Nidek CV-24000 Service manual

NIDEK
OPHTHALMIC SURGICAL SYSTEM
[PHACOEMULSIFICATION • VITREOUS SURGERY]
Model CV-24000
SERVICE MANUAL
December 21, 2001
XCV4**RDA001B/E Total 142 pages
§
1 CAUTIONS .................................................................................................................. 1-1
1.1 Cautions in using the manual ................................................................................... 1-1
1.2 Cautions in repairing ................................................................................................ 1-1
§
2 TROUBLESHOOTING .............................................................................................. 2-1
§
3 SUB-TROUBLESHOOTING ..................................................................................... 3-1
3.1 Pilot lamp does not illuminate.................................................................................. 3-1
3.1.1 Check of P.S. board ......................................................................................... 3-2
3.2 Nothing appears on screen ....................................................................................... 3-3
3.2.1 Communication error of screen ....................................................................... 3-4
3.3 Back light can not be adjusted.................................................................................. 3-4
3.4 Touch panel does not work ....................................................................................... 3-5
3.5 Seat switch does not work ........................................................................................ 3-6
3.6 Remote control does not work ................................................................................. 3-7
3.7 Sounds do not come out ........................................................................................... 3-8
3.8 Volume can not be adjusted ...................................................................................... 3-8
3.9 Irrigation pole does not move by irrigation UP/DOWN switch ............................... 3-9
3.10 Printer does not work normally ............................................................................ 3-11
3.11 Memory inside of main body does not work correctly ......................................... 3-12
3.12 Memory card does not work correctly .................................................................. 3-12
3.13 Cassette test is not passed .................................................................................... 3-13
3.14 Cassette can not be removed ................................................................................ 3-14
3.15 SYSTEM (cataract) test is not passed .................................................................. 3-16
3.15.1 SINGLE CASSETTE .................................................................................. 3-16
3.15.2 ANT DUAL CASSETTE ............................................................................ 3-20
3.16 US test is not passed ............................................................................................. 3-24
3.17 DIA test is not passed ........................................................................................... 3-24
3.18 SYSTEM (vitrectomy) test is not passed ............................................................. 3-25
3.18.1 ANT/POST DUAL CASSETTE ................................................................. 3-25
3.18.2 POST DUAL CASSETTE .......................................................................... 3-29
3.19 PPL test is not passed ........................................................................................... 3-32
3.20 Foot switch does not work ................................................................................... 3-33
3.21 Irrigation does not move correctly ....................................................................... 3-35
3.22 Aspiration does not work correctly ...................................................................... 3-36
3.23 US oscillation is not performed correctly............................................................. 3-38
3.24 Diathermy output is abnormal .............................................................................. 3-39
3.25 Cutter does not work (for using built-in compressor) .......................................... 3-40
3.25.1 Check of air pressure piping for cutter (only for ANTERIOR) .................. 3-41
3.25.2 Check of air pressure piping for cutter (for both of ANT/POST) ............... 3-42
Page
3.26 Cutter does not work (in the case of using external air pressure) ........................ 3-44
3.27 Scissors do not work correctly ............................................................................. 3-46
3.27.1 Check of air pressure piping for using built-in compressor ........................ 3-46
3.27.2 Check of air pressure piping for using external air pressure ....................... 3-48
3.28 FGX does not work correctly ............................................................................... 3-50
3.29 Illumination does not work correctly ................................................................... 3-52
3.30 Check of power (P.S. board)................................................................................. 3-54
§
4 REPLACEMENT ......................................................................................................... 4-1
4.1 Replacement of top cover ......................................................................................... 4-1
4.2 Replacement of rear cover ........................................................................................ 4-1
4.3 Replacement of lower part of left cover ................................................................... 4-2
4.4 Replacement of lower part of right cover ................................................................. 4-2
4.5 Replacement of front cover ...................................................................................... 4-3
4.6 Replacement of inverter ........................................................................................... 4-3
4.7 Replacement of display board .................................................................................. 4-4
4.8 Replacement of main board...................................................................................... 4-4
4.9 Replacement of limit switch ..................................................................................... 4-5
4.10 Replacement of driving (AC) motor ...................................................................... 4-6
4.11 Replacement of control unit ................................................................................... 4-7
4.12 Replacement of P.S. board...................................................................................... 4-8
4.13 Replacement of US unit ......................................................................................... 4-9
4.14 Replacement of compressor ................................................................................. 4-10
4.15 Replacement of printer ......................................................................................... 4-10
4.16 Replacement of noise filter .................................................................................. 4-11
4.17 Replacement of power switch .............................................................................. 4-11
4.18 Replacement of transformer ................................................................................. 4-12
4.19 Replacement of BU CONNECT board (BA06) ................................................... 4-12
4.20 Replacement of VR board (BA21) ....................................................................... 4-13
4.21 Replacement of RECEIVER board (BA02) ......................................................... 4-13
4.22 Replacement of speaker ....................................................................................... 4-14
4.23 Replacement of irrigation pole UP/DOWN switch (EA56) ................................. 4-15
4.24 Replacement of PRINT/FEED switch (BA29)..................................................... 4-15
4.25 Replacement of POLE CONTROL board (BA07) ............................................... 4-16
4.26 Replacement of microswitch (SW56, 58) ............................................................ 4-16
4.27 Replacement of EJECT switch board (BA29) ..................................................... 4-17
4.28 Replacement of TUBE SENS board (BA28) ....................................................... 4-17
4.29 Replacement of DRIVER board (BA05) .............................................................. 4-18
4.30 Replacement of CASSETTE SENS board (BA27) .............................................. 4-18
4.31 Replacement of FGX/SCIS board (BA08) ........................................................... 4-19
4.32 Replacement of ILL/SCIS SELECT board (BA09) ............................................. 4-19
4.33 Replacement of solenoid valve ............................................................................ 4-20
4.34 Replacement of pressure sensor for control (EA86) ............................................ 4-21
4.35 Replacement of pressure sensor (EA84) .............................................................. 4-22
4.36 Replacement of condenser (C1) for AC motor (driving motor) ........................... 4-22
4.37 Replacement of VACUUM SENS board (BA26) ................................................ 4-23
4.38 Replacement of CARD board (BA24) ................................................................. 4-23
4.39 Replacement of valve unit (1 set and 2 sets) ........................................................ 4-24
4.40 Replacement of pressure sensor for FGX output (EA88) .................................... 4-24
4.41 Replacement of seat switch .................................................................................. 4-25
4.42 Replacement of load cell unit (EA31) .................................................................. 4-25
4.43 Replacement of aspiration motor (EA39) ............................................................ 4-26
4.44 Replacement of pilot lamp ................................................................................... 4-27
§
5 ADJUSTMENT ............................................................................................................ 5-1
5.1 Transformer output (check at secondary side) ......................................................... 5-1
5.2 Output voltage check of P.S. board (BA04) ............................................................. 5-1
5.3 Output voltage check of BU CONNECT board (BA06) .......................................... 5-1
5.4 Output voltage check of MAIN board (BA03)......................................................... 5-2
5.5 Output voltage check of DRIVER board (BA05) .................................................... 5-2
5.6 Output voltage check of FGX/SCIS board (BA08) .................................................. 5-2
5.7 Output voltage check of US unit (E100) .................................................................. 5-2
5.8 Voltage adjustment of FGX/SCIS board (BA08) ..................................................... 5-3
5.9 Output voltage check of DISPLAY board (BA02) ................................................... 5-3
5.10 Adjustment of irrigation height detection potentiometer ....................................... 5-4
5.11 Adjustment of control unit ..................................................................................... 5-4
5.12 Adjustment of FGX/SCIS mode ............................................................................ 5-5
5.13 Adjustment of pinch valve ..................................................................................... 5-6
5.14 Adjustment of load cell positioning ....................................................................... 5-6
5.15 Adjustment of aspiration motor (EA39) positioning.............................................. 5-7
§
6 WIRING DIAGRAM ................................................................................................... 6-1
6.1 Block diagram .......................................................................................................... 6-1
6.2 Display unit diagram ................................................................................................ 6-2
6.3 Main unit diagram .................................................................................................... 6-3
6.4 P.S. unit diagram....................................................................................................... 6-4
6.5 Driver unit diagram .................................................................................................. 6-5
6.6 BU CONNECT unit diagram ................................................................................... 6-6
6.7 POLE CONTROL unit diagram ............................................................................... 6-7
6.8 FGX/SCIS unit diagram ........................................................................................... 6-8
6.9 ILL/FGX SELECT unit diagram .............................................................................. 6-9
§
7 TOOLS AND SOLVENTS .......................................................................................... 7-1
7.1 Tools ......................................................................................................................... 7-1
7.2 Solvents .................................................................................................................... 7-1
§
8 ERROR CODE ............................................................................................................. 8-1
§
1
CAUTIONS
1.1 Cautions in using the manual
• This manual is for the service work of the Phacoemulsification System, Model CV-24000.
• Before using this manual, thoroughly read and understand the operator’s manual for CV-24000.
• This manual does not contain the contents written in the operator’s manual.
• The specifications are subject to change without notice for improvement. For the important change, see the technical bulletin issued for the change concurrently.
1.2 Cautions in repairing
• For repairing, make sure to wear rubber gloves to protect from the infection of bacteria.
• Never touch the things which are used for surgery and not yet sterilized.
• When checking the motion, ensure to use the sterilized handpiece.
• Repairing the system is prohibited except the personnel who received the training.
• For maintenance, turn OFF the power and disconnect the socket from the power unless this manual specifies “Turn ON the power”.
• Prepare a case for removed screws and parts not to miss or drop them into the system.
• For loosening and tightening threadlocker-applied screws, apply the threadlocker again.
§
2
Turn ON the power.
TROUBLESHOOTING
Is an error indicated?
YES
Does the pilot lamp illuminate?
YES
Does an operation screen appear?
YES
Can the back light be adjusted by the knob on the rear panel?
YES
Does the touch panel work?
NO
Check the indicated error.
NO
3.1 Pilot lamp does not illuminate
NO
3.2 Nothing appears on screen
NO
3.3 Back light can not be adjusted
NO
3.4 Touch panel does not work
YES
Does the set switch work?
YES
Does the remote control work?
YES
Do sounds come out of speaker?
YES
NO
3.5 Seat switch does not work
NO
3.6 Remote control does not work
NO
3.7 Sounds do not come out
2 - 2
Can the volume of speaker be adjusted by the volume control on the rear panel?
YES
Does the irrigation pole move by irrigation pole UP/DOWN switch on the rear panel?
YES
Does the printer work normally?
YES
Is data stored or read to the main body correctly?
YES
Is data stored or read to memory card correctly?
NO
3.8 Volume can not be adjusted
NO
3.9 Irrigation pole does not move by irrigation pole UP/DOWN switch
NO
3.10 Printer does not work normally
NO
3.11 Memory inside of main body does not work correctly
NO
3.12 Memory card does not work correctly
YES
Does the system start a cassette test and pass it by inserting cassette?
YES
Can the cassette be removed by EJECT switch while the cassette is inserted?
YES
Start the SYSTEM (cataract) test. Is it passed?
YES
Start the US test. Is it passed?
YES
NO
3.13 Cassette test is not passed
NO
3.14 Cassette can not be removed
NO
3.15 SYSTEM (cataract) test is not passed
NO
3.16 US test is not passed
Start the DIA test. Is it passed?
YES
NO
3.17 DIA test is not passed
2 - 3
Does the foot switch work?
YES
Does the irrigation work correctly?
YES
Does the aspiration work correctly?
YES
Is US oscillation performed correctly?
YES
Is the diathermy output correctly?
NO
3.20 Foot switch does not work
NO
3.21 Irrigation does not move correctly
NO
3.22 Aspiration does not work correctly
NO
3.23 US oscillation is not performed correctly
NO
3.24 Diathermy output is abnormal
YES
Does the cutter (ANT) work correctly?
YES
Is it a surgical instrument for cataract and vitrectomy?
Surgical instrument for cataract and vitrectomy
YES
Select vitrectomy surgery screen.
Start the SYSTEM (vitrectmy) test. Is it passed?
YES
NO
3.25 Cutter (ANT) does not work
NO
Is the power correct?
YES NO
End
NO
3.18 SYSTEM (vitrectomy) test is not passed
3.30 Check of power (P.S.board)
2 - 4
Start the PPL test. Is it passed?
YES
Start the DIA test. Is it passed?
YES
Does the cutter work correctly?
YES
Do scissors work correctly?
YES
Does FGX work correctly?
NO
3.19 PPL test is passed
NO
3.17 DIA test is passed
NO
3.26 Cutter (POST) does not work
NO
3.27 Scissors do not work correctly
NO
3.28 FGX does not work correctly
YES
Does the illumination work correctly?
YES
Is the power correct?
YES
End
NO
3.29 Illumination does not work correctly
NO
3.30 Check of the power (P.S. board)
§
3
SUB-TROUBLESHOOTING
3.1 Pilot lamp does not illuminate
Is power voltage proper? Voltage 100V/115V/230
YES
Is the power code connected correctly?
YES
Is the fuse blown? Disconnect the power cord from a wall outlet for checking the fuse.
NO
Are voltages between the power input terminals of the noise filter and the power switch (turned OFF) same as the power voltage?
YES
Turn ON the power switch. Is the power voltage supplied to the transformer? (Check by TB1, a terminal block.)
NO
NO
YES
NO
NO
Use a proper power voltage. Usable input voltage is written on the rating plate.
Connect the power cord.
Repalce the fuse with the new one. Disconnect the power cord from a wall outlet for checking the fuse.
Failure of the noise filter is a probable cause. Replace it with the new one. (See 4.16.)
Replace the power switch with the new one. (See 4.17.)
YES
Is the secondary side voltage of transformer normal? Check the secondary side voltage with no load.
Between YEL and YEL : AC33V±10% Between ORN and ORN : AC33V±10% Between BLK and BLU : AC100V±10% Between GRY and BLU : AC66V±10% Between VIO and BLU : AC33V±10%
YES
Is the voltage on the P.S. board (BA04) normal? See 3.1.1 Check of P.S. board.
YES
Disconnect the P618 from the BU CONNECT board (BA06). Is the voltage between 1st and 2nd pins of J618 +5V?
YES
NO
Replace the transformer with the new one. (See 4.18.)
NO
Replace the P.S. board (BA04) with the new one. (See 4.12.)
NO
Failure of the BU CONNECT board (BA06). Replace it with the new one. (See 4.19.)
Failure of the pilot lamp. Replace it with the new one. (See 4.44.)
3.1.1 Check of P.S. board
Is the input voltage to the board normal? Disconnect terminals from the terminal block (TB401) on the board and check voltages. Between YEL and YEL : AC35V±10% Between ORN and ORN : AC35V±10%
YES
Connect terminals disconnected from the terminal block (TB401). Disconnect all connectors.
Is the input voltage to the board normal? Check voltages of terminal block (TB401) on the board. Between 1 and 2 : AC35V±10% Between 3 and 4 : AC35V±10%
Check the input voltage of transformer. If it is
NO
normal, replace the transformer with the new one. (See 4.17.)
Failure of the P.S. board (BA04) is a probable
NO
cause. Replace the board with the new one. (See 4.12.)
3 - 2
YES
Check voltages between GND and each TP while connecting only TB401. Is voltages on the P.S. board (BA04) normal? TP2 +12V (+11.5 +12.5V TP3 +28V (+27.5 +29.0V TP4 -28V (-27.5 -29.0V TP6 +12V (+11.9 +12.1V TP7 -12V (-11.9 -12.1V TP8 +6V (+5.0 +6.0V TP9 +5V (+4.8 +5.2V TP10 +24V (+22.0 +26.0V
YES
Connect all removed connectors and check voltages between GND and each TP. Is voltages on the P.S. board (BA04) normal? TP2 +12V (+11.5 +12.5V TP3 +28V (+27.5 +29.0V TP4 -28V (-27.5 -29.0V TP6 +12V (+11.9 +12.1V TP7 -12V (-11.9 -12.1V TP8 +6V (+5.0 +6.0V TP9 +5V (+4.8 +5.2V TP10 +24V (+22.0 +26.0V
NO
Failure of the P.S. board (BA04) is a probable cause. Replace it with the new one. (See 4.12.)
Not only the failure of P.S. board (BA04) but
NO
other failures of board or unit are probable causes. Check other symptoms.
YES
The P.S. board (BA04) is normal.
3.2 Nothing appears on screen
3 - 3
Does the irrigation pole move up and down by the irrigation pole UP/DOWN switch on the rear
NO
3.2.1 Communication error of screen
panel?
YES
YES
Can the error beep be heard? 3.2.1 Communication error of screen
NO
Does the back light illuminate?
YES
The screen becomes black if the back light does not illuminate.
NO
Disconnect the connector (P201) from the DISPLAY board and check the voltage. Is the voltage normal?
Between 1 and 2 (0V) of P201 : +5V Between 3 and 4 (0V) of P201 : +12V
YES
NO
Failure of the power is a probable cause. Check the cable and the P.S. board. (See 3.1.1 Check of P.S. board.)
Are J205 (DISPLAY board) and CN1 (inverter) connected by a cable?
YES
Are CN2 and 3 of inverter connected to the LCD back light?
YES
Failure of the LCD back light or the inverter. Replace either of them with the new one. (See 4.6.)
Is the cable between the LCD and the display board connected?
YES
Failure of the LCD or the display board are probable causes. Replace either of them with the new one. (See 4.7.)
NO
Connect them.
NO
Connect them.
NO
Connect them.
3.2.1 Communication error of the screen
Turn OFF the dip switch (SW1) on the MAIN board and turn ON the power again.
3 - 4
Is an error indicated? Check the indicated error.
NO
Is the flat cable connected to the main body correctly?
MAIN board (P304) DRIVER board (P502) DRIVER board (P503) DRIVER board (P202)
YES
Failure of the MAIN board or the DISPLAY board are probable causes. Replace either of them with the new one. (See 4.7 and 4.8.)
YES
NO
Connect the flat cable correctly.
3.3 Back light can not be adjusted
Does the brightness of back light change by turning the back light control knob on the rear panel?
YES
Normal
NO
Does the voltage of TP7 (BACKLIGHT) change by turning the back light control knob?
YES
Does the voltage of TP9 change by turning the back light control knob?
YES
Failure of the inverter or the LCD back light are probable causes. Replace either of them with the new one. (See 4.6.)
Failure of cables among back light control
NO
volume, the VR board and the DISPLAY board. Replace the VR board or the cable with the new one.
NO
Failure of the DISPLAY board. Replace it with the new one. (See 4.7.)
3.4 Touch panel does not work
3 - 5
Does the touch panel work normally? Is there no torn, worn or dented part on the
NO
Replace the LCD unit with the new one.
surface?
YES
NO
Is PJ204 connected correctly? Connect it correctly.
YES
Turn OFF the power to check.
NO
Is the resistance between TP2 and 3 several kΩ?
YES
Is the resistance between TP5 and 6 several kΩ?
YES
Failure of the touch panel is a probable cause. Replace the LCD unit with the new one.
NO
Failure of the touch panel is a probable cause. Replace the LCD unit with the new one.
Is the resistance between TP1 and 6 more than
10 k? Does it change to several k by pressing
the touch panel?
YES
Is the resistance between TP3 and 4 more than
10 k? Does it change to several k by pressing
the touch panel?
YES
The touch panel is probably normal. Replace the DISPLAY board with the new one. (See 4.7.)
NO
Failure of the touch panel is a probable cause. Replace the LCD unit with the new one.
NO
Failure of the touch panel is a probable cause. Replace the LCD unit with the new one.
3.5 Seat switch does not work
NO
Does the touch panel work?
YES
3 - 6
Failure of the DISPLAY board or other units. Replace the DISPLAY board or check other items. (See 4.7.)
Does a mode change or the beep sound by operating the switch on the seat switch? If a volume is at low level, the beep can not be heard.
NO
Is the connection terminal of seat switch connected to the connector (J206) on the DISPLAY board?
YES
Replace the seat switch with the new one. Is it possible to operate by the seat switch?
YES
The cause was failure of the seat switch.
YES
NO
NO
Normal. Refer to the operator's manual and check the operating procedures.
Connect the terminal of seat switch to the connector.
Failure of the DISPLAY board. Replace it with the new one. (See 4.7.)
3.6 Remote control does not work
Failure of DISPLAY board or the other units.
NO
Does the touch panel work?
YES
Is the power of remote control turned ON by pressing the power switch of remote control?
YES
Are codes of remote control and RC of the main body same?
Replace the DISPLAY board or check the other items. (See 4.7.)
Replace batteries of remote control with the new
NO
one. If the power is not turned ON even replacing batteries, it shows failure of remote control itself.
Use a same cord for remote control and RC of the main body.
NO
The cord for remote control is set by the dip switch in batteries of remote control. The cord for the main body is set on the CUSTOM screen.
3 - 7
YES
Are there any other infrared light on the photoreceptor of the display unit?
NO
Is the cable of RECEIVER board connected to J210 on the DISPLAY board?
YES
Connect the photoreceptive signal between R5 and R6 on the DISPLAY board to the oscilloscope. Does the photoreceptor receive remote control signal and input the signal to the oscilloscope?
YES
Failure of the DISPLAY board. Replace it with the new one. (See 4.7.)
YES
NO
NO
Eliminate the infrared light.
Connect the cable.
Failure of the remote control or the RECEIVER board on the display unit are probable causes. Replace either of them with the new one. (See 4.21.)
3.7 Sounds do not come out
3 - 8
Is the volume control knob at proper position? No sounds can be heard if the knob is turned to the lowest level.
YES
YES
Can the beep sound be heard by pressing switches?
NO
Is the cable from speaker connected to J308 on the MAIN board?
YES
Is the resistance between 1st and 2nd pins of
P308 connector of speaker 7 to 8Ω?
YES
Failure of the MAIN board. Replace it with the new one. (See 4.8.)
NO
Turn the volume control knob to proper position.
NO
Connect the cable.
NO
Failure of speaker. Replace it with the new one. (See 4.22.)
Is sound output set ON? Check the sound output setting by VOICE on the CUSTOM screen.
YES
Input sound data from the sound card on the service screen. Do sounds become normal?
YES
End
3.8 Volume can not be adjusted
Does the voltage of TP8 (VOLUME) on the DISPLAY board change by turning the volume control knob?
YES
NO
Set the sound setting to ON.
NO
Failure of the MAIN board. (See 4.8.)
NO
Failure of the VR board. Replace it with the new one. (See 4.20.)
Failure of the volume control circuit on the MAIN board is a probable cause. Replace the MAIN board with the new one. (See 4.8.)
3 - 9
3.9 Irrigation pole does not move by irrigation UP/DOWN
switch
NO
Does the irrigation pole move up and down by the LCD touch panel?
YES
Disconnect P310 from the MAIN board and check switch operation. Is switch contact turned ON or OFF correctly by operating the irrigation pole UP/DOWN switch?
YES
Failure of the MAIN board. Replace it with the new one.
Is the limit switch (micro switch) on the upper side or the lower side of the irrigation pole turned ON?
NO
NO
YES
Failure of the irrigation pole UP/DOWN switch. Replace it with the new one. (See 4.23.)
The irrigation pole does not move since it comes to the limit position. Check the limit switch or adjust the pole position. (See 4.9.) Moreover, the pole can not move since the upper side or lower side of the irrigation pole is bitten in some cases.
Is the signal of limit switch of the irrigation pole turned ON? MAIN board UP LIMIT SWIC20-2 pin 0V=ON, 5V=OFF DOWN LIMIT SW IC20-3 pin 0V=ON, 5V=OFF
NO
Is the indication of irrigation height within the working range?
YES
Is the control signal output by pressing the irrigation pole UP/DOWN switch? MAIN board UP SW :IC21-19 pin 0V=ON、5V=OFF DOWN SW IC20-18 pin 0V=ON5V=OFF BU CONNECT board UP SW :IC2-12 pin 0V=ON、12V=OFF DOWN SW IC2-11 pin 0V=ON12V=OFF
YES
YES
NO
NO
Failure of the limit switch is a probable cause. Check the limit switch or the signal line.
The irrigation pole moves up and down only within the working range. Ajdust the potentiometer if the height position is not correct. (See 5.10.)
Failure of the MAIN board or the BU CONNECT board are probable causes. Replace the board with the new one. (See 4.8. and 4.19.)
3 - 10
Is AC100V supplied to the POLE CONTROL board? Check the voltage between 1st and 3rd pins of P712 is AC100V.
YES
Is the connector on the POLE CONTROL board connected correctly?
YES
Press the irrigation pole UP switch. Is the relay (RY1) on the POLE CONTROL board turned ON? Is the AC100V supplied between 1st and 4th pins of J706 connector?
YES
Press the irrigation pole DOWN switch. Is the relay (RY2) on the POLE CONTROL board turned ON? Is the AC100V supplied between 2nd and 4th pins of J706 connector?
NO
NO
NO
NO
The transformer or the input power is abnormal. Check the input power or replace the transformer with the new one. (See 4.18.)
Connect it correctly.
Failure of the POLE CONTROL board is a probable cause. Replace the board with the new one. (See 4.25.)
Failure of the POLE CONTROL board is a probable cause. Replace the board with the new one. (See 4.25.)
YES
Does the AC motor rotate by pressing the irrigation pole UP/DOWN switch?
YES
Does the irrigation pole move up and down if the AC motor for the irrigation pole rotate?
YES
Normal.
Failure of the AC motor.
NO
Replace the AC motor or the condenser for motor (C1) with the new one. (See 4.10 and 4.36.)
The gear or the belt of the motor may be
NO
abnormal. Check them.
3.10 Printer does not work normally
3 - 11
Is paper loaded to the printer? Load paper to the printer.
NO
YES
YES
Is it possible to print by the PRINT switch on the rear panel?
NO
Is it possible to print by the PRINT switch on the touch panel?
NO
Is the connector on the MAIN board (J312) connected correctly?
YES
YES
NO
Replace the PRINT/FEED switch (BA29) or the cable (EA41) with the new one. (See 4.24.)
Connect it.
Failure of the printer, the board (BA08) or the MAIN board are probable causes. Replace the printer or the board with the new one. If the symptom is not solved, replace the MAIN board with the new one. (See 4.8. and 4.14.)
Are date and time printed correctly?
YES
Failure of the printer, the board (BA08) or the MAIN board are probable causes. Replace the printer or the board with the new one. If the symptom is not solved, replace the MAIN board with the new one. (See 4.8 and 4.14.)
NO
Set time on the CUSTOM screen.
3 - 12
3.11 Memory inside of main body does not work correctly
Is the Store switch pressed to store the contents of program?
YES
Failure of the MAIN board or the flash memory are probable causes. Replace the board with the new one. (See 4.8.)
NO
Press the Store switch to store the program to the flash memory.
3.12 Memory card does not work correctly
Is the memory card for programming inserted? Insert the memory card for programming.
YES
Does the memory card operation screen appear by pressing the memory card switch?
YES
NO
NO
The memory card is not recognised for use in programming. Check the memory card.
Is it possible to write to the memory card?
YES
Replace the memory card with the new one. Does it work normally?
NO
Is the cable between the MAIN board and the CARD board connected correctly?
YES
Failure of the MAIN board (BA03) or the CARD board (BA24) are probable causes. Replace the board with the new one. (See 4.8 and 4.38.)
NO
YES
NO
The memory card may be protected. Unprotect the card.
Breakage of the memory card is a probable cause.
Connect it correctly.
3.13 Cassette test is not passed
NO
Insert the cassette. Does the cassette loading start?
YES
3 - 13
Replace the cassette with the new one. Is the cassette test passed?
NO
Does an error occur when the cassette test is finished?
NO
Normal
YES
Does the screen change to the cassette test by inserting the cassette?
NO
Is the continuity between 1st and 2nd pins of P515 connected to the DRIVER board turned ON from OFF by inserting the cassette?
YES
YES
NO
Failure of the cassette is a probable cause.
Check the cause of error from the error code.
Failure of the microswitch (SW56) is a probable cause. Replace it or check soldering. (See 4.26.)
YES
Is the continuity between 1st and 2nd pins of P516 connected to the DRIVER board turned ON from OFF by inserting the cassette?
YES
Failure of the DRIVER board is a probable cause. Replace it with the new one. (See 4.29.)
Failure of the DRIVER board is a probable cause. Replace the board with the new one. (See 4.29.)
NO
Failure of the microswitch (SW58) is a probable cause. Replace it or check soldering. (See 4.26.)
3.14 Cassette can not be removed
3 - 14
YES
Is the continuity of the cable between the DRIVER board and the EJECT switch normal?
YES
Does the 3rd pin of IC3 on the DRIVER board change from 5V to 0V by pressing the EJECT switch?
NO
Disconnect the P524. Does the continuity between 5th and 7th pins of P524 change by pressing the EJECT switch?
YES
Does the CASSETTE EJECT LED change from
YES
lighting up to blinking by pressing the EJECT siwtch?
NO
YES
Is the state of the poistion O indicated while the foot switch is not pressed on the screen?
NO
NO
Replace or repair the cable.
NO
Replace the board of the EJECT switch with the new one. (See 4.27.)
Disconnect the connector of the foot switch and press the EJECT switch. Is it possible to remove the cassette?
YES
Failure of the foot switch.
Does the DC motor of EA38 rotate?
YES
Does the cassette move down?
YES
Are all pinch valves turned up?
NO
Do pinch valves move by replacing the unmoving valves? If there is no pinch valve to replace, connect to the connector of other pinch valve and check the operation.
YES
NO
NO
NO
YES
NO
Failure of the cable and the board (BU CONNECT board) between the foot switch connector and the MAIN board. (See 4.19.)
Replace the control unit (18214-1130) with the new one. (See 4.11.)
Replace the control unit (18214-1130) with the new one. (See 4.11.)
Replace the pinch valve with the new one. (See 4.39.)
3 - 15
Does the 24V of voltage occur between 1st and 2nd pins of J513 while ejecting the cassette by the EJECT switch?
NO
Is DC 24V supplied to the 7th pin of P501?
YES
Is DC 5V supplied to the 1st pin of P501?
YES
Failure of the DRIVER board, the MAIN board or the cable between boards. Replace either of them with the new one. (See 4.8 and 4.29.)
YES
Failure of the DC motor (EA38) or the loading mechanism. Replace the control unit (18214-
1130) with the new one. (See 4.11.)
Failure of the power cable (EA34) to the DRIVER
NO
board, or the power. Replace the cable with the new one or check the power following the item of power. (See 3.30.)
Failure of the power cable (EA34) to the DRIVER
NO
board, or the power. Replace the cable with the new one or check the power following the item of power. (See 3.30.)
3.15 SYSTEM (cataract) test is not passed
3.15.1 SINGLE CASSETTE
3 - 16
Does the ERROR CODE E200 No.17 appear on the screen?
NO
Does the ERROR CODE E200 No.31 appear on the screen?
NO
A
YES
Is the connection set removed during a test?
YES
Eject the cassette once. Start SYSTEM TEST again after connecting the connection set.
YES
Are each conductors of EA44 continuous?
NO
Replace the EA44 with the new one.
Shade the each photointerrupters of BA28. Do the 2nd, 3rd and 4th pins of IC4 on the DRIVER board (BA05) change ON or OFF?
NO
Replace the TUBE SENS board (BA28) with the new one. (See 4.28.)
Replace the DRIVER board (BA05) with the new one. (See 4.29.)
YES
Does the ERROR CODE E307 appear simultaneously?
YES
Failure of EA39 or the DRIVER board (BA05). (See 4.29.)
NO
YES
NO
YES
Does the aspiration motor rotate while all pinch valves are closed?
NO
Failure of the pinch valve. Check the pinch valve following the item of irrigation and aspiration. (See 3.21 and 3.22.)
An air leakage of the cassette is a probable cause. Replace the cassette with the new one.
Select a mode on the service mode. Does the aspiration
YES
pressure change by pressing the aspiration sensor with a hand?
NO
Is DC5V supplied to the 1st pin of J262 on the VACUUM SENS board (BA26)? (See 4.37.)
NO
YES
Are each conductors of EA46 continuous?
NO
Failure of EA46.
Failure of the VACUUM SENS board (BA26). (See 4.37.)
Select a mode in the service mode while the cassette is
YES
not inserted and press the load cell with a hand. Does the voltage of TP1 on the VACUUM SENS board (BA26) change?
NO
Failure of the EA31. Replace it with the new one.
Failure of the VACUUM SENS board (BA26) or EA31. Replace or readjust either of them. (See 4.37 and 5.11.)
YES
3 - 17
A
Does the ERROR CODE E200 No.32 appear on the screen?
NO
Does the ERROR CODE E200 No.33 appear on the screen?
NO
YES
An air leakage of the cassette is a probable cause. Replace the cassette with the new one.
YES
Does the VENT VALVE work?
NO
Failure of the pinch valve. Check the pinch valve following the item of irrigation and aspiration. (See 3.21 and 3.22.)
YES
Is the flow rate control of transfusion tube closed?
NO
Is the opening for air intake of spike of transfusion tube closed?
YES
Open the opening for air intake or insert the needle. And then, perform SYSTEM TEST again.
The cassette is probably clogged. Replace the cassette with the new one.
YES
NO
Does the ERROR CODE E200 No.34 appear on the screen?
NO
B
Open the flow rate control and perform SYSTEM TEST again.
Are the aspiration line, the irrigation line,
YES
handpiece and the test chamber connected securely?
NO
Connect the aspiration line, the irrigation line, handpiece and the test chamber securely.
YES
Does the IRR1 valve work normally?
NO
Check the pinch valve following the item of irrigation. (See 3.21.)
The cassette is probably clogged. Replace the cassette with the new one.
YES
3 - 18
B
YES
Does ERROR CODE E200 No.35 appear? Does the ASP1 valve work normally?
NO
NO
Check the pinch valve following the item of aspiration. (See 3.22.)
The cassette or handpiece is probably clogged.
YES
Does ERROR CODE E200 No.36 appear?
NO
Is the connection set inserted securely?
NO
Insert the connection set securely and perform SYSTEM TEST again.
An air leakage of cassette is a probable cause. Replace the cassette with the new one.
YES
Does ERROR CODE E200 No.42 appear?
NO
Does the cassette loader work?
NO
Check the control unit following the item that cassette can not be removed. (See 3.14.)
C
YES
Are each conductors of EA44 continuous securely?
YES
YES
YES
NO
Replace the EA44 with the new one.
Failure of the CASSETTE SENS board (BA27) or the adjustment failure of the control unit. Replace the control unit or BA27, and readjust the replaced one. (See 4.11, 4.30 and 5.11.)
3 - 19
C
Does the ERROR CODE E200 No.43 appear?
NO
Does the ERROR CODE E200 No.44 appear?
NO
YES
Does the drainage bag come to the limit?
YES
Replace the cassette with the new one.
NO
Is the problem solved by replacing the cassette?
YES
Failure of the cassette. Readjust the control unit if the error frequently occurs. (See 4.11.)
Readjust the control unit. (See 5.11.)
YES
Does the cassette loader work?
NO
Check the control unit following the item that cassette can not be removed. (See 3.14.)
NO
YES
Does the ERROR CODE E210 appear?
NO
If the switch of SYSTEM TEST can not be pressed, it means occurrence of an error. Follow the indicated error code to solve it.
YES
Are each conductors of EA44 continuous securely?
NO
Replace the EA44 with the new one.
Failure of the CASSETTE SENS board (BA27) or adjustment failure of the control unit. Replace the control unit or BA27, and readjust the replaced one. (See 4.11, 4.30 and 5.11.)
YES
Check the ultrasound oscillation part following the item of TEST US.
3.15.2 ANT DUAL CASSETTE
3 - 20
Does the ERROR CODE E200 No.17 appear?
NO
Does the ERROR CODE E200 No.31 appear?
NO
A
YES
Is the connection set removed during TEST?
YES
Eject the cassette once. Perform SYSTEM TEST again while connecting the connecting set.
YES
Are the each conductors of EA44 continuous?
NO
Replace EA44 with the new one.
Shade the each photointerrupters of BA28. Does the 2nd, 3rd and 4th pins of IC4 of BA05 change to ON or OFF?
NO
Replace BA28 with the new one. (See 4.28.)
Replace BA05 with the new one. (See 4.29.)
YES
Does ERROR CODE E307 appear simultaneously?
YES
Failure of EA39 or BA05. (See 4.29.)
NO
YES
NO
YES
Does the aspiration motor rotate while the pinch valve is fully opened?
NO
Failure of the pinch valve. Check the pinch valve following the item of irrigation and aspiration. (See 3.21 or 3.22.)
An air leakage of the cassette is a probable cause. Replace the cassette with the new one.
Select a mode in the service mode and press the
YES
aspiration sensor with a hand. Does the aspiration pressure change?
NO
Is DC5V supplied to the 1st of J262 of BA26?
NO
YES
Are each conductors of EA46 continuous?
NO
Failure of EA46.
Failure of BA26. (See 4.37.)
Select a mode without inserting a cassette and
YES
press the load cell with a hand. Does the voltage of TP1 of BA26 change?
NO
Failure of EA31. Replace it with the new one.
Failure of BA26 or EA31. Replace or readjust either of them (See 4.37 and 5.11.)
YES
Loading...
+ 113 hidden pages