installation. Welch Allyn recommends that the customer contact their Biomedical Engineering Department or maintenance service to ensure professional installation for safety and reliability of any mounting accessory.
WARNING The Spot Vital Signs LXi consists of high-quality precision parts. Protect it from severe impact and shock. A qualified service technician must check any Spot Vital Signs LXi that is dropped or damaged for proper operation prior to further use. Do not use the Spot Vital Signs LXi if you notice any signs of damage. Contact the Welch Allyn Technical Support Department for assistance.
WARNING Do not use an SpO2finger clip sensor and a blood pressure cuff simultaneously on the same limb. Doing so may result in inaccurate pulse rate and perfusion readings.
WARNING All signal input and output (I/O) connectors are intended for connection of only devices complying with IEC 60601-1, or other IEC standards (for example, IEC 60950) as appropriate to the device. Connecting additional devices to the Spot LXi may increase leakage currents. To maintain operator and patient safety, it is necessary to consider the requirements of IEC 60601-1-1.
WARNING For proper patient electrical isolation, use only a Welch Allyn power supply (4500-101A) to charge Spot Vital Signs LXi.
WARNING Do not use the manufacturer supplied charger for the weight scale while it is attached to Spot LXi.
Service Manual Safety Summary 3

Blood Pressure Warnings

These warnings pertain to the Spot Vital Signs LXi blood pressure feature.
WARNING Spot Vital Signs LXi is not intended to measure BLOOD PRESSURE on neonatal patients. The AAMI SP10:2002 standard defines neonates as children 28 days or less of age if born at term (37 weeks gestation or more); otherwise up to 44 gestational weeks.
WARNING To ensure pediatric blood pressure accuracy and safety, the Child Reusable Two-Piece Blood Pressure Cuff (4500-01), Infant Durable One-Piece Cuff (5082-82-4MQ), and the Infant Disposable One-Piece Cuff (5082-92-4MQ) are the smallest cuffs approved for use with young children and infants. The child’s arm must fit within the range markings on the cuff.
WARNING Avoid compression of the blood pressure hose or cuff tubing of Spot Vital Signs LXi. This may cause system errors to occur in the device.
WARNING Patients who are experiencing moderate to severe arrhythmias may give inaccurate blood pressure measurements.
WARNING Spot Vital Signs LXi does not operate effectively on patients who are experiencing convulsions or tremors.
WARNING Use only Welch Allyn blood pressure cuffs and/or hoses. Using other manufacturers’ blood pressure cuffs and/or hoses may produce inaccurate blood pressure readings.
WARNING When several blood pressure measurements are taken on the same patient, regularly check the cuff site and extremity for possible ischemia, purpura, and/or neuropathy.
WARNING Do not place the cuff on any extremity that is used for intravenous infusions or any area where circulation is compromised.
WARNING Excessive cuff tightness may cause venous congestion and discoloration of the limb.
WARNING Wrapping the cuff too loosely (preventing proper inflation) may result in errors.
WARNING Do not change the connector(s) on the blood pressure cuff tubing of this device to luer type. Luer type connectors are commonly used in intravenous infusion systems. Using the luer connectors on blood pressure cuff tubing creates the risk that the blood pressure tubing could be mistakenly connected to a patient's intravenous line, resulting in the introduction of air into the patient's circulatory system.
4 Safety Summary Welch Allyn Spot Vital Signs LXi

Temperature Warnings

These warnings pertain to the Spot Vital Signs LXi temperature feature.
SureTemp®Plus
These warnings are specific to the SureTemp Plus thermometer option.
WARNING Use only Welch Allyn probe covers. Using other manufacturers’ probe covers or no probe cover may produce temperature measurement errors and/or inaccuracy.
WARNING Always use a probe cover whenever coming into contact with a patient.
WARNING Long-term continuous monitoring beyond three to five minutes is not recommended in any mode.
WARNING Oral/axillary probes (blue ejection button at top of probe) and blue removable probe wells are used for taking oral and axillary temperatures only. Rectal probes (red ejection button) and red removable probe wells are used for taking rectal temperatures only. Use of the probe at the wrong site will result in temperature errors. Use of the incorrect removable probe well could result in patient cross-contamination.
WARNING The thermometer connectors and probe are not waterproof. Do not immerse or drip fluids on these items. Should this occur, dry the connectors and probe with warm air. Check all functions for proper operation and accuracy.
WARNING Do not take an axillary temperature through patient’s clothing. Direct probe cover to skin contact is required.
WARNING Do not autoclave.
WARNING Use Welch Allyn single-use disposable probe covers to limit patient
cross-contamination.
WARNING Incorrect insertion of probe can cause bowel perforation.
WARNING Washing hands greatly reduces the risk of cross-contamination and
nosocomial infection.
WARNING To ensure optimal accuracy, always confirm that the correct mode is selected.
Service Manual Safety Summary 5
Braun Thermoscan PRO 4000
These warnings are specific to the Braun ThermoScan PRO 4000 thermometer option.
WARNING Keep the probe window clean, dry, and undamaged at all times to ensure accurate measurements. To protect the probe window, always keep the thermometer in the storage cover while transporting or when not in use.
WARNING Only use Braun ThermoScan probe covers with this thermometer. Using other manufacturer’s probe covers or no probe cover may produce temperature measurement errors and/or inaccuracies. If the thermometer is used without a probe cover attached, clean the lens (see “Braun ThermoScan PRO 4000 Thermometer” on page 44).
WARNING Do not autoclave.
WARNING The thermometer is not waterproof. Do not immerse or drip fluids
on it. Should this occur, dry the thermometer with warm air. Check for proper operation and accuracy.

SpO2Warnings

These warnings pertain to the Spot Vital Signs LXi SpO2feature.
WARNING Only use Spot Vital Signs LXi with Masimo or Nellcor SpO2option with Masimo or Nellcor brand sensors and accessories, respectively. Using the wrong or unapproved sensors or cables may cause improper performance.
WARNING The SpO2sensor and extension cables are intended for use only for pulse oximetry measurements. Do not attempt to connect these cables to a PC or any similar device.
WARNING Before using, carefully read the sensor Directions for Use, including all warnings, cautions, and instructions.
WARNING Do not use a damaged sensor or pulse oximetry cable or a sensor with exposed optical components.
WARNING Incorrect application or a long duration of use of an SpO2sensor may cause tissue damage. Inspect the sensor site periodically as directed in the sensor’s Directions for Use.
WARNING Certain ambient environmental conditions, sensor application errors, and certain patient conditions may affect SpO2readings and pulse signal.
WARNING Do not immerse the sensor or patient cables in water, solvents, or cleaning solutions (the sensors and connections are not waterproof). Do not use irradiation, steam, or ethylene oxide for sterilization.
WARNING The SpO2in the Spot Vital Signs LXi device is not intended for use as an apnea monitor.
WARNING Consider the SpO2an early warning device. As a trend toward patient deoxygenation is indicated, use laboratory instruments to analyze blood samples to completely understand the patient’s condition.
6 Safety Summary Welch Allyn Spot Vital Signs LXi
WARNING Tissue damage can be caused by incorrect application or duration of use of a Nellcor OxiMax sensor. Inspect the sensor site as directed in the sensor Directions for Use.
WARNING Do not use the sensors during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns. The MS board pulse oximeter may affect the MRI image, and the MRI unit may affect the accuracy of the oximetry measurements.
WARNING Carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING Failure to cover the Nellcor OxiMax sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.

General Cautions

A caution statement in this manual identifies a condition or practice, which if not corrected or discontinued immediately, could lead to equipment failure, equipment damage, or data loss.
These cautions pertain to the entire Spot Vital Signs LXi device.
Caution If the accuracy of any measurement is in question, check the patient’s vital sign(s) with an alternate method and then check to verify the device is functioning properly.
Caution Place the device on a secure surface or use one of the optional mounting accessories.
Caution Do not place fluids on or near the device.
Caution It is recommended that the device is used within stated operating
temperature ranges (see “Environmental” on page 93). The device will not meet its performance specifications if used outside these temperatures ranges.
Caution Always unplug the AC power transformer from the outlet before moving the mobile stand to a new location.
Caution The basket has a three-pound weight limit. Take care not to exceed this limit.
Caution When connecting a Health o meter®weight scale, only operate the scale with battery power. Remove and discard the scale's AC power cord.
Caution When using the Health o meter scale, remove the two hex nuts on the RS-232 cable, as supplied, before screwing the cable into the scale base.
Service Manual Safety Summary 7

Blood Pressure Cautions

These cautions pertain to the Spot Vital Signs LXi blood pressure feature.
Caution Minimize extremity and cuff motion during blood pressure readings.
Caution If the blood pressure cuff is not at heart level, note the difference in
reading due to the hydrostatic effect. Add the value of 1.80 mmHg (.2 kPa) to the displayed reading for every inch (2.5 cm) above heart level. Subtract the value of
1.80 mmHg (.2 kPa) from the displayed reading for every inch (2.5 cm) below heart level.
Caution Proper blood pressure cuff size and placement is essential to the accuracy of the blood pressure determination. See Reusable Two-Piece Cuff Measurements or Durable One-Piece Cuff Measurements of the Directions for Use for sizing information.
Caution The position and physiologic condition of the subject can affect a blood pressure reading.

Temperature Cautions

These cautions pertain to the Spot Vital Signs LXi temperature feature.

SpO2Cautions

These cautions pertain to the Spot Vital Signs LXi SpO2feature.
Caution The SureTemp Plus feature only operates with the probe well in place.
Caution Do not use alkaline batteries in the Braun ThermoScan PRO 4000.
Caution Biting the probe tip may result in damage to the probe.
Caution The pulse oximeter is calibrated to determine the percentage of
arterial oxygen saturation of functional hemoglobin. Significant levels of dysfunctional hemoglobin such as carboxyhemoglobin or methemoglobin may affect the accuracy of the measurement.
Caution Some sensors may not be appropriate for a particular patient. If at least 10 seconds of perfusion pulses cannot be observed for a given sensor, change sensor location or sensor type for perfusion to resume.
Caution Physiological conditions, medical procedures, or external agents that may interfere with the pulse oximeter’s ability to detect and display measurements include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents such as nail polish, dye, or pigmented cream.
Caution When selecting a sensor, consider the patient’s weight and activity level, the adequacy of perfusion, the available sensor sites, the need for sterility, and the anticipated duration of monitoring.
8 Safety Summary Welch Allyn Spot Vital Signs LXi

Electrostatic Discharge (ESD)

Electrostatic discharge is a sudden current flowing from a charged object to another object or to ground. Electrostatic charges can accumulate on common items such as foam drinking cups, cellophane tape, synthetic clothing, untreated foam packaging material, and untreated plastic bags and work folders, to name only a few.
Electronic components and assemblies, if not properly protected against ESD, can be permanently damaged or destroyed when near or in contact with electrostatically charged objects. When you handle components or assemblies that are not in protective bags and you are not sure whether they are static-sensitive, assume that they are static-sensitive and handle them accordingly.
Perform all service procedures in a static-protected environment. Always use techniques and equipment designed to protect personnel and equipment from electrostatic discharge.
Remove static-sensitive components and assemblies from their static-shielding bags only at static-safe workstations - a properly grounded table and grounded floor mat ­and only when you are wearing a grounded wrist strap (with a resistor of at least 1 megohm in series) or other grounding device.
Use only grounded tools when inserting, adjusting, or removing static-sensitive components and assemblies.
Remove or insert static-sensitive components and assemblies only with monitor power turned off.
Insert and seal static-sensitive components and assemblies into their original static­shielding bags before removing them from static-protected areas.
Always test your ground strap, bench mat, conductive work surface, and ground cord before removing components and assemblies from their protective bags and before beginning any disassembly or assembly procedures.
Service Manual Safety Summary 9

Symbols

The following symbols are associated with the Spot Vital Signs LXi.

Safety Symbols

Button Symbols

Identifies information within the manual to avoid injury.
Caution: consult accompanying documents
Handle with Care Transport Temperature
Storage Humidity Recycle
MAX
Class II Equipment Equipment is not protected against
Type BF Equipment On/Off
Recycle the product separate from other disposables, see “Product Disposal” on page 48.
Mode of Operation: Continuous DC Power In
Pb
IPXØ
Identifies information within the manual to avoid equipment failure.
Internally Powered, Lead Acid Battery
the ingress of liquid.
Non-ionizing radiation (RF transmitter)

Connection Symbols

Navigation Buttons Select
Blood Pressure Memory
MEM
Power On/Off
USB Connection Serial Port Connection
10 Safety Summary Welch Allyn Spot Vital Signs LXi

Agency Symbols

CONFORMS TO: UL STD 60601-1
IEC 60601-1
The CE mark on this product indicates that it has been tested to and conforms with the provisions noted within the 93/42/EEC Medical Device Directive.
0297
Regulatory Affairs Representative Welch Allyn Limited Navan Business Park Dublin Road Navan, County Meath, Republic of Ireland
11
2

Overview

Purpose and Scope

The Spot Vital Signs LXi Service Manual is intended as a reference for maintenance and repair to the field replaceable unit (FRU) level and are listed on page 121.
Note
This manual provides the technical qualified service person with troubleshooting information, repair procedures, and calibration and performance verification instructions. A technical overview of the Spot Vital Signs LXi subsystems is provided as an introduction to the device’s circuitry and pneumatics.
This manual is intended for the technical qualified service personnel. Service training classes on Welch Allyn’s products are available. Contact Welch Allyn Technical Support for information.
Opening the Spot Vital Signs device case requires recalibration of specific functions. This includes any internal components that were replaced. Calibration requires the use of the Welch Allyn Spot LXi Tool.

Other Applicable Documents

The Spot Vital Signs LXi Directions for Use manual is also available. Refer to this document for information other than maintenance and repair.
Welch Allyn 9600 Plus Calibration Tester Directions for Use - for all models.
Braun ThermoScan PRO 4000 User’s Guide - for models 450E0, 45NE0, 45ME0.
Masimo Directions for Use - for models 45MT0, 45ME0.
Nellcor Directions for Use - for models 45NT0, 45NE0.
MOUNT KIT
a
12 Overview Welch Allyn Spot Vital Signs LXi

Miscellaneous Mounting Accessories

Wall Mount Kit

2
EF: 4701-62
terial #: 711890 Rev. B
1
3
Attach bracket to wall before
assembling basket.
4
}
Welch Allyn is NOT responsible for the integrity of any wall mounting interface.
5
hardware to mount the device to a surface. recommends that the customer contact their Biomedical Engineering Department or maintenance service to ensure professional installation, safety, and reliability of any mounting accessory.
Ensure you are using the appropriate
Welch Allyn
Service Manual Overview 13
14 Overview Welch Allyn Spot Vital Signs LXi

Mobile Stand Kit

Service Manual Overview 15
16 Overview Welch Allyn Spot Vital Signs LXi

Radio Mounting Accessory

A
2
B
1
3
5
4
Service Manual Overview 17

Barcode Reader Mounting Accessory

18 Overview Welch Allyn Spot Vital Signs LXi

Warranty

Spot Vital Signs LXi

Welch Allyn warrants that the Spot Vital Signs LXi Radio products meet the labeled specifications of the products and will be free from defects in materials and workmanship that occur within one year after the date of purchase.
The date of purchase is: 1) the date specified in our records if you purchased the Product directly from us, 2) the date specified in the warranty registration card that we ask you to send to us, or 3) the date of purchase of product from the authorized Welch Allyn distributor as documented from a receipt from said distributor.
This warranty does not cover damage caused by: 1) handling during shipping, 2) use or maintenance contrary to labeled instructions, 3) alteration or repair by anyone not authorized by Welch Allyn, and 4) accidents.
If a Product or accessory covered by this warranty is determined to be defective because of defective materials, components, or workmanship, and the warranty claim is made within the warranty period described above, Welch Allyn will, at its discretion, repair or replace the defective Product or accessory free of charge.
You must obtain a service notification number from Welch Allyn to return your Product before you send it to Welch Allyn’s designated service center for repair. Contact Welch Allyn Technical Support.
THIS WARRANTY IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. WELCH ALLYN’S OBLIGATION UNDER THIS WARRANTY IS LIMITED TO REPAIR OR REPLACEMENT OF PRODUCTS CONTAINING A DEFECT. WELCH ALLYN IS NOT RESPONSIBLE FOR ANY INDIRECT OR CONSEQUENTIAL DAMAGES RESULTING FROM A PRODUCT DEFECT COVERED BY THE WARRANTY.

Accessories

Refer to the manufacturer’s Directions for Use for the Masimo and Nellcor finger sensor and cable warranty.
Refer to the Directions for Use provided with the Welch Allyn Blood Pressure Cuff for warranty information.
The SureTemp Plus probe is covered by a one-year warranty and the SureTemp Plus probe well is covered by a 90-day warranty against original defects in material and workmanship. Probe covers are intended for single-use only.
The Braun ThermoScan PRO 4000 is covered by a three-year warranty against original defects in material or workmanship.
The printer is covered by a one-year warranty against original defects in material or workmanship.
The barcode scanner is covered by a five-year warranty against original defects in material or workmanship.
Service Manual Overview 19

Service

Caution Unauthorized repairs will void the warranty.
All repairs on products under warranty must be completed by Welch Allyn or Welch Allyn certified service personnel in accordance with warranty guidelines. See the Welch Allyn Partners In Care™web site (http://www.welchallyn.com/promotions/services/default.htm) for a description of the different levels of services offered.

User Serviceable Parts

For user serviceable parts available see “Supplies and Accessories” on page 113.

Technical Assistance

If you have an equipment problem that you cannot resolve, call the Welch Allyn Service Center nearest you for assistance. Technical service telephone support is available on normal business days.
If you are advised to return a product to Welch Allyn for repair or routine maintenance, schedule the repair with the service center nearest you.
Before returning a product for repair, you must obtain authorization from Welch Allyn. Service personnel will give you a Service Notification number. Please note this number on the outside of your shipping box. Returns without a Service Notification number will not be accepted for delivery.

Field Replacement Units

See “Field Replaceable Units” on page 121.

Service Loaners

Service loaners or exchanges may be provided depending on the specific Partners In Care™agreement in place for the device. Loaners for products repaired while under the original warranty, may be provided free of charge and are shipped within 48 hours of notification of need.
For service repairs outside of warranty or outside of service agreement are available for a nominal charge and shipment is subject to availability. Loaners are shipped pre-paid; however, this charge is added to the service charges.

Service Intervals

Calibrate the Spot Vital Signs LXi on an annual basis for blood pressure, temperature, and SpO2 accuracy.
20 Overview Welch Allyn Spot Vital Signs LXi

Spot Vital Signs LXi Configurations

Table 1. Available Versions of Spot Vital Signs LXi
REF Description
450T0 SureBP with SureTemp Plus Thermometer
450E0 SureBP with Braun ThermoScan PRO 4000 Thermometer
45MT0 SureBP with Masimo SpO
45ME0 SureBP with Masimo SpO
45NT0 SureBP with Nellcor SpO
45NE0 SureBP with Nellcor SpO
and SureTemp Plus Thermometer
2
and Braun ThermoScan PRO 4000 Thermometer
2
and SureTemp Plus Thermometer
2
and Braun ThermoScan PRO 4000 Thermometer
2

Controls

Figure 1. Spot Vital Signs LXi Front Panel with SureTemp Plus Thermometer
SureTemp Plus thermometer:
remove the probe from the probe well to
BP
SYS
mm g
DIA
mm g
TEMP
SpO2
9
21
7
8
4 WEST
7
Reading
01
AM
3
F
5
PULSE
%
/min
take a temperature.
Removable probe well: store the
temperature probe here when not in use; remove to clean or replace to eliminate cross-contamination.
62
Patient ID 0 2 4
Navigation button: move throughout the options in the Navigation Window or to increase/decrease parameter units.
Weight Height Respiration
Pain Clear Send/Next Reading
Enter
Probe cover storage area:
storage space for one box of probe covers.
Select button:
accepts the current option.
Power button: controls
power to the device.
Charging LED: indicates when device is plugged in and charging the battery.
Blood Pressure Start/Stop button:
initiates a new blood pressure cycle. Pressing again aborts an active blood pressure measurement.
Memory button:
recalls up to 50 most recent readings.
Service Manual Overview 21
Figure 2. Spot Vital Signs LXi Front Panel with Braun ThermoScan PRO 4000 Thermometer
Braun ThermoScan PRO 4000 thermometer: store the probe
covers toward the back of the thermometer housing; remove the
BP
SYS
mm g
DIA
mm g
TEMP
SpO2
9
21
7
8
4 WEST
7
Reading
01
AM
3
F
5
PULSE
%
/min
thermometer to access the covers.
Navigation button: move throughout the options in the Navigation Window or to increase/decrease parameter units.
Power button: controls power to the device.
Patient ID 0 2 4
Weight Height Respiration
Charging LED: indicates when device is plugged in and charging the battery.
Pain Clear Send/Next Reading
62
Enter
BRAUn
Memory button: recalls up to 50 most recent measurements.
Blood Pressure Start/Stop button:
initiates a new blood pressure cycle. Pressing again aborts an active blood pressure measurement.
Select button:
accepts the current option.
Charging LED: indicates the thermometer battery is charging.
22 Overview Welch Allyn Spot Vital Signs LXi

Display Window

Note
Before using Spot Vital Signs LXi for the first time, you must program it using the initial configuration screen.
1. Press the Power button. The display window shows the initial configuration screen.
Figure 3. Initial Configuration Screen
Change Local Defaults Menu
Model No.: 4 NTO
Serial No. 200 040004
Language:
BP Units:
Temperature Units:
eight Units:
Weight Units:
Date Format:
Date:
Time Format:
Time:
Select
2. Use the Select button to access the options and accept the entries, and use the Navigation buttons to move through the menu.
3. The word “Exit” appears at the bottom of the list after you have programmed all items in the menu. You must program all items before you can start to use the device.
4. Scroll to Exit and press the Select button to save the entries.