Siemens Mammomat 1000 User manual

MAMMOMAT 1000/3000
Service
Service Program
SP
© Siemens AG 1999
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Register 5 English Print No.: SPB7-230.114.02.01.02 Doc . Gen. Date: 03.99
Replaces: n.a.
65 19 792
0 - 2 Revision
Chapter Revision
All 01
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 2 of 4 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Contents 0 - 1
Page
This document is valid from serial No. 6500 and from FW-version 3.0.
1 _______Working with the service PC _____________________________________1 - 1
Description of the syntax used in these instructions . . . . . . . . . . . . . . . . . . 1 - 1
Connecting the service PC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 - 2
2 _______Adjustment and service PC programs _____________________________2 - 1
Configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 1
Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 37
Normal mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 47
Te st DUEP communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 55
Backup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 - 56
3 _______Error messages________________________________________________3 - 1
General . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 1
Error messages of the master Er 0XX. . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Error messages of the control deck Er 3XX. . . . . . . . . . . . . . . . . . . . . . . 3 - 7
Error messages of the IONTOMAT PM Er 5XX . . . . . . . . . . . . . . . . . . . . .3 - 8
Error messages of the power pack Er 6XX . . . . . . . . . . . . . . . . . . . . . . 3 - 11
Error messages of the stand Er 8XX . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 15
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 1 of 2 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
0 - 2 Contents
Page
This page intentionally left blank.
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 2 of 2 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden

Working with the service PC 1

Description of the syntax used in these instructions 1
< ..... > The indication of which function keys to press is given between
these characters, for example <ENTER>, <ESC> etc.
CAPITALS Capital letters indicate data which must be entered unchanged,
for example the name of a register, files etc.
1 - 1
Italics
[ ..... ] Square brackets enclose additions to commands which may be
Bold Data relating to formats, user entries etc., which is important for
xx yy zz Data can be entered in place of
{...} Curved brackets indicate that out of several terms listed one be-
NOTICE
* * * * * When the password is entered, only these characters are shown. Menu Selection: When several menus, programs, fi les etc. are presented f or selec-
Italics represent data in whi ch a value should be ent ered, e. g. for user name, the name of the technician should be entered.
optionally entered.
the following entry, is shown in bold as it appears on the monitor screen.
This character indicates that at this point the space key must be pressed.
"x, y, z"
(e.g. day´s date).
low the other, one must be selected. Important remarks are indicated with this box.
tion, they are shown in a box (program window). Selection is made with the <> and <> keys.
The module selected is highligthed in the display. <ENTER> Every entry must be confirmed with the <ENTER> key. <ESC> ESC allows paging back through the program. <xx> + <y> Some functions are selected by pressing two keys
simultaneously. Procedure: Press, for example, the <Shift> key
and keep it depressed, press the <*> key and then release both
keys. <F1> Key <F1> calls up a selective help text. <F10> Key <F10> exits the program.
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 1 of 6 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
1 - 2 Working with the service PC
XXX - - XXX - ­XXX - - XXX
By this means, the way is shown how to call up a particular sub­routine (display windows).
For example:
Main menu – Configuration – Iontomat – Corr . curve –
Connecting the service PC
The service PC must be connected with connecti ng cable part no. 99 00 440 RE999 to p.c. board in the generator (do not i nsert the diskette in the drive yet). Use port COM 1 on the
service PC. Starting up and using the service PC
1. Switch on generator and service PC. After initialization, the s ervice PC shows: C:\>
2. Now insert the discette with th e service program.
3. Select the appropriate drive and then press <ENTER>. The screen shows: A:\> or B:\>
{A:} {B:}
4. Start the service program by typing ’SERVICE’, then press <ENTER>. The program asks for the use’sr name: Type the name of the technician, for examp le NN, and then press <ENTER>.
5. The program asks for the password: Type the password (********) and then pr ess <ENTER>.
6. The display window shows: If
Mode: Normal Mode: Stand-alone
if press <ENTER>. The display window shows:
7. Select the program part to be used with keys < ↑> and/or <↓>, then press <ENTER>. The program part selected is shown with a b ackground: If necessary, additional sub routines can be similarly selected here.
8. Make the necessary entries in the appr opriate part of the program. Save the entered data with <F2>. Page back in the pro gram with <ESC>. The appropriate instructions are shown on the monit or.
End the procedure with the service PC with <F10>.
is the correct choice then press < ENTER> or
PRGRAM-MODE; Mode: Normal
is wanted press the <SPACE> key once and then
Main menu
Your name, please
Password, please
Configuration
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 2 of 2 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden

Adjustment and Service PC Programs 2

2 - 1
Configuration 3
System type 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
System type
NOTE: If changing system type, always check the power setting and, when changing from M1000 to M3000, also the anode selection. Collimator choice is valid only for M1000. Changes in the ’System type’ and ’Collimator’ above become active after next power up of the system.

Configuration

System type
Anode Show configuration file Save configuration file Iontomat Miscellaneous Filament
System type Collimator
M3000 Automatic
<ESC> to exit, <space> to toggle entry
1 Help 2 Save 3 45678910Quit
Fig. 1
System type indicates type of stand, generator and control panel. Collimator indicates type of collimator in the stand. Collimator choice (Automatic or Maual) valid only for MAMMOMAT 1000.
When MAMMOMAT 1000 is selected, the tungsten anode is automatically deselected, and furthermore, the
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 1 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
Anode
menu cannot be selected.
MAM00449
2 - 2 Adjustment and Service PC Programs
NOTICE
Anode 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Anode
Configuration
System type Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Tungsten anode
NOTE: When changing the anode selection, always check the power setting.
<ESC> to exit, <space to toggle entry>
1 Help 2 Save 3 45678910Quit
Fig. 2
Enable
Activation and deactivation of the tungsten anode. When deactivated, it will not be possible to select W/Rh alternative on the control panel.
This menu is not selectable when t he s ystem type i s set t o M1000.
MAM00454
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 2 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 3
NOTICE
Show configuration file 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Show file mammo.cfg
Configuration
System type Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Fig. 3
System M3000 Power
Molly large focus 3.75 kW Tungsten large focus 4.70 kW Molly small focus 0.70 kW Tungsten small focus 0.85 kW
Filament current
Molly large focus 6500 mA Tungsten large focus 6500 mA
Molly small focus 6500 mA Tungsten small focus 6500 mA Tube current reduction ON Power supply OFF Sensitivity Wing 1
H 11.0
D 11.0 Correction curve A Sensitivity correction
no grid
grid
magnification Optimized compression
OPCOMP 18*24 and 24*30 ON
OPCOMP mag 15 and 18 OFF Auto limits P1 30mm P2 45mm P3 60mm
Any key to continue
1 Help 2 3 45678910Quit
Tungsten Anode
H AD
ENABLED
Reduction time: 100 ms
Wing 2
11.0
11.0
Wing 1 0 0 0 0
Wing 2 0 0
Collimator AUTOMATIC
This menu displays the system configuration as stored in the “mammo.cfg” file.
MAM00450
All data are examples and may vary for the unit in question.
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 3 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 4 Adjustment and Service PC Programs
Save configuration file 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Save configuration -> mammo.cfg
Configuration
System type Anode Show configuration file Save configuration file Iontomat Miscellaneous Filament
System M3000 Tungsten Anode Power
Molly large focus 3.75 kW Tungsten large focus 4.70 kW Molly small focus 0.70 kW Tungsten small focus 0.85 kW
Filament current
Molly large focus 6500 mA Tungsten large focus 6500 mA
Molly small focus 6500 mA Tungsten small focus 6500 mA Tube current reduction ON Power supply OFF Sensitivity Wing 1
H 11.0
D 11.0 Correction curve A Sensitivity correction
no grid
grid
magnification Optimized compression
OPCOMP 18*24 and 24*30 ON
OPCOMP mag 15 and 18 OFF Auto limits P1 30mm P2 45mm P3 60mm
<ESC> to exit, <F2> save list in file mammo.cfg
1 Help 2 Save 3 45678910Quit
Fig. 4
H AD
ENABLED
Reduction time: 100 ms
Wing 1 0 0 00
Collimator AUTOMATIC
Wing 2
11.0
11.0 Wing 2
0 0
MAM00451
This menu is used for storing information about the system confi guration in the
“mammo.cfg” file. The file will be stor ed on the hard disk of the service PC at current directory, if the service
PC program is running from this. If the service PC program is running from a floppy disk, the file wil l be stored on this.
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 4 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 5
NOTICE
IONTOMAT PM, sensitivity 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
IONTOMAT PM - Sensitivity
Wing 1 11.0 Wing 2 11.0 11.0
Configuration
System type Anode Show configuration file Save configuration file Iontomat Miscellaneous Filament
IONTOMAT PM
Sensitivity Sensitivity correction Correction curve
HD
11.0
Valid entries from 0 to 31 in 1/2 E.P. Input of decimal point not required!
<ESC> to exit, <TAB> move to next entry field
1 Help 2 Save 3 45678910Quit
Fig. 5
Indicates the sensiti vity of the IONTOMAT channel for high speed (H) and high resolution screen (D) respectively.
Values from 0 to 31 can be entered in 1/2 exposure points.
If only one type of film-screen-combination is used, set both val ­ues (for H and D) to the same level.
Also see “Installation and Start-Up Inst ructions”.
MAM00455
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 5 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 6 Adjustment and Service PC Programs
NOTICE
IONTOMAT PM, sensitivity correction 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
IONTOMAT PM - Sensitivity correction
Without grid 0 With grid 0 Magnification 0
Configuration
System type Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
IONTOMAT PM
Sensitivity Sensitivity correction Correction curve
Wing 1 Wing 2
0 0 0
Valid entries from -8 to 8 in 1/2 E.P.
For pos entries 1st digit is <space> for neg entries type <->.
Input of decimal point not required!
<ESC> to exit, <TAB> move to next entry field
1 Help 2 Save 3 45678910Quit
Fig. 6
Indicates the sensitivity correction for different types of object tables.
All data are examples and may vary for the unit in question.
MAM00456
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 6 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 7
IONTOMAT PM, correction curve 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
IONTOMAT PM - Correction curve
Configuration
System type
Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Curve number
Speed H A Speed D A
IONTOMAT PM
Sensitivity Sensitivity correction Correction curve
<ESC> to exit, <TAB> move to next entry field
1 Help 2 Save 3 45678910Quit
Fig. 7
Indicates which correction curve shall be used for high speed (H) and high resolution (D) screen respectively.
For further information, please refer to the installation manual, chapter “Start-up and func­tional test of the IONTOMAT PM”.
MAM00457
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 7 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 8 Adjustment and Service PC Programs
NOTICE
Miscellaneous, DLF switch 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Dynamic learn filament
Configuration
System type Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
Switch is ON
<ESC> to exit, <space> to toggle entry
1 Help 2 Save 3 45678910Quit
Fig. 8
Activation and deactivation of the filament current adapti on. When activated, the adapti on takes place after a total of ten exposures has been made wit h the actual focus, i f the devi­ation of the average value from the nominal value is within -15% to + 5%.
Without any exceptional reason, the DLF-switch shall be “ON”.
MAM00458
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 8 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 9
Miscellaneous, reduction of power 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Reduction of power depending on line quality
Configuration
System type Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
Tube is P40MoW Select nominal line voltage
240 V <-- with cursor keys and <Enter>
Type in line impedance Power is 4.70 kW TUNGSTEN
3.75 kW MOLLY
<ESC> to exit
1 Help 2 Save 3 45678910Quit
Fig. 9
Calculates the maximum power allowed with regard to nominal voltage and line imped­ance.
Select the line voltage a nd type in t he line impeda nce you have meas ured. The proce ssor will calculate the power and wi ll indicat e whe n the power must be redu ced be cause of the line impedance.
Finally press <F2> to save.
MAM00459
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 9 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 10 Adjustment and Service PC Programs
Miscellaneous, tube current reduction 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Tube current reduction (TCR) at short exposure time (100ms)
Configuration
System type Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
TCR switch is ON Exposure time with 50 mA tube current is 100 ms
<ESC> to exit, <space> to toggle entry
1 Help 2 Save 3 45678910Quit
Fig. 10
Activation and deactivation of TCR. When this function is activated, the tube current is reduced to 50 mA, in IONTOMAT
mode, at the start of exposure. The time for the reduction is always 100 ms.
MAM00460
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 10 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 11
If TCR switch is “ON”, the tube current will look like this: 3
50 mA
approx. 100 ms
MAM00461
Fig. 11
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 11 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 12 Adjustment and Service PC Programs
Miscellaneous, printer 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Configuration
System type Anode Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
Printer type selection
Actual setting
<ESC> to exit, <space> to toggle entry
1 Help 2 Save 3 45678910Quit
Fig. 12
No print
Select no print, RS 232C or PC print. RS 232C is used when the printer is connected directly to the generator. PC print is used when the printer is connected to the system via a PC connected to the
generator.
MAM00462
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 12 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 13
Miscellaneous, power supply 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Power supply
Configuration
System type Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
Switch is
<ESC> to exit, <space> to toggle entry
1 Help 2 Save 3 45678910Quit
Fig. 13
OFF
Indicates whether the Mammomat is powered by an external power supply (e.g. Power Aid) or not.
MAM00463
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 13 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 14 Adjustment and Service PC Programs
Miscellaneous, potentiometer check 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Potentiometer check
Configuration
System type
Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
A/D value
<ESC> to exit
1 Help 2 Save 3 45678910Quit
Fig. 14
032A
Sampled calibration value for the return current through the four potentiometers “preset force”, “preset angle”, “thickness” and “angle”. This retur n current is continuously moni­tored by the system. Should the return current deviate too much from the calibration value, this is probably due to a faulty potentiometer or due to a contact failure with one of the potentiometers, which results in error code 825 or 826.
MAM00464
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 14 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 15
NOTICE
NOTICE
Miscellaneous, panel programming 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Panel programming
Configuration
System type
Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
Switch is
<ESC> to exit, <space> to toggle entry
1 Help 2 Save 3 45678910Quit
Fig. 15
ON
Activation and deactivation of the Memory button on the control panel, making it possible to change the program settings for the four programs.
With the switch “OFF”, any change of data is blocked.
Should be switched “OFF” only if customer demands.
This menu is not selectable when the sys tem t ype is set to M1000.
MAM00465
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 15 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 16 Adjustment and Service PC Programs
NOTICE
Miscellaneous, auto limits 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Auto limits
Configuration
System type Anode Show configuration file Save configuration file Iontomat Miscellaneous Filament
Auto limits Program 1 Program 2 45 mm Program 3 60 mm
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
30 mm
Valid entries from 0 to 180 mm
<ESC> to exit, <TAB> move to next entry field
1 Help 2 Save 3 45678910Quit
Fig. 16
Set the upper thickness limits for the programs 1-3. Program 1 has a lower limit of 0 and Program 4 has a higher limit of 180 mm. Supposing the above used values are stored in auto mode:
The program key P1 is automatically selected with a compression thickness between 0
and 30 mm.
The program key P2 is automatically selected with a compression thickness between 31
and 45 mm.
The program key P3 is automatically selected with a compression thickness between 46
and 60 mm.
The program key P4 is automatically selected with a compression thickness more than
60 mm.
MAM00466
This menu is not selectable when t he s ystem type i s set t o M1000.
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 16 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 17
Miscellaneous, cassette loaded check 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Cassette loaded check
Configuration
System type Anode Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
Switch is
<ESC> to exit, <SPACE> to toggle entry
1 Help 2 Save 3 45678910Quit
Fig. 17
ON
Activate or deactivate the check that a cassette is loaded/reloaded. With this function activa ted, the casset te symbol on the contr ol panel will light up when the
inserted cassette has been exposed or if no cassette is insert ed a t all. Exposure release is in this case blocked. The film cassette must be exchanged after each exposure.
During service it might be helpful to set the switch to “OFF”. When finishing service, do not forget to set it back to the original setting.
MAM00467
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 17 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 18 Adjustment and Service PC Programs
Miscellaneous, illumination time 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Illumination time
Configuration
System type
Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
Illumination time
<ESC> to exit
1 Help 2 Save 3 45678910Quit
Fig. 18
20 s
Duration of switched-on field light (collimator lamp). 1-99 s.
Duration should be set according to customer’s wishes.
MAM00468
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 18 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 19
Miscellaneous, version info 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Version info
Configuration
System type
Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
Miscellaneous
DLF switch Reduction of power Tube mA reduction Printer Power supply Potentiometer check Panel programming Auto limits Cassette loaded check Illumination time Version info
MASTER PANEL STAND Version: Date: 94-02-15 93-11-23 94-03-05
Any key to continue
1 Help 2 3 45678910Quit
Fig. 19
2.1 1.2 1.4
Shows software versions and dates for the three computers in the system.
MAM00469
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 19 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
2 - 20 Adjustment and Service PC Programs
NOTICE
Filament 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Filament
Filament current @ 25 kV Large focus molly 6900
Large focus tungsten 6500 Small focus molly 7500 Small focus tungsten 6500
Configuration
System type
Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament
<ESC> to exit, <TAB> to move to next entry field
1 Help 2 Save 3 45678910Quit
Fig. 20
Learn
Indicates the adapted filament current of both anode materials for large and small focus. Adaption exposures for the filament current are also made in this menu. Pressing F3 sets the exposure parameters on the control panel, and subsequently press-
ing Alt-F6 starts the exposure. Repeat this procedure until the tube-current deviation is
within ±5%. If the tube has been replaced, program the filament current value 200 mA lower than the
value stated in the test certificat e as a preadjust ment. (e.g. If tube value is 6900, pr ogram
6700.) If a new filament current ha s been entered you must sav e the new value (press F2) before
the learn filament is performed.
The Dynamic-Learn-Filament may change this adjustment. But the DLF will only learn when the exposure time is longer than 60 ms.
MAM00470
Mammomat 1000/3000 Register 5 SPB7-230.114.02 Page 20 of 56 Siemens-Elema AB Service Program Rev. 01 03.99 SPS-UD Solna, Sweden
Adjustment and Service PC Programs 2 - 21
Power 3
Main menu
Configuration Service Normal mode Test DUEP Communic. Backup Quit
Configuration
System type
Anode
Show configuration file Save configuration file Iontomat Miscellaneous Filament Power Clock Compression Lift Rotation Grid speed Beam limiting device
Power
Power
Large focus molly 3.75 kW Large focus tungsten 4.70 kW Small focus molly .70 kW Small focus tungsten .85 kW
Input of decimal point not required!
<ESC> to exit, <TAB> move to next entry field
1 Help 2 Save 3 45678910Quit
Fig. 21
Indicates the maximum power for large and small focus of both anode materials. The values programmed must not exceed the values stated above. These maximum values are only possible when the line impedance is not too high.
MAM00471
Siemens-Elema AB Register 5 SPB7-230.114.02 Page 21 of 56 Mammomat 1000/3000 Solna, Sweden Rev. 01 03.99 SPS-UD Service Program
Loading...
+ 61 hidden pages