SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD. (hereinafter called Mindray)
owns all rights to this unpublished work and intends to maintain this work as confidential.
Mindray may also seek to maintain this work as an unpublished copyright. This publication is to
be used solely for the purposes of reference, operation, maintenance, or repair of Mindray
equipment. No part of this can be disseminated for other purposes.
In the event of inadvertent or deliberate publication, Mindray intends to enforce its rights to this
work under copyright laws as a published work. Those having access to this work may not
copy, use, or disclose the information in this work unless expressly authorized by Mindray to
do so.
All information contained in this publication is believed to be correct. Mindray shall not be liable
for errors contained herein nor for incidental or consequential damages in connection with the
furnishing, performance, or use of this material. This publication may refer to information and
protected by copyrights or patents and does not convey any license under the patent rights of
Mindray, nor the rights of others. Mindray does not assume any liability arising out of any
infringements of patents or other rights of third parties.
Content of this manual is subject to changes without prior notice.
PROPERTY OF SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.
ALL RIGHTS RESERVED
Responsibility on the manufacturer party
Mindray is responsible for safety, reliability and performance of this equipment only in the
condition that:
•all installation, expansion, change, modification and repair of this equipment are conducted
by Mindray qualified personnel;
• applied electrical appliance is in compliance with relevant National Standards;
• the monitor is operated under strict observance of this manual.
Warning
For continued safe use of this equipment, it is necessary that the listed instructions are
followed. However, instructions listed in this manual in no way supersede established medical
practices concerning patient care.
zDo not rely only on audible alarm system to monitor patient. When monitoring
I
Service Manual
adjusting the volume to very low or completely muting the sound may result in the
disaster to the patient. The most reliable way of monitoring the patient is at the
same time of using monitoring equipment correctly, manual monitoring should be
carried out.
z This multi-parameter patient monitor is intended for use only by medical
professionals in health care institutions.
z T o avoid electrical shock , you shall not open any cover by yourself. Service must be
carried out by qualified personnel.
z Use of this device may affect ultrasonic imaging system in the presence of the
interfering signal on the screen of ultrasonic imaging system. Keep the distance
between the monitor and the ultrasonic imaging system as far as possible.
z It is dangerous to expose electrical contact or applicant coupler to normal saline,
other liquid or conductive adhesive. Electrical contact and coupler such as cable
connector, power supply and parameter module socket-inlet and frame must be
kept clean and dry. Once being polluted by liquid, they must be thoroughly dried. If
to further remove the pollution, please contact your biomedical department or
Mindray.
It is important for the hospital or organization that employs this equipment to carry out a
reasonable maintenance schedule. Neglect of this may result in machine breakdown or injury
of human health.
II
Service Manual
Warranty
THIS WARRANTY IS EXCLUSIVE AND IS IN LIEU OF ALL OTHER WARRANTIES,
EXPRESSED OR IMPLIED, INCLUDING WARRANTIES OF MERCHANT ABILITY OR
FITNESS FOR ANY PARTICULAR PURPOSE.
Exemptions
Mindray's obligation or liability under this warranty does not include any transportation or other
charges or liability for direct, indirect or consequential damages or delay resulting from the
improper use or application of the product or the substitution upon it of parts or accessories not
approved by Mindray or repaired by anyone other than a Mindray authorized representative.
This warranty shall not extend to any instrument which has been subjected to misuse,
negligence or accident; any instrument from which Mindray's original serial number tag or
product identification markings have been altered or removed, or any product of any other
manufacturer.
Safety, Reliability and Performance
Mindray is not responsible for the effects on safety, reliability and performance of the
PM-7000 Patient Monitor if:
■ assembly operations, extensions, re-adjusts, modifications or repairs are carried out
by persons other than those authorized by Mindray.
■ the PM-7000 is not used in accordance with the instructions for use, or the electrical
installation of the relevant room does not comply with NFPA 70: National Electric
Code or NFPA 99: Standard for Health Care Facilities (Outside the United States, the
relevant room must comply with all electrical installation regulations mandated by the
local and regional bodies of government).
III
Service Manual
Return Policy
Return Procedure
In the event that it becomes necessary to return a unit to Mindray, the following procedure
should be followed:
1. Obtain return authorization. Contact the Mindray Service Department and obtain a
Customer Service Authorization (Mindray) number. The Mindray number must appear on
the outside of the shipping container. Return shipments will not be accepted if the Mindray
number is not clearly visible. Please provide the model number, serial number, and a brief
description of the reason for return.
2. Freight policy. The customer is responsible for freight charges when equipment is shipped
to Mindray for service (this includes customs charges).
In this manual, the signal wordsDANGER, WARNING, andCAUTION are used
regarding safety and other important instructions. The signal words and their meanings are
defined as follows. Please understand their meanings clearly before reading this manual.
Signal word Meaning
DANGER
WARNING
CAUTION
Indicates an imminently hazardous situation which, if not
avoided, will result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided,
could result in death or serious injury.
Indicates a potentially hazardous situation which, if not avoided,
may result in minor or moderate injury.
2. Meaning of Safety Symbols
Symbol Description
Type-BF applied part
"Attention" (Refer to the operation manual.)
Safety Precautions
Please observe the following precautions to ensure the safety of service engineers as well as
operators when using this system.
DANGER: Do not use flammable gases such as anesthetics, or flammable
liquids such as ethanol, near this product, because there is danger
of explosion.
WARNING: Do not connect this system to outlets with the same circuit
breakers and fuses that control current to devices such as
life-support systems. If this system malfunctions and
generates an over current, or when there is an instantaneous
V
Service Manual
current at power ON, the circuit breakers and fuses of the
building’s supply circuit may be tripped.
CAUTION:1. Malfunctions due to radio waves
(1) Use of radio-wave-emitting devices in the proximity of this
kind of medical electronic system may interfere with its
operation. Do not bring or use devices which generate radio
waves, such as cellular telephones, transceivers, and radio
controlled toys, in the room where the system is installed.
(2) If a user brings a device which generates radio waves near the
system, they must be instructed to immediately turn OFF the
device. This is necessary to ensure the proper operation of
the system.
2. Do not allow fluids such as water to contact the system or
peripheral devices. Electric shock may result.
VI
Symbols
Equipotential grounding
terminal
Be Careful
Service Manual
CE mark 93/42/EEC a
directive of the European
Economic Community
Protective earth ground
Direct current and alternating
current (DC&AC)
Alternating current (AC)
ESD sensitivity
Network connector
Indicates that the instrument is IEC-60601-1 Type CF equipment. The unit
displaying this symbol contains an F-Type isolated (floating) patient
applied part providing a high degree of protection against shock, and is
suitable for use during defibrillation.
Direct current (DC)
Battery indicator
Power ON/OFF
High voltage
VII
Service Manual
Contents
Chapter 1 About the Product .....................................................................................................1
* Conditions for measurements in typical precision:
The measurement is started after the preheating mode of the module;
Ambient pressure: 750mmHg to 760mmHg; room temperature: 22 to 28;℃℃
The gas under test is dry, and the balance gas is N2;
The deflation rate is 150ml/min, the respiration rate is no greater than 30BrPM, with a
fluctuation less than ±3BrPM, and the inhale interval/exhale interval is 1:2;
In other conditions, the measurement precision should meet the requirements of EN864 or
ISO9918: ±4mmHg (0 to 40mmHg) or ±12% of the reading (41 to 99mmHg)
1.1.5. Oridion CO
Specifications
2
measurement range 0 to 99mmHg
CO
2
0 to 38 mmHg:
Precision*
39 to 99 mmHg:
Waveform:
Resolution
Value:
7.5
−
50
Flow rate
15
+
ml/min
Initialization time 30s (typical)
Response time 2.9s (typical)
Delay time 2.7s (typical)
AwRR measurement range 0 to 150 BrPM
0 to 70BrPM:
AwRR measurement
precision
70 to 120BrPM:
121 to 150BrPM:
±2mmHg
±5% + 0.08%× (reading - 38mmHg)
0.1mmHg
1mmHg
±1BrPM
±2BrPM
±3BrPM
Apnea alarm delay AwRR: 10 to 40s
* Precision applies for breath rates of up to 80 bpm. For breath rates above 80 bpm, accuracy
complies with EN 864/ISO 9918 (4 mmHg or ±12% of reading whichever is greater) for EtCO2
values exceeding 18 mmHg. To achieve the specified accuracies for breath rates above 60
breaths/minute, the Microstream® FilterLine H Set for Infant/Neonatal (p/n 006324) must be
used.
The accuracy specification is maintained to within 4% of the values indicated in the above table
in the presence of interfering gases according to EN864 Section Eleven, Part 101.
42
Chapter 3 Product Specifications
1.1.6. Welch Allyn CO
Specifications
2
CO2 measurement range 0 to 99mmHg
Precision*
0 to 40 mmHg:
41 to 76 mmHg:
77 to 99 mmHg:
±2mmHg
±5%
±10%
Resolution 1mmHg
Refreshing rate 1s
Start-up time
< 80s (ambient temperature: 25; preheating power of ℃
transducer: 5W)
Response time 100ms (10% to 90 %)
Calibration Daily calibration is unnecessary
Calibration stability
There is a difference (< 1%) from the precison criteria after a
12-month continuous service time
Alarm range 0 to 99 mmHg
AwRR measurement range 0 to 150 BrPM
AwRR alarm range 0 to 150 BrPM
Apnea alarm delay AwRR: 10 to 40 s
* Precision specification is based upon the following standard airway conditions: sensor 42; ℃
airway adapter temperature 33; water vapor pressure 38 mmHg; standard gas mixture equals ℃
CO2 in balance air; fully hydrated at 33; atmospheric pressure 760 mmHg; airway flow rate ℃
60 cc/min.
43
Chapter 3 Product Specifications
3.16 AG Specifications
Measurement technique Infrared absorption
Measurement mode Side stream
AG functions
Meets requirements of ISO9918, ISO11196, EN12598 and
In case of faults causing white screen or other abnormal screens,
■ Check whether the LCD connection wires are in good contact;
■ Replace the LCD connection wires, or replace the LCD if necessary;
■ Replace the main control board if the fault still exists.
4.2.3 Encoder Faults
1. If all other functions (indicator, alarm, buttons) of the button panel are normal,
proceed to step 2; otherwise, replace the button panel;
2. Check whether short-circuit or abnormal open-circuit occurs in the encoder pad;
3. Replace the encoder.
4.2.4 No Audio Alarm
1. Check whether the audio alarm function is disabled in the software settings;
2. Replace the speaker;
3. Replace the button panel.
4.2.5 Printing Failure
1. Check whether there is any alarm about the recorder. If any, eliminate it;
2. Check whether the recorder indictor is on;
3. If not, check the connection wire for inputting signals to the recorder;
4. Check whether the recorder module is enabled in the maintenance menu;
5. Check the power cord of the recorder (including the recorder power PCB);
6. Replace the recorder module.
4.2.6 Abnormal Paper Drive
1. Check whether there are blocks on the paper roller of the recorder;
2. Check whether there are blocks in the gear cluster of thermal assembly of the
recorder;
3. Check whether the voltage input of the recorder is larger than 17.6V.
54
Chapter 5 Test and Material List
Chapter 5 Test and Material List
5.1 Test Procedure
5.1.1 Connection and Checking
Connect the simulators, power supply and test fixture properly to the PM-7000, and power
it on. Then, the patient monitor displays the start-up screen on the TFT screen and enters
the system screen.
5.1.2 Functions of Buttons
Press every button on the button panel to check their functions as specified in PM-7000
Patient Monitor Operation Manual. Rotate the control knob to check its functions.
5.1.3 ECG/RESP
The TFT screen displays the standard ECG waveform, and the error between the heart
rate and the set value of the simulator is no more than ±1, namely 60±1; the RESP
waveform is smooth, and the respiration rate is 20±1.
1. Select all leads in order, including Cal, select all the four gains and AUTO, ensure
the waveforms are displayed properly, and check whether the 50Hz/60Hz
interference can be filtered.
2. Check, in all the above-mentioned cases, the consistency between the
heartbeats, the flashes of the red heart-like indicator, and the R-wave.
3. The gain has no impact on the message “ECG signal over weak” in the HR
calculation.
4. Verify the range and precision: Suppose that the amplitude of the GCG signal of
the simulator is 1mV, the heart rates are respectively 30, 60, 120, 200, 240 and
300. Check leads I, II and III. The results should meet 29-31, 59-61, 119-121,
198-202, 238-242, and 297-303.
5. PACE pulse test: Set the simulator to PACE. You should be able to view the pace.
Change PACE amplitude to ±8 – 700mv, and pulse width to 0.1ms – 2ms. The
PACE should be legible, and LEAD OFF is displayed properly.
6. RESP measurement: Set the baseline impedance to 1K, the respiration
impedance to 0.5Ω and 3Ω, and the respiration rate to 30 and 120. The
respiration rate should be 29 – 31, 118 –122.
7. PVC test: Set the simulator to the PVC mode, and set the occurrence times. The
relevant PVCS should be obtained.
8. Set the simulator as follows: RR: 40, baseline impedance: 2KΩ, RESP waveform:
3:1. Open the apnea alarm, set the respiration resistance to 0Ω, and set various
alarm time. Alarms should be given.
55
Chapter 5 Test and Material List
5.1.4 Temperature
1.
YSI probe
Select YSI probe from the manufacturer menu, select YSI temperature probe as
the test fixture, set the analog resistance to 1.471K, 1.355K and 1.249K. Then
the TEMP parameter should be 35±0.1 , 37±0.1 and 39±0.1.℃℃℃
2. CY-F1 probe
Select CY-F1 probe from the manufacturer menu, select CY-F1 temperature
probe as the test fixture, set the analog resistance to 6.534K, 6.018K and
5.548K. Then the TEMP parameter should be 35±0.1, 37±0.1℃ and ℃
39±0.1 .℃
5.1.5 NIBP
Connect the NIBP simulator, adult cuff and accessories, and then connect the module
CUFF and clockwise screw it tightly.
1. After the simulator self-test, press <ENT> to enter the ADULT analog blood
pressure mode. Set the blood pressure to the 255/195/215 mmHg level, SHIFT
to +15, and the HR to 80BPM. Set PM-7000 to the adult mode. Press <START>.
Then the results will be obtained in about 30s. The measured results should be
respectively 270±8mmHg, 210±8mmHg and 230±8mmHg.
2. Press <ESC> and <↓> on the simulator to enter the NEONATE mode. Set the
blood pressure to the 120/80/90 mmHg level, HR to 120bmp, and PM-7000 to
the pediatric mode. Press <START>. Then the results will be obtained in about
30s. The measured results should be respectively 120±8mmHg, 80±8mmHg
and 90±8mmHg.
3. Press <ESC> and <↓> on the simulator to enter the NEONATE mode. Set the
blood pressure to the 60/30/40 mmHg level, SHIFT to -20, HR to 120bmp, and
PM-7000 to the neonate mode. Change the simulator accessory to the neonatal
cuff. Press <START>. Then the results will be obtained in about 30s. The
measured results should be respectively 40±8mmHg, 10±8mmHg and
20±8mmHg.
5.1.6 SpO2
Select PLETH as the HR source of PM-7000, and put the finger into the SpO2 sensor.
The screen should display the PR and SpO
values normally. The normal SpO2 value is
2
above 97%.
5.1.7 IBP
1. Test fixture
Physiological signal simulator
2. Test procedure
IBP1 test:①
Set the BP sensitivity of the ECG simulator to 5uv/v/mmHg, BP to 0mmHG, and the
IBP channel 1 to ART. Enter the IBP PRESSURE ZERO menu of the PM-7000, zero
Channel 1, and then return to the main screen. Set the BP of the simulator to
200mmHg. Enter the IBP PRESSURE CALIBRATE menu of the PM-7000, conduct
56
Chapter 5 Test and Material List
calibration, and then exit the IBP PRESSURE CALIBRATE menu.
Set the BP value of the simulator respectively to 40mmHg, 100mmHg and
200mmHg. Then the screen of the PM-7000 should display 40±1mmHg,
100±2mmHg and 200±4mmHg.
Set the simulator output to ART wave. Then the screen of the PM-7000 should
display relevant waveform properly.
Unplug the IBP probe. Then the screen should prompt “IBP: Transducer 1 OFF!”
and “IBP: Transducer 2 OFF!”
Plug the OHMEDA cable to the IBP1 channel. Then the prompting message “IBP:
Transducer 1 OFF!” disappears.
IBP2 test:②
Plug the IBP cable to the IBP2 channel, and repeat the procedure in Section .①
5.1.8 CO
Test fixture
1.
Physiological signal simulator
2. Test procedure
Injectate and blood temperature test: Assemble the TB and TI test fixture, output
three TB temperature values: 36 , 37 and 38. Then TB should be respectively ℃℃℃
36.0±0.1 , 37.0±℃0.1 and 38.0±0.1. Set the injectate switch to ON, output two ℃℃
TI temperature values: 0 and 2. Then the screen should display 0±0.1 and ℃℃℃
2.0±0.1 .℃
CO measurement: Set the CO.CONST and T
to the default values: 0.542 and 0 , ℃
I
set the injectate switch to OFF, and then press START. Then the simulator will
output 0 , 2.5L/M and 0, 5L/M within 2s. The CO values should be 2.5±0.25L/M ℃℃
and 5±0.5L/M.
5.1.9 CO2
1. Test fixture
CO
steel bottle (containing 10% CO2)
2
2. Test procedure
Mainstream CO①
measurement: Set the calculation compensation of PM-7000 to
2
COMMON.
Plug the mainstream transducer to the CO
with the CO
the interval of 3s. The CO
steel bottle, and open/close the valve of the CO2 steel bottle based on
2
value should be the calibration gas pressure value:
2
socket, connect the windpipe connector
2
76±5%mmHg. When the valve is opened permanently, the patient monitor prompts
“APNEA ALARM”.
Unplug the transducer. The patient monitor prompts “CO
transducer OFF” on
2
the main screen. Plug the transducer again. The patient monitor prompts “CO
transducer pre-heated”.
Sidestream CO②
measurement: Set the calculation compensation of PM-7000 to
2
COMMON.
Plug the water trap to the water trap socket, connect the sampling tube with the CO
steel bottle, and open//close the valve of the CO
3s. The CO
value should be the calibration gas pressure value: 76±5%mmHg. When
2
steel bottle based on the interval of
2
2
2
57
Chapter 5 Test and Material List
the valve is opened permanently, the patient monitor prompts “APNEA ALARM”.
Unplug the water trap. The patient monitor prompts “CO
water trap again. The prompting message disappears.
When the measured value exceeds the high limit of CO③
prompts “CO
too high” on the main screen. When the measured value is lower
2
than the low limit, the patient monitor prompts “CO
5.1.10 Recorder
1. Print the ECG waveform. The recorder should print it normally and clearly. Set
the recorder to the fault of lack of paper and abnormal clip. There should be relevant
prompting messages on the main screen. When the fault is cleared, the patient
monitor should become normal.
2. Print the alarm messages of all parameters. Set the alarm print switch to ON for
all parameters, and set different alarm limits. Then the recorder should print the
alarm message in case of an alarm
.
5.1.12 Power Supply
When the patient monitor is supplied with the external AC power, the CHARGE
indicator becomes ON. When it is disconnected from the external AC power, the
CHARGE indicator becomes OFF. After the patient monitor is started without
assembling the batteries, “x” is displayed in the battery indication frame on the main
screen. After the batteries are assembled, the battery electricity is displayed in the
battery indication frame on the main screen. The patient monitor can work normally
with or without batteries. It, however, should give an alarm when the batteries are
exhausted.
water trap OFF”. Plug the
2
, the patient monitor
2
too low”.
2
5.1.13 Clock
Verify the correctness of the clock in the system test, and then set the clock to the
current time.
5.1.14 System Te st
Load all parameters, and conduct operations respectively on the loaded parameters.
During the synchronization, no exceptions (for example, mutual interference) occur.
Set all parameter setups in menus to the default values which are those at the time of
software loading, and conduct operations on the menus, for example, managing the
patient information, recalling data, and so on. All the operations should be done
normally, and the corresponding functions should be correct and meet the product
requirements.
58
Chapter 5 Test and Material List
5.2 NIBP Calibration
Figure 5-1 NIBP Calibration
Calibration method:
Based on the precision of 50mmHg (6.7kPa), increase the pressure step by step. The
maximum error at any pressure point within the NIBP measurement range of the patient
monitor should be no more than ±3mmHg (±0.4kPa). Decrease the pressure step by step.
The maximum error at any pressure point within the NIBP measurement range of the patient
monitor should be no more than ±3mmHg (±0.4kPa).
5.3 IBP CALIBRATE
5.3.1 IBP Transducer Zero
Press the ZERO button on the IBP module to call up IBP PRESSURE ZERO menu as
shown below:
Figure 5-2 IBP PRESSURE ZERO
59
Chapter 5 Test and Material List
Zero Calibration of Transducer
Select CH1, the system will zero IBP1. Select CH2, the system will zero IBP2.
Cautions:( Use the PM-6000 IBP module as a example)
Turn off patient stopcock before you start the zero procedure.
The transducer must be vented to atmospheric pressure before the zero procedure.
The transducer should be placed at the same height level with the heart, approximately
mid-axially line.
Zero procedure should be performed before starting the monitoring and at least once a
day after each disconnect-and-connect of the cable.
Figure 5-3 IBP Zero
5.3.2 IBP Calibration
Press CAL button on the IBP module to call up the IBP PRESSURE CALIBRATE menu as
shown below:
60
Chapter 5 Test and Material List
Figure 5-4 IBP Calibration Menu
Calibrate the transducer:
Turn the knob to select the item CH1 CAL VALUE, press and turn the knob to select the
pressure value to be calibrated for channel 1. Then turn the knob to select the item
CALIBRATE to start calibrating channel 1.
Turn the knob to select the item CH2 CAL VALUE, press and turn the knob to select the
pressure value to be calibrated for channel 2. Then turn the knob to select the item
CALIBRATE to start calibrating channel 2.
The pressure calibration of PM-7000:
Pressure
transducer
3-way
stopcock
T-shape connector
Hydrargyrum
pressure meter
Figure 5-5 IBP Calibration
You will need the following pieces of equipment:
• Standard sphygmomanometer
• 3-way stopcock
• Tubing approximately 25 cm long
Pressure transducer
interface cable
Monitor
61
Chapter 5 Test and Material List
The Calibration Procedure:
1. Close the stopcock that was open to atmospheric pressure for the zero calibration.
2. Attach the tubing to the sphygmomanometer.
3. Ensure that connection that would lead to patient is off.
4. Connect the 3-way connector to the 3-way stopcock that is not connected to the patient
catheter.
5. Open the port of the 3-way stopcock to the sphygmomanometer. .
6. Select the channel to be calibrated in the menu and select the pressure value to which the
IBP is to be adjusted.
7. Inflate to make the mercury bar rise to the setup pressure value.
8. Adjust repeatedly until the value in the menu is equal to the pressure value shown by the
mercury calibration.
9. Press the Start button, the device will begin calibrating.
10. Wait for the calibrated result. You should take corresponding measures based on the
prompt information.
11. After calibration, disassemble the blood pressure tubing and the attached 3-way valve.
Via the PM-7000’s system and maintain menus you are prompted for a password for entering
the factory key. After entering the password you get access to the pump rate settings and to
check the accuracy of the CO2 measurement. Using the below test set up to verify the
accuracy of the CO2 module.
Figure 5-6 Sidestream test set up
62
Chapter 5 Test and Material List
Note Neither the mainstream nor the sidestream module can be calibrated. Only the
overall performance and accuracy is checked. If the Co2 module fails the tests it should be
replaced.
Figure 5-7 Factory Maintain Menu
Figure 5-8 CO2 check menu
63
Chapter 5 Test and Material List
5.5 AG CALIBRATE
5.5.1 AG Check1、Using T-piece to connect the watertrap and Agent steel bottle well.
One of the T-piece ports must be vented to atmospheric pressure.
2、Select ‘MEASURE’ from work mode item in “AG SETUP” menu, then set pump rate ‘low’
and wait for 10 minutes after the warm up information disappears.
3、Enter ‘CALIBRATE’ menu, then open AG bottle and press the ‘VERIFY ACCURACY’ item.
Figure 5-9 AG Check Menu
4、Observe the display value after 1 minute. The agent concentration accurate should be less than
±5%.
、Choose other pump rate ‘middle’ or ‘high’,and repeat the previous procedures.
5
(pump rate definition: three pump rate under adult mode: 100/150/200ml/min; neonate:
70/90/110 ml/min)
6、If the accurate over range, please press ‘START CAL’.
64
Chapter 5 Test and Material List
5.5.2 AG CALIBRATE
(Agent>1.5%, CO2>1.5%, N2O>40%, O2>40% )
1、Press ‘START CAL’, then input password ‘MINDRAY’ entering ‘CALIBRATE’ menu.
Note: Make sure the AG in ‘Measure’ mode not ‘Standby’ mode before you do calibrate.
Figure 5-10
Figure 5-11
2、Input each gas concentration value according to the label on the AG bottle label.
Note: If your monitor do not have O2 module, input ‘0.0’ in O2 item.
3、Open AG cover, wait for the display value stabilization.
4、If the display value does not accord with the input value, please press ‘CALIBRATE’ item
and exit.
AG concentration must fit the following requirements:
Agent>1.5%, CO2>1.5%, N2O>40%, O2>40%
65
Chapter 5 Test and Material List
5.6 Bill of Materials for PM-7000 Main Unit
SN P/N Description Quantity
1 M04-004012--- Cross-head screw M3*6 39
2 TR6C-30-16650 Recorder drive board1
5 7000-20-24422 Main support (with lithium battery) 1
9 7001-30-67438 Power board for lead acid battery 1
10 M05-302R3R--- Lead acid battery 1
11 0010-10-12329 Lithium battery 1
12 DA8K-20-14520 Insulation sheet for ECG board 1
13 812A-30-08557 812A ECG board 1
14 M04-060009--- Bolt M3*14 1
15 DA8K-20-14426 SPO2 board mount 1
16 9006-30-33900 SPO2 board 1
17 630D-30-09121 630D pump 1
18 7000-30-24511 Nellcor SP02 module 1
19 0010-10-12274 MASIMO SPO2 module 1
20 7000-20-24387 Insulation sheet for power board 1
21 7001-30-67439 Power board for lithium battery 1
22 7001-30-67464 AG module (with O2) 1
23 7001-30-67461 AG module (without O2) 1
24 7001-30-67458 Mainstream CO2 module 1
25 9000-20-07289 Mainstream CO2 module mounting board 1
26 7001-30-67441 Button board 1
27 9200-30-10701 Alarm light board 1
28 6200-30-09774 Display encoder mounting board 1
29 7001-20-67403 CO2 module mounting board 1
30 M05-010R03--- Button cell battery 1
31 7001-30-67456 Display assembly 1
32 M08A-30-34703 Digiboard for 12 lead ECG module 1
33 M08A-30-34701 Analog board for 12 lead ECG module 1
34 7001-30-67436 Oridion CO2 module 1
35 7000-30-24574 Mindray CO2 module 1
66
Chapter 6 Maintenance and Cleaning
Chapter 6 Maintenance and Cleaning
6.1 Maintenance
6.1.1Checking Before Using
■ Check the patient monitor for mechanical damages;
■ Check all exposed conductors, connectors and accessories;
■ Check all functions that are possibly enabled for the monitored patient, and
ensure the device is in good working status.
In case of any damage, stop using this patient monitor, and contact biomedical
engineers of the hospital or Mindray maintenance engineers.
6.1.2 Regular Checking
An all-around check, including the safety check, should be done by qualified personnel
every 6-12 months or after maintenance each time.
All checks in which the patient monitor should be disassembled should be done by
qualified maintenance personnel. The safety and maintenance checks can be done by
Mindray engineers. The local office of Mindray at your region will be pleased to provide
you with the information about the maintenance contract.
6.2Cleaning
Do switch off the patient monitor and disconnect the AC power supply before
cleaning it or the probes.
The PM-7000r should be dust free. To clean the surface of its enclosure and screen,
use the cleaning agent that is not corrosive, for example, soap and water.
1. Do not use strong solvent, such as acetone;
2. Most cleaning agents must be diluted before being used, so conduct dilution under
the instruction of manufacturers;
3. Do not use any erosive material (such as steel wool or polishing agent);
4. Prevent the ingress of any liquid to the enclosure and any part of the device;
5. Ensure no residue of cleaning liquid on the surface of the device.
6.3 Cleaning Reagent
1. Diluted aqua ammonia
2. Diluted sodium hypochlorite (bleaching powder for washing)
3. Hydrogen peroxide 3%
4. Ethanol
5. Isopropyl alcohol
67
Chapter 6 Maintenance and Cleaning
6.4 Sterilization
To avoid the long-time damage to the patient monitor, we recommend you
9 To conduct only sterilization which is considered necessary in your maintenance
plan;
9 To clean the patient monitor before the sterilization;
9 To sterilize the patient monitor with specified sterilization agent: Ethylate, and
Acetaldehyde.
Caution
Conduct dilution or use the liquid of the possibly-lowest concentration
under the instructions by the manufacturer.
Prevent the ingress of liquid to the enclosure.
Prevent any part of the system from being dipped.
In sterilization, do not spill the liquid to the patient monitor.
Ensure no residue of sterilization agent on the surface of the patient
monitor. Clean it if any.
6.5Disinfection
To avoid the long-time damage to the patient monitor, we recommend you
9 To conduct only disinfection which is considered necessary in your maintenance
plan;
9To clean the patient monitor before the disinfection;
Gas (EtO) or formaldehyde are forbidden for the disinfection of the patient
monitor.
68
P/N: 7001-20-67407(1.0)
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.