English
Print No.: SPB7-230.841.03.02.02 Doc. Gen. Date:04.04
Replaces: SPB7-230.841.03.01.02(66 47 015)
0 - 2Revision
ChapterPageRevision
AllAll02
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel
who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or
one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or
one of its affiliates before attempting installation or service procedures.
M300/1000/3000/3000 NovaSPB7-230.841.03Page 2 of 4Siemens AG
ServiceRev. 0204.04CS PS 24Medical Solutions
4 _______Changes to previous version _____________________________________ 4 - 1
Siemens AGSPB7-230.841.03Page 3 of 4 M300/1000/3000/3000 Nova
Medical SolutionsRev. 0204.04CS PS 24Service
0 - 4Contents
M300/1000/3000/3000 NovaSPB7-230.841.03Page 4 of 4Siemens AG
ServiceRev. 0204.04CS PS 24Medical Solutions
Prerequisites1
General1
This instruction describes:
in Chapter 2 - how to replace the compression mechanism in the compression unit of the
mammography systems MAMMOMAT 300 and MAMMOMAT 1000/3000/3000 Nova.
in Chapter 3 - how to replace the potentiometer R863 - the position transducer of the
compression unit.
Components concerned1
Components concerned are the compression units/devices normally used on
MAMMOMAT 1000/3000/3000 Nova systems, on MAMMOMAT 300 systems and - with
some modifications - also used on MAMMOMAT Novation DR systems.
Material numbers
063 86 051 (used for the Sales Item)
065 21 681 (used for the Sales Item)
064 31 725 (used as Spare Part with REP-Status)
If you encounter compression units with material numbers that are not listed here, apply
this instruction appropriately. If necessary, contact CS HSC 24.
1 - 1
Tools required1
Standard toolkit (including Allen keys with ball-shaped heads).
Resistance Measuring Instrument (e.g. Fluke - with Ohm-meter function).
Documents required1
The technical documentation delivered with the corresponding MAMMOMAT system.
Mainly:Maintenance Instruction SPB7-230.105...
Service Program SPB7-230.114...
NOTICE
Strictly observe all safety precautions - as described in the
relevant technical documents - during work that involves
assembly, disassembly, removal of covers or work under voltage.
Siemens AGSPB7-230.841.03Page 1 of 2M300/1000/3000/3000 Nova
Medical SolutionsRev. 0204.04CS PS 24Service
1 - 2Prerequisites
Parts included1
Repair kit "Compression Mechanism and Potentiometer", Material Number
065 76 214 mainly contains:
(Details in Design Specification 65 76 214 EDS-01S-00)
ItemDescriptionAmountRemark
1Compression mechanism1Assembly
2Tool for compression unit165 76 222
3Spring pin 2.5 x 14 ISO 8752260 53 656, if needed, because
original damaged, see Pos. 98
on Fig. 5.
4Potentiometer, 2 kOhm 1W, 10% lin.162 35 118
5Allen key 1.3 mm1
6This repair instruction1SPB7-230.841.03.02
Time required1
Approximately 1.5 hours for 1 person for exchange of the compression mechanism.
Approximately 3.0 hours for 1 person for exchange of the potentiometer.
M300/1000/3000/3000 NovaSPB7-230.841.03Page 2 of 2Siemens AG
ServiceRev. 0204.04CS PS 24Medical Solutions