Olympus CX-31, CX-41 Service manual

Page 1
Maintenance Manual
for maintenance engineer
CX31/41
Thank you for purchasing the Olympus microscope CX31/41. In order to fully utilize its performance and secure optimum condition, please read this manual before maintenance work. Please also keep it at hand during maintenance as well as for future reference.
All rights reserved, Reproduction in whole or in part without written permission is prohibited.
Page 2

INTRODUCTION

The purpose of this manual is to satisfy any requirements for maintenance material that maybe considered as necessary to professionals in the maintenance field, such as Maintenance engineer in Hospitals. It is intended to be used as a tool for performing basic maintenance procedures if needed or when required as per the recommended maintenance schedule.
The sections from this manual only cover procedure pertaining maintenance ’s that are considered to be easily performed. For more involved maintenance’s or repairs, it is recommended that you contact a qualified service engineer from your local Authorized Olympus dealer.
Maintenance parts, grease, and other items specified in the manual can be ordered from your Authorized Olympus dealer, and subject to change without notice.
The recommended maintenance schedule is shown below as reference. ( * Necessary item)
Portion Cleaning Optical/mechanical check Maintenance schedule
Optical components
1) Outer surface Eyepiece, objective , condenser lens, filter, photo eyepiece
2) Inner part Prism, internal lenses
Appearance
Microscope frame, mechanical part
Observation tube
Mechanical part
Focusing unit, stage, revolving nosepiece, aperture/field iris diaphragm
*
*
*
*
Optical adjustment :
1) Optical axis (standard)
2) Left/right optical axis
3) Revolving axis
4) Parfocality *
Mechanical movement :
Grease replacement Mechanical adjustment
Once in a year (If dirt is conspicuous or oil immersion objec tive is used, cleaning should be made after every use.)
Once in a year
Once in a year (If dirt is conspicuous, cleaning should be made after every use.) Once in a year
Once in two to three years
Page 3

CONTENTS

CHAPTER 1 MAINTENANCE PROCEDU RE........................................................................ 1
1. Maintenance of Microscope............................................................................................ 1
2. Guide to Maintenance.................................................................................................... 2
2-1 Overview of maintenance....................................................................................... 2
2-2 Cleaning method for the optical components.......................................................... 3
3. Preparing for Inspection................................................................................................. 6
4. CX31/41 Inspection Sheet.............................................................................................. 7
CHAPTER 2 INSPECTION PROCEDURE............................................................................. 8
1. Checking Performance of Microscope............................................................................ 8
2. Checking Dirty Portion.................................................................................................... 8
2-1 Image influence caused by dirt on each component............................................... 8
2-2 How to find dirty portion through observation.......................................................... 9
2-3 How to check cleaning condition ..........................................................................10
CHAPTER 3 REPAIR PROCEDURE.....................................................................................11
1. Optical Adjustment ........................................................................................................11
2. Mechanical Adjustment................................................................................................. 14
2-1 Preparation for the tension adjustment of Y-wire................................................... 14
2-2 Adjustment method for the tension of Y-wire......................................................... 15
2-3 Confirmation of the Y-wire tension ........................................................................15
2-4 Final adjustment................................................................................................... 16
3. Replacing Grease for Fine Focus Adjustment Knob Ass’y............................................. 17
4. Electrical Adjustment.................................................................................................... 18
4-1 CX31/41 wiring diagram ....................................................................................... 18
4-2 Replacing circuit board / rheostat ass’y ................................................................18
4-3 Preparation ..........................................................................................................19
4-4 Voltage adjustments............................................................................................. 19
CHAPTER 4 JIGS AND TOOLS / GREASES AND ADHESIVES .......................................20
1. List of Jigs and Tools.................................................................................................... 20
2. List of Greases............................................................................................................. 20
3. List of Adhesives.......................................................................................................... 20
CHAPTER 5 MAINTENANCE PARTS................................................................................. 21
1. List of Maintenance Parts............................................................................................. 21
Page 4
CX31/41 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
CHAPTER 1
MAINTENANCE PROCEDURE

1. Maintenance of Microscope

1) Fundamental handling
a. Read the instruction manual thoroughly, handle the microscope correctly. b. Be sure to make a usual cleaning, especially after every use of microscope. c. When handling the microscope, do not expose it to shock, moisture, heat and dust. d. If the problem occurs, do not treat it in self-judgement.
2) Using and storing conditions
a. The microscope should be used under no vibration.
If it is placed in such environment as vibration, this causes disturbance and fatigue in observation and affects the photomicrography.
b. It should not be stored in high humidity.
Such condition causes fungus, corrosion on lens and rust on metal part. Therefore, special caution is exercised if stored in a long period of time.
c. Intense temperature change should be avoided.
Be careful not to place the microscope near window exposed to direct sunlight and air conditioner. When bringing it into warm room from cold storage location in winter, condensation occurs on lens and metal part, it causes fungus and rust.
d. It should be kept clean.
The microscope is required to keep away from dust that causes contamination, fungus on lens and failure of sliding part in the frame. Be careful not to place it in the environment where the corrodent chemicals such as hydrogen sulfide, hydrogen fluoride and acid are handled.
<Reference> fungus: occurrence conditions
Temperature
Dry blue aspergillus
Degree(C)
40
30
Humidity
%
60 70
- 1 -
Aspergillus
80
20
10
90
100
Page 5
CX31/41 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
with the soft brush. If there are stains on
lens

2. Guide to Maintenance

2-1 Overview of maintenance

1)
3 4
5
2
6
1
Set your correct interpupillary distance. Note any areas suggesting a need for mechanical and/or optical maintenance by operating it or observing a specimen.
4)
4
3
5
Plastic part
2
6
1
2)
3 4
5
2
6
1
Sweep off dust on the outer surfaces the microscope frame, clean them with
neutral detergent.
5)
3)
3 4
5
2
6
*removed if necessary
1
*Attachment
Remove the optical components and the specimen holder. (The attachment lens is removed from the condenser with the condenser turned upside down.)
6)
4
3
5
2
1
6
plastic part
Wipe off any dirt spots on the surface of lenses, filter, glass and stage. The plastic parts should be cleaned with neutral detergent.
7)
3 4
5
2
6
1
Polish all plastic components and the frame with silicon cloth.
Tool required:
Clean all exposed optics such as lenses of eyepieces and objectives.
8)
3 4
5
2
6
1
Return the interpupillary distance to original condition and prepare for the inspection. (see page 6)
Mount the removed components to the microscope frame.
9)
Do a final check following the inspection sheet in this manual. (See page 7)
Lens tissue Neutral detergent
(For plastic part or
frame)
Cleaning solution
(For lens or filter
etc.)
- 2 -
Cotton swab
or tweezers
etc.
Blower Silicon cloth
(For finishing)
Page 6
CX31/41 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
Put it on the desk

2-2 Cleaning method for the optical components

Required tools:
1) Lens tissue
2) Cotton swab or tweezers etc.
3) Blower
4) Magnifier (Eyepiece is possible to be used by turning it upside down. Refer to page 10.)
5) Cleaning solution: e.g. Alcohol Before cleaning: Lightly brush the lens surface or blow with the blower before wiping with tissue.
This removes particles that may scratch the lens surface. (to protect the lens coating)
HOW TO CLEAN THE OBJECTIVE LENS
1 2 3 4
Moisten the tip of cotton swab with cleaning solution.
When the lens size is large and difference in level between the lens and the lens frame is small:
Fold the lens tissue several times and moisten it with cleaning solution. After that, apply the folded line edge to the center of lens, push it with index finger and turn the objective by the other hand to clean the lens while moving it from the center towards the periphery.
With a circular motion, wipe the top lens surface with the cotton swab, to thoroughly remove any oil or dirt from the lens.
Dip a new cotton swab in the cleaning solution and shake vigorously to remove any excess cleaning solution.
Cleaning the immersion objective:
Wipe off the immersion oil while absorbing it with lens tissue that is not moistened. After that, clean the lens as the same manner mentioned on the left. When the top surface of lens frame is higher than that of lens and remained dirty potion on the periphery of lens can not be wiped off, clean the lens referring the above figures, 1 to 4.
Wipe the objective lens from the center towards the periphery, while rotating the lens.
- 3 -
Page 7
CX31/41 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
HOW TO CLEAN THE FILTER
Fold the lens tissue into two or three layers and moisten its shaded part with cleaning solution.
Hold the filter at its edge and fold the lens paper from the lens center to outside as illustrated. Move the lens tissue gradually to outside while turning the filter by left hand.
HOW TO CLEAN THE PRISM
Hold a sheet of lens tissue between your middle and index fingers, then fold and wrap it around your index finger. Hold the tissue down with your thumb and moisten it with sufficient cleaning solution.
1)
2) 3)
4)
Wipe the prism surfaces from front to backward at a stroke, applying even pressure.
5)
6)
- 4 -
Page 8
CX31/41 MAINTENANCE MANUAL MAINTENANCE PROCEDURE
HOW TO CLEAN THE EYEPIECE
1 2
Wrap a sheet of lens tissue around a cotton swab as illustrated. If the area to be cleaned is large, wrap the lens tissue looser and thicker. Otherwise, make a thin, tight wrap.
Dip the wrapped lens tissue in the cleaning solution, and wipe the eyepiece from the center towards the periphery in a circular motion.
Important notes:
1) Never rub the lens surface strongly.
2) Do not use the same lens tissue to clean more than one lens .
3) Do not moisten the lens tissue with an excessive amount of cleaning solution.
4) When cleaning with tweezers, be careful not to protrude its tip from the lens tissue.
- 5 -
Page 9
CX31/41 MAINTENANCE MANUAL MAINTENANCE PROCEDURE

3. Preparing for Inspection

ADJUSTMENT OF KOEHLER ILLUMINATION
1) Set the main switch “A” to “I” (ON) and adjust the brightness by turning the adjustment knob “B” .
2) Place a specimen on the stage.
3) Engage the 10X objective in the light path.
4) Turn the condenser height adjustment knob “C” to raise the condenser to its upper limit.
5) Looking through the eyepiece in the right sleeve without the diopter adjustment ring, turn the coarse and fine focus adjustment knobs “D” to bring the specimen into focus.
6) Looking through the eyepiece in the left sleeve with the diopter adjustment ring, turn only the diopter adjustment ring “E” to focus on the specimen. (At this time, adjust the interpupillary distance so that the binocular visions on the left and right fields of view coincide completely.)
7) Turn the field iris diaphragm ring “F” counterclockwise so that the iris diaphragm image comes inside the field of view.
8) Manipulate the condenser height adjustment knob “C” to focus on the iris diaphragm image. (See Fig.1)
9) Turn the two centering screws “G” of attachment lens to move the iris diaphragm image to the center of field of view. (See Fig. 2)
10) Gradually open the field iris diaphragm. The condenser is properly centered if the iris diaphragm image is centered and inscribed in the field of view. (See Fig.3) (During actual use, open the field iris diaphragm slightly until its image circumscribes the field of view. See Fig.4) Reference: Since the contrast of microscope specimens is ordinary low, setting the condenser
aperture iris diaphragm to between 70% and 80% of the N.A. of the objective in use is usually recommended. If necessary, adjust the ratio by removing the eyepiece and looking into the eyepiece sleeve while adjusting the aperture iris diaphragm ring “H”.
(E)
(H)
(G)
(F)
(C)
(A)
4
3
5
2
6
1
(B)
(D)
Field iris diaphragm image
Fig.1
Fig.2
Fig.3
Fig.4
- 6 -
Page 10
CX31/41 MAINTENANCE MANUAL MAINTENANCE PROCEDURE

4. CX31/41 Inspection Sheet

Model : Check Date : S/N : Checking by :
Check Point Check Contents Result Ref. Page
1. Electrical unit 1) When the power switch is turned on, the lamp is lit and the brightness can be varied by adjustment knob.
2. Coarse/fine focus
adjustment knob
3. Stage
4. Observation tube
5. Revolving
nosepiece
7. Illumination 1) The field/aperture iris diaphragm ring is moved
8. Visibility
(Observation)
Remarks:
1) The coarse/fine focus adjustment knob is smoothly moved without any stress or unevenness.
2) The tension of coarse focus adjustment knob can be adjusted by the adjustment ring.
3) CX31: The upper limit is changed by turning the
ring of stopper on the front upper side.
CX41: The coarse upper limit lock is effective.
1) The stage should not fall spontaneously. OK / NO
2) A specimen is held securely by the specimen holder.
3) The X/Y movement is smooth without unevenness, backlash or slipping.
1) The interpupillary distance adjustment can be operated smoothly in working range.
2) When changing the interpupillary distance, the displacement of optical axis is not apparent.
3) The diopter adjustment ring is moved smoothly in working range.
4) The optical axis of left side coincides with that of right side.
5) CX41: The tilting angle is easily adjusted and its
angle is stable. (when the U-TBI3 is used: )
6) CX41: When changing the tilting angle, the
optical axis displacement is not apparent. (when the U-TBI3 is used: )
1) The revolving nosepiece can be rotated smoothly and stops at the click position.
1) The vertical movement of condenser is smooth. OK / NO6. Condenser
2) The centering of field iris diaphragm can be adjusted with the centering knobs of attachment lens(CX-AL).
normally.
1) Observation image is normal. Without flares /ghosts / vignetting /uneven illumination
2) When photographed, there is no unevenness or shading on the film surface.
3) Dust and dirt are not noticeable in observation or photomicrography.
OK / NO 18, 19
OK / NO 17
OK / NO
OK / NO
OK / NO
OK / NO 15, 16
OK / NO
OK / NO
OK / NO
OK / NO 11, 12, 13
OK / NO OK / NO
OK / NO
OK / NO
OK / NO
OK / NO
OK / NO
OK / NO 8, 9
- 7 -
Page 11
CX31/41 MAINTENANCE MANUAL INSPECTION PROCEDURE
CHAPTER 2
INSPECTION PROCEDURE

1. Checking Performance of Microscope

Using the CX31/41 inspection sheet (P.7), check the electrical unit, mechanical and optical performance.

2. Checking Dirty Portion

2-1 Image influence caused by dirt on each component
The following figure shows the influence of image on each optical component if stains or dust is adhered to that portion. In general, the microscope image or photographing is largely affected by dirt adhered on the nearer portion to a specimen and image surfaces. Therefore, the optical components should be kept clean and dust-free.
C Prism
B Eyepiece
D Objective
OLYMPUS
CX41
A Specimen
B Condenser lens
B Attachment lens
(Lens and filter)
B or A Glass
304
1
5
6
C Frosted glass
C Collector lens
A: Dirt is clearly seen. B: Blurred contours of dirt is seen. C: Dirt is seen when the aperture iris diaphragm is stopped down. D: Dirt is not directly seen, but contrast of image deteriorates
- 8 -
Page 12
CX31/41 MAINTENANCE MANUAL INSPECTION PROCEDURE

2-2 How to find dirty portion through observation

1) Close the aperture iris diaphragm. (When the aperture iris diaphragm is closed, it facilitates finding the dirt particles
because the depth of focus increases and the dirt position bring into focus. However, very small dirt particle may not be found in this method.)
2) Observe a specimen through the eyepiece. If dirt is seen by observing it, look for the portion where stains or dust is adhered by
moving the following components as well as a specimen. For TV camera, check it by method shown below.
Photo eyepiece
Check it by rotating the photo eyepiece. (The dust particles will be focused on the film plane.)
Eyepiece Check it by rotating the eyepiece. (Wipe carefully and gently so as not to damage the lens coating.)
Objective Dirt is not directly seen, but dirt and dust particles affect the microscope image.
Condenser Check it by loosening the fixing screw and turning the condenser.
Collector lens Check it if conspicuous dirt is not seen. (Dust and dirt particles here have minimal effect on the image. However, the overall appearance requires inside lens cleaning.)
OL
YMPUS
NFK5XLD 125
TV camera Dust particles appearing on the monitor screen: if the particles move only when the specimen or condenser is moved, clean the specimen and/or condenser. If they do not moved when the TV camera is rotated, clean the camera tip.
4
3
5
2
6
1
Mirror for Large format attachment
Dust and dirt particles here affect pictures. Difficult to clean without damaging the mirror surface. Please contact your Authorized Olympus dealer.
Note: If dirt particles do not move by moving the above components, it is assumed that
internal lenses are contaminated. In this case, please contact your Authorized Olympus dealer.
- 9 -
Page 13
CX31/41 MAINTENANCE MANUAL INSPECTION PROCEDURE

2-3 How to check cleaning condition

1) When a large lens is checked, look at the lens while putting it toward bright side or breathe on the lens and observe the condition that the haze on the whole surface of the lens disappears evenly.
Light
Dust becomes conspicuous when looking at it with the lens inclination changed.
2) For a small lens such as top lens of objective, check it by magnifier.
If there is a dirty part or a remained part that is not cleaned completely, the haze of this part will disappear slower than that of the other part.
An eyepiece can be substituted for magnifier by turning the eyepiece upside down.
- 10 -
Page 14
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE
CHAPTER 3
REPAIR PROCEDURE

1. Optical Adjustment

PREPARATION
*1 WHB10X
*1 It is necessary to set the cross micrometer disk to the above eyepiece.
Adjusting the left/right optical axis
If the left/right optical axis is remarkably displaced at checking, perform the following adjustment.
*Insert the cross eyepiece into the right sleeve.
Objective 4X-10X
specimen
whose center is identified (Concentric circles etc.)
Align the specimen center with the cross center of the
WHB10X eyepiece by turning the control knob of the stage.
Before adjustment
After adjustment
*Adjust the interpupillary distance to about 62mm (See the illustration)
Position of approx. 62mm
75
70 60 50
Cross center of WHB10X
specimen center
Control knob
- 11 -
Page 15
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE
ADJUSTMENT THE LEFT/RIGHT OPTICAL AXIS
(1) Moving the cross eyepiece to the left sleeve
Work Image seen through the cross eyepiece
Move the cross eyepiece to the left sleeve.
If the optical axis between left and right sleeve is deviated, the center of the
75
70
60 50
specimen and the cross center of eyepiece are also deviated.
(2) Aligning the cross center of eyepiece with the specimen center
Work Image seen through the cross eyepiece
1. Loosen the two screws slightly which
Image at first
secure the left sleeve.
75
70 60
50
2. Align the center of eyepiece with the specimen center while observing through the WHB10X. (Change the left sleeve position by hand.)
75
70 60 50
The center is aligned.
- 12 -
Page 16
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE
3. Firmly tighten the screws which secure
Image at the end of adjustment
the left sleeve.
75
70
60
50
* If this adjustment could not be done, inside mechanism may be damaged.
Please contact your Authorized dealer.
- 13 -
Page 17
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE

2. Mechanical Adjustment

2-1 Preparation for the tension adjustment of Y-wire

If a specimen image is moved when the stage is brought into the desired position of specimen, it is necessary to adjust the wire tension of stage.
1)
Back plate ass'y
*1
Remove the back plate ass’y from the frame. (For the screw positions, refer to the above figure.)
: Attachment screws to the frame
CUK3X4SA 5pcs. CUKHWB3X6SA 1pc. (lower left: *1)
2)
a
(up)
b
(down)
CN1
Inlet cable
of power
switch
CN3
CN2
b
(White)
a
(Black)
Disconnect the cable connectors from the circuit board (CN1-CN3) and also the inlet able connectors from the power switch terminals.
3)
*Stage holder reference direction
: Stage holder attaching screws
ABSK3X10SA 4pcs.
Remove the screws on the back inner section and the stage holder. (For the screw positions, refer to the above figure.)
4)
4
3
5
2
6
1
Separate the stage from the frame, turn over the stage and put on the table. (When assembling, install the stage holder while pushing it to the lower right as seen from the rear. *See the left figure.)
- 14 -
Page 18
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE

2-2 Adjustment method for the tension of Y-wire

A007
*1
*2
(A)
1) Loosen the screws (*1) securing the holder.
Screws: CUK 3X6SA (*1) 2pcs. (clamping)
(B)
2) Adjust the tension of the wire by turning the screw (*2) clockwise to tight or counter­clockwise to release.
Screw: CSK3X6SA (*2) 1pc.(adjustment)
3) Temporarily tighten the screws (*1).

2-3 Confirmation of the Y-wire tension

1) Check that the Y-knob (A) is rotated following the movement of lower stage (B) when moving the lower stage in stroke by a hand.
2) After confirming the above condition, Install the stage to the frame,set the microscope at observation state and check that the image is brought into the desired position without backlash (within 2 microns). If it is out of standard, perform the adjustment referring to 2-4 on page16.
3) Repeat the above procedures until the condition is satisfied. After adjustment, tighten the screws (*1) firmly and apply adhesive to three screws
(*1, 2).
Adhesive: OT1378 (Solvent-based adhesive)
Note:
1) If the wire is too tight, image backlash may occur. If the wire is loose, slip may occur between the knob and wire deteriorating the image movement.
2) In case where the stage movement is heavy due to hardening of grease or the X- wire adjustment is necessary, please contact your Authorized Olympus dealer because disassembling the stage is required for grease replacement or that adjustment.
- 15 -
Page 19
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE

2-4 Final adjustment

Image backlash adjustment:
1) Under observation state (with 100X objec­tive), move the stage to the desired image position by turning the Y-knob (A). At that stop position, check image backlash. If it is over 2 microns, conduct the following adjustment.
2) When adjusting the Y-movement, loosen the two screws (*1) and turn the Y-knob (A) to bring backlash within 2 microns.
* After turning the knob and temporarily
tighten the screws, check image backlash in the observation state. Repeat the adjustment until image backlash is within the standard.
A009
(B)
A008
(A)
*1
*2
Screws: AWU3 X4SA (*1) 2pcs.
3) For the X-movement, check that the backlash can be adjusted by turning the X-knob (B) in the same manner as the Y-knob adjustment.
* If it could not be adjusted, contact your
Authorized dealer because X-wire need to be adjusted.
Screws: AWU3 X4SA (*2) 2pcs.
- 16 -
Page 20
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE
3. Replacing Grease for Fine Focus Adjustment Knob Ass’y
If the fine focus adjustment knob is not turned smoothly, replace grease on the shaft of fine focus adjustment ass’y in the following procedure. (In case where the coarse focus adjustment knob is not turned evenly, please contact your Authorized Olympus dealer because it is necessary to disassemble the coarse focus adjustment knob ass’y and/or guide unit.)
1) Peel off the plates on the left and right sides of fine focus adjustment knob. Insert Allen wrenches (2.5mm) into the both side screws.
Screws: ABS3X8SA 2pcs. (*1)
2) Turn the screw (*1) on the right side with Allen wrench while the left side screw is held with it, and pull out the fine focus adjustment knob ass’y.
(At this time, the fine focus adjustment knob ass’y (right ) is removed as an assembly of
shaft (a), gear (b) and fine focus adjustment knob (c).)
3) Confirm that the gear is not damaged. If it is damaged, replace the gear or the fine focus adjustment knob ass’y as a whole.
4) Remove grease (OT2008) on the shaft by cleaning solution and replace it. *When applying the grease, refer to the explanation shown below.
5) Assemble the components in reverse order of disassembly.
4
3
5
2
1
A006
H
Applying grease (OT2008) to the shaft (a) of the fine focus adj.knob ass’y and the sliding surface: Apply grease to the outer surface of the shaft. First, spread the grease by inserting from the left while rotating the fine focus adj. knob ass’y. After removing the fine adj. knob ass’y, place the shaft at the opposite side mouth (to the right). Again apply grease to the shaft and then insert into the right side. Use the lens tissue to wipe away any grease exuding from the left and right.
Fine focus adjustment knob ass'y
6
A002 (a) *Apply grease(OT2008) to the shaft.
Adhesive(OT1378) has been applied to the screw thread of the shaft on the right side.
A004
A003 (b)
A005 (c)
*1
Plate
A001
- 17 -
Page 21
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE
(C)

4. Electrical Adjustment

4-1 CX31/41 wiring diagram

U003
CN2
U004
1
2
A012
(AQ802400)
6V Halogen
bulb
U001
INPUT
100-120/ 220-240VAC 50/60Hz
H
Voltage Adjustment
A010
U002
S001
X001
N
L
E
X002
1
CN1
3
Main Board
A011
Rheostat
Ass'y
CN3
Dimmer Cont
The circuit board ass’y (AQ802400) consists of the circuit board (DZ290300) and rheostat ass’y (DZ290400). In case where the above parts are replaced as AQ802400, the voltage adjustments for the circuit board ass’y are not necessary. However, voltage adjustments are necessary when individually replacing either the circuit board (DZ290300) or rheostat ass’y (DZ290400). The following explains procedures for replac ing, preparation and adjustment of the minimum and maximum voltages .
4-2 Replacing circuit board / rheostat ass’y
1) Remove the screws securing the BACK
*1
PLATE ASS’Y (A).
Screws: CUK3X4SA (*1) 5 pcs.
CUKHWB3X6SA (*2) 1 pc.
(A)
*3
2) Disconnect the cables from the CIRCUIT BOARD (B) connectors.
*2
*4
(B)
CN1
a
CN1: Output side connector CN2: Lamp connector CN3: Rheostat ass’y connector
3) Disconnect the inlet CABLES (C) from the power switch terminals, and then take off
CN3
CN2
a
the BACK PLATE ASS’Y (A). (No need to disconnect the “a” connection)
4) Remove the CIRCUIT BOARD (B).
Screws: CUK3X4SA (*3) 3 pcs.
(D)
CUKHWB3X6SA (*4) 1 pc.
5) Replace the CIRCUIT BOARD (B) / RHEOSTAT ASS’Y (D). (Rheostat ass’y (D): Please contact your Authorized dealer.)
- 18 -
Page 22
CX31/41 MAINTENANCE MANUAL REPAIR PROCEDURE

4-3 Preparation

1) Connect the cables to each connector
Inlet terminals
(CN1-CN3).
a
(up)
b
(down)
(A)
CN1
CN3
CN2

4-4 Voltage adjustments

CN1
b
(white)
a
(black)
Halogen bulb
2) Connect the cables to the power switch inlet terminals. Upper side: black Lower side: white
3) Set the digital multimeter so that the voltage of the CN2 2-pin can be measured.
H
Use a new Halogen bulb when making voltage adjustment
Minimum Voltage Adjustment
1) Turn ON the power. Turn the light intensity control knob (A) counterclockwise to lower the lamp brightness to its lowest level.
VR21
VR23
CN3
CN2
VR22
* Do not touch the inlet terminals with
power turned on because it could cause an electric shock. Be sure to turn off power during assembly and reassembly.
* Do not turn the trimmer VR22 because it
is adjusted to the prescribed current value. (VR22 is set for overcurrent protection.)
2) Rotate the circuit board’s trimmer VR21 to adjust so that the lamp output voltage between the CN2’s 1 and 2 pins is within the standard shown below using a digital multimeter.
Standard: DC1.10~1.15V
(adjustment target: 1.123V)
Maximum Voltage Adjustment
1) Turn ON the power. Turn the light intensity control knob (A) clockwise to increase the lamp brightness to its highest level.
2) Rotate the circuit board’s trimmer VR23 to adjust so that the lamp output voltage between the CN2’s 1 and 2 pins is within the standard shown below using a digital multimeter.
Standard: DC 5.65~5.75V
(adjustment target: 5.70V)
- 19 -
Page 23
CX31/41 MAINTENANCE MANUAL JIGS AND TOOLS / GREASES AND ADHESIVES
CHAPTER 4
JIGS AND TOOLS / GREASES AND ADHESIVES

1. List of Jigs and Tools

No. Description Ref. page
Cleaning tools 2, 3, 4, 5
*1 WHB10X
(It is necessary to set the cross micrometer disk.) Digital multimeter 19 Philips screwdriver 12, 13, 14, 15, 18 Allen screwdriver (2.5mm) 14 Allen wrench (2mm, 2.5mm) 16, 17 Precision screwdriver 19 Dust cover
*1 Contact your Authorized dealer.
11, 12, 13

2. List of Greases

No. Description Ref. page
OT2008 Grease (medium) 17

3. List of Adhesives

No. Description Ref. page
OT1378 Solvent-based adhesive (transparent)
(For the fine focus adjustment knob ass’y, use adhesive only when it is disassembled.)
15, 17
- 20 -
Page 24
CX31/41 MAINTENANCE MANUAL MAINTENANCE PARTS
CHAPTER 5
MAINTENANCE PARTS

1. List of Maintenance Parts

Index No. Order No. Description Ref. page
*1 CX31LBSF Instruction manual
*1 CX41 Instruction manual A001 AD071500 Plate A002 AC732500 Shaft A003 AA782600 Gear A004 AQ153900 Fine focus adjustment knob ass’y
(when replaced as an assembly:) A005 AB467500 Fine focus adjustment knob (right) A006 AB059300 Fine focus adjustment knob (left) A007 AD411200 Holder (for Y-wire) 15 A008 AC888300 X-knob A009 AC888400 Y-knob A010 DZ290300 Circuit board A011 DZ290400 Rheostat ass’y A012 AQ802400 Circuit board ass’y
(when replaced as a set :)
*1 6V30W halogen bulb 19
17
16
18
H
The index No. are shown on the figures in “REPAIR PROCEDURE” instead of parts No. *1 Contact your Authorized dealer.
Page 25
OLYMPUS OPTICAL CO., LTD
1-22-2, Nishi shinjuku shinjuku-ku, Tokyo, Japan
Issued by Marketing Dept. Printed in Japan 2001 11 SK9214
Loading...