VEGA DIS61 User Manual

Ope
rating Instructions
VEGADIS 61
Document ID:
27720
Indic
and adjustment
ation

Contents

1 About this document
2 For your safety
3 Product description
4 Mounting
5 Connecting to the sensor
ts
1.1 Function. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.2 Target group . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.3 Symbolism used . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.1 Authorised personnel . . . . . . . . . . . . . . . . . . . . . . . .
2.2 Appropriate use . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.3 Warning about misuse . . . . . . . . . . . . . . . . . . . . . . .
2.4 General safety instructions . . . . . . . . . . . . . . . . . . . .
2.5 Safety label on the instrument . . . . . . . . . . . . . . . . . .
2.6 CE conformity . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2.7 NAMUR recommendations . . . . . . . . . . . . . . . . . . . .
2.8 Safety instructions for Ex areas . . . . . . . . . . . . . . . . .
2.9 Environmental instructions. . . . . . . . . . . . . . . . . . . . .
3.1 Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Principle of operation . . . . . . . . . . . . . . . . . . . . . . . .
3.3 Packaging, transport and storage . . . . . . . . . . . . . . .
4.1 General instructions . . . . . . . . . . . . . . . . . . . . . . . . .
4.2 Instructions for installation . . . . . . . . . . . . . . . . . . . . .
5.1 Preparing the connection . . . . . . . . . . . . . . . . . . . . .
5.2 Connection procedure. . . . . . . . . . . . . . . . . . . . . . . .
5.3 Wiring plan. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4 4 4
5 5 5 5 6 6 6 6 6
7 8
10
11 11
13 13 15
6 Set up with the indicating and adjustment module
6.1 Short description . . . . . . . . . . . . . . . . . . . . . . . . . . .
6.2 Insert indicating and adjustment module. . . . . . . . . . .
6.3 Adjustment system . . . . . . . . . . . . . . . . . . . . . . . . . .
6.4 Setup steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7 Maintenance and fault rectication
7.1 Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2 Remove interferences . . . . . . . . . . . . . . . . . . . . . . . .
7.3 Instrument repair . . . . . . . . . . . . . . . . . . . . . . . . . . .
8 Dismounting
8.1 Dismounting steps . . . . . . . . . . . . . . . . . . . . . . . . . .
8.2 Disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9 Supplement
9.1 Technical data . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
9.2 Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
2 VE
18 18 19 20
21 21 21
22 22
23 25
GADIS 61
27720-EN-120329
Contents
plementary documentation
Sup
tion:
Informa
Supplementary documents appropriate to the ordered version come with the delivery. You can nd them listed in chapter "Product
description".
Editing status: 2012-03-15
27720-EN-120329
VEGADIS 61 3
1 Abou
t this document
1 Abou
t this document

1.1 Function

This operating instructions manual provides all the information you need for mounting, connection and setup as well as important instructions for maintenance and fault rectication. Please read this information before putting the instrument into operation and keep this manual accessible in the immediate vicinity of the device.

1.2 Target group

This operating instructions manual is directed to trained qualied personnel. The contents of this manual should be made available to these personnel and put into practice by them.

1.3 Symbolism used

Inform
ation, tip, note
This symbol indicates helpful additional information.
Cauti
on: If this warning is ignored, faults or malfunctions can
result. Warning: If this warning is ignored, injury to persons and/or serious damage to the instrument can result. Danger: If this warning is ignored, serious injury to persons and/or destruction of the instrument can result.
applications
Ex
This symbol indicates special instructions for Ex applications.
l List
The dot set in front indicates a list with no implied sequence.
à Action
Th
is arrow indicates a single action.
1 Sequence
Numbers set in front indicate successive steps in a procedure.
4 VE
27720-EN-120329
GADIS 61
your safety
2 For
or your safety
2 F

2.1 Authorised personnel

All operations described in this operating instructions manual must be carried out only by trained specialist personnel authorised by the plant operator.
During work on and with the device the required personal protective equipment must always be worn.

2.2 Appropriate use

VEGADIS 61 is an external indicating and adjustment unit for VEGA
®
sensors.
plics
You can nd detailed information on the application range in chapter
"Product description".
Operational reliability is ensured only if the instrument is properly used according to the specications in the operating instructions manual as well as possible supplementary instructions.
For safety and warranty reasons, any invasive work on the device beyond that described in the operating instructions manual may be carried out only by personnel authorised by the manufacturer. Arbitrary conversions or modications are explicitly forbidden.

2.3 Warning about misuse

Inappropriate or incorrect use of the instrument can give rise to application-specic hazards, e.g. vessel overll or damage to system components through incorrect mounting or adjustment.

2.4 General safety instructions

This is a high-tech instrument requiring the strict observance of standard regulations and guidelines. The user must take note of the safety instructions in this operating instructions manual, the country­specic installation standards as well as all prevailing safety regulations and accident prevention rules.
The instrument must only be operated in a technically awless and reliable condition. The operator is responsible for trouble-free operation of the instrument.
During the entire duration of use, the user is obliged to determine the compliance of the necessary occupational safety measures with the current valid rules and regulations and also take note of new regulations.
27720-EN-120329
VEGADIS 61 5
2 For
your safety
2.5 Safety
The safety approval markings and safety tips on the device must be observed.
label on the instrument

2.6 CE conformity

The device fullls the legal requirements of the applicable EC guidelines. With the CE mark, we provide conrmation of successful testing.
You can nd the conformity ce rticate in the download section of our homepage.

2.7 NAMUR recommendations

NAMUR is the automation technology user association in the process industry in Germany. The published NAMUR recommendations are accepted as the standard in eld instrumentation.
The device fullls the requirements of the following NAMUR recommendations:
l NE 21 – Electromagnetic compatibility of equipment l NE 53 – Compatibility of eld devices and indicating/adjustment
components

2.8 Safety instructions for Ex areas

Please note the Ex-specic safety information for installation and operation in Ex areas. These safety instructions are part of the operating instructions manual and come with the Ex-approved instruments.

2.9 Environmental instructions

Protection of the environment is one of our most important duties. That is why we have introduced an environment management system with the goal of continuously improving company environmental protection.
The environment management system is certied according to DIN EN ISO 14001.
Please help us full this obligation by observing the environmental
instructions in this manual:
l Chapter "Packaging, transport and storage" l Chapter "Disposal"
6 VE
27720-EN-120329
GADIS 61
1
3
2
3 Produc
t description
pe of delivery
Sco
Constituent parts
3 Produc
t description

3.1 Structure

The scope of delivery encompasses:
l Indicating and adjustment unit VEGADIS 61 l Cable entry M20x1 for the sensor l Documentation
- this operating instructions manual
- Operating instructions manual 27835 "Indicating and adjust-
ment module PLICSCOM"
- Supplementary instructions manual 31708 "Heating for in-
dicating and adjustment module PLICSCOM" (optional)
- Ex-specic "Safety instructions" (with Ex versions)
- if necessary, further certicates
The VEGADIS 61 consists of the components:
l Mounting element (depending on order specication, mounting
plate, clip or strap)
l Housing with electronics l Housing cover with integrated indicating and adjustment module
The components are available in dierent versions.
Fig. 1: VEGADIS 61 with
1 Housing cover with integrated indicating and adjustment module 2 Housing with electronics 3 Mounting plate
plastic housing and mounting plate for wall mounting
27720-EN-120329
VEGADIS 61 7
2 3
1
5
6
4
7
11
9
10
8
3 Produc
t description
Type
label
e type label on the housing contains the most important data for
Th identication and use of the instrument:
Fig. 2: Structure
1 Instrument type 2 Product code 3 Approval 4 Voltage supply 5 Protection rating 6 Ambient temperature 7 Order number 8 Seriennummer des Gerätes 9 ID numbers, instrument documentation 10 Note to observe the instrument documentation 11 Notizierte Stelle für die CE-Kennzeichnung 12 Zulassungsrichtlinie
of the type label (example)
Seria
l number
With the serial number of the instrument on the type label you have access to the following data on our homepage:
l Article number of the instrument (HTML) l Delivery date (HTML) l Order-specic instrument features (HTML) l Operating instructions at the time of shipment (PDF)
Go to
www.vega.co
m, "Service" "VEGA Tools" and "serial number
search".

3.2 Principle of operation

The VEGADIS 61 is an external indicating and adjustment unit for VEGA plics
®
sensors. The device enables remote measured value
indication and adjustment of sensors at measuring sites that are dicult to access.
8 VE
27720-EN-120329
GADIS 61
2
4
1
3
2
5
4
2
4
1
3
2
5
4
6
cation examples
Appli
3 Produc
t description
Fig. 3: Connection and sensor adjustment are carried out via the integrated indicating and adjustment module
1 Voltage supply/Signal output sensor 2 Sensor 3 Connection cable VEGADIS 61 - Sensor 4 VEGADIS 61 5 Indicating/adjustment module
Fig. 4: Connection carried out via VEGACONNECT and PC with PACTware
1 Voltage supply/Signal output sensor 2 Sensor 3 Connection cable VEGADIS 61 - Sensor 4 VEGADIS 61 5 VEGACONNECT 6 PC with PACTware
of VEGADIS 61 to the sensor. The measured value indication
of VEGADIS 61 to the sensor. The sensor adjustment is
27720-EN-120329
VEGADIS 61 9
3 Produc
t description
Conne
ction
Packaging
Transport
Transport inspection
Storage
Storage and transport
temp
erature
e VEGADIS 61 is connected with a four-wire cable to the sensor.
Th Via this cable, the communication and power supply of VEGADIS 61 is
carried out through the sensor. A dditional external energy is not required. Permissible cable lengths, see chapter "Technical data".

3.3 Packaging, transport and storage

Your instrument was protected by packaging during transport. Its capacity to handle normal loads during transport is assured by a test according to DIN EN 24180.
The packaging of standard instruments consists of environment­friendly, recyclable cardboard. For special versions, PE foam or PE foil is also used. Dispose of the packaging material via specialised recycling companies.
Transport must be carried out under consideration of the notes on the transport packaging. Nonobservance of these instructions can cause damage to the device.
The delivery must be checked for completeness and possible transit damage immediately at receipt. Ascertained transit damage or concealed defects must be appropriately dealt with.
Up to the time of installation, the packages must be left closed and stored according to the orientation and storage markings on the outside.
Unless otherwise indicated, the packages must be stored only under the following conditions:
l Not in the open l Dry and dust free l Not exposed to corrosive media l Protected against solar radiation l Avoiding mechanical shock and vibration
l Storage and transport temperature see chapter "Supplement -
Technical data - Ambient conditions"
l Relative humidity 20 85 %
10 VE
27720-EN-120329
GADIS 61
62 mm (2.44")
12 mm (0.47")
7 mm
(0.28")
85 mm (3.35")
4 M
ounting
Mounting position
Moisture
4 Moun
ting

4.1 General instructions

VEGADIS 61 functions in any installation position. The housing can be rotated by 330° without the use of any tools.
Use the recommended cables (see chapter "Connecting to power supply") and tighten the cable gland.
You can give your instrument additional protection against moisture penetration by leading the connection cable downward in front of the cable entry. Rain and condensation water can thus drain off. This applies mainly to outdoor mounting as well as installation in areas where high humidity is expected (e.g. through cleaning processes) or on cooled or heated vessels.
Fig. 5: Measures
against moisture penetration

4.2 Instructions for installation

Wall

mounting

27720-EN-120329
VEGADIS 61 11
VEGADIS 61 for wall mounting is supplied with a mounting socket.
Fig. 6: VEGADIS 61 for
1 Drilling dimensions
wall mounting, bottom view of mounting plate.
1 2
132
4 M
ounting
r rail mounting
Carrie
Tube mounting
VEGADIS 61 for
mounting on carrier rail is supplied with a mounting
adapter.
Fig. 7: VEGADIS 61 for
1 Adapter plate with screw M4 x 6 2 Carrier rail
carrier rail mounting
VEGADIS 61 for tube mounting is supplied with a meas. instrument holder and four screws M5 x 12 as mounting accessory. The meas. instrument holder is mounted to the socket of VEGADIS 61.
Fig. 8: VEGADIS 61 for
1 4 screws M5 x 12 2 Measuring instrument holder 3 Tube
mounting on tubes up to 1" diameter
12 VE
27720-EN-120329
GADIS 61
ecting to the sensor
5 Conn
Safety instructions
4 20 mA/HART senso
PA/FF sensor
Cable screening and
groun
ding
5 Conn
ecting to the sensor

5.1 Preparing the connection

Always keep in mind the following safety instructions:
l Connect only in the complete absence of line voltage
VEGADIS 61 is connected to the sensor with standard four-wire cable
r
with screen.
Use cable with round cross-section. A cable outer diameter of 5 … 9 mm (0.2 0.35 in) ensures the seal eect of the cable gland. If you are
using cable with a dierent diameter or cross-section, exchange the seal or use a suitable cable gland.
Ti
p:
We recommend using the matching, ready-made connection cable from the VEGA product line.
VEGADIS 61 is connected with a screened four-wire special cable of up to 25 m length with connection socket M12 x 1 to the sensor.
Informa
tion:
The special cable is part of the scope of delivery of PA/FF sensors with connection plug M12 x 1 for VEGADIS 61.
This special cable is absolutely necessary for reliable functioning of VEGADIS 61 as well as the sensor.
Connect the cable screen on both ends to ground potential. In VEGADIS 61 and in the sensor, the screen must be connected directly
to the internal ground terminal. The ground terminal on the outside of the housing must be connected to the potential equalisation (low impedance).

5.2 Connection procedure

Proceed as follows:
1 Unscrew the housing cover 2 Remove the indicating and adjustment module by turning it slightly
to the left
3 Loosen compression nut of the cable entry 4 Remove approx. 10 cm of the cable mantle, strip approx. 1 cm
insulation from the individual wires
5 Insert the cable into VEGADIS 61 through the cable entry 6 Lift the opening levers of the terminals with a screwdriver (see
following illustration)
7 Insert the wire ends into the open terminals according to the wiring
plan
27720-EN-120329
VEGADIS 61 13
ecting to the sensor
5 Conn
8 Pres
s down the opening levers of the terminals, you will hear the
terminal spring closing
9 Check the hold of the wires in the terminals by lightly pulling on
them
10 Connect the screen to the ground terminal 11 Connect the ground terminal outside on the housing according to
specication (low impedance)
12 Tighten the compression nut of the cable entry. The seal ring must
completely encircle the cable
13 Screw the housing cover back on
The electrical connection is nished.
Fig. 9: Connection
steps 6 and 7
14 VE
27720-EN-120329
GADIS 61
3
2
4
Display
I2C
5 6 7 8
1
Display
5 6 7 8
4
1
2
ecting to the sensor
5 Conn
nics and connec-
Electro tion
compartment
Wir
ing plan 4 20 mA/
HART
5.3 Wir
Fig. 10: Electronics
1 Contact pins for the indicating and adjustment module 2 Spring-loaded terminals for connection to the sensor 3 Plug connector for interface converter 4 Ground terminal for connection of the cable screen
ing plan
and connection compartment
27720-EN-120329
VEGADIS 61 15
Fig. 11: Wiring
1 To the sensor 2 Ground terminal for connection of the cable screen
1)
Connect housing as prescribed. The two terminals are galvanically connected.
plan VEGADIS 61 for 4 20 mA/HART sensors
here. Connect ground terminal on the outside of the
screen
1)
Display
5 6 7 8
4
1
ecting to the sensor
5 Conn
Wiring plan PA/FF
ction via standard cable
Conne
Connection between VEGADIS 61 and sensor is carried out according to the chart:
VEGADIS 61 Sensor
Terminal 5 Terminal 5
Terminal 6 Terminal 6
Terminal 7 Terminal 7
Terminal 8 Terminal 8
Connection via ready-made cable with 4-pin socket M12 x 1
Connection between VEGADIS 61 and sensor with 4-pin plug M12 x 1 is carried out according to the following chart. The specication of the wire colours refers to the ready-made cable from our own product line
Wire colour VEGADIS 61
Brown Terminal 5
White Terminal 6
Blue Terminal 7
Black Terminal 8
2)
plan VEGADIS 61 for PA/FF sensors
Fig. 12: Wiring
1 To the sensor
Connection via ready-made cable with 5-pin socket M12 x 1
The connection between VEGADIS 61 and sensor with 5-pin plug M12 x 1 is carried out according to the following chart. The
specication of the wire colours refers to the ready-made special cable supplied with this sensor version.
2)
Plug M12 x 1, optionally
3)
Plug M12 x 1, optional with
3)
double chamber housing.
with
®
single and double chamber housing.
plics
16 VE
27720-EN-120329
GADIS 61
5 Conn
colour VEGADIS 61
Wire
Black Terminal 5
White Terminal 6
Blue Terminal 7
Brown Terminal 8
ecting to the sensor
27720-EN-120329
VEGADIS 61 17
up with the indicating and adjustment module
6 Set
Function/Con
Mount/Dismount
ting and adjustment mo­dule
guration
indica-
6 Set
up with the indicating and adjustment
module

6.1 Short description

The indicating and adjustment module is used for measured value display, adjustment and diagnosis. It can be mounted in the following housing versions and instruments:
l All continuously measuring sensors in single as well as double
chamber housing (optionally in the electronics or connection compartment)
l External indicating and adjustment unit
Note:
You
can nd detailed information on the adjustment in the operating
instructions manual "Indicating and adjustment module".

6.2 Insert indicating and adjustment module

The indicating and adjustment module can be inserted into VEGADIS
61 and removed again at any time. It is not necessary to interrupt the power supply.
For mounting proceed as follows:
1 Unscrew the housing cover 2 Place the indicating and adjustment module in the desired position
on the electronics (you can choose any one of four dierent positions - each displaced by 90°)
3 Press the indicating and adjustment module onto the electronics
and turn it to the right until it snaps in.
4 Screw housing cover with inspection window tightly back on
Removal is carried out in reverse order.
The indicating and adjustment module is powered by the sensor, an
additional connection is not necessary.
18 VE
27720-EN-120329
GADIS 61
2
3
1
1.1
up with the indicating and adjustment module
6 Set
Fig. 13: Installation
of the indicating and adjustment module

6.3 Adjustment system

Fig. 14: Indicating
1 LC display 2 Indication of the menu item number 3 Adjustment keys
and adjustment elements
27720-EN-120329
VEGADIS 61 19
up with the indicating and adjustment module
6 Set
Key functions
Adjustment system
] key:
l [OK
- Move to the menu overview
- Conrm selected menu
- Edit parameter
- Save value
l [->] key to select:
- Menu change
- Select list entry
- Select editing position
l [+] key:
- Change value of the parameter
l [ESC] key:
- interrupt input
- Return to higher-ranking menu
The sensor is adjusted via the four keys of the indicating and adjustment module. The LC display indicates the individual menu items. The functions of the individual keys are shown in the above illustration. Approx. 10 minutes after the last pressing of a key, an automatic reset to measured value indication is triggered. Any values not conrmed with [OK] will not be saved.

6.4 Setup steps

Setup and adjustment of the respective sensor is carried out according to the operating instructions manual of the respective sensor.
20 VE
27720-EN-120329
GADIS 61
aintenance and fault rectication
7 M
ction when malfunc-
Rea tions
occur
24 hour service hotline
Reaction after fault rec­ti
cation
7 Maint
enance and fault rectication

7.1 Maintenance

If the instrument is used properly, no special maintenance is required in normal operation.

7.2 Remove interferences

The operator of the system is responsible for taking suitable measures to rectify faults.
Should these measures not be successful, please call in urgent cases the VEGA service hotline under the phone no. +49 1805 858550.
The hotline is available to you 7 days a week round-the-clock. Since we oer this service world-wide, the support is only available in the English language. The service is free of charge, only the standard telephone costs will be charged.
Depending on the reason for the fault and the measures taken, the steps described in chapter "Set up" may have to be carried out again.

7.3 Instrument repair

If a repair is necessary, please proceed as follows:
You can download a return form (23 KB) from our homepage at
vega.com
By doing this you help us carry out the repair quickly and without having to call back for needed information.
under: "Downloads - Forms and certicates - Repair form".
www.
l Print and ll out one form per instrument l Clean the instrument and pack it damage-proof
l Attach the completed form and, if need be, also a safety data
sheet outside on the packaging
l Please ask the agency serving you for the address of your return
shipment. You can nd the respective agency on our website www.vega.com
27720-EN-120329
VEGADIS 61 21
under: "Company - VEGA worldwide"
8 Dismoun
ting
8 Dismou
nting

8.1 Dismounting steps

Warning:
Before dismounting, be aware of dangerous process conditions such as e.g. pressure in the vessel, high temperatures, corrosive or toxic products etc.
Take note of chapters "Mounting" and "Connecting to power supply" and carry out the listed steps in reverse order.

8.2 Disposal

The instrument consists of materials which can be recycled by specialised recycling companies. We use recyclable materials and have designed the electronics to be easily separable.
WEEE directive 2002/96/EG
This instrument is not subject to the WEEE directive 2002/96/EG and the respective national laws. Pass the instrument directly on to a specialised recycling company and do not use the municipal collecting points. These may be used only for privately used products according to the WEEE directive.
Correct disposal avoids negative eects on humans and the environ­ment and ensures recycling of useful raw materials.
Materials: see chapter "Technical data"
If you have no way to dispose of the old instrument properly, please
contact us concerning return and disposal.
22 VE
27720-EN-120329
GADIS 61
9 Sup
plement
9 Supp
lement

9.1 Technical data

General data
316L corresponds to 1.4404 or 1.4435, 316Ti corresponds to 1.4571
Materials VEGADIS 61
- Housing plastic PBT, Alu die-casting, 316L
- Inspection window in housing cover for
indicating and adjustment module
- Ground terminal 316Ti/316L
-
Materials mounting elements
- Mounting plate Aluminium, 316L
- Mounting adapter PVDF
- Measuring instrument holder 316Ti/316L
-
Weight without mounting element 0.35 kg (0.772 lbs)
Ambient conditions
Ambient temperature -15 +70 °C (+5 +158 °F) Storage and transport temperature -40 +80 °C (-40 +176 °F)
Electromechanical data
Cable gland 1 x cable gland M20 x 1.5 (cable: ø 5 … 9 mm), 1 x
Spring-loaded terminals for wire cross-sec-
tion up to
Polycarbonate (UL-746-C listed)
blind stopper M20 x 1.5 or 1 x closing cap ½ NPT,
1 x blind stopper ½ NPT
2.5 mm² (AWG 14)
Indicating and adjustment module
Voltage supply and data transmission through sensor via gold-plated sliding contacts (I²C
Indication LC display in dot matrix
Measured value indication
- Number of digits 5
- Size of digits (sensors with software
4.0.0, hardware 2.0.0)
- Size of digits (sensors with software
3.99, hardware < 2.0.0)
Adjustment elements 4 keys
Protection rating
- unassembled IP 20
27720-EN-120329
VEGADIS 61 23
bus)
W x H = 7 x 13 mm
W x H = 5 x 8 mm
9 Suppl
ement
- mounte
Materials
- Housing ABS
- Inspection window Polyester foil
Adjustment circuit
Connection to the respective sensor Data transmission digital (I²C-Bus) Conguration connection cable 4-wire, screened
Cable length max.
- Sensors with software 4.0.0, hardware
- Sensors with software 3.99, hard-
Electrical protective measures
Protection rating
- Housing plastic IP 66/IP 67
- Housing Aluminium, stainless steel IP 66/IP 68 (0.2 bar)
Overvoltage category III
Protection class II
Approvals
Instruments with approvals can have dierent technical data depending on the version.
That's why the associated approval documents have to be noted with these instruments. They are
part of the delivery or can be downloaded under number search" as well as via "Downloads" and "Approvals".
d into VEGADIS 61 without cover IP 40
2.0.0
ware< 2.0.0
50 m
25 m
www.vega.co
m via "VEGA Tools" and "serial
24 VE
27720-EN-120329
GADIS 61
imensions
1 2 3
4
123 mm (4.84")
128 mm (5.04")
112 mm (4.41")
125 mm (4.92")
~ 69 mm
(2.72")
ø 77 mm
(3.03")
~ 69 mm
(2.72")
~ 59 mm
(2.32")
~ 116 mm (4.57")
ø 77 mm
(3.03")
ø 80 mm
(3.15")
ø 84 mm (3.31")
M20x1,5
M20x1,5/ ½ NP
T
M20x1,5/ ½ NPT
M20x1,5/ ½ NPT
M20x1,5/ ½ NPT
9.2 D
VEGADIS 61 housing
9 Sup
plement
27720-EN-120329
Fig. 15: VEGADIS 61 housing
1 Plastic housing 2 Stainless steel housing - precision casting 3 Stainless steel housing, electropolished 4 Aluminium housing
VEGADIS 61 25
versions
5mm
(
13
/
64
")
16mm
(
5
/
8
")
84mm (3
5
/
16
")
70mm (2
3
/
4
")
R32mm
(1
17
/
64
")
M6
67mm (2
41
/
64
")
~ 111mm (4
31
/
8
")
3mm
(
1
/
8
")
11mm
(
7
/
16
")
60mm (2
23
/
64
")
1 2 3
9 Suppl
ement
VEGADIS 61 mounting
Fig. 16: VEGADIS 61 mounting
elements
elements
1 Mounting plate for wall mounting 2 Clip for carrier rail mounting 3 Strap for tube mounting
26 VE
GADIS 61
27720-EN-120329
9 Sup
plement
9.3 Indus
VEGA product lines are global protected by industrial property rights. Further information see http://www.vega.com. Only in U.S.A.: Further information see patent label at the sensor
housing.
VEGA Produktfamilien sind weltweit geschützt durch gewerbliche Schutzrechte.
Nähere Informationen unter http://www.vega.com.
Les lignes de produits VEGA sont globalement protégées par des droits de propriété intellectuelle.
Pour plus d'informations, on pourra se référer au site http://www.vega.
com.
VEGA lineas de productos están protegidas por los derechos en el campo de la propiedad industrial.
Para mayor información revise la pagina web http://www.vega.com.
Линии продукции фирмы ВЕГА защищаются по всему миру правами на интеллектуальную собственность.
Дальнейшую информацию смотрите на сайте http://www.vega.com.
VEGA 站<http://www.vega.com>
trial property rights

9.4 Trademark

All the brands as well as trade and company names used are property of their lawful proprietor/originator.
27720-EN-120329
VEGADIS 61 27
VEGA Grieshaber KG
ISO 9001
Am Hohenstein 113 77761 Schiltach
Germany Phone +49 7836 50-0 Fax +49 7836 50-201 E-mail: info.de@vega.com
www.vega.com
Printing date:
statements concerning scope of delivery, application,
All
practical use and operating conditions of the sensors and
processing systems correspond to the information avail-
able at the time of printing.
© VEGA Grieshaber KG, Schiltach/Germany 2012
Subject to change without prior notice 27720-EN-120329
Loading...