Siemens Mammomat Novation DR Installation Manual

Page 1
Opdimar
Installation
Installation and Start-Up
SP
on MAMMOMAT Novation
English Print No.: SPB7-230.814.01.01.02 Doc. Gen. Date: 10.04
Replaces: n.a. (66 31 944)
DR
© Siemens AG 2004
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Page 2
0 - 2 Revision
Chapter Page Revision
All All 01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revi­sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service procedures.
Opdima SPB7-230.814.01 Page 2 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 3
Contents 0 - 3
Page
1 _______Prerequisites __________________________________________________1 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
Training of customer support engineers. . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Text emphasis . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Documents required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Meters and appliances required. . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 3
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Additional installation material required . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Start-up, DHHS and X-ray decree area of application. . . . . . . . . . . . . . . . . .1 - 3
Notes on start-up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 3
Checking and recording for the area of application of the X-ray decree
(§16 Germany). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 4
Checking and data recording in the USA . . . . . . . . . . . . . . . . . . . . . . 1 - 4
2 _______Protective measures for CCD camera______________________________2 - 1 3 _______Protective measures ____________________________________________3 - 1 4 _______Preparatory work_______________________________________________ 4 - 1
Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 1
Unpacking . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4 - 2
Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Biopsy controller, miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
Optional equipment and update kits. . . . . . . . . . . . . . . . . . . . . . . . . 4 - 2
5 _______Installation ____________________________________________________5 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 1
Sub-assemblies, overview. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Removal of MAMMOMAT NovationDR stand covers . . . . . . . . . . . . . . . . . .5 - 2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 2
Installation of biopsy controller holders . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
Connection of cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 5
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 6
Cables to CCD camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 6
Installation of cable duct. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
Modification of cable outlet cover . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
Installation of workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 9
Preparing the workstation table (option). . . . . . . . . . . . . . . . . . . . . . .5 - 9
Connecting the workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12
Strain relieving the cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 17
Installation of input/output devices. . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Installation of monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 19
Fine adjusting the monitor settings . . . . . . . . . . . . . . . . . . . . . . . . 5 - 19
Connecting the workstation to the network . . . . . . . . . . . . . . . . . . . . 5 - 21
Siemens AG SPB7-230.814.01 Page 3 of 4 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 4
0 - 4 Contents
Page
Installation of Opdima printer (option) . . . . . . . . . . . . . . . . . . . . . . .5 - 22
Functional test (workstation table) . . . . . . . . . . . . . . . . . . . . . . . . .5 - 23
6 ______ Final procedures _______________________________________________6 - 1
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Starting up the MAMMOMAT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Starting up the workstation and login. . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Starting up the MO unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Installation of software on OPDIMA WORKSTATION . . . . . . . . . . . . . . . 6 - 2
Mounting of covers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Filling in of IVK list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Performing of DHHS maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
7 ______ Changes to previous version _____________________________________7 - 1
Opdima SPB7-230.814.01 Page 4 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 5

Prerequisites 1

General 1

This installation instruction is valid for MAMMOMAT Novation > 1100.
DR
with serial numbers
1 - 1
The wall socket for the power supply for the computers of Opdimar must be connected
DR
to the same branch circuit as the MAMMOMAT Novation
power supply
This Installation and Start-Up Instruction is valid for Opdimar systems:
- with SUN workstation part No. 83 83 106/ Sun Blade 150 No.W 10000
with ASW 3.1 and with serial
Siemens AG SPB7-230.814.01 Page 1 of 4 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 6
1 - 2 Prerequisites

Training of customer support engineers 1

Due to the technology used in this equipment, setup, service and maintenance may only be carried out by a customer support engineer who has attended a training workshop or has participated in at least one installation.

Text emphasis 1

DANGER
WARNING
CAUTION
NOTICE
NOTE
DANGER indicates when there is an immediate danger that leads to death or serious physical injury.
WARNING indicates a risk of danger that may lead to death or to serious physical injury.
CAUTION used with the safety alert symbol indicates a risk of danger that leads to slight or moderate physical injury and/or damage to property.
NOTICE used without the safety alert symbol indicates a risk of danger that if disregarded leads or may lead to a potential situa­tion which may result in an undesirable result or state other than death, physical injury or damage to property.
NOTE contains information provided with special emphasis to facilitate proper use of the equipment or proper execution of a procedure, i.e. hints, tips
.
Opdima SPB7-230.814.01 Page 2 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 7
Prerequisites 1 - 3

Documents required 1

Supplement to the Instructions for Use MAMMOMAT Novation
(included in the Opdimar delivery)
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
(included in the Opdimar delivery)
MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar DHHS Maintenance Instructions (included in the Opdimar delivery)
MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy
Attachment and Opdimar Measurement Certificates (included in the Opdimar delivery)
MAMMOMAT Novation
MAMMOMAT Novation
MAMMOMAT Novation
DR
- Opdimar Planning Guide
DR
- Opdimar Installation and Start-Up
DR
- Opdimar Wiring Diagram
MAMMOMAT 3000 Installation and Setting Instructions Update Kit for Stereotactic
Biopsy Attachment (where applicable)
DR
- Opdimar

Meters and appliances required 1

Protective ground wire tester (44 15 899 RV090)
Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to
generator (part No. 99 00 440 RE999)
Stereo calibration phantom (included in the delivery) (part No.64 30 701)
AEC calibration plexiglass, four plates measuring 150 mm x 150 mm x 19 mm and one
plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224 respectively
Resolution phantom with at least 10 line pairs per mm

Tools required 1

Standard installation tools
Electric drill with 5 mm drill

Additional installation material required 1

Extension mains cord with at least three outlets, of protective earth type

Start-up, DHHS and X-ray decree area of application 1

Notes on start-up 1

Opdimar is adjusted, programmed and tested in the factory, leaving only the connection to the on-site mains voltage, the functional tests and customer dependent adjustments to be performed.
Siemens AG SPB7-230.814.01 Page 3 of 4 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 8
1 - 4 Prerequisites

Checking and recording for the area of application of the X-ray decree (§16 Germany) 1

An X-ray unit can only be put into operation if an acceptance test has been performed according to §16 RoeV. It is recommended that the responsible Customer Service Engineer is present to assist the person responsible for this acceptance test.

Checking and data recording in the USA 1

Maintenance measurements must be made according to the instructions in MAMMOMAT 300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdimar DHHS Maintenance Instructions.
Sections
Required labels
Reproducibility
Automatic exposure control (AEC)
The result must be recorded in the document MAMMOMAT 300/3000, MAMMOMA T 3000 Modular incl. Stereotactic Biopsy Attachment and Opdimar Measurement Certificates.
Opdima SPB7-230.814.01 Page 4 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 9

Protective measures for CCD camera 2

2 - 1
CAUTION
The camera shall be used within 10230i C. The camera shall be transported or stored within 0240i C.
The CCD camera is very sensitive to mechanical shocks and tem­perature changes. In the camera shock and temperature sensors are integrated.
Risk of damaging the equipment. The camera has to be handled with extreme care.
When disconnected from the biopsy controller, the camera shall always be stored in the attaché case delivered with the system. Do not touch the pins in the contacts of the camera.
Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 10
2 - 2 Protective measures for CCD camera
This page intentionally left blank.
Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 11

Protective measures 3

It is very important that any intervention in the equipment shall start with disconnecting it from the power supply with the main circuit breaker. To prevent accidental triggering of high voltage and radiation, set the switch S2 (SS) on board D750 to OFF.
3 - 1
CAUTION
WARNING
WARNING
When switching off the workstation use the power off procedure described in the Supplement to the Instructions for Use MAMMO­MAT Novation
Switching off the workstation before the software has been closed down may cause damage to the files on the hard disc.
If the system is only switched off at the control panel or with the circuit breaker CB S700, there is still voltage present within the generator. See the wiring diagram for MAMMOMAT Novation
Life-threatening electric shock hazard exists. Disconnect mains cable and comply with the information on this
page.
After shut-down of the system, there may still be about 380 V DC present on the intermediate circuit of the MAMMOMAT generator.
Life-threatening electric shock hazard exists.
DR
- Opdimar.
DR
.
CAUTION
A bleader resistor is connected over the output DC capacitors. The DC voltage will discharge below 60 Volt in 3 minutes. The voltage level is indicated by LEDs H3 and H4.
Observe the currently valid guidelines for handling electronics endangered by electrostatic discharge.
Use ESD-equipment, ground prior to making contact and place the components on a conductive surface.
The boards contain electrostatic highly sensitive components requiring particular care in their handling.
Risk of damaging components.
Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 12
3 - 2 Protective measures
This page intentionally left blank.
Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 13

Preparatory work 4

Scope of delivery 4

The following parts are always included in the delivery of your equipment:
Technical Manual comprising information for installation, operation and service
CCD camera with a fixed cable
Biopsy controller with five cables and cable duct
Stereo calibration phantom, 64 30 701 (see page 1-3)
Workstation including:
- power supply cable
- network cable
- KVM switch with cables
- MO unit with one power supply cable and one data cable
- Main unit including CD drive
- In some deliveries, the monitor, keyboard and mouse may be delivered. In such case the KVM switch is not needed.
If ordered, the following is delivered:
Biopsy unit
4 - 1
Compression plate for biopsy examination
Printed circuit board D702
MAMMOMAT software V2.2 or higher
Isolation p.c. board D707 kit
DICOM license key
Workstation table
Printer
Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 14
4 - 2 Preparatory work

Unpacking 4

General 4

NOTE
Follow the directional marks on the box during transport, storage and unpacking.

Workstation 4

The workstation is packed in a card board box which also contains technical documentation and CDs for application software and operating system
1
.

Biopsy controller, miscellaneous 4

The biopsy controller is packed in a card board box together with cables, miscellaneous installation material.
CCD camera
CAUTION
The CCD camera is very sensitive to mechanical shocks and tem­perature changes. In the camera shock and temperature sensors are integrated.
Risk of damaging the equipment. The camera has to be handled with extreme care.
When disconnected from the biopsy controller, the camera shall always be stored in the attaché case delivered with the system. Do not touch the pins in the contacts of the camera.
The CCD camera is packed in an attaché case. Keep the case for future use.

Optional equipment and update kits 4

If optional equipment and/or update kits are ordered, these are packed in separate boxes.
1. Only needed if a software reinstallation is necessary (see MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions). Store the documentation and software CDs in a proper place for future service use.
Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 15

Installation 5

General 5

5 - 1
Before starting the installation, a backup of the installation parameters of the
DR
MAMMOMAT Novation
can be carried out. This may be omitted, if backup already
available.
1. Connect a PC to the MAMMOMAT Novation
2. Select
Main menu => Backup => Copy installation area to disk => All
DR
and start the service program.
.
This will store the installation parameters on the hard disk of the PC.
3. Select
Main menu => Backup => Copy installation area to floppy => All
.
This will store the installation parameters on the service diskette.
4. Switch off the MAMMOMAT.

Sub-assemblies, overview 5

The components shall be placed according to the figure below.
Biopsy unit
X
mm
X
X
mm
Y
mm
Z
Biopsy controller
Y
Y
Z
Z
X
Z
Y
CCD camera
Stand with X-ray unit
Workstation
Workstation table
Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdimar
NOTE
The workstation shall be placed at least 1.5 m from the object table of the stand.
Generator
OPD00597
Siemens AG SPB7-230.814.01 Page 1 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 16
5 - 2 Installation

Removal of MAMMOMAT NovationDR stand covers 5

5
WARNING
The right/rear side cover should be removed. The holes for fastening of the biopsy controller BC and of the cable tray below the BC
have been drilled in the factory. There are two cable trays of different lenghts delivered, use the suitable one.
High voltage! Life-threatening electric shock hazard exists. Switch off the power to the system with the main circuit breaker.
Opdima SPB7-230.814.01 Page 2 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 17
Installation 5 - 3

Installation of biopsy controller holders 5

The biopsy controller is attached to the rear cover of the stand. To install the holders proceed as follows:
1. Install the three holders (2/Fig. 2) by using the nuts (3/Fig. 2).
2. Install the two standoffs (4/Fig. 2) by using the screws (5/Fig. 2) and the washers (6/Fig. 2).
UP
985 – 5 mm
2
Fig. 2 Rear cover, from the back side
A
3
4
B
6
5
1
OPD00267
Siemens AG SPB7-230.814.01 Page 3 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 18
5 - 4 Installation
3. Install the rear cover (2/Fig. 3) on the stand and attach the biopsy controller (1/Fig. 3) to the holders. Switch on the power-on switch on the biopsy controller.
2
1
OPD00152
Fig. 3 Biopsy controller
Opdima SPB7-230.814.01 Page 4 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 19
Installation 5 - 5

Connection of cables 5

The biopsy controller is the central unit for the cabl e connections of OPDIMA. The cables are connected to the biopsy controller according to the figure 6 on page 2-6, see wiring diagram doc No: SPB7-230.051.09.03.02.
Perform the connection of cables according to data in the wiring diagram.
Siemens AG SPB7-230.814.01 Page 5 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 20
5 - 6 Installation
X201
X202
X203
Ferrite
X204
OPD00453
5
X205 X206 X207 Mains cable
Fig. 4 Connection of cables to the biopsy controller
Protective earth cable

Cables to CCD camera 5

The CCD camera including cable with connectors marked X201 and X202 is connected to the biopsy controller.
1. Connect the cable to X201 and X202 on the biopsy controller, see Fig. 4.
CAUTION
2. Remove the screws (1/Fig. 5) and remove the cover (2/Fig. 5) of the object table arm.
Sensitive electronic equipment! Risk of damaging the camera. Do not touch the pins in the contacts of the CCD camera.
1 2
Fig. 5 Object table arm before installation
Opdima SPB7-230.814.01 Page 6 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
OPD00161
Page 21
Installation 5 - 7
1. Install a new thread (1/Fig. 6).
1
Fig. 6 Installation of thread
OPD00162
2. Install the holder (1/Fig. 7) with the pull relief device, using included Allen screws (2/Fig. 7).
21
OPD00163
Fig. 7 Object table arm after installation
Siemens AG SPB7-230.814.01 Page 7 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 22
5 - 8 Installation

Installation of cable duct 5

Four cables are laid from the bottom of the biopsy controller in a cable duct to the cable outlet cover (3/Fig. 8). To install the cable duct on the rear side of the stand proceed as follows:
1. Attach the cable duct (2/Fig. 8) to the rear cover by removing the protective strips from the attached Velcro strips and place the duct flush with the biopsy controller (1/Fig. 8). Secure the cable duct to the two standoffs (5/Fig. 8) with the delivered screws and washers.
UP
1
4
5
2
3
OPD00164
Fig. 8 Cable duct
50mm
25mm

Modification of cable outlet cover 5

To allow the cables to pass through the cable outlet cover, the cover has to be modified according to the following:
1. Cut an opening centrally on the cable outlet cover (4/Fig. 8). The opening should be approximately 50 x 25 mm.
2. Attach the supplied plastic strip around the opening of the cable outlet cover (4/Fig. 8).
3. Install the cable outlet cover (4/Fig. 8).
4. Mount the previously removed stand covers.
Opdima SPB7-230.814.01 Page 8 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 23
Installation 5 - 9

Installation of workstation 5

Preparing the workstation table (option) 5

The workstation table (option), a motorized vertically adjustable table , shall be used to put the workstation components to the Opdimar system on. The vertical adjustment device includes a transformer, which is not mounted on delivery.
CAUTION
Handschalter
5
Fig. 9 Transformer
The workstation table must not be overloaded. Any additional equipment is added at your own risk and the maxi-
mum load of 80 kg must not be exceeded.
2
3
1
4
Opd00494
1. Remove the screws on the underside of the tabletop and use them when mounting the transformer (1/Fig. 9). Place the transformer so that the connectors are turned to the column.
2. Connect the cable from the control unit (2/Fig. 9) for the vertically adjustment to the connector marked “Handschalter” (3/Fig. 9).
3. Connect the cable from the motor (4/Fig. 9) for the vertically adjustment to the other connector (5/Fig. 9).
Siemens AG SPB7-230.814.01 Page 9 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 24
5 - 10 Installation
4. Place the workstation table outside the patient environment of safety reasons. Patient environment is determined by IEC 60601-1-1 as 1.5 m away from the patient.
Cables 8 m
Cable 1,5 m
Biopsy controller
m
5
,
1
.
n
i
M
Cable (workstation table) 2 m
CCD camera holder
Stand
Biopsy unit Generator
Workstation
150
CCD camera
Cable duct
Fig. 10 System site overview
NOTE
OPD00599
There must be enough space around the workstation table so that it safely can be vertically adjusted.
Opdima SPB7-230.814.01 Page 10 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 25
Installation 5 - 11
5. If the workstation table stands on an uneven floor, it can be adjusted with the adjustable screws on the framework.
BUSY
POWER
Fig. 11 MO Shelf
6. Place the workstation, the monitor, the keyboard and the mouse on the tabletop. Make sure that all the equipment is within the edges of the tabletop.
1
EJECT
OPD00611
7. Let the cables from the workstation, the monitor, the keyboard and the mouse pass through the hole at the back of the tabletop.
8. Mount the MO unit shelf on the right below the tabletop (1/Fig. 11).
9. Place the MO unit to the right on the shelf (1/Fig. 11).
Siemens AG SPB7-230.814.01 Page 11 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 26
5 - 12 Installation

Connecting the workstation 5

1. Connect workstation cable marked X203 to the biopsy controller connector X203, see Fig. 12, and the other end (X101) of the cable to the workstation connector (2B/Fig. 26).
CAUTION
Connecting the cable to the wrong connector may cause severe damage to the equipment.
2. Attach ferrite on the cable on the biopsy controller side, see Fig. 12.
X201
X202
X203
Ferrite
X204
OPD00453
X205 X206 X207 Mains cable
Fig. 12 Biopsy controller
Protective earth cable
3. Connect the workstation cable marked X204 to the biopsy controller connector X204 in Fig. 12 and the other end of the cable to the workstation connector A (8/Fig. 17) and B (2A/Fig. 17).
4. Attach ferrite on the cable on the biopsy controller side, see Fig. 12.
5. Mount the touch protection over the connectors X203 and X204.
-Place the plastic plate (1/Fig. 13) behind the connectors X203 and X204.
-Place the plastic cover (2/Fig. 13) on the other side of the connectors.
-Fasten the plastic screws (3/Fig. 13).
3
1
2
Biopsy controller
Fig. 13 Mounting the touch protection
OPD0
Opdima SPB7-230.814.01 Page 12 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 27
Installation 5 - 13
6. Connect the MO unit SCSI cable to the workstation SCSI connector (2C/Fig. 17).
7. Connect the MO unit cable to one of the two connectors of the MO unit. Fit the terminator to the other connector of the MO unit. For switch settings, SCSI ID and termination see Paragraph “Installing the MO unit” in Chapter “Removal and replacement of sub-assemblies” in Mammomat 1000/3000/3000 Nova - Opdimar Service Instructions.
CAUTION
Make sure that the SCSI cables are properly connected and fixed to the units and that the cable is not bent too much just behind the MO unit. Also, check that the MO unit is terminated correctly.
If disregarded, this might generate a lot of problems e.g. failure to write images to MO disk, corrupted MO disk or corrupted hard disk.
8. Connect the mouse to the keyboard.
9. Connect keyboard and mouse (3/Fig. 17).
10. Connect monitor cable to the workstation connector (6/Fig. 17).
11. Connect Ethernet cable (4/Fig. 17).
12. Mark the wall socket connected to the same branch circuit as the MAMMOMAT (Fig. 14) with label according to Fig. 15.
The power distribution to the system is shown in Fig. 14.
1-phase
L1 N
PE
10A
2-phase
L1L2
10A
N
PE
Off On
Power On-Off breaker
Fig. 14 Power distribution
Work­station
Power connection
in MAMMOMAT 3000-stand
Off On
Power On-Off breaker
Power connection
in MAMMOMAT 3000-stand
Work­station
OPD00166
Siemens AG SPB7-230.814.01 Page 13 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 28
5 - 14 Installation
13. If a voltage selector switch exists, ensure the line voltage selector switch is set to the appropriate setting: 115V or 230V.
CAUTION
Plugging a 115V power cord into a 230V connector will severely damage the system.
14. Connect the AC power cords of the main unit, the monitor, the MO unit, the local Opdimar printer (if present) and the MAMMOMAT printer (if present) to the wall socket marked with label.
It is advisable to use an extension cord with at least three outlets.
CAUTION
The workstation and the printer must be grounded to the same potential as the MAMMOMAT.
Connect the wall socket for the workstation and printer power supply to the same branch circuit as the MAMMOMAT power sup­ply.
(The workstation table can be connected to any branch circuit.)
Opdima mains power supply outlet!
OPD00600
Fig. 15 Mains marking label
Opdima SPB7-230.814.01 Page 14 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 29
Installation 5 - 15
Front panel description
Position Explanation
1 Power switch 2 Power-indicator LED 3 Smart card reader (not used in Opdima) 4 3.5-inch diskette drive (not used in Opdima) 5 5.25-inch CD drive
2
1
Fig. 16 Front panel overview
Back panel description and connector symbols
Position Explanation Connector
1 Mains power supply None 2A Biopsy controller interface (X204) PCI-3 2B Camera interface via biopsy controller (X203) PCI-2 2C SCSI interface PCI-1 3 Universal serial bus (USB) connectors (four)
(keyboard and mouse interface) 4 Network connector 5 Not applicable (IEEE 1394 (two) connectors), not used 6 Monitor interface 7 Parallel port to local printer 8 Mammomat interface via biopsy controller (X204) 9 Audio module headphones connector, not used 9 Audio module line-out connector, not used 9 Audio module line-in connector, not used 9 Audio module microphone connector, not used
3
4
5
OPD00613
symbols
SERIAL
OPD00612
1
2A
Fig. 17 Back panel overview
Siemens AG SPB7-230.814.01 Page 15 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
2B
2C
4356789
Page 30
5 - 16 Installation
68
Fig. 18 Port extenders to the back panel of the workstation
OPD00629
Opdima SPB7-230.814.01 Page 16 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 31
Installation 5 - 17

Strain relieving the cables 5

If the workstation table (option) is used with the Opdima system, the cables from the workstation components and from the MAMMOMAT must be strain relieved by cable clamps at the back of the MO unit shelf according to the figure below.
To ensure that the cables have enough length for the vertical adjustment, let the
workstation table be in its highest possible position when strain relieving the cables.
Use cable ties to arrange the cables in an organized way and make sure that the cables
can run without a risk of getting caught.
Fig. 19
MO AC
power
WS Mains power supply
Biopsy
controller Printer/ Network connector
A1BCDEFGH
SCSI ID
0
SCSI CONNECTOR
Camera controller
AC IN
F.GND
Opd00630
Siemens AG SPB7-230.814.01 Page 17 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 32
5 - 18 Installation

Installation of input/output devices 5

Since the MAMMOMAT Novation DR and the Opdimar use monitors, keyboards and mouse, it was decided to install:
- one monitor
- one keyboard and
- one mouse. There is a KVM switch delivered connecting these three input/output devices alternatingly
either to the WH AWS of the MAMMOMAT Novation DR or to the SUN-Blade 150 com­puter workstation of the Opdimar.
If reprogramming of the KVM switch is necessary follow the description below.
Opdima SPB7-230.814.01 Page 18 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 33
Installation 5 - 19

Installation of monitor 5

Monitor type: Siemens TFT-LCD monitor 1897 Material required: Mains cable, Cable BNC to HD15 or 13W3(Sun). Proceed as follows to install the monitor:
1. Unpack the monitor and place it on top of the workstation.
2. Connect the monitor to the workstation with the accompanying cables to the connectors on the backside of the monitor according to Fig. 20.
Sun Ultra 1, Sun Ultra 10: 13W3-4 BNC, Siemens Part. no. 66 33 684
R: Red cable G: Green cable B: Blue cable H (H/W): Sync, black cable
SunBlade 150: HD15(VGA)-5 BNC, Siemens Part. no. 66 33 676
Fig. 20 Connectors between workstation and monitor
3. Start up the workstation and log in as service, select service and start Advanced service 3 Software Upgrade 6 Miscellaneous 5 Select Monitor type: 67 Hz SunBlade 150. Select Apply and close to close down the rest of the windows.
4. Log out and log in again as service.

Fine adjusting the monitor settings 5

Read the monitor manual “High-Resolution 18” LCD Monitor SCD 1898-0” (delivered with the monitor) and especially the chapter for OSD (On Screen Display).
R: Red cable G: Green cable B: Blue cable H (S): Horizontal/(vertical) sync, white cable V: Vertical sync, black cable
Network cable: 230 Volt
OPD00631
Use the SMPTE test pattern image as reference when calibrating the image geometry of the monitor (see the Opdima Quality Control Manual on how to do so). To compensate for movement in the grid pattern, use frequency/phase under others (settings).
Siemens AG SPB7-230.814.01 Page 19 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 34
5 - 20 Installation
If the monitor is wrongly adjusted:
Make an autoadjust under the heading Auto. Control that the monitor has the correct settings for vertical sync under the Others set-
tings, see Fig. 21.
OPD00632
Fig. 21 Monitor settings
The settings shall be:
Brightness Contrast:
Brightness 50 Contrast 50 Backlight 70 Color 2
Position Zoom:
Fill all
Source:
Analog RGB
Auto:
Auto brightness: off Auto position zoom: on Auto frequency phase: on Auto all off
Language:
English
Opdima SPB7-230.814.01 Page 20 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 35
Installation 5 - 21
Others:
select Frequency phase, Frequency: 64*
Phase 160* * Can be adjusted for optimal image reproduction, see the monitor manual “High-Resolu-
tion 18” LCD Monitor SCD 1898-0” (delivered with the monitor). Use the SMPTE test pat­tern image for fine adjustments of the image.
select Sharpness, 3 select Video/PIP settings, off select PIP Source, auto select OSD Horizontal: Max
Vertical: Max Background: Translucent Timeout: 20 Sec.
Make a final evaluation of the monitor settings with the clinical image as reference. See Quality Control Manual.

Connecting the workstation to the network 5

The Opdima workstation can be connected to a network, partly for the use of DICOM and also for networked printers.
1. Connect one end of the network cable (included in the delivery) to the workstation.
2. Connect the other end of the network cable to a network socket.
3. Log in with the user name “service”. The password is obtained from Siemens Service Centre.
4. Press the Service button in the mode selection dialog, see Selection of mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
5. Press the Advanced service button in the service dialog, see Use of Advanced service functions in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
6. In Advanced service configure the network parameters, see Network settings in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
For physical limitations of the network see Mammomat 1000/3000/3000 Nova ­Opdimar Planning Guide.
Siemens AG SPB7-230.814.01 Page 21 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 36
5 - 22 Installation

Installation of Opdima printer (option) 5

Stand alone Opdima system:
When the Opdima system is stand alone (not connected to a network) it is possible to install a Codonics NP-1660 or NP-1600 printer to Opdimar as a stand alone printer connected to the network connector.
NOTE
The following equipment is needed:
It is also possible to connect a stand alone postscript printer to the parallel or network connector of the workstation, see Use of printer setup in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
Codonics printer NP-1660 or NP-1600
Crossover network cable
Proceed as follows to install the printer:
1. Connect one end of the network cable to the workstation (2C/Fig. 17).
2. Connect the other end of the network cable to the printer.
3. Connect the mains cable of the printer to the wall socket marked with label, see Fig. 15.
CAUTION
The printer must be grounded to the same potential as the MAMMOMAT.
Connect the printer power supply to the same branch circuit as the MAMMOMAT power supply.
4. Set the printer IP address to 10.10.10.2 see the User’s Manual for the Codonics NP-1660 or NP-1600.
5. Select Scaled in printer setup in the advanced service dialog, see Stand alone Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
Networked Opdima system:
When the Opdimar system is networked it is possible to configure the system to print to a printer connected to the network or a local postscript printer connected to the parallel interface of the workstation.
NOTE
It is also possible to connect a stand alone postscript printer to the parallel connector of the workstation, see Select printer in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
Opdima SPB7-230.814.01 Page 22 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 37
Installation 5 - 23
Proceed as follows to install the network printer:
1. Add access to the network printer in the advanced service dialog, see Networked Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
2. Select printer in the advanced service dialog, see Networked Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.

Functional test (workstation table) 5

Before the installed workstation table (option) is handed over to the customer, check that the workstation table can be vertically adjusted between the highest and lowest possible positions properly and safety.
Siemens AG SPB7-230.814.01 Page 23 of 24 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 38
5 - 24 Installation
This page intentionally left blank.
Opdima SPB7-230.814.01 Page 24 of 24 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 39

Final procedures 6

General 6

The following procedures shall be performed:
Starting up the workstation, see Page 6 - 2
Country settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Test of units,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Turn screen lock on/of f, software upgrade, miscellaneous,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
CCD camera calibration,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Calibration of the biopsy unit,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
6 - 1
Verifying the calibration of the biopsy unit,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Check of Opdimar AEC,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Check of resolution,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Network settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Activate DICOM (optional),
see DICOM option in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova ­Opdimar Service Instructions
Printer setup (optional),
see Use of printer setup in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova
- Opdimar Service Instructions
Mounting of covers, see Page 6 - 3
Protective earth measurement,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Fill in the IVK list or report the IVK components to your responsible uptime service center
(USC).
Siemens AG SPB7-230.814.01 Page 1 of 4 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 40
6 - 2 Final procedures
Performing of DHHS maintenance, see Page 6 - 3
NOTICE
The Opdimar external diaphragm must be used for tests and cali­brations involving radiation.

Starting up the MAMMOMAT 6

Switch on the MAMMOMAT.

Starting up the workstation and login 6

The workstation is started according to instructions in Starting up the workstation and login in Supplement to the Instructions for Use MAMMOMAT Novation
Use the user name “service” and the password obtained from Siemens service center when entering the system.
NOTE
To obtain optimal picture, adjust the monitor settings as follow:
Any adjustments of the monitor settings shall be performed by qualified personell who are employed by Siemens or one of its affiliates or who are otherwise autorized by Siemens or one of its affiliates to provide such services.
DR
- Opdimar.
Brightness: 50 Contrast: 50 Backlight: 70

Starting up the MO unit 6

OPD00601
Fig. 1 Magneto-optical drive (SONY)
1. Insert the MO disk, type 9.1 GB MO disk
2. For switch settings and SCSI ID, see Paragraph “ Installing the MO unit” in Chapter “Removal and replacement of sub-assemblies” in MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions.

Installation of software on OPDIMA WORKSTATION 6

The software is installed in the factory before deliv ery . If the softw are for an y reason needs to be reinstalled, refer to chapter “Reinstallation of software” in MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions.
Opdima SPB7-230.814.01 Page 2 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 41
Final procedures 6 - 3

Mounting of covers 6

Mount the covers previously removed.

Filling in of IVK list 6

Fill in the IVK list or report the IVK components to your responsible uptime service center (USC).

Performing of DHHS maintenance 6

In the USA, perform maintenance according to DHHS Maintenance Instructions for MAMMOMAT 1000/ 3000 Nova incl. Stereotactic Biopsy Attachment and Opdimar and fill in DHHS Measurement Certificates for MAMMOMAT 1000/ 3000 Nova incl. Stereotactic Biopsy Attachment and Opdimar.
NOTE
The DHHS Maintenance Instructions and DHHS Measurement Certificates delivered with the Opdimar also includes the meas­urements to be performed on the MAMMOMAT. Therefore the DHHS Maintenance Instructions supplied with the MAMMOMAT can be removed. However, the DHHS Measurement Certificates supplied with the MAMMOMAT should be kept for future refer­ence, if it contains previous measurement results.
Siemens AG SPB7-230.814.01 Page 3 of 4 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 42
6 - 4 Final procedures
This page intentionally left blank.
Opdima SPB7-230.814.01 Page 4 of 4 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions
Page 43

Changes to previous version 7

Chapter Title Changes to previous version
All n.a., new version!
7 - 1
Siemens AG SPB7-230.814.01 Page 1 of 2 Opdima Medical Solutions Rev. 01 10.04 CS PS 24 Installation
Page 44
7 - 2 Changes to previous version
This page intentionally left blank.
Opdima SPB7-230.814.01 Page 2 of 2 Siemens AG Installation Rev. 01 10.04 CS PS 24 Medical Solutions