0 - 2Revision
ChapterPageRevision
AllAll01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
Opdima SPB7-230.814.01Page 2 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
7 ______ Changes to previous version _____________________________________7 - 1
Opdima SPB7-230.814.01Page 4 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 5
Prerequisites1
General1
This installation instruction is valid for MAMMOMAT Novation
> 1100.
DR
with serial numbers
1 - 1
• The wall socket for the power supply for the computers of Opdimar must be connected
DR
to the same branch circuit as the MAMMOMAT Novation
power supply
• This Installation and Start-Up Instruction is valid for Opdimar systems:
- with SUN workstation part No. 83 83 106/ Sun Blade 150
No.W 10000
with ASW 3.1 and with serial
Siemens AGSPB7-230.814.01Page 1 of 4Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 6
1 - 2Prerequisites
Training of customer support engineers1
Due to the technology used in this equipment, setup, service and maintenance may only
be carried out by a customer support engineer who has attended a training workshop or
has participated in at least one installation.
Text emphasis1
DANGER
WARNING
CAUTION
NOTICE
NOTE
DANGER indicates when there is an immediate danger that leads
to death or serious physical injury.
WARNING indicates a risk of danger that may lead to death or to
serious physical injury.
CAUTION used with the safety alert symbol indicates a risk of
danger that leads to slight or moderate physical injury and/or
damage to property.
NOTICE used without the safety alert symbol indicates a risk of
danger that if disregarded leads or may lead to a potential situation which may result in an undesirable result or state other than
death, physical injury or damage to property.
NOTE contains information provided with special emphasis to
facilitate proper use of the equipment or proper execution of a
procedure, i.e. hints, tips
.
Opdima SPB7-230.814.01Page 2 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 7
Prerequisites1 - 3
Documents required1
• Supplement to the Instructions for Use MAMMOMAT Novation
(included in the Opdimar delivery)
• MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions
Attachment and Opdimar Measurement Certificates
(included in the Opdimar delivery)
• MAMMOMAT Novation
• MAMMOMAT Novation
• MAMMOMAT Novation
DR
- Opdimar Planning Guide
DR
- Opdimar Installation and Start-Up
DR
- Opdimar Wiring Diagram
• MAMMOMAT 3000 Installation and Setting Instructions Update Kit for Stereotactic
Biopsy Attachment (where applicable)
DR
- Opdimar
Meters and appliances required1
• Protective ground wire tester (44 15 899 RV090)
• Service PC (e.g. Siemens-Nixdorf PCD 3NSX/20 or similar) with connecting cable, PC to
generator (part No. 99 00 440 RE999)
• Stereo calibration phantom (included in the delivery) (part No.64 30 701)
• AEC calibration plexiglass, four plates measuring 150 mm x 150 mm x 19 mm and one
plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224
respectively
• Resolution phantom with at least 10 line pairs per mm
Tools required1
• Standard installation tools
• Electric drill with 5 mm drill
Additional installation material required1
• Extension mains cord with at least three outlets, of protective earth type
Start-up, DHHS and X-ray decree area of application1
Notes on start-up1
Opdimar is adjusted, programmed and tested in the factory, leaving only the connection
to the on-site mains voltage, the functional tests and customer dependent adjustments to
be performed.
Siemens AGSPB7-230.814.01Page 3 of 4Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 8
1 - 4Prerequisites
Checking and recording for the area of application of the X-ray decree
(§16 Germany)1
An X-ray unit can only be put into operation if an acceptance test has been performed
according to §16 RoeV. It is recommended that the responsible Customer Service
Engineer is present to assist the person responsible for this acceptance test.
Checking and data recording in the USA1
Maintenance measurements must be made according to the instructions in MAMMOMAT
300/3000, MAMMOMAT 3000 Modular incl. Stereotactic Biopsy Attachment and Opdimar
DHHS Maintenance Instructions.
Sections
• Required labels
• Reproducibility
• Automatic exposure control (AEC)
The result must be recorded in the document MAMMOMAT 300/3000, MAMMOMA T 3000
Modular incl. Stereotactic Biopsy Attachment and Opdimar Measurement Certificates.
Opdima SPB7-230.814.01Page 4 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 9
Protective measures for CCD camera2
2 - 1
CAUTION
The camera shall be used within 10230i C.
The camera shall be transported or stored within 0240i C.
The CCD camera is very sensitive to mechanical shocks and temperature changes. In the camera shock and temperature sensors
are integrated.
Risk of damaging the equipment.
The camera has to be handled with extreme care.
When disconnected from the biopsy controller, the camera shall
always be stored in the attaché case delivered with the system.
Do not touch the pins in the contacts of the camera.
Siemens AGSPB7-230.814.01Page 1 of 2Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 10
2 - 2Protective measures for CCD camera
This page intentionally left blank.
Opdima SPB7-230.814.01Page 2 of 2Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 11
Protective measures3
It is very important that any intervention in the equipment shall start with disconnecting it
from the power supply with the main circuit breaker. To prevent accidental triggering of
high voltage and radiation, set the switch S2 (SS) on board D750 to OFF.
3 - 1
CAUTION
WARNING
WARNING
When switching off the workstation use the power off procedure
described in the Supplement to the Instructions for Use MAMMOMAT Novation
Switching off the workstation before the software has been
closed down may cause damage to the files on the hard disc.
If the system is only switched off at the control panel or with the
circuit breaker CB S700, there is still voltage present within the
generator. See the wiring diagram for MAMMOMAT Novation
Life-threatening electric shock hazard exists.
Disconnect mains cable and comply with the information on this
page.
After shut-down of the system, there may still be about 380 V DC
present on the intermediate circuit of the MAMMOMAT generator.
Life-threatening electric shock hazard exists.
DR
- Opdimar.
DR
.
CAUTION
A bleader resistor is connected over the output DC capacitors.
The DC voltage will discharge below 60 Volt in 3 minutes.
The voltage level is indicated by LEDs H3 and H4.
Observe the currently valid guidelines for handling electronics
endangered by electrostatic discharge.
Use ESD-equipment, ground prior to making contact and place
the components on a conductive surface.
The boards contain electrostatic highly sensitive components
requiring particular care in their handling.
Risk of damaging components.
Siemens AGSPB7-230.814.01 Page 1 of 2Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 12
3 - 2Protective measures
This page intentionally left blank.
Opdima SPB7-230.814.01Page 2 of 2Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 13
Preparatory work4
Scope of delivery4
The following parts are always included in the delivery of your equipment:
• Technical Manual comprising information for installation, operation and service
• CCD camera with a fixed cable
• Biopsy controller with five cables and cable duct
• Stereo calibration phantom, 64 30 701 (see page 1-3)
• Workstation including:
- power supply cable
- network cable
- KVM switch with cables
- MO unit with one power supply cable and one data cable
- Main unit including CD drive
- In some deliveries, the monitor, keyboard and mouse may be delivered. In such case
the KVM switch is not needed.
If ordered, the following is delivered:
• Biopsy unit
4 - 1
• Compression plate for biopsy examination
• Printed circuit board D702
• MAMMOMAT software V2.2 or higher
• Isolation p.c. board D707 kit
• DICOM license key
• Workstation table
• Printer
Siemens AGSPB7-230.814.01 Page 1 of 2Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 14
4 - 2Preparatory work
Unpacking4
General4
NOTE
Follow the directional marks on the box during transport, storage
and unpacking.
Workstation4
The workstation is packed in a card board box which also contains technical
documentation and CDs for application software and operating system
1
.
Biopsy controller, miscellaneous4
The biopsy controller is packed in a card board box together with cables, miscellaneous
installation material.
CCD camera
CAUTION
The CCD camera is very sensitive to mechanical shocks and temperature changes. In the camera shock and temperature sensors
are integrated.
Risk of damaging the equipment.
The camera has to be handled with extreme care.
When disconnected from the biopsy controller, the camera shall
always be stored in the attaché case delivered with the system.
Do not touch the pins in the contacts of the camera.
The CCD camera is packed in an attaché case. Keep the case for future use.
Optional equipment and update kits4
If optional equipment and/or update kits are ordered, these are packed in separate boxes.
1. Only needed if a software reinstallation is necessary (see MAMMOMAT 1000/3000 Nova - Opdimar Service
Instructions). Store the documentation and software CDs in a proper place for future service use.
Opdima SPB7-230.814.01Page 2 of 2Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 15
Installation5
General5
5 - 1
Before starting the installation, a backup of the installation parameters of the
DR
MAMMOMAT Novation
can be carried out. This may be omitted, if backup already
available.
1.Connect a PC to the MAMMOMAT Novation
2.Select
Main menu => Backup => Copy installation area to disk => All
DR
and start the service program.
.
This will store the installation parameters on the hard disk of the PC.
3.Select
Main menu => Backup => Copy installation area to floppy => All
.
This will store the installation parameters on the service diskette.
4.Switch off the MAMMOMAT.
Sub-assemblies, overview5
The components shall be placed according to the figure below.
Biopsy unit
X
mm
X
X
mm
Y
mm
Z
Biopsy
controller
Y
Y
Z
Z
X
Z
Y
CCD camera
Stand with
X-ray unit
Workstation
Workstation table
Fig. 1 MAMMOMAT 3000 Nova with digital biopsy and spot imaging system Opdimar
NOTE
The workstation shall be placed at least 1.5 m from the object
table of the stand.
Generator
OPD00597
Siemens AGSPB7-230.814.01 Page 1 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 16
5 - 2Installation
Removal of MAMMOMAT NovationDR stand covers5
5
WARNING
The right/rear side cover should be removed.
The holes for fastening of the biopsy controller BC and of the cable tray below the BC
have been drilled in the factory.
There are two cable trays of different lenghts delivered, use the suitable one.
High voltage!
Life-threatening electric shock hazard exists.
Switch off the power to the system with the main circuit breaker.
OpdimaSPB7-230.814.01Page 2 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 17
Installation5 - 3
Installation of biopsy controller holders5
The biopsy controller is attached to the rear cover of the stand. To install the holders
proceed as follows:
1.Install the three holders (2/Fig. 2) by using the nuts (3/Fig. 2).
2.Install the two standoffs (4/Fig. 2) by using the screws (5/Fig. 2) and the washers
(6/Fig. 2).
UP
985 – 5 mm
2
Fig. 2 Rear cover, from the back side
A
3
4
B
6
5
1
OPD00267
Siemens AGSPB7-230.814.01 Page 3 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 18
5 - 4Installation
3.Install the rear cover (2/Fig. 3) on the stand and attach the biopsy controller
(1/Fig. 3) to the holders. Switch on the power-on switch on the biopsy controller.
2
1
OPD00152
Fig. 3 Biopsy controller
OpdimaSPB7-230.814.01Page 4 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 19
Installation5 - 5
Connection of cables5
The biopsy controller is the central unit for the cabl e connections of OPDIMA. The cables
are connected to the biopsy controller according to the figure 6 on page 2-6, see wiring
diagram doc No: SPB7-230.051.09.03.02.
Perform the connection of cables according to data in the wiring diagram.
Siemens AGSPB7-230.814.01 Page 5 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 20
5 - 6Installation
X201
X202
X203
Ferrite
X204
OPD00453
5
X205X206X207Mains cable
Fig. 4 Connection of cables to the biopsy controller
Protective earth cable
Cables to CCD camera5
The CCD camera including cable with connectors marked X201 and X202 is connected to
the biopsy controller.
1.Connect the cable to X201 and X202 on the biopsy controller, see Fig. 4.
CAUTION
2.Remove the screws (1/Fig. 5) and remove the cover (2/Fig. 5) of the object table
arm.
Sensitive electronic equipment!
Risk of damaging the camera.
Do not touch the pins in the contacts of the CCD camera.
12
Fig. 5 Object table arm before installation
OpdimaSPB7-230.814.01Page 6 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
OPD00161
Page 21
Installation5 - 7
1.Install a new thread (1/Fig. 6).
1
Fig. 6 Installation of thread
OPD00162
2.Install the holder (1/Fig. 7) with the pull relief device, using included Allen screws
(2/Fig. 7).
21
OPD00163
Fig. 7 Object table arm after installation
Siemens AGSPB7-230.814.01 Page 7 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 22
5 - 8Installation
Installation of cable duct5
Four cables are laid from the bottom of the biopsy controller in a cable duct to the cable
outlet cover (3/Fig. 8). To install the cable duct on the rear side of the stand proceed as
follows:
1.Attach the cable duct (2/Fig. 8) to the rear cover by removing the protective strips
from the attached Velcro strips and place the duct flush with the biopsy controller
(1/Fig. 8). Secure the cable duct to the two standoffs (5/Fig. 8) with the delivered
screws and washers.
UP
1
4
5
2
3
OPD00164
Fig. 8 Cable duct
50mm
25mm
Modification of cable outlet cover5
To allow the cables to pass through the cable outlet cover, the cover has to be modified
according to the following:
1.Cut an opening centrally on the cable outlet cover (4/Fig. 8). The opening should
be approximately 50 x 25 mm.
2.Attach the supplied plastic strip around the opening of the cable outlet cover
(4/Fig. 8).
3.Install the cable outlet cover (4/Fig. 8).
4.Mount the previously removed stand covers.
OpdimaSPB7-230.814.01Page 8 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 23
Installation5 - 9
Installation of workstation5
Preparing the workstation table (option)5
The workstation table (option), a motorized vertically adjustable table , shall be used to put
the workstation components to the Opdimar system on. The vertical adjustment device
includes a transformer, which is not mounted on delivery.
CAUTION
Handschalter
5
Fig. 9 Transformer
The workstation table must not be overloaded.
Any additional equipment is added at your own risk and the maxi-
mum load of 80 kg must not be exceeded.
2
3
1
4
Opd00494
1.Remove the screws on the underside of the tabletop and use them when
mounting the transformer (1/Fig. 9).
Place the transformer so that the connectors are turned to the column.
2.Connect the cable from the control unit (2/Fig. 9) for the vertically adjustment to
the connector marked “Handschalter” (3/Fig. 9).
3.Connect the cable from the motor (4/Fig. 9) for the vertically adjustment to the
other connector (5/Fig. 9).
Siemens AGSPB7-230.814.01 Page 9 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 24
5 - 10Installation
4.Place the workstation table outside the patient environment of safety reasons.
Patient environment is determined by IEC 60601-1-1 as 1.5 m away from the
patient.
Cables
8 m
Cable
1,5 m
Biopsy
controller
m
5
,
1
.
n
i
M
Cable (workstation table) 2 m
CCD camera holder
Stand
Biopsy unitGenerator
Workstation
150
CCD camera
Cable duct
Fig. 10 System site overview
NOTE
OPD00599
There must be enough space around the workstation table so that
it safely can be vertically adjusted.
OpdimaSPB7-230.814.01Page 10 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 25
Installation5 - 11
5.If the workstation table stands on an uneven floor, it can be adjusted with the
adjustable screws on the framework.
BUSY
POWER
Fig. 11 MO Shelf
6.Place the workstation, the monitor, the keyboard and the mouse on the tabletop.
Make sure that all the equipment is within the edges of the tabletop.
1
EJECT
OPD00611
7.Let the cables from the workstation, the monitor, the keyboard and the mouse
pass through the hole at the back of the tabletop.
8.Mount the MO unit shelf on the right below the tabletop (1/Fig. 11).
9.Place the MO unit to the right on the shelf (1/Fig. 11).
Siemens AGSPB7-230.814.01 Page 11 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 26
5 - 12Installation
Connecting the workstation5
1.Connect workstation cable marked X203 to the biopsy controller connector X203,
see Fig. 12, and the other end (X101) of the cable to the workstation connector
(2B/Fig. 26).
CAUTION
Connecting the cable to the wrong connector may cause severe
damage to the equipment.
2.Attach ferrite on the cable on the biopsy controller side, see Fig. 12.
X201
X202
X203
Ferrite
X204
OPD00453
X205X206 X207Mains cable
Fig. 12 Biopsy controller
Protective earth cable
3.Connect the workstation cable marked X204 to the biopsy controller connector
X204 in Fig. 12 and the other end of the cable to the workstation connector
A (8/Fig. 17) and B (2A/Fig. 17).
4.Attach ferrite on the cable on the biopsy controller side, see Fig. 12.
5.Mount the touch protection over the connectors X203 and X204.
-Place the plastic plate (1/Fig. 13) behind the connectors X203 and X204.
-Place the plastic cover (2/Fig. 13) on the other side of the connectors.
-Fasten the plastic screws (3/Fig. 13).
3
1
2
Biopsy controller
Fig. 13 Mounting the touch protection
OPD0
OpdimaSPB7-230.814.01Page 12 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 27
Installation5 - 13
6.Connect the MO unit SCSI cable to the workstation SCSI connector (2C/Fig. 17).
7.Connect the MO unit cable to one of the two connectors of the MO unit. Fit the
terminator to the other connector of the MO unit.
For switch settings, SCSI ID and termination see Paragraph “Installing the MO
unit” in Chapter “Removal and replacement of sub-assemblies” in
Mammomat 1000/3000/3000 Nova - Opdimar Service Instructions.
CAUTION
Make sure that the SCSI cables are properly connected and fixed
to the units and that the cable is not bent too much just behind
the MO unit. Also, check that the MO unit is terminated correctly.
If disregarded, this might generate a lot of problems e.g. failure to
write images to MO disk, corrupted MO disk or corrupted hard
disk.
8.Connect the mouse to the keyboard.
9.Connect keyboard and mouse (3/Fig. 17).
10.Connect monitor cable to the workstation connector (6/Fig. 17).
11.Connect Ethernet cable (4/Fig. 17).
12.Mark the wall socket connected to the same branch circuit as the MAMMOMAT
(Fig. 14) with label according to Fig. 15.
The power distribution to the system is shown in Fig. 14.
1-phase
L1 N
PE
10A
2-phase
L1L2
10A
N
PE
Off
On
Power On-Off breaker
Fig. 14 Power distribution
Workstation
Power connection
in MAMMOMAT 3000-stand
Off
On
Power On-Off breaker
Power connection
in MAMMOMAT 3000-stand
Workstation
OPD00166
Siemens AGSPB7-230.814.01 Page 13 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 28
5 - 14Installation
13.If a voltage selector switch exists, ensure the line voltage selector switch is set to
the appropriate setting: 115V or 230V.
CAUTION
Plugging a 115V power cord into a 230V connector will severely
damage the system.
14.Connect the AC power cords of the main unit, the monitor, the MO unit, the local
Opdimar printer (if present) and the MAMMOMAT printer (if present) to the wall
socket marked with label.
It is advisable to use an extension cord with at least three outlets.
CAUTION
The workstation and the printer must be grounded to the same
potential as the MAMMOMAT.
Connect the wall socket for the workstation and printer power
supply to the same branch circuit as the MAMMOMAT power supply.
(The workstation table can be connected to any branch circuit.)
Opdima mains power
supply outlet!
OPD00600
Fig. 15 Mains marking label
OpdimaSPB7-230.814.01Page 14 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 29
Installation5 - 15
Front panel description
PositionExplanation
1Power switch
2Power-indicator LED
3Smart card reader (not used in Opdima)
43.5-inch diskette drive (not used in Opdima)
55.25-inch CD drive
2
1
Fig. 16 Front panel overview
Back panel description and connector symbols
PositionExplanationConnector
1Mains power supplyNone
2ABiopsy controller interface (X204)PCI-3
2BCamera interface via biopsy controller (X203)PCI-2
2CSCSI interfacePCI-1
3Universal serial bus (USB) connectors (four)
(keyboard and mouse interface)
4Network connector
5Not applicable (IEEE 1394 (two) connectors), not used
6Monitor interface
7Parallel port to local printer
8Mammomat interface via biopsy controller (X204)
9Audio module headphones connector, not used
9Audio module line-out connector, not used
9Audio module line-in connector, not used
9Audio module microphone connector, not used
3
4
5
OPD00613
symbols
SERIAL
OPD00612
1
2A
Fig. 17 Back panel overview
Siemens AGSPB7-230.814.01 Page 15 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
2B
2C
4356789
Page 30
5 - 16Installation
68
Fig. 18 Port extenders to the back panel of the workstation
OPD00629
OpdimaSPB7-230.814.01Page 16 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 31
Installation5 - 17
Strain relieving the cables5
If the workstation table (option) is used with the Opdima system, the cables from the
workstation components and from the MAMMOMAT must be strain relieved by cable
clamps at the back of the MO unit shelf according to the figure below.
• To ensure that the cables have enough length for the vertical adjustment, let the
workstation table be in its highest possible position when strain relieving the cables.
• Use cable ties to arrange the cables in an organized way and make sure that the cables
can run without a risk of getting caught.
Fig. 19
MO AC
power
WS Mains
power supply
Biopsy
controller
Printer/
Network
connector
A1BCDEFGH
SCSI ID
0
SCSI CONNECTOR
Camera
controller
AC IN
F.GND
Opd00630
Siemens AGSPB7-230.814.01 Page 17 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 32
5 - 18Installation
Installation of input/output devices5
Since the MAMMOMAT Novation DR and the Opdimar use monitors, keyboards and
mouse, it was decided to install:
- one monitor
- one keyboard and
- one mouse.
There is a KVM switch delivered connecting these three input/output devices alternatingly
either to the WH AWS of the MAMMOMAT Novation DR or to the SUN-Blade 150 computer workstation of the Opdimar.
If reprogramming of the KVM switch is necessary follow the description below.
OpdimaSPB7-230.814.01Page 18 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 33
Installation5 - 19
Installation of monitor5
Monitor type: Siemens TFT-LCD monitor 1897
Material required: Mains cable, Cable BNC to HD15 or 13W3(Sun).
Proceed as follows to install the monitor:
1.Unpack the monitor and place it on top of the workstation.
2.Connect the monitor to the workstation with the accompanying cables to the
connectors on the backside of the monitor according to Fig. 20.
Sun Ultra 1, Sun Ultra 10:
13W3-4 BNC, Siemens Part. no. 66 33 684
R:Red cable
G:Green cable
B: Blue cable
H (H/W):Sync, black cable
Fig. 20 Connectors between workstation and monitor
3.Start up the workstation and log in as service, select service and
start Advanced service
3 Software Upgrade
6 Miscellaneous
5 Select Monitor
type: 67 Hz SunBlade 150.
Select Apply and close to close down the rest of the windows.
4.Log out and log in again as service.
Fine adjusting the monitor settings5
Read the monitor manual “High-Resolution 18” LCD Monitor SCD 1898-0” (delivered with
the monitor) and especially the chapter for OSD (On Screen Display).
R:Red cable
G:Green cable
B: Blue cable
H (S):Horizontal/(vertical) sync, white cable
V:Vertical sync, black cable
Network cable: 230 Volt
OPD00631
Use the SMPTE test pattern image as reference when calibrating the image geometry of
the monitor (see the Opdima Quality Control Manual on how to do so). To compensate for
movement in the grid pattern, use frequency/phase under others (settings).
Siemens AGSPB7-230.814.01 Page 19 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 34
5 - 20Installation
If the monitor is wrongly adjusted:
Make an autoadjust under the heading Auto.
Control that the monitor has the correct settings for vertical sync under the Others set-
tings, see Fig. 21.
OPD00632
Fig. 21 Monitor settings
The settings shall be:
Brightness Contrast:
Brightness50
Contrast50
Backlight70
Color2
Position Zoom:
Fill all
Source:
Analog RGB
Auto:
Auto brightness:off
Auto position zoom:on
Auto frequency phase:on
Auto alloff
Language:
English
OpdimaSPB7-230.814.01Page 20 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 35
Installation5 - 21
Others:
select Frequency phase,
Frequency:64*
Phase160*
* Can be adjusted for optimal image reproduction, see the monitor manual “High-Resolu-
tion 18” LCD Monitor SCD 1898-0” (delivered with the monitor). Use the SMPTE test pattern image for fine adjustments of the image.
Vertical:Max
Background:Translucent
Timeout:20 Sec.
Make a final evaluation of the monitor settings with the clinical image as reference. See
Quality Control Manual.
Connecting the workstation to the network5
The Opdima workstation can be connected to a network, partly for the use of DICOM and
also for networked printers.
1.Connect one end of the network cable (included in the delivery) to the workstation.
2.Connect the other end of the network cable to a network socket.
3.Log in with the user name “service”. The password is obtained from Siemens
Service Centre.
4.Press the Service button in the mode selection dialog, see Selection of mode in
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
5.Press the Advanced service button in the service dialog, see Use of Advanced
service functions in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions.
6.In Advanced service configure the network parameters, see Network settings in
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
For physical limitations of the network see Mammomat 1000/3000/3000 Nova Opdimar Planning Guide.
Siemens AGSPB7-230.814.01 Page 21 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 36
5 - 22Installation
Installation of Opdima printer (option)5
Stand alone Opdima system:
When the Opdima system is stand alone (not connected to a network) it is possible to
install a Codonics NP-1660 or NP-1600 printer to Opdimar as a stand alone printer
connected to the network connector.
NOTE
The following equipment is needed:
It is also possible to connect a stand alone postscript printer to
the parallel or network connector of the workstation, see Use of
printer setup in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions.
• Codonics printer NP-1660 or NP-1600
• Crossover network cable
Proceed as follows to install the printer:
1.Connect one end of the network cable to the workstation (2C/Fig. 17).
2.Connect the other end of the network cable to the printer.
3.Connect the mains cable of the printer to the wall socket marked with label,
see Fig. 15.
CAUTION
The printer must be grounded to the same potential as the
MAMMOMAT.
Connect the printer power supply to the same branch circuit as
the MAMMOMAT power supply.
4.Set the printer IP address to 10.10.10.2 see the User’s Manual for the Codonics
NP-1660 or NP-1600.
5.Select Scaled in printer setup in the advanced service dialog, see Stand alone
Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions.
Networked Opdima system:
When the Opdimar system is networked it is possible to configure the system to print to a
printer connected to the network or a local postscript printer connected to the parallel
interface of the workstation.
NOTE
It is also possible to connect a stand alone postscript printer to
the parallel connector of the workstation, see Select printer in
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions.
OpdimaSPB7-230.814.01Page 22 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 37
Installation5 - 23
Proceed as follows to install the network printer:
1.Add access to the network printer in the advanced service dialog, see Networked
Opdima system in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions.
2.Select printer in the advanced service dialog, see Networked Opdima system in
MAMMOMAT 1000/3000/3000 Nova - Opdimar Service Instructions.
Functional test (workstation table)5
Before the installed workstation table (option) is handed over to the customer, check that
the workstation table can be vertically adjusted between the highest and lowest possible
positions properly and safety.
Siemens AGSPB7-230.814.01 Page 23 of 24Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 38
5 - 24Installation
This page intentionally left blank.
OpdimaSPB7-230.814.01Page 24 of 24Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 39
Final procedures6
General6
The following procedures shall be performed:
• Starting up the workstation, see Page 6 - 2
• Country settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• Test of units,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• CCD camera calibration,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• Calibration of the biopsy unit,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
6 - 1
• Verifying the calibration of the biopsy unit,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Check of Opdimar AEC,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Check of resolution,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Network settings,
see chapter Service mode in MAMMOMAT 1000/3000/3000 Nova - Opdimar Service
Instructions
• Activate DICOM (optional),
see DICOM option in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova Opdimar Service Instructions
• Printer setup (optional),
see Use of printer setup in chapter Service mode in MAMMOMAT 1000/3000/3000 Nova
- Opdimar Service Instructions
• Mounting of covers, see Page 6 - 3
• Protective earth measurement,
see chapter Measures after service in MAMMOMAT 1000/3000/3000 Nova - Opdimar
Service Instructions
• Fill in the IVK list or report the IVK components to your responsible uptime service center
(USC).
Siemens AGSPB7-230.814.01 Page 1 of 4Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 40
6 - 2Final procedures
• Performing of DHHS maintenance, see Page 6 - 3
NOTICE
The Opdimar external diaphragm must be used for tests and calibrations involving radiation.
Starting up the MAMMOMAT6
Switch on the MAMMOMAT.
Starting up the workstation and login6
The workstation is started according to instructions in Starting up the workstation and
login in Supplement to the Instructions for Use MAMMOMAT Novation
Use the user name “service” and the password obtained from Siemens service center
when entering the system.
NOTE
To obtain optimal picture, adjust the monitor settings as follow:
Any adjustments of the monitor settings shall be performed by
qualified personell who are employed by Siemens or one of its
affiliates or who are otherwise autorized by Siemens or one of its
affiliates to provide such services.
DR
- Opdimar.
Brightness: 50
Contrast: 50
Backlight: 70
Starting up the MO unit6
OPD00601
Fig. 1 Magneto-optical drive (SONY)
1.Insert the MO disk, type 9.1 GB MO disk
2.For switch settings and SCSI ID, see Paragraph “ Installing the MO unit” in
Chapter “Removal and replacement of sub-assemblies” in
MAMMOMAT 1000/3000 Nova - Opdimar Service Instructions.
Installation of software on OPDIMA WORKSTATION6
The software is installed in the factory before deliv ery . If the softw are for an y reason needs
to be reinstalled, refer to chapter “Reinstallation of software” in MAMMOMAT 1000/3000
Nova - Opdimar Service Instructions.
Opdima SPB7-230.814.01Page 2 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 41
Final procedures6 - 3
Mounting of covers6
Mount the covers previously removed.
Filling in of IVK list6
Fill in the IVK list or report the IVK components to your responsible uptime service center
(USC).
Performing of DHHS maintenance6
In the USA, perform maintenance according to DHHS Maintenance Instructions for
MAMMOMAT 1000/ 3000 Nova incl. Stereotactic Biopsy Attachment and Opdimar and fill
in DHHS Measurement Certificates for MAMMOMAT 1000/ 3000 Nova incl. Stereotactic
Biopsy Attachment and Opdimar.
NOTE
The DHHS Maintenance Instructions and DHHS Measurement
Certificates delivered with the Opdimar also includes the measurements to be performed on the MAMMOMAT. Therefore the
DHHS Maintenance Instructions supplied with the MAMMOMAT
can be removed. However, the DHHS Measurement Certificates
supplied with the MAMMOMAT should be kept for future reference, if it contains previous measurement results.
Siemens AGSPB7-230.814.01 Page 3 of 4Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 42
6 - 4Final procedures
This page intentionally left blank.
Opdima SPB7-230.814.01Page 4 of 4Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions
Page 43
Changes to previous version7
ChapterTitleChanges to previous version
Alln.a., new version!
7 - 1
Siemens AGSPB7-230.814.01 Page 1 of 2Opdima
Medical SolutionsRev. 0110.04CS PS 24Installation
Page 44
7 - 2Changes to previous version
This page intentionally left blank.
Opdima SPB7-230.814.01Page 2 of 2Siemens AG
InstallationRev. 0110.04CS PS 24Medical Solutions