Siemens Mammomat 1000 Update Instructions

Page 1
Mammomat 1000
SP
Update Instructions
Title:
Reason for Update: Performance
Urgency: Immediate Within 6 months Update material required? Yes No Materials free of charge? Yes No Return of parts? Yes (See page 3) No Estimated completion time: 1 h Number of CSE’s: one (1) Customer application training? Yes No
Systems/Products affected/System identifying IVK
Name Material No. Serial No.
Exchange of Diaphragm
X X X
SP017/02/P
X
X
X
Mammomat 1000 61 34 709 8601 - 9131
Remark: Affected units delivered to different countries within the serial number inter­val, see table on page 2.
Components affected/to be modified
Name Material No. Serial No. Component status
affected
n.a. n.a. n.a. n.a.
Remark:
© Siemens AG 2002
Chg. Ref. No.: RBN S2050785 Name: Engman/Sundkvist/Lindström Dept.: CS PS 24/SPS
Replaces: n.a. 66 31 803 Page 1 of 6
The reproduction, transmission or use of this document or its contents is not permitted without express written authority. Offenders will be liable for damages. All rights, including rights created by patent grant or registration of a utility model _or_ design,_are_ reserved.
Page 2
2 Update Instructions SP017/02/P
Document Revision Level 0
This document corresponds to the version/revision level effective at the time of system delivery. Revisions to hardcopy documentation are not automatically distributed.
Please contact your local Siemens office to order current revision levels.
Disclaimer 0
The installation and service of equipment described herein is to be performed by qualified personnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by Siemens or one of its affiliates to provide such service.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or authorized by Siemens or one of its affiliates are directed to contact one of the local offices of Siemens or one of its affiliates before attempting installation or service pro­cedures.
Systems/Products Affected 0
Mammomat 1000 (mat. no. 61 34 709) within serial number interval 8601 - 9131. See table below for detailed designation of affected units.
Country Number of
Update Kits
Austria 1 9054
France 1 8636
Germany 6 8680
India 1 8611
Italy 1 8725
Japan 2 9033
Pakistan 1 9108
Philippines 1 9038
Portugal 1 9053
South Korea 1 9037
Spain 3 8601
The Netherlands 1 8699
Turkey 4 8640
Uruguay 1 9008
Tab. 1 Designation of affected units
Serial No.
Mammomat 1000
9014
9059
9060
9083
9105
9051
8639
8641
9056
9057
9131
Mammomat 1000 SPB7-230.896.11 Page 2 of 6 Siemens AG Rev. 01 10.02 CS PS 24/SPS Medical Solutions
Page 3
Update Instructions SP017/02/P 3
Reason for the Update 0
A number of faulty diaphragms have been delivered together with the Mammomat 1000 in the mentioned serial number interval.
These diaphragms can be mounted in the wrong position giving a white stripe along the chest wall.
These faulty diaphragms must be exchanged to the ones delivered in the Update Kit.
18 x 24 CM
SID 65 CM
MODEL No.:
SERIAL/LOT No.:
This product complies with DHHS regulations 21 CFR Subchapter J, applicable at date of manufacture: Manufactured:
Siemens Aktiengesellschaft
Wittelsbacherplatz 2. D-8 Munchen 2
Germany
R 8 mm x 2
R 4 mm x 2
18 x 24 CM
SID 65 CM
MODEL No.:
SERIAL/LOT No.:
This product complies with DHHS regulations 21 CFR Subchapter J, applicable at date of manufacture: Manufactured:
Siemens Aktiengesellschaft
Wittelsbacherplatz 2. D-8 Munchen 2
Germany
R 8 mm x 4
MAM00950
Fig. 1 Difference between right and faulty diaphragm
Prerequisites 0
No special prerequisites.
Special Tools / Documents 0
No special tools needed.
Ordering Information 0
The following update kit has to be ordered from CSML (SAP Distribution Chan nel, factory
2050): Update Kit Diaphragm M1000 66 33 148 X041E
Siemens AG SPB7-230.896.11 Page 3 of 6 Mammomat 1000 Medical Solutions Rev. 01 10.02 CS PS 24/SPS
Page 4
4 Update Instructions SP017/02/P
Contents of the Update Kit 0
The Update Kit 66 33 148 contains the following parts:
Pos. Quan. Material No. Name
1 1 64 83 692 Diaphragm, 18x24, deep 2 1 64 83 700 Diaphragm, Magnification, deep 3 1 66 31 803 Update Instructions, UI SP017/02/P
Tab. 2 Contents of the Update Kit
Return of Parts 0
There is no requirement on return of faulty diaphragms back to CSML for scrapping. The diaphragms can be scrapped by the CSE at the Regional Unit.
Work Steps 0
Exchange the faulty diaphragms at site with the ones delivered with the Update kit.
Scrap the faulty diaphragms.
Customer Information 0
n.a.
Final Work Steps 0
Update the system documentation.
Update the revision level, the operating instructions and the technical documentation. Fill out, and if needed, make a copy of the attached "Completion Protocol/ Update Completion Form" and file it in the corresponding System Binder/User Handbook.
This update includes an IVK.
- Changes in the IVK structure must be reported to the management system to correct the installed volume.
Updates that have already been completed prior to publication of this UI must also be
reported.
The update is reported as follows:
- The modification reply cards (Type 606) previously distributed with the publication of updates no longer apply.
- The modification reply report has to be prepared by authorized personnel using an application on the Intranet.
Mammomat 1000 SPB7-230.896.11 Page 4 of 6 Siemens AG Rev. 01 10.02 CS PS 24/SPS Medical Solutions
Page 5
Update Instructions SP017/02/P 5
Completion Protocol/Update Completion Form 0
The update with the number SP017/02/P has been completed.
Material Number: 61 34 709
Serial Number: ................................
Customer: ................................ Site: ................................
Customer No.: ................................
Name ( CSE ): ................................ Telephone: ................................
Country: ................................ Location: ................................
Date: ................................ Signature: ................................
Remark: ....................................................................................................
....................................................................................................
NOTE
Siemens AG SPB7-230.896.11 Page 5 of 6 Mammomat 1000 Medical Solutions Rev. 01 10.02 CS PS 24/SPS
After completing the update, make a copy of this page, fill it out and file it in the corresponding System Binder/User Handbook
.
Page 6
6 Update Instructions SP017/02/P
This page intentionally left blank.
Mammomat 1000 SPB7-230.896.11 Page 6 of 6 Siemens AG Rev. 01 10.02 CS PS 24/SPS Medical Solutions