Mitutoyo 516S, 503S, 501S, LSM-500S, 506S, 512S User Manual
No. 99MBC094
A
SERIES No. 544
LSM-500S/501S/503S/
506S/512S/516S
Laser Scan
Micrometer
(Measuring Unit)
User’s Manual
Read this User’s Manual thoroughly
before operating the instrument. After reading,
retain it close at hand for future reference.
[MEMO]
CONVENTIONS USED IN USER’S MANUAL
Mitutoyo manuals use various safety signs. The visual cues and the contents of
description appended to each cue are described below.
Safety Precautions
To ensure that instruments are operated correctly and safely, Mitutoyo manuals use safety
signs (Signal Words and Safety Alert Symbols) to identify and warn against hazards and
potential accidents.
The following signs indicate general warnings:
Indicates an imminently hazardous situation which, if not avoided, will result in serious
injury or death.
DANGER
WARNING
CAUTION
Indicates a potentially hazardous situation which, if not avoided, could result in serious
injury or death.
Indicates a potentially hazardous situation which, if not avoided, may result in minor or
moderate injury or property damage.
The following signs indicate specific warnings or prohibited actions, or indicate a
mandatory action:
Alerts the user to a specific hazardous situation. The given example means “Caution, risk
of electric shock”.
Prohibits a specific action. The given example means “Do not disassemble”.
No. 99MBC094A
Specifies a required action. The given example means “Ground”.
i
CONVENTIONS USED IN USER’S MANUAL
Various Types of Notes
The following types of notes are provided to help the operator obtain reliable
measurement data through correct instrument operation.
IMPORTANT
NOTE
TIP
• An important note is a type of note that provides information essential to the completion
of a task. You cannot disregard this note to complete the task.
• An important note is a type of precaution, which if neglected could result in a loss of
data, decreased accuracy or instrument malfunction/failure.
A note emphasizes or supplements important points of the main text. A note supplies
information that may only apply in special cases (e.g., Memory limitations, equipment
configurations, or details that apply to specific versions of a program).
A tip is a type of note that helps the user apply the techniques and procedures described
in the text to his or her specific needs.
It also provides reference information associated with the topic being discussed.
Mitutoyo assumes no liability to any party for any loss or damage, direct or
indirect, caused by use of this instrument not conforming to this manual.
Information in this document is subject to change without notice.
Before exporting this product confirm the final purpose of use at the export destination to
prevent the product from being used for developing weapons of mass destruction or
military affairs. In the case of export to the U.S., this product requires an application for
prior approval of CDRH (Center for Devices and Radiological Health) in FDA (Food and
Drug Administration). An additional cost generates in application procedure. For detailed
information consult a Mitutoyo sales office. Also, if this product is exported with it
incorporated in equipment, the final product requires an application for FDA approval. If
this is the case, note that the client must file an application for approval.
No. 99MBC094A
iii
INSTALLATION CONDITIONS
The Mitutoyo Laser Scan Micrometer LSM-500H series is an instrument for indoor use.
Also, this series is a precision optical instrument and a precision electronic instrument.
Therefore, it must be carefully installed and the following conditions must be taken into
account to attain the highest possible accuracy. Mitutoyo assumes no liability for any
accident or damage caused by use of this instrument not conforming to these conditions.
1. Vibration
Install the LSM in a place with minimum vibration. If the LSM is subjected to vibrations for
a long period of time, precision components in the system may malfunction, thus affecting
the measuring accuracy.
If the vibrations are significant, minimize them by laying a vibration damping rubber pad,
etc.
2. Dust/Grit
Dust, if present in the installation site, will affect the optical components such as
protection glass (in the measuring unit) and electronic components (in the display unit).
Install the LSM in a place that is as free as possible from dust/grit.
3. Direct sunlight
If this instrument is exposed to direct sunlight, the heat may deform the instrument and
adversely affect the measuring accuracy.
If this instrument must be placed by a window where it will be subject to direct sunlight,
protect the instrument by shading it with curtains, etc.
4. Blow from the air conditioner
If the measuring place is exposed to hot or cool air blow from the air conditioner, the
laser beam may be refracted due to the difference in air density and adversely affect the
measuring accuracy.
If this is the case, block direct air flow with curtains, etc.
5. Ambient temperature and humidity
Avoid installing this instrument in an environment where there is rapid temperature or
humidity change. Otherwise, it may reduce measuring accuracy. This instrument must be
operated in the following ambient conditions.
Temperature: 0°C to 40°C
Humidity: 35%RH to 85%RH
iv
No. 99MBC094A
WARRANTY
This instrument has been manufactured under rigorous Mitutoyo quality control. Should
it malfunction under normal use within one year from the date of original purchase, it will
be repaired or replaced, at our option, free of charge. Contact your dealer or the nearest
Mitutoyo representative for more information.
However, the following damages may be subject to a repair charge even if damage
occurs within the warranty period:
1. Unit malfunction or damage arising from improper handling, or unauthorized retrofit or
repair by the user.
2. Unit malfunction or damage as the result of moving, dropping, or transporting after
purchase.
3. Unit malfunction or damage due to fire, salt, gas, abnormal voltage, or natural
catastrophe.
This warranty is not transferable and is only valid within the country of the original
purchase.
No. 99MBC094A
v
CONTENTS
CONVENTIONS USED IN USER’S MANUAL....................................................................................................i
NOTES FOR EXPORTING................................................................................................................................iii
5.2 Safety Precaution labels located on LSM-500S Series.............................................................5-2
SERVICE NETWORK
No. 99MBC094A
vii
MEMO
viii
No. 99MBC094A
1 INTRODUCTION
1
1.1 Outline
The Laser Scan Micrometer is a high-precision laser measuring system that performs
non-contact dimensional measurements with a high-speed scanning laser beam.
With non-contact measurement capability, this system features high-precision
measurement of workpieces that are difficult for conventional measuring systems to
measure, including hot workpieces, brittle or elastic workpieces, workpieces that must be
kept free from contamination, and soft workpieces subject to the measuring force.
1.2 Introduction
This chapter describes the maintenance and inspecton of the Laser
Scan Micrometer (LSM) measuring unit.
This measuring unit is used with the display unit of the LSM-6200. Since this User’s
Manual primarily describes the specifications of the measuring unit, please refer to the
User’s Manual of the LSM-6200 for details about connection with a display unit, functions,
and measurement procedures.
The measuring unit is provided with a shutter or cap on the emission unit and reception
unit. Laser equipment has such safety precautions.
During measurement: Open the shutter by sliding it as shown in the below left diagram
of fig-a), or remove the lens cap as shown in the below left diagram of fig-b).
During no-measurement: Close the shutter by sliding it as shown in the below right
diagram of fig-a), or mount the lens cap as shown in the below right diagram of fig-b).
Laser beam
Open Close
a) LSM-501S/503S/506S/512S/516S
window
Shutter
window
Shutte
Laser beam
window
Open Close
b) LSM-500S
Lens cap
No. 99MBC094A
1-1
1.3 Inspection and Maintenance
This section describes the inspection and maintenance procedures of the measuring unit.
1.3.1 Laser emission display LED
The Laser Emission Display LED that is provided on the rear side lights while the laser
is being emitted to call the operator’s attention for safety. Do not look the laser beam.
Laser emission
display LED
1.3.2 Cleaning optical parts
To clean the optical parts, turn off the power switch and disconnect the signal cable for
safety.
Always keep the protective glass of the emission unit and reception unit clean. Soiled
protective glass will not only result in reduced measurement accuracy, but possibly
produce erroneous measurements due to dust and foreign particles being treated as part
of the workpiece.
For cleaning, use a blower brush or use gauze slightly dampened with ethyl alcohol to
gently wipe clean soiled portions using very light pressure.
z Checking the contamination of the protective glass using an oscilloscope
A monitor connector, “SCAN SIG.-1”, of the reception light signal is provided on the
rear panel of the display unit. Check the reception light signal by connecting the probe of
the oscilloscope with the monitor connector. Oscilloscope setup is as follows:
z Vertical sensitivity: 0.1V/DIV (with a probe of 10:1)
z Horizontal sensitivity: 100 µs/DIV
If the waveform generated by the oscilloscope looks like either of the ones shown in (b)
or (c), clean the protective glass to reduce the disorder of waveform to less than 0.3V.
Approx.
310μS
(a) Normal
Approx.
4V
(b) Soiled glass
(c) Soiled glass
NOTE
1-2
・The protective glass of the windows is a precision optical part. Handle with care so as
not to scratch the glass.
No. 99MBC094A
1.3.3 Replacing the protective glass and the condenser lens
Follow the procedures below to remove the protective glass for replacement or cleaning.
To replace the protective glass, first turn off the power switch, then disconnect the signal
cable for safety.
Model LSM-501S/LSM-503S/LSM-506S/LSM-512S/LSM-516S
Removing steps Unscrew the screws that secure the protective glass.
Remove the protective glass and the rubber packing.
Remounting steps Remount the protective glass by reversing the steps above.
NOTE Dust will enter the enclosure if the protective glass is not re-installed. In a
humid environment damp air will enter the enclosure, causing reduced
optical clarity. So, ensure that the work is done in an air-conditioned room
that is free from dust.
Model LSM-500S
Removing steps Remove the protective glass and the condenser lens by rotating them
counterclockwise.
Remove the O-ring.
Remounting steps Attach the O-ring.
Mount the protective glass and the condenser lens in place by rotating them
clockwise, as shown below. If mounted in the wrong order, measurement
cannot be taken.
NOTE Dust will enter the enclosure if the protective glass is not re-installed. In a
humid environment damp air will enter the enclosure, causing reduced
optical clarity. So, ensure that the work is done in an air-conditioned room
that is free from dust.
1. INTRODUCTION
Protective glass
No. 99MCA094A
Condenser lens
1-3
MEMO
1-4
No. 99MBC094A
GAGE
GAGE
2 CALIBRATION
2
The accuracy of the instrument has been verified by two-point calibration using two
reference gages.
High-precision calibration is possible and simple to perform, as mentioned in the User’s
Manual attached to the display unit. Calibrate the instrument properly. Types and sizes of
the calibration gages are as follows:
Model LSM-500S LSM-501SLSM-503SLSM-506SLSM-512S LSM-516S
Size φ0.1mm φ 1mm φ 20mm LOW CAL
Type With-holder Stepped
Size φ 2mm φ 10mm φ 30mm φ 60mm φ 120mm φ 160mm HIGH CAL
Type With-holder Straight Stepped
With-holder type
No. 99MBC094A
Straight type Stepped type
2-1