5.10.2 Circuit Breaker Installation23
I/0 (ON/OFF) Power Switch Replacement23
5.11.1 I/0 (ON/OFF) Power Switch Removal23
5.11.2 I/0 (ON/OFF) Power Switch Installation24
Buzzer Replacement24
Hour Meter Replacement24
Flow Meter Replacement24
5.14.1 Flow Meter Removal24
5.14.2 Flow Meter Installation24
Power Cord Replacement24
Section 6.0
Troubleshooting
Air Pressure Test (P1)25
6.1.1 High Air Pressure25
6.1.2 Low Air Pressure26
Product Pressure Test (P2)26
6.2.1 High Product Pressure26
6.2.2 Low Product Pressure27
General Troubleshooting27
Troubleshooting Chart28-32
Tool Kit and Pressure Test Gauge33
Page 4 of 592010-2225 Rev - B
Appendices
Exploded
Parts Listing
Maintenance
Drawings
A1: Base Assembly
A1.1: Base Assembly
A2: Main Structure Assembly – Front View36
A3: Main Structure Assembly – Rear View37
A4: Regulator Assembly38
A5: Compressor Assembly
A5.1: Compressor Assembly
A6: Control Panel Assembly; OCSI Unit - Front41
A7: Control Panel Assembly; OCSI Unit - Rear42
A6.1: Control Panel Assembly; OCSI Unit - Front43
A7.1: Control Panel Assembly; OCSI Unit - Rear44
A8: Control Panel Assembly; Non OCSI Unit Front
A9: Control Panel Assembly; Non OCSI Unit –
Rear
A10: Air Tank Assembly47
A10.1: Air Tank Assembly48
A11: Front Cabinet Assembly49
A11.1: Front Cabinet Assembly50
A12: Rear Cabinet Assembly51
A12: Rear Cabinet Assembly52
A13: Module Assembly53
34
35
39
40
45
46
Record
A14: Replacement Parts Listing54-57
A15: Maintenance Log58-59
2010-2225 Rev - BPage 5 of 59
General Safety Instructions
Production and use of oxygen
Oxygen is not a flammable gas, but accelerates the combustion of materials. To
prevent fire risks, the MARK5 NUVO should be kept away from flames,
incandescent sources or sources of heat (cigarettes) and combustible products
such as oil, grease, solvents, aerosols, etc.
Do not use in an explosive atmosphere.
Prevent oxygen from accumulating on upholstered seats or any other fabric. If the
concentrator operates without being administered to a patient, locate it so that the
flow of gas generated is dissipated into the air.
Locate the equipment in a free space (filter to the rear and below) which is well
ventilated and free of fumes or atmospheric pollution.
Use and Maintenance of the Device
Use the electric cable provided and check that the voltage of the mains socket used
complies with the electrical characteristics of the appliance indicated on the
manufacturers plate on the rear of the appliance.
Do not use an extension cord or multiple sockets which can create sparks and
therefore pose a fire risk.
Use of the MARK5 NUVO must be restricted solely to oxygen therapy on medical
prescription in compliance with the daily rate and duration.
Use in other circumstances may represent a hazard to patient health.
Do not use in a specifically magnetic environment (MRI, etc.).
The MARK5 Nuvo has an audible alarm intended to warn the user of any
problems. The user must determine the maximum distance away from the
Nuvobased on the on the sound levels in the environment, to ensure that the
alarm is always audible.
Standards & Regulations
In compliance with UL60601-1 [EN60601-1] (para 6.82.b):
“The manufacturer, assembler, installer or importer are not considered to be
responsible for consequences or the safety, reliability and characteristics of a
device unless,
● the assembly, extensions, adjustments modifications or repairs have been
performed by persons authorized by the manufacturer,
Page 6 of 592010-2225 Rev - B
● the electrical installation of the corresponding premises complies with appropriate
regulations and codes,
● the device is used in accordance with the instructions for its use.
If the replacement parts used for periodic servicing by an approved technician do
not comply with the manufacturer’s specifications, the manufacturer is absolved of
all liability in the event of an incident.
Do not open the equipment when it is powered on: risk of electrocution.
This device complies with the requirements of the FDA Quality System Regulation
and EU Directive 93/42/EEC, but its operation may be affected by the use in the
surrounding area of appliances such as diathermy, high frequency electro-surgical
instruments, defibrillators, short wave treatment appliances, cell-phones, CB
devices and other portables, microwave ovens, induction hot plates or remote
control toys, and more generally, by electromagnetic interference exceeding the
levels specified in standard IEC(EN) 60601-1-2.
1.0Introduction
1.1Home Service Provider Responsibility
All HomeServiceProviders oftheNidekMedical Mark5Nuvo®Oxygen
Concentrator must assume responsibilities for handling, operational check-out,
patient instruction, and maintenance. These responsibilities are outlined below and
throughout this manual.
WARNING
Mark5 Nuvo units must not be used for or with any life-supporting or life sustaining
applications.Patients unable to communicate discomfort while using this device
may require additional monitoring. Advise patients to immediately notify their Home
Service Provider(s) and/or physician(s) in case of an alarm or any discomfort.
As a Home Service Provider, you must do all of the following:
● Inspect the condition of each Mark5 Nuvo® unit immediately upon
delivery to your business location. Note any sign of damage, external or
internal, on the delivery receipt, and report it directly to both the freight
company and Nidek Medical Products, Inc. immediately.
● Check the operation of each Mark5 Nuvo® before delivery to a patient.
Always operate the unit for a reasonable length of time and check that the
oxygen concentration level is within specifications as referred to in
Section 2.4. Test the battery and power disconnect alarm as described in
Section 2.3 of this manual.
2010-2225 Rev - BPage 7 of 59
● Deliver Mark5 Nuvo® units only to patients authorized by a physician’s
hazard or unsafe practice that can result
Describes a hazard or unsafe practice that can result
Provides information important enough to emphasize
prescription. The Mark5 Nuvo® must not be used as a life-supporting or
life sustaining device. A backup supply of oxygen must be available.
● Instruct patients how to use the Mark5 Nuvo® in conjunction with the
Patient Manual.
● Instruct patients to notify their physicians and/or Home Service Providers
if they experience any signs of discomfort.
● Instruct each patient how to perform routine maintenance of the cabinet
air filter and how to check the alarm system battery. (Refer to Section
3.2.)
Be available to service each patient at any time. Maintain the Mark5
Nuvo® in accordance with Section 4.0.
Repair components and replace parts only as outlined in this manual. Use only
Nidek Medical parts for replacement in Mark5 Nuvo® Oxygen Concentrators.
● Refer to the Mark5 Nuvo® Product Warranty if parts replacement is
required within the warranty period.
1.2Important Notice and Symbol Explanations
As you read the manual, pay special attention to the WARNING, CAUTION, and
NOTE messages. They identify safety guidelines or other important information as
follows:
WARNING:
CAUTION:
NOTE:
The following harmonized symbols (pictograms), used for non-English language
countries, will be located on the exterior of the Mark5 Nuvo® unit:
Store, ship and use the device in an upright condition.
Describes a
in severe bodily injury or death.
in minor bodily injury or property damage.
or repeat.
No smoking within five feet of this device, oxygen-carrying tubing, or
accessories.
Indicates an alarm signal.
Page 8 of 592010-2225 Rev - B
Do not use any oil or grease on or near the device
1.3Functional Specifications
Dimensions: 15.5 in. long, 15.6 in. wide, 27.8 in. tall
(394 mm long, 396 mm wide, 706 mm high)
Weight:50 to 54 Ib (22.5 to 24.5 kg) depending on sound attenuation;
Electrical120 VAC, 60 Hz, 4 amps, 410 watts
Requirements:220 VAC, 50 Hz, 2 amps, 420 watts
Capacity:5 liters per minute at 90% oxygen
(Based on 70°F [21°C] at sea level)
Accuracy:Flow meter ±5% full scale. (Based on pre-calibrated analysis.)
Concentration:1 - 3 liters per minute at 95% + 0.5 / - 3%
4 liters per minute at 92% + 3.5 / - 3%
5 liters per minute at 90% + 5.5 / - 3%
Response Time:Acceptable concentration is normally achieved in about
90 seconds; allow 5 minutes to attain full concentration.
Positioning:Operate the unit in an upright position, maintaining at
least six inches of open space on all sides for ventilation.
2.0Operational Check and Concentration Test
2.1Description of Operation
Air enters the Mark5 Nuvo® Oxygen Concentrator through an external cooling air
intake filter. This filtered air enters the compressor via a suction resonator and fine
filter, which quiets the suction sounds made by the compressor. Pressurized air
then exits the compressor and passes through a heat exchanger into an air tank.
The heat exchanger reduces the temperature of the compressed air and the air tank
stores air when feed is interrupted. Next, a rotary poppet feed valve directs the air
into one of two sieve beds that contain molecular sieve. The special characteristic
property of molecular sieve is that it physically attracts (adsorbs) nitrogen when air
passes through this material, thus enabling the production of high purity oxygen.
There are two sieve beds or adsorbent columns; while one produces high purity
oxygen, the other is purged of the nitrogen it adsorbed (collected) while it was
producing oxygen. Each column produces oxygen for approximately four seconds
and delivers it to the product storage volume in the end of the column. Oxygen exits
the adsorbent column through a pressure regulator, flow control valve, flow meter,
and final product filter. The flow control valve, which is part of the flow meter,
controls the flow rate of oxygen delivered to the patient. The Mark5 Nuvo® unit
2010-2225 Rev - BPage 9 of 59
delivers up to 95% oxygen concentration at flow rates from 1 to 5 l/min.The
remaining constituents of the product gas stream are nitrogen and argon, both of
which are part of the air we breathe, are inert and are completely safe.
2.2Operational Check
Nidek Medical runs each device through a burn in period and tests every Mark5
Nuvo® Oxygen Concentrator thoroughly after manufacture before releasing for
shipment. As the home service provider, it is your responsibility to perform the
following test to ensure that no damage occurred in shipping or handling.
1.Open and inspect all concentrator cartons upon receipt. Unpack each
unit and remove it from its carton. Inspect the unit itself for damage. If
the exterior of the carton is damaged, or the unit itself is damaged,
note it on the freight bill signed by the driver.
● Before plugging the unit into the wall outlet, actuate the I/0
(ON/OFF) switch to the I (ON) position and note that the audible
alarm sounds continuously. (See Section 2.3). Move the switch to
the 0 (OFF) position.
2.Plug in the power cord of the unit, and set the I/0 (ON/OFF) switch to
the I (ON) position. Check to see that the following occurs:
● The compressor runs, listen for the sound.
● Exhaust air flows out of the bottom of the unit.
● OPTIONAL for Units Equipped with Oxygen Concentration Status
Indicator (OCSI): The OCSI green light remains off until the oxygen
concentration reaches 85% ± 3% (82% ± 2% for 50 Hz units)
(approximately two minutes).
3.Turn the flow meter adjustment knob counterclockwise until it stops
(wide open). The flow meter should register in excess of 5.5 liters/min.
If not, refer to Section 5.8 to adjust the product regulator.
4Perform an oxygen concentration test, as described in Section 2.4.
2.3Alarm System
The Mark5 Nuvo® Oxygen Concentrator is equipped with a battery-powered alarm
system, which sounds a continuous and loud alarm when a power failure occurs or
when one or more cycle variables are not within specification. It sounds an alarm if
the high or low pressure indicators are activated or if the optional OCSI detects
lower than predetermined levels of oxygen concentration. The alarm remains on
until you correct the alarm condition or you set the I/0 (ON/OFF) switch to the 0
(OFF) position. Refer to Section 6.0 for a list of probable alarm causes,
Page 10 of 592010-2225 Rev - B
2.3.1 Battery Test
The Mark5 Nuvo® battery can be checked by observing that the continuous alarm
sounds when the I/0 (ON/OFF) switch is moved to the I (ON) position when the unit
is unplugged from the wall outlet.
2.3.2 Power Failure Alarm Test
To test the power failure alarm, perform the following actions:
Unplug the power cord from the wall outlet, and set the I/0 (ON/OFF) switch to the I
(ON) position.
This should immediately activate the audible alarm. If it does not, refer to the
troubleshooting chart in Section 6.0 of this manual.
2.4Oxygen Concentration Test and Specification
To ensure that the output of oxygen from the device is within specification, you must
perform an oxygen concentration test. Test the unit upon delivery to a patient and at
periodicintervals.HomeServiceProviders,basedontheirexpertiseand
documentation, may establish and implement their own plans for checking oxygen
concentration. The interval established may be longer or shorter than 90 days,
which is Nidek Medical’s default time period for providers who do not choose to
establish their own plan.
1.If an oxygen humidifier bottle is used, remove it from the oxygen outlet.
2.Connect a calibrated oxygen concentration analyzer to the oxygen outlet.
3.Set the I/0 (ON/OFF) power switch to the I (ON) position. (It takes
approximately five minutes for the oxygen concentration to stabilize.) Take oxygen
concentration readings over a period of several minutes to reduce any cyclic
variations
4.Verify that the product flow rate delivered by the unit matches the patient’s
prescription and does not exceed the capacity of the unit.
5.Disconnect the oxygen analyzer, and reconnect the humidifier bottle (if used)
and any other equipment / accessories that may be required.
6.Adjust the flow meter adjustment knob to the prescribed flow rate.
NOTE
Do not measure oxygen concentration output after the product stream passes
through a humidifier bottle, erroneous readings will result and your oxygen
concentration measuring device might be damaged.
2010-2225 Rev - BPage 11 of 59
Nidek Medical Mark5 Nuvo® Concentration Specifications
air filter from the back of the
with household
Reinstall the clean air filter on the grille in the back of
Liter FlowSpecificationWithin Specification
1 to 3 l/min95% + 0.5 / - 3%92% or higher
4 l/min92% + 3.5 / - 3%89% or higher
5 l/min90% + 5.5 / - 3%87% or higher
3.0Patient Instructions
3.1General Instructions
It is important that patients thoroughly understand how to operate the Nidek Medical
Mark5 Nuvo® unit. This enables proper treatment as prescribed by a qualified,
licensed physician. You must explain that the purpose of this therapy is to alleviate
symptoms. If patients experience any discomfort or the unit alarms, they must notify
their Home Service Provider and/or physician immediately.You, as the Home
Service Provider, are responsible to see that each patient receives the Patient
Manual. Explain each step in the operation of the unit to the patient in reference to
this manual.
3.2Routine Maintenance by the Patient
To ensure accurate output and efficient operation of the unit, the patient must
perform two simple routine maintenance tasks:
•Clean the cabinet air filter
•Check the alarm system battery
3.2.1 Cleaning the Cabinet Air Filter
NOTE:
The patient must clean this filter weekly, as described below. The filter
may require daily cleaning if the Mark5 Nuvo® unit operates in a harsh
environment such as a house heated by wood, kerosene, or oil, or one
with excessive cigarette smoke.
1Remove the dirty cabinet
Mark5 Nuvo® unit.
2Wash the dirty filter in warm water
detergent, and rinse.
3Use a soft absorbent towel to remove excess water.
4
the unit.
Page 12 of 592010-2225 Rev - B
3.2.2 Checking the Alarm System Battery
See Procedure described in Paragraph 2.3.1
4.0Home Service Provider Maintenance
4.1Routine Maintenance
The Mark5 Nuvo® unit has two filters and a 9-volt battery that require scheduled
maintenance and replacement.
To ensure that the output of oxygen from the unit is within specification, you must
perform an oxygen concentration test. Test the unit upon delivery to a patient and at
periodicintervals.HomeServiceProviders,basedontheirexpertiseand
documentation, may establish and implement their own practices for checking
oxygen concentration. Consult Nidek Medical’s Service and Maintenance Log (A-
15) for the recommended intervals for testing.
Nidek Medical does not require preventive maintenance on the concentrator. You
do not need to perform any maintenance as long as the Mark5 Nuvo® unit remains
withinspecificationsattheprescribedflowrate.(Refer toSection2.4)
Cabinet Air Filter
The external cabinet air filter is located on the back of the unit, You can easily
remove it by hand. Instruct the patient to clean this filter weekly. (Refer to Section
3.2.1.)
NOTE
The filter may require more frequent cleaning if the Mark5 Nuvo® unit operates in a
harsh environment such as a house heated by wood, kerosene, or oil, or one with
excessive cooking, cigarette smoke or atmospheric dust.
4.1.2 Final Product Filter Replacement
The final product filter does not require periodic replacement; it needs to be
replaced only if it restricts oxygen flow. It is suggested that it be replaced whenever
the sieve module is repaired or replaced and after the compressor is rebuilt.
1.Set the I/0 (ON/OFF) switch to the 0 (OFF) position, and unplug the power
cord.
2.Remove the cabinet back to locate the final product filter.
NOTE: Observe the position of the filter before removal.
3.Separate the silicone tubing from both sides of the filter.
4.Install the new filter with the inlet side in the same position as before. Push
the tubing together so that it overlaps the barbs of the final product filter
2010-2225 Rev - BPage 13 of 59
connections.
5.Record information about the final product filter replacement in Appendix 15
of this manual and online at www.nidekmedical.com under the ‘Maintenance
Log’ tab.
6.Reinstall the cabinet back.
4.1.3 Inlet Air Filter Replacement
The inlet air filter requires inspection at each patient visit. The filter should be
replaced annually, or more often depending on environment.
1.Set the unit I/0 (ON/OFF) switch to the 0 (OFF) position, and unplug the
power cord.
2.Remove the rear cabinet access panel to locate the inlet air filter.
3.Remove filter from the unit, and replace with a new filter.
4.Record information about the filter replacement in Appendix 15 of this
manual and online at www.nidekmedical.com under the ‘Maintenance Log’
tab.
5.Reinstall the access panel.
4.1.4 Battery Replacement
Each time the standard Mark5 Nuvo® unit is turned on, the alarm must sound loudly
for approximately five seconds to indicate a good battery. An alarm that does
anything other than sound loudly for five seconds indicates a weak battery and
requires replacement. For units with OCSI, actuate the switch with the power cord
unplugged to test the battery.
To replace the battery, take the following steps:
1.Set the I/0 (ON/OFF) switch to the 0 (OFF) position, and unplug the power
cord.
2.Remove the access panel in the cabinet back.
3.Pull the battery from the mounting slot and remove the connection wires.
4.Attach the new battery to the connector and reinsert it into the mounting slot.
5.Set the I/0 (ON/OFF) switch to the I (ON) position to test the alarm.
6.Record the battery replacement information in Appendix 15 of this manual
and online at www.nidekmedical.com under the ‘Maintenance Log’ tab.
7.Reinstall the access panel.
4.1.5 Wick Replacement
The wick is installed in the process air stream of the unit, connected to the air tank,
to assist with removal of moisture from the air.This helps prevent premature
contamination of the sieve beds, particularly in humid climates. The wick should be
replaced at each compressor service / module replacement; it is suggested that
more frequent replacement be done in very humid climates. The wick is located in
a bushing located just to the right of the gearmotor. It is installed in a quick release
fitting. Pull the fitting from the bushing by hand – it may require several pounds of
force, remove the wick and install a new one in its place. Re-insert the wick end
and fitting into the bushing so the wick is totally below the base plate surface.
Page 14 of 592010-2225 Rev - B
4.1.6 Recording Maintenance
As the Home Service Provider, it is suggested that you record all routine
maintenance and repairs performed on the Mark5 Nuvo® unit, including hours and
dates of service in Appendix 15 of this manual and online at www.nidekmedical.com
under the ‘Maintenance Log’ tab.
4.2Cleaning and Infection Control
With the growing concern about possible cross infection from home oxygen
equipment (i.e.. oxygen concentrators) from one home care patient to another,
appropriate cautions should be exercised.
The organisms of most concern are M. Tuberculosis, HIV, and Viral Hepatitis.
These are potentially pathogenic.
Tuberculosis can survive outside of the human body, but its mode of transmission is
by droplet nuclei. When infected individuals cough, they release droplet nuclei into
the air, and these carry the Tuberculosis organism. Another person may breathe in
these droplet nuclei, but prolonged exposure to the infected person is usually
necessary for infection to occur.
HIV and Viral Hepatitis are both viruses, which are not living cells themselves but
which can duplicate when in a living “host” cell. Both of these organisms are usually
passed on by person-to-person contact, and both need to be in the human body to
survive. Once outside the body, viruses can survive for only a short period of time.
4.2.1 Preparing for New Patient Use
When you remove the Mark5 Nuvo® from a patient’s home, always dispose of the
used nasal cannula and humidifier bottle. Clean the exterior of the Mark5 Nuvo®
with a soapy water solution or commercial detergent cleaner to remove any gross
debris, organic or otherwise. Be careful not to get any liquid into the interior of the
unit.
Next, clean the exterior with either a common chemical disinfectant or a bleach
solution* and allow it to air dry.For the bleach solution, wear eye and skin
protection to prevent exposure to the chlorine. Retest the Mark5 Nuvo® before you
return it to your inventory.
Replace the cabinet air filter between each patient’s use or clean with warm soapy
water if it is in good condition. Clean this filter at least once per week or more
frequently if operated in a dusty environment.
*Make the bleach solution a 1:100 dilution of 5.25% sodium hypochlorite. Mix one
part household bleach (e.g. Clorox) with 99 parts cold tap water. To measure the
2010-2225 Rev - BPage 15 of 59
solution easily, take 1/4 cup of household bleach, and mix it with a gallon of cold tap
water. Allow the mixture to sit on potentially contaminated surfaces for 10 minutes.
5.0 Service
5.1Components
The design of the Nidek Medical Mark5 Nuvo® Oxygen Concentrator allows for
easy access and removal of most components. This allows you to perform
scheduled maintenance, repair, and replacement of parts with minimal time and
effort.The inlet air filter and battery are conveniently located inside the small
access panel in the cabinet back.
CAUTION:
For your safety, be sure to set the I/0 (ON/OFF) switch to the 0 (OFF)
position and unplug the power cord before you service the Mark5 Nuvo®
Oxygen Concentrator.
NOTE: Record all scheduled maintenance. (Refer to Section 4.1.6.)
5.2Cabinet Removal
5.2.1 Removing Cabinet Back
To remove the cabinet back, remove two screws near the base (one on each side)
and two inset screws at the top just under the handle.
5.2.2 Removing Cabinet Front
To remove the cabinet front, remove two screws near the base (one on each side)
adjacent to the rear cabinet screws and two inset screws at the top just under the
handle. It is necessary to remove the cabinet back to access the top screws.
5.2.3 Removing Control Panel
Eight screws hold the control panel in place; the front cover must be removed to
access these screws. It is necessary to remove the electrical wiring and pneumatic
tubing to fully release the control panel.
5.2.4 Main Structure
The weight and forces of the internal components rest solely on four parts: the main
structure, the compressor plate, the base-plate and the base. These parts were
specially designed and formed. They should never require replacement under
normal use.
Page 16 of 592010-2225 Rev - B
5.2.5 Caster Replacement
The casters are a push in type that does not require any fastener. Lay the device
on its back to access the casters from the bottom. Pull them straight out away from
the bottom.
5.3Compressor
The compressor is the pump within the oxygen concentrator that supplies air to the
separation process performed by the sieve beds. The pressure generated by the
compressor forces oxygen to flow out of the top of the sieve columns.
The compressor is the likely cause of two potential specific problems:
a. An insufficient amount of air is supplied to the process, and
b. An excessive sound level.
● Air Supply
Compressor output refers to how much compressed air the compressor can
produce. This depends upon the model of the compressor, length of stroke, piston
diameter, speed of rotation and condition of seals. The cup seals form the seal
between the piston and the cylinder wall. As the cup seals wear, the output begins
to gradually decrease. This reduction in compressor output results in less air, and
thus less oxygen, entering the sieve beds. Therefore, the production of oxygen
decreases.
Because this drop in oxygen production occurs over a long period of time,
preventive maintenance on the compressor is not required.
You can continue a patient’s therapy on the Mark5 Nuvo® unit as long as the
oxygen concentration level at the prescribed liter flow rate is within Nidek Medical’s
specification limits. Refer to Section 2.4.
The sound level is largely determined by the condition of the compressor’s
bearings.
There are four bearings located within the compressor that allow the inner
components of the compressor to rotate. If the bearings wear to the point that they
become loose and noisy, the compressor becomes noticeably loud and needs
servicing. The life of a compressor is determined primarily by its operating
temperature. It is extremely important that the cabinet air filters are cleaned and
replaced as required.
5.3.1 Compressor Replacement
Remove Compressor Assembly
To remove the compressor assembly for exchange, follow the steps listed below:
1.Set the unit’s I/0 (ON/OFF) switch to the 0 (OFF) position, and unplug the
power cord.
2.Remove the cabinet back and front.
3.Disconnect the suction tube.
4.Disconnect the two compressor power cable leads and the two leads to the
capacitor at the inline connector.
5.Remove the three screws that connect the compressor plate to the base of
the unit, lift and slide out the compressor assembly.
6.Remove compressor from the plate by removing the four compressor bolts.
7.Remove tie wrap from compressor plate wiring harness.
8.Remove the heat exchanger tubing and fittings.
Page 18 of 592010-2225 Rev - B
Loading...
+ 41 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.