Installtion Instructions, OPCOMP 0 - 1
Table of contents, Revision status
PageRev.
Installation of OPCOMP ......................................................... 1 - 1
Exchanging the display film ................................................................1 - 1
Siemens-Elema AB SRX-MK MAMMOMAT 300
Solna, Sweden
RX B7-120.033.02.01.02
Rev. 01 04.94
0 - 2Installation Instructions, OPCOMP
Siemens-Elema AB SRX-MK MAMMOMAT 300
Solna, Sweden
RX B7-120.033.02.01.02
Rev. 01 04.94
Installation Instructions, OPCOMP1 - 1
Installation of OPCOMP
Exchanging the display film
Display film
Fig. 1.1
•Remove the display film from the stand display.
•Affix the new display film on the stand display.
Siemens-Elema AB SRX-MK MAMMOMAT 300
Solna, Sweden
RXB7-120.033.02.01.02
Rev. 01 04.94
1 - 2Installation Instructions, OPCOMP
Installing OPCOMP
Operation of the service PC is described in
Instructions"
RX B7-120.033.01...
"Installation and Start-Up
Connecting the service PC
•Connect service PC to p.c. board D702 in the generator using connection
cable part no. 96 60 978 RE999. Switch S3 on board D702 must be set to
upper position (ON).
•Insert service disk.
•Start service program.
•Generator "ON"
DisplayEntry requiredRemarks
c:\>{A:}According to service
{B:}<Enter>PC disk drive names
a:\>SERVICE<Enter>
Your name please
Password please
Main menu< ↓ > < ↑ ><Enter>Select "M2000 Config"
Configuration< ↓ > < ↑ ><Enter>Select "Best com-
USER'S NAME
xxxxx
<Enter>
<Enter>
subroutine
pression" subroutine
Siemens-Elema AB SRX-MK MAMMOMAT 300
Solna, Sweden
RXB7-120.033.02.01.02
Rev. 01 04.94
Installation Instructions, OPCOMP1 - 3
Calibrating for OPCOMP
When calibrating for OPCOMP, following will be needed:
•a 3 – 5 cm thick, non-compressible object,
•the 18 x 24 cm and the 24 x 30 cm compression plates for normal
screening,
•the 18 x 24 cm and the 24 x 30 cm object tables.
•Check that the x-ray tube assembly is in 0°-position.
•Choose
program, see fig. 1.2
•Check that the values in the Best compression table are as follows:
best comp mod00
best comp timeout1E
force per thickness0055
divider05
best comp force lim1C
loop count compare01
sample time diff0A
Save all changes with <F2>.
•Attach the 18 x 24 cm compression plate and the 18 x 24 cm object table.
•Place the non-compressible object on the object table.
"Best compression"
from the Installation table in the Service
Siemens-Elema AB SRX-MK MAMMOMAT 300
Solna, Sweden
RXB7-120.033.02.01.02
Rev. 01 04.94
1 - 4Installation Instructions, OPCOMP
Main menu
Configuration
Service
SAve config file
SHow config file
Normal mode
M2000 Config
M2000 Test
Quit
Installation tables
Stand installation
Calibrate compression
Rotation calibration
Installation default
Lift parameters
Best compression
Rotation parameters
Best compression table
best comp mod00
best comp timeout1E
force per thickness0055
divider05
best comp force lim1C
loop count compare01
sample time diff0A
1 Help 2 Save 3 Calibrate 5 6 7 8 9 10 Quit
Fig. 1.2