Welch Allyn Spot LXI Repair Tool Firmware Upgrade Procedure User Manual

Spot Vital Signs LXi -
Technical Service Dept. 4619 Jordan Road
Welch Allyn, Inc.
DIR #80012398 Ver. D
Repair Tool
PO Box 187 Skaneateles Falls, NY 13153-0187, USA ATTN: Product Service Department
Firmware Upgrade Procedure
Spot Vital Signs LXi Firmware Upgrade

Setup

The Spot Vital Signs LXi Firmware Upgrade resets all user-defined Internal Configuration settings (e.g., language, bp units, temperature units, height units, weight units, date, time, etc.) to the factory defaults. It is recommended that prior to executing the Firmware Upgrade that you record the Spot Vital Signs LXi settings. After the Firmware Upgrade, re-set the Spot Vital Signs LXi settings manually or through the use of the repair tool.

Connecting the hardware

During the Firmware Upgrade, Welch Allyn recommends using a USB cable for a more efficient,
S
N O I T C
t
r
faster upgrade.
1. Connect the USB cable to the Spot Vital Signs LXi and the computer port.
Note
If you must use a serial cable or recover from a failed installation, please contact Welch
o
6
5 9 3 0 7
: T A M
I
I
I
Allyn Technical Support for further assistance.
2. Connect the AC power transformer to the Spot Vital Signs LXi and plug it in.

Operation

1. Press the Power button to turn the Spot Vital Signs LXi on.
2. Open the Firmware Upgrade Utility on the computer. Click Start>All Programs>Welch Allyn>Spot LXi Repair Software >Spot LXi
Repair.
3. Select To o l s >Options.
4. Select the USB0 port, and then click the Te s t button. If a successful communication message displays, click OK.
5. Click OK to return to the main screen.
6. Select Unit>Upgrade Firmware.
7. Click Upgrade, if enabled. If Current Version is the same as Loading Version, click Close. This returns you to the main screen.
C O LINK
M
P
M
OWE
R
Note
The progress bar indicates the upgrade is occurring and the remaining time to complete. When complete, the Status changes to Success! The device powers off upon completion and follow any computer prompts.
If the battery charger is left unplugged, a pop-up displays “Warning, Charger has to be plugged in to continue.” Verify the charger is plugged in and functional. If the charger is not plugged in or is not functional, a new popup displays “Charger is not plugged in, cannot continue.” Click OK, and abort the upgrade.
8. Power on the Spot Vital Signs LXi and it displays internal post errors. To continue select Accept. Spot Vital Signs LXi displays the Set Defaults screen. Reset any changed defaults and click Exit.
9. Power off and then power on the Spot Vital Signs LXi in the Internal Configuration Mode. To enter the Internal Configuration Mode, press and hold the Spot Vital Signs LXi Select and Power buttons for 5 seconds. The Spot Vital Signs LXi enters the Internal Configuration mode and the Configuration Menu displays. Press Select on Version Numbers to verify the version information is correct.
10. Turn the Spot Vital Signs LXi off and click Close.
11. Go to the Spot LXi repair software and click the printer icon. Print two copies of the report.
12. Return one signed document to the customer and mail the other signed document to the address listed above.
Figure 1. Firmware Upgrade Report Sample
--------------------------------------------------------------------------­UPGRADE FIRMWARE
Serial Number: 20050800001 / Model Number: 45NT0
---------------------------------------------------------------------------
---------------------------------------------------------------------------
--------------------------------------------------------------------------­Calibration Settin gs Prior To Firmware Upgrad e
--------------------------------------------------------------------------­Serial Number: 20050800001
Model Number: 45NT0
Software Versions:
Unit : 5.7
BP Module: 1.02
Temperature Mod ule: 1.2
SpO2 Module: 1.9.0.1
Pressure:
Zero: -96
Gain: 18871 Cal Signature: ATE2 Cal Date/Time: 11/22/2005 07:57:24 PM
Voltage:
Cal Factor: 16608 Cal Signature: ATE2 Cal Date/Time: 11/22/2005 07:56:43 PM
---------------------------------------------------------------------------
--------------------------------------------------------------------------­Filename: C:\SOS\Spot LXi Repair Tools\Applications\ultra_prod_5_7.mot
C:\Program Files\Welch Allyn\Spot LXi Repair Software\ultra_prod_6_0.mot
Highest address loaded 0xA3297 Erasing Sector 4 Erasing Sector 5 Erasing Sector 6 Erasing Sector 7 Erasing Sector 8 Erasing Sector 9 Erasing Sector 10 Erasing Sector 11 Erasing Sector 12 Erasing Sector 13 PROM CRCssent: 4278684E D738B593 Received CRCs 4278684E 4278684ED738B593 D738B593 CRC checksums good!
---------------------------------------------------------------------------
---------------------------------------------------------------------------
Calibration Settings Aft er Firmware Upgrade
---------------------------------------------------------------------------
Serial Number: 20050800001
Model N umber: 45NT 0
Software Vers ions:
•Unit : 5.7
BP Module: 1.02
Temperature Module: 1.2
SpO2 M odule: 1.9.0. 1
Pressure:
•Zero: -96
Gain: 18871
Cal Signature: ATE2
Cal Date/Time: 11/22/2005 07:57:24 PM
•Voltage:
Cal Factor: 166 08
Cal Signature: ATE2
Cal Date/Time: 11/22/2005 07:56:43 PM
The above checks are d esigned to:
1.) Verify that the device stored data was not lost.
2.) Verif y Firmware insta llation was comp leted and acc urate during th e upgrade proc ess.
The Devi ce should have Ca libration and O perational per formance ver ified prior to first use.
---------------------------------------------------------------------------
---------------------------------------------------------------------------
Signature Block
Welch Allyn Representative __________________ __________________ ______
Print ed Name S ignature D ate
Customer Representative __________________ __________________ ______
Print ed Name S ignature D ate
---------------------------------------------------------------------------
---------------------------------------------------------------------------