English
Print No.: SPB7-420.812.05.01.02 Doc . Gen. Date:04.03
Replaces: n.a.
66 54 003
Page 2
0 - 2Revision
ChapterPageRevision
AllAll01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision le vels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempti ng installation or service procedures.
MR Plus - M3000 NovaSPB7-420.812.05Page 2 of 4Siemens-Elema AB
Siemens-Elema ABSPB7-420.812.05Page 3 of 4MR Plus - M3000 Nova
Solna, SwedenRev. 0104.03SPS-UD
Page 4
0 - 4Contents
This page intentionally left blank.
MR Plus - M3000 NovaSPB7-420.812.05Page 4 of 4Siemens-Elema AB
Rev. 0104.03SPS-UDSolna, Sweden
Page 5
Prerequisites1
Purpose1
1 - 1
This document describes extra settings for a system comprising the MAMMOMAT 3000
Nova, MammoReport
Plus
, Fuji CR Console and FCR5000 MA reader.
Important note1
After the instructions in this document have been performed, the installation protocol at
the end of this document must be signed in by the CSE and a Fuji Technician to confirm
the installation. A copy of the signed installation protocol shall be sent to:
Siemens AG
Medical Solutions
SP WH - Product Quality Group
Henkestrasse 127
D - 91052 Erlangen
Validity1
The instructions in this document must only be perfomed for a system comprising the
MAMMOMAT 3000 Nova, MammoReport
The worksteps have to be performed by a Fuji Technician.
Plus
, Fuji CR Console and FCR5000 MA reader.
Components included1
The package contains the following parts:
Part no.Name
66 53 997Workflow instructions, Supplement to the instruc-
tions for use
66 54 003This instruction
Siemens-Elema ABSPB7-420.812.05Page 1 of 2MR Plus - M3000 Nova
Solna, SwedenRev. 0104.03SPS-UD
Page 6
1 - 2Prerequisites
This page intentionally left blank.
MR Plus - M3000 NovaSPB7-420.812.05Page 2 of 2Siemens-Elema AB
Rev. 0104.03SPS-UDSolna, Sweden
Page 7
Worksteps2
Settings2
The following settings shall be made for the Fuji CR Console in order to function properly
with the other units in the system. The settings shall be performed by a Fuji Technician.
• The following six menus shall be setup:
- CC left
- CC right
-MLO left
- MLO right
-ML left
- ML right
• The six menus shall be based on the menus 0 307 and 0317, but with the following image
processing parameters:
- PEM off
2 - 1
- DRC on
- TAS off
- GA=1.8, GT=Q, GC=0.6, GS=0
- MRB=G, MRT=P, MRE=0.4
- EDR=AUTO
Final procedures2
After the settings have been performed, fill in the Installation protocol at the end of this
document and send it to:
Siemens AG
Medical Solutions
SP WH - Product Quality Group
Henkestrasse 127
D - 91052 Erlangen
Siemens-Elema ABSPB7-420.812.05Page 1 of 2MR Plus - M3000 Nova
Solna, SwedenRev. 0104.03SPS-UD
Page 8
2 - 2Worksteps
This page intentionally left blank.
MR Plus - M3000 NovaSPB7-420.812.05Page 2 of 2Siemens-Elema AB
Rev. 0104.03SPS-UDSolna, Sweden
Page 9
Installation pr oto co l3
Settings3
PassedFailed
The six menues hav e been setup:
The image processing parameters have been setup:
Date of installation:
Name and signature,
Fuji Technician:
System environment3
PassedFailed
3 - 1
A Hardcopy printer or an Archiv e system (e.g. PACS) is
connected to the system:
Name of Hard co py pr inter or Archive syste m :
Name and signature, CSE:
MAMMOMAT upgrade3
PassedFailed
The AEC tables of the MAMMOMAT are calibrated with
the Upgrade kit AEC (Fuji) for MAMMOMAT:
Name and signature, CSE:
When this protocol has been completed, file a copy with the MammoReport
binder and send this protocol to:
Plus
system
Siemens AG
Medical Solutions
SP WH - Product Quality Group
Henkestrasse 127
D - 91052 Erlangen
Siemens-Elema ABSPB7-420.812.05Page 1 of 2MR Plus - M3000 Nova
Solna, SwedenRev. 0104.03SPS-UD
Page 10
3 - 2Installation protocol
This page intentionally left blank.
MR Plus - M3000 NovaSPB7-420.812.05Page 2 of 2Siemens-Elema AB
Rev. 0104.03SPS-UDSolna, Sweden