Siemens MAMMOMAT 3000 Nova Service Instructions

MAMMOMAT 1000/3000 Nova - Opdimar
Service
Service Instructions
ASW 3.0
© Siemens AG 2002
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Register 5 English Print No.: SPB7-230.061.10.01.02 Doc . Gen. Date: 09.02
Replaces: n.a. 66 31 175
0 - 2 Revision
Chapter Page Revision
All 01
Document revision level
The document corresponds to the version/revision level effective at the time of system delivery. Revi­sions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or autho­rized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempti ng installation or service procedures.
Opdimar Register 5 SPB7-230.061.10 Page 2 of 6 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Contents 0 - 3
Page
1 _______Prerequisites __________________________________________________1 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Training of customer support engineers. . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Documents required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
Ultra 10 CD-ROMs requir e d. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Meters and appliances required. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
2 _______Functional description __________________________________________2 - 1
Block diagram . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Biopsy controller. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
Biopsy unit control . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
RS-232 switching . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
CCD camera power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 2
LEDs. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 2
Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
Application software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 3
CCD camera communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Control of MAMMOMAT and biopsy unit . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Monitor. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
CCD camera. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
Biopsy unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 4
3 _______Protective measures for CCD camera______________________________3 - 1 4 _______Protective measures____________________________________________4 - 1 5 _______Service mode__________________________________________________5 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Advanced service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 1
Selection of mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 2
Calibration of the biopsy unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 3
Setting up needles. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 6
Editing needle values . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 7
Backup of temporary storage media . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 8
Use of Advanced service functions . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 9
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 9
Main menu. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 9
Country settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 11
Test of units . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 12
Software upgrade . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 13
New software version . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 14
Enabling the DICOM option . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 15
Instruction for obtaining the systems host ID and hostname . . . . . . . . . . . 5 - 15
Network settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 17
Miscellaneous . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 18
Restoring data from MO disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 - 20
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 3 of 6 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
0 - 4 Contents
Page
CCD camera calibration and maintenance . . . . . . . . . . . . . . . . . . . . . .5 - 21
Grid table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 21
Non grid table. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 22
Disk cache settings. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 23
Use of printer setup. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 24
Stand alone Opdima system . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 24
Networked Opdima system. . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 25
Log administration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 27
Log inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5 - 28
6 ______ Removal and replacement of sub-assemblies _______________________ 6 - 1
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
MO unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Removing the MO unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 1
Installing the MO unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Tests and Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 2
Biopsy controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Removal of biopsy controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 3
Removal of biopsy controller cover . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 4
Replacement of components in the biopsy control ler. . . . . . . . . . . . . . . . 6 - 4
Cables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 6
Biopsy unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 7
Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 8
Connectors . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 8
CCD camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 - 9
Reinstallation of software. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 - 10
Installation of Operating Solaris 8 Instal lation CD-ROM for,
(Mat. No. 66 33 049) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 - 10
Installation of Opdima Software Insta ll ation CD-ROM (Mat. No. 66 33 064) . . . .6 - 11
Restore hostname and data . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 - 11
Final procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6 - 13
7 ______ Messages _____________________________________________________7 - 1
Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 1
Control panel . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 - 3
8 ______ Fault isolation chart_____________________________________________8 - 1
Fault isolation chart. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 1
Explanations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 - 2
9 ______ Troubleshooting guide __________________________________________9 - 1
General. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 1
Biopsy unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Biopsy unit not responding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Problem with calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 2
Initialization of BC nvram on the D200 board. . . . . . . . . . . . . . . . . . . . 9 - 2
Camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 4
Opdimar Register 5 SPB7-230.061.10 Page 4 of 6 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Contents 0 - 5
Page
Camera not responding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 4
Cannot use camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 4
Workstation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 5
If “Bogus file system” appears. . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 5
Problem with date in database mode . . . . . . . . . . . . . . . . . . . . . . . .9 - 5
Opdima user interface does not appear at log in . . . . . . . . . . . . . . . . . . 9 - 5
No images displayed on monitor . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 5
Tests. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 - 5
Image Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 7
Problems during calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 7
Quadrant difference on patient images . . . . . . . . . . . . . . . . . . . . . . . 9 - 7
Quadrant missing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 7
White line/dot . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 7
MAMMOMAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 8
Cannot use MAMMOMAT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 8
Network problem . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 9
Printer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 10
Printer connected directly to Opdima . . . . . . . . . . . . . . . . . . . . . . . 9 - 10
Printer connected to network . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 10
MO unit / CD drive. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 11
All problems with MO disk . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 - 11
If the message “Failed to store on MO disk” appears . . . . . . . . . . . . . . . 9 - 11
If the message “Cannot read disk Prepare disk?” appears . . . . . . . . . . . . 9 - 11
Cannot communicate with MO unit . . . . . . . . . . . . . . . . . . . . . . . . 9 - 11
10 ______Measures after service _________________________________________10 - 1
Verifying the calibration of the biopsy unit. . . . . . . . . . . . . . . . . . . . . . . 10 - 1
Check of Opdima AEC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 2
Performance Criteria. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 3
Check of resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 4
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 4
Performance Criteria. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 4
Protective earth measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 5
Biopsy unit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 5
Biopsy controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 5
CCD camera. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 5
Biopsy controller cable duct . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10 - 6
11 ______Changes to previous version____________________________________11 - 1 12 ______Appendix 1__________________________________________________ A1 - 1
Database log file. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A1 - 1
13 ______Appendix 2__________________________________________________ A2 - 1
Customer specific data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 1
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 5 of 6 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
0 - 6 Contents
Test protocol repetitiveness. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 1
Biopsy calculations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 1
Test protocol CCD camera calibration . . . . . . . . . . . . . . . . . . . . . . . . A2 - 2
Test protocol image quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 3
AEC function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 3
Resolution . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 5
Test protocol protective earth measurement . . . . . . . . . . . . . . . . . . . . . A2 - 6
Biopsy unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 6
Biopsy controller . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 6
CCD camera . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 6
Biopsy controller cable duct . . . . . . . . . . . . . . . . . . . . . . . . . . . A2 - 6
Opdimar Register 5 SPB7-230.061.10 Page 6 of 6 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden

Prerequisites 1

General 1

Valid for Opdimar system ASW 3.0 on SUN workstation, part No. 64 30 453. This document is valid for an Ultra 10 (SUN workstation with serial No.W1201).

Training of customer support engineers 1

Due to the technology used in this equipment, setup, service and maintenance may only be carried out by a customer support engineer who has attended a training workshop or has participated in at least one installation.

Documents required 1

Supplement to the Instructions for Use MAMMOMAT 3000 - Opdimar
(included in the Opdimar delivery)
MAMMOMAT 1000/3000/3000 Nova - Opdimar Mai ntenance Instructions
(included in the Opdimar delivery)
MAMMOMAT 3000 - Opdimar Wiring Diagram
(included in the Opdimar delivery)
1 - 1
MAMMOMAT 1000/3000 Nova Wiring Diagram

Ultra 10 CD-ROMs required 1

Ultra 10 Service Manual, part No. 65 27 670 (included in t he Opdimar delivery)
Reading Ultra 10 Service Manual CD-ROM:
You may view the HTML document and watch the video clips on a PC with a CD dri ve
running Windows 95 or later by using Netsca pe Navigator
Internet Explorer, version 4.0 or later. Refer to the browser documentati on about
configuring your browser to view MPEG-1 files.
OPDIMA Solaris 8 Installation CD v1.0, part No. 6 6 33 049
(included in the Opdimar delivery)
OPDIMA software installation cd, versi on 3.0, part No. 66 33 064
TM
, version 4.0 or later, or
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 1 of 2 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
1 - 2 Prerequisites

Meters and appliances required 1

Protective ground wire tester (44 15 899 RV090)
Stereo calibration phantom (included in t he Opdimar delivery) (part No. 64 30 701)
AEC calibration plexiglass, three plates measuring 150 mm x 150 mm x 19 mm and one
plate measuring 150 mm x 150 mm x 9,7 mm, part No. 65 61 232 and 65 61 224 respectively
Resolution phantom with at least 10 line pairs per mm
4,5 cm PMMA
PC with a CD drive running Windows 95 or later by using Netsc ape Navigator
4.0 or later, or Internet Explor er, version 4.0 or later (to be abl e to read Ultra 10 Service Manual)

Tools required 1

Standard service tools
PROM extractor for PLCC 32
TM
, version
Opdimar Register 5 SPB7-230.061.10 Page 2 of 2 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden

Functional description 2

Block diagram 2

The block diagram below shows the function of the Opdima r system.
2 - 1
M3000 printer / service PC
Workstation
System monitor
Graphics
acc.
RAM
Incoming mains 100 - 120V AC or 220 - 240V AC
CPU
Switch selectable
2 x RS-232
D101 Camera interface board
PCI BUS
PCI BUS
Printer (option)
RS-232
A
exp. sync.
B
biopsy data
X101 12-bit image
data
RS-422
X101 control data
M3000 stand
Biopsy controller
D200 Biopsy control board
X205
X204
X203
X203
X206
DTR
POW_ON
X202
X111 X111 X112X110,X111
X202
RS-232
AC
DC
5V_COM
230V AC
Mains cable
DC
AC
7V_COM
POW_ON
X202 X201
DC
CPU
DC
7V
20V
AC
30V
DC
5V
ADC
CCD camera
D110 CCD camera board
X207
X207
X885
X885
M3000
Biopsy unit
X Y Z
Fig. 1 Block diagram
= Galvanic separation
CCD
control
logic
CCD sandwich
ADC
DC
DC
5V
C_CL
A_CL
CCD_R
SRC_CL
OPD00465
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 1 of 4 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
2 - 2 Functional description

Biopsy controller 2

The biopsy controller has three main funct ions:

Biopsy unit control

Switching of RS-232 from MAMMOMAT to either MAMMOMAT printer/service PC or
workstation

CCD camera power supply

Biopsy unit control 2
The biopsy functions are controlled by the micro controller. In the biopsy unit there are three potentiometers indicating the current needle position.
The values from the poten tiometers ar e A/D con v erted and sent to the micro contr oller. To check the converted values from the A/D converter, the potentiometer values are converted in paralle l by the micro controller.

RS-232 switching 2

To enable RS-232 communication between MAMMOMAT and either MAMMOMAT printer/ service PC or workstation, a relay is used. The rel ay is controlled by the workstation.
CCD camera power supply 2
The biopsy controller conv erts the mains AC voltage to DC voltages. The DC voltages are used by the CCD camera.

LEDs 2

On the biopsy controller printed circuit board , D200, there are a number of LEDs with the following meaning:
Indication of RAM error
Indication of PROM error
Indication of analog/digital conver ter
Indication of NVM error
Indication of TxD error
Selection of RS-232 communication to MAMMOMAT printer/servic e PC or workstation
(SELECT_WS)
Supply voltages (e.g. 5V_COM, 30V_CC)
For more information on LEDs, see MAMMOMAT 1000/3000/3000 Nova - Opdima r Wiring Diagram.
Opdimar Register 5 SPB7-230.061.10 Page 2 of 4 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Functional description 2 - 3

Workstation 2

The workstation includes the follo w ing parts:
Main unit, including temporary storage medi a i.e. hard disk
Monitor
Keyboard and mouse
Internal CD drive
External MO unit i.e. magneto-optical dri ve, used for the permanent storage media and
backup
The workstation is used for:
Running application software
Communication with the CCD camera
Control of MAMMOMAT and biopsy unit

Application software 2

The application software run by the workstation is the main communication interface with the user and units of the Opdimar system.
Stereo examination
mode
Fig. 2 Application software logic flow
Start-up and login
Mode selection
Spot examination
mode
Database
mode
Service mode
Error handlingHelp (n.a.)Data storage
OPD00595
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 3 of 4 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
2 - 4 Functional description

CCD camera communication 2

The camera interface in the workstation main unit sends comma nds to the CCD camera from the workstation and receives the image from the camera. The interface adapts the workstation data bus to RS-422.
The camera interface also supplies t he CCD camera opto couplers with power.

Control of MAMMOMAT and biopsy unit 2

The workstation controls the MAMMOMAT via RS-232 communication. Biopsy calculations are performed by the workstation and target coordinates are sent to
the biopsy unit.

Monitor 2

If the settings of the monitor has changed, reset the default values according to the Moni­tor manual.
Ex.
NOTICE
If a 21” monitor is inst a lled be sure n ot t o change t he proporti ons between length and width. If this has been done anyway, press the auto sizing and centring button marked with a symbol (Fig. 3) to restore the previous settings.
Fig. 3 Symbol

CCD camera 2

The CCD camera consists of a printed circuit board, CCD sandwich and RS-42 2 interface. When the x-ray beams reach the CCD sandwich, electric energy proportional to the x-ray energy is produced. The produced electrical current is A/D conv erted to digital information and sent to the workstation RAM via an RS-422 interface.

Biopsy unit 2

The biopsy unit is used for performing biopsy examinati ons. The biopsy unit consists of a needle positioning device which is firmly attached to an
18 cm x 24 cm object table with a cut-out contour that is superimposed on the CCD. External stereo diaphragm, needle support and compression plate are also included in the biopsy unit. The biopsy unit can easily be attached to the swiv el arm of the MAMMOMAT.
There are three displays on the biopsy unit front. The displays show spatial deviation of needle tip from the suspect point in x-, y- and z- axis calculated by the workstation.
Three potentiometers monitors the actual position of the needle tip.
Opdimar Register 5 SPB7-230.061.10 Page 4 of 4 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden

Protective measures for CCD camera 3

3 - 1
CAUTION
CAUTION
The camera shall be used within 10230i C. The camera shall be transported or stored within 0240i C.
The CCD camera has to be handled with extreme care, it is very sensitive to mechanical shocks and temperat ure. When not con­nected do not touch the pins in the camera contacts. Shock and temperature sensors are integrated in the camera.
The CCD camera is sensitive to mechanical shock and shall always be stored in the attaché case, delivered with the system, when disconnected from the biopsy controller.
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 1 of 2 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
3 - 2 Protective measures for CCD camera
This page intentionally left blank.
Opdimar Register 5 SPB7-230.061.10 Page 2 of 2 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden

Protective measures 4

It is very important that any intervention in the equipment shall start with disconnecting it from the power supply with the main circuit br eaker. To prevent accidental triggering of high voltage and radiat ion, set the switch S2 (SS) on board D702 to OFF (lower position, no triggering of the SS relay).
4 - 1
CAUTION
WARNING
WARNING
When switching off the w o rkstation use the power off procedure described in the Supplement to the Instructions for Use MAMMO­MAT 3000 - Opdimar.
Switching off the workstation before the software has been closed down may cause damage to the files on the hard disc.
If the system is only switched off at the control panel or with S2/ D711 in the MAMMOMAT generator, line voltage will still be present at the generator line connection, line filter Z1, Z2, trans­former T1, transformer T10 and board D711 (see MAMMOMAT 1000/3000 Nova Wiring Diagram). The Opdimar is switched off separately.
After shut-down of the system, there may still be about 380 V DC present on the intermediate circuit of the MAMMOMAT generator. This will be indicated by LED V24 on board D710. The voltage will drop to less than 30 V within about 3 minutes, the LED goes out at about 30 V.
CAUTION
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 1 of 2 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
Observe the currently valid guidelines for handling elect ronics endangered by electrostatic discharge.
Use ESD-equipment, ground prior to making contact and place the components on a conductive surface.
The boards contain electrostatic highly sensitive components requiring particular care in their handl ing.
Risk of damaging components.
4 - 2 Protective measures
This page intentionally left blank.
Opdimar Register 5 SPB7-230.061.10 Page 2 of 2 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden

Service mode 5

General 5

There are three different user levels in the Opdimar software.
Regular user
Administration user
Service user
Administration users use the Service mode to perform the following tasks:
Calibration of the biopsy unit
Setting up needles
User administration
Network setup, DICOM nodes
Backup functions

Advanced service 5

Service users can additionally access advanced service functions. To get access you have to l og in with the user name “service”. The pass w ord is obtained from Sie mens Head Quarter Support Center.
5 - 1
The following advanced service functions are included:
Country settings
Unit tes ts
Software upgrade
Restore disk
Camera calibration/maintenance
Disk cache
Printer setup
Log administration
View log
NOTICE
The Opdimar external diaphragm m ust be used for tests and cali­brations in volving radiation.
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 1 of 28 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
5 - 2 Service mode

Selection of mode 5

After a successful login, the mode selection dial og is di splayed.
1. Press the Service button. Press Logout to cancel.
Mode Selection
Press a button to enter a user mode.
Stereo Spot
Database Service
Logout
Fig. 1 Mode selection dialog
Help
OPD00063
2. Select desired service function f rom the service dialog.
Service
Biopsy Calibration Needle Setup
User Administration
Advanced Service
Network Setup
Change Password
Backup
Close Help
OPD00433
Fig. 2 Service dialog
Opdimar Register 5 SPB7-230.061.10 Page 2 of 28 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Service mode 5 - 3

Calibration of the biopsy unit 5

Calibration of the biopsy unit is carried out by using a stereo calibration phantom with targets at fixed positions. By releasing two st ereo pair exposures with a fine needle adjusted to different targets , the system is automatically cal ibrated. F ol low the inst ructions given on the screen during the calibration procedure.
1. Press the Biopsy Calibration button i n the service dialog. The following is shown on the scre en:
Pixel x 138 value 1139 Gray scale center 1000 width 400y 152
123-45-6789
03/26/97
Help >>
i
Z
Z
1) Place and compress the stereo calibration phantom and insert a fine needle of suitable length.
2) Move the needle holder to the first calibration position. 3) Remove the needle. 4) Acquire a stereo pair of images.
Fig. 3 Dialogs during calibration
2. Place the stereo calibration phantom on t he biopsy table and compress. With the targets facing the pati ent side, the phantom fits in the openi ng of the
stereo compression plate, see Fig. 4. Tar get 2 shall be positioned 25228 mm in negative x-axis direction an d 12 mm from the object table side.
1.00x
Message
Help
Default Needle
Num. X Y Z
Close Help
Targets
Display a stereo pair!
Core 100
[mm]
[mm]
Delete
Transmit
Control
Image Tools
Windows
Pointer
Magnify
Ruler
Delete
Histogram
Filters
2
Control
Layout
Invert
Zoom
Pan
Restore
Screen Dump
Help
Acquire
Single Double
Targets
Exit
OPD00569
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 3 of 28 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
5 - 4 Service mode
3. Select a fine needle to be used and measure the length. Measuring the needle length is described on Page 5 - 7.
NOTICE
We recommend to use a needle that measures 90 mm, this will make it possible to avoid having to change needle while calibrat­ing.
4. Choose Other in the Needle option menu and enter t he selected needle length (minimum 90 mm).
5. Insert needle guides corresponding wit h the needle diameter.
6. Insert the fine needle into t he needle guides.
7. Move the needle to Target 1 of the stereo calibration phanto m, see Fig. 4, by using the adjustment knobs. Position the needl e according to Fig. 5.
NOTICE
Make sure the needle is moved to Target 1.
Fig. 4 Stereo calibration phantom
Target
Fig. 5 Position of needle
8. Remove the needle.
9. Acquire and release a stereo pair of images . Set the exposure parameters to 25 kV and 28 mAs in manual mode.
Target 1
Target 2
Fine needle Needle channel center
OPD00512
OPD00175
Opdimar Register 5 SPB7-230.061.10 Page 4 of 28 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Service mode 5 - 5
10. Check the reference marks and adjust if necessary, mark Target 1 and press Transmit.
NOTICE
When performing the biopsy calibration set the magnification to
0.7 and change Contrast/brightness in order to find all targets in the phantom. Be sure to mark the target at which the needle tip was positioned.
11. Insert the fine needle int o the needle support of the biopsy unit.
12. Move the needle to Target 2 of th e stereo calibration phantom by using the adjustment knobs.
13. Remove the needle.
14. Acquire and release a stereo pair of images. Set the exposure parameters to 25 kV and 28 mAs in manual mode.
15. Check the reference marks and adjust if necessary, mark Target 2 and press Transmit.
When the calibration is successfu lly calibrated, the followi ng message is displayed:
Message
The biopsy unit has now been calibrated. Press EXIT!
OK
Fig. 6 Information message
Help
OPD00176
16. Perform a final check according to Ver ifying the calibration of the bio psy unit on Page 10 - 1 to make sure that the biopsy unit wor ks properly.
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 5 of 28 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
5 - 6 Service mode

Setting up needles 5

1. Press the Needle Setup button in the serv ice dialog.
Needle Setup
Type Fine
Core
Close Help
Fig. 7 Setting up needles dialog
L1 (mm) L2 (mm)
90.5 -
110.0 5
2. Press Add... .
You can only add needles which match the needle guides wi th the fixed diameters: 0.7, 0.9, 1.2, 1.65, or 2.1 mm.
Add...
Edit...
Delete
OPD00087
Needles
Enter needle values.
L1
L2
OK Help
Fig. 8 Needle values dialog
Cancel
3. Select needle type (fine or core) fr om the option menu and enter needle length in the text field.
L1 shall be 30 to 175 mm when selecting fine needle, L2 is not applicable for fine needle. When selecting core, L1 shall be set to <170 mm.
CoreType
(mm)
(mm)
OPD00088
Opdimar Register 5 SPB7-230.061.10 Page 6 of 28 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Service mode 5 - 7
CAUTION
Fine needle
Ensure by measuring the length of the specified needle or core gun with needle that the right val ues are entered.
If the stroke length of the core gun is changed, a new needle has to be selected from the Targets dialog. The core needle length L1 shall be measured with the core needle in outer position and mounted in the core gun.
Core gun with coax
L1
L2
L1
L1
Core gun
L1
L2
L1L2
Needle channel center
Fig. 9 Measuring of needle length
NOTICE
When using core gun, the system safety margin to avoid hitting
OPD00089
the biopsy table is 5 mm.

Editing needle values 5

1. Select a needle from the needle setup dialog.
2. Press Edit... .
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 7 of 28 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
5 - 8 Service mode

Backup of temporary storage media 5

As a service user it is possible to make a complete backup of all folders existing on the temporary storage media (the hard disk).
1. Press the Backup button in the servic e dialog.
Backup
Backup / restore
Backup to MO
Setup for backup
Backup enabled
OK Help
Fig. 10 Backup dialog
Restore from MO
2. In the setup for backup section dis able the backup.
3. Press the Backup to MO button.
4. Follow the instruction giv en in the message dialogs.
5. Enable the backup in the backup dialog.
WARNING
Backup of temporary storage media cannot replace the regular use of backup described in Using the Backup function in Supple­ment to the Instructions for Use MAMMOMAT 3000 - Opdimar.
OPD00326
Backup of temporary storage media only saves t he content present on the hard disk at the actual moment. Images may have been erased by the disk cache system (see Page 5 - 23) or soft­ware reinstallation.
NOTICE
This backup might take a long time and require several MO disks since all folders on hard disk will be copied to backup MO.
Opdimar Register 5 SPB7-230.061.10 Page 8 of 28 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Service mode 5 - 9

Use of Advanced service functions 5

General 5

To get access to the advanced service menu you have to log in with the user name “service”. The password is obtained from Siemens Service Centre.
The advanced service is carried out by using a number of text dialogs. Selection of an item in a dialog can be done in three different ways:
Use the up/down arrows of the keyboard to step thr ough the fields (underlined)
Use the space bar to step through the fields
Type the corresponding number or letter of the item
Press Enter to execute.

Main menu 5

The main menu is used to access all the adv anced service functions.
1. Press the Advanced service button in the ser vice dialog.
2. Select a function from the menu.
Advanced Service
SERVICES
1.
Country Settings
2.
Unit Tests
3.
Software Upgrade
4.
Restore Disk
5.
Camera Calibration/Maintenance
6.
Disk Cache
7.
Printer Setup
8.
Log Administration
9.
View Log Close Help
Select: Close
Fig. 11 Advanced service dialog
Country Settings Function for selection of country of installation (your country). Unit Tests Function for performing tests of subassemblies.
OPD00409
Software Upgrade Function for performing upgrade of software and for
modifications of software. Restore Disk Function for restoring data from MO disk. Camera Calibration/
Maintenance
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 9 of 28 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
Function for performing installation and maintenance of CCD camera.
5 - 10 Service mode
Disk Cache Function for disk cache settin gs. Printer Setup Function for setup of printer. Log Administration Function for log settings. View Log Function for inspection of logs. Close Function for leaving the advanced service mode.
Opdimar Register 5 SPB7-230.061.10 Page 10 of 28 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Service mode 5 - 11
Country settings 5
National parameters are set in the country settings dialog. On delivery it contains default settings.
Advanced Service

COUNTRY SETTINGS

Language: English Time Zone: MET Time Settings: Year: 2000 Month: 10 Day: 24 Hour: 09 Minute: 21 Id Pattern: 111-11-1111 Date Pattern: m/d/y Service Center: Undefined Institution Name: Undefined Department Name: Undefined Action: Close Close/Help/Apply
OPD00513
Fig. 12 Country settings dialog
Available Languages Shows the languages currently available by the system. Language Selection of language in the dialogs. Select with space bar. If language
is changed, log out and log in to make the change take effect. Time Zone A selection of the time zone used by the system. Select with space bar .
For faster selection, type the first letter of the desired time zone name. Time Settings Value of system clock. If changed, Apply will reboot.
NOTICE! The database is updated every time an examination is
performed or when an image is loaded from a MO disk. The sys-
tem clock cannot be set to a point earlier than the last update.
ID Pattern Selection of ID-number structure. Type pattern with the keyboard: “1”
for digit, “a” for letter and an arbitrary character for punctuation mark.
For example: 111-11-1111. Date Pattern Selection of date structure. T ype pattern with the keyboard: “y” for year,
“m” for month, “d” for day and an arbitrary character for punctuation
mark.
For example: m/d/y or y-m-d (do not type mm/dd/yy or yy-mm-dd). Service Center Type the appropriate service center. Institution Name Type the appropriate name for the institution. Department Name Type the appropriate name for the department.
1. Select Country Settings in t he advanced service dialog.
2. Set the values for Language, Time Zone, ID Pattern, Dat e Pattern, Service Center, Institution Name and Departmen t Name in each respective fields.
3. Select Apply and press Enter to execute the chang es.
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 11 of 28 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service
5 - 12 Service mode

Test of units 5

Advanced Service
UNIT TESTS
Unit
Test
Mammomat
Yes
Camera
Yes
Biopsy Controller
Yes
Modem
Yes
MO-Disk
Yes
Database
Yes
Action: Run Close/Help/Run
Fig. 13 Un it tests dialog
NOTICE
Status OK
OK OK N/A N/A OK
OPD00171
Before performing unit tests, the biopsy unit needs to be mounted to the MAMMOMAT. Otherwise the biopsy controller test will fail.
1. Select Unit Tests in the advanced ser vice dialog.
2. Select Yes or No to select/deselect the parts to be included in the tes t.
3. Select Run and press Enter.
Each unit that passes the test will be indicated with an OK mess age. If it does not pass, a Fai led message will appear. Units that have not been tested are indicated with an Untested message.
The units are tested according to the following:
MAMMOMAT - test if there is a connection and if the power is on
CCD camera - test if there is a connection and i f the power is on
Biopsy controller - test if there is a connection and if the power is on
Database - reading and writing in the database
Opdimar Register 5 SPB7-230.061.10 Page 12 of 28 Siemens-Elema AB Service Rev. 01 09.02 SPS-UD Solna, Sweden
Service mode 5 - 13
Software upgrade 5
The software upgrade dial og is used to upgrade the current softwar e version with a new version from CD-ROM and to do modifications of the software.
Advanced Service

SOFTWARE UPGRADE

1. New software version
2. DICOM option
3. Network settings
4. Restore hostname
5. Update eeprom
6. Miscellaneous Close Help
Select: Close
OPD00514
Fig. 14 Software upgrade dialog
New software version Upgrade with new version from CD-ROM. DICOM option Enabling the DICOM option.
NOTICE! A license key is necessary to install DICOM.
Network settings Function for defining the network settings for the Opdimar sys-
tem.
Restore hostname The hostname of the original workstation can be restored to a
new work station.
Restore hostname from most recently used local/backup MO disk. Update eeprom Not applicable. Miscellaneous Database and MO disk ut iliti es.
Siemens-Elema AB Register 5 SPB7-230.061.10 Page 13 of 28 Opdimar Solna, Sweden Rev. 01 09.02 SPS-UD Service