The reproduction, transmission or use
of this document or its contents is not
permitted without express written
authority. Offenders will be liable for
damages. All rights, including rights
created by patent grant or registration
of a utility model or design, are
reserved.
English
Doc. Gen. Date: 08.04
2004
2Revision / Disclaimer
1Revision / Disclaimer
Document revision level
The document corresponds to the version/revision level effective at the time of system
delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified
personnel who are employed by Siemens or one of its affiliates or who are otherwise
authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated
with or authorized by Siemens or one of its affiliates are directed to contact one of the
local offices of Siemens or one of its affiliates before attempting installation or service procedures.
6 _______ Changes since last version_______________________________________ 35
MAMMOMAT NovationSPB7-250.812.03.02.02Siemens AG
08.04CS SD 24
Page 4 of 36
Medical Solutions
Prerequisites5
1Prerequisites
2-
Configuration0
The configuration of the MAMMOMAT NovationDR installation depends on the customer"s
choice.
Tab. 1Configuration
MAMMOMAT Novation
DR
⇓
Integrated Generator
⇓
Molybdenum/tungsten anode X-ray
tube assembly with filter disc
Mo 0.03/Rh 0.025/Rh 0.050
⇓
Dose Calculation System
⇓
Control console
with integrated radiation
shield (optional)
⇓
Acquisition Workstation
Unless otherwise stated, these instructions describe the
•MAMMOMAT Novation
DR
stand with the integrated generator,
•the radiation shield with the integrated control console
•the detector and
•the syngo Acquisition Workstation.
Depending on the installation configuration, some points may be omitted.
Siemens AGSPB7-250.812.03.02.02MAMMOMAT Novation
Medical Solutions
08.04CS SD 24
Page 5 of 36
6Prerequisites
Tools0
ItemRemarks
Standard installation and service tools
Torque wrench for bolting the stand/console
to the floor
Electrical screwdriver with adjustable torque
is recommended
MAMMOMAT NovationSPB7-250.812.03.02.02Siemens AG
08.04CS SD 24
Page 6 of 36
Medical Solutions
Important Notes7
2Important Notes
3-
Symbols0
XThis symbol indicates exposure of X-ray radiation.
PThis symbol indicates that the measured values have to be entered in the Installation Pro-
tocol (SPB7-250.813.01...).
Siemens AGSPB7-250.812.03.02.02MAMMOMAT Novation
Medical Solutions
08.04CS SD 24
Page 7 of 36
8Important Notes
Factory adjustments0
The MAMMOMAT NovationDR is adjusted, programmed and tested in the factory, leaving the adaptation to the on-site mains voltage, the adjustment of the AEC and the functional tests to be performed.
When the measurements to be made (kV, mAs, etc.) are within the tolerances stated in
these instructions, this confirms that the settings made in the factory have not changed
and the equipment is fully serviceable.
The equipment is set to 400 V, 2-phase on delivery.
MAMMOMAT NovationSPB7-250.812.03.02.02Siemens AG
08.04CS SD 24
Page 8 of 36
Medical Solutions
Important Notes9
Protective measures0
System power supply0
It is very important that any intervention in the equipment starts with disconnecting it from
the power supply with the main circuit-breaker. Before removing or inserting any of the
printed circuit boards, switch off the equipment.
WARNING
WARNING
If the system is only switched off at the control panel, the line voltage will still be present at the generator line connection (see wiring
diagram SPB7-250.844.01...).
A life-threatening hazard of electric shock exists.
¹ Disconnect the mains cable and comply with the infor-
mation on this page.
After shut-down of the system, there may still be 380 V DC present
on the intermediate circuit.
A life-threatening hazard of electric shock exists.
¹ Disconnect the mains cable and comply with the infor-
mation on this page.
PC boards0
CAUTION
The p.c. boards contain electrostatic highly sensitive components.
If disregarded, the components could be damaged.
¹ Use ESD equipment, ground prior to making contact and
place the components on a conductive surface.
Metal curtain0
WARNING
Siemens AGSPB7-250.812.03.02.02MAMMOMAT Novation
Medical Solutions
The edges of the metal curtain of the stand are very sharp.
They may cause severe injury.
¹ Apply the protective strips as mentioned in section(Link
chapter) after removing the covers from the stand.
Remove the protective strips only when the covers are to
be mounted or when vertical adjustment of the swivel
arm system is necessary.
Page 9 of 36
08.04CS SD 24
10Important Notes
Measurements with the oscilloscope0
WARNING
The existing ground conductor in the mains cable must under no
circumstances be disconnected when operating the oscilloscope.
A life-threatening hazard of electric shock exists.
¹ For measurements in which a resulting ground loop may
falsify the measuring result, use the differential amplifier
(difference measurement).
MAMMOMAT NovationSPB7-250.812.03.02.02Siemens AG
08.04CS SD 24
Page 10 of 36
Medical Solutions
Unpacking the system components11
3Unpacking the system com ponents
4-
Scope of delivery0
The MAMMOMAT NovationDR is normally packed in two crates and one cardboard box
(the number of packages depends on the customer"s choice):
DR
The MAMMOMAT Novation
1055T3, 1000 kg/m2 are permissible.
Crate (length: 210 cm, width: 80 cm, height: 150 cm), 360 kg plus 60 kg (pallet)
The crate contains the MAMMOMAT NovationDR stand.
Crate (lenght: 210 cm, width: 80 cm, height: 120 cm)
The crate contains:
•Accessories, e.g. cover panels, WorkStation, compression plates, documentation, in-
stallation material
•Radiation shield
Cardboard boxes (length: 90 cm, width: 80 cm, height: 60 cm), approximately 60 kg
This cardboard box contains the detector.
has an area loading of 800 kg/m2; according to DIN
NOTE
NOTE
The detector package is temperature sensitive, from 10° to 40° Celsius.
The detector package has to be disposed locally, after the installation.
Siemens AGSPB7-250.812.03.02.02MAMMOMAT Novation
Medical Solutions
08.04CS SD 24
Page 11 of 36