MAMMOMAT 3000/3000 Nova
Service
Modification Instructions
Drill a keying hole in the biopsy unit
© Siemens AG 2000
The reproduction, transmission or use of this docu men t or its con tent s is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
English
Print No.: SPB7-230.092.18.01.02 Doc . Gen. Date: 04.00
Replaces: n.a. 65 69 193
Page 2
0 - 2 Revision
Chapter Page Revision
all all 01
Mammomat 3000/3000 Nova SPB7-230.092.18 Page 2 of 4 Siemens-Elema AB Service Rev. 01 04.00 SPS-UD Solna, Sweden
Page 3
Contents 0 - 3
Page
1 _______Prerequisites __________________________________________________1 - 1
Purpose . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1 - 1
Tools required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Time required . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
Important notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 1
2 _______Procedure_____________________________________________________2 - 1
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .2 - 1
Siemens-Elema AB SPB7-230.092.18 Page 3 of 4 Mammomat 3000/3000 Nova Solna, Sweden Rev . 01 04.00 SPS-UD Service
Page 4
0 - 4 Contents
This page intentionally left blank.
Mammomat 3000/3000 Nova SPB7-230.092.18 Page 4 of 4 Siemens-Elema AB Service Rev. 01 04.00 SPS-UD Solna, Sweden
Page 5

Prerequisites 1

Purpose 1

This instruction describes how to dril l a keyi ng hole in a biopsy unit, belonging to a Mammomat 3000 with serial no. 6849 or lower, so that it can be used with a Mammomat 3000/3000 Nova with serial no. from 6850.

Tools required 1

Standard tool kit
Drilling machine
Set of drills containing a 8 mm drill

Time require d 1

Approximately 1 h for 1 person.

Important notes 1

While dismounting do not use excessive f orce or the object table top might be da maged
1 - 1
After drilling the keying hole, be sure to remove all metal chip s
Siemens-Elema AB SPB7-230.092.18 Page 1 of 2 Mammomat 3000/3000 Nova Solna, Sweden Rev . 01 04.00 SPS-UD Service
Page 6
1 - 2 Prerequisites
This page intentionally left blank.
Mammomat 3000/3000 Nova SPB7-230.092.18 Page 2 of 2 Siemens-Elema AB Service Rev. 01 04.00 SPS-UD Solna, Sweden
Page 7

Procedure 2

NOTICE

Procedure 2

2 - 1
mm
mm
X
Y
Z
X
Y
Z
X
mm
Y
Z
Z
X
Y
Object table top
1
2
3
Fig. 1
Guiding pins
1. Remove the four screws holding the front edge of the object table top.
2. Carefully lift the front edge of the object table top to reveal the four s crew placed behind it.
Do not use excessive force or the object table top might be damaged.
Remove the revealed screws, the two nuts and t he two guiding pins.
3. Remove the bottom plate by unscrewing the eig ht remaining screws:
- Two screws at either side of the biopsy uni t.
- Four screws at the bottom of the biopsy uni t.
Nuts
Siemens-Elema AB SPB7-230.092.18 Page 1 of 2 Mammomat 3000/3000 Nova Solna, Sweden Rev . 01 04.00 SPS-UD Service
Page 8
2 - 2 Procedure
4. Drill a 8 mm keying hole in the biopsy unit. The distance shall be 23 mm from the top and 60 mm from the connector (Fig. 2). Fig. 3 can be used as a drilling template .
MAM00739
Fig. 2
23 mm
60 mm
5. Remove all metal chips .
6. Reassemble the biopsy unit.
7. Check that the biopsy unit fits a Mammomat wi th serial no. from 6850.
Fold here
Cut out here
Fig. 3
Mammomat 3000/3000 Nova SPB7-230.092.18 Page 2 of 2 Siemens-Elema AB Service Rev. 01 04.00 SPS-UD Solna, Sweden