The following pictograms are used in these
directions for use or on the device itself:
Read the directions for use!
Please observe the warning and
safety instructions!
Risk of explosion!
V
mA
2015-01
6 GB
Direct current (type of current and voltage)
Milliampere
Danger to life and danger of accidents
for infants and children!
Medical device type BF
Year and month of manufacture
Dispose of the packaging and appliance
in an environmentally friendly way!
Forehead & Ear Thermometer
Introduction
Congratulations on the purchase of your new
appliance. You have clearly decided in favour
of a quality product.
These operating instructions are a part of this product.
They contain important information in regard to safety,
use and disposal. Before starting the device up for the fi rst
time, familiarise yourself with its functions. Read the following directions for use. Please keep these instructions in a
safe place. When handing the device on to a third party,
hand all instructions on, too.
Intended use
The device is intended for indoor use for measuring the
body temperature at the ear or temple/forehead, measuring the room temperature and measuring surface temperatures in the household. In no way does it replace an examination by a qualifi ed doctor. This appliance is solely
intended for private use. Any use other than that previously described or any change to the product is not permitted
and may lead to injuries and / or damage to the product.
The manufacturer does not assume any liability for damage arising from improper use.
7 GB
Scope of delivery
1 x Forehead and ear thermometer
(incl. battery and protective head guard
1 x Directions for use
Description of parts
Measuring head
1
“EAR” button
2
(Temperature measurement in the ear / scan mode)
“FOREHEAD” button
3
(Temperature measurement at forehead / temple)
LC display
4
Red LED: Fever 38.0 °C - 42.9 °C
5
(100.4 °F - 109.2 °F)
Orange LED: slightly elevated temperature
6
37.5 °C - 37.9 °C (99.5 °F - 100.3 °F)
Green LED: no fever 34.0 °C - 37.4 °C
7
(93.2 °F - 99.4 °F)
Protective cap
8
“MEM / SET” button (Interrogate memory etc.)
9
Battery compartment lid
10
Battery
11
)
8
8 GB
LC display
Battery change symbol
12
Year display symbol
13
Date display symbol
14
Ear mode symbol
15
Standard mode symbol
16
Scan mode symbol
17
Temperature display
18
Forehead mode symbol
19
Cleaning symbol
20
Date and time in standard mode
21
Sleep mode symbol
22
PM display
23
Setup mode symbol
24
Memory mode symbol
25
Technical data
Measurement
method: Infrared (non-contact)
Electrical
power supply: Lithium battery 3 V
CR2032
Power consumption: < 15 mA
Memory: 9 temperature measurements with
details of date, time and measurement
method (ear or forehead measurement)
9 GB
Temperature scale: °C (°Celsius)
°F (°Fahrenheit)
Time display: 12 / 24-hour format
Automatic switch-off : Automatic switch-off after 1 minute
Measurement range
forehead & ear mode: 32°C - 42.9°C (89.6°F - 109.2°F)
Measurement
accuracies: ± 0.2°C (± 0.4°F) in the range
(50.0 °F - 104.0 °F)
Forehead mode: +15.0 °C - +40.0 °C
(59.0 °F - 104.0 °F)
10 GB
Storage temperature: –10.0 °C - +55.0 °C
(14.0°F - 131.0°F) at
20 % RH* - 85 % RH*
Atmospheric pressure: 700 ~ 1060 hPa
Dimensions: approx. 135 x 35 x 58 mm (L x B x H)
Weight: approx. 54 g
*RH = relative humidity
Safety
PLEASE KEEP ALL SAFETY INFORMATION AND INSTRUCTIONS FOR
FUTURE USE!
General safety instructions
WARNING!
DANGER
TO LIFE AND DANGER OF
ACCIDENTS FOR INFANTS
AND CHILDREN! Never leave
children alone and unsupervised with
the packaging material. Danger of suff ocation. Children often underestimate
risks. Always keep children away
from the product.
11 GB
WARNING!
DANGER TO LIFE!
Batteries can be swallowed, which may
represent a danger to life. If a battery
has been swallowed, medical help is
required immediately.
This device may be used by children
aged 8 over and by persons with reduced physical, sensory or mental capabilities or lack of experience and knowledge, provided that they are under
supervision or have been told how to use
the device safely and are aware of the
potential risks. Children must not play
with the device. Cleaning and user maintenance tasks must not be carried out by
children unless they are supervised.
Do not expose the device to
- extreme temperatures,
- strong vibrations,
- heavy mechanical loads,
- direct sunlight.
Otherwise there is a danger that the device will be damaged.
Please note that the guarantee does not
cover damage caused by incorrect handling, non-compliance with the operating
instructions or interference with the device by unauthorised individuals.
12 GB
Under no circumstances should you take
the device apart. Improper repairs may
place the user in considerable danger.
Repairs should only be carried out by
specialist personnel.
Do not use the device if it is damaged.
Otherwise the consequence may be injuries and / or inaccurate readings.
Please consult a doctor if the thermome-
ter indicates that the body temperature
measured is high. With this measuring
method, a fever is indicated if the temperature is above 38.0 °C (100.4 °F).
The red LED 5 is a warning. Please take
several measurements at intervals of 0.5
to 1 hour in order to follow the progression of the temperature. The values you
measure will help your doctor make a
diagnosis.
Do not use the temperature under ex-
treme temperature conditions and / or
outdoors. This will result in inaccurate
measuring results. Please observe the details regarding
ambient and storage tem-
peratures (see “Technical data”).
The measurement sensor should not be
touched directly or be blown on. This will
result in inaccurate measuring results.
13 GB
Mobile telephones should not be used in
the immediate vicinity when the thermometer is being used. Otherwise the consequence may be inaccurate measurement
results.
Safety instructions
This appliance is solely intended for pri-
vate use. That is why it is not subject to
mandatory verifi cation.
Discontinue using the appliance if it is
damaged or if you feel that the measured values are unrealistic. Consult the
chapter “Functions/Error analysis - Troubleshooting” for advice or contact our
service centre.
Safety Instructions
for the Battery
Remove the battery from the device if it
is not going to be used for a prolonged
period.
CAUTION! DANGER OF
EXPLOSION!
The battery
must never be recharged!
14 GB
Ensure correct polarity when inserting
the battery! This is indicated in the battery compartment.
If necessary, clean the battery and de-
vice contacts before inserting the battery.
Batteries must not be disposed of in the
normal domestic waste!
Every consumer is legally obliged to dis-
pose of batteries in the proper manner!
CAUTION! RISK OF EX-
PLOSION! Keep batteries
away from children; do not
throw batteries into a fi re; do not
short-circuit the batteries and do
not take them apart.
Avoid contact with the skin, eyes and
mucous membranes. Any parts of the
body coming into contact with battery
acid should be rinsed off with copious
amounts of water and / or a doctor
contacted if necessary!
Avoid contact with skin, eyes and mu-
cous membranes. Upon contact with
battery acid, rinse the aff ected area
with lots of water and / or consult a
doctor!
15 GB
Before getting started
Removing the battery fuse
Push the battery compartment lid 10 downwards.
Pull the tab of the protective fi lm out of the battery
compartment.
Push the battery compartment lid 10 onto the battery
compartment until it clicks into place (Fig. A).
– When it is switched on, an internal alignment and a
self-test of the LC display are carried out.
– To verify that all of the bars of the numbers are being
managed correctly, when you switch on check the
display of the LC display
4
on your appliance
against the example (Fig. B).
Operation
Setting the date / time (Fig. C)
You only have to set the date and time when you operate
the appliance for the fi rst time and after you change the
battery.
1. Press and hold the “MEM / SET” button
down for
9
two seconds in standard mode. An acoustic signal
will sound. The device is now in set-up mode.
2. Press the “MEM / SET” button
to choose between
9
the 12-hour or 24-hour format.
3. Press the “FOREHEAD” button
and can be corrected with the “MEM / SET” button
16 GB
. The hours will fl ash
3
9
.
4. Press the “FOREHEAD” button 3 again. The minutes
will fl ash and can be corrected with the “MEM / SET”
button
.
9
Repeat steps 3 and 4 to set the values for the year,
month and day or to switch the sleep mode on or off .
Finally, press the “FOREHEAD” button 3. The settings will
be taken over and the appliance reverts to standard mode.
Temperature measurements
Measuring room temperature (standard mode)
When you have fi nished setting the time and date and do
not carry out any further procedures, the device switches
to sleep mode after 1 minute in order to save the battery.
In the condition as supplied to the customer the sleep
mode is activated, as shown by the
display
is completely switched off . When the sleep
4
symbol, and the LC
mode is deactivated, the device remains switched on,
while the time and room temperature are still shown and
updated once a minute. The LC display
shows
4
the time and date alternately (Illustration D).
Measuring body temperature (ear mode)
IMPORTANT! It is advisable to carry out 3 meas-
urements in a row on the same ear and to take the highest
temperature as the measured value. A new measurement
can be taken when the display “
” goes out.
Do not use the thermometer if pain occurs when the meas-
urement is being taken. This could damage the ear canal.
17 GB
Do not use the thermometer with external ear disorders
(e.g. infl ammation of the ear canal, injuries to the outer
ear). The condition of the aff ected areas could worsen.
Do not use the thermometer straight after bathing or
swimming when the ear canal is wet. This could damage the ear canal.
Due to the risk of contamination of the measuring
head
, clean the appliance after every use.
1
In this mode, only measure the temperature in the ear
and at no other place on the body.
1. If the sleep mode has been activated, activate the
thermometer again by pressing any button before
taking the measurement.
2. Carefully press on the protective cap
and remove it
8
with a gentle rotary motion and without using any force.
3. Check that the measuring head
and the measuring
1
sensors are clean.
4. Carefully clean the ear canal with a cotton bud.
5. Pull the outer ear, tilted slightly upwards/to the rear,
towards the back of his head, so that the ear canal is
freely accessible and there are no impediments between the measurement sensor and the eardrum.
6. Please keep the outer ear slightly extended and now
carefully insert the measuring head
7. Briefl y press the “EAR” button
appear in the LC display
4
into the ear canal.
1
. The symbol will
2
and the measurement
will start after an acoustic signal.
8. The end of the measurement is indicated by means of
2 acoustic signals.
18 GB
9. Now you can read off the measurement result on the
LC display
. At the same time, one of the 3 LEDs
4
will come on.
Note: LED green
: no fever 34.0 °C - 37.4 °C
7
(93.2 °F - 99.4 °F)
LED orange
: slight increase in temperature
6
37.5 °C - 37.9 °C (99.5 °F - 100.3 °F)
LED red
: fever 38.0 °C - 42.9 °C
5
(100.4 °F - 109.2 °F)
10. Values can be displayed in °C as well as in °F
(see “Functions / error analysis - selecting the
temperature unit”).
11. Wait until the display “
” goes out before you
start taking a new measurement.
Note: The appliance automatically switches off if it is not
operated for longer than 1 minute. When the sleep mode
has been activated, the LC display
switches itself off .
4
When the sleep mode has been deactivated, the appliance
switches to standard mode. If the thermometer has been
operated under normal conditions, the result of the last
measurement taken before the appliance was switched off
is stored. The measured value is available again along
with information on the date, time and measurement mode
when the “MEM / SET” button
is pressed again. The
9
device has 9 storage spaces (see section on “Functions /
error analysis - the memory function”).
19 GB
Measuring the body temperature
(forehead mode)
IMPORTANT! It is advisable to carry out 3 meas-
urements in a row at the same place and to take the
highest temperature as the measured value.
With diff erent room temperatures, let the appliance
lie in the room where the temperature is to be taken
for at least 30 minutes to “acclimatise” it.
Before measuring your body temperature, you should
relax for about 5 minutes.
If you have done any physical activity or taken a
bath, wait for at least 30 minutes before taking a
measurement in order to get a measurement that is
as accurate as possible.
Ensure that the forehead or temple is clean, free of
injuries and not covered with sweat, cosmetics or
ointment.
Take a note of what measurement method you used to
measure the temperature and inform your doctor of it.
1. If the sleep mode has been activated, activate the
thermometer again by pressing any button before
taking the measurement.
2. Carefully press on the protective cap
and remove it
8
with a gentle rotary motion and without using any force.
3. Check that the measuring head
and the measuring
1
sensors are clean.
4. Hold the measuring sensor 1 cm away from the
forehead or temple.
20 GB
5. To start the measurement, press on the “FOREHEAD”
button
4
. The symbol will appear in the LC display
3
and the measurement will start after an acoustic
signal.
6. The end of the measurement is indicated by 2 acoustic
signals.
7. Now you can read off the measurement result on the
LC display
. At the same time, one of the 3 LEDs
4
will come on.
Note: LED green
no fever 34.0 °C - 37.4 °C
7
(93.2 °F - 99.4 °F)
LED orange
slight increase in temperature
6
37.5 °C - 37.9 °C (99.5 °F - 100.3 °F) LED red fever
38.0 °C - 42.9 °C (100.4 °F - 109.2 °F)
5
8. Values can be displayed in °C as well as in °F
(see “Functions / error analysis - selecting the
temperature unit”).
9. Repeat steps 3 to 6 to carry out further measurements.
Note: The appliance automatically switches off if it is not
operated for longer than 1 minute. When the sleep mode
has been activated, the LC display
switches itself off .
4
When the sleep mode has been deactivated, the appliance
switches to standard mode. If the thermometer has been
operated under normal conditions, the result of the last
measurement taken before the appliance was switched off
is stored. The measured value is available again along
with information on the date, time and measurement mode
when the “MEM / SET” button
is pressed again.
9
The appliance has 9 storage spaces (see “Functions /
error analysis - the memory function”).
21 GB
Use as a household thermometer (scan mode)
The thermometer can also be used in the house for
non-contact temperature measurements in the range of
–22.0 °C - 80.0 °C (–7.6 °F - 176.0 °F).
Attention! Do not use this mode for measuring
body temperature.
– With this measurement, the current surface temperature
of the object is displayed. It can be diff erent from its
internal temperature, especially if the surface is exposed
to direct sunlight or a draught!
Due to its design, the thermometer measures the precise
value when its casing has adjusted to room temperature.
That is why you should not hold it in the hand for a
long time when you want to take a measurement and
why you should keep it away from direct sunlight!
Do not use the appliance immediately after cleaning
the measuring head
. The evaporative cooling eff ect
1
of the detergent may falsify the measuring result.
Sample applications:
- The temperature of milk in baby bottles
- The temperature of bathwater
1. If the sleep mode has been activated, activate the
thermometer again by pressing any button before
taking a measurement.
22 GB
2. Carefully press on the protective cap 8 and remove
it with a gentle rotary motion and without using any
force.
3. Make sure that the measuring head
and the
1
measuring sensor are clean.
4. Press the “FOREHEAD” button
2
button
repeatedly at the same time until “Food”
is shown in the LC display
3
and the “EAR”
4
(Illustration E).
5. Hold the measuring head at a distance of 1 cm from
the liquid and / or surface whose temperature you
want to measure.
6. Press the “EAR” button
to start the measurement.
2
7. Now you can read off the measurement result on the
LC display
(Illustration E).
4
8. Values can be displayed in °C as well as in °F
(see “Functions / error analysis - selecting the
temperature unit”).
9. Press the “EAR” button
again to carry out further
2
measurements.
10. The appliance automatically switches off if it is not
operated for longer than 1 minute. This application is
not stored in the device.
In order to return to standard mode, proceed as follows:
1. Press the “FOREHEAD” button
2
button
display
repeatedly at the same time until the LC
4
again shows the room temperature, date
and time, as well as the
3
and the “EAR”
symbol (Illustration D).
2. You are now back in standard mode.
23 GB
Functions / error analysis
Selecting the temperature unit
Press and hold down the “EAR” button 2 and the
“FOREHEAD” button
3
at the same time until “°F”
appears behind the temperature display. Switching
from “°F” to “°C” is possible with the same procedure.
The memory function (memory mode)
IMPORTANT!
The current measurement is always stored in the last storage
space. When all storage spaces have been occupied, the
oldest measurement is always deleted from the memory.
Measurements taken in ear and forehead mode are
automatically stored in the appliance. The measured
values are available again along with information on the
date, time and measurement mode when the “MEM / SET”
button
SET” button
is operated in standard mode. Press the “MEM /
9
once to retrieve the value of the last
9
measurement. The “Memory” symbol (memory mode), the
number of the measurement, date, time and measurement
mode will appear in the LC display
the “MEM / SET” button
the next storage space along
9
(Fig. F). By pressing
4
with its associated measured value will appear. A maximum
of 9 measurements can be stored.
In order to return to standard mode, proceed as follows:
24 GB
1. Press the “FOREHEAD” button 3 and the “EAR”
2
button
display
and time, as well as the
repeatedly at the same time until the LC
4
again shows the room temperature, date
symbol (Illustration D).
2. You are now in standard mode again.
Changing the battery
The appliance is fi tted with a lithium battery (CR2032,
3 V
) that allows a lot of measurements to be taken. If
the battery is empty, you will be made aware of that by
means of an appropriate error message (see “Functions /
error analysis - problem-solving”).
Push the battery compartment lid 10 downwards.
Lever the battery
out with a small screwdriver.
11
Avoid short-circuiting the battery at all cost. A short-
circuit can occur if the screwdriver comes into contact
with the battery and the live parts in the battery
compartment at the same time.
Insert the new battery into the battery compartment.
Note: Ensure correct polarity. The plus sign on the
battery must point upwards.
Push the new battery under the metal nose and press
it down until it clicks into place.
Push the battery compartment lid 10 onto the battery
compartment until it clicks into place (Fig. A).
25 GB
Problem-solving
If the appliance cannot take a measurement in the
designated range, it will generate an error message.
This message should help to solve the problem.
ErrorProblemSolution
Appliance
does not react / automatically resets
itself when
protective fi lm
is removed
Empty battery?Insert a new
battery
Wrong battery
polarity?
Removing battery and insert it
correctly
Faulty battery
contact?
Battery symbol on the LC
display
4
“Lo” on the
LC display
(measured
value below
32.0°C or
89.6°F)
26 GB
Weak batteryInsert a new
Thermometer
correctly posi-
4
tioned in the ear
or on the forehead/temple?
battery
Pay heed to the
operating in-
structions in re-
gard to position-
ing in the ear or
on the fore-
head/temple.
ErrorProblemSolution
“Hi” on the
LC display
4
(measured
value above
42.9°C or
109.2°F)
“ErrE” on the
LC display
4
Check the measurement mode.
Ambient temperature outside the
range of 15.0 °C
- 40.0 °C /
59.0 °F -
104.0 °F (forehead mode) and
10.0 °C -
40.0 °C /50.0 °F
- 104.0 °F (ear
and scan mode).
Please observe
the directions
for use with re-
gard to taking
correct measure-
ments.
Please ensure an
ambient tempera-
ture in the range
of 15.0 °C -
40.0 °C /
59.0 °F -
104.0 °F (fore-
head mode) and
10.0 °C -
40.0 °C/
50.0 °F -
104.0 °F (ear
and scan mode)
“ErrU” on the
LC display
4
No measurement result after
measurement.
Please observe
the directions
for use with re-
gard to taking
correct measure-
ments.
“ErrH” on the
LC display
4
“ErrP” in LC
display
4
Error detected
during self-test.
Hardware
problem
Contact the
service centre.
Contact the
service centre.
27 GB
Cleaning and care
The device reminds you to clean the measuring sensor after taking several temperature measurements. When this
occurs, the cleaning symbol
and acoustic signals are heard. In order to avoid con-
4
appears in the LC display
20
tamination with germs, the lens has to be rubbed after
every use with a soft cloth that has been moistened with
70% medical alcohol. The use of hygiene attachments is,
therefore, not necessary. After that, wait for at least
30 minutes before taking another measurement. Please
keep in mind that due to the evaporative cooling eff ect of
the alcohol, even the smallest amounts can falsify the result.
Clean the housing with a soft, slightly damp cloth. In the
case of more severe soiling a mild detergent can also be
added to the water.Keep the appliance in a dry environment and protect from direct sunlight. The appliance can
be stored at temperatures of –10.0 °C - 55.0 °C
(14.0 °F - 131.0 °F). Temperatures over 80.0 °C are to
be avoided! In order to measure with it, however, it is necessary to let the appliance adjust to room temperature for
a suffi cient length of time - otherwise an error message
will be generated (see “Functions / Fault Analysis - Correction of Problems”). A room temperature of 5.0°C - 59.9°C
(41.0°F - 139.8°F) should also be maintained for the purposes of measurement. If the temperature is outside this
range, an error message is shown (see “Functions / Fault
Analysis - Correction of Problems”).
Remove the battery when not using the appliance for any
length of time.