Nidek Nuvo 8 Service manual

Nidek Medical Products, Inc Nuvo 8 Oxygen Concentrator
Service Manual
Nidek Medical Products, Inc. 3949 Valley East Industrial Drive
Birmingham, Alabama 35217 USA
Telephone: (205) 856-7200 • 24-Hour Fax: (205) 856-0533
General Safety
Instructions
Section 1.0
Introduction
1.1
1.2
1.3
Section 2.0
Operational Check and Concentration Test
2.1
Section 3.0
Patient Instructions
3.1
Section 4.0
Home Service Provider Maintenance
4.1
4.2
Section 5.0
Service
5.1
5.2
2-4
Production and Use of Oxygen 5 Use and Maintenance of the Device 5 Standards and Regulations 5-6
Home Service Provider Responsibility 6-7 Important Notice and Symbol Explanations 7 Functional Specifications 8
2.2
2.3
2.4
3.2
Description of Operation Operation Check Alarm System
2.3.1 Battery Test
2.3.2 Power Failure Alarm test Oxygen Concentration Test and Specification
General Instructions Routine Maintenance by the Patient
3.2.1 Cleaning the Cabinet Air Filter
3.2.2 Checking the Alarm System Battery
Routine Maintenance 12
4.1.1 Cabinet Air Filter 12
4.1.2 Final Product Filter Replacement 12
4.1.3 Inlet Air Filter Replacement 13
4.1.4 Battery Replacement 13
4.1.5 Recording Maintenance 13 Cleaning and Infection Control 13-14
4.2.1 Preparing for New Patient Use 14
10 10
10-11
11 11 11 12
8 9 9
Page 2 of 52 2010-2229 Rev - B
Components 14-15 Cabinet Removal 15
5.2.1 Removing Cabinet Back 15
5.2.2 Removing Cabinet Front 15
5.2.3 Removing Control Panel 15
5.2.4 Main Structure 15
5.2.5 Caster Replacement 15
5.3
Compressor 15
5.4
5.5
5.6
5.7
5.8
5.9
5.10
5.11
5.12
5.13
5.14
5.15
6.1
6.2
6.3
6.4
6.5
5.3.1 Compressor Replacement 15-16
5.3.2 Capacitor Replacement 17 Process Control Valve 17
Sieve Module Replacement 18
5.5.1 Sieve Module Removal 28
5.5.2 Sieve Module Installation 18 Cabinet Fan Replacement 19 Circuit Board Replacement 19-20
5.7.1 Circuit Board Removal 20
5.7.2 Circuit Board Installation 20
5.7.3 Timing Board Removal 20
5.7.4 Timing Board Installation 21 Product Regulator Check and Setting 21
5.8.1 Product regulator cleaning or rebuilding 21 Pressure Switch Replacement 22 Circuit Breaker Replacement 22
5.10.1 Circuit Breaker Removal 22
5.10.2 Circuit Breaker Installation 22 I/0 (ON/OFF) Power Switch Replacement 22
5.11.1 I/0 (ON/OFF) Power Switch Removal 22
5.11.2 I/0 (ON/OFF) Power Switch Installation 23 Buzzer Replacement 23 Hour Meter Replacement 23 Flow Meter Replacement 23
5.14.1 Flow Meter Removal 23
5.14.2 Flow Meter Installation 23 Power Cord Replacement 23-24
Section 6.0
Troubleshooting
Air Pressure Test (P1) 24-25
6.1.1 High Air Pressure 26
6.1.2 Low Air Pressure 26 Product Pressure Test (P2) 26-27
6.2.1 High Product Pressure 27
6.2.2 Low Product Pressure 27 General Troubleshooting 27-29 Troubleshooting Chart 30-32 Tool Kit and Pressure Test Gauge 33
2010-2229 Rev - B Page 3 of 52
Appendices
Exploded
Drawings
A-1: Pneumatic Flow Chart 34 A-2: Electrical Schematic, 230 Volt Unit 35
A-3: Electrical Schematic, 115 Volt Unit 36 A-4: Base Assembly 37 A-5: Main Structure Assembly Front View A-5.1: Main Structure Assembly Rear View A-6: Regulator Assembly 40 A-7: Compressor Assembly, 115 Volt Unit 41 A-7.1: Compressor Assembly, 230 Volt Unit 42 A-8: Control Panel Assembly; Front View 43 A-8.1: Control Panel Assembly; Rear View 44 A-9: Front Cabinet Assembly 45 A-10: Rear Cabinet Assembly 46 A-11: Module Assembly 47 A-12: Maintenance Log 48-49
38 39
Page 4 of 52 2010-2229 Rev - B
General Safety Instructions
Production and use of oxygen
Oxygen is not a flammable gas, but accelerates the combustion of materials. To prevent fire risks, the Nuvo 8 should be kept away from flames, incandescent sources or sources of heat (cigarettes) and combustible products such as oil, grease, solvents, aerosols, etc.
Do not use in an explosive atmosphere. Prevent oxygen from accumulating on upholstered seats or any other fabric. If the
concentrator operates without being administered to a patient, locate it so that the flow of gas generated is dissipated into the air.
Locate the equipment in a free space (filter to the rear and below) which is well ventilated and free of fumes or atmospheric pollution.
Use and Maintenance of the Device
Use the electric cable provided and check that the voltage of the mains socket used complies with the electrical characteristics of the appliance indicated on the manufacturers plate on the rear of the appliance.
Do not use an extension cord or multiple sockets which can create sparks and therefore pose a fire risk.
Use of the Nuvo 8 must be restricted solely to oxygen therapy on medical prescription in compliance with the daily rate and duration.
Use in other circumstances may represent a hazard to patient health. Do not use in a specifically magnetic environment (MRI, etc.).
The Nuvo 8 has an audible alarm intended to warn the user of any problems. The user must determine the maximum distance away from the Nuvo 8based on the
sound levels in the environment, to ensure that the alarm is always audible.
Standards & Regulations In compliance with UL60601-1 [EN60601-1] (para 6.82.b):
“The manufacturer, assembler, installer or importer are not considered to be responsible for consequences or the safety, reliability and characteristics of a device unless,
the assembly, extensions, adjustments modifications or repairs have been
performed by persons authorized by the manufacturer,
2010-2229 Rev - B Page 5 of 52
the electrical installation of the corresponding premises complies with
appropriate regulations and codes,
the device is used in accordance with the instructions for its use.
If the replacement parts used for periodic servicing by an approved technician do not comply with the manufacturer’s specifications, the manufacturer is absolved of all liability in the event of an incident.
Do not open the equipment when it is powered on: risk of electrocution.
This device complies with the requirements of the FDA Quality System Regulation and EU Directive 93/42/EEC, but its operation may be affected by the use in the surrounding area of appliances such as diathermy, high frequency electro-surgical instruments, defibrillators, short wave treatment appliances, cell-phones, CB devices and other portables, microwave ovens, induction hot plates or remote control toys, and more generally, by electromagnetic interference exceeding the levels specified in standard IEC(EN) 60601-1-2.
1.0 Introduction
1.1 Home Service Provider Responsibility
All Home Service Providers of the Nidek Medical Nuvo 8 Oxygen Concentrator must assume responsibilities for handling, operational check-out, patient instruction, and maintenance. These responsibilities are outlined below and throughout this manual.
WARNING Nuvo 8 units must not be used for or with any life-supporting or life sustaining
applications. Patients unable to communicate discomfort while using this device may require additional monitoring. Advise patients to immediately notify their Home Service Provider(s) and/or physician(s) in case of an alarm or any discomfort.
As a Home Service Provider, you must do all of the following:
● Inspect the condition of each Nuvo 8 unit immediately upon delivery to your business location. Note any sign of damage, external or internal, on the delivery receipt, and report it directly to both the freight company and Nidek Medical Products, Inc. immediately.
● Check the operation of each Nuvo 8 before delivery to a patient. Always operate the unit for a reasonable length of time and check that the oxygen concentration level is within specifications as referred to in Section 2.4. Test the battery and power disconnect alarm as described in Section 2.3 of this manual.
● Deliver Nuvo 8 units only to patients authorized by a physician’s prescription. The Nuvo 8 must not be used as a life-supporting or life sustaining device. A backup supply of oxygen must be available.
Page 6 of 52 2010-2229 Rev - B
● Instruct patients how to use the Nuvo 8 in conjunction with the Patient
a hazard or unsafe practice that can result
Describes a hazard or unsafe practice that can result
Provides information important enough to emphasize
Manual.
● Instruct patients to notify their physicians and/or Home Service Providers
if they experience any signs of discomfort.
● Instruct each patient how to perform routine maintenance of the cabinet
air filter and how to check the alarm system battery, if equipped. (Refer to Section 3.2.) Be available to service each patient at any time. Maintain the Nuvo 8 in accordance with Section 4.0.
Repair components and replace parts only as outlined in this manual. Use only Nidek Medical parts for replacement in Nuvo 8 Oxygen Concentrators.
● Refer to the Nuvo 8 Product Warranty if parts replacement is required within the warranty period.
1.2 Important Notice and Symbol Explanations
As you read the manual, pay special attention to the WARNING, CAUTION, and NOTE messages. They identify safety guidelines or other important information as follows:
WARNING:
CAUTION:
NOTE:
The following harmonized symbols (pictograms), used for non-English language countries, will be located on the exterior of the Nuvo 8 unit:
Read the accompanying documents; particularly the patient operating guide.
Store, ship and use the device in an upright condition.
No smoking within five feet of this device, oxygen-carrying tubing, or accessories.
Indicates an alarm signal.
Do not use any oil or grease on or near the device
Describes in severe bodily injury or death.
in minor bodily injury or property damage.
or repeat.
2010-2229 Rev - B Page 7 of 52
1.3 Functional Specifications
Dimensions: 15.5 in. long, 15.6 in. wide, 27.8 in. tall
(394 mm long, 396 mm wide, 706 mm high) Weight: 50 to 54 Ib (22.5 to 24.5 kg) Electrical 120 VAC, 60 Hz, 4.4-5.0 amps, 500 watts avg
Requirements: 230 VAC, 50 Hz, 1.8-2.1 amps, 490 watts avg
230 VAC, 60 Hz, 2.4-2.7 amps, 585 watts avg Capacity: 8 liters per minute at 90% oxygen
(Based on 70°F [21°C] at sea level) Accuracy: Flow meter ±5% full scale. (Based on pre-calibrated
analysis.) Concentration: 2-8 liters per minute at 90% + 6.5 / - 3% Response Time: Acceptable concentration is normally achieved in about
90 seconds; allow 5 minutes to attain full concentration. Positioning: Operate the unit in an upright position, maintaining at
least six inches of open space on all sides for ventilation.
2.0 Operational Check and Concentration Test
2.1 Description of Operation
Air enters the Nuvo 8 Oxygen Concentrator through an external cooling air intake filter. This filtered air enters the compressor via a suction resonator and fine filter, which quiets the suction sounds made by the compressor. Pressurized air then exits the compressor and passes through a heat exchanger, which reduces the temperature of the compressed air. Next, an electronic valve system directs the air into one of two sieve beds that contain molecular sieve. The special characteristic property of molecular sieve is that it physically attracts (adsorbs) nitrogen when air passes through this material, thus enabling the production of high purity oxygen.
There are two sieve beds or adsorbent columns; while one produces high purity oxygen, the other is purged of the nitrogen it adsorbed (collected) while it was producing oxygen. Each column produces oxygen for approximately 5.5 seconds and delivers it to the product storage volume in the end of the column. Oxygen exits the adsorbent column through a pressure regulator, flow control valve, flow meter, and final product filter. The flow control valve, which is part of the flow meter, controls the flow rate of oxygen delivered to the patient. The Nuvo 8 unit delivers up to 95% oxygen concentration at flow rates from 2 to 8 l/min. The remaining constituents of the product gas stream are nitrogen and argon, both of which are part of the air we breathe, are inert and are completely safe.
Page 8 of 52 2010-2229 Rev - B
2.2 Operational Check
Nidek Medical runs each device through a burn in period and tests every Nuvo 8 Oxygen Concentrator thoroughly after manufacture before releasing for shipment. As the home service provider, it is your responsibility to perform the following test to ensure that no damage occurred in shipping or handling.
1. Open and inspect all concentrator cartons upon receipt. Unpack each
unit and remove it from its carton. Inspect the unit itself for damage. If the exterior of the carton is damaged, or the unit itself is damaged, note it on the freight bill signed by the driver.
● Before plugging the unit into the wall outlet, actuate the I/0
(ON/OFF) switch to the I (ON) position and note that the audible alarm sounds continuously. (See Section 2.3). Move the switch to the 0 (OFF) position.
2. Plug in the power cord of the unit, and set the I/0 (ON/OFF) switch to
the I (ON) position. Check to see that the following occurs:
● The compressor runs, listen for the sound.
● Exhaust air flows out of the bottom of the unit.
● OPTIONAL for Units Equipped with Oxygen Concentration Status
Indicator (OCSI): The OCSI green light remains off until the oxygen concentration reaches 85% @ 60 Hz and 83% @ 50 Hz. (approximately two minutes).
3. Turn the flow meter adjustment knob counterclockwise until it stops
(wide open). The flow meter should register in excess of 8.5 liters/min. If not, refer to Section 5.8 to adjust the product regulator.
4 Perform an oxygen concentration test, as described in Section 2.4.
2.3 Alarm System
The Nuvo 8 Oxygen Concentrator is equipped with an alarm system, which sounds a continuous and loud alarm when a power failure occurs or when one or more cycle variables are not within specification. It sounds an alarm if the high or low pressure indicators are activated (standard models only) or if the optional OCSI detects lower than predetermined levels of oxygen concentration. The alarm remains on until you correct the alarm condition or you set the I/0 (ON/OFF) switch to the 0 (OFF) position. Refer to Section 6.0 for a list of probable alarm causes.
2010-2229 Rev - B Page 9 of 52
2.3.1 Battery Test
If equipped, the Nuvo 8 battery should be checked by observing that the continuous alarm sounds when the I/0 (ON/OFF) switch is moved to the I (ON) position when the unit is unplugged from the wall outlet.
2.3.2 Power Failure Alarm Test
To test the power failure alarm, perform the following actions: Unplug the power cord from the wall outlet, and set the I/0 (ON/OFF) switch to the I
(ON) position. This should immediately activate the audible alarm. If it doesn’t, refer to the
troubleshooting chart in Section 6.0 of this manual.
2.4 Oxygen Concentration Test and Specification
To ensure that the output of oxygen from the device is within specification, you must perform an oxygen concentration test. Test the unit upon delivery to a patient and at periodic intervals. Home Service Providers, based on their expertise and documentation, may establish and implement their own plans for checking oxygen concentration. Consult Nidek Medical’s Service and Maintenance Log (A-12) for the recommended intervals for testing.
1. If an oxygen humidifier bottle is used, remove it from the oxygen outlet.
2. Connect a calibrated oxygen concentration analyzer to the oxygen outlet.
3. Set the I/0 (ON/OFF) power switch to the I (ON) position. (It takes approximately five minutes for the oxygen concentration to stabilize.) Take oxygen concentration readings over a period of several minutes to reduce any cyclic variations
4. Verify that the product flow rate delivered by the unit matches the patient’s prescription and does not exceed the capacity of the unit.
5. Disconnect the oxygen analyzer, and reconnect the humidifier bottle (if used) and any other equipment / accessories that may be required.
6. Adjust the flow meter adjustment knob to the prescribed flow rate.
NOTE:
Do not measure oxygen concentration output after the product stream passes through a humidifier bottle, erroneous readings will result and your oxygen concentration measuring device might be damaged.
Page 10 of 52 2010-2229 Rev - B
Nidek Medical Nuvo 8 Concentration Specifications
Remove the dirty cabinet air filter from the back of the
Wash the dirty filter in warm water with household
Reinstall the clean cabinet air filter on the grille in the
Liter Flow Specification Within Specification 2 to 8 l/min 90% + 6.5 / - 3% 87% or higher
3.0 Patient Instructions
3.1 General Instructions
It is important that patients thoroughly understand how to operate the Nidek Medical Nuvo 8 unit. This enables proper treatment as prescribed by a qualified, licensed physician. You must explain that the purpose of this therapy is to alleviate symptoms. If patients experience any discomfort or the unit alarms, they must notify their Home Service Provider and/or physician immediately. You, as the Home Service Provider, are responsible to see that each patient receives the Patient Manual. Explain each step in the operation of the unit to the patient in reference to this manual.
3.2 Routine Maintenance by the Patient
To ensure accurate output and efficient operation of the unit, the patient must perform two simple routine maintenance tasks:
Clean the cabinet air filter
Check the alarm system
3.2.1 Cleaning the Cabinet Air Filter
NOTE:
The patient must clean this filter weekly, as described below. The filter may require daily cleaning if the Nuvo 8 unit operates in a harsh environment such as a house heated by wood, kerosene, or oil, or one with excessive cigarette smoke.
1
Nuvo 8 unit.
2
detergent, and rinse. 3 Use a soft absorbent towel to remove excess water. 4
back of the unit.
3.2.2 Checking the Alarm System Battery
See Procedure described in Paragraph 2.3.1 2010-2229 Rev - B Page 11 of 52
4.0 Home Service Provider Maintenance
4.1 Routine Maintenance
The Nuvo 8 unit has two filters and a 9-volt battery, if equipped, that require scheduled maintenance and replacement.
To ensure that the output of oxygen from the unit is within specification, you must perform an oxygen concentration test. Test the unit upon delivery to a patient and at periodic intervals. Home Service Providers, based on their expertise and documentation, may establish and implement their own practices for checking oxygen concentration. The interval established may be longer or shorter than 90 days, which is the default time period recommended for providers who do not choose to establish their method.
Nidek Medical does not require preventive maintenance on the concentrator. You do not need to perform any maintenance as long as the Nuvo 8 unit remains within specifications at the prescribed flow rate. (Refer to Section 2.4)
4.1.1 Cabinet Air Filter
The external cabinet air filter is located on the back of the unit; it is easily removed by hand. Instruct the patient to clean this filter weekly. (Refer to Section 3.2.1.)
NOTE:
The filter may require more frequent cleaning if the Nuvo 8 unit operates in a harsh environment such as a house heated by wood, kerosene, or oil, or one with excessive cooking, cigarette smoke or atmospheric dust.
4.1.2 Final Product Filter Replacement
The final product filter does not require periodic replacement; it needs to be replaced only if it restricts oxygen flow. It is suggested that it be replaced whenever the sieve module is repaired or replaced and after the compressor is rebuilt.
1. Set the I/0 (ON/OFF) switch to the 0 (OFF) position, and unplug the power cord.
2. Remove the cabinet back to locate the final product filter. NOTE: Observe the flow direction of the filter before removal.
3. Separate the silicone tubing from both sides of the filter.
4. Install the new filter with the inlet side in the same position as before. Push the tubing together so that it overlaps the barbs of the final product filter connections.
5. Record information about the final product filter replacement in Appendix 12 of this manual and online at www.nidekmedical.com under the ‘Maintenance Log’ tab.
6. Reinstall the cabinet back.
Page 12 of 52 2010-2229 Rev - B
4.1.3 Inlet Air Filter Replacement
The inlet air filter requires inspection at each patient visit. The filter should be replaced annually, or more often depending on the environment.
1. Set the unit I/0 (ON/OFF) switch to the 0 (OFF) position, and unplug the power cord.
2. Remove the rear cabinet access panel to locate the intake filter.
3. Remove filter from the unit, and replace with a new filter.
4. Record information about the filter replacement in Appendix 12 of this manual and online at www.nidekmedical.com under the ‘Maintenance Log’ tab.
5. Reinstall the access panel.
4.1.4 Battery Replacement
Each time the Nuvo 8 unit is turned on, the alarm must sound loudly for approximately five seconds to indicate a good battery. An alarm that does anything other than sound loudly for five seconds indicates a weak battery and requires replacement. For units with OCSI, actuate the switch with the power cord unplugged to test the battery. To replace the battery, take the following steps:
1. Set the I/0 (ON/OFF) switch to the 0 (OFF) position, and unplug the power cord.
2. Remove the access panel in the cabinet back.
3. Pull the battery from the mounting slot and remove the connection wires.
4. Attach the new battery to the connector and reinsert it into the mounting slot.
5. Set the I/0 (ON/OFF) switch to the I (ON) position to test the alarm.
6. Record the battery replacement information in Appendix 12 of this manual and online at www.nidekmedical.com under the ‘Maintenance Log’ tab.
7. Reinstall the access panel.
4.1.5 Recording Maintenance
As the Home Service Provider, it is suggested that you record all routine maintenance and repairs performed on the Nuvo 8 unit, including hours and dates of service in Appendix 12 of this manual and online at www.nidekmedical.com under the ‘Maintenance Log’ tab.
4.2 Cleaning and Infection Control
With the growing concern about possible cross infection from home oxygen equipment (i.e.. oxygen concentrators) from one home care patient to another, appropriate cautions should be exercised. The organisms of most concern are M. Tuberculosis, HIV, and Viral Hepatitis. These are potentially pathogenic.
Tuberculosis can survive outside of the human body, but its mode of transmission is by droplet nuclei. When infected individuals cough, they release droplet nuclei into
2010-2229 Rev - B Page 13 of 52
the air, and these carry the Tuberculosis organism. Another person may breathe in these droplet nuclei, but prolonged exposure to the infected person is usually necessary for infection to occur.
HIV and Viral Hepatitis are both viruses, which are not living cells themselves but which can duplicate when in a living “host” cell. Both of these organisms are usually passed on by person-to-person contact, and both need to be in the human body to survive. Once outside the body, viruses can survive for only a short period of time.
4.2.1 Preparing for New Patient Use
When you remove the Nuvo 8 from a patient’s home, always dispose of the used nasal cannula and humidifier bottle. Clean the exterior of the Nuvo 8 with a soapy water solution or commercial detergent cleaner to remove any debris, organic or otherwise. Be careful not to get any liquid into the interior of the unit.
Next, clean the exterior with either a common chemical disinfectant or a bleach solution* and allow it to air dry. For the bleach solution, wear eye and skin protection to prevent exposure to the chlorine. Retest the Nuvo 8 before you return it to your inventory.
Replace the cabinet air filter between each patient’s use or clean with warm soapy water if it is in good condition. Clean this filter at least once per week or more frequently if operated in a dusty environment.
*Make the bleach solution a 1:100 dilution of 5.25% sodium hypochlorite. Mix one part household bleach (e.g.. Clorox) with 99 parts cold tap water. To measure the solution easily, take 1/4 cup of household bleach, and mix it with a gallon of cold tap water. Allow the mixture to sit on potentially contaminated surfaces for 10 minutes.
5.0 Service
5.1 Components
The design of the Nidek Medical Nuvo 8 Oxygen Concentrator allows for easy access and removal of most components. This allows you to perform scheduled maintenance, repair, and replacement of parts with minimal time and effort. The inlet air filter and battery are conveniently located inside the small access panel in the cabinet back.
CAUTION:
For your safety, be sure to set the I/0 (ON/OFF) switch to the 0 (OFF) position and unplug the power cord before you service the Nuvo 8 Oxygen Concentrator.
NOTE:
Record all scheduled maintenance. (Refer to Section 4.1.5.)
Page 14 of 52 2010-2229 Rev - B
5.2 Cabinet Removal
5.2.1 Removing Cabinet Back
To remove the cabinet back, remove two screws near the base (one on each side) and two inset screws at the top just under the handle.
5.2.2 Removing Cabinet Front
To remove the cabinet front, remove two screws near the base (one on each side) adjacent to the rear cabinet screws and two inset screws at the top just under the handle. It is necessary to remove the cabinet back to access the top screws.
5.2.3 Removing Control Panel
Eight screws hold the control panel in place; the front cover must be removed to access these screws. It is necessary to remove the electrical wiring and pneumatic tubing to fully release the control panel.
5.2.4 Main Structure
The weight and forces of the internal components rest solely on four parts: the main structure, the compressor plate, the base-plate and the base. These parts were specially designed and formed. They should never require replacement under normal use.
5.2.5 Caster Replacement
The casters are a push in type that do not require any fasteners. Lay the device on its back to access the casters from the bottom. Pull them straight out away from the bottom.
5.3 Compressor
The compressor is the pump within the oxygen concentrator that supplies air to the separation process performed by the sieve beds. The pressure generated by the compressor forces oxygen to flow out of the top of the sieve columns.
The compressor is the likely cause of two potential specific problems:
a. An insufficient amount of air is supplied to the process. b. An excessive sound level.
● Air Supply
Compressor output refers to how much compressed air the compressor can produce. This depends upon the model of the compressor, length of stroke, piston diameter, speed of rotation and condition of seals. The cup seals form the seal between the piston and the cylinder wall. As the cup seals wear, the output begins
2010-2229 Rev - B Page 15 of 52
to gradually decrease. This reduction in compressor output results in less air, and thus less oxygen, entering the sieve beds. Therefore, the production of oxygen decreases.
Because this drop in oxygen production occurs over a long period of time, preventive maintenance on the compressor is not required.
You can continue a patient’s therapy on the Nuvo 8 unit as long as the oxygen concentration level at the prescribed liter flow rate is within Nidek Medical’s specification limits. Refer to Section 2.4.
_______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______ _______
Air Flow
Meter
0 to 100 l/min
Pressure
Restrictor
1 bar
Pressure
Gauge
Flow
Compressor
Air
Filter
● Sound Level
The sound level is largely determined by the condition of the compressor’s bearings.
There are four bearings located within the compressor that allow the inner components of the compressor to rotate. If the bearings wear to the point that they become loose and noisy, the compressor becomes noticeably loud and needs servicing. The life of a compressor is determined primarily by its operating temperature. It is extremely important that the inlet cooling air filters are cleaned and replaced as required.
Page 16 of 52 2010-2229 Rev - B
Loading...
+ 37 hidden pages