0 - 2Revision
ChapterPageRevision
AllAll01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Opdima SPB7-230.814.01Page 2 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
7 ______ Changes to previous version _____________________________________7 - 1
Opdima SPB7-230.814.01Page 4 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Prerequisites1
General1
This installation instruction is valid for MAMMOMAT Novation
> 1100.
DR
with serial numbers
1 - 1
• The wall socket for the power supply for the computers of Opdimar must be connected
DR
to the same branch circuit as the MAMMOMAT Novation
power supply
• This Installation and Start-Up Instruction is valid for Opdimar systems:
- with SUN workstation part No. 83 83 106/ Sun Blade 150
No.W 10000
with ASW 3.1 and with serial
Siemens AGSPB7-230.814.01Page 1 of 4Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
1 - 2Prerequisites
Training of customer support engineers1
Due to the technology used in this equipment, setup, service and maintenance may only
be carried out by a customer support engineer who has attended a training workshop or
has participated in at least one installation.
Text emphasis1
DANGER
WARNING
CAUTION
NOTICE
NOTE
DANGER indicates when there is an immediate danger that leads
to death or serious physical injury.
WARNING indicates a risk of danger that may lead to death or to
serious physical injury.
CAUTION used with the safety alert symbol indicates a risk of
danger that leads to slight or moderate physical injury and/or
damage to property.
NOTICE used without the safety alert symbol indicates a risk of
danger that if disregarded leads or may lead to a potential situation which may result in an undesirable result or state other than
death, physical injury or damage to property.
NOTE contains information provided with special emphasis to
facilitate proper use of the equipment or proper execution of a
procedure, i.e. hints, tips
.
Opdima SPB7-230.814.01Page 2 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Prerequisites1 - 3
Documents required1
• Supplement to the Instructions for Use MAMMOMAT Novation
(included in the Opdimar delivery)
• MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Attachment and Opdimar Measurement Certificates
(included in the Opdimar delivery)
• MAMMOMAT Novation
• MAMMOMAT Novation
• MAMMOMAT Novation
DR
- Opdimar Planning Guide
DR
- Opdimar Installation and Start-Up
DR
- Opdimar Wiring Diagram
• MAMMOMAT 3000 Installation and Setting Instructions Update Kit for Stereotactic
Biopsy Attachment (where applicable)
DR
- Opdimar
Meters and appliances required1
• Protective ground wire tester (44 15 899 RV090)
• Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to
generator (part No. 99 00 440 RE999)
• Stereo calibration phantom (included in the delivery) (part No.64 30 701)
• AEC calibration plexiglass, four plates measuring 150 mm x 150 mm x 19 mm and one
plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224
respectively
• Resolution phantom with at least 10 line pairs per mm
Tools required1
• Standard installation tools
• Electric drill with 5 mm drill
Additional installation material required1
• Extension mains cord with at least three outlets, of protective earth type
Start-up, DHHS and X-ray decree area of application1
Notes on start-up1
Opdimar is adjusted, programmed and tested in the factory, leaving only the connection
to the on-site mains voltage, the functional tests and customer dependent adjustments to
be performed.
Siemens AGSPB7-230.814.01Page 3 of 4Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
1 - 4Prerequisites
Checking and recording for the area of application of the X-ray decree
(§16 Germany)1
An X-ray unit can only be put into operation if an acceptance test has been performed
according to §16 RoeV. It is recommended that the responsible Customer Service
Engineer is present to assist the person responsible for this acceptance test.
Checking and data recording in the USA1
Maintenance measurements must be made according to the instructions in MAMMOMAT
300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdimar
DHHS Maintenance Instructions.
Sections
• Required labels
• Reproducibility
• Automatic exposure control (AEC)
The result must be recorded in the document MAMMOMAT 300/3000, MAMMOMA T 3000
Modular incl. Stereotactic Biopsy Attachment and Opdimar Measurement Certificates.
Opdima SPB7-230.814.01Page 4 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Protective measures for CCD camera2
2 - 1
CAUTION
The camera shall be used within 10230i C.
The camera shall be transported or stored within 0240i C.
The CCD camera is very sensitive to mechanical shocks and temperature changes. In the camera shock and temperature sensors
are integrated.
Risk of damaging the equipment.
The camera has to be handled with extreme care.
When disconnected from the biopsy controller, the camera shall
always be stored in the attaché case delivered with the system.
Do not touch the pins in the contacts of the camera.
Siemens AGSPB7-230.814.01Page 1 of 2Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
2 - 2Protective measures for CCD camera
This page intentionally left blank.
Opdima SPB7-230.814.01Page 2 of 2Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Protective measures3
It is very important that any intervention in the equipment shall start with disconnecting it
from the power supply with the main circuit breaker. To prevent accidental triggering of
high voltage and radiation, set the switch S2 (SS) on board D750 to OFF.
3 - 1
CAUTION
WARNING
WARNING
When switching off the workstation use the power off procedure
described in the Supplement to the Instructions for Use MAMMOMAT Novation
Switching off the workstation before the software has been
closed down may cause damage to the files on the hard disc.
If the system is only switched off at the control panel or with the
circuit breaker CB S700, there is still voltage present within the
generator. See the wiring diagram for MAMMOMAT Novation
Life-threatening electric shock hazard exists.
Disconnect mains cable and comply with the information on this
page.
After shut-down of the system, there may still be about 380 V DC
present on the intermediate circuit of the MAMMOMAT generator.
Life-threatening electric shock hazard exists.
DR
- Opdimar.
DR
.
CAUTION
A bleader resistor is connected over the output DC capacitors.
The DC voltage will discharge below 60 Volt in 3 minutes.
The voltage level is indicated by LEDs H3 and H4.
Observe the currently valid guidelines for handling electronics
endangered by electrostatic discharge.
Use ESD-equipment, ground prior to making contact and place
the components on a conductive surface.
The boards contain electrostatic highly sensitive components
requiring particular care in their handling.
Risk of damaging components.
Siemens AGSPB7-230.814.01 Page 1 of 2Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
3 - 2Protective measures
This page intentionally left blank.
Opdima SPB7-230.814.01Page 2 of 2Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Preparatory work4
Scope of delivery4
The following parts are always included in the delivery of your equipment:
• Technical Manual comprising information for installation, operation and service
• CCD camera with a fixed cable
• Biopsy controller with five cables and cable duct
• Stereo calibration phantom, 64 30 701 (see page 1-3)
• Workstation including:
- power supply cable
- network cable
- KVM switch with cables
- MO unit with one power supply cable and one data cable
- Main unit including CD drive
- In some deliveries, the monitor, keyboard and mouse may be delivered. In such case
the KVM switch is not needed.
If ordered, the following is delivered:
• Biopsy unit
4 - 1
• Compression plate for biopsy examination
• Printed circuit board D702
• MAMMOMAT software V2.2 or higher
• Isolation p.c. board D707 kit
• DICOM license key
• Workstation table
• Printer
Siemens AGSPB7-230.814.01 Page 1 of 2Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
4 - 2Preparatory work
Unpacking4
General4
NOTE
Follow the directional marks on the box during transport, storage
and unpacking.
Workstation4
The workstation is packed in a card board box which also contains technical
documentation and CDs for application software and operating system
1
.
Biopsy controller, miscellaneous4
The biopsy controller is packed in a card board box together with cables, miscellaneous
installation material.
CCD camera
CAUTION
The CCD camera is very sensitive to mechanical shocks and temperature changes. In the camera shock and temperature sensors
are integrated.
Risk of damaging the equipment.
The camera has to be handled with extreme care.
When disconnected from the biopsy controller, the camera shall
always be stored in the attaché case delivered with the system.
Do not touch the pins in the contacts of the camera.
The CCD camera is packed in an attaché case. Keep the case for future use.
Optional equipment and update kits4
If optional equipment and/or update kits are ordered, these are packed in separate boxes.
1. Only needed if a software reinstallation is necessary (see MAMMOMAT 1000/3000 Nova - Opdimar Service
Instructions). Store the documentation and software CDs in a proper place for future service use.